BASICS OF FLUOROSCOPY

Size: px
Start display at page:

Download "BASICS OF FLUOROSCOPY"

Transcription

1 Medical Physics Residents Training Program BASICS OF FLUOROSCOPY Dr. Khalid Alyousef, PhD Department of Medical Imaging King Abdulaziz Medical City- Riyadh

2 Edison examining the hand of Clarence Dally with a fluoroscope 1896

3 Fluoroscopy for tuberculosis

4 The purpose of fluoroscopy is to provide a real-time x-ray imaging system. The two principle components are the image intensifier and the CC-TV image system. Image intensifiers have increased in size from the older 15cm systems to the modern 40cm systems. It s now possible to view an entire abdomen of the patient on the image intensifier. The development of High Definition TV will also help to improve the resolution of fluoroscopy systems.

5 Since the procedures requiring fluoroscopy may take some time, the exposure to the patient must be low. The x-ray exposure rate from fluoroscopy is several orders of magnitude less than that of general x-ray. Since the duration of the exposure is much longer and the penetration (kv) must still be maintained, it is the tube current that must be reduced. X-ray tube currents are usually much less for fluoroscopy, usually times less. The exposure rate for fluoroscopy is therefore also much less.

6 Relatively few x-rays are used to produce a fluoroscopic image. The resolution will therefore be less than that for radiographic imaging. Relative to general x-ray, the speed of the fluoroscopy systems are much much higher. The speed of the system will also have an effect on the rate at which the image can be refreshed. Frame rates of 30 frames per second are usually the limit.

7 REAL TIME IMAGING Dynamic studies require an imaging device that can follow events at a frame rate that is high enough so that a diagnosis can be made. A barium swallow study for example requires a fast enough frame rate to follow the peristaltic motion of the esophagus. Dynamic imaging in fluoroscopy provides adequate temporal resolution (frames per second) to visualize this function. Even higher frames rates are used for cardiac imaging, but this also requires a higher exposure rate.

8 POSITIONING Fluoroscopy is most commonly used for positioning. Following the progress of barium through the GI system or catheters as they pass through vessels are the procedures most commonly supported by fluoroscopy. Whilst the basic design of fluoroscopic systems is similar, systems can be designed for particular types of procedures. Small facilities will require a general purpose unit for a wide range of procedures whilst large facilities may have systems that are specifically designed for specialist tasks (e.g. cardiac, vascular, neuro, etc.)

9 FLUOROSCOPY EQUIPMENT

10 } More sensitive detector system than film. Low Current: ma Long exposure Controlled by foot switch } Tube capable of prolonged current

11 Image Intensification Tube Components Input screen and photocathode Electrostatic lenses Magnification tubes Anode Output phosphor

12 IMAGE INTENSIFIER This device is the major component of the fluoroscopy system. Its function is to convert the small number of x-rays incident upon it into visible light images. The primary technique is to amplify the small signal into a signal that can be viewed. The very early systems used a primitive x-ray fluorescent screen onto which x-rays were directed through the patient and viewed from behind. These very old systems produced huge radiation doses to the patient and the radiologist.

13 The modern image intensifier has 4 principle components:- 1. Vacuum bottle to keep all air out 2. Input layer to convert the incident x-rays into electrons 3. Electronic lenses to focus the electrons 4. Output phosphor to convert the energy of the electrons into visible light.

14 Focusing Electrodes Output Phosphur Incoming X-rays Electron Beam Anode Vacuum Tube

15 IMAGE INTENSIFIER Vacuum Bottle Whilst the task of the vacuum bottle seems simple, the design of the bottle is complex. The input phosphor may be up to 40cm in diameter and is able to withstand up to a ton of pressure from the outside atmosphere. Whilst maintaining the vacuum the bottle must also allow x- rays to penetrate. The loss of vacuum is usually the cause of degradation in the performance of the Image Intensifier. The Image Intensifier is a high cost item.

16 Input Layer The input layer is actually 4 layers. The first is a thin aluminum window built into the vacuum bottle. This window has a curved design to withstand the high pressures of the vacuum. This layer is usually about 1mm thick. The second is a substrate layer that supports the input phosphor but is also able to allow the penetration of as many x-rays as possible. This is also the first component of the electronic focusing system and is made from a 0.5mm layer of aluminum.

17 Input Layer The third layer is the input phosphor. It s function is to convert x-rays into visible light. This is similar in function to the screen in a film/screen radiographic system. It is also subject to the same compromising situation with regard to it s thickness (i.e. resolution vs sensitivity). Most Input Phosphors are made from CsI. CsI is constructed of thin needle-shaped light pipes, ideal for directing the light with minimal spreading. The K-edges of Cesium (36keV) and Iodine (33keV) allow a high x-ray absorption efficiency.

18 X-Ray Support CsI Needles Light Photocathode Electrons

19 IMAGE INTENSIFIER Input Layer The fourth layer is the photocathode. The primary function of this layer is convert light photons from the phosphor to photoelectrons. This layer is usually constructed by a thin alloy layer of antimony and alkali metals. The efficiency of the photocathode is approximately 10-20%. For a single 60keV x-ray photon absorbed in the phosphor, approximately 400 electrons are released from the photocathode.

20 Focusing Electrodes Aluminium Cover Output Phosphur Input Phosphur Anode Light Pipe Photocathode Vacuum Tube

21 IMAGE INTENSIFIER Electronic Lens System The purpose of the lens system is to focus the electrons produced by the photocathode and accelerate them towards the output phosphor. Electrons are released from the photocathode with little energy so a relatively large potential is applied to accelerate the electrons to the output phosphor. As each electron strikes each electrode in the lens system more electrons are produced, increasing the flux of electrons that finally strike the output phosphor. In order to maintain an equal path length for all electrons, the surface of the photocathode must be curved.

22 Output Phosphor The purpose of the output phosphor is to convert the electrons striking it into visible light. The output phosphor is usually made of zinc cadmium sulfide and emits green light at a wavelength of 530nm The output phosphor is thin (4-8 mm) to maintain the highest possible spatial resolution. Each electron striking the output phosphor will produce approximately 1000 light photons. The image produced by the output phosphor is highly minified. The input phosphor is 30-40cm in diameter while the output phosphor is just 2.5cm diameter.

23 Output Window The final stage of the image intensifier is the output window. The output window is part of the vacuum bottle and must allow as much light as possible to pass from the output phosphor. The glass window must transmit as much light as possible with as little reflected light as possible. Reflected light can cause a loss of image contrast and is called veiling glare. Output windows are now designed to transmit light and trap the small fraction of reflected light before it can travel back to the output phosphor.

24 IMAGE INTENSIFIER CHARACTERISTICS 1. Conversion Factor The conversion factor is a measurement of the light output relative to the input radiation exposure. Light Output is measured in units of Candela/m 2 (Cd/m 2 ) Input Exposure Rate is measured in units of mr/sec Conversion Factor = Light Input/Exposure Rate (Cd-s/mR-m 2 ) Typical conversion factors are in the range of The conversion factor will degrade with time and is another indication for the replacement of the II.

