Annual Report on CCD Imaging at the OAN-SPM 2007
|
|
- Amos Jackson
- 5 years ago
- Views:
Transcription
1 Annual Report on CCD Imaging at the OAN-SPM 2007 Michael Richer & Alan Watson November Introduction This is a report on the state of CCDs and small telescopes of the OAN-SPM. It is based on measurements performed between 26 and 31 October We have published similar reports every year since 2000 [1, 2, 3, 4, 5, 6, 7]. We are grateful to Benjamín García and Fernando Quirós for their help during this run, for their efficiency, and good humor in the face of so many instrument changes. 2 Measurements 2.1 CCD Characteristics We characterized the SITe1, SITe3, and Thomson 2k CCDs. We were unable to characterize the Marconi as it had to be sent to Ensenada for repairs. Table 1 shows the gain and read noise for each CCD. We report these figures for both gain modes for all of the CCDs. We measured read noises and gains using pairs of bias and flat field exposures and the findgain task in IRAF. The read noises are effective read noises, and as such include contributions from amplifier read noise, quantization noise, and spurious charge. Table 2 shows the linearity for each CCD determined from graduated flats. Fig. 1 presents the results graphically. As a measure of non-linearity, we fit a line to the ratio of observed to expected signals as a function of signal level for signals less than ADU. We then multiply the slope of this line by to obtain an estimate of the non-linearity of the detector over its full dynamic range. The non-linearities measured in this way are typically of order 1% or less, except for gain mode 1 with the Thomson 2k CCD, which consistently presents a slight nonlinearity. The single linearity test of the SITe1 at the 1.5-m telescope in gain mode 4 apparently resulted in a nearly 2% nonlinearity. However, it was linear in this gain mode at the 84-cm during at least three tests, so we suspect that the nonlinearity at the 1.5-m is spurious. 1
2 Compared to previous years, the linearity tests contain more scatter about unity. We found this effect last year for the Marconi CCD and attributed it to gain variations. This year, however, it appears in many of the tests. Fig. 2 compares linearity tests based upon long and short exposures. There is much less scatter in the test based upon long exposures. Perhaps, part of the scatter in the other tests is a result of lamp variations on short time scales that the monitor exposures are unable to remove. Another possibility is that shutter errors are contributing to some of this scatter (see Section 2.2). Table 3 shows the CCD operating temperatures and dark currents (per physical pixel). In the past, the temperature reading for the SITe1 controller was unreliable, but this has perhaps now been corrected. During our tests at both the 84-cm and 1.5-m telescopes (in that order), the temperature was stable at the first value indicated in Table 3. Subsequently, after the CCD was moved back to the 84-cm telescope for an observing run, the temperature was unstable and so was raised to the second value presented, resulting in a higher dark current. The difference in dark current at the two temperature settings is reasonable given the difference in the temperature settings. (Typically, the dark current doubles for a temperature change of 6-8 K.) Figure 3 shows the dark current patterns for the SITe3 CCD. This pattern is visible in images when the background is very faint. Our experience indicates that resetting the CCD controller or running the PMIS setup macro has no effect on this (as expected if it is dark current). Table 4 shows the read times (the time to read a full frame to the CCD control computer). For the SITe1, SITe3, and Thomson CCDs, two control interfaces are available, PMIS and the new Linux graphical interface. We were able to test the SITe1 with both interfaces at the 84-cm telescope, but only with the Linux graphical interface at the 1.5-m telescope. At the 84-cm telescope, the read times were shorter with the PMIS interface (these are the ones presented in Table 4). Oddly, the read times with the Linux graphical interface were significantly longer at the 1.5-m telescope than at the 84-cm telescope. Figure 4 presents the bit frequencies for the CCDs tested. These bit frequencies are derived from well-exposed flat field images (> 40,000 ADU). Given the number of counts, the lower bits should sample pure Poisson noise while the higher bits reflect the structure in the image. To illustrate, if the signal level exceeds 40,000 ADU, the Poisson noise will exceed 200/ gain ADU, a value that falls between 100 and 200 ADU for the gains available. Since these values correspond approximately to 2 7 = 128 and 2 8 = 256, equal numbers of bits 0 to 7 or 8 should have values of zero and one for a perfect analog-to-digital converter (ADC). In Fig. 4, the fraction of bits with a value of one is plotted as a function of the bit number. The frequency of 1 s for the lower bits is within 0.6% of the result expected for a perfect ADC for all of the CCD controllers. 2.2 Shutter Errors Figure 5 and Table 5 show the shutter error (the additional exposure time above that requested). These were calculated by comparing long and short flat-field exposures. The variations are as expected for five bladed iris shutters with travel times of ms. However, there are significant pedestals, which are presumably the result of timing 2
3 Table 1: CCD Electronic Characteristics CCD Telescope Mode Binning Gain Read noise Bias SITe1 84-cm e 29.5 e e e e e 8.7 e e e e 624 SITe1 1.5-m e 16.8 e e e e e 7.8 e e e e 660 SITe3 1.5-m e 10.2 e e e e e 12.4 e e e e 1462 SITe3 84-cm e 9.5 e e e e e 10.6 e e e e 1463 Thomson 84-cm e 6.0 e e e e e 5.5 e e e e 412 3
4 84-cm 1.5-m observed signal/expected signal observed signal/expected signal signal signal (ADU) 84-cm 1.5-m observed signal/expected signal observed signal/expected signal signal signal Thomson 84-cm 1.05 observed signal/expected signal signal Figure 1: A comparison of the observed/expected count rates for all of the CCDs. 4
