1 Blood Vessel Tree Reconstruction in Retinal OCT Data Gazárek J, Kolář R, Jan J, Odstrčilík J, Taševský P Department of Biomedical Engineering, FEEC, Brno University of Technology xgazar03@stud.feec.vutbr.cz Initial study of blood vessels reconstruction in the retinal optical coherence tomography (OCT) data is reported in this paper. Reconstruction and segmentation of the blood vessel tree is an important task for image registration, because it carries spatial information about the OCT B-scans positions. The dark shadows created by blood-vessels in the retinal pigment epithelium (RPE) shown in the OCT scans offer a possibility for blood vessel tree reconstruction. The main part of this paper is therefore focused on RPE layer detection and alignment. Promising results have been obtained for OCT data concerning several patients in this initial study. 1 Introduction The optical coherence tomography (OCT) has found many applications in medical and biological sciences, particularly in clinical ophthalmology. One of the most important applications is the early glaucoma diagnosis based either on the examination of the retinal nerve fibre layer from cross-sectional scans [1] or based on the segmentation of optic disc cup and rim [2]. Another important clinical application is OCT-based diagnosis of the age related macular degeneration [3]. Nowadays, the OCT devices work in the spectral domain, which enables faster acquisition of particular cross-sectional tomographic images (B-scans) of retinal structures [4]. A data set from a single measurement is therefore three dimensional (3D). Although the data acquisition is fast, the patient and eye movement can cause artefacts, which can disturb the workflow of the following data processing and analysis. This paper is focused on the retinal blood vessel tree reconstruction and segmentation in the OCT data [5]. The blood vessels create important features for monomodal or multimodal registration of retinal images; for example OCT data and images taken by the fundus camera [6] or scanning laser ophthalmoscopy (SLO) images [7]. 2 Methods The positions of the blood vessels can be recognized in each B-scan as dark stripes, which are caused by shadows generated by blood vessels in outer part of retina. This property can be used for detection. The highest contrast between the blood vessels shadows and the background can be observed in the retinal pigment epithelium (RPE) and surrounding bright layers. Therefore, the main segmentation task is the RPE detection. The proposed approach for RPE segmentation is depicted in Fig. 1. Fig. 1. The flowchart of proposed method for RPE detection 287
2 The typical B-scan without significant distortion with several marked structures is shown in Fig. 2. The retinal pigment epithelium (RPE) is an intraretinal layer with high optical reflectivity, which can be relatively easily detected in a case of low axial warping. Fig. 2. One B-scan with several retinal structures important for presented application. ILM: internal limiting membrane, RNFL: retinal nerve fibre layer, RPE: retinal pigment epithelium The proposed approach was tested on 8 volume sets from 4 eyes. The volume B-scans had been obtained by OCT (Heidelberg Engineering Spectralis) and has resolution pixels. They were acquired with high density B-mode scanning (from 61 to 169 B-scans per field of view). Fig. 3 shows two distorted B-scans with axially warped retinal layers. This kind of distortion is caused by the high frequency eye movement, which is impossible to track and eliminate [8]. Fig. 3. One B-scan with low level of spatial warping; An example of B-scan with high spatial distortion. 2.1 Preprocessing The preprocessing step consists of median filtering for speckle noise elimination. The mask size has been determined experimentally to obtain sharp edges: 21x21 pixels. This filter has been applied two times to obtain more noise-free images. The filtered image was consequently filtered with convolution kernel approximating the gradient in vertical direction. 288
3 Fig. 4. The gradient images corresponding to images in Fig RPE localization It has been observed that the strongest edge in each column arises from the internal limiting membrane (ILM) or RPE layer for majority of A-scans. Therefore two strongest reflections can be easily detected taking the maximum intensity for each A-scan in the gradient image. The second strongest edge is considered if the gradient is above detected threshold (0.01), which helps to eliminate detection of the weak edges inside the optic disc. The next step consists of moving averaging of the spatial position of detected RPE edge. The size of the averaging window has been set to 5 pixels. The next step helps to eliminate those edge pixels or edge fragments, which don t belong to RPE layer. A priori information about the continuity of RPE is used defining the local tangent (LT) as: LTk = positionx k +1 positionx k, k = 1,2 K n 1 positiony k +1 positiony k, where n is a number of connected edge points (fragments), positionxk+1 (positionyk+1) is X (Y) coordinate of the nearest fragment to the neighbouring fragment pixel kpositionx (kpositiony). If LT value is higher than 1.5, the shorter fragment is erased. This algorithm works iteratively until the LT values are lower than 1.5. Two results of this step are shown in Fig. 5. Fig. 5. Two examples of RPE localization. The presence of optic disc ( complicates the RPE detection. 2.3 Postprocessing After removing the outlying detected points, the cubic interpolation is used to compute the missing points in RPE layer. For smoothing the polynomial approximation of 20th order is used. This higher order is necessary for successful tracking of high spatial RPE variation in distorted images. The results are discussed in the next section. 289
