05e28 Di-electric Strength (High Voltage) Test and Leakage Current Test Proficiency Testing Program

Size: px
Start display at page:

Download "05e28 Di-electric Strength (High Voltage) Test and Leakage Current Test Proficiency Testing Program"

Transcription

1 instructions Di-electric Strength (High Voltage) Test and Leakage Current Test Proficiency Testing Program Please read the instructions carefully and perform the testing as directed. Should you have any comments or questions, please refer them in writing to us by fax or Fax: mail to: Ingrid Flemming At the conclusion of testing, please your completed result sheet to Margaret Page: Mailing Results Please note that submission of these results and the questionnaire must be completed by the staff performing the test. Results are due on September Copyright IFM Quality Services This document not to be distributed or used other than as intended by the authors. 1 of 10

2 instructions The artefact is a sealed resistive voltage divider of known ratio. The serial number of your test piece is on a label attached to your sample. The target standard of this program is IEC clauses 13 and 16. There are other standards to which this program are equally applicable, particularly IEC clause 5. Copyright IFM Quality Services This document not to be distributed or used other than as intended by the authors. 2 of 10

3 instructions Dielectric Strength Test This is a test of the high voltage test equipment used by the laboratory. Whilst based on IEC , any available equipment may be applied. As mains power is not required, the connection of fig 8 of the standard is not applicable and the device should be stabilised at normal laboratory temperature of 20 C±5K You are asked to apply the voltage generated by your normal High Voltage Test set, equivalent to test voltage for basic insulation. Please apply the test between High and Common (V1) and measuring the resultant voltage between Low and Common using a high input impedance (1Megohm or higher) volt meter (V2). The AC voltage to be applied for the electric strength test is 1000V. V2 should be measured by a true rms meter. Provision is given to specify the type of Voltmeter used for both V1 and V2 in the results sheet. The applied voltage (V1) is to be measured by the normal method for determining the electric strength voltage. If the laboratory uses a DC HV source as permitted in other standards, the measurement may also be made at DV. Measure the voltages at the stabilised value applied at the beginning of the test and one minute later at the end of the test. Record leakage current if available as part of the test rig. Calculate and record the ratio of V2/V1. Copyright IFM Quality Services This document not to be distributed or used other than as intended by the authors. 3 of 10

4 Results Sheet Name of Laboratory Your Assigned Code Number Please Specify Your status within the CB Scheme CB Testing Laboratory Please Specify your relevant National Certification Body, or parent organisation, if applicable. Please specify the standard which the operator completing the program would normally be following when performing this test (if other, please specify) Copyright IFM Quality Services This document not to be distributed or used other than as intended by the authors. 4 of 10

5 Di-electric Strength Test Test Voltage Applied (as measured by test set) Measured Low Voltage Calculated Leakage Current (Provide answer only if able to be measured with dielectric strength test set up. See separate section for specific leakage current determination.) High-common (1) Low -common (2) ratio (2)/(1) (ma) Initial Final 1 AC 2 DC 3 AC 4 DC v1 v2 r lc *if test equipment has ac and dc capability measure in both modes Copyright IFM Quality Services This document not to be distributed or used other than as intended by the authors. 5 of 10

6 Voltmeter Data 5 Impedance AC (MΩ) 6 Impedance DC (MΩ) (if applicable) 7 Sensing type (eg true rms, average rectified) Applied voltage meter (V1) Not required Not required Measured voltage meter (V2) 8 Brand (Not required if part of test set) 9 Model No. (Not required if part of test set) 10 Rated Accuracy (%reading) Question A: How was leakage current measured in this section? Copyright IFM Quality Services This document not to be distributed or used other than as intended by the authors. 6 of 10

7 High Voltage Test Set data Maximum output volts rated output current trip current setting Specified harmonic distortion (ac) or ripple (dc) Supply Mains Harmonic distortion 11 - AC 12 - DC The estimated uncertainty of calculated ratio at 95%confidence limit is The temperature of the ambient environment in the vicinity (within 200mm) of the test equipment is: AC DC Copyright IFM Quality Services This document not to be distributed or used other than as intended by the authors. 7 of 10

8 Leakage Current Test The rated voltage for the purposes of this clause is 230 volts. You are asked to apply 1.06 times rated voltage between the High and Common connections with your normal leakage meter, complying with the requirements of IEC annex G, connected between Low and Common. Copyright IFM Quality Services This document not to be distributed or used other than as intended by the authors. 8 of 10

9 Record the current measured by your normal method of measuring or calculating leakage current. Measured Leakage Current (ma) Applied Voltage (V rms) Mains frequency (Hz) Leakage Current Meter Manufacturer Leakage Current Meter Model Number Leakage Current Meter Input Impedence Question B: the operational principle of the leakage current meter in your laboratory is based on: (tick appropriate) current measurement voltage measurement 21 The temperature of the ambient environment within 200mm of the leakage current test was ( C) Copyright IFM Quality Services This document not to be distributed or used other than as intended by the authors. 9 of 10

10 If you have any comments, please make them here. Copyright IFM Quality Services This document not to be distributed or used other than as intended by the authors. 10 of 10

LOVAG GENERAL INSTRUCTION G2 MEASUREMENT UNCERTAINTY. This instruction of a general nature and does not relate to specific standards.

