ORIGINAL ARTICLE. ESTHER MORENO-BARRIUSO, PhD, SUSANA MARCOS, PhD, RAFAEL NAVARRO, PhD, and STEPHEN A. BURNS, PhD
|
|
- Diana Kellie Bennett
- 6 years ago
- Views:
Transcription
1 /01/ /0 VOL. 78, NO. 3, PP OPTOMETRY AND VISION SCIENCE Copyright 2001 American Academy of Optometry ORIGINAL ARTICLE Comparing Laser Ray Tracing, the Spatially Resolved Refractometer, and the Hartmann- Shack Sensor to Measure the Ocular Wave Aberration ESTHER MORENO-BARRIUSO, PhD, SUSANA MARCOS, PhD, RAFAEL NAVARRO, PhD, and STEPHEN A. BURNS, PhD Instituto de Optica Daza de Valdés, Consejo Superior de Investigaciones Científicas (C.S.I.C.), Serrano 121, Madrid, Spain (EM- B, SM, RN) and Schepens Eye Research Institute, Boston Massachusetts (SAB) ABSTRACT: Purpose. To compare quantitatively three techniques to measure the optical aberrations of the human eye: laser ray tracing (LRT), the Hartmann-Shack wavefront sensor (H-S), and the spatially resolved refractometer (SRR). LRT and H-S are objective imaging techniques, whereas SRR is psychophysical. Methods. Wave aberrations were measured in two normal subjects with all three techniques implemented in two different laboratories. Results. We compared the experimental variability of the results obtained with each technique with the overall variability across the three methods. For the two subjects measured (RMS wavefront error 0.5 m and 0.9 m, respectively), we found a close agreement; the average standard deviation of the Zernike coefficients within a given method was 0.07 m, whereas the average global standard deviation across techniques was 0.09 m, which is only slightly higher. Conclusions. There is a close match between the Zernike coefficients obtained by LRT, H-S, and SRR. Thus, all three techniques provide similar information concerning wave aberration when applied to normal human eyes. However, the methods are operationally different, and each has advantages and disadvantages depending on the particular application. (Optom Vis Sci 2001;78: ) Key Words: ocular aberrations, laser ray tracing, Hartmann-Shack, spatially resolved refractometer, aberrometer In recent years, there has been a renewed effort to develop techniques for estimating the wave aberration of the human eye, 1 7 partly stimulated by the interest in assessing the changes in optical quality produced by refractive surgery 8, 9 and by the desire to compensate the ocular aberrations to achieve diffraction-limited optics Although a large number of techniques are now available, little work has been done so far to assess their equivalence or to establish which technique is best suited for a particular situation. In the current study, three different techniques, laser ray tracing (LRT), the spatially resolved refractometer (SRR), and the Hartmann-Shack wavefront sensor (H-S), have been compared in two subjects. Their principle and optical setups have been thoroughly described elsewhere. 3, 4, 16 LRT and H-S are objective techniques, whereas SRR is psychophysical. They all share a common basic principle 17 : the pupil is sampled at a set ( i, i ), for i 1,...nof locations, forming a given pattern. The raw data represent the set of ray aberrations [ x ( i, i ), y ( i, i )] corresponding to those pupil locations a and are proportional to the slope, or partial derivatives, of the wave aberration (W) 18 : x 1 W(, ) R p ; y 1 W(, ) R p where /R P, /R P are dimensionless canonical pupil coordinates and R p is the pupil radius. Ray aberrations are also given in dimensionless tangent units. If they are small, the tangent is equal to the angle. The joint representation of all the raw data [ x ( i, i ), y ( i, i )], for i 1,...n constitutes the spot diagram, which can be taken as a rough estimate of the shape of the retinal point spread a The raw data correspond to the average slope of the wave aberration over this sample area. This value is associated to the pupil location of the center of the sample beam for further data fitting. This is a good approximation in normal eyes. See reference 2 (Sections 2F and 3) for further details. (1)
2 Measuring the Ocular Wave Aberration Moreno-Barriuso et al. 153 function. 4 Instead of directly integrating Equation 1 to estimate W, it is numerically more stable to consider an expansion of W in terms of Zernike polynomials. Here we have considered a seventhorder approximation; that is, 35 terms 19 : W(, ) 35 k 1 Z k P k (, ) (2) where Z k are the coefficients of the expansion, in microns, and P k are dimensionless Zernike polynomials. By computing the partial derivatives of Equation 2 and substituting them in Equation 1 the wave aberration is estimated by a standard least-squares fitting of the raw data to the partial derivatives of the Zernike polynomials for each pupil location sampled. 20 METHODS The experimental approach differs between techniques. Fig. 1 shows simplified schematic diagrams of the setups. In LRT (Fig. 1a), a set of parallel laser Gaussian pencils is sequentially delivered FIGURE 1. Schematic diagrams of the basic setup for the three techniques: laser ray tracing (a), spatially resolved refractometer (b), Hartmann-Shack wavefront sensor (c). onto the eye through different pupil locations ( i, i ). Each ray forms a Gaussian spot on the retina (assuming that for the size of the entry beam, the effective pupil is diffraction limited) and an aerial image of the retina (which is the cross-correlation of the retinal point spread function with the Gaussian spot 21 ) is then relayed onto a CCD camera. The image is recorded, and its centroid is subsequently computed as a maximum-likelihood estimate of its position (X ( i, i ), Y ( i, i )). The joint plot of all the centroids constitutes the spot diagram. In the presence of aberrations, the spots (located in (X,Y )) are displaced from their reference positions (X X 0 ;Y Y 0 ), and the geometrical aberration is defined by the magnitude of these displacements (given here in dimensionless tangent units): x ( i, i ) [X ( i, i ) X 0 ( i, i )] y ( i, i ) [Y ( i, i ) Y 0 i, i )] (3) We use a unique reference in LRT (X 0,Y 0 ), which is the location of the spot for the chief ray. In LRT, the first pass is the measurement pass, whereas in the second pass, all rays are affected equally by the overall aberrations (assuming isoplanatism). SRR (Fig. 1b) is a psychophysical technique. As in LRT, the pupil is sampled by sequentially projecting diffraction-limited apertures (drilled in a spinning wheel) at different locations ( i, i )on the pupil. A focusing block (not shown in Fig 1b) allows for defocus compensation. Due to the