Technical Paper CONSISTENT PRESENTATION OF MEDICAL IMAGES

Size: px
Start display at page:

Download "Technical Paper CONSISTENT PRESENTATION OF MEDICAL IMAGES"

Transcription

1 Technical Paper CONSISTENT PRESENTATION OF MEDICAL IMAGES A REPORT ON MEDICAL WORKSTATION CALIBRATION By Tom Schulte In years past, the quality of medical images was strictly controlled during the exposure, development, and viewing of hard-copy film. With today s digital imaging systems, the acquisition of digital images is still tightly controlled, following most of the same practices used for film. However, with modern softcopy (computer) display systems, image quality control is often neglected during the display and viewing process. Medical personnel need to know that each soft-copy image they view is an accurate depiction of the tissue and bone that were scanned. The presentation of each image must be repeatable to ensure that every image is shown consistently every time it is viewed. A digital image which has good diagnostic value when viewed on one viewing station should not look different when viewed on another station. This could result in difficulty in reading images, compromised diagnostic value, or even misdiagnosis. Reading difficulty and misdiagnosis due to inconsistent presentation of images can easily be avoided, however, by applying existing technical standards with the medical display calibration tools available today. Types of Medical Images There are two broad categories of patient images used in medical practices; radiology images and medical video images. Radiology images are scanner-sourced images that use some non-visible energy (e.g. X-ray, nuclear, ultrasound, magnetic) to form an interior view of human body parts. Medical video images are camera-sourced images that use visible light to form a surface view of human body parts or samples. Medical video may be motion video or still video images used in endoscopy, surgical, ophthalmology, dermatology, and microscopy applications. Due to differing characteristics between scannersourced radiology images and camera-sourced medical video images, there are different performance requirements for displays used to render the two types of images. In the following sections, we first cover aspects of display requirements for radiology imaging. We then cover medical video. Scanned Radiology Images X-ray scanning systems are designed to produce image brightness levels that are seen to have a linear relationship to the internal densities of the tissue and bone being scanned. However, these systems produce digital image data that has an approximately logarithmic relationship to the internal densities of the scanned tissue or bone. The contrast of darker areas of scanned images presented on electronic monitors must be enhanced to achieve perceptual linearity. If the logarithmic image data produced by these scanning systems were to be displayed on a computer monitor with a logarithmic luminance response*, the monitor would then produce image brightness levels that would be seen to have an approximately linear relationship to the scanned object densities. *A monitor s luminance response is the relationship between its digital drive levels (e.g ) and the resulting screen luminance. SpectraCal, Inc. Page 1

2 There are two important factors to consider, however, when displaying radiology images on computer monitors. 1. Computer monitors do not natively exhibit a logarithmic luminance response. Monitors usually have a luminance response that is approximately a power function, with an exponent (gamma) of 2.2, which is the industry srgb standard for presenting web and other photographic images. Of course, dedicated medical monitors are initially calibrated with the appropriate luminance response for radiology images. 2. The Human Visual System (HVS) exhibits an approximate logarithmic luminance response over much of its range, but not at lower brightness levels. The contrast sensitivity of the HVS depends on factors that include the absolute luminance level, the luminance difference, the ambient lighting environment, among others. Including all factors, human contrast sensitivity is approximately logarithmic above an image luminance of about 20 cd/m 2 (Figure 1). Figure 1: Above approximately 20 cd/m 2 (in yellow), human contrast sensitivity is almost purely logarithmic (linear plot with the log luminance scale). Below 20 cd/m 2 (in blue), contrast sensitivity decreases. At luminance levels below about 20 cd/m 2, however, the contrast sensitivity of the HVS diverges from its near-logarithmic response, with decreasing contrast sensitivity at lower luminance levels (Figure 1). Thus, the contrast of darker areas of scanned images presented on electronic monitors must be enhanced to achieve perceptual linearity between scanned object densities and image brightness. (Medical films on traditional light boxes produce luminance levels that are mostly above 20 cd/m 2 and don t require this contrast enhancement.) Therefore, for accurate presentation of scanned images, dedicated medical monitors are initially calibrated with the appropriate luminance response for radiology images. To display radiology images, a monitor s luminance response should be approximately logarithmic at higher luminance levels, with additional compensation for reduced HVS contrast sensitivity at lower luminance levels. Inconsistent image presentation from one monitor to another causes inconsistent patient diagnoses and healthcare decisions. Radiology Imaging Standards A wide range of monitor types are used to display scanned radiology images that are used for making patient diagnoses and healthcare decisions. Since a given radiology image may be viewed at different locations within a healthcare organization, on monitors with different native characteristics, images produced by the same image file can present completely different visual appearance, information, and characteristics on those different display devices. This inconsistent image rendering from one electronic display to another in different viewing environments, if uncorrected, causes inconsistent clinical perception. Because the human visual system adapts to overall luminance, different display devices in different locations, with different maximum luminance values, can produce consistent presentation of grayscale images, provided that they have the same luminance ratio and the same luminance response function. This consistent presentation of images across different display devices is the goal of imaging standards for medical radiology displays. SpectraCal, Inc. Page 2

3 A number of technical standards and practice guidelines for medical imaging digital displays have been developed in the United States over the past 30+ years. Although these standards haven t had regulatory authority until recently in a few locales, they have been largely accepted by those medical facilities that have implemented quality control programs for their radiology workstations. DICOM Standard The American College of Radiology (ACR) and the National Electrical Manufacturers Association (NEMA) formed a committee in 1984 that produced the Digital Imaging and Communications in Medicine (DICOM) standard. [1] Part 14 of the DICOM standard specifies a luminance response model for scanned images known as the Grayscale Standard Display Function (GSDF). The intent was to cause radiology images to be displayed on any GSDF-conforming monitor with consistent contrast across the brightness range. The DICOM GSDF is the standard luminance response model for viewing scanned medical images. It is designed to produce approximately the same viewing experience on digital monitors that we would experience when viewing radiology film. Figure 2: The DICOM GSDF function, showing the luminance on a log scale as a function of the JND index, as obtained from Barten s model of human contrast sensitivity. Note that the same perceived contrast (JND range) can result from different absolute luminance differences. The GSDF maps scanned image values to digital driving levels (signal bit levels), such that equal changes in image values produce equal changes in perceived brightness. The brightness changes are measured in terms of Just-Noticeable Differences (JNDs) of brightness. The DICOM GSDF matches a monitor s luminance response to the nonlinear perceptual characteristics of our vision, based on Barten s model of human contrast sensitivity[2,3]. At luminance levels below about 20 cd/m 2, the DICOM GSDF departs from a near-logarithmic model to match the nonlinear perceptual characteristics of the human visual system (Figure 2). Each equal step in image values should produce a perceptually equal brightness change. Figure 3: Measured luminance response of typical uncalibrated monitor, compared to response of DICOM GSDF standard. SpectraCal, Inc. Page 3

