The iris plays an important role in angle closure disease. Relationship Between Iris Surface Features and Angle Width in Asian Eyes.

Size: px
Start display at page:

Download "The iris plays an important role in angle closure disease. Relationship Between Iris Surface Features and Angle Width in Asian Eyes."

Transcription

1 Glaucoma Relationship Between Iris Surface Features and Angle Width in Asian Eyes Elizabeth Sidhartha, 1,2 Monisha Esther Nongpiur, 1,3 Carol Y. Cheung, 1,3 Mingguang He, 4 Tien Yin Wong, 1 3 Tin Aung, 1,2 and Ching-Yu Cheng 1 3,5 1 Singapore Eye Research Institute, Singapore National Eye Centre, Singapore 2 Department of Ophthalmology, Yong Loo Lin School of Medicine, National University of Singapore and National University Health System, Singapore 3 Duke-NUS Graduate Medical School, Singapore 4 State Key Laboratory of Ophthalmology, Zhongshan Ophthalmic Center, Sun Yat-sen University, Guangzhou, China 5 Graduate School for Integrative Sciences and Engineering, National University of Singapore, Singapore Correspondence: Ching-Yu Cheng, Singapore Eye Research Institute, 20 College Road, The Academia, Level 6, Discovery Tower, Singapore ; ching-yu_cheng@nuhs.edu.sg. Submitted: August 5, 2014 Accepted: September 26, 2014 Citation: Sidhartha E, Nongpiur ME, Cheung CY, et al. Relationship between iris surface features and angle width in Asian eyes. Invest Ophthalmol Vis Sci. 2014;55: DOI: /iovs PURPOSE. To examine the associations between iris surface features with anterior chamber angle width in Asian eyes. METHODS. In this prospective cross-sectional study, we recruited 600 subjects from a large population-based study, the Singapore Epidemiology of Eye Diseases (SEED) study. We obtained standardized digital slit-lamp iris photographs and graded the iris crypts (by number and size), furrows (by number and circumferential extent), and color (higher grade denoting darker iris). Vertical and horizontal cross-sections of anterior chamber were imaged using anterior segment optical coherence tomography. Angle opening distance (AOD), angle recess area (ARA), and trabecular-iris space area (TISA) were measured using customized software. Associations of the angle width with the iris surface features in the subject s right eyes were assessed using linear regression analysis. RESULTS. A total of 464 eyes of the 464 subjects (mean age: years) had complete and gradable data for crypts and color, and 423 eyes had gradable data for furrows. After adjustment for age, sex, ethnicity, pupil size, and corneal arcus, higher crypt grade was independently associated with wider AOD750 (b [change in angle width per grade higher] ¼ 0.018, P ¼ 0.023), ARA750 (b ¼ 0.022, P ¼ 0.049), and TISA750 (b ¼ 0.011, P ¼ 0.019), and darker iris was associated narrower ARA750 (b ¼ 0.025, P ¼ 0.044) and TISA750 (b ¼ 0.013, P ¼ 0.011). CONCLUSIONS. Iris surface features, assessed and measured from slit-lamp photographs, correlated well with anterior chamber angle width; irises with more crypts and lighter color were associated with wider angle. These findings may provide another imaging modality to assess angle closure risk based on iris surface features. Keywords: iris, angle closure, glaucoma anterior segment The iris plays an important role in angle closure disease. Many features of the iris, including its geometric measurements and dynamic features, have been linked to angle closure. 1 7 Irises that are thicker, have larger cross-sectional area, and more convexity are associated with higher risk of angle closure. 5,6,8 Furthermore, greater iris volume expansion on pupil dilation also has been linked to angle closure. 1 However, the assessment of some of these iris parameters (e.g., thickness, cross-sectional area) requires technical expertise and costly instruments, such as anterior segment optical coherence tomography (AS-OCT) or ultrasound biomicroscopy. In contrast, the surface of the iris has distinctive clinical features, such as crypts, furrows, and color, which can be more readily captured and assessed by standard slit-lamp examination or slit-lamp photography. 9 These features may be novel markers of angle closure risk. The current assessment methods of iris surface features 10,11 were developed primarily for lighter Caucasian eyes. However, because Asian individuals are at a higher risk of developing primary angle closure glaucoma, 12 measuring and profiling the iris of Asian eyes is more important and clinically useful. We recently developed standardized methods to measure iris crypts and furrows in Asian eyes, and showed that these features are correlated with iris thickness. 9 In this study, we determined the association between surface features of the iris and anterior chamber angle width as measured using AS-OCT. METHODS Subjects of this study were enrolled from the Singapore Epidemiology of Eye Diseases (SEED) study, which is a population-based cohort study of eye diseases in Chinese, Malay, and Indian populations aged 40 to 80 years in Singapore The baseline examination was conducted between 2004 and 2011 and a follow-up examination has been conducted since January 2011, 6 years after the baseline examination. For this study, we prospectively recruited 600 Copyright 2014 The Association for Research in Vision and Ophthalmology, Inc. j ISSN:

2 Iris Features and Angle Width IOVS j December 2014 j Vol. 55 j No. 12 j 8145 consecutive subjects from SEED participants who attended the follow-up examination between August 2011 and April We excluded eyes with previous ocular surgery (including cataract surgery and filtering surgery) or with previous iris laser treatment, or under IOP-lowering medication, as these conditions may have changed the iris color or morphology in some eyes. Furthermore, eyes with corneal opacity or marked corneal arcus covering 50% or more of the iris area were excluded, as these conditions may affect iris grading. Participants recruited in the current study underwent standardized ophthalmic examination according to the study protocol, 13,15 including slit-lamp examination by trained ophthalmologists and AS-OCT imaging. In addition, standardized slit-lamp iris photography was done according to a standardized protocol developed specifically for this study purpose. 9 Written informed consent was obtained from all participants after explanation of the nature of the study. The study adhered to the tenets of the Declaration of Helsinki, and ethics committee approval was obtained from the Singapore Eye Research Institute Institutional Review Board. Iris Photography and Grading A slit-lamp digital camera (DC3; Topcon Corporation, Tokyo, Japan) was used to take color photographs of the iris at 316 magnification. Photographs were taken in a dark room (20 lux). The slit-lamp light beam was kept at full width (>20 mm) and height (14 mm) at 30% of the maximum brightness, and angled at 458 temporally. No flash was used. For grading purposes, photographs were viewed on a /60-Hz resolution screen, using the viewing software ACDSee Photo Manager Version 11.0 (ACD Systems, Seattle, WA, USA). We developed a grading scheme for iris crypts, contraction furrows, and iris color as described previously (Supplementary Material). 9 In brief, for grading of crypts, irises were given an integer grade between 1 and 5 based on the number and size of crypts present, as follows: grade 1, no crypts; grade 2, one to three crypts; grade 3, at least four crypts 1 mm or smaller in diameter; grade 4, at least four crypts 1 mm or larger in diameter; and grade 5, numerous crypts 1 mm or larger in diameter, covering nearly the entire iris. Furrows were graded based on their number and circumferential extent: grade 1, no furrow; grade 2, five or fewer furrows present extending 1808 or less); and grade 3, five or more furrows present extending 1808 or more). Iris color was graded from 1 to 5 based on the overall color of the iris, in comparison with the reference photographs. If an iris was deemed to fall between two consecutive grades, the higher grade was assigned. As furrows are mostly peripherally located, the grading of furrows was challenging in eyes with extensive corneal arcus. Therefore, eyes with corneal arcus covering 50% or more of the iris area were excluded from further analysis. We previously reported that this grading system has good intra- and intergrader agreements ( and , respectively) as well as excellent intra- and intervisit repeatability ( and , respectively). 9 Anterior Segment OCT Imaging Anterior segment OCT imaging (Visante; Carl Zeiss Meditec, Jena, Germany) was performed on all participants under standardized dark conditions (20 lux). We used the Anterior Segment Single-Scan protocol to obtain cross-sectional images of the anterior segment along the horizontal axis (nasaltemporal angles at ) and vertical axis (superior-inferior angles at ). Images were processed using the customized software, the Zhongshan Angle Assessment Program (Guangzhou, China). 16 Following the user input on the location of the two scleral spurs, the algorithm automatically measures the following angle parameters: angle opening distance (AOD250, AOD500, and AOD750), angle recess area (ARA750), and trabecular-iris space area (TISA500 and TISA750). Angle opening distance 250, AOD500, and AOD750 were the lengths of a line from the anterior iris to the corneal endothelium perpendicular to a line drawn along the trabecular meshwork at 250, 500, and 750 lm from the scleral spur, respectively. 17 Angle recess area 750 was defined as the area bordered by the anterior iris surface, corneal endothelium, and a line perpendicular to the corneal endothelium drawn to the iris surface from a point at 750 lm anterior to the scleral spur. 18 Trabecular-iris space area 500 and TISA750 were the areas modified by not including the area posterior to a line drawn from the scleral spur to the anterior iris perpendicular to the plane of the inner scleral wall. Among the three AOD and two TISA measures, AOD750 and TISA750 have been shown to have the highest diagnostic performance for angle closure, respectively. 19 In addition, the software was also used to measure iris thickness, iris curvature (I-Curv), and pupil size. Iris thickness was measured at 2000 lm from the scleral spur. I-Curv was measured by drawing a perpendicular line from the iris pigment epithelium at the point of greatest iris convexity, to the line extending from the most peripheral to the most central points of iris pigment epithelium. 8 The average of measured values of the four quadrants (nasal, temporal, superior, and inferior) was used for the analysis. Anterior segment OCT image analysis was done without the knowledge of the iris grading results. Statistical Analysis Statistical analysis was performed using MedCalc Version 12.3 (MedCalc Software, Ostend, Belgium) and SPSS Version 20.0 (SPSS, Inc., Chicago, IL, USA). Data from both eyes were collected but only the right eye was used so as to eliminate intereye correlation issue in analysis. The association between iris surface features (independent variables) and angle width (dependent variables) was assessed by linear regression models, adjusted for potential confounders, such as age, sex, ethnicity, pupil size, and corneal arcus. Corneal arcus might make the iris color appear lighter, and might partially or completely block the view of the peripheral iris. In addition to excluding eyes with extensive corneal arcus as mentioned earlier, because 59.1% of our study populations had mild corneal arcus, we included presence of corneal arcus in our multiple regression models as a covariate to account for its potential effect on iris grading. We did not adjust for iris thickness, because iris thickness is likely an intermediate variable on the causal pathway from iris surface texture to angle width and, hence, including it in the regression models may lead to overadjustment, thus obscuring the association between iris surface textures and angle width. RESULTS A total of 600 subjects were recruited for this study. Among them, 136 eyes were excluded for the following reasons: pseudophakic eyes (25 eyes), history of laser peripheral iridotomy (2 eyes), inability to complete AS-OCT imaging (6 eyes), poor-quality AS-OCT images in at least one quadrant (63 eyes), unidentifiable scleral spur in at least one quadrant (23 eyes), software demarcation error (13 eyes), and corneal opacity (4 eyes). Furthermore, 41 eyes were ungradable for furrows because of marked or extensive corneal arcus covering 50% or more of the entire iris area. None of the subjects were under any IOP-lowering medication or had undergone filtering surgery. Table 1 shows that compared with the excluded

3 Iris Features and Angle Width IOVS j December 2014 j Vol. 55 j No. 12 j 8146 TABLE 1. Comparison of Demographic and Clinical Characteristics Between Included and Excluded Eyes Characteristics Included, n ¼ 464 Excluded, n ¼ 136 P Value* Age, y, mean (SD) 57.5 (8.6) 65.8 (9.5) Sex, female (%) 256 (53.9) 69 (50.7) Ethnicity Malay 234 (50.4) 71 (52.3) Indian 163 (35.1) 52 (38.2) Chinese 67 (14.4) 13 (9.5) IOP, mm Hg, mean (SD) (3.09) (3.65) Vertical cup-disc ratio, mean (SD) 0.41 (0.13) 0.40 (0.14) * Based on independent sample t-test for continuous variables or v 2 /Fisher s exact test for categorical variables. TABLE Anterior Segment Characteristics of Study Participants, n ¼ Characteristics Mean (SD) or n (%) Corneal arcus, present 274 (59.1) Pupil size, mm 4.01 (0.75) Iris thickness, mm 0.44 (0.10) AOD750, mm 0.40 (0.18) ARA750, mm (0.19) TISA750, mm (0.10) I-Curv, mm 0.25 (0.13) subjects, the included subjects were younger and had a higher proportion of females, but there was no difference in IOP and vertical cup-to-disc ratio between the two groups. In total, 464 eyes were graded for crypts and color, and 423 eyes were graded for furrows and included in the analysis. The mean age of the 464 subjects was years, and of them, 53.9% were women. Ocular anterior segment characteristics of the study population are shown in Table 2, and the distribution of grades of iris surface features in the study population is shown in Figure 1. Table 3 presents the association between the three iris surface features and three different measures of angle width, namely AOD750, ARA750, and TISA750. Increasing crypt grade was significantly associated with wider AOD750 (change in width in millimeters per one grade higher, b ¼ 0.021, P ¼ 0.007) and TISA750 (b ¼ 0.012, P ¼ 0.010; Model 1). After adjusting for age, sex, presence of corneal arcus, ethnicity, and pupil size (Model 2), the association between increasing crypt grade and wider AOD750 (b ¼ 0.018, P ¼ 0.023) and TISA750 (b ¼ 0.011, P ¼ 0.019) remained significant, and the association between increasing crypt grade and larger ARA750 (b ¼ 0.022, P ¼ 0.049) became marginally significant. Increasing furrow grade did not show significant association with AOD750, ARA750 and TISA750 before or after adjustment of the aforementioned potential confounders (all P > 0.05). Iris color was negatively associated with all the angle width measures; darker color was associated with narrower AOD750 (b ¼ 0.017, P ¼ 0.048), ARA750 (b ¼ 0.035, P ¼ 0.005), and TISA750 (b ¼ 0.015, P ¼ 0.003). After adjustment for age, sex, ethnicity, corneal arcus, and pupil size, darker color was still significantly associated with smaller ARA750 (b ¼ 0.025, P ¼ 0.044) and TISA750 (b ¼ 0.013, P ¼ 0.011). Figure 2 further illustrates the trend observed between iris surface features and TISA750, independent of age, sex, ethnicity, pupil size, and corneal arcus, showing the increase in TISA750 with increasing crypt grade, and the decrease in TISA750 with increasing color grade. Table 4 shows the relationship between iris surface features and three additional measures of angle width, namely AOD250, AOD500, and TISA500. Similar trends were observed; higher crypt grade and lower color grade were associated with wider angle, whereas furrow grade was not associated with angle width. DISCUSSION The surface features of the iris are easy to observe clinically and can be captured with slit-lamp photography. Previously we developed a grading system tailored for Asian eyes to quantify iris surface features and showed that more iris crypts were associated with thinner iris thickness, whereas irises with more furrows tended to be thicker. 9 In this study, we further examined the associations of these surface features with angle width, and showed that iris crypt and color were significantly associated with angle width. We showed that eyes with larger and more iris crypts had wider angle, and eyes with darker irises had narrower angle. Iris crypts are regions of iris stroma hypoplasia or atrophy. 10 On slit-lamp examination, they appear as wide pits on the iris surface, often revealing less dense stroma at the base of the pits. FIGURE 1. Distribution of grades of crypts, furrows, and color in the study population.