aberrations, the subject will perceive the spots in different retinal locations. For each entry pupil, the task consists of having the patient tilt the incoming beam a given angle along both axes ( x ( i, i ), y ( i, i )) by means of a joystick until the spot is perceived to be aligned to a reference cross, viewed through the center of the pupil (reference channel in Fig. 1b). The tilt equals the wavefront slope, but with opposite sign so that the ray aberration is cancelled out. The ray aberration [ x ( i, i ), y ( i, i )] is computed from Equation 3 using (X ( i, i ), Y ( i, i )) ( x ( i, i ), y ( i, i )) and considering the center of the pupil as the aberration-free reference. In the H-S wavefront sensor (Fig. 1c), a monolithic microlens array, located in a plane conjugate to the exit pupil, samples the wavefront in parallel. Each lenslet selects a portion of the wavefront of the size of the lenslet and forms a spot of light at its focal plane. A single snapshot provides a grid of image spots, which is imaged onto a CCD camera. Each spot is the cross-correlation between the point spread function of the eye (first pass) and that of a single microlens (second, measurement pass). For an aberration-free eye, the emerging wavefront is flat, and thus the grid of image spots has a spatial distribution (X 0 ( i, i ), Y 0 ( i, i )) identical to that of the microlenses themselves. In the presence of aberrations, the image spots (X ( i, i ), Y ( i, i )) are shifted from their reference position. This shift is proportional to the average slope of the wavefront across the microlens pupil (Equation 1). By applying Equation 3, subtracting the test from the reference coordinates of the centroids, we obtain the geometrical aberrations [ x( i, i ), y( i, i )] and the spot diagram. Table 1 summarizes the characteristic features of each of the three techniques, both the ones inherent to the method itself and the ones due to the particular setup used for this experiment (but that could be modified in alternative implementations). For this comparative study, two different laboratories were involved: we
3 154 Measuring the Ocular Wave Aberration Moreno-Barriuso et al. TABLE 1. Specifications for Laser Ray Tracing, Spatially Resolved Refractometer, and Hartmann-Shack Wavefront Sensor implemented for this study LRT a SRR H-S Objective/psychophysical Objective Psychophysical Objective Parallel/sequential Sequential Sequential Parallel No. of samples taken b Sample diameter (mm) b Sampling step (mm) b Sampling pattern (shape) Rectangular Rectangular Hexagonal Effective pupil sampled (mm) c Measurement in first/second pass First First Second Entry pupil size (mm) Exit pupil size (mm) 3 d N/A 0.5 e Wavelength (nm) No. of runs averaged for each subject Mydriasis 1 Drop cyclopentolate 1% No Mydriasis required 1 Drop cyclopentolate 1% in these measurements 1 Drop cyclopentolate 1% Light level required 10 W 0.1 W (Photopic) f 30 W Duration of the measurement 5 s 3 or4min 2 s Area available for each aerial image (pixels) N/A Zernike terms fitted a Abbreviations: LRT, Laser Ray Tracing; SRR, Spatially Resolved Refractometer; H-S, Hartmann-Shack Wavefront Sensor. b Specifications of the microlens array: focal length f 50 mm; clear aperture 500 m; lenslet interdistance 600 m; hexagonal arrangement. c The effective pupil diameter is the distance from the pupil center required for all the light from each aperture to enter (or exit) the eye. For each technique, the wave aberration was computed for the corresponding effective pupil size, and then the Zernike coefficients are recalculated for the smallest pupil (6.6 mm) to allow direct comparison. d Artificial pupil located in a plane conjugate to the eye s pupil. e Equal to the microlens diameter. f Retinal illuminance is about 10 td when considering both the measurement beam and the extended background field. used the SRR built at the Schepens Eye Research Institute (Boston, MA), 2 whereas the H-S sensor and the LRT technique were implemented at the Institute of Optics (Madrid, Spain). 14 The experimental conditions were kept as similar as possible, taking into account the inherent differences among methods. Two subjects, one male aged 41 years (RN) and one female aged 25 years (EM), participated in the study, both using their right eye. In all three types of measurements, defocus was left uncorrected ( 1.5 D for RN and 0.5D for EM). Measurements were carried out over a 2-year period. For each instrument, sessions consisted of four runs of measurements from which the set of average Zernike coefficients (for comparison between techniques) and the corresponding standard deviation (for variability estimation within each technique) were computed. RESULTS The resulting Zernike coefficients are shown in Fig. 2 for subjects RN (a) and EM (b). Solid triangles represent values for SRR, open circles represent LRT, and solid squares represent H-S. Each symbol is the average value from four runs. The error bars (representing the standard deviation of each technique) have not been depicted for the sake of clarity because they are all the same size as the symbol or smaller. The ordering for the Zernike polynomials corresponds to that recommended by the Optical Society of America Standardization Committee. 22 We assessed the variability within a technique by computing the standard deviation of each Zernike coefficient across runs in a session and then taking the average across coefficients. The standard deviation within each technique was compared with the average standard deviation across techniques. For this purpose, we averaged the set of Zernike Coefficients obtained with LRT, SRR, and H-S, computed the standard deviation of each individual coefficient, and then averaged these values across the coefficients. Averaging was done either for coefficients 3 to 35 (orders 2 7) or for coefficients 3 to 14 (orders 2 4) to analyze possible changes across orders. There was a close match among the three techniques: For RN (RMS wavefront error 0.5 m) the average standard deviation across data (Zernike coefficients) from an individual technique was m (0.066 m for H-S, for LRT, and for SRR), whereas the average standard deviation across the three techniques was 0.09 m. For subject EM (RMS wavefront error 0.9 m), the average standard deviation of data from individual techniques was m, whereas the average standard deviation across the three techniques was again 0.09 m. The main contribution to the slight mismatch across techniques was due to the low-order aberrations, which present the largest aberration values.
