Compton Scattering. MIT Department of Physics

Size: px
Start display at page:

Download "Compton Scattering. MIT Department of Physics"

Transcription

1 Compton Scattering MIT Department of Physics You will observe the scattering of kev photons by electrons and measure the energies of the scattered gamma rays as well as the energies of the recoil electrons. The results can be compared with the formulas for Compton scattering, considering the quantized photon like a particle. PREPARATORY QUESTIONS Please visit the Compton Scattering chapter on the course website to review the background material for this experiment. Answer all questions found in the chapter. Work out the solutions in your laboratory notebook; submit your answers on the course website. [Note: Not available to OCW users] SUGGESTED SCHEDULE Day 1: Familiarize yourself with the equipment and do Section III.1. Day 2: Repeat Section III.1 and do Section III.2. By the end of this session you should have verified that E scat + E recoil 662 kev for at least two angular positions. Day 3: Perform Section III.3. Day 4: Repeat portions of the experiment that need improvement. I. INTRODUCTION By 192 the successes of the quantum theories of blackbody spectra (Planck, 191), the photoelectric effect (Einstein, 195) and the hydrogen spectrum (Bohr, 1913) had established the idea that interactions between electromagnetic radiation of frequency ν and matter occur through the emission or absorption of discrete quanta of energy E = hν. The next crucial step in the development of the modern concept of the photon as the particle of electromagnetic radiation was taken by Arthur Compton in the interpretation of experiments he initiated in 192 to measure with precision the wavelengths of X-rays scattered from electrons in materials of (low atomic number) [1, 3]. The phenomena of X-ray scattering had already been studied intensively. It was known that the penetrating power of X-rays decreases with increasing wavelength and that X-rays are less penetrating after being scattered than before, which indicated that the scattering process somehow increases their wavelength. Compton s idea was to use the recently developed technique of high-resolution X-ray spectrometry, based on measurement of the angle of Bragg reflection of X-rays from crystals, to measure precisely the wavelengths of the scattered X-rays [4]. Irradiating a carbon target with an intense collimated beam of monochromatic molybdenum K α X-rays and using an ionization chamber as the detector in his spectrometer, Compton found that the spectrum of scattered X-rays had two distinct spectral lines, one at the wavelength of the incident X-rays and another at a wavelength that was longer by an amount that depends on the angle of scattering. The scattering without a wavelength shift was readily explained by the classical theory of coherent scattering of electromagnetic waves from electrons bound in atoms. However, the classical theory provided no explanation of the wavelength-shifting incoherent scattering process. The phenomenon of Bragg reflection used in Compton s measurements was a clear demonstration of the wavelike character of the X-rays. Nevertheless, Compton put forward the apparently contradictory idea that X-rays, known to be electromagnetic radiation of very short wavelength, interact with electrons like particles of zero rest mass so that their energy E = hν = hc/λ and momentum p are related by the relativistic equation for particles of zero rest mass, namely p = E/c. He calculated according to relativistic mechanics the relations between the initial and final energies and momenta of an X-ray quantum and a free electron involved, like billiard balls, in an elastic collision. In this way he arrived at the formula for the Compton shift in the wavelength of incoherently scattered X-rays, namely (see [2], Appendix A) h Δλ = (1 cos θ), (1) mc where m is the mass of the electron and θ is the angle between the trajectories of the incident and scattered photon. The agreement between Compton s experimental results confronted physicists and philosophers with the conceptual dilemma of the particle-wave duality of electromagnetic radiation. In particular, how was one to understand how each particle of light in a Young interference experiment goes simultaneously through two slits? The dilemma was further compounded when, in 1924, de Broglie put forward the idea that material particles (i.e. electrons, protons, atoms, etc.) should exhibit wavelike properties characterized by a wavelength λ related to their momentum p by the same formula as that for light quanta, namely λ = h/p. In 1927 the wave-like properties of electrons were discovered by Davisson and Germer in experiments on the reflection of electrons from crystals. A particularly interesting account of these developments is given by Compton and Allison in their classic treatise [5]. Id: 1.compton.tex,v /8/27 19:2:55 spatrick Exp

2 2 I.1. Experimental Goals 1. Kinematics and theory of Compton scattering; In this experiment you will measure: 1. the energies of Compton-scattered gamma-ray photons and recoil electrons, 2. the frequency of scattering as a function of angle θ, and 3. the total cross section of electrons for Compton scattering. The results of the energy measurements will be compared with the predictions of Compton kinematics. The results of the scattering rate and cross section measurements will be compared with the predictions of the classical (Thomson) theory of X-ray scattering and with the Klein-Nishina formula derived from relativistic quantum theory. The experimental setup employs a radioactive source of kev photons from 137 Cs, and two NaI scintillation counters. One counter serves as the scattering target and measures the energy of the recoil electrons; the other counter detects the scattered photons and measures the energy they deposit. Both the target and scatter counters have a scintillator consisting of a 2 2 cylinder of Thallium-activated sodium iodide optically coupled to a photomultiplier. A kev photon traversing sodium iodide has about equal probability of undergoing photoelectric absorption and Compton scattering. A photoelectric absorption event is a quantum mechanical transition from an initial state consisting of an incident photon of energy E and a neutral atom to a final state. This event therefore consists of an excited ion with a vacancy in one of its inner shells where the electron that was formerly bound and a free photoelectron with kinetic energy E W, where W is the binding energy of the electron in the neutral atom. The energetic photoelectron loses its energy within a small fraction of a microsecond by multiple Coulomb interactions with electrons and nuclei in the sodium iodide crystal, and a certain fraction of the lost energy is converted into scintillation light. Meanwhile, the ion from which the photoelectron was ejected loses its energy of excitation through a cascade of transitions that fill the vacancy in its electronic structure and give rise to lower energy photons or electrons that, in turn, lose their energy in the crystal and generate additional scintillation light. The total amount of scintillation light produced is closely proportional to the total amount of energy dissipated in all these processes. As a consequence, the pulse height spectrum of a NaI scintillation counter exposed to monoenergetic gamma rays shows a distinct photoelectric peak which facilitates accurate calibration in terms of pulse height versus energy deposited. The counters can then be used to measure the energies of scattered photons and recoil electrons. The following background topics should be studied in [2, 6]: 2. Properties of 137 Cs; 3. Scintillation counters and the interpretation of pulse height spectra (see preparatory question 3); 4. Radiation safety; 5. Parameter and error estimation [6]. II. EXPERIMENTAL ARRANGEMENT The experimental arrangement for the Compton experiment is shown schematically in Figure 1. The 2 x2 cylindrical recoil electron or target scintillator detector (Canberra model 82-3/27), is irradiated by a beam of kev photons emitted by about 1 µci ( 1 microgram!) of 137 Cs located at the end of a hole in a large lead brick which acts as a gamma-ray howitzer. If a photon entering the target scintillator scatters from a loosely bound, effectively free, electron, the resulting recoil electron may lose all of its energy in the target, causing a scintillation pulse with an amplitude proportional to the energy of the recoil electron. If the scattered photon emerges from the target scintillator without further interaction, and if its trajectory passes through the NaI crystal of the scattered photon detector, it will have a substantial probability of depositing all of its energy by a single photoelectric interaction or by a sequence of Compton scatterings and photoelectric interactions. This will produce a scintillation pulse that contributes to the photopeak of the pulse height spectrum. The median channel of the photopeak in the multichannel analyzer (MCA) display is a good measure of the median energy of the detected photons. If the scattered photon undergoes a Compton scattering in the scatter scintillator and then escapes from the scintillator, the resulting pulse, with a size proportional to the energy of the Compton recoil electron, will be registered in the Compton recoil continuum of the spectrum. Several experimental details are considered in [7 11]. In the experimental arrangement described by [2] (pg. 256) the target is inert material (not a detector). The scatter counter suffers background due to gamma-ray photons leaking through the lead from the 137 Cs source, cosmic rays, and radioactive isotopes in the environment which is suppressed by the use of very heavy ( 2 lb) lead shielding around the source and the scatter counter. Without such shielding and without any electronic tricks the pulses produced by Compton scattered photons would be buried in the background. In a better way, the Compton scattered signal can be discriminated from the background using time coincidence techniques, as described in the following paragraphs. In our experiment (see Fig. 1) both the recoil electron detector and the scattered photon detector are scintillation detectors. Circuits are arranged so that a pulse

3 3 FIG. 1. Schematic diagram of the experimental arrangement for measuring Compton Scattering. from a detector is accepted for pulse height analysis by the multichannel analyzer (MCA) only if it is coincident (within a fraction of a microsecond) with a pulse from the other detector. Coincident pulses occur when a gamma ray photon produces a Compton-recoil electron in the recoil electron detector, and the Compton-scattered photon is subsequently absorbed in the scattered photon detector. The two nearly simultaneous pulses produced in such an event are amplified and fed through inverters (Ortec model 533) to constant fraction discriminators (CFDs) (Canberra model 2126) whose threshold is user selectable between -5 mv and -1 V through a 1-turn potentiometer. (The inverters are required because the amplifiers are designed to put out positive pulses, whereas the CFDs are designed for negative input pulses.) CFDs employ a historically interesting method to achieve precise timing described in Appendix B. Your experiment does not require this accuracy, and it is superseded by modern electronics.

