Effect of an aspheric intraocular lens on the ocular wave-front adjusted for pupil size and capsulorhexis size

Size: px
Start display at page:

Download "Effect of an aspheric intraocular lens on the ocular wave-front adjusted for pupil size and capsulorhexis size"

Transcription

1 Effect of an aspheric intraocular lens on the ocular wave-front adjusted for pupil size and capsulorhexis size Alja Crnej, 1 Wolf Buehl, 2 Roman Greslechner, 2 Nino Hirnschall 1,3 and Oliver Findl 1,3 1 Moorfields Eye Hospital NHS Foundation Trust, London, UK 2 Department of Ophthalmology, Medical University of Vienna, Vienna, Austria 3 Department of Ophthalmology, VIROS Vienna Institute for Research in Ocular Surgery (VIROS), a Karl Landsteiner Institute, Vienna, Austria ABSTRACT. Purpose: To compare the ocular wave-front of eyes with silicone Intraocular lens (IOLs) with aspheric and spherical optics after cataract surgery, taking into account the patient s pupil size under reading conditions and after pupil dilatation. Methods: In this institutional prospective, randomized, controlled, patient and examiner masked, bilateral trial with intra-individual comparison, 60 eyes of 30 patients with bilateral age-related cataract were included. Each patient received a spherical IOL (CeeOn Edge, 911A, AMO, Santa Ana, CA, USA) in one eye and an aspheric IOL (Tecnis, Z9000, AMO) in the contra-lateral eye. Exclusion criteria were other ocular pathologies, capsular changes or zonular weakness. The main outcome variable was spherical aberration of the ocular wave-front under mesopic pupil conditions measured 2 years after surgery. Additional outcome variables were visual acuity and photopic and mesopic contrast sensitivity. Results: There was no significant difference in visual acuity between the two IOL types under physiological pupil conditions and also not after pupil dilation. However, spherical aberrations were significantly lower with the aspheric IOL (SA: spherical 0.38 lm, SD: 0.11 lm; aspheric 0.10 lm, SD: 0.13 lm; p < 0.01), and there was a significant difference in contrast sensitivity at 12 cycles/degree. Conclusions: This is the first randomized and masked trial on visual function and ocular wave-front after implantation with this silicone aspheric IOL, taking the patients own pupil size into account. The effect on visual function was detectable for mesopic contrast sensitivity, but there was no difference in visual acuity. The SA was found to be significantly lower under physiological pupil conditions as well as when recalculated for the rhexis size and under pharmacological dilatation. Key words: aspheric intraocular lens pupil size wave-front Acta Ophthalmol. 2014: 92: e353 e357 ª 2014 Acta Ophthalmologica Scandinavica Foundation. Published by John Wiley & Sons Ltd doi: /aos Introduction A goal of cataract surgery today is not only to restore visual acuity but also to provide the best quality of vision possible. The cornea typically induces positive spherical aberration, which in young people is compensated for by the negative spherical aberration of the crystalline lens. As the eye ages, spherical aberration of the lens shifts towards less negative or even positive values (Smith et al. 2001), which when combined with the positive corneal spherical aberration (Glasser & Campbell 1998; Guirao et al. 2002; Packer, Fine & Hoffman 2004), results in an increased total spherical aberration and decreased ocular optical quality (Brunette et al. 2003; He et al. 2003; Montes- Mico, Ferrer-Blasco & Cervino 2009). Intraocular lens (IOL) implantation after cataract surgery provides excellent visual acuity outcomes, but contrast sensitivity in pseudophakic eyes is lower than in normal, elderly presbyopic eyes. Aspheric IOLs with a modified optic geometry induce negative spherical aberration and were designed to compensate for the spherical aberration of the cornea to improve optical performance. The aim of this study was to compare the optical performance of an aspheric IOL with a modified anterior surface to that of a standard spherical IOL of an otherwise similar design. Outcome variables were visual acuity, photopic and mesopic contrast sensitivity as well as measurements of the ocular wave-front. Patients and Methods This prospective randomized singlecentre study was approved by the local ethics committee and adhered to the e353

2 guidelines of the tenets of Declaration of Helsinki. The aspheric IOL (AIOL) (Tecnis Z9000; AMO, Santa Ana, CA, USA) was compared to the spherical IOL (SIOL) (CeeOn Edge 911A, AMO) in this randomized, controlled, doublemasked, bilateral trial with intra-individual comparison. Both IOLs are foldable, three-piece, open-loop silicone IOLs with a 6.0 mm, biconvex optic and square-edge design. The only difference between the IOLs is a modified anterior surface of the optic in the aspheric lens, which has a flatter curvature in the periphery. Thirty patients with bilateral agerelated cataract were included after providing consent. Exclusion criteria consisted of retinal pathology with a likely expected postoperative visual acuity outcome of 0.8 Snellen or worse, as well as any capsular changes that would increase the risk of postoperative decentration or tilt of the IOL, such as pseudoexfoliation syndrome or zonular weakness. The first eye to be operated on was randomly assigned to receive either an AIOL or a SIOL using a sealed-envelope randomization method. The fellow eye received the alternate IOL type and had surgery within the next 2 weeks. Surgery was performed in a standardized way by an experienced surgeon. The surgical technique included application of topical anaesthesia, a 3.0-mm temporal limbal incision, capsulorhexis, phacoemulsification, irrigation/aspiration of cortical remnants and IOL implantation in the capsular bag with an injector. In addition to the regular check-up, a followup examination was performed 2 years postoperatively. At 2 years, the patient s distant corrected visual acuity (DCVA) was determined using the backlit Early Treatment Diabetic Retinopathy Study (ETDRS) chart (Lighthouse International, New York, NY, USA) at 4.0 m with a 100% contrast chart and then with a 5% contrast chart. Under these conditions, pupil size was measured using a Colvard pupillometer (Oasis Medical, Glendora, USA). While measuring pupil size in one eye, the other eye was fixating on the ETDRS chart. To evaluate the effect of the pupil size on visual acuity, the tests were repeated as described above following pharmacological dilatation of the pupil, using phenylephrine 2.5% and tropicamide 0.5%. Contrast sensitivity (CS) was measured using the FACT chart (Ginsburg Box, VSCR-CST-6500; Vision Science Research Corporation, Walnut Creek, CA, USA) under mesopic lighting conditions (6 cd/m²). The FACT chart consists of five rows (A to E) of sinewave gratings with increasing frequency values of 1.5, 3, 6, 12 and 18 cycles/degree and nine columns with decreasing contrast levels. The patient reports the orientation of the gratings: right, up or left. The last correct grating seen for each spatial frequency is plotted on a contrast sensitivity curve. Wave-front aberrations were measured with a Hartmann Shack sensor (WASCA Analyzer; Carl Zeiss Meditec, Jena, Germany) and calculated for three different analysis diameters. All measurements were performed under maximal mydriasis, and the wave-front errors were described according to the OSA (Optical Society of America) notation for Zernike polynomials. At least three measurements were taken per eye, and the mean values of the Zernike coefficients (up to the 5th order) and root mean square (RMS) [of the total and higher order aberrations (HOA s)] were calculated. Concerning the diameter of analysis, Zernike polynomials were recomputed for each patient s individual pupil diameter (natural pupil size) measured during the ETDRS testing, as this was the main outcome measure. The pupil diameter was measured in each eye under visual acuity measurement conditions and then taken as diameter of analysis for the respective wave-front measurement. The Zernike polynomials were also recomputed for the individual rhexis diameters as measured from the retroillumination photograph to investigate the influence of the IOL/ rhexis overlap on the ocular wavefront. Finally, the wave-front was measured under maximal pharmacological pupil dilatation. Although dilated pupil size and rhexis size are not physiological diameters of the pupil, these diameters were included to assess aberration scores under extreme conditions. However, the main outcome measure was determined at physiological pupil size. Slitlamp examination included investigation of the occurrence of capsular folds, IOL decentration, buttonholing (none, partial, half or full), extent of anterior fibrosis (ACO) and posterior capsule opacification (PCO). The PCO intensity was assessed using digital retroilluminated photographs and Automated Quantification of After-Cataract (AQUA) (Findl 2003) software. From the photos taken, rhexis diameter and rhexis area were analysed. For statistical analysis, Microsoft Excel 2011 for Mac (Microsoft, Redmond, WA, USA) with a Statplus:mac version plug-in (AnalystSoft, Alexandria, VA, USA) and a Xlstat 2012 plug-in (Addinsoft, Paris, France) were used. Descriptive data are always shown as mean, standard deviation (SD) and range, if not stated otherwise. A 2-sided paired t-test was used to assess significance in differences between both IOLs in BDVA. P-values < 0.05 were considered statistically significant. Multiple testing with a Bonferroni correction was used to assess difference in contrast sensitivity and wave-front aberrations. The Bonferroni-corrected significance value was Results Thirty patients with bilateral cataracts were recruited for this trial. The average age of the patients was 76.0 years (range: years). Twenty-one patients came for the 2-year examination. Of the nine patients that were lost to follow-up, three were previously excluded due to exclusion criteria, one patient had died, three could not be contacted and two refused to come for the examination. Concerning visual acuity, no significant difference was found between the two IOLs either under natural or pharmacological dilation conditions (Table 1). Distance low-contrast visual acuity with the 5% ETDRS chart showed a trend for better acuity with the AIOL under both pupil size conditions; however, this was not statistically significant (Table 1 and Fig. 1). Average contrast sensitivity under mesopic lighting conditions (6 cd/m²) showed no significant differences between the IOLs (p 1.5 = 0.957; p 3 = 0.520; p 6 = 0.453; p 12 = 0.391; p 18 = 0.789). However, comparing single pairs of each spatial frequency showed a significant difference at 12 cycles per degree (p (Wilcoxon rank test)=0.02) (Fig. 2.) e354

