Operating manual. Digital Transducer-Electronics AED 9301A. Part 1, Profibus basic device
|
|
- Allyson Phillips
- 5 years ago
- Views:
Transcription
1 Operating manual Digital Transducer-Electronics AED 9301A Part 1, Profibus basic device
2 1/24 Part 1 Description of the hardware of the AED9301A-basic device Contents Page Safety Notes Introduction and se Mechanical construction Electrical connection Transducer connect. AED9301A with AD103 or AD101B Connection of the operating supply voltage Profibus connection Connection digital inputs/outputs Cable connection AED 9301A via PG unions Technical data of the basic device AED9301A Part 2 Description of the hardware and the functions of the measurement amplifier board AD103 or AD101B Part 3 Description of the commands for the serial communication with the measurement amplifier board AD103 or AD101B Part 4 Description of the commands for dosing / filling control with the measurement amplifier board AD103 Part 5 Description of the Profibus communication
3 2/24 Important notes Without our express approval, the device must not be modified in terms of its design and/or technical safety aspects. Any modification excludes any and all liability on our part for any damage resulting therefrom. In particular, it is prohibited to carry out any repairs, soldering work etc whatsoever on the boards and to exchange components. Repairs may only be carried out by HBM. The complete factory settings are stored nonvolatile in the measurement board AD101B and cannot be deleted or overwritten; they can be reset at any time by means of the command TDD0. You will find further information in the chapter Individual descriptions of the commands AD101B, AD103; Part 3. The manufacturing number set in the factory should also not be changed. The transducer must always be connected up (connect transducer or bridge model).
4 3/24 Safety Notes Normally, the product does not represent any hazard if the notes and instructions for project planning, assembly, use within specifications and maintenance are complied with. The relevant accident prevention regulations, as applicable in each individual case, must be complied with. Assembly and commissioning may only be carried out exclusively by qualified personnel. Prevent dirt and humidity from ingressing into the inside of the equipment. When connecting lines implement measures against electrostatic discharges which may cause damage to the electronics. For power supply, a small voltage ( V) with safe disconnection from mains is required. When connecting additional units, the safety regulations according to EN ) must be complied with. se shielded lines for Profibus and transducer connections. The screen is to be connected flush to ground on both sides. Lines for connecting the power supply as well as the digital I/Os are to be implemented as shielded lines only if a cable length of 30m is exceeded or if lines are laid outside enclosed buildings. 1) "Safety regulations for electrical measurement, control, regulatory, and laboratory units"
5 4/24 1 Introduction and se The digital transducer electronic units AED9301A belong to the family of AED components which digitally condition and network as bus-capable signals from mechanical measured value transducers. This includes digital measurement amplifier boards, base units with RS-485 or Profibus DP interface and intelligent sensors with integrated signal processing. The objective of these components is the digitization and conditioning of the measuring signals directly at the transducer location. Via the digital transducer electronics AED9301A, SG 1) transducers can be connected directly to a Profibus DP within a full bridge circuit. This enables you to build up complete measurement chains fast and at low cost. AED9301basic: = Basic device AED9301A + Amplifier AD101B AED9301plus: = Basic device AED9301A + Amplifier AD103 The digital inputs and outputs offer the controling of processes via two limit values (LIV1/2), the detrmination of triggered measing values (MAV), and the controling of a filling or dosing process (only AD103) The transducer electronic units AD101B and AD103 are also abbreviated with AED in the following text. 1) Strain Gages
6 5/24 2 Mechanical construction The basic device AED9301A offers the following functions: Degree of (mechanical) protection (IP65) Slot for measurement amplifier board AD101B or Ad103 Voltage supply for the measurement amplifier board and transducer supply (potential-isolated) Total bridge resistance transducer >80Ω Ω Profibus DP interface (potential-isolated) from measurement amplifier and from the digital inputs / outputs) Digital inputs/outputs (potential-isolated from measurement amplifier and from Profibus) EMC protection The measuring amplifier board is designed as a plug-in type board, which can be plugged into the carrier board of the basic unit via a 25pin D connector. The basic device comprises the terminals for transducer, power supply unit, digital inputs outputs and profibus connection, the slide switches for bus connection and end-resistors, as well as the voltage stabilizer. The connection cables are fed out of the side of the housing, via PG boltings.
7 6/24 Connector meas. Amplifier Voltage supply and digital I/O LED 4: red LED 3: yellow LED 2: green LED 1 Profibus- LEDs Transducerconnection Adresssetting Profibusconnection Bus termination and diagnosis switch Fig. 2-1: Mechanical construction AED9301A (without measurement amplifier)
8 7/24 3 Electrical connection A connection diagram is enclosed with the basic unit AED 9301A. When making connections, please ensure that the cable wires do not protrude beyond the connection terminals (risk of short circuits). Please also ensure that the cable shield is correctly connected to the PG bolting (see Chapter 5.5). 3.1 Transducer connect. AED9301A with AD103 or AD101B Important note The transducer must always be connected up (connect transducer). AED9301A with AD101B It is possible to connect strain gage transducers in a full bridge circuit with a total bridge resistance of R B = Ω. The bridges are supplied by means of the basic device AED9301A (5V DC ). Bridge supply voltage (+) Sense line (+) Meas. signal (-) Meas. signal (+) Sense line (-) Bridge supply voltage (-) Cable shield on housing Wire colour blue green red white grey black Pin Fig : Transducer connection in 6-wire technology (colour coding HBM) The 6 wire connection avoids a long cable influencing the measured value. When using several transducers and a junction box, the 6-wire technology is routed through to the junction box.
