What are the steps to use the Axoclamp-2B for whole-cell patch clamp in continuous single-electrode voltage clamp (csevc) mode?

Size: px
Start display at page:

Download "What are the steps to use the Axoclamp-2B for whole-cell patch clamp in continuous single-electrode voltage clamp (csevc) mode?"

Transcription

1 What are the steps to use the Axoclamp-2B for whole-cell patch clamp in continuous single-electrode voltage clamp (csevc) mode? The Axoclamp-2B only performs csevc with ME1. The best overall headstage for this purpose is the HS-2A x0.1lu. For more information regarding the choice of headstages, see the Axoclamp-2B manual Reference section. Initial Amplifier Settings All controls refer to the ME1 controls, or to the controls in the VOLTAGE CLAMP section on the front panel of the AxoClamp. To begin, put the Axoclamp in BRIDGE mode. Turn the following settings on the amplifier to minimum (fully counter-clockwise): GAIN, PHASE LAG, PHASE LAG MULTIPLIER, CAPACITANCE NEUTRALIZATION, ANTI-ALIAS FILTER, and BRIDGE BALANCE. STEP COMMAND: +10.0; DESTINATION: ME1; Switch: EXT. Ensure that the H1 switch beneath the I(nA) meter is correctly positioned depending on the gain of the headstage. For an HS-2A x0.1 LU, the switch should be in the x0.1 position. Select Im on the I DISPLAY SELECT switch. Immediately to the left of the BRIDGE button is a switch for CONT. SEVC and DISCONT. SEVC mode. Put this switch to CONT. SEVC. The DC CURRENT COMMAND +/OFF/- switch should be OFF. Set the amplifier OUTPUT BANDWIDTH to 10 khz (also called the FILTER in the following sections). BNC Connections to the amplifier (assuming a Digidata digitizer) Amplifier Outputs: o Im to Analog In 0 o 10 Vm to Analog In 1 Amplifier Inputs: o EXT VC COMMAND (rear panel) to Analog Out 0 o STEP ACTIVATE (rear panel) to Digital Out 0

2 Model Cell Although the Axoclamp-2B was originally supplied with a choice of model cells, the PATCH-1U model cell is used for the following tests. This is the most appropriate model for whole-cell patch-clamp experiments. PATCH-1U models are available for purchase, and a description can be found in KB article #868. Other model cells may be used for this tutorial, but the details of the response may vary depending on the model cell parameters. Whole-cell voltage clamp procedure 1. Zero the Pipette Offset. Connect the PATCH-1U model cell BATH input to the headstage. Ground the model cell to the yellow socket on the headstage. Turn the INPUT OFFSET potentiometer on the front panel of the AxoClamp until the Vm meter on the amplifier reads 0 mv. 2. Measure the Pipette Resistance. When a TTL (5 V) pulse is applied to the STEP ACTIVATE input, a step pulse is delivered to the electrode as determined by the settings of the STEP COMMAND. Initially the amplifier is in BRIDGE mode, so that with the above settings, 1 na is applied to the pipette on ME1 for the duration of the TTL pulse. Therefore, use Clampex, a similar program, or a step-function generator to apply a short TTL pulse to STEP ACTIVATE. The resulting waveform should be a step in both current and voltage:

3 From this recording the pipette resistance can be calculated. Here, the model cell in the BATH position has a resistance of 10 mv / 1000 pa = 10 MΩ. 2. Set the Holding Position to Zero mv. Turn the HOLDING POSITION until the two RMP BALANCE lights are equally dim. The value on the HOLDING POSITION dial should be close to 5. This will set the holding command of the amplifier to 0 mv when the amplifier is switched to csevc mode. 3. Switch to csevc Mode. Press SEVC to put the amplifier into csevc mode. Turn the DESTINATION switch to VC(mV). Set the TTL pulse to STEP ACTIVATE to run repetitively, for example, at 10 Hz. Turn the pulses back on. The resulting output now resembles the pulses below.

4 (Sampling rate in Clampex = 20 khz, FILTER = 10 khz) Note that even though the STEP COMMAND is set to 10 mv, the resulting step is only ~7 mv. This is because the voltage clamp GAIN setting is set to its lowest value. The voltage clamp is unable to change the pipette potential to the desired command potential because the feedback loop gain is too low. 4. Neutralize the Pipette Capacitance. Increase the CAPACITANCE NEUTRALIZATION until the small overshoot on the current trace is removed, leaving a slightly rounded corner at the top of the current step. 5. Increase the Feedback GAIN. While repeatedly applying the test pulse, first increase the MULTIPLIER switch to 10. Now increase the GAIN slowly. The voltage step will increase, approaching 10 mv, while the current response also grows. Eventually the current becomes noisy, and then oscillates out of control. In order to achieve the optimal setting of GAIN, adjust the PHASE LAG and CAPACITANCE NEUTRALIZATION in turn, while increasing the GAIN and observing the trace. In general, as the trace becomes noisy and begins to show oscillation,

5 increase the PHASE LAG slowly until the trace begins to overshoot; then increase CAPACITANCE NEUTRALIZATION to reduce the overshoot. After adjustment of GAIN, PHASE LAG, and CAPACITANCE NEUTRALIZATION: (GAIN = 3, PHASE LAG = 0.3 x 10, CAP. NEUTRALIZATION adjusted; FILTER = 10 khz, Sampling = 20 khz) GAIN must be set high enough so that the amplifier can maintain the command potential. Therefore, the optimal settings of GAIN, PHASE LAG, and CAPACITANCE NEUTRALIZATION for a specific experiment must be chosen based upon the parameters of the pipette. Larger currents (lower pipette resistance) will require higher GAIN. 6. Form a Seal. In a real experiment, at this point the pipette would be advanced on the cell, a holding potential applied to the pipette, and a GΩ seal formed. As the seal forms, a test pulse is delivered to measure the seal resistance; with Clampex, this is easily done in the Membrane Test (or Seal Test in pclamp 8 and 9). The increasing resistance during seal formation can destabilize the voltage clamp. To avoid this problem, reduce the GAIN to 0.7 before beginning seal formation, and set MULTIPLIER to OFF (thereby turning off PHASE LAG). The voltage clamp may temporarily fail at this low GAIN, but as the seal forms, the current decreases and the clamp will return to the command potential. For example, the PATCH setting on the PATCH-1U model cell models a 10GΩ seal. The figure below shows its response to a 10 mv test pulse.