4 3 Results An example of two results is presented in Fig. 6a,b. Successfully detected spatial variation can be seen, which corresponds to an approximate position of the RPE layer. Nevertheless, the precise detection of the RPE is not the main task of this work. The main reason for the RPE detection is the reconstruction and segmentation of the blood vessels, which is simpler for the aligned B-scans (see Fig. 6c,d). c) d) Fig. 6. a, Examples of two B-scans and detected shape of spatial warping; c, d) corresponding images after warping elimination. B-scan on c) presents an example of blood vessel detection. The reconstructed blood vessel tree local luminosity can be obtained by column-wise averaging of 10 pixels belonging to RPE layer. This operation eliminates the noise and in the end increases the blood vessel contrast in the reconstructed image. The results, i.e. blood-vessel images are presented here for two different eyes. Fig. 7 shows the successfully reconstructed tree without any artefacts and the corresponding colour fundus image. Fig. 8 shows the results with several uncorrected moving artefacts, which are marked by red arrows. These artefacts are caused by false RPE detection due to several effects, as: - higher level of noise in case that all B-scans do not have the same signal-to-noise ratio, which is probably caused by different number of individual B-scans used for averaging [4]; - absence of RPE - in the region of optic disc or macula; - detection of false edges corresponding to different retinal layers with strong reflection; - wrong B-scan acquisition the spatial warping can be so high, that the retinal layers lie outside of the acquired image depth range, in some cases; 290
5 Fig.7. Reconstructed blood-vessel OCT image and the corresponding colour fundus image Fig.8. Another example of reconstructed blood-vessel image with the corresponding colour fundus image 291
6 4 Conclusions A new method for blood vessel tree reconstruction based on 2D detection of RPE in OCT B-scans was proposed. The algorithm was tested on 8 cases of eye volume scans with 783 B-scans in total. The detected RPE layer has been subjectively evaluated and classified into 4 classes: 1. correct RPE detection: 92.46% (724 B-scans) 2. small distortion of RPE detection: 3,32% (26 B-scans) 3. high distortion of RPE detection: 1,92% (15 B-scans) 4. low-quality B-scans 2,30% (18 B-scans) The obtained 2D image of the vessel tree enables registering it with the corresponding fundus-camera retinal image, and consequently it is possible to find a position of each OCT B-scan in the fundus-camera image. This way, the objective quantitative evaluation of the retinal neural layer (RNL) obtained by the OCT method may be compared with the qualitative conclusion on the local presence of the RNL provided by the texture analysis based detection in the fundus-camera image. A future development of this method will cover 3D RPE segmentation, which will utilize also the correlation along the 3 rd (transversal) dimension. Acknowledgement This work has been supported by the national research center DAR (Data, Algorithms and Decision making) project no. 1M0572 coordinated by the Institute of Information Theory and Automation, Academy of Science, Czech Rep. and partly also by the institutional research frame no. MSM ; both grants sponsored by the Ministry of Education of the Czech Republic. The authors highly acknowledge the cooperation with University Eye Hospital Erlangen-Nurnberg (R.Tornow, R. Laemmer, Ch. Mardin), through which also the test set of images was provided. References [1] Trip S.A, Schlottmann P.G, Jones S.J et al. Retinal nerve fibber layer axonal loss and visual dysfunction in optic neuritis. American Journal of Ophthalmology, Vol. 140, No. 6, [2] Lee K, Niemeijer M, Garvin M.K et al. Segmentation of the optic disc in 3-d oct scans of the optic nerve head. Medici Imaging, IEEE Transactions on, Vol. 29, No. 1, pp , Jan [3] Coscas G, Coscas F, Zourdani A et al. Optical coherence tomography and armd. J Fr Ophtalmol., Vol. 27, pp. 3S7 3S30, [4] Drexler W, Fujimoto J.G. Optical Coherence Tomography. Springer, [5] Fercher A.F, Drexler W, Hitzenberger C.K et al. Optical coherence tomography - principles and applications Reports on Progress in Physics, Vol.66, pp , [6] Kolar R, Tasevsky P. Registration of 3D Retinal Optical Coherence Tomography Data and 2D Fundus Images. Accepted to Workshop on Biomedical Image Registration 2010, Luebeck, Germany. [7] Ricco S, Chen M, Ishikawa H et al., Correcting motion artifacts in retinal spectral domain optical coherence tomography via image registration. Medical Image Computing and Computer-Assisted Intervention MICCAI 2009, pp , October [Online]. Available: [8] Fuller A.R, Zawadzki R.J, Choi S et al., Segmentation of three-dimensional retinal image data. IEEE Transactions on Visualization and Computer Graphics, Vol. 13, No. 6, pp , November
Going beyond the surface of your retina OCT-HS100 OPTICAL COHERENCE TOMOGRAPHY
Going beyond the surface of your retina OCT-HS100 OPTICAL COHERENCE TOMOGRAPHY Automatic functions make examinations short and simple. Perform the examination with only two simple mouse clicks! 1. START
