DENTAL RADIOGRAPHY KAMARUL AMIN BIN ABU BAKAR
|
|
- Kimberly Lang
- 6 years ago
- Views:
Transcription
1 DENTAL RADIOGRAPHY KAMARUL AMIN BIN ABU BAKAR
2 Components of the Dental X-Ray Machine Dental x-ray machines may vary somewhat in size and appearance, but all machines will have three primary components: The tubehead An extension arm The control panel
3 Fig Dental x-ray machine and arm. A, Tubehead. B, Position-indicator device (PID). C, Extension arm.
4 The Tubehead The x-ray tubehead is tightly sealed; heavy metal housing contains the x-ray tube that produces dental x-rays.
5 Fig Diagram of the dental x-ray tubehead. (From Haring J, Jansen L: Dental radiography: principles and techniques, ed 2, Philadelphia, 2000, Saunders.)
6 Fig X-ray tube. (Courtesy of Xintek Inc.)
7 Components of the Tubehead The metal body of the tubehead that houses the x-ray tube is called the metal housing. It is filled with insulating oil. Insulating oil surrounds the x-ray tube and prevents overheating by absorbing the heat created in the production of x-rays. Tubehead seal is made of leaded glass or aluminum. It seals the oil in the tubehead. The x-ray tube is the heart of the x-ray generating unit. (Cont d)
8 Components of the Tubehead (Cont d) The transformer alters the voltage of incoming electric current. The aluminum filter is aluminum sheets 0.5 mm thick. The lead collimator is a metal disc with a small opening in the center to control the size and shape of the x-ray beam as it leaves the tubehead. The PID is the open-ended, lead-lined cylinder that extends from the opening of the metal housing of the tubehead. It is used to aim the x-ray beam.
9 The X-Ray Tube The vacuum environment allows the electrons to flow with minimum resistance between the electrodes: Cathode Anode
10 The Cathode The cathode consists of a tungsten filament in a focusing cup made of molybdenum. The purpose of the cathode is to supply the electrons necessary to generate x-rays. Electrons are generated in the x-ray tube at the cathode. The hotter the filament becomes, the more electrons are produced.
11 The Anode The anode is the target for the electrons. It is composed of a tungsten target (a small block of tungsten) that is embedded in the larger copper stem. The copper around the target conducts heat away from the target, reducing wear and tear on the target. The purpose of the tungsten target is to serve as a focal spot and convert the bombarding electrons into x-ray photons. The x-rays at the center of this beam are known as the central ray.
12 Fig The production of dental x-rays occurs in the x-ray tube. (From Haring J, Jansen L: Dental radiography: principles and techniques, ed 2, Philadelphia, 2000, Saunders.)
13 The PID The open end of the lead-lined PID is placed against the patient s face during film exposure. The PID may be cylindrical or rectangular. The rectangular PID limits the size of the beam to that of a dental film. PIDs used in dentistry are usually 8, 12, or 16 inches long. The choice of length is determined by the radiographic technique being used. The long (12- to 16-inch) PID is more effective in reducing exposure to the patient than a short (8-inch) PID because there is less divergence (separation) of the beam.
14 The Extension Arm The extension arm encloses the wire between the tubehead and the control panel. It also has an important function in positioning the tubehead. The extension arm folds up and can be swiveled from side to side. If the extension arm is left in an extended position when the machine is not in use, the weight of the tubehead can cause it to become loose, and the tubehead will drift (slip out of position) after it is positioned for an exposure.
15 The Control Panel The control panel of an x-ray unit contains: The master switch Indicator light Exposure button Indicator light Control devices (time, milliamperage [ma] selector, and kilovoltage [kvp] selector) A single centrally located control panel may be used to operate several tubeheads located in separate treatment rooms.
