ANALYTICAL MICRO X-RAY FLUORESCENCE SPECTROMETER

Size: px
Start display at page:

Download "ANALYTICAL MICRO X-RAY FLUORESCENCE SPECTROMETER"

Transcription

1 Copyright(c)JCPDS-International Centre for Diffraction Data 2001,Advances in X-ray Analysis,Vol ANALYTICAL MICRO X-RAY FLUORESCENCE SPECTROMETER ABSTRACT William Chang, Jonathan Kerner, and Edward Franc0 ARACOR 425 Lakeside Drive Sunnyvale, CA An analytical micro-spectrometer has been developed for materials characterization using X-ray fluorescence with a spatial resolution between 1 and 25 microns. The key component of this system is a monocapillary X-ray optic that has been optimized for intensity throughput. INTRODUCTION Monocapillary X-ray optics are a promising technology for producing intense microbeams of X rays. When coupled with laboratory X-ray sources, these optics are an enabling technology for the implementation of X-ray-based micro-spectrometer with spatial resolutions ranging from 1 to 25 microns. These instruments use X-ray fluorescence to determine the composition and thickness of materials with a spatial resolution that is comparable to the beam size. This paper concentrates on the measured performance of these optics and their incorporation into a generalpurpose micro-spectrometer. SYSTEM OVERVIEW An overview of the micro-spectrometer is shown in Figure 1. The primary components are 1) a high-magnification microscope to image the sample; 2) a precision translation stage; 3) a low power X-ray tube and capillary optic; 4) an energy dispersive X-ray detector; and, 5) a computer to control the data acquisition and stage position. The X-ray tube, with a tungsten anode and a focal spot size of about 30 microns, can be operated up to a maximum voltage of 80 kv and at 50 W.2 A significant advantage of this source is the extremely close coupling (less than 5 mm) that can be achieved between the X-ray focal spot and the entrance of the capillary optic. The X-ray detector is a Si PIN diode with an energy resolution of about 240 ev.3 The user interface for acquiring the spectra was adapted from commercially available software. X-RAY OPTICS Tapered monocapillary optics are used to concentrate the X-ray beam and produce a small spot. These optics are all standardized to a common length which allows us to rapidly interchange the optics to produce a range of beam sizes, depending on the sample requirements. The optics were all fabricated at ARACOR using an advanced capillary puller that can be programmed to reproducibly fabricate optics with almost any desired shape. We pull the glass dynamically, moving the location of the heat zone, to shape the glass over a total length of about 30 cm. For this application, we fabricated conical optics that are optimized for close-coupled anode geometry of the X-ray tube. Typical entrance diameters for this application range from 50 to 250 microns, with an exit diameter between 1 and 100 microns, over a lo-cm length of glass.

2 This document was presented at the Denver X-ray Conference (DXC) on Applications of X-ray Analysis. Sponsored by the International Centre for Diffraction Data (ICDD). This document is provided by ICDD in cooperation with the authors and presenters of the DXC for the express purpose of educating the scientific community. All copyrights for the document are retained by ICDD. Usage is restricted for the purposes of education and scientific research. DXC Website ICDD Website -

3 Copyright(c)JCPDS-International Centre for Diffraction Data 2001,Advances in X-ray Analysis,Vol f Capillary Stage Low Power X-ray Tube Video Microscope Monocapillary Image Plane Y 1 Precision Stage Figure 1. Schematic of the micro-spectrometer. Figure 2 shows a representative capillary optic and a standardized capillary holder. The capillary, which was designed for a different application, is about 15 cm long. The capillary holder is designed to locate the tapered optic in a reproducible manner so that a common optical axis exists for a wide variety of optic geometries. It also contains shielding to reduce the leakage of radiation that passes through the glass walls of the optic. Prior to their installation into the system, these optics are characterized on a laboratory bench to determine their efficiency, throughput, beam size and divergence. We have the choice of using copper, molybdenum, or tungsten X-ray tubes. Data using molybdenum and tungsten tubes indicate that the range of absolute gains in intensity (over a pinhole located at the exit) are from 40 to 120X. The variability in the gain is due to differences in the optic design that affect the entrance solid angle and the exit aperture. Figure 2. Tapered monocapillary optic and holder. Typical beam size data are shown in Figure 3, obtained by scanning a smooth wire through the beam and recording the resulting attenuation as a function of distance from the tip of the capillary. The beam from the capillary appears to be well behaved for distances from tip to sample of about 10 mm or less. This can be seen from the smooth fall-off in the intensity as a function of the wire position. Also, the width of the beam is comparable when the X-ray tube is operated at 20 or 45 kv, indicating that the beam size does not significantly depend on the

4 Copyright(c)JCPDS-International Centre for Diffraction Data 2001,Advances in X-ray Analysis,Vol energy of the incident X-ray beam. This is a consequence of the fact that we have achieved an essentially straight centerline and that energy-dependent leakage is not a significant factor. Finally, the divergence of the beam is approximately 4 milliradians, which means that the beam grows by approximately 4 microns per mm of separation between the exit of the capillary and the sample. Thus at a distance of 1 mm, 90 percent of the beam is contained within 15 microns, while at a distance of 6 mm, 90 percent of the beam is contained within 35 microns g Position (urn) Distance from tip (mm) Figure 3. Measured beam profiles and sizes (595%) as a function of distance from the tip. Once these capillaries are installed into the system, an edge is scanned through the beam while recording the resulting X-ray fluorescence intensity as a function of position. This indicates that the sample is correctly positioned within the instrument and provides a direct measurement of the actual X-ray beam size used in the examination. PERFORMANCE The signal levels from various bulk samples were examined to obtain an indication of the performance of the system. Typical results for a potential of 45 kv and a measured beam size of about 10 microns were obtained with the micro-spectrometer. A 13 mm2 Si PIN detector (300 microns thick) located at a distance of 15 mm from the sample with a take-off angle of 45 degrees, was used to record the resulting spectra. Under these conditions, we measured a Cu-Ka count rate of 120 cps/w, a MO-Ka count rate of 8 cps/w, and a Sn-Ka count rate of 3 cps/w. The background, due to scatter of the incident X-ray beam, is generally two orders of magnitude lower than the signal. The Cu count rate is significantly higher than the MO and Sn count rates. This is due to the efficient excitation of the Cu-Ka X rays by the characteristic W-L lines, produced in the W anode, and by the almost 100% detection efficiency of the Si PIN detector for