4 Iris Features and Angle Width IOVS j December 2014 j Vol. 55 j No. 12 j 8147 TABLE 3. Changes in Angle Width Per Grade Increase in Iris Surface Features Model 1* Model 2 b (95% CI) P for Trend b (95% CI) P for Trend Crypt, n ¼ 464 AOD750, mm ( ) ( ) ARA750, mm ( ) ( ) TISA750, mm ( ) ( ) Furrow, n ¼ 423 AOD750, mm ( ) ( ) ARA750, mm ( ) ( ) TISA750, mm ( ) ( ) Color, n ¼ 464 AOD750, mm ( to 0.001) ( ) ARA750, mm ( to 0.011) ( to 0.001) TISA750, mm ( to 0.005) ( to 0.003) b, change in AOD (in mm) or area (in mm 2 ) for one grade higher in iris surface features; CI, confidence interval. * Model 1: unadjusted models. Model 2: adjusted for age, sex, ethnicity, pupil size, and corneal arcus. FIGURE 2. Relationship between iris features and TISA750. Increasing crypt grade and decreasing color grade were associated with wider TISA750, whereas furrow grade did not show any trend with TISA750. Data shown are after adjustment for age, sex, ethnicity, pupil size, and corneal arcus. Our results showed that higher crypt grade was associated with wider angle. We speculated that the presence of more and wider crypts may indicate a more porous iris stroma, which permits better flow of aqueous humor across the iris. Such flow may help maintain a low pressure gradient between the posterior chamber and anterior chamber, and in turn reduce the likelihood of the iris bowing forward and narrowing the anterior chamber angle. In addition, we also found that eyes with darker irises had narrower angle. This might be because darker irises have higher melanocytes and melanin content, 10,20 which makes the iris stroma thicker and thus less permeable to aqueous humor. However, further adjustment for iris curvature in our analysis had no effects on the association results (data not shown). This indicates that there may be another mechanism TABLE 4. Features Changes in Additional Measures of Angle Width (AOD at 250 lm and 500 lm, and TISA at 500 lm) per Grade Increase in Iris Surface Model 1* Model 2 b (95% CI) P for Trend b (95% CI) P for Trend Crypt, n ¼ 464 AOD ( ) ( ) AOD ( ) ( ) TISA ( ) ( ) Furrow, n ¼ 423 AOD ( ) ( ) AOD ( ) ( ) TISA ( ) ( ) Color, n ¼ 464 AOD ( ) ( to 0.003) AOD ( to 0.003) ( to 0.001) TISA ( to 0.001) ( ) * Model 1: unadjusted models. Model 2: adjusted for age, sex, ethnicity, pupil size, and corneal arcus.

5 Iris Features and Angle Width IOVS j December 2014 j Vol. 55 j No. 12 j 8148 underlying the association between iris crypt or color and angle width that is independent of iris bowing. We did not observe any statistically significant association between iris furrows and angle width. Furrows indicate sites on the iris that are frequently folded when the pupil dilates. 10 They appear as protuberant lines on the iris periphery. It could be expected that such protrusions would contribute to the narrowing of the angle; however, this association was not observed. The presence of corneal arcus may have hindered an accurate assessment of the peripherally located furrows. In addition, the exclusion of 41 eyes in which the furrows were ungradable may have reduced our power to detect the association between furrows and angle width. Slit-lamp examination of iris surface features is potentially advantageous over the use of AS-OCT. Slit-lamp photography is a simpler method to assess the entire iris globally, permitting the assessment of overall angle width, which is beneficial over the use of the cross-sectional images taken by AS-OCT. Although AS-OCT gives a more direct and precise measurement of angle width, it allows only point measurements of angle width from several cross-sectional images, thus may not represent the whole angle width. Current emerging technology, such as swept-source OCT, allows cross-sectional imaging of the whole 3608 view of the anterior chamber, which may allow more accurate measurement of angle width. 21 However, such instruments are costly and require technical expertise to operate, as opposed to the slit-lamp biomicroscope. There are limitations to our study. First, the subjective grading of iris surface features may have introduced grading variability and measurement bias. Nevertheless, the inter- and intragrading agreement was high for all three features in our study. 9 In addition, the graders were masked from a subject s clinical characteristics while performing iris grading. A more objective, automated method of quantifying iris surface features could be developed to achieve even more precise and reliable measurement of these features. Second, our results may not be directly applied to populations of European descent, whose eyes have much higher crypt frequency 10,22 and lighter color than Asian eyes. 11,23 In conclusion, iris surface features, particularly iris crypts and color, can provide information on angle width status. We showed that irises with more crypts and lighter color are associated with wider angle. Further studies involving angle closure patients would be warranted to directly assess whether these iris surface features could provide additional predictive power for the risk of angle closure, above and beyond the known risk factors such as shallow anterior chamber depth and shorter axial length. Acknowledgments Supported by grants from the National Medical Research Council, Singapore (NMRC R607/28/2008 and NMRC/NIG/1069/2012). The funding organization had no role in the design or conduct of this research. Disclosure: E. Sidhartha, None; M.E. Nongpiur, None; C.Y. Cheung, None; M. He, None; T.Y. Wong, None; T. Aung, None; C.-Y. Cheng, None References 1. Aptel F, Denis P. Optical coherence tomography quantitative analysis of iris volume changes after pharmacologic mydriasis. Ophthalmology. 2010;117: Ganeshrao SB, Mani B, Ulganathan S, et al. Change in iris parameters with physiological mydriasis. Optom Vis Sci. 2012; 89: He M, Lu Y, Liu X, et al. Histologic changes of the iris in the development of angle closure in chinese eyes. J Glaucoma. 2008;17: Quigley HA, Silver DM, Friedman DS, et al. Iris cross-sectional area decreases with pupil dilation and its dynamic behavior is a risk factor in angle closure. J Glaucoma. 2009;18: Wang B, Sakata LM, Friedman DS, et al. Quantitative iris parameters and association with narrow angles. Ophthalmology. 2010;117: Wang BS, Narayanaswamy A, Amerasinghe N, et al. Increased iris thickness and association with primary angle closure glaucoma. Br J Ophthalmol. 2011;95: Zheng C, Cheung CY, Aung T, et al. In vivo analysis of vectors involved in pupil constriction in chinese subjects with angle closure. Invest Ophthalmol Vis Sci. 2012;53: Ishikawa H, Liebmann JM, Ritch R. Quantitative assessment of the anterior segment using ultrasound biomicroscopy. Curr Opin Ophthalmol. 2000;11: Sidhartha E, Gupta P, Liao J, et al. Assessment of iris surface features and their relationship with iris thickness in Asian eyes. Ophthalmology. 2014;121: Larsson M, Pedersen NL. Genetic correlations among texture characteristics in the human iris. Mol Vis. 2004;10: Sturm RA, Larsson M. Genetics of human iris colour and patterns. Pigment Cell Melanoma Res. 2009;22: Cedrone C, Mancino R, Cerulli A, et al. Epidemiology of primary glaucoma: prevalence, incidence, and blinding effects. Prog Brain Res. 2008;173: Lavanya R, Jeganathan VS, Zheng Y, et al. Methodology of the Singapore Indian Chinese Cohort (SICC) eye study: quantifying ethnic variations in the epidemiology of eye diseases in Asians. Ophthalmic Epidemiol. 2009;16: Rosman M, Zheng Y, Wong W, et al. Singapore Malay eye study: rationale and methodology of 6-year follow-up study (SIMES-2). Clin Experiment Ophthalmol. 2012;40: Foong AW, Saw SM, Loo JL, et al. Rationale and methodology for a population-based study of eye diseases in Malay people: The Singapore Malay Eye Study (SIMES). Ophthalmic Epidemiol. 2007;14: Console JW, Sakata LM, Aung T, et al. Quantitative analysis of anterior segment optical coherence tomography images: The Zhongshan angle assessment program. Br J Ophthalmol. 2008;92: Pavlin CJ, Harasiewicz K, Foster FS. Ultrasound biomicroscopy of anterior segment structures in normal and glaucomatous eyes. Am J Ophthalmol. 1992;113: Ishikawa H, Esaki K, Liebmann JM, et al. Ultrasound biomicroscopy dark room provocative testing: a quantitative method for estimating anterior chamber angle width. Jpn J Ophthalmol. 1999;43: Narayanaswamy A, Sakata LM, He MG, et al. Diagnostic performance of anterior chamber angle measurements for detecting eyes with narrow angles: an anterior segment OCT study. Arch Ophthalmol. 2010;128: Imesch PD, Bindley CD, Khademian Z, et al. Melanocytes and iris color. Electron microscopic findings. Arch Ophthalmol. 1996;114: Liu S, Yu M, Ye C, Lam DS, Leung CK. Anterior chamber angle imaging with swept-source optical coherence tomography: an investigation on variability of angle measurement. Invest Ophthalmol Vis Sci. 2011;52: Spierer A, Isenberg SJ, Inkelis SH. Characteristics of the iris in 100 neonates. J Pediatr Ophthalmol Strabismus. 1989;26: Edwards M, Gozdzik A, Ross K, Miles J, Parra EJ. Technical note: quantitative measures of iris color using high resolution photographs. Am J Phys Anthropol. Jan 2012;147:

Primary angle-closure glaucoma is a major cause of visual

Primary angle-closure glaucoma is a major cause of visual Glaucoma Dynamic Analysis of Iris Configuration with Anterior Segment Optical Coherence Tomography Carol Yim-lui Cheung, 1 Shu Liu, 1 Robert N. Weinreb, 2 Jing Liu, 1 Haitao Li, 1 Dexter Yu-lung Leung,

More information

ANTERIOR CHAMBER ANGLE MEASUREMENT USING OPTICAL COHERENCE TOMOGRAPHY IMAGE

ANTERIOR CHAMBER ANGLE MEASUREMENT USING OPTICAL COHERENCE TOMOGRAPHY IMAGE ANTERIOR CHAMBER ANGLE MEASUREMENT USING OPTICAL COHERENCE TOMOGRAPHY IMAGE M.Prabha Mr.A.Rajan ME-Applied Electronics Assistant Professor (Sr)/ECE Shri Andal Alagar College of Engineering Shri Andal Alagar

More information

Please feel to contact Sonomed for any suggestions, comments or request about this bulletin.

Please feel to contact Sonomed for any suggestions, comments or request about this bulletin. Sonomed Quarterly Bulletin Volume One, January 2008 Dear Sonomed Colleagues: Sonomed is proud to announce the VuMAX II Ultrasound Biomicroscope is experiencing huge success in the global ophthalmic markets.

More information

11/10/2015. Haag Streit Topcon Zeiss Kowa Add On Systems- OIS/Escalon and Others. The Original Design. Photo Slit lamp Systems. Who Makes Them?

11/10/2015. Haag Streit Topcon Zeiss Kowa Add On Systems- OIS/Escalon and Others. The Original Design. Photo Slit lamp Systems. Who Makes Them? The Original Design Photo Slit lamp Systems Who Makes Them? 1862-1930 Alvar Gullstrand Inventor of the Slit lamp illuminator - 1911 Swedish ophthalmologist, recipient of the 1911 Nobel Prize for Physiology

More information

Gonioscopy Wet-Lab. About Me. About Me Gonioscopy. Indications for Gonioscopy. Billing Gonioscopy 12/13/2012. Code: 92020

Gonioscopy Wet-Lab. About Me. About Me Gonioscopy. Indications for Gonioscopy. Billing Gonioscopy 12/13/2012. Code: 92020 About Me Gonioscopy Wet-Lab Marcus Gonzales, OD, FAAO Cedar Springs Eye Clinic (formerly BridgeBuilders Eye Clinic) About Me Gonioscopy Evaluates the anterior chamber angle responsible for the outflow

More information

Optical Coherence Tomography Retina Scan Duo

Optical Coherence Tomography Retina Scan Duo Optical Coherence Tomography Retina Scan Duo High Definition OCT & Fundus Imaging in One Compact System The Retina Scan Duo is a combined OCT and fundus camera system that is a user friendly and versatile

More information

Kodak (ASA 400). image by computer. A given point (xi, yi) was. xi is the distance from the axis in the optical image,

Kodak (ASA 400). image by computer. A given point (xi, yi) was. xi is the distance from the axis in the optical image, British Journal of Ophthalmology, 1986, 70, 668-672 Measurement method of the anterior chamber volume by image analysis TAKEHISA KONDO,' MASAO MIURA,' AND MASATSUGU IMAMICHIP From the 'Department of Ophthalmology,

More information

Classification Of Types Of Glaucoma Based On Anterior Chamber Region Assessment Of Hd-Oct Images. S. J. Grace Shoba et al.,,,

Classification Of Types Of Glaucoma Based On Anterior Chamber Region Assessment Of Hd-Oct Images. S. J. Grace Shoba et al.,,, Classification Of Types Of Glaucoma Based On Anterior Chamber Region Assessment Of Hd-Oct Images. S. J. Grace International Journal of Technology and Engineering System (IJTES) Vol 7. No.3 2015 Pp. 230-238

More information

OCULAR MEDIA* PHOTOGRAPHIC RECORDING OF OPACITIES OF THE. development by the control of diabetes, the supply of a deficient hormone

OCULAR MEDIA* PHOTOGRAPHIC RECORDING OF OPACITIES OF THE. development by the control of diabetes, the supply of a deficient hormone Brit. J. Ophthal. (1955) 39, 85. PHOTOGRAPHIC RECORDING OF OPACITIES OF THE OCULAR MEDIA* BY E. F. FINCHAM Institute of Ophthalmology, University of London THE value of photography for recording pathological

More information

Department of Ophthalmology, Perelman School of Medicine at the University of Pennsylvania

Department of Ophthalmology, Perelman School of Medicine at the University of Pennsylvania Yuanjie Zheng 1, Dwight Stambolian 2, Joan O'Brien 2, James Gee 1 1 Penn Image Computing & Science Lab, Department of Radiology, 2 Department of Ophthalmology, Perelman School of Medicine at the University

More information

The TRC-NW8F Plus: As a multi-function retinal camera, the TRC- NW8F Plus captures color, red free, fluorescein

The TRC-NW8F Plus: As a multi-function retinal camera, the TRC- NW8F Plus captures color, red free, fluorescein The TRC-NW8F Plus: By Dr. Beth Carlock, OD Medical Writer Color Retinal Imaging, Fundus Auto-Fluorescence with exclusive Spaide* Filters and Optional Fluorescein Angiography in One Single Instrument W

More information

case report Scleral lens fit based on OCT data

case report Scleral lens fit based on OCT data Page 1 of 5 I-site Amsterdam Netherlands i-site@netherlens.com Home Archive September 2009 Downloads case report Scleral lens fit based on OCT data Key words: scleral lens, ocular coherence tomography,

More information

Simple method of determining the axial length of the eye

Simple method of determining the axial length of the eye Brit. Y. Ophthal. (1976) 6o, 266 Simple method of determining the axial length of the eye E. S. PERKINS, B. HAMMOND, AND A. B. MILLIKEN From the Department of Experimental Ophthalmology, Institute of Ophthalmology,

More information

Optical Coherence Tomography. RS-3000 Advance / Lite

Optical Coherence Tomography. RS-3000 Advance / Lite Optical Coherence Tomography RS-3000 Advance / Lite 12 mm wide horizontal scan available with the RS-3000 Advance allows detailed observation of the vitreous body, retina, and choroid from the macula to

More information

Clinical evaluation and management of glaucoma is largely

Clinical evaluation and management of glaucoma is largely Macular Segmentation with Optical Coherence Tomography Hiroshi Ishikawa, 1,2 Daniel M. Stein, 1 Gadi Wollstein, 1,2 Siobahn Beaton, 1,2 James G. Fujimoto, 3 and Joel S. Schuman 1,2 PURPOSE. To develop

More information

Effects of Pupil Center Shift on Ocular Aberrations

Effects of Pupil Center Shift on Ocular Aberrations Visual Psychophysics and Physiological Optics Effects of Pupil Center Shift on Ocular Aberrations David A. Atchison and Ankit Mathur School of Optometry & Vision Science and Institute of Health & Biomedical

More information

Image Database and Preprocessing

Image Database and Preprocessing Chapter 3 Image Database and Preprocessing 3.1 Introduction The digital colour retinal images required for the development of automatic system for maculopathy detection are provided by the Department of

More information

Headline. IOLMaster. Subline. The gold standard in biometry

Headline. IOLMaster. Subline. The gold standard in biometry Headline IOLMaster Subline The gold standard in biometry The rapid evolution of IOL technology promises superior outcomes in cataract surgery, and it necessarily raises the bar for pre-operative biometry.

