Update on Fluoroscopy Physics AAPM MO-A-210A-1 Stephen Balter, Ph.D.
|
|
- Susan Davis
- 6 years ago
- Views:
Transcription
1 Update on Fluoroscopy Physics Stephen Balter, PhD Columbia University Draft for MO-A-210A AAPM Educational objectives Understand dosimetric concepts relating to interventional fluoroscopy Characterize the dosimetric features and performance of a modern fluoroscope Be able to set up a clinical dose recording and reporting policy that will meet clinical and JC requirements. SB0907 UFP AAPM - 1 SB0907 UFP AAPM - 2 Outline Notation This presentation uses the notation given in ICRU-74 (2006) The notation has been extended to encompass additional quantities specifically related to fluoroscopy Organizations that have accepted or are in the process of accepting this notation include the IAEA, IEC, NCRP and NEMA. SB0907 UFP AAPM - 3 SB0907 UFP AAPM - 4 Air KERMA (K a ) Measurements are made under lowscatter conditions Metric usually used by calibration laboratories to report instrument factors. Unless the detector is air, there may be additional corrections needed to account for instrument construction and X-ray spectrum. From the lab to the clinic K a,i is the air kerma at the location of the patient s skin in the absence of the patient. K a,e is the air kerma at the location of the patient s skin with the patient present. SB0907 UFP AAPM - 5 SB0907 UFP AAPM - 6
2 Backscatter factors K a,i K a,e Reference Point Air Kerma (K a,r ) Reference point originally defined in IEC (2000) Included in 2005 FDA fluoro regulations. Intent is to approximate the location of the patient s skin Reported as K a without scatter IEC, FDA ± 35 % SB0907 UFP AAPM - 7 SB0907 UFP AAPM - 8 Reference Point Air Kerma (K a,r ) Reference Point FDA Point Total air kerma accumulated at the reference point from the beginning of the procedure. (called cumulative dose in the regs) Displayed to operators at the working position and in the control room. Defined under low scatter conditions. Table top and mattress attenuation? Not standardized. Propose measurement without attenuation unless the attenuators are always in the beam. SID = Any FDA Dose Compliance Point 15 cm Image Receptor Isocenter IEC & FDA K a,r Reference Point Focal Spot SB0907 UFP AAPM - 9 SB0907 UFP AAPM - 10 Reference point skin Conversion of K a,r to D skin D max,skin is seldom numerically equal to K a,r For teaching purposes, assume that the beam does not move during the procedure. K a,i = K a,r * (SRD/SSD) 2 where SRD is the source to reference point distance K a,e = K a,i * Backscatter f(kvp, filter, field size) D skin = K a,e * (µ skin / µ air ) SB0907 UFP AAPM - 11 SB0907 UFP AAPM - 12
3 Kerma Area Product (P KA ) P KA is independent of distance! Total KAP accumulated from the beginning of the procedure. Displayed to operators at the working position and in the control room. Can be used to estimate K a,r (need to know field size at reference point) Defined under low scatter conditions. Table top and mattress attenuation? Where should DAP be measured? Area = a * d 2 Output = k *d -2 P KA = a * d 2 * k*d -2 = a * k SB0907 UFP AAPM - 13 SB0907 UFP AAPM - 14 Outline Geometry (generic system) Min SSD = 38 cm (<) SID 80 cm 125 cm SAD = cm (focus to isocenter) Table top height - 30 to + 20 cm (relative to isocenter) 110º LAO 110º RAO 45º CRA 45º CAU FS cm 2 (at image receptor) SB0907 UFP AAPM - 15 SB0907 UFP AAPM - 16 Radiation technique range Table-Top Outputs FP K a,i range <1 4,000 mgy/min kvp Fixed filtration 2 6 mm Al Added Filter mm Cu Brand A : Added Cu fluoro filter independent of kvp Brands B,C: Added Cu fluoro filter tends to decrease as kvp increases All Brands: Less or no Cu for acquisition modes New systems do not provide physicist level control of basic X-ray factors mgy/min Chamber from FP Minimum SID 25NF 20NF 15NF 25CI 20CI 15CI 38 mm Al +0.5 mm +1.0 mm +1.5 mm +2.0 mm +3 mm Cu +4 mm Cu 6 mm Cu +10 mm Cu Cu Cu Cu Cu Attenuator 0902D SB0907 UFP AAPM - 17 SB0907 UFP AAPM - 18
4 Observed values Dose rates are down; BUT!!! SB0907 UFP AAPM - 19 SB0907 UFP AAPM - 20 Conventional Wisdom FDA limits normal-mode table-top outputs to 87 mgy/min (10 R/min) Using more magnification increases patient skin-dose-rate Using more magnification increases limiting spatial resolution Source to Image Receptor Tracking FDA regulations limit normal mode fluoroscopy to less than 87 mgy/min at from the image receptor. Source to image receptor distance (SID) is not mentioned in the regulations. Most interventional limit output to less than 87 mgy/min at any SID What happens at the table top when SID is varied? SB0907 UFP AAPM - 21 SB0907 UFP AAPM - 22 Minimum SID Table-Top K a,i Maximum SID Table-Top K a,i Minimum SID Maximum SID Flu 75 mgy/m Cin 1700 mgy/m Flu 75 mgy/m Flu 230 mgy/m Cin 1700 mgy/m SB0907 UFP AAPM - 23 SB0907 UFP AAPM - 24