More informationGoing beyond the surface of your retina OCT-HS100 OPTICAL COHERENCE TOMOGRAPHY
Going beyond the surface of your retina OCT-HS100 OPTICAL COHERENCE TOMOGRAPHY Full Auto OCT High specifications in a very compact design Automatic functions make examinations short and simple. Perform
More informationImage Database and Preprocessing
Chapter 3 Image Database and Preprocessing 3.1 Introduction The digital colour retinal images required for the development of automatic system for maculopathy detection are provided by the Department of
More informationFourier Domain (Spectral) OCT OCT: HISTORY. Could OCT be a Game Maker OCT in Optometric Practice: A THE TECHNOLOGY BEHIND OCT
Could OCT be a Game Maker OCT in Optometric Practice: A Hands On Guide Murray Fingeret, OD Nick Rumney, MSCOptom Fourier Domain (Spectral) OCT New imaging method greatly improves resolution and speed of
More informationGoing beyond the surface of your retina
Going beyond the surface of your retina OCT-HS100 Optical Coherence Tomography Canon s expertise in optics and innovative technology have resulted in a fantastic 3 μm optical axial resolution for amazing
More informationCHAPTER 4 LOCATING THE CENTER OF THE OPTIC DISC AND MACULA
90 CHAPTER 4 LOCATING THE CENTER OF THE OPTIC DISC AND MACULA The objective in this chapter is to locate the centre and boundary of OD and macula in retinal images. In Diabetic Retinopathy, location of
More informationAutomatic No-Reference Quality Assessment for Retinal Fundus Images Using Vessel Segmentation
Automatic No-Reference Quality Assessment for Retinal Fundus Images Using Vessel Segmentation Thomas Köhler 1,2, Attila Budai 1,2, Martin F. Kraus 1,2, Jan Odstrčilik 4,5, Georg Michelson 2,3, Joachim
More informationThe TRC-NW8F Plus: As a multi-function retinal camera, the TRC- NW8F Plus captures color, red free, fluorescein
The TRC-NW8F Plus: By Dr. Beth Carlock, OD Medical Writer Color Retinal Imaging, Fundus Auto-Fluorescence with exclusive Spaide* Filters and Optional Fluorescein Angiography in One Single Instrument W
More informationFovea and Optic Disc Detection in Retinal Images with Visible Lesions
Fovea and Optic Disc Detection in Retinal Images with Visible Lesions José Pinão 1, Carlos Manta Oliveira 2 1 University of Coimbra, Palácio dos Grilos, Rua da Ilha, 3000-214 Coimbra, Portugal 2 Critical
More informationDrusen Detection in a Retinal Image Using Multi-level Analysis
Drusen Detection in a Retinal Image Using Multi-level Analysis Lee Brandon 1 and Adam Hoover 1 Electrical and Computer Engineering Department Clemson University {lbrando, ahoover}@clemson.edu http://www.parl.clemson.edu/stare/
More informationA new method for segmentation of retinal blood vessels using morphological image processing technique
A new method for segmentation of retinal blood vessels using morphological image processing technique Roya Aramesh Faculty of Computer and Information Technology Engineering,Qazvin Branch,Islamic Azad
More informationOCT - Anatomy of a Scan. OCT - Anatomy of a Scan. OCT Imaging. OCT Imaging
OCT - Anatomy of a Scan Timothy J. Bennett, CRA, OCT-C, FOPS Penn State Eye Center Hershey, PA OCT - Anatomy of a Scan A systematic approach to understanding what we see in retinal OCT images including
More informationComputer analysis of optic disc images. Comparison with HRT data
Computer analysis of optic disc images. Comparison with HRT data Mihai Bîscă, Liliana Voinea, Radu Burcin, Mădălina Voicu University Hospital Bucureşti, Ophthalmology Clinic, Oftalux Medical Center 1.
More informationSUPPLEMENTARY INFORMATION
Computational high-resolution optical imaging of the living human retina Nathan D. Shemonski 1,2, Fredrick A. South 1,2, Yuan-Zhi Liu 1,2, Steven G. Adie 3, P. Scott Carney 1,2, Stephen A. Boppart 1,2,4,5,*
More informationMedical imaging has long played a critical role in diagnosing
Three-Dimensional Optical Coherence Tomography (3D-OCT) Image Enhancement with Segmentation-Free Contour Modeling C-Mode Hiroshi Ishikawa, 1,2 Jongsick Kim, 1,2 Thomas R. Friberg, 1,2 Gadi Wollstein, 1
More informationOptical Coherence Tomography Retina Scan Duo
Optical Coherence Tomography Retina Scan Duo High Definition OCT & Fundus Imaging in One Compact System The Retina Scan Duo is a combined OCT and fundus camera system that is a user friendly and versatile
More informationFundus Photograph Reading Center
Spectral Domain Optical Coherence Tomography (SD-OCT) Heidelberg Spectralis 8010 Excelsior Drive, Suite 100, Madison WI 53717 Telephone: (608) 410-0560 Fax: (608) 410-0566 Table of Contents 1. Heidelberg
More informationVolumetric microvascular imaging of human retina using optical coherence tomography with a novel motion contrast technique
Volumetric microvascular imaging of human retina using optical coherence tomography with a novel motion contrast technique Jeff Fingler 1,*, Robert J. Zawadzki 2, John S. Werner 2, Dan Schwartz 3, Scott
More informationGaussian and Fast Fourier Transform for Automatic Retinal Optic Disc Detection
Gaussian and Fast Fourier Transform for Automatic Retinal Optic Disc Detection Arif Muntasa 1, Indah Agustien Siradjuddin 2, and Moch Kautsar Sophan 3 Informatics Department, University of Trunojoyo Madura,
More informationAutomatic Detection Of Optic Disc From Retinal Images. S.Sherly Renat et al.,
International Journal of Technology and Engineering System (IJTES) Vol 7. No.3 2015 Pp. 203-207 gopalax Journals, Singapore available at : www.ijcns.com ISSN: 0976-1345 AUTOMATIC DETECTION OF OPTIC DISC
More informationSECTION I - CHAPTER 2 DIGITAL IMAGING PROCESSING CONCEPTS