3 Table 1. Visual acuity (VA) as the number of correctly read letters on the ETDRS chart and distance low-contrast VA with the 5% ETDRS chart. No significant difference was found between IOLs VA Distance low-contrast VA Pupil size Physiological Dilated Physiological Dilated SIOL 58.9 (SD: 5.5, from 52.0 to 68.0) AIOL 60.6 (SD: 5.2, from 51.0 to 68.0) 57.0 (SD: 6.3, from 51.0 to 67.0) 57.1 (SD: 5.7, from 47.0 to 67.0) 44.1 (SD: 6.3, from 34.0 to 53.0) 44.2 (SD: 5.7, from 29.0 to 53.0) 34.2 (SD: 10.1, from 10.0 to 47.0) 37.9 (SD: 7.3, from 19.0 to 53.0) Fig. 1. Visual acuity under physiological pupil conditions and with dilated pupils for spherical and aspheric IOL on the 100% ETDRS (a) and the 5% ETDRS (b) chart. Circles ( ) are outliers, asterisks ( ) are extreme values. The mean pupil size under visual acuity testing in dim room light was 4.05 mm (SD: 0.85 mm, range from 3.00 to 6.50 mm) in SIOL group and 4.07 (SD: 0.87 mm, range from 3.00 to 6.50 mm) in AIOL group. Under those natural pupil conditions, only spherical aberration (SA) was lower in the aspheric eyes when assessing HOA (Table 2). The mean rhexis diameter was 4.12 mm (SD: 0.53 mm range from 3.50 to 5.75 mm) for the spherical eyes and 4.06 mm (SD: 0.56 mm range from 2.84 to 5.25 mm) for the aspheric Fig. 2. Contrast sensitivity under mesopic lighting conditions (6 cd/m²) for spherical and aspheric IOL (a) and intra-patient difference in contrast sensitivity for both IOLs (b). eyes. Using this analysis diameter, in which the anterior capsule does not influence wave-front analysis, significant differences were found for SA (Table 2). Wave-front errors were largest when measured in mydriasis. Significant differences between the two IOL types were found for SA (Table 2 and Fig. 3), but not when analysing the data for each module, such as for the combined horizontal and vertical trefoil as well as horizontal and vertical coma. Concerning the slitlamp examination, capsular folds were found in only one eye that had received an AIOL. Fig. 3. Differences in all higher order aberrations (HOAs) for natural pupil size (a) and dilated pupils (b). A statistically significant difference was found only in spherical aberration (SA). Circles ( ) are outliers, asterisks ( ) are extreme values. None of the IOLs in either of the groups was visibly decentered. Opticrhexis buttonholing occurred once in each group. Anterior fibrosis, especially of mild extent, occurred in more than 90% of all eyes (spherical group: 19 eyes; AIOL group: 20 eyes), and no significant differences were found between the two IOL groups. Regenerative PCO was either not visible or mild, with a mean severity score (range 0 10, measured with the software) of 0.38 and 0.26 for spherical and aspheric eyes, respectively [p = 0.46]. Discussion The key findings from similar previous trials and the present one comparing e355

4 Table 2. Significant differences that were found in measuring wave-front errors under physiological pupil diameter (PPD), rhexis diameter (RD) and mydriasis (My) aspheric and spherical IOLs are summarized in Table 3. Similar to the majority of published data, this study found no difference in DCVA between spheric (CeeOn Edge, 911A) and aspheric (Tecnis Z9000) IOLs. There was no statistically significant difference in low-contrast DCVA and, surprisingly, no difference in DCVA under mydriasis. A significantly better outcome of contrast sensitivity was found only at a spatial frequency of 12 cycles/degree for the aspheric eyes in mesopic conditions. For all the other spatial frequencies, the aspheric lens showed better contrast sensitivity, but none of the differences was statistically significant. Only in the study by Denoyer et al. comparing the same IOLs, a statistically significant difference between the two lenses was found at all spatial frequencies. Two-thirds of studies SIOL AIOL p-value PPD SA 0.10 lm 0.01 lm < Z(4,0) SD: 0.11 lm SD: 0.02 lm RD SA 0.10 lm 0.01 lm < Z(4,0) SD: 0.07 lm SD: 0.02 lm My SA 0.38 lm 0.10 lm < Z(4,0) SD: 0.11 lm SD: 0.13 lm AIOL, aspheric IOL; SIOL, spherical IOL; and SA, spherical aberration. investigating similar (but not the same) lens design did not report statistical significant difference in contrast sensitivity between the IOLs (Table 3). In assessing the difference in HOA between AIOLs and SIOLs, all studies found statistically significant differences in SA for all pupil diameters, which is consistent with our results. The study by Mu~noz et al. (2006), Denoyer et al. (2007), on the other hand, is the only study reporting a significant difference for total coma for 4- and 6-mm pupil sizes. Studies are also reporting different outcomes for HOA RMS. Three studies found a significant difference for all tested pupil diameters, one only for a large pupil diameter and two, including this study, did not find any statistical difference between the IOLs. Significant differences in SA but only a minor impact on contrast sensitivity are also represented in a post hoc power analysis, which showed that the power of this study was >90% for differences in spherical aberrations between the groups, but only 20% for differences in contrast sensitivity. In a review article about possible benefits of AIOLs by Montes-Mico et al. 2009; most studies calculated the wave-front for a fixed pupil size and not for the individual pupil diameter, as we have carried out in the present study. The influence of pupil size on optical quality of aspheric IOLs was shown by Tabernero et al. 2006, 2009; in only a small percentage of patients. Therefore, the current study confirms those findings suggesting that aspheric IOLs show good results in most patients regardless of pupil size. Standardized measurements for evaluating AIOLs would be necessary to allow a better comparison of results among different trials. Moreover, accurate measurements of tilt and decentration are needed to assess the effect of lens and haptic design on wave-front aberrations. Aspheric IOLs were developed to correct asphericity in eyes with a normal amount of HOAs. So, we would ideally measure HOAs of each patient preoperatively to identify those patients who show very low HOAs and would not benefit from an aspheric IOL, or even be at risk of induction of even worse HOAs. In summary, the improvement in contrast sensitivity in this study was not as pronounced as Table 3. Summary of key findings from similar previous trials and the present study comparing aspheric and spherical IOLs CS HOA BCVA Mesopic Photopic Scotopic Pupil diameter SA Trefoil Total coma Tetrafoil HOA RMS Kasper et al. (2006) NS 3,5 and 6 mm S NS NS NS S (6)* Denoyer et al. (2007) NS S NS 5 mm S Bellucci et al. (2007a) 4 mm S NS S The Moorfields IOL study group (2007) NS NS 4 mm S S Mester et al. (2003) S S S 4 mm S Ohtani et al. (2009) NS NS NS S 4 mm S NS NS Rocha et al. (2006) NS Ph S NS NS NS NS Awwad et al. (2008) NS S** NS 4,5 and 6 mm S NS NS NS S Bellucci et al. (2007b) S S S Mu~noz et al. (2006) NS NS NS 4 and 6 mm S NS S NS Crnej et al. (2011) NS S*** Ph, Rh, My S NS NS NS NS S, significant difference; NS, no significant difference; BCVA, best corrected visual acuity; CS, contrast sensitivity; HOA, higher order aberrations; SA, spherical aberration; RMS, root mean square; Ph, physiological pupil size; Rh, rhexis diameter; and My, mydriasis. * Only for certain pupil sizes. ** At 12 and 18 cycles/degree. *** At 12 cycles/degree. Studies comparing the same IOLs as the current study. e356