More information

Lecture 2 Slit lamp Biomicroscope

Lecture 2 Slit lamp Biomicroscope Lecture 2 Slit lamp Biomicroscope 1 Slit lamp is an instrument which allows magnified inspection of interior aspect of patient s eyes Features Illumination system Magnification via binocular microscope

More information

Computer analysis of optic disc images. Comparison with HRT data

Computer analysis of optic disc images. Comparison with HRT data Computer analysis of optic disc images. Comparison with HRT data Mihai Bîscă, Liliana Voinea, Radu Burcin, Mădălina Voicu University Hospital Bucureşti, Ophthalmology Clinic, Oftalux Medical Center 1.

More information

Image Modeling of the Human Eye

Image Modeling of the Human Eye Image Modeling of the Human Eye Rajendra Acharya U Eddie Y. K. Ng Jasjit S. Suri Editors ARTECH H O U S E BOSTON LONDON artechhouse.com Contents Preface xiiii CHAPTER1 The Human Eye 1.1 1.2 1. 1.4 1.5

More information

1. Introduction to Anatomy of the Eye and its Adnexa

1. Introduction to Anatomy of the Eye and its Adnexa 1. Introduction to Anatomy of the Eye and its Adnexa Fig 1: A Cross section of the human eye. Let us imagine we are traveling with a ray of light into the eye. The first structure we will encounter is

More information

Headline. SL 130 Slit Lamp. Subline. Maximum quality for optimum performance

Headline. SL 130 Slit Lamp. Subline. Maximum quality for optimum performance Headline SL 130 Slit Lamp Subline Maximum quality for optimum performance Versatility is the outstanding feature of the SL 130 slit lamp from Carl Zeiss. Combined with an advanced, user-focused operating

More information

Segmentation of Blood Vessel in Retinal Images and Detection of Glaucoma using BWAREA and SVM

Segmentation of Blood Vessel in Retinal Images and Detection of Glaucoma using BWAREA and SVM Segmentation of Blood Vessel in Retinal Images and Detection of Glaucoma using BWAREA and SVM P.Dhivyabharathi 1, Mrs. V. Priya 2 1 P. Dhivyabharathi, Research Scholar & Vellalar College for Women, Erode-12,

More information

Reports. \ $mm>-j \ksiiimj

Reports. \ $mm>-j \ksiiimj Reports A scanning slit optical microscope. DAVID M. MAURICE. A transparent tissue is illuminated with a narrow slit of light which is formed by light transmitted down one-half of a microscope objective,

More information

Fourier Domain (Spectral) OCT OCT: HISTORY. Could OCT be a Game Maker OCT in Optometric Practice: A THE TECHNOLOGY BEHIND OCT

Fourier Domain (Spectral) OCT OCT: HISTORY. Could OCT be a Game Maker OCT in Optometric Practice: A THE TECHNOLOGY BEHIND OCT Could OCT be a Game Maker OCT in Optometric Practice: A Hands On Guide Murray Fingeret, OD Nick Rumney, MSCOptom Fourier Domain (Spectral) OCT New imaging method greatly improves resolution and speed of

More information

Visual Optics. Visual Optics - Introduction

Visual Optics. Visual Optics - Introduction Visual Optics Jim Schwiegerling, PhD Ophthalmology & Optical Sciences University of Arizona Visual Optics - Introduction In this course, the optical principals behind the workings of the eye and visual

More information

MEASUREMENTS OF THE SAGITTAL AXIS OF THE HUMAN

MEASUREMENTS OF THE SAGITTAL AXIS OF THE HUMAN Brit. J. Ophthal. (1968) 52, 81 MEASUREMENTS OF THE SAGITTAL AXIS OF THE HUMAN EYE IN VIVO DURING APPLANATION OF THE CORNEA*t$ BY From the Second Eye Clinic, University of Vienna, Austria (Director: Univ.

More information

Upper and Lower Tear Menisci on Contact Lenses METHODS

Upper and Lower Tear Menisci on Contact Lenses METHODS Upper and Lower Tear Menisci on Contact Lenses Jianhua Wang, 1,2 Ian Cox, 3 and William T. Reindel 3 PURPOSE. The purpose of this study was to measure upper and lower tear menisci on contact lenses using

More information

Technicians & Nurses Program

Technicians & Nurses Program ASCRS ASOA Symposium & Congress Technicians & Nurses Program April 17-21, 2015 San Diego, California What Do All These Colors Mean And who s paying for all the ink? BLUF: Red is bad. Frank W. Scribbick,

More information

What s Fundus photography s purpose? Why do we take them? Why do we do it? Why do we do it? Why do we do it? 11/3/2014. To document the retina

What s Fundus photography s purpose? Why do we take them? Why do we do it? Why do we do it? Why do we do it? 11/3/2014. To document the retina What s Fundus photography s purpose? To document the retina Photographers role to show the retina Document other ocular structures Why do we take them? Why do we do it? We as photographers help the MD

More information

Image formation in fundus cameras

Image formation in fundus cameras Image formation in fundus cameras Oleg Pomerantzeff, Robert H. Webb, and Francois C. Delori Imaging in a fundus camera depends more on design of the system stop than on correction of the first fundus image

More information

This is the author s version of a work that was submitted/accepted for publication in the following source:

This is the author s version of a work that was submitted/accepted for publication in the following source: This is the author s version of a work that was submitted/accepted for publication in the following source: Atchison, David A. & Mathur, Ankit (2014) Effects of pupil center shift on ocular aberrations.

More information

Normal Wavefront Error as a Function of Age and Pupil Size

Normal Wavefront Error as a Function of Age and Pupil Size RAA Normal Wavefront Error as a Function of Age and Pupil Size Raymond A. Applegate, OD, PhD Borish Chair of Optometry Director of the Visual Optics Institute College of Optometry University of Houston

More information

Posterior capsule opacification (PCO) is the most frequent

Posterior capsule opacification (PCO) is the most frequent Retroillumination versus Reflected-Light Images in the Photographic Assessment of Posterior Capsule Opacification Monica Camparini, Claudio Macaluso, Luca Reggiani, and Giovanni Maraini PURPOSE. To investigate

More information

Smart-Device Ophthalmic Photography

Smart-Device Ophthalmic Photography Smart-Device Ophthalmic Photography Austin Holmes, M.P.H. Clifford M. Terry, M.D. Financial Interest Disclosure: Unrestricted Grant From Terry Eye Institute Overview Creation of Smart-Device Photography

More information

MULTIPLE SENSORS LENSLETS FOR SECURE DOCUMENT SCANNERS

MULTIPLE SENSORS LENSLETS FOR SECURE DOCUMENT SCANNERS INFOTEH-JAHORINA Vol. 10, Ref. E-VI-11, p. 892-896, March 2011. MULTIPLE SENSORS LENSLETS FOR SECURE DOCUMENT SCANNERS Jelena Cvetković, Aleksej Makarov, Sasa Vujić, Vlatacom d.o.o. Beograd Abstract -