5 Low Table/Max SID Table-Top K a,i Table Top Maximum Air Kerma Rate Maximum SID 2.75 Table Top (chamber) fixed 55 cm from FS Flu 75 mgy/m Flu 270 mgy/m Cin 2000 mgy/m Relative Table Top Output Fluoroscopy Table Top Rate Inverse Square Cine Table Top Rate SB0907 UFP AAPM - 25 Minimum SSD Source to Image Receptor Distance (cm) 0907 Alpha SB0907 UFP AAPM - 26 Effect of zoom on EERII Field of view considerations An image intensifier zooms a variable size input field onto a constant output field. Unmodified: Dose per frame 1/ FOV 2 Programmed: Dose per frame 1/ FOV A first generation flat-panel electronically zooms a variable number of pixels onto a full size image display. Dose per frame constant Dose per frame 1/ FOV Dose per frame 1/ FOV 2 SB0907 UFP AAPM - 27 SB0907 UFP AAPM - 28 Measured Phantom EER vs. f(fov) Image intensifier zoom Program kvp Beam filter Dose Rate Constraint Normalized Phantom Input (fluoroscopy) 6 Base 38 Al 0.5 Cu Cu 2.0 Cu 4.0 Cu 8.0 Cu 1/r 4 1/r cm 42 cm 32 cm 22 cm 16 cm 11cm Field of View - Flat Panel 0805 BETA SB0907 UFP AAPM - 29 SB0907 UFP AAPM - 30
6 Pixel size image intensifier zoom DR Indirect Detector X-ray Light Structured X-ray phosphor (CsI) S G D + Photodiode Storage capacitor Charge X-rays to light to electrons to electronic signal: a -Si TFT/ CsI phosphor SB0907 UFP AAPM - 31 SB0907 UFP AAPM - 32 Pixel size: flat panel zoom (1 st gen) Smaller FOV = Fewer Pixels SB0907 UFP AAPM - 33 SB0907 UFP AAPM - 34 Pixel size: zoom comparison Rebinning SB0907 UFP AAPM - 35 SB0907 UFP AAPM - 36
7 Display Display Processing Zoomed FP images are digitally transformed from the detected pixel matrix to the full display matrix. All images are highly processed to enhance perception of detail. Dose per frame is programmed by the system to standardize the perception of noise. These measures are designed to maximize perception High Contrast Objects (small interventional devices < 1 mm) Low Contrast Objects (relatively large tissue blush > 5 mm) Different acquisition and display optimization is needed for different clinical procedures. Outline SB0907 UFP AAPM - 37 SB0907 UFP AAPM - 38 What instrument? Reference Point Locations If you are super-fussy (and want to try to stay below the ICRU-74 tolerance of 7%) the functionally closer to a free air chamber the better. For most work, energy compensation of solid-state detectors is adequate Absolute accuracy should improve when appropriate calibration beams are defined and implemented. Consider the influence of any instrument on the fluoroscope s ADRC. Caution: You, your service engineers, and your regulators may be using different types of instrumentation. SID = Any FDA Dose Compliance Point 15 cm Image Receptor Isocenter IEC & FDA K a,r Reference Point Focal Spot SB0907 UFP AAPM - 39 SB0907 UFP AAPM - 40 Compliance Setup Compliance minimum SID 50% Equipment placed on table top Attenuators 15 cm above ion chamber Minimum SID Adjust table height for between center of chamber to entrance surface of image receptor Collimate to just inside smallest FOV Test common modes at all FOVs. SB0907 UFP AAPM - 41 SB0907 UFP AAPM - 42
8 FL FN CL CN Update on Fluoroscopy Physics AAPM MO-A-210A-1 Attenuator protocol Acquisition modes 19 mm Al (children) 38 mm Al (small adult) 38 mm Al mm Cu (medium adult) 38 mm Al mm Cu (large adult) 38 Al + 2 Cu + 3 mm Pb (maximum) or 38 mm Al mm Cu (maximum) There is reluctance to test acquisition modes at full output due to concerns about damaging the fluoroscope. This is reflected in regulatory requirements Modern systems are actually clinically operated at or near full output a significant fraction of the time. Maximum outputs are often more than double the 38 Al + 2 Cu outputs SB0907 UFP AAPM - 43 SB0907 UFP AAPM - 44 Clinical kvp in one laboratory Typical compliance data Compliance Image Intensifier - 17 cm FOV FDA Geometry kvp Cine n = 1856 Fluoro n = 6594 fkvp ckvp mgy/min Al Cu + 1 Cu + 2 Cu + 4 Cu + 8 Cu + 8Cu+3Pb Attenuator Display = Cine actual/2, fc actual mgy/min cine S fluoro mm Al + X mm Cu 80 Flat Panel Detector Image Intensifier % 5% 10% 20% 25% 30% 40% 50% 60% 70% 75% 80% 90% 95% 99% Both have variable Cu beam filters Percentile SB0907 UFP AAPM - 45 SB0907 UFP AAPM - 46 Dose instrumentation Dose displays in the lab Locations Integrated Interfaced External Measurement technologies None Calculated from settings P KA chamber + calculation K a,r chamber + calculation P KA & K a,r dual channel chamber SB0907 UFP AAPM - 47 The same RPDose will display as 5678 on all three systems The same KAP will display as on GE and SIEMENS and on PHILIPS SB0907 UFP AAPM - 48
9 Verification of displayed dose Accuracy of display IEC ± 50% (RPDose &KAP) FDA ± 35% (RPDose) Stability usually better Usually validated at factory Seldom validated by installers When verified as part of QA constancy should be ± 5% Stability of K a,r instruments AVE STDEV Jul-Dec 05Jan-Jun 06Jul-Dec 06Jan-Jun 07Jul-Dec 07 A B C DA DB E F GB GA SB0907 UFP AAPM - 49 SB0907 UFP AAPM - 50 K a,i and P KA at isocenter Field size measurement SB0907 UFP AAPM - 51 SB0907 UFP AAPM - 52 Outline Skin dose map and beam position Illustration courtesy of Siemens SB0907 UFP AAPM - 53 SB0907 UFP AAPM - 54