5 reported in previous studies; however, a significant improvement could be shown for the spatial frequency of 12 cycles/degree, and the study confirmed the majority of data already published that eyes with implanted aspheric IOLs don t have better visual acuity than spherical IOLs. Acknowledgements Funding and financial disclosures The authors indicate no financial support. None of the authors has any financial disclosures. OF is a member of a scientific advisory board of Abott Medical Optics. He has no financial interest in any of the products mentioned. References Awwad ST, Warmerdam D, Bowman RW, Dwarakanathan S, Cavanagh HD & McCulley JP (2008): Contrast sensitivity and higher order aberrations in eyes implanted with Acrysof IQ SN60WF and Acrysof SN60AT intraocular lenses. J Refract Surg 24: Bellucci R, Morselli S & Pucci V (2007a): Spherical aberration and coma with an aspheric and a spherical intra-ocular lens in normal age-matched eyes. J Cataract Refract Surg 33: Bellucci R, Scialdone A, Buratto L, Morselli S, Chierego C, Criscuoli A, Moretti G & Piers P (2007b): Visual acuity and contrast sensitivity comparison between Tecnis and Acry- Sof SA60AT intraocular lens: A multicenter randomized study. J Cataract Refract Surg 31: Brunette I, Bueno JM, Parent M, Hamam H & Simonet P (2003): Monochromatic aberrations as a function of age, from childhood to advanced age. Invest Ophthalmol Vis Sci 44: Crnej A, Hirnschall N, Nishi Y, Gangwani V, Tabernero J, Artal P & Findl O (2011): Impact of intraocular lens haptic design and orientation on decentration and tilt. J Cataract Refract Surg 37: Denoyer A, Le Lez ML, Majzoub S & Pisella PJ (2007): Quality of vision after cataract surgery after Tecnis Z9000 intraocular lens implantation; Effect of contrast sensitivity and wave-front aberration improvements on the quality of daily vision. J Cataract Refract Surg 33: Findl O, Buehl W, Menapace R, Georgopoulos M, Rainer G, Siegl H, Kaider A & Prinz A (2003): Comparison of 4 methods for quantifying posterior capsule opacification. J Cataract Refract Surg 29: Glasser A & Campbell MC (1998): Presbyopia and the optical changes in the human crystalline lens with age. Vision Res 38: Guirao A, Redondo M, Geraghty E, Piers P, Norrby S & Artal P (2002): Corneal optical aberrations and retinal image quality in patients in whom monofocal intraocular lenses were implanted. Arch Ophthalmol 120: He JC, Gwiazda J, Thorn F & Held R (2003): Wave-front aberrations in the anterior corneal surface and the whole eye. J Opt Soc Am A Opt Image Sci Vis 20: Kasper T, B uhren J & Kohnen T (2006): Intraindividual comparison of higher-order aberrations after implantation of aspheric and spherical intraocular lenses as a function of pupil diameter. J Cataract Refract Surg 32: Mester U, Dillinger P & Anterist N (2003): Impact of a modified optic design on visual function: Clinical comparative study. J Cataract Refract Surg 29: Montes-Mico R, Ferrer-Blasco T & Cervi~no A (2009): Analysis of the possible benefits of aspheric intraocular llenses: review of the literature. J Cataract Refract Surg 35: Mu~noz G, Albarran-Diego C, Montes-Mico R, Rodriguez-Galietero A & Alio JL (2006): Spherical aberration and contrast sensitivity after cataract surgery with the Tecnis Z9000 intraocular lens. J Cataract Refract Surg 32: Ohtani S, Gekka S, Honbou M, Kataoka Y, Minami K, Miyata K & Oshika T (2009): One-year prospective intra-patient comparison of aspherical and spherical intraocular lenses in patients with bilateral cataract. Am J Ophthalmol 147: Packer M, Fine IH & Hoffman RS (2004): Wave-front technology in cataract surgery. Curr Opin Ophthalmol 15: Rocha KM, Soriano ES, Chalita MR, Yamada AC, Bottos K, Bottos J, Morimoto L & Nose W (2006): Wave-front analysis and contrast sensitivity of aspheric and spherical intraocular lenses: a randomized prospective study. Am J Ophthalmol 142: Smith G, Cox MJ, Calver R & Garner LF (2001): The spherical aberration of the crystalline lens of the human eye. Vision Res 41: Tabernero J, Piers P, Benito A, Redondo M & Artal P (2006): Predicting the optical performance of eyes implanted with IOLs to correct spherical aberration. Invest Ophthalmol Vis Sci 47: Tabernero J, Atchoson DA & Makrwell EL (2009): Aberrations and Pupil location under corneal topography and Hartmann- Shack illumination conditions. Invest Ophthalmol Vis Sci 50: The Moorfields IOL study group (2007): Binocular implantation of the Tecnis Z9000 or Acrysof MA60AC intraocular lens in routine cataract surgery. Prospective randomized controlled trial comparing VF-14 scores. J Cataract Refract Surg 33: Received on April 24th, Accepted on December 10th, Correspondence: Oliver Findl, MD, MBA Department of Ophthalmology VIROS Vienna Institute for Research in Ocular Surgery (VIROS), a Karl Landsteiner Institute Heinrich-Collin-Straße Vienna Austria Tel: Fax: oliver@findl.at e357

Visual Outcomes of Two Aspheric PCIOLs: Tecnis Z9000 versus Akreos AO

Visual Outcomes of Two Aspheric PCIOLs: Tecnis Z9000 versus Akreos AO Visual Outcomes of Two Aspheric PCIOLs: Tecnis Z9000 versus Akreos AO Ahmad-Reza Baghi, MD; Mohammad-Reza Jafarinasab, MD; Hossein Ziaei, MD; Zahra Rahmani, MD Shaheed Beheshti Medical University, Tehran,

More information

University of Groningen. Young eyes for elderly people van Gaalen, Kim

University of Groningen. Young eyes for elderly people van Gaalen, Kim University of Groningen Young eyes for elderly people van Gaalen, Kim IMPORTANT NOTE: You are advised to consult the publisher's version (publisher's PDF) if you wish to cite from it. Please check the

More information

Objective and subjective outcomes in comparing three different aspheric intraocular lens implants with their spherical counterparts

Objective and subjective outcomes in comparing three different aspheric intraocular lens implants with their spherical counterparts (2009) 23, 877 883 & 2009 Macmillan Publishers Limited All rights reserved 0950-222X/09 $32.00 www.nature.com/eye Objective and subjective outcomes in comparing three different aspheric intraocular lens

