Assessing Visual Quality With the Point Spread Function Using the NIDEK OPD-Scan II

Size: px
Start display at page:

Download "Assessing Visual Quality With the Point Spread Function Using the NIDEK OPD-Scan II"

Transcription

1 Assessing Visual Quality With the Point Spread Function Using the NIDEK OPD-Scan II Edoardo A. Ligabue, MD; Cristina Giordano, OD ABSTRACT PURPOSE: To present the use of the point spread function (PSF) metric pre- and postoperatively for the assessment of visual quality in cataract and refractive surgery. METHODS: Case examples of cataract and refractive surgery and the effect of accommodation are presented. All PSF measurements were obtained using the NIDEK OPD-Scan II, and corneal internal and total aberrations were simulated using the NIDEK OPD-Station software. All eyes underwent corneal topography, wavefront aberrometry, autorefraction, keratometry, and pupillometry measurements pre- and postoperatively using the OPD- Scan II. The PSF was used to assess visual quality. RESULTS: Four case examples including refractive surgery, aspheric multifocal intraocular lens (IOL) implantation, toric IOL implantation, and the difference in PSF due to accommodation are presented. CONCLUSIONS: The PSF in refractive and cataract surgery is a clinically valuable metric to evaluate the patient s visual quality pre- and postoperatively. The OPD-Scan II can measure the effect of accommodation and distant fi xation providing an assessment of changes in visual quality. [J Refract Surg. 2009;25:S104-S109.] T he measurement of high-contrast visual acuity is not an adequate representation of the patient s visual quality. Visual acuity can be described by the following expression 1 : V A = 1/, where represents the minimum angle of resolution. Snellen acuity is a measure of the ability of the eye to resolve details, and the recorded value provides a metric of the functional integrity of the optical system. However, vision is a complex neural process, and the perception of scotopic symptoms such as halos and glare cannot be described by a single value metric. Theoretically, a perfect optical system would focus all incoming light rays into one single point that would not cause halos or glare. In the human eye, however, light is distorted, and this distortion can be measured using wavefront aberrometry. 2 The wave theory of light can be used to explain interference and the diffraction primarily responsible for less-than-perfect optics in the human eye that form the basis of the point spread function (PSF). 3 For example, when the crest of one light wave meets the crest of another, the crests combine and undergo constructive interference. Waves that combine in this manner are considered in phase. Alternately, when the crest of one light wave meets the trough of another, the light waves cancel and undergo destructive interference. These waves are considered out of phase. Diffraction is defined as any deviation of light rays from a rectilinear path which cannot be interpreted as reflection or refraction. 2 Diffraction occurs when light waves meet an aperture (eg, the pupil in the eye) or an obstacle (eg, the diffractive steps in a multifocal intraocular lens [IOL]). A small aperture causes greater diffraction compared with a large aperture. In the eye, the pupil serves as the aperture, and optics of the eye determine the wavefront aberrations that can be used to simulate objective visual quality using the PSF. 4 From CDI Eye Clinic, Milan, Italy. The authors have no financial interest in the materials presented herein. Correspondence: Edoardo A. Ligabue, MD, CDI Eye Clinic, Corso di Porta Vittoria 5, Milan, Italy. Tel: ; Fax: ; edoliga@tiscali.it S104 journalofrefractivesurgery.com

2 Geometric optics defines the image of a point source as a point. Physical optics states that light waves from a point source can interfere. Optically, these effects determine the pattern of the PSF. The PSF for a perfect optical system is the Airy Disk that is calculated using Fraunhofer s approximation. 5 The PSF is the image of a point source of light formed on the retina by the eye s optical system. Every object can be considered a combination of a number of point sources of light. The mathematical process of overlapping all of these points forming an object is called a convolution, from which the PSF is calculated. 6 In a normal eye, the PSF is generally dependent on pupils because as the pupil diameter increases, so do the wavefront aberrations that directly affect the PSF. To assess visual quality, the PSF, modulation transfer function (MTF), Strehl ratio, and wavefront maps can be evaluated. The MTF represents perhaps the most intuitive representation of visual quality. Basically, it is picture of how the optics of an eye distorts a point source of light. The MTF represents the visual performance of the eye across the range of spatial frequencies used in daily life. Image quality is improved when the area under the MTF is increased or maximized. The Strehl ratio is the ratio of the peak height of the PSF being measured to the peak height of a perfect optical system. The Strehl ratio range is from 0 to 1. In a normal eye, the Strehl ratio does not reach 1, because diffraction and the optical aberrations limit the visual resolution of the eye. A typical higher order Strehl ratio value for a 5-mm pupil is The higher the Strehl ratio, the better the potential image quality. The assessment of visual quality is important because of the introduction of newer IOL and excimer laser technologies that purport to increase visual quality compared with conventional models. In addition, a more informed and more demanding patient base makes it necessary to assess the visual quality preoperatively for treatment planning. We report the use of these image quality metrics in cases of cataract and refractive surgery and assess changes due to accommodation. Clinically, the surgeon will be able to understand image perception and visual disturbances that make the patient symptomatic. PATIENTS AND METHODS The OPD-Scan II with OPD-Station software (NIDEK Co Ltd, Gamagori, Japan) was used in four cases that included toric IOL implantation, preoperative LASIK candidate assessment, aspheric multifocal IOL implantation, and the effect of accommodation on visual quality. The OPD-Scan II measures ocular wavefront, corneal topography, autorefraction, autokeratometry, and pupillometry. All measurements were performed on the same optical axis without moving the instrument or the patient; hence, the corneal and whole eye wavefront data can be directly correlated. The OPD-Station software separates corneal (anterior corneal surface) from the internal aberrations (posterior cornea surface, lenticular, vitreous, and retina) to determine the effect of each optical element in the visual system separately. The PSF, early treatment of diabetic retinopathy study (ETDRS) chart, and MTF can be simulated for the total eye or the lenticular or corneal components of the eye. The OPD-Scan and OPD-Station have been described previously. 7-9 CASE REPORTS CASE 1 Refractive Surgery Screening and Postoperative Changes. A 30-year-old woman presented for LASIK. The right eye is presented as an example. The manifest refraction was The OPD-Station evaluation is presented in Figure 1. The Holladay summary displays the total, corneal, and internal wavefront maps of the eye along with the corresponding PSF measurements. All wavefront maps in Figure 1 present the effect of lower and higher order aberrations of the eye. Using the Holladay summary, one can determine whether the corneal or internal components (or a combination) of the eye are detrimental to visual quality. In this case, the total PSF and internal PSF patterns were similar, indicating that the majority of refractive error and visual quality was determined by the internal aberrations of the eye. The wavefront maps show a bull s eye pattern consistent with myopia. The patient underwent LASIK. The corneal wavefront map did not show significant aberration 3 months postoperatively compared with preoperatively. The postoperative corneal PSF was slightly worse compared with preoperative corneal PSF, likely due to the lamellar cut. Postoperatively, the total eye PSF resembled a point of light indicating successful reduction of refractive error and higher order aberrations. The Strehl ratio for the total wavefront increased from 0 preoperatively to postoperatively, indicating an increase in visual quality. CASE 2 Cataract Screening and Toric IOL Implantation Assessment. A 65-year-old man reporting reduction in vision along with ghosting at night presented for routine assessment. Only the left eye is presented for illustrative purposes. The manifest refraction was with 20/30 vision. Slit-lamp examination revealed a cataract in both eyes. Preoperatively, the total and in- Journal of Refractive Surgery Volume 25 January (Suppl) 2009 S105

