2-2 Power Meter Calibration Power Meter Calibration 1 (1 mw, 50 ohm)

Size: px
Start display at page:

Download "2-2 Power Meter Calibration Power Meter Calibration 1 (1 mw, 50 ohm)"

Transcription

1 esearch and evelopment of Calibration Technology - ower eter Calibration --1 ower eter Calibration 1 (1 mw, 50 ohm) Tsutomu UGIYAA, ojiro AAI, ouichi EBATA, Iwao NIHIYAA, and atsumi FUJII NICT performs calibration services for power meters with coaxial sensors in the frequency range from 100 khz to 50 GHz, and for those with waveguide sensors from 50 GHz up to 170 GHz. In order to ensure the accuracy, the simultaneous comparison and substitution method is adopted as the principle of calibration. The calibration system has the expanded of 0.66 % (coverage factor k=) at 100 Hz. In this report, we describe the calibration methods, systems, and uncertainties in the frequency range 100 khz to 110 GHz. 1 Introduction The output levels and modulation methods are established for TV and radio broadcast waves, and various radio waves produced from wireless devices such as mobile phones and radar. For measurement of their output levels, high-frequency power meters (hereinafter, ower eters ) are mainly used. ower eters are also used when calibrating measurement instruments, such as signal generators and spectrum analyzers. ower eters are important measuring instruments used in evaluating characteristics of wireless devices, and in calibration of various measuring instruments, so their calibration is required in order to measure accurately. As of November 016, NICT is calibrating ower eters in frequencies from 100 khz to 170 GHz. For sensors (Type-N 50 Ω, 3.5 mm,.4 mm) with various coaxial connectors as input terminals (hereinafter abbreviated as Coaxial ensors ), NICT is calibrating from 100 khz to 50 GHz. For sensors with waveguides as input terminals (V-Band (50 75 GHz), W-Band ( GHz), -Band ( GHz)) (hereinafter abbreviated as Waveguide ensors), NICT is calibrating up to 170 GHz. This report describes calibration methods for ower eters from 100 khz to 110 GHz frequencies (1 mw or less, Coaxial ensors and Waveguide ensors), and calibration systems and calculation methods. efinitions General ower eters have a structure divided into the indicator part and sensor. Both are connected by a cable (Fig. 1). Equation (1) shows the relationship between the instruction value and incident power in into the sensor. (1) in Here, is the calibration factor, and calibration is seeking the value of this. General ower eters contain a reference signal source (50 Hz, 1 mw), and when using a ower eter, a sensor is first attached to the reference signal source, and the 1 mw value is aligned. Therefore, the calibration factor also includes the precision of this reference signal source, and at NICT, the indicator part and sensor are calibrated as one unit. 3 Calibration methods 1 Basic components of high-frequency ower eters 3.1 imultaneous comparison and substitution method There are various ower eter calibration methods: comparison method, simultaneous comparison method, simultaneous comparison and substitution method, etc. The simultaneous comparison and substitution method has 13

2 esearch and evelopment of Calibration Technology ower plitter EF, ignal ource #1 # #3 T UT substitute imultaneous comparison and substitution method UT ort # T ort #3 advantages: it is not affected by reflection of signal source, is strong against output variation of the signal source, etc. NICT uses the simultaneous comparison and substitution method for calibration of ower eters (1 mw or less, Coaxial ensors and Waveguide ensors). Figure shows a conceptual diagram of the simultaneous comparison and substitution method. The signal from the signal source is connected to ort #1 of the power splitter, and power that goes through the power splitter is distributed into Test ort # and eference ort #3. After that, calibration is done by the steps below. 1) Connect the reference device (EF) to eference ort #3 (do not remove until calibration ends) ) Connect the standard device (T) to Test ort # 3) Adjust the signal source connected to ort #1 so the T measurement value of Test ort # is 1 mw. 4) eek the ratio ) of T measurement ( value vs. EF measurement value at that time 5) Change T connected to Test ort # to evice Under Test (UT) 6) ake the signal source the same output as 3) ( 7) eek the ratio ) of UT measurement value vs. EF measurement value at that time 8) From 4) and 7), use Equation () to determine the calibration factor of UT. [1] () where, is the calibration factor of T used by an upperlevel calibration organization. (3) W ignal Generator (4) Here, mismatch is expressed in the following equation from the equivalent signal source reflection coefficient Γg, the reflection coefficient Γ of T, and the reflection coefficient Γ of UT. 1 g (5) 1 g Γg can be obtained from parameters of the power splitter. 1 g 3 (6) plitter 3 Calibration system 3. Calibration of power meter using coaxial sensor For a Coaxial ensor, it is possible to calibrate ower eters with 1 mw power, 100 khz to 50 GHz frequency [1]. Figure 3 shows a calibration system photo, with a block diagram in Fig. 4. This calibration system uses the simultaneous comparison and substitution method. Output from the signal generator is input via a switch into the power splitter (ort #1), and the power splitter splits it into eference ort #3 and Test ort #. The calibration steps are the steps explained in the steps of the simultaneous comparison and 14 Journal of the National Institute of Information and Communications Technology Vol. 63 No. 1 (016)

3 --1 ower weter CalierCelon 1 (1 mw, 50 ohm) ignal Generator W Control witch #1 ower ensor (EF) #3 ower plitter ower eter (EF) ort # # ort #3 ower plitter ower ensor (T) ower ensor (UT) substitute 4 Calibration system block diagram ower eter (T) ower eter (UT) ignal Generator ource odule irectional Coupler Isolator substitution method. One advantage of this calibration system is it is designed with the calibration system s eference ort #3 and Test ort # facing upward, so each port plane and sensor s connector plane contact equally. oreover, by switching the switches, one can select the power splitter corresponding to the frequency range and the output port (test port or reference port) corresponding to the sensor s connector, and one can thereby calibrate various connectors (Type-N 50 Ω, 3.5 mm,.4 mm) from 100 Hz to 50 GHz. When connecting each connector, to ensure reproducibility, we use a torque wrench, and always tighten to the same torque. 3.3 Calibration of power meter using waveguide sensor In the case of a Waveguide ensor, calibration of the ower eter is possible with 1 mw V-Band, and 0.1 mw W-Band[]. There are different calibration systems for V-Band and for W-Band. Figure 5 shows a photo for the W-Band system, and Fig. 6 shows its block diagram. This system also uses the simultaneous comparison and substitution method for its calibration method. However, the Coaxial ensor s calibration system used a power splitter, but this system uses a directional coupler (degree of coupling: 6 db) instead of a power splitter. Output from the signal generator is multiplied via a multiplier, multiplied 4 times in the case of V-Band, and multiplied 6 times in the case of W-Band, then input into the directional coupler (ort #1). In the directional coupler, the signal is split into each, and in the direction of the traveling waves, two isolators are also connected, and that output terminal is Test ort #. Also, eference ort #3 is a terminal in the direction of the reflected wave of the directional coupler. Like the Coaxial ensor calibration system, an advantage of this calibration system is that it is designed with the calibration system s eference ort #3 and Test ort # ignal Generator 5 Calibration system (W-band) ource odule 4 (V-Band) 6 (W-Band) EF #3 irectional Coupler (6dB) 6 Calibration system (W-band) block diagram T or UT # Isolator 0 db Isolator 0 db facing upward, so each port and Waveguide ensors waveguides contact equally. oreover, the calibration system s waveguides are very delicate, so it has fixtures to secure the Waveguide ensor to prevent shaking. 3.4 Calibration value Calibration is done by the steps described above. and are each measured 100 times, that average is obtained, and Equation () is used to calculate the calibration value. The calibration factor is calculated with as 1, and the effect due to considering as 1 is evaluated as. However, at high frequencies, we can no longer ignore s effect on calibration value, so Γg, Γ and Γ (all of them complex quantities) are used to calculate the value of defined in Equation (5), that result is applied in Equation (), and the corrected calibration factor is calculated. Figure 7 is a graph of values of when UT (.4 mm connector, 1 to 50 GHz) is calibrated in a Coaxial ensor calibration system. As shown in the figure, especially when more than #1 15