More information

Aberrations Before and After Implantation of an Aspheric IOL

Aberrations Before and After Implantation of an Aspheric IOL Ocular High Order Aberrations Before and After Implantation of an Aspheric IOL Fabrizio I. Camesasca, MD Massimo Vitali, Orthoptist Milan, Italy I have no financial interest to disclose Wavefront Measurement

More information

Comparison of Visual Acuity, Contrast Sensitivity and Spherical Aberration after Implantation of Aspheric and Spheric Intraocular Lenses

Comparison of Visual Acuity, Contrast Sensitivity and Spherical Aberration after Implantation of Aspheric and Spheric Intraocular Lenses Comparison of Visual Acuity, Contrast Sensitivity and Spherical Aberration after Implantation of Aspheric and Spheric Intraocular Lenses Mohammad Nasser Hashemian, MD 1 Morteza Movassat, MD 2 Abdolreza

More information

Comparison of higher order aberrations with spherical and aspheric IOLs compared to normal phakic eyes

Comparison of higher order aberrations with spherical and aspheric IOLs compared to normal phakic eyes European Journal of Ophthalmology / Vol. 18 no. 5, 2008 / pp. 728-732 Comparison of higher order aberrations with spherical and aspheric IOLs compared to normal phakic eyes M. RĘKAS, K. KRIX-JACHYM, B.

More information

MODERN CATARACT SURGERY AND LENS REplacement

MODERN CATARACT SURGERY AND LENS REplacement Wavefront Analysis and Contrast Sensitivity of Aspheric and Spherical Intraocular Lenses: A Randomized Prospective Study KAROLINNE MAIA ROCHA, MD, EDUARDO S. SORIANO, MD, MARIA REGINA CHALITA, MD, ANA

More information

Analysis of the possible benefits of aspheric intraocular lenses: Review of the literature

Analysis of the possible benefits of aspheric intraocular lenses: Review of the literature REVIEW/UPDATE Analysis of the possible benefits of aspheric intraocular lenses: Review of the literature Robert Montés-Micó, OD, MPhil, PhD, Teresa Ferrer-Blasco, OD, MSc, PhD, Alejandro Cerviño, OD, PhD

More information

Role of Asphericity in Choice of IOLs for Cataract Surgery

Role of Asphericity in Choice of IOLs for Cataract Surgery Role of Asphericity in Choice of IOLs for Cataract Surgery Delhi J Ophthalmol 2015; 25 (3): 185-189 DOI: http://dx.doi.org/10.7869/djo.105 Aman Khanna, Rebika Dhiman, Rajinder Khanna, Yajuvendra Singh

More information

University of Groningen. Young eyes for elderly people van Gaalen, Kim

University of Groningen. Young eyes for elderly people van Gaalen, Kim University of Groningen Young eyes for elderly people van Gaalen, Kim IMPORTANT NOTE: You are advised to consult the publisher's version (publisher's PDF) if you wish to cite from it. Please check the

More information

Quality of vision after cataract surgery after Tecnis Z9000 intraocular lens implantation

Quality of vision after cataract surgery after Tecnis Z9000 intraocular lens implantation J CATARACT REFRACT SURG - VOL 33, FEBRUARY 27 Quality of vision after cataract surgery after Tecnis Z9 intraocular lens implantation Effect of contrast sensitivity and wavefront aberration improvements

More information

Abetter understanding of the distribution of aberrations in

Abetter understanding of the distribution of aberrations in Predicting the Optical Performance of Eyes Implanted with IOLs to Correct Spherical Aberration Juan Tabernero, 1 Patricia Piers, 2 Antonio Benito, 1 Manuel Redondo, 3 and Pablo Artal 1 PURPOSE. To use

More information

Corneal Asphericity and Retinal Image Quality: A Case Study and Simulations

Corneal Asphericity and Retinal Image Quality: A Case Study and Simulations Corneal Asphericity and Retinal Image Quality: A Case Study and Simulations Seema Somani PhD, Ashley Tuan OD, PhD, and Dimitri Chernyak PhD VISX Incorporated, 3400 Central Express Way, Santa Clara, CA

More information

Unique Aberration-Free IOL: A Vision that Patients

Unique Aberration-Free IOL: A Vision that Patients Unique Aberration-Free IOL: A Vision that Patients Can Appreciate An Aspheric Optic for Improved Quality of Vision n Traditional spherical IOLs create Bilateral implantation study spherical aberration

More information

This is the author s version of a work that was submitted/accepted for publication in the following source:

This is the author s version of a work that was submitted/accepted for publication in the following source: This is the author s version of a work that was submitted/accepted for publication in the following source: Atchison, David A. & Mathur, Ankit (2014) Effects of pupil center shift on ocular aberrations.

More information

Postoperative Wavefront Analysis and Contrast Sensitivity of a Multifocal Apodized Diffractive IOL (ReSTOR) and Three Monofocal IOLs

Postoperative Wavefront Analysis and Contrast Sensitivity of a Multifocal Apodized Diffractive IOL (ReSTOR) and Three Monofocal IOLs Postoperative Wavefront Analysis and Contrast Sensitivity of a Multifocal Apodized Diffractive IOL (ReSTOR) and Three Monofocal IOLs Karolinne Maia Rocha, MD; Maria Regina Chalita, MD; Carlos Eduardo B.

More information

Clinical Evaluation 3-month Follow-up Report

Clinical Evaluation 3-month Follow-up Report Clinical Evaluation 3-month Follow-up Report Of SeeLens HP Intraocular Lens 27 December 2010 version 1.1 1of 16 Table of Contents TABLE OF CONTENTS... 1 OBJECTIVES... 2 EFFICACY AND SAFETY ASSESSMENTS...

More information

Effects of Pupil Center Shift on Ocular Aberrations

Effects of Pupil Center Shift on Ocular Aberrations Visual Psychophysics and Physiological Optics Effects of Pupil Center Shift on Ocular Aberrations David A. Atchison and Ankit Mathur School of Optometry & Vision Science and Institute of Health & Biomedical

More information

NOW. Approved for NTIOL classification from CMS Available in Quar ter Diopter Powers. Accommodating. Aberration Free. Aspheric.

NOW. Approved for NTIOL classification from CMS Available in Quar ter Diopter Powers. Accommodating. Aberration Free. Aspheric. NOW Approved for NTIOL classification from CMS Available in Quar ter Diopter Powers Accommodating. Aberration Free. Aspheric. Accommodation Meets Asphericity in AO Merging Innovation & Proven Design The

More information

The Aberration-Free IOL:

The Aberration-Free IOL: The Aberration-Free IOL: Advanced Optical Performance Independent of Patient Profile Griffith E. Altmann, M.S., M.B.A.; Keith H. Edwards, BSc FCOptom Dip CLP FAAO, Bausch & Lomb Some of these results were

More information

Update on Aspheric IOL Technology

Update on Aspheric IOL Technology Peer-Reviewed Literature: Update on Aspheric IOL Technology Editor: Ming Wang, MD, PhD, Clinical Associate Professor of Ophthalmology at the University of Tennessee and Director of the Wang Vision Institute

More information

Multifocal Intraocular Lenses for the Treatment of Presbyopia: Benefits and Side-effects

Multifocal Intraocular Lenses for the Treatment of Presbyopia: Benefits and Side-effects Published on Points de Vue International Review of Ophthalmic Optics () Home > Multifocal Intraocular Lenses for the Treatment of Presbyopia: Benefits and Side-effects Multifocal Intraocular Lenses for

More information

Dr. Magda Rau Eye Clinic Cham, Germany

Dr. Magda Rau Eye Clinic Cham, Germany 3 and 6 Months clinical Results after Implantation of OptiVis Diffractive-refractive Multifocal IOL Dr. Magda Rau Eye Clinic Cham, Germany Refractive zone of Progressive power for Far to Intermediate