LOVAG GENERAL INSTRUCTION G2 MEASUREMENT UNCERTAINTY. This instruction of a general nature and does not relate to specific standards. LTI-G2 LOVAG GENERAL INSTRUCTION G2 MEASUREMENT UNCERTAINTY This instruction of a general nature and does not relate to specific standards. It provides additional information ensuring a suitable degree

More information

Test Report CESI A6/ Approved Page 1. MF Trasformatori s.r.l. - Calcinato (BS) Italy

Test Report CESI A6/ Approved Page 1. MF Trasformatori s.r.l. - Calcinato (BS) Italy A6/028659 Test Report CESI Approved Page 1 Client MF Trasformatori s.r.l. - Calcinato (BS) Italy Tested equipment Non-enclosed three-phase dry-type power transformer with three windings, with encapsulated

More information

5322A Electrical Safety Tester Calibrator

5322A Electrical Safety Tester Calibrator TECHNICAL DATA 5322A Electrical Safety Tester Calibrator Calibrate to international standards UK 17th Edition BS 7671 17th Edition - Requirements for Electrical Installations, IET Wiring Regulations The

More information

Organisation Internationale de Métrologie Légale

Organisation Internationale de Métrologie Légale Organisation Internationale de Métrologie Légale INTERNATIONAL RECOMMENDATION Sound level meters Sonomètres OIML R 58 Edition 1998 (E) CONTENTS Foreword... 3 1 Scope... 4 2 Construction and maximum permissible

More information

1507/1503. Insulation Testers. Technical Data. Features and benefits:

1507/1503. Insulation Testers. Technical Data. Features and benefits: 1507/1503 Insulation Testers Technical Data The Fluke 1507 and 1503 Insulation Testers are compact, rugged, reliable, and easy to use. With their multiple test voltages, they are ideal for many troubleshooting,

More information

OIML R 130 RECOMMENDATION. Edition 2001 (E) ORGANISATION INTERNATIONALE INTERNATIONAL ORGANIZATION. Octave-band and one-third-octave-band filters

OIML R 130 RECOMMENDATION. Edition 2001 (E) ORGANISATION INTERNATIONALE INTERNATIONAL ORGANIZATION. Octave-band and one-third-octave-band filters INTERNATIONAL RECOMMENDATION OIML R 130 Edition 2001 (E) Octave-band and one-third-octave-band filters Filtres à bande d octave et de tiers d octave OIML R 130 Edition 2001 (E) ORGANISATION INTERNATIONALE

More information

Title Substation Auxiliary Transformer from Rectifier Transformer Secondary. Reference Number PDS 01 (RIC Standard: EP SP)

Title Substation Auxiliary Transformer from Rectifier Transformer Secondary. Reference Number PDS 01 (RIC Standard: EP SP) Discipline Engineering Standard NSW Category Electrical Title Substation Auxiliary Transformer from Rectifier Transformer Secondary Reference Number PDS 01 (RIC Standard: EP 05 00 00 01 SP) Document Control

More information

NATIONAL TELECOMMUNICATIONS AGENCY RESOLUTION Nº 238, NOVEMBER

NATIONAL TELECOMMUNICATIONS AGENCY RESOLUTION Nº 238, NOVEMBER NATIONAL TELECOMMUNICATIONS AGENCY RESOLUTION Nº 238, NOVEMBER 9 2000 Approving the Regulation for Certifying Telecommunication Equipment Electrical Safety Aspects THE DIRECTING BOARD OF THE NATIONAL TELECOMMUNICATIONS

More information

Technical Requirements for Connecting Small Scale PV (sspv) Systems to Low Voltage Distribution Networks

Technical Requirements for Connecting Small Scale PV (sspv) Systems to Low Voltage Distribution Networks 2014 Technical Requirements for Connecting Small Scale PV (sspv) Systems to Low Voltage Distribution Networks This document specifies the technical requirement for connecting sspv to the low voltage distribution

More information

SAMSUNG ELECTRO-MECHANICS CO., LTD

SAMSUNG ELECTRO-MECHANICS CO., LTD Model Name : Features : ㅡ Suitable for LED lighting application ㅡ UL8750, UL1310 Class 2 ㅡ For use in Dry and Damp Location ㅡ Full-protection provided: OCP/OVP/OTP/SCP ㅡ 0-10V Dimming Control ㅡ IP67 ㅡ

More information

P3065 LOW PROFILE LINE MATCHING TRANSFORMER PRODUCT DATA SHEET. IEC 950, UL 1950 and EN certified UL Recognized Component

P3065 LOW PROFILE LINE MATCHING TRANSFORMER PRODUCT DATA SHEET. IEC 950, UL 1950 and EN certified UL Recognized Component PRODUCT DATA SHEET LOW PROFILE MATCHING TRANSFORMER P3065 Features Low Distortion Low Profile (11mm) Vacuum encapsulated IEC 950, UL 1950 and EN 60950 certified UL Recognized Component BABT Certificate

More information

Test Report. File No.: SHV01023/16 Test Report No.: Taizhou, Zhejiang , P.R. China

Test Report. File No.: SHV01023/16 Test Report No.: Taizhou, Zhejiang , P.R. China Applicant... : Manufacturer... : Order No.... : Zhejiang ERA Solar Technology Co., Ltd. Sihai Road, Huangyan Economic Development Zone Taizhou, Zhejiang 318020, P.R. China Zhejiang ERA Solar Technology

More information

P3801 LOW DISTORTION LINE MATCHING TRANSFORMER PRODUCT DATA SHEET. V.90 and V.92 modems V.34 modems

P3801 LOW DISTORTION LINE MATCHING TRANSFORMER PRODUCT DATA SHEET. V.90 and V.92 modems V.34 modems PRODUCT DATA SHEET LOW DISTORTION MATCHING TRANSFORMER P3801 Features Leadfree (Pbfree) RoHS compliant Low distortion Environmentally tested to IEC 68 CERT reliability tested 12.6mm (0.5") seated height

More information

SPECIFICATION. SilverStone. 300W SFX Switching Power Supply FSP300-60GHS Meet 80 Plus Active PFC Circuit Full Range Input Small Size

SPECIFICATION. SilverStone. 300W SFX Switching Power Supply FSP300-60GHS Meet 80 Plus Active PFC Circuit Full Range Input Small Size SPECIFICATION SilverStone 00W SFX Switching Power Supply FSP00-60GHS Meet 80 Plus Active PFC Circuit Full Range Input Small Size. GENERAL DESCRIPTION AND SCOPE This is the specification of Model UFSP00-60GHSU;