4 esearch and evelopment of Calibration Technology TTable 30 GHz, is far different from 1, and we cannot ignore the effect of on calibration value (in Fig. 7, at 50 GHz, it has about 1% effect on calibration factor). 4 Traceability 7 graph 1 JC certification range Calibration and measurement Frequency(GHz) capability (%) (Level of confidence approximately 95%) , 0.03, 0.04, 0.05, 0.06, 0.07, 0.08, 0.09, 0.1, 0.15, 0., 0.5, , 0.4, 0.5, 0.6, 0.7, 0.8, 0.9, 1.0, , 1.5, 1.6, , 3.0, , All ower eter calibrations are traceable at the National etrology Institute of Japan (NIJ), which sets Japan s national standards (secondary standard device). Among those, ower eters (1 mw, Type-N 50 Ω sensor) with frequencies from 10 Hz to 18 GHz can be calibrated by JC (international A compatible) based on IO/IEC1705[3] [5]. JC is the easurement Act traceability system based on Japan s easurement Act, with the National Institute of Technology and Evaluation managing the registration system for calibration laboratories. These registration criteria are whether there is compliance with the easurement Act related regulations and items required by IO/IEC1705. Table 1 shows the frequencies and calibration and measurement capabilities of JC certification. For calibration and measurement capability, it shows the smallest value when UT is calibrated, for when reflection Γe of UT sensors is 0; this is written in the registration certificate during JC registration. 5 Uncertainty 5.1 Coaxial sensor Equation (7) shows the propagation equation for of simultaneous comparison and substitution method[6]. u( s ) is the of calibration used by an upper-level calibration organization, and u(x) shows the standard for x. u C u u u u s where, the T calibration factor, and due to the power splitter, 1, 1 (7) oreover, if the calibration factor is calculated with equal to 1 (corrections are not made using ), then the estimates of,, and in Equation (7) are all 1, so the sensitivity coefficient for each factor of is 1 or -1, which are 1 when squared, so the relative standard is obtained using the following equation. u( ) u( ) u( ) u( ) u( ) s( ) (8) In the case where corrections are not made using, the factors of are (1) Uncertainty of T, () ifference of ambient temperature during UT calibration vs. during T calibration conducted by an upper-level organization, (3) Change over time of T, (4) Number of digits displayed of UT, (5) Number of digits displayed of T, (6) isalignment between UT and power splitter and between T and power splitter, (7) easurement 16 Journal of the National Institute of Information and Communications Technology Vol. 63 No. 1 (016)

5 --1 ower weter CalierCelon 1 (1 mw, 50 ohm) TTable Factor of Uncertainty budget example (100 Hz frequency) (No correction) Uncertainty istribution ivisor tandard ensitivity coefficient Contribution to u(x) c(x) c(x) u(x) Upper-level calibration % Normal Temperature change 0.4% 0.4% Normal Change over time 0 0 Normal s UT resolution Uniform T resolution Uniform ismatch U eproducibility easure 5 times Normal Combined tandard Uncertainty Expanded relative (k= ) Frequency 0.1 GHz, = % variability. (1) uses the value (differs by frequency) shown in the calibration certification of T (normal distribution). () is obtained by actual measurements of the effect on calibration value due to the difference between the temperature shown in the calibration certification of T (3 ± 1ºC), and the ambient temperature during calibration (3 ± ºC) (normal distribution). (3) is obtained from fluctuations in calibration value measured by an upper-level calibration organization over a one-year period (normal distribution). (4) and (5) are determined from the digits read of UT and T during calibration (round to the nearest 4th decimal) (uniform distribution). (6) is calculated from actual measurements of the reflection coefficients of T and UT and the parameter of Test ort # (U distribution). In (7), calibration is repeated n times, and the variability is calculated. However, the surface joining T and Test ort # is taken as a fixed size, while the position of circumference direction of the surface joining UT and Test ort # divided into n pieces (Type-N 50 Ω connector s female core conductor is divided an even number (4 or 6) of times[7], so it is desirable that n is 3 or a higher odd number) is measured repeatedly over the entire circumference (360/n degrees x n times), and the variability is measured (normal distribution). Generally, sensors are designed so the reflection coefficients Γ and Γ are sufficiently small. And from Equation (6), we know that Γg is also sufficiently small, so g 1 and g 1 (9) are true. Therefore, Equation (5) becomes 1 g 1 g g 1 g (10) Therefore, generated when is approximately 1 is expressed by the following equation. u g g (11) Table shows an example (100 Hz frequency) of an budget of calibration, from the factors listed above, in the case where correction due to is not done Also, Equation (7) is used to obtain the in the case where correction due to is done[8]. In this case, of differs from Equation (11); it adds of the measurements Γg, Γ, and Γ: u u 4 u u 4 (1) g Also, from Equation (6), we get: g 1 3 u u u u 3 1 (13) u g Here, u(xx) is the standard when xx is measured. The value of is calculated using measurements of parameters, by a calculation formula of the company that manufactures the Vector Network Analyzer (VNA). As shown in Fig. 7, especially when exceeding 30 GHz, is far different from 1, so in the case of ower eters using Coaxial ensors, NICT currently does correction using at 30 GHz or greater. Table 3 and 4 show examples 17

6 esearch and evelopment of Calibration Technology TTable Factor of 3 Uncertainty budget example (50 GHz frequency) (With correction) Uncertainty istribution ivisor tandard ensitivity coefficient Contribution to u(x) c(x) c(x) u(x) Upper-level calibration % Normal Temperature change 0.4% 0.4% Normal Change over time 0 0 Normal s UT resolution Uniform T resolution Uniform ismatch U eproducibility easure 5 times Normal Combined tandard Uncertainty Expanded relative (k= ) Frequency 50 GHz, = % Factor of TTable 4 Uncertainty budget example (100 GHz frequency) tandard Uncertainty istribution ivisor ensitivity coefficient Contribution to u(x) c(x) c(x) u(x) Upper-level calibration % Normal Temperature change 0.4% 0.4% Normal Change over time 0 0 Normal s UT resolution Uniform T resolution Uniform ismatch U eproducibility easure 4 times Normal Combined tandard Uncertainty Expanded relative (k= ) Frequency 100 GHz, = % of the budget of calibration when is corrected. 5. Waveguide sensor Also in the case of a Waveguide ensor, like in the case of Coaxial ensors, Equation (7) is used to obtain. Also, for Equation (6), in cases where 1 3 (14) We focus on ability to approximate (15) g As shown in Fig. 6, in a calibration system for Waveguide ensors, signals output from the multiplier are input into the directional coupler. The directional coupler s degree of coupling is 6 db (uniform value), so the parameter () between ort #1 and eference ort #3 becomes approximately 6 db. On the other hand, two isolators are inserted between ort #1 and Test ort #, and due to the loss when passing through these, and the loss when passing through the directional coupler, the parameter (1) between ort #1 and Test ort # becomes approximately 6 db 18 Journal of the National Institute of Information and Communications Technology Vol. 63 No. 1 (016)