More information

Comparison of 4 methods for quantifying posterior capsule opacification

Comparison of 4 methods for quantifying posterior capsule opacification Comparison of 4 methods for quantifying posterior capsule opacification Oliver Findl, MD, Wolf Buehl, MD, Rupert Menapace, MD, Michael Georgopoulos, MD, Georg Rainer, MD, Hannes Siegl, MSc, Alexandra Kaider,

More information

*Simulated vision. **Individual results may vary and are not guaranteed. Visual Performance When It s Needed Most

*Simulated vision. **Individual results may vary and are not guaranteed. Visual Performance When It s Needed Most Simulated vision. Individual results may vary and are not guaranteed. Visual Performance When It s Needed Most The aspheric design of the AcrySof IQ IOL results in improved clarity and image quality. The

More information

In Vitro Strehl Ratios with Spherical, Aberration-Free, and customized spherical aberration-correcting METHODS

In Vitro Strehl Ratios with Spherical, Aberration-Free, and customized spherical aberration-correcting METHODS In Vitro Strehl Ratios with Spherical, Aberration-Free, Average, and Customized Spherical Aberration-Correcting Intraocular Lenses Stefan Pieh, 1 Werner Fiala, 1 Andre Malz, 2 and Wilhelm Stork 2 PURPOSE.

More information

The design is distinctive. The outcomes are clear. Defocus tolerance 1 Glistening-free performance 1,2 Predictable outcomes 1

The design is distinctive. The outcomes are clear. Defocus tolerance 1 Glistening-free performance 1,2 Predictable outcomes 1 The design is distinctive. The outcomes are clear. Defocus tolerance 1 Glistening-free performance 1,2 Predictable outcomes 1 The clear choice for consistent visual excellence. For over 165 years Bausch

More information

Long-term quality of vision is what every patient expects

Long-term quality of vision is what every patient expects Long-term quality of vision is what every patient expects Innovative combination of HOYA technologies provides: 1-piece aspheric lens with Vivinex hydrophobic acrylic material Unique surface treatment

More information

Crystalens AO: Accommodating, Aberration-Free, Aspheric Y. Ralph Chu, MD Chu Vision Institute Bloomington, MN

Crystalens AO: Accommodating, Aberration-Free, Aspheric Y. Ralph Chu, MD Chu Vision Institute Bloomington, MN Crystalens AO: Accommodating, Aberration-Free, Aspheric Y. Ralph Chu, MD Chu Vision Institute Bloomington, MN Financial Disclosure Advanced Medical Optics Allergan Bausch & Lomb PowerVision Revision Optics

More information

Design of a Test Bench for Intraocular Lens Optical Characterization

Design of a Test Bench for Intraocular Lens Optical Characterization Journal of Physics: Conference Series Design of a Test Bench for Intraocular Lens Optical Characterization To cite this article: Francisco Alba-Bueno et al 20 J. Phys.: Conf. Ser. 274 0205 View the article

More information

Mechanism of compensation of aberrations in the human eye

Mechanism of compensation of aberrations in the human eye 3274 J. Opt. Soc. Am. A/ Vol. 24, No. 10/ October 2007 Tabernero et al. Mechanism of compensation of aberrations in the human eye Juan Tabernero,* Antonio Benito, Encarna Alcón, and Pablo Artal Laboratorio

More information

10/25/2017. Financial Disclosures. Do your patients complain of? Are you frustrated by remake after remake? What is wavefront error (WFE)?

10/25/2017. Financial Disclosures. Do your patients complain of? Are you frustrated by remake after remake? What is wavefront error (WFE)? Wavefront-Guided Optics in Clinic: Financial Disclosures The New Frontier November 4, 2017 Matthew J. Kauffman, OD, FAAO, FSLS STAPLE Program Soft Toric and Presbyopic Lens Education Gas Permeable Lens

More information

Evolution of Diffractive Multifocal Intraocular Lenses

Evolution of Diffractive Multifocal Intraocular Lenses Evolution of Diffractive Multifocal Intraocular Lenses Wavefront Congress February 24, 2007 Michael J. Simpson, Ph.D. Alcon Research, Ltd., Fort Worth, Texas Presentation Overview Multifocal IOLs two lens

More information

THE ASPHERIC PRELOADED INJECTION

THE ASPHERIC PRELOADED INJECTION New from STAAR for Safe and Easy Implantation THE ASPHERIC PRELOADED INJECTION S Y S T E M PRELOADED INJECTION SYSTEM FOR SIMPLE AND EASY IOL DELIVERY SAFE The KS3-Ai features an aspheric IOL preloaded

More information

Comparison of Contrast Sensitivity, Higher- Order Aberrations and Subjective Visual Function after Different Aspheric Intraocular Lenses Implantation

Comparison of Contrast Sensitivity, Higher- Order Aberrations and Subjective Visual Function after Different Aspheric Intraocular Lenses Implantation Thai J Ophthalmol Vol. 25 No. 2 July-December 20 9 Original Article/π æπ åμâπ Comparison of Contrast Sensitivity, Higher- Order Aberrations and Subjective Visual Function after Different Aspheric Intraocular

More information

Wavefront Aberrations in Eyes With Acrysof Monofocal Intraocular Lenses

Wavefront Aberrations in Eyes With Acrysof Monofocal Intraocular Lenses Wavefront Aberrations in Eyes With Acrysof Monofocal Intraocular Lenses Prema Padmanabhan, MS; Geunyoung Yoon, PhD; Jason Porter, PhD; Srinivas K. Rao, FRCSEd; Roy J, MSc; Mitalee Choudhury, BS ABSTRACT

More information

Raise your expectations. Deliver theirs.

Raise your expectations. Deliver theirs. 66 EXTENDED RANGE OF VISION MONOFOCAL-LIKE DISTANCE Raise your expectations. Deliver theirs. Now you can give your patients the best of both worlds with the first and only hybrid designed monofocal-multifocal

More information

Evaluation of the Impact of Intraocular Lens Tecnis Z9000 Misalignment on the Visual Quality Using the Optical Eye Modeling

Evaluation of the Impact of Intraocular Lens Tecnis Z9000 Misalignment on the Visual Quality Using the Optical Eye Modeling Evaluation of the Impact of Intraocular Lens Tecnis Z9000 Misalignment on the Visual Quality Using the Optical Eye Modeling Azam Asgari 1 Ali Asghar Parach 1 Keykhosro Keshavarzi 2 Abstract Purpose: The

More information

Contrast Sensitivity after Refractive Lens Exchange with A Multifocal Diffractive Aspheric Intraocular Lens

Contrast Sensitivity after Refractive Lens Exchange with A Multifocal Diffractive Aspheric Intraocular Lens Contrast Sensitivity after Refractive Lens Exchange with A Multifocal Diffractive Aspheric Intraocular Lens Teresa Ferrer-Blasco, PhD Santiago García-Lázaro, PhD César Albarrán-Diego, MSc 3 Lurdes Belda-Salmerón,

More information

Visual Simulation: application to monofocal intraocular lens analysis

Visual Simulation: application to monofocal intraocular lens analysis ARTICLE Visual Simulation: application to monofocal intraocular lens analysis Alberto Domínguez Vicent, OD; Cari Pérez-Vives, MSc; Lurdes Belda-Salmerón, MSc; César Albarrán-Diego, MSc; Santiago García-Lázaro,

More information

Comparison of Contrast Sensitivity in Eyes with Toric and Aspheric Intraocular Lenses

Comparison of Contrast Sensitivity in Eyes with Toric and Aspheric Intraocular Lenses ARTICLE Comparison of Contrast Sensitivity in Eyes with Toric and Aspheric Intraocular Lenses Francisco Pastor-Pascual, PhD 1 ; Aitor Lanzagorta-Aresti, MD 1 ; Pablo Alcocer-Yuste, MD 1 ; Cristina Puchades-Lladró

