LCM plus. Operating Instructions

Size: px
Start display at page:

Download "LCM plus. Operating Instructions"

Transcription

1 LCM plus Operating Instructions

2 Dear client, Congratulations for the purchase of this high-quality instrument from Dr. Lange. LCM plus is a modern and sophisticated instrument for your lab. LCM plus includes state-of-the-art technology and a very efficient user surface for easy familiarization. Still we recommend to read these operating instructions thoroughly, as they give important information you should know to handle this product adequately. Thank you for your confidence, and enjoy working with your new LCM plus. Yours, BYK-Gardner GmbH Concise Operating Instructions 1) Connect LCM plus and printer LD500, if desired, to the mains. Then plug the printer cable that came with the printer into the RS232-port of LCM plus. Turn LCM plus on by pressing key OK and the printer by toggling the green switch to the right. 2) After zero measurement wait for the display of LCM plus to prompt you to insert the calibration cuvette. 3) Fill a cuvette with distilled water and put it into the cuvette compartment. The display reads e.g.: 4) Take the water cuvette out and put one containing sample liquid into the cuvette compartment. 5) After measuring, the colour values of the sample are indicated in the configured colour system. 6) Now press the -key until the desired colour system is displayed. Zero meas.m. Please wait Pleas ins. cuv. for calibration Calibration Please wait Pleas ins. cuv. for sample Iodine color value 70.5 ÐDIM Conf 11

3 CONTENTS A B C D E Instrument Description A1 General...4 A2 Instrument Design...5 A3 Instrument Set-Up...6 A4 Reference Standards...6 Operation B1 Control And Functional Elements...7 B2 Initial Start-Up of the Instrument...9 B3 Printer Start-Up...9 B4 Battery Mode...10 B5 Turning-Off...11 B6 Photometric Preparation...11 Colour Measurement C1 Calibration...13 C2 Colour Measurement...15 C3 Iodine Number...16 C4 Hazen Colour Number (APHA-Method)...16 C5 Gardner Colour Number...17 C6 Mineral Oil Colour Number ASTM D 1500 And ISO C7 Saybolt Colour Number (ASTM D 156)...19 C8 Configuration...20 Technical Notes D1 Error Messages...23 D2 Maintenande and Test Media Control...24 D3 Lamp Exchange...25 D4 RS232C-Interface of LCM plus...25 D5 Technical Data...26 Annex E1 Spares and Accessories...27 E2 Regulations and Standards...27

4 A Instrument Description A1 General LCM plus is a universal colorimeter for liquids suitable for manifold industrial applications. It is easy to handle and has a sturdy design. But what is really the reason why objective colour measurement of liquids is so important? The human eye is able to perceive about 100 spectral colours in approximately 100 degrees of brightness and saturation, which results in about 1 million perceptible shades. Therefore, hues are difficult to describe by subjective visual comparison. The specific colour of a product is often regarded as a sign of purity and quality, although that is not necessarily true. Clients often interpret deviations from the shade of an earlier production as defective product or supply and reject it preventively. In order to avoid such trouble, objective colour assessment, a uniform colorimetric system and reliable instrument are required for production and quality control of transparent coloured liquids. New Standard EN 1557 EN 1557 (Colorimetric characterization of transparent coloured liquids) now also defines colour measurement of liquids on the basis of the time-tested colorimetric standards of DIN The principle is to measure the X-, Y- and Z-transmissions. The sample is illuminated with a halogen lamp corresponding to standard illuminant C. On this basis, the LCM plus was developped for exact and reproducible measurements of transparent yellow-cast liquids. 4

5 A2 Instrument Design LCM plus measures transmissions of all transparent media. It operates with a halogen lamp producing standard illuminant C and uses the 2 standard observer as defined by DIN-standard A reference beam path compensates measuring values modified by driftings of lamps and temperatures during operation. Therefore, LCM plus provides an excellent long-term constancy and reproducible measuring results. The universal cuvette compartment suitable for 50 x 10mm square cuvettes and 11mm round glass cuvettes permits an optimum adaptation of LCM plus to the measurement tasks. In just one measurement, LCM plus evaluates by to the latest ISO-standards giving e.g. Iodine, Hazen or Gardner values. The basical instrument includes one optional colour number with the possibility of adding-on a random combination of colour numbers later or at once. These colour systems are available: Iodine number, Hazen number, Gardner number, Saybolt number and ASTM D number. To add-on more colour systems later, a program-module-box (bit-box) is supplied. The description of this bit-box is given in chapter D6. Fig. 1: Measuring principle of LCM plus 5

6 A3 Instrument Set-Up Unwrap the instrument carefully and check it for missing parts (delivery note). Please keep the packing box to pack the instrument safely in case it has to be transported again later. LCM plus should be placed on a level desk surface. The instrument comes with a power supply unit and adapter for the socket ( Volt / Hz) and is turned on by either pressing the OK-key or the grey menu key. Do not turn on and off too fast. Wait for 20 seconds before turning on again not to damage the electronic or mechanic system of the instrument. The blue lamp protection cover on the upper left part of the instrument serves to carry the heat off the lamp and must not be covered. As LCM plus is a sophisticated optical instrument, avoid the following at any rate: Do not use the instrument in direct sunlight. Do not allow water, inflammable or metallic objects to enter the instrument. Do not put filled cuvettes or heavy objects on the housing. Do not use the instrument in excessively moist, dusty or vibrating surroundings. Do not open the instrument. There are no parts in the housing for the user to check. If you have any questions of problems, please contact with our offices in Düsseldorf or one of our service stations (see page Error! Bookmark not defined.). A4 Reference Standards In general, an absolutely transparent medium similar to the sample serves as a calibration or reference standard for coloriemtry on the transmission principle. To measure the colour of liquids, distilled water is filled into a cuvette similar to the measuring cuvette and used as a reference standards. In some exceptional cases where the density of the substance to be emasured differs from that of the distilled water, calibration is made against a colourless measuring medium. Calibration assigns the reference standard a transmission of 100% (or Iodine number = 0, Hazen = 0, etc.). As this assignment never changes, 100 no calibration value needs to be set for LCM. Instead, just fill a cuvette with the reference standard and put it into the cuvette compartment to perform calibration. 6

7 B Operation B1 Control And Functional Elements Figures Error! Reference source not found. and Error! Reference source not found. show the control and functional elements of LCM plus. Display Cuvette compartment Lamp cover Cursor keys Fig. 2: Front view OK-key Menu key Cal- and Measure key Battery cover Fig. 3: Back view Printer port Mains socket Bit-Box compartment 7

8 LCM plus is operated with 8 keys. The following table shows the keys and displays in the operating instructions. Cursor keys Calibration and Measure key Menu key OK-Key Display and input area Definition Press left, right, upper or lower cursor key Press blue key (Calibration) Press green key (Measurement) Turn instrument on Quit submenu option Turn instrument on Confirm or call the selected action Display of current values, data, submenus and input areas of the single options. The input area is marked by a window inviting inputs. Representation CAL calibration key (blue) Meas measure key (green) Menu key (grey) OK Aut.shut off: 30 min Menu selection Error messages OK e.g..:ðconfig, another window opens offering more options: It is necessary to acknowledge every error message by pressing the grey menu key, or the instrument stops operating. Switch Off Config Printer Min. limit: 10 8

9 B2 Initial Start-Up of the Instrument For start-up plug-in the mains plug and press the OK-key or grey menu key. The port for the power supply unit is at the rear of the instrument (page Error! Bookmark not defined., Error! Reference source not found., Items 10 and 8). When the instrument is on, the display indicates instrument type and number of EPROM-version while the self-check is running. When self-check and zero measurement are over (approx. 20 sec.) you are prompted to insert a calibration or sample cuvette. LCM plus V1.00 Pleas ins. cuv. for sample LCM plus V : B3 Printer Start-Up The supply voltage for the printer is turned on and off with the green toggle switch 1 on the right of the printer s front. The printer is on when the black key 2 is in neutral position (1) and off when key 2 is at (0). Paper feed is done pressing the ' ' part of key 2 next to the printing unit. B3.1 Connection to LCM plus The printer is connected by plugging the supplied connection cable to the RS232-port on the rear side of LCM plus. B3.2 Change paper Turn the printer on and open the flap 3 behind the printing unit. Put a new reel of paper into the paper container and unroll approx. 10cm of it. Then lead the free beginning of the paper into the slot until perceiving a slight resistance. Press the paper feed key and push the paper on carefully. Hold the key down until the paper passes the tear bar. Finally close the flap again. Fig. 4: Printer LD 500 9