7 --1 ower weter CalierCelon 1 (1 mw, 50 ohm) Also, considering the parameter between Test ort # and eference ort #3, there is 40 db isolation due to two isolators, and the directionality of the directional coupler used ( 0log 1 ) is 0 db or greater. Therefore, 3 becomes approximately 60 db smaller than the reflection of Test ort # (). Therefore, in the calibration system shown in Fig. 6, the relationship of Equation (14) holds true, so we can use the Equation (15) approximation formula. Thus it is not necessary to use VNA to measure all parameters of the calibration system. o we can obtain Γg by only measuring (possible by -ort VNA) the reflection coefficient of Test ort # (). The factors of are similar to those for calibration of Coaxial ensors, but variability of measurements are evaluated by measuring n times (n is an even number) with the direction faced (waveguide) changed 180 degrees for Test ort # and UT. Table 4 shows an example of an budget of calibration (100 GHz frequency). 8 ame UT calibration results (10 Hz to 18 GHz) 6 Changes over years In order to check the calibration system and validity of calibration results, the same calibration of UT is done each year, and changes in calibration results over the years are evaluated. Figure 8 shows ower eter calibration results from 011 to 014 (Type-N 50 Ω sensors, 10 Hz to 18 GHz frequency). Figure 9 shows their calibration results (W-Band) from 01 to 016. However, frequencies that were not calibrated are not marked. It was decided to evaluate changes over the years by the number of En [9]. The number of En is used in evaluations such as round robin tests. En is expressed by the following equation. E n LAB EF (16) U U LAB EF where, LAB : easurement value from a participating calibration organization EF : easurement value from a reference calibration organization ULAB : Expanded from a participating calibration organization (k = ) UEF : Expanded from a reference calibration organization (k = ) and the evaluation is satisfactory if n, not satisfactory if n. 9 ame UT calibration results (75 to 110 GHz) Evaluations were performed using Equation (16) with the calibration value of the reference year (final year) as EF ( UEF), and each year s calibration value as LAB ( ULAB). The number of Eon is obtained for each frequency, and the highest value over the past four years is obtained. However, the calibration value of T by an upper-level calibration organization differs slightly each year, but only the calibration results are evaluated, so the ratio vs. the reference year (final year s T calibration value/each year s T calibration value) is multiplied by each calibration value, and the T change portion is removed. The evaluation results are that the maximum number of Eon is 0.4 for ower eters using Type-N 50 Ω sensors (10 Hz to 18 GHz frequency), 0.15 Eon for.4 mm sensors (1 to 50 GHz frequency), 0. Eon for V-Band sensors, and 0.11 Eon for W-Band sensors. All are evaluated as satisfactory. 19

8 esearch and evelopment of Calibration Technology As described above, calibration results over the past four years by this system using the simultaneous comparison and substitution method show that the maximum number of Eon within the evaluation range is less than 1, and stable calibration results are obtained. 7 Conclusion This described a method for calibration of ower eters until 110 GHz frequency, a calibration system, and a method for calculating, that enable very precise calibration using the simultaneous comparison and substitution method. This also showed that stable results were obtained for calibration of ower eters by this calibration system. epresentative expanded uncertainties (coverage factor k = ) were 0.66% for 100 Hz, 3.% for 50 GHz, and 3.7% for 100 GHz. The method of ower eter calibration using the simultaneous comparison and substitution method can also be applied to absolute values of calibration of spectrum analyzers, and is actually being used in calibrations. Appendix. erivation of Equation () In the calibration system with simultaneous comparison and substitution method shown in Fig., when a standard device is connected to Test ort #, and it is expressed using parameters, the following equation is obtained. b1 b b G Gb a1 a a 3 (A.1) a a (A.) a a (A.3) b (A.4) 3 b 3 Here, the matrix expresses characteristics of a three port circuit shown by the dashed lines containing the power splitter. ag is the source power of signal source, ΓG is the reflection coefficient of signal source, Γ is the reflection coefficient of reference device, and Γ is the reflection coefficient of the standard device. From these equations, the power input into the standard device, and the power input into the reference device, are each obtained by in ()(11) b3 ag (A.6) where: 1 11G 1 13 det 1G 1 3 G (A.7) 1 3 det (A.8) ( 1)(11) det 3 3 (A.9) ( )(11) 1 Here, det[a] expresses the matrix formula of matrix A. Now, if we simultaneously measure two incident powers and obtain their ratio, then by (A.5) and (A.6), we obtain: 1 T 33 3 in (1)(11) 1 1 in ()(11) (A.10) T Here, the relationship between in and in was used. and are each the calibration factors of the standard device and reference device. Next, if instead of the standard device, a device under test (UT) is connected to Test ort #, then measuring the input power obtains: 1 UT 33 3 in 1 1 (A.11) in Here, Γ is the reflection coefficient of the device under test (UT), where the relationships of and in were used. is the calibration factor of the device under test. Now, if we calculate the ratio of Equation (A.10) and Equation (A.11), we get (A.1) If we transform the equation, we obtain the following Equation (). T in (1)(11) b ag (A.5) 0 Journal of the National Institute of Information and Communications Technology Vol. 63 No. 1 (016)

9 --1 ower weter CalierCelon 1 (1 mw, 50 ohm) (A.13) ouichi EBATA enior esearcher, Electromagnetic Compatibillity Laboratory, Applied Electromagnetic esearch Institute Calibration of easuring Instruments and Antennas for adio Equipment, geodesy In the equation s derivation process, the power splitter s 1 and, the calibration factor and reflection coefficient Γ of the reference device, and matrix formula shown in Equation (A.7), are all deleted, so there is no need to actually obtain them. Iwao NIHIYAA Electromagnetic Compatibillity Laboratory, Applied Electromagnetic esearch Institute Calibration of easuring Instruments and Antennas for adio Equipment eference 1. FUJII, T. UGIYAA, A. UZUI, T. HINOZUA, and Y. YAANAA, evelopment of a ower eter Calibration ystem for illimeter Wave Frequencies, IEICE Technical eport, ECJ006-57, W , pp.37 41, Oct FUJII, T. UGIYAA, A. UZUI, T. HINOZUA, and Y. YAANAA, evelopment of a ower eter Calibration ystem for the use of V and W bands, IEICE Technical eport, ECJ007-, pp.7 1, April IO/IEC 1705:005, General requirements for the competence of testing and calibration laboratories, IWAA,. FUJII, H. AUZAWA,. OIE,. AAAI, A. UZUI, Y. IYAZAWA, Y. YAANAA, and T. HINOZUA, evelopments of IO/ IEC1705 Calibration ystems in Wireless Communications epartment, Journal of the NICT, vol.53, no.1, pp.43 57, National Institute of Technology and Evaluation, JC Technical Guide for Estimation Examples of Uncertainty: ower ensor, JCG111-06, June 010. (in Japanese) 7 eysight Technologies, icrowave and illimeter-wave coaxial connector, ec (in Japanese) 8. INOHITA,. HIAOA, and. OIYAA, Comparison method for the calibration of microwave power meters and its analysis, NIJ Technical eport, vol.6, no.3, pp , ept JIQ 17043: 011 (IO/IEC 17043: 010) Conformity Assessment-General requirements for proficiency testing. atsumi FUJII, r. Eng. esearch anager, Electromagnetic Compatibility Laboratory, Applied Electromagnetic esearch Institute Calibration of easuring Instruments and Antennas for adio Equipment, Electromagnetic Compatibility Tsutomu UGIYAA enior esearcher, Electromagnetic Compatibillity Laboratory, Applied Electromagnetic esearch Institute Calibration of easuring Instruments and Antennas for adio Equipment ojiro AAI Technical Expert, Electromagnetic Compatibility Laboratory, Applied Electromagnetic esearch Institute Calibration of easuring Instruments and Antennas for adio Equipment 1

2-3 Calibration of Standard Voltage and Current Generator

2-3 Calibration of Standard Voltage and Current Generator 2-3 Calibration of Standard Voltage and Current Generator Katsumi FUJII, Kojiro SAKAI, Tsutomu SUGIYAMA, Kouichi SEBATA, and Iwao NISHIYAMA This paper describes the calibration method of standard voltage

More information

2-2-3 Power Meter Calibration for 10 W (10 W, 50 Ω/75 Ω )

