1 Nuclear Associates Digital Subtraction Angiography Phantom Users Manual March 2005 Manual No Rev , 2005 Fluke Corporation, All rights reserved. Printed in U.S.A. All product names are trademarks of their respective companies
2 Fluke Biomedical Radiation Management Services 6045 Cochran Road Cleveland, Ohio
3 Introduction Introduction Introduction Section 1 Introduction This unique QA test tool provides a reliable, accurate, and easy means of evaluating the digital functions of DSA equipment. * Important performance parameters are measured by a system of five clear-acrylic building blocks: 1 step wedge, 1 bone block, 1 resolution block, 1 artery block, and 1 slot block. Each has a specific function. However, some of the modules can be combined to perform additional tests. For example, the step wedge has two functions: (1) It serves as a contrast measurement wedge, and (2) it "folds" on itself to become a 3-inch thick solid body block. This system versatility assures maximum performance with minimal effort and cost. 1.2 System Components The phantom consists of five clear-acrylic building blocks: 1 step wedge, 1 bone block, 1 resolution block, 1 artery block, and 1 slot block (see Figure 1). Figure 1-1. System Building Blocks Step Wedge This module tests a machine's contrast range by its ability to visualize all steps. The six acrylic steps are each 2.5 cm high. The top three steps can be "folded" over the bottom three to form a solid block, 20 x 20 x 7.5 cm thick. When used with the slot block, it simulates the patient's body. *Non-digital functions of DSA systems can be checked with conventional fluoroscopic test equipment such as: Fluoroscopic System Resolution Test Tools (07-601/07-619) Rotatable Spoke Test Pattern (07-629) Fluoroscopic Quality Control Test Tool (07-637) Fluoro-Test Tool (07-645) Patient Phantom/Penetrometer System (07-706) 1-1
4 Nuclear Associates Operators Manual Bone Block This 20 x 20 x 2.5 cm thick piece contains simulated calcium bones (20 x 2.5 cm) in three thicknesses-- 0.5, 1.0 and 1.5 cm. When the bone block is positioned over the other blocks, the subtraction function can be tested by the degree to which the simulated bones remain visible in the final subtracted images Resolution Block There are two precision test patterns in this 15 x 20 x 2.5 cm module. The patterns have line-pair groups, from 0.6 to 5.0 LP/mm, with a 0.01 mm lead thickness (Test Pattern ) Artery Block This 15 x 45 x 2.5 cm piece contains three iodine filled simulated arteries whose width and depth are 1, 2 and 4 mm respectively. Each artery includes simulated stenoses and aneurysms that are one-fourth, onehalf, and three-fourths of the individual artery's width. The iodine concentration is 15 mg/ ML. The module also has three circles 1 cm diameter, each with a different iodine content (1.5, 3 and 6 mg/cm 2 ) for testing linearity. Iodine concentrations of 150 mg per ML and 300 mg per ML are also available Slot Block The center of this 20 x 20 x 7.5 cm thick module has a slot that holds and positions the resolution and artery sections. The block is also used to increase "body" thickness by 7.5 cm Blank Block This is not a separate module but part of the elongated artery block. The blank portion can be positioned in the slot block to make the mask image. 1.3 Directions General 1. Program a delay of 10 to 15 seconds between the mask and image exposures. Use this same delay for all images. 2. The first time you use the phantom, find and record the best technique to use for each test. Include all details such as aperture, geometry, gain, kilovoltage, current and exposure time. Continue to use these same values when repeating the tests. 3. Use the 6-inch image intensifier mode, or whichever mode is used most often. Always use the same mode for testing Wedge Images For testing video system stability, contrast, dynamic range and system noise. 1. Place the step wedge (6 steps) on the table (see Figure 1-2), with steps at 45 to the video scan (raster) lines. Figure 1-2. Step Wedge (unfolded). 1-2
5 Introduction Directions 1 2. Make sure the image intensifier grid is in place. 3. Use a standard, repeatable geometry for the exposure (e. g., an 80 cm source-to-image-intensifier distance and a 20 cm object-to-image-intensifier distance). Position the wedge, using the fluoro mode. A "scout" or test image will indicate when the correct video level has been set. 4. Make two exposures of the wedge--one for the mask and one for the test. Subtract one from the other. Make a film of the subtracted image. Note that all films can be made at one time--at the end of the testing, if desired Resolution Test Images For checking low-contrast spatial resolution: 1. Place the phantom on the table as shown in Figure 1-3. Do not use the bone block. 2. Adjust the object-to-image-intensifier distance to 15 cm and center the phantom. 3. Make sure that the image intensifier grid is in place. Figure 1-3. Blocks Arranged for Testing Resolution Figure 1-4. Artery Block Patterns 1-3