10 What can be measured in the lab? Dose displays in the lab Fluoroscopy time Number of acquisition Runs Frames Kerma Area Product P KA Reference Point Air Kerma K a,r Peak Skin Dose Dose maps SB0907 UFP AAPM - 55 The same RPDose will display as 5678 on all three systems The same KAP will display as on GE and SIEMENS and on PHILIPS SB0907 UFP AAPM - 56 In the control room SIR Standard of Practice Record all available data for every interventional procedure Order of Precedence Skin dose map (not currently available) Reference Point Air Kerma K a,r Kerma Area Product P KA Fluoro time and number of images Uses for the data Patient pre discharge instruction Dose reconstruction when necessary Clinical QA Physics QA against guidance levels SB0907 UFP AAPM - 57 SB0907 UFP AAPM - 58 Manual log A New IHE Profile IHE Radiation Exposure Monitoring Profile Integration of systems reporting dose and systems which receive, store, or process those reports Modalities, PACS, RIS, Workstations, Registries Technician hand writes data to log Data retyped into clinical records or dose data base Time lags and overflow can be problems Process is error prone Facilitate compliance with Euratom 97/43, ACR Guidelines, etc. Directly based on DICOM Dose Reports Creation, Collection, Distribution, Processing SB0907 UFP AAPM - 59 IHE 2008 Webinar Series
11 Outline JC Radiation Sentinel Event An unexpected occurrence involving death or serious physical or psychological injury, or the risk thereof. Serious injury specifically includes loss of limb or function. The phrase or the risk thereof includes any process variation for which a recurrence would carry a significant chance of a serious adverse outcome. Such events are called sentinel because they signal the need for immediate investigation and response. Prolonged fluoroscopy with cumulative dose >1500 rads to a single field or any delivery of radiotherapy to the wrong body region or >25% above the planned radiotherapy dose Can you demonstrate to the surveyors that Sentinel Events did not occur? How to be accurate enough? (not easy) SB0907 UFP AAPM - 61 SB0907 UFP AAPM - 62 MedPhys list May 2007 A facility without a full-time physicist needs a quick, easy number that they can refer to, so we use fluoro time, as archaic and inaccurate as it might be. If you want to know if your patient received more than the threshold for skin injuries, look at the patient's skin, not the KAP meter or fluoro time! Etiology not identified for more than one year SB0907 UFP AAPM - 63 Fluoro 35 minutes Cine 12 minutes What was covered SB0907 UFP AAPM - 64
I. PERFORMANCE OF X-RAY PRODUCTION COMPONENTS FLUOROSCOPIC ACCEPTANCE TESTING: TEST PROCEDURES & PERFORMANCE CRITERIA
FLUOROSCOPIC ACCEPTANCE TESTING: TEST PROCEDURES & PERFORMANCE CRITERIA EDWARD L. NICKOLOFF DEPARTMENT OF RADIOLOGY COLUMBIA UNIVERSITY NEW YORK, NY ACCEPTANCE TESTING GOALS PRIOR TO 1st CLINICAL USAGE
More informationClinical Experiences with a Patient Skin Dose Monitoring and Tracking Program
Clinical Experiences with a Patient Skin Dose Monitoring and Tracking Program Allen R. Goode, MS, DABR Chief Diagnostic Medical Physicist Department of Radiology & Medical Imaging University of Virginia
More informationKey words: fluoroscopy, dose-area-product, kerma-area-product, calibration of KAP meters, patient exposure
Accuracy and calibration of integrated radiation output indicators in diagnostic radiology: A report of the AAPM Imaging Physics Committee Task Group 190 Pei-Jan P. Lin a) Virginia Commonwealth University
More informationJoint ICTP/IAEA Advanced School on Dosimetry in Diagnostic Radiology and its Clinical Implementation May 2009
2033-6 Joint ICTP/IAEA Advanced School on Dosimetry in Diagnostic Radiology and its Clinical Implementation 11-15 May 2009 Dosimetry for Fluoroscopy Basics Renato Padovani EFOMP Joint ICTP-IAEA Advanced
More informationInterventional Radiological Equipment selection and installation
Interventional Radiological Equipment selection and installation Renato Padovani ICTP Learning objectives To understand the main components of an interventional radiology equipment To understand the relevance
More informationDisclosures. Outline 7/31/2017. Current Implementation Status of IEC Standard : Exposure Index (EI) for Digital Radiography
Current Implementation Status of IEC Standard 62494-1: Exposure Index (EI) for Digital Radiography July 31, 2017 Ryan Fisher, PhD, DABR Katie Hulme, MS, DABR None Disclosures Outline Review of IEC Standard
More informationPractical Medical Physics Session: TG-151 Dose Monitoring. August 5, 2013 Katie Hulme, M.S.
Practical Medical Physics Session: TG-151 Dose Monitoring August 5, 2013 Katie Hulme, M.S. Digital Imaging and Dose Creep Images courtesy of Agfa Healthcare Under-Exposed Over-Exposed Freedman et al.,
More informationExposure Indices and Target Values in Radiography: What Are They and How Can You Use Them?