6 (GAIN = 0.7; MULTIPLIER = OFF; CAPACITANCE NEUTRALIZATION adjusted; FILTER = 10 khz; Sampling = 20 khz; HOLDING POSITION = -50 mv) The seal resistance is too high to accurately measure because of the noise in the current trace. 7. Break into the Cell. At this point the portion of cell membrane sealed to the pipette is ruptured, typically by further application of suction. This establishes the whole-cell voltage clamp. Before breaking into the cell, make sure that the GAIN setting is 0.7 or lower, to avoid oscillations which could damage the cell. For the purposes of this tutorial, switch the PATCH-1U model cell to CELL position. With the test pulse off, turn HOLDING POSITION until Vm is 50 mv. Alternatively, Clampex may be used to set the holding potential. The response to a 10 mv step is shown below.

7 (GAIN = 0.3; MULTIPLIER = OFF; HOLDING POSITION = -50 mv; FILTER = 10 khz, Sampling rate = 20 khz) Once again, since the GAIN is low, the amplifier is unable to keep the holding potential at 50mV, or apply a square voltage command. Set the MULTIPLIER to 10. Now increase the GAIN slowly, while observing the step response. As you see overshoots in the voltage response, increase the PHASE LAG. It may also be necessary to adjust the CAPACITANCE NEUTRALIZATION to achieve the best response, similar to that shown below.

8 (GAIN = 50; PHASE LAG = 0.95x10; CAP. NEUT. Adjusted; FILTER = 10 khz; Sampling = 20 khz) The appropriate final GAIN setting depends on the currents to be measured. Larger currents require higher GAIN so that the membrane potential will remain clamped. For this reason, it is recommended that Vm be recorded together with Im, in order to verify the validity of the voltage clamp. 8. Apply Series Resistance Correction. The current response is somewhat slow because the series resistance of the pipette is not corrected. In csevc mode, the BRIDGE BALANCE control acts to correct series resistance. While applying a repetitive test pulse, increase the BRIDGE BALANCE until the current response becomes sharper and the voltage step shows a slight overshoot. Increasing the series resistance correction too far, however, will result in severe oscillations that can damage a cell. Only increase the series resistance correction (BRIDGE BALANCE) until ringing is observed following the step on the voltage trace; then, reduce series resistance slightly to give a margin of safety.

9 With the model cell, the following traces show the response before (black) and after (red) series resistance correction was applied. (Filter = 10 khz, Sampling = 20 khz) In this case, BRIDGE BALANCE was advanced to 0.7. In csevc mode, with H = x0.1, this indicates that 0.7/0.1 = 7 MΩ of series resistance is corrected. Looking back at the measurement of the pipette in the bath, 7 MΩ / 10 MΩ = 70% of the series resistance was corrected in this case. This is the typical maximum series resistance compensation possible with the Axoclamp-2B. With a real cell, determining the percentage compensation is more complex, because the series resistance is not identical to the pipette resistance measured in the bath. Instead, the true series resistance must be measured in BRIDGE mode after breaking into the cell, and then compared to the corrected series resistance value. For details, see KB article # Choose a FILTER Setting. At this point, the cell is voltage clamped. Depending on the experiment, the FILTER can be set to a lower corner frequency in order to remove some

10 high frequency noise. The appropriate FILTER setting depends on the type of current being measured.

USING THE AXOCLAMP-2B TUTORIALS

USING THE AXOCLAMP-2B TUTORIALS USING THE AXOCLAMP-2B Page 9 USING THE AXOCLAMP-2B TUTORIALS It is recommended that you set up and test the electronics using the model cell supplied, or one of your own design that will mimic the cell

More information

MultiClamp 700B COMPUTER-CONTROLLED MICROELECTRODE AMPLIFIER. Theory and Operation. Part Number Rev D March 2005 Printed in USA

MultiClamp 700B COMPUTER-CONTROLLED MICROELECTRODE AMPLIFIER. Theory and Operation. Part Number Rev D March 2005 Printed in USA MultiClamp 700B COMPUTER-CONTROLLED MICROELECTRODE AMPLIFIER Theory and Operation Part Number 2500-0157 Rev D March 2005 Printed in USA Copyright 2005 Axon Instruments / Molecular Devices Corp. No part

More information

Series Resistance Compensation

Series Resistance Compensation Series Resistance Compensation 1. Patch clamping Patch clamping is a form of voltage clamping, a technique that uses a feedback circuit to set the membrane potential, V m, of a cell to a desired command

More information

Single Electrode Voltage Clamping

Single Electrode Voltage Clamping Plymouth Microelectrode Techniques Workshop Single Electrode Voltage Clamping Alasdair Gibb Research Department of Neuroscience, Physiology & Pharmacology, University College London ciona intestinalis

More information

IR-183A & IR-283A Intracellular Recording Amplifier Instruction Manual

IR-183A & IR-283A Intracellular Recording Amplifier Instruction Manual IR-183A & IR-283A Intracellular Recording Amplifier Instruction Manual 2002 Cygnus Technology, Inc. Cygnus Technology, Inc. P.O.Box 219 Delaware Water Gap, PA 18327 Tel: (570) 424-5701 Fax: (570) 424-5630

More information

BIO 365L Neurobiology Laboratory. Training Exercise 1: Introduction to the Computer Software: DataPro

BIO 365L Neurobiology Laboratory. Training Exercise 1: Introduction to the Computer Software: DataPro BIO 365L Neurobiology Laboratory Training Exercise 1: Introduction to the Computer Software: DataPro 1. Don t Panic. When you run DataPro, you will see a large number of windows, buttons, and boxes. In

More information

Warner Instruments 1125 Dixwell Avenue, Hamden, CT (800) / (203) (203) fax

Warner Instruments 1125 Dixwell Avenue, Hamden, CT (800) / (203) (203) fax PC-505B Manual, Rev. 120716.2 Whole Cell/Patch Clamp Amplifier Model PC-505B 1125 Dixwell Avenue, Hamden, CT 06514 (800) 599-4203 / (203) 776-0664 (203) 776-1278 fax www.warneronline.com PC-505B Manual,

More information

VE-2 Voltage Clamp User's Guide. Copyright Alembic Instruments. All rights reserved.

