LASIK for Myopia and Astigmatism Using the SCHWIND AMARIS Excimer Laser: An International Multicenter Trial

Size: px
Start display at page:

Download "LASIK for Myopia and Astigmatism Using the SCHWIND AMARIS Excimer Laser: An International Multicenter Trial"

Transcription

1 ORIGINAL ARTICLES LASIK for Myopia and Astigmatism Using the SCHWIND AMARIS Excimer Laser: An International Multicenter Trial Maria Clara Arbelaez, MD; Ioannis M. Aslanides, MD; Carmen Barraquer, MD; Francesco Carones, MD; Alena Feuermannova, MD; Tobias Neuhann, MD; Pavel Rozsival, MD ABSTRACT PURPOSE: To assess the effi cacy, predictability, and safety of LASIK for the surgical correction of low to moderate myopia with astigmatism using the SCHWIND AMARIS excimer laser. METHODS: Six international study sites enrolled 358 eyes with a manifest refraction spherical equivalent (MRSE) from 0.50 to 7.38 diopters (D) (mean sphere: D) with up to 5.00 D of astigmatism (mean: D). All eyes underwent treatment with the nonwavefront-guided aspheric algorithm of the SCHWIND AMARIS excimer laser. All eyes were targeted for emmetropia. Refractive outcomes and corneal higher order aberrations were analyzed pre- and postoperatively. Visual quality was assessed using photopic and mesopic contrast sensitivity. Six-month postoperative outcomes are reported. RESULTS: At 6 months postoperative, the MRSE for all eyes was D, and 96% (343/358) of eyes had MRSE within 0.50 D. Uncorrected visual acuity was 20/20 or better in 98% (351/358) of eyes, and no eyes lost 2 or more lines of best spectacle-corrected visual acuity. The total corneal higher order aberrations root-mean-square increased by 0.09 µm, spherical aberration increased by 0.08 µm, and coma increased by 0.04 µm postoperatively. Photopic and mesopic contrast sensitivity did not change 6 months postoperatively. CONCLUSIONS: Treatment of myopia with astigmatism using the SCHWIND AMARIS excimer laser is safe, effi cacious, predictable, and maintains visual quality. [J Refract Surg. 2010;26:88-98.] doi: / x W avefront, aspheric, and conventional LASIK for the treatment of myopia and myopic astigmatism is safe and effective. 1,2 Recent advances in excimer laser technology, such as the use of aspheric ablation profiles, incorporation of higher order aberration treatment, and eye trackers, have presumably led to better refractive outcomes and reduced higher order aberration induction postoperatively. 3,4 Although most laser manufacturers incorporate the suite of products mentioned above, 1 the use of high repetition rates (500 Hz or higher) to reduce treatment times has not been widely adopted. This is likely due to technological constraints and the increased thermal effect concomitant with the use of higher repetition rates. 5,6 If left unaddressed, the thermal effect can cause tissue damage 5,6 and potentially reduce refractive outcomes. 7 The reduction of treatment times may be beneficial for the following reasons: 1) shorter treatment times may result in less stromal dehydration, and 2) patients are less likely to lose fixation during the ablation. 8,9 The SCHWIND AMARIS (SCHWIND eye-tech-solutions GmbH & Co KG, Kleinostheim, Germany) is a commercially available flying spot excimer laser that utilizes a 500-Hz repetition rate with an aspheric ablation algorithm for refractive treatments. One ablation algorithm, the Aberration-Free algorithm, available in this platform, aims at maintaining the preoperative levels of higher order aberrations. 10 Reasonable reductions in higher order aberrations after wavefront-guided treatments on aberrated eyes and reason- From Muscat Eye Laser Center, Muscat, Oman (Arbelaez); Emmetropia Mediterranean Eye Institute, Heraklion, Crete, Greece (Aslanides); Clinica Barraquer, Bogotá, Colombia (Barraquer); Centro Oftalmo-Chirurgico Carones, Milano, Italy (Carones); Klinika Oftalmologicka, Lekarske Fakulty Univ. Karlovy, Kralove, Czech Republic (Feuermannova, Rozsival); and Augenklinik am Marienplatz, München, Germany (Neuhann). The authors have no proprietary interest in the materials presented herein. Correspondence: Maria Clara Arbelaez, MD, Muscat Eye Laser Center, PO Box 938 PC, 117 Wadi al Kabir, Muscat, Oman. Tel: ; Fax: ; drmaria@omantel.net.om Received: May 19, 2008; Accepted: February 10, 2009 Posted online: April 15,

2 able changes in higher order aberrations after wavefront-optimized treatments have been reported. 11,12 However, ocular wavefront-guided and conventional treatments can increase higher order aberrations by 100% postoperatively. 3 A significant number of refractive surgery patients may not benefit from ocular wavefront-guided treatment as the induction of higher order aberrations is related to baseline levels of higher order aberrations. 3,13 For example, higher order aberrations tend to be induced in patients with 0.30 µm and reduced in patients with 0.30 µm of higher order aberrations. 3,13 Physiologic optical aberrations may be required to maintain the optical quality of the eye. 14,15 Based on these studies 3,13-15 it seems that customized ablation algorithms in any form (ocular wavefrontguided, corneal wavefront-guided, topography-guided, etc) may not be appropriate for the entire refractive surgery population (ie, specific population groups have specific demands and deserve specific treatment solutions). No one-size-fits-all concept can be applied. Our definition of customization is conceptually different and can be stated as: The planning of the most optimum ablation pattern specifically for each individual eye is based on its diagnosis and visual demands. It is often the case that the best approach for planning an ablation is a sophisticated pattern, which can still be simply described in terms of sphere, cylinder, and orientation (axis). The present study evaluated the safety, predictability, efficacy, aberrations, and contrast sensitivity of LASIK to correct low to moderate myopia with astigmatism using the aspheric algorithm of the SCHWIND AMARIS excimer laser. PATIENTS AND METHODS PATIENT POPULATION AND EXAMINATIONS Six investigative sites were involved in this prospective, international multicenter study of the treatment of myopia with astigmatism using the SCHWIND AMARIS excimer laser. The first three AMARIS laser machines fabricated by SCHWIND were used in this study and moved between the investigative sites to perform the treatments. This study followed the tenets of the Declaration of Helsinki. Three hundred fifty-eight eyes of 179 patients undergoing bilateral LASIK for myopia with astigmatism were enrolled. Mean patient age was 28 7 years (range: 18 to 60 years). All patients included in the study were aged 18 years and had a manifest refraction spherical equivalent (MRSE) ranging from 0.50 D to 7.38 diopters (D) with up to 5.00 D of astigmatism. Patients were enrolled in the study Journal of Refractive Surgery Vol. 26, No. 2, 2010 if they had best spectacle-corrected visual acuity (BSCVA) of 20/40 or better using the Early Treatment of Diabetic Retinopathy Study (ETDRS) chart, stable refraction for 1 year prior to the study, and discontinued contact lens wear for at least 1 week to 4 weeks (depending on contact lens type) prior to the preoperative evaluation. Patients were required to have normal keratometry and topography with 0.50 equivalent defocus 16 of corneal wavefront aberration in root-mean-square (RMS) analyzed at a 6-mm analysis diameter or the maximum possible analysis diameter minus 0.5 mm for each topography if it was smaller than 6 mm measured with the SCHWIND Corneal Wavefront Analyzer (SCHWIND eye-tech-solutions GmbH & Co KG). For comparative analyses, all aberrations were analyzed for a 6-mm diameter. For our analysis, the concept of equivalent defocus has been used as a metric to minimize the differences in the Zernike coefficients due to different pupil sizes. Equivalent defocus is the amount of defocus required to produce the same wavefront variance as found in one or more high order aberrations. 16 A simple formula computes equivalent defocus in diopters from the wavefront variance in the Zernike modes in question: M e = 16 3 RMS PD (1) 2 where M e is the equivalent defocus in diopters, RMS is the wavefront variance in the Zernike modes in question, and PD is the diameter considered for the wavefront aberration analysis. On virgin eyes, equivalent defocus as proposed by Thibos et al 16 seems to be relatively insensitive to different analysis diameters. 17 Patients who suffered from an acute illness or who had a calculated postoperative corneal bed thickness 250 µm after ablation, preoperative central corneal thickness 475 µm, previous ocular surgery, or abnormal corneal topography were excluded from the study. Baseline examinations included measurement of uncorrected visual acuity (UCVA), BSCVA, MRSE, contrast sensitivity, corneal topography, corneal wavefront, ultrasound corneal pachymetry, pupillometry, slit-lamp examination of the anterior segment, and a dilated fundus examination. Pre- and postoperative refraction were based on obtaining patient subjective manifest refraction under photopic conditions and crosschecked with the objective refraction analyzed at 4 mm measured by the SCHWIND Ocular Wavefront Analyzer (SCHWIND eye-tech-solutions GmbH & Co KG)