16
17
18
19
20
21
22 End of Chapter
X-rays. X-rays are produced when electrons are accelerated and collide with a target. X-rays are sometimes characterized by the generating voltage
X-rays Ouch! 1 X-rays X-rays are produced when electrons are accelerated and collide with a target Bremsstrahlung x-rays Characteristic x-rays X-rays are sometimes characterized by the generating voltage
More informationLECTURE 1 The Radiographic Image
LECTURE 1 The Radiographic Image Prepared by:- KAMARUL AMIN ABDULLAH @ ABU BAKAR UiTM Faculty of Health Sciences Medical Imaging Department 11/23/2011 KAMARUL AMIN (C) 1 Lesson Objectives At the end of
More informationP R E S E N T E D B Y. K A M A R U L A M I N A B D U L L A H Dip. MED. IMG., BSc. MED. IMG. (UiTM)
+ - P R E S E N T E D B Y K A M A R U L A M I N A B D U L L A H Dip. MED. IMG., BSc. MED. IMG. (UiTM) 1 I N T R O D U C T I O N : An x-ray generator is a device that Supplies electrical power to x-ray
More information1-1. GENERAL 1-2. DISCOVERY OF X-RAYS
1-1. GENERAL Radiography is a highly technical field, indispensable to the modern dental practice, but presenting many potential hazards. The dental radiographic specialist must be thoroughly familiar
More informationVeterinary Science Preparatory Training for the Veterinary Assistant. Floron C. Faries, Jr., DVM, MS
Veterinary Science Preparatory Training for the Veterinary Assistant Floron C. Faries, Jr., DVM, MS Radiology Floron C. Faries, Jr., DVM, MS Objectives Determine the appropriate machine settings for making
More informationBeam-Restricting Devices
Beam-Restricting Devices Three factors contribute to an increase in scatter radiation: Increased kvp Increased Field Size Increased Patient or Body Part Size. X-ray Interactions a some interact with the
More information- KiloVoltage. Technique 101: Getting Back to Basics
Why do I need to know technique? Technique 101: Getting Back to Basics Presented by: Thomas G. Sandridge, M.S., M.Ed., R.T.(R) Program Director Northwestern Memorial Hospital School of Radiography Chicago,
More informationMammography is a radiographic procedure specially designed for detecting breast pathology Approximately 1 woman in 8 will develop breast cancer over
Mammography is a radiographic procedure specially designed for detecting breast pathology Approximately 1 woman in 8 will develop breast cancer over a lifetime Breast cancer screening programs rely on
More informationNuclear Associates
Nuclear Associates 07-591 Focal Spot Test Tool Users Manual February 2005 Manual No. 07-591-1 Rev. 2 2004, 2005 Fluke Corporation, All rights reserved. Printed in U.S.A. All product names are trademarks
More informationUsing Carbon Nano-Tube Field Emitters to Miniaturize X-Ray Tubes
Using Carbon Nano-Tube Field Emitters to Miniaturize X-Ray Tubes Authors: Martin Pesce, RT(R), Xiaohui Wang, PhD, Peter Rowland X-rays are produced by the impact of an accelerated electron beam on a tungsten
More informationCHAPTER 6 QC Test For Fluoroscopic Equipment. Prepared by:- Kamarul Amin bin Abu Bakar School of Medical Imaging KLMUC
CHAPTER 6 QC Test For Fluoroscopic Equipment Prepared by:- Kamarul Amin bin Abdullah @ Abu Bakar School of Medical Imaging KLMUC Lesson Outcomes Describe the objectives of each QC test done. Identify QC
More informationMultiple Choice Identify the letter of the choice that best completes the statement or answers the question.
RA110 test 3 Multiple Choice Identify the letter of the choice that best completes the statement or answers the question. 1. An object 35 cm in width is radiographed at 100 cm SID and at a 50 cm SOD. What
More informationX-RAY IMAGING EE 472 F2017. Prof. Yasser Mostafa Kadah
X-RAY IMAGING EE 472 F2017 Prof. Yasser Mostafa Kadah www.k-space.org Recommended Textbook Stewart C. Bushong, Radiologic Science for Technologists: Physics, Biology, and Protection, 10 th ed., Mosby,
More informationX-ray Tube and Generator Basic principles and construction
X-ray Tube and Generator Basic principles and construction Dr Slavik Tabakov - Production of X-rays and Patient Dose OBJECTIVES - X-ray tube construction - Anode - types, efficiency - Classical X-ray generator
More informationCHAPTER 8 HDR102 PHYSICS FOR RADIOGRAPHERS 1 RECTIFICATION MR KAMARUL AMIN BIN ABDULLAH SCHOOL OF MEDICAL IMAGING FACULTY OF HEALTH SCIENCES