25 2. Brightness Gain This is a measure of the increase in light output from the output phosphor. The Brightness Gain is the product of the Flux Gain and the Minification Gain. Typically times. Brightness Gain = Flux Gain x Minification Gain The Flux Gain is the increase in the number of light photons emitted by the output phosphor compared to the input phosphor. Usually a factor of Minification Gain is the geometric change in the image size from the input to output phoshors. Minification Gain = (d input /d output ) 2

26 3. Contrast Ratio This is the ability of the II do identify contrast of a large object. Contrast is primarily determined by the amount of veiling glare, modern systems have reduced this and therefore enhanced the contrast of fluoroscopy systems. Contrast Ratio is determined by measuring the light output at the center of the FOV, with and without a lead disk to filter to block x-rays incident on the II. In an ideal system the light behind the lead disk should be zero and the contrast ratio should be infinite. In modern systems the actual values are in the range of

27 A Contrast Ratio = A/B B

28 4. Field of View (Magnification Modes) The physical size of the input phosphor of most modern II s is in the range of 23-40cm. The larger the II diameter, the more expensive the system. The larger systems are very useful for imaging the larger sections of the body, i.e. abdomen. It is however also useful for be able to magnify sections of the image to smaller fields of view. This is made possible by the electric field of the electric lens system. Modifying the voltage on this system effectively reduces the field of view.

29 5. Detected Quantum Efficiency The Detected Quantum Efficiency (DQE) is a parameter that relates the statistical quality of the output signal to that of the input x-ray signal. A perfect system would have a DQE of 100%. Noise will always reduce DQE. The DQE will also be highly dependant on the x-ray spectrum of the incident radiation. Typical DQE values are in the range of 60-80%. The advantage of a high DQE is that low noise images can be produced at the same exposure rate.

30 6. Limiting Spatial Resolution The resolution of a modern II ranges from 4-5lp/mm for a 23cm FOV up to 7lp/mm for higher magnification systems. The limiting spatial resolution is easily measured with a line/pair phantom. Whilst this measurement may be subjective, it is still a useful guide in the clinical setting.

31 Field Size Input Field size typically in the range 23-57cm. Size can be reduced further electronically to 14 17cm. Large field sizes for imaging the abdomen. Resolution decreases with increasing field size 23cm (5lp/mm) to 57cm (3lp/mm). Output Phosphor typically 2.5cm diameter. Input Phosphor µm thick Cesium Iodide (CsI) Like screens, resolution decreases as phosphur thickness increases.

32 Output Phosphor 4-8µm thick to preserve the spatial resolution of the image Zinc Cadmium Sulphide Usually emits light in the green section of the visible spectrum (530nm) Each electron striking the output phosphor produces about 1000 light photons.

33 Image Intensifier (II) Light emitted by the input phosphor is absorbed by the photocathode to produce photoelectrons. Photoelectrons are absorbed in the output phosphor and converted to light Image brightness increased by minimizing the image and producing many light photons for each x-ray absorbed. Brightness Gain = Minimization Gain x Flux Gain Minimization Gain = Flux gain = Therefore, Brightness Gain =

34 Image Quality Resolution in the range 3-5lp/mm Resolution will decrease at the edges Since fluoroscopy is performed at low doses, images are noisy. Artifacts Lag due to continued luminescence (minor problem) Veiling Glare Light scatter in the output image adds to the total background of the image. Pincushion distortion of the image at the edges of the II Vignetting low brightness at the edges of the II

35 Optical System TV Camera Photospot or Cine Camera

36 Radiation Exposure Whilst the radiation dose rate from fluoroscopy is much lower than from conventional x-ray, the exposure times from fluoroscopy is much longer. The radiation dose rate to the skin can be quite high and is limited to 100mGy/minute. Hence, most fluoroscopy systems will have an alarm to warn the user of excessive radiation doses.

37 Digital Fluoroscopy An analogue-to-digital converter can be used to digitize the TV output of the system. The digital images can be stored on a computer system and the images can be further processed Multiple images can be stored as a series at frame rates of up to 30 frames/second on a digital system. Digital Subtraction Angiography (DSA) involves the subtraction of a background image from an angiographic image.

38 Digital Fluoroscopy Image matrix of 1024x1024 pixels

39 Digital Subtraction Angiography A Post-Contrast Image is subtracted from a Pre-Contrast Mask Image. The result is an image of the contrast only and gives the best visualisation of vessels. Post processing is possible on the resulting images. Image matrix sizes usually 512 x 512 and 1024 x 1024

40 Image Intensifier Artifacts 1. Veiling Glare This is predominantly the result of light scatter from the output window of the Image Intensifier. This scattered light will have an affect on the entire area of the field of view and adds to the entire background of the image. Modern II s have a design that reduces this effect and there is little that can be done by the technologist to reduce this. Measuring the Contrast Ratio will indicate the particular amount of Veiling Glare resulting from the output window design.

41 2. Pincushion Distortion This is a form of spatial distortion that warps the appearance of the image. This is due to mapping the image from a curved input phosphor to the flat output phosphor. This distortion results in a slightly higher magnification at the edges of the field of view. Viewing the affect of pincushion distortion can best be done using a grid or screen with a regular spacing of lines.

42 Input Image Output Image

43 3. Vignetting The brightness of the entire field of view will vary from the center to the edge. The brightest section of the image will be at the center. This is brought about in a similar way to that of the pincushion artifact. Magnification will not only create the pincushion effect but also reduce the intensity of the image at the edges. Even though the image is minified by the electron focusing system, the loss of brightness at the edges will still be apparent.

44 4. S Distortion The S-Distortion is a spatial warping of the image along an s- shaped axis through the image. This type of distortion is subtle and the effect of pincushion distortion is usually more pronounced. S-distortion is the result of stray magnetic fields that affect the electron focusing lens system. When electrons are accelerated they will be influenced by magnetic fields. If the image intensifier is rotated the s-distortion will follow the direction will change position and align with the magnetic field.

45 OPTICAL COUPLING Distribution Mechanism. The image coming from the Output Phosphor requires a pair of lenses to focus the image onto the TV system. In simple systems this is carried out by a single fully enclosed connection. In more complex systems, the image from the Output Phosphor may be directed to more than one device. An Optical Distributor is used to direct the image through a mirror system to multiple devices. This mirror system is called a beam splitter. No mechanical devices are used in the beam splitter, it is simply a system of mirrors.

46 OPTICAL COUPLING Aperture Effect on Dose Rate and Noise. The light output of the Output Phosphor varies greatly as the exposure rate of the II changes. The image brightness that the TV system can view is limited. An aperture is placed between the optical coupling mechanism and the TV system to adjust the light input to the TV system. Adjusting the aperture will require an adjustment to the input exposure to the II in order to obtain good image quality. Hence this will also affect the radiation dose and image noise.

47 Video Camera Lens Accessory Port Aperture Mirror Output Window Output Phosphor

48 Optical System TV Camera Photospot or Cine Camera

49 VIDEO CAMERAS The output image of the optical distributor is connected to a CC-TV system on most modern fluoroscopy systems. The major disadvantage of using CC-TV for viewing is that the image resolution of the CC-TV system will be less than that of the Output Phosphor. Disadvantage: Since CC-TV resolution is low, this image is only used for positioning, permanent images are recorded using a different system attached to the accessory port.

50 Advantages of video camera: The CC-TV monitor is large and can be viewed from a large distance. The CC-TV monitor will remain stationary, even if the II is moved relative to the patient. A number of monitors can be set-up to view the same image from the CC-TV camera in multiple locations. The CC-TV image can easily be recorded on video tape for later viewing The distance between the II and the CC-TV monitor can be large, thereby reducing crowding in the fluoroscopy suite.