3 For crosschecking refraction with ocular wavefront, patients were asked to see-through-the-target (forcing a disaccommodated state) instead of focusing-at-thetarget (which is, however, fogged with 1.50 D) when performing aberrometry. Ocular wavefront was analyzed for a 4-mm diameter, and the Zernike refraction based on second order parabolic terms was recorded as the objective refraction value. This objective refraction value (rounded to a quarter of a diopter for sphere and cylinder, and to 5 for the astigmatism axis) was used as the initial refraction to be tested at the phoropter. Pre- and postoperative contrast sensitivity using the CST 1800 D (Vision Sciences Research Corp, San Ramon, Calif) was measured at 3, 6, 12, and 18 cycles per degree (cpd) after correcting the refractive error with spectacles. Photopic contrast sensitivity measurement was done with illumination of 85 cd/m 2 without any glare and mesopic contrast sensitivity with an illumination of 6 cd/m 2 with 35 lux glare. Log values of the contrast sensitivity scores were used for statistical analysis. At 1 day postoperatively, UCVA was measured and the patient underwent slit-lamp examination of the anterior segment. The same measurements as the baseline examination (with the exception of dilated funduscopy and pupillometry unless warranted and contrast sensitivity at 6 months only) were performed at 1 week and 1, 3, and 6 months postoperatively. TREATMENT PLAN All aspheric treatments were planned using the SCHWIND CAM (Customized Ablation Manager, SCHWIND eye-tech-solutions GmbH & Co KG) treatment planning software. The SCHWIND CAM software integrates aspheric ablation profiles that compensate for the peripheral loss of energy due to an increased angle of incidence on the cornea 19 and for biomechanical changes induced during LASIK. The treatment of ocular or corneal wavefront aberrations was not intended in this study. The sphere and cylinder values entered into the laser were based on the manifest refraction without nomogram adjustment. Further, the flat and steep keratometry readings at 3-mm diameter as measured by the topographer were used for the compensation of the loss of ablation efficiency when the laser hits the cornea in non-normal incidence. 19 All eyes underwent the refractive treatment using to 7.00-mm diameter optical zones based on the preoperative scotopic pupil diameter. For each treatment, the SCHWIND CAM calculated the size of the optimal transition zone, depending on the preoperative refraction and optical zone. The total ablation zone ranged from 6.90 to 8.51 mm. All eyes were targeted for emmetropia. 90 Maintenance of the preoperative corneal higher order aberrations required a change in the corneal curvature postoperatively based on proprietary ablation algorithms considering astigmatic condition, spot overlapping, tissue removal, and tissue remodeling, which are calculated by the SCHWIND CAM. Retreatments were not permitted during the course of this study. Once finalized, the treatment plan was directly entered or transferred via secure digital memory card to the SCHWIND AMARIS excimer laser. SURGERY Eyes undergoing surgery were prepared as customary for each center. One or two drops of topical anesthetic were instilled and a sterile drape was used to isolate the surgical field. A lid speculum was inserted to allow maximum exposure of the globe. Two additional drops of topical anesthetic were instilled in the upper and lower fornices. Flaps with a superior hinge were made using the Carriazo-Pendular microkeratome (SCHWIND eye-tech-solutions GmbH & Co KG) using 110- or 130-µm heads (336 eyes) and flaps with a nasal hinge were made using the Amadeus (Ziemer Ophthalmic Systems, Port, Switzerland) using 120-µm heads (22 eyes). Proper alignment of the eye with the laser was achieved with a 1050-Hz infrared eye tracker with simultaneous limbus, pupil, and torsion 20 tracking integrated into the laser system and centered on the pupil. The eye tracker had an average response time of 2.9 milliseconds. The flap was lifted and the excimer laser ablation was delivered to the stroma. Patients were requested to fixate on a fixation light throughout the ablation. The flap was repositioned and the interface was irrigated with balanced salt solution, removing any debris. Patients received topical fluoroquinolone antibiotic, corticosteroid drops four times per day for 5 days, and ocular lubricants as needed. EXCIMER LASER The laser ablation algorithm used a flying spot laser delivery system that operates at 500 Hz with a super- Gaussian beam profile (Fig 1) of 0.54 mm Full Width Half Maximum. Depending on the planned refractive correction, approximately 80% of the corneal ablation is performed with a high fluence level ( 400 mj/cm 2 ), thus decreasing treatment times. Fine correction is performed for the remaining ~20% of the treatment using a low fluence level ( 200 mj/cm 2 ), which reduces the ablation volume per pulse delivered to smooth out the ablated area. Spot placement is randomized to prevent heat buildup between laser pulses. Additionally, an aspiration system with laminar flow dynamics is incorporated to reduce debris and heat buildup.

4 RESULTS A PREOPERATIVE Mean preoperative MRSE was D (range: 0.50 to 7.38 D), mean sphere was D (range: 7.25 to 0.00 D), and mean cylinder was D (range: 5.00 to 0.00 D). All 358 eyes underwent corneal wavefront measurements; however, only 223 eyes showed a topographical map free of artifacts in the central 6.5-mm region. Thus, a reliable corneal wavefront map of 6-mm analysis diameter was only obtained in these 223 eyes. Figure 1. A) Beam energy and B) ablative spot profiles of Gaussian, Super-Gaussian, and Flat-top profiles. Compared to a Gaussian profile, the Super-Gaussian beam has the same amplitude but a larger Full Width at Half Maximum. DATA ANALYSIS Refractive outcomes, changes in higher order aberrations, and contrast sensitivity were analyzed using Microsoft Excel software (Microsoft Corp, Redmond, Wash) and Datagraph (Datagraph-Med GmbH, Wendelstein, Germany) refractive outcomes analysis software. LogMAR visual acuity was converted to Snellen acuity for data reporting purposes only. The paired single-sided t test and unpaired t test were used to determine statistically significant changes in corneal wavefront aberrations and contrast sensitivity, respectively. A P value.05 was considered statistically significant. Data for 1 day, 1 week, and 1, 3, and 6 months after LASIK are reported. B POSTOPERATIVE Predictability and Refractive Stability. The attempted versus achieved MRSE for all eyes is plotted in Figure 2. At 6 months postoperatively, the mean MRSE for the entire study population was D (range: 0.88 to 0.50 D). Mean postoperative sphere was D (range: 0.75 to 1.00 D) and mean postoperative cylinder was D (range: 1.25 to 0.00 D) at 6 months. For the entire cohort, 73% (261/358) of eyes had MRSE within 0.25 D (Fig 3). Refractive stability postoperatively was evaluated by assessing the mean rate of change in MRSE over time (0.04 D/month). The mean MRSE did not change from 1 week to 1 month postoperatively (Fig 4). The mean rate of change in MRSE between 1 week and 3 months was 0.02 D/month, and the rate of change between 3 and 6 months was 0.01 D/month (see Fig 4). Mean residual astigmatism magnitude did not change significantly from 1 to 6 months postoperatively. Figure 5 plots the change in astigmatism 6 months postoperatively. Visual Acuity. Six months after surgery, 98% (351/358) of eyes had UCVA of 20/20 or better and Journal of Refractive Surgery Vol. 26, No. 2,

5 Refractive Outcome Within Attempted Correction (D) Figure 2. Scattergram of attempted versus achieved manifest refraction spherical equivalent (SEQ) of 358 eyes 6 months after LASIK for myopia and myopic astigmatism using the SCHWIND AMARIS excimer laser. Figure 3. Refractive outcomes 6 months postoperatively of 358 eyes that underwent LASIK for myopia and myopic astigmatism using the SCHWIND AMARIS excimer laser. Diopters Figure 4. Change in manifest refraction spherical equivalent over time of eyes that underwent LASIK for myopia and myopic astigmatism with the SCHWIND AMARIS excimer laser. 65% (233/358) had UCVA of 20/16 or better (Fig 6). At 6 months postoperatively, there was a higher percentage of eyes with better UCVA than preoperative BSCVA at the 20/20 or better level of acuity (Fig 7). Clinically significant loss of BSCVA is considered a loss of 2 or more Snellen lines. Six months after surgery, no eye lost 2 or more lines of BSCVA (Fig 8). Best spectaclecorrected visual acuity increased by 2 or more lines in 8.4% (30/358) of eyes 6 months after surgery (see Fig 8). Corneal Wavefront Aberrations. At 6 months postoperatively, all eyes underwent corneal wavefront measurements, of those only 259 eyes showed a topographical map free of artifacts in the central 6.5-mm region. Thus, a reliable corneal wavefront map of 6-mm analysis diameter was only obtained in these 259 eyes. 92 This discrepancy means that higher order aberrations may not have been measured in the same eyes pre- and postoperatively. To ensure inductions of corneal wavefront aberrations were not masked due to the differences in the number of eyes preoperatively (223 eyes) and 6 months postoperatively (259 eyes), all eyes with corresponding preoperative and 6-month postoperative corneal wavefront were analyzed (162 eyes) (Fig 9). Mean change in total corneal higher order aberrations increased by 0.09 µm 6 months postoperatively (P.01). The mean change in coma 0.04 µm (P=.02) and spherical aberration 0.08 µm (P.01) was also statistically significant (see Fig 9). Figure 10 plots the change in the attempted spherical equivalent correction compared to the change in various corneal higher order aberrations for all eyes that completed preoperative and 6-month postoperative corneal wavefront measurements. A statistically significant increase was noted in spherical aberration, coma, and total higher order aberrations as the attempted spherical equivalent correction increased (Fig 10). The change in BSCVA compared to the change in various corneal higher order aberrations is shown in Figure 11. In addition, predictability, refractive stability, visual acuity, and safety analyses did not show different results when analyzed for the complete sample of treatments (358 eyes) compared with the analysis restricted for the group with corresponding preoperative and 6-month postoperative corneal wavefront (162 eyes). Contrast Sensitivity. Six months postoperatively, no statistically significant change was noted in photopic or mesopic contrast sensitivity (P.05) (Fig 12). Complications. Intraoperatively, in one eye, a par-

