Effect of pressure, temperature and humidity in air on photon fluence and air kerma values at low photon energies

Size: px
Start display at page:

Download "Effect of pressure, temperature and humidity in air on photon fluence and air kerma values at low photon energies"

Transcription

1 ARTICLE IN PRESS Radiation Physics and Chemistry 68 (2003) Effect of pressure, temperature and humidity in air on photon fluence and air kerma values at low photon energies M. Assiamah, D. Mavunda, T.L. Nam, R.J. Keddy* Health Physics Service, Schonland Research Institute for Nuclear Sciences, University of the Witwatersrand, Private Bag 3, Wits 2050, Johannesburg, South Africa Received 3 April 2003; accepted 19 May 2003 Abstract An investigation into the effect of pressure, temperature and humidity in air on photon fluence at a typical mammography, low bremsstrahlung energy (25 kvp), has been carried out. Pressure values corrected for humidity at varying temperatures were employed to determine the density of moist air. Using the corrected moist air densities, the X-ray photon fluence remaining after air attenuation has been computed assuming an X-ray focal spot-detector distance of 65 cm. Bremsstrahlung spectral distributions and the effects of pressure, temperature and humidity on the photon fluence from molybdenum and tungsten targets are illustrated. Comparative results suggest that such effects could be significant and need to be considered when calculating exposure levels from low-energy photons. The investigation showed that air kerma values from an X-ray spectrum that has significant lower-energy components is likely to be more sensitive to changes in pressure, temperature and humidity than the air kerma from an X-ray spectrum with lower-energy components less pronounced. Non-negligible air kerma values are involved. r 2003 Elsevier Ltd. All rights reserved. Keywords: X-ray photon fluence; Attenuation; Mammography; Bremsstrahlung; Air kerma 1. Introduction Mammography X-ray examinations are typically performed at a focal-spot detector distance (FSD) of cm. In the calculation of the photon fluence reaching the detector for dose or exposure measurements at mammography energies, the attenuation due to air for the specific FSD must be taken into consideration. In many such calculations, the air is normally assumed to be dry; hence it is common to find the density of dry air at sea level being applied in such computations. However, sea level atmospheric pressures do not apply with many measurements being conducted at centres at significant altitudes. Air normally contains water vapour and is classified as dry or humid based on *Corresponding author. Tel.: ; fax: address: keddy@src.wits.ac.za (R.J. Keddy). its water vapour content. The humidity is specified using partial pressures of water in the air. The relative humidity of the water content in the air is not constant, varying from a minimum of zero to a maximum of 100%, determined by the saturated vapour pressure of water at the given air temperature. The density of air is therefore affected by changes in environmental conditions, i.e. temperature, pressure and moisture or humidity. This study sets out to determine the effect of these physical properties on the photon fluence and air kerma values of mammography bremsstrahlung beams. 2. Methodology The photon fluence (photons/cm 2 ) used for this study was obtained from a computer generated X-ray spectrum code namely: molybdenum anode spectra model using interpolation polynomial (MASMIP) and X/03/$ - see front matter r 2003 Elsevier Ltd. All rights reserved. doi: /s x(03)

2 708 ARTICLE IN PRESS M. Assiamah et al. / Radiation Physics and Chemistry 68 (2003) tungsten anode spectra model using interpolation polynomial (TASMIP) developed by Boone et al. (1997). The X-ray spectrum for a 25 kvp tube potential was generated from both the MASMIP and TASMIP models. A nominal filtration of 0.03 mm molybdenum was applied to the molybdenum target spectrum, whilst a 0.05 mm rhodium filtration was applied to the tungsten target spectrum. The spectral shaping (filters) materials applied are routinely used on clinical mammography systems. A focal spot-detector distance of 65 cm (an FSD commonly used at clinical mammography settings) was assumed for the calculation. For computational purposes the bremsstrahlung spectrum was divided into 0.5 kev energy bins. The photon fluence remaining in each bin after air attenuation was calculated by applying, to each bin, the energy-dependent Lambert Beer s law equation f ¼ f 0 expð meþ; ð1þ where f is the photon fluence (photon/cm 2 ) in each bin after attenuation by air of thickness e (g/cm 2 ) with mass attenuation coefficients m (cm 2 /g) and f 0 is the initial photon fluence before attenuation (Archer and Wagner, 1982; Attix, 1986; Wolbarst, 1993). The thickness e is given by the product of the density r in g/cm 3 and the focal spot-detector distance x in cm. The density of air used in Eq. (1) has been corrected for temperature, pressure and humidity. The corrected air densities used for this investigation are those given in the CRC Handbookof Chemistry and Physics ( ) for a temperature range of 5 35 C and corrected pressure range of mmhg. The densities for moist air at temperatures 40 C, 45 C and 50 C, not given in the reference for the specified corrected pressures, were calculated using Eq. (2) and Table II of section F of Weast et al. ( ). The density r T;P;H of moist air at absolute temperature T, barometric pressure P in mmhg, is given by r T;P;H ¼ 1: :13 T P 0:3783e 760 ; ð2þ where is the density of dry air in g/l at standard temperature and pressure (s.t.p.) i.e. 0 C/ K and 760 mmhg/101.3 kpa, respectively. The parameter e appearing in Eq. (2) is the vapour pressure of the moisture in the air in millimetres. Employing the dewpoint method and the vapour pressure of water at different dew points, the humidity H of the air has been accounted for (Marion and Hornyak, 1982). The corrected pressure values at various temperatures were then employed to determine the density of moist air for the calculated relative humidity. Relative humidities of 0 100% at intervals of 10% were used in the calculations. Using the photon fluence f iðcorrþ ; that which remains after 65 cm air attenuation, exposures X i ðeþ T;P;H in roentgen (R) at absolute temperature T, barometric pressure P and humidity H was calculated employing X i ðeþ T;P;H ¼ 1: f iðcorrþ E i ðm en rþ airi r T;P;H : ð3þ The subscript i refers to the ith photon energy interval E i (kev), m en, is the mass energy-absorption coefficient (cm 2 /g) of air at s.t.p., r and r T;P;H are the densities of air at s.t.p. and at absolute temperature T, barometric pressure P in mmhg, and relative humidity H, respectively (Cember, 1996). From the exposure X i (E) values, the air kerma K i (E) values in cgy were calculated using the exposure and air kerma relationship given by Wolbarst (Wolbarst, 1993). K i ðeþ ¼0:873X i ðeþ: The mass attenuation and mass energy-absorption coefficients data used for this study are those compiled by Hubbell and Seltzer (1996) and reworked into an extremely user-friendly windows program, WinXCom, (Gerward et al., 2001). To obtain data covering the energy of interest, a fitting algorithm was developed to obtain mass energy-absorption coefficient values with negligible deviation from the original published data (Assiamah et al., 2003). Both the mass attenuation and mass energy coefficients data were corrected to s.t.p. In analysing the results, the photon fluence data at temperature 0 C, 760 mmhg corrected pressure and 0% relative humidity was chosen as a reference data point. The results of the study are presented in tabular and graphical forms. 3. Results and discussion The effects of pressure on the photon fluence are illustrated in Figs. 1 and 2 for the molybdenum target spectrum and in Figs. 3 and 4 for the tungsten target. The temperature effects on the photon fluence are presented in Figs. 5 and 6 for the molybdenum target and in Figs. 7 and 8 for the tungsten target. The effects of humidity on the photon fluence are illustrated in Figs. 9 and 10 for the molybdenum target and in Figs. 11 and 12 for the tungsten target. A comparison of a typical photon fluence (photons/ cm 2 ) spectrum from molybdenum and tungsten targets at tube voltage of 25 kvp, 760 mmhg, 0 C and 0% relative humidity (reference spectrum) with a spectrum at the same tube voltage but corrected for air density at a pressure of 600 mmhg, temperature 35 C and 50% relative humidity are shown in Figs. 13, 14 and 15, respectively. Table 1 presents a comparison of the contributions of three arbitrary energy segments to the overall photon fluence from the reference and corrected molybdenum spectra. The comparison of the photon ð4þ

3 ARTICLE IN PRESS M. Assiamah et al. / Radiation Physics and Chemistry 68 (2003) Fig. 1. Deviation of the photon fluence due to pressure variation from the photon fluence calculated at 0 C, 760 mmhg and 0% relative humidity for a molybdenum target at 25 kvp. Fig. 2. Overall deviation (%) due to pressure on the integrated photon fluence from the photon fluence calculated at 0 C, 760 mmhg and 0% relative humidity for a molybdenum target spectrum at 25 kvp.