3 Figure 1. Case 1. A) Pre- and B) postoperative OPD-Station evaluation of a patient who underwent LASIK. Postoperatively, the point spread function (PSF) due to all components (PSF/OPD/Tot) is significantly smaller and less distorted than preoperatively, indicating better uncorrected visual quality. WF/OPD/Tot denotes the higher and lower order aberrations of all components of the eye combined; WF/Corn/Tot denotes the higher and lower order wavefront aberrations of the corneal front surface; and WF/Int/Tot denotes the higher and lower order wavefront aberration of the internal (primarily lenticular) aberrations of the eye. All measurements and simulations were performed for a 4-mm pupil diameter to the eighth Zernike order. ternal aberration patterns were similar, indicating the internal aberrations were causing the reduction in visual acuity (Fig 2A). The patient underwent SN60T D AcrySof toric IOL implantation (Alcon Laboratories Inc, Ft Worth, Tex). Figure 2B presents the OPD-Station evaluation 1 month postoperatively. The wavefront refraction shows 0.50 D of astigmatism postoperatively with an excellent wavefront total PSF despite a total wavefront error µm. The discrepancy between the wavefront error and PSF highlights the fact that one metric or one map cannot suffice in clinical assessment using wavefront aberrometry and that multiple indices and maps are required for a thorough assessment. The pre- and postoperative corneal PSF were somewhat different, likely because of placement of the incision. The internal total wavefront was reduced to half the preoperative value after IOL implantation. The internal wavefront map showed a band of little to no optical distortion (green), indicating that the IOL is positioned on axis without significant tilt (see Fig 2B). Significant tilt and IOL decentration lead to coma-like patterns in the wavefront maps. CASE 3 Changes in Visual Quality Due to Accommodation. During accommodation, the crystalline lens changes shape so that it is thicker centrally than peripherally. This peripheral flattening increases negative longitudinal spherical aberration. 1 This changes the asphericity and high order aberrations of the crystalline lens, which cause a change in the visual quality (Fig 3). For example, measurement of a 24-year-old patient looking at distance then near induces a myopic shift, a significant distortion in the PSF, and a 10-fold decrease in the Strehl ratio from at distance to at near (see Fig 3). During cataract formation, the amplitude of accommodation declines, causing little or no change in the PSF (Fig 4). For example, a patient with an incipient cataract shows little change in the PSF or Strehl ratio when looking at distance or near. The measurement of near and distance visual quality can be used to evaluate the effect of accommodative IOLs. For example, the PSF of this patient who received an accommodative IOL shows no difference when looking at distance ver- S106 journalofrefractivesurgery.com

4 Figure 2. Case 2. A) Pre- and B) postoperative OPD-Station evaluation of a patient who underwent toric IOL implantation. Postoperatively, the point spread function (PSF) due to all components (PSF/OPD/Tot) is significantly smaller and less distorted than preoperatively, indicating better uncorrected visual quality. Postoperatively, the internal PSF also became significantly smaller and less distorted. WF/OPD/Tot denotes the higher and lower order aberrations of all components of the eye combined; WF/Corn/Tot denotes the higher and lower order wavefront aberrations of the corneal front surface; and WF/Int/Tot denotes the higher and lower order wavefront aberration of the internal (primarily lenticular) aberrations of the eye. All measurements and simulations were performed for a 4.8-mm pupil diameter to the eighth Zernike order. sus near, indicating that the IOL remains stable when presented with a near stimulus (Fig 5). CASE 4 Aspheric IOL Implantation. Figure 6 presents OPD- Scan II evaluation of a patient who underwent implantation of an aspheric AcrySof ReStor IOL (Alcon Laboratories Inc) in which some residual positive spherical aberration was targeted in the right eye. Postoperatively, the ocular spherical aberration was µm at 5.8 mm of pupil, and corneal spherical aberration remained unchanged at µm at 5.8 mm of pupil. The PSF showed some glare, likely due to a combination of the residual astigmatism optics of the IOL. The A/B ratio on the MTF graph was 50.3%, indicating the visual performance of the patient s eyes without correction compared with a pair of emmetropic eyes with excellent visual acuity and contrast sensitivity. The H/B ratio was 62.4%; this ratio indicates the best corrected visual performance of the patient s eyes compared with the emmetropic eyes. Based on this result, only a 12% increase in visual performance will be achieved by eliminating the residual refractive error (primarily astigmatism). This small increase did not warrant an IOL exchange procedure. The patient remained asymptomatic and was satisfied with the unaided distance and near vision 6 months postoperatively. DISCUSSION Visual quality simulations and metrics represent an indication of the success of corneal refractive and cataract surgeries. Wavefront aberrometry allows the surgeon to plan the procedure and determine whether the origin of visual symptoms is optical. For example, aspheric IOLs could be selected on the basis of corneal spherical aberration values provided in the OPD-Station. The surgeon could select the IOL that leaves the patient with a slight residual spherical aberration for depth of focus or completely negate the spherical aberration depending on the patient s requirements. As with Journal of Refractive Surgery Volume 25 January (Suppl) 2009 S107

5 Figure 3. Case 3. A) Point spread function (PSF) at distance and B) PSF due to accommodation of a 24-year-old emmetropic patient. The change in higher and lower order ocular PSF (PSF/OPD/Tot) is significant when the patient is presented with a near target. All measurements and simulations were performed for a 5.8-mm pupil diameter to the eighth Zernike order. Figure 4. Case 3. A) Point spread function (PSF) at distance and B) PSF due to accommodation of a patient with an incipient cataract. The change in higher and lower order ocular PSF (PSF/OPD/Tot) is not significantly different when the patient is presented with a near stimulus of 3.00 D. Distance measurement and simulation were performed for a 2.9-mm pupil diameter to the sixth Zernike order and near measurement and simulation for a 4-mm pupil diameter to the sixth Zernike order. Figure 5. Case 3. A) Point spread function (PSF) at distance and B) PSF due to accommodation for a patient with an accommodative IOL implant. The changes in higher and lower order ocular PSF (PSF/OPD/Tot) do not differ significantly when the patient is presented with a near stimulus of 3.00 D. Distance measurement and simulation were performed for a 5.6-mm pupil diameter to the eighth Zernike order and near measurement and simulation for a 5.8-mm pupil diameter to the eighth Zernike order. LASIK candidates, patients with cataracts demand excellent uncorrected visual acuity with little loss of visual quality. Patient needs and the development of premium IOLs have changed cataract surgery from functional surgery to refractive surgery; hence, the preoperative workup should be similar to that for LASIK for which wavefront aberrometers are routinely used. The ability to measure visual quality both at near and far will allow the surgeon to determine the effects of a multifocal IOL on near visual quality. In summary, the OPD-Scan II and S108 journalofrefractivesurgery.com