9 8/24 Load cell KL B GN RD WH GY BK Fig : Transducer connection KL 4 in the basic device AED9301A as 6-wire connection (Wire colours: B blue, GN - green, RD red, WH white, GY grey, BK black) For transducers that are calibrated in four-wire-technology, there are two types of connection: Connection via a six wire extension cable; sensor line bridged in transducer connector ) Connector Transducer bl rt ws sw bl gn rt ws gr sw Cable to transducer electronics Cable shield Fig : Transducer connection in four-wire technology via six-wire cable extension (HBM colours)
10 9/24 Connection without an extension cable; sensor line bridged on the transducer electronics ( connect 3 to 3 and 2 to 2 ) When connecting several transducers, it is recommended to use a junction box VKK4 or VKK6 by HBM. In general the supply lines to the AED should be implemented as shielded lines. Notes on connection type, cable length and cable cross-section: Depending on the bridge resistance of the transducer used, cable length, and cable cross-section of the load cell connection cable, voltage drops arise that lead to a reduction in the bridge supply voltage. Additionally, the voltage drop on the connection cable is also temperature-dependent ( copper resistance ). The load cell output signal also changes in proportion to the bridge supply voltage. Effect on the measurement result with ATOCAL (command ACL1) activated: 6-wire circuit (standard operation mode): With the 6-wire-circuit the return lines 2 and 3 in the load cell ( for load cells with 6-wire-connection) or directly on the load cell connection ( for load cells with 4-wire circuit) with the supply lines 2 and 3 bridged (Fig ). This corrects all influences. Even a change of cable length following a calibration does not lead to any deviations in the measurement results. 4-wire circuit: With the 6-wire-circuit the return lines 2 and 3 on the connection terminals 2 and 3 are connected directly to the AED (Fig ). As the correction by ATOCAL can only be effected up to the return points 2, 3, all changes in cable resistance are reflected in the measurement result. That is, even if the 4- wire cable used for a calibration is no longer changed, there still result measurement errors in conditions with changing temperatures, caused by the temperature-dependent cable resistance and possibly also by transitory resistances in the connectors.
11 10/24 Replacement circuit for the bridge with bridge resistance RB and the supply lines with line resistances R L1 and R L2 : R L1 = R L2 : = (4 x ρ C /π) x (l[m] / A[mm 2 ] R L1 ρ C = 0,0178 [Ωmm 2 /m] for copper R B R L2 BR π = 3.14 l = line length, A = line cross section R L1 = R L2 = 1.6Ω at l = 10m and A = 0.14mm 2 From the bridge resistance R B and the line length I, line cross-section A and the bridge supply voltage, the voltage across the supply lines of the bridge can be determined: B + RL1 + RL2 = BR For R B = 80Ω, R L1 = R L2 = 1.6Ω (l = 10m) and BR = 5V there results a a supply current I BR = BR / (R L1 + R L2 + R B ) = 60mA and thus a voltage drop across the two line resistors of a total approx. 0.2V ( Bridge = 4.8V). For R B = 80Ω, R L1 = R L2 = 16Ω (l = 100m) and BR = 5V there results a a supply current I BR = BR / (R L1 + R L2 + R B ) = 45mA and thus a voltage drop across the two line resistors of a total approx. 1.4V ( Bridge = 3.6V). For the 6-wire circuit, this is without significance as the voltage drop is taken into account via the sensor lines in the measurement signal.
12 11/24 However, for the 4-wire circuit, the dependence of the copper resistance of the lines directly influences the measurement result via the temperature, as the bridge supply voltage Bridge changes: R L(t) := R L20 (1 + α (t 20 C)), with R L20 being the line resistance at 20 C and α being the temperature coefficient of copper R L20 Calculation see page 11, α C := [1/K] For a line length of l= 100m and a temperature difference of 10 C the following line resistance results R L1(t) = R L2(t) := 16 ( ) = 16.6 Ω This changes the bridge supply voltage from Bridge = 3.6V (at 20 C) to Bridge = 3.53V. This change in the bridge supply voltage directly at the transducer changes the measured signal of the bridge by 2% ( = 100% (1 3.53V / 3.6V)). This exemplary calculation shows that for long line lengths only 6-wire technology may be used.
13 12/ Connection of the operating supply voltage The voltage supply must meet the following requirements: AED9301A supply voltage range B = +18V V (DC) Current consumption = 200mA + current of control outputs OT1..4 (for 80Ω Bridge / 24V supply, see technical data Chapter 6) 0V 24VDC external PS N I B B B KL1 Digital I/ O KL2 B B V V other 0 V units Fig : Connection of voltage supply B The AED supply voltage can be connected at the terminals KL1 or KL2. The three ground terminals as well as the two voltage terminals, respectively, are internally connected. The supply voltage coming from the power supply unit is connected to terminal KL1; at terminal KL2 this supply voltage can be fed on to other units. The potential-isolated digital outputs OT1..OT4 are also supplied from this voltage. Potential-separation is effected in the direction of the AED. Thus, the units controlled by OT1..4 can also be supplied from B (see Section 5.4)
14 13/24 The control inputs IN1 and IN2 are initially potential-separated from the supply voltage B. The two grounds (ground B and ground IN) can be connected to terminal KL1, if required (see Section 5.4). 3.3 Profibus connection The Profibus DP is completely standardized in accordance with IEC / EN for universal automation; this provides for the troublefree connection of standard components. In the case of the AED9301A, the Profibus is a DP Slave according to DIN It offers a simple and fast to implement way in which to connect electro-mechanical measured value transducers with automation systems such as e.g. SIEMENS SIMATIC S7 or PCs. The Profibus is potentail-isolated separated from the measuring system and from the supply voltage, the transfer process is RS485. The maximum possible baud rate for RS485 is 12 Mbaud. The subscriber address can be set by means of two BCD coded rotary switches S5 and S4. The Profibus connection KL3 is equipped with four terminals so that it can be fed through to the next bus subscriber. The connection lines should be implemented as shielded and twisted 2-wire lines (see Profibus cable specification). At both ends of the Profibus line, the bus termination must be activated (with the AED switch S2 = on). For all other bus subscribers this switch is to be set to 'off'. For diagnosis purposes the module can be disconnected from the Profibus (S3 = off), the works setting is on.
15 14/24 S S Profibus KL3 IN OT A B A B Bus conn. on off S3 S2 Bus term. off on from Master next unit Fig : Profibus - connection via terminal KL3 Maximum cable length in relation to the baud rate: Baud rate [kbit/s] Max. Cable length [m] Setting the Profibus address: Address = S5 x 10 + S4 (permitted adress range: ) The address is to be set with the unit in a deactivated condition and is read in from the AED when supply voltage is applied. The address is set to 03 at the factory.