6 Nuclear Associates Operators Manual 4. With the blank end of the artery block inserted in the slot, make a mask exposure. You have only 10 or 15 seconds to remove the artery block and insert the resolution block so that the patterns are centered in the slot block. Allow the image exposure to be made. 5. Subtract the mask image from the resolution image. Make a film (enlarge the image with the zoom control, if desired). Note that the image perpendicular to the scan or raster lines is the image to be examined. Because of the visibility of the raster lines, the image of the pattern that runs parallel to them may give false readings. Nevertheless, such a reading may be valuable Artery Detectability Images For checking the low-contrast iodine image visibility and linearity and quality of subtraction (see Figure 1-4). 1. Set the standard phantom up as in Figure 1-3, and add the bone block to the top of the phantom. Be sure that the direction of the simulated bones is perpendicular to that of the iodine patterns and parallel to the long dimension of the insert block itself. 2. Keep the image intensifier grid in place if it is normally used clinically. If not, make sure it is removed. 3. With the blank end of the artery-iodine block inserted in the slot, make a mask exposure. You have 10 or 15 seconds to push toe block through so that the artery pattern is in place. Allow the image exposure to be made. 4. Subtract the mask image from the artery image. Make a film (enlarge various parts of the image with the zoom control, if desired). 5. The lack of prominence of the bone images in the subtracted image is a measure of the subtraction effectiveness. 6. Make another subtraction image, exactly as in steps 1 to 3, but without the bone block. 7. With a densitometer, measure the film densities of the centers of the three 1cm circles that contain 1.5, 3 and 6 mg/cm 2 iodine. 8. Subtract the density readings of each iodine concentration from that of the next higher concentration, obtaining D(6) - D(3), and D(3) - D(1.5), where "D" is the density reading. 9. The two differences would be equal (approx. 0.3) were it not for unavoidable non-linearity such as beam hardening by the iodine, scattering, and film H-D curve non-linearity, among other factors. However, such measurements (made as part of a QA procedure) can be compared with reference measurements made earlier, when the system is known to be operating satisfactorily. Of course, the measurements must be made under precisely the same conditions of technique, using the same film, etc. Note that problems with either the hard copy camera and or the film processor might also result in changes in the density differences Exposure Measurements Phantom entrance and exit exposure measurements can be make using the RAD-CHECK PLUS X-Ray Exposure Meter or an equivalent instrument. Use the technique (kvp, tube current, and exposure time) that is normally used for quality control measurements. 1. Screw the four steel legs into the threaded holes in the slot block, and stand the block on a table. Place the blank end of either the artery or the resolution block into the slot. Finally, position the folded step wedge on the slot block. Do not use the bone block. 2. Set the technique at 80 kvp and 10 mas--or at your normal clinical settings, as desired. Do not use the grid. 3. Entrance and exit exposures are made with the RAD-CHECK or ion chamber placed above and below the phantom. The location of the radiation source will determine whether the measurement is the entrance or the exit exposure. 1-4
7 (Blank Page)
8 Fluke Biomedical Radiation Management Services 6045 Cochran Road Cleveland, Ohio
Nuclear Associates
Nuclear Associates 07-591 Focal Spot Test Tool Users Manual February 2005 Manual No. 07-591-1 Rev. 2 2004, 2005 Fluke Corporation, All rights reserved. Printed in U.S.A. All product names are trademarks
More informationNuclear Associates
Nuclear Associates 07-706 Patient Phantom/Penetrometer System Users Manual March 2005 Manual No. 07-706-1 Rev. 2 2004, 2005 Fluke Corporation, All rights reserved. Printed in U.S.A. All product names are
More informationNuclear Associates EZ CR-DIN Phantoms
Nuclear Associates 07-605-7777 EZ CR-DIN Phantoms Users Manual August 2006 Manual No. 07-605-7777-1 Rev. 4 2006 Fluke Corporation, All rights reserved. Printed in U.S.A. All product names are trademarks
More informationNuclear Associates
Nuclear Associates 07-649 CDRH Fluoroscopic Phantom Users Manual March 2005 Manual No. 07-649-1 Rev. 2 2004, 2005 Fluke Corporation, All rights reserved. Printed in U.S.A. All product names are trademarks
More informationNuclear Associates
Nuclear Associates 07-647 R/F QC Phantom Operators Manual March 2005 Manual No. 07-647-1 Rev. 2 2004, 2005 Fluke Corporation, All rights reserved. All product names are trademarks of their respective companies
More informationNuclear Associates
Nuclear Associates 07-644 Grid Alignment Test Tool Users Manual March 2005 Manual No. 07-644-1 Rev. 2 2004, 2005 Fluke Corporation, All rights reserved. Printed in U.S.A. All product names are trademarks
More informationNuclear Associates , , CT Head and Body Dose Phantom
Nuclear Associates 76-414,76-414-4150,76-415 CT Head and Body Dose Phantom Users Manual March 2005 Manual No. 76-414-1 Rev. 2 2004, 2005 Fluke Corporation, All rights reserved. Printed in U.S.A. All product
More informationI. PERFORMANCE OF X-RAY PRODUCTION COMPONENTS FLUOROSCOPIC ACCEPTANCE TESTING: TEST PROCEDURES & PERFORMANCE CRITERIA
FLUOROSCOPIC ACCEPTANCE TESTING: TEST PROCEDURES & PERFORMANCE CRITERIA EDWARD L. NICKOLOFF DEPARTMENT OF RADIOLOGY COLUMBIA UNIVERSITY NEW YORK, NY ACCEPTANCE TESTING GOALS PRIOR TO 1st CLINICAL USAGE
More informationNuclear Associates , &
Nuclear Associates 76-810, 76-814 76-815 & 76-818 Bar Phantoms and Test Patterns Operators Manual March 2005 Manual No. 76-810-1 Rev. 2 2004, 2005 Fluke Corporation, All rights reserved. Printed in U.S.A.