More information

P3065 LOW PROFILE LINE MATCHING TRANSFORMER PRODUCT DATA SHEET

P3065 LOW PROFILE LINE MATCHING TRANSFORMER PRODUCT DATA SHEET PRODUCT DATA SHEET LOW PROFILE MATCHING TRANSFORMER P3065 Features Low Distortion Leadfree (Pbfree) RoHS compliant Low Profile (11mm) Vacuum encapsulated IEC 60950 and UL 60950 certified UL Recognized

More information

TEST REPORT. Energy consumption / efficiency of electrical and electronic products

TEST REPORT. Energy consumption / efficiency of electrical and electronic products MEANPR23 TEST REPORT Energy consumption / efficiency of electrical and electronic products Report Reference No. : 3139541.10 Tested by (name + signature) Approved by (name + signature) : Tony Chen : Cheery

More information

Test Report. File No.: SHV04007/15-02 Test Report No.: , P.R. China

Test Report. File No.: SHV04007/15-02 Test Report No.: , P.R. China Applicant... : Manufacturer... : Order No.... : Zhejiang ERA Solar Technology Co., Ltd. Sihai Road, Huangyan Economic Development Zone, Taizhou, Zhejiang 318020, P.R. China Zhejiang ERA Solar Technology

More information

Introduction. AC or DC? Insulation Current Flow (AC) 1. TECHNICAL BULLETIN 012a Principles of Insulation Testing. Page 1 of 10 January 9, 2002

Introduction. AC or DC? Insulation Current Flow (AC) 1. TECHNICAL BULLETIN 012a Principles of Insulation Testing. Page 1 of 10 January 9, 2002 Page 1 of 10 January 9, 2002 TECHNICAL BULLETIN 012a Principles of Insulation Testing Introduction Probably 80% of all testing performed in electrical power systems is related to the verification of insulation

More information

INTERNATIONAL OIML R 103 RECOMMENDATION

INTERNATIONAL OIML R 103 RECOMMENDATION INTERNATIONAL OIML R 103 RECOMMENDATION Edition 1992 (E) Measuring instrumentation for human response to vibration (with reference to International Standards 8041 and 5347 of the International Organization

More information

Distribution transformers Part 2 Ground mounted transformers not closecoupled

Distribution transformers Part 2 Ground mounted transformers not closecoupled PRODUCED BY THE OPERATIONS DIRECTORATE OF ENERGY NETWORKS ASSOCIATION Technical Specification 35-1 Distribution transformers Part 2 Ground mounted transformers not closecoupled www.energynetworks.org PUBLISHING

More information

RAP Ripple Control Coupling for parallel injection of RC signals in Medium and High Voltage Networks

RAP Ripple Control Coupling for parallel injection of RC signals in Medium and High Voltage Networks RAP Ripple Control Coupling for parallel injection of RC signals in Medium and High Voltage Networks We reserve all rights to this document, and the subject-matter it deals with. Duplication, dissemination

More information

SPECIFICATION ESD FSP250-60HNC. Released Date:2012/08/15-21:58:49

SPECIFICATION ESD FSP250-60HNC. Released Date:2012/08/15-21:58:49 SPECIFICATION ESD12047757 FSP250-60HNC Released Date:2012/08/15-21:58:49 1/9 禁止翻印外洩 ESD12047757-R1.pdf 發行時間 :2012/8/15 全漢企業股份有限公司 FSP TECHNOLOGY INC. 台灣桃園市建國東路 22 號 NO.22,Jianguo E, Rd.,Taoyuan City,Taiwan,R.O.C.

More information

SPECIFICATION FOR APPROVAL

SPECIFICATION FOR APPROVAL Scanning LIB. SPECIFICATION FOR APPROVAL CUSTOMER : ITEM : Power Supply Unit. DESCRIPTION : LCD & LED Monitor Power Supply. CUSTOMER P/NO : SUPPLIER P/NO : BRK-3500 DATE : 2015-10-05 CUSTOMER EN GR CHKD

More information

P3304 LOW INSERTION LOSS LINE MATCHING TRANSFORMER PRODUCT DATA SHEET

P3304 LOW INSERTION LOSS LINE MATCHING TRANSFORMER PRODUCT DATA SHEET PRODUCT DATA SHEET LOW INSERTION LOSS LINE MATCHING TRANSFORMER P3304 Features Leadfree (Pbfree) RoHS compliant Low Insertion Loss,

More information

Transformer Basics AN05-10ST. Application Note. innovation in wire wound magnetic technology. January 09 Rev 1

Transformer Basics AN05-10ST. Application Note. innovation in wire wound magnetic technology. January 09 Rev 1 innovation in wire wound magnetic technology Transformer Basics January 09 Rev 1 AN05-10ST Isolation Transformers Increase Safety of Electronic Systems Application Note Isolation Transformers Increase

More information

PRODUCT TEST MANUAL 2SY110K28 SYNCHRONISM CHECK RELAY

PRODUCT TEST MANUAL 2SY110K28 SYNCHRONISM CHECK RELAY Sheet 1 of 7 Order Number Serial Number PRODUCT TEST MANUAL 2SY110K28 SYNCHRONISM CHECK RELAY Issue Date Level D 12/12/1996 Initial issue. Summary of changes Due to RMS continuous product improvement policy

More information

P2001 LINE MATCHING TRANSFORMER PRODUCT DATA SHEET. 12.6mm (0.5") seated height

P2001 LINE MATCHING TRANSFORMER PRODUCT DATA SHEET. 12.6mm (0.5) seated height PRODUCT DATA SHEET MATCHING TRANSFORMER P2001 Features Ultra Low Distortion 12.6mm (0.5") seated height Vacuum encapsulated IEC 950, UL 1950 and EN 60950 Certified UL Recognized Component BABT Certificate

More information

Test Report. File No.: SHV01032/18-02 Test Report No.:

Test Report. File No.: SHV01032/18-02 Test Report No.: File No.: SHV01032/18-02 No.: 492011003.001 Applicant... : Manufacturer 1... : Manufacturer 2... : Order No.... : Suzhou Akcome Optronics Science &Technology Co., Ltd. Jintang Rd, Zhangjiagang Economic

More information

VHF LAND MOBILE SERVICE

VHF LAND MOBILE SERVICE RFS21 December 1991 (Issue 1) SPECIFICATION FOR RADIO APPARATUS: VHF LAND MOBILE SERVICE USING AMPLITUDE MODULATION WITH 12.5 khz CARRIER FREQUENCY SEPARATION Communications Division Ministry of Commerce

More information

Organisation Internationale de Métrologie Légale

Organisation Internationale de Métrologie Légale Organisation Internationale de Métrologie Légale INTERNATIONAL RECOMMENDATION Integrating-averaging sound level meters Sonomètres intégrateurs-moyenneurs OIML R 88 Edition 1998 (E) CONTENTS Foreword...