5 Copyright(c)JCPDS-International Centre for Diffraction Data 2001,Advances in X-ray Analysis,Vol Cu Ka X rays. In contrast, only bremsstrahlung radiation above the Sn-K edge (at kev) can produce characteristic Sn X rays, and their detection efficiency is only about 10%. Figure 4 shows representative images, obtained in l-3 hours, that illustrate the performance and usefulness of the micro-spectrometer. The diameter of each copper wire in the grid is 60 microns. The other two images are representative of applications in mineralogy and in contamination analysis. The mineral sample depicts an iron silicate species that is on the order of 50 microns in diameter contained in a fluorapatite sample. The image of the indium bump (60 microns square) shows a Zn-containing particle, estimated to be about 10 microns in diameter, on the alignment cross that was introduced during deposition. The high spatial resolution achieved with the capillary optic allowed the identification of the composition of this particle. Copper grid Mineral sample Figure 4. Example elemental images. Indium bump SUMMARY We have shown that resolution in the range of 10 microns is achievable for elemental mapping with a monocapillary optic and a laboratory X-ray source. These optics produce a gain in beam intensity, and can produce a beam size that is normally not feasible for pinhole geometries. The X-ray micro-spectrometer is complementary to other microbeam technologies. ACKNOWLEDGEMENTS This effort was partially supported by the Department of Energy under DE-FG03-96ER82133 and the National Institutes of Health under 2R42GM We would like to thank Don Bilderback, CHESS, and Brian Cross, METARA, for many helpful discussions. REFERENCES 1) Bilderback D. H. and E. D. France, Total Reflection Optics: Single Capillaries, in Handbook of Optics Vol. III, Optical Society of America (2001), pages ) XTG UltraBright Microfocus x-ray source from Oxford Instruments, Scotts Valley, CA. 3) Si-PIN photodiode detector XR-1 OOCR from Amptek, Bedford, MA.

MICRO XRF OF LIGHT ELEMENTS USING A POLYCAPILLARY LENS AND AN ULTRA THIN WINDOW SILICON DRIFT DETECTOR INSIDE A VACUUM CHAMBER

MICRO XRF OF LIGHT ELEMENTS USING A POLYCAPILLARY LENS AND AN ULTRA THIN WINDOW SILICON DRIFT DETECTOR INSIDE A VACUUM CHAMBER Copyright JCPDS - International Centre for Diffraction Data 2005, Advances in X-ray Analysis, Volume 48. 229 MICRO XRF OF LIGHT ELEMENTS USING A POLYCAPILLARY LENS AND AN ULTRA THIN WINDOW SILICON DRIFT

More information

X-RAY OPTICS FOR TWO-DIMENSIONAL DIFFRACTION

X-RAY OPTICS FOR TWO-DIMENSIONAL DIFFRACTION Copyright (c)jcpds-international Centre for Diffraction Data 2002, Advances in X-ray Analysis, Volume 45. 332 ABSTRACT X-RAY OPTICS FOR TWO-DIMENSIONAL DIFFRACTION Bob B. He and Uwe Preckwinkel Bruker

More information

WIDE ANGLE GEOMETRY EDXRF SPECTROMETERS WITH SECONDARY TARGET AND DIRECT EXCITATION MODES

WIDE ANGLE GEOMETRY EDXRF SPECTROMETERS WITH SECONDARY TARGET AND DIRECT EXCITATION MODES Copyright(C)JCPDS-International Centre for Diffraction Data 2000, Advances in X-ray Analysis, Vol.42 11 Copyright(C)JCPDS-International Centre for Diffraction Data 2000, Advances in X-ray Analysis, Vol.42

More information

USING A CHARGE-COUPLED DEVICE (CCD) TO GATHER X-RAY FLUORESCENCE (XRF)AND X-RAY DIFFRACTION (XRD) INFORMATION SIMULTANEOUSLY

USING A CHARGE-COUPLED DEVICE (CCD) TO GATHER X-RAY FLUORESCENCE (XRF)AND X-RAY DIFFRACTION (XRD) INFORMATION SIMULTANEOUSLY Copyright(c)JCPDS-International Centre for Diffraction Data 2001,Advances in X-ray Analysis,Vol.44 343 USING A CHARGE-COUPLED DEVICE (CCD) TO GATHER X-RAY FLUORESCENCE (XRF)AND X-RAY DIFFRACTION (XRD)

More information

Advancing EDS Analysis in the SEM Quantitative XRF. International Microscopy Congress, September 5 th, Outline

Advancing EDS Analysis in the SEM Quantitative XRF. International Microscopy Congress, September 5 th, Outline Advancing EDS Analysis in the SEM with in-situ Quantitative XRF Brian J. Cross (1) & Kenny C. Witherspoon (2) 1) CrossRoads Scientific, El Granada, CA 94018, USA 2) ixrf Systems, Inc., Houston, TX 77059,

More information

DEVELOPMENT OF A WAVELENGTH DISPERSIVE X-RAY FLUORESCENCE SPECTROMETER USING A MULTI-CAPILLARY X-RAY LENS FOR X-RAY DETECTION

DEVELOPMENT OF A WAVELENGTH DISPERSIVE X-RAY FLUORESCENCE SPECTROMETER USING A MULTI-CAPILLARY X-RAY LENS FOR X-RAY DETECTION Copyright JCPDS - International Centre for Diffraction Data 2003, Advances in X-ray Analysis, Volume 46. 346 DEVELOPMENT OF A WAVELENGTH DISPERSIVE X-RAY FLUORESCENCE SPECTROMETER USING A MULTI-CAPILLARY

More information

DOUBLE MULTILAYER MONOCHROMATOR WITH FIXED EXIT GEOMETRY. H.Gatterbauer, P.Wobrauschek, F.Hegediis, P.Biini, C.Streli

DOUBLE MULTILAYER MONOCHROMATOR WITH FIXED EXIT GEOMETRY. H.Gatterbauer, P.Wobrauschek, F.Hegediis, P.Biini, C.Streli Copyright (C) JCPDS International Centre for Diffraction Data 1999 379 DOUBLE MULTILAYER MONOCHROMATOR WITH FIXED EXIT GEOMETRY H.Gatterbauer, P.Wobrauschek, F.Hegediis, P.Biini, C.Streli Atominsitut der

More information

MINIATURE X-RAY SOURCES AND THE EFFECTS OF SPOT SIZE ON SYSTEM PERFORMANCE

MINIATURE X-RAY SOURCES AND THE EFFECTS OF SPOT SIZE ON SYSTEM PERFORMANCE 228 MINIATURE X-RAY SOURCES AND THE EFFECTS OF SPOT SIZE ON SYSTEM PERFORMANCE D. CARUSO, M. DINSMORE TWX LLC, CONCORD, MA 01742 S. CORNABY MOXTEK, OREM, UT 84057 ABSTRACT Miniature x-ray sources present