4 4 FIG. 2. Schematic representation of the proper time relation between the photomultiplier tube signal (amplified and inverted), whose pulse height is to be measured, and the coincident gate pulse produced by the Gate and Delay Generator NIM module. The coincident gate pulse needs to arrive before the peak of the pulse and persist for.5 µs after the pulse peak. II.1. Coincidence Techniques A true Compton event will produce coincident (within 2 µs) signals in both detectors. If the logic (square) discriminator outputs overlap, the coincidence ( AND ) produces an output YES signal to the gate generator, which in turn notifies the MCA in the computer to analyze the pulse heights generated by this event. Most likely it is a Compton event, however there is a chance that two different background events accidentally occurred within the overlap time 2τ given by the discriminator outputs. From a scatter counter with rate n s and target counter with rate n t the accidental rate n acc = 2τn s n t (2) is small unless you have noise increasing n s or n t. A good method to set up the coincidence uses a 22 Na source placed in between the two detectors. But for Compton coincidences, consider that for small θ, the recoil (target) energy and hence signal are small. If the signals are below your discriminator threshold, you fail to see coincident Compton events at small θ. II.2. Gating the MCA In principle, the gate should tell the MCA ahead of time that a good event is coming. With the present electronics, this is not possible so we try to catch the majority of the signal (see Fig. 2) including the peak. The analog pulse shown is derived from the unipolar output of the amplifier. The bipolar has positive and negative peaks and is more tricky to use. Nuclear physicists use it at high counting rates. When you think you have everything working correctly, and before you commit to a long series of measurements, check the performance at the extreme values of θ, say and 15. At both positions the peaks in the spectra due to the scattered photons and the recoil electrons should stand out clearly. If they do not, then more adjustments must be made to achieve correct operation of the coincidence gate logic. In coincidence mode, the MCA histogram is not updated unless an appropriate gate pulse arrives at the gate input during the rise of the signal pulse. During the dead time required for each pulse height analysis ( 8 µs for the MCAs used in Junior Lab) the input to the pulse height analyzer is held closed by a disable gate so that the arrival of a second pulse will not interfere with the current analysis. More details of the specific MCAs (Perkin-Elmer model Trump-PCI) used in Junior Lab are available in the Junior Lab Filing Cabinet or in the online e-library. III. EXPERIMENTAL PROCEDURE Keep the lead door of the gamma-ray howitzer closed when not in use. The purpose of the procedures described below is to acquaint you with the operation of the equipment, particularly the coincidence scheme which permits the experiment to achieve a good signal/background ratio with a minimum amount of shielding. The steps below suggest a sequence of tests and adjustments to achieve that purpose, but you should feel free to devise your own. III.1. Experimental Calibration Set up and test the experimental arrangement with 511 kev annihilation photons from a 22 Na calibration source. 22 Na, a radioactive isotope of sodium, decays to an excited state of 22 Na by emission of an anti-electron, i.e. a positron. The excited neon nucleus, in turn, decays quickly to its ground state by emission of a 1.27 MeV photon. Meanwhile, the positron comes to rest in the material in which the sodium isotope is embedded, combines with an electron to form an electrically neutral electronic atom called positronium (discovered in 1947 by Prof. Martin Deutsch of MIT). The latter lives about 1 7 s until the electron and positron annihilate yielding two 511 kev (= m e c 2 ) photons traveling in opposite directions. Thus 22 Na is a handy source of pairs of monoenergetic photons traveling in opposite directions which can be used to test a coincidence detector system. 1. Set up the NaI recoil electron and scattered photon scintillation detectors as illustrated in Figure

5 5 1. Turn on the power to the NIM bin, HV supplies, MCA and oscilloscope. Keep the door on the 137 Cs source howitzer closed to reduce the intensity of kev photons at the detectors. 2. Adjust the gain of the scattered photon detector and calibrate it. Apply high voltage to the scatter counter photomultiplier ( +1 VDC). Using the Maestro-32 MCA software, turn off the coincidence gate requirement on the MCA. Place the 22 Na calibration source near the scattered photon detector and examine the output of the scattered photon amplifier signal (following the pre-amplifier AND the amplifier in series) with the oscilloscope. Adjust the gain so that the amplitude of the pulses produced by 511 kev annihilation gamma rays is about +7 volts. (You might see that a few very high energy pulses saturate the amplifier resulting in square peaked signals. Since the Compton scattering experiment will measure only energies < kev, it is okay to discriminate against these pulses.) Determine the gain setting of the primary amplifier at which non-linear response begins, as indicated by a flattening of the top of the pulse as viewed on the oscilloscope. Then reduce the primary amplifier gain sufficiently to assure a linear response up to energies safely above the kev energy of the 137 Cs photons. Now feed the output pulses from the amplifier to the input of the MCA and acquire about 3 seconds of pulse height data using the acquisition presets sub-menu. Examine the Energy spectrum/histogram of pulses from the scattered photon detector and adjust the gain of the amplifier so that the median channel of the photopeak for 511 kev photons is 2/3 of the full scale MCA range (check that this is set to 248 channels). Calibrate the counter with other plastic rod calibration gamma-ray sources. A good set consists of 137 Cs (661.6 kev), 22 Na (511 kev), and 133 Ba (356, 32, and 81 kev plus other lines) each of which produce easily identified photopeaks. Be aware that too high a counting rate will spoil the spectrum by overlapping signals and amplifier saturation. Write in your lab notebook a succinct description of what you are doing along with the data so that you can later recall exactly what you did and repeat it quickly. You will have to redo the calibration at the start of each lab session because other users will have changed the settings. When you have identified the photopeaks and are confident about the counting rates, measure the median channels of the photopeaks and plot the average median photopeak channel numbers versus the photon energies on millimeter graph paper to provide yourself with a calibration curve. The simplest way to estimate the random error in a measurement of the median is to repeat it five or more times and then evaluate the root mean square (RMS) deviation of the successive measurements from their mean value. 3. Adjust and calibrate the recoil electron target detector in the same way. 4. Send the amplified recoil electron and scattered photon signals to the Inverter module and then send the resultant negative polarity pulses into the two discriminators. Examine the positive output pulses generated by the units. 5. Send the two discriminator outputs into channels 1 and 2 of the coincidence module and examine its output. (It is easiest if you temporarily use a single detector signal to trigger both discriminators for this test.) 6. Send the output from the coincidence circuit into the gate generator unit and explore its operation. Adjust the delay of the delay amplifier so the coincidence gate pulse arrives at the MCA gate input during the rise of the analog signal pulse (see Figure 2). A good way to verify the action of the gate pulse is to feed the pulses from one of the amplifiers to both CFD inputs. This will ensure that a coincidence pulse is generated at relatively high rates. (Students should play with the gate generator controls, specifically delay and width, to verify the operation of the MCA card coincidence circuitry as described in Figure 2.) 7. Test the coincidence logic. Place the 22 Na source between the two scintillators so that pairs of annihilation photons traveling in precisely opposite directions can produce coincident pulses in the two counters. For this test it is interesting to reduce the gain by a factor of 2 so that the photopeak of the 1.27 MeV photons can be seen on the MCA spectrum display. Compare the spectrum with and without the coincidence requirement, and observe the effect of moving the source in and out of the straight line between the scintillators in order to observe hits by a pair of annihilation photons. Take careful notes and be sure you understand what you observe. Do not hesitate to ask for help if you need it. In each of the above steps it is very useful to sketch a picture (or generate a bitmap using the Agilent digital oscilloscopes) of the waveforms taking particular note of their amplitudes, rise and fall times and temporal relationships with other signals, e.g. gates. III.2. Compton Scattering Using 137 Cs This section details the measurement of the energies of the scattered gamma rays and the recoil electrons as functions of the scattering angle θ.