6 Figure 5. Double angle scatterplot of astigmatism (minus cylinder notation) of 358 eyes that underwent LASIK with the SCHWIND AMARIS excimer laser: A) preoperatively, B) 6 months postoperatively, C) normalized preoperative plot, and D) 6-month postoperative plot relative to the preoperative axis of astigmatism. tial flap was created; however, the surgeon elected to continue the treatment including laser ablation. Two eyes of two patients developed grade 1 diffuse lamellar keratitis (DLK) at day 1 postoperatively. Both eyes were treated with topical steroids and antibiotics four times daily on a tapering schedule over 1 month. The DLK resolved by 1 month postoperatively without adversely affecting BSCVA. DISCUSSION The present investigation of LASIK using the SCHWIND AMARIS excimer laser with a 500-Hz repetition rate found this laser platform to be safe and predictable and to provide stable results for the treatment of low to moderate myopia with up to 5.00 D of astigmatism. For example, 73% of eyes were within 0.25 D of intended correction (see Fig 3). Safety was demonstrated with no eyes losing more than one line of BSCVA 6 months after surgery (see Fig 7). Stability was achieved quickly, with a mean rate of change of 0.02 D/month from 1 week to 3 months postoperatively (see Fig 4). Compared to other studies using nonwavefront-guided profiles, we found less amounts of induced higher order aberrations in our sample. Kohnen et al 21 measured induced corneal higher order aberrations after conventional treatments. Using a 6-mm analysis diameter, they found an increase of 0.09 µm for coma, 0.13 µm for spherical aberration, and 0.17 µm RMS for higher order aberrations. Lombardo et al, 22 in a study analyzing corneal wavefront for a 6-mm diameter after conventional PRK, reported an increase of 0.08 µm and 0.11 µm for coma and spherical aberration, respectively, for low myopic corrections, and an increase of 0.14 µm and 0.37 µm for coma and spherical aberration, respectively, for high myopic corrections. After conventional LASIK, de Ortueta et al 10 found an increase of 0.10 µm for coma, 0.17 µm for spherical aberration, and 0.20 µm RMS for higher order aberrations for a 6-mm diameter and an increase of 0.07 µm for coma, 0.09 µm for spherical aberration, and 0.02 µm RMS for higher order aberrations after Aberration- Free LASIK with a different laser platform. These values are comparable to our findings of 0.04 µm for coma, 0.08 µm for spherical aberration, and 0.09 µm RMS for higher order aberrations for a 6-mm diameter. Marcos et al 23 found that ocular and corneal aberrations increased statistically significantly after myopic LASIK surgery by a factor of 1.92 (ocular) and 3.72 (corneal), on average. They found a good correlation (P.0001) between the aberrations induced in the entire optical system and those induced in the anterior corneal surface. However, anterior corneal aberrations increased more than ocular aberrations, suggesting changes in the posterior corneal surface as well. Lee et al 24 found that after laser refractive surgery, anterior corneal aberration and ocular aberration increased equally and showed statistically significant correlations. They found no statistically significant differences of internal optics aberration values in coma, spherical aberration, and RMS for higher order aberrations. Arbelaez et al 25 found that when comparing corneal Journal of Refractive Surgery Vol. 26, No. 2,

7 UCVA Figure 6. Cumulative uncorrected visual acuity (UCVA) of eyes that underwent LASIK for myopia and myopic astigmatism using the SCHWIND AMARIS excimer laser. Visual Acuity Figure 7. Preoperative best spectacle-corrected visual acuity (prebscva) compared to 6-month postoperative uncorrected visual acuity (UCVA) (6 m) after LASIK for myopia and myopic astigmatism using the SCHWIND AMARIS excimer laser. and ocular aberrations, the amount of induced aberrations was similar for spherical aberration and coma. For the RMS for higher order aberrations, corneal induced aberrations were moderately higher, although not statistically significant, than ocular induced aberrations. In our study of 716 analyzed topographies, 482 (67%) showed a 6.5-mm diameter disk available for corneal wavefront analyses. The reason why not all topographies showed reliable corneal wavefront aberrations of at least 6-mm diameter is attributable to the fact that we do not allow the videokeratoscope to extrapolate corneal topography to areas with no real data, which are affected by potential artifacts (eg, shadows from eyelashes, etc), during the measurements. Such artifacts may be due to disruption of the tear film, interference 94 Change in BSCVA Figure 8. Change in best spectacle-corrected visual acuity (BSCVA) of 358 eyes 6 months after LASIK for myopia and myopic astigmatism using the SCHWIND AMARIS excimer laser. Figure 9. Change in mean higher order aberrations (HO) from preoperatively to 6 months postoperatively for eyes that underwent LASIK for myopia and myopic astigmatism using the SCHWIND AMARIS excimer laser. All data are reported for a 6.00-mm diameter. The paired single sided t test was used to determine statistically significant difference as data for all eyes were available pre- and postoperatively. A P value.05 was considered statistically significant. SphAb = spherical aberration, RMS = root-mean-square of the eyelashes or eyelids within the corneal area, or difficult recognition of the pupil contour (especially in dark colored eyes). Finally, 223 preoperative topographies (62%) and 259 postoperative topographies (72%) provided reliable corneal wavefront aberration information of at least 6 mm in size. This discrepancy means that corneal aberrations may not have been analyzed in the same eyes pre- and postoperatively. To ensure inductions of corneal wavefront aberrations were not masked due to the differences in the number of eyes preoperatively (223 eyes) and 6 months postoperatively (259 eyes), only eyes with corresponding

8 reliable preoperative and 6-month postoperative corneal wavefront (162 eyes) were reported in the analysis (see Fig 9). Compared with other studies of ocular wavefrontguided LASIK treatment for myopia with astigmatism that found wavefront-guided LASIK to be safe and effective, 2-4 the outcomes of the present study are better. Previous studies of myopic corneal wavefrontguided treatments using the ESIRIS laser (SCHWIND eye-tech-solutions GmbH & Co KG) with smaller sample sizes and lower preoperative astigmatism than our study found that there was no clinically significant loss of BSCVA. 2,26 In the present study, Figure 2 demonstrates little scatter, indicating excellent accuracy despite the lack of nomogram refinement. Perhaps the incorporation of center-specific nomograms would further refine accuracy. The outcomes of the present study confirm that aspheric LASIK with a 500-Hz excimer laser is clinically viable for the treatment of primary myopia and astigmatism. The efficacy reflects the predictability and accuracy reported herein. For example, within 6 months after surgery, 98% of eyes had UCVA of 20/20 or better without correction (see Fig 7). Six months postoperatively, UCVA exceeded preoperative BSCVA in more than 46% of eyes at the 20/16 or better levels. Best spectacle-corrected visual acuity was preserved in 99.4% (356/358) of eyes (see Fig 8). A contralateral eye study comparing aspheric and wavefront-guided treatments using the WaveLight laser (WaveLight AG, Erlangen, Germany) reported BSCVA was preserved in 83% of eyes. 27 A prospective, randomized comparison of the Wave- Light and CustomCornea (Alcon Laboratories Inc, Ft Worth, Tex) platforms reported 80% of eyes achieved UCVA of 20/ A study by Binder and Rosenshein 3 comparing three wavefront platforms found approximately 5% of eyes that underwent spherocylindrical Journal of Refractive Surgery Vol. 26, No. 2, 2010 A Figure 10. Preoperative to 6 months after LASIK for myopia and myopic astigmatism using the SCHWIND AMARIS excimer laser showing change versus attempted spherical equivalent refraction (SEQ) for A) corneal spherical aberration (SphAb), B) corneal coma, and C) corneal higher order aberrations (HO). P.05 is considered statistically significant. RMS = root-mean-square treatments gained clinically significant BSCVA, which is less than our result of 8.5%. 3 Comparing the outcomes in the present study with the aforementioned studies indicates the outcomes with the SCHWIND AMARIS are equivalent to or exceed those reported for wavefront-guided treatments and aspheric treatments despite treating higher levels of preoperative astigmatism. In our study, 0.5% (two eyes) lost one line of BSCVA. Because one line is the day-to-day variability in measuring visual acuity, clinically relevant loss of visual acuity is reported for two or more lines. Moreover, the two eyes losing one line of BSCVA were eyes with preoperative super vision (20/16) that resulted in normal vision (20/20) postoperatively. On a more conceptual basis, our study population is relatively young, with a mean age of 28 years, so most of the patients can accommodate, thus, a hyperopic B C 95

9 96 Change in Visual Acuity Figure 11. Change in corneal spherical aberration (SphAB), corneal coma, and corneal higher order aberrations (HO) of eyes that lost or gained best spectacle-corrected visual acuity 6 months after LASIK with the SCHWIND AMARIS excimer laser. Two eyes lost 1 line, 63 eyes remained unchanged, 83 eyes gained 1 line, and 14 eyes gained 2 lines. RMS = root-mean-square overcorrection of 0.50 D would still produce excellent UCVA despite a non-plano outcome. However, because the pre- and postoperative determined subjective manifest refraction of the patient was crosschecked with the objective refraction analyzed at 4 mm, the influence of accommodation is minimal. Our study used an aspheric ablation algorithm independent of wavefront customization that aimed to maintain the preoperative levels of higher order aberrations (Aberration-Free algorithm). McLellan et al 14 reported a beneficial effect on visual quality due to pre-existing higher order aberrations. Evidence of neural adaption to the baseline wavefront profile exists. 15 The interaction between higher order aberrations can be beneficial to visual quality regardless of the magnitude of higher order aberrations. 29,30 To date, the induction of wavefront aberrations postoperatively is random and the wavefront profile postoperatively cannot be predicted. Based on the random nature of higher order aberration induction and current research, 13-15,29,30 it may be beneficial to maintain the preoperative wavefront profile for a significant number of refractive surgery candidates. Using aspheric treatments, we found mild increases in corneal higher order aberration RMS (see Fig 9) and a statistically significant increase in spherical aberration and coma based on attempted correction (see Fig 10). Ocular wavefront measurements were only used to crosscheck manifest refraction measurements, and not for assessing higher order aberration at this time, which is why the higher order aberrations are reported in terms of induced corneal wavefront. Most studies report results in terms of induced or postoperative ocular aberrations, hence direct comparisons to the induced corneal aberrations in our study are not possible. Because we used the ocular wavefront measurement to objectively assess the refraction, we did not pay special care to pupil dilation beyond 6.00 mm or at least 5.50 mm. In addition, illumination conditions provided for ocular wavefront were not identical in the six centers, therefore, different maximum analysis diameters were available across the sample. Comparing different pupil sizes is not the standard method for comparison, and comparison of aberrations in diameters 5.50 mm has low clinical relevance. However, other studies in which direct comparison of induced aberrations in terms of corneal wavefront and ocular wavefront over the same sample was studied, showed that the induction of anterior corneal aberrations was always, at least, as high as the induction of ocular wavefront aberrations for the entire eye. The increase in spherical aberration can be partially explained by the biomechanical response and corneal epithelial remodeling. 31,32 The mild induction of higher order aberrations, spherical aberration, and coma may explain the maintenance of visual quality postoperatively found in the current study. For example, at 6 months postoperatively, there was negligible change in photopic and mesopic contrast sensitivity. This outcome may indicate that the induction of higher order aberrations reported here may require less neural adaption (if any at all). Although not statistically significant, a trend in Figure 10 indicated the lower the change in spherical aberration, the greater the increase in visual acuity postoperatively. However, the results of our study show no significant improvement of contrast sensitivity but improved postoperative visual outcomes. However, a separate study specifically designed to address neural adaption with nonwavefront-guided aspheric treatments is warranted to address this topic. Based on the results presented, we are not postulating that customized ablation algorithms in any form (ocular wavefront-guided, corneal wavefront-guided, topography-guided, etc) are not useful. Rather, specific populations with specific demands deserve specific treatment solutions. Aspheric treatments aiming for preservation of the preoperative higher order aberrations show their strengths in patients with preoperative BSCVA of 20/20 or better, or in patients in whom the visual degradation cannot be attributed to the presence of clinically relevant higher order aberrations. The use of a 500-Hz repetition rate did not create any postoperative complications. Thermal effects due