51 Components of video camera: The components of the CC-TV system are the TV camera, TV monitor and the camera control unit. The camera converts the 2-dimensional image into a single signal of varying voltage. The voltage is proportional to the brightness of the output image in the same relative location of the filed of view of the camera. This video signal is transmitted along a cable to the TV monitor and the one-dimensional signal is converted back to a 2- dimensional image by the monitor.

52 VIDEO CAMERAS Image Production: TV camera design has aimed to develop a system that reduces image lag. Image Lag becomes a problem when large bright objects move quickly across the field if view, leaving a comet tail artifact. Modern cameras have largely overcome this problem.

53 VIDEO CAMERAS New Developments Modern CCD (charge couple device) cameras use an array to record an image instead of an electron beam. CCD camera produce images similar to a computer matrix with a 512x512 or 1024x1024 pixel image. CCD cameras are very rugged and this technology is used in many of the modern home video cameras. CCD cameras produce virtually no lag. The introduction of HDTV will also benefit fluoroscopy. Higher resolution TV systems will be able to record the high resolution output phosphor images.

54 AUTOMATIC BRIGHTNESS CONTROL The exposure rate of a fluoroscopic system is controlled by the Automatic Exposure Control to ensure the brightness of the image remains constant. The exposure rate will vary greatly depending on the body part being imaged. The system monitors the brightness of the output phosphor and controls the kv and ma of the x-ray generator to maintain constant brightness. A light sensitive device (usually a photodiode) is used to measure the light output of the output phosphor. The signal from the light sensitive device feeds back to the x- ray generator.

55 Whilst the Automatic Brightness Control has the ability to change both the kv and ma, most systems have a predetermined sequence for modifying kv and ma. An increase in kv will result in a decrease in image contrast so if extra x-rays are required, ma is usually altered first before kv. It is also important to maintain low patient doses. Lower patient doses are achieved by limiting the increase in ma before increasing kv. It is always a trade-off between image contrast and patient dose.

56 SYSTEM RESOLUTION The overall resolution of a complex system is the product of the resolution of the components that make-up the system. The component with the poorest resolution will have the largest detrimental effect on overall system resolution. Of the three major components of the fluoroscopic imaging system (Image Intensifier, Optics and TV camera), the TV camera has the poorest resolution. Optics have the least effect on resolution, ahead of the Image Intensifier. Plotting the Modulation Transfer Function of each component will indicate the contribution of each component to spatial resolution.

57 IMAGE ACCESSORIES FOR FLUOROSCOPY Photospot Camera This camera is attached to the accessory port of the optical distributor. The camera records the image on radiographic film, usually in a small 100x100mm square format. The image is created from a single radiographic pulse of x-rays and the film ejected after exposure. The resolution of the photospot film will actually be higher than that of the TV image. This is generally an old technology replaced now by digital devices.

58 IMAGE ACCESSORIES FOR FLUOROSCOPY Digital Photospot This is a high resolution TV camera system which digitizes the image and stores it in a computer system. The image is created by a single radiographic exposure from the x-ray generator The resulting image is displayed on a computer screen. The resolution of this system is that of the TV System and is usually 1023 lines vertical. Hardcopy imaging can be printed on a laser printer from the computer image. Digital photospot cameras are in common use on modern fluoroscopy systems.

59 IMAGE ACCESSORIES FOR FLUOROSCOPY Cine Radiography A cine camera is attached to the accessory port of the optical distributor. The camera can record a very rapid sequence of images, usually on 35mm film. Cine radiography is primarily used in cardiology where frames rates of up to 120 frames/second are required. Images are recorded from sequences of short radiographic pulses from specially designed x-ray generators. X-ray tube loading is usually quite high for these systems.

60 RADIATION EXPOSURE Skin entrance exposure is the primary concern in fluoroscopy. Organ doses are reduced by the attenuation of overlying soft tissue. Skin doses of 5-50mGy/min are typical skin doses for fluoroscopy. Maximum allowable doses vary from country to country from mGy/min. Skin erythema is a possible outcome from high dose fluoroscopy. Patient doses can be reduced by reducing fluoroscopy time and increasing the field of view.

Fluoroscopy - Chapter 9

Fluoroscopy - Chapter 9 Fluoroscopy - Chapter 9 Kalpana Kanal, Ph.D., DABR Lecturer, Diagnostic Physics Dept. of Radiology UW Medicine a copy of this lecture may be found at: http://courses.washington.edu/radxphys/physicscourse04-05.html

More information

10/15/2012 SECTION III - CHAPTER 6 DIGITAL FLUOROSCOPY RADT 3463 COMPUTERIZED IMAGING

10/15/2012 SECTION III - CHAPTER 6 DIGITAL FLUOROSCOPY RADT 3463 COMPUTERIZED IMAGING RADT 3463 - COMPUTERIZED IMAGING Section III: Chapter 6 RADT 3463 Computerized Imaging 1 SECTION III - CHAPTER 6 DIGITAL FLUOROSCOPY RADT 3463 COMPUTERIZED IMAGING Section III: Chapter 6 RADT 3463 Computerized

More information

Introduction. Chapter 16 Diagnostic Radiology. Primary radiological image. Primary radiological image

Introduction. Chapter 16 Diagnostic Radiology. Primary radiological image. Primary radiological image Introduction Chapter 16 Diagnostic Radiology Radiation Dosimetry I Text: H.E Johns and J.R. Cunningham, The physics of radiology, 4 th ed. http://www.utoledo.edu/med/depts/radther In diagnostic radiology

More information

X-RAY FLUOROSCOPY IMAGING SYSTEMS. Dr Slavik Tabakov. Luminescence: Dept. Medical Eng. & Physics King s College London

X-RAY FLUOROSCOPY IMAGING SYSTEMS. Dr Slavik Tabakov. Luminescence: Dept. Medical Eng. & Physics King s College London X-RAY FLUOROSCOPY IMAGING SYSTEMS Dr Slavik Tabakov OBJECTIVES - Image Intensifier construction - Input window - Accelerating and focusing electrodes - Output window - Conversion factor - II characteristics

More information

10/26/2015. Study Harder

10/26/2015. Study Harder This presentation is a professional collaboration of development time prepared by: Rex Christensen Terri Jurkiewicz and Diane Kawamura Study Harder CR detection is inefficient, inferior to film screen

More information

Do you have any other questions? Please call us at (Toll Free) or , or

Do you have any other questions? Please call us at (Toll Free) or , or INSTRUCTIONS Read the appropriate course/ textbook. This is an open book test. A score of 75% or higher is needed to receive CE credit. You will have a maximum of three attempts to pass this course. Please

More information

I. PERFORMANCE OF X-RAY PRODUCTION COMPONENTS FLUOROSCOPIC ACCEPTANCE TESTING: TEST PROCEDURES & PERFORMANCE CRITERIA

I. PERFORMANCE OF X-RAY PRODUCTION COMPONENTS FLUOROSCOPIC ACCEPTANCE TESTING: TEST PROCEDURES & PERFORMANCE CRITERIA FLUOROSCOPIC ACCEPTANCE TESTING: TEST PROCEDURES & PERFORMANCE CRITERIA EDWARD L. NICKOLOFF DEPARTMENT OF RADIOLOGY COLUMBIA UNIVERSITY NEW YORK, NY ACCEPTANCE TESTING GOALS PRIOR TO 1st CLINICAL USAGE

More information

10/3/2012. Study Harder

10/3/2012. Study Harder This presentation is a professional collaboration of development time prepared by: Rex Christensen Terri Jurkiewicz and Diane Kawamura Study Harder CR detection is inefficient, inferior to film screen