6 6 1. Align the primary gamma-ray beam from the 137 Cs source in the howitzer. Position the howitzer and the scatter counter so that the source (at the inside of the hole in the lead howitzer) and the scatter counter scintillator are at opposite ends of a diameter of the round platform, with the source 4 cm from the center of the scintillator. (Keep careful notes about the exact geometry you choose for the setup so you can return to it if necessary at a later time.) Open the door of the howitzer and measure the rate of pulses in the photopeak from the scatter counter as a function of its position angle. In other words, measure the profile of the beam and determine θ =. Check that the beam profile is symmetric about a line from the source through the center of the turntable to insure that the target counter will be uniformly illuminated when it is placed at the center. 2. Adjust the discriminators of the CFDs to insure that they will generate coincident pulses whenever a Compton scattering occurs in the target scintillator with a scattering angle in the range you wish to cover. Position the NaI target counter over the center of the turntable at a height such that the axis of the beam intersects the mid-plane of the scintillator. For a preliminary test, connect the (inverted) output of the target amplifier to the inputs of both CFDs. Simultaneously connect the output of the target amplifier to the INPUT BNC connector on the MCA card in the PC. Connect the output of the coincidence circuit to the gate generator module and the output of the gate generator to the GATE BNC on the MCA card. Using the Maestro- 32 software package, set the discriminators as low as possible without getting into the rapidly rising spectrum of photomultiplier noise such that the accidental coincidence rate would rival the true coincidence rate. Start a spectrum accumulation and test the effects of changing the delay of the gate pulse with respect to the signal pulse. 3. Measure the median energy E of scattered γ photons and the median energy E e of recoil electrons as functions of the position angle θ of the scattered photon detector. With both the source in the howitzer and the scintillator of the scattered photon detector located 4 cm from the recoil electron detector, set the position angle of the scatter counter at 9. With the coincidence requirement enabled, accumulate a spectrum of scattered photon detector pulses on the scatter MCA and a spectrum of recoil electron detector pulses on the target MCA and measure the median channel of the pulses in the photoelectric peaks. Repeat this dual measurement at several position angles from just outside the primary beam profile to as near to 18 as you can get. A good strategy is to take data at widely separated position angles, say 9, 3, 12, 15 and then fill in when you have time. Plot the raw results (median energy against position angle) as you go along to guide your measurement strategy. It is wise to check the calibration between each measurement by switching the (software based) gate setting to OFF and recording in your notebook the median channels of the photopeaks produced by kev photons from the 137 Cs calibration source, the 511 kev photons from 22 Na, and the 356 kev and 81 kev photons of 133 Ba. Record the integration time of each measurement to find the rate as a function of θ to later compare with Thomson s classical prediction. III.3. Scattering Cross Section Measurements In this part of the experiment, you will measure the total scattering cross section per electron in plastic. The plastic scintillator blocks used as absorbers in this measurement are made from polyvinyltoluene (C 1 H 11, density 1.32 g/cm 3, composed of almost equal numbers of carbon and hydrogen atoms. The most tightly bound electrons are in the K-shells of the carbon atoms where they are bound by.277 kev, which is small compared to kev. In fact, many other materials can be used providing approximately free electrons. This material resembles most modern plastic scintillators. You will determine their quantum efficiency, e.g. how often they see γ-rays. Place the scatter counter at and remove the target counter from the beam. Measure the counting rates in the kev photopeak of the NaI scatter counter with no absorber and with three or more different thicknesses of plastic scintillator placed just in front of the exit hole of the howitzer. Determine the thicknesses (in g/cm 2 ) of the scintillator blocks by measuring their dimensions and mass. Plot the natural logarithm of the measured rate as a function of the thickness and check the validity of your data by observing whether the data points fall along a straight line as expected for exponential attenuation. Question: Why is it better to place the absorbers as near as possible to the source rather than just in front of the detector? IV. ANALYSIS 1. Plot E target + E scatter vs. θ. 2. Compare your results on the angular dependence of energies of the scattered photons and the recoil electrons with the predictions of the Compton formula. Since the relation between the scattered photon energy Eγ and the scattering angle θ, is non-linear, one can anticipate that a plot of the

7 7 measured energy versus position angle could be fitted only by a curve. However, an easy manipulation of the Compton formula yields a linear relation between 1/Eγ and 1-cos θ. Plot 1/Eγ against (1 cos θ) with appropriate error estimates. On the same graph plot the calculated curve of 1/Eγ against (1 cos θ) based on the Compton formula. Do a similar analysis of the recoil electron data on a separate graph. Plot 1/E e against a function of θ such that the expected relation is linear. Comment on the degree of conformance of the prediction to your experimental results and discuss possible causes of any systematic differences. 3. Plot the measured rate as a function of θ and compare to the classical Thomson prediction. 4. Compute the total interaction cross section per electron. The attenuation of a collimated beam of particles by interactions in a slab of material of thickness x (in centimeters) is described by the formula µx I(x) = I e, (3) where I is the initial intensity and µ is the total linear attenuation coefficient (cm 1 ). In plastic scintillator the attenuation of kev photons is due almost entirely to Compton scattering. Under this circumstance the total attenuation coefficient is related to the Compton scattering cross section per electron, which we call σ total, by the equation µ σ total =, (4) n e where n e is the number of electrons/cm 3 of the material. To find µ from your data, you should fit an exponential directly to the data, determining the best-fit parameters and their errors. Alternatively, you may plot the natural log of the measured values of I/I against x. Then determine the slope (and its error) of the straight line that best fits the data by linear regression. In either case, calculate the number of electrons/cm 3 of scintillator (CH) n, and find σ total. Compare to the classical Thomson calculation and to the theoretical result of Klein- Nishina [12]. IV.1. Suggested Theoretical Topics 1. Relativistic mechanics and the Compton shift. 2. Derivation of the Thomson differential and total cross section for unpolarized photons. 3. Discuss the correct Klein-Nishina prediction. [1] A. Compton, Phys. Rev. 21, 483 (1923), brother of K.T. Compton, MIT President. [2] A. Melissinos and J. Napolitano, Experiments in Modern Physics, 2nd ed. (Academic Press, Orlando, 23). [3] A. Compton, Phys. Rev. 22, 49 (1923). [4] A. Compton, Nobel Prize Lecture(1927). [5] A. Compton and S. Allison, X-Rays in Theory and Experiment (Van Nostrand, Princeton, 1992) qa481.c738 Physics Department Reading Room. [6] P. Bevington and D. Robinson, Data Reduction and Error Analysis for the Physical Sciences, 3rd ed. (McGraw- Hill, 23). [7] G. Knoll, Radiation Detection and Measurement, 3rd ed. (John Wiley, Orlando, 1999) ISBN [8] R. D. N.H. Lazar and P. Bell, Nucleonics 14, 52 (1956). [9] R. Garner and K. Verghese, Nuclear Instruments and Methods 93, 163 (1971). [1] R. Hofstadter and J. McIntyre, Physical Review 78, 619 (195). [11] J. Higbie, American Journal of Physics 42, 642 (1974). [12] D. J. Griffiths, Introduction to Quantum Mechanics (Prentice-Hall, 1995).

8 8 h ν/c h ν /c FIG. 3. Schematic representation of Compton scattering. Appendix A: Derivation of the Compton Scattering Formula Consider the interaction between a photon with momentum vector p γ (and zero rest mass) and an electron initially at rest with rest mass m e as depicted in Figure 3. Call p γ and p e the momenta of the photon and the electron after the interaction, respectively. By conservation of momentum, the relation between the initial and final momenta is P φ θ Appendix B: Introduction to Timing A timing instrument marks or measures the precise occurrence of nuclear events. In practice, the release of electrons, which result from the detector s absorption of a particle s energy, occurs after the actual event; how long after is dependent upon the detector and how it is employed. The electronics processing this output further modifies its characteristics. All of these factors must be accounted for, in addition to the problem of noise, when making precise and highly resolved timing measurements. Sharp timing (2τ overlap) reduces the number of accidentals. The primary limitations in this (and other) experiments is the time slew. A pulse of lower amplitude reaches the threshold of the discriminator later. The coincidence time has to be enlarged to ensure overlap with the other normal (earlier) signal. The time slew is large because the rise time our amplifiers is slow. To overcome timing inaccuracies from slew and differing pulse shapes, the constant fraction technique is used. e p γ = p γ + p. (A1) Rearranging and squaring both sides we obtain 2 2 γ p γ + p γ 2p γ p 2 = p. e (A2) The total relativistic energy E and momentum p of a particle are related to its rest mass m by the invariant 2 4 relation (p p)c 2 E 2 = m c. By conservation of energy, q m 2 c 2 + p 2 p γ + m e c = p. (A3) γ + Rearranging and squaring both sides, we obtain e e FIG. 4. Manipulation of input signal for constant fraction discrimination p γ + p γ 2p γ p γ + 2m e c(p γ p γ ) + m c = m c + p. Constant fraction timing involves the inversion, delay (A4) and recombination of the signal to create a zero crossing Subtraction of Equation (A2) from Equation (A4) and mark. However, rather than adding the inverted and derearrangement yields layed signal to the original to achieve cancellation, some fractional part of the original is used to insure baseline m e c(p γ p γ ) = p γ p γ p γ p γ. (A5) crossover. As is evident in Figure 4, the crossover point is independent of amplitude of the input signal. The con- Finally, dividing both sides by m e cp γ pγ, we obtain for stant fraction discriminator (CFD) fixes its reference at the relation between the energies of the incident and the this point. scattered photons the equation 2 2 e e e = (1 cos θ), (A6) E m e c2 γ E γ 1. Constant Fraction Discrimination where θ is the angle between their final momentum vec- tors. CFDs have an advantage over simple discriminators in experiments that require the detection of coincidences with high time resolution. A simple discriminator produces a logic output pulse that starts at some fixed time