6 Figure 6. Case 4. Postoperative OPD-Scan II measurement of a patient who received a multifocal aspheric IOL. Axial denotes axial corneal topography (top left); OPD map (top right) is a refractive wavefront map plotting the distribution of refractive power across the pupil due to all wavefront aberrations of the eye. The WF/OPD/Gp map (middle) plots the longitudinal spherical aberration of the eye, the PSF/OPD/Tot plots the PSF of the eye. The MFT/OPD/Tot plots the modulation transfer function of the eye. The corneal and ocular spherical aberration is shown in the red circle. All measurements and simulations were performed for a 5.9-mm pupil diameter to the eighth Zernike order. OPD-Station are important instruments for the pre- and postoperative assessment of point spread function in a variety of patients. REFERENCES 1. Visual acuity and contrast sensitivity. In: Bennett AG, Rabbetts RB, eds. Clinical Visual Optics. 3rd ed. Oxford, United Kingdom: Butterworth Heinemann; 1998: Williams DR, Porter J, Yoon G, Guirao A, Hofer H, Chen L, Cox I, MacRae SM. How far can we extend the limits of human vision? In: Krueger RR, Applegate RA, Macrae SM, eds. Wavefront Customized Visual Correction: The Quest for Super Vision II. Thorofare, NJ: SLACK Incorporated; 2004: Freeman MH, Hull CC. The nature of light sources and detectors. In: Bennett AG, Rabbetts RB, eds. Clinical Visual Optics. 11th ed. Oxford, United Kingdom: Butterworth Heinemann; 2003: Diffraction. Available at: Accessed June Ligabue E, Giglio S, Ort AA, Giordano C. Objective and subjective evaluations in vision quality after diffractive IOLs implant. Presented at: Refractive Surgery online; September 14-17, 2006; Milan, Italy. 6. Giordano C. PSF e qualità visiva: La diagnostica del segmento anteriore. Presented at: 88 Congresso Nazionale Società Oftalmologica Italiano; April 18, 2008; Rome, Italy. 7. MacRae S, Fujieda M. Slit skiascopic-guided ablation using the NIDEK laser. J Refract Surg. 2000;16:S576-S Gatinel D, Hoang-Xuan T. Measurement of combined corneal, internal, and total ocular optical quality analysis in anterior segment pathology with the OPD-scan and OPD-station. J Refract Surg. 2006;22:S1014-S Klyce SD, Karon MD, Smolek MK. Screening patients with the corneal navigator. J Refract Surg. 2005;21:S617-S622. Journal of Refractive Surgery Volume 25 January (Suppl) 2009 S109

Aberrations Before and After Implantation of an Aspheric IOL

Aberrations Before and After Implantation of an Aspheric IOL Ocular High Order Aberrations Before and After Implantation of an Aspheric IOL Fabrizio I. Camesasca, MD Massimo Vitali, Orthoptist Milan, Italy I have no financial interest to disclose Wavefront Measurement

More information

Optical Path Difference Scanning System OPD-Scan II ARK-10000

Optical Path Difference Scanning System OPD-Scan II ARK-10000 Optical Path Difference Scanning System OPD-Scan II ARK-10000 Optical Path Difference Scanning System OPD-Scan II ARK-10000 Accurate and Reliable Data for Optic Diagnostics The OPD-Scan II provides information

More information

Refractive Power / Corneal Analyzer. OPD-Scan III

Refractive Power / Corneal Analyzer. OPD-Scan III Refractive Power / Corneal Analyzer OPD-Scan III Comprehensive Vision Analysis and NIDEK, a global leader in ophthalmic and optometric equipment, has created the OPD-Scan III, the third generation aberrometer

More information

NOW. Approved for NTIOL classification from CMS Available in Quar ter Diopter Powers. Accommodating. Aberration Free. Aspheric.

NOW. Approved for NTIOL classification from CMS Available in Quar ter Diopter Powers. Accommodating. Aberration Free. Aspheric. NOW Approved for NTIOL classification from CMS Available in Quar ter Diopter Powers Accommodating. Aberration Free. Aspheric. Accommodation Meets Asphericity in AO Merging Innovation & Proven Design The

More information

10/25/2017. Financial Disclosures. Do your patients complain of? Are you frustrated by remake after remake? What is wavefront error (WFE)?

10/25/2017. Financial Disclosures. Do your patients complain of? Are you frustrated by remake after remake? What is wavefront error (WFE)? Wavefront-Guided Optics in Clinic: Financial Disclosures The New Frontier November 4, 2017 Matthew J. Kauffman, OD, FAAO, FSLS STAPLE Program Soft Toric and Presbyopic Lens Education Gas Permeable Lens

More information

Corneal Asphericity and Retinal Image Quality: A Case Study and Simulations

Corneal Asphericity and Retinal Image Quality: A Case Study and Simulations Corneal Asphericity and Retinal Image Quality: A Case Study and Simulations Seema Somani PhD, Ashley Tuan OD, PhD, and Dimitri Chernyak PhD VISX Incorporated, 3400 Central Express Way, Santa Clara, CA

More information

Dr. Magda Rau Eye Clinic Cham, Germany

Dr. Magda Rau Eye Clinic Cham, Germany 3 and 6 Months clinical Results after Implantation of OptiVis Diffractive-refractive Multifocal IOL Dr. Magda Rau Eye Clinic Cham, Germany Refractive zone of Progressive power for Far to Intermediate

More information

Principles and clinical applications of ray-tracing aberrometry (Part II)

Principles and clinical applications of ray-tracing aberrometry (Part II) UPDATE/REVIEW Principles and clinical applications of ray-tracing aberrometry (Part II) Alfredo Castillo Gómez, MD, PhD 1 ; Antonio Verdejo del Rey, OD 2 ; Carlos Palomino Bautista, MD 3 ; Ana Escalada

More information

Customized intraocular lenses

Customized intraocular lenses Customized intraocular lenses Challenges and limitations Achim Langenbucher, Simon Schröder & Timo Eppig Customized IOL what does this mean? Aspherical IOL Diffractive multifocal IOL Spherical IOL Customized

More information

Choices and Vision. Jeffrey Koziol M.D. Thursday, December 6, 12

Choices and Vision. Jeffrey Koziol M.D. Thursday, December 6, 12 Choices and Vision Jeffrey Koziol M.D. How does the eye work? What is myopia? What is hyperopia? What is astigmatism? What is presbyopia? How the eye works How the Eye Works 3 How the eye works Light rays

More information

Página 1 de 9 TopPage > Eye Care > Diagnostic > Wave-Front Analyzer KR-1W Wave-Front Analyzer KR-1W Perfection for Professionals : KR-1W Topcon, with its wealth of experience in designing and manufacturing

More information

Choices and Vision. Jeffrey Koziol M.D. Friday, December 7, 12

Choices and Vision. Jeffrey Koziol M.D. Friday, December 7, 12 Choices and Vision Jeffrey Koziol M.D. How does the eye work? What is myopia? What is hyperopia? What is astigmatism? What is presbyopia? How the eye works Light rays enter the eye through the clear cornea,

More information

Subjective Image Quality Metrics from The Wave Aberration

Subjective Image Quality Metrics from The Wave Aberration Subjective Image Quality Metrics from The Wave Aberration David R. Williams William G. Allyn Professor of Medical Optics Center For Visual Science University of Rochester Commercial Relationship: Bausch