10 Figure 12. Preoperative and 6-month postoperative A) photopic and B) mesopic contrast sensitivity of eyes that underwent LASIK for myopia and myopic astigmatism using the SCHWIND AMARIS excimer laser. to excimer laser ablation and associated plume debris have been reported previously. 5,6,33,34 One study reported better refractive outcomes after cooling the cornea prior to the ablation, which likely reduced the thermal buildup. 7 The use of a particle aspirator coupled with random placement of the laser spots in this study seems to efficiently speed up the treatment without increasing the thermal load in the cornea. The 6-month results presented in this study indicate that Aberration-Free treatment of myopia with and without astigmatism using the SCHWIND AMARIS excimer laser is safe, provides excellent refractive and visual acuity results, and maintains preoperative contrast sensitivity. AUTHOR CONTRIBUTIONS Study concept and design (F.C.); data collection (M.C.A., I.A., C.B., F.C., A.F., T.N., P.R.); analysis and interpretation of data (F.C.); drafting of the manuscript (M.C.A.); critical revision of the manuscript (M.C.A., I.A., C.B., F.C., A.F., T.N., P.R.); administrative, technical, or material support (M.C.A., I.A., C.B., A.F., T.N., P.R.) REFERENCES 1. Netto MV, Dupps W Jr, Wilson SE. Wavefront-guided ablation: evidence for efficacy compared to traditional ablation. Am J Ophthalmol. 2006;141: Zhou C, Jin M, Wang X, Ren Q. Corneal wavefront-guided ablation with the Schwind ESIRIS laser for myopia. J Refract Surg. 2007;23: Binder PS, Rosenshein J. Retrospective comparison of 3 laser platforms to correct myopic spheres and spherocylinders using conventional and wavefront-guided treatments. J Cataract Refract Surg. 2007;33: Hori-Komai Y, Toda I, Asano-Kato N, Ito M, Yamamoto T, Tsubota K. Comparison of LASIK using the NIDEK EC-5000 optimized aspheric transition zone (OATz) and conventional ablation profile. J Refract Surg. 2006;22: Ishihara M, Arai T, Sato S, Morimoto Y, Obara M, Kikuchi M. Temperature measurement for energy-efficient ablation by thermal radiation with a microsecond time constant from the corneal surface during ArF excimer laser ablation. Front Med Biol Eng. 2001;11: Bende T, Seiler T, Wollensak J. Side effects in excimer corneal surgery. Corneal thermal gradients. Graefes Arch Clin Exp Ophthalmol. 1988;226: Amoils SP. Using a Nidek excimer laser with a rotary epithelial brush and corneal chilling: clinical results. J Cataract Refract Surg. 1999;25: Kim WS, Jo JM. Corneal hydration affects ablation during laser in situ keratomileusis surgery. Cornea. 2001;20: Iseli HP, Mrochen M, Hafezi F, Seller T. Clinical photoablation with a 500-Hz scanning spot excimer laser. J Refract Surg. 2004;20: de Ortueta D, Arba Mosquera S, Baatz H. Comparison of standard and aberration-neutral profiles for myopic LASIK with the SCHWIND ESIRIS laser. J Refract Surg. 2009;25: Stonecipher KG, Kezirian GM. Wavefront-optimized versus wavefront-guided LASIK for myopic astigmatism with the ALLEGRETTO WAVE: three-month results of a prospective FDA trial. J Refract Surg. 2008;24:S424-S Padmanabhan P, Mrochen M, Basuthkar S, Viswanathan D, Joseph R. Wavefront-guided versus wavefront-optimized laser in situ keratomileusis: contralateral comparative study. J Cataract Refract Surg. 2008;34: Subbaram MV, MacRae SM. Customized LASIK treatment for myopia based on preoperative manifest refraction and higher order aberrometry: the Rochester nomogram. J Refract Surg. 2007;23: McLellan JS, Marcos S, Prieto PM, Burns SA. Imperfect optics may be the eye s defence against chromatic blur. Nature. 2002;417: Artal P, Chen L, Fernández EJ, Singer B, Manzanera S, Williams DR. Neural compensation for the eye s optical aberrations. J Vis. 2004;4: Thibos LN, Hong X, Bradley A, Cheng X. Statistical variation of aberration structure and image quality in a normal population of healthy eyes. J Opt Soc Am A Opt Image Sci Vis. 2002;19: Thibos L. VSIA-2001 wavefront data reporting and terminology. Presented at: Optical Society of America Meeting, VSIA, Wavefront Sensing Congress; February 9-10, 2001; Monterey, CA. Journal of Refractive Surgery Vol. 26, No. 2,

11 18. Salmon TO, West RW, Gasser W, Kenmore T. Measurement of refractive errors in young myopes using the COAS Shack-Hartmann aberrometer. Optom Vis Sci. 2003;80: Arba Mosquera S, de Ortueta D. Geometrical analysis of the loss of ablation efficiency at non-normal incidence. Opt Express. 2008;17: Arba Mosquera S, Merayo-Lloves J, de Ortueta D. Clinical effects of pure cyclotorsional errors during refractive surgery. Invest Ophthalmol Vis Sci. 2008;49: Kohnen T, Mahmoud K, Bühren J. Comparison of corneal higherorder aberrations induced by myopic and hyperopic LASIK. Ophthalmology. 2005;112: Lombardo M, Lombardo G, Manzulli M, Palombi M, Serrao S. Relative contribution of central and peripheral aberrations to overall high order corneal wavefront aberration. J Refract Surg. 2006;22: Marcos S, Barbero S, Llorente L, Merayo-Lloves J. Optical response to LASIK surgery for myopia from total and corneal aberration measurements. Invest Ophthalmol Vis Sci. 2001;42: Lee JM, Lee DJ, Jung WJ, Park WC. Comparison between anterior corneal aberration and ocular aberration in laser refractive surgery. Korean J Ophthalmol. 2008;22: Arbelaez MC, Vidal C, Al Jabri B, Arba Mosquera S. Threemonth clinical outcomes in LASIK for myopia with aspheric aberration neutral ablations using the ESIRIS laser system. J Refract Surg. 2009;25: de Ortueta D, Magnago T, Flugge J. Schwind ESIRIS excimer laser. In: Roy HF, Benjamin L, eds. Surgical Techniques in Ophthalmology: Refractive Surgery. Philadelphia, PA: Elsevier Inc; 2008: Koller T, Iseli HP, Hafezi F, Mrochen M, Seiler T. Q-factor customized ablation profile for the correction of myopic astigmatism. J Cataract Refract Surg. 2006;32: Brint SF. Higher order aberrations after LASIK for myopia with Alcon and Wavelight lasers: a prospective, randomized trial. J Refract Surg. 2005;21:S799-S Applegate RA, Sarver EJ, Khemsara V. Are all aberrations equal? J Refract Surg. 2002;18:S556-S Applegate RA, Marsack JD, Ramos R, Sarver EJ. Interaction between aberrations to improve or reduce visual performance. J Cataract Refract Surg. 2003;29: Roberts C. Biomechanics of the cornea and wavefront-guided laser refractive surgery. J Refract Surg. 2002;18:S589-S Huang D, Tang M, Shekhar R. Mathematical model of corneal surface smoothing after laser refractive surgery. Am J Ophthalmol. 2003;135: Jaycock PD, O Brart DP, Rajan MS, Marshall J. 5-year follow-up of LASIK for hyperopia. Ophthalmology. 2005;112: Zaldivar R, Oscherow S, Bains HS. Five techniques for improving outcomes of hyperopic LASIK. J Refract Surg. 2005;21:S628-S

Corneal Asphericity and Retinal Image Quality: A Case Study and Simulations

Corneal Asphericity and Retinal Image Quality: A Case Study and Simulations Corneal Asphericity and Retinal Image Quality: A Case Study and Simulations Seema Somani PhD, Ashley Tuan OD, PhD, and Dimitri Chernyak PhD VISX Incorporated, 3400 Central Express Way, Santa Clara, CA

More information

Transferring wavefront measurements to ablation profiles. Michael Mrochen PhD Swiss Federal Institut of Technology, Zurich IROC Zurich

Transferring wavefront measurements to ablation profiles. Michael Mrochen PhD Swiss Federal Institut of Technology, Zurich IROC Zurich Transferring wavefront measurements to ablation profiles Michael Mrochen PhD Swiss Federal Institut of Technology, Zurich IROC Zurich corneal ablation Calculation laser spot positions Centration Calculation

More information

The influence of the aspheric profiles for transition zone on optical performance of human eye after conventional ablation

The influence of the aspheric profiles for transition zone on optical performance of human eye after conventional ablation J. Europ. Opt. Soc. Rap. Public. 9, 4060 (204) www.jeos.org The influence of the aspheric profiles for transition zone on optical performance of human eye after conventional ablation L. Fang fanglh7@26.com

More information

SCHWIND AMARIS. We have redefined perfection for you

SCHWIND AMARIS. We have redefined perfection for you SCHWIND AMARIS We have redefined perfection for you 2 SCHWIND AMARIS the TotalTech Laser Not only can it do anything it can do it outstandingly well, too. The SCHWIND AMARIS is a TotalTech Laser. It is

More information

Normal Wavefront Error as a Function of Age and Pupil Size

Normal Wavefront Error as a Function of Age and Pupil Size RAA Normal Wavefront Error as a Function of Age and Pupil Size Raymond A. Applegate, OD, PhD Borish Chair of Optometry Director of the Visual Optics Institute College of Optometry University of Houston

More information

Aberration Interaction In Wavefront Guided Custom Ablation

Aberration Interaction In Wavefront Guided Custom Ablation Aberration Interaction In Wavefront Guided Custom Ablation Scott M. MacRae MD Professor of Ophthalmology Professor of Visual Science University of Rochester Collaborators and Disclosures: Manoj Subbaram

More information

What s New in Ocular Biomechanics?