HDR102 PHYSICS FOR RADIOGRAPHERS 1 CHAPTER 8 RECTIFICATION PREPARED BY: MR KAMARUL AMIN BIN ABDULLAH SCHOOL OF MEDICAL IMAGING FACULTY OF HEALTH SCIENCES LEARNING OUTCOMES At the end of the lesson, the
More informationCIRCLEX 0.3/0.8P324&0.6/1.2P324DK-85
PD53-012 p Rotating Anode X-ray tube Assembly 0.3/0.8P32&0.6/1.2P32DK-85 GENERAL The Shimadzu 0.3/0.8P32DK-85 & 0.6/1.2P32DK-85, Rotating Anode X-ray tube assemblies are rated to 150kV and feature a 100mm
More informationCourse Instructions: Check your for your CE certification of completion (please check your junk/spam folder as well). About SMS CE courses:
2017 Course #4 Self-Study Course Contact Us: Phone 614-292-6737 Toll Free 1-888-476-7678 Fax 614-292-8752 E-mail smsosu@osu.edu Web dentistry.osu.edu/sms The Ohio State University College of Dentistry
More informationVersion 1.0. TechnicVR. Student Guide
Version 1.0 TechnicVR s h a d e r w a r e. c o m Student Guide TechnicVR s h a d e r w a r e. c o m Student Guide shaderware 2008 PO Box 103 Saltburn Cleveland TS12 1WP w w w. s h a d e r w a r e. c o
More informationExposure System Selection
Principles of Imaging Science II (RAD120) Exposure Systems Exposure System Selection Radiographic exposure is a very complex process Best technique systems manipulate one variable while holding others
More informationContrast. Contrast: the difference in density on adjacent areas of a radiograph or other image receptor. Subjective. Long Scale (Low Contrast)
Contrast Contrast: the difference in density on adjacent areas of a radiograph or other image receptor. Subject Subjective Radiographic Long Scale (Low Contrast) Short Scale (High Contrast) Factors affecting
More informationX-RAY. Lecture No.4. Image Characteristics:
Lecture No.4 X-RAY أ.م.د. اسامة مراد ابراهيم Image Characteristics: *Radiographic density: It s the degree of blackness of the film. when a film is exposed by an x-ray beam (or by light in case of screenfilm
More informationSTUDENT REVIEW QUESTION SET K CR/DR CONTENT AREA
STUDENT REVIEW QUESTION SET K CR/DR CONTENT AREA RADT 2913 COMPREHENSIVE REVIEW 1 The CR cassette is backed by aluminum that: A. reflects x-rays B. absorbs x-rays C. captures the image D. transmits x-rays
More informationX-RAYS - NO UNAUTHORISED ENTRY
Licencing of premises Premises Refer Guidelines A radiation warning sign and warning notice, X-RAYS - NO UNAUTHORISED ENTRY must be displayed at all entrances leading to the rooms where x-ray units are
More informationTechnical data CAMARGUE CS-VH50/300. VARIABLE Height Bucky Table With Ceiling Suspension
Technical data VARIABLE Height Bucky Table With Ceiling Suspension Model Variations CAMARGUE FH (Fixed Height) CAMARGUE FH Tomo CAMARGUE FH Ceiling suspension CAMARGUE VH (Variable Height) CAMARGUE VH
More informationRADIOGRAPHIC EXPOSURE
RADIOGRAPHIC EXPOSURE Receptor Exposure Receptor Exposure the that interacts with the receptor. Computed Radiography ( ) requires a. Direct Digital Radiography (DR) requires a. Exposure Indicators Exposure
More informationX-ray Tube and Generator Basic principles and construction
X-ray Tube and Generator Basic principles and construction Dr Slavik Tabakov - Production of X-rays OBJECTIVES - X-ray tube construction - Anode - types, efficiency - X-ray tube working characteristics
More informationNuclear Associates
Nuclear Associates 07-649 CDRH Fluoroscopic Phantom Users Manual March 2005 Manual No. 07-649-1 Rev. 2 2004, 2005 Fluke Corporation, All rights reserved. Printed in U.S.A. All product names are trademarks
More information1. Carlton, Richard R., and Arlene M. Adler. Principles of Radiographic Imaging: An Art and a Science, 5th edition (2013).
CODE: RADT 151 INSTITUTE: Health Science TITLE: Radiographic Exposure DEPARTMENT: Radiologic Technology COURSE DESCRIPTION: This course covers the principles of radiographic exposure selection and manipulation
More informationProduct Information PI
Product Information PI TOSHIBA X-RAY TUBE D-045 Stationary Anode X-ray Tube Especially designed for dental radiography unit. This tube has focus 0.4, and is available for maximum tube voltage 70 kv. Installed
More informationProX Intraoral X-ray. PLANMECA is proud to introduce a new intraoral X-ray unit to its comprehensive collection of imaging products- the ProX.
