The operation manual of spotlight 300 IR microscope

Size: px
Start display at page:

Download "The operation manual of spotlight 300 IR microscope"

Transcription

1 The operation manual of spotlight 300 IR microscope Make sure there is no sample under the microscope and then click spotlight on the desktop to open the software. You can do imaging with the image mode and line scanning and mapping with the point mode. Imaging NOTE: Spotlight must be in Image mode before you can scan an image. To switch to Image mode select Image in the Control window. 1. Load the sample. In the Stage Control window, click, or select Load Position from GoTo menu in the Stage Control Window, to drive the sample stage to the loading position. Position the sample in the center of the sample stage. Then click, or display the GoTo menu and select Center of Stage to drive the sample stage so that the sample is in the center. 2. Look at the sample and focus it. The Monitor Visible window displays a visible image. To focus this image, use the controls on the Imaging System. 3. Collect a visible image survey. The Monitor Visible window displays only the field of view of the Imaging System, which is a small area. Select a default survey size from the View menu in the Stage Control window. (To get an overall picture of your sample, click in the Stage Control window.) A visible image is collected and displayed in the Stage Control window. 4. Collect an image background. Using the joystick, move to a clear area that you wish to use for the background. In the Stage Control window, click. Check the settings in the Scan Image Background dialog and click Start to begin scanning. 5. Scan the image. Using the joystick, move back to the region you want to scan. In either the Monitor window or the Stage Control window, draw a grow box around the area you want to scan. In the Stage Control window, click. The Start Image dialog is displayed. Enter a filename for the new image and click Start Image. Data collection begins and the new image is displayed in a new window. Point Mode In Point mode, you can collect individual spectra (either at the current position or at positions you have marked). You can also collect spectra at regularly spaced intervals along a line (line-scanning) or within a marked area (mapping). In Point mode, you can choose whether the spectra you collect are in ratio or single-beam mode. If you choose ratio mode, the sample spectra are automatically ratioed against a background spectrum, so you must collect a background spectrum before you begin scanning (see Collecting a

2 Background Spectrum in Point mode). However, if you collect spectra in single-beam mode, you can ratio them against a background spectrum afterwards, (see Ratioing Single-beam Spectra). on this marker In Point mode, you can change the size and shape of the infrared aperture. The red box which is displayed at the current position represents the current aperture (see Setting the Infrared Aperture). Double-clicking displays the Infrared Aperture dialog. Collecting a Background Spectrum in Point Mode If you want to collect spectra in ratio mode, you must first collect a background spectrum. 1. Move the sample stage so that the area on the sample from which you want to collect spectra is in the current aperture. 2. Set the infrared aperture. For mapping, it is usual to set the aperture to the size of the smallest feature (see Setting an Aperture). Setting an aperture does not apply for Imaging. If you are using Spotlight in Image mode, refer to Collecting an Image Background. 3. Move the current position to the area of the sample that you want to use to collect the background spectrum. In the Stage Control window, display the Options menu and choose Set Background Position. This position is marked as the background position. 4. If you are working in transmittance you must maximize the energy reaching the detector, see maximizing the energy reaching the detector in transmittance mode. If you are working in reflectance, you do not need to maximize the energy provided that the infrared focus is the same as the visible focus. 5. In the Stage Control window, display the Scan menu and choose Background or click on. The Scan Background dialog is displayed. 6. Choose a Display Mode.

3 If you choose Overlay, the spectrum is displayed in an existing view window, if you choose New Window, the spectrum is displayed in a new view window 7. Type the Scan Parameters that you require (see Scan Parameters). 8. If the ATR is required, select Use ATR. 9. Choose OK. If the sample stage is not already at the background position, a dialog is displayed. 10. Choose Yes to collect the background spectrum at the position you previously set, or No to collect the background spectrum at the current position. If the ATR is in use, dialogs will prompt you to raise or lower the ATR as required. When the FT-IR has finished scanning if the sample stage has moved to collect the background spectrum, it returns to its previous position. The Imaging System returns to view mode and the resulting background spectrum is displayed. Again, if the ATR is in use, dialogs will prompt you to raise or lower the ATR as required. Collecting a Spectrum at the Current Aperture Position 1. Move the current position to the part of the sample where you want to collect a spectrum. 2. In the Stage Control window, display the Scan menu and choose Sample Ratio or Sample Single Beam, depending on whether you want to collect spectra in Ratio Mode or Single-beam Mode. OR Click on (ratio) or (single-beam). The Scan Sample Ratio or Scan Sample Single dialog is displayed. The Imaging System automatically changes to IR mode, and the Monitor Visible window becomes black. 3. Choose the Display Mode. If you choose Overlay, the spectrum is displayed in an existing view window, if you choose New Window, the spectrum is displayed in a new view window. 4. Type the Scan Parameters that you require. 5. If you are working in transmittance you must maximize the energy reaching the detector, see Maximizing the Energy in Transmittance.

4 6. Choose OK. When the FT-IR has finished scanning, the results are displayed and the Imaging System returns to view mode. Defining a Line Scan When you have decided on your scan parameters, and on the position of the line, you are ready to define the line scan. 1. In the Stage Control window, use the mouse to zoom in on the area where you want to position the line (see Rescaling the Stage Control Window). 2. In the Stage Control window, display the Edit menu and choose Setup Line Scan. Note: The Setup Line Scan command is not available when Spotlight is in Image mode. Switch to Point mode before Line Scanning. The Setup Line Scan dialog is displayed. 3. Choose Define LineScan. The Define Line Scan dialog is displayed. While the Define Line Scan dialog is displayed, you can use the mouse to change the size and orientation of the line. 4. Choose New. A new line scan marker is displayed, which crosses the central part of the Stage Control window, and which has X and Y intervals that are the same as the current X and Y dimensions of the infrared aperture. 5. Use the mouse to position the line scan marker on the part of the sample that you want to scan, and to change its size so that it crosses the features that you are interested in. You can move and size the line scan marker in exactly the same way as a Section Locator. Use the Monitor Visible window to position the line scan marker - its relation to visible features can then be seen clearly. The coordinates of the line scan and the number of points are updated in the Define Line Scan dialog. Notice that the interval between scan points stays the same when you change the size of the line.