4 Measuring the Ocular Wave Aberration Moreno-Barriuso et al. 155 Considering only orders 2 to 4 (coefficients 3 to 14), the average standard deviation across techniques was 0.16 m, and the standard deviation within a technique was 0.09 m for RN; the standard deviations were 0.14 m and 0.08 m, respectively, for EM. The main contribution to the larger intertechnique standard deviation was due to the defocus term, which is the one most likely to change across sessions due to the variation of the accommodative state reached by the subject after instillation of the mydriatic agent. Fig. 3 shows contour plots of the wave aberration for RN for each of the three methods. Tilt terms (Z 1 and Z 2 ) and defocus (Z 4 ) have been cancelled in these plots to enhance the high-order features of the wave aberration, showing also the equivalence of the three methods for the measurement of optical aberrations in normal subjects. Nevertheless, given their different nature, each one can present advantages or limitations depending on the specific application. DISCUSSION FIGURE 2. Zernike coefficients for the three techniques for subjects RN (upper) and EM (lower). Triangles represent SRR, open circles represent LRT, and solid squares represent H-S. Each symbol is the average across four runs. Errors bars (not shown) are the same size or smaller than the symbols in all cases. FIGURE 3. Contour plots of the wave aberration for the right eye of subject RN for H-S (left), LRT (center), and SRR (right) measurements. Left represents temporal side, and right represents nasal side. Pupil coordinates range from 3.25 mm to 3.25 mm. Step between adjacent contour lines is 0.5 m. Tilt terms (Z 1 and Z 2 ) and defocus (Z 4 ) have been cancelled. The results of this study and the main differences among the three techniques are summarized below. We emphasize both issues that are inherent to the nature of each method and issues related to our particular implementation. First, for the two normal subjects presented here, the three methods are equivalent. However, some differences may arise when measuring non-normal eyes. For example, for eyes with considerable amounts of wave aberrations, LRT and SSR may be more robust than H-S because of their sequential nature. That is, because they are sequential, they do not suffer from indexing problems, 8 which may appear in H-S if the grid of image spots is too distorted. Alternatively, the computation of the centroid may be affected by speckle noise and the limited number of pixels/spot in H-S 14 ; when the spots are too distorted (for highly aberrated eyes) and the boundary between adjacent spots is not clear, the centroid computations become difficult, even for a diffraction-limited first pass. 14 This may limit its range of application, especially in some clinically interesting cases. Alternative light sources (such as superluminescent diodes) have been suggested, to overcome the speckle noise problem in H-S. This limitation is reduced in LRT, in which the entire CCD is used for each spot. In SRR, the only limitation is the ability of the subject to align the image spot and the dynamic range of the gimbaled mirror. Second, measurements with LRT and H-S rely on a good retinal reflection, which might be a problem in abnormal retinas. 23 This is not a hindrance for SRR; however, SRR requires proper light capture by the photoreceptors and good fixation. These requirements could be a disadvantage for some classes of patients. Third, the sequential nature of LRT and SRR limits their application for real-time dynamic measurements. H-S is the fastest technique (1 or 2 s), and speckle removal techniques can be applied to take faster real-time dynamic recordings. 24 In the present configuration, the duration of each exposure in LRT can be reduced to ~50 ms without being limited by speckle noise, 14 the main constraint in duration of a measurement being the speed of the imaging board. This implementation requires 5 s for 37 rays. An SRR run takes 3 to 4 min due to its psychophysical nature. Although eye movements, fluctuations of accommodation, etc., may be thought to be more important problems for the slowest techniques, we did not find a trend for increased variability with a particular technique: for RN, SRR was the noisiest (0.083 m); for EM, H-S was the noisiest (0.09 m). The following two aspects are related to the particular implementation of the technique used in the study and thus could be overcome in future versions of the system. First, all three systems operate with visible light (although some versions of H-S with infrared light have already been used by other groups 24 ), and as a result, pupil dilation is required for H-S and LRT to avoid reflex pupil constriction. SRR uses light levels sufficiently low to operate with a natural pupil for most subjects (although for this work, a
5 156 Measuring the Ocular Wave Aberration Moreno-Barriuso et al. midriatic was used for comparison with the other methods. In this sense, it can therefore be more comfortable for the subject. Second, H-S uses a fixed monolithic microlens array; the sampling configuration cannot be modified unless different microlens arrays are used. In the SRR version used in this work, the sampling pattern provided by the wheel was also fixed, although programmable masks are possible. 25 In LRT, the laser scanner can be flexibly programmed to change the shape of the sampling pattern (rectangular, hexagonal, and polar). This allows the sampling pattern to be adapted to individual cases (i.e., normal vs. refractive surgery patients) or even the use of a nonhomogeneous sampling pattern. 26 Even if the sampling step size can be changed easily, the sample size (given by the beam waist size) remains unchanged, which affects the filling factor (fraction of the pupil covered by the sampling pattern). In summary, we have demonstrated experimentally the equivalence of three different techniques for measuring ocular aberrations in normal human eyes: the laser ray tracing, the spatially resolved refractometer and the Hartmann-Shack wavefront sensor. The particular features of each technique (see table) make one or the other more convenient depending on the specific problem or application. ACKNOWLEDGMENTS We thank Mike Hutley for providing the microlens array used in the experiments. This research was supported by grant TIC98 to 0925-C02 to 01 from the Comisión Interministerial de Ciencia y Tecnología of Spain, grant EY04395 from the National Eye Institute, National Institutes of Health, and by the Human Frontier Science Program LT-542/97. Received August 21, 2000; revision received December 7, REFERENCES 1. Walsh G, Charman WN, Howland HC. Objective technique for the determination of monochromatic aberrations of the human eye. J Opt Soc Am (A) 1984;1: He JC, Marcos S, Webb RH, Burns SA. Measurement of the wavefront aberration of the eye by a fast psychophysical procedure. J Opt Soc Am (A) Opt Image Sci Vis 1998;15: Liang J, Grimm B, Goelz S, Bille JF. Objective measurement of wave aberrations of the human eye with the use of a Hartmann-Shack wave-front sensor. J Opt Soc Am (A) 1994;11: Navarro R, Losada MA. Aberrations and relative efficiency of light pencils in the living human eye. Optom Vis Sci 1997;74: Liang J, Williams