10 B4 Battery Mode LCM plus can be used with power supply unit or rechargeable batteries. For the latter mode, open the cover of the battery compartment and insert the batteries as follows: 1. Insert the battery with the switch to the right into the holder until only one contact remains visible. 2. Shift the battery from the left side over the contacts, passing a slight resistance. 3. A beep indicates that the batteries are placed and ready for operation (only when the instrument is not connected to the mains). Battery Battery switch Fig. 5: Insertion of rechargeable batteries Contact Attention: The contacts must not get into contact with metal or greasy parts. To recharge the batteries, insert them into the compartment and connect LCM plus to the mains with the power supply unit. We recommend a continuous loading time of 14 hours before the first use. To exploit the battery capacity best, charge and discharge it always entirely. The display of LCM plus informs about the current discharge degree of the battery when the instrument is turned on. When the battery is discharged, a beep and a display message are given. The message must be acknowledged with the grey menu key to go on working for a short while. The display message battery empty means that the battery is definitely discharged and needs recharging. 10

11 B5 Turning-Off LCM plus turns off automatically by a timer function. Manual turning-off is possible, too. Manual Turning-Off Press Cursor-Key in the measuring menu and select option Switch Off with the OK-key. Then the instrument turns off. LCM plus Switch Off Config Printer V1.05 Turning-Off Via Timer On delivery, the timer of LCM plus is set to automatical turning-off time of 30 minutes. The automatical turning-off is 14: only active if no measurement was made within the given period and no cuvette is in the compartment. The setting can be modified within the tolerances of 10 to 60 minutes (cp. C8.2). B6 Photometric Preparation The photometric preparation requires utmost care, because careless sample preparation is the major source of trouble, and not the handling of the instrument itself. Cleanliness is of vital importance in photometry. Therefore, clean the glass cuvettes immediately after use. Employ only optically perfect samples for measuring. They must be clean and not show any signs of opaqueness. Air bubbles in the sample or at the cuvette sides influence the measuring result. They can be avoided by filling the liquid slowly into the cuvette. Pasty or solid products should be melted and filled-in without bubbles, if they cannot be measured as they are. B6.1 Sampling And Sample Preparation Take a sample from the product you want to measure in accordance with DIN part 1. Eliminate turbidities by heating or filtration. Heat partly solid samples before measuring in order to dissolve the solid share in the liquid one. The preparation must not cause any chemical changes in the sample. During the measurement there should be no bubbles in the sample. 1 There are several cuvette types available for colour measurement with LCM plus, differing by material (glass, polystyrene or PMMA 2 ) and path length (11mm and 50mm). The program calculates automatically Iodine, Hazen, Gardner, Saybolt and ASTM D 1500-numbers under consideration of the cuvette type employed. 1 2 Acc- to DIN 6162, ed , para. 5 a. 6 Polymethylmethacrylate 11

12 For the 11mm disposable round glass cuvettes, a dry thermostat is available to warm up to 15 cuvettes to 100 C or 148 C. 12

13 C Colour Measurement Whenever the instrument is turned on, a zero measurement is made, taking the empty value (i.e. the pure transmission values without cuvette inserted) of the optical beam path and comparing it with the empty value obtained in the previous zero measurement. If the difference between both empty values is near zero, you can start measuring without calibration. Calibration is only required in case of considerable deviations of the empty value due to e.g. fluctuations of temperature. The calibration remains valid until the next recalibration. LCM plus features a calibration memory for both path lengths employed (50mm square cuvette and 11mm round cuvette). If you have to use two different path lengths, owing to the nature of your samples (e.g. bright samples with Hazen values below 50 or strongly coloured samples with greater Gardner values), it is possible to calibrate both path lengths and then toggle between them. LCM plus uses the corresponding set of calibration data automatically. C1 Calibration LCM plus features a calibration memory for 11mm Dr. Lange round cuvettes and 50mm square cuvettes. To be on the safe side, calibrate LCM plus at least every day. In the calibration, the employed medium is assigned a transmission of 100% over the whole spectral range. Observe the following rules for calibration: The cuvette type selected for calibration must be identical to the one used for measurement, otherwise wrong Iodine, Hazen and Gardner values are given. Use always clean cuvettes and fresh distilled water for calibration. Instrument calibration for colour measurement 1. Turn-on instrument and printer, if desired. After self-check and zero measurement, LCM plus might ask for calibration. If LCM plus is already on, make sure the cuvette compartment is empty and press the blue CAL - key. The display reads: Zero meas.m. Please wait Pleas ins. cuv. for calibration 2. Fill a clean cuvette (material and path length to suit your samples) with distilled water and put it into the measuring compartment. Calibration Please wait 3. LCM plus detects the cuvette type automatically. 4. Now press the blue CAL - key again to trigger calibration. 13

14 5. When the calibration is over and the printer is on, date, time and selected cuvette type are printed, e.g. : Calibration with 11 mm circular cuv. ok! 12: Then LCM plus prompts Please insert sample cuvette. Now you can insert your sample cuvette directly. As soon as you have done so, LICO starts measuring automatically. To calibrate one more path length, remove the cuvette and press the blue CAL -key again. iodine color value Please calibrate carefully to avoid mismeasurements caused by faulty calibration ÐDIM Conf 11 14

15 C2 Colour Measurement The colours of all transparent liquids can be measured, using 50mm square cuvettes or 11mm round glass cuvettes, depending on their transmittances. If required, select another colour system for printer output in the configuration menu (cp. page Error! Bookmark not defined.). Instrument preparation 1. Turn instrument and printer on. 2. Calibrate the instrument as described in chapter. C1. Measuring procedure 1. Fill the sample liquid into a cuvette and insert it into the corresponding compartment. Measurement starts automatically. 2. After measuring, the colour values of the sample will appear on the display iodine color value 2.6 and be printed, e.g. ÐDIM Conf Iodine color value 507 Hazen color value 3. Now press the -key (DIM) until the desired colour system appears on the display. 4. The colour system will be displayed during the measurements to follow. Hazen color value 507 ÐDIM Conf 11 15

16 C3 Iodine Number DIN 6162 defines the Iodine colour number as mg of iodine per 100ml of potassium iodide solution. Colour matching by the Iodine number serves to judge the shade depth of clear liquids like e.g. tensides, plasticizers, resins, oils and fatty acids, whose colours resemble that of an equally layered solution of potassium iodide. For Iodine numbers less than or about 1, the Hazen number according to DIN-ISO 6271 should be preferred. 3 Sampling and preparation are described in chapter B6. A satisfactory precision of the Iodine value is achieved with a 11mm round glass cuvette. Measuring procedure 1. Fill the sample liquid into a cuvette and insert it into the corresponding compartment. Measurement starts automatically. 2. After the measurement, the colour value is indicated on the display. iodine color value 70.5 ÐDIM Conf Now press the -key (DIM) until the Iodine number appears on the display. C4 Hazen Colour Number (APHA-Method) The Hazen colour number (DIN-ISO 6271, also known as "APHA-Method or Platinum-Cobaltscale ) is defined as mg of platinum per 1l of solution. The Hazen parent solution is composed of 1.246g potassium-hexachloroplatinate (IV) and 1.00g cobaltous (II)-chloride dissolved in 100ml of hydrochloric acid and filled up with distilled water until reaching 1000ml. The Hazen colour scale serves to determine the colours of nearly water-clear products. It has a narrower gradation in the light yellow range than the Iodine scale and reaches even water-clear casts. Attention: Samples with Hazen values between 50 and 1000 can be measured with sufficient accuracy in 11mm round cuvettes. Hazen values < 50 should be mesured in 50mm cuvettes, Hazen values < 10 should be measured with the cuvette which was used for calibration, because nearly water-clear products are susceptible to giving faulty measuring results due to cuvette tolerances acc. to DIN 6162, Edition 12.81, para. 1