16 15/24 Function of light emitting diodes (LEDs) Different Profibus states are indicated by 4 LEDs (Fig. 2-1) LED1: Voltage supply (green, r/h next to the diagnosis switch) supply voltage of the RS485 driver is applied, if LED 1 is constantly illuminated in green. LED2: Profibus Data Exchange (green) The 'Data Exchange' state is indicated by the green LED 2 in cyclic data exchange. LED3: Profibus Diagnosis (yellow) The yellow LED 3 illuminates if there is an internal fault. The measured values are possibly invalid. LED4: Profibus Error (red) In the event of a bus error, LED 4 will illuminate in red for as long as the error state exists. Possible causes of the fault: - Incorrect wiring (A and B poss. mixed up) - Profibus Master does not (yet) operate Installation of the Profibus: 1. Connection of the supply lines to the AED when deactivated (acc. to Chapters 5.1 to 5.5). Via KL3, the Profibus line is fed through from one unit to the next. 2. Setting the subscriber address via S5, S4. 3. Set switch for connecting in the bus termination (S2) to 'on' only for the first and last units; otherwise leave in position 'off'. If the first or last bus subscriber is not an AED, then it is usually possible to connect in the bus termination by means of a slide switch on the housing of the Profibus connector. 4. Check whether the switch for the bus connection (S3) is set to on. 5. Switch-on of the supply voltage B. 6. Configuration and parameterization of the Profibus subscriber with the respective tools.
17 16/24 The configuration and parameterization of the Profibus telegram by means of a configuration tool and GSD file is described in the operating instructions, Part 5 (communication via Profibus with AED9301A), as well as the bus coupling via PC and the HBM Profibus Panel Program. For diagnosis purposes the Profibus can be disconnected from the AED (bus connection (S3) to 'off') 3.4 Connection digital inputs/outputs At the terminals KL1 and KL2, the digital control inputs and control outputs and their associated reference grounds are connected up. The amplifier is allways disconnected potentially from supply voltage B and the igital inputs and outputs. Control inputs: The control inputs IN1 and IN2 are initially isolated from the supply voltage B and the measurement ground, the reference potential is ground IN on terminal KL1 on the right-hand side of the inputs. PS 2 0V V DC PS 1 IN1 IN2 0V 24 V DC KL1 Digital I/O 1 N 2 N I N I I B B B V KL2 T1 O T2 O T3 O T4 O B B external load Fig : Connection digital inputs and outputs, inputs and external power supply unit 1 potential isolated
18 17/24 Logic level: IN1: Trigger: Break Dosing: Quiescent level = High, active Flanke = High Low- edge Quiescent level = Low, activat. = Low High Low-Impuls (duration 20ms) IN2: Quiescent level = Low, activat. = Low High Low- Impuls (duration 20ms) Function: Tare or Start dosing Note nused inputs remain open. If the input circuit is also supplied via B, then the ground of the inputs and the ground of B must be connected. PS IN1 IN2 0V 24V DC Digital I/O O O O O External load Fig : Connection digital inputs and outputs with one external power supply unit (i.e. IN1=Trigger)
19 18/24 Control outputs: The digital outputs OT1..4 are potential-isolated towards the AED and are supplied via the external supply voltage B. They are designed as H-sideswitches. Thus the loads need to be connected against ground. The outputs are short-circuit proof and can drive ohmic and inductive loads with currents up to approx. 0.5 A per output. Logic level: OT inactive voltage is low (H-side-switch deactivated) OT active voltage is high (H-side-switch activated The functions of the digital inputs and outputs are different according to the measuring amplifier AED9301basic The measurement amplifier board AD101B features two inputs (IN1 and IN2) and two outputs (OT1 and OT2, do not connect OT3,4). The functions are determined via the commands IMD, LIV (see operating instructions, Part 3): Input functions: IMD0: Input functions deactivated, status can be read in via the command POR. IMD1: IN1 = Tare, IN2 = external trigger for the trigger function (TRC). IMD2: not selectable Output functions: LIV1 deactivated: LIV2 deactivated: LIV1 activated: LIV2 activated: Control OT1 via POR command Control OT1 via POR command Limit value LIV1 controls output OT1 Limit value LIV2 controls output OT2 The outputs OT3/4 cannot be driven by the AD101B.
20 19/ AED9301plus The measurement amplifier board AD103 features two inputs (IN1 and IN2) and 4 outputs (OT1...4). The functions are determined via the commands IMD, LIV, OMD (see operating instructions, Part 3): Input functions: IMD0: Input functions deactivated, status can be read in via the command POR. IMD1: IN1 = external trigger for the trigger function (TRC), IN2 = Tare, IMD2: IN1 = stop filling, IN2 = start filling (Dosing function) Output functions: IMD < 2 (no dosing mode, as AD101B): LIV1 deactivated: Control OT1 via POR command LIV2 deactivated: Control OT2 via POR LIV1 activated: Limit value LIV1 controls output OT1 LIV2 activated: Limit value LIV2 controls output OT2 The outputs OT3/4 cannot be driven. IMD = 2 (dosing mode, only for AD103, see manual part 4) OT1: Coarse flow - control OT2: Fine flow - control OT3: Ready signal or emptying control (see operating instructions AD103, Part 4, dosing) OT4: Tolerance fault or alarm (see operating instructions AD103, Part 4, dosing)
21 20/ Cable connection AED 9301A via PG unions As a connection line between the AED 9301A and transducer or Profibus DP, only a connection line with a shield grounded on both sides may be used. The shield is to be full-surface applied to the PG union. The cable shield is to be fullsurface connected to ground (housing). If large ground potential differences exist between AED9301A and the partner device, an equipotential bonding conductor is to be provided additionally. Dep. on req. wire length L remove ext. cable sheath Strip off wire ends and tin. Slide cable unions with sealing ring and press. rings over the cable ends Shorten cable shields and strip off strands. Ground sleeve Fan out cable shield radially. Slide ground sleeve between strands and cable shield to stop pos., press shield against sleeve flange, cut off protrusions. Slide cable through intermediate supports on the housing to stop pos., adv. cable union and bolt securely. Fig : Cable connection via PG unions