4 Each equal step in image values should produce a perceptually equal brightness change, across the monitor s entire luminance range. This equal JND spacing of image brightness steps is called perceptual linearization. The DICOM Grayscale Standard Display Function (GSDF) insures that each image level captured by an X-ray scanner is uniquely perceivable on a GSDFconforming monitor. By standardizing and mapping displays to the contrast sensitivity of the human eye, DICOM calibration insures that a given difference in gray levels appears the same from one display to the next. This is consistent presentation of images. Note that the DICOM Standard defines the GSDF luminance response model for radiology monitors, but doesn t specify any test procedures or acceptance criteria. AAPM and ACR fulfill those needs. The DICOM GSDF insures that each image level captured by an X-ray scanner is uniquely perceivable on a GSDF-conforming monitor. AAPM Guidelines Task Group No. 18 of the AAPM published their On-Line Report No Assessment of Display Performance for Medical Imaging Systems [4]. The intent of the report was To provide standard guidelines to practicing medical physicists, engineers, researchers, and radiologists for the performance evaluation of electronic displays intended for medical use. The scope of the guidelines was limited to display devices that are used to display monochrome medical images. The guidelines provide test procedures and tools for evaluating the performance of radiology imaging displays. These include procedures for qualitative (visual) assessments of display performance as well as procedures for quantitative (measured) assessments of display performance. The guidelines also provide a set of test patterns for the qualitative and quantitative assessments. The TG18 guidelines also recommended quantitative (measured) QC acceptance criteria that were achievable by CRT displays. ACR Standards The American College of Radiology (ACR), in collaboration with the American Association of Physicists in Medicine (AAPM) and the Society for Imaging Informatics in Medicine (SIIM), has published technical standards [5] that, in part, cover radiology diagnostic and clinical displays. The ACR acceptance criteria for radiology diagnostic, mammography, and clinical displays is covered below, in the Luminance Factors section. The ACR Practice Guidelines & Technical Standards recommends QC acceptance criteria that are now achievable by LCD displays. Color in Radiology Images Although grayscale monitors are most commonly used on radiology reading room workstations, color monitors (both medical and commercial grades) are now being heavily used for clinical workstations and for some diagnostic workstations. The use of color has become significant for clinical images obtained from radiography, ultrasound, nuclear, CT, and MRI, and the color information is being used more and more for decision-making. Even when color is not required, commercial grade color monitors, rather than medical grade grayscale monitors, are often deployed in primary diagnostic and clinical review imaging stations for purposes of economy. Using color monitors in these formerly grayscale applications does add the need to test for consistent and accurate presentation of both grayscale and color, to not degrade the diagnostic value of either grayscale or color images presented on these color monitors. Color Standards There is no standard yet for how color radiology images are to be rendered on medical monitors (as opposed to camera-sourced color medical video images). Medical grade color radiology monitors are calibrated using DICOM GSDF. However, DICOM calibration doesn't correct a monitor s color performance. And, DICOM calibration decreases the SpectraCal, Inc. Page 4

5 contrast of camera-sourced color medical video images. As the DICOM GSDF does for grayscale levels, there needs to be a method to create perceptual linearity between color levels in medical images. Practice guidelines and standards for rendering color on medical displays are being established. Just as the JND metric is currently used to create perceptually uniform grayscale steps, the delta E 2000 color difference metric will be used to create perceptually uniform color steps. DICOM, AAPM, and ICC are currently coordinating efforts to develop methods and standards for consistently displaying color images on medical displays and CalMED is ready to implement those standards. Luminance Factors The luminance due to ambient light reflected from a monitor screen reduces the ability to distinguish individual image levels in the darker parts of a scanned image, if the luminance of the darkest image level is close to that of the ambient luminance. The ambient luminance effectively compresses or washes out the darker parts of the image and reduces the contrast in the image (Figure 4). lux of diffuse ambient light be used in diagnostic reading rooms. This is generally sufficient to avoid most screen reflections and still provide sufficient light for the human visual system to adapt to the surrounding environment and the displays. Minimum Luminance, Ambient Offset The lowest luminance elements of an image (Lmin) will not be visible if they are darker than the ambient reflected luminance (Lamb). To insure accurate visual perception of all image luminance levels, the luminance produced on a particular monitor by an image s darkest level needs to be significantly higher than that monitor s ambient luminance level. This means that a monitor s native black level normally needs to be increased, to enable accurate perception of the darkest image elements. Because digital RGB monitors don t usually have a black level control, the image black level needs to be increased before the image pixel values are applied to the monitor as digital drive levels. This is done with a lookup table (LUT) of correction values written to the workstation graphics adapter. According to the latest ACR technical standards for electronic medical imaging, the lowest luminance level from a scanned image (Lmin) needs to be at least four times the luminance level due to reflected ambient light (Lamb). Also, the lowest image luminance level plus the ambient luminance level (L min)*, needs to be at least 1.2 cd/m 2 for mammography displays, 1.0 cd/m 2 for primary diagnostic displays, and at least 0.8 cd/m 2 for all other radiology displays. *The presence of the prime symbol in L min indicates the inclusion of ambient light. Figure 4: Ambient light, when added to a monitor s native (uncalibrated) luminance response, causes the darkest image levels to be compressed, blending them together. However, a certain amount of ambient room light is required in a reading environment, to avoid eyestrain and to facilitate movement and interaction with other items in the area. It is recommended that about 20 Note that higher levels of ambient light can be tolerated when viewing medical films on light boxes, because, as noted above, the lowest luminance levels produced by trans-illuminated films are mostly above 20 cd/m 2 ; much higher than the lowest luminance levels produced by computer monitors displaying scanned medical images. Maximum Luminance SpectraCal, Inc. Page 5

6 The maximum luminance of a scanned radiology image (Lmax) needs to be high enough to yield sufficient image contrast (JND range) to visually detect each image luminance level difference. According to the ACR technical standards, a monitor s maximum luminance needs to be at least 420 cd/m 2 for diagnostic X-ray mammography, 350 cd/m 2 for diagnostic X-ray radiography, and at least 250 cd/m 2 for all other modalities. Multiple monitors on a radiology workstation should have no greater than a 10% difference in maximum luminance between the monitors. Luminance Ratio A monitor s luminance ratio (LR) is the ratio of the monitor s maximum luminance to its minimum luminance, including ambient (L min). The monitor s LR is basically its contrast ratio in the presence of ambient light. Because the human visual system adapts to the average luminance of a still image, and is not able to optimally adapt to very bright and very dark regions at the same time, we need to limit an image s overall luminance ratio to a value for which the HVS has reasonably good adapted contrast response. If we reproduce images with a luminance ratio above about 500, no contrast will be perceived in dark regions of the images. If we reproduce images with a luminance ratio below about 250, however, the image will lack sufficient contrast between individual image levels and will appear washed out. According to the ACR technical standards, a monitor s luminance ratio needs to be about 350 for optimum contrast (on diagnostic and mammography monitors), and at least 250 for acceptable contrast (on clinical review monitors). The luminance ratio should be consistent on all radiology workstations throughout a medical facility to assure consistent presentation of images. Note that a monitor s ability to produce a desired luminance ratio depends primarily on the monitor s maximum luminance capability, along with its minimum luminance level (increased to compensate for ambient light). Medical and high end commercial grade monitors will likely be able to produce a suitable luminance ratio, but many commoditypriced commercial grade monitors may have marginal or insufficient maximum luminance (measured in cd/m 2, inaccurately called brightness) to produce a desired luminance ratio. Different display devices can give consistent presentation of images provided that they have: 1. The same luminance ratio (LR), and 2. The same luminance response model (DICOM GSDF). Bottom Line Because the human visual system adapts to overall brightness, different display devices, with different maximum luminance values, can give consistent presentation of images provided that they have: 3. The same luminance ratio (LR), and 4. The same luminance response model (DICOM GSDF). Grades of Viewing Monitors RGB monitors of various performance grades are used to view scanned images on radiology viewing stations, depending upon each viewing station s purpose, be it for primary diagnostic, clinical review, or technologist modality workstation. Monitors used for medical imaging may be purpose-built medical grade monitors or they may be commercial grade monitors. Both purpose-built medical grade monitors and commercial grade monitors are able to fill the needs for rendering radiology scanned images in various use cases. Commercial Grade Monitors Commercial grade monitors typically have lower luminance output and no luminance stabilization feature, compared to medical grade monitors. They are typically set to produce their maximum luminance output, typically cd/m 2, during initial setup. Without automatic luminance stabilization, commercial monitors are unable to SpectraCal, Inc. Page 6