More information

Prospective sual evaluation of apodized diffractive intraocular lenses

Prospective sual evaluation of apodized diffractive intraocular lenses See discussions, stats, and author profiles for this publication at: https://www.researchgate.net/publication/6251759 Prospective sual evaluation of apodized diffractive intraocular lenses ARTICLE in JOURNAL

More information

HOYA aspherical IOL with ABC (Aspheric Balanced Curve) Design

HOYA aspherical IOL with ABC (Aspheric Balanced Curve) Design HOYA aspherical IOL with ABC (Aspheric Balanced Curve) Design Contents Basics of asphericity Visual quality and aspheric IOL Features of HOYA ABC Design 2 What is asphericity? Deviating from the spherical

More information

Clinical Update for Presbyopic Lens Options

Clinical Update for Presbyopic Lens Options Clinical Update for Presbyopic Lens Options Gregory D. Searcy, M.D. Erdey Searcy Eye Group Columbus, Ohio The Problem = Spherical Optics Marginal Rays Spherical IOL Light Rays Paraxial Rays Spherical Aberration

More information

IOL Types. Hazem Elbedewy. M.D., FRCS (Glasg.) Lecturer of Ophthalmology Tanta university

IOL Types. Hazem Elbedewy. M.D., FRCS (Glasg.) Lecturer of Ophthalmology Tanta university IOL Types Hazem Elbedewy M.D., FRCS (Glasg.) Lecturer of Ophthalmology Tanta university Artificial intraocular lenses are used to replace the eye natural lens when it has been removed during cataract surgery.

More information

Clinical Study Effect of Spherical Aberration on the Optical Quality after Implantation of Two Different Aspherical Intraocular Lenses

Clinical Study Effect of Spherical Aberration on the Optical Quality after Implantation of Two Different Aspherical Intraocular Lenses Hindawi Ophthalmology Volume 2017, Article ID 8039719, 6 pages https://doi.org/10.1155/2017/8039719 Clinical Study Effect of Spherical Aberration on the Optical Quality after Implantation of Two Different

More information

Correlation of pupil size with visual acuity and contrast sensitivity after implantation of an apodized diffractive intraocular lens

Correlation of pupil size with visual acuity and contrast sensitivity after implantation of an apodized diffractive intraocular lens ARTICLE Correlation of pupil size with visual acuity and contrast sensitivity after implantation of an apodized diffractive intraocular lens José F. Alfonso, MD, PhD, Luis Fernández-Vega, MD, PhD, M. Begoña

More information

COMPARISON OF THE MEDICONTUR 860FAB

COMPARISON OF THE MEDICONTUR 860FAB COMPARISON OF THE MEDICONTUR 860FAB HYDROPHOBIC IOL AND THE ACRYSOF IQ LONG TERM FOLLOW UP Péter Vámosi MD, Amanda Argay MD, Zsófia Rupnik MD, János Fekete Péterfy Sándor Hospital Budapest, Hungary PREFERENCE

More information

Customized Correction of Wavefront Aberrations in Abnormal Human Eyes by Using a Phase Plate and a Customized Contact Lens

Customized Correction of Wavefront Aberrations in Abnormal Human Eyes by Using a Phase Plate and a Customized Contact Lens Journal of the Korean Physical Society, Vol. 49, No. 1, July 2006, pp. 121 125 Customized Correction of Wavefront Aberrations in Abnormal Human Eyes by Using a Phase Plate and a Customized Contact Lens

More information

Quality of Vision With Multifocal Progressive Diffractive Lens: Two-Year Follow-up

Quality of Vision With Multifocal Progressive Diffractive Lens: Two-Year Follow-up Quality of Vision With Multifocal Progressive Diffractive Lens: Two-Year Follow-up Antonio Mocellin, MD & Matteo Piovella, MD CMA, Centro di Microchirurgia Ambulatoriale Monza (Milan) Italy Dr Piovella

More information

SEE BEYOND WITH FULLRANGE OPTICS. Developed by Hanita Lenses

SEE BEYOND WITH FULLRANGE OPTICS. Developed by Hanita Lenses SEE BEYOND WITH FULLRANGE OPTICS Developed by Hanita Lenses SEE beyond with FullRange optics FullRange optic lenses are proven, highlyreliable and safe intraocular lenses designed to provide a solution

More information

CONSISTENT ADVANCES IN CATARACT SURGICAL

CONSISTENT ADVANCES IN CATARACT SURGICAL Evaluation of the Aspheric Tecnis Multifocal Intraocular Lens: One-Year Results from the First Cohort of the Food and Drug Administration Clinical Trial MARK PACKER, Y. RALPH CHU, KEVIN L. WALTZ, ERIC

More information

CLINICAL SCIENCES. Corneal Optical Aberrations and Retinal Image Quality in Patients in Whom Monofocal Intraocular Lenses Were Implanted

CLINICAL SCIENCES. Corneal Optical Aberrations and Retinal Image Quality in Patients in Whom Monofocal Intraocular Lenses Were Implanted CLINICAL SCIENCES Corneal Optical Aberrations and Retinal Image Quality in Patients in Whom Monofocal Intraocular Lenses Antonio Guirao, PhD; Manuel Redondo, PhD; Edward Geraghty; Patricia Piers; Sverker

More information

Choices and Vision. Jeffrey Koziol M.D. Thursday, December 6, 12

Choices and Vision. Jeffrey Koziol M.D. Thursday, December 6, 12 Choices and Vision Jeffrey Koziol M.D. How does the eye work? What is myopia? What is hyperopia? What is astigmatism? What is presbyopia? How the eye works How the Eye Works 3 How the eye works Light rays

More information

Maximum Light Transmission. Pupil-independent Light Distribution. 3.75D Near Addition Improved Intermediate Vision

Maximum Light Transmission. Pupil-independent Light Distribution. 3.75D Near Addition Improved Intermediate Vision Multifocal Maximum Light Transmission Pupil-independent Light Distribution Better Visual Quality Increased Contrast Sensitivity 3.75D Near Addition Improved Intermediate Vision Visual Performance After

More information

Impact of scattering and spherical aberration in contrast sensitivity

Impact of scattering and spherical aberration in contrast sensitivity Journal of Vision (2009) 9(3):19, 1 10 http://journalofvision.org/9/3/19/ 1 Impact of scattering and spherical aberration in contrast sensitivity Guillermo M. Pérez Silvestre Manzanera Pablo Artal Laboratorio

More information

Normal Wavefront Error as a Function of Age and Pupil Size

Normal Wavefront Error as a Function of Age and Pupil Size RAA Normal Wavefront Error as a Function of Age and Pupil Size Raymond A. Applegate, OD, PhD Borish Chair of Optometry Director of the Visual Optics Institute College of Optometry University of Houston

More information

Roadmap to presbyopic success

Roadmap to presbyopic success Roadmap to presbyopic success Miltos O Balidis MD, PhD, FEBOphth, ICOphth Early experience with Presbyopic correction 2003 Binocular Distance-Corrected Intermediate and Near Vision Binocular Distance-Corrected

More information

CLINICAL SCIENCE INTRODUCTION

CLINICAL SCIENCE INTRODUCTION CLINICS 2009;64(10):953-60 CLINICAL SCIENCE A CONTRALATERAL EYE STUDY COMPARING APODIZED DIFFRATIVE AND FULL DIFFRATIVE LENSES: WAVEFRONT ANALYSIS AND DISTANCE AND NEAR UNCORRECTED VISUAL ACUITY Marcony

More information

TRANSLATIONAL SCIENCE. Effect of Crystalline Lens Aberrations on Adaptive Optics Simulation of Intraocular Lenses

TRANSLATIONAL SCIENCE. Effect of Crystalline Lens Aberrations on Adaptive Optics Simulation of Intraocular Lenses TRANSLATIONAL SCIENCE Effect of Crystalline Lens Aberrations on Adaptive Optics Simulation of Intraocular Lenses Eloy A. Villegas, PhD; Silvestre Manzanera, PhD; Carmen M. Lago, MSc; Lucía Hervella, MSc;