22 21/24 4 Technical data of the basic device AED9301A Type Transducer Measured signal input Bridge excitation voltage DC SG transducer (1...4 full bridges, each 350Ω), R B mv/v V Ω AED9301A ±3, nominal ±2 5 >80 Transducer connection Transducer cable length Profibus DP Protocol Baud rate, max. Subscriber address, can be set by switch Interface cable length Profibus Control inputs (potential-isolated) Number Input voltage range, LOW Input voltage range, HIGH Input current, typ., High-level = 24 V Control outputs (potential-isolated) Number max. output current per output Short circuit current, typ., B = 24 V; R L < 0,1 Ω Short circuit duration Output current at Low-level Output current at High-level Insulation voltage, typ. (DC) Supply Operating voltage (DC) Current consumption (with load cell (R B = 80Ω) plus output currents of control outputs Iout1..4) m Mbit/s m V V ma A A ma V V V ma 6-wire circuit 100 Profibus-DP Slave, acc. DIN (for 9,6 / 19,2 / 93,75 kbit/s) 1000 (for 187,5 kbit/s) 400 (for 500 kbit/s) 200 (for 1,5 Mbit/s) 100 (for 12 Mbit/s) unlimited <2 >15 at I max (Calculation see below)
23 22/24 Technical Data, Continuation Type Temperature range Nominal temperature range Service temperature range Storage temperature range In addition Dimensions (L x W x H) Gewicht Degree of protection DIN40050 (IEC529) C C C mm g AED9301A x 100 x 70 approx. 925 IP65 Calculation of the total current consumption (for 80Ω bridge) Current consumption at 18V supply: 250mA + I OT Current consumption at 24V supply: 200mA + I OT Current consumption at 30V supply: 170mA + I OT I OT = Current of the control outputs The technical data of the measuring amplifier AD101B or AD103 are included in the operating manual AD101B, AD103;Part 2. Within the shielded structure (see Sect. 3), the entire measuring chain including the AED is insensitive against HF interference and conducted interference in accordance with OIMLR76, EN45501 or EN55011B (interference emissions) and EN
24 Modifications reserved. All details describe our products in general form. They are not to be understood as express warranty and do not constitute any liability whatsoever. AED9301A_P1_e_ Hottinger Baldwin Messtechnik GmbH Postfach D Darmstadt Im Tiefen See 45 D Darmstadt Tel.: +49/6151/803-0 Fax: +49/6151/
User Manual AED9501A. Digital Transducer Electronics Basic device CANOpen/DeviceNet. I en
User Manual Digital Transducer Electronics Basic device CANOpen/DeviceNet I2331-1.2 en Table of content Content Content......................................................................... 1 Typographical
More informationCANHEAD. Data Sheet. Special features. Distributed measurement acquisition. B en
CANHEAD Data Sheet Special features 10channel amplifier modules for installation close to measuring points Measured data transmission to communication master via field bus Base modules for individual SGs,
More informationOperating manual. 600 Hz Measuring Amplifier ME30. A en
Operating manual 600 Hz Measuring Amplifier ME30 ME30 Contents 3 Page Safety instructions............................................. 4 1 Measuring Amplifiers on Eurocards ME30.................... 8
More informationOperating manual. Force transducer with strain gage measuring system C9B. A en. hvordan man bruger astrologi med tibetanske medicin
Operating manual Force transducer with strain gage measuring system C9B hvordan man bruger astrologi med tibetanske medicin C9B Contents 3 Page Safety instructions..............................................
More informationData sheet. P3 Top Class BlueLine. P3MBP BlueLine. Ultra-high pressure transducers to 15,000 bar. Special features
Ultra-high pressure transducers to 1,000 bar Data sheet Special features P Top Class BlueLine PMBP BlueLine - For static and dynamic pressure variance, pressure peaks and pressure fluctuations - Principle
More informationE *) F 15. H 12 h9. 5.5f7 8h6. a./f approx. 0.5mm Free stroke. Displacement. Initial stroke
Displacement probe Plunger WA Inductive Standard Displacement Transducers Special features Displacement probe and transducer with detachable plunger Good thermal stability in the event of temperature gradients
More informationSpecifications Type WA2 WA10 WA20 WA50 WA100 WA200 WA300 WA500 Nominal displacement mm Nomi
Displacement probe WA Inductive Standard Displacement Transducers Special features Data Sheet Plunger Displacement probe and transducer with detachable plunger Good thermal stability in the event of temperature
More informationInductive Standard Displacement Transducers
WA Inductive Standard Displacement Transducers Special features - Available as displacement probe or with detachable plunger - Good thermal stability in the event of temperature gradients - Space saving,
More informationelectronics Data Sheet Industrial Amplifier Special features Dimensions (in mm; 1 mm= inches) B en
C l i p electronics Industrial Amplifier Data Sheet Special features Amplifier and additional units for strain gauge full bridges and inductive half and full bridges Modules for mounting onto support rails
More informationMDEX. on customized sensors. Technical information. Torque transducer. CS-T-04-MDEX Special features
MDEX Torque transducer CS-T-04-MDEX Special features - Explosion proof acc. to ATEX 95 - Exx II 2 G Ex deq IIC T4 - Nominal (rates) torque of 100kNm, 130kNm and 150kNm - HBM Accuracy Class 0,1 - Digital
More informationSUNSTAR 传感与控制
SUNSTAR 传感与控制 http://www.sensor-ic.com/ HBM Mess- und Systemtechnik GmbH Electrical measurement of mechanical quantities Operating manual Force transducer with strain gauge measuring system C9B B 20.C9B.20
More informationMultipoint measuring units UPM100 and Centipede100
OTTINGER ALDWIN ESSTECHNIK HBM Mess- und Systemtechnik GmbH Multipoint measuring units UPM100 and Centipede100 S 33.UPM100.10 e Multipoint measuring units UPM100 and Centipede100 HBM s multipoint measurement
More informationT40FM. Data Sheet. Torque flange. Special features. Overall concept. B en
T40FM Torque flange Special features Data Sheet - Nominal (rated) torque: 15 kn m, 20 kn m, 25 kn m, 30 kn m, 40 kn m, 50 kn m, 60 kn m, 70 kn m and 80 kn m - Nominal (rated) rotational speed up to 8000
More informationT10F. Data Sheet. Torque Flange. Special features. Installation example T10F. B en
T10F Torque Flange Data Sheet Special features Extremely short design High permissible dynamic loads High permissible transverse forces and bending moments Very high torsional stiffness Contactless Selectable
More informationT10FS. Data Sheet. Torque Flange. Special features. Installation example T10FS. B en