7 maintain constant luminance during backlight warmup time. When the monitor is first turned on, luminance output tends to be lower until the backlight warms up. They slowly increase to maximum luminance over the first 30 to 60 minutes of run time. Also, a bright commercial monitor that produces 350 cd/m 2 when put into service decays to 250 cd/m 2 in about 18 months. Since cd/m 2 is considered the appropriate luminance range for viewing radiography images, the brightest commercial grade monitors maintain acceptable luminance for only about 18 months. All monitors used to display scanned radiology images require both initial and annual calibration and periodic conformance testing to insure conformance to industry guidelines. Medical Grade Radiology Monitors Medical grade radiology monitors, which are usually used on diagnostic stations in hospital radiology reading rooms, have several features that make them most suitable for use on critical diagnostic reading stations. Medical grade monitors are generally capable of much higher luminance than commercial monitors, typically producing cd/m 2. Medical grade monitors typically have a backlight capable of producing much more luminance than required when the monitor is new. This surplus luminance capability is beneficial because medical grade monitors have a luminance stabilization circuit, with feedback from a backlight sensor, which adjusts the backlight luminance down to the desired output level ( cd/m 2 ). Since a medical grade monitor has extra initial luminance capability, it adjusts its backlight output to compensate for natural luminance decay, keeping the luminance output constant over time. This backlight feedback circuit is capable of maintaining constant luminance not only over the long term life of the monitor (usually about 5 years), but also in the short term, during monitor warm-up. An advantage of brighter displays is that they allow radiologists to work in brighter reading environments, reducing viewing fatigue. The display s minimum luminance should always be regulated, though, to produce the optimum luminance ratio. Per ACR: For brighter monitors, L'min, should be proportionately larger to maintain the same LR. Medical grade radiology monitors may also have a front calibration sensor. The sensor will be integrated with an auto-calibration function within the monitor. Medical Grade Mammography Monitors Medical grade monitors designed for viewing mammography images are generally capable of higher luminance than even general radiology medical monitors, some over 1500 cd/m 2. This is desirable for displaying mammography images, which tend to be much darker than other X-ray images. Like general radiology medical monitors, mammography monitors have a luminance stabilization circuit, with feedback from a backlight sensor, which adjusts the backlight luminance down to the desired output level (420 cd/m 2 for mammography). The front calibration sensor that is optional on general radiology medical monitors is standard on all mammography monitors. An auto-calibration function within the monitor uses the front sensor for DICOM conformance tests and calibration. All monitors used to display scanned radiology images require both initial and annual calibration and periodic conformance testing to insure conformance to industry performance guidelines. Medical grade monitors, with their automatic luminance stabilization circuits, usually require quarterly conformance testing and only annual recalibration, while commercial grade monitors may require quarterly or monthly recalibration to assure continued conformance. SpectraCal, Inc. Page 7

8 Radiology Workstation Test Schedule Both AAPM and ACR recommend a regular schedule of workstation testing. Following is a typical test schedule that ensures continued compliance to industry standards, high reading efficiency, and no contribution to misdiagnoses due to poor image quality. Annual Tests - Performed by medical physicist Test ambient lighting Calibrate display(s) Run integrated calibration software, if available, or Calibrate with CalMED software and certified external meter. Validate display conformance Perform full range of visual tests with test patterns. Conformance test with CalMED software and certified external meter. Document Results Save CalMED conformance report. Quarterly/Monthly Tests - Performed by medical physicist or QC tech Test ambient lighting Analyze display conformance Perform luminance response visual tests (review TG18-QC and TG18-CT patterns). Conformance test with CalMED software and certified external meter. If poor conformance, calibrate display(s) Run integrated calibration software, if available, or Calibrate with CalMED software and certified external meter. Validate display conformance Conformance test with CalMED software and certified external meter. Document Results Save CalMED conformance report. Daily/Weekly Tests - Performed by workstation user Perform luminance response visual tests (review TG18-QC and/or TG18-CT patterns). Radiology Imaging CalMED Calibration The Radiology Imaging Technical QC workflow that is provided with CalMED guides us through the process of calibrating a radiology imaging display by implementing industry standard radiology imaging guidelines from DICOM, AAPM, and ACR. An outline of the workflow (Figure 5) is viewable when you open the left hand CalMED panel. Figure 5: An outline of the Radiology Imaging workflow appears when you open the left hand CalMED panel. Sections 1 and 2 of the workflow, Initial Setup and Workstation Analysis are used to test whether a display conforms to industry standards. Sections 3 and 4, Display Optimization and Workstation Calibration, are used to calibrate a display for conformance. The workflow first guides us through measuring the ambient luminance, setting our minimum luminance target for the display, and then testing the display for conformance to the standard acceptance criteria. If the test shows the display to be in conformance, we can save, view, and/or print a conformance report for the display. If the conformance test shows that the display is not in conformance, we can then proceed to the additional workflow steps to optimize any controls on the display and then calibrate the workstation. This would be followed by a repeat of the SpectraCal, Inc. Page 8

9 conformance test, to validate and document the display s conformance to the standard acceptance criteria. For a detailed walkthrough of the CalMED Radiology Imaging workflow, see the paper Medical Display Calibration Walkthrough, available for download at: pers/medicaldisplaycalibrationwalkthrough.pdf Medical Video Standards Images viewed on medical video displays originate from cameras, rather than from scanners. These endoscopic and still image cameras are used in surgery, ER, clinics, dermatology and pathology, among others. Camera-sourced medical images conform to the 1999 srgb standard from the International Electrotechnical Commission (IEC). Consumer digital cameras and internet images also follow the srgb standard, which was designed to be appropriate for brighter viewing environments. The srgb luminance response model starts with a linear section just above black (gamma of 1), then transitions to a power function at higher levels, with a gamma of 2.4. The gamma over the entire luminance range averages to about 2.2. The srgb white point (D65) and color gamut (primary colors) are the same as for studio monitors and HDTV. Medical Video CalMED Calibration The Medical Video workflow that is provided with CalMED guides us through the process of calibrating a medical video display by implementing the srgb industry standard. An outline of the workflow (Figure 6) is viewable when you open the left hand CalMED panel. Figure 6: An outline of the Medical Video workflow appears when you open the left hand CalMED panel. Sections 1 and 2 of the workflow, Initial Setup and Workstation Analysis are used to test whether a display conforms to industry standards. Sections 3 and 4, Display Optimization and Workstation Calibration, are used to calibrate a display for conformance. The workflow first guides us through testing the medical video display for conformance to the srgb standard. If the test shows the display to be in conformance, we can save, view, and/or print a conformance report for the display. If the conformance test shows that the display is not in conformance, we can then proceed to the additional workflow steps to optimize any controls on the display and then calibrate the workstation. This would be followed by a repeat of the conformance test, to validate and document the display s conformance to the srgb standard. For a detailed walkthrough of the CalMED Medical Video workflow, see the paper Medical Display Calibration Walkthrough, available for download at: pers/medicaldisplaycalibrationwalkthrough.pdf SpectraCal, Inc. Page 9