Exposure Indices and Target Values in Radiography: What Are They and How Can You Use Them? Definition and Validation of Exposure Indices Ingrid Reiser, PhD DABR Department of Radiology University of Chicago
More informationCalculating the peak skin dose resulting from fluoroscopically guided interventions. Part I: Methods
JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, VOLUME 12, NUMBER 4, fall 2011 Calculating the peak skin dose resulting from fluoroscopically guided interventions. Part I: Methods A. Kyle Jones, 1a and Alexander
More informationMinnesota Rules, Chapter 4732 X-ray Revision
Minnesota Rules, Chapter 4732 X-ray Revision DRAFT FLUOROSCOPIC X-RAY SYSTEMS, 1.0 Subpart 1. Applicability. Subpart 2. Limitation of the useful beam. Subpart 3. Measuring compliance; primary protective
More informationAcceptance Testing of a Digital Breast Tomosynthesis Unit
Acceptance Testing of a Digital Breast Tomosynthesis Unit 2012 AAPM Spring Clinical Meeting Jessica Clements, M.S., DABR Objectives Review of technology and clinical advantages Acceptance Testing Procedures
More informationNEMA XR X-ray Equipment for Interventional Procedures User Quality Control Mode
NEMA XR 27-2012 X-ray Equipment for Interventional Procedures User Quality Control Mode Published by: National Electrical Manufacturers Association 1300 North 17th Street, Suite 1752 Rosslyn, Virginia
More informationDigital radiography (DR) post processing techniques for pediatric radiology
Digital radiography (DR) post processing techniques for pediatric radiology St Jude Children s Research Hospital Samuel Brady, MS PhD DABR samuel.brady@stjude.org Purpose Review common issues and solutions
More informationTest Equipment for Radiology and CT Quality Control Contents
Test Equipment for Radiology and CT Quality Control Contents Quality Control Testing...2 Photometers for Digital Clinical Display QC...3 Primary Workstations...3 Secondary Workstations...3 Testing of workstations...3
More informationNuclear Associates
Nuclear Associates 07-706 Patient Phantom/Penetrometer System Users Manual March 2005 Manual No. 07-706-1 Rev. 2 2004, 2005 Fluke Corporation, All rights reserved. Printed in U.S.A. All product names are
More informationBreast Tomosynthesis. Bob Liu, Ph.D. Department of Radiology Massachusetts General Hospital And Harvard Medical School
Breast Tomosynthesis Bob Liu, Ph.D. Department of Radiology Massachusetts General Hospital And Harvard Medical School Outline Physics aspects of breast tomosynthesis Quality control of breast tomosynthesis
More information3/31/2011. Objectives. Emory University. Historical Development. Historical Development. Historical Development
Teaching Radiographic Technique in a Digital Imaging Paradigm Objectives 1. Discuss the historical development of digital imaging. Dawn Couch Moore, M.M.Sc., RT(R) Assistant Professor and Director Emory
More information2012 :15th SESSION of ESMP
2012 :15th SESSION of ESMP Lecture presented in Archamps (Salève Building) by : Elly CASTELLANO (London) Patient dosimetry in x-ray imaging and CT Elly Castellano Objectives measurable dose quantities
More informationC-ARM FLUOROSCOPIC AND SPOT-FILM SYSTEMS
PART X C-ARM FLUOROSCOPIC AND SPOT-FILM SYSTEMS FORM FD 3260 REPRINTED APRIL 2000 ROUTINE COMPLIANCE TESTING C-ARM FLUOROSCOPES (Test Procedure CFA - Use Form FDA 3260) 1.0 GENERAL GUIDANCE 1.1 This procedure
More informationTESTING FLAT-PANEL IMAGING SYSTEMS: What the Medical Physicist Needs to Know. JAMES A. TOMLINSON, M.S., D.A.B.R. Diagnostic Radiological Physicist
TESTING FLAT-PANEL IMAGING SYSTEMS: What the Medical Physicist Needs to Know JAMES A. TOMLINSON, M.S., D.A.B.R. Diagnostic Radiological Physicist Topics Image Uniformity and Artifacts Image Quality - Detail
More informationTruly flexible to meet your clinical needs
Truly flexible to meet your clinical needs 2 Adapting to meet your needs Flexible Fast and responsive Excellent image quality Designed with ergonomic efficiency Equipped with dose management tools 3 Three
More informationThe Current State of EPID-Based Linear Accelerator Quality Assurance. Disclosures. Purpose of this First Talk 8/3/2017
The Current State of EPID-Based Linear Accelerator Quality Assurance Timothy Ritter, PhD, DABR, FAAPM 1 Disclosures Employed by the Veterans Health Administration Faculty appointment with the University
More informationFeatures and Weaknesses of Phantoms for CR/DR System Testing
Physics testing of image detectors Parameters to test Features and Weaknesses of Phantoms for CR/DR System Testing Spatial resolution Contrast resolution Uniformity/geometric distortion Dose response/signal
More informationSPECIFICATION. Kilovoltage X-ray calibration system for protection and diagnostic level dosimetry. Prepared by
SPECIFICATION Kilovoltage X-ray Prepared by Igor Gomola, Technical Officer, Project ECU6023, Date 2015-Oct-06 Revision Date Status Comments 0.1 2015-Oct-06 Draft Igor Gomola Page 1 of 12 1. Scope This
More informationAN ABSTRACT OF THE THESIS OF. W. Scott Helms for the degree of Master of Science in Radiation Health Physics
AN ABSTRACT OF THE THESIS OF W. Scott Helms for the degree of Master of Science in Radiation Health Physics presented on November 24, 2014 Title: A Quantitative Comparison of Cardiovascular Imaging Systems
More informationDigital Imaging started in the 1972 with Digital subtraction angiography Clinical digital imaging was employed from the 1980 ~ 37 years ago Amount of
Digital Imaging started in the 1972 with Digital subtraction angiography Clinical digital imaging was employed from the 1980 ~ 37 years ago Amount of radiation to the population due to Medical Imaging
More informationCalibration of KAP meters
Calibration of KAP meters Alexandr Malusek! Division of Radiological Sciences Department of Medical and Health Sciences Linköping University! 2014-04-15 1 Outline 1. KAP meter construction 2. Air kerma-area
More informationLearning Objectives: What s my motivation? (unknown screen actor) Workshop Overview
Practical Medical Physics Adapting Traditional Clinical Medical Physics to Digital Radiography Charles E. Willis, Ph.D., DABR Associate Professor Department of Imaging Physics The University of Texas M.D.