T10FS Torque Flange Data Sheet Special features Nominal (rated) torques: 100 NVm, 200 NVm, 500 NVm, 1 knvm, 2 knvm, 3 knvm, 5 knvm, 10 knvm Nominal speed from 12,000 rpm to 24,000 rpm Low rotor weights
More informationMX440A. Data Sheet. Universal amplifier. Special features. Block diagram Digital platform. B en
MX440A Universal amplifier Data Sheet Special features 4 individually configurable inputs (electrically isolated) Connection of more than 12 transducers technologies Data rate: up to 19,200 Hz 24 bit A/D
More informationMX1615. Data sheet. Bridge/strain gauge amplifier
MX1615 Bridge/strain gauge amplifier Data sheet Special features 16 individually configurable inputs Connection of strain gauges in quarter-, half- and full-bridge Bridge excitation : DC or carrier frequency
More informationT40FH. Torque flange. Special features. Data sheet
T40FH Torque flange Special features - Nominal (rated) torques: 100kNm, 125kNm, 150kNm, 200kNm, 250kNm, 300kNm - Nominal (rated) rotational speed of 2000 rpm up to 3000 rpm - Compact design - Version for
More informationU10M. Force Transducer. Special features. Data sheet. Mounting dimensions of the connection variants in mm [inch]
U10M Force Transducer Special features - Precise and robust tensile/compressive force transducer for static and dynamic measurement tasks - High lateral force and bending moment stability, the effect of
More informationT22. Data Sheet. Torque transducer. Special features. Installation example with two bellows couplings. B en
T22 Torque transducer Data Sheet Special features - Nominal (rated) torques N m, 10 N m, 20 N m, 0 N m, 100 N m, 200 N m, 00 N m and 1 kn m - Nominal (rated) rotational speeds of 9000 rpm to 16 000 rpm
More informationMX1615B R. Data Sheet. Ultra-rugged Bridge Amplifier
MX1615B R Ultra-rugged Bridge Amplifier Data Sheet Special features 16 individually configurable inputs Connection of strain gages in quarter-, half- or full-bridge Bridge excitation: DC or carrier frequency
More informationData Sheet. Absolute pressure transducer Nominal (rated) pressure 10 bar to 3000 bar. Special features
P3 Top Class P3MB version with fixed cable P3MBP version with plug connection TEDS: Transducer Electronic Data Sheet P3 Absolute pressure transducer Nominal (rated) pressure 10 bar to 3000 bar Special
More informationMX840B R. Data sheet. Ultra-rugged universal amplifier CAN. Special features. Block diagram. B en
MX840B R Ultra-rugged universal amplifier Data sheet Special features 8 individually configurable inputs (electrically isolated) Connection of more than 16 technologies per channel Sampling rates of up
More informationC10. Data sheet. Force transducer. Special features. B de
C10 Force transducer Data sheet Special features Force transducer for static and dynamic applications Made of non-rusting materials Precise (accuracy class from 0.03) Numerous options (double bridge, TEDS,
More informationMX1601B. Data Sheet. Universal amplifier. Special features. Block diagram Digital platform. B en
MX1601B Universal amplifier Data Sheet Special features 16 individually configurable inputs (electrically isolated) Connection of of standard signals (60, 10, 100 m, 20 ma, IEPE) Individual data rates
More informationU2B. Data sheet. Force Transducer. Special features. B en
U2B Force Transducer Special features Data sheet Tensile/compressive force transducer Non-rusting, degree of protection IP67 Can be configured with different cable lengths, plug fitting and TEDS on request
More informationMX1615B-R. Ultra-rugged Bridge Amplifier. Special features. Data sheet. Block diagram
MX1615B-R Ultra-rugged Bridge Amplifier Special features - 16 individually configurable inputs - Connection of strain gages in half-, full- or quarter-bridge (120 Ohm or 350 Ohm) - Bridge excitation: DC
More informationT12. Data sheet. Digital Torque Transducer. Special features. Signal flow block diagram. B en
T12 Digital Torque Transducer Data sheet Special features Nominal (rated) torques 100 N m, 200 N m, 500 N m, 1 kn m, 2 kn m, 3 kn m, 5 kn m and 10 kn m Nominal (rated) rotational speeds from 10 000 rpm
More informationSingle point load cell
PW15AH Single point load cell Special features - Nominal load 10 kg... 100 kg - Stainless steel - High ratio of minimum verification interval Y - Industrial Footprint (SP4M) - Degree of Protection IP68;
More informationU2B. Force Transducer. Special features. Data sheet
U2B Force Transducer Special features - Tensile/compressive force transducer - Non rusting, degree of protection IP67 - Can be configured with different cable lengths, plug fitting and TEDS on request
More informationU15. Force transducer. Special features. Data sheet. Installed dimensions of connection variants
U15 Force transducer Special features - Tensile/compressive force transducer - Nominal (rated) forces 2.5 kn to 2.5 MN - lass 0.5 to ISO 376 in force measurement range between 10% and 100% of the nominal
More informationCPU. BM40, BM40PB, BM40IE Industrial amplifier. Special features. Data sheet. Block diagram
Industrial amplifier Special features - Freely configurable measurement channel with TEDS channel parameterization - Connection of 7 transducer technologies with a 19.2 KHz sample rate - Accuracy class
More informationT40B. Torque Flange. Special features. Data sheet. Overall concept
T40B Torque Flange Special features - Nominal (rated) torques 50 N m, 0 N m, 200 N m, 500 N m, 1 kn m, 2 kn m, 3 kn m, 5 kn m and kn m - Nominal rated rotational speed up to 24000 rpm (depending on nominal
More informationT22. Torque transducer. Special features. Data sheet
T22 Torque transducer Special features - Nominal (rated) torques 0.5 Nm, 1 Nm, 2 Nm, 5 Nm, Nm, Nm, 50 Nm, 0 Nm, 0 Nm, 500 Nm and 1 knm - Nominal (rated) rotational speed up to 000 rpm (depending on measuring
More informationSpecifications (in accordance with DIN 16086) Equipment family Digibar II Type 1-PE50A1/...; K-PE PE00A1/...; K-PE00... Class of accuracy 0.2 0
Digibar II: PE00, PE50 Digital pressure gage Special features Data Sheet PE00 PE50 Operated as manometer with battery operation and Operated as pressure measurement transmitter with 420 current output,
More informationStrain Gauge Measuring Amplifier BA 662
Strain Gauge Measuring Amplifier BA 662 Original of the Manual Table of Contents 1. Safety precautions... 2 1.1. Field of application... 2 1.2. Installation... 2 1.3. Maintenance... 2 2. Function... 2
More informationSIMATIC ET 200SP. Digital input module DI 8xNAMUR (6ES7131-6TF00-0CA0) Manual. Answers for industry.