More information

CONFOCAL GRADED d-spacing MULTILAYER BEAM CONDITIONING OPTICS

CONFOCAL GRADED d-spacing MULTILAYER BEAM CONDITIONING OPTICS Copyright(C)JCPDS-International Centre for Diffraction Data 2000, Advances in X-ray Analysis, Vol.42 321 Copyright(C)JCPDS-International Centre for Diffraction Data 2000, Advances in X-ray Analysis, Vol.42

More information

LONG TERM STATISTICS OF X-RAY SPECTROMETERS

LONG TERM STATISTICS OF X-RAY SPECTROMETERS 403 LONG TERM STATISTICS OF X-RAY SPECTROMETERS J. F. Dlouhy*, D. Mathieu Department of the Environment, Environmental Technology Center, River Road, Ottawa, Ontario, Canada Kl A OH3 K. N. Stoev Bulgarian

More information

MOXTEK S NEW ULTRA-LITE X-RAY SOURCES: PERFORMACE CHARACTERIZATIONS

MOXTEK S NEW ULTRA-LITE X-RAY SOURCES: PERFORMACE CHARACTERIZATIONS Copyright JCPDS-International Centre for Diffraction Data 2013 ISSN 1097-0002 202 MOXTEK S NEW ULTRA-LITE X-RAY SOURCES: PERFORMACE CHARACTERIZATIONS S. Cornaby, S. Morris, J. Smith, D. Reynolds, K. Kozaczek

More information

GUNSHOT RESIDUE INVESTIGATIONS USING TXRF

GUNSHOT RESIDUE INVESTIGATIONS USING TXRF 299 GUNSHOT RESIDUE INVESTIGATIONS USING TXRF Alexander Wastl 1, Bettina Bogner 2, Peter Kregsamer 1, Peter Wobrauschek 1, Christina Streli 1 1 Atominstitut, Vienna University of Technology, Vienna, Austria

More information

SIMULTANEOUS XRD/XRF WITH LOW-POWER X-RAY TUBES

SIMULTANEOUS XRD/XRF WITH LOW-POWER X-RAY TUBES Copyright (c)jcpds-international Centre for Diffraction Data 2002, Advances in X-ray Analysis, Volume 45. 34 SIMULTANEOUS XRD/XRF WITH LOW-POWER X-RAY TUBES S. Cornaby 1, A. Reyes-Mena 1, P. W. Moody 1,

More information

ON THE ENERGY DEPENDENCE OF THE DETECTOR EFFICIENCY OF A Si-PIN DIODE

ON THE ENERGY DEPENDENCE OF THE DETECTOR EFFICIENCY OF A Si-PIN DIODE Copyright(C)JCPDS-International Centre for Diffraction Data 2, Advances in X-ray Analysis, Vol.42 36 Copyright(C)JCPDS-International Centre for Diffraction Data 2, Advances in X-ray Analysis, Vol.42 36

More information

Applications of New, High Intensity X-Ray Optics - Normal and thin film diffraction using a parabolic, multilayer mirror

Applications of New, High Intensity X-Ray Optics - Normal and thin film diffraction using a parabolic, multilayer mirror Applications of New, High Intensity X-Ray Optics - Normal and thin film diffraction using a parabolic, multilayer mirror Stephen B. Robie scintag, Inc. 10040 Bubb Road Cupertino, CA 95014 Abstract Corundum

More information

(Refer Slide Time: 00:10)

(Refer Slide Time: 00:10) Fundamentals of optical and scanning electron microscopy Dr. S. Sankaran Department of Metallurgical and Materials Engineering Indian Institute of Technology, Madras Module 03 Unit-6 Instrumental details

More information

X-RAY BACKSCATTER IMAGING: PHOTOGRAPHY THROUGH BARRIERS

X-RAY BACKSCATTER IMAGING: PHOTOGRAPHY THROUGH BARRIERS Copyright JCPDS-International Centre for Diffraction Data 2006 ISSN 1097-0002 X-RAY BACKSCATTER IMAGING: PHOTOGRAPHY THROUGH BARRIERS 13 Joseph Callerame American Science & Engineering, Inc. 829 Middlesex

More information

OPTIMIZING THE ELEMENTAL SENSITIVITY AND FOCAL SPOT SIZE OF A MONOLITHIC POLYCAPILLARY OPTIC USING MICRO-X-RAY FLUORESCENCE

OPTIMIZING THE ELEMENTAL SENSITIVITY AND FOCAL SPOT SIZE OF A MONOLITHIC POLYCAPILLARY OPTIC USING MICRO-X-RAY FLUORESCENCE , Copyright(C)JCPDS-International Centre for Diffraction Data 2000, Advances in X-ray Analysis, Vol.42 26 Copyright(C)JCPDS-International Centre for Diffraction Data 2000, Advances in X-ray Analysis, Vol.42

More information

ARTAX. Innovation with Integrity. Portable Micro-XRF Spectrometer. Micro-XRF

ARTAX. Innovation with Integrity. Portable Micro-XRF Spectrometer. Micro-XRF ARTAX Portable Micro-XRF Spectrometer Innovation with Integrity Micro-XRF ARTAX Elemental Analysis for the Art Community and More The ARTAX is the first portable X ray fluorescence (XRF) spectrometer designed

More information

research papers First results from a macromolecular crystallography system with a polycapillary collimating optic and a microfocus X-ray generator

research papers First results from a macromolecular crystallography system with a polycapillary collimating optic and a microfocus X-ray generator Journal of Applied Crystallography ISSN 0021-8898 First results from a macromolecular crystallography system with a polycapillary collimating optic and a microfocus X-ray generator Received 7 September

More information

Phase Imaging Using Focused Polycapillary Optics

Phase Imaging Using Focused Polycapillary Optics Phase Imaging Using Focused Polycapillary Optics Sajid Bashir, Sajjad Tahir, Jonathan C. Petruccelli, C.A. MacDonald Dept. of Physics, University at Albany, Albany, New York Abstract Contrast in conventional

More information

MICROANALYSIS WITH A POLYCAPILLARY IN A VACUUM CHAMBER

MICROANALYSIS WITH A POLYCAPILLARY IN A VACUUM CHAMBER THE RIGAKU JOURNAL VOL. 20 / NO. 2 / 2003 MICROANALYSIS WITH A POLYCAPILLARY IN A VACUUM CHAMBER CHRISTINA STRELI a), NATALIA MAROSI, PETER WOBRAUSCHEK AND BARBARA FRANK Atominstitut der Österreichischen

More information

Supplementary Figure S1. Schematic representation of different functionalities that could be