More informationNuclear Associates , , , , , ,
Nuclear Associates 57-411, 57-412, 57-413 57-426, 57-431, 57-432 57-433, 57-435, 57-436 CLEAR-Pb Transparent X-Ray Compensation Filters Users Manual March 2005 Manual No. 57-XXX-1 Rev. 2 2003, 2005 Fluke
More informationNEMA. Cardiology Phantom Nuclear Associates Model Diagnostic Imaging. Introduction. Features. Applications
NEMA Cardiology Phantom Nuclear Associates Model 07-680 Diagnostic Imaging! Phantom and test procedures simulate a range of fluoroscopically-guided invasive and interventional procedures! Provides simultaneous
More informationCHAPTER 6 QC Test For Fluoroscopic Equipment. Prepared by:- Kamarul Amin bin Abu Bakar School of Medical Imaging KLMUC
CHAPTER 6 QC Test For Fluoroscopic Equipment Prepared by:- Kamarul Amin bin Abdullah @ Abu Bakar School of Medical Imaging KLMUC Lesson Outcomes Describe the objectives of each QC test done. Identify QC
More informationNuclear Associates , &
Nuclear Associates 76-823, 76-824 & 76-825 PET/SPECT Phantom Source Tank, Phantom Inserts and Cardiac Insert Users Manual March 2005 Manual No. 76-823-1 Rev. 2 2004, 2005 Fluke Corporation, All rights
More informationNuclear Associates
Nuclear Associates 05-582 PRIMA 7 Transport Index Radiation Monitor Operators Manual March 2005 Manual No. 135011 Rev. 3 2004, 2005 Fluke Corporation, All rights reserved. Printed in U.S.A. All product
More informationIntroduction. Chapter 16 Diagnostic Radiology. Primary radiological image. Primary radiological image
Introduction Chapter 16 Diagnostic Radiology Radiation Dosimetry I Text: H.E Johns and J.R. Cunningham, The physics of radiology, 4 th ed. http://www.utoledo.edu/med/depts/radther In diagnostic radiology
More informationExposure System Selection
Principles of Imaging Science II (RAD120) Exposure Systems Exposure System Selection Radiographic exposure is a very complex process Best technique systems manipulate one variable while holding others
More informationNuclear Associates
Nuclear Associates 07-424 Digital Densitometer II Operators Manual March 2005 Manual No. 112111 Rev. 4 2003, 2005 Fluke Corporation, All rights reserved. Printed U.S.A. All product names are trademarks
More informationNuclear Associates &
Nuclear Associates 06-526 & 06-526-2200 RAD-CHECK PLUS Operators Manual March 2005 Manual No. 136201 Rev. 5 2003, 2005 Fluke Corporation, All rights reserved. Printed in U.S.A. All product names are trademarks
More informationBeam-Restricting Devices
Beam-Restricting Devices Three factors contribute to an increase in scatter radiation: Increased kvp Increased Field Size Increased Patient or Body Part Size. X-ray Interactions a some interact with the
More informationY11-DR Digital Radiography (DR) Image Quality
Y11-DR Digital Radiography (DR) Image Quality Image quality is stressed for all systems in Safety Code 35. In the relevant sections Health Canada s advice is the manufacturer s recommended test procedures
More informationX-RAYS - NO UNAUTHORISED ENTRY
Licencing of premises Premises Refer Guidelines A radiation warning sign and warning notice, X-RAYS - NO UNAUTHORISED ENTRY must be displayed at all entrances leading to the rooms where x-ray units are
More informationRAD 150 RADIOLOGIC EXPOSURE TECHNIQUE II
RAD 150 RADIOLOGIC EXPOSURE TECHNIQUE II APPROVED 12/O2/2011 EFFECTIVE SPRING 2013-14 Prefix & Number RAD 150 Course Title: Radiologic Exposure Technique II & Lab Purpose of this submission: New Change/Updated
More informationLudlum Medical Physics
Ludlum Medical Physics Medical Imaging Radiology QA Test Tools NEW LUDLUM PRODUCT LINE Medical Physics Products Medical Physics Products What are they? Products used to measure radiation output and to
More informationSPRINGFIELD TECHNICAL COMMUNITY COLLEGE ACADEMIC AFFAIRS
SPRINGFIELD TECHNICAL COMMUNITY COLLEGE ACADEMIC AFFAIRS Course Number: RADG 112 Department: Radiography Course Title: Image Production & Eval. Semester: Spring Year: 1997 Objectives/ Unit One: Introduction
More informationNuclear Associates and
Nuclear Associates 76-410-4130 and 76-411 AAPM CT Performance Phantom Users Manual March 2005 Manual No. 76-410-4130-1 Rev. 2 2004, 2005 Fluke Corporation, All rights reserved. Printed in U.S.A. All product
More informationNuclear Associates
Nuclear Associates 07-443 Clamshell Densitometer Users Manual February 2005 Manual No. 166001 Rev. 5 2004, 2005 Fluke Corporation, All rights reserved. Printed in U.S.A. All product names are trademarks
More informationMultiple Choice Identify the letter of the choice that best completes the statement or answers the question.