More information

Calibration Specialists Ltd

Calibration Specialists Ltd The National Technological Park, Castletroy, Limerick Calibration Laboratory Registration number: 001C is accredited by the Irish National Board (INAB) to undertake calibration as detailed in the Schedule

More information

TPS 49 EDITION 2 JUNE 2009

TPS 49 EDITION 2 JUNE 2009 TPS 49 EDITION 2 JUNE 2009 Interim arrangements & guidance on the interpretation of IEC 61672 Sound Level Meters - Periodic Tests CONTENTS SECTION PAGE 1 Purpose and duration 2 2 Statement 2 3 Specific

More information

Integrated calibration system for electronic and pneumatic instrumentation model CBS-SC2000

Integrated calibration system for electronic and pneumatic instrumentation model CBS-SC2000 Calibration technology Integrated calibration system for electronic and pneumatic instrumentation model CBS-SC2000 WIKA data sheet CT 92.32 Applications Calibration and testing of pneumatic and electronic

More information

Industrial Electricity. Answer questions and/or record measurements in the spaces provided.

Industrial Electricity. Answer questions and/or record measurements in the spaces provided. Industrial Electricity Lab 10: Building a Basic Power Supply ame Due Friday, 3/16/18 Answer questions and/or record measurements in the spaces provided. Measure resistance (impedance actually) on each

More information

OPERATING AND MAINTENANCE MANUAL

OPERATING AND MAINTENANCE MANUAL 5Hz to 1MHz WIDE RANGE FULLY AUTOMATIC DISTORTION ANALYZER MODEL 6900B SERIAL NO. OPERATING AND MAINTENANCE MANUAL Unit 4, 15 Jonathan Drive, Brockton, MA 02301-5566 Tel: (508) 580-1660; Fax: (508) 583-8989

More information

P2001 LINE MATCHING TRANSFORMER PRODUCT DATA SHEET

P2001 LINE MATCHING TRANSFORMER PRODUCT DATA SHEET PRODUCT DATA SHEET MATCHING TRANSFORMER P2001 Features Ultra Low Distortion Leadfree (Pbfree) RoHS compliant 12.6mm (0.5") seated height Vacuum encapsulated IEC 60950 and UL 60950 Certified UL Recognized

More information

IECEE OPERATIONAL DOCUMENT

IECEE OPERATIONAL DOCUMENT IECEE OD-2025-A Edition 1.1 IECEE OPERATIONAL DOCUMENT IEC System of Conformity Assessment Schemes for Electrotechnical Equipment and Components (IECEE System) CTF - Customers' Testing Facility CTF (CTF

More information

VIM Series. 90 & 60 W, Efficient, CV Class 2 LED Drivers ORDERING INFORMATION

VIM Series. 90 & 60 W, Efficient, CV Class 2 LED Drivers ORDERING INFORMATION 060W-12 100W-24 Nominal Input Voltage 90 &, Efficient, CV Class 2 Max. Output Power Nominal Output Voltage Max. Output Current 120 & 277 Vac 12, 24 Vdc 5, 3.75 A Efficiency up to 90% typical Max. Case

More information

ETSI EN V1.2.1 ( )

ETSI EN V1.2.1 ( ) EN 300 132-3 V1.2.1 (2003-08) European Standard (Telecommunications series) Environmental Engineering (EE); Power supply interface at the input to telecommunications equipment; Part 3: Operated by rectified

More information

201AP Charge Amplifier User Manual

201AP Charge Amplifier User Manual Trig-Tek 201AP Charge Amplifier User Manual Publication No. 980996 Rev. A Astronics Test Systems Inc. 4 Goodyear, Irvine, CA 92618 Tel: (800) 722-2528, (949) 859-8999; Fax: (949) 859-7139 atsinfo@astronics.com

More information

AS/NZS :2013 (IEC , , MOD)

AS/NZS :2013 (IEC , , MOD) AS/NZS 60076.7:2013 (IEC 60076-7, Ed. 1.0 2005, MOD) Australian/New Zealand Standard Power transformers Part 7: Loading guide for oil-immersed power transformers AS/NZS 60076.7:2013 AS/NZS 60076.7:2013

More information

Model 4402B. Ultra-Pure Sinewave Oscillator 1Hz to 110kHz Typical Distortion of % Serial No. Operating Manual

Model 4402B. Ultra-Pure Sinewave Oscillator 1Hz to 110kHz Typical Distortion of % Serial No. Operating Manual Model 4402B Ultra-Pure Sinewave Oscillator 1Hz to 110kHz Typical Distortion of 0.0005% Serial No. Operating Manual 15 Jonathan Drive, Unit 4, Brockton, MA 02301 U.S.A. Tel: (508) 580-1660; Fax: (508) 583-8989

More information

Butler Technologies. Unit 14, Block G, Maynooth Business Campus, Co. Kildare, Ireland. Calibration Laboratory Registration number: 256C

Butler Technologies. Unit 14, Block G, Maynooth Business Campus, Co. Kildare, Ireland. Calibration Laboratory Registration number: 256C Unit 14, Block G, Maynooth Business Campus, Co. Kildare, Ireland Calibration Laboratory Registration number: 256C is accredited by the Irish National Board (INAB) to undertake calibration as detailed in