17 Measuring procedure Hazen color value 1. Fill the sample liquid into a cuvette and insert it into the corresponding compartment. Measurement starts automatically. 2. After the measurement, the colour value is indicated on the display. 20 ÐDIM Conf Now press the -key (DIM) until the Hazen number appears on the display. C5 Gardner Colour Number The Gardner colour number is defined in DIN ISO The lighter Gardner colour standards (1 to 8) are based on potassium-chloroplatinate solutions, colour numbers 9 to 18 on ferric(iii)- chloride, cobaltous(ii)-chloride and hydrochloric acid solutions. LCM plus determines the Gardner number at all cuvette path lengths (in 50mm cuvettes only values between 0 and 4), with 11mm cuvettes providing sufficient accuracy. Measuring procedure 1. Fill the sample liquid into a cuvette and insert it into the corresponding compartment. 2. After the measurement, the colour value is indicated on the display. 3. Now press the -key (DIM) until the Gardner number appears on the display. Gardner color value 6.5 ÐDIM Conf 11 C6 Mineral Oil Colour Number ASTM D 1500 And ISO 2049 The Mineral Oil Colour Scale is used for matching the colours of mineral oil products like lubricating oil, fuel oil, diesel oil and paraffins. The colour scale begins at 0 for water-clear, colourless oils and ends at 8 for very dark brown oils. Visual matching is in steps of 0.5 ( ). LCM displays the results in steps of 0.1. For samples of stronger hue than colour glass 8, colour value D8 is displayed. 17

18 Owing to the intensive colour gradation of the ASTM 4 D 1500 colour number, this is only determined for the 11mm round cuvette. There is no determination of path length 50mm American Society for Testing and Materials

19 Measuring procedure 1. Fill the sample liquid into a cuvette and insert it into the corresponding compartment. 2. After the measurement, the colour value is indicated on the display. 3. Now press the -key (DIM) until the ASTM- number appears on the display. ASTM D ÐDIM Conf 11 C7 Saybolt Colour Number (ASTM D 156) The Saybolt colour scale is used to match colours of refined oil products like fuel and kerosene or also petroleum waxes and pharmaceutical white oils. Its colorimetric characteristics are comparable with those of the Hazen scale (APHA). The Saybolt scale begins at +30 (brightest coloration, corresponding to approx Hazen) and ends at -16 (stongest coloration, corresponding to approx. 350 Hazen). Measurements can be made with cuvettes of 11mm or 50mm path lengths. The greater the path length, the better the measuring accuracy (50mm recommended). 1. Fill the sample liquid into a cuvette and insert it into the corresponding compartment. 2. After the measurement, the colour value is indicated in the configured colour system on the display. 3. Now press the -key (DIM) until the Saybolt number appears on the display. Saybolt color value 24 ÐDIM Conf 50 19

20 C8 Configuration The configuration serves to select the colour systems for the print-out and general instrument settings like e.g. date/time, language, etc. Press cursor-key for configuration. The selection menu shown at the right opens. Move the bar to the menu options with cursor-key or. The option in reverse presentation can be selected by pressing key OK. LCM Switch Offplus Config Printer V1.05 C8.1 Turning-off After selecting option Turn-off and pressing key plus is turned off. OK, LCM 14: Switch LCM Offplus Config Printer V1.05 C8.2 Configure After selecting option Config. and pressing key OK, another menu window opens. The small arrow in the frame window at the right indicates that there are more options to follow. Langue selection These languages are at choice: German (D), English (GB), French (F), Swedish (S), Spanish (E), Dutch (NL), Italian (I), Polish (PL) and Danish (DK). Move the bar to the desired language and select it with key OK. Date and time After selecting option Date and pressing key OK, a window opens to set date and time. Move the blinking cursor to the digit you want to change by cursor-key or. Cursor-keys or increase or decrease the blinking digit by one. Finish the setting with the OK -key. 14: Switch LCM Offplus Config Printerer V1.05 Language Date Timer 14:50 Display LCM Turn-offplus Language select: D Config. GB Printer F S E Turn-off Date LCM Time plus Config Printer V :50 14:

21 Timer Selecting option Timer and pressing key OK opens a window to set timer functions "Aut.shut off" and "delay meas.". Set the desired value with cursor-keys and. Auto shut off time Selecting option "Aut.shut off" and pressing key OK opens a window to set the turn-off time. Move the blinking cursor to the digit you want to change by cursor-key or Cursor-keys or increase or decrease the blinking digit by one. Finish the setting with the OK -key. Measurement delay Selecting option "delay meas." and pressing key OK opens a window to set the Measurement delay. Move the blinking cursor to the digit you want to change by cursor-key or. Cursor-keys or increase or decrease the blinking digit by one. Finish the setting with the OK -key. Display Selecting option "Display and pressing key OK opens a window for display settings of "Contrast" and "Lighting". Make your choice with the cursor-keys and and confirm it with key OK. Printer At this option, you can set more colour systems for printer output. The system selected at display is marked and will be printed in the first line. To add more colour systems, move the bar to the desired system with cursor-keys and and confirm your choice with the OK - key. When the cursor passes the upper or lower margin of the menu, it pages on and shows more options. Here you can add date and time to the measurement. To finish the configuration of the printer output, select "back" and press key OK. LICO 100 Adjust Turn-off timer: Config Aut.shut off Printer delay. meas. 14: LICO 100 V1.00 Aut.shut off: 30 min 14: LICO 100 V1.00 delay. meas. 00 sec 14: LICO 100 Adjust Turn-off display: Contrast... Config V1.00 Lighting Printer 14: Turn-off LICO Iodine number 100 Config Hazen number Printer V1.00 Gardner no. Saybolt no. 14: Turn-off LICO ASTM D Config Date / Time Printer back V : The program returns to the beginning of the configuration. 21

22 22

23 C8.3 Instrument check LCM plus can be checked with a certified test solution set addista -color in compliance with ISO 9000ff. D Technical Notes D1 Error Messages The following display messages indicate errors of operator or instrument. *** Result outside the measuring range. Please remove cuvette! There is a cuvette left in the cuvette compartment. Insert paper Error! Calibration No Signal Lamp? Put new reel of paper into LD 500. The lamp is probably broken. Exchange it as described in chapter Error! Reference source not found.. If one of these error messages appears, please check whether e.g. the printer is connected properly, calibration is correct and the sample cuvette at its place. If you still cannot eliminate the error, contact our service department or field service. 23

24 D2 Maintenande and Test Media Control Even a perfected product like LCM plus requires careful handling. Use a dry cloth or a mild household cleaner for the housing, but never solvents like naphta, acetone or the like. The optical elements should be cleaned by our maintenance staff. A certified set of test solutions is available for test media control according to ISO 9000ff with order no The test set comprises six bottles containing 50ml of test solution each, a quality control certificate according to DIN part 18, and a test protocol for the documentation of reference, actual and tolerance values. LCM plus is tested as described in enclosed test instruction. The Fig. 6: Test solution addista-color instrument should be tested in regular intervals of 1 to 2 weeks, depending on its application and frequency of use. 24

25 D3 Lamp Exchange The lamp compartment is left of the display and covered with a slotted cap on the upper side of the housing. If the lamp is broken, it should be exchanged by Dr. Lange service staff, because we recommend a general instrument check-up after an operating period of 1-2 years. If you insist on doing it yourself, proceed as follows: R e m o v e p o w e r s u p p l y u n i t o r b a t t e r y! Remove the grey cover of the bit-box. Loosen the fixture of the bit-box by pushing the locking clamp 1 and put it apart. Unplug contact 2 from the lamp. Loosen the metal clamps 3 at the upper end one be one from their stay and hinge them away. Take the lamp 4 off and put in a new one with the marking pointing to the left. Attention: Do not touch the glass bulb of the lamp. 1 Fig. 7: Removing the fixture of the bitbox Hinge the metal clamps back to their places and fix them to their stay. Connect the new lamp with the plug-in contact 2 and do the same steps as for dismantling, but in reverse sense. Only when the instrument is mounted completely, turn it on and test it. Fig. 8: Lamp exchange D4 RS232C-Interface of LCM plus LCM plus has a serial interface at its rear side (page Error! Bookmark not defined., Error! Reference source not found.) corresponding to RS 232 C and V24-standard. 25