More information

Blood Vessel Tree Reconstruction in Retinal OCT Data

Blood Vessel Tree Reconstruction in Retinal OCT Data Blood Vessel Tree Reconstruction in Retinal OCT Data Gazárek J, Kolář R, Jan J, Odstrčilík J, Taševský P Department of Biomedical Engineering, FEEC, Brno University of Technology xgazar03@stud.feec.vutbr.cz

More information

ABO Certification Training. Part I: Anatomy and Physiology

ABO Certification Training. Part I: Anatomy and Physiology ABO Certification Training Part I: Anatomy and Physiology Major Ocular Structures Centralis Nerve Major Ocular Structures The Cornea Cornea Layers Epithelium Highly regenerative: Cells reproduce so rapidly

More information

Smartphone Imaging in Ophthalmology: A Comparison with Traditional Methods on the Reproducibility and Usability for Anterior Segment Imaging

Smartphone Imaging in Ophthalmology: A Comparison with Traditional Methods on the Reproducibility and Usability for Anterior Segment Imaging 6 Smartphone Imaging in Ophthalmology David ZY Chen et al Original Article Smartphone Imaging in Ophthalmology: A Comparison with Traditional Methods on the Reproducibility and Usability for Anterior Segment

More information

Retinal stray light originating from intraocular lenses and its effect on visual performance van der Mooren, Marie Huibert

Retinal stray light originating from intraocular lenses and its effect on visual performance van der Mooren, Marie Huibert University of Groningen Retinal stray light originating from intraocular lenses and its effect on visual performance van der Mooren, Marie Huibert IMPORTANT NOTE: You are advised to consult the publisher's

More information

Optical Coherence Tomography. RS-3000 Advance

Optical Coherence Tomography. RS-3000 Advance Optical Coherence Tomography RS-3000 Advance See it in Advance See it in high resolution with the AngioScan* image. SLO Superficial capillary OCT-Angiography (3 x 3 mm) Deep capillary OCT-Angiography (3

More information

EYE ANATOMY. Multimedia Health Education. Disclaimer

EYE ANATOMY. Multimedia Health Education. Disclaimer Disclaimer This movie is an educational resource only and should not be used to manage your health. The information in this presentation has been intended to help consumers understand the structure and

More information

Medical imaging has long played a critical role in diagnosing

Medical imaging has long played a critical role in diagnosing Three-Dimensional Optical Coherence Tomography (3D-OCT) Image Enhancement with Segmentation-Free Contour Modeling C-Mode Hiroshi Ishikawa, 1,2 Jongsick Kim, 1,2 Thomas R. Friberg, 1,2 Gadi Wollstein, 1

More information

Simultaneously measuring ocular aberration and anterior segment biometry during accommodation

Simultaneously measuring ocular aberration and anterior segment biometry during accommodation Journal of Innovative Optical Health Sciences Vol. 8, No. 2 (2015) 1550005 (6 pages) #.c The Authors DOI: 10.1142/S1793545815500054 Simultaneously measuring ocular aberration and anterior segment biometry

More information

THE EYE. People of Asian descent have an EPICANTHIC FOLD in the upper eyelid; no functional difference.

THE EYE. People of Asian descent have an EPICANTHIC FOLD in the upper eyelid; no functional difference. THE EYE The eye is in the orbit of the skull for protection. Within the orbit are 6 extrinsic eye muscles, which move the eye. There are 4 cranial nerves: Optic (II), Occulomotor (III), Trochlear (IV),

More information

The angle of the anterior chamber needs to be assessed

The angle of the anterior chamber needs to be assessed Thomas George MS Common Tests Simplified Gonioscopy The angle of the anterior chamber needs to be assessed in all glaucomas. It usually holds the secrets to pathogenesis in that particular case. It may

More information

Introduction. Chapter Aim of the Thesis

Introduction. Chapter Aim of the Thesis Chapter 1 Introduction 1.1 Aim of the Thesis The main aim of this investigation was to develop a new instrument for measurement of light reflected from the retina in a living human eye. At the start of

More information

What s New in Ocular Biomechanics?

What s New in Ocular Biomechanics? What s New in Ocular Biomechanics? The International Congress of Wavefront Sensing & Optimized Refractive Corrections Wavefront Course January 28, 2006 Torrence A. Makley Research Professor Department

More information

PERIMETRY A STANDARD TEST IN OPHTHALMOLOGY

PERIMETRY A STANDARD TEST IN OPHTHALMOLOGY 7 CHAPTER 2 WHAT IS PERIMETRY? INTRODUCTION PERIMETRY A STANDARD TEST IN OPHTHALMOLOGY Perimetry is a standard method used in ophthalmol- It provides a measure of the patient s visual function - performed

More information

Optical Coherence Tomography. RS-3000 Advance / Lite

Optical Coherence Tomography. RS-3000 Advance / Lite Optical Coherence Tomography RS-3000 Advance / Lite See it in Advance See it in high resolution with the AngioScan* image. SLO Superficial capillary OCT-Angiography (3 x 3 mm) Deep capillary OCT-Angiography

More information

THRESHOLD AMSLER GRID TESTING AND RESERVING POWER OF THE POTIC NERVE by MOUSTAFA KAMAL NASSAR. M.D. MENOFIA UNIVERSITY.

THRESHOLD AMSLER GRID TESTING AND RESERVING POWER OF THE POTIC NERVE by MOUSTAFA KAMAL NASSAR. M.D. MENOFIA UNIVERSITY. THRESHOLD AMSLER GRID TESTING AND RESERVING POWER OF THE POTIC NERVE by MOUSTAFA KAMAL NASSAR. M.D. MENOFIA UNIVERSITY. Since Amsler grid testing was introduced by Dr Marc Amsler on 1947and up till now,

More information

An Automatic System for Classification of Nuclear Sclerosis from Slit-Lamp Photographs

An Automatic System for Classification of Nuclear Sclerosis from Slit-Lamp Photographs An Automatic System for Classification of Nuclear Sclerosis from Slit-Lamp Photographs Shaohua Fan 1, Charles R. Dyer 1, Larry Hubbard 2, Barbara Klein 2 {shaohua, dyer}@cs.wisc.edu, hubbard@rc.ophth.wisc.edu,

More information

AHEAD Superior technology, thoughtfully designed with you in mind for an intelligent approach to cataract surgery.

AHEAD Superior technology, thoughtfully designed with you in mind for an intelligent approach to cataract surgery. ALWAYS THINKING AHEAD Superior technology, thoughtfully designed with you in mind for an intelligent approach to cataract surgery. The LENSAR Laser System was designed with your efficiency in mind, so

More information

Objectives. 3. Visual acuity. Layers of the. eye ball. 1. Conjunctiva : is. three quarters. posteriorly and

Objectives. 3. Visual acuity. Layers of the. eye ball. 1. Conjunctiva : is. three quarters. posteriorly and OCULAR PHYSIOLOGY (I) Dr.Ahmed Al Shaibani Lab.2 Oct.2013 Objectives 1. Review of ocular anatomy (Ex. after image) 2. Visual pathway & field (Ex. Crossed & uncrossed diplopia, mechanical stimulation of

More information

Instruments Commonly Used For Examination of the Eye

Instruments Commonly Used For Examination of the Eye Instruments Commonly Used For Examination of the Eye There are many instruments that the eye doctor might use to evaluate the eye and the vision system. This report presents some of the more commonly used

More information

EYE: THE PHOTORECEPTOR SYSTEM. Prof. Dr. Huda Al Khateeb

EYE: THE PHOTORECEPTOR SYSTEM. Prof. Dr. Huda Al Khateeb EYE: THE PHOTORECEPTOR SYSTEM Prof. Dr. Huda Al Khateeb Lecture 1 The eye ball Objectives By the end of this lecture the student should: 1. List the layers and chambers of the eye ball 2. Describe the

More information

NOW. Approved for NTIOL classification from CMS Available in Quar ter Diopter Powers. Accommodating. Aberration Free. Aspheric.