DR. Aberrations and retinal image quality of the normal human eye. J Opt Soc Am (A) 1997;14: Iglesias I, Berrio E, Artal P. Estimates of the ocular wave aberration from pairs of double-pass retinal images. J Opt Soc Am (A) 1998;15: Mierdel P, Krinke HE, Wiegand W, Kaemmerer M, Seiler T. Measuring device for determining monochromatic aberration of the human eye. Ophthalmologe 1997;94: Thibos LN, Hong X. Clinical applications of the Shack-Hartmann aberrometer. Optom Vis Sci 1999;76: Oshika T, Klyce SD, Applegate RA, Howland HC, El Danasoury MA. Comparison of corneal wavefront aberrations after photorefractive keratectomy and laser in situ keratomileusis. Am J Ophthalmol 1999;127: Liang J, Williams DR, Miller DT. Supernormal vision and highresolution retinal imaging through adaptive optics. J Opt Soc Am (A) 1997;14: Zhu L, Sun P, Bartsch D, Freeman WR, Fainman Y. Adaptive control of a micromachined continous-membrane deformable mirror for aberration compensation. Appl Opt 1999;38: Navarro R, Moreno-Barriuso E, Bará S, Mancebo T. Phase-plates for wave-aberration compensation in the human eye. Opt Lett 2000;25: Salmon TO, Thibos LN, Bradley A. Comparison of the eye s wavefront aberration measured psychophysically and with the Shack- Hartmann wave-front sensor. J Opt Soc Am (A) 1998;15: Moreno-Barriuso E, Navarro R. Laser Ray Tracing versus Hartmann- Shack sensor for measuring optical aberrations in the human eye. J Opt Soc Am (A) Opt Image Sci Vis 2000;17: Thibos L, Hong X. Comparison of monochromatic aberrations of the human eye measured with the Howland crossed-cylinder aberroscope and the Shack-Hartmann aberrometer. In: Optical Society of America Annual Meeting, Technical Digest Series,1999. Washington, DC: Optical Society of America,1999: Webb RH, Penney CM, Thompson KP. Measurement of ocular wavefront distortion with a spatially resolved refractometer. Appl Opt 1992;31: Smirnov MS. Measurement of the wave aberration of the human eye. 1961: Born M, Wolf E. Principles of Optics: Electromagnetic theory of Propagation, Intererence and Diffraction of Light, 6th ed. Oxford: Pergamon Press, Malacara D. Optical Shop Testing, 2nd ed. New York: Wiley, Southwell WH. Wave-front estimation from wave-front slope measurements. J Opt Soc Am (A) 1980;70: Navarro R, Losada MA. Phase transfer and point-spread function of the human eye determined by a new asymmetric double-pass method. J Opt Soc Am (A) 1995;12: Thibos LN, Applegate RA, Schwiegerling JT, Webb R, Burns S, VSIA Standards Taskforce Members. Report from the VSIA Taskforce on standards for the reporting of the optical aberrations of the eye. In: Vision Science and its Applications, Technical Digest Series. Washington, DC: Optical Society of America, 2000: Elsner AE, Burns SA, Weiter JJ, Delori FC. Infrared imaging of sub-retinal structures in the human ocular fundus. Vision Res 1996; 36: Hofer HJ, Porter J, Williams DR. Dynamic measurement of the wave aberration of the human eye. Invest Ophthalmol Vis Sci 1998;39: S Webb RH, Burns SA, Penney M. Coaxial spatially resolved refractometer. US Patent 6,000,800, Rios S, Acosta E, Bara S. Hartmann sensing with Albrecht grids. Opt Comm 1997;133: Esther Moreno Instituto de Optica Daza de Valdés (C.S.I.C.) Serrano Madrid, Spain e.moreno@io.cfmac.csic.es
The pupil of the eye is a critical limiting factor in the optics
Pupil Location under Mesopic, Photopic, and Pharmacologically Dilated Conditions Yabo Yang, 1,2 Keith Thompson, 3 and Stephen A. Burns 1 PURPOSE. To determine whether there are systematic changes in pupil
More informationCustomized Correction of Wavefront Aberrations in Abnormal Human Eyes by Using a Phase Plate and a Customized Contact Lens
Journal of the Korean Physical Society, Vol. 49, No. 1, July 2006, pp. 121 125 Customized Correction of Wavefront Aberrations in Abnormal Human Eyes by Using a Phase Plate and a Customized Contact Lens
More information4th International Congress of Wavefront Sensing and Aberration-free Refractive Correction ADAPTIVE OPTICS FOR VISION: THE EYE S ADAPTATION TO ITS
4th International Congress of Wavefront Sensing and Aberration-free Refractive Correction (Supplement to the Journal of Refractive Surgery; June 2003) ADAPTIVE OPTICS FOR VISION: THE EYE S ADAPTATION TO
More informationA new approach to the study of ocular chromatic aberrations
Vision Research 39 (1999) 4309 4323 www.elsevier.com/locate/visres A new approach to the study of ocular chromatic aberrations Susana Marcos a, *, Stephen A. Burns b, Esther Moreno-Barriusop b, Rafael
More informationRon Liu OPTI521-Introductory Optomechanical Engineering December 7, 2009
Synopsis of METHOD AND APPARATUS FOR IMPROVING VISION AND THE RESOLUTION OF RETINAL IMAGES by David R. Williams and Junzhong Liang from the US Patent Number: 5,777,719 issued in July 7, 1998 Ron Liu OPTI521-Introductory
More informationMonochromatic aberrations and point-spread functions of the human eye across the visual field
2522 J. Opt. Soc. Am. A/Vol. 15, No. 9/September 1998 Navarro et al. Monochromatic aberrations and point-spread functions of the human eye across the visual field Rafael Navarro, Esther Moreno, and Carlos
More informationORIGINAL ARTICLE. Aberrations of the Human Eye in Visible and Near Infrared Illumination
1040-5488/03/8001-0026/0 VOL. 80, NO. 1, PP. 26 35 OPTOMETRY AND VISION SCIENCE Copyright 2003 American Academy of Optometry ORIGINAL ARTICLE Aberrations of the Human Eye in Visible and Near Infrared Illumination
More informationRAYMOND A. APPLEGATE,
1040-5488/03/8001-0015/0 VOL. 80, NO. 1, PP. 15 25 OPTOMETRY AND VISION SCIENCE Copyright 2003 American Academy of Optometry ORIGINAL ARTICLE Comparison of Monochromatic Ocular Aberrations Measured with
More informationIs Aberration-Free Correction the Best Goal
Is Aberration-Free Correction the Best Goal Stephen Burns, PhD, Jamie McLellan, Ph.D., Susana Marcos, Ph.D. The Schepens Eye Research Institute. Schepens Eye Research Institute, an affiliate of Harvard
More informationNormal Wavefront Error as a Function of Age and Pupil Size
RAA Normal Wavefront Error as a Function of Age and Pupil Size Raymond A. Applegate, OD, PhD Borish Chair of Optometry Director of the Visual Optics Institute College of Optometry University of Houston
More informationThe reduction in photopic contrast sensitivity with age 1 3
Age-Related Changes in Monochromatic Wave Aberrations of the Human Eye James S. McLellan, 1 Susana Marcos, 1,2 and Stephen A. Burns 1 PURPOSE. To investigate the relations between age and the optical aberrations
More informationIn this issue of the Journal, Oliver and colleagues
Special Article Refractive Surgery, Optical Aberrations, and Visual Performance Raymond A. Applegate, OD, PhD; Howard C. Howland,PhD In this issue of the Journal, Oliver and colleagues report that photorefractive
More informationInvestigating sources of variability of monochromatic and transverse chromatic aberrations across eyes
Vision Research 41 (2001) 3861 3871 www.elsevier.com/locate/visres Investigating sources of variability of monochromatic and transverse chromatic aberrations across eyes Susana Marcos a,b, *, Stephen A.
More informationVision Research at. Validation of a Novel Hartmann-Moiré Wavefront Sensor with Large Dynamic Range. Wavefront Science Congress, Feb.