2-2-3 Power Meter Calibration for 10 W (10 W, 50 Ω/75 Ω ) Research and evelopment of Calibration Technology --3 ower eter Calibration for 10 W (10 W, 50 Ω/75 Ω ) ouichi EBATA, ojiro AAI, Tsutomu UGIYAA, Iwao NIHIYAA, and atsumi FUJII NICT performs calibration

More information

RF power measurement in. three-mixer method

RF power measurement in. three-mixer method RF power measurement in D-band using downconverter calibrated by three-mixer method Katsumi Fujii a), Toshihide Tosaka, Kaori Fukunaga, and Yasushi Matsumoto National Institute of Information and Communications

More information

Evaluating VNA post-calibration residual errors using the ripple technique at millimetre wavelengths in rectangular waveguide

Evaluating VNA post-calibration residual errors using the ripple technique at millimetre wavelengths in rectangular waveguide Evaluating VNA post-calibration residual errors using the ripple technique at millimetre wavelengths in rectangular waveguide Abstract C P Eiø and N M Ridler RF & Microwave Guided Wave Metrology Group,

More information

Determination of Uncertainty for Dielectric Properties Determination of Printed Circuit Board Material

Determination of Uncertainty for Dielectric Properties Determination of Printed Circuit Board Material Determination of Uncertainty for Dielectric Properties Determination of Printed Circuit Board Material Marko Kettunen, Kare-Petri Lätti, Janne-Matti Heinola, Juha-Pekka Ström and Pertti Silventoinen Lappeenranta

More information

Contents. CALIBRATION PROCEDURE NI PXIe-5668R 14 GHz and 26.5 GHz Signal Analyzer

Contents. CALIBRATION PROCEDURE NI PXIe-5668R 14 GHz and 26.5 GHz Signal Analyzer CALIBRATION PROCEDURE NI PXIe-5668R 14 GHz and 26.5 GHz Signal Analyzer This document contains the verification procedures for the National Instruments PXIe-5668R (NI 5668R) vector signal analyzer (VSA)

More information

FieldFox Handheld Education Series Part 3: Calibration Techniques for Precise Field Measurements

FieldFox Handheld Education Series Part 3: Calibration Techniques for Precise Field Measurements FieldFox Handheld Education Series Part 3: Calibration Techniques for Precise Field Measurements FieldFox Handheld Education Series Interference Testing Cable and Antenna Measurements Calibration Techniques

More information

N8480 Series Thermocouple Power Sensors DATA SHEET

N8480 Series Thermocouple Power Sensors DATA SHEET N8480 Series Thermocouple Power Sensors DATA SHEET Introduction The Keysight Technologies, Inc. N8480 Series thermocouple power sensors are amongst the most accurate and reliable sensors, plus they include

More information

772D coaxial dual-directional coupler 773D coaxial directional coupler. 775D coaxial dual-directional coupler 776D coaxial dual-directional coupler

772D coaxial dual-directional coupler 773D coaxial directional coupler. 775D coaxial dual-directional coupler 776D coaxial dual-directional coupler 72 772D coaxial dual-directional coupler 773D coaxial directional coupler 775D coaxial dual-directional coupler 776D coaxial dual-directional coupler 777D coaxial dual-directional coupler 778D coaxial

More information

VSWR MEASUREMENT APPLICATION NOTE ANV004.

VSWR MEASUREMENT APPLICATION NOTE ANV004. APPLICATION NOTE ANV004 Bötelkamp 31, D-22529 Hamburg, GERMANY Phone: +49-40 547 544 60 Fax: +49-40 547 544 666 Email: info@valvo.com Introduction: VSWR stands for voltage standing wave ratio. The ratio

More information

Voltage Sensors URV5-Z

Voltage Sensors URV5-Z Data sheet Version 05.00 Voltage Sensors URV5-Z May 2005 Universal voltage measurements from RF to microwaves The voltage sensors of the URV5-Z series are indispensable tools in RF and microwave laboratories,

More information

N432A Thermistor Power Meter DATA SHEET

N432A Thermistor Power Meter DATA SHEET N432A Thermistor Power Meter DATA SHEET Why Keysight s Power Meters and Sensors? Keysight s only power meter that supports thermistor mount with useful enhancements for metrology and calibration lab environments.

More information

3-2 Evaluation of Uncertainty of Horn Antenna Calibration with the Frequency range of 1 GHz to 18 GHz.

3-2 Evaluation of Uncertainty of Horn Antenna Calibration with the Frequency range of 1 GHz to 18 GHz. 3-2 Evaluation of Uncertainty of Horn Antenna Calibration with the Frequency range of 1 GHz to 18 GHz. SAKASAI Makoto, MASUZAWA Hiroshi, FUJII Katsumi, SUZUKI Akira, KOIKE Kunimasa, and YAMANAKA Yukio

More information

This is a preview - click here to buy the full publication

This is a preview - click here to buy the full publication TECHNICAL REPORT IEC TR 63170 Edition 1.0 2018-08 colour inside Measurement procedure for the evaluation of power density related to human exposure to radio frequency fields from wireless communication

More information

Supplementary Information

Supplementary Information Supplementary Information Standards Laboratory Calibration Order 1-999999999-1 Model 8481A Manufacturer Agilent Technologies Description Power Sensor, 10 MHz to 18 GHz, -30 to +20 dbm Serial 2 Options

More information

Contents. CALIBRATION PROCEDURE NI PXIe-5698

Contents. CALIBRATION PROCEDURE NI PXIe-5698 CALIBRATION PROCEDURE NI PXIe-5698 This document contains the verification and adjustment procedures for the National Instruments PXIe-5698 (NI 5698). See ni.com/calibration for more information about

More information

Final Report. Bilateral Comparison on Electric Field Measurements Between TÜBİTAK UME and SASO NMCC GULFMET.EM.RF-S1. UME-EM-D

Final Report. Bilateral Comparison on Electric Field Measurements Between TÜBİTAK UME and SASO NMCC GULFMET.EM.RF-S1. UME-EM-D Final Report Bilateral Comparison on Electric Field Measurements Between TÜBİTAK UME and SASO NMCC GULFMET.EM.RF-S1 UME-EM-D3-2.23.6.a Çağlar ASLAN Abdullah M. ALROBAISH Osman ŞEN (Rev. 0) July 25, 2017

More information

Reducing Uncertainty in EMC Measurements

Reducing Uncertainty in EMC Measurements Reducing Uncertainty in EMC Measurements Uncertainty In general, a standardized EMC test must be developed such that reproducible results are obtained if different parties perform the same test with the

More information

Preliminary Users Manual for the Self Contained Return Loss and Cable Fault Test Set with Amplified Wideband Noise Source Copyright 2001 Bryan K.

Preliminary Users Manual for the Self Contained Return Loss and Cable Fault Test Set with Amplified Wideband Noise Source Copyright 2001 Bryan K. Preliminary Users Manual for the Self Contained Return Loss and Cable Fault Test Set with Amplified Wideband Noise Source Copyright 2001 Bryan K. Blackburn Self Contained Test Set Test Port Regulated 12

More information

Microwave Circuit Design and Measurements Lab. INTRODUCTION TO MICROWAVE MEASUREMENTS: DETECTION OF RF POWER AND STANDING WAVES Lab #2

Microwave Circuit Design and Measurements Lab. INTRODUCTION TO MICROWAVE MEASUREMENTS: DETECTION OF RF POWER AND STANDING WAVES Lab #2 EE 458/558 Microwave Circuit Design and Measurements Lab INTRODUCTION TO MICROWAVE MEASUREMENTS: DETECTION OF RF POWER AND STANDING WAVES Lab #2 The purpose of this lab is to gain a basic understanding

More information

Bill Ham Martin Ogbuokiri. This clause specifies the electrical performance requirements for shielded and unshielded cables.