More information

Suppliers' Information Note. BT Metallic Path Facility. Interface Description

Suppliers' Information Note. BT Metallic Path Facility. Interface Description SIN 349 Issue 2.5 August 2015 Suppliers' Information Note For The BT Network BT Metallic Path Facility Interface Description Each SIN is the copyright of British Telecommunications plc. Reproduction of

More information

5790A Automated AC Measurement Standard

5790A Automated AC Measurement Standard 5790A Automated AC Measurement Standard Technical Data Accuracy that s easy to use The 5790A is a complete automated ac measurement standard designed for the most demanding calibration applications. It

More information

WDG 71 - Technical Data Sheet

WDG 71 - Technical Data Sheet HV 804 R WDG 71 - Technical Data Sheet FRAME HV 804 R SPECIFICATIONS & OPTIONS STANDARDS Cummins Generator Technologies industrial generators meet the requirements of BS EN 60034 and the relevant sections

More information

Model Range code System Power supply Option. variables. Housing for panel mounting, (front) protection degree IP50 and optional RS485 serial output.

Model Range code System Power supply Option. variables. Housing for panel mounting, (front) protection degree IP50 and optional RS485 serial output. Power analyzers and Energy Meters Multifunction indicator Type WM12-96 Accuracy ±0.5 F.S. (current/voltage) Multifunction indicator Display of instantaneous variables: 3x3 digit Variable system and phase

More information

WDG 61 - Technical Data Sheet

WDG 61 - Technical Data Sheet HV 804 W WDG 61 - Technical Data Sheet FRAME HV 804 W SPECIFICATIONS & OPTIONS STANDARDS STAMFORD AC generators are designed to meet the performance requirements of IEC EN 60034-1. Other international

More information

Low Pass Harmonic Filters

Low Pass Harmonic Filters Exclusive e-rated Provider PRODUCT SHEET HARMITIGATOR TM Low Pass Harmonic Filters A solution for electrical distribution systems that require stable, reliable power, characterized by unparalleled power

More information

WDG 83 - Technical Data Sheet

WDG 83 - Technical Data Sheet HV 804 R WDG 83 - Technical Data Sheet FRAME HV 804 R SPECIFICATIONS & OPTIONS STANDARDS STAMFORD AC generators are designed to meet the performance requirements of IEC EN 60034-1. Other international

More information

Provisional Specifications

Provisional Specifications Provisional Specifications Category: ELECTRICITY Provisional Specification: PS-E-15 Page: 1 of 7 of Type of Electricity Meters, Instrument Transformers and Auxiliary Devices, Provisional Specifications

More information

LED Driver Linear / area fixed output

LED Driver Linear / area fixed output Driver LC 5W 5/3/35mA fixc lp SNC essence series Product description Fixed output built-in LED Driver Constant current LED Driver Output current 5, 3 and35 ma Max. output power 5 W Up to 91 % efficiency

More information

Model Operating Manual

Model Operating Manual Model 7500 DC to 1MHz Wideband Power Amplifier Operating Manual Copyright 2004. All rights reserved. Contents of this publication may not be reproduced in any form without the written permission of Krohn-Hite

More information

TR CRITERIA FOR LABORATORY ACCREDITATION IN THE FIELD OF ELECTRICAL DCLF METROLOGY. Approved By: Senior Manager: Mpho Phaloane Revised By:

TR CRITERIA FOR LABORATORY ACCREDITATION IN THE FIELD OF ELECTRICAL DCLF METROLOGY. Approved By: Senior Manager: Mpho Phaloane Revised By: CRITERIA FOR LABORATORY ACCREDITATION IN THE FIELD OF ELECTRICAL LF METROLOGY Approved By: Senior Manager: Mpho Phaloane Revised By: Specialist Technical Committee Date of Approval: 2015-08-26 Date of

More information

LICENSING GUIDELINES AND PROCEDURES FOR SHORT-RANGE RADIO SERVICE (SRRS). For the purpose of this Circular, the following terms are defined hereunder:

LICENSING GUIDELINES AND PROCEDURES FOR SHORT-RANGE RADIO SERVICE (SRRS). For the purpose of this Circular, the following terms are defined hereunder: MEMORANDUM CIRCULAR NO. 01-01-98 SUBJECT: LICENSING GUIDELINES AND PROCEDURES FOR SHORT-RANGE RADIO SERVICE (SRRS). Pursuant to the provisions of Republic Act No.7925 Act No. 3846 as amended, Executive

More information

Understanding and Optimizing Electromagnetic Compatibility in Switchmode Power Supplies

Understanding and Optimizing Electromagnetic Compatibility in Switchmode Power Supplies Understanding and Optimizing Electromagnetic Compatibility in Switchmode Power Supplies 1 Definitions EMI = Electro Magnetic Interference EMC = Electro Magnetic Compatibility (No EMI) Three Components

More information

ELECTRICAL NETWORKS SPECIFICATION TECHNICAL SPECIFICATION FOR A 230V/110V AND 400V/110V TRANSFORMER

ELECTRICAL NETWORKS SPECIFICATION TECHNICAL SPECIFICATION FOR A 230V/110V AND 400V/110V TRANSFORMER Approval Amendment Record Approval Date Version Description 03/05/2017 1 Initial issue PRINTOUT MAY NOT BE UP-TO-DATE; REFER TO METRO INTRANET FOR THE LATEST VERSION Page 1 of 13 Table of Contents 1. Purpose...