Supplementary Figure S1. Schematic representation of different functionalities that could be Supplementary Figure S1. Schematic representation of different functionalities that could be obtained using the fiber-bundle approach This schematic representation shows some example of the possible functions

More information

BRUKER ADVANCED X-RAY SOLUTIONS. SPECTROMETRY SOLUTIONS ARTAX mxrf SPECTROMETER

BRUKER ADVANCED X-RAY SOLUTIONS. SPECTROMETRY SOLUTIONS ARTAX mxrf SPECTROMETER BRUKER ADVANCED X-RAY SOLUTIONS SPECTROMETRY SOLUTIONS ARTAX mxrf SPECTROMETER Microanalysis ARTAX Elemental Analysis for the Art Community and More Non-destructive elemental analysis is strictly required

More information

BIFOCAL MINIATURE TOROIDAL SHAPED X-RAY MIRRORS

BIFOCAL MINIATURE TOROIDAL SHAPED X-RAY MIRRORS 1 BIFOCAL MINIATURE TOROIDAL SHAPED X-RAY MIRRORS Sterling Cornaby 1,2, Detlef-M. Smilgies 2, and Donald H. Bilderback 1,2 1 Cornell High Energy Synchrotron Source (CHESS) 2 School of Applied and Engineering

More information

X-rays. X-rays are produced when electrons are accelerated and collide with a target. X-rays are sometimes characterized by the generating voltage

X-rays. X-rays are produced when electrons are accelerated and collide with a target. X-rays are sometimes characterized by the generating voltage X-rays Ouch! 1 X-rays X-rays are produced when electrons are accelerated and collide with a target Bremsstrahlung x-rays Characteristic x-rays X-rays are sometimes characterized by the generating voltage

More information

Zaidi Embong and Husin Wagiran Physics Department, University Of Technology Malaysia, P.O Box 791, 80990, Johor Baharu

Zaidi Embong and Husin Wagiran Physics Department, University Of Technology Malaysia, P.O Box 791, 80990, Johor Baharu MY9800971 Optimization of a Spectrometry for Energy -Dispersive X-ray Fluorescence Analysis by X-ray Tube in Combination with Secondary Target for Multielements Determination of Sediment Samples. Zaidi

More information

Soft X-Ray Silicon Photodiodes with 100% Quantum Efficiency

Soft X-Ray Silicon Photodiodes with 100% Quantum Efficiency PFC/JA-94-4 Soft X-Ray Silicon Photodiodes with 1% Quantum Efficiency K. W. Wenzel, C. K. Li, D. A. Pappas, Raj Kordel MIT Plasma Fusion Center Cambridge, Massachusetts 2139 USA March 1994 t Permanent

More information

ON THE DETECTION LIMIT OF TEY (TOTAL ELECTRON YIELD) Maria F. Ebel, Horst Ebel and Robert Svagera

ON THE DETECTION LIMIT OF TEY (TOTAL ELECTRON YIELD) Maria F. Ebel, Horst Ebel and Robert Svagera Copyright(C)JCPDS-International Centre for Diffraction Data 2, Advances in X-ray Analysis, Vol.42 91 Copyright(C)JCPDS-International Centre for Diffraction Data 2, Advances in X-ray Analysis, Vol.42 91

More information

TOWARDS SUB-100 NM X-RAY MICROSCOPY FOR TOMOGRAPHIC APPLICATIONS

TOWARDS SUB-100 NM X-RAY MICROSCOPY FOR TOMOGRAPHIC APPLICATIONS Copyright -International Centre for Diffraction Data 2010 ISSN 1097-0002 89 TOWARDS SUB-100 NM X-RAY MICROSCOPY FOR TOMOGRAPHIC APPLICATIONS P. Bruyndonckx, A. Sasov, B. Pauwels Skyscan, Kartuizersweg

More information

M4 TORNADO PLUS. Innovation with Integrity. Super Light Element Micro-XRF Spectrometer. Micro-XRF

M4 TORNADO PLUS. Innovation with Integrity. Super Light Element Micro-XRF Spectrometer. Micro-XRF M4 TORNADO PLUS Super Light Element Micro-XRF Spectrometer Innovation with Integrity Micro-XRF M4 TORNADO PLUS - A New Era in Micro-XRF M4 TORNADO PLUS is the world's first Micro-XRF spectrometer that

More information

TOWARDS FAST RECIPROCAL SPACE MAPPING

TOWARDS FAST RECIPROCAL SPACE MAPPING Copyright JCPDS - International Centre for Diffraction Data 2005, Advances in X-ray Analysis, Volume 48. 165 ABSTRACT TOWARDS FAST RECIPROCAL SPACE MAPPING J.F. Woitok and A. Kharchenko PANalytical B.V.,

More information

Mammography is a radiographic procedure specially designed for detecting breast pathology Approximately 1 woman in 8 will develop breast cancer over

Mammography is a radiographic procedure specially designed for detecting breast pathology Approximately 1 woman in 8 will develop breast cancer over Mammography is a radiographic procedure specially designed for detecting breast pathology Approximately 1 woman in 8 will develop breast cancer over a lifetime Breast cancer screening programs rely on

More information

Various beams for RBS at IFIN-HH

Various beams for RBS at IFIN-HH OVERWIEW RBS E sc : function of K and de/dx (mass and depth of target nucleus); Measured spectrum is the sum for all contribution of constitutive elements; Specific Analysis Performed COMPOSITION ANALYSIS:

More information

Copyright -International Centre for Diffraction Data 2010 ISSN

Copyright -International Centre for Diffraction Data 2010 ISSN 234 BRIDGING THE PRICE/PERFORMANCE GAP BETWEEN SILICON DRIFT AND SILICON PIN DIODE DETECTORS Derek Hullinger, Keith Decker, Jerry Smith, Chris Carter Moxtek, Inc. ABSTRACT Use of silicon drift detectors

More information

Preliminary Comparison of Monolithic and Aperture Optics for XRMF. George J. Havrilla CST-8 Christopher G. Worley CST-8

Preliminary Comparison of Monolithic and Aperture Optics for XRMF. George J. Havrilla CST-8 Christopher G. Worley CST-8 Title: Author(s): Submitted to: 1998327 63 LosAlamos NATIONAL LABORATORY Preliminary Comparison of Monolithic and Aperture Optics for XRMF George J. Havrilla CST-8 Christopher G. Worley CST-8 RECEIVED

More information

DESIGN OF SINGLE-BOUNCE MONOCAPILLARY X-RAY OPTICS

DESIGN OF SINGLE-BOUNCE MONOCAPILLARY X-RAY OPTICS 194 DESIGN OF SINGLE-BOUNCE MONOCAPILLARY X-RAY OPTICS Sterling Cornaby 2, Thomas Szebenyi 1, Rong Huang 3 and Donald H Bilderback 1,2 ABSTRACT 1 Cornell High Energy Synchrotron Source (CHESS) 2 School

More information

photolithographic techniques (1). Molybdenum electrodes (50 nm thick) are deposited by

photolithographic techniques (1). Molybdenum electrodes (50 nm thick) are deposited by Supporting online material Materials and Methods Single-walled carbon nanotube (SWNT) devices are fabricated using standard photolithographic techniques (1). Molybdenum electrodes (50 nm thick) are deposited

More information

A PORTABLE X-RAY APPARATUS FOR BOTH STRESS MEASUREMENT AND PHASE ANALYSIS UNDER FIELD CONDITIONS.