More information

Q-value Adjusted Ablation PRK PRK. Allegretto Randomized control trial : .(Corneal asphericity) (PRK) Photo refractive keratectomy

Q-value Adjusted Ablation PRK PRK. Allegretto Randomized control trial : .(Corneal asphericity) (PRK) Photo refractive keratectomy 89/8/16 : 89/6/14 : 1389 /115 / Q-value Adjusted Ablation PRK PRK Allegretto 2 1 1 4 3 Q-value adjusted PRK (PRK) Standard Photorefractive Keratectomy :. Allegretto Eye-Q 75. Randomized control trial :

More information

*Simulated vision. **Individual results may vary and are not guaranteed. Visual Performance When It s Needed Most

*Simulated vision. **Individual results may vary and are not guaranteed. Visual Performance When It s Needed Most Simulated vision. Individual results may vary and are not guaranteed. Visual Performance When It s Needed Most The aspheric design of the AcrySof IQ IOL results in improved clarity and image quality. The

More information

Clinical Update for Presbyopic Lens Options

Clinical Update for Presbyopic Lens Options Clinical Update for Presbyopic Lens Options Gregory D. Searcy, M.D. Erdey Searcy Eye Group Columbus, Ohio The Problem = Spherical Optics Marginal Rays Spherical IOL Light Rays Paraxial Rays Spherical Aberration

More information

Improving Lifestyle Vision. with Small Aperture Optics

Improving Lifestyle Vision. with Small Aperture Optics Improving Lifestyle Vision with Small Aperture Optics The Small Aperture Premium Lens Solution The IC-8 small aperture intraocular lens (IOL) is a revolutionary lens that extends depth of focus by combining

More information

Optics of Wavefront. Austin Roorda, Ph.D. University of Houston College of Optometry

Optics of Wavefront. Austin Roorda, Ph.D. University of Houston College of Optometry Optics of Wavefront Austin Roorda, Ph.D. University of Houston College of Optometry Geometrical Optics Relationships between pupil size, refractive error and blur Optics of the eye: Depth of Focus 2 mm

More information

Aberrations and Visual Performance: Part I: How aberrations affect vision

Aberrations and Visual Performance: Part I: How aberrations affect vision Aberrations and Visual Performance: Part I: How aberrations affect vision Raymond A. Applegate, OD, Ph.D. Professor and Borish Chair of Optometry University of Houston Houston, TX, USA Aspects of this

More information

Quality of Vision With Multifocal Progressive Diffractive Lens: Two-Year Follow-up

Quality of Vision With Multifocal Progressive Diffractive Lens: Two-Year Follow-up Quality of Vision With Multifocal Progressive Diffractive Lens: Two-Year Follow-up Antonio Mocellin, MD & Matteo Piovella, MD CMA, Centro di Microchirurgia Ambulatoriale Monza (Milan) Italy Dr Piovella

More information

Refractive Surgery: Vance Thompson, MD, FACS Refractive Surgeon. Oculeve Wavetec Zeiss Mynosys LRG Equinox Precision Lens ORA Amaken EXCELLens

Refractive Surgery: Vance Thompson, MD, FACS Refractive Surgeon. Oculeve Wavetec Zeiss Mynosys LRG Equinox Precision Lens ORA Amaken EXCELLens Refractive Surgery: My Way Vance Thompson, MD, FACS Refractive Surgeon Vance Thompson Vision Sioux Falls, SD Disclosures Abbott Medical Optics Alcon Avedro Calhoun Euclid Systems EyeBrain Medical Forsight

More information

Causes of refractive error post premium IOL s 3/17/2015. Instruction course: Refining the Refractive Error After Premium IOL s.

Causes of refractive error post premium IOL s 3/17/2015. Instruction course: Refining the Refractive Error After Premium IOL s. Instruction course: Refining the Refractive Error After Premium IOL s. Senior Instructor: Mounir Khalifa, MD Instructors: David Hardten,MD Scott MacRea,MD Matteo Piovella,MD Dr. Khalifa: Causes of refractive

More information

4/2/2015. Bonnie An Henderson MD Clinical Professor of Ophthalmology Tufts University School of Medicine Ophthalmic Consultants of Boston

4/2/2015. Bonnie An Henderson MD Clinical Professor of Ophthalmology Tufts University School of Medicine Ophthalmic Consultants of Boston Imaging Modalities That Help Maximize Uncorrected Visual Outcomes After Cataract Surgery Dr.BonnieAn Henderson, MD Dr. Kalpana Narendran, DNB Dr.Prabhu vijayaraghavan, M.S, FICO Dr. Sandra Chandramouli,DNB

More information

Explanation of Aberration and Wavefront

Explanation of Aberration and Wavefront Explanation of Aberration and Wavefront 1. What Causes Blur? 2. What is? 4. What is wavefront? 5. Hartmann-Shack Aberrometer 6. Adoption of wavefront technology David Oh 1. What Causes Blur? 2. What is?

More information

Trust your eyes. Presbyopic treatment methods on the cornea. PresbyMAX Decision criteria and patient s acceptance

Trust your eyes. Presbyopic treatment methods on the cornea. PresbyMAX Decision criteria and patient s acceptance Trust your eyes. Directory Presbyopic treatment methods on the cornea PresbyMAX The Principle PresbyMAX Expectations and Key Factors PresbyMAX Decision criteria and patient s acceptance PresbyMAX Upcoming

More information

Multifocal Progressive Diffractive Lens to Improve Light Distribuition and Avoid Light Loss: Two Years Clinical Results

Multifocal Progressive Diffractive Lens to Improve Light Distribuition and Avoid Light Loss: Two Years Clinical Results Multifocal Progressive Diffractive Lens to Improve Light Distribuition and Avoid Light Loss: Two Years Clinical Results Matteo Piovella MD & Barbara Kusa MD CMA, Centro di Microchirurgia Ambulatoriale

More information

Optical Connection, Inc. and Ophthonix, Inc.

Optical Connection, Inc. and Ophthonix, Inc. Optical Connection, Inc. and Ophthonix, Inc. Partners in the delivery of nonsurgical vision optimization www.opticonnection.com www.ophthonix.com The human eye has optical imperfections that can not be

More information

Optical aberrations and the eye Part 3

Optical aberrations and the eye Part 3 clinical 22 Optical aberrations and the eye Part 3 In the final part of our series, Alejandro Cerviño and Dr Shehzad Naroo discuss the methods of correction required for low and high order wavefront aberrations

More information

Roadmap to presbyopic success

Roadmap to presbyopic success Roadmap to presbyopic success Miltos O Balidis MD, PhD, FEBOphth, ICOphth Early experience with Presbyopic correction 2003 Binocular Distance-Corrected Intermediate and Near Vision Binocular Distance-Corrected

More information

Crystalens AO: Accommodating, Aberration-Free, Aspheric Y. Ralph Chu, MD Chu Vision Institute Bloomington, MN

Crystalens AO: Accommodating, Aberration-Free, Aspheric Y. Ralph Chu, MD Chu Vision Institute Bloomington, MN Crystalens AO: Accommodating, Aberration-Free, Aspheric Y. Ralph Chu, MD Chu Vision Institute Bloomington, MN Financial Disclosure Advanced Medical Optics Allergan Bausch & Lomb PowerVision Revision Optics