5 6. Type values for either the Interval between scanning points or the Number of Points in each direction, and choose Update in the appropriate box. The line is updated to show the new intervals between line scanning points. The Update button that you choose affects the spacing of the scan points. If you choose Update in the Interval section, the interval between points stays the same, and the number of points on each axis is recalculated. If you choose Update in the Number of Points section, the number of points on each axis stays the same, and the interval is recalculated. 7. When the line is defined as you require, press OK. The Setup Line Scan dialog is displayed. 8. Choose Ratio or Single Beam mode. 9. Type the Scan Parameters that you require (see Scan Parameters). 10. If you want to automatically collect new background spectra at regular intervals during line scanning, choose Scan background every, and type a number of points in the text box. 11. If you want to automatically refocus the sample at regular intervals during line scanning choose Autofocus every, and type a number of points in the text box. If you choose Autofocus every, at the interval you have set: the Imaging System changes to visible mode; the image in the Monitor Visible window is automatically focused; the Imaging System changes to infrared mode; line scanning continues. 12. If you are collecting an ATR line scan using the ATR accessory, choose ATR Mode. See ATR Line Scanning 13. Choose OK. You are now ready to start the line scan. Starting a Line Scan

6 1. In the Stage Control window, display the Scan menu and choose Start Line Scan. Note: The Start Line Scan command is not available when Spotlight is in Image mode. Switch to Point mode before Line Scanning. The Start Line Scan dialog is displayed. The Imaging System automatically changes to IR mode, and the Monitor Visible window becomes black. 2. Type a filename for the results, or choose Select and choose a file for the results. 3. If you want to monitor the energy before you start line scanning, choose Monitor; if you want to scan a new background, choose Background. 4. Choose Summary. 5. The Line Scan Summary dialog is displayed. Make sure that the line scan setup is correct. This dialog contains a Print button, which enables you to print the summary information. 6. Choose Close. The Line Scan Summary dialog closes. 7. Choose OK in the Start Line Scan dialog. The Line Scan Information dialog is displayed. 8. If the line scan duration is too long or the file size is too large, choose Cancel, and change one or more of the settings that affect scanning time and file size. The scan parameters (Setup Line Scan dialog), the number of points (Define Line Scan dialog) and the FT-IR settings all affect the scanning time (see Scan Parameters and FT-IR settings). The Imaging System automatically changes to IR mode, and the Monitor Visible window becomes black. 9. If you want to be able to see the visible image while you are line scanning, survey the visible image first. Line scanning starts, and the progress is indicated by blue crosses, which appear on the line when each spectrum has been collected. 10. To halt the line scan before it is completed, display the Scan menu and choose Halt.

7 OR Click on. The data that has been collected so far is saved. The Imaging System returns to view mode. When the line scan is completed a message box is displayed and the Imaging System returns to view mode. 11. Choose OK. A 3D view of the data is displayed. Defining a Map 1. In the Stage Control window, use the mouse to zoom in on the area that you want to map (see Rescaling the Stage Control Window). 2. In the Stage Control window, display the Edit menu and choose Setup Map. Note: The Setup Map command is only available when Spotlight is in Point mode. The Setup Map dialog is displayed. 3. Choose Define Map. The Define Map dialog is displayed. While the Define Map dialog is displayed, you can use the mouse to change the size of the map boundary. 4. Choose New. A new map boundary is displayed, which fills the central part of the Stage Control window, and which has X and Y intervals that are the same as the current X and Y dimensions of the infrared aperture. 5. Use the mouse to position the map boundary on the part of the sample that you want to map, and to change its size so that it includes the features that you are interested in. You can move and size the map boundary in exactly the same way as a grow box. Use the Monitor Visible window to position the edges of the map boundary - the relation of the map boundary to visible features can then be seen clearly.

8 The coordinates of the map boundary and the number of points are updated in the Define Map dialog. Notice that the interval between scan points stays the same when you change the size of the map. 6. Type values for either the Interval between scanning points or the Number of Points in each direction, and choose Update in the appropriate box. The map is updated to show the new intervals between mapping points. The Update button that you choose affects the spacing of the scan points. If you choose Update in the Interval section, the interval between points stays the same, and the number of points on each axis is recalculated. If you choose Update in the Number of Points section, the number of points on each axis stays the same, and the interval is recalculated. 7. When the map is defined as you require, press OK. The Setup Map dialog is displayed. 8. Choose Ratio or Single Beam mode. For ratio mode, you must have a valid background spectrum. 9. Type the Scan Parameters that you require (see Scan Parameters). 10. If you want to automatically collect new background spectra at regular intervals during mapping, choose Scan background every, and type a number of points in the text box. 11. If you want to automatically refocus the sample at regular intervals during mapping choose Autofocus every, and type a number of points in the text box. If you choose Autofocus every, at the interval you have set: the Imaging System changes to visible mode; the image in the Monitor Visible window is automatically focused; the Imaging System changes to infrared mode; mapping continues. 12. If you are collecting an ATR map scan using the ATR accessory, choose ATR Mode. See ATR Mapping. 13. Choose OK.

9 You are now ready to start the map. Starting a Map 1. In the Stage Control window, display the Scan menu and choose Start Map. Note: The Start Map command is only available when Spotlight is in Point mode. The Start Map dialog is displayed. The Imaging System automatically changes to IR mode, and the Monitor Visible window becomes black. If you want to be able to see the visible image while you are mapping, transfer the visible image to a new view window (see Visible Images). 2. Type a filename for the results, or choose Select and choose a file for the results. 3. If you want to monitor the energy before you start mapping choose Monitor; or if you want to collect a new background spectrum, choose Background. 4. Choose Summary. The Map Summary dialog is displayed. 5. Make sure that the map setup is correct. This dialog contains a Print button, which enables you to print the summary information. 6. Choose Close. The Map Summary dialog closes. 7. Choose OK in the Start Map dialog. The Mapping Information dialog is displayed. 8. If the map duration is too long or the file size is too large, choose Cancel, and change one or more of the settings that affect scanning time and file size. The scan parameters (Setup Map dialog), the number of points (Define Map dialog) and the FT-IR settings all affect the scanning time (see Scan Parameters and FT-IR settings). Mapping starts, and the progress is indicated by blue crosses, which appear on the map when each spectrum has been collected. 9. To halt the map before it is completed, display the Scan menu and choose Halt.