10 Technical Paper Conclusion It is critical to provide consistent presentation of medical images, in order to optimize reading efficiency and diagnostic value and to avoid misdiagnoses due to compromised image quality. The medical industry has developed, and continues to develop, technical standards and practice guidelines that follow the characteristics of the human visual system when rendering images on soft-copy electronic monitors. By maintaining a consistent luminance ratio and luminance response model plus appropriate luminance levels across all displays, you can ensure that medical images will be displayed consistently and accurately on all imaging workstations throughout a medical network. By maintaining conformance to the appropriate industry standards, both radiology imaging and medical video workstations will consistently render their respective images for efficient and accurate reading. The CalMED Radiology Imaging and Medical Video workflows provide easy to use processes for routinely testing workstation conformance to standards and, when necessary, provide fast and accurate calibration processes to bring workstations into conformance. Both CalMED workflows provide conformance reports to fully document a workstation quality assurance program. Display Profiling Solutions Page 10 A Special Report for Imaging Professionals

11 References 1. NEMA: Digital Imaging and Communications in Medicine (DICOM), DICOM PS a - Grayscale Standard Display Function. Available at 2. Barten PGJ. Physical model for contrast sensitivity of the human eye. Proc SPIE Int Soc Opt Eng.1992; 1666: pp Barten PGJ. Contrast Sensitivity of the Human Eye and Its Effects on Image Quality. Bellingham, WA: SPIE Press; Assessment of Display Performance for Medical Imaging Systems, American Association of Physicists in Medicine (AAPM), Task Group 18. Available at 5. ACR AAPM SIIM Technical Standard for Electronic Practice of Medical Imaging, Res Available at 6. ACR AAPM SIIM Practice Parameter for Determinants of Image Quality in Digital Mammography Res , Amended 2014 (Res. 39). Available at SpectraCal, Inc. Page 11

12 About the Author: Tom Schulte is Product Development Manager at SpectraCal. Tom has extensive experience in electronic systems test, calibration and service, as well as electronics test instrument design and usage and has authored numerous technical white papers. Tom was previously an Application Engineer at Sencore for over twenty years, where he was involved in video, audio, and RF test instrument design, plus training and support for electronic test equipment users. About SpectraCal: SpectraCal specializes in the tools and training necessary to achieve images representative of the original subjects for environments from low to high ambient light while achieving the dynamic range, contrast, and colorimetry necessary for the image to have the proper presentation to the viewer. SpectraCal CalMED software was developed to support medical personnel in the step by step process of workstation calibration. The foundation of medical workstation calibration is to understand the elements in a display that require adjustment and how each element inter-relates to the others. From its inception, CalMED has earned rave reviews and has become the preeminent medical workstation calibration software on the market, compatible with virtually all color meters available today. As the medical industry and display technology evolves, CalMED continues to provide the first choice for medical workstation calibration solutions. More Information: For more information on CalMED medical workstation calibration: Visit: info@spectracal.com Phone: SpectraCal, Inc Midvale Avenue N., Suite 100 Shoreline, WA info@spectracal.com SpectraCal, Inc. Page 12

Image Display and Perception

Image Display and Perception Image Display and Perception J. Anthony Seibert, Ph.D. Department of Radiology UC Davis Medical Center Sacramento, California, USA Image acquisition, display, & interpretation X-rays kvp mas Tube filtration

More information

White Paper. The Advantages of Hybrid Gamma PXL. 1 Introduction Outline of Hybrid Gamma PXL Study Results...

White Paper. The Advantages of Hybrid Gamma PXL. 1 Introduction Outline of Hybrid Gamma PXL Study Results... White Paper The Advantages of Hybrid Gamma PXL CONTENTS 1 Introduction... 2 2 Outline of Hybrid Gamma PXL... 4 3 Study... 5 4 Results... 7 5 Conclusions... 10 No.17-001 Revision A November 2017 Product

More information

SUSPENSION CRITERIA FOR IMAGE MONITORS AND VIEWING BOXES.

SUSPENSION CRITERIA FOR IMAGE MONITORS AND VIEWING BOXES. SUSPENSION CRITERIA FOR IMAGE MONITORS AND VIEWING BOXES. Tingberg, Anders Published in: Radiation Protection Dosimetry DOI: 10.1093/rpd/ncs302 Published: 2013-01-01 Link to publication Citation for published

More information

Increasing the Number of Gray Shades in Medical Display Systems How Much is Enough?

Increasing the Number of Gray Shades in Medical Display Systems How Much is Enough? Increasing the Number of Gray Shades in Medical Display Systems How Much is Enough? Tom Kimpe and Tom Tuytschaever Medical images produced by x-ray detectors, computed tomography (CT) scanners, and other

More information

Barco medical display systems. Product catalog

Barco medical display systems. Product catalog Barco medical display systems Product catalog TECHNOLOGy THAT MAkES A DIFFERENCE Backlight Output Stabilization (BLOS) Backlight Output Stabilization (BLOS) is an image stabilization system driven by a

More information

A Novel 12-bit Grayscale Topology PACSmate Revolutionary GrayBoost Technology

A Novel 12-bit Grayscale Topology PACSmate Revolutionary GrayBoost Technology A Novel 12-bit Grayscale Topology PACSmate Revolutionary GrayBoost Technology Author: Kenny Jan Date: August 3, 2009 Presented by PACSmate Technology Inc. Table of Contents 1. Introduction... 2 2. GrayBoost

More information

Practical guidelines for color calibration and quality assurance of medical displays

Practical guidelines for color calibration and quality assurance of medical displays Practical guidelines for color calibration and quality assurance of medical displays Poster No.: C-1140 Congress: ECR 2017 Type: Educational Exhibit Authors: T. Kimpe, J. Rostang, G. Van Hoey, A. Xthona

More information

ABSTRACT 1. PURPOSE 2. METHODS

ABSTRACT 1. PURPOSE 2. METHODS Perceptual uniformity of commonly used color spaces Ali Avanaki a, Kathryn Espig a, Tom Kimpe b, Albert Xthona a, Cédric Marchessoux b, Johan Rostang b, Bastian Piepers b a Barco Healthcare, Beaverton,

More information

Things you may want to know

Things you may want to know Things you may want to know 1 / 9 Last updated 18-Nov-05 Why medical monitors are special? Unlike commercial displays, Medical displays are used in the hospital and they are displaying contain life critical

More information

DICOM Conformance. DICOM Detailed Specification for Diagnostic Labs and Radiology Center Connectivity

DICOM Conformance. DICOM Detailed Specification for Diagnostic Labs and Radiology Center Connectivity DICOM Detailed Specification for Diagnostic Labs and Radiology Center Connectivity Authored by Global Engineering Team, Health Gorilla April 10, 2014 Table of Contents About Health Gorilla s Online Healthcare

More information

Chin J Radiol 2003; 28:

Chin J Radiol 2003; 28: Chin J Radiol 2003; 28: 175-180 175 modality console Image acquisition Quality assurance ISO Quality control PACS Picture Archiving and Communication System - Communication in Medicine [1] service WorkList

More information

Outline. Digital Radiography. Understanding Digital Modalities: Image Quality and Dose. Image Quality. Dose Control

Outline. Digital Radiography. Understanding Digital Modalities: Image Quality and Dose. Image Quality. Dose Control Understanding Digital Modalities: Image Quality and Dose S. Jeff Shepard, M.S. University of Texas M. D. Anderson Cancer Center Houston, Texas Special Acknowledgement: Stephen K. Thompson, M.S. William

More information

Clinical and management aspects of digital imaging and PACS

Clinical and management aspects of digital imaging and PACS Clinical and management aspects of digital imaging and PACS พญ.จามร เช อเพชระโสภณ นายกร งส ว ทยาสมาคมแห งประเทศไทย chamareec@gmail.com www.radiologythailand.org Digital imaging Abbreviation and terminology

More information

It should also be noted that with modern cameras users can choose for either

It should also be noted that with modern cameras users can choose for either White paper about color correction More drama Many application fields like digital printing industry or the human medicine require a natural display of colors. To illustrate the importance of color fidelity,