9 9 after the voltage at the input rises above a set value. Therefore pulses of similar shape but different peak amplitudes trigger a simple discriminator at different times relative to the occurrence times of the events that generate the input pulses. A CFD also produces a logic output pulse if the voltage of the input pulse exceeds a set value; its special virtue is that the output pulse starts at a fixed time after the input pulse has risen to a certain constant fraction of its peak value, independent of the magnitude of that peak value. If the input pulses have a fixed shape, but different amplitudes, then the delay between the event that produces the input pulse and the logic output pulse from the CFD will be a constant, which can be adjusted by the WALK control. The positive +5 VDC outputs of the CFDs are fed to a coincidence circuit that produces a logic pulse when all of its enabled input ports see pulses within the selected resolving time (user selectable within two ranges: fast coincidence 1 1 ns or slow coincidence 1 1 ns). The output pulses from the coincidence circuit are fed to a gate generator and delay module, after which, the suitably stretched and delayed gate signal is sent to the GATE input of the MCA. The amount of delay and the gate width in the delay module should be adjusted to meet the requirements of the coincident gate mode of the MCA. This requirement is that the gate pulse begin before the peak of the analog input signal and last for at least.5 µs beyond the peak. The coincidence gate requirement suppresses background events and makes the Compton scatter events stand out clearly in the resulting pulse height spectrum without the use of heavy shielding. Appendix C: Scintillation Counter Efficiency Calculations A significant complication in the analysis of the data on the differential cross section is the fact that the target is not thin, i.e. the mean free paths of the incident and scattered photons suffer substantial attenuation as it passes through the target material, and the scattered photons suffer substantial attenuation in their passage out of the target. The effects of these attenuations on the counting rate obviously depend on the position of the detector. A count must therefore be taken of these thick target effects in order to derive accurate values of the differential cross section from the measured counting rates at various scattering angles. As in all cross section measurements, the necessarily finite sizes of source, target and detector introduce into the analysis geometrical factors that require laborious multidimensional integrations for their evaluations. In our experiment we have a divergent beam of photons from a radioactive source of finite size that can interact at various depths in the target within certain ranges of solid angle to produce scattered photons that traverse the detector would be a convolution of the differential cross section with geometrical and attenuation factors represented by a multiple integral over at least ten variables with hideous limits of integration. The only practical way around this kind of mess in particle physics experiments where the highest possible precision is required is the Monte Carlo technique in which the sequence of physical processes from emission to detection is sampled by use of appropriate distribution functions to represent the various possible processes, and random numbers to select the outcome of each process. Here we outline an approximate analytical treatment of our problem based on several simplifying assumptions. The formulation is suitable for numerical computation on a modest computer. We will consider only single scattering events. We call R and H the radius and length, respectively, of the cylindrical scintillators which are the sensitive elements of the target and scatter counters. We assume that the source is so far from the target scintillator that the incident beam may be considered uniform and parallel (i.e. unidirectional). We assume that the distance D of the scatter counter scintillator from the target is so large that the detected portion of the scattered radiation may be considered to comprise a parallel (but not uniform) beam. Consider the scatterings that occur within a solid element of volume rdφdrdz at a position in the target scintillator with cylindrical coordinates r, φ, z. As can be deduced with the aid of Figure 5, the distance x of this element from the point of entry of the incident photons into the target scintillator is given by the expression x = R{[1 (q sin φ) 2 ] 1 2 q cos φ}, (C1) where q = r/r. Similarly, for photons scattered from the point P at the angle θ, the distance y of the element from the point of exit is y = R{[1 (q sin ψ) 2 ] 1 2 q cos ψ}, (C2) where ψ = π θ + φ. We call α the total linear at- FIG. 5. Diagram of the Compton scattering geometry. tenuation coefficient (units of cm 1 ) of the target plastic

10 1 scintillator for the incident photons, β(θ) the total linear attenuation coefficient of the plastic for photons scattered at angle θ, and γ(θ) the total linear attenuation coefficient of the scatter counter NaI scintillator for those scattered photons. The efficiency for detection of scattering events that occur at a given position in the target scintillator is exp( βy)[1 exp( γh)] which represents the probability that the scattered photon will escape from the target counter multiplied by the probability that the scattered photon interacts in the scatter counter. We must now find the average of this efficiency over the target scintillator. Recalling the assumption that the incident beam is parallel and uniform, the weighting factor is I exp( αx) which is the intensity of the incident beam at the point of interaction inside the target scintillator. We can now write for the average efficiency the expression R exp( αx) exp( βy)[1 exp( γh)]dv η(θ) = R, (C3) exp( αx)dv where the integrals are computed over the volume of the target scintillator. We note that η 1 as α, β. The total attenuation coefficients are functions of the energy of the scattered photon and the energy of the scattered photon is a function of the scattering angle. According to the Compton theory, E E γ (θ) =. (C4) 1 + (E /mc 2 )(1 cos θ) In the energy range from.2 MeV to.7 MeV, the total attenuation coefficient in sodium iodide can be fairly represented by the formula " # E E µ = cm 1, 1 kev 1 kev (C5) and in plastic by the formula E µ = kev.37 cm 1. (C6) The quantity η can be evaluated as a function of θ by numerical integration.

11 MIT OpenCourseWare Experimental Physics I & II "Junior Lab" Fall Spring 217 For information about citing these materials or our Terms of Use, visit:

Experimental Physics I & II "Junior Lab" Fall Spring 2008

Experimental Physics I & II Junior Lab Fall Spring 2008 MIT OpenCourseWare http://ocw.mit.edu 8.13-14 Experimental Physics I & II "Junior Lab" Fall 2007 - Spring 2008 For information about citing these materials or our Terms of Use, visit: http://ocw.mit.edu/terms.

More information

COMPTON SCATTERING. Purpose. Introduction. Fundamentals of Experiment

COMPTON SCATTERING. Purpose. Introduction. Fundamentals of Experiment COMPTON SCATTERING Purpose The purpose of this experiment is to verify the energy dependence of gamma radiation upon scattering angle and to compare the differential cross section obtained from the data

More information

Physics Laboratory Scattering of Photons from Electrons: Compton Scattering

Physics Laboratory Scattering of Photons from Electrons: Compton Scattering RR Oct 2001 SS Dec 2001 MJ Oct 2009 Physics 34000 Laboratory Scattering of Photons from Electrons: Compton Scattering Objective: To measure the energy of high energy photons scattered from electrons in

More information

Gamma Ray Spectroscopy with NaI(Tl) and HPGe Detectors

Gamma Ray Spectroscopy with NaI(Tl) and HPGe Detectors Nuclear Physics #1 Gamma Ray Spectroscopy with NaI(Tl) and HPGe Detectors Introduction: In this experiment you will use both scintillation and semiconductor detectors to study γ- ray energy spectra. The

More information

Ph 3324 The Scintillation Detector and Gamma Ray Spectroscopy

Ph 3324 The Scintillation Detector and Gamma Ray Spectroscopy Ph 3324 The Scintillation Detector and Gamma Ray Spectroscopy Required background reading Attached are several pages from an appendix on the web for Tipler-Llewellyn Modern Physics. Read the section on

More information

Physics 342 Laboratory. Scattering of Photons from Free Electrons: Compton Scattering

Physics 342 Laboratory. Scattering of Photons from Free Electrons: Compton Scattering RR Oct 2001 SS Dec 2001 Physics 342 Laboratory Scattering of Photons from Free Electrons: Compton Scattering Objective: To measure the energy of high energy photons scattered from electrons in a brass

More information

PHYSICS ADVANCED LABORATORY I COMPTON SCATTERING Spring 2002

PHYSICS ADVANCED LABORATORY I COMPTON SCATTERING Spring 2002 PHYSICS 334 - ADVANCED LABORATORY I COMPTON SCATTERING Spring 00 Purposes: Demonstrate the phenomena associated with Compton scattering and the Klein-Nishina formula. Determine the mass of the electron.