5 Table 2: CCD Linearity CCD Mode Non-linearity SITe1 1 < 1% 4 < 1% SITe3 1 < 1% 4 < 1% Thomson 1 1.3% 4 < 1% Table 3: CCD Operating Temperatures and Dark Currents CCD Temperature Dark Current SITe1 a 99.1 C 1.9 e /pix/h SITe1 b 88.5 C 7.2 e /pix/h SITe C 4.3 e /pix/h Thomson 95.2 C 1.0 e /pix/h SITe 84-cm, observed signal/expected signal , long exposures, short exposures signal level (ADU) Figure 2: A comparison of linearity tests based upon long and short exposures. 5
6 Figure 3: The dark current pattern for the SITe3 CCD. This pattern is visible in images with very faint backgrounds (e.g., spectroscopy) or especially when used with 2 2 binning. The dark current for the SITe1 and Thomson 2k CCDs appears uniform. 6
7 84-cm 1.5-m frequency of 1's fraction of 1's bit 0 bit 1 bit 2 bit 3 bit 4 bit 5 bit 6 bit 7 bit 8 bit 9 bit 10 bit 11 bit 12 bit 13 bit 14 bit bit 0 bit 1 bit 2 bit 3 bit 4 bit 5 bit 6 bit 7 bit 8 bit 9 bit 10 bit 11 bit 12 bit 13 bit 14 bit 15 bit number bits 84-cm 1.5-m fraction of 1's fraction of 1's bit 0 bit 1 bit 2 bit 3 bit 4 bit 5 bit 6 bit 7 bit 8 bit 9 bit 10 bit 11 bit 12 bit 13 bit 14 bit 15 bit 0 bit 1 bit 2 bit 3 bit 4 bit 5 bit 6 bit 7 bit 8 bit 9 bit 10 bit 11 bit 12 bit 13 bit 14 bit 15 bit bits Thomson 84-cm fraction of 1's bit 0 bit 1 bit 2 bit 3 bit 4 bit 5 bit 6 bit 7 bit 8 bit 9 bit 10 bit 11 bit 12 bit 13 bit 14 bit 15 bit Figure 4: The frequency of 1 s as a function of the bit number. A perfect ADC would have a frequency of 0.5 for bits 0 to 7 or 8. All of the CCD controllers are within 0.6% this value. 7
8 Table 4: CCD Read Times CCD Gain mode Binning Read Time Read Time Linux PMIS SITe s 10 s s 5 s s 4 s s 4 s s 18 s s 7 s s 5 s s 5 s SITe s 31 s s 11 s s 8 s s 6 s s 58 s s 18 s s 10 s s 8 s Thomson s s s s s s s s Table 5: CCD Shutter Errors CCD Telescope Center Edge SITe1 84-cm +66 ms +40 ms SITe1 1.5-m +11 ms -5 ms SITe3 84-cm +61 ms +40 ms SITe3 1.5-m +40 ms +18 ms Thomson 84-cm +73 ms +38 ms 8
9 Figure 5: The shutter error (median-smoothed) for the SITe1 CCD at the 1.5-m telescope. The scale is from -10 ms to +40 ms. 9
10 errors on the part of the CCD control system. In the past, we have found that these pedestals appear to depend upon where the data are acquired for the Marconi, SITe1, and Thomson CCDs [7, 6]. This year, we have found variable pedestals for a given CCD at a given telescope on a given night. The values given in Table 5 should be interpreted as typical values, not absolute values. To minimize problems in photometry, exposures of at least 10 seconds are recommended. 2.3 Fringes In previous years we have investigated this issue [5, 9], and there is nothing new to report. Briefly, the Thomson suffers from severe fringing in I and in narrowband filters longward of 6000 Å and moderate fringing in R. The Marconi and SITe3 suffer from fringing in I as well as in narrowband filters longwards of 6000 Å, but of much lower amplitude than the Thomson, and with the SITe3 being the least affected of the two. We have never found evidence that the SITe1 suffers from fringing. 2.4 Zero Points Table 6 shows the zero points for various broadband filters. The zero point is the number of electrons per second expected from a star with a Vega-mag of 0 at 1 airmass and zero colour. 2.5 UBVRI filter sets See the 2005 report [6] for a comparison of the relative efficiencies of the three filter sets. The 2006 report [7] discusses colour terms for broad-band photometry. 3 Comments 3.1 Telescope Reflectivities Figures 6 and 7 show the efficiencies of the 1.5-m and 84-cm telescopes, respectively, for The reflectivity of the 1.5-m telescope is similar to that found last year (when it was last aluminized). The reflectivity of the 84-cm telescope, however, is lower than was measured last year by 20-25%. This telescope was also aluminized last year. Clearly, it would be profitable to re-aluminize the 84-cm telescope at the first opportunity. 3.2 SITe1 CCD The electronics associated with this CCD were damaged by a lightning strike at the beginning of September 2007 and only returned to the observatory just before our run. 10
11 In gain mode 1, the readnoise measured this year is substantially higher than that measured previously. In gain mode 4, however, the gain, readnoise, and spurious charge are similar to values previously found. The readnoise in both gain modes is lower at the 1.5-m than at the 84-cm telescope. The temperature reading is now stable and a colder temperature is maintained. The dark current is lower and its pattern is now uniform. We recommend using this detector in gain mode 4 only. 3.3 SITe3 CCD The dark current is higher than we measured last year [6], though this could easily arise as a result of being operated at a higher temperature. Otherwise, its electronic characteristics are similar to those measured last year. 3.4 Thomson 2k CCD It has been two years since we evaluated this CCD. Nothing seems to have changed. It has the best electronic characteristics of any of the CCDs, but is slow to read out, suffers strong fringing in the red, and is less sensitive than the other CCDs in U and I. 3.5 Shutter Errors These shutter errors have important implications for standard stars. If you place a star at the center of the CCD and ask for a 1 second exposure, the actual exposure time will exceed this by 3 7% (Table 5), meaning that your photometry will also be systematically wrong by this same amount. To get 1% photometry, we recommend taking standard star exposures that are no shorter than 10 seconds at the 84-cm and 6 seconds at the 1.5-m. However, it is probably worthwhile deriving the shutter shading on each run; simply compare the difference between a single 10-second dome flat and ten 1-second dome flats. 3.6 The Tcl/Tk graphical interface This year, the graphical user interface was available for all of the CCDs and was used with the SITe1 and SITe3 CCDs. As Table 4 indicates, this interface reads these CCDs more slowly than does the PMIS interface. We found that the read times with the graphical interface were longer at the 1.5-m than at the 84-cm. Also, when used with the SITe1 CCD, the first 10 physical columns are apparently discarded, since the overscan region begins in columns 1015, 508, 339, and 254 with binnings 1 1, 2 2, 3 3, and 4 4, respectively. This effect was seen at both the 84-cm and 1.5-m telescopes with the SITe1 CCD, but not with the SITe3 CCD at the 1.5-m. 11