Wavefront Science Congress, Feb. 2008 Validation of a Novel Hartmann-Moiré Wavefront Sensor with Large Dynamic Range Xin Wei 1, Tony Van Heugten 2, Nikole L. Himebaugh 1, Pete S. Kollbaum 1, Mei Zhang
More informationEffect of rotation and translation on the expected benefit of an ideal method to correct the eye s higher-order aberrations
Guirao et al. Vol. 18, No. 5/May 2001/J. Opt. Soc. Am. A 1003 Effect of rotation and translation on the expected benefit of an ideal method to correct the eye s higher-order aberrations Antonio Guirao
More informationAlthough the presence of optical imperfections
Validation of the Estimation of Corneal Aberrations From Videokeratography in Keratoconus Sergio Barbero, BSc; Susana Marcos, PhD; Jesus Merayo-Lloves, MD, PhD; Esther Moreno-Barriuso, PhD ABSTRACT PURPOSE:
More informationCorrecting Highly Aberrated Eyes Using Large-stroke Adaptive Optics
Correcting Highly Aberrated Eyes Using Large-stroke Adaptive Optics Ramkumar Sabesan, BTech; Kamran Ahmad, MS; Geunyoung Yoon, PhD ABSTRACT PURPOSE: To investigate the optical performance of a large-stroke
More informationHartmann-Shack sensor ASIC s for real-time adaptive optics in biomedical physics
Hartmann-Shack sensor ASIC s for real-time adaptive optics in biomedical physics Thomas NIRMAIER Kirchhoff Institute, University of Heidelberg Heidelberg, Germany Dirk DROSTE Robert Bosch Group Stuttgart,
More informationIn recent years there has been an explosion of
Line of Sight and Alternative Representations of Aberrations of the Eye Stanley A. Klein, PhD; Daniel D. Garcia, PhD ABSTRACT Several methods for representing pupil plane aberrations based on wavefront
More informationCalculated impact of higher-order monochromatic aberrations on retinal image quality in a population of human eyes: erratum
ERRATA Calculated impact of higher-order monochromatic aberrations on retinal image quality in a population of human eyes: erratum Antonio Guirao* Laboratorio de Optica, Departamento de Física, Universidad
More informationGeneration of third-order spherical and coma aberrations by use of radially symmetrical fourth-order lenses
López-Gil et al. Vol. 15, No. 9/September 1998/J. Opt. Soc. Am. A 2563 Generation of third-order spherical and coma aberrations by use of radially symmetrical fourth-order lenses N. López-Gil Section of
More informationphone extn.3662, fax: , nitt.edu ABSTRACT
Analysis of Refractive errors in the human eye using Shack Hartmann Aberrometry M. Jesson, P. Arulmozhivarman, and A.R. Ganesan* Department of Physics, National Institute of Technology, Tiruchirappalli
More informationShape of stars and optical quality of the human eye
R. Navarro and M. A. Losada Vol. 14, No. 2/February 1997/J. Opt. Soc. Am. A 353 Shape of stars and optical quality of the human eye Rafael Navarro and M. Angeles Losada Instituto de Optica Daza de Valdés,
More informationThis is the author s version of a work that was submitted/accepted for publication in the following source:
This is the author s version of a work that was submitted/accepted for publication in the following source: Atchison, David A. & Mathur, Ankit (2014) Effects of pupil center shift on ocular aberrations.
More informationPantoscopic tilt induced higher order aberrations characterization using Shack Hartmann wave front sensor and comparison with Martin s Rule.
Research Article http://www.alliedacademies.org/ophthalmic-and-eye-research/ Pantoscopic tilt induced higher order aberrations characterization using Shack Hartmann wave front sensor and comparison with
More informationWaveMaster IOL. Fast and Accurate Intraocular Lens Tester
WaveMaster IOL Fast and Accurate Intraocular Lens Tester INTRAOCULAR LENS TESTER WaveMaster IOL Fast and accurate intraocular lens tester WaveMaster IOL is an instrument providing real time analysis of
More informationWaveMaster IOL. Fast and accurate intraocular lens tester
WaveMaster IOL Fast and accurate intraocular lens tester INTRAOCULAR LENS TESTER WaveMaster IOL Fast and accurate intraocular lens tester WaveMaster IOL is a new instrument providing real time analysis
More informationCorneal and total optical aberrations in a unilateral aphakic patient
Corneal and total optical aberrations in a unilateral aphakic patient Sergio Barbero, Susana Marcos, PhD, Jesús Merayo-Lloves, MD, PhD Purpose: To measure corneal and total optical aberrations in the normal
More informationORIGINAL ARTICLE. On-Eye Measurement of Optical Performance of Rigid Gas Permeable Contact Lenses Based on Ocular and Corneal Aberrometry
1040-5488/03/8002-0115/0 VOL. 80, NO. 2, PP. 115 125 OPTOMETRY AND VISION SCIENCE Copyright 2003 American Academy of Optometry ORIGINAL ARTICLE On-Eye Measurement of Optical Performance of Rigid Gas Permeable
More informationImage Quality of the Human Eye. Susana Marcos, Ph.D.
Image Quality of the Human Eye Susana Marcos, Ph.D. Factors Contributing to Retinal Image Degradation The eye is an optical instrument that projects scenes of the visual world onto the retina. It has been
More informationTransferring wavefront measurements to ablation profiles. Michael Mrochen PhD Swiss Federal Institut of Technology, Zurich IROC Zurich
Transferring wavefront measurements to ablation profiles Michael Mrochen PhD Swiss Federal Institut of Technology, Zurich IROC Zurich corneal ablation Calculation laser spot positions Centration Calculation
More informationOcular Shack-Hartmann sensor resolution. Dan Neal Dan Topa James Copland
Ocular Shack-Hartmann sensor resolution Dan Neal Dan Topa James Copland Outline Introduction Shack-Hartmann wavefront sensors Performance parameters Reconstructors Resolution effects Spot degradation Accuracy
More informationOcular aberrations as a function of wavelength in the near infrared measured with a femtosecond laser
Ocular aberrations as a function of wavelength in the near infrared measured with a femtosecond laser Enrique J. Fernández Department of Biomedical Engineering and Physics, Medical University of Vienna,
More informationEffects of Pupil Center Shift on Ocular Aberrations
Visual Psychophysics and Physiological Optics Effects of Pupil Center Shift on Ocular Aberrations David A. Atchison and Ankit Mathur School of Optometry & Vision Science and Institute of Health & Biomedical
More informationORIGINAL ARTICLE. Metrics of Retinal Image Quality Predict Visual Performance in Eyes With 20/17 or Better Visual Acuity
1040-5488/06/8309-0635/0 VOL. 83, NO. 9, PP. 635 640 OPTOMETRY AND VISION SCIENCE Copyright 2006 American Academy of Optometry ORIGINAL ARTICLE Metrics of Retinal Image Quality Predict Visual Performance
More informationPaper Synopsis. Xiaoyin Zhu Nov 5, 2012 OPTI 521
Paper Synopsis Xiaoyin Zhu Nov 5, 2012 OPTI 521 Paper: Active Optics and Wavefront Sensing at the Upgraded 6.5-meter MMT by T. E. Pickering, S. C. West, and D. G. Fabricant Abstract: This synopsis summarized
More informationExtended source pyramid wave-front sensor for the human eye
Extended source pyramid wave-front sensor for the human eye Ignacio Iglesias, Roberto Ragazzoni*, Yves Julien and Pablo Artal Laboratorio de Optica, Departamento de Física, Universidad de Murcia, Murcia,
More informationAdaptive Optics. Adaptive optics for imaging. Adaptive optics to improve. Ocular High order Aberrations (HOA)
Effect of Adaptive Optics Correction on Visual Performance and Accommodation Adaptive optics for imaging Astromomy Retinal imaging Since 977, Hardy et al, JOSA A Since 989, Dreher et al. Appl Opt Susana
More informationTesting Aspherics Using Two-Wavelength Holography
Reprinted from APPLIED OPTICS. Vol. 10, page 2113, September 1971 Copyright 1971 by the Optical Society of America and reprinted by permission of the copyright owner Testing Aspherics Using Two-Wavelength
More informationExplanation of Aberration and Wavefront
Explanation of Aberration and Wavefront 1. What Causes Blur? 2. What is? 4. What is wavefront? 5. Hartmann-Shack Aberrometer 6. Adoption of wavefront technology David Oh 1. What Causes Blur? 2. What is?