A PORTABLE X-RAY APPARATUS FOR BOTH STRESS MEASUREMENT AND PHASE ANALYSIS UNDER FIELD CONDITIONS. Copyright(c)JCPDS-International Centre for Diffraction Data 2000,Advances in X-ray Analysis,Vol.43 66 A PORTABLE X-RAY APPARATUS FOR BOTH STRESS MEASUREMENT AND PHASE ANALYSIS UNDER FIELD CONDITIONS. V.

More information

FAST ELEMENTAL MAPPING WITH MICRO-XRF

FAST ELEMENTAL MAPPING WITH MICRO-XRF 286 FAST ELEMENTAL MAPPING WITH MICRO-XRF Haschke, M.; Rossek, U.; Tagle, R.; Waldschläger, U. Bruker Nano GmbH, 12489 Berlin, Schwarzschildstr.12 ABSTRACT X-Ray optics are now in common use for concentrating

More information

Spectral distribution from end window X-ray tubes

Spectral distribution from end window X-ray tubes Copyright ISSN (C) 1097-0002, JCPDS-International Advances in X-ray Centre Analysis, for Volume Diffraction 41 Data 1999 393 Spectral distribution from end window X-ray tubes N. Broll 1, P. de Chateaubourg

More information

Horiba Jobin-Yvon LabRam Raman Confocal Microscope (GERB 120)

Horiba Jobin-Yvon LabRam Raman Confocal Microscope (GERB 120) Horiba Jobin-Yvon LabRam Raman Confocal Microscope (GERB 120) Please contact Dr. Amanda Henkes for training requests and assistance: 979-862-5959, amandahenkes@tamu.edu Hardware LN 2 FTIR FTIR camera 1

More information

Scanning electron microscope

Scanning electron microscope Scanning electron microscope 6 th CEMM workshop Maja Koblar, Sc. Eng. Physics Outline The basic principle? What is an electron? Parts of the SEM Electron gun Electromagnetic lenses Apertures Chamber and

More information

MC SIMULATION OF SCATTER INTENSITIES IN A CONE-BEAM CT SYSTEM EMPLOYING A 450 kv X-RAY TUBE

MC SIMULATION OF SCATTER INTENSITIES IN A CONE-BEAM CT SYSTEM EMPLOYING A 450 kv X-RAY TUBE MC SIMULATION OF SCATTER INTENSITIES IN A CONE-BEAM CT SYSTEM EMPLOYING A 450 kv X-RAY TUBE A. Miceli ab, R. Thierry a, A. Flisch a, U. Sennhauser a, F. Casali b a Empa - Swiss Federal Laboratories for

More information

STEM Spectrum Imaging Tutorial

STEM Spectrum Imaging Tutorial STEM Spectrum Imaging Tutorial Gatan, Inc. 5933 Coronado Lane, Pleasanton, CA 94588 Tel: (925) 463-0200 Fax: (925) 463-0204 April 2001 Contents 1 Introduction 1.1 What is Spectrum Imaging? 2 Hardware 3

More information

NanoSpective, Inc Progress Drive Suite 137 Orlando, Florida

NanoSpective, Inc Progress Drive Suite 137 Orlando, Florida TEM Techniques Summary The TEM is an analytical instrument in which a thin membrane (typically < 100nm) is placed in the path of an energetic and highly coherent beam of electrons. Typical operating voltages

More information

I. PERFORMANCE OF X-RAY PRODUCTION COMPONENTS FLUOROSCOPIC ACCEPTANCE TESTING: TEST PROCEDURES & PERFORMANCE CRITERIA

I. PERFORMANCE OF X-RAY PRODUCTION COMPONENTS FLUOROSCOPIC ACCEPTANCE TESTING: TEST PROCEDURES & PERFORMANCE CRITERIA FLUOROSCOPIC ACCEPTANCE TESTING: TEST PROCEDURES & PERFORMANCE CRITERIA EDWARD L. NICKOLOFF DEPARTMENT OF RADIOLOGY COLUMBIA UNIVERSITY NEW YORK, NY ACCEPTANCE TESTING GOALS PRIOR TO 1st CLINICAL USAGE

More information

Applications of Steady-state Multichannel Spectroscopy in the Visible and NIR Spectral Region

Applications of Steady-state Multichannel Spectroscopy in the Visible and NIR Spectral Region Feature Article JY Division I nformation Optical Spectroscopy Applications of Steady-state Multichannel Spectroscopy in the Visible and NIR Spectral Region Raymond Pini, Salvatore Atzeni Abstract Multichannel

More information

Spatially Resolved Backscatter Ceilometer

Spatially Resolved Backscatter Ceilometer Spatially Resolved Backscatter Ceilometer Design Team Hiba Fareed, Nicholas Paradiso, Evan Perillo, Michael Tahan Design Advisor Prof. Gregory Kowalski Sponsor, Spectral Sciences Inc. Steve Richstmeier,

More information

Development of X-ray Tool For Critical- Dimension Metrology

Development of X-ray Tool For Critical- Dimension Metrology Development of X-ray Tool For Critical- Dimension Metrology Boris Yokhin, Alexander Krokhmal, Alexander Dikopoltsev, David Berman, Isaac Mazor Jordan Valley Semiconductors Ltd., Ramat Gabriel Ind. Zone,

More information

Diffraction-enhanced X-ray Imaging (DEXI) Medical Solutions. More information using less radiation

Diffraction-enhanced X-ray Imaging (DEXI) Medical Solutions. More information using less radiation Diffraction-enhanced X-ray Imaging (DEXI) Medical Solutions More information using less radiation Medical Small Animal Security NDE/NDT Diffraction-Enhanced X-ray Imaging Medical Solutions Safe non-invasive