10 OR Click on. The data that has been collected so far is saved. The Imaging System returns to view mode. When mapping is completed a message box is displayed. 10. Choose OK. Default views of the data are displayed. The default views are of a Total absorbance image displayed in false color format, with a crosshairs cursor linked to a spectrum view (see View Systems).

Seized Drugs Operational Guidelines for the Thermo FTIR Comparative and Analytical Division

Seized Drugs Operational Guidelines for the Thermo FTIR Comparative and Analytical Division Operational Guidelines for the Thermo FTIR Comparative and Analytical Division THERMO FOURIER TRANSFORM INFRARED (FTIR) SPECTROMETER Instrument Nicolet 4700 Series FTIR spectrometer (Serial Number AFZ0400253)

More information

Quick Start MMRC Nicolet 6700 Revised 10/24/2018

Quick Start MMRC Nicolet 6700 Revised 10/24/2018 Do Not Remove Quick Start MMRC Nicolet 6700 Revised 10/24/2018 1. Fill out the logbook before you begin. 2. Before you start the Nicolet should be on with purge gas flowing. 3. Check that the purge gas

More information

Renishaw InVia Raman microscope

Renishaw InVia Raman microscope Laser Spectroscopy Labs Renishaw InVia Raman microscope Operation instructions 1. Turn On the power switch, system power switch is located towards the back of the system on the right hand side. Wait ~10

More information

SolidWorks Tutorial 1. Axis

SolidWorks Tutorial 1. Axis SolidWorks Tutorial 1 Axis Axis This first exercise provides an introduction to SolidWorks software. First, we will design and draw a simple part: an axis with different diameters. You will learn how to

More information

PANalytical X pert Pro Gazing Incidence X-ray Reflectivity User Manual (Version: )

PANalytical X pert Pro Gazing Incidence X-ray Reflectivity User Manual (Version: ) University of Minnesota College of Science and Engineering Characterization Facility PANalytical X pert Pro Gazing Incidence X-ray Reflectivity User Manual (Version: 2012.10.17) The following instructions

More information

RENISHAW INVIA RAMAN SPECTROMETER

RENISHAW INVIA RAMAN SPECTROMETER STANDARD OPERATING PROCEDURE: RENISHAW INVIA RAMAN SPECTROMETER Purpose of this Instrument: The Renishaw invia Raman Spectrometer is an instrument used to analyze the Raman scattered light from samples

More information

Guide to Operating the Bruker FT-IR Microscope I. Basic Spectrum Acquisition with Vertex 70

Guide to Operating the Bruker FT-IR Microscope I. Basic Spectrum Acquisition with Vertex 70 Guide to Operating the Bruker FT-IR Microscope I. Basic Spectrum Acquisition with Vertex 70 Susheng Tan, Ph.D. NanoScale Fabrication and Characterization Facility, University of Pittsburgh SB60 Benedum

More information

LumaSpec 800S User Manual

LumaSpec 800S User Manual LumaSpec 800S User Manual Worldwide distribution VERSION 09112014 Prior Scientific, Ltd Cambridge, UK Prior Scientific, Inc Rockland, MA. USA Prior Scientific, GmbH Jena, Germany Prior Scientific KK Tokyo,

More information

Spotlight 150 and 200 FT-IR Microscopy Systems

Spotlight 150 and 200 FT-IR Microscopy Systems S P E C I F I C A T I O N S Spotlight 150 and 200 FT-IR Microscopy Systems FT-IR Microscopy Spotlight 200 with Frontier FT-IR Spectrometer Introduction PerkinElmer Spotlight FT-IR Microscopy Systems are

More information

OPT3: Operating Procedure for Horiba Jobin Yvon LabRam Aramis Raman/PL System See LabSpec_6_2 General User Quick Start Guide on the computer desktop

OPT3: Operating Procedure for Horiba Jobin Yvon LabRam Aramis Raman/PL System See LabSpec_6_2 General User Quick Start Guide on the computer desktop OPT3: Operating Procedure for Horiba Jobin Yvon LabRam Aramis Raman/PL System See LabSpec_6_2 General User Quick Start Guide on the computer desktop 1. Log in usage using the SMIF web site 2. Turn power

More information

Figure 1 The Raith 150 TWO

Figure 1 The Raith 150 TWO RAITH 150 TWO SOP Figure 1 The Raith 150 TWO LOCATION: Raith 150 TWO room, Lithography area, NanoFab PRIMARY TRAINER: SECONDARY TRAINER: 1. OVERVIEW The Raith 150 TWO is an ultra high resolution, low voltage

More information

ECG Analysis using the Offline Averaging Mode

ECG Analysis using the Offline Averaging Mode BIOPAC Systems, Inc. 42 Aero Camino Goleta, Ca 93117 Ph (805)685-0066 Fax (805)685-0067 www.biopac.com info@biopac.com ECG Analysis using the Offline Averaging Mode For years, cardiologists examined paper

More information

BEAMAGE KEY FEATURES AVAILABLE MODELS. CMOS Beam Profiling Cameras

BEAMAGE KEY FEATURES AVAILABLE MODELS. CMOS Beam Profiling Cameras BEAM DIAGNOS TICS Beam Profiling Cameras KEY FEATURES SPECIAL PRODUCTS OEM DETECTORS THZ DETECTORS PHOTO DETECTORS HIGH POWER SOLUTIONS POWER DETECTORS ENERGY DETECTORS MONITORS AVAILABLE MODELS Beamage-3.0

More information

Information & Instructions

Information & Instructions KEY FEATURES 1. USB 3.0 For the Fastest Transfer Rates Up to 10X faster than regular USB 2.0 connections (also USB 2.0 compatible) 2. High Resolution 4.2 MegaPixels resolution gives accurate profile measurements

More information

Your first NMR measurement

Your first NMR measurement Your first NMR measurement Introduction Select 10mM water in D2O as NMR sample. The NMR spectrum of such sample consists of only two signals: the water signal and the peak of the reference (TSP). Follow