More information

Theoretical Considerations

Theoretical Considerations Bedeutung der Linsenposition für die optische Abbildugsqualität- Ein neues Messverfahren mittels Purkinje Reflexbildern SA(-) Acri.Tec 35A Hoya FY60AD Tecnis ZCB00 AcrySof IQ U. Mester, T. Sauer, H.Kaymak

More information

Customized intraocular lenses

Customized intraocular lenses Customized intraocular lenses Challenges and limitations Achim Langenbucher, Simon Schröder & Timo Eppig Customized IOL what does this mean? Aspherical IOL Diffractive multifocal IOL Spherical IOL Customized

More information

Accommodating IOL s History and Clinical Management

Accommodating IOL s History and Clinical Management Accommodating IOL s History and Clinical Management Bausch & Lomb Surgical Aliso Viejo, CA Genesis of an Accommodating IOL 1 Observations with Plate IOLs Stuart Cumming 1989 Some plate haptic IOL patients

More information

Comparative Study Between the High Order Aberrations Before and After Cataract Surgery Using Two Different Types of IOLS

Comparative Study Between the High Order Aberrations Before and After Cataract Surgery Using Two Different Types of IOLS Cronicon OPEN ACCESS EC OPHTHALMOLOGY Research Article Comparative Study Between the High Order Aberrations Before and After Cataract Surgery Using Two Different Types of IOLS Ahmed Abdelkareem Elmassry*,

More information

Clinical Evaluation 3-month Follow-up Report

Clinical Evaluation 3-month Follow-up Report Clinical Evaluation 3-month Follow-up Report Of SeeLens AF Intraocular Lens 30 June 2010 version 1.1 1 of 21 Table of Contents: Objectives 2 Medical device specification and administration 4 Methods 9

More information

Causes of refractive error post premium IOL s 3/17/2015. Instruction course: Refining the Refractive Error After Premium IOL s.

Causes of refractive error post premium IOL s 3/17/2015. Instruction course: Refining the Refractive Error After Premium IOL s. Instruction course: Refining the Refractive Error After Premium IOL s. Senior Instructor: Mounir Khalifa, MD Instructors: David Hardten,MD Scott MacRea,MD Matteo Piovella,MD Dr. Khalifa: Causes of refractive

More information

Theoretical Comparison of Aberrationcorrecting Customized and Aspheric Intraocular Lenses

Theoretical Comparison of Aberrationcorrecting Customized and Aspheric Intraocular Lenses Theoretical Comparison of Aberrationcorrecting Customized and Aspheric Intraocular Lenses Patricia A. Piers; Henk A. Weeber; Pablo Artal, PhD; Sverker Norrby, PhD ABSTRACT PURPOSE: To assess the performance

More information

Improving Lifestyle Vision. with Small Aperture Optics

Improving Lifestyle Vision. with Small Aperture Optics Improving Lifestyle Vision with Small Aperture Optics The Small Aperture Premium Lens Solution The IC-8 small aperture intraocular lens (IOL) is a revolutionary lens that extends depth of focus by combining

More information

Although, during the last decade, peripheral optics research

Although, during the last decade, peripheral optics research Visual Psychophysics and Physiological Optics Comparison of the Optical Image Quality in the Periphery of Phakic and Pseudophakic Eyes Bart Jaeken, 1 Sandra Mirabet, 2 José María Marín, 2 and Pablo Artal

More information

Cataract surgery using phacoemulsification and implantation

Cataract surgery using phacoemulsification and implantation Visual quality of spherical and aspherical IOLs Clinical Research Evaluation of visual quality of spherical and aspherical intraocular lenses by Optical Quality Analysis System Yan Chen 1, Xue Wang 1,

More information

Comparison of contrast sensitivity and color discrimination after clear and yellow intraocular lens implantation

Comparison of contrast sensitivity and color discrimination after clear and yellow intraocular lens implantation J CATARACT REFRACT SURG - VOL 31, SEPTEMBER 2005 Comparison of contrast sensitivity and color discrimination after clear and yellow intraocular lens implantation Antonio Rodríguez-Galietero, MD, PhD, FEBO,

More information

Choices and Vision. Jeffrey Koziol M.D. Friday, December 7, 12

Choices and Vision. Jeffrey Koziol M.D. Friday, December 7, 12 Choices and Vision Jeffrey Koziol M.D. How does the eye work? What is myopia? What is hyperopia? What is astigmatism? What is presbyopia? How the eye works Light rays enter the eye through the clear cornea,

More information

AT LISA tri 839MP and AT LISA tri toric 939MP from ZEISS The innovative trifocal IOL concept providing True Living Vision to more patients

AT LISA tri 839MP and AT LISA tri toric 939MP from ZEISS The innovative trifocal IOL concept providing True Living Vision to more patients Premium Trifocal MICS OVDs IOLs AT LISA tri 839MP and AT LISA tri toric 939MP from ZEISS The innovative trifocal IOL concept providing True Living Vision to more patients The moment you help your patients

More information

Comparison of retinal image quality with spherical and customized aspheric intraocular lenses

Comparison of retinal image quality with spherical and customized aspheric intraocular lenses Comparison of retinal image quality with spherical and customized aspheric intraocular lenses Huanqing Guo,* Alexander V. Goncharov, and Chris Dainty Applied Optics Group, School of Physics, National University

More information

ROTATIONAL STABILITY MAKES THE DIFFERENCE

ROTATIONAL STABILITY MAKES THE DIFFERENCE The Bi-Flex platform the proven platform of Excellence 01 Proven Stability less than 02 Optimal biomaterials 2 degrees long term rotation 03 Posterior Toric Lens surface with marks indicating the flat

More information

Research Article In Vitro Aberrometric Assessment of a Multifocal Intraocular Lens and Two Extended Depth of Focus IOLs

Research Article In Vitro Aberrometric Assessment of a Multifocal Intraocular Lens and Two Extended Depth of Focus IOLs Hindawi Ophthalmology Volume 2017, Article ID 7095734, 7 pages https://doi.org/10.1155/2017/7095734 Research Article In Vitro Aberrometric Assessment of a Multifocal Intraocular Lens and Two Extended Depth

More information

Refractive Power / Corneal Analyzer. OPD-Scan III

Refractive Power / Corneal Analyzer. OPD-Scan III Refractive Power / Corneal Analyzer OPD-Scan III Comprehensive Vision Analysis and NIDEK, a global leader in ophthalmic and optometric equipment, has created the OPD-Scan III, the third generation aberrometer

More information

Effects of intraocular lenses with different diopters on chromatic aberrations in human eye models

Effects of intraocular lenses with different diopters on chromatic aberrations in human eye models Song et al. BMC Ophthalmology (2016) 16:9 DOI 10.1186/s12886-016-0184-6 RESEARCH ARTICLE Open Access Effects of intraocular lenses with different diopters on chromatic aberrations in human eye models Hui

More information

Diffractive Optics. Multifocal Lenses. Correction of Pseudophakic Presbyopia with Multifocal IOLs. Basic Designs

Diffractive Optics. Multifocal Lenses. Correction of Pseudophakic Presbyopia with Multifocal IOLs. Basic Designs Correction of Pseudophakic Presbyopia with Multifocal IOLs GEORGE H.H. BEIKO, B.M.,B.Ch.,FRCSC ST. CATHARINES, CANADA ASSIST PROF, MCMASTER UNIV george.beiko@sympatico.ca AMO Tecnis MFIOL Alcon ReSTOR

More information

NEW. AT LISA tri 839MP and AT LISA tri toric 939MP from ZEISS The innovative trifocal IOL concept providing True Living Vision to more patients

NEW. AT LISA tri 839MP and AT LISA tri toric 939MP from ZEISS The innovative trifocal IOL concept providing True Living Vision to more patients Premium Trifocal MICS OVDs IOLs NEW AT LISA tri 839MP and AT LISA tri toric 939MP from ZEISS The innovative trifocal IOL concept providing True Living Vision to more patients Trifocal toric IOL The moment