What s New in Ocular Biomechanics? What s New in Ocular Biomechanics? The International Congress of Wavefront Sensing & Optimized Refractive Corrections Wavefront Course January 28, 2006 Torrence A. Makley Research Professor Department

More information

Retrospective analysis of changes in the anterior corneal surface after Q value guided LASIK and LASEK in high myopic astigmatism for 3 years

Retrospective analysis of changes in the anterior corneal surface after Q value guided LASIK and LASEK in high myopic astigmatism for 3 years Huang et al. BMC Ophthalmology 2012, 12:15 RESEARCH ARTICLE Open Access Retrospective analysis of changes in the anterior corneal surface after Q value guided LASIK and LASEK in high myopic astigmatism

More information

Assessing Visual Quality With the Point Spread Function Using the NIDEK OPD-Scan II

Assessing Visual Quality With the Point Spread Function Using the NIDEK OPD-Scan II Assessing Visual Quality With the Point Spread Function Using the NIDEK OPD-Scan II Edoardo A. Ligabue, MD; Cristina Giordano, OD ABSTRACT PURPOSE: To present the use of the point spread function (PSF)

More information

Schwind Amaris 1050 Smart Pulse Technology

Schwind Amaris 1050 Smart Pulse Technology Schwind Amaris 1050 Smart Pulse Technology Paolo Vinciguerra, MD 1, 2 Samuel Arba Mosquera 3 PhD 1 Dept of Ophthalmology, Istituto Clinico Humanitas 2 Columbus, Ohio State University 3 SCHWIND eye-tech-solutions

More information

Visual outcomes and higherorder aberrations of wavefront vs. combined wavefront aspheric myopic LASIK

Visual outcomes and higherorder aberrations of wavefront vs. combined wavefront aspheric myopic LASIK PHILIPPINE JOURNAL OF Ophthalmology Vol. 36 No. 1 Ja n ua r y June 211 ORIGINAL ARTICLE Robert Edward T. Ang, MD 1,2 Aimee Rose A. Icasiano-Ramirez, MD 2 Gladness Henna A. Martinez, MD 1,2 Emerson M. Cruz,

More information

The Aberration Structure of the Keratoconic Eye

The Aberration Structure of the Keratoconic Eye The Aberration Structure of the Keratoconic Eye Geunyoung Yoon, Ph.D. Department of Ophthalmology Center for Visual Science Institute of Optics Department of Biomedical Engineering University of Rochester

More information

Q-value Adjusted Ablation PRK PRK. Allegretto Randomized control trial : .(Corneal asphericity) (PRK) Photo refractive keratectomy

Q-value Adjusted Ablation PRK PRK. Allegretto Randomized control trial : .(Corneal asphericity) (PRK) Photo refractive keratectomy 89/8/16 : 89/6/14 : 1389 /115 / Q-value Adjusted Ablation PRK PRK Allegretto 2 1 1 4 3 Q-value adjusted PRK (PRK) Standard Photorefractive Keratectomy :. Allegretto Eye-Q 75. Randomized control trial :

More information

Aberrations Before and After Implantation of an Aspheric IOL

Aberrations Before and After Implantation of an Aspheric IOL Ocular High Order Aberrations Before and After Implantation of an Aspheric IOL Fabrizio I. Camesasca, MD Massimo Vitali, Orthoptist Milan, Italy I have no financial interest to disclose Wavefront Measurement

More information

This is the author s version of a work that was submitted/accepted for publication in the following source:

This is the author s version of a work that was submitted/accepted for publication in the following source: This is the author s version of a work that was submitted/accepted for publication in the following source: Atchison, David A. & Mathur, Ankit (2014) Effects of pupil center shift on ocular aberrations.

More information

10/25/2017. Financial Disclosures. Do your patients complain of? Are you frustrated by remake after remake? What is wavefront error (WFE)?

10/25/2017. Financial Disclosures. Do your patients complain of? Are you frustrated by remake after remake? What is wavefront error (WFE)? Wavefront-Guided Optics in Clinic: Financial Disclosures The New Frontier November 4, 2017 Matthew J. Kauffman, OD, FAAO, FSLS STAPLE Program Soft Toric and Presbyopic Lens Education Gas Permeable Lens

More information

Surgical data reveals that Q-Factor is important for good surgical outcome

Surgical data reveals that Q-Factor is important for good surgical outcome Surgical data reveals that Q-Factor is important for good surgical outcome Michael Mrochen, PhD Michael Bueeler, PhD Tobias Koller, MD Theo Seiler, MD, PhD IROC AG Institut für Refraktive und Ophthalmo-Chirurgie

More information

Effects of Pupil Center Shift on Ocular Aberrations

Effects of Pupil Center Shift on Ocular Aberrations Visual Psychophysics and Physiological Optics Effects of Pupil Center Shift on Ocular Aberrations David A. Atchison and Ankit Mathur School of Optometry & Vision Science and Institute of Health & Biomedical

More information

Customized Correction of Wavefront Aberrations in Abnormal Human Eyes by Using a Phase Plate and a Customized Contact Lens

Customized Correction of Wavefront Aberrations in Abnormal Human Eyes by Using a Phase Plate and a Customized Contact Lens Journal of the Korean Physical Society, Vol. 49, No. 1, July 2006, pp. 121 125 Customized Correction of Wavefront Aberrations in Abnormal Human Eyes by Using a Phase Plate and a Customized Contact Lens

More information

Optimized Profiles for Astigmatic Refractive Surgery

Optimized Profiles for Astigmatic Refractive Surgery 14 Optimized Profiles for Astigmatic Refractive Surgery Samuel Arba-Mosquera 1,, Sara Padroni 3, Sai Kolli 4 and Ioannis M. Aslanides 3 1 Grupo de Investigación de Cirugía Refractiva y Calidad de Visión,

More information

4th International Congress of Wavefront Sensing and Aberration-free Refractive Correction ADAPTIVE OPTICS FOR VISION: THE EYE S ADAPTATION TO ITS

4th International Congress of Wavefront Sensing and Aberration-free Refractive Correction ADAPTIVE OPTICS FOR VISION: THE EYE S ADAPTATION TO ITS 4th International Congress of Wavefront Sensing and Aberration-free Refractive Correction (Supplement to the Journal of Refractive Surgery; June 2003) ADAPTIVE OPTICS FOR VISION: THE EYE S ADAPTATION TO

More information

Corneal laser surgery is currently shifting its

Corneal laser surgery is currently shifting its Correlation Between Corneal and Total Wavefront Aberrations in Myopic Eyes Michael Mrochen, PhD; Mirko Jankov, MD; Michael Bueeler, MS; Theo Seiler, MD, PhD ABSTRACT PURPOSE: Corneal topography data expressed

More information

Aspheric Refractive Correction of Irregular Astimatism

Aspheric Refractive Correction of Irregular Astimatism 7 Aspheric Refractive Correction of Irregular Astimatism Massimo Camellin 1 and Samuel Arba-Mosquera 2,3 1 SEKAL Rovigo Microsurgery Centre, Rovigo, 2 Grupo de Investigación de Cirugía Refractiva y Calidad

More information

Comparison of higher order aberrations with spherical and aspheric IOLs compared to normal phakic eyes

Comparison of higher order aberrations with spherical and aspheric IOLs compared to normal phakic eyes European Journal of Ophthalmology / Vol. 18 no. 5, 2008 / pp. 728-732 Comparison of higher order aberrations with spherical and aspheric IOLs compared to normal phakic eyes M. RĘKAS, K. KRIX-JACHYM, B.

More information

The Impact of New Generation Aspherical Soft Contact Lenses on Quality of Vision: A Comparison with Spherical Contact Lenses and Spectacle Correction

The Impact of New Generation Aspherical Soft Contact Lenses on Quality of Vision: A Comparison with Spherical Contact Lenses and Spectacle Correction Deniz Oral, Maryo C. Kohen, Melda Yenerel, Ebru Gorgun, Sule Ziylan, Ferda Ciftci Yeditepe University Faculty of Medicine, Department of Ophthalmology, Istanbul Introduction The correction of higher order

More information

NON-LINEAR ASPHERIC ABLATION PROFILE FOR PRESBYOPIC CORNEAL TREATMENT USING THE MEL80/90 AND CRS MASTER PRESBYOND MODULE

NON-LINEAR ASPHERIC ABLATION PROFILE FOR PRESBYOPIC CORNEAL TREATMENT USING THE MEL80/90 AND CRS MASTER PRESBYOND MODULE NON-LINEAR ASPHERIC ABLATION PROFILE FOR PRESBYOPIC CORNEAL TREATMENT USING THE MEL80/90 AND CRS MASTER PRESBYOND MODULE Dan Z Reinstein, MD MA(Cantab) FRCSC DABO FRCOphth FEBO 1,2,3,4 Timothy J Archer,

More information

An Interesting Use of Bausch and Lomb s KeraSoft IC Lens

An Interesting Use of Bausch and Lomb s KeraSoft IC Lens An Interesting Use of Bausch and Lomb s KeraSoft IC Lens Nate Schlotthauer, OD 2012 Michigan College of Optometry Cornea and Contact Lens Resident Introduction: The KeraSoft IC lens, introduced to the