More information

Measurement calibration in Video ToolBox Software. An example using a ruler

Measurement calibration in Video ToolBox Software. An example using a ruler Measurement calibration in Video ToolBox Software An example using a ruler Video ToolBox is capable of precise measurement using a number of different measurement tools. To use the measurement tools it

More information

Getting Started. Spectra Acquisition Tutorial

Getting Started. Spectra Acquisition Tutorial Getting Started Spectra Acquisition Tutorial ABB Bomem Inc. All Rights Reserved. This Guide and the accompanying software are copyrighted and all rights are reserved by ABB. This product, including software

More information

RAITH e-line OPERATING INSTRUCTIONS

RAITH e-line OPERATING INSTRUCTIONS RAITH e-line OPERATING INSTRUCTIONS 1) LOADING A SAMPLE a. Start the system i. On the Column PC (Right side monitor [R]), select the SmartSEM icon to on the desktop to begin the column software. ii. On

More information

LAB 2: Sampling & aliasing; quantization & false contouring

LAB 2: Sampling & aliasing; quantization & false contouring CEE 615: Digital Image Processing Spring 2016 1 LAB 2: Sampling & aliasing; quantization & false contouring A. SAMPLING: Observe the effects of the sampling interval near the resolution limit. The goal

More information

Working With Drawing Views-I

Working With Drawing Views-I Chapter 12 Working With Drawing Views-I Learning Objectives After completing this chapter you will be able to: Generate standard three views. Generate Named Views. Generate Relative Views. Generate Predefined

More information

Training Guide for Carl Zeiss AxioZoom V16 Stereo Microscope

Training Guide for Carl Zeiss AxioZoom V16 Stereo Microscope Training Guide for Carl Zeiss AxioZoom V16 Stereo Microscope ZEN 2012 Optical Imaging & Vital Microscopy Core Baylor College of Medicine (2017) Power ON Routine 1 2 If you require fluorescence imaging,

More information

Confocal imaging on the Leica TCS SP8. 1) Turn the system on. 2) Use TCS user account. 3) Start LAS X software:

Confocal imaging on the Leica TCS SP8. 1) Turn the system on. 2) Use TCS user account. 3) Start LAS X software: Confocal imaging on the Leica TCS SP8 1) Turn the system on. 2) Use TCS user account. 3) Start LAS X software: 4) Do not touch the microscope while the software is initializing. Choose your options: Turn

More information

You can easily print images using the Capture NX print function. Here we will explain the process for printing

You can easily print images using the Capture NX print function. Here we will explain the process for printing Printing - Print Size Request How do you print images to fit on particular paper sizes. Response You can easily print images using the Capture NX print function. Here we will explain the process for printing

More information

Measurement Method of High Absorbance (Low Transmittance) Samples by UH4150 INTRODUCTION

Measurement Method of High Absorbance (Low Transmittance) Samples by UH4150 INTRODUCTION INTRODUCTION With UH4150, a detector can be selected depending on the analysis purpose. When analyzing a solid sample which doesn t contain any diffuse components, by selecting the direct light detector,

More information

Exercise 2-2. Four-Wire Transmitter (Optional) EXERCISE OBJECTIVE DISCUSSION OUTLINE. Ultrasonic level transmitter DISCUSSION

Exercise 2-2. Four-Wire Transmitter (Optional) EXERCISE OBJECTIVE DISCUSSION OUTLINE. Ultrasonic level transmitter DISCUSSION Exercise 2-2 Four-Wire Transmitter (Optional) EXERCISE OBJECTIVE Become familiar with HART point-to-point connection of a four-wire transmitter. DISCUSSION OUTLINE The Discussion of this exercise covers

More information

minicad5 QuickStart Tutorial Stage: 4 Plot the Job

minicad5 QuickStart Tutorial Stage: 4 Plot the Job minicad5 QuickStart Tutorial Stage: 4 Plot the Job Aim: Plot the job to both printer / plotter and DXF file. Figure 1 - The finished product Plotting and export to DXF We wish to produce a final paper

More information

Software Manual for the Economy LBIC Demo

Software Manual for the Economy LBIC Demo printed organic photovoltaics solar testing euipment specialty materials Software Manual for the Economy LBIC Demo Introduction This demonstration program is meant to show the capabilities of the infinitypv

More information

11Beamage-3. CMOS Beam Profiling Cameras

11Beamage-3. CMOS Beam Profiling Cameras 11Beamage-3 CMOS Beam Profiling Cameras Key Features USB 3.0 FOR THE FASTEST TRANSFER RATES Up to 10X faster than regular USB 2.0 connections (also USB 2.0 compatible) HIGH RESOLUTION 2.2 MPixels resolution

More information

How to Make a Run Chart in Excel

How to Make a Run Chart in Excel How to Make a Run Chart in Excel While there are some statistical programs that you can use to make a run chart, it is simple to make in Excel, using Excel s built-in chart functions. The following are

More information

Principles and Practice

Principles and Practice Principles and Practice An Integrated Approach to Engineering Graphics and AutoCAD 2011 Randy H. Shih Oregon Institute of Technology SDC PUBLICATIONS www.sdcpublications.com Schroff Development Corporation

More information

Automated Double Aperture Accessory

Automated Double Aperture Accessory For the Cary 1, 3, 100, 300, 4, 5, 400, 500, 500i, 4000, 5000, 6000i, Deep UV Installation Category II Pollution Degree 2 Equipment Class I Table of Contents Introduction Theory Operation Installation

More information

Introduction to Simulink

Introduction to Simulink EE 460 Introduction to Communication Systems MATLAB Tutorial #3 Introduction to Simulink This tutorial provides an overview of Simulink. It also describes the use of the FFT Scope and the filter design

More information

BEAMAGE-3.0 KEY FEATURES BEAM DIAGNOSTICS AVAILABLE MODELS MAIN FUNCTIONS SEE ALSO ACCESSORIES. CMOS Beam Profiling Cameras