More informationDISC QC/QA Program for Digital Imaging Systems using the DR Radchex Plus Meter
DISC QC/QA Program for Digital Imaging Systems using the DR Radchex Plus Meter Revision Date: January 5th, 2017 www.disc-imaging.com Table of Contents Section A: Preliminary Setup Requirements... 4 Tools
More informationPhilips XPER FD10C R7.0.4
Philips XPER FD10C R7.0.4 Reconditioned 2005 System- Upgraded to R7 in Oct 2010 The Allura Xper FD10 (Ceiling) single-plane cardiovascular system is comprised of a ceiling mounted C-arm stand and digital
More informationMammography: Physics of Imaging
Mammography: Physics of Imaging Robert G. Gould, Sc.D. Professor and Vice Chair Department of Radiology and Biomedical Imaging University of California San Francisco, California Mammographic Imaging: Uniqueness
More informationInitial Survey Results on the Accuracy of Dose Data in the DICOM Objects
Initial Survey Results on the Accuracy of Dose Data in the DICOM Objects Renato Padovani, Annalisa Trianni University Hospital S. Maria della Misericordia Udine, Italy INTRODUCTION The EC MED has forced
More informationNew spectral benefi ts, proven low dose
New spectral benefi ts, proven low dose Philips MicroDose mammography SI, technical data sheet Philips MicroDose SI with single-shot spectral imaging is a fullfi eld digital mammography solution that delivers
More informationX-ray Imaging. PHYS Lecture. Carlos Vinhais. Departamento de Física Instituto Superior de Engenharia do Porto
X-ray Imaging PHYS Lecture Carlos Vinhais Departamento de Física Instituto Superior de Engenharia do Porto cav@isep.ipp.pt Overview Projection Radiography Anode Angle Focal Spot Magnification Blurring
More informationY11-DR Digital Radiography (DR) Image Quality
Y11-DR Digital Radiography (DR) Image Quality Image quality is stressed for all systems in Safety Code 35. In the relevant sections Health Canada s advice is the manufacturer s recommended test procedures
More information7/24/2014. Image Quality for the Radiation Oncology Physicist: Review of the Fundamentals and Implementation. Disclosures. Outline
Image Quality for the Radiation Oncology Physicist: Review of the Fundamentals and Implementation Image Quality Review I: Basics and Image Quality TH-A-16A-1 Thursday 7:30AM - 9:30AM Room: 16A J. Anthony
More informationRAD 150 RADIOLOGIC EXPOSURE TECHNIQUE II
RAD 150 RADIOLOGIC EXPOSURE TECHNIQUE II APPROVED 12/O2/2011 EFFECTIVE SPRING 2013-14 Prefix & Number RAD 150 Course Title: Radiologic Exposure Technique II & Lab Purpose of this submission: New Change/Updated
More informationPractical Aspects of Medical Physics Surveys of Mammography Equipment and Facilities
Practical Aspects of Medical Physics Surveys of Mammography Equipment and Facilities Melissa Martin, M.S., FAAPM, FACR, FACMP AAPM Annual Meeting - Philadelphia July 19, 2010 MO-B-204C-1 Educational Objectives
More informationNuclear Associates
Nuclear Associates 07-649 CDRH Fluoroscopic Phantom Users Manual March 2005 Manual No. 07-649-1 Rev. 2 2004, 2005 Fluke Corporation, All rights reserved. Printed in U.S.A. All product names are trademarks
More informationSYLLABUS. TITLE: Equipment Operation I. DEPARTMENT: Radiologic Technology
CODE: RADT 156 INSTITUTE: Health Science TITLE: Equipment Operation I DEPARTMENT: Radiologic Technology COURSE DESCRIPTION: This course covers the principles of equipment operation and maintenance of radiographic
More informationEmerging Technology: Real-Time Monitoring of Treatment Delivery EPID Exit Dose QA
Emerging Technology: Real-Time Monitoring of Treatment Delivery EPID Exit Dose QA Arthur Olch, PhD, FAAPM AAPM Spring Clinical Meeting, March 21, 2017 Or.. What Dose are the Patients Really Getting???
More informationOutline. Digital Radiography. Understanding Digital Modalities: Image Quality and Dose. Image Quality. Dose Control
Understanding Digital Modalities: Image Quality and Dose S. Jeff Shepard, M.S. University of Texas M. D. Anderson Cancer Center Houston, Texas Special Acknowledgement: Stephen K. Thompson, M.S. William
More informationAIR KERMA RATES MEASUREMENT IN AN INTERVENTIONAL CARDIOLOGY SUITE
AIR KRMA RATS MASURMNT IN AN INTRVNTIONAL ARIOLOGY SUIT anevaro, L.V. 1 ; Gamarra Sanchez, M..; Pereira, L.S.; Maurício,.L.P. Instituto de Radioproteção e osimetria. omissão Nacional de nergia Nuclear.