More informationAgilEye Manual Version 2.0 February 28, 2007
AgilEye Manual Version 2.0 February 28, 2007 1717 Louisiana NE Suite 202 Albuquerque, NM 87110 (505) 268-4742 support@agiloptics.com 2 (505) 268-4742 v. 2.0 February 07, 2007 3 Introduction AgilEye Wavefront
More informationORIGINAL ARTICLES. Image Metrics for Predicting Subjective Image Quality
1040-5488/05/8205-0358/0 VOL. 82, NO. 5, PP. 358 369 OPTOMETRY AND VISION SCIENCE Copyright 2005 American Academy of Optometry ORIGINAL ARTICLES Image Metrics for Predicting Subjective Image Quality LI
More informationImpact of scattering and spherical aberration in contrast sensitivity
Journal of Vision (2009) 9(3):19, 1 10 http://journalofvision.org/9/3/19/ 1 Impact of scattering and spherical aberration in contrast sensitivity Guillermo M. Pérez Silvestre Manzanera Pablo Artal Laboratorio
More informationORIGINAL ARTICLE. Correlation between Optical and Psychophysical Parameters as a Function of Defocus
1040-5488/02/7901-0001/0 VOL. 79, NO. 1, PP. 60-67 OPTOMETRY AND VISION SCIENCE Copyright 2002 American Academy of Optometry A schematic view of the apparatus used is shown in Fig. 1. It is a double-pass
More informationWavefront Sensing In Other Disciplines. 15 February 2003 Jerry Nelson, UCSC Wavefront Congress
Wavefront Sensing In Other Disciplines 15 February 2003 Jerry Nelson, UCSC Wavefront Congress QuickTime and a Photo - JPEG decompressor are needed to see this picture. 15feb03 Nelson wavefront sensing
More informationTheoretical modeling and evaluation of the axial resolution of the adaptive optics scanning laser ophthalmoscope
Journal of Biomedical Optics 9(1), 132 138 (January/February 2004) Theoretical modeling and evaluation of the axial resolution of the adaptive optics scanning laser ophthalmoscope Krishnakumar Venkateswaran
More informationOpen-loop performance of a high dynamic range reflective wavefront sensor
Open-loop performance of a high dynamic range reflective wavefront sensor Jonathan R. Andrews 1, Scott W. Teare 2, Sergio R. Restaino 1, David Wick 3, Christopher C. Wilcox 1, Ty Martinez 1 Abstract: Sandia
More informationOptics of Wavefront. Austin Roorda, Ph.D. University of Houston College of Optometry
Optics of Wavefront Austin Roorda, Ph.D. University of Houston College of Optometry Geometrical Optics Relationships between pupil size, refractive error and blur Optics of the eye: Depth of Focus 2 mm
More informationHigh contrast imaging lab
High contrast imaging lab Ay122a, November 2016, D. Mawet Introduction This lab is an introduction to high contrast imaging, and in particular coronagraphy and its interaction with adaptive optics sytems.
More informationCorneal laser surgery is currently shifting its
Correlation Between Corneal and Total Wavefront Aberrations in Myopic Eyes Michael Mrochen, PhD; Mirko Jankov, MD; Michael Bueeler, MS; Theo Seiler, MD, PhD ABSTRACT PURPOSE: Corneal topography data expressed
More informationWavefront sensing by an aperiodic diffractive microlens array
Wavefront sensing by an aperiodic diffractive microlens array Lars Seifert a, Thomas Ruppel, Tobias Haist, and Wolfgang Osten a Institut für Technische Optik, Universität Stuttgart, Pfaffenwaldring 9,
More informationOdd aberrations and double-pass measurements of retinal image quality
Artal et al. Vol. 12, No. 2/February 1995/J. Opt. Soc. Am. A 195 Odd aberrations and double-pass measurements of retinal image quality Pablo Artal Laboratorio de Optica, Departamento de Física, Universidad
More informationAdaptive Optics for LIGO
Adaptive Optics for LIGO Justin Mansell Ginzton Laboratory LIGO-G990022-39-M Motivation Wavefront Sensor Outline Characterization Enhancements Modeling Projections Adaptive Optics Results Effects of Thermal
More informationDETERMINING CALIBRATION PARAMETERS FOR A HARTMANN- SHACK WAVEFRONT SENSOR
DETERMINING CALIBRATION PARAMETERS FOR A HARTMANN- SHACK WAVEFRONT SENSOR Felipe Tayer Amaral¹, Luciana P. Salles 2 and Davies William de Lima Monteiro 3,2 Graduate Program in Electrical Engineering -
More informationAn analysis of retinal receptor orientation
An analysis of retinal receptor orientation IV. Center of the entrance pupil and the center of convergence of orientation and directional sensitivity Jay M. Enoch and G. M. Hope In the previous study,
More informationAnalysis of Hartmann testing techniques for large-sized optics
Analysis of Hartmann testing techniques for large-sized optics Nadezhda D. Tolstoba St.-Petersburg State Institute of Fine Mechanics and Optics (Technical University) Sablinskaya ul.,14, St.-Petersburg,
More informationOff-axis wave front measurements for optical correction in eccentric viewing
Journal of Biomedical Optics 10(3), 03400 (May/June 005) Off-axis wave front measurements for optical correction in eccentric viewing Linda Lundström Peter Unsbo Royal Institute of Technology Biomedical
More informationPopulation study of the variation in monochromatic aberrations of the normal human eye over the central visual field
Population study of the variation in monochromatic aberrations of the normal human eye over the central visual field Matthew T. Sheehan a, Alexander V. Goncharov b, Veronica M. O Dwyer a, Vincent Toal
More informationAberrations and adaptive optics for biomedical microscopes
Aberrations and adaptive optics for biomedical microscopes Martin Booth Department of Engineering Science And Centre for Neural Circuits and Behaviour University of Oxford Outline Rays, wave fronts and
More informationCorneal Asphericity and Retinal Image Quality: A Case Study and Simulations
Corneal Asphericity and Retinal Image Quality: A Case Study and Simulations Seema Somani PhD, Ashley Tuan OD, PhD, and Dimitri Chernyak PhD VISX Incorporated, 3400 Central Express Way, Santa Clara, CA
More informationCone spacing and waveguide properties from cone directionality measurements
S. Marcos and S. A. Burns Vol. 16, No. 5/May 1999/J. Opt. Soc. Am. A 995 Cone spacing and waveguide properties from cone directionality measurements Susana Marcos and Stephen A. Burns Schepens Eye Research