BEAMAGE-3.0 KEY FEATURES BEAM DIAGNOSTICS AVAILABLE MODELS MAIN FUNCTIONS SEE ALSO ACCESSORIES. CMOS Beam Profiling Cameras BEAM DIAGNOSTICS BEAM DIAGNOSTICS SPECIAL PRODUCTS OEM DETECTORS THZ DETECTORS PHOTO DETECTORS HIGH POWER DETECTORS POWER DETECTORS ENERGY DETECTORS MONITORS CMOS Beam Profiling Cameras AVAILABLE MODELS

More information

Contents STARTUP MICROSCOPE CONTROLS CAMERA CONTROLS SOFTWARE CONTROLS EXPOSURE AND CONTRAST MONOCHROME IMAGE HANDLING

Contents STARTUP MICROSCOPE CONTROLS CAMERA CONTROLS SOFTWARE CONTROLS EXPOSURE AND CONTRAST MONOCHROME IMAGE HANDLING Operations Guide Contents STARTUP MICROSCOPE CONTROLS CAMERA CONTROLS SOFTWARE CONTROLS EXPOSURE AND CONTRAST MONOCHROME IMAGE HANDLING Nikon Eclipse 90i Operations Guide STARTUP Startup Powering Up Fluorescence

More information

J. La Favre Fusion 360 Lesson 2 April 19, 2017

J. La Favre Fusion 360 Lesson 2 April 19, 2017 In this lesson, you will create a round plate with 12 counter-bored holes to fit 6-32 socket head screws. A counter-bored hole has two diameters, one to fit the threaded part of the screw and the other

More information

The Discussion of this exercise covers the following points: Differential-pressure transmitter. Differential-pressure transmitter

The Discussion of this exercise covers the following points: Differential-pressure transmitter. Differential-pressure transmitter Exercise 2-1 Two-Wire Transmitter EXERCISE OBJECTIVE Become familiar with HART point-to-point connection of a two-wire transmitter. DISCUSSION OUTLINE The Discussion of this exercise covers the following

More information

JEOL 6500 User Manual

JEOL 6500 User Manual LOG IN to your session on the computer to the left of the microscope. Starting Conditions 1. Press Ctrl-Alt-Del and log on to the microscope computer. Click on JEOL PC SEM 6500 icon. Click yes if message

More information

ERS KEY FEATURES BEAM DIAGNOSTICS MAIN FUNCTIONS AVAILABLE MODEL. CMOS Beam Profiling Camera. 1 USB 3.0 for the Fastest Transfer Rates

ERS KEY FEATURES BEAM DIAGNOSTICS MAIN FUNCTIONS AVAILABLE MODEL. CMOS Beam Profiling Camera. 1 USB 3.0 for the Fastest Transfer Rates POWER DETECTORS ENERGY DETECTORS MONITORS SPECIAL PRODUCTS OEM DETECTORS THZ DETECTORS PHOTO DETECTORS HIGH POWER DETECTORS CAMERA PROFIL- CMOS Beam Profiling Camera KEY FEATURES ERS 1 USB 3.0 for the

More information

Telephone: NICOLET ( ) or

Telephone: NICOLET ( ) or The information in this publication is provided for reference only. All information contained in this publication is believed to be correct and complete. Nicolet Instrument Corporation shall not be liable

More information

Simplified Instructions: Olympus Widefield Microscope S1230

Simplified Instructions: Olympus Widefield Microscope S1230 Contents General Microscope Operation Simple Image Capture Multi-Wavelength Capture Z-Series Timelapse Combining Capture Modes Synopsis of Other Functions Pages 2-23 24-40 41-47 48-56 57-59 60-68 69-83

More information

Automatic Tool Changer (ATC) for the prolight A Supplement to the prolight 1000 User s Guide

Automatic Tool Changer (ATC) for the prolight A Supplement to the prolight 1000 User s Guide Automatic Tool Changer (ATC) for the prolight 1000 A Supplement to the prolight 1000 User s Guide 1 1995 Light Machines Corporation All rights reserved. The information contained in this supplement (34-7221-0000)

More information

R I T. Title: Wyko RST Plus. Semiconductor & Microsystems Fabrication Laboratory Revision: A Rev Date: 05/23/06 1 SCOPE 2 REFERENCE DOCUMENTS

R I T. Title: Wyko RST Plus. Semiconductor & Microsystems Fabrication Laboratory Revision: A Rev Date: 05/23/06 1 SCOPE 2 REFERENCE DOCUMENTS Approved by: Process Engineer / / / / Equipment Engineer 1 SCOPE The purpose of this document is to detail the use of the Wyko RST Plus. All users are expected to have read and understood this document.

More information

Training Guide for Leica SP8 Confocal/Multiphoton Microscope

Training Guide for Leica SP8 Confocal/Multiphoton Microscope Training Guide for Leica SP8 Confocal/Multiphoton Microscope LAS AF v3.3 Optical Imaging & Vital Microscopy Core Baylor College of Medicine (2017) Power ON Routine 1 2 Turn ON power switch for epifluorescence

More information

PANalytical X pert Pro High Resolution Specular and Rocking Curve Scans User Manual (Version: )

PANalytical X pert Pro High Resolution Specular and Rocking Curve Scans User Manual (Version: ) University of Minnesota College of Science and Engineering Characterization Facility PANalytical X pert Pro High Resolution Specular and Rocking Curve Scans User Manual (Version: 2012.10.17) The following

More information

Create styles that control the display of Civil 3D objects. Copy styles from one drawing to another drawing.

Create styles that control the display of Civil 3D objects. Copy styles from one drawing to another drawing. NOTES Module 03 Settings and Styles In this module, you learn about the various settings and styles that are used in AutoCAD Civil 3D. A strong understanding of these basics leads to more efficient use

More information

High Point Communications Authorized dealer for USA Fleet Services

High Point Communications Authorized dealer for USA Fleet Services Summary of Display Features Idle Alert - Idle alerts can be sent every 10 minutes (up to 60 minutes) Stop Duration Alert Stop Duration alerts can be sent every 10 minutes (up to 60 minutes), after 24 hours,

More information

Draw IT 2016 for AutoCAD

Draw IT 2016 for AutoCAD Draw IT 2016 for AutoCAD Tutorial for System Scaffolding Version: 16.0 Copyright Computer and Design Services Ltd GLOBAL CONSTRUCTION SOFTWARE AND SERVICES Contents Introduction... 1 Getting Started...