More information

ORTEC Experiment 13. Gamma-Gamma Coincidence with Angular Correlation. Equipment Required

ORTEC Experiment 13. Gamma-Gamma Coincidence with Angular Correlation. Equipment Required ORTEC Experiment 13 Equipment Required Two 905-3 2-in. x 2-in. NaI(Tl) Scintillation Detector Assemblies. Two 266 Photomultiplier Tube Bases. Two 113 Scintillation Preamplifiers. Two 556 High Voltage Power

More information

ORTEC Experiment 3. Gamma-Ray Spectroscopy Using NaI(Tl) Equipment Required. Purpose. Gamma Emission

ORTEC Experiment 3. Gamma-Ray Spectroscopy Using NaI(Tl) Equipment Required. Purpose. Gamma Emission ORTEC Experiment 3 Equipment Required Electronic Instrumentation o SPA38 Integral Assembly consisting of a 38 mm x 38 mm NaI(Tl) Scintillator, Photomultiplier Tube, and PMT Base with Stand o 4001A/4002D

More information

Instructions for gg Coincidence with 22 Na. Overview of the Experiment

Instructions for gg Coincidence with 22 Na. Overview of the Experiment Overview of the Experiment Instructions for gg Coincidence with 22 Na 22 Na is a radioactive element that decays by converting a proton into a neutron: about 90% of the time through β + decay and about

More information

K 223 Angular Correlation

K 223 Angular Correlation K 223 Angular Correlation K 223.1 Aim of the Experiment The aim of the experiment is to measure the angular correlation of a γ γ cascade. K 223.2 Required Knowledge Definition of the angular correlation

More information

Positron Emission Tomography

Positron Emission Tomography Positron Emission Tomography UBC Physics & Astronomy / PHYS 409 1 Introduction Positron emission tomography (PET) is a non-invasive way to produce the functional 1 image of a patient. It works by injecting

More information

Physics Experiment N -17. Lifetime of Cosmic Ray Muons with On-Line Data Acquisition on a Computer

Physics Experiment N -17. Lifetime of Cosmic Ray Muons with On-Line Data Acquisition on a Computer Introduction Physics 410-510 Experiment N -17 Lifetime of Cosmic Ray Muons with On-Line Data Acquisition on a Computer The experiment is designed to teach the techniques of particle detection using scintillation

More information

Atomic and Nuclear Physics

Atomic and Nuclear Physics Atomic and Nuclear Physics Nuclear physics -spectroscopy LEYBOLD Physics Leaflets Detecting radiation with a scintillation counter Objects of the experiments Studying the scintillator pulses with an oscilloscope

More information

LIFETIME OF THE MUON

LIFETIME OF THE MUON Muon Decay 1 LIFETIME OF THE MUON Introduction Muons are unstable particles; otherwise, they are rather like electrons but with much higher masses, approximately 105 MeV. Radioactive nuclear decays do

More information

GAMMA-GAMMA CORRELATION Latest Revision: August 21, 2007

GAMMA-GAMMA CORRELATION Latest Revision: August 21, 2007 C1-1 GAMMA-GAMMA CORRELATION Latest Revision: August 21, 2007 QUESTION TO BE INVESTIGATED: decay event? What is the angular correlation between two gamma rays emitted by a single INTRODUCTION & THEORY:

More information

Scintillation Counters

Scintillation Counters PHY311/312 Detectors for Nuclear and Particle Physics Dr. C.N. Booth Scintillation Counters Unlike many other particle detectors, which exploit the ionisation produced by the passage of a charged particle,

More information

ORTEC. AN34 Experiment 14 Nuclear Lifetimes and the Coincidence Method. Equipment Needed from ORTEC. Equipment Required from Other Manufacturers

ORTEC. AN34 Experiment 14 Nuclear Lifetimes and the Coincidence Method. Equipment Needed from ORTEC. Equipment Required from Other Manufacturers Equipment Needed from ORTEC Two 113 Scintillation Preamplifiers Two 266 Photomultiplier Tube Bases 4001A/4002D Bin and Power Supply 414A Fast Coincidence Two 551 Timing Single-Channel Analyzers 567 Time-to-Amplitude

More information

Nuclear Experiment Phys 318/317 Room 208 Instructor Richard Lindgren Room 302 Ext

Nuclear Experiment Phys 318/317 Room 208 Instructor Richard Lindgren Room 302 Ext Revised Aug 28, 2008 Nuclear Experiment Phys 318/317 Room 208 Instructor Richard Lindgren Room 302 Ext 2-2691 ral5q@virginia.edu 1 NOTE: Some of the figures referred to in this document can be found in

More information

Purpose This experiment will use the coincidence method for time correlation to measure the lifetime in the decay scheme of 57

Purpose This experiment will use the coincidence method for time correlation to measure the lifetime in the decay scheme of 57 Equipment Required Two 113 Scintillation Preamplifiers Two 266 Photomultiplier Tube Bases 4001A/4002D Bin and Power Supply 414A Fast Coincidence Two 551 Timing Single-Channel Analyzers 567 Time-to-Amplitude

More information

AN ABSTRACT ON THE THESIS OF. David C. Vasquez for the degree of Master of Science in Radiation Health Physics presented on February 26, 2010.

AN ABSTRACT ON THE THESIS OF. David C. Vasquez for the degree of Master of Science in Radiation Health Physics presented on February 26, 2010. AN ABSTRACT ON THE THESIS OF David C. Vasquez for the degree of Master of Science in Radiation Health Physics presented on February 26, 2010. Title: The Design, Use and Implementation of Digital Radiation

More information

ORTEC Experiment 19. Gamma-Ray Decay Scheme and Angular Correlation for 60 Co. Equipment Required. Purpose. Introduction

ORTEC Experiment 19. Gamma-Ray Decay Scheme and Angular Correlation for 60 Co. Equipment Required. Purpose. Introduction ORTEC Experiment 19 Equipment Required Two 905-3 NaI(Tl) 2- x 2-in. Detectors with Phototubes. Two 266 PMT Bases. Two 556 High Voltage Power Supplies. Two 113 Scintillation Preamplifiers. Two 575A Amplifiers.

More information

Fast first practical help -- detailed instructions will follow- preliminary Experiment F80

Fast first practical help -- detailed instructions will follow- preliminary Experiment F80 Fast first practical help -- detailed instructions will follow- preliminary Experiment F80 Measurement Methods of Nuclear and Particle Physics Introduction: This experiment is going to introduce you to

More information

Electronic Instrumentation for Radiation Detection Systems

Electronic Instrumentation for Radiation Detection Systems Electronic Instrumentation for Radiation Detection Systems January 23, 2018 Joshua W. Cates, Ph.D. and Craig S. Levin, Ph.D. Course Outline Lecture Overview Brief Review of Radiation Detectors Detector

More information

80 Physics Essentials Workbook Stage 2 Physics

80 Physics Essentials Workbook Stage 2 Physics 80 Physics Essentials Workbook Stage 2 Physics the thickness of the tissue: Obviously, the thicker the tissue through which the X-rays have to pass the more they will be absorbed from the beam passing

More information

CAEN Tools for Discovery

CAEN Tools for Discovery Viareggio 5 September 211 Introduction In recent years CAEN has developed a complete family of digitizers that consists of several models differing in sampling frequency, resolution, form factor and other

More information

1 Purpose of This Lab Exercise:

1 Purpose of This Lab Exercise: Physics 4796 - Experimental Physics Temple University, Spring 2010-11 C. J. Martoff, Instructor J. Tatarowicz, TA Physics 4796 Lab Writeup Hunting for Antimatter with NaI Spectroscopy 1 Purpose of This

More information

Scintillators as an external trigger for cathode strip chambers

Scintillators as an external trigger for cathode strip chambers Scintillators as an external trigger for cathode strip chambers J. A. Muñoz Department of Physics, Princeton University, Princeton, NJ 08544 An external trigger was set up to test cathode strip chambers

More information

Modern Physics Laboratory MP4 Photoelectric Effect

Modern Physics Laboratory MP4 Photoelectric Effect Purpose MP4 Photoelectric Effect In this experiment, you will investigate the photoelectric effect and determine Planck s constant and the work function. Equipment and components Photoelectric Effect Apparatus

More information

Silicon Photomultiplier Evaluation Kit. Quick Start Guide. Eval Kit SiPM. KETEK GmbH. Hofer Str Munich Germany.