RADT 3463 - COMPUTERIZED IMAGING Section I: Chapter 2 RADT 3463 Computerized Imaging 1 SECTION I - CHAPTER 2 DIGITAL IMAGING PROCESSING CONCEPTS RADT 3463 COMPUTERIZED IMAGING Section I: Chapter 2 RADT
More informationLocalization of Optic Disc and Macula using Multilevel 2-D Wavelet Decomposition Based on Haar Wavelet Transform
Localization of Optic Disc and Macula using Multilevel 2-D Wavelet Decomposition Based on Haar Wavelet Transform Deepali D. Rathod MS Ramesh R. Manza MS ogesh M. Rajput MS Manjiri B. Patwari Institute
More informationVisualization of human retinal micro-capillaries with phase contrast high-speed optical coherence tomography
Visualization of human retinal micro-capillaries with phase contrast high-speed optical coherence tomography Dae Yu Kim 1,2, Jeff Fingler 3, John S. Werner 1,2, Daniel M. Schwartz 4, Scott E. Fraser 3,
More informationSegmentation approaches of optic cup from retinal images: A Survey
I J C T A, 10(8), 2017, pp. 377-382 International Science Press ISSN: 0974-5572 Segmentation approaches of optic cup from retinal images: A Survey Niharika Thakur* and Mamta Juneja** ABSTRACT Eye is a
More informationUltrahigh speed volumetric ophthalmic OCT imaging at 850nm and 1050nm
Ultrahigh speed volumetric ophthalmic OCT imaging at 850nm and 1050nm The MIT Faculty has made this article openly available. Please share how this access benefits you. Your story matters. Citation As
More informationBlood Vessel Tracking Technique for Optic Nerve Localisation for Field 1-3 Color Fundus Images
Blood Tracing Technique for Optic Nerve Localisation for Field 1-3 Color Fundus Images Hwee Keong Lam, Opas Chutatape School of Electrical and Electronic Engineering Nanyang Technological University, Nanyang
More informationAutomatic and manual segmentation of healthy retinas using high-definition optical coherence tomography
Automatic and manual segmentation of healthy retinas using high-definition optical coherence tomography Isabelle Golbaz, 1 Christian Ahlers, 1 Nina Goesseringer, 2 Geraldine Stock, 1 Wolfgang Geitzenauer,
More informationImage Processing Of Oct Glaucoma Images And Information Theory Analysis
University of Denver Digital Commons @ DU Electronic Theses and Dissertations Graduate Studies 1-1-2009 Image Processing Of Oct Glaucoma Images And Information Theory Analysis Shuting Wang University of
More informationThe First True Color Confocal Scanner on the Market
The First True Color Confocal Scanner on the Market White color and infrared confocal images: the advantages of white color and confocality together for better fundus images. The infrared to see what our
More informationSegmentation of Blood Vessel in Retinal Images and Detection of Glaucoma using BWAREA and SVM
Segmentation of Blood Vessel in Retinal Images and Detection of Glaucoma using BWAREA and SVM P.Dhivyabharathi 1, Mrs. V. Priya 2 1 P. Dhivyabharathi, Research Scholar & Vellalar College for Women, Erode-12,
More informationOPTIC DISC LOCATION IN DIGITAL FUNDUS IMAGES
OPTIC DISC LOCATION IN DIGITAL FUNDUS IMAGES Miss. Tejaswini S. Mane 1,Prof. D. G. Chougule 2 1 Department of Electronics, Shivaji University Kolhapur, TKIET,Wrananagar (India) 2 Department of Electronics,
More informationRetinal blood vessel extraction
Retinal blood vessel extraction Surya G 1, Pratheesh M Vincent 2, Shanida K 3 M. Tech Scholar, ECE, College, Thalassery, India 1,3 Assistant Professor, ECE, College, Thalassery, India 2 Abstract: Image
More informationDepartment of Ophthalmology, Perelman School of Medicine at the University of Pennsylvania
Yuanjie Zheng 1, Dwight Stambolian 2, Joan O'Brien 2, James Gee 1 1 Penn Image Computing & Science Lab, Department of Radiology, 2 Department of Ophthalmology, Perelman School of Medicine at the University
More informationThe First True-Color Wide-Field Confocal Scanner
The First True-Color Wide-Field Confocal Scanner 2 Company Profile CenterVue designs and manufactures highly automated medical devices for the diagnosis and management of ocular pathologies, including
More informationAn Efficacious Method of Cup to Disc Ratio Calculation for Glaucoma Diagnosis Using Super pixel
An Efficacious Method of Cup to Disc Ratio Calculation for Glaucoma Diagnosis Using Super pixel Dr.G.P.Ramesh 1, M.Malini 2, Professor 1, PG Scholar 2, St.Peter s University, TN, India. Abstract: Glaucoma
More informationImpressive Wide Field Image Quality with Small Pupil Size
Impressive Wide Field Image Quality with Small Pupil Size White color and infrared confocal images: the advantages of white color and confocality together for better fundus images. The infrared to see
More informationOptical Coherence Tomography. RS-3000 Advance / Lite
Optical Coherence Tomography RS-3000 Advance / Lite 12 mm wide horizontal scan available with the RS-3000 Advance allows detailed observation of the vitreous body, retina, and choroid from the macula to
More informationSPECTRALIS Training Guide
SPECTRALIS Training Guide SPECTRALIS Diagram 1 SPECTRALIS Training Guide Table of Contents 1. Entering Patient Information & Aligning the Patient a. Start Up/Shut Down the System... 4 b. Examine a New
More informationCSC 320 H1S CSC320 Exam Study Guide (Last updated: April 2, 2015) Winter 2015
Question 1. Suppose you have an image I that contains an image of a left eye (the image is detailed enough that it makes a difference that it s the left eye). Write pseudocode to find other left eyes in
More informationBetter diagnosis and treatment all-in-one.