26 D5 Technical Data Viewing geometry: 0 /180 (transmission) Display/operation: Backlit LC-display (128 x 64 Pixels) Reference standard: Distilled water Spectral range: 8 spectral wavelengths ranged from 400nm to 700nm Wavelength accuracy: ± 2 nm Diffuse light: < 1 % Light source: Krypton lamp Warming-up: approx. 2 minutes Reproducibility: ± 2 Hazen at 50mm path width Documentation: External printer LD 500 Control: Integated microprocessor Measuring time: approx. 10 sec. Program functions: Iodine number Hazen-(APHA) numberl Gardner colour scale ASTM D 1500 number Saybolt-number Cuvettes compartment: 50mm x 10mm square cuvette 11 mm round cuvette with automatic detection Control elements: 8 Foil keys Data interfaces: RS 232 C for LD 500-Printer Mains connection: power supply unit V, 47/63 Hz Dimensions: 260mm x 90mm x 200mm (WxHxD) Weight: approx. 1350g Subject to technical modifications. 26

27 E Annex E1 Spares and Accessories Identification: Quantity/Pcs. Order number: Square plastic cuvettes 10mm * 50mm Square glass cuvette 10mm * 50mm Round glass cuvettes 11mm Cuvette holder for 16 round cuvettes Printer paper 4 reels 6388 Exchange lamp Test solutions addista-color Rechargeable battery 1.4Ah E2 Regulations 5 and Standards EN 1557 Colorimetric characterization of optically clear coloured liquids (products) as X, Y, Z - tristimulus values in transmission, 10.'92 (DIN 53995) DIN 5033 Colour measurement (ASTM E 179, ISO 7724) DIN 6174 Colorimetric evaluation of colour differences (ASTM E 308) DIN 6162 ISO 6271 ISO 4630 ASTM D1500 Determination of iodine colour number Clear liquids;estimation of colour by the platinum-cobalt-scale (Hazen colour number, APHA colour number, also ASTM D , ASTM D , ASTM D , BS 2690:1956.) Estimation of colour of clear liquids by the Gardner colour scale, also ASTM D Standard Test Method for ASTM Color of Petroleum Products (ASTM Color Scale), also DIN/ISO ASTM D156 Standard Test Method for Saybolt Color of Petroleum Products (DIN 51411). ASTM D1925 ASTM D 6045 Yellowness-Index Standard Test Method for Color of Petroleum Products by the Automatic Tristimulus Method 5 Standards can be ordered at: Beuth Verlag GmbH, Burggrafenstraße 6, D Berlin, /

28 Notizen/Notes: 28

29 29

Instrument Description

Instrument Description A Instrument Description A1 General LCM plus is a universal colorimeter for liquids suitable for manifold industrial applications. It is easy to handle and has a sturdy design. But what is really the reason

More information

K13250 LICO 100 Portable Color Measurement System

K13250 LICO 100 Portable Color Measurement System K13250 LICO 100 Portable Color Measurement System Operating Instructions Koehler Instrument Company, Inc. 1595 Sycamore Ave Bohemia, New York 11716 http://www.koehlerinstrument.com e-mail: info@koehlerinstrument.com

More information

Photometric Colorimetry

Photometric Colorimetry Photometric Colorimetry Photometric colorimetry is used in water analytics as well as in industrial production and is usually used to determine quality. In practice, different types of colorimetry have

More information

Lovibond Tintometer PFXi-195/1 & PFXi-195/2

Lovibond Tintometer PFXi-195/1 & PFXi-195/2 Lovibond Tintometer PFXi-195/1 & PFXi-195/2 Consistent and reliable color data Comprehensive and flexible choice of standard color scales Remote upgrade facility for adding scales once in service Allows

More information

UV / VIS Spectrophotometer

UV / VIS Spectrophotometer UV / VIS Spectrophotometer Single Beam Double Beam NANO EMCLAB Instruments GmbH EMCLAB Instruments GmbH Kulturstrasse 55 47055 Duisburg Germany Fon: +49 203 3064042 Fax: +49 203 41504809 E-mail: info@emc-lab.de

More information

User manual. paper moisture meter RH5.1 with sword-sensor

User manual. paper moisture meter RH5.1 with sword-sensor User manual paper moisture meter RH5.1 with sword-sensor version 2.0_en Schaller GmbH 2012 User manual Positioning the instrument Insert the sword-sensor into the stack for only approx. 10 cm, and push

More information

MANUAL BMS SPEC5000 Spectrophotometer

MANUAL BMS SPEC5000 Spectrophotometer MANUAL BMS SPEC5000 Spectrophotometer BMS Art. Nr. 17020 1 PRINCIPLE, APPLICATION AND CHARACTERISTICS 1.1 Principle The SPEC5000 spectrophotometer is a single beam spectrophotometer with a basic structure

More information

User instructions Compound laboratory microscope

User instructions Compound laboratory microscope KERN & Sohn GmbH Ziegelei 1 D-72336 Balingen E-mail: info@kern-sohn.com User instructions Compound laboratory microscope Tel: +49-[0]7433-9933-0 Fax: +49-[0]7433-9933-149 Internet: www.kern-sohn.com KERN

More information

User instructions Compound laboratory microscope

User instructions Compound laboratory microscope KERN & Sohn GmbH Ziegelei 1 D-72336 Balingen E-mail: info@kern-sohn.com User instructions Compound laboratory microscope Tel: +49-[0]7433-9933-0 Fax: +49-[0]7433-9933-149 Internet: www.kern-sohn.com KERN

More information

RQflex 2. Manual English

RQflex 2. Manual English RQflex 2 Manual English Contents 1 Introduction 3 2 Description of the instrument 5 display, keys 6 technical data 8 3 Setting up the instrument 9 insertion of batteries 10 setting the time and date 11

More information

spectro-guide Total Appearance Control color and gloss in one unit Easy to use and handle Solid Color

spectro-guide Total Appearance Control color and gloss in one unit Easy to use and handle Solid Color spectro-guide Total Appearance Control color and gloss in one unit The overall appearance of a product is influenced by color and gloss. A sample of the same color but higher gloss level is visually perceived

More information

Thermo Scientific SPECTRONIC 200 Education

Thermo Scientific SPECTRONIC 200 Education molecular spectroscopy Thermo Scientific SPECTRONIC 200 Education Part of Thermo Fisher Scientific Designed for the Teaching Laboratory Classroom Friendly Sample Compartment Whether you measure in 10 mm

More information

PD-3000UV. INSTRUCTION MANUAL UV-VIS Spectrophotometer OMA IMPORTANT!

PD-3000UV. INSTRUCTION MANUAL UV-VIS Spectrophotometer OMA IMPORTANT! INSTRUCTION MANUAL UV-VIS Spectrophotometer PD-3000UV IMPORTANT! READ THIS OPERATION MANUAL! *Prior to operation, please read this manual carefully and retain it for future reference. Failure to read,

More information

USER MANUAL FOR VISIBLE SPECTROPHOTOMETER

USER MANUAL FOR VISIBLE SPECTROPHOTOMETER USER MANUAL FOR VISIBLE SPECTROPHOTOMETER 1 Table of Contents 1. MAIN USAGES...3 2. WORKING ENVIRONMENT...3 3. MAIN TECHNICAL DATA AND SPECIFICATIONS...4 4. WORKING PRINCIPLE...5 5. OPTICAL PRINCIPLE...6

More information

Photonic Micro Sensors for Color and Spectral Characterization of Transparent Liquids in Laboratories and In-Field

Photonic Micro Sensors for Color and Spectral Characterization of Transparent Liquids in Laboratories and In-Field 10 th Anniversary 10. International Conference on Measurement May 25-28, 2015, Smolenice Castle, Slovakia Dietrich Hofmann Randolf Margull Paul-Gerald Dittrich Daniel Kraus Photonic Micro Sensors for Color

More information

Instruction manual SUN-PM100L. SUN-PM100L Polishing Machine English

Instruction manual SUN-PM100L. SUN-PM100L Polishing Machine English Instruction manual SUN-PM100L SUN-PM100L Polishing Machine English Table: SUN-UM-TL-PM001 Version: A/0 Fiber Optic Solutions Provider Ⅰ Safety Precautions To ensure safe operation and maximize product