More information

DIGITAL IMAGE PROCESSING IN X-RAY IMAGING

DIGITAL IMAGE PROCESSING IN X-RAY IMAGING DIGITAL IMAGE PROCESSING IN X-RAY IMAGING Shalini Kumari 1, Bachan Prasad 2,Aliya Nasim 3 Department of Electronics And Communication Engineering R.V.S College of Engineering & Technology, Jamshedpur,

More information

AN ABSTRACT OF THE THESIS OF. W. Scott Helms for the degree of Master of Science in Radiation Health Physics

AN ABSTRACT OF THE THESIS OF. W. Scott Helms for the degree of Master of Science in Radiation Health Physics AN ABSTRACT OF THE THESIS OF W. Scott Helms for the degree of Master of Science in Radiation Health Physics presented on November 24, 2014 Title: A Quantitative Comparison of Cardiovascular Imaging Systems

More information

Seminar 8. Radiology S8 1

Seminar 8. Radiology S8 1 Seminar 8 Radiology Medical imaging. X-ray image formation. Energizing and controlling the X-ray tube. Image detectors. The acquisition of analog and digital images. Digital image processing. Selected

More information

Joint ICTP/IAEA Advanced School on Dosimetry in Diagnostic Radiology and its Clinical Implementation May 2009

Joint ICTP/IAEA Advanced School on Dosimetry in Diagnostic Radiology and its Clinical Implementation May 2009 2033-6 Joint ICTP/IAEA Advanced School on Dosimetry in Diagnostic Radiology and its Clinical Implementation 11-15 May 2009 Dosimetry for Fluoroscopy Basics Renato Padovani EFOMP Joint ICTP-IAEA Advanced

More information

Radiology Physics Lectures: Digital Radiography. Digital Radiography. D. J. Hall, Ph.D. x20893

Radiology Physics Lectures: Digital Radiography. Digital Radiography. D. J. Hall, Ph.D. x20893 Digital Radiography D. J. Hall, Ph.D. x20893 djhall@ucsd.edu Background Common Digital Modalities Digital Chest Radiograph - 4096 x 4096 x 12 bit CT - 512 x 512 x 12 bit SPECT - 128 x 128 x 8 bit MRI -

More information

X-ray Tube and Generator Basic principles and construction

X-ray Tube and Generator Basic principles and construction X-ray Tube and Generator Basic principles and construction Dr Slavik Tabakov - Production of X-rays and Patient Dose OBJECTIVES - X-ray tube construction - Anode - types, efficiency - Classical X-ray generator

More information

X-rays in medical diagnostics

X-rays in medical diagnostics X-rays in medical diagnostics S.Dolanski Babić 2017/18. History W.C.Röntgen (1845-1923) discovered a new type of radiation Nature, Jan. 23. 1896.; Science, Feb.14. 1896. X- rays: Induced the ionization

More information

Unit thickness. Unit area. σ = NΔX = ΔI / I 0

Unit thickness. Unit area. σ = NΔX = ΔI / I 0 Unit thickness I 0 ΔI I σ = ΔI I 0 NΔX = ΔI / I 0 NΔX Unit area Δx Average probability of reaction with atom for the incident photons at unit area with the thickness of Delta-X Atom number at unit area

More information

Radiographic sensitivity improved by optimized high resolution X -ray detector design.

Radiographic sensitivity improved by optimized high resolution X -ray detector design. DIR 2007 - International Symposium on Digital industrial Radiology and Computed Tomography, June 25-27, 2007, Lyon, France Radiographic sensitivity improved by optimized high resolution X -ray detector

More information

COMPUTED TOMOGRAPHY 1

COMPUTED TOMOGRAPHY 1 COMPUTED TOMOGRAPHY 1 Why CT? Conventional X ray picture of a chest 2 Introduction Why CT? In a normal X-ray picture, most soft tissue doesn't show up clearly. To focus in on organs, or to examine the

More information

X-ray Imaging. PHYS Lecture. Carlos Vinhais. Departamento de Física Instituto Superior de Engenharia do Porto

X-ray Imaging. PHYS Lecture. Carlos Vinhais. Departamento de Física Instituto Superior de Engenharia do Porto X-ray Imaging PHYS Lecture Carlos Vinhais Departamento de Física Instituto Superior de Engenharia do Porto cav@isep.ipp.pt Overview Projection Radiography Anode Angle Focal Spot Magnification Blurring

More information

Amorphous Selenium Direct Radiography for Industrial Imaging

Amorphous Selenium Direct Radiography for Industrial Imaging DGZfP Proceedings BB 67-CD Paper 22 Computerized Tomography for Industrial Applications and Image Processing in Radiology March 15-17, 1999, Berlin, Germany Amorphous Selenium Direct Radiography for Industrial

More information

Examination of Pipe Welds by Image Plate Based Computed Radiography System

Examination of Pipe Welds by Image Plate Based Computed Radiography System Examination of Pipe Welds by Image Plate Based Computed Radiography System Sanjoy Das, M.S.Rana, Benny Sebastian, D. Mukherjee and K.K. Abdulla Atomic Fuels Division Bhabha Atomic Research Centre Mumbai

More information

Medical Imaging. X-rays, CT/CAT scans, Ultrasound, Magnetic Resonance Imaging

Medical Imaging. X-rays, CT/CAT scans, Ultrasound, Magnetic Resonance Imaging Medical Imaging X-rays, CT/CAT scans, Ultrasound, Magnetic Resonance Imaging From: Physics for the IB Diploma Coursebook 6th Edition by Tsokos, Hoeben and Headlee And Higher Level Physics 2 nd Edition

More information

Acquisition, Processing and Display

Acquisition, Processing and Display Acquisition, Processing and Display Terri L. Fauber, R.T. (R)(M) Department of Radiation Sciences School of Allied Health Professions Virginia Commonwealth University Topics Image Characteristics Image

More information

Overview. Professor Roentgen was a Physicist!!! The Physics of Radiation Oncology X-ray Imaging

Overview. Professor Roentgen was a Physicist!!! The Physics of Radiation Oncology X-ray Imaging The Physics of Radiation Oncology X-ray Imaging Charles E. Willis, Ph.D. DABR Associate Professor Department of Imaging Physics The University of Texas M.D. Anderson Cancer Center Houston, Texas Overview

More information

SECTION I - CHAPTER 1 DIGITAL RADIOGRAPHY: AN OVERVIEW OF THE TEXT. Exam Content Specifications 8/22/2012 RADT 3463 COMPUTERIZED IMAGING

SECTION I - CHAPTER 1 DIGITAL RADIOGRAPHY: AN OVERVIEW OF THE TEXT. Exam Content Specifications 8/22/2012 RADT 3463 COMPUTERIZED IMAGING RADT 3463 - COMPUTERIZED IMAGING Section I: Chapter 1 RADT 3463 Computerized Imaging 1 SECTION I - CHAPTER 1 DIGITAL RADIOGRAPHY: AN OVERVIEW OF THE TEXT RADT 3463 COMPUTERIZED IMAGING Section I: Chapter

More information

SECTION I - CHAPTER 2 DIGITAL IMAGING PROCESSING CONCEPTS

SECTION I - CHAPTER 2 DIGITAL IMAGING PROCESSING CONCEPTS RADT 3463 - COMPUTERIZED IMAGING Section I: Chapter 2 RADT 3463 Computerized Imaging 1 SECTION I - CHAPTER 2 DIGITAL IMAGING PROCESSING CONCEPTS RADT 3463 COMPUTERIZED IMAGING Section I: Chapter 2 RADT