12 m 1.20 efficiency/efficiency(2006) U2 B2 V2 R2 I2 filter Figure 6: Relative efficiency of the 1.5-m telescope for the past few years, based upon the zero points from the SITe1 calibrations ( ). All data are normalized to those from 2006, which is when the telescope was most recently aluminized. The mirrors were also aluminized in Table 6: Zero Points Telescope Filter SITe1 SITe3 Thomson 2k 84-cm U B V R I m U B V R I
13 84-cm 1.2 efficiency/efficiency(2006) U3 B3 V3 R3 I3 filter Figure 7: Relative efficiency of the 84-cm telescope for the past few years, based upon the zero points from the Marconi (2006, 2005) and SITe1 (2007, 2005, 2004) calibrations. All data are normalized to those from 2006, when he mirrors were last aluminized. The mirrors were also aluminized in
14 References [1] Reporte de la temporada de ingeniería del de septiembre de 2000, Michael Richer, Alan Watson & Henri Plana, 20 September [2] Las respuestas fotométricas de los CCDs, Michael Richer & Alan Watson, 11 February [3] Reporte de caracterización de los CCDs y de los telescopios de 1.5m y 84cm, Michael Richer, Alan Watson, Simon Kemp, & Sandra Ayala, 9 January [4] Reporte de caracterización de los CCDs y de los telescopios de 1.5m y 84cm, Sandra Ayala, Michael Richer, Alan Watson, & Simon Kemp, 6 January [5] Annual Report on CCD Imaging at the OAN, Alan Watson, Michael Richer, Arturo Godínez, & Teresa García, August [6] Annual Report on CCD Imaging at the OAN 2005, Michael Richer, Alan Watson, Gabriela Montes, & Georgina Bénitez de la Mora, August [7] Annual Report on CCD Imaging at the OAN 2006, Alan Watson & Michael Richer, September [8] CCD-OAN-E2V. Programa para la adquisición de las imágenes de un CCD, Enrique Colorado & Leonel Gutiérrez, December 2004 [9] Franjas de interferencia con el CCD Marconi, Michael Richer, March
CCD reductions techniques
CCD reductions techniques Origin of noise Noise: whatever phenomena that increase the uncertainty or error of a signal Origin of noises: 1. Poisson fluctuation in counting photons (shot noise) 2. Pixel-pixel
More informationProperties of a Detector
Properties of a Detector Quantum Efficiency fraction of photons detected wavelength and spatially dependent Dynamic Range difference between lowest and highest measurable flux Linearity detection rate
More informationPadova and Asiago Observatories
ISSN 1594-1906 Padova and Asiago Observatories The Echelle E2V CCD47-10 CCD H. Navasardyan, M. D'Alessandro, E. Giro, Technical Report n. 22 September 2004 Document available at: http://www.pd.astro.it/
More informationThe 0.84 m Telescope OAN/SPM - BC, Mexico
The 0.84 m Telescope OAN/SPM - BC, Mexico Readout error CCD zero-level (bias) ramping CCD bias frame banding Shutter failure Significant dark current Image malting Focus frame taken during twilight IR
More informationAstronomy 341 Fall 2012 Observational Astronomy Haverford College. CCD Terminology
CCD Terminology Read noise An unavoidable pixel-to-pixel fluctuation in the number of electrons per pixel that occurs during chip readout. Typical values for read noise are ~ 10 or fewer electrons per
More information"Internet Telescope" Performance Requirements
"Internet Telescope" Performance Requirements by Dr. Frank Melsheimer DFM Engineering, Inc. 1035 Delaware Avenue Longmont, Colorado 80501 phone 303-678-8143 fax 303-772-9411 www.dfmengineering.com Table
More informationWFC3 TV2 Testing: UVIS Shutter Stability and Accuracy
Instrument Science Report WFC3 2007-17 WFC3 TV2 Testing: UVIS Shutter Stability and Accuracy B. Hilbert 15 August 2007 ABSTRACT Images taken during WFC3's Thermal Vacuum 2 (TV2) testing have been used
More informationTemperature Reductions to Mitigate the WF4 Anomaly
Instrument Science Report WFPC2 2007-01 Temperature Reductions to Mitigate the WF4 Anomaly V. Dixon, J. Biretta, S. Gonzaga, and M. McMaster April 18, 2007 ABSTRACT The WF4 anomaly is characterized by
More informationMAOP-702. CCD 47 Characterization
Doc # : MAOP702 Date: 2013Apr03 Page: 1 of 14 MAOP702 Prepared By: Name(s) and Signature(s) Date Jared R. Males Approved By Name and Signature Title Laird Close PI Victor Gasho Program Manager Date Revision
More informationCerro Tololo Inter-American Observatory. CHIRON manual. A. Tokovinin Version 2. May 25, 2011 (manual.pdf)
Cerro Tololo Inter-American Observatory CHIRON manual A. Tokovinin Version 2. May 25, 2011 (manual.pdf) 1 1 Overview Calibration lamps Quartz, Th Ar Fiber Prism Starlight GAM mirror Fiber Viewer FEM Guider
More informationFLAT FIELD DETERMINATIONS USING AN ISOLATED POINT SOURCE
Instrument Science Report ACS 2015-07 FLAT FIELD DETERMINATIONS USING AN ISOLATED POINT SOURCE R. C. Bohlin and Norman Grogin 2015 August ABSTRACT The traditional method of measuring ACS flat fields (FF)
More informationObservation Data. Optical Images
Data Analysis Introduction Optical Imaging Tsuyoshi Terai Subaru Telescope Imaging Observation Measure the light from celestial objects and understand their physics Take images of objects with a specific
More informationAperture Photometry with CCD Images using IRAF. Kevin Krisciunas
Aperture Photometry with CCD Images using IRAF Kevin Krisciunas Images must be taken in a sensible manner. Ask advice from experienced observers. But remember Wallerstein s Rule: Four astronomers, five
More informationCharge Coupled Devices. C. A. Griffith, Class Notes, PTYS 521, 2016 Not for distribution.