4 710 ARTICLE IN PRESS M. Assiamah et al. / Radiation Physics and Chemistry 68 (2003) Fig. 3. Deviation of the photon fluence due to pressure variation from the photon fluence calculated at 0 C, 760 mmhg and 0% relative humidity for a tungsten target at 25 kvp. Fig. 4. Overall deviation (%) due to pressure on the integrated photon fluence from the photon fluence calculated at 0 C, 760 mmhg and 0% relative humidity for a tungsten target spectrum at 25 kvp.

5 ARTICLE IN PRESS M. Assiamah et al. / Radiation Physics and Chemistry 68 (2003) Fig. 5. Deviation of the photon fluence due to temperature variation from the photon fluence calculated at 0 C, 760 mmhg and 0% relative humidity for a molybdenum target at 25 kvp. Fig. 6. Overall deviation (%) due to temperature on the integrated photon fluence from the photon fluence calculated at 0 C, 760 mmhg at 0% relative humidity for a molybdenum target at 25 kvp.

6 712 ARTICLE IN PRESS M. Assiamah et al. / Radiation Physics and Chemistry 68 (2003) Fig. 7. Deviation of the photon fluence due to temperature variation from the photon fluence calculated at 0 C, 760 mmhg and 0% relative humidity for a tungsten target spectrum at 25 kvp. Fig 8. Overall deviation (%) due to temperature on the integrated photon fluence from the photon fluence calculated at 0 C, 760 mmhg and 0% relative humidity for a tungsten target spectrum at 25 kvp.

7 ARTICLE IN PRESS M. Assiamah et al. / Radiation Physics and Chemistry 68 (2003) Fig. 9. Deviation of photon fluence due to a humidity variation from the photon fluence calculated at 0 C and 760 mmhg for a molybdenum target at 25 kvp. Fig. 10. Overall deviation (%) due to humidity on the integrated photon fluence from the photon fluence calculated at 0 C and 760 mmhg for a molybdenum target at 25 kvp.

8 714 ARTICLE IN PRESS M. Assiamah et al. / Radiation Physics and Chemistry 68 (2003) Fig. 11. Deviation of photon fluence due to a humidity variation from the photon fluence calculated at 0 C and 760 mmhg for a tungsten target at 25 kvp. Fig. 12. Overall deviation (%) due to humidity on the integrated photon fluence from the photon fluence calculated at 0 C, 760 mmhg at 0% relative humidity for a tungsten target at 25 kvp.

9 ARTICLE IN PRESS M. Assiamah et al. / Radiation Physics and Chemistry 68 (2003) Fig. 13. A comparison of the photon fluence spectrum from a molybdenum target at 25 kvp. (A) Spectrum at 760 mmhg, 0 C and 0% relative humidity. (B) Spectrum corrected for temperature (35 C), pressure (600 mmhg) and relative humidity (50%). Fig. 14. A comparison of the air kerma spectrum from a molybdenum target at 25 kvp. (A) Spectrum at 760 mmhg, 0 C and 0% relative humidity. (B) Spectrum corrected for temperature (35 C), pressure (600 mmhg) and relative humidity (50%).

10 716 ARTICLE IN PRESS M. Assiamah et al. / Radiation Physics and Chemistry 68 (2003) Fig. 15. A comparison of the photon fluence spectrum from a tungsten target at 25 kvp. (A) Spectrum at 760 mmhg, 0 C and 0% relative humidity. (B) Spectrum corrected for temperature (35 C), pressure (600 mmhg) and 50% relative humidity. Table 1 Comparison of contributions of three arbitrary energy segments to the overall photon fluence (photons/cm 2 ) from a molybdenum spectrum at 25 kvp, 0 C, 760 mmhg and 0% relative humidity with the molybdenum target spectrum at the same tube voltage and corrected for temperature, pressure and humidity Contribution to photon fluence by energy range 4 8 kev Contribution to photon fluence by energy range 8 15 kev Contribution to photon fluence by energy range kev Photon fluence at Total contribution 4.73E E E mmhg, 0 C, 0% RH ~ Percent contribution (%) Photon fluence at Total contribution 6.12E E E mmhg 35 C, 50% RH Percent contribution (%) % difference in photon fluence due to P, T, H Y correction The overall percentage difference in the photon fluence due to the combined effects of temperature, pressure and humidity is also shown. Overall difference in photon fluence due to P, T and H correction=3.98%. ~ RH is relative humidity. Y P, T and H are the pressure, temperature and humidity, respectively. fluence from the reference and corrected tungsten spectra are similarly shown in Table 3. A comparison of the contributions of three arbitrary energy segments to the overall air kerma values for both the reference and corrected molybdenum and tungsten spectra are presented in Tables 2 and 4. For purposes of illustrating the weights of contributions from various regions of the spectrum, three arbitrary energy segments

11 ARTICLE IN PRESS M. Assiamah et al. / Radiation Physics and Chemistry 68 (2003) Table 2 Comparison of contributions of three arbitrary energy segments to the overall air kerma in cgy from a molybdenum spectrum at 25 kvp, 0 C, 760 mmhg and 0% relative humidity with the molybdenum target spectrum at the same tube voltage and corrected for temperature, pressure and humidity Contribution to air kerma by energy range 4 8 kev Contribution to air kerma by energy range 8 15 kev Contribution to air kerma by energy range kev Air kerma (cgy) at Total contribution mmhg, Percent contribution (%) C, 0% RH Air kerma (cgy) at Total contribution mmhg Percent contribution (%) C, 40% RH % difference in air kerma due to P, T, H correction The overall percentage due to the combined effects of temperature, pressure and humidity is also shown. Overall difference in air kerma due to P, T and H correction=51.85%. Table 3 Comparison of contributions of three arbitrary energy segments to the overall photon fluence from tungsten spectrum at 25 kvp, 0 C, 760 mmhg and 0% relative humidity with the tungsten target spectrum at the same tube voltage and corrected for temperature, pressure and humidity Contribution to photon fluence by energy range 6 8 kev Contribution to photon fluence by energy range 8 15 kev Contribution to photon fluence by energy range kev Photon fluence at Total contribution 1.15E E E mmhg, Percent contribution (%) C, 0% RH Photon fluence at Total contribution 1.47E E E mmhg Percent contribution (%) C, 40% RH % difference in photon fluence due to P, T, H correction The overall percentage due to the combined effects of temperature, pressure and humidity is also shown. Overall difference in estimated photon fluence due to P, T and H correction=2.68%. were selected: 4 8, 8 15 and kev for the molybdenum spectrum and 6 8, 8 15 and kev for the tungsten spectrum. The overall difference in the estimated air kerma values of the entire molybdenum and tungsten spectra due to the combined effects of temperature, pressure and humidity correction is also shown in Tables 2 and Pressure As expected, the effect of pressure was more pronounced at the lower energies than at higher energies (Fig. 1). For the molybdenum spectrum, a deviation range from less than % was observed, between the energies 4 and 10 kev, for a pressure range of mmhg. For the same energy range, the deviation was 2 50% for pressure values mmhg. Above 10 kev, the deviations due to pressure effects ranged from less than 1% to 10% for the entire pressure range of mmhg. The overall effect of pressure on the integrated photon fluence is minimal with a maximum of 3% at 600 mmhg pressure and a minimum of 0.33% at 780 mmhg (Fig. 2). The low overall deviation of pressure on the photon fluence in spite of the high deviations at the lower energies is due to the fact that after filtration of the spectrum, the lower-energy