More information

Wavefront Aberrations in Eyes With Acrysof Monofocal Intraocular Lenses

Wavefront Aberrations in Eyes With Acrysof Monofocal Intraocular Lenses Wavefront Aberrations in Eyes With Acrysof Monofocal Intraocular Lenses Prema Padmanabhan, MS; Geunyoung Yoon, PhD; Jason Porter, PhD; Srinivas K. Rao, FRCSEd; Roy J, MSc; Mitalee Choudhury, BS ABSTRACT

More information

Surgical data reveals that Q-Factor is important for good surgical outcome

Surgical data reveals that Q-Factor is important for good surgical outcome Surgical data reveals that Q-Factor is important for good surgical outcome Michael Mrochen, PhD Michael Bueeler, PhD Tobias Koller, MD Theo Seiler, MD, PhD IROC AG Institut für Refraktive und Ophthalmo-Chirurgie

More information

Aberrometry in Clinical Practice

Aberrometry in Clinical Practice Aberrometry in Clinical Practice Aravind Roy, M.S L V Prasad Eye Institute KVC Campus, Vijayawada, India No financial disclosures No conflicts of interest What is your position? Poll Question 1 1. Ophthalmologist

More information

Headline. Pseudophakic Implants, Aspherical Optics, Quality of Vision for Cataract Patients Subline. Damien Gatinel, MD, PhD

Headline. Pseudophakic Implants, Aspherical Optics, Quality of Vision for Cataract Patients Subline. Damien Gatinel, MD, PhD Headline Pseudophakic Implants, Aspherical Optics, Quality of Vision for Cataract Patients Subline Damien Gatinel, MD, PhD Introduction Vision is a complex phenomenon involving a sequence of events that

More information

Wave-Front Analyzer KR-1W

Wave-Front Analyzer KR-1W Wave-Front Analyzer Perfection for Professionals: Topcon, with its wealth of experience in designing and manufacturing refractometers and other diagnostic equipment for over 50 years, introduces a new

More information

Author Contact Information: Erik Gross VISX Incorporated 3400 Central Expressway Santa Clara, CA, 95051

Author Contact Information: Erik Gross VISX Incorporated 3400 Central Expressway Santa Clara, CA, 95051 Author Contact Information: Erik Gross VISX Incorporated 3400 Central Expressway Santa Clara, CA, 95051 Telephone: 408-773-7117 Fax: 408-773-7253 Email: erikg@visx.com Improvements in the Calculation and

More information

ROTATIONAL STABILITY MAKES THE DIFFERENCE

ROTATIONAL STABILITY MAKES THE DIFFERENCE The Bi-Flex platform the proven platform of Excellence 01 Proven Stability less than 02 Optimal biomaterials 2 degrees long term rotation 03 Posterior Toric Lens surface with marks indicating the flat

More information

Maximum Light Transmission. Pupil-independent Light Distribution. 3.75D Near Addition Improved Intermediate Vision

Maximum Light Transmission. Pupil-independent Light Distribution. 3.75D Near Addition Improved Intermediate Vision Multifocal Maximum Light Transmission Pupil-independent Light Distribution Better Visual Quality Increased Contrast Sensitivity 3.75D Near Addition Improved Intermediate Vision Visual Performance After

More information

Aberration Interaction In Wavefront Guided Custom Ablation

Aberration Interaction In Wavefront Guided Custom Ablation Aberration Interaction In Wavefront Guided Custom Ablation Scott M. MacRae MD Professor of Ophthalmology Professor of Visual Science University of Rochester Collaborators and Disclosures: Manoj Subbaram

More information

Treatment of Presbyopia during Crystalline Lens Surgery A Review

Treatment of Presbyopia during Crystalline Lens Surgery A Review Treatment of Presbyopia during Crystalline Lens Surgery A Review Pierre Bouchut Bordeaux Ophthalmic surgeons should treat presbyopia during crystalline lens surgery. Thanks to the quality and advancements

More information

NIDEK ADVANCED VISION EXCIMER LASER SYSTEM

NIDEK ADVANCED VISION EXCIMER LASER SYSTEM NIDEK ADVANCED VISION EXCIMER LASER SYSTEM NIDEK ADVANCED VISION EXCIMER LASER SYSTEM Delivering Ultimate Solutions Today NIDEK delivers NAVEX Quest, the evolutionary customized refractive surgery platform.

More information

Unique Aberration-Free IOL: A Vision that Patients

Unique Aberration-Free IOL: A Vision that Patients Unique Aberration-Free IOL: A Vision that Patients Can Appreciate An Aspheric Optic for Improved Quality of Vision n Traditional spherical IOLs create Bilateral implantation study spherical aberration

More information

What is Wavefront Aberration? Custom Contact Lenses For Vision Improvement Are They Feasible In A Disposable World?

What is Wavefront Aberration? Custom Contact Lenses For Vision Improvement Are They Feasible In A Disposable World? Custom Contact Lenses For Vision Improvement Are They Feasible In A Disposable World? Ian Cox, BOptom, PhD, FAAO Distinguished Research Fellow Bausch & Lomb, Rochester, NY Acknowledgements Center for Visual

More information

Long-term quality of vision is what every patient expects

Long-term quality of vision is what every patient expects Long-term quality of vision is what every patient expects Innovative combination of HOYA technologies provides: 1-piece aspheric lens with Vivinex hydrophobic acrylic material Unique surface treatment

More information

Advanced Technology IOLs

Advanced Technology IOLs Introduction Advanced Technology IOLs Stephen V. Scoper, MD Virginia Eye Consultants 2013 Cataract surgery has a refractive element Patient expectations are increased Close is no longer good enough The

More information

day night convinced supreme contrast sensitivity THE IOL FOR DAY & NIGHT

day night convinced supreme contrast sensitivity THE IOL FOR DAY & NIGHT day supreme contrast sensitivity night convinced THE IOL FOR DAY & NIGHT The IOL for DAY & NIGhT UnIQUE form and features Innovative blue light filtering Excellent quality of vision Maximum depth of focus

More information

DesiGneD and ManUFaCTUreD in italy The sharpest vision.

DesiGneD and ManUFaCTUreD in italy The sharpest vision. The sharpest vision. DESIGNED AND MANUFACTURED in ItalY 2 A complete diagnostic station used in clinical practice and research to analyze the optical environment of ocular aberration. Its functions are:

More information

Ron Liu OPTI521-Introductory Optomechanical Engineering December 7, 2009

Ron Liu OPTI521-Introductory Optomechanical Engineering December 7, 2009 Synopsis of METHOD AND APPARATUS FOR IMPROVING VISION AND THE RESOLUTION OF RETINAL IMAGES by David R. Williams and Junzhong Liang from the US Patent Number: 5,777,719 issued in July 7, 1998 Ron Liu OPTI521-Introductory

More information

Basics Of Retinal Image Quality

Basics Of Retinal Image Quality Slide 2 Basics Of Retinal Image Quality Slide 3 The optics of the eye are the first stage of vision. It is an extremely important stage but not the only stage. Slide 4 Broadly There Are Two Components