More information

Thermo Scientific SPECTRONIC 200

Thermo Scientific SPECTRONIC 200 molecular spectroscopy Thermo Scientific SPECTRONIC 200 Part of Thermo Fisher Scientific The New Standard for Routine Measurements Robust, Multifunction Sample Compartment Whether you measure in 10 mm

More information

Thermo Scientific SPECTRONIC 200 Visible Spectrophotometer. The perfect tool. for routine measurements

Thermo Scientific SPECTRONIC 200 Visible Spectrophotometer. The perfect tool. for routine measurements Thermo Scientific SPECTRONIC 200 Visible Spectrophotometer The perfect tool for routine measurements The Standard for Routine Measurements Thermo Scientific SPECTRONIC spectrophotometers have served as

More information

UV / VIS Spectrophotometer EMCLAB Instruments GmbH

UV / VIS Spectrophotometer EMCLAB Instruments GmbH A Professional Manufacturer UV / VIS Spectrophotometer EMCLAB Instruments GmbH EMCLAB Instruments GmbH Bismarckstrasse 120 47057 Duisburg Germany Phone: +49 203 3064042 Fax: +49 203 3064044 E-mail: info@emc-lab.de

More information

Bedienungsanleitung Instruction Manual. testo 815 Sound Level Meter

Bedienungsanleitung Instruction Manual. testo 815 Sound Level Meter Bedienungsanleitung Instruction Manual testo 815 Sound Level Meter de en 18 General Information General Information Please read this documentation through carefully and familiarise yourself with the operation

More information

User instructions Metallurgical microscope

User instructions Metallurgical microscope KERN & Sohn GmbH Ziegelei 1 D-72336 Balingen E-Mail: info@kern-sohn.com User instructions Metallurgical microscope Tel: +49-[0]7433-9933-0 Fax: +49-[0]7433-9933-149 Internet: www.kern-sohn.com KERN OKM-1

More information

Thermo Scientific SPECTRONIC 200 Visible Spectrophotometer. The perfect. teaching instrument

Thermo Scientific SPECTRONIC 200 Visible Spectrophotometer. The perfect. teaching instrument Thermo Scientific SPECTRONIC 200 Visible Spectrophotometer The perfect teaching instrument Designed for the Teaching Laboratory Thermo Scientific SPECTRONIC spectrophotometers have served as core analytical

More information

Heiland electronic GmbH TD / TD1 / TD2. B&W-Densitometers. USERS MANUAL Version 5

Heiland electronic GmbH TD / TD1 / TD2. B&W-Densitometers. USERS MANUAL Version 5 Heiland electronic GmbH TD / TD1 / TD2 B&W-Densitometers USERS MANUAL Version 5 2 Table of Contents 1. GENERAL INFORMATION...4 2. SAFETY REGULATIONS...5 3. AREA OF APPLICATIONS...5 4. INSTRUMENT DESCRIPTION...6

More information

INDEX OPERATING PRINCIPLE... 3 Power Supply... 4 Replacing the Batteries... 4 GENERAL PHOTOMETER OPERATION... 4 Operating Modes... 4 SYSTEM MODE...

INDEX OPERATING PRINCIPLE... 3 Power Supply... 4 Replacing the Batteries... 4 GENERAL PHOTOMETER OPERATION... 4 Operating Modes... 4 SYSTEM MODE... 9300 PHOTOMETER The YSI 9300 direct-reading photometer is designed to give long and troublefree operation. To ensure the best results, please read these instructions carefully and follow the procedures

More information

INSTRUCTION MANUAL Milwaukee Lab Photometer Mi490 Peroxide value

INSTRUCTION MANUAL Milwaukee Lab Photometer Mi490 Peroxide value www.milwaukeetesters.com INSTRUCTION MANUAL Milwaukee Lab Photometer Mi490 Peroxide value www.milwaukeeinst.com 1 Instruction Manual Mi490 PEROXIDE VALUE Photometer for edible oils TABLE OF CONTENTS FUNCTIONAL

More information

User Manual. Digital Compound Binocular LED Microscope. MicroscopeNet.com

User Manual. Digital Compound Binocular LED Microscope. MicroscopeNet.com User Manual Digital Compound Binocular LED Microscope Model MD82ES10 MicroscopeNet.com Table of Contents i. Caution... 1 ii. Care and Maintenance... 2 1. Components Illustration... 3 2. Installation...

More information

SPECTRONIC Standards User Guide

SPECTRONIC Standards User Guide SPECTRONIC Standards User Guide The information in this publication is provided for reference only. All information contained in this publication is believed to be correct and complete. Thermo Fisher Scientific

More information

User instructions Metallurgical inverted microscope

User instructions Metallurgical inverted microscope KERN & Sohn GmbH Ziegelei 1 D-72336 Balingen E-mail: info@kern-sohn.com Tel: +49-[0]7433-9933-0 Fax: +49-[0]7433-9933-149 Internet: www.kern-sohn.com User instructions Metallurgical inverted microscope

More information

Educational Spectrophotometer Accessory Kit and System OS-8537 and OS-8539

Educational Spectrophotometer Accessory Kit and System OS-8537 and OS-8539 GAIN 1 10 Instruction Manual with Experiment Guide and Teachers Notes 012-06575C *012-06575* Educational Spectrophotometer Accessory Kit and System OS-8537 and OS-8539 100 CI-6604A LIGHT SENSOR POLARIZER

More information

ProfiScale MULTI Multimeter

ProfiScale MULTI Multimeter 1,5 V 9V 200 mv 600 V 200 ma 1/10 A ProfiScale MULTI Multimeter en Operating instructions BURG-WÄCHTER KG Altenhofer Weg 15 58300 Wetter Germany Introduction Want the reassurance of knowing whether current

More information

Durst HL 2506 AF. Durst HL 2506 AF

Durst HL 2506 AF. Durst HL 2506 AF Durst HL 2506 AF Durst HL 3506 AF Professional horizontal enlarger for colour and BW-enlargements from film formats up to 25 x 25 cm (10 x 10 in.) with computer driven Permanent Closed Loop light monitoring

More information

Technical description

Technical description STAHLWILLE Standard Manoskop 721 Service Manoskop 730 List of contents Technical description... 27 ã=important safety points... 30 Operation... 32 Maintenance... 42 Cleaning the Manoskop... 47 Accessories...

More information

Getting started with your NutriOpt On-site Adviser. Better feed choices today

Getting started with your NutriOpt On-site Adviser. Better feed choices today Getting started with your NutriOpt On-site Adviser Better feed choices today This manual explains how to get started with the NutriOpt On-site Adviser. To make the most of this innovative solution, please

More information

impact VC-500LR Monolight INSTRUCTIONS

impact VC-500LR Monolight INSTRUCTIONS impact lighting equipment and accessories VC-500LR Monolight INSTRUCTIONS Congratulations on your purchase of the Impact VC-500LR Monolight. We feel that it will contribute much to your photographic skill

More information

Sarspec, Lda. - Rua Camilo Castelo Branco, 965 PQ Vila Nova de Gaia Phone:

Sarspec, Lda. - Rua Camilo Castelo Branco, 965 PQ Vila Nova de Gaia Phone: 2 3 IMPORTANT SAFETY NOTE: Before operating this device, please read carefully this User Manual and be familiar with its contents prior to using this equipment. To help avoid potential serious injury to

More information

Visit us at

Visit us at testo 816 Sound Level Meter Instruction Manual 99 Washington Street Melrose, MA 02176 Phone 781-665-1400 Toll Free 1-800-517-8431 Visit us at www.testequipmentdepot.com 20 General Information General Information

More information

USB. Part No Wavelength range. Spectral bandwidth 5 nm 4 nm Optical system

USB. Part No Wavelength range. Spectral bandwidth 5 nm 4 nm Optical system Visible range spectrophotometers V-1100 and VR-2000 V-1100 MODEL WITH MANUAL WAVELENGTH SETTINGS AND AUTOMATIC BLANK. VR-2000 MODEL WITH AUTOMATIC WAVELENGTH SETTINGS AND BLANK. V-1100 Part no. 4120025