RA110 test 3 Multiple Choice Identify the letter of the choice that best completes the statement or answers the question. 1. An object 35 cm in width is radiographed at 100 cm SID and at a 50 cm SOD. What
More informationNuclear Associates
Nuclear Associates 07-453 Digital X-Ray Pulse Counter/Timer Operators Manual February 2005 Manual No. 07-453-1 Rev. 2 2004, 2005 Fluke Corporation, All rights reserved. Printed in U.S.A. All product names
More informationSeminar 8. Radiology S8 1
Seminar 8 Radiology Medical imaging. X-ray image formation. Energizing and controlling the X-ray tube. Image detectors. The acquisition of analog and digital images. Digital image processing. Selected
More informationTESTING FLAT-PANEL IMAGING SYSTEMS: What the Medical Physicist Needs to Know. JAMES A. TOMLINSON, M.S., D.A.B.R. Diagnostic Radiological Physicist
TESTING FLAT-PANEL IMAGING SYSTEMS: What the Medical Physicist Needs to Know JAMES A. TOMLINSON, M.S., D.A.B.R. Diagnostic Radiological Physicist Topics Image Uniformity and Artifacts Image Quality - Detail
More informationVeterinary Science Preparatory Training for the Veterinary Assistant. Floron C. Faries, Jr., DVM, MS
Veterinary Science Preparatory Training for the Veterinary Assistant Floron C. Faries, Jr., DVM, MS Radiology Floron C. Faries, Jr., DVM, MS Objectives Determine the appropriate machine settings for making
More informationDose Reduction and Image Preservation After the Introduction of a 0.1 mm Cu Filter into the LODOX Statscan unit above 110 kvp
Dose Reduction and Image Preservation After the Introduction of a into the LODOX Statscan unit above 110 kvp Abstract: CJ Trauernicht 1, C Rall 1, T Perks 2, G Maree 1, E Hering 1, S Steiner 3 1) Division
More informationApplications Note. Successfully using the Model 8000 NERO dental x-ray machine timer performance. max to evaluate
Successfully using the Model 8000 NERO dental x-ray machine timer performance max to evaluate Information in this article applies to Model 8000 NERO max firmware release 2.3 or higher. Introduction The
More informationSYLLABUS. 1. Identification of Subject:
SYLLABUS Date/ Revision : 30 January 2017/1 Faculty : Life Sciences Approval : Dean, Faculty of Life Sciences SUBJECT : Biophysics 1. Identification of Subject: Name of Subject : Biophysics Code of Subject
More informationMAN Revision 002. InSight Mini C-arm Imaging System Technical Reference Manual
MAN-00668 Revision 002 InSight Mini C-arm Imaging System Technical Reference Manual April 2007 The information contained in this Manual is confidential and proprietary to Hologic, Inc. This information
More informationJoint ICTP/IAEA Advanced School on Dosimetry in Diagnostic Radiology and its Clinical Implementation May 2009
2033-6 Joint ICTP/IAEA Advanced School on Dosimetry in Diagnostic Radiology and its Clinical Implementation 11-15 May 2009 Dosimetry for Fluoroscopy Basics Renato Padovani EFOMP Joint ICTP-IAEA Advanced
More informationNuclear Associates
Nuclear Associates 07-492 Mammographic Digital kvp Meter Operators Manual March 2005 Manual No. 164001 Rev. 3 2003, 2005 Fluke Corporation, All rights reserved. Printed in U.S.A. All product names are
More informationBASICS OF FLUOROSCOPY
Medical Physics Residents Training Program BASICS OF FLUOROSCOPY Dr. Khalid Alyousef, PhD Department of Medical Imaging King Abdulaziz Medical City- Riyadh Edison examining the hand of Clarence Dally with
More informationEffect of patient support pads on image quality and dose in fluoroscopy a)
Effect of patient support pads on image quality and dose in fluoroscopy a) William R. Geiser, Walter Huda, b) and Nikolaos A. Gkanatsios Department of Radiology, University of Florida, Gainesville, Florida
More informationC-ARM FLUOROSCOPIC AND SPOT-FILM SYSTEMS
PART X C-ARM FLUOROSCOPIC AND SPOT-FILM SYSTEMS FORM FD 3260 REPRINTED APRIL 2000 ROUTINE COMPLIANCE TESTING C-ARM FLUOROSCOPES (Test Procedure CFA - Use Form FDA 3260) 1.0 GENERAL GUIDANCE 1.1 This procedure
More informationDISC QC/QA Program for Digital Imaging Systems using the DR Radchex Plus Meter
DISC QC/QA Program for Digital Imaging Systems using the DR Radchex Plus Meter Revision Date: January 5th, 2017 www.disc-imaging.com Table of Contents Section A: Preliminary Setup Requirements... 4 Tools