More information

Trust your eyes. Presbyopic treatment methods on the cornea. PresbyMAX Decision criteria and patient s acceptance

Trust your eyes. Presbyopic treatment methods on the cornea. PresbyMAX Decision criteria and patient s acceptance Trust your eyes. Directory Presbyopic treatment methods on the cornea PresbyMAX The Principle PresbyMAX Expectations and Key Factors PresbyMAX Decision criteria and patient s acceptance PresbyMAX Upcoming

More information

Optical Profile Following High Hyperopia Correction With a 500-Hz Excimer Laser System

Optical Profile Following High Hyperopia Correction With a 500-Hz Excimer Laser System ORIGINAL ARTICLE Optical Profile Following High Hyperopia Correction With a 5-Hz Excimer Laser System Ana B. Plaza-Puche, MSc; Amr El Aswad, MD; Samuel Arba-Mosquera, PhD; Dominika Wróbel-Dudzinska, MD;

More information

Dr. Magda Rau Eye Clinic Cham, Germany

Dr. Magda Rau Eye Clinic Cham, Germany 3 and 6 Months clinical Results after Implantation of OptiVis Diffractive-refractive Multifocal IOL Dr. Magda Rau Eye Clinic Cham, Germany Refractive zone of Progressive power for Far to Intermediate

More information

MEL 80 Excimer Laser. When you want to see better performance

MEL 80 Excimer Laser. When you want to see better performance MEL 80 Excimer Laser When you want to see better performance Reward your practice Invest in the very latest refractive excimer technology! The MEL 80 makes vision correction even safer, more patient-friendly

More information

Optical aberrations and the eye Part 3

Optical aberrations and the eye Part 3 clinical 22 Optical aberrations and the eye Part 3 In the final part of our series, Alejandro Cerviño and Dr Shehzad Naroo discuss the methods of correction required for low and high order wavefront aberrations

More information

*Simulated vision. **Individual results may vary and are not guaranteed. Visual Performance When It s Needed Most

*Simulated vision. **Individual results may vary and are not guaranteed. Visual Performance When It s Needed Most Simulated vision. Individual results may vary and are not guaranteed. Visual Performance When It s Needed Most The aspheric design of the AcrySof IQ IOL results in improved clarity and image quality. The

More information

NOW. Approved for NTIOL classification from CMS Available in Quar ter Diopter Powers. Accommodating. Aberration Free. Aspheric.

NOW. Approved for NTIOL classification from CMS Available in Quar ter Diopter Powers. Accommodating. Aberration Free. Aspheric. NOW Approved for NTIOL classification from CMS Available in Quar ter Diopter Powers Accommodating. Aberration Free. Aspheric. Accommodation Meets Asphericity in AO Merging Innovation & Proven Design The

More information

Wavefront Aberrations in Eyes With Acrysof Monofocal Intraocular Lenses

Wavefront Aberrations in Eyes With Acrysof Monofocal Intraocular Lenses Wavefront Aberrations in Eyes With Acrysof Monofocal Intraocular Lenses Prema Padmanabhan, MS; Geunyoung Yoon, PhD; Jason Porter, PhD; Srinivas K. Rao, FRCSEd; Roy J, MSc; Mitalee Choudhury, BS ABSTRACT

More information

NIDEK ADVANCED VISION EXCIMER LASER SYSTEM

NIDEK ADVANCED VISION EXCIMER LASER SYSTEM NIDEK ADVANCED VISION EXCIMER LASER SYSTEM NIDEK ADVANCED VISION EXCIMER LASER SYSTEM Delivering Ultimate Solutions Today NIDEK delivers NAVEX Quest, the evolutionary customized refractive surgery platform.

More information

Focus on performance. The SCHWIND AMARIS product family TREAT

Focus on performance. The SCHWIND AMARIS product family TREAT Focus on performance The SCHWIND AMARIS product family TREAT AMARIS Product Family Our SCHWIND AMARIS product family The absolute best performance from the leader in technology The ever-increasing pace

More information

The Effect of Phenylephrine and Cyclopentolate on Objective Wavefront Measurements

The Effect of Phenylephrine and Cyclopentolate on Objective Wavefront Measurements The Effect of Phenylephrine and Cyclopentolate on Objective Wavefront Measurements Mirko R. Jankov II, MD; Hans Peter Iseli, MD; Michael Bueeler, PhD; Paulo Schor, MD, PhD; Theo Seiler, MD, PhD; Michael

More information

Optical Connection, Inc. and Ophthonix, Inc.

Optical Connection, Inc. and Ophthonix, Inc. Optical Connection, Inc. and Ophthonix, Inc. Partners in the delivery of nonsurgical vision optimization www.opticonnection.com www.ophthonix.com The human eye has optical imperfections that can not be

More information

Focus on performance. The SCHWIND AMARIS product family TREAT

Focus on performance. The SCHWIND AMARIS product family TREAT Focus on performance The SCHWIND AMARIS product family TREAT AMARIS Product Family Our SCHWIND AMARIS product family The absolute best performance from the leader in technology The ever-increasing pace

More information

Optical Path Difference Scanning System OPD-Scan II ARK-10000

Optical Path Difference Scanning System OPD-Scan II ARK-10000 Optical Path Difference Scanning System OPD-Scan II ARK-10000 Optical Path Difference Scanning System OPD-Scan II ARK-10000 Accurate and Reliable Data for Optic Diagnostics The OPD-Scan II provides information

More information

THE BEST OF BOTH WORLDS Dual-Scheimpflug and Placido Reaching a new level in refractive screening

THE BEST OF BOTH WORLDS Dual-Scheimpflug and Placido Reaching a new level in refractive screening THE BEST OF BOTH WORLDS Dual-Scheimpflug and Placido Reaching a new level in refractive screening Clinical Applications Corneal Implant Planning The comes with a licensable corneal inlay software module

More information

Wave Front Topography. ReSeeVit Evolution Topography Module for Modi Topographer

Wave Front Topography. ReSeeVit Evolution Topography Module for Modi Topographer Wave Front Topography ReSeeVit Evolution Topography Module for Modi Topographer Introduction The aberrations in the central optical zone have a greater effect than those closer to the edge. From an optical

More information

Limits of Higher Order Correction based on Spot Size, Ablation Depth, and Tracker Responsiveness

Limits of Higher Order Correction based on Spot Size, Ablation Depth, and Tracker Responsiveness Limits of Higher Order Correction based on Spot Size, Ablation Depth, and Tracker Responsiveness Michael Bueeler a,b, Michael Mrochen a,b, Theo Seiler b a Swiss Federal Institute of Technology Zurich,

More information

Refractive Power / Corneal Analyzer. OPD-Scan III

Refractive Power / Corneal Analyzer. OPD-Scan III Refractive Power / Corneal Analyzer OPD-Scan III Comprehensive Vision Analysis and NIDEK, a global leader in ophthalmic and optometric equipment, has created the OPD-Scan III, the third generation aberrometer

More information

Aspheric photorefractive keratectomy for myopia and myopic astigmatism with the SCHWIND AMARIS laser: 2 years postoperative outcomes

Aspheric photorefractive keratectomy for myopia and myopic astigmatism with the SCHWIND AMARIS laser: 2 years postoperative outcomes Journal of Optometry (2013) 6, 9-17 www.journalofoptometry.org ORIGINAL ARTICLE Aspheric photorefractive keratectomy for myopia and myopic astigmatism with the SCHWIND AMARIS laser: 2 years postoperative

More information

Unique Aberration-Free IOL: A Vision that Patients

Unique Aberration-Free IOL: A Vision that Patients Unique Aberration-Free IOL: A Vision that Patients Can Appreciate An Aspheric Optic for Improved Quality of Vision n Traditional spherical IOLs create Bilateral implantation study spherical aberration

More information

ORIGINAL ARTICLE. Dan Z. Reinstein, MD, MA(Cantab), FRCOphth; Marine Gobbe, MST(Optom), PhD; Timothy J. Archer, MA(Oxon), DipCompSci(Cantab)

ORIGINAL ARTICLE. Dan Z. Reinstein, MD, MA(Cantab), FRCOphth; Marine Gobbe, MST(Optom), PhD; Timothy J. Archer, MA(Oxon), DipCompSci(Cantab) ORIGINAL ARTICLE Coaxially Sighted Corneal Light Reflex Versus Entrance Pupil Center Centration of Moderate to High Hyperopic Corneal Ablations in Eyes With Small and Large Angle Kappa Dan Z. Reinstein,

More information

Choices and Vision. Jeffrey Koziol M.D. Thursday, December 6, 12

Choices and Vision. Jeffrey Koziol M.D. Thursday, December 6, 12 Choices and Vision Jeffrey Koziol M.D. How does the eye work? What is myopia? What is hyperopia? What is astigmatism? What is presbyopia? How the eye works How the Eye Works 3 How the eye works Light rays

More information

Auto Ref/ Keratometer / Auto Refractometer ARK-1 / AR-1

Auto Ref/ Keratometer / Auto Refractometer ARK-1 / AR-1 Auto Ref/ Keratometer / Auto Refractometer ARK-1 / AR-1 The Superior Auto and Auto What is the superior auto ref / The ARK-1 / AR-1 series speak for themselves, and auto refractometer generating greater

More information

WITH ACCUMULATING EXPERIENCE AND CONtinuing

WITH ACCUMULATING EXPERIENCE AND CONtinuing Comparison of Corneal Wavefront Aberrations After Photorefractive Keratectomy and Laser In Situ Keratomileusis TETSURO OSHIKA, MD, STEPHEN D. KLYCE, PHD, RAYMOND A. APPLEGATE, OD, PHD, HOWARD C. HOWLAND,

More information

Auto Ref/ Keratometer / Auto Refractometer ARK-1 / AR-1

Auto Ref/ Keratometer / Auto Refractometer ARK-1 / AR-1 Auto Ref/ Keratometer / Auto Refractometer ARK-1 / AR-1 The Superior Auto and Auto What is the superior auto ref / The ARK-1 / AR-1 series speak for themselves, and auto refractometer generating greater

More information

In this issue of the Journal, Oliver and colleagues

In this issue of the Journal, Oliver and colleagues Special Article Refractive Surgery, Optical Aberrations, and Visual Performance Raymond A. Applegate, OD, PhD; Howard C. Howland,PhD In this issue of the Journal, Oliver and colleagues report that photorefractive

More information

Update on Aspheric IOL Technology

Update on Aspheric IOL Technology Peer-Reviewed Literature: Update on Aspheric IOL Technology Editor: Ming Wang, MD, PhD, Clinical Associate Professor of Ophthalmology at the University of Tennessee and Director of the Wang Vision Institute

More information

What is Wavefront Aberration? Custom Contact Lenses For Vision Improvement Are They Feasible In A Disposable World?