More information

X-RAY IMAGING EE 472 F2017. Prof. Yasser Mostafa Kadah

X-RAY IMAGING EE 472 F2017. Prof. Yasser Mostafa Kadah X-RAY IMAGING EE 472 F2017 Prof. Yasser Mostafa Kadah www.k-space.org Recommended Textbook Stewart C. Bushong, Radiologic Science for Technologists: Physics, Biology, and Protection, 10 th ed., Mosby,

More information

PLD5600A High Frequency Digital Gastrointestinal &DR System(630mA)

PLD5600A High Frequency Digital Gastrointestinal &DR System(630mA) PLD5600A High Frequency Digital Gastrointestinal &DR System(630mA) Application: Full support perspective, gastrointestinal spot film, GI (barium meal, barium enema), orthopedic photography, pediatrics

More information

X-ray generation by femtosecond laser pulses and its application to soft X-ray imaging microscope

X-ray generation by femtosecond laser pulses and its application to soft X-ray imaging microscope X-ray generation by femtosecond laser pulses and its application to soft X-ray imaging microscope Kenichi Ikeda 1, Hideyuki Kotaki 1 ' 2 and Kazuhisa Nakajima 1 ' 2 ' 3 1 Graduate University for Advanced

More information

TESTING FLAT-PANEL IMAGING SYSTEMS: What the Medical Physicist Needs to Know. JAMES A. TOMLINSON, M.S., D.A.B.R. Diagnostic Radiological Physicist

TESTING FLAT-PANEL IMAGING SYSTEMS: What the Medical Physicist Needs to Know. JAMES A. TOMLINSON, M.S., D.A.B.R. Diagnostic Radiological Physicist TESTING FLAT-PANEL IMAGING SYSTEMS: What the Medical Physicist Needs to Know JAMES A. TOMLINSON, M.S., D.A.B.R. Diagnostic Radiological Physicist Topics Image Uniformity and Artifacts Image Quality - Detail

More information

Technical information. X-RAY PRODUCTS/Integrated solutions

Technical information. X-RAY PRODUCTS/Integrated solutions X-RAY PRODUCTS/Integrated solutions MTHF C-arm mobile Units with image intensifier Mobile C-arm units for fluoroscopy and radiography, with high frequency generator. Complete with digital image memories.

More information

PD233: Design of Biomedical Devices and Systems

PD233: Design of Biomedical Devices and Systems PD233: Design of Biomedical Devices and Systems (Lecture-8 Medical Imaging Systems) (Imaging Systems Basics, X-ray and CT) Dr. Manish Arora CPDM, IISc Course Website: http://cpdm.iisc.ac.in/utsaah/courses/

More information

Digital radiography: Practical advantages of Digital Radiography. Practical Advantages in image quality

Digital radiography: Practical advantages of Digital Radiography. Practical Advantages in image quality Digital radiography: Digital radiography is set to become the most common form of processing radiographic images in the next 10 years. This is due to a number of practical and image quality issues. Practical

More information

STUDENT REVIEW QUESTION SET K CR/DR CONTENT AREA

STUDENT REVIEW QUESTION SET K CR/DR CONTENT AREA STUDENT REVIEW QUESTION SET K CR/DR CONTENT AREA RADT 2913 COMPREHENSIVE REVIEW 1 The CR cassette is backed by aluminum that: A. reflects x-rays B. absorbs x-rays C. captures the image D. transmits x-rays

More information

INTRODUCTION TO FLEXIBLE BRONCHOSCOPY. Fluoroscopy Synopsis HENRI G COLT MD SECOND EDITION THE BRONCHOSCOPY EDUCATION PROJECT SERIES

INTRODUCTION TO FLEXIBLE BRONCHOSCOPY. Fluoroscopy Synopsis HENRI G COLT MD SECOND EDITION THE BRONCHOSCOPY EDUCATION PROJECT SERIES SECOND EDITION INTRODUCTION TO FLEXIBLE BRONCHOSCOPY Fluoroscopy Synopsis HENRI G COLT MD With contributions from Dr. S. Murgu THE BRONCHOSCOPY EDUCATION PROJECT SERIES FLUOROSCOPY SYNOPSIS The purpose

More information

Flat-panel detectors: how much better are they?

Flat-panel detectors: how much better are they? Pediatr Radiol (2006) 36 (Suppl 2): 173 181 DOI 10.1007/s00247-006-0208-0 ALARA J. Anthony Seibert Flat-panel detectors: how much better are they? Published online: 22 July 2006 # Springer-Verlag 2006

More information

CHAPTER 9 POSITION SENSITIVE PHOTOMULTIPLIER TUBES

CHAPTER 9 POSITION SENSITIVE PHOTOMULTIPLIER TUBES CHAPTER 9 POSITION SENSITIVE PHOTOMULTIPLIER TUBES The current multiplication mechanism offered by dynodes makes photomultiplier tubes ideal for low-light-level measurement. As explained earlier, there

More information

Dose Reduction and Image Preservation After the Introduction of a 0.1 mm Cu Filter into the LODOX Statscan unit above 110 kvp

Dose Reduction and Image Preservation After the Introduction of a 0.1 mm Cu Filter into the LODOX Statscan unit above 110 kvp Dose Reduction and Image Preservation After the Introduction of a into the LODOX Statscan unit above 110 kvp Abstract: CJ Trauernicht 1, C Rall 1, T Perks 2, G Maree 1, E Hering 1, S Steiner 3 1) Division

More information

TSBB09 Image Sensors 2018-HT2. Image Formation Part 1

TSBB09 Image Sensors 2018-HT2. Image Formation Part 1 TSBB09 Image Sensors 2018-HT2 Image Formation Part 1 Basic physics Electromagnetic radiation consists of electromagnetic waves With energy That propagate through space The waves consist of transversal

More information

Veterinary Science Preparatory Training for the Veterinary Assistant. Floron C. Faries, Jr., DVM, MS

Veterinary Science Preparatory Training for the Veterinary Assistant. Floron C. Faries, Jr., DVM, MS Veterinary Science Preparatory Training for the Veterinary Assistant Floron C. Faries, Jr., DVM, MS Radiology Floron C. Faries, Jr., DVM, MS Objectives Determine the appropriate machine settings for making

More information

High collection efficiency MCPs for photon counting detectors

High collection efficiency MCPs for photon counting detectors High collection efficiency MCPs for photon counting detectors D. A. Orlov, * T. Ruardij, S. Duarte Pinto, R. Glazenborg and E. Kernen PHOTONIS Netherlands BV, Dwazziewegen 2, 9301 ZR Roden, The Netherlands

More information

Digital Detector Array Image Quality for Various GOS Scintillators

Digital Detector Array Image Quality for Various GOS Scintillators Digital Detector Array Image Quality for Various GOS Scintillators More info about this article: http://www.ndt.net/?id=22768 Brian S. White 1, Mark E. Shafer 2, William H. Russel 3, Eric Fallet 4, Jacques

More information

Observational Astronomy

Observational Astronomy Observational Astronomy Instruments The telescope- instruments combination forms a tightly coupled system: Telescope = collecting photons and forming an image Instruments = registering and analyzing the

More information

ARCO Rk.5 Digital Mobile C-Arm

ARCO Rk.5 Digital Mobile C-Arm Product Data ARCO Rk.5 Digital Mobile C-Arm Product Data 1 Easy to move INTRODUCTION Designed with the latest technology throughout, the ARCO Rk.5 set new standards for the excellence image quality ease