Accessories Options duct Specifications hs-on control of the slit lamp without disturbing r view of the retina. solid state diode cavity yellow-red configuration: 5 nm 70 nm green-red configuration: 53
More informationA Study On Preprocessing A Mammogram Image Using Adaptive Median Filter
A Study On Preprocessing A Mammogram Image Using Adaptive Median Filter Dr.K.Meenakshi Sundaram 1, D.Sasikala 2, P.Aarthi Rani 3 Associate Professor, Department of Computer Science, Erode Arts and Science
More informationimagespectrum ADVANCED DIGITAL IMAGE MANAGEMENT SYSTEM Get a Better Handle on the Big Picture
ADVANCED DIGITAL IMAGE MANAGEMENT SYSTEM Get a Better Handle on the Big Picture SECURELY STREAMLINE YOUR PRACTICE WORKFLOW imagespectrum enables eye care practices, clinics, and even entire hospital departments
More informationLarge-field high-speed polarization sensitive spectral domain OCT and its applications in ophthalmology
Large-field high-speed polarization sensitive spectral domain OCT and its applications in ophthalmology Stefan Zotter, 1* Michael Pircher, 1 Teresa Torzicky, 1 Bernhard Baumann, 1 Hirofumi Yoshida, 3 Futoshi
More informationABSTRACT 1. INTRODUCTION
High-resolution retinal imaging: enhancement techniques Mircea Mujat 1*, Ankit Patel 1, Nicusor Iftimia 1, James D. Akula 2, Anne B. Fulton 2, and R. Daniel Ferguson 1 1 Physical Sciences Inc., Andover
More informationSegmentation Of Optic Disc And Macula In Retinal Images
Segmentation Of Optic Disc And Macula In Retinal Images Gogila Devi. K #1, Vasanthi. S *2 # PG Student, K.S.Rangasamy College of Technology Tiruchengode, Namakkal, Tamil Nadu, India. * Associate Professor,
More informationVersion 1.0. th March 2011
Optical Coherence Tomography Scan and Retinal Imagingg Version 1.0 http://www.ukbiobank.ac.uk/ 5 th March 2011 This manual details the procedure for Scan and Retinal Imagingg at an Assessment Centre of
More informationOptical Coherence Tomography. RS-3000 Advance
Optical Coherence Tomography RS-3000 Advance See it in Advance See it in high resolution with the AngioScan* image. SLO Superficial capillary OCT-Angiography (3 x 3 mm) Deep capillary OCT-Angiography (3
More informationVLSI Implementation of Impulse Noise Suppression in Images
VLSI Implementation of Impulse Noise Suppression in Images T. Satyanarayana 1, A. Ravi Chandra 2 1 PG Student, VRS & YRN College of Engg. & Tech.(affiliated to JNTUK), Chirala 2 Assistant Professor, Department
More informationFPGA IMPLEMENTATION OF RSEPD TECHNIQUE BASED IMPULSE NOISE REMOVAL
M RAJADURAI AND M SANTHI: FPGA IMPLEMENTATION OF RSEPD TECHNIQUE BASED IMPULSE NOISE REMOVAL DOI: 10.21917/ijivp.2013.0088 FPGA IMPLEMENTATION OF RSEPD TECHNIQUE BASED IMPULSE NOISE REMOVAL M. Rajadurai
More informationAutomated sub-retinal fluid detection comprising RPE-region using neighbouring pixel connectivity paradigm.