More information

Data. microcat +SPECT

Data. microcat +SPECT Data microcat +SPECT microcat at a Glance Designed to meet the throughput, resolution and image quality requirements of academic and pharmaceutical research, the Siemens microcat sets the standard for

More information

AutoMATE II. Micro-area X-ray stress measurement system. Highly accurate micro area residual stress

AutoMATE II. Micro-area X-ray stress measurement system. Highly accurate micro area residual stress AutoMATE II Micro-area X-ray stress measurement system Highly accurate micro area residual stress The accuracy of an R&D diffractom dedicated residua In the past, if you wanted to make highly accurate

More information

A Parallel Radial Mirror Energy Analyzer Attachment for the Scanning Electron Microscope

A Parallel Radial Mirror Energy Analyzer Attachment for the Scanning Electron Microscope 142 doi:10.1017/s1431927615013288 Microscopy Society of America 2015 A Parallel Radial Mirror Energy Analyzer Attachment for the Scanning Electron Microscope Kang Hao Cheong, Weiding Han, Anjam Khursheed

More information

Lab Report XRF 441 Elemental distribution analysis on geological samples with the M4 TORNADO

Lab Report XRF 441 Elemental distribution analysis on geological samples with the M4 TORNADO Bruker Nano Spectrum Geological sample M4 TORNADO Quantification Lab Report XRF 441 Elemental distribution analysis on geological samples with the M4 TORNADO Geological samples are inhomogeneous. The distribution

More information

M6 JETSTREAM. Innovation with Integrity. Large Area Micro X-ray Fluorescence Spectrometer. Micro-XRF

M6 JETSTREAM. Innovation with Integrity. Large Area Micro X-ray Fluorescence Spectrometer. Micro-XRF M6 JETSTREAM Large Area Micro X-ray Fluorescence Spectrometer Innovation with Integrity Micro-XRF Spatially Resolved Elemental Analysis of Large Objects The Bruker M6 JETSTREAM is designed for the nondestructive

More information

Material analysis by infrared mapping: A case study using a multilayer

Material analysis by infrared mapping: A case study using a multilayer Material analysis by infrared mapping: A case study using a multilayer paint sample Application Note Author Dr. Jonah Kirkwood, Dr. John Wilson and Dr. Mustafa Kansiz Agilent Technologies, Inc. Introduction

More information

SUPPLEMENTAL MATERIAL

SUPPLEMENTAL MATERIAL SUPPLEMENTAL MATERIAL 1 - Folios and areas of analysis Figure S1.1. Folio 4, areas of analysis for microxrf ( ), FORS ( ), micro-samples for Raman and FTIR ( ) and Raman in-situ ( ). Figure S1.2. Folio

More information

VISUAL PHYSICS ONLINE DEPTH STUDY: ELECTRON MICROSCOPES

VISUAL PHYSICS ONLINE DEPTH STUDY: ELECTRON MICROSCOPES VISUAL PHYSICS ONLINE DEPTH STUDY: ELECTRON MICROSCOPES Shortly after the experimental confirmation of the wave properties of the electron, it was suggested that the electron could be used to examine objects

More information

Atomic and nuclear physics

Atomic and nuclear physics Atomic and nuclear physics X-ray physics Physics of the atomic shell LEYBOLD Physics Leaflets Investigating the energy spectrum of an x-ray tube as a function of the high voltage and the emission current

More information

SPECIFICATION. Kilovoltage X-ray calibration system for protection and diagnostic level dosimetry. Prepared by

SPECIFICATION. Kilovoltage X-ray calibration system for protection and diagnostic level dosimetry. Prepared by SPECIFICATION Kilovoltage X-ray Prepared by Igor Gomola, Technical Officer, Project ECU6023, Date 2015-Oct-06 Revision Date Status Comments 0.1 2015-Oct-06 Draft Igor Gomola Page 1 of 12 1. Scope This

More information

WITec Alpha 300R Quick Operation Summary October 2018

WITec Alpha 300R Quick Operation Summary October 2018 WITec Alpha 300R Quick Operation Summary October 2018 This document is frequently updated if you feel information should be added, please indicate that to the facility manager (currently Philip Carubia,

More information

Test procedures Page: 1 of 5

Test procedures Page: 1 of 5 Test procedures Page: 1 of 5 1 Scope This part of document establishes uniform requirements for measuring the numerical aperture of optical fibre, thereby assisting in the inspection of fibres and cables

More information

Applications of Optics

Applications of Optics Nicholas J. Giordano www.cengage.com/physics/giordano Chapter 26 Applications of Optics Marilyn Akins, PhD Broome Community College Applications of Optics Many devices are based on the principles of optics

More information

Applications of Micro XRF for the Analysis of Traditional Japanese "Ainu" Glass Beads and other Artifacts

Applications of Micro XRF for the Analysis of Traditional Japanese Ainu Glass Beads and other Artifacts 161 161 Applications of Micro XRF for the Analysis of Traditional Japanese "Ainu" Glass Beads and other Artifacts K.Sugihara 1, M.Satoh 1, Y.Hayakawa 2, A.Saito 3 and T.Sasaki 4 1 Seiko Instruments Inc.,

More information

PROCEEDINGS OF A SYMPOSIUM HELD AT THE CAVENDISH LABORATORY, CAMBRIDGE, Edited by

PROCEEDINGS OF A SYMPOSIUM HELD AT THE CAVENDISH LABORATORY, CAMBRIDGE, Edited by X - R A Y M I C R O S C O P Y A N D M I C R O R A D I O G R A P H Y PROCEEDINGS OF A SYMPOSIUM HELD AT THE CAVENDISH LABORATORY, CAMBRIDGE, 1956 Edited by V. E. COSSLETT Cavendish Laboratory, University

More information

PRECISION APERTURES PRECISION APERTURES PRECISION APERTURES PRECISION APERTURES PRECISION APERTURES PRECISION APERTURES

PRECISION APERTURES PRECISION APERTURES PRECISION APERTURES PRECISION APERTURES PRECISION APERTURES PRECISION APERTURES PRECISION APERTURES PRECISION APERTURES PRECISION APERTURES PRECISION APERTURES PRECISION APERTURES PRECISION APERTURES HIGH PRECISION APERTURES Pinholes with High Edge Definition and Roundness Slits with

More information

X-ray investigation of crystal structures / Laue method with digital X-ray detector (XRIS) (Item No.: P )

X-ray investigation of crystal structures / Laue method with digital X-ray detector (XRIS) (Item No.: P ) X-ray investigation of crystal structures / Laue method with digital X-ray detector (XRIS) (Item No.: P2541602) Curricular Relevance Area of Expertise: Physik Education Level: Hochschule Topic: Moderne