12 718 ARTICLE IN PRESS M. Assiamah et al. / Radiation Physics and Chemistry 68 (2003) Table 4 Comparison of contributions of three arbitrary energy segments to the overall air kerma from a tungsten spectrum at 25 kvp, 0 C, 760 mmhg and 0% relative humidity with the tungsten target spectrum at the same tube voltage and corrected for temperature, pressure and humidity Contribution to air kerma (cgy) by energy range 6 8 kev Contribution to air kerma (cgy) by energy range 8 15 kev Contribution to air kerma (cgy) by energy range kev Air kerma (cgy) at Total contribution 1.32E mmhg, Percent contribution (%) C, 0% RH Air kerma (cgy) at Total contribution 2.42E mmhg Percent contribution (%) C, 40% RH % difference in air kerma due to P, T, H correction The overall percentage due to the combined effects of temperature, pressure and humidity is also shown. Overall difference in estimated air kerma (cgy) due to P, T and H correction=49.2%. component is removed hence its contribution towards the entire spectrum is reduced. The deviation pattern is similar for both molybdenum and tungsten targets (Figs. 3 and 4). The figures involved are however different. The effects on the tungsten target spectrum being lower than that from the molybdenum. The difference in the pressure effect on the molybdenum and tungsten targets is as a result of the molybdenum X-ray spectrum having many more lower-energy components when compared to the X-ray spectrum from tungsten Temperature The temperature effect on the photon fluence increases with increasing temperature. Below 10 kev, and for temperature range of 5 25 C, a deviation of 1 70% was calculated for the molybdenum target (Fig. 5). For the same energies, the deviation was about 5 120% at temperatures C. At temperatures above 40 C, deviations of 5% to more than 150% were observed for the same energy range. At energies greater than 10 kev, the deviations were much lower and ranged from less than 1 10% for the entire temperature range, 5 50 C. The overall temperature effect on the photon fluence from a molybdenum target yielded a deviation of % (Fig. 6). The patterns of the temperature effect on both the molybdenum and tungsten targets were similar (Figs. 7 and 8). The temperature effect was however lower for the tungsten target than for the molybdenum target. This may be attributed to the proportionally larger contribution from the low-energy components that are present in the molybdenum spectrum when compared to tungsten X-ray spectrum Humidity The humidity effects on the photon fluence for the molybdenum and tungsten targets were very low compared to the pressure and temperature effects. The deviations due to humidity effects on the entire energy spectrum of both the molybdenum and tungsten targets were less than 5% (Figs. 9 and 11) for the full relative humidity range, 0 100%, at all temperatures considered Pressure, temperature and humidity The photon fluence comparison (Table 1) showed that the effect of the pressure, temperature and humidity correction on the three arbitrary energy ranges chosen for the molybdenum target varied with the effect being highest (29.5%) in the lower energy range; 6.6% and 2.7% for the energy ranges 8 15 and kev, respectively. The overall difference in photon fluence of the molybdenum spectrum due to the correction for P, T and H was 4%. The overall difference was low in view of the fact that the contribution to the lower-energy segment of the spectrum from molybdenum is reduced due to filtration. The photon fluence comparison for the tungsten spectrum (Table 3) followed a similar pattern to the molybdenum spectrum with the overall difference even smaller. The results from the air kerma calculations of the molybdenum spectra (Table 2) indicate that at the energy range 4 8 kev, the combined effects of pressure, temperature and humidity is significant. The contribution of air kerma for this energy range to the air kerma of the entire spectrum is however negligible. The contribution to the calculated air kerma in the energy range 8 15 kev was 52% for the molybdenum spectrum

13 ARTICLE IN PRESS M. Assiamah et al. / Radiation Physics and Chemistry 68 (2003) corrected for air density and 50% for the reference molybdenum spectrum. For the same energy range, the contribution to the calculated air kerma was 26% and 25% respectively, for the corrected and reference tungsten spectra (Table 4). The overall percentage difference in the estimated air kerma value due to the pressure, temperature and humidity correction at this energy range was 55% and 53%, respectively, for the molybdenum and tungsten spectra (Tables 2 and 4). For the energy range kev, the contribution to the estimated air kerma was 53% and 54%, respectively, for the corrected and reference molybdenum spectra; 79% and 80%, respectively, for the corrected and reference tungsten spectra. The overall difference in the estimated air kerma at this energy range, due to pressure, temperature and humidity correction was 48% for both the molybdenum and tungsten spectra. The overall difference in the estimated air kerma of the entire spectrum due to the combined effects of pressure, temperature and humidity was 52% and 49%, respectively, for the molybdenum and tungsten spectra at 0 C, 760 mmhg and 0% relative humidity shown in Figs. 14 and 16 (Tables 2 and 4), respectively. For the molybdenum spectrum, the air kerma difference contrasts significantly with the 45% overall difference obtained routinely by correcting for pressure and temperature only (Table 5). 4. Conclusion The investigation has shown that the effect of pressure on photon fluence of an X-ray spectrum from both molybdenum and tungsten targets are significant at Fig. 16. A comparison of the air kerma spectrum from a tungsten target at 25 kvp. (A) Spectrum at 760 mmhg, 0 C and 0% relative humidity. (B) Spectrum corrected for temperature (35 C), pressure (600 mmhg) and 50% relative humidity. Table 5 Comparison of change in the air kerma values obtained using normal temperature and pressure correction with values obtained from the correction procedure outlined in this study X-ray target material Predicted air kerma change (%) Expected air kerma change after normal P, T correction (%) Predicted Expected (%) Mo W

14 720 ARTICLE IN PRESS M. Assiamah et al. / Radiation Physics and Chemistry 68 (2003) photon energies less than 10 kev. Lower-pressure values showed higher deviations on the photon fluence at this energy range than higher-pressure values. Notwithstanding this observation, the contribution of the lowerenergy segment to the entire spectrum is reduced after filtration; hence, the overall deviation due to pressure of the integrated photon fluence was found to be small. It has also been established from this study that the temperature effect on photon fluence for both target materials are significant at the lower-energy segment of the X-ray spectrum. The temperature effect on the tungsten spectrum is however less than for the molybdenum target. The moisture content of the air has been found not to have any appreciable effect on photon fluence at any energy segment of the spectrum unlike temperature and pressure. The study has established that for high-altitude areas where air pressures are low, there could be a 4% difference in photon fluence when temperatures are high. It has also been established that for the same atmospheric conditions (low pressures and high temperatures) there could be a 50% overall difference in estimated air kerma values in mammography or low bremsstrahlung energy beam measurements unless the air density is corrected for pressure and temperature. The results presented suggest that unless air density is corrected for atmospheric conditions at the time of measurements, the overall difference in the estimated air kerma could be large. In the case of an X-ray spectrum with a high proportion of low-energy contributions, unless the effect of temperature, pressure and humidity on the photon flux is considered over the entire energy range, errors in kerma values significantly larger than values obtained after normal routine temperature and pressure corrections could result. This can partly be attributed, to changes in the photon fluence spectrum resulting from changes in atmospheric conditions and partly due to the influence of the energy and the linear energy-absorption product term ðe i Þðm en rþ; shown in Eq. (3). Acknowledgements The authors wish to express their gratitude to Dr. J. M. Boone, Dr. T. R. Fewell and Dr. R. J. Jennings who made available their molybdenum and tungsten anode spectral models used for this study. References Archer, B.R., Wagner, L.K., Med. Phys. 9, Assiamah, M., Mavunda, D., Nam, T.L., Keddy, R.J., Radiat. Phys. Chem. 67, 1 6. Attix, F.H., Introduction of Radiological Physics and Radiation Dosimetry. Wiley-Interscience, New York, pp Boone, J.M., Fewell, T.R., Jennings, R.J., Med. Phys. 24, Cember, H., Introduction to Health Physics (3rd Edition), McGraw-Hill, New York, pp Gerward, L., Guilbert, N., Jensen, K.B., Levring, H., Rad. Phys. Chem. 60, Hubbell, J.H., Seltzer, S.M., Tables of X-ray Mass Attenuation Coefficients and Mass Energy-Absorption Coefficients from 1 kev to 20 MeV for Elements Z=1 to 92 and 48 Additional Substances of Dosimetric Interest. National Institute of Standards and Technology, US Department of Commerce, Gaithersburg, MD Marion, J.B., Hornyak, W.F., Physics for Science and Engineering, Part 1. Saunders College Publishing, New York, pp Weast, R.C., Astle, M.J., Beyer, W.H. (Eds.), CRC Handbookof Chemistry and Physics, CRC Press, Boca Raton, FL, pp. F-9, F-10. Wolbarst, A.B., Physics of Radiology, International Edition. Appleton and Lange, Connecticut, USA, pp. 96,