Charge Coupled Devices C. A. Griffith, Class Notes, PTYS 521, 2016 Not for distribution. 1 1. Introduction While telescopes are able to gather more light from a distance source than does the naked eye,
More informationSpectral Analysis of the LUND/DMI Earthshine Telescope and Filters
Spectral Analysis of the LUND/DMI Earthshine Telescope and Filters 12 August 2011-08-12 Ahmad Darudi & Rodrigo Badínez A1 1. Spectral Analysis of the telescope and Filters This section reports the characterization
More informationOptical Photometry. The crash course Tomas Dahlen
The crash course Tomas Dahlen Aim: Measure the luminosity of your objects in broad band optical filters Optical: Wave lengths about 3500Å 9000Å Typical broad band filters: U,B,V,R,I Software: IRAF & SExtractor
More informationObserving*Checklist:*A3ernoon*
Ay#122a:# Intro#to#Observing/Image#Processing# (Many&slides&today& c/o&m.&bolte)& Observing*Checklist:*A3ernoon* Set*up*instrument*(verify*and*set*filters,*gra@ngs,*etc.)* Set*up*detector*(format,*gain,*binning)*
More informationWFC3 Post-Flash Calibration
Instrument Science Report WFC3 2013-12 WFC3 Post-Flash Calibration J. Biretta and S. Baggett June 27, 2013 ABSTRACT We review the Phase II implementation of the WFC3/UVIS post-flash capability, as well
More informationStellar Photometry: I. Measuring. Ast 401/Phy 580 Fall 2014
What s Left (Today): Introduction to Photometry Nov 10 Photometry I/Spectra I Nov 12 Spectra II Nov 17 Guest lecture on IR by Trilling Nov 19 Radio lecture by Hunter Nov 24 Canceled Nov 26 Thanksgiving
More informationWhat an Observational Astronomer needs to know!
What an Observational Astronomer needs to know! IRAF:Photometry D. Hatzidimitriou Masters course on Methods of Observations and Analysis in Astronomy Basic concepts Counts how are they related to the actual
More informationCross-Talk in the ACS WFC Detectors. II: Using GAIN=2 to Minimize the Effect
Cross-Talk in the ACS WFC Detectors. II: Using GAIN=2 to Minimize the Effect Mauro Giavalisco August 10, 2004 ABSTRACT Cross talk is observed in images taken with ACS WFC between the four CCD quadrants
More informationThe Observation Summary of South Galactic Cap U band Sky Survey (SGCUSS)
The Observation Summary of South Galactic Cap U band Sky Survey (SGCUSS) Zhen-Yu Wu, Xu Zhou and Zhou Fan 1. process September: In this month, Xu, Zhou, Zhen-Yu, Mike, and ED went to Bok telescope and
More informationErrata to First Printing 1 2nd Edition of of The Handbook of Astronomical Image Processing
Errata to First Printing 1 nd Edition of of The Handbook of Astronomical Image Processing 1. Page 47: In nd line of paragraph. Following Equ..17, change 4 to 14. Text should read as follows: The dark frame
More informationPixel Response Effects on CCD Camera Gain Calibration
1 of 7 1/21/2014 3:03 PM HO M E P R O D UC T S B R IE F S T E C H NO T E S S UP P O RT P UR C HA S E NE W S W E B T O O L S INF O C O NTA C T Pixel Response Effects on CCD Camera Gain Calibration Copyright
More informationNonlinearity in the Detector used in the Subaru Telescope High Dispersion Spectrograph
Nonlinearity in the Detector used in the Subaru Telescope High Dispersion Spectrograph Akito Tajitsu Subaru Telescope, National Astronomical Observatory of Japan, 650 North A ohoku Place, Hilo, HI 96720,
More informationThe Noise about Noise
The Noise about Noise I have found that few topics in astrophotography cause as much confusion as noise and proper exposure. In this column I will attempt to present some of the theory that goes into determining
More informationTemperature Dependent Dark Reference Files: Linear Dark and Amplifier Glow Components
Instrument Science Report NICMOS 2009-002 Temperature Dependent Dark Reference Files: Linear Dark and Amplifier Glow Components Tomas Dahlen, Elizabeth Barker, Eddie Bergeron, Denise Smith July 01, 2009
More informationINTRODUCTION TO CCD IMAGING
ASTR 1030 Astronomy Lab 85 Intro to CCD Imaging INTRODUCTION TO CCD IMAGING SYNOPSIS: In this lab we will learn about some of the advantages of CCD cameras for use in astronomy and how to process an image.