More information

Multifocal Progressive Diffractive Lens to Improve Light Distribuition and Avoid Light Loss: Two Years Clinical Results

Multifocal Progressive Diffractive Lens to Improve Light Distribuition and Avoid Light Loss: Two Years Clinical Results Multifocal Progressive Diffractive Lens to Improve Light Distribuition and Avoid Light Loss: Two Years Clinical Results Matteo Piovella MD & Barbara Kusa MD CMA, Centro di Microchirurgia Ambulatoriale

More information

The Blueprint for Improved Image Quality

The Blueprint for Improved Image Quality Insert to January 2009 The Blueprint for Improved Image Quality The function and design of the AcrySof IQ IOL. Supported by an educational grant from Alcon Laboratories, Inc. The Function and Design of

More information

Improved functional vision with a modified prolate intraocular lens

Improved functional vision with a modified prolate intraocular lens Improved functional vision with a modified prolate intraocular lens Mark Packer, MD, I. Howard Fine, MD, Richard S. Hoffman, MD, Patricia A. Piers, MSc Purpose: To evaluate whether the Tecnis Z9000 intraocular

More information

ASPIRA PUBLICATION LIST

ASPIRA PUBLICATION LIST ASPIRA PUBLICATION LIST ASPIRA-aA / MC 6125 AS PR2 PR3 Eppig T, Scholz K, Löffler A, et al. Effect of decentration and tilt on the image quality of aspheric intraocular lens designs in a model eye J Cataract

More information

The Impact of New Generation Aspherical Soft Contact Lenses on Quality of Vision: A Comparison with Spherical Contact Lenses and Spectacle Correction

The Impact of New Generation Aspherical Soft Contact Lenses on Quality of Vision: A Comparison with Spherical Contact Lenses and Spectacle Correction Deniz Oral, Maryo C. Kohen, Melda Yenerel, Ebru Gorgun, Sule Ziylan, Ferda Ciftci Yeditepe University Faculty of Medicine, Department of Ophthalmology, Istanbul Introduction The correction of higher order

More information

Centre Hospitalier Universitaire et Psychiatrique de Mons-Borinage. B-Flex Multifocal. Dr Emmanuel Van Acker Belgium

Centre Hospitalier Universitaire et Psychiatrique de Mons-Borinage. B-Flex Multifocal. Dr Emmanuel Van Acker Belgium Centre Hospitalier Universitaire et Psychiatrique de Mons-Borinage B-Flex Multifocal Dr Emmanuel Van Acker Belgium Comparison of clinical outcomes and patient satisfaction after implantation of two different

More information

Sulcoflex. For when perfection is the only option! Pseudophakic Sulcus Fixated Secondary IOLs. Sulcoflex Aspheric. Sulcoflex Toric

Sulcoflex. For when perfection is the only option! Pseudophakic Sulcus Fixated Secondary IOLs. Sulcoflex Aspheric. Sulcoflex Toric Sulcoflex Pseudophakic Sulcus Fixated Secondary IOLs Sulcoflex Aspheric Sulcoflex Toric Sulcoflex Multifocal For when perfection is the only option! Sulcoflex Pseudophakic Sulcus Fixated Secondary IOLs

More information

In this issue of the Journal, Oliver and colleagues

In this issue of the Journal, Oliver and colleagues Special Article Refractive Surgery, Optical Aberrations, and Visual Performance Raymond A. Applegate, OD, PhD; Howard C. Howland,PhD In this issue of the Journal, Oliver and colleagues report that photorefractive

More information

Optical Characteristics of Next Generation Dual Optic IOL

Optical Characteristics of Next Generation Dual Optic IOL Optical Characteristics of Next Generation Dual Optic IOL Scott Evans, MD Sanjeev Kasthurirangan, PhD Val Portney, PhD Financial Disclosures Scott Evans is an employee of Abbott Medical Optics Inc. Sanjeev

More information

final corrected draft

final corrected draft Incisions, IOLs and aberrations 1. Archived at the Flinders Academic Commons http://dspace.flinders.edu.au/dspace/ This is the author s final corrected draft of this article. It has undergone peer review.

More information

Página 1 de 9 TopPage > Eye Care > Diagnostic > Wave-Front Analyzer KR-1W Wave-Front Analyzer KR-1W Perfection for Professionals : KR-1W Topcon, with its wealth of experience in designing and manufacturing

More information

The eye-care profession s advancement from cataract

The eye-care profession s advancement from cataract Cataract surgery and optimal spherical aberration: as simple as you think? Shamrozé Khan,* BSc, OD; Guillermo Rocha, MD, FRCSC ABSTRACT RÉSUMÉ This paper reviews the optics of higher-order and spherical

More information

Visual function after bilateral implantation of apodized diffractive aspheric multifocal intraocular lenses with a D3.0 D addition

Visual function after bilateral implantation of apodized diffractive aspheric multifocal intraocular lenses with a D3.0 D addition ARTICLE Visual function after bilateral implantation of apodized diffractive aspheric multifocal intraocular lenses with a D3.0 D addition Thomas Kohnen, MD, Rudy Nuijts, MD, Pierre Levy, MD, Eduard Haefliger,

More information

Optical aberrations of intraocular lenses measured in vivo and in vitro

Optical aberrations of intraocular lenses measured in vivo and in vitro Barbero et al. Vol. 20, No. 10/October 2003/J. Opt. Soc. Am. A 1841 Optical aberrations of intraocular lenses measured in vivo and in vitro Sergio Barbero and Susana Marcos Instituto de Óptica, Consejo

More information

day night convinced supreme contrast sensitivity THE IOL FOR DAY & NIGHT

day night convinced supreme contrast sensitivity THE IOL FOR DAY & NIGHT day supreme contrast sensitivity night convinced THE IOL FOR DAY & NIGHT The IOL for DAY & NIGhT UnIQUE form and features Innovative blue light filtering Excellent quality of vision Maximum depth of focus

More information

Energy balance in apodized diffractive multifocal intraocular lenses

Energy balance in apodized diffractive multifocal intraocular lenses Energy balance in apodized diffractive multifocal intraocular lenses Francisco Alba-Bueno *a, Fidel Vega a, María S. Millán a a Optics and Optometry Department, Polytechnic University of Catalonia. ABSTRACT

More information

Principles and clinical applications of ray-tracing aberrometry (Part II)

Principles and clinical applications of ray-tracing aberrometry (Part II) UPDATE/REVIEW Principles and clinical applications of ray-tracing aberrometry (Part II) Alfredo Castillo Gómez, MD, PhD 1 ; Antonio Verdejo del Rey, OD 2 ; Carlos Palomino Bautista, MD 3 ; Ana Escalada

More information

Visual Performance after AcrySof ReSTOR Aspheric Intraocular Lens Implantation

Visual Performance after AcrySof ReSTOR Aspheric Intraocular Lens Implantation ORIGINAL ARTICLE J Optom 2008;1:30-35 Visual Performance after AcrySof ReSTOR Aspheric Intraocular Lens Implantation José F. Alfonso 1,2, Luis Fernández-Vega 1,2, Beatriz Valcárcel 1 and Robert Montés-Micó

More information

Assessing Visual Quality With the Point Spread Function Using the NIDEK OPD-Scan II

Assessing Visual Quality With the Point Spread Function Using the NIDEK OPD-Scan II Assessing Visual Quality With the Point Spread Function Using the NIDEK OPD-Scan II Edoardo A. Ligabue, MD; Cristina Giordano, OD ABSTRACT PURPOSE: To present the use of the point spread function (PSF)

More information

Treatment of Presbyopia during Crystalline Lens Surgery A Review

Treatment of Presbyopia during Crystalline Lens Surgery A Review Treatment of Presbyopia during Crystalline Lens Surgery A Review Pierre Bouchut Bordeaux Ophthalmic surgeons should treat presbyopia during crystalline lens surgery. Thanks to the quality and advancements

More information