More information

Seite 2 Standard-Empfänger. EN Operation Instructions

Seite 2 Standard-Empfänger. EN Operation Instructions sign lux DE Bedienungsanleitung Seite 2 Standard-Empfänger EN Operation Instructions Page 10 Standard receiver FR Mode d emploi Page 18 Récepteur standard NL Gebruiksaanwijzing Pagina 26 Standaard ontvanger

More information

Advanced User s Guide

Advanced User s Guide Advanced User s Guide DCP-J72W Version 0 ARL/NZ User's Guides and where do I find them? Which Guide? What's in it? Where is it? Product Safety Guide Quick Setup Guide Basic User s Guide Advanced User s

More information

S E L E C T I O N. Upper Back. User manual

S E L E C T I O N. Upper Back. User manual and S E L E C T I O N T H E S T R E N G T H E V O L U T I O N User manual and and The identification plate of the and manufacturer, affixed to the frame on the side opposite the padded rest, gives the

More information

1. Introduction 2. Objectivation of Inspections via Measurements 3. Practical Application of Micro Sensor Systems 4. Summary & Conclusions

1. Introduction 2. Objectivation of Inspections via Measurements 3. Practical Application of Micro Sensor Systems 4. Summary & Conclusions , San Francisco, United States Mobile Devices and Multimedia: Enabling Technologies, Algorithms and Applications 2015 Session 2: Emerging Mobile Applications and Enabling Technologies Wednesday 11 February

More information

EBA 430 E EBA 430 EP. Operating Instructions

EBA 430 E EBA 430 EP. Operating Instructions EBA 430 E EBA 430 EP Operating Instructions EBA 430 E EBA 430 EP Dear customer, Thank you for choosing a paper cutting machine from EBA. With the purchase of this quality product you can be sure you have

More information

just below the screen. Data collection will begin, and a graph will show your data being plotted in real time.

just below the screen. Data collection will begin, and a graph will show your data being plotted in real time. To Collect Additional Data To start a second data collection run, tap the file should now see Run 2 displayed with a blank graph. cabinet in the upper right corner. You just below the screen. Data collection

More information

S E L E C T I O N. Arm Curl. User manual

S E L E C T I O N. Arm Curl. User manual S E L E C T I O N T H E S T R E N G T H E V O L U T I O N User manual The identification plate of the and manufacturer, affixed behind the seat, gives the following details: A Name and address of the manufacturer

More information

Sodium Thiosulphate and hydrochloric acid

Sodium Thiosulphate and hydrochloric acid Sodium Thiosulphate and hydrochloric acid Introduction The rate of reaction can be altered by varying temperature or concentration, or by changing the surface area of a solid reactant, or by adding a catalyst.

More information

SFP-550. Operating Manual MATSUHIDAI, MATSUDO-CITY, CHIBA JAPAN TEL: FAX:

SFP-550. Operating Manual MATSUHIDAI, MATSUDO-CITY, CHIBA JAPAN TEL: FAX: SFP-550 Operating Manual 296-1 MATSUHIDAI, MATSUDO-CITY, CHIBA 270-2214 JAPAN TEL: +81-47-388-6111 FAX: +81-47-388-4477 Safety Precautions To ensure safe operation and maximize product service life, observe

More information

User manual. Air humidity and temperature moisture meter humimeter RH2. Version 2.3_en Schaller GmbH 2016

User manual. Air humidity and temperature moisture meter humimeter RH2. Version 2.3_en Schaller GmbH 2016 User manual Air humidity and temperature moisture meter humimeter RH2 Version 2.3_en Schaller GmbH 2016 Distributore autorizzato : Geass S.r.l. - Torino - Tel.: +39.011.22.91.578 - mail : info@geass.com

More information

INDEX. Accessories and Components System Unit and Joystick Assembly and Charging the Battery Using with LED System...

INDEX. Accessories and Components System Unit and Joystick Assembly and Charging the Battery Using with LED System... USER GUIDE INDEX Accessories and Components... 4 System Unit and Joystick... 6 Assembly and Charging the Battery... 9 Using with LED System... 11 What is Ground Setting and How It Is Done... 14 Ground

More information

Heiland electronic GmbH TRD 2. B&W-Densitometer. User manual Version 6.4

Heiland electronic GmbH TRD 2. B&W-Densitometer. User manual Version 6.4 Heiland electronic GmbH TRD 2 B&W-Densitometer User manual Version 6.4 Table of Contents 1. GENERAL INFORMATION... 3 2. SAFETY REGULATIONS... 3 3. AREA OF APPLICATIONS... 3 4. INSTRUMENT DESCRIPTION 4.1.

More information

INTRODUCTION CHAPTER 1

INTRODUCTION CHAPTER 1 TABLE OF CONTENTS Introduction... 3 Package Contents... 5 Design... 7 Key Functions... 9 Grain Species and Measuring Ranges....11 First Steps...13 Measurement...15 Modification...19 Setup...21 Replacing

More information

Style 5 Style 7. INDEP. QUARTZ MIN.-1/2 MIN. Synchro radio INSTALLATION AND OPERATING INSTRUCTIONS. Réf.: I

Style 5 Style 7. INDEP. QUARTZ MIN.-1/2 MIN. Synchro radio INSTALLATION AND OPERATING INSTRUCTIONS. Réf.: I Style 5 Style 7 INDEP. QUARTZ MIN.-1/2 MIN. Synchro radio INSTALLATION AND OPERATING INSTRUCTIONS B.P. 1 49340 TRÉMENTINES FRANCE Tél. : 02 41 71 72 00 Fax : 02 41 71 72 01 www.bodet.com Réf.: 605808 I

More information

TABLE OF CONTENTS. Electromatic Equipment Co., Inc. -1-

TABLE OF CONTENTS. Electromatic Equipment Co., Inc. -1- TABLE OF CONTENTS 1.0 Introduction 2 1.1 Calibration curves 1.2 Explanation of calibration curves 2.0 Operating the instrument. 3 3.0 View of the instrument. 4 4.0 Menu.. 5 5.0 Other symbols and key functions..

More information

Fluke 125. Getting Started. Industrial ScopeMeter

Fluke 125. Getting Started. Industrial ScopeMeter Fluke 125 Industrial ScopeMeter Getting Started GB Dec 2006, Rev. 1, 09/2009 2006, 2009 Fluke Corporation, All rights reserved. Printed in The Netherlands All product names are trademarks of their respective

More information

Instruction Manual. DIGIPRO F2 Exposure Meter for Flash and Ambient Light /11-12

Instruction Manual. DIGIPRO F2 Exposure Meter for Flash and Ambient Light /11-12 Instruction Manual DIGIPRO F2 Exposure Meter for Flash and Ambient Light 15482 1/11-12 Swivel head Socket to connect the synchronising cable Measuring button M Buttons to adjust the values Display Buttons

More information

Enlarging Meter - Timer Instruction Manual DLGelectronics.com. Enlarging Meter - Timer

Enlarging Meter - Timer Instruction Manual DLGelectronics.com. Enlarging Meter - Timer Enlarging Meter - Timer 1 1 CONTENTS 1 Contents... 2 2 Important Note... 3 3 Introduction... 4 4 Setting Up the Enlarging Meter-Timer... 5 4.1 Precautions... 5 4.2 Connections... 5 4.2.1 Power Connections...