Bill Ham Martin Ogbuokiri. This clause specifies the electrical performance requirements for shielded and unshielded cables. 098-219r2 Prepared by: Ed Armstrong Zane Daggett Bill Ham Martin Ogbuokiri Date: 07-24-98 Revised: 09-29-98 Revised again: 10-14-98 Revised again: 12-2-98 Revised again: 01-18-99 1. REQUIREMENTS FOR SPI-3

More information

Fast and Accurate Simultaneous Characterization of Signal Generator Source Match and Absolute Power Using X-Parameters.

Fast and Accurate Simultaneous Characterization of Signal Generator Source Match and Absolute Power Using X-Parameters. Fast and Accurate Simultaneous Characterization of Signal Generator Source Match and Absolute Power Using X-Parameters. April 15, 2015 Istanbul, Turkey R&D Principal Engineer, Component Test Division Keysight

More information

Today s modern vector network analyzers

Today s modern vector network analyzers DISTORTION INHERENT TO VNA TEST PORT CABLE ASSEMBLIES Fig. 1 VNA shown with a flexible test port cable assembly on. Today s modern vector network analyzers (VNA) are the product of evolutionary advances

More information

Return Loss Bridge Basics

Return Loss Bridge Basics 1.0 Introduction Return loss bridges have many useful applications for the two-way radio technician These bridges are particularly helpful when used with the tracking generator feature of many service

More information

Ultra High Frequency Measurements

Ultra High Frequency Measurements Ultra High Frequency Measurements Desmond Fraser desmond@rheintech.com 703.689.0368 360 Herndon Parkway Suite 1400 Herndon, VA 20170 IEEE EMC DC / N. VA Chapter 31 January 2012 Overview We ll review Millimeter

More information

SWR/Return Loss Measurements Using System IIA

SWR/Return Loss Measurements Using System IIA THE GLOBAL SOURCE FOR PROVEN TEST SWR/Return Loss Measurements Using System IIA SWR/Return Loss Defined Both SWR and Return Loss are a measure of the divergence of a microwave device from a perfect impedance

More information

Agilent N1911A/N1912A P-Series Power Meters and N1921A/N1922A Wideband Power Sensors. Data sheet

Agilent N1911A/N1912A P-Series Power Meters and N1921A/N1922A Wideband Power Sensors. Data sheet Agilent N1911A/N191A P-Series Power Meters and N191A/N19A Wideband Power Sensors Data sheet Specification Definitions There are two types of product specifications: Warranted specifications are specifications

More information

PNA Family Microwave Network Analyzers (N522x/3x/4xB) CONFIGURATION GUIDE

PNA Family Microwave Network Analyzers (N522x/3x/4xB) CONFIGURATION GUIDE PNA Family Microwave Network Analyzers (N522x/3x/4xB) CONFIGURATION GUIDE Table of Contents PNA Family Network Analyzer Configurations... 05 Test set and power configuration options...05 Hardware options...

More information

PXIe Contents CALIBRATION PROCEDURE

PXIe Contents CALIBRATION PROCEDURE CALIBRATION PROCEDURE PXIe-5632 This document contains the verification and adjustment procedures for the PXIe-5632 Vector Network Analyzer. Refer to ni.com/calibration for more information about calibration

More information

PM Series Microwave Power Calibration System

PM Series Microwave Power Calibration System PM Series Microwave Power Calibration System Supports Sensors from most major manufacturers from 6 khz to 50 GHz Faster than direct compare method Lowest total uncertainty National Metrology Institute

More information

MICROWAVE MICROWAVE TRAINING BENCH COMPONENT SPECIFICATIONS:

MICROWAVE MICROWAVE TRAINING BENCH COMPONENT SPECIFICATIONS: Microwave section consists of Basic Microwave Training Bench, Advance Microwave Training Bench and Microwave Communication Training System. Microwave Training System is used to study all the concepts of

More information

(a) The insertion loss is the average value of the transmission coefficient, S12 (db), in the passband (Figure 1 Label A)

(a) The insertion loss is the average value of the transmission coefficient, S12 (db), in the passband (Figure 1 Label A) Lab 6-1: Microwave Multiport Circuits In this lab you will characterize several different multiport microstrip and coaxial components using a network analyzer. Some, but not all, of these components have

More information

Microwave. Accessories for Microwave Scalar and System Analyzers

Microwave.  Accessories for Microwave Scalar and System Analyzers Microwave Accessories for Microwave Scalar and System Analyzers The following optional accessories are designed for use with the 6200B series of Microwave Test Sets, the 6820 series Scalar Analyzers and

More information

Vector Network Analyzer

Vector Network Analyzer Vector Network Analyzer VNA Basics VNA Roadshow Budapest 17/05/2016 Content Why Users Need VNAs VNA Terminology System Architecture Key Components Basic Measurements Calibration Methods Accuracy and Uncertainty

More information

SCOPE OF ACCREDITATION TO ISO/IEC 17025:2005 & ANSI/NCSL Z

SCOPE OF ACCREDITATION TO ISO/IEC 17025:2005 & ANSI/NCSL Z SCOPE OF ACCREDITATION TO ISO/IEC 17025:2005 & ANSI/NCSL Z540-1-1994 ETS-LINDGREN INC. 1301 Arrow Point Drive Cedar Park, TX 78613 Ron Bethel Phone: 512 531 6400 CALIBRATION Valid To: December 31, 2018

More information

Waveguide Calibration with Copper Mountain Technologies VNA

Waveguide Calibration with Copper Mountain Technologies VNA Clarke & Severn Electronics Ph: +612 9482 1944 BUY NOW www.cseonline.com.au Introduction Waveguide components possess certain advantages over their counterpart devices with co-axial connectors: they can

More information

EE 3324 Electromagnetics Laboratory

EE 3324 Electromagnetics Laboratory EE 3324 Electromagnetics Laboratory Experiment #10 Microstrip Circuits and Measurements 1. Objective The objective of Experiment #8 is to investigate the application of microstrip technology. A precision

More information

FCC 47 CFR PART 15 SUBPART B

FCC 47 CFR PART 15 SUBPART B FCC 47 CFR PART 15 SUBPART B TEST REPORT For DC TO DC CONVERTER Model : D100 Series Issued for MicroPower Direct, LLC 232 Tosca Drive Stoughton, MA02072 U.S.A. Issued by Compliance Certification Services

More information

Agilent PNA Series RF Network Analyzers

Agilent PNA Series RF Network Analyzers Agilent PNA Series RF Network Analyzers Configuration Guide E8356A/E8801A/N3381A E8357A/E8802A/N3382A E8358A/E8803A/N3383A 300 khz to 3 GHz 300 khz to 6 GHz 300 khz to 9 GHz System configuration summary

More information

Amateur Extra Manual Chapter 9.4 Transmission Lines

Amateur Extra Manual Chapter 9.4 Transmission Lines 9.4 TRANSMISSION LINES (page 9-31) WAVELENGTH IN A FEED LINE (page 9-31) VELOCITY OF PROPAGATION (page 9-32) Speed of Wave in a Transmission Line VF = Velocity Factor = Speed of Light in a Vacuum Question

More information

Agilent N4000A, N4001A, N4002A SNS Series Noise Sources 10 MHz to 26.5 GHz

Agilent N4000A, N4001A, N4002A SNS Series Noise Sources 10 MHz to 26.5 GHz Agilent N4000A, N4001A, N4002A SNS Series Noise Sources 10 MHz to 26.5 GHz Technical Overview Advances in Noise Figure Accuracy N4000A Used for low noise figure devices or devices sensitive to mismatch

More information

Agilent 8703B Lightwave Component Analyzer Technical Specifications. 50 MHz to GHz modulation bandwidth

Agilent 8703B Lightwave Component Analyzer Technical Specifications. 50 MHz to GHz modulation bandwidth Agilent 8703B Lightwave Component Analyzer Technical Specifications 50 MHz to 20.05 GHz modulation bandwidth 2 The 8703B lightwave component analyzer is a unique, general-purpose instrument for testing