More informationWFC3 Thermal Vacuum Testing: UVIS Broadband Flat Fields
WFC3 Thermal Vacuum Testing: UVIS Broadband Flat Fields H. Bushouse June 1, 2005 ABSTRACT During WFC3 thermal-vacuum testing in September and October 2004, a subset of the UVIS20 test procedure, UVIS Flat
More informationHST and JWST Photometric Calibration. Susana Deustua Space Telescope Science Institute
HST and JWST Photometric Calibration Susana Deustua Space Telescope Science Institute Charge On the HST (and JWST) photometric calibrators, in particular the white dwarf standards including concept for
More informationCalibrating VISTA Data
Calibrating VISTA Data IR Camera Astronomy Unit Queen Mary University of London Cambridge Astronomical Survey Unit, Institute of Astronomy, Cambridge Jim Emerson Simon Hodgkin, Peter Bunclark, Mike Irwin,
More informationA Test of non-standard Gain Settings for the NICMOS Detectors
Instrument Science Report NICMOS 23-6 A Test of non-standard Gain Settings for the NICMOS Detectors Chun Xu & Torsten Böker 2 May, 23 ABSTRACT We report on the results of a test program to explore the
More informationTHE CCD RIDDLE REVISTED: SIGNAL VERSUS TIME LINEAR SIGNAL VERSUS VARIANCE NON-LINEAR
THE CCD RIDDLE REVISTED: SIGNAL VERSUS TIME LINEAR SIGNAL VERSUS VARIANCE NON-LINEAR Mark Downing 1, Peter Sinclaire 1. 1 ESO, Karl Schwartzschild Strasse-2, 85748 Munich, Germany. ABSTRACT The photon
More informationMiCPhot: A prime-focus multicolor CCD photometer on the 85-cm Telescope
Research in Astron. Astrophys. 2009 Vol. 9 No. 3, 349 366 http://www.raa-journal.org http://www.iop.org/journals/raa Research in Astronomy and Astrophysics MiCPhot: A prime-focus multicolor CCD photometer
More informationUsing CCDAuto (last update: 06/21/05)
(last update: 06/21/05) (1) Table of Contents I. Overview...3 A. Program...3 B. Observatory...3 i. Specifications...3 ii. Instruments...4 iii. Using the UCI Student Observatory...7 II. Acquiring Calibration
More informationHigh Contrast Imaging using WFC3/IR
SPACE TELESCOPE SCIENCE INSTITUTE Operated for NASA by AURA WFC3 Instrument Science Report 2011-07 High Contrast Imaging using WFC3/IR A. Rajan, R. Soummer, J.B. Hagan, R.L. Gilliland, L. Pueyo February
More informationOmegaCAM calibrations for KiDS
OmegaCAM calibrations for KiDS Gijs Verdoes Kleijn for OmegaCEN & KiDS survey team Kapteyn Astronomical Institute University of Groningen A. Issues common to wide field imaging surveys data processing
More informationWFC3 SMOV Program 11433: IR Internal Flat Field Observations
Instrument Science Report WFC3 2009-42 WFC3 SMOV Program 11433: IR Internal Flat Field Observations B. Hilbert 27 October 2009 ABSTRACT We have analyzed the internal flat field behavior of the WFC3/IR
More informationLow Light Level CCD Performance and Issues
Low Light Level CCD Performance and Issues Nagaraja Bezawada UK Astronomy Technology Centre 04 July 2007 Overview of the Talk Introduction to L3CCD (EM CCD) ULTRASPEC Performance and Issues New L3 CCD
More informationVATTSpec Instructions Rev. 10/23/2015
VATTSpec Instructions Rev. 10/23/2015 Introduction VATTSpec is a medium resolution CCD range spectrograph with a skinny chip having excellent cosmetics. Its UA ITL chip, Serial Number 8228, has a gain
More informationCCD Characteristics Lab
CCD Characteristics Lab Observational Astronomy 6/6/07 1 Introduction In this laboratory exercise, you will be using the Hirsch Observatory s CCD camera, a Santa Barbara Instruments Group (SBIG) ST-8E.
More informationProposal for a research project to be carried. out in Physics 400 (Senior Research) and. IDIS 493 (Honors Thesis).
Proposal for a research project to be carried out in Physics 400 (Senior Research) and IDIS 493 (Honors Thesis). Variable Star CCD Photometry and Analysis Amber L. Stuver Submitted in fulfillment of requirements
More informationThe DECam System: Technical Characteristics
The DECam System: Technical Characteristics Alistair R. Walker DECam Instrument Scientist DECam Community Workshop 1 Contents Status & Statistics A selective look at some DECam & Blanco technical properties
More informationMaster sky images for the WFC3 G102 and G141 grisms
Master sky images for the WFC3 G102 and G141 grisms M. Kümmel, H. Kuntschner, J. R. Walsh, H. Bushouse January 4, 2011 ABSTRACT We have constructed master sky images for the WFC3 near-infrared G102 and
More informationWFC3 Thermal Vacuum Testing: UVIS Science Performance Monitor
WFC3 Thermal Vacuum Testing: UVIS Science Performance Monitor H. Bushouse and O. Lupie May 24, 2005 ABSTRACT During WFC3 thermal-vacuum testing in September and October 2004, the UVIS28 test procedure,
More information4k CCD Observers Software Observers manual for BOAO 4k CCD camera system Byeong-Gon Park KASI Optical Astronomy Division Fri. Oct. 28.