More information

Visual Requirements for High-Fidelity Display 1

Visual Requirements for High-Fidelity Display 1 Michael J Flynn, PhD Visual Requirements for High-Fidelity Display 1 The digital radiographic process involves (a) the attenuation of x rays along rays forming an orthographic projection, (b) the detection

More information

Acceptance Testing of a Digital Breast Tomosynthesis Unit

Acceptance Testing of a Digital Breast Tomosynthesis Unit Acceptance Testing of a Digital Breast Tomosynthesis Unit 2012 AAPM Spring Clinical Meeting Jessica Clements, M.S., DABR Objectives Review of technology and clinical advantages Acceptance Testing Procedures

More information

Practical Aspects of Medical Physics Surveys of Mammography Equipment and Facilities

Practical Aspects of Medical Physics Surveys of Mammography Equipment and Facilities Practical Aspects of Medical Physics Surveys of Mammography Equipment and Facilities Melissa Martin, M.S., FAAPM, FACR, FACMP AAPM Annual Meeting - Philadelphia July 19, 2010 MO-B-204C-1 Educational Objectives

More information

Digital radiography (DR) post processing techniques for pediatric radiology

Digital radiography (DR) post processing techniques for pediatric radiology Digital radiography (DR) post processing techniques for pediatric radiology St Jude Children s Research Hospital Samuel Brady, MS PhD DABR samuel.brady@stjude.org Purpose Review common issues and solutions

More information

Soft Proofing Page: 1

Soft Proofing Page: 1 Page: 1 The following instructions will help you understand the concept and practice of soft proofing as well as step you through how to soft proof through different applications. General Philosophy &

More information

Photography and graphic technology Extended colour encodings for digital image storage, manipulation and interchange. Part 4:

Photography and graphic technology Extended colour encodings for digital image storage, manipulation and interchange. Part 4: Provläsningsexemplar / Preview TECHNICAL SPECIFICATION ISO/TS 22028-4 First edition 2012-11-01 Photography and graphic technology Extended colour encodings for digital image storage, manipulation and interchange

More information

Calibration. Kent Messamore 7/23/2013. JKM 7/23/2013 Enhanced Images 1

Calibration. Kent Messamore 7/23/2013. JKM 7/23/2013 Enhanced Images 1 Calibration Kent Messamore 7/23/2013 JKM 7/23/2013 Enhanced Images 1 Predictable Consistent Results? How do you calibrate your camera? Auto White Balance in camera is inconsistent Amateur takes a single

More information

Display Profiling. How to Calibrate a. Introduction. 4 C s of Color Management. Step 1: Consistency. Step 2: Calibration

Display Profiling. How to Calibrate a. Introduction. 4 C s of Color Management. Step 1: Consistency. Step 2: Calibration Introduction Monitor previews that match the print can save a considerable amount of time and money for anyone in the photo and printing industry. Photographers and designers who need to adjust images

More information

Color Management User Guide

Color Management User Guide Color Management User Guide Edition July 2001 Phase One A/S Roskildevej 39 DK-2000 Frederiksberg Denmark Tel +45 36 46 01 11 Fax +45 36 46 02 22 Phase One U.S. 24 Woodbine Ave Northport, New York 11768

More information

Assessment of Color Displays. Michael Flynn Radiology Research Henry Ford Health System Detroit, MI

Assessment of Color Displays. Michael Flynn Radiology Research Henry Ford Health System Detroit, MI Assessment of Color Displays Learning Objectives 1. Human color vision 2. CIE color spaces 3. ICC color management 4. Color measurements 5. Medical standards for color Michael Flynn Radiology Research

More information

Digital Image Management: the Basics

Digital Image Management: the Basics Digital Image Management: the Basics Napapong Pongnapang, Ph.D. Department of Radiological Technology Faculty of Medical Technology Mahidol University Outline From screen/film to digital radiography PACS/Tele

More information

Quality Control of Full Field Digital Mammography Units

Quality Control of Full Field Digital Mammography Units Quality Control of Full Field Digital Mammography Units Melissa C. Martin, M.S., FACMP, FACR, FAAPM Melissa@TherapyPhysics.com 310-612-8127 ACMP Annual Meeting Virginia Beach, VA May 2, 2009 History of

More information

ISO 3664 INTERNATIONAL STANDARD. Graphic technology and photography Viewing conditions

ISO 3664 INTERNATIONAL STANDARD. Graphic technology and photography Viewing conditions INTERNATIONAL STANDARD ISO 3664 Third edition 2009-04-15 Graphic technology and photography Viewing conditions Technologie graphique et photographie Conditions d'examen visuel Reference number ISO 3664:2009(E)

More information

For Your Diagnostic Confidence. Product Brochure

For Your Diagnostic Confidence. Product Brochure For Your Diagnostic Confidence Product Brochure ABOUT US KOSTEC specializes in designing, developing and manufacturing industrial image display devices, particularly medical imaging display monitors. Since

More information

PSSA Calibration and Colour Management

PSSA Calibration and Colour Management PSSA Calibration and Colour Management Background A large portion of PSSA activities is focused around the display of images. PSSA understand the importance of consistency when displaying images. Even

More information

HP DREAMCOLOR USE CASES

HP DREAMCOLOR USE CASES HP DREAMCOLOR USE CASES ACHIEVING ACCURATE COLORS ACROSS DIVERSE WORKFLOWS TECHNOLOGy SPOTLIGHT HP PROFESSIONAL INNOVATIONS FOR business NOTEbOOKS WHEN COLOR IS EVERyTHING For creative professionals, consistent

More information

Effects of Pixel Density On Softcopy Image Interpretability

Effects of Pixel Density On Softcopy Image Interpretability Effects of Pixel Density On Softcopy Image Interpretability Jon Leachtenauer ERIM-International, Arlington, Virginia Andrew S. Biache and Geoff Garney Autometric Inc., Springfield, Viriginia Abstract Softcopy

More information

Small Animal Radiographic Techniques and Positioning COPYRIGHTED MATERIAL

Small Animal Radiographic Techniques and Positioning COPYRIGHTED MATERIAL Small Animal Radiographic Techniques and Positioning COPYRIGHTED MATERIAL Section 1 Theory and Equipment 1 Introduction to Digital Imaging Small animal radiography has changed dramatically in the past

More information

Working with Wide Color Gamut and High Dynamic Range in Final Cut Pro X. New Workflows for Editing

Working with Wide Color Gamut and High Dynamic Range in Final Cut Pro X. New Workflows for Editing Working with Wide Color Gamut and High Dynamic Range in Final Cut Pro X New Workflows for Editing White Paper Contents Introduction 3 Background 4 Sources of Wide-Gamut HDR Video 6 Wide-Gamut HDR in Final

More information

Improving Color Image Quality in Medical Photography

Improving Color Image Quality in Medical Photography White Paper #46 Level: Intermediate Date: April 2017 Improving Color Image Quality in Medical Photography Introduction Medical images have played an important role in the development of modern medicine.

More information

Digital Imaging and Communications in Medicine (DICOM) Part 14: Grayscale Standard Display Function

Digital Imaging and Communications in Medicine (DICOM) Part 14: Grayscale Standard Display Function PS 3.14-2006 Digital Imaging and Communications in Medicine (DICOM) Part 14: Grayscale Standard Display Function Published by National Electrical Manufacturers Association 1300 N. 17th Street Rosslyn,

More information

The Missed Opportunity of Soft Proofing

The Missed Opportunity of Soft Proofing #COLOR19 The Missed Opportunity of Soft Proofing Don Hutcheson HutchColor.com Based on slides by Dave Dezzutti & Joe Marin, PIA Content Soft proofing basics Basic LCD/LED display concepts Why should you

More information

Images and Displays. Lecture Steve Marschner 1

Images and Displays. Lecture Steve Marschner 1 Images and Displays Lecture 2 2008 Steve Marschner 1 Introduction Computer graphics: The study of creating, manipulating, and using visual images in the computer. What is an image? A photographic print?