NOW. Approved for NTIOL classification from CMS Available in Quar ter Diopter Powers. Accommodating. Aberration Free. Aspheric. NOW Approved for NTIOL classification from CMS Available in Quar ter Diopter Powers Accommodating. Aberration Free. Aspheric. Accommodation Meets Asphericity in AO Merging Innovation & Proven Design The

More information

A reduction of visual fields during changes in the background image such as while driving a car and looking in the rearview mirror

A reduction of visual fields during changes in the background image such as while driving a car and looking in the rearview mirror Original Contribution Kitasato Med J 2012; 42: 138-142 A reduction of visual fields during changes in the background image such as while driving a car and looking in the rearview mirror Tomoya Handa Department

More information

Sensory receptors External internal stimulus change detectable energy transduce action potential different strengths different frequencies

Sensory receptors External internal stimulus change detectable energy transduce action potential different strengths different frequencies General aspects Sensory receptors ; respond to changes in the environment. External or internal environment. A stimulus is a change in the environmental condition which is detectable by a sensory receptor

More information

Going beyond the surface of your retina OCT-HS100 OPTICAL COHERENCE TOMOGRAPHY

Going beyond the surface of your retina OCT-HS100 OPTICAL COHERENCE TOMOGRAPHY Going beyond the surface of your retina OCT-HS100 OPTICAL COHERENCE TOMOGRAPHY Automatic functions make examinations short and simple. Perform the examination with only two simple mouse clicks! 1. START

More information

Going beyond the surface of your retina OCT-HS100 OPTICAL COHERENCE TOMOGRAPHY

Going beyond the surface of your retina OCT-HS100 OPTICAL COHERENCE TOMOGRAPHY Going beyond the surface of your retina OCT-HS100 OPTICAL COHERENCE TOMOGRAPHY Full Auto OCT High specifications in a very compact design Automatic functions make examinations short and simple. Perform

More information

Optical Coherence Tomography. RS-3000 Advance / Lite

Optical Coherence Tomography. RS-3000 Advance / Lite Optical Coherence Tomography RS-3000 Advance / Lite See it in Advance See it in high resolution with the AngioScan* image. OCT-Angiography of choroidal neovascularization * AngioScan (OCT-Angiography)

More information

Anterior and Posterior Axial Lens Displacement and Human Aqueous Outflow Facility

Anterior and Posterior Axial Lens Displacement and Human Aqueous Outflow Facility Investigative Ophthalmology & Visual Science, Vol. 29, No. 7, July 1988 Copyright Association for Research in Vision and Ophthalmology Anterior and Posterior Axial Lens Displacement and Human Aqueous Outflow

More information

Going beyond the surface of your retina

Going beyond the surface of your retina Going beyond the surface of your retina OCT-HS100 Optical Coherence Tomography Canon s expertise in optics and innovative technology have resulted in a fantastic 3 μm optical axial resolution for amazing

More information

Choosing the Proper Power for the IOL. Brannon Aden, MD Miles H. Friedlander, MD, FACS

Choosing the Proper Power for the IOL. Brannon Aden, MD Miles H. Friedlander, MD, FACS Choosing the Proper Power for the IOL Brannon Aden, MD Miles H. Friedlander, MD, FACS Goal s of Surgery Have Changed. In past the goal was good visual outcome Now an equal goal is a good refractive outcome

More information

THE BEST OF BOTH WORLDS Dual-Scheimpflug and Placido Reaching a new level in refractive screening

THE BEST OF BOTH WORLDS Dual-Scheimpflug and Placido Reaching a new level in refractive screening THE BEST OF BOTH WORLDS Dual-Scheimpflug and Placido Reaching a new level in refractive screening Clinical Applications Corneal Implant Planning The comes with a licensable corneal inlay software module

More information

Optical Coherence Tomography. RS-3000 Advance 2

Optical Coherence Tomography. RS-3000 Advance 2 Optical Coherence Tomography RS-3000 Advance 2 -Providing a comprehensive solution for retina and glaucom Retina Analysis Retinal mode Glaucoma Analysis Choroidal mode Image courtesy of Hokkaido University

More information

Macula centred, giving coverage of the temporal retinal. Disc centred. Giving coverage of the nasal retina.

Macula centred, giving coverage of the temporal retinal. Disc centred. Giving coverage of the nasal retina. 3. Field positions, clarity and overall quality For retinopathy screening purposes in England two images are taken of each eye. These have overlapping fields of view and between them cover the main area

More information

Corneal Asphericity and Retinal Image Quality: A Case Study and Simulations

Corneal Asphericity and Retinal Image Quality: A Case Study and Simulations Corneal Asphericity and Retinal Image Quality: A Case Study and Simulations Seema Somani PhD, Ashley Tuan OD, PhD, and Dimitri Chernyak PhD VISX Incorporated, 3400 Central Express Way, Santa Clara, CA

More information

studies on the cup/disc ratio

studies on the cup/disc ratio British Journal of Ophthalmology, 1983, 67, 356-361 An optic disc grid: its evaluation in reproducibility studies on the cup/disc ratio R. A. HITCHINGS, C. GENIO, S. ANDERTON, AND P. CLARK* From the Glaucoma

More information

Patient-Assisted Compression Impact on Image Quality and Workflow

Patient-Assisted Compression Impact on Image Quality and Workflow Patient-Assisted Compression Impact on Image Quality and Workflow Senographe Pristina In 2017, GE Healthcare s Senographe Pristina ( Pristina ) was approved by the FDA using the standard technologist-controlled

More information

4Basic anatomy and physiology

4Basic anatomy and physiology Hene_Ch09.qxd 8/30/04 6:51 AM Page 348 348 4Basic anatomy and physiology The eye is a highly specialized organ with an average axial length of 24 mm and a volume of 6.5 ml. Except for its anterior aspect,

More information

INVIVO PATTERN RECOGNITION AND DIGITAL IMAGE ANALYSIS OF SHEAR STRESS DISTRIBUTION IN HUMAN EYE

INVIVO PATTERN RECOGNITION AND DIGITAL IMAGE ANALYSIS OF SHEAR STRESS DISTRIBUTION IN HUMAN EYE INVIVO PATTERN RECOGNITION AND DIGITAL IMAGE ANALYSIS OF SHEAR STRESS DISTRIBUTION IN HUMAN EYE S. Joseph Antony School of Chemical and Process Engineering Faculty of Engineering, University of Leeds,

More information

Iris based Human Identification using Median and Gaussian Filter

Iris based Human Identification using Median and Gaussian Filter Iris based Human Identification using Median and Gaussian Filter Geetanjali Sharma 1 and Neerav Mehan 2 International Journal of Latest Trends in Engineering and Technology Vol.(7)Issue(3), pp. 456-461

More information

In the following diagram the parts of the eye are visualized and labeled for you.