The premium intraoral X-ray unit... ProX Intraoral X-ray PLANMECA is proud to introduce a new intraoral X-ray unit to its comprehensive collection of imaging products- the ProX. This advanced unit provides
More informationSPRINGFIELD TECHNICAL COMMUNITY COLLEGE ACADEMIC AFFAIRS
SPRINGFIELD TECHNICAL COMMUNITY COLLEGE ACADEMIC AFFAIRS Course Number: RADG 212 Department: Radiography Course Title: Equip. Operation & Maint. Semester: Spring Year: 1997 Objectives/ Unit One: The X-ray
More informationOPERATOR'S INSTRUCTIONS
MODEL 096 DENTAL X-RAY OPERATOR'S INSTRUCTIONS (for USA & Canada) Table of contents Page Layout of Control Box... 1 Function of Controls... 2 Operating Procedures... 4 Error Codes... 5 Maintenance... 6
More informationPD233: Design of Biomedical Devices and Systems
PD233: Design of Biomedical Devices and Systems (Lecture-8 Medical Imaging Systems) (Imaging Systems Basics, X-ray and CT) Dr. Manish Arora CPDM, IISc Course Website: http://cpdm.iisc.ac.in/utsaah/courses/
More informationIntroduction: Why electrons?
Introduction: Why electrons? 1 Radiations Visible light X-rays Electrons Neutrons Advantages Not very damaging Easily focused Eye wonderful detector Small wavelength (Angstroms) Good penetration Small
More informationABOVETABLE X-RAY SOURCE FLUOROSCOPIC AND SPOT-FILM SYSTEMS
PART VI ABOVETABLE X-RAY SOURCE FLUOROSCOPIC AND SPOT-FILM SYSTEMS FORM FDA 3069 REPRINTED April 2000 ROUTINE COMPLIANCE TESTING ABOVETABLE X-RAY SOURCE FLUOROSCOPIC AND SPOT-FILM SYSTEMS (Test Procedure
More informationRotating Anode X-ray Tube
Datasheet RADII KL74-1.0/2.0-125 Rotating Anode X-ray Tube Rotating anode X-ray tube for the purpose of general diagnostic X-ray procedures. Specially processed Tungsten faced molybdenum target of 74 mm
More informationX-ray Imaging. PHYS Lecture. Carlos Vinhais. Departamento de Física Instituto Superior de Engenharia do Porto
X-ray Imaging PHYS Lecture Carlos Vinhais Departamento de Física Instituto Superior de Engenharia do Porto cav@isep.ipp.pt Overview Projection Radiography Anode Angle Focal Spot Magnification Blurring
More informationARU-01CF Portable X-Ray NSN:
ARU-01CF Portable X-Ray NSN: 6525-01-468-1672 COMMAND l AIR OPERATION & MAINTENANCE MANUAL Table of Contents Introduction................................................................3 Purpose.............................................................3
More informationDose Reduction and Image Preservation After the Introduction of a 0.1 mm Cu Filter into the LODOX Statscan unit above 110 kvp
Dose Reduction and Image Preservation After the Introduction of a into the LODOX Statscan unit above 110 kvp Abstract: CJ Trauernicht 1, C Rall 1, T Perks 2, G Maree 1, E Hering 1, S Steiner 3 1) Division
More informationDistributed by. Ecotron Anyview Series. The Intelligent C-Arm System
Distributed by Ecotron Anyview Series The Intelligent C-Arm System The Ecotron Anyview Series is a new standard for C-arm imaging Excellent radiographic imaging at an economical price Advanced X-ray Generator
More informationPrinciple of X-Ray Systems
Principle of X-Ray Systems Hossein Ebrahimi Nasab PHYSICS OF X-RAYS Nature of X-rays Energy unit Interaction with matter INTERACTION WITH THE MATTER In vacuum: photon move along a straight line In materials,
More informationProduct Information PI
Rotating Anode X-ray Tube Assembly Product Information PI ROTANODE The TOSHIBA XH-183 housing is developed for use in surgical C-arm application. It provides ease for use at a nominal patient table height.
More informationRadiology. Radiograph: Is the image of an object made with use of X- ray instead of light.