VE-2 Voltage Clamp User's Guide. Copyright Alembic Instruments. All rights reserved. VE-2 Voltage Clamp User's Guide Copyright 2003. Alembic Instruments. All rights reserved. Copyright 2003. Alembic Instruments. All rights reserved. VE-2 Voltage Clamp User's Guide. Questions? Call (514)

More information

OPERATING INSTRUCTIONS AND SYSTEM DESCRIPTION FOR THE ELC-03XS

OPERATING INSTRUCTIONS AND SYSTEM DESCRIPTION FOR THE ELC-03XS OPERATING INSTRUCTIONS AND SYSTEM DESCRIPTION FOR THE ELC-03XS UNIVERSAL AMPLIFIER with switchable x10 / x100 headstage for EXTRA & INTRACELLULAR RECORDING, SINGLE CELL STIMULATION and ELECTROPORATION

More information

OPERATING INSTRUCTIONS AND SYSTEM DESCRIPTION FOR THE TEC-B-01M VOLTAGE CLAMP MODULE FOR EPMS SYSTEMS. VERSION 1.2 npi 2014

OPERATING INSTRUCTIONS AND SYSTEM DESCRIPTION FOR THE TEC-B-01M VOLTAGE CLAMP MODULE FOR EPMS SYSTEMS. VERSION 1.2 npi 2014 OPERATING INSTRUCTIONS AND SYSTEM DESCRIPTION FOR THE TEC-B-01M VOLTAGE CLAMP MODULE FOR EPMS SYSTEMS VERSION 1.2 npi 2014 npi electronic GmbH, Bauhofring 16, D-71732 Tamm, Germany Phone +49 (0)7141-9730230;

More information

OPERATING INSTRUCTIONS AND SYSTEM DESCRIPTION FOR THE. ISO-STIM 01D STIMULUS ISOLATION UNIT ±100 V / ±10 ma, bipolar output

OPERATING INSTRUCTIONS AND SYSTEM DESCRIPTION FOR THE. ISO-STIM 01D STIMULUS ISOLATION UNIT ±100 V / ±10 ma, bipolar output OPERATING INSTRUCTIONS AND SYSTEM DESCRIPTION FOR THE ISO-STIM 01D STIMULUS ISOLATION UNIT ±100 V / ±10 ma, bipolar output VERSION 4.0 npi 2014 npi electronic GmbH, Bauhofring 16, D-71732 Tamm, Germany

More information

OPERATING INSTRUCTIONS AND SYSTEM DESCRIPTION FOR THE RS-01R

OPERATING INSTRUCTIONS AND SYSTEM DESCRIPTION FOR THE RS-01R OPERATING INSTRUCTIONS AND SYSTEM DESCRIPTION FOR THE RS-01R INSTRUMENT FOR MEASURING SERIES RESISTANCE IN TEVC RECORDINGS VERSION 1.2 npi 2014 npi electronic GmbH, Bauhofring 16, D-71732 Tamm, Germany

More information

OPERATING INSTRUCTIONS AND SYSTEM DESCRIPTION OF THE SEC-05X SINGLE-ELECTRODE CLAMP AMPLIFIER

OPERATING INSTRUCTIONS AND SYSTEM DESCRIPTION OF THE SEC-05X SINGLE-ELECTRODE CLAMP AMPLIFIER OPERATING INSTRUCTIONS AND SYSTEM DESCRIPTION OF THE SEC-05X SINGLE-ELECTRODE CLAMP AMPLIFIER VERSION 2.0 npi 2015 npi electronic GmbH, Bauhofring 16, D-71732 Tamm, Germany Phone +49 (0)7141-9730230; Fax:

More information

INSTRUCTION MANUAL FOR CELL SIMULATOR MODEL 2410

INSTRUCTION MANUAL FOR CELL SIMULATOR MODEL 2410 Cell Simulator INSTRUCTION MANUAL FOR CELL SIMULATOR MODEL 2410 A-M Systems PO Box 850 Carlsborg, WA 98324 U.S.A. 360-683-8300 800-426-1306 FAX: 360-683-3525 http://www.a-msystems.com Disclaimer THIS EQUIPMENT

More information

DC motor control using arduino

DC motor control using arduino DC motor control using arduino 1) Introduction: First we need to differentiate between DC motor and DC generator and where we can use it in this experiment. What is the main different between the DC-motor,

More information

Sampling and Reconstruction

Sampling and Reconstruction Experiment 10 Sampling and Reconstruction In this experiment we shall learn how an analog signal can be sampled in the time domain and then how the same samples can be used to reconstruct the original

More information

OPERATING INSTRUCTIONS AND SYSTEM DESCRIPTION FOR THE ELC-01X AMPLIFIER FOR EXTRACELLULAR RECORDING AND ELECTROPORATION

OPERATING INSTRUCTIONS AND SYSTEM DESCRIPTION FOR THE ELC-01X AMPLIFIER FOR EXTRACELLULAR RECORDING AND ELECTROPORATION OPERATING INSTRUCTIONS AND SYSTEM DESCRIPTION FOR THE ELC-01X AMPLIFIER FOR EXTRACELLULAR RECORDING AND ELECTROPORATION VERSION 1.1 npi 2009 npi electronic GmbH, Hauptstrasse 96, D-71732 Tamm, Germany

More information

Agilent 33220A Function Generator Tutorial

Agilent 33220A Function Generator Tutorial Contents UNIVERSITY OF CALIFORNIA AT BERKELEY College of Engineering Department of Electrical Engineering and Computer Sciences EE105 Lab Experiments Agilent 33220A Function Generator Tutorial 1 Introduction

More information

e-physiology equipment

e-physiology equipment e-physiology equipment Section Page No. Patch Clamp Patch Clamp, PC-505B 152-153 Patch Clamp, PC-501A 154-154 Bilayer Workstation Complete Bilayer Workstation 156 Bilayer Workstation Component list 157