BOAO_4kCCD_SW_001E_20111028 4k CCD Observers Software Observers manual for BOAO 4k CCD camera system Byeong-Gon Park KASI Optical Astronomy Division Fri. Oct. 28. 2011 Byeong-Gon Park email: bgpark@kasi.re.kr
More informationWFC3 SMOV Program 11427: UVIS Channel Shutter Shading
Instrument Science Report WFC3 2009-25 WFC3 SMOV Program 11427: UVIS Channel Shutter Shading B. Hilbert June 23, 2010 ABSTRACT A series of internal flat field images and standard star observations were
More informationSTIS CCD Saturation Effects
SPACE TELESCOPE SCIENCE INSTITUTE Operated for NASA by AURA Instrument Science Report STIS 2015-06 (v1) STIS CCD Saturation Effects Charles R. Proffitt 1 1 Space Telescope Science Institute, Baltimore,
More informationPhotometry of the variable stars using CCD detectors
Contrib. Astron. Obs. Skalnaté Pleso 35, 35 44, (2005) Photometry of the variable stars using CCD detectors I. Photometric reduction. Š. Parimucha 1, M. Vaňko 2 1 Institute of Physics, Faculty of Natural
More informationDBSP Observing Manual
DBSP Observing Manual I. Arcavi, P. Bilgi, N.Blagorodnova, K.Burdge, A.Y.Q.Ho June 18, 2018 Contents 1 Observing Guides 2 2 Before arrival 2 2.1 Submit observing setup..................................
More informationThe. FIES Camera. equipped with E2V CCD B83, ser. no Pre-commissioning characterisation
The FIES Camera equipped with E2V CCD42-40-1-B83, ser. no. 01064-17-04 Pre-commissioning characterisation Anton Norup Sørensen Copenhagen University Observatory October 2003 Contents 1 Introduction 2 2
More informationWFC3/UVIS TV3 Post-flash Results
Technical Instrument Report WFC3 2012-01 WFC3/UVIS TV3 Post-flash Results S. Baggett and T. Wheeler March 29, 2012 Abstract Given recent interest in potentially reviving the WFC3 post-flash capability,
More informationARRAY CONTROLLER REQUIREMENTS
ARRAY CONTROLLER REQUIREMENTS TABLE OF CONTENTS 1 INTRODUCTION...3 1.1 QUANTUM EFFICIENCY (QE)...3 1.2 READ NOISE...3 1.3 DARK CURRENT...3 1.4 BIAS STABILITY...3 1.5 RESIDUAL IMAGE AND PERSISTENCE...4
More informationAstro-photography. Daguerreotype: on a copper plate
AST 1022L Astro-photography 1840-1980s: Photographic plates were astronomers' main imaging tool At right: first ever picture of the full moon, by John William Draper (1840) Daguerreotype: exposure using
More informationPhotometric Calibration for Wide- Area Space Surveillance Sensors
Photometric Calibration for Wide- Area Space Surveillance Sensors J.S. Stuart, E. C. Pearce, R. L. Lambour 2007 US-Russian Space Surveillance Workshop 30-31 October 2007 The work was sponsored by the Department
More informationA Quantix monochrome camera with a Kodak KAF6303E CCD 2-D array was. characterized so that it could be used as a component of a multi-channel visible
A Joint Research Program of The National Gallery of Art, Washington The Museum of Modern Art, New York Rochester Institute of Technology Technical Report March, 2002 Characterization of a Roper Scientific
More informationPhotometry. Variable Star Photometry
Variable Star Photometry Photometry One of the most basic of astronomical analysis is photometry, or the monitoring of the light output of an astronomical object. Many stars, be they in binaries, interacting,
More informationSOAR Integral Field Spectrograph (SIFS): Call for Science Verification Proposals
Published on SOAR (http://www.ctio.noao.edu/soar) Home > SOAR Integral Field Spectrograph (SIFS): Call for Science Verification Proposals SOAR Integral Field Spectrograph (SIFS): Call for Science Verification
More informationa simple optical imager
Imagers and Imaging a simple optical imager Here s one on our 61-Inch Telescope Here s one on our 61-Inch Telescope filter wheel in here dewar preamplifier However, to get a large field we cannot afford
More informationSARG: The Graphical User Interface Manual
1/28 SARG: The Graphical User Interface Manual Document: TNG-SARG-001 Issue: 1.0 Prepared by : Name: S. Scuderi Institute: INAF-Osservatorio Astrofisico di Catania Date : Approved by : Name: R. Cosentino
More informationHANDBOOK OF CCD ASTRONOMY
HANDBOOK OF CCD ASTRONOMY STEVE B. HOWELL ASTROPHYSICS GROUP, PLANETARY SCIENCE INSTITUTE, TUCSON, AND DEPARTMENT OF PHYSICS AND ASTRONOMY, UNIVERSITY OF WYOMING published by the press syndicate of the
More informationPhotometry. La Palma trip 2014 Lecture 2 Prof. S.C. Trager
Photometry La Palma trip 2014 Lecture 2 Prof. S.C. Trager Photometry is the measurement of magnitude from images technically, it s the measurement of light, but astronomers use the above definition these
More informationProperties of CCD #4. installed in dewar no. 2. Operated as HiRAC science camera. Anton Norup Srensen & Michael I. Andersen
Properties of CCD #4 installed in dewar no. 2 Operated as HiRAC science camera. Anton Norup Srensen & Michael I. Andersen Copenhagen University Observatory December 1996 1 Introduction The BroCamII (dewar
More informationSPACE TELESCOPE SCIENCE INSTITUTE Operated for NASA by AURA
SPACE TELESCOPE SCIENCE INSTITUTE Operated for NASA by AURA Instrument Science Report WFC3 2010-08 WFC3 Pixel Area Maps J. S. Kalirai, C. Cox, L. Dressel, A. Fruchter, W. Hack, V. Kozhurina-Platais, and