Radiology د. اريج Lec. 3 X Ray Films Radiograph: Is the image of an object made with use of X- ray instead of light. Dental x- ray film: Is a recording media on which image of the object was made by exposing
More informationStationary Anode X-Ray Tube. General Data
Technical Data TD TOSHIBA X-RAY TUBE D-041 D-041S D-041SB Stationary Anode X-Ray Tube Especially designed for dental radiography unit. Provided with an insulation cylinder (D-041S) and lead cylinder. (D-041SB)
More informationSPECIFICATION. Kilovoltage X-ray calibration system for protection and diagnostic level dosimetry. Prepared by
SPECIFICATION Kilovoltage X-ray Prepared by Igor Gomola, Technical Officer, Project ECU6023, Date 2015-Oct-06 Revision Date Status Comments 0.1 2015-Oct-06 Draft Igor Gomola Page 1 of 12 1. Scope This
More informationSeminar 8. Radiology S8 1
Seminar 8 Radiology Medical imaging. X-ray image formation. Energizing and controlling the X-ray tube. Image detectors. The acquisition of analog and digital images. Digital image processing. Selected
More informationRadiographic Testing (RT) [10]
Radiographic Testing (RT) [10] Definition: An NDT method that utilizes x-rays or gamma radiation to detect discontinuities in materials, and to present their images on recording medium. 1> Electromagnetic
More informationHIGH FREQUENCY X-RAY MONOBLOCK 3.5kW
HIGH FREQUENCY X-RAY MONOBLOCK 3.5kW The high frequency X-Ray monoblock consists of an inverter and oil-filled tank which houses the high voltage transformer and tube. It has been designed and developed
More informationRULES OF TENNESSEE DEPARTMENT OF ENVIRONMENT AND CONSERVATION DIVISION OF RADIOLOGICAL HEALTH CHAPTER USE OF X-RAY APPARATUS
RULES OF TENNESSEE DEPARTMENT OF ENVIRONMENT AND CONSERVATION DIVISION OF RADIOLOGICAL HEALTH CHAPTER 0400-20-06 USE OF X-RAY APPARATUS TABLE OF CONTENTS 0400-20-06-.01 Purpose 0400-20-06-.06 Veterinary
More information2014 CINDE Toronto Portable X-ray & X-Ray Production
2014 CINDE Toronto Portable X-ray & X-Ray Production Nils Hase, Business Development Manager Portable X-Ray business unit Copenhagen and USA Technology with Passion 3 COMET Headquarters Flamatt Switzerland
More informationNuclear Associates
Nuclear Associates 07-644 Grid Alignment Test Tool Users Manual March 2005 Manual No. 07-644-1 Rev. 2 2004, 2005 Fluke Corporation, All rights reserved. Printed in U.S.A. All product names are trademarks
More informationHIGH FREQUENCY X-RAY MONOBLOCK 3.5kW
HIGH FREQUENCY X-RAY MONOBLOCK 3.5kW The high frequency X-Ray monoblock consists of an inverter and oil-filled tank which houses the high voltage transformer and tube. It has been designed and developed
More information(12) Patent Application Publication (10) Pub. No.: US 2017/ A1
(19) United States US 201701 35653A1 (12) Patent Application Publication (10) Pub. No.: US 2017/0135653 A1 Ren et al. (43) Pub. Date: May 18, 2017 (54) TOMOSYNTHESIS WITH SHIFTING FOCAL SPOT AND OSCILLATING
More informationThe X-ray circuit: part II
The X-ray circuit: part II By Dr. Mohsen Dashti 357 Radiologic Processing & Procedure Lecture notes #2 Key issues Types of x-ray equipment. Power for x-ray generator. A basic x-ray circuit. Generators.
More informationCh. 223 VETERINARY MEDICINE CHAPTER 223. VETERINARY MEDICINE GENERAL PROVISIONS X-RAYS RADIOACTIVE MATERIAL. Authority
Ch. 223 VETERINARY MEDICINE 25 223.1 CHAPTER 223. VETERINARY MEDICINE Sec. 223.1. Purpose and scope. 223.2. [Reserved]. 223.2a. Definitions. 223.3 223.6. [Reserved]. 223.7. Structural shielding. 223.8.
More informationIntraoral X-Ray For Film and Digital Systems
Intraoral X-Ray For Film and Digital Systems ImageWorks Generations of Imaging With the full line of Endos intraoral X-ray units, everything has been designed to maximize operating efficiency while allowing
More informationMaxRay Handheld X-ray Systems Operator Training Exam
MaxRay Handheld X-ray Systems Operator Training Exam Employee: Instructor: ate: Score: Instructions Read each question carefully and choose the best answer. 1) LR is 2) 3) 4) a. a safety principle meant
More informationDedicated Veterinary Imaging Solutions Digital, CR and Analog Imaging Solutions for any size patient and any size budget.
by Dedicated Veterinary Imaging Solutions Digital, CR and Analog Imaging Solutions for any size patient and any size budget. Serving the Veterinary Profession for Over 75 Years. ... We See Things Differently
More informationRADIOGRAPHY TERMS TO KNOW SELF STUDY DENTALELLE TUTORING
RADIOGRAPHY TERMS TO KNOW SELF STUDY DENTALELLE TUTORING PLEASE NOTE You DO NOT need to study these for the board exam if this is why you bought our Radiography course, however if you come across any terms
More informationRotating Anode X-Ray Tube Housing Assembly. General Data
Display Devices & Components Company TECHNICAL DATA ROTANODE TM E7252X Rotating Anode X-Ray Tube Housing Assembly! High speed rotating anode X-ray tube housing assembly for high energy radiographic and
More informationMXHF-1500RF is controlled by Digital key panel console that displays KV, ma and mas with APR menu programmed.