Silicon Photomultiplier Evaluation Kit. Quick Start Guide. Eval Kit SiPM. KETEK GmbH. Hofer Str Munich Germany. KETEK GmbH Hofer Str. 3 81737 Munich Germany www.ketek.net info@ketek.net phone +49 89 673 467 70 fax +49 89 673 467 77 Silicon Photomultiplier Evaluation Kit Quick Start Guide Eval Kit Table of Contents

More information

Traditional analog QDC chain and Digital Pulse Processing [1]

Traditional analog QDC chain and Digital Pulse Processing [1] Giuliano Mini Viareggio April 22, 2010 Introduction The aim of this paper is to compare the energy resolution of two gamma ray spectroscopy setups based on two different acquisition chains; the first chain

More information

Electron Spin Resonance v2.0

Electron Spin Resonance v2.0 Electron Spin Resonance v2.0 Background. This experiment measures the dimensionless g-factor (g s ) of an unpaired electron using the technique of Electron Spin Resonance, also known as Electron Paramagnetic

More information

Atomic and nuclear physics

Atomic and nuclear physics Atomic and nuclear physics X-ray physics Physics of the atomic shell LEYBOLD Physics Leaflets Investigating the energy spectrum of an x-ray tube as a function of the high voltage and the emission current

More information

Experiment 10. The Speed of Light c Introduction Apparatus

Experiment 10. The Speed of Light c Introduction Apparatus Experiment 10 The Speed of Light c 10.1 Introduction In this experiment you will measure the speed of light, c. This is one of the most fundamental constants in physics, and at the same time the fastest

More information

Energy Measurements with a Si Surface Barrier Detector and a 5.5-MeV 241 Am α Source

Energy Measurements with a Si Surface Barrier Detector and a 5.5-MeV 241 Am α Source Energy Measurements with a Si Surface Barrier Detector and a 5.5-MeV 241 Am α Source October 18, 2017 The goals of this experiment are to become familiar with semiconductor detectors, which are widely

More information

PMT tests at UMD. Vlasios Vasileiou Version st May 2006

PMT tests at UMD. Vlasios Vasileiou Version st May 2006 PMT tests at UMD Vlasios Vasileiou Version 1.0 1st May 2006 Abstract This memo describes the tests performed on three Milagro PMTs in UMD. Initially, pulse-height distributions of the PMT signals were

More information

SAMPLE QUESTION PAPER CLASS-XII. Physics(Theory)

SAMPLE QUESTION PAPER CLASS-XII. Physics(Theory) SAMPLE QUESTION PAPER CLASS-XII Time allowed: 3 Hrs Physics(Theory) Maximum Marks: 70 GENERAL INSTRUCTIONS: 1. All questions are compulsory. 2. There are 29 questions in total. Questions 1 to 8 are very

More information

OPERATING CHARACTERISTICS OF THE GEIGER COUNTER

OPERATING CHARACTERISTICS OF THE GEIGER COUNTER OPERATING CHARACTERISTICS OF THE GEIGER COUNTER OBJECTIVE The objective of this laboratory is to determine the operating voltage for a Geiger tube and to calculate the effect of the dead time and recovery

More information

and N(t) ~ exp(-t/ ),

and N(t) ~ exp(-t/ ), Muon Lifetime Experiment Introduction Charged and neutral particles with energies in excess of 10 23 ev from Galactic and extra Galactic sources impinge on the earth. Here we speak of the earth as the

More information

Gamma Spectrometer Initial Project Proposal

Gamma Spectrometer Initial Project Proposal Gamma Spectrometer Initial Project Proposal Group 9 Aman Kataria Johnny Klarenbeek Dean Sullivan David Valentine Introduction There are currently two main types of gamma radiation detectors used for gamma

More information

Chapter Ray and Wave Optics

Chapter Ray and Wave Optics 109 Chapter Ray and Wave Optics 1. An astronomical telescope has a large aperture to [2002] reduce spherical aberration have high resolution increase span of observation have low dispersion. 2. If two

More information

A Measurement of the Photon Detection Efficiency of Silicon Photomultipliers

A Measurement of the Photon Detection Efficiency of Silicon Photomultipliers A Measurement of the Photon Detection Efficiency of Silicon Photomultipliers A. N. Otte a,, J. Hose a,r.mirzoyan a, A. Romaszkiewicz a, M. Teshima a, A. Thea a,b a Max Planck Institute for Physics, Föhringer

More information

--- preliminary Experiment F80

--- preliminary Experiment F80 --- preliminary Experiment F80 Measurement Methods of Nuclear and Particle Physics Introduction: This experiment is going to introduce you to important counting and measuring techniques of nuclear and

More information

DOE FUNDAMENTALS HANDBOOK INSTRUMENTATION AND CONTROL Volume 2 of 2

DOE FUNDAMENTALS HANDBOOK INSTRUMENTATION AND CONTROL Volume 2 of 2 DOE-HDBK-1013/2-92 JUNE 1992 DOE FUNDAMENTALS HANDBOOK INSTRUMENTATION AND CONTROL Volume 2 of 2 U.S. Department of Energy Washington, D.C. 20585 FSC-6910 Distribution Statement A. Approved for public

More information

Laser Locking with Doppler-free Saturated Absorption Spectroscopy

Laser Locking with Doppler-free Saturated Absorption Spectroscopy Laser Locking with Doppler-free Saturated Absorption Spectroscopy Paul L. Stubbs, Advisor: Irina Novikova W&M Quantum Optics Group May 12, 2010 Abstract The goal of this project was to lock the frequency

More information

I = I 0 cos 2 θ (1.1)

I = I 0 cos 2 θ (1.1) Chapter 1 Faraday Rotation Experiment objectives: Observe the Faraday Effect, the rotation of a light wave s polarization vector in a material with a magnetic field directed along the wave s direction.

More information

ORTEC. Research Applications. Pulse-Height, Charge, or Energy Spectroscopy. Detectors. Processing Electronics

ORTEC. Research Applications. Pulse-Height, Charge, or Energy Spectroscopy. Detectors. Processing Electronics ORTEC Spectroscopy systems for ORTEC instrumentation produce pulse height distributions of gamma ray or alpha energies. MAESTRO-32 (model A65-B32) is the software included with most spectroscopy systems

More information

ORTEC. Time-to-Amplitude Converters and Time Calibrator. Choosing the Right TAC. Timing with TACs

ORTEC. Time-to-Amplitude Converters and Time Calibrator. Choosing the Right TAC. Timing with TACs ORTEC Time-to-Amplitude Converters Choosing the Right TAC The following topics provide the information needed for selecting the right time-to-amplitude converter (TAC) for the task. The basic principles

More information

Homework Set 3.5 Sensitive optoelectronic detectors: seeing single photons

Homework Set 3.5 Sensitive optoelectronic detectors: seeing single photons Homework Set 3.5 Sensitive optoelectronic detectors: seeing single photons Due by 12:00 noon (in class) on Tuesday, Nov. 7, 2006. This is another hybrid lab/homework; please see Section 3.4 for what you

More information

Today s Outline - January 25, C. Segre (IIT) PHYS Spring 2018 January 25, / 26

Today s Outline - January 25, C. Segre (IIT) PHYS Spring 2018 January 25, / 26 Today s Outline - January 25, 2018 C. Segre (IIT) PHYS 570 - Spring 2018 January 25, 2018 1 / 26 Today s Outline - January 25, 2018 HW #2 C. Segre (IIT) PHYS 570 - Spring 2018 January 25, 2018 1 / 26 Today

More information

Experiment 6: Franck Hertz Experiment v1.3

Experiment 6: Franck Hertz Experiment v1.3 Experiment 6: Franck Hertz Experiment v1.3 Background This series of experiments demonstrates the energy quantization of atoms. The concept was first implemented by James Franck and Gustaf Ludwig Hertz

More information

The Influence of Crystal Configuration and PMT on PET Time-of-Flight Resolution

The Influence of Crystal Configuration and PMT on PET Time-of-Flight Resolution The Influence of Crystal Configuration and PMT on PET Time-of-Flight Resolution Christopher Thompson Montreal Neurological Institute and Scanwell Systems, Montreal, Canada Jason Hancock Cross Cancer Institute,

More information

Detecting and Suppressing Background Signal

Detecting and Suppressing Background Signal Detecting and Suppressing Background Signal Valerie Gray St. Norbert College Advisors: Dr. Michael Wiescher Freimann Professor Nuclear Physics University of Notre Dame Dr. Ed Stech Associate Professional

More information

PH2510 Nuclear Physics Laboratory Use of Scintillation Counters (NP5)

PH2510 Nuclear Physics Laboratory Use of Scintillation Counters (NP5) Physics Department Royal Holloway University of London PH2510 Nuclear Physics Laboratory Use of Scintillation Counters (NP5) 1. Introduction 1.1 Object of the Experiment The object of this experiment is

More information

FIBER OPTICS. Prof. R.K. Shevgaonkar. Department of Electrical Engineering. Indian Institute of Technology, Bombay. Lecture: 37

FIBER OPTICS. Prof. R.K. Shevgaonkar. Department of Electrical Engineering. Indian Institute of Technology, Bombay. Lecture: 37 FIBER OPTICS Prof. R.K. Shevgaonkar Department of Electrical Engineering Indian Institute of Technology, Bombay Lecture: 37 Introduction to Raman Amplifiers Fiber Optics, Prof. R.K. Shevgaonkar, Dept.