More information

OCULUS Pentacam AXL Always an Axial Length Ahead

OCULUS Pentacam AXL Always an Axial Length Ahead OCULUS Pentacam AXL Always an Axial Length Ahead EFFICIENCY AND BETTER WORKFLOW Your Cataract Workstation! The new Pentacam AXL is an alliance of the time-tested Pentacam technology with high-precision

More information

Evolution of Diffractive Multifocal Intraocular Lenses

Evolution of Diffractive Multifocal Intraocular Lenses Evolution of Diffractive Multifocal Intraocular Lenses Wavefront Congress February 24, 2007 Michael J. Simpson, Ph.D. Alcon Research, Ltd., Fort Worth, Texas Presentation Overview Multifocal IOLs two lens

More information

Special Publication: Ophthalmochirurgie Supplement 2/2009 (Original printed issue available in the German language)

Special Publication: Ophthalmochirurgie Supplement 2/2009 (Original printed issue available in the German language) Special Publication: Ophthalmochirurgie Supplement 2/2009 (Original printed issue available in the German language) LENTIS Mplus - The one -and and-only Non--rotationally Symmetric Multifocal Lens Multi-center

More information

THE XTRAFOCUS IS AN ELEGANT SOLUTION TO COMPLEX CASES.

THE XTRAFOCUS IS AN ELEGANT SOLUTION TO COMPLEX CASES. XtraFocus THE XTRAFOCUS IS AN ELEGANT SOLUTION TO COMPLEX CASES. CONTENT Based on the well-established principle of pinhole optics, this intraocular implant represents an innovative alternative for the

More information

Clinical Evaluation 3-month Follow-up Report

Clinical Evaluation 3-month Follow-up Report Clinical Evaluation 3-month Follow-up Report Of SeeLens HP Intraocular Lens 27 December 2010 version 1.1 1of 16 Table of Contents TABLE OF CONTENTS... 1 OBJECTIVES... 2 EFFICACY AND SAFETY ASSESSMENTS...

More information

Evaluate Your Patient s Total Visual System With the OPD-Scan III

Evaluate Your Patient s Total Visual System With the OPD-Scan III 0811_OMD553-3MH.qxd:Layout 1 8/9/11 3:08 PM Page 1 SEPTEMBER 2011 Evaluate Your Patient s Total Visual System With the OPD-Scan III New features benefit you and your patients Complete data for IOL selection

More information

Raise your expectations. Deliver theirs.

Raise your expectations. Deliver theirs. 66 EXTENDED RANGE OF VISION MONOFOCAL-LIKE DISTANCE Raise your expectations. Deliver theirs. Now you can give your patients the best of both worlds with the first and only hybrid designed monofocal-multifocal

More information

Multifocal Intraocular Lenses for the Treatment of Presbyopia: Benefits and Side-effects

Multifocal Intraocular Lenses for the Treatment of Presbyopia: Benefits and Side-effects Published on Points de Vue International Review of Ophthalmic Optics () Home > Multifocal Intraocular Lenses for the Treatment of Presbyopia: Benefits and Side-effects Multifocal Intraocular Lenses for

More information

HOYA aspherical IOL with ABC (Aspheric Balanced Curve) Design

HOYA aspherical IOL with ABC (Aspheric Balanced Curve) Design HOYA aspherical IOL with ABC (Aspheric Balanced Curve) Design Contents Basics of asphericity Visual quality and aspheric IOL Features of HOYA ABC Design 2 What is asphericity? Deviating from the spherical

More information

Design of a Test Bench for Intraocular Lens Optical Characterization

Design of a Test Bench for Intraocular Lens Optical Characterization Journal of Physics: Conference Series Design of a Test Bench for Intraocular Lens Optical Characterization To cite this article: Francisco Alba-Bueno et al 20 J. Phys.: Conf. Ser. 274 0205 View the article

More information

4th International Congress of Wavefront Sensing and Aberration-free Refractive Correction ADAPTIVE OPTICS FOR VISION: THE EYE S ADAPTATION TO ITS

4th International Congress of Wavefront Sensing and Aberration-free Refractive Correction ADAPTIVE OPTICS FOR VISION: THE EYE S ADAPTATION TO ITS 4th International Congress of Wavefront Sensing and Aberration-free Refractive Correction (Supplement to the Journal of Refractive Surgery; June 2003) ADAPTIVE OPTICS FOR VISION: THE EYE S ADAPTATION TO

More information

Retinal stray light originating from intraocular lenses and its effect on visual performance van der Mooren, Marie Huibert

Retinal stray light originating from intraocular lenses and its effect on visual performance van der Mooren, Marie Huibert University of Groningen Retinal stray light originating from intraocular lenses and its effect on visual performance van der Mooren, Marie Huibert IMPORTANT NOTE: You are advised to consult the publisher's

More information

Customized Correction of Wavefront Aberrations in Abnormal Human Eyes by Using a Phase Plate and a Customized Contact Lens

Customized Correction of Wavefront Aberrations in Abnormal Human Eyes by Using a Phase Plate and a Customized Contact Lens Journal of the Korean Physical Society, Vol. 49, No. 1, July 2006, pp. 121 125 Customized Correction of Wavefront Aberrations in Abnormal Human Eyes by Using a Phase Plate and a Customized Contact Lens

More information

NEW THE WORLD S FIRST AND ONLY SINUSOIDAL TRIFOCAL IOL

NEW THE WORLD S FIRST AND ONLY SINUSOIDAL TRIFOCAL IOL NEW THE WORLD S FIRST AND ONLY SINUSOIDAL TRIFOCAL IOL ALL TRIFOCAL IOLS ARE NOT THE SAME! Seamless Vision Near Intermediate Far Light Figure 1: Comparison of MTF Values 1,2 THE WORLD S FIRST AND ONLY

More information

Comparison of higher order aberrations with spherical and aspheric IOLs compared to normal phakic eyes

Comparison of higher order aberrations with spherical and aspheric IOLs compared to normal phakic eyes European Journal of Ophthalmology / Vol. 18 no. 5, 2008 / pp. 728-732 Comparison of higher order aberrations with spherical and aspheric IOLs compared to normal phakic eyes M. RĘKAS, K. KRIX-JACHYM, B.

More information

Accommodating IOL s History and Clinical Management

Accommodating IOL s History and Clinical Management Accommodating IOL s History and Clinical Management Bausch & Lomb Surgical Aliso Viejo, CA Genesis of an Accommodating IOL 1 Observations with Plate IOLs Stuart Cumming 1989 Some plate haptic IOL patients

More information

Corporate Perspective Alcon Unanswered Technical Challenges that Still Need to be Overcome

Corporate Perspective Alcon Unanswered Technical Challenges that Still Need to be Overcome Corporate Perspective Alcon Unanswered Technical Challenges that Still Need to be Overcome Ronald Krueger, MD Refractive Industry Challenges Diagnostic Improvement Optimal Laser Performance Corneal Factors

More information

FOR PRECISE ASTIGMATISM CORRECTION.