Predicted image quality of a CMOS APS X-ray detector across a range of mammographic beam qualities

Predicted image quality of a CMOS APS X-ray detector across a range of mammographic beam qualities Journal of Physics: Conference Series PAPER OPEN ACCESS Predicted image quality of a CMOS APS X-ray detector across a range of mammographic beam qualities Recent citations - Resolution Properties of a

More information

Evaluation of X-Ray Beam Quality Based on Measurements and Estimations Using SpekCalc and Ipem78 Models

Evaluation of X-Ray Beam Quality Based on Measurements and Estimations Using SpekCalc and Ipem78 Models Original Article Evaluation of X-Ray Beam Quality Based on Measurements and Estimations Using SpekCalc and Ipem78 Models Suk Chiang Chen, Wei Loong Jong, Ahmad Zaky Harun Submitted: 20 Aug 2011 Accepted:

More information

Mammography is a radiographic procedure specially designed for detecting breast pathology Approximately 1 woman in 8 will develop breast cancer over

Mammography is a radiographic procedure specially designed for detecting breast pathology Approximately 1 woman in 8 will develop breast cancer over Mammography is a radiographic procedure specially designed for detecting breast pathology Approximately 1 woman in 8 will develop breast cancer over a lifetime Breast cancer screening programs rely on

More information

Update on NIST X-Ray Air-Kerma Standards and Calibrations

Update on NIST X-Ray Air-Kerma Standards and Calibrations CCRI(I)/03-19 Update on NIST X-Ray Air-Kerma Standards and Calibrations Michelle O Brien and Stephen M. Seltzer Ionizing Radiation Division National Institute of Standards and Technology Gaithersburg,

More information

Derivation of total filtration thickness for diagnostic x-ray source assembly

Derivation of total filtration thickness for diagnostic x-ray source assembly Physics in Medicine & Biology PAPER OPEN ACCESS Derivation of total filtration thickness for diagnostic x-ray source assembly To cite this article: Michiharu Sekimoto and Yoh Katoh 2016 Phys. Med. Biol.

More information

Imaging Technique Optimization of Tungsten Anode FFDM System

Imaging Technique Optimization of Tungsten Anode FFDM System Imaging Technique Optimization of Tungsten Anode FFDM System Biao Chen a*, Andrew P. Smith b, Zhenxue Jing a, Elena Ingal a a Hologic, Inc. 600 Technology Drive, DE 1970 b Hologic, Inc. 35 Crosby Drive,

More information

X-rays. X-rays are produced when electrons are accelerated and collide with a target. X-rays are sometimes characterized by the generating voltage

X-rays. X-rays are produced when electrons are accelerated and collide with a target. X-rays are sometimes characterized by the generating voltage X-rays Ouch! 1 X-rays X-rays are produced when electrons are accelerated and collide with a target Bremsstrahlung x-rays Characteristic x-rays X-rays are sometimes characterized by the generating voltage

More information

I. Introduction.

I. Introduction. JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, VOLUME 15, NUMBER 1, 2014 Accuracy of measuring half- and quarter-value layers and appropriate aperture width of a convenient method using a lead-covered case

More information

Optimization of Energy Modulation Filter for Dual Energy CBCT Using Geant4 Monte-Carlo Simulation

Optimization of Energy Modulation Filter for Dual Energy CBCT Using Geant4 Monte-Carlo Simulation Original Article PROGRESS in MEDICAL PHYSICS 27(3), Sept. 2016 http://dx.doi.org/10.14316/pmp.2016.27.3.125 pissn 2508-4445, eissn 2508-4453 Optimization of Energy Modulation Filter for Dual Energy CBCT

More information

Dosepix Detector as kvp-meter in Radiology and Mammography: First steps

Dosepix Detector as kvp-meter in Radiology and Mammography: First steps Dosepix Detector as kvp-meter in Radiology and Mammography: First steps F.Bisello, I.Ritter, F.Tennert, A.Zang MediPix Collaboration Meeting, 19th February 2014, CERN Protect, Enhance, and Save Lives -

More information

SpekCalc: a program to calculate photon spectra from tungsten anode x-ray tubes

SpekCalc: a program to calculate photon spectra from tungsten anode x-ray tubes NOTE SpekCalc: a program to calculate photon spectra from tungsten anode x-ray tubes 5 10 G Poludniowski 1, G Landry 2, F DeBlois 2,3, P M Evans 1 and F Verhaegen 2,4 1 Joint Department of Physics, Institute

More information

MEASURING CONDITIONS AND UNCERTAINTIES FOR THE COMPARISON AND CALIBRATION OF NATIONAL DOSIMETRIC STANDARDS AT THE BIPM *

MEASURING CONDITIONS AND UNCERTAINTIES FOR THE COMPARISON AND CALIBRATION OF NATIONAL DOSIMETRIC STANDARDS AT THE BIPM * MEASURING CONDITIONS AND UNCERTAINTIES FOR THE COMPARISON AND CALIBRATION OF NATIONAL DOSIMETRIC STANDARDS AT THE BIPM * C. Kessler and D.T. Burns December 2018 BUREAU INTERNATIONAL DES POIDS ET MESURES

More information

Calibration of KAP meters

Calibration of KAP meters Calibration of KAP meters Alexandr Malusek! Division of Radiological Sciences Department of Medical and Health Sciences Linköping University! 2014-04-15 1 Outline 1. KAP meter construction 2. Air kerma-area

More information

Determination of the tube voltage from clinic mammographic system using two types of detectors

Determination of the tube voltage from clinic mammographic system using two types of detectors BJRS BRAZILIAN JOURNAL OF RADIATION SCIENCES 03-1A (2015) 01-06 Determination of the tube voltage from clinic mammographic system using two types of detectors J. S. Barreira a ; V. Vivolo a a Gerência

More information

Mammography: Physics of Imaging

Mammography: Physics of Imaging Mammography: Physics of Imaging Robert G. Gould, Sc.D. Professor and Vice Chair Department of Radiology and Biomedical Imaging University of California San Francisco, California Mammographic Imaging: Uniqueness

More information

Preliminary studies of a new monitor ionization chamber

Preliminary studies of a new monitor ionization chamber 1 Preliminary studies of a new monitor ionization chamber Maíra T. Yoshizumi, Vitor Vivolo and Linda V. E. Caldas Instituto de Pesquisas Energéticas e Nucleares (IPEN-CNEN) Comissão Nacional de Energia

More information

Collimation Assessment Using GAFCHROMIC XR-M2

Collimation Assessment Using GAFCHROMIC XR-M2 Collimation Assessment Using GAFCHROMIC XR-M2 I. Introduction A method of collimation assessment for GE Senographe full-field digital mammography (FFDM) systems is described that uses a self-developing

More information

Date: July 31, 2017 Title: Design of an Aperture Assembly for X-Ray Diffraction Student: Christina Schmidt Mentor: Dr. Peter Ko

Date: July 31, 2017 Title: Design of an Aperture Assembly for X-Ray Diffraction Student: Christina Schmidt Mentor: Dr. Peter Ko Date: July 31, 2017 Title: Design of an Aperture Assembly for X-Ray Diffraction Student: Christina Schmidt Mentor: Dr. Peter Ko Abstract: Dr. Ko and I designed and built a new X-Ray diffraction aperture

More information

X-RAY IMAGING EE 472 F2017. Prof. Yasser Mostafa Kadah

X-RAY IMAGING EE 472 F2017. Prof. Yasser Mostafa Kadah X-RAY IMAGING EE 472 F2017 Prof. Yasser Mostafa Kadah www.k-space.org Recommended Textbook Stewart C. Bushong, Radiologic Science for Technologists: Physics, Biology, and Protection, 10 th ed., Mosby,

More information

Dose Reduction and Image Preservation After the Introduction of a 0.1 mm Cu Filter into the LODOX Statscan unit above 110 kvp