More information

Contents. CALIBRATION PROCEDURE NI PXIe GHz and 14 GHz RF Vector Signal Analyzer

Contents. CALIBRATION PROCEDURE NI PXIe GHz and 14 GHz RF Vector Signal Analyzer CALIBRATION PROCEDURE NI PXIe-5665 3.6 GHz and 14 GHz RF Vector Signal Analyzer This document contains the verification procedures for the National Instruments PXIe-5665 (NI 5665) RF vector signal analyzer

More information

SCOPE OF ACCREDITATION TO ISO/IEC 17025:2005 ANSI/NCSL Z & ANSI/NCSLI Z

SCOPE OF ACCREDITATION TO ISO/IEC 17025:2005 ANSI/NCSL Z & ANSI/NCSLI Z SCOPE OF ACCREDITATION TO ISO/IEC 17025:2005 ANSI/NCSL Z540-1-1994 & ANSI/NCSLI Z540.3-2006 ANRITSU COMPANY MORGAN HILL CALIBRATION SERVICES 490 Jarvis Drive Morgan Hill, CA 95037 Yeou-Song (Brian) Lee

More information

A Study of Conducted-Emission Stable Source Applied to the EMC US and EU Standards

A Study of Conducted-Emission Stable Source Applied to the EMC US and EU Standards Fourth LACCEI International Latin American and Caribbean Conference for Engineering and Technology (LACCEI 2006) Breaking Frontiers and Barriers in Engineering: Education, Research and Practice, 21-23

More information

Format of S-parameter entries in CMC database

Format of S-parameter entries in CMC database Format of S-parameter entries in CMC database M. Zeier, METAS 10.03.2015 CCEM GT-RF Meeting, BIPM 1 S-parameters Scattering parameters are fundamental quantities in RF & MW metrology. They are measured

More information

Radio Frequency Exposure Test Report

Radio Frequency Exposure Test Report Radio Frequency Exposure EN 62311 January 2008 Assessment of electronic and electrical equipment related to human exposure restrictions for electromagnetic fields (0Hz 300GHz) (IEC 62311:2007, modified)

More information

Broadband Current Probe Series Operation Manual

Broadband Current Probe Series Operation Manual Broadband Current Probe Series Operation Manual 1 TABLE OF CONTENTS WARRANTY 3 INTRODUCTION 4 GENERAL INFORMATION 5 OPERATING INSTRUCTIONS 6 FORMULAS 7 MAINTENANCE 8 2 WARRANTY INFORMATION A.H. Systems

More information

Precision Validation of Radar System Performance in the Field

Precision Validation of Radar System Performance in the Field Precision Validation of Radar System Performance in the Field August 19, 2015 Tom Hoppin Application Specialist Component Test Division Keysight Technologies Keysight Technologies 2015 1 Precision Validation

More information

SCOPE OF ACCREDITATION TO ISO/IEC 17025:2005 ANSI/NCSL Z & ANSI/NCSL Z540.3

SCOPE OF ACCREDITATION TO ISO/IEC 17025:2005 ANSI/NCSL Z & ANSI/NCSL Z540.3 SCOPE OF ACCREDITATION TO ISO/IEC 17025:2005 ANSI/NCSL Z540-1-1994 & ANSI/NCSL Z540.3 KEYSIGHT TECHNOLOGIES, INC. SERVICE CENTERS 1346 Yellowwood Rd Kimballton, IA 51543 Brandt Langer Phone: 712 254 5100

More information

L.S. Compliance, Inc. W66 N220 Commerce Court Cedarburg, WI

L.S. Compliance, Inc. W66 N220 Commerce Court Cedarburg, WI L.S. Compliance, Inc. W66 N220 Commerce Court Cedarburg, WI 53012 262-375-4400 COMPLIANCE TESTING OF: Quartex Synchronization Transmitter Model FM-72 PREPARED FOR: Quartex, Division of Primex, Inc. 965

More information

CALIBRATION TYPES & CONSIDERATIONS

CALIBRATION TYPES & CONSIDERATIONS CALIBRATION TYPES & CONSIDERATIONS 03/12/2018 Introduction One of the most frequently asked questions we receive at Copper Mountain Technologies sales and support departments goes something like this:

More information

Noise Figure Measurement Accuracy: The Y-Factor Method

Noise Figure Measurement Accuracy: The Y-Factor Method APPLICATION NOTE Noise Figure Measurement Accuracy: The Y-Factor Method 1.6 1.4 Measurement uncertainty (db) 1.2 1 0.8 0.6 0.4 0.2 0 DUT noise figure (db) 2 4 6 30 25 20 DUT gain (db) 15 10 5 Table of

More information

Grundlagen der Impedanzmessung

Grundlagen der Impedanzmessung Grundlagen der Impedanzmessung presented by Michael Benzinger Application Engineer - RF & MW Agenda Impedance Measurement Basics Impedance Basics Impedance Dependency Factors Impedance Measurement Methods

More information

Scattered thoughts on Scattering Parameters By Joseph L. Cahak Copyright 2013 Sunshine Design Engineering Services

Scattered thoughts on Scattering Parameters By Joseph L. Cahak Copyright 2013 Sunshine Design Engineering Services Scattered thoughts on Scattering Parameters By Joseph L. Cahak Copyright 2013 Sunshine Design Engineering Services Scattering parameters or S-parameters (aka Spars) are used by RF and microwave engineers

More information

MA24104A. Inline High Power Sensor. True-RMS, 600 MHz to 4 GHz

MA24104A. Inline High Power Sensor. True-RMS, 600 MHz to 4 GHz Product Brochure MA24104A Inline High Power Sensor True-RMS, 600 MHz to 4 GHz A Standalone, Compact, and Highly Accurate Inline High Power Sensor for your RF Power Measurement Needs MA24104A at a Glance

More information

TECHNICAL INFORMATION

TECHNICAL INFORMATION TECHNICAL INFORMATION TECHNOLOGY Y-Junction circulator PORT 1 PORT 2 PORT 3 FIG. 1 The Y-junction circulator uses spinel ferrites or garnet ferrites in the presence of a magnetic bias field, to provide

More information

SCOPE OF ACCREDITATION TO ISO/IEC 17025:2005 AMETEK CO., LTD. NAGOYA OFFICE. (Main Laboratory) Onna, Atsugi-shi, Kanagawa-ken, Japan

SCOPE OF ACCREDITATION TO ISO/IEC 17025:2005 AMETEK CO., LTD. NAGOYA OFFICE. (Main Laboratory) Onna, Atsugi-shi, Kanagawa-ken, Japan SCOPE OF ACCREDITATION TO ISO/IEC 17025:2005 AMETEK CO., LTD. NAGOYA OFFICE (Main Laboratory) 5-1-1 Onna, Atsugi-shi, Kanagawa-ken, 243-0032 Japan (Commercial Office) 1-329 Kifune, Meito-ku Nagoya-shi,

More information

LB679A CW and Pulse (Modulation) USB PowerSensor+ Data Sheet

LB679A CW and Pulse (Modulation) USB PowerSensor+ Data Sheet Key PowerSensor+ Specifications 50 MHz to 20 GHz - 40 dbm to +20 dbm 2.8% Total Error* 1.20:1 VSWR (-21 db Return Loss) * Measuring a well matched DUT (-20 dbm @ 2 GHz) Key PowerSensor+ Capability Test

More information

RF and Microwave Design Solutions. Bob Alman (707)

RF and Microwave Design Solutions. Bob Alman (707) RF and Microwave Design Solutions Bob Alman (707) 529-8481 Bob@AlmanEngineering.com Santa Rosa, CA About Bob Alman Bob Alman acts as an extension of your engineering team by providing guidance, application

More information

Practical Considerations for Radiated Immunities Measurement using ETS-Lindgren EMC Probes

Practical Considerations for Radiated Immunities Measurement using ETS-Lindgren EMC Probes Practical Considerations for Radiated Immunities Measurement using ETS-Lindgren EMC Probes Detectors/Modulated Field ETS-Lindgren EMC probes (HI-6022/6122, HI-6005/6105, and HI-6053/6153) use diode detectors

More information

ACCREDITED LABORATORY. LIBERTY LABS, INC. Kimballton, IA for technical competence in the field of Calibration

ACCREDITED LABORATORY. LIBERTY LABS, INC. Kimballton, IA for technical competence in the field of Calibration THE AMERICAN ASSOCIATION FOR LABORATORY ACCREDITATION ACCREDITED LABORATORY A2LA has accredited LIBERTY LABS, INC. Kimballton, IA for technical competence in the field of Calibration The accreditation

More information

ME7220A. Radar Test System (RTS) Target Simulation & Signal Analysis for Automotive Radar Exceptional Performance at an Affordable Price.