More information

Optimizing throughput with Machine Vision Lighting. Whitepaper

Optimizing throughput with Machine Vision Lighting. Whitepaper Optimizing throughput with Machine Vision Lighting Whitepaper Optimizing throughput with Machine Vision Lighting Within machine vision systems, inappropriate or poor quality lighting can often result in

More information

19 Setting Up Your Monitor for Color Management

19 Setting Up Your Monitor for Color Management 19 Setting Up Your Monitor for Color Management The most basic requirement for color management is to calibrate your monitor and create an ICC profile for it. Applications that support color management

More information

Color Management. Photographers Thomas Zuber.

Color Management. Photographers Thomas Zuber. Color Management For Color and Black & White Photographers 2010 Thomas Zuber Agenda Scope of Presentation Three characteristics of light What is/is not Color Management Color Management for Cameras Review:

More information

Understand brightness, intensity, eye characteristics, and gamma correction, halftone technology, Understand general usage of color

Understand brightness, intensity, eye characteristics, and gamma correction, halftone technology, Understand general usage of color Understand brightness, intensity, eye characteristics, and gamma correction, halftone technology, Understand general usage of color 1 ACHROMATIC LIGHT (Grayscale) Quantity of light physics sense of energy

More information

k m = 683 lumens/watt Display QC QC Measures for Color Displays joules watts lumens SAM Imaging Education Course

k m = 683 lumens/watt Display QC QC Measures for Color Displays joules watts lumens SAM Imaging Education Course Display QC SAM Imaging Education Course MO-GH-205-3 DisplayQuality Test Image TG18-pQC QC Measures for Color Displays a) Color coordinate systems (5) b) Devices for measuring color coordinates (11) c)

More information

IHE Radiology Technical Framework Supplement. Stereotactic Mammography Image (SMI) Trial Implementation

IHE Radiology Technical Framework Supplement. Stereotactic Mammography Image (SMI) Trial Implementation Integrating the Healthcare Enterprise 5 IHE Radiology Technical Framework Supplement 10 Stereotactic Mammography Image (SMI) 15 Trial Implementation 20 25 Date: June 11, 2013 Author: IHE Radiology Technical

More information

Display profiling and Color Management

Display profiling and Color Management Display profiling and Color Management Andrew Rodney aka The Digital Dog www.digitaldog.net andrew@digitaldog.net Email me (andrew@digitaldog.net) if you need this presentation in PDF form. Most of the

More information

Perceptual Rendering Intent Use Case Issues

Perceptual Rendering Intent Use Case Issues White Paper #2 Level: Advanced Date: Jan 2005 Perceptual Rendering Intent Use Case Issues The perceptual rendering intent is used when a pleasing pictorial color output is desired. [A colorimetric rendering

More information

A Module for Visualisation and Analysis of Digital Images in DICOM File Format

A Module for Visualisation and Analysis of Digital Images in DICOM File Format A Module for Visualisation and Analysis of Digital Images in DICOM File Format Rumen Rusev Abstract: This paper deals with design and realisation of software module for visualisation and analysis of digital

More information

Graphic technology Prepress data exchange Preparation and visualization of RGB images to be used in RGB-based graphics arts workflows

Graphic technology Prepress data exchange Preparation and visualization of RGB images to be used in RGB-based graphics arts workflows Provläsningsexemplar / Preview INTERNATIONAL STANDARD ISO 16760 First edition 2014-12-15 Graphic technology Prepress data exchange Preparation and visualization of RGB images to be used in RGB-based graphics

More information

MEDICAL DISPLAYS. NEC 90 Series. NEC MD Series

MEDICAL DISPLAYS. NEC 90 Series. NEC MD Series MEDICAL DISPLAYS NEC MD Series Are you really sure your display is as accurate as your diagnosis? In medical environments such as radiology and emergency rooms, the right equipment and technology can help

More information

Yagi Digital Microscope Calibration

Yagi Digital Microscope Calibration Yagi Digital Microscope Calibration Method summary, assessment and suggestions for improvement W Craig Revie, International Color Consortium Introduction In the area of pathology, a type of digital microscope

More information

SIMPLE. VERSATILE. VERSATILE PRINTING ON FILM OR PAPER. DRYVIEW CHROMA Imaging System

SIMPLE. VERSATILE. VERSATILE PRINTING ON FILM OR PAPER. DRYVIEW CHROMA Imaging System DRYVIEW CHROMA Imaging System VERSATILE PRINTING ON FILM OR PAPER One of the most trusted names in medical printing now offers a highly versatile printer that empowers physicians and other healthcare professionals

More information

The Science Seeing of process Digital Media. The Science of Digital Media Introduction

The Science Seeing of process Digital Media. The Science of Digital Media Introduction The Human Science eye of and Digital Displays Media Human Visual System Eye Perception of colour types terminology Human Visual System Eye Brains Camera and HVS HVS and displays Introduction 2 The Science

More information

COLORMUNKI DISPLAY & i1display PRO

COLORMUNKI DISPLAY & i1display PRO Now supports Mobile Calibration with ColorTRUE app. COLORMUNKI DISPLAY & i1display PRO Display and Projector Calibration Solutions for all Types of Color Perfectionists Color Perfectionists Unite! Is your

More information

Imaging Informatics. Standards Support for Color in Medical Imaging. SAM Imaging Education Course. Photometric Standards Colorimetric Standards

Imaging Informatics. Standards Support for Color in Medical Imaging. SAM Imaging Education Course. Photometric Standards Colorimetric Standards Imaging Informatics SAM Imaging Education Course AAPM Standards Support for Color in Medical Imaging Photometric Standards Colorimetric Standards Michael J. Flynn, PhD Henry Ford Health System, Detroit,

More information

ISO 3664 INTERNATIONAL STANDARD. Graphic technology and photography Viewing conditions

ISO 3664 INTERNATIONAL STANDARD. Graphic technology and photography Viewing conditions INTERNATIONAL STANDARD ISO 3664 Third edition 2009-04-15 Graphic technology and photography Viewing conditions Technologie graphique et photographie Conditions d'examen visuel Reference number ISO 3664:2009(E)

More information

SECTION I - CHAPTER 1 DIGITAL RADIOGRAPHY: AN OVERVIEW OF THE TEXT. Exam Content Specifications 8/22/2012 RADT 3463 COMPUTERIZED IMAGING

SECTION I - CHAPTER 1 DIGITAL RADIOGRAPHY: AN OVERVIEW OF THE TEXT. Exam Content Specifications 8/22/2012 RADT 3463 COMPUTERIZED IMAGING RADT 3463 - COMPUTERIZED IMAGING Section I: Chapter 1 RADT 3463 Computerized Imaging 1 SECTION I - CHAPTER 1 DIGITAL RADIOGRAPHY: AN OVERVIEW OF THE TEXT RADT 3463 COMPUTERIZED IMAGING Section I: Chapter

More information

srgb: A Standard for Color Management

srgb: A Standard for Color Management srgb: A Standard for Color Management Introduction Over the years, magazines, newspapers, television, computers and, now, the Internet have all made the transition from black and white to color. With the

More information

SECTION I - CHAPTER 2 DIGITAL IMAGING PROCESSING CONCEPTS

SECTION I - CHAPTER 2 DIGITAL IMAGING PROCESSING CONCEPTS RADT 3463 - COMPUTERIZED IMAGING Section I: Chapter 2 RADT 3463 Computerized Imaging 1 SECTION I - CHAPTER 2 DIGITAL IMAGING PROCESSING CONCEPTS RADT 3463 COMPUTERIZED IMAGING Section I: Chapter 2 RADT

More information

Remote Director. Sony 23 SDM-P232W. Collaborative Soft Proofing using the I. MANUFACTURER INTRODUCTION. SWOP Application Data Sheet

Remote Director. Sony 23 SDM-P232W. Collaborative Soft Proofing using the I. MANUFACTURER INTRODUCTION. SWOP Application Data Sheet SWOP Application Data Sheet Remote Director Collaborative Soft Proofing using the Sony 23 SDM-P232W The SWOP Review Committee has approved the use of off-press proofs as input material to publications.