FOR PRECISE ASTIGMATISM CORRECTION. WHY TORIC INTRAOCULAR LENSES? FOR PRECISE ASTIGMATISM CORRECTION. PATIENT INFORMATION Cataract treatment OK, I HAVE A CATARACT. NOW WHAT? WE UNDERSTAND YOUR CONCERNS WE CAN HELP. Dear patient, Discovering

More information

IOL Types. Hazem Elbedewy. M.D., FRCS (Glasg.) Lecturer of Ophthalmology Tanta university

IOL Types. Hazem Elbedewy. M.D., FRCS (Glasg.) Lecturer of Ophthalmology Tanta university IOL Types Hazem Elbedewy M.D., FRCS (Glasg.) Lecturer of Ophthalmology Tanta university Artificial intraocular lenses are used to replace the eye natural lens when it has been removed during cataract surgery.

More information

THE BEST OF BOTH WORLDS Dual-Scheimpflug and Placido Reaching a new level in refractive screening

THE BEST OF BOTH WORLDS Dual-Scheimpflug and Placido Reaching a new level in refractive screening THE BEST OF BOTH WORLDS Dual-Scheimpflug and Placido Reaching a new level in refractive screening Clinical Applications Corneal Implant Planning The comes with a licensable corneal inlay software module

More information

Soft CL Multifocals Design and Fitting. Soft Multifocal Lens Designs. Issues Surrounding Multifocals. Blur Interpretation. Simultaneous Vision Designs

Soft CL Multifocals Design and Fitting. Soft Multifocal Lens Designs. Issues Surrounding Multifocals. Blur Interpretation. Simultaneous Vision Designs Soft CL Multifocals Design and Fitting Mark Andre, FAAO Associate Professor of Optometry Pacific University Mark Andre, FAAO is affiliated with CooperVision, as a consultant. Issues Surrounding Multifocals

More information

Internal Aberrations and Optical Quality After Femtosecond Laser Anterior Capsulotomy in Cataract Surgery

Internal Aberrations and Optical Quality After Femtosecond Laser Anterior Capsulotomy in Cataract Surgery ORIGINAL ARTICLES Internal Aberrations and Optical Quality After Femtosecond Laser Anterior Capsulotomy in Cataract Surgery Kata Miháltz, MD; Michael C. Knorz, MD; Jorge L. Alió, MD, PhD; Ágnes I. Takács,

More information

Financial Disclosure. Acufocus. Presbyopia Surgery. Inlay Concept 8/14/17. Presbyopia Correction: The Holy Grail of Ophthalmology

Financial Disclosure. Acufocus. Presbyopia Surgery. Inlay Concept 8/14/17. Presbyopia Correction: The Holy Grail of Ophthalmology Acufocus Financial Disclosure I have no financial interest in any subject presented Presbyopia Correction: The Holy Grail of Ophthalmology Presbyopia Surgery Inlay Concept First conceived in 1949 by Dr.

More information

THE SHARPEST VISION. DESIGNED AND MANUFACTURED IN ITALY

THE SHARPEST VISION. DESIGNED AND MANUFACTURED IN ITALY THE SHARPEST VISION. DESIGNED AND MANUFACTURED IN ITALY 2 A complete diagnostic station used in clinical practice and research to analyze the optical environment of ocular aberration. Its functions are:

More information

The Aberration Structure of the Keratoconic Eye

The Aberration Structure of the Keratoconic Eye The Aberration Structure of the Keratoconic Eye Geunyoung Yoon, Ph.D. Department of Ophthalmology Center for Visual Science Institute of Optics Department of Biomedical Engineering University of Rochester

More information

ATLAS Corneal Topography System

ATLAS Corneal Topography System ATLAS Corneal Topography System Simply accurate for maximum productivity Model 9000 The New ATLAS Take your practice to the next level Carl Zeiss Meditec has taken the world s leading corneal topography

More information

Cataract Information. victoriaeye.com

Cataract Information. victoriaeye.com Cataract Information victoriaeye.com Iris Cilliary Body Sclera Choroid Retina Fovea Centralis Pupil Cornea Optic Disk (blind spot) Blood Vessels Lens Suspensory Ligament Optic nerve Understanding the eye.

More information

In this issue of the Journal, Oliver and colleagues

In this issue of the Journal, Oliver and colleagues Special Article Refractive Surgery, Optical Aberrations, and Visual Performance Raymond A. Applegate, OD, PhD; Howard C. Howland,PhD In this issue of the Journal, Oliver and colleagues report that photorefractive

More information

Subjective refraction OPTICS OF HUMAN EYE & REFRACTIVE ERRORS

Subjective refraction OPTICS OF HUMAN EYE & REFRACTIVE ERRORS Subjective refraction OPTICS OF HUMAN EYE & REFRACTIVE ERRORS Dr. Ali Abusharha Optics of human eye Eye as a camera Components Schematic eye and reduced eyes Axes and visual angles Optical aberrations

More information

Quality Testing of Intraocular Lenses. OptiSpheric IOL Family and WaveMaster IOL 2

Quality Testing of Intraocular Lenses. OptiSpheric IOL Family and WaveMaster IOL 2 Quality Testing of Intraocular Lenses OptiSpheric IOL Family and WaveMaster IOL 2 LEADING TO THE FUTURE OF OPTICS Optical systems have changed the world. And they will continue to do so. TRIOPTICS is significantly

More information

Corneal laser surgery is currently shifting its

Corneal laser surgery is currently shifting its Correlation Between Corneal and Total Wavefront Aberrations in Myopic Eyes Michael Mrochen, PhD; Mirko Jankov, MD; Michael Bueeler, MS; Theo Seiler, MD, PhD ABSTRACT PURPOSE: Corneal topography data expressed

More information

Quality Testing of Intraocular Lenses. OptiSpheric IOL Family and WaveMaster IOL 2

Quality Testing of Intraocular Lenses. OptiSpheric IOL Family and WaveMaster IOL 2 Quality Testing of Intraocular Lenses OptiSpheric IOL Family and WaveMaster IOL 2 LEADING TO THE FUTURE OF OPTICS Optical systems have changed the world. And they will continue to do so. TRIOPTICS is significantly

More information

AT LISA tri 839MP and AT LISA tri toric 939MP from ZEISS The innovative trifocal IOL concept providing True Living Vision to more patients

AT LISA tri 839MP and AT LISA tri toric 939MP from ZEISS The innovative trifocal IOL concept providing True Living Vision to more patients Premium Trifocal MICS OVDs IOLs AT LISA tri 839MP and AT LISA tri toric 939MP from ZEISS The innovative trifocal IOL concept providing True Living Vision to more patients The moment you help your patients

More information

LASIK & Refractive Surgery

LASIK & Refractive Surgery LASIK & Refractive Surgery LASIK PRK ICL RLE Monovision + + + Understanding the Basics: Why You Need Vision Correction What is a refraction and refractive error? First and foremost, we should give you

More information

Visual outcomes and higherorder aberrations of wavefront vs. combined wavefront aspheric myopic LASIK

Visual outcomes and higherorder aberrations of wavefront vs. combined wavefront aspheric myopic LASIK PHILIPPINE JOURNAL OF Ophthalmology Vol. 36 No. 1 Ja n ua r y June 211 ORIGINAL ARTICLE Robert Edward T. Ang, MD 1,2 Aimee Rose A. Icasiano-Ramirez, MD 2 Gladness Henna A. Martinez, MD 1,2 Emerson M. Cruz,