Biomedical Research 2018; Special Issue: S108-S112 ISSN 0970-938X www.biomedres.info Automated sub-retinal fluid detection comprising RPE-region using neighbouring pixel connectivity paradigm. Piyush Mishra
More informationQuantitative Analysis of Noise Suppression Methods of Optical Coherence Tomography (OCT) Images
Quantitative Analysis of Noise Suppression Methods of Optical Coherence Tomography (OCT) Images Chandan Singh Rawat 1, Vishal S. Gaikwad 2 Associate Professor, Dept. of Electronics and Telecommunications,
More informationSEGMENTATION OF BRIGHT REGION OF THE OPTIC DISC FOR EYE DISEASE PREDICTION
RAHUL JADHAV AND MANISH NARNAWARE: SEGMENTATION OF BRIGHT REGION OF THE OPTIC DISC FOR EYE DISEASE PREDICTION DOI: 10.21917/ijivp.2018.0239 SEGMENTATION OF BRIGHT REGION OF THE OPTIC DISC FOR EYE DISEASE
More informationA Review on Image Enhancement Technique for Biomedical Images
A Review on Image Enhancement Technique for Biomedical Images Pankaj V.Gosavi 1, Prof. V. T. Gaikwad 2 M.E (Pursuing) 1, Associate Professor 2 Dept. Information Technology 1, 2 Sipna COET, Amravati, India
More informationClinical evaluation and management of glaucoma is largely
Macular Segmentation with Optical Coherence Tomography Hiroshi Ishikawa, 1,2 Daniel M. Stein, 1 Gadi Wollstein, 1,2 Siobahn Beaton, 1,2 James G. Fujimoto, 3 and Joel S. Schuman 1,2 PURPOSE. To develop
More informationIntroduction. Chapter Aim of the Thesis
Chapter 1 Introduction 1.1 Aim of the Thesis The main aim of this investigation was to develop a new instrument for measurement of light reflected from the retina in a living human eye. At the start of
More informationAn Efficient Pre-Processing Method to Extract Blood Vessel, Optic Disc and Exudates from Retinal Images
An Efficient Pre-Processing Method to Extract Blood Vessel, Optic Disc and Exudates from Retinal Images 1 K. Priya, 2 Dr. N. Jayalakshmi 1 (Research Scholar, Research & Development Centre, Bharathiar University,
More informationSegmentation of Blood Vessels and Optic Disc in Fundus Images
RESEARCH ARTICLE Segmentation of Blood Vessels and Optic Disc in Fundus Images 1 M. Dhivya, 2 P. Jenifer, 3 D. C. Joy Winnie Wise, 4 N. Rajapriya, Department of CSE, Francis Xavier Engineering College,
More informationImage Modeling of the Human Eye
Image Modeling of the Human Eye Rajendra Acharya U Eddie Y. K. Ng Jasjit S. Suri Editors ARTECH H O U S E BOSTON LONDON artechhouse.com Contents Preface xiiii CHAPTER1 The Human Eye 1.1 1.2 1. 1.4 1.5
More informationA Novel Method for Enhancing Satellite & Land Survey Images Using Color Filter Array Interpolation Technique (CFA)
A Novel Method for Enhancing Satellite & Land Survey Images Using Color Filter Array Interpolation Technique (CFA) Suma Chappidi 1, Sandeep Kumar Mekapothula 2 1 PG Scholar, Department of ECE, RISE Krishna
More informationDigital Retinal Images: Background and Damaged Areas Segmentation
Digital Retinal Images: Background and Damaged Areas Segmentation Eman A. Gani, Loay E. George, Faisel G. Mohammed, Kamal H. Sager Abstract Digital retinal images are more appropriate for automatic screening
More informationHybrid Method based Retinal Optic Disc Detection
Hybrid Method based Retinal Optic Disc Detection Arif Muntasa 1, Indah Agustien Siradjuddin, and Moch Kautsar Sophan 3 Informatics Department, University of Trunojoyo Madura, Bangkalan Madura Island, Indonesia
More informationAutomatic High Dynamic Range Image Generation for Dynamic Scenes
Automatic High Dynamic Range Image Generation for Dynamic Scenes IEEE Computer Graphics and Applications Vol. 28, Issue. 2, April 2008 Katrien Jacobs, Celine Loscos, and Greg Ward Presented by Yuan Xi
More informationOptical Coherence Tomography. RS-3000 Advance / Lite
Optical Coherence Tomography RS-3000 Advance / Lite See it in Advance See it in high resolution with the AngioScan* image. SLO Superficial capillary OCT-Angiography (3 x 3 mm) Deep capillary OCT-Angiography
More informationPreprocessing and Segregating Offline Gujarati Handwritten Datasheet for Character Recognition
Preprocessing and Segregating Offline Gujarati Handwritten Datasheet for Character Recognition Hetal R. Thaker Atmiya Institute of Technology & science, Kalawad Road, Rajkot Gujarat, India C. K. Kumbharana,
More informationOptical Coherence Tomography. RS-3000 Advance / Lite
Optical Coherence Tomography RS-3000 Advance / Lite See it in Advance See it in high resolution with the AngioScan* image. OCT-Angiography of choroidal neovascularization * AngioScan (OCT-Angiography)
More informationRefined Slanted-Edge Measurement for Practical Camera and Scanner Testing
Refined Slanted-Edge Measurement for Practical Camera and Scanner Testing Peter D. Burns and Don Williams Eastman Kodak Company Rochester, NY USA Abstract It has been almost five years since the ISO adopted
More informationIntroduction Approach Work Performed and Results
Algorithm for Morphological Cancer Detection Carmalyn Lubawy Melissa Skala ECE 533 Fall 2004 Project Introduction Over half of all human cancers occur in stratified squamous epithelia. Approximately one
More informationOPTICAL COHERENCE TOMOGRAPHY: OCT supports industrial nondestructive depth analysis
OPTICAL COHERENCE TOMOGRAPHY: OCT supports industrial nondestructive depth analysis PATRICK MERKEN, RAF VANDERSMISSEN, and GUNAY YURTSEVER Abstract Optical coherence tomography (OCT) has evolved to a standard
More informationDevelopment of a Semi-Automatic Segmentation Method for Retinal OCT Images Tested in Patients with Diabetic Macular Edema
Development of a Semi-Automatic Segmentation Method for Retinal OCT Images Tested in Patients with Diabetic Macular Edema Yijun Huang 1,2 *, Ronald P. Danis 1,2, Jeong W. Pak 1, Shiyu Luo 1, James White
More informationSpeckle Noise Reduction for the Enhancement of Retinal Layers in Optical Coherence Tomography Images
Iranian Journal of Medical Physics Vol. 12, No. 3, Summer 2015, 167-177 Received: February 25, 2015; Accepted: July 8, 2015 Original Article Speckle Noise Reduction for the Enhancement of Retinal Layers
More informationOur vision is foresight
Our vision is foresight iseries OCT Systems The Optovue iseries Improving OCT performance with ease Who ever said advanced OCT scanning had to be complicated? When an OCT design puts user experience first,
More informationKeywords Fuzzy Logic, ANN, Histogram Equalization, Spatial Averaging, High Boost filtering, MSE, RMSE, SNR, PSNR.