More information

Introduction to Analog Interfacing. ECE/CS 5780/6780: Embedded System Design. Various Op Amps. Ideal Op Amps

Introduction to Analog Interfacing. ECE/CS 5780/6780: Embedded System Design. Various Op Amps. Ideal Op Amps Introduction to Analog Interfacing ECE/CS 5780/6780: Embedded System Design Scott R. Little Lecture 19: Operational Amplifiers Most embedded systems include components that measure and/or control real-world

More information

EPC 10 USB Patch Clamp Amplifier family

EPC 10 USB Patch Clamp Amplifier family Electrophysiology Electrochemistry EPC 10 USB Patch Clamp Amplifier family EPC 10 USB Single EPC 10 USB Double EPC 10 USB Triple EPC 10 USB Quadro EPC 10 USB Double HEKA provides the finest instruments

More information

Neuronal Signals - NBDS 5161 Session 5: Data Acquisition & Analysis. Abdallah HAYAR. Lectures can be downloaded from

Neuronal Signals - NBDS 5161 Session 5: Data Acquisition & Analysis. Abdallah HAYAR. Lectures can be downloaded from Neuronal Signals - NBDS 5161 Session 5: Data Acquisition & Analysis Abdallah HAYAR Lectures can be downloaded from http://hayar.net/nbds5161 Updated Tentative Schedule for Neuronal Signals (NBDS 5161)

More information

Neuroprobe Amplifier

Neuroprobe Amplifier Neuroprobe Amplifier INSTRUCTION MANUAL FOR NEUROPROBE AMPLIFIER MODEL 1600 Serial # Date PO Box 850 Carlsborg, WA 98324 U.S.A. 360-683-8300 800-426-1306 FAX: 360-683-3525 http://www.a-msystems.com Version

More information

Dual operational amplifier

Dual operational amplifier DESCRIPTION The 77 is a pair of high-performance monolithic operational amplifiers constructed on a single silicon chip. High common-mode voltage range and absence of latch-up make the 77 ideal for use

More information

IE-251A Manual, Rev Warner Instruments 1125 Dixwell Avenue, Hamden, CT (800) / (203) (203) fax

IE-251A Manual, Rev Warner Instruments 1125 Dixwell Avenue, Hamden, CT (800) / (203) (203) fax IE-251A Manual, Rev 050816.1 Intracellular Electrometer Model IE-251A 1125 Dixwell Avenue, Hamden, CT 06514 (800) 599-4203 / (203) 776-0664 (203) 776-1278 fax IE-251A Manual, Rev 050816.1 2 Table of Contents

More information

AxoGraph X Data Acquisition Manual

AxoGraph X Data Acquisition Manual AxoGraph X Data Acquisition Manual PLEASE NOTE: For the best figure quality when reading this document onscreen, the zoom setting should be 147 %. If the zoom setting has changed, type 147 % into the zoom

More information

AC LAB ECE-D ecestudy.wordpress.com

AC LAB ECE-D ecestudy.wordpress.com PART B EXPERIMENT NO: 1 AIM: PULSE AMPLITUDE MODULATION (PAM) & DEMODULATION DATE: To study Pulse Amplitude modulation and demodulation process with relevant waveforms. APPARATUS: 1. Pulse amplitude modulation

More information

ELC-03XS. Versatile All-in-One Amplifier

ELC-03XS. Versatile All-in-One Amplifier ELC-03XS Versatile All-in-One Amplifier ELC amplifiers are multi technique systems where researchers can combine traditional patch and microelectrode recording with electrical stimulation, dye application

More information

Technical Service Bulletin DCX AGC Setup Procedure for MODAP Exciters

Technical Service Bulletin DCX AGC Setup Procedure for MODAP Exciters Technical Service Bulletin 030624 DCX AGC Setup Procedure for MODAP Exciters This service bulletin provides the procedure to properly install and align the total system automatic gain control (AGC) upgrade

More information

Lab 1: Basic Lab Equipment and Measurements

Lab 1: Basic Lab Equipment and Measurements Abstract: Lab 1: Basic Lab Equipment and Measurements This lab exercise introduces the basic measurement instruments that will be used throughout the course. These instruments include multimeters, oscilloscopes,

More information

CHAPTER 6. Motor Driver

CHAPTER 6. Motor Driver CHAPTER 6 Motor Driver In this lab, we will construct the circuitry that your robot uses to drive its motors. However, before testing the motor circuit we will begin by making sure that you are able to

More information

250mA HIGH-SPEED BUFFER

250mA HIGH-SPEED BUFFER ma HIGH-SPEED BUFFER FEATURES HIGH OUTPUT CURRENT: ma SLEW RATE: V/µs PIN-SELECTED BANDWIDTH: MHz to MHz LOW QUIESCENT CURRENT:.mA (MHz ) WIDE SUPPLY RANGE: ±. to ±V INTERNAL CURRENT LIMIT THERMAL SHUTDOWN

More information

Warner Instruments 1125 Dixwell Avenue, Hamden, CT (800) / (203) (203) fax

Warner Instruments 1125 Dixwell Avenue, Hamden, CT (800) / (203) (203) fax IE-210 Manual, Rev 061221 Intracellular Electrometer Model IE-210 1125 Dixwell Avenue, Hamden, CT 06514 (800) 599-4203 / (203) 776-0664 (203) 776-1278 fax IE-210 Manual, Rev 061221 2 Table of Contents

More information

FREQUENCY SYNTHESIZERS, SIGNAL GENERATORS

FREQUENCY SYNTHESIZERS, SIGNAL GENERATORS SYNTHESIZED SIGNAL GENERATOR MG3641A/MG3642A 12 khz to 1040/2080 MHz NEW New Anritsu synthesizer technology permits frequency to be set with a resolution of 0.01 Hz across the full frequency range. And

More information

Applications of the LM392 Comparator Op Amp IC

Applications of the LM392 Comparator Op Amp IC Applications of the LM392 Comparator Op Amp IC The LM339 quad comparator and the LM324 op amp are among the most widely used linear ICs today. The combination of low cost, single or dual supply operation