SIMATIC ET 200SP Digital input module DI 8xNAMUR (6ES7131-6TF00-0CA0) Manual Edition 02/2014 Answers for industry. DI 8xNAMUR HF digital input module (6ES7131-6TF00-0CA0) SIMATIC ET 200SP DI 8xNAMUR HF
More information8V General information. 2 Order data 8V
8V80.00-8V80.00- General information Modular mechanical design using plug-in modules Integrated line filter Integrated or optional external braking resistor All connections are made using plug-in connectors
More informationPHOENIX CONTACT - 09/2009
Electronic miniature circuit-breaker CLIPLINE Data sheet 03906_en_0 PHOENIX CONTACT - 09/2009 Description The EC-E... electronic miniature circuit-breaker selectively protects all 24 V DC load circuits
More informationU9C. Data sheet. Force Transducer. Special features. B en
U9C Force Transducer Data sheet Special features Tensile/compressive force transducer Accuracy class 0.2 Nominal (rated) forces 50 N 50 KN Non-rusting, protection class IP67 Configurable with different
More informationCharacteristics and functioning
Characteristics and functioning /25 ENOD4 PRODUCT RANGE:... 4. General presentation:... 4.2 enodview software tool... 4.3 Versions and options:... 4.3. Versions:...4.3.2 Options :...4.4 Dimensions:...
More informationonlinecomponents.com
PSM-ME-/-P. Short Description The rail-mountable and compact RS-485 repeater, PSM-ME-/-P, was specially developed for the high demands for use in fieldbus systems. The device is snapped onto standard EN
More informationMODEL AT-10 ANALOG TRANSMITTER
MODEL AT-10 ANALOG TRANSMITTER INSTALLATION & OPERATING MANUAL Industrial Weighing Systems 9 Richmond Street Picton, Ontario K0K 2T0 Phone: 613-921-0397 Fax: 613-476-5293 Web: www.iwsystems.ca info@iwsystems.ca
More informationMINI MCR-SL-UI-f(-SP)
Configurable Analog Frequency Transducer INTERFACE Data Sheet PHOENIX CONTACT - 02/2006 Description The configurable analog frequency transducer MINI MCR-SL-UI-f(-SP) is used to convert analog standard
More informationMoxa ICF-1280I Series Industrial PROFIBUS-to-Fiber Converter
Moxa ICF-1280I Series Industrial PROFIBUS-to-Fiber Converter Hardware Installation Guide First Edition, August 2013 2013 Moxa Inc. All rights reserved. P/N: 1802012800011 Introduction The ICF-1280I series
More informationCharacteristics and functioning
Characteristics and functioning 1/36 enod4-d Characteristics and functioning NU-eNod4D-ETH-E-1014_216710-A 1 ENOD4 PRODUCT RANGE... 5 1.1 General presentation... 5 1.2 Versions and options... 5 1.2.1 Versions...
More informationSCHMIDT Flow Sensor SS 25.60
Product description Product advantages Form 107/06/09/1000/3 Art. No. 511 047.02 A Data subject to change The flow sensor is a mass flow sensor (thermal anemometer), designed for use in oxygen. Due to
More informationMP85A/ MP85ADP. FASTpress. Data Sheet. Special features MP85A / MP85ADP. B en
MP85A/ MP85ADP FASTpress Data Sheet Special features 100% quality control of the production process Powerful algorithms for monitoring fitting, testing and press fitting processes Universal twin channel
More informationT12. Digital Torque Transducer. Data Sheet. Special features. Block diagram signal flow. B en
T12 Digital Torque Transducer Data Sheet Special features Nominal (rated) torques of 100 NVm, 200 NVm, 500 NVm, 1 knvm, 2 knvm, 3 knvm, 5 knvm and 10 knvm Nominal (rated) speeds from 10,000 rpm to 18,000
More information8V General information. 2 Order data 8V
8V05.00-8V05.00- General information Modular mechanical design using plug-in modules Integrated line filter Integrated braking resistor All connections are made using plug-in connectors Integrated electronic
More informationMX1615B. Strain gauge bridge amplifier. Special features. Data sheet. Block diagram
MX1615B Strain gauge bridge amplifier Special features - 16 individually configurable inputs - Connection of strain gauges in full-, half-, or quarter-bridge (120 or 350 Ohm) - Bridge excitation : DC or
More informationSingle point load cell
PW15AH Single point load cell Special features - Nominal load 10 kg... 100 kg - Stainless steel - High ratio of minimum verification interval Y - Industrial Footprint (SP4M) - Degree of Protection IP68;
More informationTechnical manual. Microstep driver SMC11. NANOTEC ELECTRONIC GmbH & Co. KG Gewerbestraße 11 D Landsham near Munich, Germany
Technical manual Microstep driver NANOTEC ELECTRONIC GmbH & Co. KG Gewerbestraße 11 D-85652 Landsham near Munich, Germany Tel. +49 (0)89-900 686-0 Fax +49 (0)89-900 686-50 info@nanotec.de Editorial Editorial
More informationMX1616B. Strain gauge bridge amplifier. Special features. Data sheet. Block diagram
MX1616B Strain gauge bridge amplifier Special features - 16 individually configurable inputs - Connection of strain gauges in full-, halfand quarter-bridge (350 or 1,000 Ohm) - Bridge excitation : DC or
More informationInstallation- / Monitoring Technique
Installation- / Monitoring Technique VARIMETER Insulation fault locator RR 5887 0269248 Product Description 4-channel design The locating current generator RR 5886 in connection with the insulation fault
More informationNOVOHALL Rotary Sensor non-contacting. Series RSC2800 digital SSI, SPI, Incremental
NOVOHALL Rotary Sensor non-contacting Series RSC2800 digital SSI, SPI, Incremental The RSC 2800 sensor utilizes a contactless magnetic measurement technology to determine the measured angle. Unlike conventional
More informationPROFIBUS HUB REPEATER
USER S MANUAL PROFIBUS HUB REPEATER RHP303 R H P 3 0 3 M E smar www.smar.com Specifications and information are subject to change without notice. Up-to-date address information is available on our website.