More information

Considerations of HDR Program Origination

Considerations of HDR Program Origination SMPTE Bits by the Bay Wednesday May 23rd, 2018 Considerations of HDR Program Origination L. Thorpe Canon USA Inc Canon U.S.A., Inc. 1 Agenda Terminology Human Visual System Basis of HDR Camera Dynamic

More information

China Resources Wandong Medical Equipment Co., Ltd. High Frequency 50kW Digital RF System - HF51-5

China Resources Wandong Medical Equipment Co., Ltd. High Frequency 50kW Digital RF System - HF51-5 China Resources Wandong Medical Equipment Co., Ltd. High Frequency 50kW Digital RF System - HF51-5 #3, No.9, Jiuxianqiaodong Road, Chaoyang District, Beijing 100015, P.R. China E-mail: international@wandong.com.cn

More information

Test Equipment for Radiology and CT Quality Control Contents

Test Equipment for Radiology and CT Quality Control Contents Test Equipment for Radiology and CT Quality Control Contents Quality Control Testing...2 Photometers for Digital Clinical Display QC...3 Primary Workstations...3 Secondary Workstations...3 Testing of workstations...3

More information

Spyder 5ELITE Professional photographers, studios, and calibration perfectionists seeking ultimate control of the color workflow.

Spyder 5ELITE Professional photographers, studios, and calibration perfectionists seeking ultimate control of the color workflow. Spyder5 User Guide Spyder 5ELITE Professional photographers, studios, and calibration perfectionists seeking ultimate control of the color workflow. Next Generation Spyder5 colorimeter with dual purpose

More information

Nathan Childress, Ph.D., DABR

Nathan Childress, Ph.D., DABR Nathan Childress, Ph.D., DABR Introduction TG-142 is a comprehensive QA protocol Covers nearly every aspect of machine and safety QA Recommends quantitative results Recommends high testing frequencies

More information

A comparison between medical-grade liquid crystal display (LCD) and ipad color imaging

A comparison between medical-grade liquid crystal display (LCD) and ipad color imaging A comparison between medical-grade liquid crystal display (LCD) and ipad color imaging Poster No.: C-1377 Congress: ECR 2014 Type: Scientific Exhibit Authors: K. Yoshimura 1, K. Shibata 2, T. Nihashi 1,

More information

WORKING WITH COLOR Monitor Placement Place the monitor at roughly right angles to a window. Place the monitor at least several feet from any window

WORKING WITH COLOR Monitor Placement Place the monitor at roughly right angles to a window. Place the monitor at least several feet from any window WORKING WITH COLOR In order to work consistently with color printing, you need to calibrate both your monitor and your printer. The basic steps for doing so are listed below. This is really a minimum approach;

More information

Appearance Match between Soft Copy and Hard Copy under Mixed Chromatic Adaptation

Appearance Match between Soft Copy and Hard Copy under Mixed Chromatic Adaptation Appearance Match between Soft Copy and Hard Copy under Mixed Chromatic Adaptation Naoya KATOH Research Center, Sony Corporation, Tokyo, Japan Abstract Human visual system is partially adapted to the CRT

More information

Image Processing Lecture 4

Image Processing Lecture 4 Image Enhancement Image enhancement aims to process an image so that the output image is more suitable than the original. It is used to solve some computer imaging problems, or to improve image quality.

More information

PACS Fundamentals. By: Eng. Valentino T. Mvanga Ministry of Health and Social Welfare Tanzania

PACS Fundamentals. By: Eng. Valentino T. Mvanga Ministry of Health and Social Welfare Tanzania PACS Fundamentals By: Eng. Valentino T. Mvanga Ministry of Health and Social Welfare Tanzania 1 Learning Goals To Understand the importance of PACS To Understand PACS infrastructure requirement Introduction

More information

Slide 1. Slide 2. Slide 3 ACR CT Accreditation. Multi-Slice CT Artifacts and Quality Control. What are the rules or recommendations for CT QC?

Slide 1. Slide 2. Slide 3 ACR CT Accreditation. Multi-Slice CT Artifacts and Quality Control. What are the rules or recommendations for CT QC? Slide 1 Multi-Slice CT Artifacts and Quality Control Dianna Cody, Ph.D. Chief, Radiologic Physics UT MD Anderson Cancer Center Houston, TX Slide 2 What are the rules or recommendations for CT QC? AAPM

More information

H34: Putting Numbers to Colour: srgb

H34: Putting Numbers to Colour: srgb page 1 of 5 H34: Putting Numbers to Colour: srgb James H Nobbs Colour4Free.org Introduction The challenge of publishing multicoloured images is to capture a scene and then to display or to print the image

More information

Preliminary Assessment of High Dynamic Range Displays for Pathology Detection Tasks. CIS/Kodak New Collaborative Proposal

Preliminary Assessment of High Dynamic Range Displays for Pathology Detection Tasks. CIS/Kodak New Collaborative Proposal Preliminary Assessment of High Dynamic Range Displays for Pathology Detection Tasks CIS/Kodak New Collaborative Proposal CO-PI: Karl G. Baum, Center for Imaging Science, Post Doctoral Researcher CO-PI:

More information

Appendix A. Practical Color Management

Appendix A. Practical Color Management Appendix A. Practical Color Management Printing a photograph can be time consuming, expensive, and frustrating without color management. I would typically print a photograph many times with different printer

More information

DIGITAL IMAGING FOUNDATIONS

DIGITAL IMAGING FOUNDATIONS CHAPTER DIGITAL IMAGING FOUNDATIONS Photography is, and always has been, a blend of art and science. The technology has continually changed and evolved over the centuries but the goal of photographers

More information

Color Management - Part II Implementing Color Management Michael J. Glagola

Color Management - Part II Implementing Color Management Michael J. Glagola 6-9-07 Michael J. Glagola 2007 2 Color Management - Part II Implementing Color Management Michael J. Glagola mglagola@cox.net 703-830-6860 6-9-07 Michael J. Glagola 2007 3 Session Goals To provide an practical

More information

Conformance Statement for DICOM Viewer

Conformance Statement for DICOM Viewer MedDream Conformance Statement for DICOM Viewer (version 5.1) 2015, Softneta UAB, Kaunas All rights reserved in the event of granting of patents or registration as a utility patent. All names of companies

More information

DIGITAL IMAGING Recognise the importance of quality assurance

DIGITAL IMAGING Recognise the importance of quality assurance DIGITAL IMAGING Recognise the importance of quality assurance There are two types of digital image receptor both of which capture a two dimensional image of the three dimensional patient. These are Computed

More information

Acceptance Testing and Annual Physics Survey Recommendations for Gamma Camera, SPECT, and SPECT/CT Systems