Volume 4, Issue 1, January 2014 ISSN: 2277 128X International Journal of Advanced Research in Computer Science and Software Engineering Research Paper Available online at: www.ijarcsse.com An Image Enhancement
More informationOptical Coherence Tomography. RS-3000 Advance 2
Optical Coherence Tomography RS-3000 Advance 2 -Providing a comprehensive solution for retina and glaucom Retina Analysis Retinal mode Glaucoma Analysis Choroidal mode Image courtesy of Hokkaido University
More informationAutomation of Fingerprint Recognition Using OCT Fingerprint Images
Journal of Signal and Information Processing, 2012, 3, 117-121 http://dx.doi.org/10.4236/jsip.2012.31015 Published Online February 2012 (http://www.scirp.org/journal/jsip) 117 Automation of Fingerprint
More informationTEPZZ A_T EP A1 (19) (11) EP A1. (12) EUROPEAN PATENT APPLICATION published in accordance with Art.
(19) TEPZZ 769666A_T (11) EP 2 769 666 A1 (12) EUROPEAN PATENT APPLICATION published in accordance with Art. 13(4) EPC (43) Date of publication: 27.08.14 Bulletin 14/3 (21) Application number: 128927.3
More informationMacula centred, giving coverage of the temporal retinal. Disc centred. Giving coverage of the nasal retina.
3. Field positions, clarity and overall quality For retinopathy screening purposes in England two images are taken of each eye. These have overlapping fields of view and between them cover the main area
More informationAutomatic optical measurement of high density fiber connector
Key Engineering Materials Online: 2014-08-11 ISSN: 1662-9795, Vol. 625, pp 305-309 doi:10.4028/www.scientific.net/kem.625.305 2015 Trans Tech Publications, Switzerland Automatic optical measurement of
More informationLocating Blood Vessels in Retinal Images by Piece-wise Threshold Probing of a Matched Filter Response
Locating Blood Vessels in Retinal Images by Piece-wise Threshold Probing of a Matched Filter Response Adam Hoover, Ph.D. +, Valentina Kouznetsova, Ph.D. +, Michael Goldbaum, M.D. + Electrical and Computer
More informationDETECTION OF OPTIC DISC BY USING THE PRINCIPLES OF IMAGE PROCESSING
DETECTION OF OPTIC DISC BY USING THE PRINCIPLES OF IMAGE PROCESSING SUSHMA G 1, VENKATESHAPPA 2 ' 1 Asst professor, 2 HoD, Dept of ECE, MSEC Bangalore E-mail: sushmavasu11@gmail.com, venkat_harishith@rediffmail.com
More informationModifications of the coherence radar for in vivo profilometry in dermatology
Modifications of the coherence radar for in vivo profilometry in dermatology P. Andretzky, M. W. Lindner, G. Bohn, J. Neumann, M. Schmidt, G. Ammon, and G. Häusler Physikalisches Institut, Lehrstuhl für
More informationEnhancement of Optical Coherence Tomography Images of the Retina by Normalization and Fusion
86 F. ROSSANT, F. AMIEL, T. EA, C. MARAVER MARTINEZ, M. PAQUES, ENHANCEMENT OF OPTICAL COHERENCE... Enhancement o Optical Coherence Tomography Images o the Retina by Normalization and Fusion Florence ROSSANT,
More informationIris Recognition using Histogram Analysis
Iris Recognition using Histogram Analysis Robert W. Ives, Anthony J. Guidry and Delores M. Etter Electrical Engineering Department, U.S. Naval Academy Annapolis, MD 21402-5025 Abstract- Iris recognition
More informationLaser Sources for Frequency-Domain Optical Coherence Tomography FD-OCT
Laser Sources for Frequency-Domain Optical Coherence Tomography FD-OCT Photonic Sensing Workshop SWISSLaser.Net Biel, 17. 9. 2009 Ch. Meier 1/ 20 SWISSLASER.NET Ch. Meier 17.09.09 Content 1. duction 2.