More informationAsk EuroSafe Imaging Tips & Tricks. Paediatric Imaging Working Group. Dose Management in Digital Radiography
Ask EuroSafe Imaging Tips & Tricks Paediatric Imaging Working Group Dose Management in Digital Radiography Raija Seuri (HUS Medical Imaging Center, FI) Cristina Almeida (Centro Hospitalar de Lisboa Central,
More informationDevelopment of the Use of Amorphous Silicon (ASi) Electronic Portal Imaging Devices as a Physics Tool for Routine Linear Accelerator QA
Development of the Use of Amorphous Silicon (ASi) Electronic Portal Imaging Devices as a Physics Tool for Routine Linear Accelerator QA Gena M.A.H 1, Ahmed L.El-Attar 2, Elbadry M. Zahran 3, Hany El-Gamal
More information10/3/2012. Study Harder
This presentation is a professional collaboration of development time prepared by: Rex Christensen Terri Jurkiewicz and Diane Kawamura Study Harder CR detection is inefficient, inferior to film screen
More informationImage Display and Perception
Image Display and Perception J. Anthony Seibert, Ph.D. Department of Radiology UC Davis Medical Center Sacramento, California, USA Image acquisition, display, & interpretation X-rays kvp mas Tube filtration
More informationWide beam CT dosimetry. Elly Castellano
Wide beam CT dosimetry Elly Castellano Outline revision: CT dose indices wide-beam CT: the end of the road for CTDI? the IEC rescue plan for CTDI 100 the american way AAPM report 111 better estimates of
More informationX-RAYS - NO UNAUTHORISED ENTRY
Licencing of premises Premises Refer Guidelines A radiation warning sign and warning notice, X-RAYS - NO UNAUTHORISED ENTRY must be displayed at all entrances leading to the rooms where x-ray units are
More informationCHAPTER 6 QC Test For Fluoroscopic Equipment. Prepared by:- Kamarul Amin bin Abu Bakar School of Medical Imaging KLMUC
CHAPTER 6 QC Test For Fluoroscopic Equipment Prepared by:- Kamarul Amin bin Abdullah @ Abu Bakar School of Medical Imaging KLMUC Lesson Outcomes Describe the objectives of each QC test done. Identify QC
More information10/26/2015. Study Harder
This presentation is a professional collaboration of development time prepared by: Rex Christensen Terri Jurkiewicz and Diane Kawamura Study Harder CR detection is inefficient, inferior to film screen
More informationWhile digital techniques have the potential to reduce patient doses, they also have the potential to significantly increase them.
In press 2004 1 2 Guest Editorial (F. Mettler, H. Ringertz and E. Vano) Guest Editorial (F. Mettler, H. Ringertz and E. Vano) Digital radiology An appropriate analogy that is easy for most people to understand
More informationDose Reduction and Image Preservation After the Introduction of a 0.1 mm Cu Filter into the LODOX Statscan unit above 110 kvp
Dose Reduction and Image Preservation After the Introduction of a into the LODOX Statscan unit above 110 kvp Abstract: CJ Trauernicht 1, C Rall 1, T Perks 2, G Maree 1, E Hering 1, S Steiner 3 1) Division
More informationDigital Image Management: the Basics
Digital Image Management: the Basics Napapong Pongnapang, Ph.D. Department of Radiological Technology Faculty of Medical Technology Mahidol University Outline From screen/film to digital radiography PACS/Tele
More informationInvestigation of the line-pair pattern method for evaluating mammographic focal spot performance
Investigation of the line-pair pattern method for evaluating mammographic focal spot performance Mitchell M. Goodsitt, a) Heang-Ping Chan, and Bob Liu Department of Radiology, University of Michigan, Ann
More informationBasis of Computed Radiography & PACS
Basis of Computed Radiography & PACS Slavik Tabakov Computed Radiography (CR) refers to new types of X-ray detectors (i.e. replaces the X-ray Film) The CR output media is a digital image, which can be
More informationNuclear Associates
Nuclear Associates 76-700 Digital Subtraction Angiography Phantom Users Manual March 2005 Manual No. 76-700-1 Rev. 2 2004, 2005 Fluke Corporation, All rights reserved. Printed in U.S.A. All product names
More informationTeaching Digital Radiography and Fluoroscopic Radiation Protection
Teaching Digital Radiography and Fluoroscopic Radiation Protection WCEC 20 th Student Educator Radiographer Conference Dennis Bowman, RT(R), CRT (R)(F) Community Hospital of the Monterey Peninsula (CHOMP)
More informationCalculati tions - BIR meth thod and NCRP 147
Calculations l - BIR method and NCRP 147 David Sutton PhD 1. Cath Lab Indicative of all C-arm applications Only secondary radiation Primary Beam is stopped by the image intensifier Geometry BIR 9.5 m Preparation
More informationCHAPTER 2 COMMISSIONING OF KILO-VOLTAGE CONE BEAM COMPUTED TOMOGRAPHY FOR IMAGE-GUIDED RADIOTHERAPY
14 CHAPTER 2 COMMISSIONING OF KILO-VOLTAGE CONE BEAM COMPUTED TOMOGRAPHY FOR IMAGE-GUIDED RADIOTHERAPY 2.1 INTRODUCTION kv-cbct integrated with linear accelerators as a tool for IGRT, was developed to
More informationABOVETABLE X-RAY SOURCE FLUOROSCOPIC AND SPOT-FILM SYSTEMS
PART VI ABOVETABLE X-RAY SOURCE FLUOROSCOPIC AND SPOT-FILM SYSTEMS FORM FDA 3069 REPRINTED April 2000 ROUTINE COMPLIANCE TESTING ABOVETABLE X-RAY SOURCE FLUOROSCOPIC AND SPOT-FILM SYSTEMS (Test Procedure
More informationEffect of slit scan imaging techniques on image quality on radiotherapy electronic portal imaging
The University of Toledo The University of Toledo Digital Repository Theses and Dissertations 2008 Effect of slit scan imaging techniques on image quality on radiotherapy electronic portal imaging Dean
More informationMammography is a radiographic procedure specially designed for detecting breast pathology Approximately 1 woman in 8 will develop breast cancer over
Mammography is a radiographic procedure specially designed for detecting breast pathology Approximately 1 woman in 8 will develop breast cancer over a lifetime Breast cancer screening programs rely on