More information

Update on Aspheric IOL Technology

Update on Aspheric IOL Technology Peer-Reviewed Literature: Update on Aspheric IOL Technology Editor: Ming Wang, MD, PhD, Clinical Associate Professor of Ophthalmology at the University of Tennessee and Director of the Wang Vision Institute

More information

NEW. AT LISA tri 839MP and AT LISA tri toric 939MP from ZEISS The innovative trifocal IOL concept providing True Living Vision to more patients

NEW. AT LISA tri 839MP and AT LISA tri toric 939MP from ZEISS The innovative trifocal IOL concept providing True Living Vision to more patients Premium Trifocal MICS OVDs IOLs NEW AT LISA tri 839MP and AT LISA tri toric 939MP from ZEISS The innovative trifocal IOL concept providing True Living Vision to more patients Trifocal toric IOL The moment

More information

7 DO IT. A SAFER WAY TO TRIFOCALITY * : ELEVATED PHASE SHIFT (EPS) ** 7 DIFFRACTIVE RINGS FOR OPTIMAL LIGHT DISTRIBUTION AND LESS DISTURBANCE

7 DO IT. A SAFER WAY TO TRIFOCALITY * : ELEVATED PHASE SHIFT (EPS) ** 7 DIFFRACTIVE RINGS FOR OPTIMAL LIGHT DISTRIBUTION AND LESS DISTURBANCE A SAFER WAY TO TRIFOCALITY * : ELEVATED PHASE SHIFT (EPS) ** 7 DIFFRACTIVE RINGS FOR OPTIMAL LIGHT DISTRIBUTION AND LESS DISTURBANCE 7 DO IT. TRIFOCAL PERFORMANCE & GLASS INDEPENDENCY UNCOMPROMISED CONTRAST

More information

Auto Ref/ Keratometer / Auto Refractometer ARK-1 / AR-1

Auto Ref/ Keratometer / Auto Refractometer ARK-1 / AR-1 Auto Ref/ Keratometer / Auto Refractometer ARK-1 / AR-1 The Superior Auto and Auto What is the superior auto ref / The ARK-1 / AR-1 series speak for themselves, and auto refractometer generating greater

More information

Auto Ref/ Keratometer / Auto Refractometer ARK-1 / AR-1

Auto Ref/ Keratometer / Auto Refractometer ARK-1 / AR-1 Auto Ref/ Keratometer / Auto Refractometer ARK-1 / AR-1 The Superior Auto and Auto What is the superior auto ref / The ARK-1 / AR-1 series speak for themselves, and auto refractometer generating greater

More information

Ocular Scatter. Rayleigh Scattering

Ocular Scatter. Rayleigh Scattering Ocular Scatter The are several sources of stray light in the eye including the cornea, transmission through the iris and the crystalline lens. Cornea tends to have Rayleigh Scatter Lens follows inverse

More information

Auto Ref/ Keratometer / Auto Refractometer ARK-1 / AR-1

Auto Ref/ Keratometer / Auto Refractometer ARK-1 / AR-1 Auto Ref/ Keratometer / Auto Refractometer ARK-1 / AR-1 The Superior Auto Ref / Keratometer and Auto Refractometer What is the superior auto ref / keratometer and auto refractometer? The ARK-1 / AR-1 series

More information

Characterizing the Wave Aberration in Eyes with Keratoconus or Penetrating Keratoplasty Using a High Dynamic Range Wavefront Sensor

Characterizing the Wave Aberration in Eyes with Keratoconus or Penetrating Keratoplasty Using a High Dynamic Range Wavefront Sensor Characterizing the Wave Aberration in Eyes with Keratoconus or Penetrating Keratoplasty Using a High Dynamic Range Wavefront Sensor Seth Pantanelli, MS, 1,2 Scott MacRae, MD, 3 Tae Moon Jeong, PhD, 2 Geunyoung

More information

Product Portfolio. Sulcoflex Pseudophakic Supplementary IOLs

Product Portfolio. Sulcoflex Pseudophakic Supplementary IOLs Product Portfolio Sulcoflex Pseudophakic Supplementary IOLs Sulcoflex Pseudophakic Supplementary IOLs For when compromise is not an option As a cataract and refractive surgeon, achieving the best possible

More information

The Impact of New Generation Aspherical Soft Contact Lenses on Quality of Vision: A Comparison with Spherical Contact Lenses and Spectacle Correction

The Impact of New Generation Aspherical Soft Contact Lenses on Quality of Vision: A Comparison with Spherical Contact Lenses and Spectacle Correction Deniz Oral, Maryo C. Kohen, Melda Yenerel, Ebru Gorgun, Sule Ziylan, Ferda Ciftci Yeditepe University Faculty of Medicine, Department of Ophthalmology, Istanbul Introduction The correction of higher order

More information

The Dysphotopsia Mystery. John J. Bussa, M.D.

The Dysphotopsia Mystery. John J. Bussa, M.D. The Dysphotopsia Mystery John J. Bussa, M.D. Cataract Surgery Cataract Surgery Desirable Traits Foldable Lens Inert (non reactive) with a memory Thin folds tight and goes through a smaller incision

More information

TRANSLATIONAL SCIENCE. Effect of Crystalline Lens Aberrations on Adaptive Optics Simulation of Intraocular Lenses

TRANSLATIONAL SCIENCE. Effect of Crystalline Lens Aberrations on Adaptive Optics Simulation of Intraocular Lenses TRANSLATIONAL SCIENCE Effect of Crystalline Lens Aberrations on Adaptive Optics Simulation of Intraocular Lenses Eloy A. Villegas, PhD; Silvestre Manzanera, PhD; Carmen M. Lago, MSc; Lucía Hervella, MSc;

More information

Forget Most Everything! The Surgical Management of Presbyopia 2/23/2016. Refraction vs. Diffraction. Presbyopic IOL s Patient Expectations

Forget Most Everything! The Surgical Management of Presbyopia 2/23/2016. Refraction vs. Diffraction. Presbyopic IOL s Patient Expectations The Surgical Management of Presbyopia Presbyopic IOL s 2011 B I L L T U L L O, O D Patient Expectations What they say is I want to be able to read The Center of a Presbyope s World What they want is Accommodation

More information

Visual Outcomes of Two Aspheric PCIOLs: Tecnis Z9000 versus Akreos AO

Visual Outcomes of Two Aspheric PCIOLs: Tecnis Z9000 versus Akreos AO Visual Outcomes of Two Aspheric PCIOLs: Tecnis Z9000 versus Akreos AO Ahmad-Reza Baghi, MD; Mohammad-Reza Jafarinasab, MD; Hossein Ziaei, MD; Zahra Rahmani, MD Shaheed Beheshti Medical University, Tehran,

More information

Wave Front Topography. ReSeeVit Evolution Topography Module for Modi Topographer

Wave Front Topography. ReSeeVit Evolution Topography Module for Modi Topographer Wave Front Topography ReSeeVit Evolution Topography Module for Modi Topographer Introduction The aberrations in the central optical zone have a greater effect than those closer to the edge. From an optical

More information

Technicians & Nurses Program

Technicians & Nurses Program ASCRS ASOA Symposium & Congress Technicians & Nurses Program May 6-10, 2016 New Orleans ADVANCED BIOMETRY AND IOL CALCULATIONS Financial Disclosures No relevant disclosures Karen Bachman, COMT, ROUB The

More information