R/F TV X-RAY SYSTEM DIAGNOSTIC RADIOGRAPHIC FLUOROSCOPIC TV SYSTEM MXHF-1500RF SYSTEM OUTLINE Product Data No. 041021-01 MXHF-1500RF is controlled by Digital key panel console that displays KV, ma and
More informationSpectrophotometer. An instrument used to make absorbance, transmittance or emission measurements is known as a spectrophotometer :
Spectrophotometer An instrument used to make absorbance, transmittance or emission measurements is known as a spectrophotometer : Spectrophotometer components Excitation sources Deuterium Lamp Tungsten
More informationDigital Imaging Considerations Computed Radiography
Digital Imaging Considerations Digital Radiography Computed Radiography o Cassette based Direct or Indirect Digital Radiography o Cassetteless Computed Radiography 1 CR Image Acquisition Most like conventional
More informationX-RAY FLUOROSCOPY IMAGING SYSTEMS. Dr Slavik Tabakov. Luminescence: Dept. Medical Eng. & Physics King s College London
X-RAY FLUOROSCOPY IMAGING SYSTEMS Dr Slavik Tabakov OBJECTIVES - Image Intensifier construction - Input window - Accelerating and focusing electrodes - Output window - Conversion factor - II characteristics
More informationCONTROL METHODS FOR THE FIELD EMISSION X-RAY SPECTRA AND THEIR APPLICATIONS
CONTROL METHODS FOR THE FIELD EMISSION X-RAY SPECTRA AND THEIR APPLICATIONS E. Sato, H. Isobe, T. Yanagisawa To cite this version: E. Sato, H. Isobe, T. Yanagisawa. CONTROL METHODS FOR THE FIELD EMISSION
More informationI. PERFORMANCE OF X-RAY PRODUCTION COMPONENTS FLUOROSCOPIC ACCEPTANCE TESTING: TEST PROCEDURES & PERFORMANCE CRITERIA
FLUOROSCOPIC ACCEPTANCE TESTING: TEST PROCEDURES & PERFORMANCE CRITERIA EDWARD L. NICKOLOFF DEPARTMENT OF RADIOLOGY COLUMBIA UNIVERSITY NEW YORK, NY ACCEPTANCE TESTING GOALS PRIOR TO 1st CLINICAL USAGE
More informationDigital Panoramic X- Ray unit. Product Data
Digital Panoramic X- Ray unit Product Data Standard examination programs Standard panoramic: adult/child panoramic exam TMJ open/close mouth: 4 slices are taken in the same image (left/right condyle, open/close
More informationPractix. Mobile Radiography. Constantly on the move, and yet never tired. Practix 160
Practix 160 Specifications Practix Mobile Radiography Constantly on the move, and yet never tired Practix 160 The mobile X-ray system Practix 160 offers ideal solutions for all routine examinations in
More informationDistributed source x-ray tube technology for tomosynthesis imaging
Distributed source x-ray tube technology for tomosynthesis imaging Authors: F. Sprenger a*, X. Calderon-Colon b, Y. Cheng a, K. Englestad a, J. Lu b, J. Maltz c, A. Paidi c, X. Qian b, D. Spronk a, S.
More information(CR) PORTABLE CHEST EXAM Baseline Definition Worksheet
Supertech P.O. Box 186. Elkhart, N 46515-0186. Phone: 1-800-654-1054 or 1-574-264-4310 Fax: 574-264-9551. Web: www.supertechx-ray.com. E-mail: sales@supertechx-ray.com (CR PORTABLE CHEST EXAM Baseline
More informationQUALITY CONTROL TESTS IN SOME DIAGNOSTIC X-RAY UNITS IN BANGLADESH
Bangladesh Journal of Medical Physics Vol. 4, No.1, 2011 QUALITY CONTROL TESTS IN SOME DIAGNOSTIC X-RAY UNITS IN BANGLADESH M. Begum 1, A. S. Mollah 2, M. A. Zaman 3 and A. K. M. M. Rahman 4 1 Health Physics
More informationMinnesota Rules, Chapter 4732 X-ray Revision
Minnesota Rules, Chapter 4732 X-ray Revision DRAFT FLUOROSCOPIC X-RAY SYSTEMS, 1.0 Subpart 1. Applicability. Subpart 2. Limitation of the useful beam. Subpart 3. Measuring compliance; primary protective
More informationGE AMX 4+ Portable X-Ray
GE AMX 4+ Portable X-Ray Typical Manufacturer s Picture GE Healthcare s AMX-4+ analog X-ray system provides high-performance in a compact, easy-to-maneuver package. The rotating arm and tube simplify positioning
More informationDental Radiography Series
Dental Radiography Series Successful Intraoral Radiography Successful Intraoral Radiography Every dental professional would like to achieve the goal of producing consistent, quality intraoral radiographs.