More informationProcedure to detect anatomical structures in optical fundus images
Procedure to detect anatomical structures in optical fundus images L. Gagnon *a, M. Lalonde *a, M. Beaulieu *a, M.-C. Boucher **b a Computer Research Institute of Montreal; b Dept. Of Ophthalmology, Maisonneuve-Rosemont
More informationGeneral Bayesian estimation for speckle noise reduction in optical coherence tomography retinal imagery
General Bayesian estimation for speckle noise reduction in optical coherence tomography retinal imagery Alexander Wong 1 *, Akshaya Mishra 1, Kostadinka Bizheva 2, and David A. Clausi 1 1 Dept. of Systems
More informationAn Improved Bernsen Algorithm Approaches For License Plate Recognition
IOSR Journal of Electronics and Communication Engineering (IOSR-JECE) ISSN: 78-834, ISBN: 78-8735. Volume 3, Issue 4 (Sep-Oct. 01), PP 01-05 An Improved Bernsen Algorithm Approaches For License Plate Recognition
More informationMultichannel Blind Deconvolution in Eye Fundus Imaging
Multichannel Blind Deconvolution in Eye Fundus Imaging Andrés G. Marrugo Dept. of Optics and Optometry Universitat Politècnica de Catalunya, Spain andres.marrugo@upc.edu Filip Šroubek UTIA Academy of Sciences
More informationA Method of Using Digital Image Processing for Edge Detection of Red Blood Cells
Sensors & Transducers 013 by IFSA http://www.sensorsportal.com A Method of Using Digital Image Processing for Edge Detection of Red Blood Cells 1 Jinping LI, Hongshan MU, Wei XU 1 Software School, East
More informationMoving from biomedical to industrial applications: OCT Enables Hi-Res ND Depth Analysis
Moving from biomedical to industrial applications: OCT Enables Hi-Res ND Depth Analysis Patrick Merken a,c, Hervé Copin a, Gunay Yurtsever b, Bob Grietens a a Xenics NV, Leuven, Belgium b UGENT, Ghent,
More informationTarget detection in side-scan sonar images: expert fusion reduces false alarms
Target detection in side-scan sonar images: expert fusion reduces false alarms Nicola Neretti, Nathan Intrator and Quyen Huynh Abstract We integrate several key components of a pattern recognition system
More informationImpulse noise features for automatic selection of noise cleaning filter
Impulse noise features for automatic selection of noise cleaning filter Odej Kao Department of Computer Science Technical University of Clausthal Julius-Albert-Strasse 37 Clausthal-Zellerfeld, Germany
More informationFrequency Domain Median-like Filter for Periodic and Quasi-Periodic Noise Removal
Header for SPIE use Frequency Domain Median-like Filter for Periodic and Quasi-Periodic Noise Removal Igor Aizenberg and Constantine Butakoff Neural Networks Technologies Ltd. (Israel) ABSTRACT Removal
More informationImage Processing for feature extraction
Image Processing for feature extraction 1 Outline Rationale for image pre-processing Gray-scale transformations Geometric transformations Local preprocessing Reading: Sonka et al 5.1, 5.2, 5.3 2 Image
More informationDespeckling vs Averaging of retinal UHROCT tomograms: advantages and limitations
Despeckling vs Averaging of retinal UHROCT tomograms: advantages and limitations Justin A. Eichel 1, Donghyun D. Lee 2, Alexander Wong 1, Paul W. Fieguth 1, David A. Clausi 1, and Kostadinka K. Bizheva
More informationCSE 564: Visualization. Image Operations. Motivation. Provide the user (scientist, t doctor, ) with some means to: Global operations:
Motivation CSE 564: Visualization mage Operations Klaus Mueller Computer Science Department Stony Brook University Provide the user (scientist, t doctor, ) with some means to: enhance contrast of local
More informationRestoration of Motion Blurred Document Images
Restoration of Motion Blurred Document Images Bolan Su 12, Shijian Lu 2 and Tan Chew Lim 1 1 Department of Computer Science,School of Computing,National University of Singapore Computing 1, 13 Computing
More informationyou can See more with fundus autofluorescence. CR-2 PLUS NON-MYDRIATIC DIGITAL RETINAL CAMERA
you can See more with fundus autofluorescence. CR-2 PLUS NON-MYDRIATIC DIGITAL RETINAL CAMERA High-performance non-mydriatic retinal camera with integrated FAF photography. Canon EOS Camera Technology
More informationMain Subject Detection of Image by Cropping Specific Sharp Area
Main Subject Detection of Image by Cropping Specific Sharp Area FOTIOS C. VAIOULIS 1, MARIOS S. POULOS 1, GEORGE D. BOKOS 1 and NIKOLAOS ALEXANDRIS 2 Department of Archives and Library Science Ionian University
More informationPostprocessing of nonuniform MRI
Postprocessing of nonuniform MRI Wolfgang Stefan, Anne Gelb and Rosemary Renaut Arizona State University Oct 11, 2007 Stefan, Gelb, Renaut (ASU) Postprocessing October 2007 1 / 24 Outline 1 Introduction
More informationFundus Photograph Reading Center
Autofluorescence Using Confocal Scanning Laser Ophthalmoscope (cslo) Instruments (AF-D) 8010 Excelsior Drive, Suite 100, Madison WI 53717 Telephone: (608) 410-0560 Fax: (608) 410-0566 Table of Contents
More information