More informationLecture 5. Telescopes (part II) and Detectors
Lecture 5 Telescopes (part II) and Detectors Please take a moment to remember the crew of STS-107, the space shuttle Columbia, as well as their families. Crew of the Space Shuttle Columbia Lost February
More informationSEAMS DUE TO MULTIPLE OUTPUT CCDS
Seam Correction for Sensors with Multiple Outputs Introduction Image sensor manufacturers are continually working to meet their customers demands for ever-higher frame rates in their cameras. To meet this
More informationX-ray Spectroscopy Laboratory Suresh Sivanandam Dunlap Institute for Astronomy & Astrophysics, University of Toronto
X-ray Spectroscopy Laboratory Suresh Sivanandam, 1 Introduction & Objectives At X-ray, ultraviolet, optical and infrared wavelengths most astronomical instruments employ the photoelectric effect to convert
More informationSink Pixels and CTE in the WFC3/UVIS Detector
Instrument Science Report WFC3 2014-19 Sink Pixels and CTE in the WFC3/UVIS Detector Jay Anderson and Sylvia Baggett June 13, 2014 ABSTRACT Post-flashed calibration products have highlighted a previously
More informationGuide to the Imaging Observation with MOIRCS
Guide to the Imaging Observation with MOIRCS For the Upgraded MOIRCS (since 2016) 2016-07-27 Introduction This document describes how to prepare an OPE file for Imaging Observation by MOIRCS. Please first
More informationWFC3 TV3 Testing: IR Channel Nonlinearity Correction
Instrument Science Report WFC3 2008-39 WFC3 TV3 Testing: IR Channel Nonlinearity Correction B. Hilbert 2 June 2009 ABSTRACT Using data taken during WFC3's Thermal Vacuum 3 (TV3) testing campaign, we have
More informationCamera Test Protocol. Introduction TABLE OF CONTENTS. Camera Test Protocol Technical Note Technical Note
Technical Note CMOS, EMCCD AND CCD CAMERAS FOR LIFE SCIENCES Camera Test Protocol Introduction The detector is one of the most important components of any microscope system. Accurate detector readings
More informationThis release contains deep Y-band images of the UDS field and the extracted source catalogue.
ESO Phase 3 Data Release Description Data Collection HUGS_UDS_Y Release Number 1 Data Provider Adriano Fontana Date 22.09.2014 Abstract HUGS (an acronym for Hawk-I UDS and GOODS Survey) is a ultra deep
More informationASTRONOMICAL STATION VIDOJEVICA: IN SITU TEST OF THE ALTA APOGEE U42 CCD CAMERA
Serb. Astron. J. 185 (2012), 65-74 UDC 520.82 13 DOI: 10.2298/SAJ1285065V Professional paper ASTRONOMICAL STATION VIDOJEVICA: IN SITU TEST OF THE ALTA APOGEE U42 CCD CAMERA O. Vince Astronomical Observatory
More informationControl of Noise and Background in Scientific CMOS Technology
Control of Noise and Background in Scientific CMOS Technology Introduction Scientific CMOS (Complementary metal oxide semiconductor) camera technology has enabled advancement in many areas of microscopy
More informationUse of the Shutter Blade Side A for UVIS Short Exposures
Instrument Science Report WFC3 2014-009 Use of the Shutter Blade Side A for UVIS Short Exposures Kailash Sahu, Sylvia Baggett, J. MacKenty May 07, 2014 ABSTRACT WFC3 UVIS uses a shutter blade with two
More informationModel ST-9XE CCD Imaging Camera SBIG ASTRONOMICAL INSTRUMENTS
Model ST-9XE CCD Imaging Camera.. SBIG ASTRONOMICAL INSTRUMENTS 1... Model ST-9XE Dual CCD Self-Guiding Camera The ST-9XE is identical to the ST-7/8/10/2000 cameras with the exception of the imaging CCD.
More informationA Stony Brook Student s Guide to Using CCDSoft By Stephanie Zajac Last Updated: 3 February 2012
A Stony Brook Student s Guide to Using CCDSoft By Stephanie Zajac Last Updated: 3 February 2012 This document is meant to serve as a quick start guide to using CCDSoft to take data using the Mt. Stony
More informationWEBCAMS UNDER THE SPOTLIGHT
WEBCAMS UNDER THE SPOTLIGHT MEASURING THE KEY PERFORMANCE CHARACTERISTICS OF A WEBCAM BASED IMAGER Robin Leadbeater Q-2006 If a camera is going to be used for scientific measurements, it is important to
More informationPhase-2 Preparation Tool
Gran Telescopio Canarias Phase-2 Preparation Tool Valid from period 2012A Updated: 6 March 2012 1 Contents 1. The GTC Phase-2 System... 3 1.1. Introduction... 3 1.2. Logging in... 3 2. Defining an observing
More informationRHO CCD. imaging and observa3on notes AST aug 2011
RHO CCD imaging and observa3on notes AST 6725 30 aug 2011 Camera Specs & Info 76 cm Telescope f/4 Prime focus (3.04 m focal length) SBIG ST- 8XME CCD Camera Kodak KAF- 1603ME Class 2 imaging CCD Built-
More informationWide-field Infrared Survey Explorer (WISE)
Wide-field Infrared Survey Explorer (WISE) Latent Image Characterization Version 1.0 12-July-2009 Prepared by: Deborah Padgett Infrared Processing and Analysis Center California Institute of Technology
More informationAstronomical Detectors. Lecture 3 Astronomy & Astrophysics Fall 2011
Astronomical Detectors Lecture 3 Astronomy & Astrophysics Fall 2011 Detector Requirements Record incident photons that have been captured by the telescope. Intensity, Phase, Frequency, Polarization Difficulty