More information10/15/2012 SECTION III - CHAPTER 6 DIGITAL FLUOROSCOPY RADT 3463 COMPUTERIZED IMAGING
RADT 3463 - COMPUTERIZED IMAGING Section III: Chapter 6 RADT 3463 Computerized Imaging 1 SECTION III - CHAPTER 6 DIGITAL FLUOROSCOPY RADT 3463 COMPUTERIZED IMAGING Section III: Chapter 6 RADT 3463 Computerized
More information3D Diode Array Commissioning: Building Confidence in 3D QA Technology
3D Diode Array Commissioning: Building Confidence in 3D QA Technology Caroline Yount, MS CANCER CENTER 3D QA The complex three-dimensional (3D) shapes of intensity modulated radiation therapy (IMRT) dose
More informationLess effort. More insight.
Less effort. More insight. Product Catalogue C O N T E N T P3~P4 Quality assurance and service of diagnostic X-ray equipment P5~P6 RaySafe X2 P7~P10 RaySafe Xi Classic P11~P13 RaySafe Phantom P14~P15 451P-DE-SI
More informationPhotons interaction with matter
ب س م هللا الر ح من الر حیم Photons interaction with matter Ionization Ionization is the process of removing an electron from an electrically neutral atom to produce an ion pair. An ion is an atom or subatomic
More informationOverview of Safety Code 35
Common Quality Control Procedures for All s Quality Control Procedures Film All s Daily Quality Control Tests Equipment Warm-up (D1) According to manufacturers instructions Can include auto calibration(d1)
More informationQC Testing for Computed Tomography (CT) Scanner
QC Testing for Computed Tomography (CT) Scanner QA - Quality Assurance All planned and systematic actions needed to provide confidence on a structure, system or component. all-encompassing program, including
More informationDigital Imaging started in the 1972 with Digital subtraction angiography Clinical digital imaging was employed from the 1980 ~ 37 years ago Amount of
Digital Imaging started in the 1972 with Digital subtraction angiography Clinical digital imaging was employed from the 1980 ~ 37 years ago Amount of radiation to the population due to Medical Imaging
More informationTest Equipment for Radiology and CT Quality Control Contents
Test Equipment for Radiology and CT Quality Control Contents Quality Control Testing...2 Photometers for Digital Clinical Display QC...3 Primary Workstations...3 Secondary Workstations...3 Testing of workstations...3
More informationSTUDENT REVIEW QUESTION SET K CR/DR CONTENT AREA
STUDENT REVIEW QUESTION SET K CR/DR CONTENT AREA RADT 2913 COMPREHENSIVE REVIEW 1 The CR cassette is backed by aluminum that: A. reflects x-rays B. absorbs x-rays C. captures the image D. transmits x-rays
More informationFluoroscopy - Chapter 9
Fluoroscopy - Chapter 9 Kalpana Kanal, Ph.D., DABR Lecturer, Diagnostic Physics Dept. of Radiology UW Medicine a copy of this lecture may be found at: http://courses.washington.edu/radxphys/physicscourse04-05.html
More informationABOVETABLE X-RAY SOURCE FLUOROSCOPIC AND SPOT-FILM SYSTEMS
PART VI ABOVETABLE X-RAY SOURCE FLUOROSCOPIC AND SPOT-FILM SYSTEMS FORM FDA 3069 REPRINTED April 2000 ROUTINE COMPLIANCE TESTING ABOVETABLE X-RAY SOURCE FLUOROSCOPIC AND SPOT-FILM SYSTEMS (Test Procedure
More informationCylindrical Ion Chambers
Cylindrical Ion Chambers Radiation Oncology ON Victoreen Model 550T Series Cylindrical Ion Chambers for use with Model 35040 and Model 560 electrometers Wide range of applications in Diagnostic X-Ray and
More informationEnhanced Functionality of High-Speed Image Processing Engine SUREengine PRO. Sharpness (spatial resolution) Graininess (noise intensity)
Vascular Enhanced Functionality of High-Speed Image Processing Engine SUREengine PRO Medical Systems Division, Shimadzu Corporation Yoshiaki Miura 1. Introduction In recent years, digital cardiovascular
More informationX-ray Imaging. PHYS Lecture. Carlos Vinhais. Departamento de Física Instituto Superior de Engenharia do Porto
X-ray Imaging PHYS Lecture Carlos Vinhais Departamento de Física Instituto Superior de Engenharia do Porto cav@isep.ipp.pt Overview Projection Radiography Anode Angle Focal Spot Magnification Blurring
More informationX-RAY FLUOROSCOPY IMAGING SYSTEMS. Dr Slavik Tabakov. Luminescence: Dept. Medical Eng. & Physics King s College London
X-RAY FLUOROSCOPY IMAGING SYSTEMS Dr Slavik Tabakov OBJECTIVES - Image Intensifier construction - Input window - Accelerating and focusing electrodes - Output window - Conversion factor - II characteristics