More information

Multiple Instrument Station Module

Multiple Instrument Station Module Multiple Instrument Station Module Digital Storage Oscilloscope Vertical Channels Sampling rate Bandwidth Coupling Input impedance Vertical sensitivity Vertical resolution Max. input voltage Horizontal

More information

DSTS-3B DEPTHSOUNDER TEST SET OPERATOR S MANUAL

DSTS-3B DEPTHSOUNDER TEST SET OPERATOR S MANUAL Page 1 1.0 INTRODUCTION DSTS-3B DEPTHSOUNDER TEST SET OPERATOR S MANUAL The DSTS-3B is a full-featured test set designed for use with all types of echo sounders from small flashers to large commercial

More information

Model Quality in Test and Measurement Since 1949 OUTSTANDING PERFORMANCE DESCRIPTION APPLICATIONS ADDITIONAL FEATURES

Model Quality in Test and Measurement Since 1949 OUTSTANDING PERFORMANCE DESCRIPTION APPLICATIONS ADDITIONAL FEATURES Quality in Test and Measurement Since 1949 DC Source/Calibrators, Tunable Electronic Filters, Wideband Power Amplifiers Precision Phasemeters, Distortion Analyzers Function Generators, RC Oscillators Model

More information

Using Circuits, Signals and Instruments

Using Circuits, Signals and Instruments Using Circuits, Signals and Instruments To be ignorant of one s ignorance is the malady of the ignorant. A. B. Alcott (1799-1888) Some knowledge of electrical and electronic technology is essential for

More information

Setup of Gain Control System (MGC/AGC)

Setup of Gain Control System (MGC/AGC) Setup of Gain Control System (MGC/AGC) Comark Optimum and Ultimate ATSC Transmitters This service bulletin provides the procedure to properly align the manual gain control (MGC) and automatic gain control

More information

M302RM OPERATING MANUAL

M302RM OPERATING MANUAL M302RM OPERATING MANUAL The Model 302RM is a Linear, high voltage, differential amplifier designed to drive a capacitive load such as Conoptics 350, 360, 370 series E.O. modulators. The amplifier is DC

More information

CHAPTER 8 PHOTOMULTIPLIER TUBE MODULES

CHAPTER 8 PHOTOMULTIPLIER TUBE MODULES CHAPTER 8 PHOTOMULTIPLIER TUBE MODULES This chapter describes the structure, usage, and characteristics of photomultiplier tube () modules. These modules consist of a photomultiplier tube, a voltage-divider

More information

Background (What Do Line and Load Transients Tell Us about a Power Supply?)

Background (What Do Line and Load Transients Tell Us about a Power Supply?) Maxim > Design Support > Technical Documents > Application Notes > Power-Supply Circuits > APP 3443 Keywords: line transient, load transient, time domain, frequency domain APPLICATION NOTE 3443 Line and

More information

OPERATIONS MANUAL MULTICHANNEL VOLTAGE-CURRENT CLAMPS SERIAL NO. PHYSIOLOGIC INSTRUMENTS, INC.

OPERATIONS MANUAL MULTICHANNEL VOLTAGE-CURRENT CLAMPS SERIAL NO. PHYSIOLOGIC INSTRUMENTS, INC. OPERATIONS MANUAL MULTICHANNEL VOLTAGE-CURRENT CLAMPS MODEL VCC MC2 MODEL VCC MC6 MODEL VCC MC8 REVISION A REVISION B REVISION A SERIAL NO. PHYSIOLOGIC INSTRUMENTS, INC. 12335 World Trade Drive Suite #6

More information

Description of options, upgrades and accessories for the laser beam stabilization system Compact

Description of options, upgrades and accessories for the laser beam stabilization system Compact Description of options, upgrades and accessories for the laser beam stabilization system Compact The basic configuration of the Compact laser beam stabilization system is fully equipped for stabilization

More information

An Electronic Watt-Watt-Hour Meter

An Electronic Watt-Watt-Hour Meter An Electronic Watt-Watt-Hour Meter The continued emphasis on energy conservation has forced designers to consider the power consumption and efficiency of their products. While equipment for the industrial

More information

iworx Sample Lab Experiment AN-2: Compound Action Potentials

iworx Sample Lab Experiment AN-2: Compound Action Potentials Experiment AN-2: Compound Action Potentials Exercise 1: The Compound Action Potential Aim: To apply a brief stimulus at the proximal end of the nerve and record a compound action potential from the distal

More information

Basic Communication Laboratory Manual. Shimshon Levy&Harael Mualem

Basic Communication Laboratory Manual. Shimshon Levy&Harael Mualem Basic Communication Laboratory Manual Shimshon Levy&Harael Mualem September 2006 CONTENTS 1 The oscilloscope 2 1.1 Objectives... 2 1.2 Prelab... 2 1.3 Background Theory- Analog Oscilloscope...... 3 1.4

More information

DC/DC Converter. Introduction

DC/DC Converter. Introduction DC/DC Converter Introduction This example demonstrates the use of Saber in the design of a DC/DC power converter. The converter is assumed to be a part of a larger system and is modeled at different levels

More information

DLVP A OPERATOR S MANUAL

DLVP A OPERATOR S MANUAL DLVP-50-300-3000A OPERATOR S MANUAL DYNALOAD DIVISION 36 NEWBURGH RD. HACKETTSTOWN, NJ 07840 PHONE (908) 850-5088 FAX (908) 908-0679 TABLE OF CONTENTS INTRODUCTION...3 SPECIFICATIONS...5 MODE SELECTOR

More information

LINEAR IC APPLICATIONS

LINEAR IC APPLICATIONS 1 B.Tech III Year I Semester (R09) Regular & Supplementary Examinations December/January 2013/14 1 (a) Why is R e in an emitter-coupled differential amplifier replaced by a constant current source? (b)

More information

ANALOG TO DIGITAL CONVERTER

ANALOG TO DIGITAL CONVERTER Final Project ANALOG TO DIGITAL CONVERTER As preparation for the laboratory, examine the final circuit diagram at the end of these notes and write a brief plan for the project, including a list of the