More informationSCHMIDT Flow Sensor SS SS FB
Product description Flow sensor with integrated microcontroller transducer for the measurement of air-flow velocity and for the determination of the flow rate and flow volume. The aerodynamically styled
More informationDigital electronic module 4DO DC24V/2A HF (6ES7132-4BD30-0AB0) SIMATIC
Digital electronic module 4DO DC24V/2A HF (6ES7132-4BD30-0AB0) SIMATIC Properties 1 Parameters 2 Diagnostics 3 ET 200S distributed I/O Digital electronic module 4DO DC24V/2A HF (6ES7132-4BD30-0AB0) Manual
More informationInstallation and Operating Manual
GB Analog Weight Transmitter PS-1000 Installation and Operating Manual LIST OF CONTENTS Section 1, General information Introduction... 3 Description... 3 Specifications... 4 Wiring connections... 5 Section
More informationMETRISO. Analog Insulation, Low Resistance and Voltage Measurement Instrument /9.14
3-349-813-3 1/9.14 Insulation measurement per EN 61557-2/VDE 413, part 2 Test voltages: 5 V, 1 V, 25 V, 5 V and 1 V Analog display of measured values and limit values Intelligent filter precise and measurement-dependent
More informationEY-EM 522, 523: Remote I/O module, ecolink522, 523
Product data sheet 94.077 EY-EM 522, 523: Remote I/O module, ecolink522, 523 How energy efficiency is improved Regulation, control, monitoring and optimisation of operational systems, e.g. room automation
More informationMX460B-R. Rugged pulse and frequency measuring module. Data sheet. Special features. Block diagram
MX460B-R Rugged pulse and frequency measuring module Special features - Four individually configurable inputs (electrically isolated) - Measure any kind of digital signals up to 1 Mhz for measurement of
More informationC9C. Force Transducer. Special features. Data sheet
C9C Force Transducer Special features - Compact design compressive force transducer - Accuracy class 0.2 - Nominal (rated) forces 50 N to 50 kn - Configurable with different cable lengths, plug assembly
More informationLED Driver Compact fixed output
w LED Driver Driver LCI 100W 900mA 1750mA TOP C TOP series Product description Fixed output built-in LED Driver Constant current LED Driver Output current settable 900 1,750 ma Max. output power 100 W
More informationMX1601B-R. Ultra rugged Standard Amplifier. Data sheet. Special features. Block diagram
MX1601B-R Ultra rugged Standard Amplifier Special features - 16 individually configurable inputs (electrically isolated) - Support of 60 V, 10 V, 100 mv, 20 ma or IEPE on all channels - Sample rate: up
More informationSection 1, General information Introduction... 3 Description... 3 Specifications... 4 Wiring connections... 5
LIST OF CONTENTS Section 1, General information Introduction... 3 Description... 3 Specifications... 4 Wiring connections... 5 Section 2, Calibration Calculations... 7 Calibration Procedure... 8 Section
More informationMultichannel amplifier Model 9235/36
wdwd Model 9235/36 2010 burster präzisionsmesstechnik gmbh & co kg All rights reserved Talstrasse 1 5 76593 Gernsbach, Germany Manufacturer: burster praezisionsmesstechnik gmbh & co kg P.O. Box 1432 76587
More informationTB2. Torque reference transducer. Special features. Data sheet
T2 Torque reference transducer pecial features tandard version - ccuracy class 0.03 - ominal (rated) s 100 m, 200 m, 500 m, 1 k m, 2 k m, 3 k m, 5 k m and 10 k m - High permissible oscillation width -
More informationTechnical Documentation
Technical Documentation PROAM-21 Power amplifier for directional valves Grundlage der Funktionsbeschreibung und zur Prüfung der Module (keine Produktdokumentation). AMCA HYDRAULIC FLUID POWER B.V. P.O.BOX
More informationSubmersible pressure sensor For water and wastewater Model LW-1
Replacement product: Model LF-1 Level measurement Submersible pressure sensor For water and wastewater Model LW-1 WIKA data sheet LM 40.03 Applications Level measurement in rivers and lakes Deep well and
More informationMRS ELECTRONIC DATASHEET CAN I/O AND PLC DESCRIPTION TECHNICAL DATA REGULATORY APPROVALS AND TESTING SOFTWARE/PROGRAMMING
DESCRIPTION The versatile CAN I/O PLC with 14 inputs and outputs impresses with its compact design and its operating voltage range of 9 to 30 volts. It provides 8 I/Os that can be configured as inputs
More informationMX1601B R. Data Sheet. Ultra-rugged Standard Amplifier
MX1601B R Ultra-rugged Standard Amplifier Data Sheet Special features 16 individually configurable inputs (electrically isolated) Support of 60 V, 10 V, 100 mv, 20 ma or IEPE on all channels Sample rate:
More informationSubmersible pressure sensor For oils and fuels Model LF-1
Level measurement Submersible pressure sensor For oils and fuels Model LF-1 WIKA data sheet LM 40.04 Applications Level measurement in vessel and storage systems for oils and fuels Overfilling and dry-run
More informationOperating Manual Series Load Pins
Operating Manual 5301 Series Load Pins Safety note Use for intended purpose The force transducers in model series F5301 are intended for the measurement of static and dynamic tensile or compressive force.
More informationForce and strain sensors. Measure. Test. Control. Edition 2016
Force and strain sensors. Measure. Test. Control. Edition 2016 Force and strain sensors by Baumer combine tried and tested technology and sophisticated innovations. 1 Contents. Introduction 1 Product key
More informationRSCC. Data Sheet. Load cells. Special features. Diagram of RSCC load cell. Design: 500 kg to 5 t. Design: 50 kg to 200 kg. B
RSCC Load cells Data Sheet Special features - Load cell with SG measurement system - aximum capacities: 50 kg to 5 t - Hermetically sealed (IP68) - Rust resistant materials - Legal for trade to 3000 divisions,
More informationEnglish. Operating manual. Temperature measuring transducer MU125. Save for later reference. Company / brands of GHM
English Operating manual Temperature measuring transducer MU125 Company / brands of GHM www.ghm-messtechnik.de Save for later reference. Table of contents Page 1. Intended use (areas of application)...