Dose Reduction and Image Preservation After the Introduction of a 0.1 mm Cu Filter into the LODOX Statscan unit above 110 kvp Dose Reduction and Image Preservation After the Introduction of a into the LODOX Statscan unit above 110 kvp Abstract: CJ Trauernicht 1, C Rall 1, T Perks 2, G Maree 1, E Hering 1, S Steiner 3 1) Division

More information

Published text: Institute of Cancer Research Repository Please direct all s to:

Published text: Institute of Cancer Research Repository   Please direct all  s to: This is an author produced version of an article that appears in: MEDICAL PHYSICS The internet address for this paper is: https://publications.icr.ac.uk/1316/ Copyright information: http://www.aip.org/pubservs/web_posting_guidelines.html

More information

K-edge subtraction X-ray imaging with a pixellated spectroscopic detector

K-edge subtraction X-ray imaging with a pixellated spectroscopic detector K-edge subtraction X-ray imaging with a pixellated spectroscopic detector Silvia Pani Department of Physics, University of Surrey Summary Hyperspectral imaging K-edge subtraction X-ray imaging for mammography

More information

MC SIMULATION OF SCATTER INTENSITIES IN A CONE-BEAM CT SYSTEM EMPLOYING A 450 kv X-RAY TUBE

MC SIMULATION OF SCATTER INTENSITIES IN A CONE-BEAM CT SYSTEM EMPLOYING A 450 kv X-RAY TUBE MC SIMULATION OF SCATTER INTENSITIES IN A CONE-BEAM CT SYSTEM EMPLOYING A 450 kv X-RAY TUBE A. Miceli ab, R. Thierry a, A. Flisch a, U. Sennhauser a, F. Casali b a Empa - Swiss Federal Laboratories for

More information

Investigation of the line-pair pattern method for evaluating mammographic focal spot performance

Investigation of the line-pair pattern method for evaluating mammographic focal spot performance Investigation of the line-pair pattern method for evaluating mammographic focal spot performance Mitchell M. Goodsitt, a) Heang-Ping Chan, and Bob Liu Department of Radiology, University of Michigan, Ann

More information

Introduction of a Single Chip TLD System for Patient Dosimetry

Introduction of a Single Chip TLD System for Patient Dosimetry Introduction of a Single Chip TLD System for Patient Dosimetry C. Hranitzky a, M. Halda a, G. Müller a, B. Obryk b, H. Stadtmann a* a Austrian Research Centers GmbH ARC, 2444 Seibersdorf, Austria. b Institute

More information

1. Carlton, Richard R., and Arlene M. Adler. Principles of Radiographic Imaging: An Art and a Science, 5th edition (2013).

1. Carlton, Richard R., and Arlene M. Adler. Principles of Radiographic Imaging: An Art and a Science, 5th edition (2013). CODE: RADT 151 INSTITUTE: Health Science TITLE: Radiographic Exposure DEPARTMENT: Radiologic Technology COURSE DESCRIPTION: This course covers the principles of radiographic exposure selection and manipulation

More information

PD233: Design of Biomedical Devices and Systems

PD233: Design of Biomedical Devices and Systems PD233: Design of Biomedical Devices and Systems (Lecture-8 Medical Imaging Systems) (Imaging Systems Basics, X-ray and CT) Dr. Manish Arora CPDM, IISc Course Website: http://cpdm.iisc.ac.in/utsaah/courses/

More information

Introduction. Chapter 16 Diagnostic Radiology. Primary radiological image. Primary radiological image

Introduction. Chapter 16 Diagnostic Radiology. Primary radiological image. Primary radiological image Introduction Chapter 16 Diagnostic Radiology Radiation Dosimetry I Text: H.E Johns and J.R. Cunningham, The physics of radiology, 4 th ed. http://www.utoledo.edu/med/depts/radther In diagnostic radiology

More information

HISTORY. CT Physics with an Emphasis on Application in Thoracic and Cardiac Imaging SUNDAY. Shawn D. Teague, MD

HISTORY. CT Physics with an Emphasis on Application in Thoracic and Cardiac Imaging SUNDAY. Shawn D. Teague, MD CT Physics with an Emphasis on Application in Thoracic and Cardiac Imaging Shawn D. Teague, MD DISCLOSURES 3DR- advisory committee CT PHYSICS WITH AN EMPHASIS ON APPLICATION IN THORACIC AND CARDIAC IMAGING

More information

Pinhole collimator design for nuclear survey system

Pinhole collimator design for nuclear survey system Annals of Nuclear Energy 29 (2002) 2029 2040 www.elsevier.com/locate/anucene Pinhole collimator design for nuclear survey system Wanno Lee*, Gyuseong Cho Department of Nuclear Engineering, Korea Advanced

More information

STEREOTACTIC BREAST BIOPSY EQUIPMENT SURVEYS

STEREOTACTIC BREAST BIOPSY EQUIPMENT SURVEYS STEREOTACTIC BREAST BIOPSY EQUIPMENT SURVEYS JAMES A. TOMLINSON, M.S. Diagnostic Radiological Physicist American Board of Radiology Certified Medical Physics Consultants, Inc. Bio 28 yrs experience 100%

More information

IBEX MATERIALS DETECTION TECHNOLOGY

IBEX MATERIALS DETECTION TECHNOLOGY WHITE PAPER: IBEX MATERIALS DETECTION TECHNOLOGY IBEX Innovations Ltd. Registered in England and Wales: 07208355 Address: Discovery 2, NETPark, William Armstrong Way, Sedgefield, TS21 3FH, UK Patents held

More information

Estimation of signal transfer property for wireless digital detector in different measurement schemes

Estimation of signal transfer property for wireless digital detector in different measurement schemes Estimation of signal transfer property for wireless digital detector in different measurement schemes Anatoli Vladimirov, Kalle Kepler Training Centre of Medical Physics, University of Tartu, Estonia 11

More information

New Detectors for X-Ray Metal Thickness Measuring

New Detectors for X-Ray Metal Thickness Measuring ECNDT 2006 - Poster 132 New Detectors for X-Ray Metal Thickness Measuring Boris V. ARTEMIEV, Alexander I. MASLOV, Association SPEKTR- GROUP, Moscow, Russia Abstract. X-ray thickness measuring instruments

More information

PATIENT EFFECTIVE DOSES IN DIAGNOSTIC RADIOLOGY, NA

PATIENT EFFECTIVE DOSES IN DIAGNOSTIC RADIOLOGY, NA Title of Paper: Patient effective doses in diagnostic radiology Authors: N.A. Gkanatsios, and W. Huda * Corresponding Author: Department of Radiology, University of Florida, P.O. Box 100374, Gainesville,

More information

Small-Animal X-ray Dose from Micro-CT

Small-Animal X-ray Dose from Micro-CT RESEARCH ARTICLE, pp. 149 158 149 Small-Animal X-ray Dose from Micro-CT John M. Boone, Orlando Velazquez, and Simon R. Cherry University of California Davis Abstract The use of micro-ct in small animals

More information

Test Equipment for Radiology and CT Quality Control Contents

Test Equipment for Radiology and CT Quality Control Contents Test Equipment for Radiology and CT Quality Control Contents Quality Control Testing...2 Photometers for Digital Clinical Display QC...3 Primary Workstations...3 Secondary Workstations...3 Testing of workstations...3

More information

The effect of compensating filter on image quality in lateral projection of thoraco lumbar radiography

The effect of compensating filter on image quality in lateral projection of thoraco lumbar radiography Journal of Physics: Conference Series OPEN ACCESS The effect of compensating filter on image quality in lateral projection of thoraco lumbar radiography To cite this article: N A A Daud et al 2014 J. Phys.:

More information

NOT FOR DISTRIBUTION JINST_083P_0914 v1

NOT FOR DISTRIBUTION JINST_083P_0914 v1 Use of XR-QA2 radiochromic films for quantitative imaging of a synchrotron radiation beam F. Di Lillo a,b, D. Dreossi c, F. Emiro a,b, C. Fedon d,e, R. Longo d,e, G. Mettivier a,b,*, L. Rigon d,e, P. Russo