More information

LED Driver Constant voltage

LED Driver Constant voltage Product description Fixed output constant voltage built-in control gear for LED in 12/24 V Input voltage range 220 240 VAC Max. output power W Connection cable with stripped cable end (300 mm ±10 mm) Polarity

More information

P5059 LOW DISTORTION LINE MATCHING TRANSFORMER PRODUCT DATA SHEET

P5059 LOW DISTORTION LINE MATCHING TRANSFORMER PRODUCT DATA SHEET PRODUCT DATA SHEET LOW DISTORTION MATCHING TRANSFORMER P5059 Features Leadfree (Pbfree) RoHS compliant Low Distortion 11.2mm (0.44 ) Seated Height Industry Standard Pinout IEC 60950 and UL 60950 Certified

More information

200ADM-P. Current Injection System with Phase Shift A 3.000s 2.000A 50.00Hz 0.0. Features

200ADM-P. Current Injection System with Phase Shift A 3.000s 2.000A 50.00Hz 0.0. Features CT ratio Power Harmonics ac+dc 200ADM-P Current Injection System with Phase Shift Features 0-200A output current True RMS metering with 1 cycle capture Variable auxiliary AC voltage/current output with

More information

Ground Bond & Optional HV & HC Scanner. Ground Bond & Functional Run Test. Ground Bond, Functional Run Test & Line Leakage Test. Features and Benefits

Ground Bond & Optional HV & HC Scanner. Ground Bond & Functional Run Test. Ground Bond, Functional Run Test & Line Leakage Test. Features and Benefits Fully-Automated, Multi-Function Electrical Safety Compliance Analyzer Multi-function safety compliance analyzers with an enhanced graphic LCD provide complete test setup and results with an easy-to-use

More information

TECHNICAL REQUIREMENTS FOR ELECTROMAGNETIC DISTURBANCES EMITTED FROM LIGHTING EQUIPMENT INSTALLED IN TELECOMMUNICATION CENTERS

TECHNICAL REQUIREMENTS FOR ELECTROMAGNETIC DISTURBANCES EMITTED FROM LIGHTING EQUIPMENT INSTALLED IN TELECOMMUNICATION CENTERS TR550004 TECHNICAL REQUIREMENTS FOR ELECTROMAGNETIC DISTURBANCES EMITTED FROM LIGHTING EQUIPMENT INSTALLED IN TELECOMMUNICATION CENTERS TR NO. 174001 EDITION 2.1 September 3 rd, 2018 Nippon Telegraph and

More information

AC/DC Current Probe GCP-100 QUICK START GUIDE ISO-9001 CERTIFIED MANUFACTURER

AC/DC Current Probe GCP-100 QUICK START GUIDE ISO-9001 CERTIFIED MANUFACTURER AC/DC Current Probe GCP-100 QUICK START GUIDE ISO-9001 CERTIFIED MANUFACTURER This manual contains proprietary information, which is protected by copyright. All rights are reserved. No part of this manual

More information

Company Directive POLICY DOCUMENT: SD4/7. Relating to 11kV and 6.6kV System Design

Company Directive POLICY DOCUMENT: SD4/7. Relating to 11kV and 6.6kV System Design Company Directive POLICY DOCUMENT: SD4/7 Relating to 11kV and 6.6kV System Design Policy Summary This document describes the standard requirements for the design of the 11kV and 6.6kV systems. Reference

More information

LED Driver Compact fixed output

LED Driver Compact fixed output Driver LC W 35/5/7mA fixc C SNC ESSENCE series Product description Fixed output built-in LED Driver Constant current LED Driver Output current 35, 5 or 7 ma Max. output power W Nominal life-time up to

More information

SPECIFICATION ESD FSP300-60GHT,(85)F. Released Date:2016/05/18-15:38:12

SPECIFICATION ESD FSP300-60GHT,(85)F. Released Date:2016/05/18-15:38:12 SPECIFICATION ESD16025495 FSP300-60GHT,(85)F Released Date:2016/05/18-15:38:12 全漢企業股份有限公司 FSP TECHNOLOGY INC. 台灣桃園市建國東路 22 號 NO.22,Jianguo E, Rd.,Taoyuan City,Taiwan,R.O.C. TEL:+886-3-375-9888 FAX:+886-3-375-6966

More information

VGM Series. 100 & 60 W, Efficient, CV Class 2 LED Drivers for Signage Applications ORDERING INFORMATION

VGM Series. 100 & 60 W, Efficient, CV Class 2 LED Drivers for Signage Applications ORDERING INFORMATION Nominal Input Voltage Max. Output Power Nominal Output Voltage Max. Output Current 120/277 Vac 12, 24, Vdc 5, 3.92 A Efficiency up to 90% typical Max. Case Temperature 100 C (measured at the hot spot)

More information

Model 7000 Low Noise Differential Preamplifier

Model 7000 Low Noise Differential Preamplifier Model 7000 Low Noise Differential Preamplifier Operating Manual Service and Warranty Krohn-Hite Instruments are designed and manufactured in accordance with sound engineering practices and should give

More information

Radio Frequency Exposure Test Report

Radio Frequency Exposure Test Report Radio Frequency Exposure EN 62311 January 2008 Assessment of electronic and electrical equipment related to human exposure restrictions for electromagnetic fields (0Hz 300GHz) (IEC 62311:2007, modified)

More information

APPROVAL SHEET 12V/2A EA1019HHUS(03) EA1019HHUS-120 APP D EDAC POWER ELECTRONICS CO., LTD. TEL: FAX:

APPROVAL SHEET 12V/2A EA1019HHUS(03) EA1019HHUS-120 APP D EDAC POWER ELECTRONICS CO., LTD. TEL: FAX: APPROVAL SHEET CUSTOMER CUSTOMER P/N DESCRIPTION EDAC MPN EDAC MODEL NO FOR SAFETY 12V/2A EA1019HHUS(03) EA1019HHUS-120 DATE 2017-12-08 REVISION 0 APPROVED DESIGN PREPARE RoHS CONCLUSION APPROVED CONDITON

More information

SAMSUNG ELECTRO-MECHANICS CO., LTD

SAMSUNG ELECTRO-MECHANICS CO., LTD Model Name : Features : ㅡ UL8750 / UL1310 Class 2 and CE Class Ⅱ ㅡ Full-protection provided: OCP/OVP/OTP/SCP ㅡ Meets safety requirements for lighting ㅡ Constant Voltage Type ㅡ 5 years warranty ㅡ IP67 Grade

More information

TEST REPORT IEC Safety requirements for electrical equipment for measurement, control, and laboratory use Part 1: General requirements