What is Wavefront Aberration? Custom Contact Lenses For Vision Improvement Are They Feasible In A Disposable World? Custom Contact Lenses For Vision Improvement Are They Feasible In A Disposable World? Ian Cox, BOptom, PhD, FAAO Distinguished Research Fellow Bausch & Lomb, Rochester, NY Acknowledgements Center for Visual

More information

Posterior corneal aberrations and their compensation effects on anterior corneal. aberrations in keratoconic eyes. Minghan Chen and Geunyoung Yoon

Posterior corneal aberrations and their compensation effects on anterior corneal. aberrations in keratoconic eyes. Minghan Chen and Geunyoung Yoon Page 1 of 34 Papers in Press. Published on July 18, 2008 as Manuscript iovs.08-1874 Posterior corneal aberrations and their compensation effects on anterior corneal aberrations in keratoconic eyes Minghan

More information

Refractive Surgery: Vance Thompson, MD, FACS Refractive Surgeon. Oculeve Wavetec Zeiss Mynosys LRG Equinox Precision Lens ORA Amaken EXCELLens

Refractive Surgery: Vance Thompson, MD, FACS Refractive Surgeon. Oculeve Wavetec Zeiss Mynosys LRG Equinox Precision Lens ORA Amaken EXCELLens Refractive Surgery: My Way Vance Thompson, MD, FACS Refractive Surgeon Vance Thompson Vision Sioux Falls, SD Disclosures Abbott Medical Optics Alcon Avedro Calhoun Euclid Systems EyeBrain Medical Forsight

More information

ORIGINAL ARTICLE. Metrics of Retinal Image Quality Predict Visual Performance in Eyes With 20/17 or Better Visual Acuity

ORIGINAL ARTICLE. Metrics of Retinal Image Quality Predict Visual Performance in Eyes With 20/17 or Better Visual Acuity 1040-5488/06/8309-0635/0 VOL. 83, NO. 9, PP. 635 640 OPTOMETRY AND VISION SCIENCE Copyright 2006 American Academy of Optometry ORIGINAL ARTICLE Metrics of Retinal Image Quality Predict Visual Performance

More information

Auto Ref/ Keratometer / Auto Refractometer ARK-1 / AR-1

Auto Ref/ Keratometer / Auto Refractometer ARK-1 / AR-1 Auto Ref/ Keratometer / Auto Refractometer ARK-1 / AR-1 The Superior Auto Ref / Keratometer and Auto Refractometer What is the superior auto ref / keratometer and auto refractometer? The ARK-1 / AR-1 series

More information

Quality of Vision With Multifocal Progressive Diffractive Lens: Two-Year Follow-up

Quality of Vision With Multifocal Progressive Diffractive Lens: Two-Year Follow-up Quality of Vision With Multifocal Progressive Diffractive Lens: Two-Year Follow-up Antonio Mocellin, MD & Matteo Piovella, MD CMA, Centro di Microchirurgia Ambulatoriale Monza (Milan) Italy Dr Piovella

More information

Corporate Perspective Alcon Unanswered Technical Challenges that Still Need to be Overcome

Corporate Perspective Alcon Unanswered Technical Challenges that Still Need to be Overcome Corporate Perspective Alcon Unanswered Technical Challenges that Still Need to be Overcome Ronald Krueger, MD Refractive Industry Challenges Diagnostic Improvement Optimal Laser Performance Corneal Factors

More information

Aberrations and Visual Performance: Part I: How aberrations affect vision

Aberrations and Visual Performance: Part I: How aberrations affect vision Aberrations and Visual Performance: Part I: How aberrations affect vision Raymond A. Applegate, OD, Ph.D. Professor and Borish Chair of Optometry University of Houston Houston, TX, USA Aspects of this

More information

Causes of refractive error post premium IOL s 3/17/2015. Instruction course: Refining the Refractive Error After Premium IOL s.

Causes of refractive error post premium IOL s 3/17/2015. Instruction course: Refining the Refractive Error After Premium IOL s. Instruction course: Refining the Refractive Error After Premium IOL s. Senior Instructor: Mounir Khalifa, MD Instructors: David Hardten,MD Scott MacRea,MD Matteo Piovella,MD Dr. Khalifa: Causes of refractive

More information

The pupil of the eye is a critical limiting factor in the optics

The pupil of the eye is a critical limiting factor in the optics Pupil Location under Mesopic, Photopic, and Pharmacologically Dilated Conditions Yabo Yang, 1,2 Keith Thompson, 3 and Stephen A. Burns 1 PURPOSE. To determine whether there are systematic changes in pupil

More information

Financial Disclosure. Acufocus. Presbyopia Surgery. Inlay Concept 8/14/17. Presbyopia Correction: The Holy Grail of Ophthalmology

Financial Disclosure. Acufocus. Presbyopia Surgery. Inlay Concept 8/14/17. Presbyopia Correction: The Holy Grail of Ophthalmology Acufocus Financial Disclosure I have no financial interest in any subject presented Presbyopia Correction: The Holy Grail of Ophthalmology Presbyopia Surgery Inlay Concept First conceived in 1949 by Dr.

More information

Principles and clinical applications of ray-tracing aberrometry (Part II)

Principles and clinical applications of ray-tracing aberrometry (Part II) UPDATE/REVIEW Principles and clinical applications of ray-tracing aberrometry (Part II) Alfredo Castillo Gómez, MD, PhD 1 ; Antonio Verdejo del Rey, OD 2 ; Carlos Palomino Bautista, MD 3 ; Ana Escalada

More information

TRANSLATIONAL SCIENCE. Effect of Crystalline Lens Aberrations on Adaptive Optics Simulation of Intraocular Lenses

TRANSLATIONAL SCIENCE. Effect of Crystalline Lens Aberrations on Adaptive Optics Simulation of Intraocular Lenses TRANSLATIONAL SCIENCE Effect of Crystalline Lens Aberrations on Adaptive Optics Simulation of Intraocular Lenses Eloy A. Villegas, PhD; Silvestre Manzanera, PhD; Carmen M. Lago, MSc; Lucía Hervella, MSc;

More information

PERSPECTIVE THE PRESENCE OF OPTICAL ABERRATIONS THAT BLUR. Making Sense Out of Wavefront Sensing

PERSPECTIVE THE PRESENCE OF OPTICAL ABERRATIONS THAT BLUR. Making Sense Out of Wavefront Sensing PERSPECTIVE Making Sense Out of Wavefront Sensing JAY S. PEPOSE, MD, PHD AND RAYMOND A. APPLEGATE, OD, PHD THE PRESENCE OF OPTICAL ABERRATIONS THAT BLUR retinal images were the subject of popular lectures

More information

Customized intraocular lenses

Customized intraocular lenses Customized intraocular lenses Challenges and limitations Achim Langenbucher, Simon Schröder & Timo Eppig Customized IOL what does this mean? Aspherical IOL Diffractive multifocal IOL Spherical IOL Customized

More information

Characterizing the Wave Aberration in Eyes with Keratoconus or Penetrating Keratoplasty Using a High Dynamic Range Wavefront Sensor

Characterizing the Wave Aberration in Eyes with Keratoconus or Penetrating Keratoplasty Using a High Dynamic Range Wavefront Sensor Characterizing the Wave Aberration in Eyes with Keratoconus or Penetrating Keratoplasty Using a High Dynamic Range Wavefront Sensor Seth Pantanelli, MS, 1,2 Scott MacRae, MD, 3 Tae Moon Jeong, PhD, 2 Geunyoung

More information

ATLAS Corneal Topography System

ATLAS Corneal Topography System ATLAS Corneal Topography System Simply accurate for maximum productivity Model 9000 The New ATLAS Take your practice to the next level Carl Zeiss Meditec has taken the world s leading corneal topography

More information

DesiGneD and ManUFaCTUreD in italy The sharpest vision.