More information

CAEN. Electronic Instrumentation. CAEN Silicon Photomultiplier Kit

CAEN. Electronic Instrumentation. CAEN Silicon Photomultiplier Kit CAEN Tools for Discovery Electronic Instrumentation CAEN Silicon Photomultiplier Kit CAEN realized a modular development kit dedicated to Silicon Photomultipliers, representing the state-of-the art in

More information

The 34th International Physics Olympiad

The 34th International Physics Olympiad The 34th International Physics Olympiad Taipei, Taiwan Experimental Competition Wednesday, August 6, 2003 Time Available : 5 hours Please Read This First: 1. Use only the pen provided. 2. Use only the

More information

Spontaneous Fission Spectrum of Neutrons from 252 Cf with Kinetic Energies Less than 1 MeV. Suraj Bastola. A senior thesis submitted to the faculty of

Spontaneous Fission Spectrum of Neutrons from 252 Cf with Kinetic Energies Less than 1 MeV. Suraj Bastola. A senior thesis submitted to the faculty of Spontaneous Fission Spectrum of Neutrons from 252 Cf with Kinetic Energies Less than 1 MeV Suraj Bastola A senior thesis submitted to the faculty of Brigham Young University in partial fulfillment of the

More information

Spectroscopy of Ruby Fluorescence Physics Advanced Physics Lab - Summer 2018 Don Heiman, Northeastern University, 1/12/2018

Spectroscopy of Ruby Fluorescence Physics Advanced Physics Lab - Summer 2018 Don Heiman, Northeastern University, 1/12/2018 1 Spectroscopy of Ruby Fluorescence Physics 3600 - Advanced Physics Lab - Summer 2018 Don Heiman, Northeastern University, 1/12/2018 I. INTRODUCTION The laser was invented in May 1960 by Theodor Maiman.

More information

Method for digital particle spectrometry Khryachkov Vitaly

Method for digital particle spectrometry Khryachkov Vitaly Method for digital particle spectrometry Khryachkov Vitaly Institute for physics and power engineering (IPPE) Obninsk, Russia The goals of Analog Signal Processing Signal amplification Signal filtering

More information

GSI Helmholtzzentrum für Schwerionenforschung. Station Coincidence

GSI Helmholtzzentrum für Schwerionenforschung. Station Coincidence GSI Helmholtzzentrum für Schwerionenforschung Station 7 --Coincidence --Coincidence Content Content... 2 The positron-electron annihilation radiation from Na-22...3 What this is about...3 E.1. The experiment...3

More information

Pulse Transmission and Cable Properties ================================

Pulse Transmission and Cable Properties ================================ PHYS 4211 Fall 2005 Last edit: October 2, 2006 T.E. Coan Pulse Transmission and Cable Properties ================================ GOAL To understand how voltage and current pulses are transmitted along

More information

PMT Calibration in the XENON 1T Demonstrator. Abstract

PMT Calibration in the XENON 1T Demonstrator. Abstract PMT Calibration in the XENON 1T Demonstrator Sarah Vickery Nevis Laboratories, Columbia University, Irvington, NY 10533 USA (Dated: August 2, 2013) Abstract XENON Dark Matter Project searches for the dark

More information

The (Speed and) Decay of Cosmic-Ray Muons

The (Speed and) Decay of Cosmic-Ray Muons The (Speed and) Decay of Cosmic-Ray Muons Jason Gross MIT - Department of Physics Jason Gross (8.13) Cosmic-Ray Muons November 4, 2011 1 / 30 Goals test relativity (time dilation) determine the mean lifetime

More information

SUPPLEMENTARY INFORMATION

SUPPLEMENTARY INFORMATION SUPPLEMENTARY INFORMATION doi:10.1038/nature10864 1. Supplementary Methods The three QW samples on which data are reported in the Letter (15 nm) 19 and supplementary materials (18 and 22 nm) 23 were grown

More information

A MONTE CARLO CODE FOR SIMULATION OF PULSE PILE-UP SPECTRAL DISTORTION IN PULSE-HEIGHT MEASUREMENT

A MONTE CARLO CODE FOR SIMULATION OF PULSE PILE-UP SPECTRAL DISTORTION IN PULSE-HEIGHT MEASUREMENT Copyright JCPDS - International Centre for Diffraction Data 2005, Advances in X-ray Analysis, Volume 48. 246 A MONTE CARLO CODE FOR SIMULATION OF PULSE PILE-UP SPECTRAL DISTORTION IN PULSE-HEIGHT MEASUREMENT

More information

Digital trigger system for the RED-100 detector based on the unit in VME standard

Digital trigger system for the RED-100 detector based on the unit in VME standard Journal of Physics: Conference Series PAPER OPEN ACCESS Digital trigger system for the RED-100 detector based on the unit in VME standard To cite this article: D Yu Akimov et al 2016 J. Phys.: Conf. Ser.

More information

Radiation Detection Instrumentation

Radiation Detection Instrumentation Radiation Detection Instrumentation Principles of Detection and Gas-filled Ionization Chambers Neutron Sensitive Ionization Chambers Detection of radiation is a consequence of radiation interaction with

More information

X-rays. X-rays are produced when electrons are accelerated and collide with a target. X-rays are sometimes characterized by the generating voltage

X-rays. X-rays are produced when electrons are accelerated and collide with a target. X-rays are sometimes characterized by the generating voltage X-rays Ouch! 1 X-rays X-rays are produced when electrons are accelerated and collide with a target Bremsstrahlung x-rays Characteristic x-rays X-rays are sometimes characterized by the generating voltage

More information

EXPERIMENT 3 THE PHOTOELECTRIC EFFECT

EXPERIMENT 3 THE PHOTOELECTRIC EFFECT EXPERIMENT 3 THE PHOTOELECTRIC EFFECT Equipment List Included Equipment 1. Mercury Light Source Enclosure 2. Track, 60 cm 3. Photodiode Enclosure 4. Mercury Light Source Power Supply 5. DC Current Amplifier

More information

ANTICOINCIDENCE LOW LEVEL COUNTING

ANTICOINCIDENCE LOW LEVEL COUNTING Med Phys 4RB3/6R3 LABORATORY EXPERIMENT #7 ANTICOINCIDENCE LOW LEVEL COUNTING Introduction This is the only experiment in this series which involves a multi- system. The low-level electronics used was

More information

NM Module Section 2 6 th Edition Christian, Ch. 3

NM Module Section 2 6 th Edition Christian, Ch. 3 NM 4303 Module Section 2 6 th Edition Christian, Ch. 3 Gas Filled Chamber Voltage Gas filled chamber uses Hand held detectors cutie pie Geiger counter Dose calibrators Cutie pie Chamber voltage in Ionization

More information

Photon Counters SR430 5 ns multichannel scaler/averager

Photon Counters SR430 5 ns multichannel scaler/averager Photon Counters SR430 5 ns multichannel scaler/averager SR430 Multichannel Scaler/Averager 5 ns to 10 ms bin width Count rates up to 100 MHz 1k to 32k bins per record Built-in discriminator No interchannel

More information

Applications of Monte Carlo Methods in Charged Particles Optics

Applications of Monte Carlo Methods in Charged Particles Optics Sydney 13-17 February 2012 p. 1/3 Applications of Monte Carlo Methods in Charged Particles Optics Alla Shymanska alla.shymanska@aut.ac.nz School of Computing and Mathematical Sciences Auckland University

More information

Attenuation length in strip scintillators. Jonathan Button, William McGrew, Y.-W. Lui, D. H. Youngblood

Attenuation length in strip scintillators. Jonathan Button, William McGrew, Y.-W. Lui, D. H. Youngblood Attenuation length in strip scintillators Jonathan Button, William McGrew, Y.-W. Lui, D. H. Youngblood I. Introduction The ΔE-ΔE-E decay detector as described in [1] is composed of thin strip scintillators,

More information

Testing the Electronics for the MicroBooNE Light Collection System

Testing the Electronics for the MicroBooNE Light Collection System Testing the Electronics for the MicroBooNE Light Collection System Kathleen V. Tatem Nevis Labs, Columbia University & Fermi National Accelerator Laboratory August 3, 2012 Abstract This paper discusses

More information

EKA Laboratory Muon Lifetime Experiment Instructions. October 2006

EKA Laboratory Muon Lifetime Experiment Instructions. October 2006 EKA Laboratory Muon Lifetime Experiment Instructions October 2006 0 Lab setup and singles rate. When high-energy cosmic rays encounter the earth's atmosphere, they decay into a shower of elementary particles.