More informationIntelligent Drive Systems, Worldwide Services SK 700E F 3070 GB
Intelligent Drive Systems, Worldwide Services SK 700E Universal application Modular High-Performance Inverter NORDAC SK 700E The NORDAC SK 700E frequency inverter is designed for the power range between
More information2.7. IT system floor-standing distribution cabinet Series -IPS-F/EDS
Dipl.-Ing. W. Bender GmbH & Co. KG Londorfer Str. 65 35305 Grünberg Tel.: 0640 807-0 Fax: 0640 807-259 IT system floor-standing distribution cabinet Series -IPS-F/EDS for supplying power to medical locations
More informationLED Driver Compact dimming
Driver LCAI 35W 350mA 900mA ECO C ECO series Product description Dimmable built-in LED Driver for LED Constant current LED Driver Output current adjustable between 350 900 ma Max. output power 35 W Nominal
More informationMX1615B. Data sheet. Bridge/strain gauge amplifier
MX1615B Bridge/strain gauge amplifier Special features 16 individually configurable inputs Connection of strain gauges in quarter-, half- and full-bridge Bridge excitation : DC or carrier frequency Internal
More informationRadio bus system Radio-universal dimming actuator, 1-gang. 1 Safety instructions. 2 Device components. Order-No. : Operating instructions
Order-No. : 1135 00 Operating instructions 1 Safety instructions Electrical equipment may only be installed and fitted by electrically skilled persons. Failure to observe the instructions may cause damage
More informationPHASED OUT. LED Driver Compact dimming. Driver LCAI 55W 900mA 1750mA ECO SR ECO series
Product description Independent dimmable LED Driver Constant current LED Driver Output current adjustable between 900 1,750 ma Max. output power 55 W Nominal life-time up to 100,000 h 5-year guarantee
More informationBrunata Optuna H Ultrasonic energy meter Type 775 Installation Guide Edition 1.2
Ultrasonic energy meter Type 775 Installation Guide Edition 1.2 UK-QB101575 / 29.05.2012 Brunata a/s is a Danish owned company. We have more than 90 years of experience within developing and producing
More informationExternal brake resistor AX2090-BW5x
Documentation External brake resistor AX2090-BW5x Accessories for Beckhoff servo drive AX5000 Version: Date: 1.3 2018-03-14 Table of content Table of content 1 Foreword... 5 1.1 Notes on the documentation...
More informationPHASED OUT. LED Driver Compact dimming. Driver LCAI 55W 900mA 1750mA ECO C ECO series
Product description Dimmable built-in LED Driver for LED Constant current LED Driver Output current adjustable between 900 1,750 ma Max. output power 55 W Nominal life-time up to 100,000 h 5-year guarantee
More informationPrecipitation Monitor ,
THE WORLD OF WEATHER DATA - THE WORLD OF WEATHER DATA - THE WORLD OF WEATHER DATA Instruction for use 021197/11/09 Precipitation Monitor 5.4103.10.000, 5.4103.10.700 ADOLF THIES GmbH & Co. KG Hauptstraße
More information21 VA (l max 20 5 ms) Connection Cable 1 m, 4 x 0.75 mm 2 Functional data Factory settings Variable Setting Torque Inhibiting torque
echnical data sheet GK4A- Communicative SuperCap rotary actuator with emergency setting function and extended functionalities for adjusting air dampers in ventilation and air-conditioning systems for building
More informationPOSICHRON position sensor in a stainless steel pressure case. Voltage: V, 3 wire Current: ma, 3 wire
Rod Profile with Analog Specifications POSICHRON position sensor in a stainless steel pressure case Protection class IP68 Compression-proof up to 15 bar Measurement range 0... 100 to 0... 5750 mm Absolute
More informationEY-EM 527: Remote I/O module, ecolink527
Product data sheet 94.078 EY-EM 527: Remote I/O module, ecolink527 How energy efficiency is improved Regulation, control, monitoring and optimisation of operational systems, e.g. room automation or HVAC
More informationElectrical devices may only be mounted and connected by electrically skilled persons.
enet Server Order No. : 5301 00 Power adapter enet Server 12 V DC Operating instructions 1 Safety instructions Electrical devices may only be mounted and connected by electrically skilled persons. Serious
More informationSmart Power Relay E I...
Smart Power Relay E-48-8I... Description The Smart Power Relay E-48-8I.- is a remotely controllable electronic load disconnecting relay with three functions in a single unit: l electronic relay l electronic
More informationMV110-8AS. Analog input module 8 channel. User guide
MV110-8AS Analog input module 8 channel User guide MV110-8AS_2016.12_0226_EN All rights reserved Subject to technical changes and misprints Contents 1 Description... 2 1.1 Function... 2 1.2 RS485 network...
More informationHARAX. Contents Technical characteristics circular connector with HARAX rapid termination 50.02
Contents Page Technical characteristics circular connector with rapid termination 50.02 Circular connector M8/M12 with rapid termination... 50.05 Han M12 panel feed-through... 50.08 Han M12 pcb adapter...
More informationOptical laser distance sensors. Dimensioned drawing
Dimensioned drawing en 02-2012/11 50118552 50 100mm 18-30 V DC We reserve the right to make changes DS_ODSL9L100_en_50118552.fm High resolution 0.01mm Reflection-independent distance information Highly
More informationLoad cell. Special features. Data sheet
Z6R Load cell Special features - Nominal (rated) loads: 20 kg 200 kg - Welded-on robust cover - Load cells made of rust-resistant materials - Easy to clean - Same dimensions like Z6F - Verifiable up to
More informationOriginal operating instructions Fail-safe inductive sensor GM504S / / 2010
Original operating instructions Fail-safe inductive sensor GM504S 704070 / 01 06 / 2010 Contents 1 Preliminary note 3 1.1 Explanation of symbols 3 2 Safety instructions 4 2.1 Safety-related requirements
More information1 Introduction. 1.1 General installation information
1 Introduction Nearly every electrical component emits electromagnetic radiation during its operation. This has effects on the quality of the useful signals especially at the communication level, in which
More informationConnection. Input EEx ia IIC. Sense. Fine tuning adjustment. Fine tuning amplifying. Amplifying. Adjustment. Output. Composition
Converter Connection 1-channel EEx ia IIC for 4-wire and 6-wire bridges Analogue output 0/4 ma... 20 ma or 4mA...-12mA Full bridge load cells and strain gauges circuit for resistance bridges up to 17 Ω
More information