More informationWelcome to: LMBR Imaging Workshop. Imaging Fundamentals Mike Meade, Photometrics
Welcome to: LMBR Imaging Workshop Imaging Fundamentals Mike Meade, Photometrics Introduction CCD Fundamentals Typical Cooled CCD Camera Configuration Shutter Optic Sealed Window DC Voltage Serial Clock
More informationSelecting the NIR detectors for Euclid
National Aeronautics and Space Administration Jet Propulsion Laboratory California Institute of Technology Selecting the NIR detectors for Euclid Stefanie Wachter Michael Seiffert On behalf of the Euclid
More informationA new Infra-Red Camera for COAST. Richard Neill - PhD student Supervisor: Dr John Young
A new Infra-Red Camera for COAST Richard Neill - PhD student Supervisor: Dr John Young The Cambridge Optical Aperture-Synthesis Telescope: COAST is a
More informationLaboratory, University of Arizona, Tucson, AZ 85721; c ImagerLabs, 1995 S. Myrtle Ave., Monrovia CA INTRODUCTION ABSTRACT
A CMOS Visible Image Sensor with Non-Destructive Readout Capability Gary R. Sims* a, Gene Atlas c, Eric Christensen b, Roger W. Cover a, Stephen Larson b, Hans J. Meyer a, William V. Schempp a a Spectral
More informationVisible Light Communication-based Indoor Positioning with Mobile Devices
Visible Light Communication-based Indoor Positioning with Mobile Devices Author: Zsolczai Viktor Introduction With the spreading of high power LED lighting fixtures, there is a growing interest in communication
More informationPhase-2 Preparation Tool
Gran Telescopio Canarias Phase-2 Preparation Tool Valid from period 2014A Updated: 5 December 2013 1 Contents 1. The GTC Phase-2 System... 3 1.1. Introduction... 3 1.2. Logging in... 3 2. Defining an observing
More informationSLICING THE UNIVERSE CCDs for MUSE
SLICING THE UNIVERSE CCDs for MUSE Roland Reiss 1, Sebastian Deiries 1, Jean Louis Lizon 1, Manfred Meyer 1, Javier Reyes 1, Roland Bacon 2, François Hénault 2, Magali Loupias 2 1 European Southern Observatory,
More informationGO Added Near-IR Fringe Flats (Rev. A)
Instrument Science Report STIS 97-15 GO Added Near-IR Fringe Flats (Rev. A) S. Baum, Harry Ferguson, J. R. Walsh, P. Goudfrooij, R. Downes, and H. Lanning December 1, 1997 (Revised November 12, 1998) ABSTRACT
More informationWide Field Camera 3: Design, Status, and Calibration Plans
2002 HST Calibration Workshop Space Telescope Science Institute, 2002 S. Arribas, A. Koekemoer, and B. Whitmore, eds. Wide Field Camera 3: Design, Status, and Calibration Plans John W. MacKenty Space Telescope
More informationNature Methods: doi: /nmeth Supplementary Figure 1. Resolution of lysozyme microcrystals collected by continuous rotation.
Supplementary Figure 1 Resolution of lysozyme microcrystals collected by continuous rotation. Lysozyme microcrystals were visualized by cryo-em prior to data collection and a representative crystal is
More informationInterpixel Capacitance in the IR Channel: Measurements Made On Orbit
Interpixel Capacitance in the IR Channel: Measurements Made On Orbit B. Hilbert and P. McCullough April 21, 2011 ABSTRACT Using high signal-to-noise pixels in dark current observations, the magnitude of
More information4.5.1 Mirroring Gain/Offset Registers GPIO CMV Snapshot Control... 14
Thank you for choosing the MityCAM-C8000 from Critical Link. The MityCAM-C8000 MityViewer Quick Start Guide will guide you through the software installation process and the steps to acquire your first
More informationF/48 Slit Spectroscopy
1997 HST Calibration Workshop Space Telescope Science Institute, 1997 S. Casertano, et al., eds. F/48 Slit Spectroscopy R. Jedrzejewski & M. Voit Space Telescope Science Institute, Baltimore, MD 21218
More informationChapter 2 DECam Imager
Chapter 2 DECam Imager Version 0.0, Aug 2011 In This Chapter Instrument Overview... 2-1 Data Products... 2-7 Calibration.(TBD) Sources of Error.(TBD) References & Further Information 2-14 NOAO DATA The
More informationTIRCAM2 (TIFR Near Infrared Imaging Camera - 3.6m Devasthal Optical Telescope (DOT)
TIRCAM2 (TIFR Near Infrared Imaging Camera - II) @ 3.6m Devasthal Optical Telescope (DOT) (ver 4.0 June 2017) TIRCAM2 (TIFR Near Infrared Imaging Camera - II) is a closed cycle cooled imager that has been
More informationIn situ CCD testing. T. M. C. Abbott. 24 May Abstract. This document describes a number of methods of CCD testing which may be used
In situ CCD testing. T. M. C. Abbott 24 May 1995 Abstract This document describes a number of methods of CCD testing which may be used at the telescope to verify the integrity of a CCD system. The rationale
More informationLight gathering Power: Magnification with eyepiece:
Telescopes Light gathering Power: The amount of light that can be gathered by a telescope in a given amount of time: t 1 /t 2 = (D 2 /D 1 ) 2 The larger the diameter the smaller the amount of time. If
More informationFlat Fields. S. Eikenberry Obs Tech
Flat Fields S. Eikenberry Obs Tech 23 Sep 2014 Review median combination Basic algorithm: Read in im1, im2, im3,, im9 Loop over 1 array dimension, index i Loop over 2 nd dimension, index j imf(i,j)=median([im1(i,j),
More information