More information

PROCEEDINGS OF A SYMPOSIUM HELD AT THE CAVENDISH LABORATORY, CAMBRIDGE, Edited by

PROCEEDINGS OF A SYMPOSIUM HELD AT THE CAVENDISH LABORATORY, CAMBRIDGE, Edited by X - R A Y M I C R O S C O P Y A N D M I C R O R A D I O G R A P H Y PROCEEDINGS OF A SYMPOSIUM HELD AT THE CAVENDISH LABORATORY, CAMBRIDGE, 1956 Edited by V. E. COSSLETT Cavendish Laboratory, University

More information

X-RAYS - NO UNAUTHORISED ENTRY

X-RAYS - NO UNAUTHORISED ENTRY Licencing of premises Premises Refer Guidelines A radiation warning sign and warning notice, X-RAYS - NO UNAUTHORISED ENTRY must be displayed at all entrances leading to the rooms where x-ray units are

More information

EE119 Introduction to Optical Engineering Spring 2003 Final Exam. Name:

EE119 Introduction to Optical Engineering Spring 2003 Final Exam. Name: EE119 Introduction to Optical Engineering Spring 2003 Final Exam Name: SID: CLOSED BOOK. THREE 8 1/2 X 11 SHEETS OF NOTES, AND SCIENTIFIC POCKET CALCULATOR PERMITTED. TIME ALLOTTED: 180 MINUTES Fundamental

More information

DETECTORS Important characteristics: 1) Wavelength response 2) Quantum response how light is detected 3) Sensitivity 4) Frequency of response

DETECTORS Important characteristics: 1) Wavelength response 2) Quantum response how light is detected 3) Sensitivity 4) Frequency of response DETECTORS Important characteristics: 1) Wavelength response 2) Quantum response how light is detected 3) Sensitivity 4) Frequency of response (response time) 5) Stability 6) Cost 7) convenience Photoelectric

More information

SECONDARY ELECTRON DETECTION

SECONDARY ELECTRON DETECTION SECONDARY ELECTRON DETECTION CAMTEC Workshop Presentation Haitian Xu June 14 th 2010 Introduction SEM Raster scan specimen surface with focused high energy e- beam Signal produced by beam interaction with

More information

SYLLABUS. 1. Identification of Subject:

SYLLABUS. 1. Identification of Subject: SYLLABUS Date/ Revision : 30 January 2017/1 Faculty : Life Sciences Approval : Dean, Faculty of Life Sciences SUBJECT : Biophysics 1. Identification of Subject: Name of Subject : Biophysics Code of Subject

More information

Basis of Computed Radiography & PACS

Basis of Computed Radiography & PACS Basis of Computed Radiography & PACS Slavik Tabakov Computed Radiography (CR) refers to new types of X-ray detectors (i.e. replaces the X-ray Film) The CR output media is a digital image, which can be

More information

LECTURE 1 The Radiographic Image

LECTURE 1 The Radiographic Image LECTURE 1 The Radiographic Image Prepared by:- KAMARUL AMIN ABDULLAH @ ABU BAKAR UiTM Faculty of Health Sciences Medical Imaging Department 11/23/2011 KAMARUL AMIN (C) 1 Lesson Objectives At the end of

More information

Applications of Optics

Applications of Optics Nicholas J. Giordano www.cengage.com/physics/giordano Chapter 26 Applications of Optics Marilyn Akins, PhD Broome Community College Applications of Optics Many devices are based on the principles of optics

More information

Multiple Choice Identify the letter of the choice that best completes the statement or answers the question.

Multiple Choice Identify the letter of the choice that best completes the statement or answers the question. RA110 test 3 Multiple Choice Identify the letter of the choice that best completes the statement or answers the question. 1. An object 35 cm in width is radiographed at 100 cm SID and at a 50 cm SOD. What

More information

Components of Optical Instruments

Components of Optical Instruments Components of Optical Instruments General Design of Optical Instruments Sources of Radiation Wavelength Selectors (Filters, Monochromators, Interferometers) Sample Containers Radiation Transducers (Detectors)

More information

RADIOGRAPHIC EXPOSURE

RADIOGRAPHIC EXPOSURE RADIOGRAPHIC EXPOSURE Receptor Exposure Receptor Exposure the that interacts with the receptor. Computed Radiography ( ) requires a. Direct Digital Radiography (DR) requires a. Exposure Indicators Exposure

More information

Nuclear Associates

Nuclear Associates Nuclear Associates 76-700 Digital Subtraction Angiography Phantom Users Manual March 2005 Manual No. 76-700-1 Rev. 2 2004, 2005 Fluke Corporation, All rights reserved. Printed in U.S.A. All product names

More information

Photons and solid state detection

Photons and solid state detection Photons and solid state detection Photons represent discrete packets ( quanta ) of optical energy Energy is hc/! (h: Planck s constant, c: speed of light,! : wavelength) For solid state detection, photons

More information

CHAPTER 6 QC Test For Fluoroscopic Equipment. Prepared by:- Kamarul Amin bin Abu Bakar School of Medical Imaging KLMUC

CHAPTER 6 QC Test For Fluoroscopic Equipment. Prepared by:- Kamarul Amin bin Abu Bakar School of Medical Imaging KLMUC CHAPTER 6 QC Test For Fluoroscopic Equipment Prepared by:- Kamarul Amin bin Abdullah @ Abu Bakar School of Medical Imaging KLMUC Lesson Outcomes Describe the objectives of each QC test done. Identify QC

More information

COPYRIGHT 2002 by Srinivasan Vedantham ALL RIGHTS RESERVED. Use or inclusion of any portion of this document in another work intended for

COPYRIGHT 2002 by Srinivasan Vedantham ALL RIGHTS RESERVED. Use or inclusion of any portion of this document in another work intended for COPYRIGHT 00 by Srinivasan Vedantham ALL RIGHTS RESERVED Use or inclusion of any portion of this document in another work intended for commercial use will require permission from the copyright owner. ii

More information

Technical Notes. Introduction. Optical Properties. Issue 6 July Figure 1. Specular Reflection:

Technical Notes. Introduction. Optical Properties. Issue 6 July Figure 1. Specular Reflection: Technical Notes This Technical Note introduces basic concepts in optical design for low power off-grid lighting products and suggests ways to improve optical efficiency. It is intended for manufacturers,

More information

CHAPTER 10. DIGITAL SUBTRACTION ANGIOGRAPHY (DSA) 10.1 Introduction

CHAPTER 10. DIGITAL SUBTRACTION ANGIOGRAPHY (DSA) 10.1 Introduction PHYSICS OF MEDICAL X-RAY IMAGING (1)Chapter 10 CHAPTER 10. DIGITAL SUBTRACTION ANGIOGRAPHY (DSA) 10.1 Introduction In this chapter we will discuss a technique that is known as digital subtraction angiography

More information

SYLLABUS. TITLE: Equipment Operation I. DEPARTMENT: Radiologic Technology

SYLLABUS. TITLE: Equipment Operation I. DEPARTMENT: Radiologic Technology CODE: RADT 156 INSTITUTE: Health Science TITLE: Equipment Operation I DEPARTMENT: Radiologic Technology COURSE DESCRIPTION: This course covers the principles of equipment operation and maintenance of radiographic