More informationInvestigation of the line-pair pattern method for evaluating mammographic focal spot performance
Investigation of the line-pair pattern method for evaluating mammographic focal spot performance Mitchell M. Goodsitt, a) Heang-Ping Chan, and Bob Liu Department of Radiology, University of Michigan, Ann
More informationCOMPUTED TOMOGRAPHY 1
COMPUTED TOMOGRAPHY 1 Why CT? Conventional X ray picture of a chest 2 Introduction Why CT? In a normal X-ray picture, most soft tissue doesn't show up clearly. To focus in on organs, or to examine the
More informationPhotomultiplier Tube
Nuclear Medicine Uses a device known as a Gamma Camera. Also known as a Scintillation or Anger Camera. Detects the release of gamma rays from Radionuclide. The radionuclide can be injected, inhaled or
More informationLuminos RF Classic. Where value meets performance.
Luminos RF Classic Where value meets performance www.siemens.com/healthcare What s good value in fluoroscopy? That s easy. Luminos RF Classic. 2 Whether for its handling convenience, outstanding image
More information10/3/2012. Study Harder
This presentation is a professional collaboration of development time prepared by: Rex Christensen Terri Jurkiewicz and Diane Kawamura Study Harder CR detection is inefficient, inferior to film screen
More informationDigital Radiography. Selected Topics
Digital Radiography Selected Topics DIGITAL RADIOGRAPHY Selected Topics Editorial Advisory Board: PETER R. ALMOND, Ph.D. University of Louisville School of Medicine Louisville, Kentucky JOHN S. CLIFTON,
More informationComparison of computed radiography and filmõscreen combination using a contrast-detail phantom
JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, VOLUME 4, NUMBER 1, WINTER 2003 Comparison of computed radiography and filmõscreen combination using a contrast-detail phantom Z. F. Lu,* E. L. Nickoloff, J.
More informationAcceptance Testing of a Digital Breast Tomosynthesis Unit
Acceptance Testing of a Digital Breast Tomosynthesis Unit 2012 AAPM Spring Clinical Meeting Jessica Clements, M.S., DABR Objectives Review of technology and clinical advantages Acceptance Testing Procedures
More information10/26/2015. Study Harder
This presentation is a professional collaboration of development time prepared by: Rex Christensen Terri Jurkiewicz and Diane Kawamura Study Harder CR detection is inefficient, inferior to film screen
More informationFig.2: Scanner VistaScan for image plates
RADIOGRAPHIC INSPECTION OF WELDINGS BY DIGITAL SENSORS H. Thiele, H.-J. Friemel RADIS GmbH, Johanniskirchen, Germany Abstract: The newly available digital sensors for radiographic inspection are suitable
More information35080M kvp Divider. Operators Manual
35080M kvp Divider Operators Manual July 2006 Manual No. 1210001000 Rev. 3 2004, 2005, 2006 Fluke Corporation, All rights reserved. Printed in U.S.A. All product names are trademarks of their respective
More informationEssentials of Digital Imaging
Essentials of Digital Imaging Module 7 Transcript 2016 ASRT. All rights reserved. Essentials of Digital Imaging Module 7 Quality 1. ASRT Animation 2. Welcome Welcome to the Essentials of Digital Imaging:
More informationMulti-Access Biplane Lab
Multi-Access Biplane Lab Advanced technolo gies deliver optimized biplane imaging Designed in concert with leading physicians, the Infinix VF-i/BP provides advanced, versatile patient access to meet the
More informationRaySafe X2. Effortless measurements of X-ray
RaySafe X2 Effortless measurements of X-ray At your fingertips We ve grown accustomed to intuitive interactions with our devices. After all, it s not the device that s most important, but what you can
More informationDo you have any other questions? Please call us at (Toll Free) or , or
INSTRUCTIONS Read the appropriate course/ textbook. This is an open book test. A score of 75% or higher is needed to receive CE credit. You will have a maximum of three attempts to pass this course. Please
More informationStudies on reduction of exposure dose using digital scattered X-ray removal processing
Studies on reduction of exposure dose using digital scattered X-ray removal processing Poster No.: C-1834 Congress: ECR 2015 Type: Scientific Exhibit Authors: K. Kashiyama, M. Funahashi, T. Nakaoka, T.