ME7220A. Radar Test System (RTS) Target Simulation & Signal Analysis for Automotive Radar Exceptional Performance at an Affordable Price. ME7220A Test System (RTS) 76 to 77 GHz Target Simulation & Signal Analysis for Automotive Exceptional Performance at an Affordable Price The Challenge The installation of collision warning and Adaptive

More information

Vector Network Analyzer Application note

Vector Network Analyzer Application note Vector Network Analyzer Application note Version 1.0 Vector Network Analyzer Introduction A vector network analyzer is used to measure the performance of circuits or networks such as amplifiers, filters,

More information

LB480A Pulse Profiling USB PowerSensor+ Data Sheet

LB480A Pulse Profiling USB PowerSensor+ Data Sheet Key PowerSensor+ Specifications 100 MHz to 8 GHz (functional to 10 GHz) -60 dbm to +20 dbm 1.95% Total Error* 1.09:1 VSWR (-27 db Return Loss) * Measuring a well matched DUT (-20 dbm @ 1 GHz) Measurement

More information

PRECISION ELECTRONICALLY COMPENSATED CURRENT TRANSFORMER

PRECISION ELECTRONICALLY COMPENSATED CURRENT TRANSFORMER RECIIO ELECTROICALLY COMEATED CURRET TRAFORMER Common Mode Rejecter for Testing Meters with Interconnected Current and Voltage Circuits * Models CMR-I 2130 CMR-I 2230 CMR-I 2330 Version 3.2b * ATETED Content

More information

RF Characterization Report

RF Characterization Report SMA-J-P-H-ST-MT1 Mated with: RF316-01SP1-01BJ1-0305 Description: 50-Ω SMA Board Mount Jack, Mixed Technology Samtec, Inc. 2005 All Rights Reserved Table of Contents Introduction...1 Product Description...1

More information

Verifying the Wideband Input of an AC Measurement Standard

Verifying the Wideband Input of an AC Measurement Standard Abstract erifying the of an AC Measurement Standard Speaker David Deaver Fluke Corporation PO Box 9090, Everett, WA, 98206 Phone: (425) 446-6434 FAX: (425) 446-5649 E-mail: david.deaver@fluke.com Authors:

More information

Optoelectronic Components Testing with a VNA(Vector Network Analyzer) VNA Roadshow Budapest 17/05/2016

Optoelectronic Components Testing with a VNA(Vector Network Analyzer) VNA Roadshow Budapest 17/05/2016 Optoelectronic Components Testing with a VNA(Vector Network Analyzer) VNA Roadshow Budapest 17/05/2016 Content Introduction Photonics & Optoelectronics components Optical Measurements VNA (Vector Network

More information

FieldFox Handheld Education Series Part 7: Precision Validation of Radar System Performance in the Field

FieldFox Handheld Education Series Part 7: Precision Validation of Radar System Performance in the Field FieldFox Handheld Education Series Part 7: Precision Validation of Radar System Performance in the Field FieldFox Handheld Education Series Interference Testing Cable and Antenna Measurements Calibration

More information

Power Dividers and Directional Couplers (7)

Power Dividers and Directional Couplers (7) Microwave Circuits 1 Power Dividers and Directional Couplers (7) The T-Junction Power Divider(7.2) Lossless Divider 1. Lossless 2. Match at the input port. 3. Mismatch at the output ports. 4. No isolation

More information

Monitoring electric and magnetic fields

Monitoring electric and magnetic fields PERSONAL MONITOR RadMan / RadMan XT DE Patent 19,726,138 US Patents 5,955,954 4,634,968 Monitoring electric and magnetic fields ranging from high frequency to microwaves Wide frequency monitoring from1

More information

E-series E9300 Average Power Sensor Specifications

E-series E9300 Average Power Sensor Specifications E-series E9300 Average Power Sensor Specifications The E-series E9300 wide dynamic range, average power sensors are designed for use with the EPM family of power meters. These specifications are valid

More information

Broadband Current Probe Series Operation Manual

Broadband Current Probe Series Operation Manual Broadband Current Probe Series Operation Manual 1 TABLE OF CONTENTS INTRODUCTION 3 GENERAL INFORMATION 4 OPERATING INSTRUCTIONS 5 FORMULAS 6 MAINTENANCE 7 WARRANTY 8 2 INTRODUCTION CURRENT PROBE SPECIFICATIONS

More information

Noise Figure Measurement Accuracy The Y-Factor Method. Application Note 57-2

Noise Figure Measurement Accuracy The Y-Factor Method. Application Note 57-2 Noise Figure Measurement Accuracy The Y-Factor Method Application Note 57-2 Table of contents 1 Introduction...4 2 Noise figure measurement...5 2.1 Fundamentals...5 2.1.1 What is noise figure?...5 2.1.2

More information

Data Sheet. Peak, CW & Average. Power Sensors. Taking performance to a new peak

Data Sheet. Peak, CW & Average. Power Sensors. Taking performance to a new peak Data Sheet Peak, CW & Average Power Sensors Taking performance to a new peak Peak, CW & Average Power Sensors The overall performance of a power meter dependents on the power sensor employed. Boonton has

More information

KULLIYYAH OF ENGINEERING

KULLIYYAH OF ENGINEERING KULLIYYAH OF ENGINEERING DEPARTMENT OF ELECTRICAL & COMPUTER ENGINEERING ANTENNA AND WAVE PROPAGATION LABORATORY (ECE 4103) EXPERIMENT NO 3 RADIATION PATTERN AND GAIN CHARACTERISTICS OF THE DISH (PARABOLIC)

More information

RF and Microwave Power Sensors/Meters Tektronix PSM3000, PSM4000, and PSM5000 Series Data Sheet

RF and Microwave Power Sensors/Meters Tektronix PSM3000, PSM4000, and PSM5000 Series Data Sheet 99 Washington Street Melrose, MA 02176 Phone 781-665-1400 Toll Free 1-800-517-8431 Visit us at www.testequipmentdepot.com RF and Microwave Sensors/Meters Tektronix PSM3000, PSM4000, and PSM5000 Series

More information

DATA SHEET. N1911A/N1912A P-Series Power Meters and N1921A/N1922A Wideband Power Sensors

DATA SHEET. N1911A/N1912A P-Series Power Meters and N1921A/N1922A Wideband Power Sensors DATA SHEET N1911A/N1912A P-Series Power Meters and N1921A/N1922A Wideband Power Sensors LXI Class-C-Compliant Power Meter A P-Series power meter is a LXI Class-C-compliant instrument, developed using LXI

More information

Keysight U8480 Series USB Thermocouple Power Sensors DC/10 MHz to 18/33/50/67/120 GHz DATA SHEET

Keysight U8480 Series USB Thermocouple Power Sensors DC/10 MHz to 18/33/50/67/120 GHz DATA SHEET Keysight U8480 Series USB Thermocouple Power Sensors DC/10 MHz to 18/33/50/67/120 GHz DATA SHEET Improve your power measurement throughput Improve your power measurement throughput with the world s fastest

More information

There is a twenty db improvement in the reflection measurements when the port match errors are removed.