More information

CR Basics and FAQ. Overview. Historical Perspective

CR Basics and FAQ. Overview. Historical Perspective Page: 1 of 6 CR Basics and FAQ Overview Computed Radiography is a term used to describe a system that electronically records a radiographic image. Computed Radiographic systems use unique image receptors

More information

Ground-based optical auroral measurements

Ground-based optical auroral measurements Ground-based optical auroral measurements FYS 3610 Background Ground-based optical measurements provides a unique way to monitor spatial and temporal variation of auroral activity at high resolution up

More information

Detectors for microscopy - CCDs, APDs and PMTs. Antonia Göhler. Nov 2014

Detectors for microscopy - CCDs, APDs and PMTs. Antonia Göhler. Nov 2014 Detectors for microscopy - CCDs, APDs and PMTs Antonia Göhler Nov 2014 Detectors/Sensors in general are devices that detect events or changes in quantities (intensities) and provide a corresponding output,

More information

Exposure System Selection

Exposure System Selection Principles of Imaging Science II (RAD120) Exposure Systems Exposure System Selection Radiographic exposure is a very complex process Best technique systems manipulate one variable while holding others

More information

Fluoroscopy & Tomography. Rhodes

Fluoroscopy & Tomography. Rhodes Fluoroscopy & Tomography Rhodes 10-526-194 1 Historical Development Dynamic examination Active Diagnosis Domain of radiologist Fluoroscope Invented by Edison in 1896 2 Fluoroscopic Imaging Chain Components

More information

Enhanced Functionality of High-Speed Image Processing Engine SUREengine PRO. Sharpness (spatial resolution) Graininess (noise intensity)

Enhanced Functionality of High-Speed Image Processing Engine SUREengine PRO. Sharpness (spatial resolution) Graininess (noise intensity) Vascular Enhanced Functionality of High-Speed Image Processing Engine SUREengine PRO Medical Systems Division, Shimadzu Corporation Yoshiaki Miura 1. Introduction In recent years, digital cardiovascular

More information

X-ray light valve (XLV): a novel detectors technology for digital mammography*

X-ray light valve (XLV): a novel detectors technology for digital mammography* X-ray light valve (XLV): a novel detectors technology for digital mammography* Sorin Marcovici, Vlad Sukhovatkin, Peter Oakham XLV Diagnostics Inc., Thunder Bay, ON P7A 7T1, Canada ABSTRACT A novel method,

More information

X-Ray Medical Imaging and Pixel detectors

X-Ray Medical Imaging and Pixel detectors X-Ray Medical Imaging and Pixel detectors PIXEL 2000 Genova, June 5-8 th 2000 J.P.Moy, TRI XELL, Moirans, France 1 OUTLINE - X-ray medical imaging. The requirements, some particular features - Present

More information

3/31/2011. Objectives. Emory University. Historical Development. Historical Development. Historical Development

3/31/2011. Objectives. Emory University. Historical Development. Historical Development. Historical Development Teaching Radiographic Technique in a Digital Imaging Paradigm Objectives 1. Discuss the historical development of digital imaging. Dawn Couch Moore, M.M.Sc., RT(R) Assistant Professor and Director Emory

More information

Advances in X-Ray Scintillator Technology Roger D. Durst Bruker AXS Inc.

Advances in X-Ray Scintillator Technology Roger D. Durst Bruker AXS Inc. Advances in X-Ray Scintillator Technology Roger D. Durst Inc. Acknowledgements T. Thorson, Y. Diawara, E. Westbrook, MBC J. Morse, ESRF C. Summers, Georgia Tech/PTCE B. Wagner, Georgia Tech/PTCE V. Valdna,

More information

Spectroscopy in the UV and Visible: Instrumentation. Spectroscopy in the UV and Visible: Instrumentation

Spectroscopy in the UV and Visible: Instrumentation. Spectroscopy in the UV and Visible: Instrumentation Spectroscopy in the UV and Visible: Instrumentation Typical UV-VIS instrument 1 Source - Disperser Sample (Blank) Detector Readout Monitor the relative response of the sample signal to the blank Transmittance

More information

MASSACHUSETTS INSTITUTE OF TECHNOLOGY Mechanical Engineering Department. 2.71/2.710 Final Exam. May 21, Duration: 3 hours (9 am-12 noon)

MASSACHUSETTS INSTITUTE OF TECHNOLOGY Mechanical Engineering Department. 2.71/2.710 Final Exam. May 21, Duration: 3 hours (9 am-12 noon) MASSACHUSETTS INSTITUTE OF TECHNOLOGY Mechanical Engineering Department 2.71/2.710 Final Exam May 21, 2013 Duration: 3 hours (9 am-12 noon) CLOSED BOOK Total pages: 5 Name: PLEASE RETURN THIS BOOKLET WITH

More information

SPRINGFIELD TECHNICAL COMMUNITY COLLEGE ACADEMIC AFFAIRS

SPRINGFIELD TECHNICAL COMMUNITY COLLEGE ACADEMIC AFFAIRS SPRINGFIELD TECHNICAL COMMUNITY COLLEGE ACADEMIC AFFAIRS Course Number: RADG 212 Department: Radiography Course Title: Equip. Operation & Maint. Semester: Spring Year: 1997 Objectives/ Unit One: The X-ray

More information

Nuclear Associates

Nuclear Associates Nuclear Associates 07-649 CDRH Fluoroscopic Phantom Users Manual March 2005 Manual No. 07-649-1 Rev. 2 2004, 2005 Fluke Corporation, All rights reserved. Printed in U.S.A. All product names are trademarks

More information

High Performance Thin Film Optical Coatings Technical Reference Document 09/13. Coatings Capabilities. Heat Control - Hot Mirror Filters

High Performance Thin Film Optical Coatings Technical Reference Document 09/13. Coatings Capabilities. Heat Control - Hot Mirror Filters Heat Control - Hot Mirror Filters A hot mirror is in essence a thin film coating applied to substrates in an effort to reflect infra-red radiation either as a means to harness the reflected wavelengths

More information

PERFORMANCE CHARACTERIZATION OF AMORPHOUS SILICON DIGITAL DETECTOR ARRAYS FOR GAMMA RADIOGRAPHY

PERFORMANCE CHARACTERIZATION OF AMORPHOUS SILICON DIGITAL DETECTOR ARRAYS FOR GAMMA RADIOGRAPHY 12 th A-PCNDT 2006 Asia-Pacific Conference on NDT, 5 th 10 th Nov 2006, Auckland, New Zealand PERFORMANCE CHARACTERIZATION OF AMORPHOUS SILICON DIGITAL DETECTOR ARRAYS FOR GAMMA RADIOGRAPHY Rajashekar

More information

X-rays. X-rays are produced when electrons are accelerated and collide with a target. X-rays are sometimes characterized by the generating voltage

X-rays. X-rays are produced when electrons are accelerated and collide with a target. X-rays are sometimes characterized by the generating voltage X-rays Ouch! 1 X-rays X-rays are produced when electrons are accelerated and collide with a target Bremsstrahlung x-rays Characteristic x-rays X-rays are sometimes characterized by the generating voltage

More information

Today s Outline - January 25, C. Segre (IIT) PHYS Spring 2018 January 25, / 26

Today s Outline - January 25, C. Segre (IIT) PHYS Spring 2018 January 25, / 26 Today s Outline - January 25, 2018 C. Segre (IIT) PHYS 570 - Spring 2018 January 25, 2018 1 / 26 Today s Outline - January 25, 2018 HW #2 C. Segre (IIT) PHYS 570 - Spring 2018 January 25, 2018 1 / 26 Today