More informationIntroduction. Chapter 16 Diagnostic Radiology. Primary radiological image. Primary radiological image
Introduction Chapter 16 Diagnostic Radiology Radiation Dosimetry I Text: H.E Johns and J.R. Cunningham, The physics of radiology, 4 th ed. http://www.utoledo.edu/med/depts/radther In diagnostic radiology
More informationEC Guideline 109: Guidance on Diagnostic Reference Levels REFERENCE LEVELS AT EUROPEAN LEVEL FOR CARDIAC INTERVENTIONAL PROCEDURES
SENTINEL WORKSHOP Delft, - April 7 EC Guideline : Guidance on Diagnostic Reference Levels REFERENCE LEVELS AT EUROPEAN LEVEL FOR CARDIAC INTERVENTIONAL PROCEDURES R. Padovani, E. Vano, A. Trianni, C. Bokou,
More informationQC Testing for Computed Tomography (CT) Scanner
QC Testing for Computed Tomography (CT) Scanner QA - Quality Assurance All planned and systematic actions needed to provide confidence on a structure, system or component. all-encompassing program, including
More informationBASICS OF FLUOROSCOPY
Medical Physics Residents Training Program BASICS OF FLUOROSCOPY Dr. Khalid Alyousef, PhD Department of Medical Imaging King Abdulaziz Medical City- Riyadh Edison examining the hand of Clarence Dally with
More informationDigital Radiography : Flat Panel
Digital Radiography : Flat Panel Flat panels performances & operation How does it work? - what is a sensor? - ideal sensor Flat panels limits and solutions - offset calibration - gain calibration - non
More informationSUBCHAPTER 14. THERAPEUTIC INSTALLATIONS
SUBCHAPTER 14. THERAPEUTIC INSTALLATIONS 7:28-14.1 Scope (a) This subchapter covers therapeutic installations used in the healing arts. These therapeutic installations include x-ray, accelerator and teletherapy
More informationMulti-Access Biplane Lab
Multi-Access Biplane Lab Advanced technolo gies deliver optimized biplane imaging Designed in concert with leading physicians, the Infinix VF-i/BP provides advanced, versatile patient access to meet the
More informationARCO Rk.5 Digital Mobile C-Arm
Product Data ARCO Rk.5 Digital Mobile C-Arm Product Data 1 Easy to move INTRODUCTION Designed with the latest technology throughout, the ARCO Rk.5 set new standards for the excellence image quality ease
More informationINTRODUCTION TO FLEXIBLE BRONCHOSCOPY. Fluoroscopy Synopsis HENRI G COLT MD SECOND EDITION THE BRONCHOSCOPY EDUCATION PROJECT SERIES
SECOND EDITION INTRODUCTION TO FLEXIBLE BRONCHOSCOPY Fluoroscopy Synopsis HENRI G COLT MD With contributions from Dr. S. Murgu THE BRONCHOSCOPY EDUCATION PROJECT SERIES FLUOROSCOPY SYNOPSIS The purpose
More informationSurveying and QC of Stereotactic Breast Biopsy Units for ACR Accreditation
Surveying and QC of Stereotactic Breast Biopsy Units for ACR Accreditation AAPM Annual Clinical Meeting Indianapolis, IN August 5, 2013 Learning Objectives Become familiar with the recommendations and
More informationof sufficient quality and quantity
of sufficient quality and quantity The patient s body attenuates the beam as it passes though the body More energy is deposited in organs located near the entry of the beam than near the exit of the beam
More informationVisualization of sources of scattered radiation from x-ray equipment used for interventional radiology
Visualization of sources of scattered radiation from x-ray equipment used for interventional radiology Poster No.: C-1190 Congress: ECR 2011 Type: Scientific Exhibit Authors: K. Chida, T. Takahashi, D.
More informationTime to take the gloves off: the use of radiation reduction gloves can greatly increase patient dose
JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, VOLUME 15, NUMBER 6, 2014 Time to take the gloves off: the use of radiation reduction gloves can greatly increase patient dose Alexander S. Pasciak, 1 and A.
More informationEstimation of signal transfer property for wireless digital detector in different measurement schemes
Estimation of signal transfer property for wireless digital detector in different measurement schemes Anatoli Vladimirov, Kalle Kepler Training Centre of Medical Physics, University of Tartu, Estonia 11
More information3D Diode Array Commissioning: Building Confidence in 3D QA Technology
3D Diode Array Commissioning: Building Confidence in 3D QA Technology Caroline Yount, MS CANCER CENTER 3D QA The complex three-dimensional (3D) shapes of intensity modulated radiation therapy (IMRT) dose
More informationRenato Padovani, University Hospital, Udine, Italy
DATAMED II Training Course Sofia, May 19-20, 2011 Patient dose assessment in fluoroscopy and interventional procedures Renato Padovani, University Hospital, Udine, Italy Content Fluoroscopy guided diagnostic
More informationDRX Plus Detectors: Going from Good to Great
DRX Plus Detectors: Going from Good to Great Authors: Karin Töpfer, Tim Wojcik Introduction Carestream s introduction in 2009 of the world s first portable, wireless, cassette-sized detector the CARESTREAM
More informationFOUR CATEGORIES OF SAFETY
OCTOBER 2013 FOUR CATEGORIES OF SAFETY DOSIMETRY PERSONAL SAFETY EQUIPMENT EQUIPMENT KNOWLEDGE PHYSICAL SAFETY DOSIMETRY THERMAL LUMINISCENT DEVICES AND FILM BADGES CNSC PERMISSIBLE DOSES WHOLE BODY DOSE
More informationCollimation Assessment Using GAFCHROMIC XR-M2
Collimation Assessment Using GAFCHROMIC XR-M2 I. Introduction A method of collimation assessment for GE Senographe full-field digital mammography (FFDM) systems is described that uses a self-developing
More informationComparison of computed radiography and filmõscreen combination using a contrast-detail phantom
JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, VOLUME 4, NUMBER 1, WINTER 2003 Comparison of computed radiography and filmõscreen combination using a contrast-detail phantom Z. F. Lu,* E. L. Nickoloff, J.