More information

DESIGN OF EDXRF EQUIPMENT FOR THE NONDESTRUCTIVE STUDY OF PRINTS

DESIGN OF EDXRF EQUIPMENT FOR THE NONDESTRUCTIVE STUDY OF PRINTS Copyright JCPDS - International Centre for Diffraction Data 2004, Advances in X-ray Analysis, Volume 47. 70 DESIGN OF EDXRF EQUIPMENT FOR THE NONDESTRUCTIVE STUDY OF PRINTS 1,3 M. Ardid, 1 J.L. Ferrero,

More information

Guide to SPEX Optical Spectrometer

Guide to SPEX Optical Spectrometer Guide to SPEX Optical Spectrometer GENERAL DESCRIPTION A spectrometer is a device for analyzing an input light beam into its constituent wavelengths. The SPEX model 1704 spectrometer covers a range from

More information

Investigation of the Near-field Distribution at Novel Nanometric Aperture Laser

Investigation of the Near-field Distribution at Novel Nanometric Aperture Laser Investigation of the Near-field Distribution at Novel Nanometric Aperture Laser Tiejun Xu, Jia Wang, Liqun Sun, Jiying Xu, Qian Tian Presented at the th International Conference on Electronic Materials

More information

RIETVELD REFINEMENT OF POWDER DATA FROM MULTILAYER OPTICS

RIETVELD REFINEMENT OF POWDER DATA FROM MULTILAYER OPTICS Copyright (c)jcpds-international Centre for Diffraction Data 2002, Advances in X-ray Analysis, Volume 45. 166 RIETVELD REFINEMENT OF POWDER DATA FROM MULTILAYER OPTICS ABSTRACT Scott T. Misture NYS College

More information

MICROFOCUSING X-RAY EQUIPMENT FOR THE LAB DIFFRACTOMETER

MICROFOCUSING X-RAY EQUIPMENT FOR THE LAB DIFFRACTOMETER 29 MICROFOCUSING X-RAY EQUIPMENT FOR THE LAB DIFFRACTOMETER Jörg Wiesmann, 1 Jürgen Graf, 1 Christian Hoffmann, 1 Carsten Michaelsen, 1 Alexandra Oehr, 1 Uwe Preckwinkel, 2 Ning Yang, 2 Holger Cordes,

More information

attocfm I for Surface Quality Inspection NANOSCOPY APPLICATION NOTE M01 RELATED PRODUCTS G

attocfm I for Surface Quality Inspection NANOSCOPY APPLICATION NOTE M01 RELATED PRODUCTS G APPLICATION NOTE M01 attocfm I for Surface Quality Inspection Confocal microscopes work by scanning a tiny light spot on a sample and by measuring the scattered light in the illuminated volume. First,

More information

Fabrication of Probes for High Resolution Optical Microscopy

Fabrication of Probes for High Resolution Optical Microscopy Fabrication of Probes for High Resolution Optical Microscopy Physics 564 Applied Optics Professor Andrès La Rosa David Logan May 27, 2010 Abstract Near Field Scanning Optical Microscopy (NSOM) is a technique

More information

AgilOptics mirrors increase coupling efficiency into a 4 µm diameter fiber by 750%.

AgilOptics mirrors increase coupling efficiency into a 4 µm diameter fiber by 750%. Application Note AN004: Fiber Coupling Improvement Introduction AgilOptics mirrors increase coupling efficiency into a 4 µm diameter fiber by 750%. Industrial lasers used for cutting, welding, drilling,

More information

Polycapillary optic source combinations for protein crystallography

Polycapillary optic source combinations for protein crystallography Journal of Applied Crystallography ISSN 0021-8898 Polycapillary optic source combinations for protein crystallography F. A. Hofmann, W. M. Gibson, C. A. MacDonald, D. A. Carter, J. X. Ho and J. R. Ruble

More information

DUANE-HUNT RELATION AND DETERMINATION OF PLANCK S CONSTANT

DUANE-HUNT RELATION AND DETERMINATION OF PLANCK S CONSTANT DUANE-HUNT RELATION AND DETERMINATION OF PLANCK S CONSTANT OBJECTIVES To determine the limit wavelength min of the bremsstrahlung continuum as a function of the high voltage U of the x-ray tube. To confirm

More information

Solid-state physics. Bragg reflection: determining the lattice constants of monocrystals. LEYBOLD Physics Leaflets P

Solid-state physics. Bragg reflection: determining the lattice constants of monocrystals. LEYBOLD Physics Leaflets P Solid-state physics Properties of crystals X-ray structural analysis LEYBOLD Physics Leaflets Bragg reflection: determining the lattice constants of monocrystals P7.1.2.1 Objects of the experiment Investigating

More information

Damage-free failure/defect analysis in electronics and semiconductor industries using micro-atr FTIR imaging

Damage-free failure/defect analysis in electronics and semiconductor industries using micro-atr FTIR imaging Damage-free failure/defect analysis in electronics and semiconductor industries using micro-atr FTIR imaging Application note Electronics and Semiconductor Authors Dr. Mustafa Kansiz and Dr. Kevin Grant

More information

Fiber Optic Communications

Fiber Optic Communications Fiber Optic Communications ( Chapter 2: Optics Review ) presented by Prof. Kwang-Chun Ho 1 Section 2.4: Numerical Aperture Consider an optical receiver: where the diameter of photodetector surface area

More information

13. Optical Instruments*

13. Optical Instruments* 13. Optical Instruments* Objective: Here what you have been learning about thin lenses is applied to make a telescope. In the process you encounter general optical instrument design concepts. The learning

More information

FIRST Newsletter March 2013, Issue 20. Elemental Distribution Analysis of a Meteorite Sample from the Rochechouart Structure with the µ-xrf M4 TORNADO

FIRST Newsletter March 2013, Issue 20. Elemental Distribution Analysis of a Meteorite Sample from the Rochechouart Structure with the µ-xrf M4 TORNADO FIRST Newsletter March 2013, Issue 20 Elemental Distribution Analysis of a Meteorite Sample from the Rochechouart Structure with the µ-xrf M4 TORNADO By Dr. Roald Tagle, Ulrich Waldschlager, Dr. Michael

More information

Introduction. Chapter 16 Diagnostic Radiology. Primary radiological image. Primary radiological image

Introduction. Chapter 16 Diagnostic Radiology. Primary radiological image. Primary radiological image Introduction Chapter 16 Diagnostic Radiology Radiation Dosimetry I Text: H.E Johns and J.R. Cunningham, The physics of radiology, 4 th ed. http://www.utoledo.edu/med/depts/radther In diagnostic radiology