More information

Precision, Low Power, Micropower Dual Operational Amplifier OP290

Precision, Low Power, Micropower Dual Operational Amplifier OP290 Precision, Low Power, Micropower Dual Operational Amplifier OP9 FEATURES Single-/dual-supply operation:. V to 3 V, ±.8 V to ±8 V True single-supply operation; input and output voltage Input/output ranges

More information

HP 34401A Specifications 8

HP 34401A Specifications 8 8 HP 34401A Specifications 8 DC Characteristics DC Characteristics Accuracy Specifications ± ( % of reading + % of range ) [ 1 ] Function Range [ 3 ] Test Current or Burden Voltage 24 Hour [ 2 ] 23 C ±

More information

EMT212 Analog Electronic II. Chapter 4. Oscillator

EMT212 Analog Electronic II. Chapter 4. Oscillator EMT Analog Electronic II Chapter 4 Oscillator Objectives Describe the basic concept of an oscillator Discuss the basic principles of operation of an oscillator Analyze the operation of RC, LC and crystal

More information

Servo Tuning. Dr. Rohan Munasinghe Department. of Electronic and Telecommunication Engineering University of Moratuwa. Thanks to Dr.

Servo Tuning. Dr. Rohan Munasinghe Department. of Electronic and Telecommunication Engineering University of Moratuwa. Thanks to Dr. Servo Tuning Dr. Rohan Munasinghe Department. of Electronic and Telecommunication Engineering University of Moratuwa Thanks to Dr. Jacob Tal Overview Closed Loop Motion Control System Brain Brain Muscle

More information

HT101V Reference Manual

HT101V Reference Manual HT101V Reference Manual Overview The HT101V from Ag-Tester is our handheld tester designed to diagnose valve components on agricultural machinery. Valves tested include: 1- Boom Control Valves 2- Servo

More information

Voltage-Clamp and Patch-Clamp Techniques. Hans Reiner Polder, Martin Weskamp, Klaus Linz, Rainer Meyer

Voltage-Clamp and Patch-Clamp Techniques. Hans Reiner Polder, Martin Weskamp, Klaus Linz, Rainer Meyer 3.4 272 Voltage-Clamp and Patch-Clamp Techniques Hans Reiner Polder, Martin Weskamp, Klaus Linz, Rainer Meyer Introduction History of Membrane Potential and Current Recordings in Cardiac Tissue This chapter

More information

OPERATING INSTRUCTIONS AND SYSTEM DESCRIPTION FOR THE EXT-02 B EXTRACELLULAR AMPLIFIER

OPERATING INSTRUCTIONS AND SYSTEM DESCRIPTION FOR THE EXT-02 B EXTRACELLULAR AMPLIFIER OPERATING INSTRUCTIONS AND SYSTEM DESCRIPTION FOR THE EXT-02 B EXTRACELLULAR AMPLIFIER VERSION 1.3 npi 2014 npi electronic GmbH, Bauhofring 16, D-71732 Tamm, Germany Phone +49 (0)7141-9730230; Fax: +49

More information

G320X MANUAL DC BRUSH SERVO MOTOR DRIVE

G320X MANUAL DC BRUSH SERVO MOTOR DRIVE G320X MANUAL DC BRUSH SERVO MOTOR DRIVE Thank you for purchasing the G320X drive. The G320X DC servo drive is warranted to be free of manufacturing defects for 3 years from the date of purchase. Any customer

More information

ECE 404 e-notes...copyright 2008 by Gregory M. Wierzba. All rights reserved...fall 2008.

ECE 404 e-notes...copyright 2008 by Gregory M. Wierzba. All rights reserved...fall 2008. ECE 404L: RF ELECTRONICS LABORATORY DEPARTMENT OF ELECTRICAL AND COMPUTER ENGINEERING MICHIGAN STATE UNIVERSITY I. TITLE: Lab III - AM/FM Radio - AM Radio II. PURPOSE: This lab will focus on soldering

More information

Built-In OVP White LED Step-up Converter in Tiny Package

Built-In OVP White LED Step-up Converter in Tiny Package Built-In White LED Step-up Converter in Tiny Package Description The is a step-up DC/DC converter specifically designed to drive white LEDs with a constant current. The device can drive up to 4 LEDs in

More information

HA-2600, HA Features. 12MHz, High Input Impedance Operational Amplifiers. Applications. Pinouts. Ordering Information

HA-2600, HA Features. 12MHz, High Input Impedance Operational Amplifiers. Applications. Pinouts. Ordering Information HA26, HA26 September 998 File Number 292.3 2MHz, High Input Impedance Operational Amplifiers HA26/26 are internally compensated bipolar operational amplifiers that feature very high input impedance (MΩ,

More information

2 : AC signals, the signal generator and the Oscilloscope

2 : AC signals, the signal generator and the Oscilloscope 2 : AC signals, the signal generator and the Oscilloscope Expected outcomes After conducting this practical, the student should be able to do the following Set up a signal generator to provide a specific

More information

Variable-Gain High Speed Current Amplifier

Variable-Gain High Speed Current Amplifier Features Transimpedance (gain) switchable from 1 x 10 2 to 1 x 10 8 V/A Bandwidth from DC up to 200 MHz Upper cut-off frequency switchable to 1 MHz, 10 MHz or full bandwidth Switchable AC/DC coupling Adjustable

More information

MPS Electrical Test Laboratory, Larsen & Toubro Limited, Mysore Campus, KIADB Industrial Area, Hebbal Hootagalli, Mysore, Karnataka

MPS Electrical Test Laboratory, Larsen & Toubro Limited, Mysore Campus, KIADB Industrial Area, Hebbal Hootagalli, Mysore, Karnataka Last Amended on - Page 1 of 6 SOURCE 1. DC VOLTAGE $ 1mV to 10 mv 0.13 % to 0.014% Using Multifunction Calibrator 10 mv to 1 V 0.014% to 0.0016% By Direct Method 1 v to 10 V 0.0016% to 0.0017% 10 V to