There is a twenty db improvement in the reflection measurements when the port match errors are removed. ABSTRACT Many improvements have occurred in microwave error correction techniques the past few years. The various error sources which degrade calibration accuracy is better understood. Standards have been

More information

Agilent U2000 Series USB Power Sensors. Data Sheet

Agilent U2000 Series USB Power Sensors. Data Sheet Agilent U2000 Series USB Power Sensors Data Sheet Features Perform power measurement without a power meter Frequency range from 9 khz to 24 GHz (sensor dependent) Dynamic range from 60 dbm to +20 dbm Internal

More information

Contents. Software Requirements. CALIBRATION PROCEDURE NI PXIe-5663E

Contents. Software Requirements. CALIBRATION PROCEDURE NI PXIe-5663E CALIBRATION PROCEDURE NI PXIe-5663E This document contains instructions for writing a manual calibration procedure for the NI PXIe-5663E (NI 5663E) RF vector signal analyzer. For more information about

More information

Double-Ridged Waveguide Horn

Double-Ridged Waveguide Horn Model 3106 200 MHz 2 GHz Uniform Gain Power Handling up to 1.6 kw Model 3115 1 GHz 18 GHz Low VSWR Model 3116 18 GHz 40 GHz Quality Construction M O D E L 3 1 0 6 Double-Ridged Waveguide Horn PROVIDING

More information

Department of Electrical and Computer Engineering ECE332. Lab 3: High Frequency Measurements

Department of Electrical and Computer Engineering ECE332. Lab 3: High Frequency Measurements Department of Electrical and Computer Engineering ECE332 Version: 1.3.1 Revised: April 30, 2011 Contents 1 Pre-Lab Assignment 2 2 Introduction 2 2.1 Vector Network Analyzer.............................

More information

Electromagnetic Compatibility Test Report FCC test results of an automatic dog brush, model EUT: Type 1 AC/DC adaptor: SYS W2E

Electromagnetic Compatibility Test Report FCC test results of an automatic dog brush, model EUT: Type 1 AC/DC adaptor: SYS W2E Electromagnetic Compatibility Test Report FCC test results of an automatic dog brush, model EUT: Type 1 AC/DC adaptor: SYS1308-1809-W2E Customer Customer's representative In the capacity of Reference number

More information

LadyBug Technologies, LLC LB5926A True-RMS Power Sensor

LadyBug Technologies, LLC LB5926A True-RMS Power Sensor LadyBug Technologies, LLC LB5926A True-RMS Power Sensor LB5926A-Rev-7 LadyBug Technologies www.ladybug-tech.com Telephone: 707-546-1050 Page 1 LB5926A Data Sheet Key PowerSensor+ TM Specifications Frequency

More information

ENGINEERING COMMITTEE Interface Practices Subcommittee AMERICAN NATIONAL STANDARD

ENGINEERING COMMITTEE Interface Practices Subcommittee AMERICAN NATIONAL STANDARD ENGINEERING COMMITTEE Interface Practices Subcommittee AMERICAN NATIONAL STANDARD ANSI/SCTE 48-2 2008 Test Procedure for Measuring Relative Shielding Properties of Active and Passive Coaxial Cable Devices

More information

Contents. Software Requirements CALIBRATION PROCEDURE NI PXI-5663

Contents. Software Requirements CALIBRATION PROCEDURE NI PXI-5663 CALIBRATION PROCEDURE NI PXI-5663 This document contains instructions for writing a manual calibration procedure for the NI PXI-5663 (NI 5663) RF vector signal analyzer. For more information about calibration,

More information

Accredited calibration of field strength meters

Accredited calibration of field strength meters Accredited calibration of field strength meters Calibration laboratory accredited in accordance with DIN EN ISO/IEC 17025:2005 by the Deutsche Akkreditierungsstelle (DAkkS) Application oriented calibration

More information

Offset Waveguide Transmission Measurements

Offset Waveguide Transmission Measurements NASA Technical Memorandum 110341 Offset Waveguide Transmission Measurements Robin L. Cravey Langley Research Center, Hampton, Virginia May 1997 National Aeronautics and Space Administration Langley Research

More information

S.E. =20log e. t P. t P

S.E. =20log e. t P. t P The effects of gaps introduced into a continuous EMI gasket When properly designed, a surface-mount EMI gasket can provide essentially the same shielding performance as continuous gasketing. THOMAS CLUPPER

More information

SCOPE OF ACCREDITATION TO ISO/IEC 17025:2005 ANSI/NCSL Z & ANSI/NCSLI Z

SCOPE OF ACCREDITATION TO ISO/IEC 17025:2005 ANSI/NCSL Z & ANSI/NCSLI Z SCOPE OF ACCREDITATION TO ISO/IEC 17025:2005 ANSI/NCSL Z540-1-1994 & ANSI/NCSLI Z540.3-2006 ANRITSU COMPANY MORGAN HILL CALIBRATION SERVICES 490 Jarvis Drive Morgan Hill, CA 95037 Yeou-Song (Brian) Lee

More information

VSWR AND ANTENNA SYSTEMS Copyright by Wayne Miller 2018 Revision 4 page 1 of 6

VSWR AND ANTENNA SYSTEMS Copyright by Wayne Miller 2018 Revision 4 page 1 of 6 VSWR AND ANTENNA SYSTEMS Wayne Miller 2018, Revision 4 BACKGROUND In the 40 years of consulting in the RF and Microwave field, I have seen so much misunderstanding about VSWR that it has prompted me to

More information

R&S NRP Power Meter Family Specifications

R&S NRP Power Meter Family Specifications R&S NRP Power Meter Family Specifications year Data Sheet Version 06.00 CONTENTS Definitions... 3 Overview of the R&S NRP power sensors... 4 Specifications in brief of the R&S NRP power sensors... 5 Multipath

More information

LadyBug Technologies, LLC LB5918L True-RMS Power Sensor

LadyBug Technologies, LLC LB5918L True-RMS Power Sensor LadyBug Technologies, LLC LB5918L True-RMS Power Sensor LB5918L-Rev-9 LadyBug Technologies www.ladybug-tech.com Telephone: 707-546-1050 Page 1 LB5918L Data Sheet Key PowerSensor+ TM Specifications Frequency

More information

Noise Figure Measurement Accuracy The Y-Factor Method. Application Note 57-2

Noise Figure Measurement Accuracy The Y-Factor Method. Application Note 57-2 Noise Figure Measurement Accuracy The Y-Factor Method Application Note 57-2 2 Table of contents 1 Introduction...4 2 Noise figure measurement...5 2.1 Fundamentals...5 2.1.1 What is noise figure?...5 2.1.2

More information

Oblique incidence measurement setup for millimeter wave EM absorbers

Oblique incidence measurement setup for millimeter wave EM absorbers Oblique incidence measurement setup for millimeter wave EM absorbers Shinichiro Yamamoto a) and Kenichi Hatakeyama Graduate School of Engineering, University of Hyogo, 2167 Shosha, Himeji-shi, Hyogo 671

More information

Transmission Line Theory and Advanced Measurements in the Field

Transmission Line Theory and Advanced Measurements in the Field Transmission Line Theory and Advanced Measurements in the Field Co-sponsored by Tom Hoppin Application Specialist On Contract to Component Test Division Keysight Technologies Keysight Technologies 2015

More information

Investigation of Two Different Techniques for Accurate Measurements of Sinusoidal Signals

Investigation of Two Different Techniques for Accurate Measurements of Sinusoidal Signals Investigation of Two Different Techniques for Accurate Measurements of Sinusoidal Signals Shereen M. El-Metwally 1, Mamdouh Halawa 2 1 Department of Systems and Biomedical Engineering, Cairo University,

More information