More information

Tutorial 3: Drawing Objects in AutoCAD 2011

Tutorial 3: Drawing Objects in AutoCAD 2011 Tutorial 3: Drawing Objects in AutoCAD 2011 Audience: Users new to AutoCAD Prerequisites: None Time to complete: 15 minutes In This Tutorial Please complete the lessons in this tutorial in order. The earlier

More information

Creo Extrude Tutorial 2: Cutting and Adding Material

Creo Extrude Tutorial 2: Cutting and Adding Material Creo Extrude Tutorial 2: Cutting and Adding Material 1. Open Creo Parametric 2. File > Open > extrudeturial (From Creo Extrude Tutorial 1) 3. Cutting Material a. Click Extrude Icon > Select the following

More information

AutoCAD LT 2012 Tutorial. Randy H. Shih Oregon Institute of Technology SDC PUBLICATIONS. Schroff Development Corporation

AutoCAD LT 2012 Tutorial. Randy H. Shih Oregon Institute of Technology SDC PUBLICATIONS.   Schroff Development Corporation AutoCAD LT 2012 Tutorial Randy H. Shih Oregon Institute of Technology SDC PUBLICATIONS www.sdcpublications.com Schroff Development Corporation AutoCAD LT 2012 Tutorial 1-1 Lesson 1 Geometric Construction

More information

AutoCAD LT 2009 Tutorial

AutoCAD LT 2009 Tutorial AutoCAD LT 2009 Tutorial Randy H. Shih Oregon Institute of Technology SDC PUBLICATIONS Schroff Development Corporation www.schroff.com Better Textbooks. Lower Prices. AutoCAD LT 2009 Tutorial 1-1 Lesson

More information

Version 6. User Manual OBJECT

Version 6. User Manual OBJECT Version 6 User Manual OBJECT 2006 BRUKER OPTIK GmbH, Rudolf-Plank-Str. 27, D-76275 Ettlingen, www.brukeroptics.com All rights reserved. No part of this publication may be reproduced or transmitted in any

More information

NIRSpec Technical Note NTN Author(s): S. Birkmann Date of Issue: September 27, 2012 Version: 1.2

NIRSpec Technical Note NTN Author(s): S. Birkmann Date of Issue: September 27, 2012 Version: 1.2 NIRSpec Technical Note NTN-2012-002 Author(s): S. Birkmann Date of Issue: September 27, 2012 Version: 1.2 estec European Space Research and Technology Centre Keplerlaan 1 2201 AZ Noordwijk The Netherlands

More information

TeleTrader FlashChart

TeleTrader FlashChart TeleTrader FlashChart Symbols and Chart Settings With TeleTrader FlashChart you can display several symbols (for example indices, securities or currency pairs) in an interactive chart. You can also add

More information

DRCR.net Image Acquisition Protocol

DRCR.net Image Acquisition Protocol DRCR.net Image Acquisition Protocol Optical Coherence Tomography Angiography (OCT-A) Using: Optovue AngioVue Version 3.0 August 14, 2017 DRCR.net OCT-A Optovue AngioVue Procedure Manual 3.0 8-14-17 Table

More information

Unit. Drawing Accurately OVERVIEW OBJECTIVES INTRODUCTION 8-1

Unit. Drawing Accurately OVERVIEW OBJECTIVES INTRODUCTION 8-1 8-1 Unit 8 Drawing Accurately OVERVIEW When you attempt to pick points on the screen, you may have difficulty locating an exact position without some type of help. Typing the point coordinates is one method.

More information

To apply proposed roadway data (vertical alignments, cross section template data, cut/fill slopes, etc.)

To apply proposed roadway data (vertical alignments, cross section template data, cut/fill slopes, etc.) That CAD Girl J ennifer dib ona Website: www.thatcadgirl.com Email: thatcadgirl@aol.com Phone: (919) 417-8351 Fax: (919) 573-0351 Roadway Design Extracting Existing Ground Cross Sections This document

More information

LSM 780 Confocal Microscope Standard Operation Protocol

LSM 780 Confocal Microscope Standard Operation Protocol LSM 780 Confocal Microscope Standard Operation Protocol Basic Operation Turning on the system 1. Sign on log sheet according to Actual start time 2. Check Compressed Air supply for the air table 3. Switch

More information

Title: Leica SP5 Confocal User Manual

Title: Leica SP5 Confocal User Manual Title: Leica SP5 Confocal User Manual Date of first issue: 23/10/2015 Date of review: Version: Admin For assistance or to report an issue Office: CG07 or 05 Email: Igmm-imaginghelpdesk@igmm.ed.ac.uk Website:

More information

WALLY ROTARY ENCODER. USER MANUAL v. 1.0

WALLY ROTARY ENCODER. USER MANUAL v. 1.0 WALLY ROTARY ENCODER USER MANUAL v. 1.0 1.MEASUREMENTS ANGULAR POSITIONING a. General Description The angular positioning measurements are performed with the use of the Wally rotary encoder. This measurement

More information

Advanced Lab LAB 6: Signal Acquisition & Spectrum Analysis Using VirtualBench DSA Equipment: Objectives:

Advanced Lab LAB 6: Signal Acquisition & Spectrum Analysis Using VirtualBench DSA Equipment: Objectives: Advanced Lab LAB 6: Signal Acquisition & Spectrum Analysis Using VirtualBench DSA Equipment: Pentium PC with National Instruments PCI-MIO-16E-4 data-acquisition board (12-bit resolution; software-controlled

More information

User s Manual Software Edition 1.7

User s Manual Software Edition 1.7 ScanView User s Manual Software Edition 1.7 The information in this document is subject to change without notice and should not be construed as a commitment by Sitronic GmbH. While every effort has been

More information

Imaging Features Available in HTML5. it just makes sense

Imaging Features Available in HTML5. it just makes sense Imaging Features Available in HTML5 it just makes sense August, 2018 Imaging Features Available in HTML5 As part of the 5.2 SP1 release, the Images functionality is now available in HTML5 and provides