TEST REPORT IEC Safety requirements for electrical equipment for measurement, control, and laboratory use Part 1: General requirements Issue Date: 2013-07-22 Page 1 of 19 Report Reference # E331788-A13-CB-1 Test Report issued under the responsibility of: TEST REPORT IEC 61010-1 Safety requirements for electrical equipment for measurement,

More information

A New Concept of Power Quality Monitoring

A New Concept of Power Quality Monitoring A New Concept of Power Quality Monitoring Victor Anunciada 1, Hugo Ribeiro 2 1 Instituto de Telecomunicações, Instituto Superior Técnico, Lisboa, Portugal, avaa@lx.it.pt 2 Instituto de Telecomunicações,

More information

Australian Standard. Electricity metering equipment (ac) Particular requirements. Part 23: Static meters for reactive energy (classes 2 and 3)

Australian Standard. Electricity metering equipment (ac) Particular requirements. Part 23: Static meters for reactive energy (classes 2 and 3) AS 62053.23 2006 IEC 62053-23, Ed.1.0 (2003) AS 62053.23 2006 Australian Standard Electricity metering equipment (ac) Particular requirements Part 23: Static meters for reactive energy (classes 2 and 3)

More information

Position Paper on IEEE Task Force on Insulator Icing Test Methods. Item # 5 Power Supply Requirements By Rudi F. Kremer Vice President

Position Paper on IEEE Task Force on Insulator Icing Test Methods. Item # 5 Power Supply Requirements By Rudi F. Kremer Vice President Position Paper on IEEE Task Force on Insulator Icing Test Methods Item # 5 Power Supply Requirements By Rudi F. Kremer Vice President 1 1. Abstract 2 AC single phase power supplies utilized for alternating

More information

Courtesy of CMA/Flodyne/Hydradyne Motion Control Hydraulic Pneumatic Electrical Mechanical (800)

Courtesy of CMA/Flodyne/Hydradyne Motion Control Hydraulic Pneumatic Electrical Mechanical (800) Primary switched-mode power supply units In standard machine production, the requirements for a power supply unit are mostly specified in advance. Here, the TRIO POWER power supply units are suitable which

More information

PRODUCT SPECIFICATIONS

PRODUCT SPECIFICATIONS FEATURES AND BENEFITS* Up to 500,000 duty cycles or 10 year life Low internal resistance High power density 1F to 150F capacitance range -40 o to 85 o C operating temperature range TYPICAL APPLICATIONS

More information

CTF - Customers' Testing Facility

CTF - Customers' Testing Facility IECEE OD-2048-F2:2017 IEC 2017 1 Ed.1.0 CTF - Customers' Testing Facility CTF Assessment Report (CTF Stages 1 and 2) CTF name CTF address, country Date of assessment: yyyy-mm-dd Ed.1.0

More information

Switching Power Supply Type SPP1 60W Enclosed type

Switching Power Supply Type SPP1 60W Enclosed type Type SPP1 60W Enclosed type Universal AC input full range Short circuit protection Internal input filter High efficiency High average efficiency (meets ErP) Low stand-by power consumption CE, TUV, and

More information

Select G3PA SSRs and G32A Cartridges to be Discontinued; Replace with G3PA-X or G3PJ Relays, G32A-A-X Cartridges

Select G3PA SSRs and G32A Cartridges to be Discontinued; Replace with G3PA-X or G3PJ Relays, G32A-A-X Cartridges NO: RL-159 PRODUCT: G3PA SSRs, G32A Cartridges DATE: January 2017 TYPE: Partial Discontinuation Notice Select G3PA SSRs and G32A Cartridges to be Discontinued; Replace with G3PA-X or G3PJ Relays, G32A-A-X

More information

Power Electronics Laboratory-2 Uncontrolled Rectifiers

Power Electronics Laboratory-2 Uncontrolled Rectifiers Roll. No: Checked By: Date: Grade: Power Electronics Laboratory-2 and Uncontrolled Rectifiers Objectives: 1. To analyze the working and performance of a and half wave uncontrolled rectifier. 2. To analyze

More information

TEST REPORT IEC Uninterruptible power systems Part 3: Method of specifying the performance and test requirements

TEST REPORT IEC Uninterruptible power systems Part 3: Method of specifying the performance and test requirements Test Report issued under the responsibility of: TEST REORT IEC 62040-3 Uninterruptible power systems art 3: Method of specifying the performance and test requirements Report reference No...: 11-EL-0026.06

More information

Lab 2: DC Circuits Lab Assignment

Lab 2: DC Circuits Lab Assignment 2 class days 1. I-V curve for various components Source: Curtis, 1.2.1. (HH 1.1, 1.2, 1.3) Lab 2: DC Circuits Lab Assignment A passive element is a two-contact device that contains no source of power or

More information

Company Directive STANDARD TECHNIQUE: SD1E/2. Technical Requirements for Customer Export Limiting Schemes

Company Directive STANDARD TECHNIQUE: SD1E/2. Technical Requirements for Customer Export Limiting Schemes Company Directive STANDARD TECHNIQUE: SD1E/2 Technical Requirements for Customer Export Limiting Schemes Policy Summary This Standard Technique specifies the requirements for customer owned Export Limitation

More information

Butler Technologies. Unit 14, Block G, Maynooth Business Campus, Co. Kildare, Ireland. Calibration Laboratory Registration number: 256C

Butler Technologies. Unit 14, Block G, Maynooth Business Campus, Co. Kildare, Ireland. Calibration Laboratory Registration number: 256C Unit 14, Block G, Maynooth Business Campus, Co. Kildare, Ireland Calibration Laboratory Registration number: 256C is accredited by the Irish National Board (INAB) to undertake calibration as detailed in

More information

APMP GUIDELINES FOR ACCEPTING A QUALITY SYSTEM(V.2.0 WD2) approved on July 2010

APMP GUIDELINES FOR ACCEPTING A QUALITY SYSTEM(V.2.0 WD2) approved on July 2010 APMP GUIDELINES FOR ACCEPTING A QUALITY SYSTEM(V.2.0 WD2) approved on July 2010 The CIPM global mutual recognition arrangement (global MRA) requires national metrology institutes (NMIs) to have in place