More information

MODEL 9041 EXTENDED SPECIFICATIONS. 25ppm TRANSPORTABLE CALIBRATOR

MODEL 9041 EXTENDED SPECIFICATIONS. 25ppm TRANSPORTABLE CALIBRATOR MODEL 9041 EXTENDED SPECIFICATIONS 25ppm TRANSPORTABLE CALIBRATOR 9041 EXTENDED SPECIFICATIONS General Specifications TRANSMILLE LTD Warm Up Time Double the time since last used up to 20 minutes maximum

More information

SOLICITUD XXVI CONGRESO NACIONAL DE METROLOGIA

SOLICITUD XXVI CONGRESO NACIONAL DE METROLOGIA SOLICITUD XXVI CONGRESO NACIONAL DE METROLOGIA Técnicas para la calibración de medidores analógicos Bill Gaviria Regional Product Manager Direct: +1.425.446.5335 Cell: +1.321.626.7845 Email: bill.gaviria@flukecal.com

More information

EXPERIMENT NUMBER 2 BASIC OSCILLOSCOPE OPERATIONS

EXPERIMENT NUMBER 2 BASIC OSCILLOSCOPE OPERATIONS 1 EXPERIMENT NUMBER 2 BASIC OSCILLOSCOPE OPERATIONS The oscilloscope is the most versatile and most important tool in this lab and is probably the best tool an electrical engineer uses. This outline guides

More information

Notes on OR Data Math Function

Notes on OR Data Math Function A Notes on OR Data Math Function The ORDATA math function can accept as input either unequalized or already equalized data, and produce: RF (input): just a copy of the input waveform. Equalized: If the

More information

SCS Automation and Control Ltd

SCS Automation and Control Ltd 1 SCS Automation and Control Ltd Dead band / Camera Position controller SCS Automation and Control Ltd Automation Centre 156 Stanley Green Road Poole Dorset England BH15 3AH 2 1) INTRODUCTION ATTENTION

More information

Precision, Low Power, Micropower Dual Operational Amplifier OP290

Precision, Low Power, Micropower Dual Operational Amplifier OP290 a FEATURES Single-/Dual-Supply Operation, 1. V to 3 V,. V to 1 V True Single-Supply Operation; Input and Output Voltage Ranges Include Ground Low Supply Current (Per Amplifier), A Max High Output Drive,

More information

Exercise 2: FM Detection With a PLL

Exercise 2: FM Detection With a PLL Phase-Locked Loop Analog Communications Exercise 2: FM Detection With a PLL EXERCISE OBJECTIVE When you have completed this exercise, you will be able to explain how the phase detector s input frequencies

More information

EE 201 Function / Arbitrary Waveform Generator and Oscilloscope Tutorial

EE 201 Function / Arbitrary Waveform Generator and Oscilloscope Tutorial EE 201 Function / Arbitrary Waveform Generator and Oscilloscope Tutorial 1 This is a programmed learning instruction manual. It is written for the Agilent DSO3202A Digital Storage Oscilloscope. The prerequisite

More information

HOME ASSIGNMENT. Figure.Q3

HOME ASSIGNMENT. Figure.Q3 HOME ASSIGNMENT 1. For the differential amplifier circuit shown below in figure.q1, let I=1 ma, V CC =5V, v CM = -2V, R C =3kΩ and β=100. Assume that the BJTs have v BE =0.7 V at i C =1 ma. Find the voltage

More information

EQUIVALENT CIRCUIT DIAGRAM

EQUIVALENT CIRCUIT DIAGRAM MP Power Operational Amplifier MP MP FEATURES LOW COST HIGH VOLTAGE - VOLTS HIGH PUURRENT- 5 AMP PULSE PUT, 5 AMP CONTINUOUS 7 WATT DISSIPATION CAPABILITY V/µS SLEW RATE 5kHz POWER BANDWIDTH APPLICATIONS

More information

SYSTEM-100 PLUG-OUT Software Synthesizer Owner s Manual

SYSTEM-100 PLUG-OUT Software Synthesizer Owner s Manual SYSTEM-100 PLUG-OUT Software Synthesizer Owner s Manual Copyright 2015 ROLAND CORPORATION All rights reserved. No part of this publication may be reproduced in any form without the written permission of

More information

Variable Gain 100 MHz Wideband Voltage Amplifier

Variable Gain 100 MHz Wideband Voltage Amplifier Features 10 to 60 db, Switchable in 10 db Steps Bandwidth DC... 100 MHz, Switchable to 10 MHz Built-In Temperature Compensation for Low Drift of 0.6 µv/k 2.5 nv/ Hz Input Noise Switchable AC/DC-Coupling

More information

Concepts to be Reviewed

Concepts to be Reviewed Introductory Medical Device Prototyping Analog Circuits Part 3 Operational Amplifiers, http://saliterman.umn.edu/ Department of Biomedical Engineering, University of Minnesota Concepts to be Reviewed Operational

More information

POWER AMPLIFIERS 4 QUADRANTS 3x500 VA to 3x1500 VA - THREE-PHASES

POWER AMPLIFIERS 4 QUADRANTS 3x500 VA to 3x1500 VA - THREE-PHASES PERFORMANCES High accuracy High stability Fast transients High inrush current facilities Wide bandwidth Very low distortion Quadrant change without transition Very low output impedance Low noise RS232

More information

Manual itev 90 Intelligent Two-Electrode Voltage Clamp

Manual itev 90 Intelligent Two-Electrode Voltage Clamp Manual 1.04 itev 90 Intelligent Two-Electrode Voltage Clamp HEKA Elektronik Phone +49 (0) 6325 / 95 53-0 Dr. Schulze GmbH Fax +49 (0) 6325 / 95 53-50 Wiesenstrasse 71 Web Site D-67466 Lambrecht/Pfalz Email

More information

Experiment 1: Amplifier Characterization Spring 2019

Experiment 1: Amplifier Characterization Spring 2019 Experiment 1: Amplifier Characterization Spring 2019 Objective: The objective of this experiment is to develop methods for characterizing key properties of operational amplifiers Note: We will be using

More information

Design Implementation Description for the Digital Frequency Oscillator

Design Implementation Description for the Digital Frequency Oscillator Appendix A Design Implementation Description for the Frequency Oscillator A.1 Input Front End The input data front end accepts either analog single ended or differential inputs (figure A-1). The input