More information

Page 1. Ground-based optical auroral measurements. Background. CCD All-sky Camera with filterwheel. Image intensifier

Page 1. Ground-based optical auroral measurements. Background. CCD All-sky Camera with filterwheel. Image intensifier Ground-based optical auroral measurements FYS 3610 Background Ground-based optical measurements provides a unique way to monitor spatial and temporal variation of auroral activity at high resolution up

More information

Visibility of Detail

Visibility of Detail Visibility of Detail Radiographic Quality Quality radiographic images represents the, and information is for diagnosis. The of the anatomic structures and the accuracy of their ( ) determine the overall

More information

Mammography is a radiographic procedure specially designed for detecting breast pathology Approximately 1 woman in 8 will develop breast cancer over

Mammography is a radiographic procedure specially designed for detecting breast pathology Approximately 1 woman in 8 will develop breast cancer over Mammography is a radiographic procedure specially designed for detecting breast pathology Approximately 1 woman in 8 will develop breast cancer over a lifetime Breast cancer screening programs rely on

More information

The importance of radiation quality for optimisation in radiology

The importance of radiation quality for optimisation in radiology Available online at http://www.biij.org/2007/2/e38 doi: 10.2349/biij.3.2.e38 biij Biomedical Imaging and Intervention Journal COMMENTARY The importance of radiation quality for optimisation in radiology

More information

Gentec Electro-Optics, Inc

Gentec Electro-Optics, Inc Gentec Electro-Optics, Inc. 2013. Accessories for Beam Diagnostics Revision 2.0 2 WARRANTY The Gentec-EO accessories for beam diagnostics carry a one-year warranty (from date of shipment) against material

More information

IMAGE SENSOR SOLUTIONS. KAC-96-1/5" Lens Kit. KODAK KAC-96-1/5" Lens Kit. for use with the KODAK CMOS Image Sensors. November 2004 Revision 2

IMAGE SENSOR SOLUTIONS. KAC-96-1/5 Lens Kit. KODAK KAC-96-1/5 Lens Kit. for use with the KODAK CMOS Image Sensors. November 2004 Revision 2 KODAK for use with the KODAK CMOS Image Sensors November 2004 Revision 2 1.1 Introduction Choosing the right lens is a critical aspect of designing an imaging system. Typically the trade off between image

More information

Digital Radiography. Selected Topics

Digital Radiography. Selected Topics Digital Radiography Selected Topics DIGITAL RADIOGRAPHY Selected Topics Editorial Advisory Board: PETER R. ALMOND, Ph.D. University of Louisville School of Medicine Louisville, Kentucky JOHN S. CLIFTON,

More information

SONIALVISION G4 Multi-purpose Digital R/F System C506-E075

SONIALVISION G4 Multi-purpose Digital R/F System C506-E075 SONIALVISION G4 Multi-purpose Digital R/F System C506-E075 Selecting the best for every examination environment BEST in CLASS Multi-purpose Digital R/F System 2 With the Sonialvision G4, Shimadzu offers

More information

SONIALVISION G4 Multi-purpose Digital R/F System C506-E075

SONIALVISION G4 Multi-purpose Digital R/F System C506-E075 SONIALVISION G4 Multi-purpose Digital R/F System C506-E075 Selecting the best for every examination environment BEST in CLASS Multi-purpose Digital R/F System With the Sonialvision G4, Shimadzu offers

More information

LlIGHT REVIEW PART 2 DOWNLOAD, PRINT and submit for 100 points

LlIGHT REVIEW PART 2 DOWNLOAD, PRINT and submit for 100 points WRITE ON SCANTRON WITH NUMBER 2 PENCIL DO NOT WRITE ON THIS TEST LlIGHT REVIEW PART 2 DOWNLOAD, PRINT and submit for 100 points Multiple Choice Identify the choice that best completes the statement or

More information

Interventional Radiological Equipment selection and installation

Interventional Radiological Equipment selection and installation Interventional Radiological Equipment selection and installation Renato Padovani ICTP Learning objectives To understand the main components of an interventional radiology equipment To understand the relevance

More information

Medical Device Manufacturing: Designing for X-ray Inspection. Gil Zweig, President Glenbrook Technologies.

Medical Device Manufacturing: Designing for X-ray Inspection. Gil Zweig, President Glenbrook Technologies. Medical Device Manufacturing: Designing for X-ray Inspection. Gil Zweig, President Glenbrook Technologies. Introduction When x-ray inspection is used as part of a quality assurance program for any assembled

More information

Vision Lighting Seminar

Vision Lighting Seminar Creators of Evenlite Vision Lighting Seminar Daryl Martin Midwest Sales & Support Manager Advanced illumination 734-213 213-13121312 dmartin@advill.com www.advill.com 2005 1 Objectives Lighting Source

More information

VISUAL PHYSICS ONLINE DEPTH STUDY: ELECTRON MICROSCOPES

VISUAL PHYSICS ONLINE DEPTH STUDY: ELECTRON MICROSCOPES VISUAL PHYSICS ONLINE DEPTH STUDY: ELECTRON MICROSCOPES Shortly after the experimental confirmation of the wave properties of the electron, it was suggested that the electron could be used to examine objects

More information

RAD 150 RADIOLOGIC EXPOSURE TECHNIQUE II

RAD 150 RADIOLOGIC EXPOSURE TECHNIQUE II RAD 150 RADIOLOGIC EXPOSURE TECHNIQUE II APPROVED 12/O2/2011 EFFECTIVE SPRING 2013-14 Prefix & Number RAD 150 Course Title: Radiologic Exposure Technique II & Lab Purpose of this submission: New Change/Updated

More information

BROADCAST ENGINEERING 5/05 WHITE PAPER TUTORIAL. HEADLINE: HDTV Lens Design: Management of Light Transmission

BROADCAST ENGINEERING 5/05 WHITE PAPER TUTORIAL. HEADLINE: HDTV Lens Design: Management of Light Transmission BROADCAST ENGINEERING 5/05 WHITE PAPER TUTORIAL HEADLINE: HDTV Lens Design: Management of Light Transmission By Larry Thorpe and Gordon Tubbs Broadcast engineers have a comfortable familiarity with electronic

More information

Mammography: Physics of Imaging

Mammography: Physics of Imaging Mammography: Physics of Imaging Robert G. Gould, Sc.D. Professor and Vice Chair Department of Radiology and Biomedical Imaging University of California San Francisco, California Mammographic Imaging: Uniqueness

More information

TECHNICAL DATA. GIOTTO IMAGE SDL/W is pre-arranged for Full Field Digital Biopsy examination with the patient in prone position.

TECHNICAL DATA. GIOTTO IMAGE SDL/W is pre-arranged for Full Field Digital Biopsy examination with the patient in prone position. Ver. 01/06/07 TECHNICAL DATA GIOTTO IMAGE SDL/W LOW DOSE, FULL FIELD DIGITAL MAMMOGRAPHY UNIT USING AMORPHOUS SELENIUM (a-se) TECHNOLOGY DETECTOR (pre-arranged for stereotactic biopsy with the same digital

More information

SONIALVISION G4 Multi-purpose Digital R/F System

SONIALVISION G4 Multi-purpose Digital R/F System C506-E075A SONIALVISION G4 Multi-purpose Digital R/F System Founded in 1875, Shimadzu Corporation, a leader in the development of advanced technologies, has a distinguished history of innovation built

More information