More information

EDUCATIONAL SPECTROPHOTOMETER ACCESSORY KIT AND EDUCATIONAL SPECTROPHOTOMETER SYSTEM

EDUCATIONAL SPECTROPHOTOMETER ACCESSORY KIT AND EDUCATIONAL SPECTROPHOTOMETER SYSTEM GAIN 1 10 100 Instruction Manual and Experiment Guide for the PASCO scientific Model OS-8537 and OS-8539 012-06575A 3/98 EDUCATIONAL SPECTROPHOTOMETER ACCESSORY KIT AND EDUCATIONAL SPECTROPHOTOMETER SYSTEM

More information

SFP-550. Operating Manual MATSUHIDAI, MATSUDO-CITY, CHIBA JAPAN TEL: FAX:

SFP-550. Operating Manual MATSUHIDAI, MATSUDO-CITY, CHIBA JAPAN TEL: FAX: SFP-550 Operating Manual 296-1 MATSUHIDAI, MATSUDO-CITY, CHIBA 270-2214 JAPAN TEL: +81-47-388-6111 FAX: +81-47-388-4477 Safety Precautions To ensure safe operation and maximize product service life, observe

More information

LED PowerBlinder 4 ORDERCODE 41320

LED PowerBlinder 4 ORDERCODE 41320 LED PowerBlinder 4 ORDERCODE 41320 Congratulations! You have bought a great, innovative product from Showtec. The Showtec LED PowerBlinder brings excitement to any venue. Whether you want simple plug-&-play

More information

1000TR. Instructions

1000TR. Instructions 1000TR ph Instructions CONTENTS 1. INTRODUCTION... 2 1.1 COMMON INTRODUCTION... 2 1.2 PARTS & ACCESSORIES... 2 2. INSTALLATION... 3 2.1 CASING... 3 2.2 MOUNTING... 3 2.3 ELECTRICAL INSTALLATION... 3 2.3.1

More information

Digital Color Printer

Digital Color Printer 3-858-635-12 (1) Digital Color Printer Operating Instructions Before operating the unit, please read this manual thoroughly and retain it for future reference. This PRINTER is designed to use PRINT PACK

More information

Nuclear Associates

Nuclear Associates Nuclear Associates 07-424 Digital Densitometer II Operators Manual March 2005 Manual No. 112111 Rev. 4 2003, 2005 Fluke Corporation, All rights reserved. Printed U.S.A. All product names are trademarks

More information

VWR SPECTROPHOTOMETERS

VWR SPECTROPHOTOMETERS VWR SPECTROPHOTOMETERS Reliable Accurate Easy to Use Your first choice for spectrophotometry VWR SPECTROPHOTOMETERS Quality Guaranteed VWR SPECTROPHOTOMETERS Single Beam Units VWR offers a choice of 5 models

More information

SL 200 instrument can process the energy intensity emitted by the solar radiation in a precise place of the earth.

SL 200 instrument can process the energy intensity emitted by the solar radiation in a precise place of the earth. Solarimeter SL 200 Table of contents 1 Introduction...4 2 General information...4 2.1 Measured units...4 2.2 Use...4 3 Operating principle...5 3.1 Keyboard presentation...5 3.2 Instrument offers 3 groups

More information

Using DIAGNOSE.EXE with a ColorQuest 45/0

Using DIAGNOSE.EXE with a ColorQuest 45/0 Insight on Color November, 1999, Vol. 11, No. 11 (Rev. 5/02) Using DIAGNOSE.EXE with a ColorQuest 45/0 The DIAGNOSE.EXE program, which may be used to assess instrument performance and diagnose potential

More information

Flash Stamp Machine T1511M User Manual

Flash Stamp Machine T1511M User Manual Flash Stamp Machine T1511M User Manual Declaration of conformity (Europe only) We, Trodat GmbH Linzer Straße 156 A-4600 Wels hereby declare that the Flash Stamp Machine model T1511M is CE certified respectively

More information

TECHNICAL INFORMATION BLUEMAXX Rechargeable Forensic Lights Catalog Nos. BM500, BM500220

TECHNICAL INFORMATION BLUEMAXX Rechargeable Forensic Lights Catalog Nos. BM500, BM500220 SIRCHIE Products Vehicles Training Copyright 2011 by SIRCHIE All Rights Reserved. TECHNICAL INFORMATION BLUEMAXX Rechargeable Forensic Lights Catalog Nos. BM500, BM500220 BLUEMAXX SYSTEMS OVERVIEW BLUEMAXX

More information

Digital Function Generator

Digital Function Generator Digital Function Generator 13654-99 PHYWE Systeme GmbH & Co. KG Robert-Bosch-Breite 10 37079 Göttingen Germany Tel. +49 (0) 551 604-0 Fax +49 (0) 551 604-107 E-mail info@phywe.de Operating Instructions

More information

MICROSCANNER TM D-Series

MICROSCANNER TM D-Series MICROSCANNER TM D-Series High Performance Models I N S T R U C T I O N M A N U A L C O R P O R A T I O N 818098 400 Pleasant Street Watertown, MA 02472 (617) 923-9900 800-442-3006 FAX: 617-923-9911 C O

More information

Digital Electronic Thermostat With RF

Digital Electronic Thermostat With RF RT300RF Manual Altech 005_89 06/05/2014 08:56 Page 1 Digital Electronic Thermostat With RF Instruction Manual Model No ALTHC015 RT300RF Manual Altech 005_89 06/05/2014 08:56 Page 2 2 ALTHC015 INSTRUCTION

More information

P20 Zoom Flash Zoom Flash P20 P20 Zoomblitz Flash con zoom P20 Flash externo P20 Flash Zoom P20

P20 Zoom Flash Zoom Flash P20 P20 Zoomblitz Flash con zoom P20 Flash externo P20 Flash Zoom P20 P20 Zoom Flash Zoom Flash P20 P20 Zoomblitz Flash con zoom P20 Flash externo P20 Flash Zoom P20 User s Guide Guide d'utilisation Benutzerhandbuch Manuale per l'utente Guía del usuario Guia do usuário 4J6021

More information

Introduction ORANGE PEEL / DOI. Structure size. Color Physical Properties Technical Service Index

Introduction ORANGE PEEL / DOI. Structure size. Color Physical Properties Technical Service Index Introduction The total appearance and the visibility of structures depend on the structure size, the observing distance and the image forming quality. Structure size Surfaces with different structure sizes

More information

Sartorius Extend The new achievers for your lab

Sartorius Extend The new achievers for your lab Sartorius Extend The new achievers for your lab Fisher Scientific The New Sartorius Extend Series The difference is... the latest technology If you compare the specifications of many lab balances on paper,

More information

Seapoint Rhodamine Fluorometer

Seapoint Rhodamine Fluorometer Seapoint Rhodamine Fluorometer User Manual Standard Version 6/00 Dimensions Open Configuration (no pump) 2.5" (6.4 cm) 6.6" (16.8 cm) 5.1" (13.0 cm) Pumped Configuration Figure 1. 1/2 Scale Drawing. Page

More information

Contents I. APPLICATIONS II. SPECIFICATIONS III. PRINCIPLES IV OPERATION V. MAINTENANCE

Contents I. APPLICATIONS II. SPECIFICATIONS III. PRINCIPLES IV OPERATION V. MAINTENANCE Contents I. APPLICATIONS...- 1 - II. SPECIFICATIONS...- 2 - III. PRINCIPLES...- 3 - IV OPERATION...- 9 - V. MAINTENANCE...- 21 - VI SUPPLEMENT...- 23 - Warning: The responsible person must be clear that

More information

S E L E C T I O N. Vertical Traction. User manual

S E L E C T I O N. Vertical Traction. User manual and S E L E C T I O N T H E S T R E N G T H E V O L U T I O N User manual and and The identification plate of the and manufacturer, affixed behind the backrest, gives the following details: A Name and

More information

TECHNICAL INFORMATION GoldPanther Forensic Light Source Kit (Patent Pending) Cat. Nos. FAL2000, FAL

TECHNICAL INFORMATION GoldPanther Forensic Light Source Kit (Patent Pending) Cat. Nos. FAL2000, FAL SIRCHIE FINGER PRINT LABORATORIES, INC. Website: www.sirchie.com E-mail: sirchie@mindspring.com 100 HUNTER PLACE YOUNGSVILLE, NC 27596 USA Phone: (919) 554-2244, (800) 356-7311 Fax: (919) 554-2266, (800)

More information

BTS256-E WiFi - mobile light meter for photopic and scotopic illuminance, EVE factor, luminous color, color rendering index and luminous spectrum.