More information

APPENDIX D DISCUSSION OF ELECTRONIC INSTRUMENTS

APPENDIX D DISCUSSION OF ELECTRONIC INSTRUMENTS APPENDIX D DISCUSSION OF ELECTRONIC INSTRUMENTS DC POWER SUPPLIES We will discuss these instruments one at a time, starting with the DC power supply. The simplest DC power supplies are batteries which

More information

User Guide. Ring Modulator - Dual Sub Bass - Mixer

User Guide. Ring Modulator - Dual Sub Bass - Mixer sm User Guide Ring Modulator - Dual Sub Bass - Mixer Thank you for purchasing the AJH Synth Ring SM module, which like all AJH Synth Modules, has been designed and handbuilt in the UK from the very highest

More information

EG medlab. Three Lead ECG OEM board. Version Technical Manual. Medlab GmbH Three Lead ECG OEM Module EG01010 User Manual

EG medlab. Three Lead ECG OEM board. Version Technical Manual. Medlab GmbH Three Lead ECG OEM Module EG01010 User Manual Medlab GmbH Three Lead ECG OEM Module EG01010 User Manual medlab Three Lead ECG OEM board EG01010 Technical Manual Copyright Medlab 2008-2016 Version 1.03 1 Version 1.03 28.04.2016 Medlab GmbH Three Lead

More information

Fluke 83V and 87V Digital Multimeters

Fluke 83V and 87V Digital Multimeters Fluke 83V and 87V Digital Multimeters Detailed Specifications For all detailed specifications: is given as ±([% of reading] + [number of least significant digits]) at 18 C to 28 C, with relative humidity

More information

1510A PRECISION SIGNAL SIMULATOR

1510A PRECISION SIGNAL SIMULATOR A worldwide leader in precision measurement solutions Portable signal source for calibrating electronic equipment and machinery monitoring systems. 1510A PRECISION SIGNAL SIMULATOR Voltage Signals Charge

More information

Applications of the LM392 Comparator Op Amp IC

Applications of the LM392 Comparator Op Amp IC Applications of the LM392 Comparator Op Amp IC The LM339 quad comparator and the LM324 op amp are among the most widely used linear ICs today The combination of low cost single or dual supply operation

More information

Technical Service Bulletin DCX AGC Setup Procedure for ADAPT IV Exciters

Technical Service Bulletin DCX AGC Setup Procedure for ADAPT IV Exciters Technical Service Bulletin 110427 DCX AGC Setup Procedure for ADAPT IV Exciters This service bulletin provides the procedure to properly install and align the total system automatic gain control (AGC)

More information

Solid State Logic S O U N D V I S I O N

Solid State Logic S O U N D V I S I O N Solid State Logic S O U N D V I S I O N SUPERANALOGUE X - R A C K Super-Analogue Outboard X-Rack Stereo Dynamics User s Guide This documentation package contains the User s Guide for your new X-Rack Stereo

More information

PAiA 4780 Twelve Stage Analog Sequencer Design Analysis Originally published 1974

PAiA 4780 Twelve Stage Analog Sequencer Design Analysis Originally published 1974 PAiA 4780 Twelve Stage Analog Sequencer Design Analysis Originally published 1974 DESIGN ANALYSIS: CLOCK As is shown in the block diagram of the sequencer (fig. 1) and the schematic (fig. 2), the clock

More information

Effects of Initial Conditions in a DRSSTC. Steven Ward. 6/26/09

Effects of Initial Conditions in a DRSSTC. Steven Ward.   6/26/09 Effects of Initial Conditions in a DRSSTC Steven Ward www.stevehv.4hv.org 6/26/09 The DRSSTC is based on the idea that the initial conditions of the tank circuit are that the primary inductor has zero

More information

TIMS: Introduction to the Instrument

TIMS: Introduction to the Instrument TIMS: Introduction to the Instrument Modules: Audio Oscillator, Speech, Adder, Wideband True RMS Meter, Digital Utilities 1 Displaying a Signal on the PicoScope 1. Turn on TIMS. 2. Computer: Start > All

More information

LABORATORY 5 v3 OPERATIONAL AMPLIFIER

LABORATORY 5 v3 OPERATIONAL AMPLIFIER University of California Berkeley Department of Electrical Engineering and Computer Sciences EECS 100, Professor Bernhard Boser LABORATORY 5 v3 OPERATIONAL AMPLIFIER Integrated operational amplifiers opamps

More information

Specifying A D and D A Converters

Specifying A D and D A Converters Specifying A D and D A Converters The specification or selection of analog-to-digital (A D) or digital-to-analog (D A) converters can be a chancey thing unless the specifications are understood by the

More information

The Discussion of this exercise covers the following points: Filtering Aperture distortion

The Discussion of this exercise covers the following points: Filtering Aperture distortion Exercise 3-1 PAM Signals Demodulation EXERCISE OBJECTIVE When you have completed this exercise you will be able to demonstrate the recovery of the original message signal from a PAM signal using the PAM

More information

Model Operating Manual

Model Operating Manual Model 7500 DC to 1MHz Wideband Power Amplifier Operating Manual Copyright 2004. All rights reserved. Contents of this publication may not be reproduced in any form without the written permission of Krohn-Hite

More information

B25A20FAC SERIES BRUSHLESS SERVO AMPLIFIERS Model: B25A20FAC 120VAC Single Supply Operation

B25A20FAC SERIES BRUSHLESS SERVO AMPLIFIERS Model: B25A20FAC 120VAC Single Supply Operation B25A20FAC Series B25A20FAC SERIES BRUSHLESS SERVO AMPLIFIERS Model: B25A20FAC 120VAC Single Supply Operation FEATURES: All connections on front of amplifier Surface-mount technology Small size, low cost,

More information

LM110 LM210 LM310 Voltage Follower

LM110 LM210 LM310 Voltage Follower LM110 LM210 LM310 Voltage Follower General Description The LM110 series are monolithic operational amplifiers internally connected as unity-gain non-inverting amplifiers They use super-gain transistors

More information