More informationStudies on reduction of exposure dose using digital scattered X-ray removal processing
Studies on reduction of exposure dose using digital scattered X-ray removal processing Poster No.: C-1834 Congress: ECR 2015 Type: Scientific Exhibit Authors: K. Kashiyama, M. Funahashi, T. Nakaoka, T.
More informationComputed Radiography
BAM Berlin Computed Radiography --INDE 2007, Kalpakkam, India -- Uwe Zscherpel, Uwe Ewert BAM Berlin, Division VIII.3 Requests Requests and and information information to: to: Dr. Dr. U. U. Zscherpel Zscherpel
More information56-6XX CLEAR-Pb Mobile Barriers
56-6XX CLEAR-Pb Mobile Barriers User Manual January 2008 Manual No. 56-6XX-1 Rev. 2 2004, 2005, 2008 Fluke Corporation, All rights reserved. All product names are trademarks of their respective companies
More informationsiemens.com/luminos-fusion Luminos Fusion The 2-in-1 system that fits your needs and fits your budget
siemens.com/luminos-fusion Luminos Fusion The 2-in-1 system that fits your needs and fits your budget Luminos Fusion The 2-in-1 system that fits your needs and fits your budget Luminos Fusion provides
More informationCylindrical Ion Chambers Victoreen Model 550 Series
Cylindrical Ion Chambers Victoreen Model 550 Series! Cylindrical Ion Chambers for use with Model 35040 and Model 530 electrometers! Wide range of applications in Diagnostic X-Ray and Radiation Oncology
More informationR&F X-ray systsem. Savings With Every Exposure
R&F X-ray systsem Savings With Every Exposure Savings with every exposure Versatility meets value Dependable performance User-friendly interface Expandable imaging capabilities Excellent image quality
More informationVisualization of sources of scattered radiation from x-ray equipment used for interventional radiology
Visualization of sources of scattered radiation from x-ray equipment used for interventional radiology Poster No.: C-1190 Congress: ECR 2011 Type: Scientific Exhibit Authors: K. Chida, T. Takahashi, D.
More informationSECTION I - CHAPTER 1 DIGITAL RADIOGRAPHY: AN OVERVIEW OF THE TEXT. Exam Content Specifications 8/22/2012 RADT 3463 COMPUTERIZED IMAGING
RADT 3463 - COMPUTERIZED IMAGING Section I: Chapter 1 RADT 3463 Computerized Imaging 1 SECTION I - CHAPTER 1 DIGITAL RADIOGRAPHY: AN OVERVIEW OF THE TEXT RADT 3463 COMPUTERIZED IMAGING Section I: Chapter
More informationLECTURE 1 The Radiographic Image
LECTURE 1 The Radiographic Image Prepared by:- KAMARUL AMIN ABDULLAH @ ABU BAKAR UiTM Faculty of Health Sciences Medical Imaging Department 11/23/2011 KAMARUL AMIN (C) 1 Lesson Objectives At the end of
More informationPD233: Design of Biomedical Devices and Systems
PD233: Design of Biomedical Devices and Systems (Lecture-8 Medical Imaging Systems) (Imaging Systems Basics, X-ray and CT) Dr. Manish Arora CPDM, IISc Course Website: http://cpdm.iisc.ac.in/utsaah/courses/
More informationAmorphous Selenium Direct Radiography for Industrial Imaging
DGZfP Proceedings BB 67-CD Paper 22 Computerized Tomography for Industrial Applications and Image Processing in Radiology March 15-17, 1999, Berlin, Germany Amorphous Selenium Direct Radiography for Industrial
More informationArcCHECK. The Ultimate 4D QA Solution. Your Most Valuable QA and Dosimetry Tools
ArcCHECK The Ultimate 4D QA Solution A 4D isotropical cylindrical detector array for arc delivery QA and Dosimetry U.S.Patent No. 8,044,359; 6,125,335 Compatible with: FFF Beams VMAT RapidArc TomoTherapy
More informationP UBLIC Law passed by the Congress in 1969 and often referred to
\OL. 117, No. A PHANTOM FOR THE EVALUATION OF TECHNIQUES AND EQUIPMENT USED FOR ROENTGENOGRAPHY OF THE CHEST* By E. DALE TROUT, D.SC., and JOHN P. KELLEY, B.S. CORVALLIS, P UBLIC Law 91-173 passed by the
More informationNPTEL NPTEL ONLINE COURSE. NPTEL Online Certification Course (NOC) NPTEL. Theory and Practice of Non Destructive Testing
NPTEL NPTEL ONLINE COURSE NPTEL Online Certification Course (NOC) NPTEL Theory and Practice of Non Destructive Testing Dr. Ranjit Bauri Dept. of Metallurgical & Materials Engineering IIT Madras, Chennai
More informationArcCHECKTM. The Ultimate 4D QA Solution. Your Most Valuable QA and Dosimetry Tools. VMAT RapidArc TomoTherapy Pinnacle 3 SmartArc Conventional IMRT