BTS256-E WiFi - mobile light meter for photopic and scotopic illuminance, EVE factor, luminous color, color rendering index and luminous spectrum. Page 1 BTS256-E WiFi - mobile light meter for photopic and scotopic illuminance, EVE factor, luminous color, color rendering index and luminous spectrum. The BTS256-E WiFi is a high-quality light meter

More information

AUTOMATIC SURFACE & INTERFACIAL TENSIOMETER CODE:

AUTOMATIC SURFACE & INTERFACIAL TENSIOMETER CODE: AUTOMATIC SURFACE & INTERFACIAL TENSIOMETER CODE: 11053-04 COMPLIANCES The instrument is based on the DuNuoy ring, Wilhelmy plate principles and confirms to various National and International standards

More information

Artisan Technology Group is your source for quality new and certified-used/pre-owned equipment

Artisan Technology Group is your source for quality new and certified-used/pre-owned equipment Artisan Technology Group is your source for quality new and certified-used/pre-owned equipment FAST SHIPPING AND DELIVERY TENS OF THOUSANDS OF IN-STOCK ITEMS EQUIPMENT DEMOS HUNDREDS OF MANUFACTURERS SUPPORTED

More information

Operating Manual. Model 721N. Visible Spectrophotometer

Operating Manual. Model 721N. Visible Spectrophotometer Operating Manual of Model 721N Visible Spectrophotometer 1 Table of Contents 1. Chief uses... 3 2. Working Conditions... 3 3. Main Specifications...3 4.Operating Principles...4 5. Optical design...4 6.

More information

Electrical Safety Tester

Electrical Safety Tester Electrical Safety Tester GPT-9600 Series QUICK START GUIDE GW INSTEK PART NO. 82PT-96030MA1 ISO-9001 CERTIFIED MANUFACTURER This manual contains proprietary information, which is protected by copyright.

More information

UV-Vis-NIR Spectrophotometer Quick Start Guide

UV-Vis-NIR Spectrophotometer Quick Start Guide UV-Vis-NIR Spectrophotometer Quick Start Guide The following instructions are provided as a Quick Start Guide for powering up, running measurements, and shutting down the Lambda 950 UV-Vis Spectrophotometer.

More information

APA3-1500R STARTER SET WIRELESS SWITCHING

APA3-1500R STARTER SET WIRELESS SWITCHING COMPATIBILITY This set can be combined with all trust smart home transmitters and receivers. TECHNICAL SPECIFICATIONS RECEIVER Power supply input Maximum load RF frequency RF range Memory addresses 3 220-240

More information

Manual Moisture determination scales PCE-MA 110

Manual Moisture determination scales PCE-MA 110 PCE Americas Inc. 711 Commerce Way Suite 8 Jupiter FL-33458 USA Tel: 410-387-7703 Fax: 410-387-7714 info@pce-americas.com www.pce-instruments.com/us www.industrial-needs.com PCE Instruments UK Ltd. Units

More information

Zoom Stereo Microscope NYMCS-360 Instruction Manual

Zoom Stereo Microscope NYMCS-360 Instruction Manual Zoom Stereo Microscope NYMCS-60 Instruction Manual This manual is written for stereo microscope NYMCS-60. To ensure the safety, obtain optimum performance and to familiarize yourself fully with the use

More information

InsuLogix T Portable One Channel Temperature Measurement System Manual

InsuLogix T Portable One Channel Temperature Measurement System Manual InsuLogix T Portable One Channel Temperature Measurement System Manual Weidmann Technologies Deutschland GmbH Washingtonstraße 16/16a D-01139 Dresden, Germany Telefon: +49 (0)351 8435990 Version 1.2 InsuLogix

More information

Digital Anemometer

Digital Anemometer THE WORLD OF WEATHER DATA - THE WORLD OF WEATHER DATA - THE WORLD OF WEATHER DATA Instruction for Use 021404/07/04 Digital Anemometer 4.3406.00.000 ADOLF THIES GmbH & Co. KG Hauptstraße 76 37083 Göttingen

More information

Non-contact Forehead IR Thermometer

Non-contact Forehead IR Thermometer User s Manual Non-contact Forehead IR Thermometer Model IR200 Introduction Congratulations on your purchase of the Model IR200 IR Thermometer. This thermometer is intended for scanning groups of individuals

More information

testo Leakage detector for refrigerants Instruction manual

testo Leakage detector for refrigerants Instruction manual testo 316-3 Leakage detector for refrigerants Instruction manual 2 1 Contents 1 Contents 1 Contents... 3 2 Safety and the environment... 4 2.1. About this document... 4 2.2. Ensure safety... 4 2.3. Protecting

More information

ATD AMP Variable Speed Reciprocating Saw Owner s Manual

ATD AMP Variable Speed Reciprocating Saw Owner s Manual ATD-10535 7 AMP Variable Speed Reciprocating Saw Owner s Manual Manufactured in China To ATD Tools, Inc. Specifications TECHNICAL SPECIFICATIONS Voltage: 120V Frequency: 60Hz Power input: 7 Amps No load

More information

C4000. inline control. Photometric Converter. Real-time photometric converter. Factory zero for scattered light sensors

C4000. inline control. Photometric Converter.  Real-time photometric converter. Factory zero for scattered light sensors C4000 Photometric Converter inline control Real-time photometric converter Factory zero for scattered light sensors Simultaneous operation of up to four inline sensors Process control functions via Remote

More information

GFL-1000 User Manual Ground Fault Locator

GFL-1000 User Manual Ground Fault Locator GFL-Series User Manual V1.1 GFL-1000 User Manual Ground Fault Locator Contents Contents... 1 1 Declaration of Conformity... 3 2 Introduction... 3 3 Equipment Information... 3 3.1 Safety Precautions...

More information

User Manual InfraRed Thermometer With Laser Pointer and High/Low Alarms Model IR270

User Manual InfraRed Thermometer With Laser Pointer and High/Low Alarms Model IR270 User Manual InfraRed Thermometer With Laser Pointer and High/Low Alarms Model IR270 Additional User Manual Translations available at www.extech.com Introduction Thank you for selecting the Extech IR270

More information

User instructions Transmitted light laboratory microscope

User instructions Transmitted light laboratory microscope KERN & Sohn GmbH Ziegelei 1 D-72336 Balingen E-Mail: info@kern-sohn.com Tel: +49-[0]7433-9933-0 Fax: +49-[0]7433-9933-149 Internet: www.kern-sohn.com User instructions Transmitted light laboratory microscope

More information

TECHNICAL DATA / COLOR PAPER July 2003 E-19

TECHNICAL DATA / COLOR PAPER July 2003 E-19 TECHNICAL DATA / COLOR PAPER July 2003 E-19 KODAK EKTACOLOR Edge 7 Paper A-00 Month 199 DESCRIPTION KODAK EKTACOLOR Edge 7 Paper is a resin-coated color paper for printing color negatives. It is designed

More information

QUANTUM Qflash T2 / X2 OPERATING INSTRUCTIONS

QUANTUM Qflash T2 / X2 OPERATING INSTRUCTIONS QUANTUM Qflash T2 / X2 OPERATING INSTRUCTIONS 1.0 DESIGNATIONS T2 AND X2 1. Removable Reflector, two positions Normal and Wide angle. 2. Flash-tube 2A. Modeling Lamp (for Model X2 only) 3. Bounce Head,

More information

GC-1032 Metal Detector OWNER S MANUAL

GC-1032 Metal Detector OWNER S MANUAL GC-1032 Metal Detector OWNER S MANUAL 1 With your GC-1032 metal detector, you can hunt for coins, relics, jewelry, gold, and silver just about anywhere. The detector comes with high sensitivity and strong

More information

OPTITEMP TT 20 Technical Datasheet

OPTITEMP TT 20 Technical Datasheet OPTITEMP TT 20 Technical Datasheet Analog PC-programmable two-wire transmitters for Pt100 Efficient PC-configuration without recalibration Very stable output Very fast response time The documentation is

More information

FAG Detector II the mobile among data collectors. Technical Product Information

FAG Detector II the mobile among data collectors. Technical Product Information FAG II the mobile among data collectors Technical Product Information Application Condition-based maintenance Principle Application The FAG II is a vibration measuring device and data collector rolled

More information

Content Spectrophotometers

Content Spectrophotometers Spectroph Content Spectrophotometers Selection table Spectrophotometers Page 163 PRIM Page 164 UviLine Page 166 162 Selection table Spectrophotometers PRIM Light/ PRIM Advanced UviLine 9100/ UviLine 9400

More information