More information

ORTEC Experiment 9. Time Coincidence Techniques Applied to Absolute Activity Measurements. Equipment Required. Purpose.

ORTEC Experiment 9. Time Coincidence Techniques Applied to Absolute Activity Measurements. Equipment Required. Purpose. ORTEC Experiment 9 Equipment Required 905-3 2-inch x 2-inch (5.08-cm x 5.08-cm) NaI(Tl) Detector and PM Tube Assembly. 266 PhotoMultiplier Tube Base. Two 113 Preamplifiers. BA-015-050-1000 Partially-Depleted

More information

EXPERIMENT 5. SCINTILLATION COUNTING AND QUENCH CORRECTION.

EXPERIMENT 5. SCINTILLATION COUNTING AND QUENCH CORRECTION. 59 EXPERIMENT 5. SCINTILLATION COUNTING AND QUENCH CORRECTION. (The report for this experiment is due 1 week after the completion of the experiment) 5.1 Introduction Liquid scintillation is the method

More information

Guided Propagation Along the Optical Fiber. Xavier Fernando Ryerson University

Guided Propagation Along the Optical Fiber. Xavier Fernando Ryerson University Guided Propagation Along the Optical Fiber Xavier Fernando Ryerson University The Nature of Light Quantum Theory Light consists of small particles (photons) Wave Theory Light travels as a transverse electromagnetic

More information

Ph 3455 The Photoelectric Effect

Ph 3455 The Photoelectric Effect Ph 3455 The Photoelectric Effect Required background reading Tipler, Llewellyn, section 3-3 Prelab Questions 1. In this experiment you will be using a mercury lamp as the source of photons. At the yellow

More information

Single Photon Interference Laboratory

Single Photon Interference Laboratory Single Photon Interference Laboratory Renald Dore Institute of Optics University of Rochester, Rochester, NY 14627, U.S.A Abstract The purpose of our laboratories was to observe the wave-particle duality

More information

KULLIYYAH OF ENGINEERING

KULLIYYAH OF ENGINEERING KULLIYYAH OF ENGINEERING DEPARTMENT OF ELECTRICAL & COMPUTER ENGINEERING ANTENNA AND WAVE PROPAGATION LABORATORY (ECE 4103) EXPERIMENT NO 3 RADIATION PATTERN AND GAIN CHARACTERISTICS OF THE DISH (PARABOLIC)

More information

ORTEC Experiment 1. Introduction to Electronic Signal Analysis in Nuclear Radiation Measurements. Equipment Required: Purpose. Electronic Circuits

ORTEC Experiment 1. Introduction to Electronic Signal Analysis in Nuclear Radiation Measurements. Equipment Required: Purpose. Electronic Circuits ORTEC Experiment 1 Equipment Required: 480 Pulser 113 Scintillation Preamplifier 4001A/4002D NIM Bin and Power Supply 575A Spectroscopy Amplifier 996 Timer and Counter 551 Timing Single-Channel Analyzer

More information

Multi-Path Fading Channel

Multi-Path Fading Channel Instructor: Prof. Dr. Noor M. Khan Department of Electronic Engineering, Muhammad Ali Jinnah University, Islamabad Campus, Islamabad, PAKISTAN Ph: +9 (51) 111-878787, Ext. 19 (Office), 186 (Lab) Fax: +9

More information

Guided Propagation Along the Optical Fiber. Xavier Fernando Ryerson Comm. Lab

Guided Propagation Along the Optical Fiber. Xavier Fernando Ryerson Comm. Lab Guided Propagation Along the Optical Fiber Xavier Fernando Ryerson Comm. Lab The Nature of Light Quantum Theory Light consists of small particles (photons) Wave Theory Light travels as a transverse electromagnetic

More information

Designing an MR compatible Time of Flight PET Detector Floris Jansen, PhD, Chief Engineer GE Healthcare

Designing an MR compatible Time of Flight PET Detector Floris Jansen, PhD, Chief Engineer GE Healthcare GE Healthcare Designing an MR compatible Time of Flight PET Detector Floris Jansen, PhD, Chief Engineer GE Healthcare There is excitement across the industry regarding the clinical potential of a hybrid

More information

Goal of the project. TPC operation. Raw data. Calibration

Goal of the project. TPC operation. Raw data. Calibration Goal of the project The main goal of this project was to realise the reconstruction of α tracks in an optically read out GEM (Gas Electron Multiplier) based Time Projection Chamber (TPC). Secondary goal

More information

Solid-State Photomultiplier in CMOS Technology for Gamma-Ray Detection and Imaging Applications

Solid-State Photomultiplier in CMOS Technology for Gamma-Ray Detection and Imaging Applications Solid-State Photomultiplier in CMOS Technology for Gamma-Ray Detection and Imaging Applications Christopher Stapels, Member, IEEE, William G. Lawrence, James Christian, Member, IEEE, Michael R. Squillante,

More information

DUANE-HUNT RELATION AND DETERMINATION OF PLANCK S CONSTANT

DUANE-HUNT RELATION AND DETERMINATION OF PLANCK S CONSTANT DUANE-HUNT RELATION AND DETERMINATION OF PLANCK S CONSTANT OBJECTIVES To determine the limit wavelength min of the bremsstrahlung continuum as a function of the high voltage U of the x-ray tube. To confirm

More information

Rayleigh scattering in X-ray polarimetry

Rayleigh scattering in X-ray polarimetry Rayleigh scattering in X-ray polarimetry Andrey Bondarev Saint-Petersburg State University, b.ndarev@gmail.com In this report we present experimental investigations of Rayleigh scattering for high energy

More information

On the initiation of lightning in thunderclouds (Instrumentation, Supplementary information)

On the initiation of lightning in thunderclouds (Instrumentation, Supplementary information) On the initiation of lightning in thunderclouds (Instrumentation, Supplementary information) Ashot Chilingarian 1,2, Suren Chilingaryan 1, Tigran Karapetyan 1, Lev Kozliner 1, Yeghia Khanikyants 1, Gagik

More information

BLACKBODY RADIATION PHYSICS 359E

BLACKBODY RADIATION PHYSICS 359E BLACKBODY RADIATION PHYSICS 359E INTRODUCTION In this laboratory, you will make measurements intended to illustrate the Stefan-Boltzmann Law for the total radiated power per unit area I tot (in W m 2 )

More information

Report on BLP Spectroscopy Experiments Conducted on October 6, 2017: M. Nansteel

Report on BLP Spectroscopy Experiments Conducted on October 6, 2017: M. Nansteel Report on BLP Spectroscopy Experiments Conducted on October 6, 2017: M. Nansteel Summary Several spectroscopic measurements were conducted on October 6, 2017 at BLP to characterize the radiant power of

More information

Vertical Tests of ILC Cavities and Detection of X-Rays from Field Emission

Vertical Tests of ILC Cavities and Detection of X-Rays from Field Emission Vertical Tests of ILC Cavities and Detection of X-Rays from Field Emission Pardis Niknejadi California State Polytechnic University, Pomona, CA 91768 Elizabeth Olhsson University of Oregon, Eugene, OR

More information

R. J. Jones College of Optical Sciences OPTI 511L Fall 2017

R. J. Jones College of Optical Sciences OPTI 511L Fall 2017 R. J. Jones College of Optical Sciences OPTI 511L Fall 2017 Active Modelocking of a Helium-Neon Laser The generation of short optical pulses is important for a wide variety of applications, from time-resolved

More information

OPAC 202 Optical Design and Instrumentation. Topic 3 Review Of Geometrical and Wave Optics. Department of

OPAC 202 Optical Design and Instrumentation. Topic 3 Review Of Geometrical and Wave Optics. Department of OPAC 202 Optical Design and Instrumentation Topic 3 Review Of Geometrical and Wave Optics Department of http://www.gantep.edu.tr/~bingul/opac202 Optical & Acustical Engineering Gaziantep University Feb

More information

1. Define attenuation of radiation, half-value layer, and tenth value layer.

1. Define attenuation of radiation, half-value layer, and tenth value layer. Objectives for Tutorial on Attenuation of Radiation 1. Define attenuation of radiation, half-value layer, and tenth value layer. 2. Explain what happens to intensity of a beam as one removes or adds HVLs

More information

The Speed of Light Laboratory Experiment 8. Introduction

The Speed of Light Laboratory Experiment 8. Introduction Exp-8-Speed of Light.doc (TJR) Physics Department, University of Windsor 64-311 Laboratory Experiment 8 The Speed of Light Introduction Galileo was right. Light did not travel instantaneously as his contemporaries

More information

Photoelectric effect

Photoelectric effect Photoelectric effect Objective Study photoelectric effect. Measuring and Calculating Planck s constant, h. Measuring Current-Voltage Characteristics of photoelectric Spectral Lines. Theory Experiments

More information