More informationOPTOMETRY REVIEW. New methods and techniques for sensing the wave aberrations of human eyes
C L I N I C A L A N D E X P E R I M E N T A L OPTOMETRY REVIEW New methods and techniques for sensing the wave aberrations of human eyes Clin Exp Optom 2009; 92: 3: 176 186 Marco Lombardo MD PhD Giuseppe
More informationAY122A - Adaptive Optics Lab
AY122A - Adaptive Optics Lab Purpose In this lab, after an introduction to turbulence and adaptive optics for astronomy, you will get to experiment first hand the three main components of an adaptive optics
More informationOptical transfer function shaping and depth of focus by using a phase only filter
Optical transfer function shaping and depth of focus by using a phase only filter Dina Elkind, Zeev Zalevsky, Uriel Levy, and David Mendlovic The design of a desired optical transfer function OTF is a
More informationFor the first time in history, it is possible to clinically
update/review Optics of aberroscopy and super vision Raymond A. Applegate, OD, PhD, Larry N. Thibos, PhD, Gene Hilmantel, OD, MS ABSTRACT This paper (1) reviews the fundamental limits to visual performance
More informationRefractive surgery and other high-tech methods
The Prospects for Perfect Vision Larry N. Thibos, PhD Refractive surgery and other high-tech methods for correcting the optical aberrations of the eye aim to make the eye optically perfect. The notion
More informationWhat is Wavefront Aberration? Custom Contact Lenses For Vision Improvement Are They Feasible In A Disposable World?
Custom Contact Lenses For Vision Improvement Are They Feasible In A Disposable World? Ian Cox, BOptom, PhD, FAAO Distinguished Research Fellow Bausch & Lomb, Rochester, NY Acknowledgements Center for Visual
More informationCoherent imaging of the cone mosaic in the living human eye
Marcos et al. Vol. 13, No. 5/May 1996/J. Opt. Soc. Am. A 897 Coherent imaging of the cone mosaic in the living human eye Susana Marcos and Rafael Navarro Instituto de Óptica, Consejo Superior de Investigaciones
More informationIAC-08-C1.8.5 OPTICAL BEAM CONTROL FOR IMAGING SPACECRAFT WITH LARGE APERTURES
IAC-08-C1.8.5 OPTICAL BEAM CONTROL FOR IMAGING SPACECRAFT WITH LARGE APERTURES Jae Jun Kim Research Assistant Professor, jki1@nps.edu Anne Marie Johnson NRC Research Associate, ajohnson@nps.edu Brij N.
More informationAccommodation with higher-order monochromatic aberrations corrected with adaptive optics
Chen et al. Vol. 23, No. 1/ January 2006/ J. Opt. Soc. Am. A 1 Accommodation with higher-order monochromatic aberrations corrected with adaptive optics Li Chen Center for Visual Science, University of
More informationCharacterizing the Wave Aberration in Eyes with Keratoconus or Penetrating Keratoplasty Using a High Dynamic Range Wavefront Sensor
Characterizing the Wave Aberration in Eyes with Keratoconus or Penetrating Keratoplasty Using a High Dynamic Range Wavefront Sensor Seth Pantanelli, MS, 1,2 Scott MacRae, MD, 3 Tae Moon Jeong, PhD, 2 Geunyoung
More informationFinite conjugate spherical aberration compensation in high numerical-aperture optical disc readout
Finite conjugate spherical aberration compensation in high numerical-aperture optical disc readout Sjoerd Stallinga Spherical aberration arising from deviations of the thickness of an optical disc substrate
More informationCLINICAL SCIENCES. Corneal Optical Aberrations and Retinal Image Quality in Patients in Whom Monofocal Intraocular Lenses Were Implanted
CLINICAL SCIENCES Corneal Optical Aberrations and Retinal Image Quality in Patients in Whom Monofocal Intraocular Lenses Antonio Guirao, PhD; Manuel Redondo, PhD; Edward Geraghty; Patricia Piers; Sverker
More informationProposed Adaptive Optics system for Vainu Bappu Telescope
Proposed Adaptive Optics system for Vainu Bappu Telescope Essential requirements of an adaptive optics system Adaptive Optics is a real time wave front error measurement and correction system The essential
More informationAdaptive Optics for Vision Science. Principles, Practices, Design, and Applications
Adaptive Optics for Vision Science Principles, Practices, Design, and Applications Edited by JASON PORTER, HOPE M. QUEENER, JULIANNA E. LIN, KAREN THORN, AND ABDUL AWWAL m WILEY- INTERSCIENCE A JOHN WILEY
More informationIndustrial quality control HASO for ensuring the quality of NIR optical components
Industrial quality control HASO for ensuring the quality of NIR optical components In the sector of industrial detection, the ability to massproduce reliable, high-quality optical components is synonymous
More informationOptical Quality of the Eye in Subjects with Normal and Excellent Visual Acuity METHODS. Subjects
Optical Quality of the ye in Subjects with Normal and xcellent Visual Acuity loy A. Villegas, ncarna Alcón, and Pablo Artal From the Laboratorio de Optica, Departamento de Fisica, Universidad de Murcia,
More informationStudy of self-interference incoherent digital holography for the application of retinal imaging
Study of self-interference incoherent digital holography for the application of retinal imaging Jisoo Hong and Myung K. Kim Department of Physics, University of South Florida, Tampa, FL, US 33620 ABSTRACT
More informationEffect of optical correction and remaining aberrations on peripheral resolution acuity in the human eye
Effect of optical correction and remaining aberrations on peripheral resolution acuity in the human eye Linda Lundström 1*, Silvestre Manzanera 2, Pedro M. Prieto 2, Diego B. Ayala 2, Nicolas Gorceix 2,
More informationVision Research 50 (2010) Contents lists available at ScienceDirect. Vision Research. journal homepage:
Vision Research 5 (2) 28 24 Contents lists available at ScienceDirect Vision Research journal homepage: www.elsevier.com/locate/visres Combining coma with astigmatism can improve retinal image over astigmatism
More informationDigital Wavefront Sensors Measure Aberrations in Eyes