More information

IBEX TECHNOLOGY APPLIED TO DIGITAL RADIOGRAPHY

IBEX TECHNOLOGY APPLIED TO DIGITAL RADIOGRAPHY WHITE PAPER: IBEX TECHNOLOGY APPLIED TO DIGITAL RADIOGRAPHY IBEX Innovations Ltd. Registered in England and Wales: 07208355 Address: Discovery 2, NETPark, William Armstrong Way, Sedgefield, UK Patents:

More information

Response characteristics of a tandem ionization chamber in standard X-ray beams

Response characteristics of a tandem ionization chamber in standard X-ray beams Applied Radiation and Isotopes 58 (2003) 495 500 Response characteristics of a tandem ionization chamber in standard X-ray beams Alessandro M. Costa*, Linda V.E. Caldas Instituto de Pesquisas Energ!eticas

More information

The importance of radiation quality for optimisation in radiology

The importance of radiation quality for optimisation in radiology Available online at http://www.biij.org/2007/2/e38 doi: 10.2349/biij.3.2.e38 biij Biomedical Imaging and Intervention Journal COMMENTARY The importance of radiation quality for optimisation in radiology

More information

diagnostic examination

diagnostic examination RADIOLOGICAL PHYSICS 2011 Raphex diagnostic examination Adel A. Mustafa, Ph.D., Editor PUBLISHED FOR: RAMPS (Radiological and Medical Physics Society of New York) preface The RAPHEX Diagnostic exam 2011

More information

X-ray Tube and Generator Basic principles and construction

X-ray Tube and Generator Basic principles and construction X-ray Tube and Generator Basic principles and construction Dr Slavik Tabakov - Production of X-rays OBJECTIVES - X-ray tube construction - Anode - types, efficiency - X-ray tube working characteristics

More information

X-ray Imaging. PHYS Lecture. Carlos Vinhais. Departamento de Física Instituto Superior de Engenharia do Porto

X-ray Imaging. PHYS Lecture. Carlos Vinhais. Departamento de Física Instituto Superior de Engenharia do Porto X-ray Imaging PHYS Lecture Carlos Vinhais Departamento de Física Instituto Superior de Engenharia do Porto cav@isep.ipp.pt Overview Projection Radiography Anode Angle Focal Spot Magnification Blurring

More information

Version 1.0. TechnicVR. Student Guide

Version 1.0. TechnicVR. Student Guide Version 1.0 TechnicVR s h a d e r w a r e. c o m Student Guide TechnicVR s h a d e r w a r e. c o m Student Guide shaderware 2008 PO Box 103 Saltburn Cleveland TS12 1WP w w w. s h a d e r w a r e. c o

More information

Acceptance Testing of a Digital Breast Tomosynthesis Unit

Acceptance Testing of a Digital Breast Tomosynthesis Unit Acceptance Testing of a Digital Breast Tomosynthesis Unit 2012 AAPM Spring Clinical Meeting Jessica Clements, M.S., DABR Objectives Review of technology and clinical advantages Acceptance Testing Procedures

More information

Radiographic Techniques, Contrast, and Noise in X-Ray Imaging

Radiographic Techniques, Contrast, and Noise in X-Ray Imaging Residents Section Physics Minimodule Huda and Abrahams Techniques, Contrast, and Noise in Radiography Residents Section Physics Minimodule Residents inradiology Walter Huda 1 R. Brad Abrahams 2 Huda W,

More information

Y11-DR Digital Radiography (DR) Image Quality

Y11-DR Digital Radiography (DR) Image Quality Y11-DR Digital Radiography (DR) Image Quality Image quality is stressed for all systems in Safety Code 35. In the relevant sections Health Canada s advice is the manufacturer s recommended test procedures

More information

Post-irradiation colouration of Gafchromic EBT radiochromic film

Post-irradiation colouration of Gafchromic EBT radiochromic film INSTITUTE OF PHYSICS PUBLISHING Phys. Med. Biol. 5 (25) N281 N285 PHYSICS IN MEDICINE AND BIOLOGY doi:1.188/31-9155/5/2/n4 NOTE Post-irradiation colouration of Gafchromic EBT radiochromic film Tsang Cheung

More information

Effective energy measurement using radiochromic film: application of a mobile scanner

Effective energy measurement using radiochromic film: application of a mobile scanner Polish Journal of Medical Physics and Engineering 2016;22(4):85-92 December 2016 doi: 10.1515/pjmpe-2016-0015 ISSN 1898-0309 Scientific Paper Effective energy measurement using radiochromic film: application

More information

Effect of patient support pads on image quality and dose in fluoroscopy a)

Effect of patient support pads on image quality and dose in fluoroscopy a) Effect of patient support pads on image quality and dose in fluoroscopy a) William R. Geiser, Walter Huda, b) and Nikolaos A. Gkanatsios Department of Radiology, University of Florida, Gainesville, Florida

More information

Seminar 8. Radiology S8 1

Seminar 8. Radiology S8 1 Seminar 8 Radiology Medical imaging. X-ray image formation. Energizing and controlling the X-ray tube. Image detectors. The acquisition of analog and digital images. Digital image processing. Selected

More information

(12) Patent Application Publication (10) Pub. No.: US 2005/ A1

(12) Patent Application Publication (10) Pub. No.: US 2005/ A1 (19) United States US 20050069086A1 (12) Patent Application Publication (10) Pub. No.: US 2005/0069086 A1 Deych et al. (43) Pub. Date: Mar. 31, 2005 (54) DYNAMIC EXPOSURE CONTROL IN RADIOGRAPHY (75) Inventors:

More information

RaySafe Xi. Specifications

RaySafe Xi. Specifications RaySafe Xi Specifications raysafe Xi General EMC tested Exposure needed Reset Temp. range According to EN 61000-6-1:2007 and en 61000-6-3:2007 One Automatic 15 35 C (59 95 F) DetECtor cable length 2 and

More information

Technical data CAMARGUE CS-VH50/300. VARIABLE Height Bucky Table With Ceiling Suspension

Technical data CAMARGUE CS-VH50/300. VARIABLE Height Bucky Table With Ceiling Suspension Technical data VARIABLE Height Bucky Table With Ceiling Suspension Model Variations CAMARGUE FH (Fixed Height) CAMARGUE FH Tomo CAMARGUE FH Ceiling suspension CAMARGUE VH (Variable Height) CAMARGUE VH

More information

New spectral benefi ts, proven low dose

New spectral benefi ts, proven low dose New spectral benefi ts, proven low dose Philips MicroDose mammography SI, technical data sheet Philips MicroDose SI with single-shot spectral imaging is a fullfi eld digital mammography solution that delivers

More information

Development of the Use of Amorphous Silicon (ASi) Electronic Portal Imaging Devices as a Physics Tool for Routine Linear Accelerator QA

Development of the Use of Amorphous Silicon (ASi) Electronic Portal Imaging Devices as a Physics Tool for Routine Linear Accelerator QA Development of the Use of Amorphous Silicon (ASi) Electronic Portal Imaging Devices as a Physics Tool for Routine Linear Accelerator QA Gena M.A.H 1, Ahmed L.El-Attar 2, Elbadry M. Zahran 3, Hany El-Gamal

More information

I. PERFORMANCE OF X-RAY PRODUCTION COMPONENTS FLUOROSCOPIC ACCEPTANCE TESTING: TEST PROCEDURES & PERFORMANCE CRITERIA

I. PERFORMANCE OF X-RAY PRODUCTION COMPONENTS FLUOROSCOPIC ACCEPTANCE TESTING: TEST PROCEDURES & PERFORMANCE CRITERIA FLUOROSCOPIC ACCEPTANCE TESTING: TEST PROCEDURES & PERFORMANCE CRITERIA EDWARD L. NICKOLOFF DEPARTMENT OF RADIOLOGY COLUMBIA UNIVERSITY NEW YORK, NY ACCEPTANCE TESTING GOALS PRIOR TO 1st CLINICAL USAGE

More information

RECOMMENDATION ITU-R P Attenuation by atmospheric gases

RECOMMENDATION ITU-R P Attenuation by atmospheric gases Rec. ITU-R P.676-6 1 RECOMMENDATION ITU-R P.676-6 Attenuation by atmospheric gases (Question ITU-R 01/3) (1990-199-1995-1997-1999-001-005) The ITU Radiocommunication Assembly, considering a) the necessity

More information

ELECTRONIC CONTROL CONCEPTS 160 Partition Street Saugerties, NY or local phone

ELECTRONIC CONTROL CONCEPTS 160 Partition Street Saugerties, NY or local phone ELECTRONIC CONTROL CONCEPTS 160 Partition Street Saugerties, NY 12477 (800)VIP-XRAY (845)247-9028 Fax or 800-847-9729 local phone 845-246-9013 http://www.eccxray.com sales@eccxray.com INSTRUCTION MANUAL

More information

Evaluation of Physical Tissue Compensation Filters

Evaluation of Physical Tissue Compensation Filters Chapter 3 Evaluation of Physical Tissue Compensation Filters The prime purpose of this project was to develop and evaluate a method to digitally alter the dynamic range of digital radiographic images.