In the following diagram the parts of the eye are visualized and labeled for you. Investigation 3.12B: The Eye In the preceding case study marker of the problem of greatest concern to you lay in finding the pupils fixed in a dilated position. But what is the pupil and what makes it

More information

Sutureless, Glueless, Scleral Fixation of Single-Piece and Toric Intraocular Lens: A Novel Technique

Sutureless, Glueless, Scleral Fixation of Single-Piece and Toric Intraocular Lens: A Novel Technique Published online: July 21, 2015 1663 2699/15/0062 0239$39.50/0 This is an Open Access article licensed under the terms of the Creative Commons Attribution-NonCommercial 3.0 Unported license (CC BY-NC)

More information

Auto Ref/ Keratometer / Auto Refractometer ARK-1 / AR-1

Auto Ref/ Keratometer / Auto Refractometer ARK-1 / AR-1 Auto Ref/ Keratometer / Auto Refractometer ARK-1 / AR-1 The Superior Auto and Auto What is the superior auto ref / The ARK-1 / AR-1 series speak for themselves, and auto refractometer generating greater

More information

CHAPTER 4 LOCATING THE CENTER OF THE OPTIC DISC AND MACULA

CHAPTER 4 LOCATING THE CENTER OF THE OPTIC DISC AND MACULA 90 CHAPTER 4 LOCATING THE CENTER OF THE OPTIC DISC AND MACULA The objective in this chapter is to locate the centre and boundary of OD and macula in retinal images. In Diabetic Retinopathy, location of

More information

The Eye. Nakhleh Abu-Yaghi, M.B.B.S Ophthalmology Division

The Eye. Nakhleh Abu-Yaghi, M.B.B.S Ophthalmology Division The Eye Nakhleh Abu-Yaghi, M.B.B.S Ophthalmology Division Coats of the Eyeball 1- OUTER FIBROUS COAT is made up of : Posterior opaque part 2-THE SCLERA the dense white part 1- THE CORNEA the anterior

More information

Focus on an optical blind spot A closer look at lenses and the basics of CCTV optical performances,

Focus on an optical blind spot A closer look at lenses and the basics of CCTV optical performances, Focus on an optical blind spot A closer look at lenses and the basics of CCTV optical performances, by David Elberbaum M any security/cctv installers and dealers wish to know more about lens basics, lens

More information

Novel 3D Computerized Threshold Amsler Grid Test CA, USA

Novel 3D Computerized Threshold Amsler Grid Test CA, USA Novel 3D Computerized Threshold Amsler Grid Test Wolfgang Fink 1,2 and Alfredo A. Sadun 2 1 California Institute of Technology, Pasadena, CA, USA 2 Doheny Eye Institute, Keck School of Medicine, University

More information

An Interesting Use of Bausch and Lomb s KeraSoft IC Lens

An Interesting Use of Bausch and Lomb s KeraSoft IC Lens An Interesting Use of Bausch and Lomb s KeraSoft IC Lens Nate Schlotthauer, OD 2012 Michigan College of Optometry Cornea and Contact Lens Resident Introduction: The KeraSoft IC lens, introduced to the

More information

Auto Ref/ Keratometer / Auto Refractometer ARK-1 / AR-1

Auto Ref/ Keratometer / Auto Refractometer ARK-1 / AR-1 Auto Ref/ Keratometer / Auto Refractometer ARK-1 / AR-1 The Superior Auto Ref / Keratometer and Auto Refractometer What is the superior auto ref / keratometer and auto refractometer? The ARK-1 / AR-1 series

More information

MEL 80 Excimer Laser. When you want to see better performance

MEL 80 Excimer Laser. When you want to see better performance MEL 80 Excimer Laser When you want to see better performance Reward your practice Invest in the very latest refractive excimer technology! The MEL 80 makes vision correction even safer, more patient-friendly

More information

Dynamic Ophthalmic Software for today s market. Automated Grading System

Dynamic Ophthalmic Software for today s market. Automated Grading System Dynamic Ophthalmic Software for today s market Automated Grading System AOS Anterior is a new device agnostic software designed to analyse digital images of the eye and automate the grading scale for (1)

More information

ATLAS Corneal Topography System

ATLAS Corneal Topography System ATLAS Corneal Topography System Simply accurate for maximum productivity Model 9000 The New ATLAS Take your practice to the next level Carl Zeiss Meditec has taken the world s leading corneal topography

More information

The First True Color Confocal Scanner on the Market

The First True Color Confocal Scanner on the Market The First True Color Confocal Scanner on the Market White color and infrared confocal images: the advantages of white color and confocality together for better fundus images. The infrared to see what our

More information

Chapter 36. Image Formation

Chapter 36. Image Formation Chapter 36 Image Formation Image of Formation Images can result when light rays encounter flat or curved surfaces between two media. Images can be formed either by reflection or refraction due to these

More information

Fitting Manual Use with kerasofttraining.com

Fitting Manual Use with kerasofttraining.com Fitting Manual Use with Fitting Manual: Contents This fitting manual is best used in conjunction with KeraSoft IC online training. To register, please visit www. 01 Kerasoft IC Design - Outlines the KeraSoft

More information

The Aberration Structure of the Keratoconic Eye

The Aberration Structure of the Keratoconic Eye The Aberration Structure of the Keratoconic Eye Geunyoung Yoon, Ph.D. Department of Ophthalmology Center for Visual Science Institute of Optics Department of Biomedical Engineering University of Rochester

More information

INTRODUCING OPTICS CONCEPTS TO STUDENTS THROUGH THE OX EYE EXPERIMENT

INTRODUCING OPTICS CONCEPTS TO STUDENTS THROUGH THE OX EYE EXPERIMENT INTRODUCING OPTICS CONCEPTS TO STUDENTS THROUGH THE OX EYE EXPERIMENT Marcela L. Redígolo redigolo@univap.br Leandro P. Alves leandro@univap.br Egberto Munin munin@univap.br IP&D Univap Av. Shishima Hifumi,

More information

An assessment of regional differences in corneal thickness in normal human eyes, using the Orbscan II or ultrasound pachymetry

An assessment of regional differences in corneal thickness in normal human eyes, using the Orbscan II or ultrasound pachymetry Optometry (2007) 78, 181-190 An assessment of regional differences in corneal thickness in normal human eyes, using the Orbscan II or ultrasound pachymetry Michael J. Doughty, Ph.D., and Sven Jonuscheit,

More information

Auto Ref / Kerato / Tono / Pachymeter TONOREFTM III

Auto Ref / Kerato / Tono / Pachymeter TONOREFTM III Auto Ref / Kerato / Tono / Pachymeter TONOREFTM III The space saving TONOREF TM III is a comfortable and efficient upgrade to your practice High Measurement Accuracy Refraction Large Pupil Zone Imaging

More information

An analysis of retinal receptor orientation

An analysis of retinal receptor orientation An analysis of retinal receptor orientation IV. Center of the entrance pupil and the center of convergence of orientation and directional sensitivity Jay M. Enoch and G. M. Hope In the previous study,

More information

Auto Ref/ Keratometer / Auto Refractometer ARK-1 / AR-1

Auto Ref/ Keratometer / Auto Refractometer ARK-1 / AR-1 Auto Ref/ Keratometer / Auto Refractometer ARK-1 / AR-1 The Superior Auto and Auto What is the superior auto ref / The ARK-1 / AR-1 series speak for themselves, and auto refractometer generating greater

More information

Raise your expectations. Deliver theirs.

Raise your expectations. Deliver theirs. 66 EXTENDED RANGE OF VISION MONOFOCAL-LIKE DISTANCE Raise your expectations. Deliver theirs. Now you can give your patients the best of both worlds with the first and only hybrid designed monofocal-multifocal

More information

Pupil: The adaptable opening at the position of the iris through which light enters the eye.

Pupil: The adaptable opening at the position of the iris through which light enters the eye. Processing and Analysis of Retinal Images in Optical Coherence Tomography (OCT) Shimaa Fawzy 1, Hossam El-Din Mostafa 2, Dalia Sabry 3, Rasheed Mokhtar 4 1 Communication Engineering Dept., Faculty of Engineering,

More information

CONTACT LENSES Discover a different world

CONTACT LENSES Discover a different world CONTACT LENSES Discover a different world Tradition and innovation Since 1858, visionary thinking and a fascination with technology have guided us to develop innovative products of outstanding reliability:

More information

Anterior Chamber Width Measurement by High-Speed Optical Coherence Tomography

Anterior Chamber Width Measurement by High-Speed Optical Coherence Tomography Anterior Chamber Width Measurement by High-Speed Optical Coherence Tomography Jason A. Goldsmith, MD, MS, 1 Yan Li, MS, 1,2 Maria Regina Chalita, MD, 1 Volker Westphal, PhD, 2 Chetan A. Patil, BS, 2 Andrew

More information