More information

Landsat 8 Pansharpen and Mosaic Geomatica 2015 Tutorial

Landsat 8 Pansharpen and Mosaic Geomatica 2015 Tutorial Landsat 8 Pansharpen and Mosaic Geomatica 2015 Tutorial On February 11, 2013, Landsat 8 was launched adding to the constellation of Earth imaging satellites. It is the seventh satellite to reach orbit

More information

Evaluation Chapter by CADArtifex

Evaluation Chapter by CADArtifex The premium provider of learning products and solutions www.cadartifex.com EVALUATION CHAPTER 2 Drawing Sketches with SOLIDWORKS In this chapter: Invoking the Part Modeling Environment Invoking the Sketching

More information

Generations Automatic Stand-Alone Lace By Bernie Griffith Generations Software

Generations Automatic Stand-Alone Lace By Bernie Griffith Generations Software We are going to create an open Italian lace. Generations software products provide advanced image processing features allowing for the creation of stand-alone lace with just a few simple techniques. A

More information

UV-Vis-NIR Spectrophotometer Quick Start Guide

UV-Vis-NIR Spectrophotometer Quick Start Guide UV-Vis-NIR Spectrophotometer Quick Start Guide The following instructions are provided as a Quick Start Guide for powering up, running measurements, and shutting down the Lambda 950 UV-Vis Spectrophotometer.

More information

Zeiss LSM 780 Protocol

Zeiss LSM 780 Protocol Zeiss LSM 780 Protocol 1) System Startup F Please note the sign-up policy. You must inform the facility at least 24 hours beforehand if you can t come; otherwise, you will receive a charge for unused time.

More information

That CAD Girl. J ennifer dib ona. Land Desktop Project Management. Website:

That CAD Girl. J ennifer dib ona. Land Desktop Project Management. Website: That CAD Girl J ennifer dib ona Website: www.thatcadgirl.com Email: thatcadgirl@aol.com Phone: (919) 417-8351 Fax: (919) 573-0351 Land Desktop Project Management When working in Land Desktop you are required

More information

Microscopy from Carl Zeiss

Microscopy from Carl Zeiss Microscopy from Carl Zeiss Contents Page Contents... 1 Introduction... 1 Starting the System... 2 Introduction to ZEN Efficient Navigation... 5 Setting up the microscope... 10 Configuring the beam path

More information

Introduction to Circular Pattern Flower Pot

Introduction to Circular Pattern Flower Pot Prerequisite Knowledge Previous knowledge of the sketching commands Line, Circle, Add Relations, Smart Dimension is required to complete this lesson. Previous examples of Revolved Boss/Base, Cut Extrude,

More information

ARCHLine.XP Interior Windows. Learning Interior. Learning material for the basics of ARCHLine.XP Interior. ARCHLine.

ARCHLine.XP Interior Windows. Learning Interior. Learning material for the basics of ARCHLine.XP Interior. ARCHLine. ARCHLine.XP Interior 2010 Windows Learning Interior Learning material for the basics of ARCHLine.XP Interior ARCHLine.XP Interior Information in this document is subject to change without notice and does

More information

Anchor Block Draft Tutorial

Anchor Block Draft Tutorial Anchor Block Draft Tutorial In the following tutorial you will create a drawing of the anchor block shown. The tutorial covers such topics as creating: Orthographic views Section views Auxiliary views

More information

How to blend, feather, and smooth

How to blend, feather, and smooth How to blend, feather, and smooth Quite often, you need to select part of an image to modify it. When you select uniform geometric areas squares, circles, ovals, rectangles you don t need to worry too

More information

Chapter 6 Title Blocks

Chapter 6 Title Blocks Chapter 6 Title Blocks In previous exercises, every drawing started by creating a number of layers. This is time consuming and unnecessary. In this exercise, we will start a drawing by defining layers

More information

1. Start with scatter plot: 2. Find corner points. 3. Capture image. 4. Corners

1. Start with scatter plot: 2. Find corner points. 3. Capture image. 4. Corners 1. Start with scatter plot: 2. Find corner points Easiest way to insert picture properly in GeoGebra is to have corner points. We see that: bottom corner is (2,10) top corner is (9,21) 3. Capture image

More information

Activity 1 Modeling a Plastic Part

Activity 1 Modeling a Plastic Part Activity 1 Modeling a Plastic Part In this activity, you will model a plastic part. When completed, your plastic part should look like the following two illustrations. While building this model, take time

More information

MicroLab 500-series Getting Started

MicroLab 500-series Getting Started MicroLab 500-series Getting Started 2 Contents CHAPTER 1: Getting Started Connecting the Hardware....6 Installing the USB driver......6 Installing the Software.....8 Starting a new Experiment...8 CHAPTER

More information

Import/Export of tendons in module StatiCa Tendon

Import/Export of tendons in module StatiCa Tendon IDEA Tutorial Import/Export of tendons in module StatiCa Tendon by using formats DXF, TXT and table format Import/Export of tendons in module StatiCa Tendon 1 Contents 1. Terminology... 2 2. Import...

More information

Module 1C: Adding Dovetail Seams to Curved Edges on A Flat Sheet-Metal Piece

Module 1C: Adding Dovetail Seams to Curved Edges on A Flat Sheet-Metal Piece 1 Module 1C: Adding Dovetail Seams to Curved Edges on A Flat Sheet-Metal Piece In this Module, we will explore the method of adding dovetail seams to curved edges such as the circumferential edge of a

More information

Supplemental Figure 1: Histogram of 63x Objective Lens z axis Calculated Resolutions. Results from the MetroloJ z axis fits for 5 beads from each

Supplemental Figure 1: Histogram of 63x Objective Lens z axis Calculated Resolutions. Results from the MetroloJ z axis fits for 5 beads from each Supplemental Figure 1: Histogram of 63x Objective Lens z axis Calculated Resolutions. Results from the MetroloJ z axis fits for 5 beads from each lens with a 1 Airy unit pinhole setting. Many water lenses

More information

AutoCAD Tutorial First Level. 2D Fundamentals. Randy H. Shih SDC. Better Textbooks. Lower Prices.