More information

Leica TCS SP8 Quick Start Guide

Leica TCS SP8 Quick Start Guide Leica TCS SP8 Quick Start Guide Leica TCS SP8 System Overview Start-Up Procedure 1. Turn on the CTR Control Box, Fluorescent Light for the microscope stand. 2. Turn on the Scanner Power (1) on the front

More information

Update on NIST X-Ray Air-Kerma Standards and Calibrations

Update on NIST X-Ray Air-Kerma Standards and Calibrations CCRI(I)/03-19 Update on NIST X-Ray Air-Kerma Standards and Calibrations Michelle O Brien and Stephen M. Seltzer Ionizing Radiation Division National Institute of Standards and Technology Gaithersburg,

More information

Exp No.(8) Fourier optics Optical filtering

Exp No.(8) Fourier optics Optical filtering Exp No.(8) Fourier optics Optical filtering Fig. 1a: Experimental set-up for Fourier optics (4f set-up). Related topics: Fourier transforms, lenses, Fraunhofer diffraction, index of refraction, Huygens

More information

Devices & Services Company

Devices & Services Company Devices & Services Company 10290 Monroe Drive, Suite 202 - Dallas, Texas 75229 USA - Tel. 214-902-8337 - Fax 214-902-8303 Web: www.devicesandservices.com Email: sales@devicesandservices.com D&S Technical

More information

PANalytical X pert Pro Gazing Incidence X-ray Reflectivity User Manual (Version: )

PANalytical X pert Pro Gazing Incidence X-ray Reflectivity User Manual (Version: ) University of Minnesota College of Science and Engineering Characterization Facility PANalytical X pert Pro Gazing Incidence X-ray Reflectivity User Manual (Version: 2012.10.17) The following instructions

More information

erosita mirror calibration:

erosita mirror calibration: erosita mirror calibration: First measurements and future concept PANTER instrument chamber set-up for XMM mirror calibration: 12 m length, 3.5 m diameter: 8m to focal plane instrumentation now: f = 1.6

More information

Partial Replication of Storms/Scanlan Glow Discharge Radiation

Partial Replication of Storms/Scanlan Glow Discharge Radiation Partial Replication of Storms/Scanlan Glow Discharge Radiation Rick Cantwell and Matt McConnell Coolescence, LLC March 2008 Introduction The Storms/Scanlan paper 1 presented at the 8 th international workshop

More information

A high energy gamma camera using a multiple hole collimator

A high energy gamma camera using a multiple hole collimator ELSEVIER Nuclear Instruments and Methods in Physics Research A 353 (1994) 328-333 A high energy gamma camera using a multiple hole collimator and PSPMT SV Guru *, Z He, JC Ferreria, DK Wehe, G F Knoll

More information

IDENTIFICATION OF PAINTING MATERIALS USED FOR MURAL PAINTINGS BY IMAGE ANALYSIS AND XRF

IDENTIFICATION OF PAINTING MATERIALS USED FOR MURAL PAINTINGS BY IMAGE ANALYSIS AND XRF IDENTIFICATION OF PAINTING MATERIALS USED FOR MURAL PAINTINGS BY IMAGE ANALYSIS AND XRF 213 Seiji SHIRONO, Yasuhiro HAYAKAWA National Research Institute for Cultural Properties, Tokyo, Japan ABSTRACT The

More information

Water-Window Microscope Based on Nitrogen Plasma Capillary Discharge Source

Water-Window Microscope Based on Nitrogen Plasma Capillary Discharge Source 2015 International Workshop on EUV and Soft X-Ray Sources Water-Window Microscope Based on Nitrogen Plasma Capillary Discharge Source T. Parkman 1, M. F. Nawaz 2, M. Nevrkla 2, M. Vrbova 1, A. Jancarek

More information

Improving the Collection Efficiency of Raman Scattering

Improving the Collection Efficiency of Raman Scattering PERFORMANCE Unparalleled signal-to-noise ratio with diffraction-limited spectral and imaging resolution Deep-cooled CCD with excelon sensor technology Aberration-free optical design for uniform high resolution

More information

Full-screen mode Popup controls. Overview of the microscope user interface, TEM User Interface and TIA on the left and EDS on the right

Full-screen mode Popup controls. Overview of the microscope user interface, TEM User Interface and TIA on the left and EDS on the right Quick Guide to Operating FEI Titan Themis G2 200 (S)TEM: TEM mode Susheng Tan Nanoscale Fabrication and Characterization Facility, University of Pittsburgh Office: M104/B01 Benedum Hall, 412-383-5978,

More information

ADVANCED OPTICS LAB -ECEN Basic Skills Lab

ADVANCED OPTICS LAB -ECEN Basic Skills Lab ADVANCED OPTICS LAB -ECEN 5606 Basic Skills Lab Dr. Steve Cundiff and Edward McKenna, 1/15/04 Revised KW 1/15/06, 1/8/10 Revised CC and RZ 01/17/14 The goal of this lab is to provide you with practice

More information

Operation of a Single Pass, Bunch-by-bunch x-ray Beam Size Monitor for the CESR Test Accelerator Research Program. October 3, 2012

Operation of a Single Pass, Bunch-by-bunch x-ray Beam Size Monitor for the CESR Test Accelerator Research Program. October 3, 2012 Operation of a Single Pass, Bunch-by-bunch x-ray Beam Size Monitor for the CESR Test Accelerator Research Program October 3, 2012 Goals Goals For This Presentation: 1.Provide an overview of the efforts

More information

DESIGN AND MEASUREMENT WITH A NEW PORTABLE X-RAY CAMERA FOR FULL-FIELD FLUORESCENCE IMAGING

DESIGN AND MEASUREMENT WITH A NEW PORTABLE X-RAY CAMERA FOR FULL-FIELD FLUORESCENCE IMAGING 14 DESIGN AND MEASUREMENT WITH A NEW PORTABLE X-RAY CAMERA FOR FULL-FIELD FLUORESCENCE IMAGING I. Ordavo 1,2, A. Bjeoumikhov 3, S. Bjeoumikhova 3, G. Buzanich 4, R. Gubzhokov 4, R. Hartmann 1, S. Ihle

More information

J.A. Casey and J.H. Irby. M.I.T. Plasma Fusion Center

J.A. Casey and J.H. Irby. M.I.T. Plasma Fusion Center March 27, 1986 PFC/JA-86-16 Thomson Scattering in the Tara Tandem Mirror Central Cell J.A. Casey and J.H. Irby M.I.T. Plasma Fusion Center I ABSTRACT: A Thomson Scattering experiment is under construction

More information