Acceptance Testing and Annual Physics Survey Recommendations for Gamma Camera, SPECT, and SPECT/CT Systems AAPM REPORT NO. 177 Acceptance Testing and Annual Physics Survey Recommendations for Gamma Camera, SPECT, and SPECT/CT Systems The Report of AAPM Task Group 177 February 2019 DISCLAIMER: This publication

More information

Images and Displays. CS4620 Lecture 15

Images and Displays. CS4620 Lecture 15 Images and Displays CS4620 Lecture 15 2014 Steve Marschner 1 What is an image? A photographic print A photographic negative? This projection screen Some numbers in RAM? 2014 Steve Marschner 2 An image

More information

Features and Weaknesses of Phantoms for CR/DR System Testing

Features and Weaknesses of Phantoms for CR/DR System Testing Physics testing of image detectors Parameters to test Features and Weaknesses of Phantoms for CR/DR System Testing Spatial resolution Contrast resolution Uniformity/geometric distortion Dose response/signal

More information

Digital Radiography using High Dynamic Range Technique

Digital Radiography using High Dynamic Range Technique Digital Radiography using High Dynamic Range Technique DAN CIURESCU 1, SORIN BARABAS 2, LIVIA SANGEORZAN 3, LIGIA NEICA 1 1 Department of Medicine, 2 Department of Materials Science, 3 Department of Computer

More information

Color Management. R. Mac Holbert

Color Management. R. Mac Holbert Color Management R. Mac Holbert Color Management Is Important! It s Relatively Inexpensive! It s Not Difficult To Understand! What is Color Management? Color Management is the name given to processes and

More information

Colour Management Workflow

Colour Management Workflow Colour Management Workflow The Eye as a Sensor The eye has three types of receptor called 'cones' that can pick up blue (S), green (M) and red (L) wavelengths. The sensitivity overlaps slightly enabling

More information

The Architecture of Medical Imaging

The Architecture of Medical Imaging University of Kansas Architecture 731 Systems and Components of Healthcare Facilities F. Zilm The Architecture of Medical Imaging Designing Healthcare Facilities for Advanced Radiologic Diagnostic and

More information

153 Shimokashiwano, Hakusan, Ishikawa Japan Phone Fax:

153 Shimokashiwano, Hakusan, Ishikawa Japan Phone Fax: A lesson on color management with 153 Shimokashiwano, Hakusan, Ishikawa 924-8566 Japan Phone +81-76-277-6792 Fax: +81-76-277-6793 www.eizo.com 2011 Eizo Nanao Corporation All product names are trademarks

More information

Medical Imaging Working Group

Medical Imaging Working Group Medical Imaging Working Group FOGRA Einsteinring 1a 85609 Ascheim Munich, Germany 26 February 2018 Craig Revie, MIWG chair, opened the meeting at 08:45 and introduced the agenda as follows: 1. Introductions

More information

How To Set Up & Calibrate Your EIZO Monitor

How To Set Up & Calibrate Your EIZO Monitor How To Set Up & Calibrate Your EIZO Monitor - A PUBLICATION OF EIZO APAC- 1 INTRODUCTION Congratulations! You ve invested in an EIZO monitor. You ve gone through the exciting unboxing process. Now what?

More information

Color Reproduction. Chapter 6

Color Reproduction. Chapter 6 Chapter 6 Color Reproduction Take a digital camera and click a picture of a scene. This is the color reproduction of the original scene. The success of a color reproduction lies in how close the reproduced

More information

Learning Objectives. Outline. Getting Started with CR. Converting the Radiology Department from Film-Screen to Digital: Making the Transition

Learning Objectives. Outline. Getting Started with CR. Converting the Radiology Department from Film-Screen to Digital: Making the Transition Converting the Radiology Department from Film-Screen to Digital: Making the Transition S. Jeff Shepard, MS, DABR University of Texas M. D. Anderson Cancer Center Houston, Texas jshepard@di.mdacc.tmc.edu

More information

Hello, welcome to the video lecture series on Digital Image Processing.

Hello, welcome to the video lecture series on Digital Image Processing. Digital Image Processing. Professor P. K. Biswas. Department of Electronics and Electrical Communication Engineering. Indian Institute of Technology, Kharagpur. Lecture-33. Contrast Stretching Operation.

More information

Nuclear Associates

Nuclear Associates Nuclear Associates 07-647 R/F QC Phantom Operators Manual March 2005 Manual No. 07-647-1 Rev. 2 2004, 2005 Fluke Corporation, All rights reserved. All product names are trademarks of their respective companies

More information

Color Management and Your Workflow. monaco

Color Management and Your Workflow. monaco Color Management and Your Workflow Problem in Matching Colors > THE RESULTS Wasted Time and Money Frustration Color Managed > THE RESULTS Save Time Money and Paper Get Great Prints Every Time The Cost

More information

Provläsningsexemplar / Preview

Provläsningsexemplar / Preview SVENSK STANDARD Fastställd 2004-08-13 Utgåva 1 Fotografi och grafisk teknik Färgkodning med utökad tonskala för digital bildlagring, manipulation och utbyte Del 1: Arkitektur och krav (ISO 22028-1:2004,

More information

FT-S5500 OUTPUT PROFILE FOR RGB OUTPUT MODE

FT-S5500 OUTPUT PROFILE FOR RGB OUTPUT MODE FT-S5500 OUTPUT PROFILE FOR RGB OUTPUT MODE Nov 2004 Recently, the digital camera has been spread and the color management system has evolved. And in additionally, the scanned image data is used in many

More information

Aspire HD. Program Manual. 2nd Edition - October 2012

Aspire HD. Program Manual. 2nd Edition - October 2012 Quality Control 1 Aspire HD Quality Control Program Manual 2nd Edition - October 2012 Overview Installation of FDR Mammography QC Program Weekly Test 2 3 4 Quarterly Test 5 Semi-annual Test 6 Annual Test

More information

CR35-SP+ Computed Radiography (CR) System. Innovations in Digital Imaging

CR35-SP+ Computed Radiography (CR) System. Innovations in Digital Imaging CR35-SP+ Computed Radiography (CR) System Innovations in Digital Imaging Solution for Quantum Medical Imaging CR35-SP+ Package An Agfa Healthcare Solution for Quantum Medical Imaging The CR35-SP+ Package

More information

Yearbook Color Management. Matthew Bernius. Rochester Institute of Technology School of Print Media

Yearbook Color Management. Matthew Bernius. Rochester Institute of Technology School of Print Media Yearbook Color Management Matthew Bernius Rochester Institute of Technology School of Print Media Topic Overview Color in Theory Color in Production Color Management Image Editing (best practices) 1 Color

More information

High Dynamic Range Displays

High Dynamic Range Displays High Dynamic Range Displays Dave Schnuelle Senior Director, Image Technology Dolby Laboratories The Demise of the CRT What was good: Large viewing angle High contrast Consistent EO transfer function Good

More information

UP-D74XRD. DICOM Digital Imager.

UP-D74XRD. DICOM Digital Imager. DICOM Digital Imager www.sonybiz.net/healthcare Supports both 8x10inch Blue Thermal Film and Thermal Printing Paper. Delivers photo-quality fi lm and paper prints with 512-step grayscale. Ensures minimal

More information

While digital techniques have the potential to reduce patient doses, they also have the potential to significantly increase them.

While digital techniques have the potential to reduce patient doses, they also have the potential to significantly increase them. In press 2004 1 2 Guest Editorial (F. Mettler, H. Ringertz and E. Vano) Guest Editorial (F. Mettler, H. Ringertz and E. Vano) Digital radiology An appropriate analogy that is easy for most people to understand