More information

Quality assurance: a comparison study of radiographic exposure for neonatal chest radiographs at 4 academic hospitals

Quality assurance: a comparison study of radiographic exposure for neonatal chest radiographs at 4 academic hospitals DOI 10.1007/s00247-011-2290-1 ORIGINAL ARTICLE Quality assurance: a comparison study of radiographic exposure for neonatal chest radiographs at 4 academic hospitals Mervyn D. Cohen & Richard Markowitz

More information

ABSORBED DOSE DISTRIBUTIONS USING THE ISODENSITOMETRIC METHOD FOR EXPOSURES WITH FILTER EMPLOYED FOR MAMMOGRAPHIES

ABSORBED DOSE DISTRIBUTIONS USING THE ISODENSITOMETRIC METHOD FOR EXPOSURES WITH FILTER EMPLOYED FOR MAMMOGRAPHIES Romanian Reports in Physics, Vol. 65, No. 1, P. 168 177, 213 ABSORBED DOSE DISTRIBUTIONS USING THE ISODENSITOMETRIC METHOD FOR EXPOSURES WITH FILTER EMPLOYED FOR MAMMOGRAPHIES F. SCARLAT 1, A. SCARISOREANU

More information

Phase Imaging Using Focused Polycapillary Optics

Phase Imaging Using Focused Polycapillary Optics Phase Imaging Using Focused Polycapillary Optics Sajid Bashir, Sajjad Tahir, Jonathan C. Petruccelli, C.A. MacDonald Dept. of Physics, University at Albany, Albany, New York Abstract Contrast in conventional

More information

Breast Tomosynthesis. Bob Liu, Ph.D. Department of Radiology Massachusetts General Hospital And Harvard Medical School

Breast Tomosynthesis. Bob Liu, Ph.D. Department of Radiology Massachusetts General Hospital And Harvard Medical School Breast Tomosynthesis Bob Liu, Ph.D. Department of Radiology Massachusetts General Hospital And Harvard Medical School Outline Physics aspects of breast tomosynthesis Quality control of breast tomosynthesis

More information

Chromatic X-Ray imaging with a fine pitch CdTe sensor coupled to a large area photon counting pixel ASIC

Chromatic X-Ray imaging with a fine pitch CdTe sensor coupled to a large area photon counting pixel ASIC Chromatic X-Ray imaging with a fine pitch CdTe sensor coupled to a large area photon counting pixel ASIC R. Bellazzini a,b, G. Spandre a*, A. Brez a, M. Minuti a, M. Pinchera a and P. Mozzo b a INFN Pisa

More information

JRPR. Microwave Dielectric Absorption Spectroscopy Aiming at Novel Dosimetry Using DNAs. Original Research. Introduction

JRPR. Microwave Dielectric Absorption Spectroscopy Aiming at Novel Dosimetry Using DNAs. Original Research. Introduction Journal of Radiation Protection and Research 2017;42(1):21-25 pissn 2508-1888 eissn 2466-2461 Microwave Dielectric Absorption Spectroscopy Aiming at Novel Dosimetry Using DNAs Yoshinobu Izumi 1, *, Makoto

More information

X-RAYS - NO UNAUTHORISED ENTRY

X-RAYS - NO UNAUTHORISED ENTRY Licencing of premises Premises Refer Guidelines A radiation warning sign and warning notice, X-RAYS - NO UNAUTHORISED ENTRY must be displayed at all entrances leading to the rooms where x-ray units are

More information

ANALYTICAL MICRO X-RAY FLUORESCENCE SPECTROMETER

ANALYTICAL MICRO X-RAY FLUORESCENCE SPECTROMETER Copyright(c)JCPDS-International Centre for Diffraction Data 2001,Advances in X-ray Analysis,Vol.44 325 ANALYTICAL MICRO X-RAY FLUORESCENCE SPECTROMETER ABSTRACT William Chang, Jonathan Kerner, and Edward

More information

Applications of a Pixellated Detection System to Digital Mammography

Applications of a Pixellated Detection System to Digital Mammography Applications of a Pixellated Detection System to Digital Mammography Valeria Rosso Dipartimento di Fisica, Universita di Pisa and Sezione INFN Pisa, Italy + valeria.rosso@pi.infn.it Outline The detection

More information

A comparison of two methods for the determination of freein-air geometric efficiency in MDCT

A comparison of two methods for the determination of freein-air geometric efficiency in MDCT A comparison of two methods for the determination of freein-air geometric efficiency in MDCT Theocharis Berris *1, Kostas Perisinakis 1,, Antonios E. Papadakis and John Damilakis 1, 1 Department of Medical

More information

THE USE OF CdTe DETECTORS FOR DENTAL X-RAY SPECTROMETRY

THE USE OF CdTe DETECTORS FOR DENTAL X-RAY SPECTROMETRY 2007 International Nuclear Atlantic Conference - INAC 2007 Santos, SP, Brazil, September 30 to October 5, 2007 ASSOCIAÇÃO BRASILEIRA DE ENERGIA NUCLEAR - ABEN ISBN: 978-85-99141-02-1 THE USE OF CdTe DETECTORS

More information

The Evaluation of Collimator Alignment of Diagnostic X-ray Tube Using Computed Radiography System

The Evaluation of Collimator Alignment of Diagnostic X-ray Tube Using Computed Radiography System The Evaluation of Collimator Alignment of Diagnostic X-ray Tube Using Computed Radiography System The Evaluation of Collimator Alignment of Diagnostic X-ray Tube Using Computed Radiography System Manus

More information

GAFCHROMIC HD-810 Radiochromic Dosimetry Film Configuration, Specifications and Performance Data

GAFCHROMIC HD-810 Radiochromic Dosimetry Film Configuration, Specifications and Performance Data GAFCHROMIC HD-810 Radiochromic Dosimetry Film Configuration, Specifications and Performance Data Description GAFCHROMIC HD-810 dosimetry film is designed for the measurement of absorbed dose of high-energy

More information

Product Information PI

Product Information PI Product Information PI TOSHIBA X-RAY TUBE D-045 Stationary Anode X-ray Tube Especially designed for dental radiography unit. This tube has focus 0.4, and is available for maximum tube voltage 70 kv. Installed

More information

Effects of Plastic Wedges on Whole Foot Radiograph in Anteroposterior and Oblique Positions

Effects of Plastic Wedges on Whole Foot Radiograph in Anteroposterior and Oblique Positions American Journal of Applied Sciences Original Research Paper Effects of Plastic Wedges on Whole Foot Radiograph in Anteroposterior and Oblique Positions Poe Lorlorm, Winit Choiprasert, Montree Tungjai

More information

Industrial X-ray Tube KL

Industrial X-ray Tube KL Description stationary anode x-ray tube is a highly popular and proven side window x-ray tube, which has focus 1.0 mm, is designed for non-destructive x-ray inspection. stationary anode x-ray tube can

More information

Explain what is meant by a photon and state one of its main properties [2]

Explain what is meant by a photon and state one of its main properties [2] 1 (a) A patient has an X-ray scan taken in hospital. The high-energy X-ray photons interact with the atoms inside the body of the patient. Explain what is meant by a photon and state one of its main properties....