Contact: Igor Lyuboshenko contact@phaseview.com Internet: www.phaseview.com Digital Measure Aberrations in Eyes 1 in Ophthalmology...2 2 Analogue...3 3 Digital...5 Figures: Figure 1. Major technology nodes
More informationAdaptive optics with a programmable phase modulator: applications in the human eye
Adaptive optics with a programmable phase modulator: applications in the human eye Pedro M. Prieto, Enrique J. Fernández, Silvestre Manzanera, Pablo Artal Laboratorio de Optica, Universidad de Murcia,
More informationOPTINO. SpotOptics VERSATILE WAVEFRONT SENSOR O P T I N O
Spotptics he software people for optics VERSALE WAVEFR SESR Accurate metrology in single and double pass Lenses, mirrors and laser beams Any focal length and diameter Large dynamic range Adaptable for
More informationFast scanning peripheral wave-front sensor for the human eye
Fast scanning peripheral wave-front sensor for the human eye Bart Jaeken, 1,* Linda Lundström, 2 and Pablo Artal 1 1 Laboratorio de Óptica, Universidad de Murcia, Campus Espinardo (Ed. CiOyN), Murcia,
More informationPablo Artal. Adaptive Optics visual simulator ( and depth of focus) LABORATORIO DE OPTICA UNIVERSIDAD DE MURCIA, SPAIN
Adaptive Optics visual simulator ( and depth of focus) Pablo Artal LABORATORIO DE OPTICA UNIVERSIDAD DE MURCIA, SPAIN 8th International Wavefront Congress, Santa Fe, USA, February New LO UM building! Diego
More informationEE119 Introduction to Optical Engineering Fall 2009 Final Exam. Name:
EE119 Introduction to Optical Engineering Fall 2009 Final Exam Name: SID: CLOSED BOOK. THREE 8 1/2 X 11 SHEETS OF NOTES, AND SCIENTIFIC POCKET CALCULATOR PERMITTED. TIME ALLOTTED: 180 MINUTES Fundamental
More informationBias errors in PIV: the pixel locking effect revisited.
Bias errors in PIV: the pixel locking effect revisited. E.F.J. Overmars 1, N.G.W. Warncke, C. Poelma and J. Westerweel 1: Laboratory for Aero & Hydrodynamics, University of Technology, Delft, The Netherlands,
More informationFigure 7 Dynamic range expansion of Shack- Hartmann sensor using a spatial-light modulator
Figure 4 Advantage of having smaller focal spot on CCD with super-fine pixels: Larger focal point compromises the sensitivity, spatial resolution, and accuracy. Figure 1 Typical microlens array for Shack-Hartmann
More informationThe Effect of Phenylephrine and Cyclopentolate on Objective Wavefront Measurements
The Effect of Phenylephrine and Cyclopentolate on Objective Wavefront Measurements Mirko R. Jankov II, MD; Hans Peter Iseli, MD; Michael Bueeler, PhD; Paulo Schor, MD, PhD; Theo Seiler, MD, PhD; Michael
More informationEstimation of centroid positions with a matched-filter algorithm: relevance for aberrometry of the eye
Estimation of centroid positions with a matched-filter algorithm: relevance for aberrometry of the eye C. Leroux and C. Dainty Applied Optics Group, School of Physics, National University of Ireland, Galway
More informationOptical Design of. Microscopes. George H. Seward. Tutorial Texts in Optical Engineering Volume TT88. SPIE PRESS Bellingham, Washington USA
Optical Design of Microscopes George H. Seward Tutorial Texts in Optical Engineering Volume TT88 SPIE PRESS Bellingham, Washington USA Preface xiii Chapter 1 Optical Design Concepts /1 1.1 A Value Proposition
More informationSimultaneously measuring ocular aberration and anterior segment biometry during accommodation
Journal of Innovative Optical Health Sciences Vol. 8, No. 2 (2015) 1550005 (6 pages) #.c The Authors DOI: 10.1142/S1793545815500054 Simultaneously measuring ocular aberration and anterior segment biometry
More informationAberrations and Visual Performance: Part I: How aberrations affect vision
Aberrations and Visual Performance: Part I: How aberrations affect vision Raymond A. Applegate, OD, Ph.D. Professor and Borish Chair of Optometry University of Houston Houston, TX, USA Aspects of this
More informationPROCEEDINGS OF SPIE. Measurement of low-order aberrations with an autostigmatic microscope
PROCEEDINGS OF SPIE SPIEDigitalLibrary.org/conference-proceedings-of-spie Measurement of low-order aberrations with an autostigmatic microscope William P. Kuhn Measurement of low-order aberrations with
More informationPERSPECTIVE THE PRESENCE OF OPTICAL ABERRATIONS THAT BLUR. Making Sense Out of Wavefront Sensing
PERSPECTIVE Making Sense Out of Wavefront Sensing JAY S. PEPOSE, MD, PHD AND RAYMOND A. APPLEGATE, OD, PHD THE PRESENCE OF OPTICAL ABERRATIONS THAT BLUR retinal images were the subject of popular lectures
More informationCardinal Points of an Optical System--and Other Basic Facts
Cardinal Points of an Optical System--and Other Basic Facts The fundamental feature of any optical system is the aperture stop. Thus, the most fundamental optical system is the pinhole camera. The image
More informationRequirements for discrete actuator and segmented wavefront correctors for aberration compensation in two large populations of human eyes
Requirements for discrete actuator and segmented wavefront correctors for aberration compensation in two large populations of human eyes Nathan Doble, 1,2, * Donald T. Miller, 3 Geunyoung Yoon, 4 and David
More informationPuntino. Shack-Hartmann wavefront sensor for optimizing telescopes. The software people for optics
Puntino Shack-Hartmann wavefront sensor for optimizing telescopes 1 1. Optimize telescope performance with a powerful set of tools A finely tuned telescope is the key to obtaining deep, high-quality astronomical
More informationBeam Profiling. Introduction. What is Beam Profiling? by Michael Scaggs. Haas Laser Technologies, Inc.
Beam Profiling by Michael Scaggs Haas Laser Technologies, Inc. Introduction Lasers are ubiquitous in industry today. Carbon Dioxide, Nd:YAG, Excimer and Fiber lasers are used in many industries and a myriad
More informationRepresentation of Wavefront Aberrations
1 4th Wavefront Congress - San Francisco - February 2003 Representation of Wavefront Aberrations Larry N. Thibos School of Optometry, Indiana University, Bloomington, IN 47405 thibos@indiana.edu http://research.opt.indiana.edu/library/wavefronts/index.htm
More information