More informationInvestigation of the line-pair pattern method for evaluating mammographic focal spot performance
Investigation of the line-pair pattern method for evaluating mammographic focal spot performance Mitchell M. Goodsitt, a) Heang-Ping Chan, and Bob Liu Department of Radiology, University of Michigan, Ann
More informationKODAK Dental Radiography Series. Successful Intraoral Radiography DENTAL PRODUCTS
KODAK Dental Radiography Series Successful Intraoral Radiography DENTAL PRODUCTS Contents Causes of Intraorai Radiography Problems... 1 How the X-ray Tube Functions... 1 Areas to Consider First... 2 Radiographic
More informationROTANODE TM E7252X E7252FX E7252GX Rotating Anode X-Ray Tube Housing Assembly. General Data. IEC Classification... Class I Electrical :
Technical Data TD ROTANODE TM E7252X E7252FX E7252GX Rotating Anode X-Ray Tube Housing Assembly High speed rotating anode X-ray tube housing assembly for high energy radiographic and cine-fluoroscopic
More informationDevelopment of a flat-panel x-ray source
Scholars' Mine Doctoral Dissertations Student Theses and Dissertations 214 Development of a flat-panel x-ray source Edwin Joseph Grant Follow this and additional works at: http://scholarsmine.mst.edu/doctoral_dissertations
More informationThe effect of compensating filter on image quality in lateral projection of thoraco lumbar radiography
Journal of Physics: Conference Series OPEN ACCESS The effect of compensating filter on image quality in lateral projection of thoraco lumbar radiography To cite this article: N A A Daud et al 2014 J. Phys.:
More information35080M kvp Divider. Operators Manual
35080M kvp Divider Operators Manual July 2006 Manual No. 1210001000 Rev. 3 2004, 2005, 2006 Fluke Corporation, All rights reserved. Printed in U.S.A. All product names are trademarks of their respective
More information(12) United States Patent
(12) United States Patent Suzuki et al. USOO6385294B2 (10) Patent No.: US 6,385,294 B2 (45) Date of Patent: May 7, 2002 (54) X-RAY TUBE (75) Inventors: Kenji Suzuki; Tadaoki Matsushita; Tutomu Inazuru,
More informationExplor-X AC / ACP. User's Manual. Release 28 April 2009 (Rev. 0)
Explor-X AC / ACP User's Manual Release 28 April 2009 Contents Contents 1. INTRODUCTION 1 1.1 Icons in the manual... 1 2. SAFETY ASPECTS 2 2.1 Warnings... 3 2.2 Protection from X-rays... 4 2.3 Environmental
More informationDate thesis is presented August 19, 1966
AN ABSTRACT OF THE THESIS OF Mark Orion Semler for the M. S. in Radiological Physics (Name) (Degree) (Major) Date thesis is presented August 19, 1966 Title DESIGN AND CONSTRUCTION OF AN X-RAY MACHINE FOR
More informationRotating Anode X-ray Tube Assembly
RADII H18X H18Y Datasheet Rotating Anode X-ray Tube Assembly The tube unit is so constructed that an X-ray tube is sealed in a diagnostic type protective tube housing of electric shockproof, radiation
More informationX-ray Imaging (XRI) Lectures on Medical Biophysics. Dept. Biophysics, Medical Faculty, Masaryk University in Brno. X-ray Imaging (XRI)
Lectures on Medical Biophysics Dept. Biophysics, Medical Faculty, Masaryk University in Brno Wilhelm Conrad Roentgen 1845-1923 Godfrey N. Hounsfield 1919-2004 X-Ray Imaging X-ray imaging (XRI) is still
More informationAssessment of Beam Alignment, Collimation and Half Value Layer of Some Selected X-Ray Machines in Plateau State, Nigeria
International Journal of Innovative Scientific & Engineering Technologies Research 5(4):-5, Oct.-Dec., 07 SEAHI PUBLICATIONS, 07 www.seahipaj.org ISSN: 60-896X Assessment of Beam Alignment, Collimation
More informationTECHNICAL SPECIFICATIONS for 300KV CONSTANT POTENTIAL X RAY EQUIPMENT (for industrial applications) S. No. PARTICULARS BHEL SPECIFICATIONS