More information

Earth Leakage Monitoring System IsoBase IsoHub IsoOut

Earth Leakage Monitoring System IsoBase IsoHub IsoOut 2014 11 21 Earth Leakage Monitoring System IsoBase IsoHub IsoOut User Manual Due to our policy of continual improvement, specifications may change without prior notice Page 2 (34) Contents Earth Leakage

More information

CMA-100. User Manual. Counter Measures Amplifier

CMA-100. User Manual. Counter Measures Amplifier CMA-100 Counter Measures Amplifier User Manual Research Electronics International, LLC 455 Security Drive, Cookeville, TN 38506 U.S.A. (800) 824-3190 (US Only) +1 931-537-6032 www.reiusa.net Copyright

More information

Test description for dry-type transformers chapter for special tests

Test description for dry-type transformers chapter for special tests Test description for dry-type transformers chapter for special tests 1. SCOPE 4 2. STANDARDS 5 3. LIGHTNING IMPULSE TEST 6 4. SOUND LEVEL MEASUREMENT 7 4.1. STANDARD 7 4.2. AIM 7 4.3. THEORETICAL PRINCIPAL

More information

Three-phase monitoring relay

Three-phase monitoring relay 1SR 430 824 F9300 Features Monitoring of three-phase mains for phase sequence and failure Powered by the measuring circuit 2 c/o (SPDT) contacts 1 LED for status indication Approvals R: yellow LED - relay

More information

Power analyzers and Energy Meters. Multifunction indicator. Type WM12-DIN. Product Description. How to order WM12-DIN AV5 3 D X.

Power analyzers and Energy Meters. Multifunction indicator. Type WM12-DIN. Product Description. How to order WM12-DIN AV5 3 D X. Power analyzers and Energy Meters Multifunction indicator Type WM12-DIN Accuracy ±0.5 F.S. (current/voltage) Multifunction indicator Display of instantaneous variables: 3x3 digit Variable system and phase

More information

BEST BMET CBET STUDY GUIDE MODULE ONE

BEST BMET CBET STUDY GUIDE MODULE ONE BEST BMET CBET STUDY GUIDE MODULE ONE 1 OCTOBER, 2008 1. The phase relation for pure capacitance is a. current leads voltage by 90 degrees b. current leads voltage by 180 degrees c. current lags voltage

More information

Specific Accreditation Criteria Calibration ISO/IEC Annex. Electrical metrology

Specific Accreditation Criteria Calibration ISO/IEC Annex. Electrical metrology Specific Accreditation Criteria Calibration ISO/IEC 17025 Annex Electrical metrology January 2018 Copyright National Association of Testing Authorities, Australia 2014 This publication is protected by

More information

Multimeter 500CVD21 RTU500 series

Multimeter 500CVD21 RTU500 series Remote Terminal Units - Data sheet Multimeter 500CVD21 RTU500 series CT/VT interface with 4 voltage and 24 current inputs for direct monitoring of 3/4 wire 0 300 V AC (line to earth), 0...500 V AC (phase

More information

22 Series Relays. Technical Data...1. Specifications...2. Model Number Structure - Relays...3. Model Number Selection

22 Series Relays. Technical Data...1. Specifications...2. Model Number Structure - Relays...3. Model Number Selection Relays 22 Technical Data...1 Specifications...2 Model Number Structure - Relays...3 Model Number Selection... 4-5 Electrical Characteristics...6 Accessories...7 Dimensions...8 Instructions...10 Safety

More information

Commissioning Manual Distribution Transformers - Three-phase Oil Filled

Commissioning Manual Distribution Transformers - Three-phase Oil Filled 1LPL000006A2110 rev. 1 Commissioning Manual Distribution Transformers - Three-phase Oil Filled 1LPL000006A2110 rev. 1 2 / 7 Table of contents 1 Commissioning and start-up... 3 1.1 Transformer survey...

More information

Testing & Calibration Lab, 204, Diamond Industrial Estate No. 2, Ketki Pada Road, (Near Dahisar Toll Naka), Dahisar (East), Mumbai, Maharashtra

Testing & Calibration Lab, 204, Diamond Industrial Estate No. 2, Ketki Pada Road, (Near Dahisar Toll Naka), Dahisar (East), Mumbai, Maharashtra Last Amended on - Page 1 of 5 SOURCE 1. DC VOLTAGE 1 mv to 32 mv 0.43 % to 0.02 % 32 mv to 32 V 0.02 % to 0.008 % 32 V to 1000 V 0.008 % to 0.02 % 2. DC CURRENT 10 µa to 320 µa 0.15 % to 0.03 % 32 µa to

More information

SPECIFICATION FOR OVERCURRENT RELAYS

SPECIFICATION FOR OVERCURRENT RELAYS SPECIFICATION FOR OVERCURRENT RELAYS 1/9 1. Procedure A manufacturer willing to classify an overcurrent relay according to this specification should provide: A complete file providing a clear, unambiguous

More information

High-Voltage Test Techniques

High-Voltage Test Techniques High-Voltage Test Techniques Dieter Kind Kurt Feser 2nd Revised and Enlarged Edition With 211 Figures and 12 Laboratory Experiments Translated from the German by Y. Narayana Rao Professor of Electrical

More information

IAEA Coordinated Research Project on Development of Harmonized QA/QC Procedures for Maintenance and Repair of Nuclear Instruments

IAEA Coordinated Research Project on Development of Harmonized QA/QC Procedures for Maintenance and Repair of Nuclear Instruments PROCEDURE: TEST PROCEDURE FOR GEIGER-MUELLER RADIATION DETECTORS Nº: MRNI-501 DECEMBER 2008 PAGE: 1 OF: 17 IAEA Coordinated Research Project on Development of Harmonized QA/QC Procedures for Maintenance

More information