DesiGneD and ManUFaCTUreD in italy The sharpest vision. The sharpest vision. DESIGNED AND MANUFACTURED in ItalY 2 A complete diagnostic station used in clinical practice and research to analyze the optical environment of ocular aberration. Its functions are:

More information

Choices and Vision. Jeffrey Koziol M.D. Friday, December 7, 12

Choices and Vision. Jeffrey Koziol M.D. Friday, December 7, 12 Choices and Vision Jeffrey Koziol M.D. How does the eye work? What is myopia? What is hyperopia? What is astigmatism? What is presbyopia? How the eye works Light rays enter the eye through the clear cornea,

More information

Rediscover quality of life thanks to vision correction with technology from Carl Zeiss. Patient Information

Rediscover quality of life thanks to vision correction with technology from Carl Zeiss. Patient Information Rediscover quality of life thanks to vision correction with technology from Carl Zeiss Patient Information 5 2 It was really w Vision defects: Light that goes astray For clear vision the eyes, cornea and

More information

Treatment of Presbyopia during Crystalline Lens Surgery A Review

Treatment of Presbyopia during Crystalline Lens Surgery A Review Treatment of Presbyopia during Crystalline Lens Surgery A Review Pierre Bouchut Bordeaux Ophthalmic surgeons should treat presbyopia during crystalline lens surgery. Thanks to the quality and advancements

More information

Ron Liu OPTI521-Introductory Optomechanical Engineering December 7, 2009

Ron Liu OPTI521-Introductory Optomechanical Engineering December 7, 2009 Synopsis of METHOD AND APPARATUS FOR IMPROVING VISION AND THE RESOLUTION OF RETINAL IMAGES by David R. Williams and Junzhong Liang from the US Patent Number: 5,777,719 issued in July 7, 1998 Ron Liu OPTI521-Introductory

More information

Corneal functional optical zone under monocular and binocular assessment

Corneal functional optical zone under monocular and binocular assessment Arba Mosquera et al. Eye and Vision (2018) 5:3 DOI 10.1186/s40662-018-0097-y RESEARCH Corneal functional optical zone under monocular and binocular assessment Samuel Arba Mosquera 1,2,3*, Diego de Ortueta

More information

FITTING GUIDE PRACTITIONER S ROSE K2 KC ROSE K2 NC ROSE K2 IC ROSE K2 PG NIPPLE CONE IRREGULAR CORNEA POST GRAFT

FITTING GUIDE PRACTITIONER S ROSE K2 KC ROSE K2 NC ROSE K2 IC ROSE K2 PG NIPPLE CONE IRREGULAR CORNEA POST GRAFT Keratoconus Nipple Cone Irregular Cornea Post Graft PRACTITIONER S FITTING GUIDE NIPPLE CONE IRREGULAR CORNEA POST GRAFT Four lens designs... One simple systematic approach to fitting Featuring Easy-to-fit

More information

Dawn of a New Era Modern LASIK Results. Christopher L. Blanton, MD April 28, 2018

Dawn of a New Era Modern LASIK Results. Christopher L. Blanton, MD April 28, 2018 Dawn of a New Era Modern LASIK Results Christopher L. Blanton, MD April 28, 2018 Financial Disclosure Paid consultant: Johnson & Johnson, Inc.- Star S4/iFS IntraLase Medical Monitor Integra LifeSciences,

More information

Although the presence of optical imperfections

Although the presence of optical imperfections Validation of the Estimation of Corneal Aberrations From Videokeratography in Keratoconus Sergio Barbero, BSc; Susana Marcos, PhD; Jesus Merayo-Lloves, MD, PhD; Esther Moreno-Barriuso, PhD ABSTRACT PURPOSE:

More information

(495) (495)

(495) (495) МЕДТЕХНИКА-СТОЛИЦА (495) 902-59-26 (495) 518-55-99 127 238, г. Москва, Дмитровское ш. 85 ATLAS Corneal Topography Product Overview Model 9000 ATLAS Model 9000 Overview Next-generation corneal topography

More information

Author Contact Information: Erik Gross VISX Incorporated 3400 Central Expressway Santa Clara, CA, 95051

Author Contact Information: Erik Gross VISX Incorporated 3400 Central Expressway Santa Clara, CA, 95051 Author Contact Information: Erik Gross VISX Incorporated 3400 Central Expressway Santa Clara, CA, 95051 Telephone: 408-773-7117 Fax: 408-773-7253 Email: erikg@visx.com Improvements in the Calculation and

More information

Wavefront-Guided Programmable Spectacles Related Metrics

Wavefront-Guided Programmable Spectacles Related Metrics Wavefront-Guided Programmable Spectacles Related Metrics Lawrence Sverdrup, Sean Sigarlaki, Jeffrey Chomyn, Jagdish Jethmalani, Andreas Dreher Ophthonix, Inc. 23rd February 2007 Outline Background on Ophthonix

More information

Página 1 de 9 TopPage > Eye Care > Diagnostic > Wave-Front Analyzer KR-1W Wave-Front Analyzer KR-1W Perfection for Professionals : KR-1W Topcon, with its wealth of experience in designing and manufacturing

More information

LASIK & Refractive Surgery

LASIK & Refractive Surgery LASIK & Refractive Surgery LASIK PRK ICL RLE Monovision + + + Understanding the Basics: Why You Need Vision Correction What is a refraction and refractive error? First and foremost, we should give you

More information

Crystalens AO: Accommodating, Aberration-Free, Aspheric Y. Ralph Chu, MD Chu Vision Institute Bloomington, MN

Crystalens AO: Accommodating, Aberration-Free, Aspheric Y. Ralph Chu, MD Chu Vision Institute Bloomington, MN Crystalens AO: Accommodating, Aberration-Free, Aspheric Y. Ralph Chu, MD Chu Vision Institute Bloomington, MN Financial Disclosure Advanced Medical Optics Allergan Bausch & Lomb PowerVision Revision Optics

More information

Optical Quality of the Eye in Subjects with Normal and Excellent Visual Acuity METHODS. Subjects

Optical Quality of the Eye in Subjects with Normal and Excellent Visual Acuity METHODS. Subjects Optical Quality of the ye in Subjects with Normal and xcellent Visual Acuity loy A. Villegas, ncarna Alcón, and Pablo Artal From the Laboratorio de Optica, Departamento de Fisica, Universidad de Murcia,

More information

What you should know about LASIK and Femto-LASIK Understanding Laser Vision Correction. Patient information

What you should know about LASIK and Femto-LASIK Understanding Laser Vision Correction. Patient information What you should know about LASIK and Femto-LASIK Understanding Laser Vision Correction Patient information The strong desire to see clearly Laser Vision Correction with a proven procedure Sharp vision

More information

Optical Zone Diameters for Photorefractive Corneal Surgery

Optical Zone Diameters for Photorefractive Corneal Surgery Optical Zone Diameters for Photorefractive Corneal Surgery Calvin W. Roberts and Charles J. Koesterf Purpose. To examine the physiological optics of photorefractive corneal surgery and to study the effect

More information

Explanation of Aberration and Wavefront

Explanation of Aberration and Wavefront Explanation of Aberration and Wavefront 1. What Causes Blur? 2. What is? 4. What is wavefront? 5. Hartmann-Shack Aberrometer 6. Adoption of wavefront technology David Oh 1. What Causes Blur? 2. What is?

More information

Visual Outcomes of Two Aspheric PCIOLs: Tecnis Z9000 versus Akreos AO

Visual Outcomes of Two Aspheric PCIOLs: Tecnis Z9000 versus Akreos AO Visual Outcomes of Two Aspheric PCIOLs: Tecnis Z9000 versus Akreos AO Ahmad-Reza Baghi, MD; Mohammad-Reza Jafarinasab, MD; Hossein Ziaei, MD; Zahra Rahmani, MD Shaheed Beheshti Medical University, Tehran,

More information

THE ASPHERIC PRELOADED INJECTION

THE ASPHERIC PRELOADED INJECTION New from STAAR for Safe and Easy Implantation THE ASPHERIC PRELOADED INJECTION S Y S T E M PRELOADED INJECTION SYSTEM FOR SIMPLE AND EASY IOL DELIVERY SAFE The KS3-Ai features an aspheric IOL preloaded

More information

THE SHARPEST VISION. DESIGNED AND MANUFACTURED IN ITALY

THE SHARPEST VISION. DESIGNED AND MANUFACTURED IN ITALY THE SHARPEST VISION. DESIGNED AND MANUFACTURED IN ITALY 2 A complete diagnostic station used in clinical practice and research to analyze the optical environment of ocular aberration. Its functions are:

More information

Centration axis in refractive surgery

Centration axis in refractive surgery Mosquera et al. Eye and Vision (2015) 2:4 DOI 10.1186/s40662-015-0014-6 REVIEW Centration axis in refractive surgery Samuel Arba Mosquera 1,2,3*, Shwetabh Verma 1 and Colm McAlinden 4,5 Open Access Abstract

More information

Multifocal Progressive Diffractive Lens to Improve Light Distribuition and Avoid Light Loss: Two Years Clinical Results

Multifocal Progressive Diffractive Lens to Improve Light Distribuition and Avoid Light Loss: Two Years Clinical Results Multifocal Progressive Diffractive Lens to Improve Light Distribuition and Avoid Light Loss: Two Years Clinical Results Matteo Piovella MD & Barbara Kusa MD CMA, Centro di Microchirurgia Ambulatoriale

More information

Trouble Shooting Guide for Ortho-K lenses

Trouble Shooting Guide for Ortho-K lenses Trouble Shooting Guide for Ortho-K lenses The basic design of the third generation e Lens for Orthokeratology 1. Optic Zone (Base curve, Compression zone, BC) width 5.6 to 6.4mm 2. Fitting curve (second

More information

The Aberration-Free IOL:

The Aberration-Free IOL: The Aberration-Free IOL: Advanced Optical Performance Independent of Patient Profile Griffith E. Altmann, M.S., M.B.A.; Keith H. Edwards, BSc FCOptom Dip CLP FAAO, Bausch & Lomb Some of these results were

More information

phone extn.3662, fax: , nitt.edu ABSTRACT

phone extn.3662, fax: , nitt.edu ABSTRACT Analysis of Refractive errors in the human eye using Shack Hartmann Aberrometry M. Jesson, P. Arulmozhivarman, and A.R. Ganesan* Department of Physics, National Institute of Technology, Tiruchirappalli

More information

OpenStax-CNX module: m Vision Correction * OpenStax

OpenStax-CNX module: m Vision Correction * OpenStax OpenStax-CNX module: m42484 1 Vision Correction * OpenStax This work is produced by OpenStax-CNX and licensed under the Creative Commons Attribution License 3.0 Abstract Identify and discuss common vision

More information

4/2/2015. Bonnie An Henderson MD Clinical Professor of Ophthalmology Tufts University School of Medicine Ophthalmic Consultants of Boston

4/2/2015. Bonnie An Henderson MD Clinical Professor of Ophthalmology Tufts University School of Medicine Ophthalmic Consultants of Boston Imaging Modalities That Help Maximize Uncorrected Visual Outcomes After Cataract Surgery Dr.BonnieAn Henderson, MD Dr. Kalpana Narendran, DNB Dr.Prabhu vijayaraghavan, M.S, FICO Dr. Sandra Chandramouli,DNB

More information