TECHNICAL SPECIFICATIONS for 300KV CONSTANT POTENTIAL X RAY EQUIPMENT (for industrial applications) BIDDER s OFFER S. No. PARTICULARS BHEL SPECIFICATIONS 1.0 Area of Application The 300kV Constant potential
More informationIndustrial X-ray Tube KL
Description stationary anode x-ray tube is a highly popular and proven side window x-ray tube, which has focus 1.0 mm, is designed for non-destructive x-ray inspection. stationary anode x-ray tube can
More informationTest Review # 9. Physics R: Form TR9.15A. Primary colors of light
Physics R: Form TR9.15A TEST 9 REVIEW Name Date Period Test Review # 9 Light and Color. Color comes from light, an electromagnetic wave that travels in straight lines in all directions from a light source
More informationDIGITAL IMAGE PROCESSING IN X-RAY IMAGING
DIGITAL IMAGE PROCESSING IN X-RAY IMAGING Shalini Kumari 1, Bachan Prasad 2,Aliya Nasim 3 Department of Electronics And Communication Engineering R.V.S College of Engineering & Technology, Jamshedpur,
More informationProduct data Remote Controlled Table
Product data Remote Controlled Table Product data Remote Controlled Table 2 BRT-20 Remote controlled table has been designed as basic equipment, compact and extremely simplified. MAIN APPLICATION Gastroenterology
More informationCT Basics: Data Acquisition Module 3
Module 3 Transcript For educational and institutional use. This transcript is licensed for noncommercial, educational inhouse or online educational course use only in educational and corporate institutions.
More informationRotating Anode X-Ray Tube Housing Assembly. General Data. IEC Classification... Class I
Technical Data TD ROTANODE E7239X E7239FX E7239GX 197 Rotating Anode X-Ray Tube Housing Assembly Rotating anode X-ray tube housing assembly for the purpose of general diagnostic X-ray procedures. Specially
More informationENDOS AC / ACP User's Manual. Release 13 July 2007 (Rev. 6)
ENDOS AC / ACP 0051 User's Manual Release 13 July 2007 Revision history Revision history Rev. Date Page/s Modification description 0 07.02.03 - Document approval. 1 11.04.03 All Introduction of KAILONG
More informationNuclear Associates
Nuclear Associates 07-706 Patient Phantom/Penetrometer System Users Manual March 2005 Manual No. 07-706-1 Rev. 2 2004, 2005 Fluke Corporation, All rights reserved. Printed in U.S.A. All product names are
More informationPrecision Performance Power
ODYSSEY HF SERIES ULTRA High Frequency X-Ray Technology Precision Performance Power Innovations in Digital Imaging. TM STEP 1 Select anatomical region STEP 2 Select anatomical view STEP 3 Ready for exposure
More informationODYSSEY HF SERIES. ULTRA High Frequency X-Ray Technology. Precision... Performance... Power. Innovations in Digital Imaging.
ODYSSEY HF SERIES ULTRA High Frequency X-Ray Technology Precision... Performance... Power Innovations in Digital Imaging. TM STEP 1 Select anatomical region STEP 2 Select anatomical view STEP 3 Ready for
More informationX-ray investigation of crystal structures / Laue method with digital X-ray detector (XRIS) (Item No.: P )
X-ray investigation of crystal structures / Laue method with digital X-ray detector (XRIS) (Item No.: P2541602) Curricular Relevance Area of Expertise: Physik Education Level: Hochschule Topic: Moderne
More informationP4ACHLETT TAN VU L. 15 NO. 1
P4ACHLETT TAN VU L. 15 NO. 1 Aluminum Tube Welding with MACHLETT ML-56513* Thermatool High Frequency Resistance Welder at JFD Electronics, continuously welds round or square tubing at 300 feet per minute.
More information(12) United States Patent (10) Patent No.: US 7,050,541 B2
US007050541B (1) United States Patent () Patent No.: Bitt (45) Date of Patent: May 3, 006 (54) X-RAY TUBE WITH LIQUID-METAL FLUID (56) References Cited BEARNG U.S. PATENT DOCUMENTS (75) Inventor: Herbert
More information