AutoCAD Tutorial First Level. 2D Fundamentals. Randy H. Shih SDC. Better Textbooks. Lower Prices. AutoCAD 2018 Tutorial First Level 2D Fundamentals Randy H. Shih SDC PUBLICATIONS Better Textbooks. Lower Prices. www.sdcpublications.com Powered by TCPDF (www.tcpdf.org) Visit the following websites to

More information

Zeiss LSM 880 Protocol

Zeiss LSM 880 Protocol Zeiss LSM 880 Protocol 1) System Startup Please note put sign-up policy. You must inform the facility at least 24 hours beforehand if you can t come; otherwise, you will receive a charge for unused time.

More information

eqwave USER MANUAL 2.21 Environmental Systems & Services Pty Ltd 8 River Street Richmond, Victoria Australia 3121

eqwave USER MANUAL 2.21 Environmental Systems & Services Pty Ltd 8 River Street Richmond, Victoria Australia 3121 eqwave USER MANUAL 2.21 Environmental Systems & Services Pty Ltd 8 River Street Richmond, Victoria Australia 3121 Phone: +61 3 8420 8999 Fax: +61 3 8420 8900 www.esands.com Table of Contents Introduction...3

More information

USER OPERATION AND MAINTENANCE MANUAL

USER OPERATION AND MAINTENANCE MANUAL 46 Robezu str. LV-1004 Riga Latvia Phone: +371-7-065-100, Fax: +371-7-065-102 Mm-wave Division in St. Petersburg, Russia Phone: +7-812-326-5924, Fax: +7-812-326-1060 USER OPERATION AND MAINTENANCE MANUAL

More information

General Measurement (CALSA) Part

General Measurement (CALSA) Part General Measurement (CALSA) Part Contests Contests 1. How to set Part conditions...1 1.1 Setting conditions... 1 1.2 Setting measurement origin and oscillation/spin conditions... 7 General Measurement

More information

Appendix B: Autocad Booklet YR 9 REFERENCE BOOKLET ORTHOGRAPHIC PROJECTION

Appendix B: Autocad Booklet YR 9 REFERENCE BOOKLET ORTHOGRAPHIC PROJECTION Appendix B: Autocad Booklet YR 9 REFERENCE BOOKLET ORTHOGRAPHIC PROJECTION To load Autocad: AUTOCAD 2000 S DRAWING SCREEN Click the start button Click on Programs Click on technology Click Autocad 2000

More information

ENSC327 Communication Systems Fall 2011 Assignment #1 Due Wednesday, Sept. 28, 4:00 pm

ENSC327 Communication Systems Fall 2011 Assignment #1 Due Wednesday, Sept. 28, 4:00 pm ENSC327 Communication Systems Fall 2011 Assignment #1 Due Wednesday, Sept. 28, 4:00 pm All problem numbers below refer to those in Haykin & Moher s book. 1. (FT) Problem 2.20. 2. (Convolution) Problem

More information

Leica Viva Image Assisted Surveying & Image Notes

Leica Viva Image Assisted Surveying & Image Notes Leica Viva Image Assisted Surveying & Image Notes Contents 1. Introduction 3. Image Notes 4. Availability 5. Summary 1. Introduction Image Assisted Surveying Camera live view of what the total station

More information

New Sketch Editing/Adding

New Sketch Editing/Adding New Sketch Editing/Adding 1. 2. 3. 4. 5. 6. 1. This button will bring the entire sketch to view in the window, which is the Default display. This is used to return to a view of the entire sketch after

More information

Zeiss 880 Training Notes Zen 2.3

Zeiss 880 Training Notes Zen 2.3 Zeiss 880 Training Notes Zen 2.3 1 Turn on the HXP 120V Lamp 2 Turn on Main Power Switch Turn on the Systems PC Switch Turn on the Components Switch. 3 4 5 Turn on the PC and log into your account. Start

More information

Minolta Scanner Plugin

Minolta Scanner Plugin Minolta Scanner Plugin For a list of Minolta digitizers and Geomagic software products with which this plugin is compatible, see Release Notes for Geomagic Minolta Plugin 7.6.0.3. Copyright 2005, Raindrop

More information

CCG Referral App - Northern Ireland

CCG Referral App - Northern Ireland CCG Referral App - Northern Ireland For practices in Northern Ireland who use the Referral Screen supplied by the CCG, the way you access the screen has been updated. A Vision App has been created to enable

More information

BIM - ARCHITECTUAL IMPORTING A SCANNED PLAN

BIM - ARCHITECTUAL IMPORTING A SCANNED PLAN BIM - ARCHITECTUAL IMPORTING A SCANNED PLAN INTRODUCTION In this section, we will demonstrate importing a plan created in another application. One of the most common starting points for a project is from

More information

Introduction to Simulink Assignment Companion Document

Introduction to Simulink Assignment Companion Document Introduction to Simulink Assignment Companion Document Implementing a DSB-SC AM Modulator in Simulink The purpose of this exercise is to explore SIMULINK by implementing a DSB-SC AM modulator. DSB-SC AM

More information

Modeling Basic Mechanical Components #1 Tie-Wrap Clip

Modeling Basic Mechanical Components #1 Tie-Wrap Clip Modeling Basic Mechanical Components #1 Tie-Wrap Clip This tutorial is about modeling simple and basic mechanical components with 3D Mechanical CAD programs, specifically one called Alibre Xpress, a freely

More information

ME scopeves Application Note #21 Calculating Responses of MIMO Systems to Multiple Forces

ME scopeves Application Note #21 Calculating Responses of MIMO Systems to Multiple Forces ME scopeves Application Note #21 Calculating Responses of MIMO Systems to Multiple Forces INTRODUCTION Driving forces and response motions of a vibrating structure are related in a very straightforward

More information

Importing and processing gel images

Importing and processing gel images BioNumerics Tutorial: Importing and processing gel images 1 Aim Comprehensive tools for the processing of electrophoresis fingerprints, both from slab gels and capillary sequencers are incorporated into

More information

FlashChart. Symbols and Chart Settings. Main menu navigation. Data compression and time period of the chart. Chart types.

FlashChart. Symbols and Chart Settings. Main menu navigation. Data compression and time period of the chart. Chart types. FlashChart Symbols and Chart Settings With FlashChart you can display several symbols (for example indices, securities or currency pairs) in an interactive chart. You can also add indicators and draw on

More information