More information

Nuclear Associates

Nuclear Associates Nuclear Associates 07-706 Patient Phantom/Penetrometer System Users Manual March 2005 Manual No. 07-706-1 Rev. 2 2004, 2005 Fluke Corporation, All rights reserved. Printed in U.S.A. All product names are

More information

熊本大学学術リポジトリ. Kumamoto University Repositor

熊本大学学術リポジトリ. Kumamoto University Repositor 熊本大学学術リポジトリ Kumamoto University Repositor Title Monte Carlo calculations of the rep correction factor, Ρ_, for cy chamber cav Author(s) Araki, Fujio CitationRadiological Physics and Technology Issue

More information

Joint ICTP/IAEA Advanced School on Dosimetry in Diagnostic Radiology and its Clinical Implementation May 2009

Joint ICTP/IAEA Advanced School on Dosimetry in Diagnostic Radiology and its Clinical Implementation May 2009 2033-6 Joint ICTP/IAEA Advanced School on Dosimetry in Diagnostic Radiology and its Clinical Implementation 11-15 May 2009 Dosimetry for Fluoroscopy Basics Renato Padovani EFOMP Joint ICTP-IAEA Advanced

More information

MILADY. Product Data. Page 1 of 8

MILADY. Product Data. Page 1 of 8 Page 1 of 8 The MILADY Mammographic Unit offers the best quality-to-price ratio to our customers worldwide. The unit advanced technology together with the application of industrial production standards,

More information

A new approach to film dosimetry for high energy photon beams: Lateral scatter filtering

A new approach to film dosimetry for high energy photon beams: Lateral scatter filtering A new approach to film dosimetry for high energy photon beams: Lateral scatter filtering Sandra E. Burch Department of Radiology, Medical College of Georgia, Augusta, Georgia 30912 Kimberlee J. Kearfott

More information

Cylindrical Ion Chambers Victoreen Model 550 Series

Cylindrical Ion Chambers Victoreen Model 550 Series Cylindrical Ion Chambers Victoreen Model 550 Series! Cylindrical Ion Chambers for use with Model 35040 and Model 530 electrometers! Wide range of applications in Diagnostic X-Ray and Radiation Oncology

More information

Monochromatic X-ray sources based on Table-top electron accelerators and X-ray tubes. A.P. Potylitsyn TPU, Tomsk, Russia

Monochromatic X-ray sources based on Table-top electron accelerators and X-ray tubes. A.P. Potylitsyn TPU, Tomsk, Russia Monochromatic X-ray sources based on Table-top electron accelerators and X-ray tubes A.P. Potylitsyn TPU, Tomsk, Russia The main radiation mechanisms in amorphous targets: Bremsstrahlung Transition radiation

More information

SPECIFICATION. Kilovoltage X-ray calibration system for protection and diagnostic level dosimetry. Prepared by

SPECIFICATION. Kilovoltage X-ray calibration system for protection and diagnostic level dosimetry. Prepared by SPECIFICATION Kilovoltage X-ray Prepared by Igor Gomola, Technical Officer, Project ECU6023, Date 2015-Oct-06 Revision Date Status Comments 0.1 2015-Oct-06 Draft Igor Gomola Page 1 of 12 1. Scope This

More information

Effect of Backscattered Radiation on X-Ray Image Contrast

Effect of Backscattered Radiation on X-Ray Image Contrast Applied Physics Research; Vol. 9, No. 1; 2017 ISSN 1916-9639 E-ISSN 1916-9647 Published by Canadian Center of Science and Education Effect of Backscattered Radiation on X-Ray Image Contrast A. T. Naji

More information

MAN Revision 002. InSight Mini C-arm Imaging System Technical Reference Manual

MAN Revision 002. InSight Mini C-arm Imaging System Technical Reference Manual MAN-00668 Revision 002 InSight Mini C-arm Imaging System Technical Reference Manual April 2007 The information contained in this Manual is confidential and proprietary to Hologic, Inc. This information

More information

PERFORMANCE MEASUREMENT OF MEDICAL IMAGING SYSTEMS BASED ON MUTUAL INFORMATION METRIC

PERFORMANCE MEASUREMENT OF MEDICAL IMAGING SYSTEMS BASED ON MUTUAL INFORMATION METRIC XIX IMEKO World Congress Fundamental and Applied Metrology September 6 11, 2009, Lisbon, Portugal PERFORMANCE MEASUREMENT OF MEDICAL IMAGING SYEMS BASED ON MUTUAL INFORMATION METRIC Eri Matsuyama 1, Du-Yih

More information

Development of the Mini-Ionization Chamber for High-Dose Real-Time Monitoring Inside a Gamma Irradiation Facility

Development of the Mini-Ionization Chamber for High-Dose Real-Time Monitoring Inside a Gamma Irradiation Facility 1098 IEEE TRANSACTIONS ON NUCLEAR SCIENCE, VOL. 50, NO. 4, AUGUST 2003 Development of the Mini-Ionization Chamber for High-Dose Real-Time Monitoring Inside a Gamma Irradiation Facility Ary de A. Rodrigues,

More information

Cylindrical Ion Chambers

Cylindrical Ion Chambers Cylindrical Ion Chambers Radiation Oncology ON Victoreen Model 550T Series Cylindrical Ion Chambers for use with Model 35040 and Model 560 electrometers Wide range of applications in Diagnostic X-Ray and

More information

- Water resistant. - Large size.

- Water resistant. - Large size. GAFCHROMIC EBT product brochure GAFCHROMIC EBT FEATURES GAFCHROMIC EBT dosimetry film has been developed specifically to address the needs of the medical physicist and dosimetrist working in the radiotherapy

More information

Phase Contrast Imaging with X-ray tube

Phase Contrast Imaging with X-ray tube Phase Contrast Imaging with X-ray tube Institute for Roentgen Optics /IRO/, Moscow Vladimir Shovkun and Muradin Kumakhov Proc. SPIE v.5943, 2005 Institute for Roentgen Optics. Vladimir Ya. Shovkun. E-mail:

More information

Your worldwide source of grids. What a difference a grid makes

Your worldwide source of grids. What a difference a grid makes Your worldwide source of grids What a difference a grid makes The right assembly to meet your every need Trust the experts A high degree of expertise is required to design and manufacture quality X-ray

More information

TITLE: Automatic Exposure Control Device for Digital Mammography

TITLE: Automatic Exposure Control Device for Digital Mammography AD Award Number: DAMD17-99-1-9429 TITLE: Automatic Exposure Control Device for Digital Mammography PRINCIPAL INVESTIGATOR: Laurie L. Fajardo, M.D. CONTRACTING ORGANIZATION: Johns Hopkins University Baltimore,

More information

X-Ray Transport, Diagnostic, & Commissioning Plans. LCLS Diagnostics and Commissioning Workshop

X-Ray Transport, Diagnostic, & Commissioning Plans. LCLS Diagnostics and Commissioning Workshop X-Ray Transport, Diagnostic, & Commissioning Plans LCLS Diagnostics and Commissioning Workshop *This work was performed under the auspices of the U.S. Department of Energy by the University of California,

More information

DENTAL RADIOGRAPHY KAMARUL AMIN BIN ABU BAKAR

DENTAL RADIOGRAPHY KAMARUL AMIN BIN ABU BAKAR DENTAL RADIOGRAPHY KAMARUL AMIN BIN ABDULLAH @ ABU BAKAR Components of the Dental X-Ray Machine Dental x-ray machines may vary somewhat in size and appearance, but all machines will have three primary

More information

Title detector with operating temperature.

Title detector with operating temperature. Title Radiation measurements by a detector with operating temperature cryogen Kanno, Ikuo; Yoshihara, Fumiki; Nou Author(s) Osamu; Murase, Yasuhiro; Nakamura, Masaki Citation REVIEW OF SCIENTIFIC INSTRUMENTS

More information

Study of increased radiation when an x-ray tube is placed in a strong magnetic field

Study of increased radiation when an x-ray tube is placed in a strong magnetic field Study of increased radiation when an x-ray tube is placed in a strong magnetic field Zhifei Wen Departments of Radiology and Physics, Stanford University, Stanford, California 94305 Norbert J. Pelc Departments

More information