TM The Ultimate 4D QA Solution A 4D isotropical cylindrical detector array for arc delivery QA and Dosimetry U.S.Patent No. 8,044,359 What is? The world s first true 4D detector array The world s first
More informationINTRODUCTION TO FLEXIBLE BRONCHOSCOPY. Fluoroscopy Synopsis HENRI G COLT MD SECOND EDITION THE BRONCHOSCOPY EDUCATION PROJECT SERIES
SECOND EDITION INTRODUCTION TO FLEXIBLE BRONCHOSCOPY Fluoroscopy Synopsis HENRI G COLT MD With contributions from Dr. S. Murgu THE BRONCHOSCOPY EDUCATION PROJECT SERIES FLUOROSCOPY SYNOPSIS The purpose
More informationTeaching Digital Radiography and Fluoroscopic Radiation Protection
Teaching Digital Radiography and Fluoroscopic Radiation Protection WCEC 20 th Student Educator Radiographer Conference Dennis Bowman, RT(R), CRT (R)(F) Community Hospital of the Monterey Peninsula (CHOMP)
More informationSTEREOTACTIC BREAST BIOPSY EQUIPMENT SURVEYS
STEREOTACTIC BREAST BIOPSY EQUIPMENT SURVEYS JAMES A. TOMLINSON, M.S. Diagnostic Radiological Physicist American Board of Radiology Certified Medical Physics Consultants, Inc. Bio 28 yrs experience 100%
More informationHIGH RESOLUTION COMPUTERIZED TOMOGRAPHY SYSTEM USING AN IMAGING PLATE
HIGH RESOLUTION COMPUTERIZED TOMOGRAPHY SYSTEM USING AN IMAGING PLATE Takeyuki Hashimoto 1), Morio Onoe 2), Hiroshi Nakamura 3), Tamon Inouye 4), Hiromichi Jumonji 5), Iwao Takahashi 6); 1)Yokohama Soei
More informationMammographic Accreditation Phantom Nuclear Associates Model
Mammographic Accreditation Phantom Nuclear Associates Model 18-220 Diagnostic Imaging! Helps ensure optimum image quality and peak performance of the mammographic system! Essential for MQSA compliance
More informationSuppression of metal artifacts using image-based monoenergetic DECT imaging
Suppression of metal artifacts using image-based monoenergetic DECT imaging Poster No.: C-0519 Congress: ECR 2011 Type: Scientific Paper Authors: B. Krauss, B. Schmidt, M. Sedlmair, T. Flohr; Forchheim/DE
More informationGAFCHROMIC. Therapy Dosimetry Media Models to
GAFCHROMIC Therapy Dosimetry Media Models 37-040 to 37-045! Superior uniformity and sensitivity! Dose rate and fractionation independent! Maps dose distribution! Provides quantitative measurements (via
More informationCh. 223 VETERINARY MEDICINE CHAPTER 223. VETERINARY MEDICINE GENERAL PROVISIONS X-RAYS RADIOACTIVE MATERIAL. Authority
Ch. 223 VETERINARY MEDICINE 25 223.1 CHAPTER 223. VETERINARY MEDICINE Sec. 223.1. Purpose and scope. 223.2. [Reserved]. 223.2a. Definitions. 223.3 223.6. [Reserved]. 223.7. Structural shielding. 223.8.
More informationX-ray Tube and Generator Basic principles and construction
X-ray Tube and Generator Basic principles and construction Dr Slavik Tabakov - Production of X-rays and Patient Dose OBJECTIVES - X-ray tube construction - Anode - types, efficiency - Classical X-ray generator
More informationHISTORY. CT Physics with an Emphasis on Application in Thoracic and Cardiac Imaging SUNDAY. Shawn D. Teague, MD
CT Physics with an Emphasis on Application in Thoracic and Cardiac Imaging Shawn D. Teague, MD DISCLOSURES 3DR- advisory committee CT PHYSICS WITH AN EMPHASIS ON APPLICATION IN THORACIC AND CARDIAC IMAGING
More information1. Carlton, Richard R., and Arlene M. Adler. Principles of Radiographic Imaging: An Art and a Science, 5th edition (2013).
CODE: RADT 151 INSTITUTE: Health Science TITLE: Radiographic Exposure DEPARTMENT: Radiologic Technology COURSE DESCRIPTION: This course covers the principles of radiographic exposure selection and manipulation
More informationAnsur TNT Users Manual. Plug-In
Ansur TNT 12000 Plug-In Users Manual August 2009, Rev. 2, 12/09 2009 Fluke Corporation. All rights reserved. Specifications are subject to change without notice. All product names are trademarks of their
More informationContrast. Contrast: the difference in density on adjacent areas of a radiograph or other image receptor. Subjective. Long Scale (Low Contrast)
Contrast Contrast: the difference in density on adjacent areas of a radiograph or other image receptor. Subject Subjective Radiographic Long Scale (Low Contrast) Short Scale (High Contrast) Factors affecting
More information