More informationNATIONWIDE EVALUATION OF X-RAY TRENDS (NEXT) SUMMARY OF 2003 FLUOROSCOPY SURVEY
CRCPD Publication E-09-5 Available Online at No Charge NATIONWIDE EVALUATION OF X-RAY TRENDS (NEXT) SUMMARY OF 2003 FLUOROSCOPY SURVEY September 2009 Published by Conference of Radiation Control Program
More informationSarah Hughes, MS, DABR Radiation Safety Officer
Sarah Hughes, MS, DABR Radiation Safety Officer 502-852-6146 sarah.hughes@louisville.edu Mo my back is burnin!!! I got it MAG the cine! Sumthin s not right. Where s his heart? Fluoroscopy http://dccwww.bumc.bu.edu/fluoroscopy/def
More informationNuclear Associates
Nuclear Associates 07-647 R/F QC Phantom Operators Manual March 2005 Manual No. 07-647-1 Rev. 2 2004, 2005 Fluke Corporation, All rights reserved. All product names are trademarks of their respective companies
More informationEffect of patient support pads on image quality and dose in fluoroscopy a)
Effect of patient support pads on image quality and dose in fluoroscopy a) William R. Geiser, Walter Huda, b) and Nikolaos A. Gkanatsios Department of Radiology, University of Florida, Gainesville, Florida
More informationInstallation und Kommissionierung des Viewray MRIdian Linac Hamburg, 28. Mai 2018 Sebastian Klüter
Installation und Kommissionierung des Viewray MRIdian Linac Hamburg, 28. Mai 2018 Sebastian Klüter MR-guided RT in Heidelberg Funded by the German Research Foundation (DFG) Heidelberg consortium received
More informationDosepix Detector as kvp-meter in Radiology and Mammography: First steps
Dosepix Detector as kvp-meter in Radiology and Mammography: First steps F.Bisello, I.Ritter, F.Tennert, A.Zang MediPix Collaboration Meeting, 19th February 2014, CERN Protect, Enhance, and Save Lives -
More informationCR Basics and FAQ. Overview. Historical Perspective
Page: 1 of 6 CR Basics and FAQ Overview Computed Radiography is a term used to describe a system that electronically records a radiographic image. Computed Radiographic systems use unique image receptors
More informationdiagnostic examination
RADIOLOGICAL PHYSICS 2011 Raphex diagnostic examination Adel A. Mustafa, Ph.D., Editor PUBLISHED FOR: RAMPS (Radiological and Medical Physics Society of New York) preface The RAPHEX Diagnostic exam 2011
More informationTechnical aspects on DAP calibration and CT calibration
EURADOS Report 2015-03 European Radiation Dosimetry Group e. V. Braunschweig, August 2015 Technical aspects on DAP calibration and CT calibration H. Järvinen, O. Turak, M. Ginjaume, J. Daures, C. Hourdakis,
More informationX-RAY FLUOROSCOPY IMAGING SYSTEMS. Dr Slavik Tabakov. Luminescence: Dept. Medical Eng. & Physics King s College London
X-RAY FLUOROSCOPY IMAGING SYSTEMS Dr Slavik Tabakov OBJECTIVES - Image Intensifier construction - Input window - Accelerating and focusing electrodes - Output window - Conversion factor - II characteristics
More informationFluoroscopy - Chapter 9
Fluoroscopy - Chapter 9 Kalpana Kanal, Ph.D., DABR Lecturer, Diagnostic Physics Dept. of Radiology UW Medicine a copy of this lecture may be found at: http://courses.washington.edu/radxphys/physicscourse04-05.html
More informationMAN Revision 002. InSight Mini C-arm Imaging System Technical Reference Manual
MAN-00668 Revision 002 InSight Mini C-arm Imaging System Technical Reference Manual April 2007 The information contained in this Manual is confidential and proprietary to Hologic, Inc. This information
More information2017 West Coast Educators Conference Orlando. Projection Geometry. 1. Review hierarchy of image qualities (amplified version):
Spatial Resolution in the Digital Age: NOTES Quinn B. Carroll, MEd, RT 2017 West Coast Educators Conference Orlando Projection Geometry 1. Review hierarchy of image qualities (amplified version): a. Maximum
More informationOverview. Professor Roentgen was a Physicist!!! The Physics of Radiation Oncology X-ray Imaging
The Physics of Radiation Oncology X-ray Imaging Charles E. Willis, Ph.D. DABR Associate Professor Department of Imaging Physics The University of Texas M.D. Anderson Cancer Center Houston, Texas Overview
More informationQuality Control for Stereotactic Breast Biopsy. Robert J. Pizzutiello, Jr., F.A.C.M.P. Upstate Medical Physics, Inc
Quality Control for Stereotactic Breast Biopsy Robert J. Pizzutiello, Jr., F.A.C.M.P. Upstate Medical Physics, Inc. 716-924-0350 Methods of Imaging Guided Breast Biopsy Ultrasound guided, hand-held needle
More informationSTEREOTACTIC BREAST BIOPSY EQUIPMENT SURVEYS
STEREOTACTIC BREAST BIOPSY EQUIPMENT SURVEYS JAMES A. TOMLINSON, M.S. Diagnostic Radiological Physicist American Board of Radiology Certified Medical Physics Consultants, Inc. Bio 28 yrs experience 100%
More information8/3/2016. The EPID Strikes Back. Novel Applications for Current EPID Technology. Joerg Rottmann, PhD. Disclosures and acknowledgements
The EPID Strikes Back Joerg Rottmann Brigham and Women s Hospital / Dana-Farber Cancer Institute Harvard Medical School Disclosures and acknowledgements Disclosures Varian MRA grant Acknowledgements Boston
More information