Visual Simulation: application to monofocal intraocular lens analysis

Size: px
Start display at page:

Download "Visual Simulation: application to monofocal intraocular lens analysis"

Transcription

1 ARTICLE Visual Simulation: application to monofocal intraocular lens analysis Alberto Domínguez Vicent, OD; Cari Pérez-Vives, MSc; Lurdes Belda-Salmerón, MSc; César Albarrán-Diego, MSc; Santiago García-Lázaro, PhD PURPOSE: Compare two monofocal intraocular lenses (IOLs) with different designs. METHOD: An adaptive optics visual simulator (crx1, Imagine Eyes, France) was used to simulate the aberration pattern of two IOLs, a spherical monofocal IOL (AcrySof SN60AT IOL, Alcon Laboratories, USA) and an aspheric monofocal IOL (AcrySof IQ SN60WF IOL Alcon Laboratories, USA). The Visual Acuity (VA) was measured at high- mediumand low- contrast when each IOL was centered, decentered and tilted different degrees. Depth-of-focus of both lenses was evaluated when each IOL was centered. The Modulation Transfer Function (MTF) and the Point Spread Function (PSF) were computed for all conditions. All measures were taken for 5 mm pupil diameter. RESULTS: There were no statistical significant differences in VA between both IOLs for centered and decentered 0.2 mm positions (P > 0.05). For these positions, both lenses showed good VA (about 20/20) at high contrast. But at 50% and 10% of contrast VA values were worse (up to 20/40). At 0.4 mm of decentering and both tilts, the SN60AT IOL has better VA values than SN60WF IOL showing statistically significant differences (P<0.05). The values of depth-of-focus were 0.25 and 0.75 dioptres (D) for aspheric and spherical IOLs, respectively. MTFs and PSFs decrease when both IOLs were decentered and tilted. The spherical IOL shows worse PSFs than aspheric IOL in centered position. CONCLUSION: It seems that spherical design is more robust and provides more depth of focus than the aspheric model. Although, aspheric model induces better optical quality in centered position. J Emmetropia 2011; 2: INTRODUCTION It is necessary to introduce some optical concepts before discuss about the different applications of adaptive optics (AO) in human vision. Wave aberration 1 is defined as the difference between the perfect (spherical) and the actual wavefront (WF) for each point over the pupil. Figure 1 shows this concept schematically. AO may be defined as an optical system which adapts to compensate aberrations which are induced by the medium between the object and the image. Submitted: 7/27/2011 Revised: 8/27/2011 Accepted: 8/29/2011 From the Optometry Research Group (GIO), Department of Optics. University of Valencia. Spain. Acknowledgements and Disclosure: This research was supported in part by Ministerio de Ciencia e Innovación Research Grant #SAF E#. The authors have no proprietary interest in any of the products mentioned in this article. Address: Alberto Domínguez Vicent. Department of Optics University of Valencia. C/ Dr Moliner, Burjassot. Spain. aldovi@alumni.uv.es Figure 1. Scheme of wavefront definition, showing the reference spherical wavefront (black dot line), the aberrated wavefront (red line) and the wave aberration SECOIR Sociedad Española de Cirugía Ocular Implanto-Refractiva ISSN:

2 144 VISUAL SIMULATION IN MONOFOCAL IOLs ANALYSIS AO is applied to the human eye because it has optical aberrations that reduce the retinal image quality. The main objective to applied AO principle to the human eye is to improve retinal image quality and retinal image visualization (retinal imaging), being the first to achieve better visual acuity (VA) and the second one for observational purposes. Improving retinal image quality To correct ocular aberrations considering AO has been recently applied to 2 contact lens (CL) and intraocular lenses (IOLs) designs or corneal refractive surgery ablation algorithms. The main idea that these solutions use to correct the whole eye aberrations is to make a CL, IOL or an anterior surface of the cornea which induces an anti-aberration of the whole eye. Consequently a perfect image on the retina, which is only affected by diffraction, may be obtained. Figure 2 shows a scheme of this concept, applied specifically to a CL 3. The others devices follow the same correction principle explained above. On the other hand, these devices must be static, centered and customized for each patient to obtain the best retinal image quality. If they are not centered, static or customized, they will induce new aberrations and the retinal image quality will deteriorates. This is the main disadvantage for ophthalmic devices. Soft CLs may be considered as the first choice for optical aberration correction because they have less movement than rigid gas-permeable CLs (RGPCLs). Scleral RGPCLs are more stable than the RGPCLs and may be also considered for this purpose. Then, soft CLs and Scleral RGPCLs must be considered as the first choice for optical aberration correction. The advantage of customized CL, in comparison with refractive surgery, is its reversibility because the CL can be removed if the visual performance is not good. One disadvantage is that CL can rotate, translate and flex with each blink, and the retinal image quality cannot be good. Figure 2. Principle of aberration correction using a contact lens. The ocular aberrations, as well as the aberration pattern induced by the contact lens are shown above. In relation to refractive surgery 1, the main objective is to customize the anterior surface of the cornea to compensate ocular aberrations with laser Excimer ablation. To obtain this, surgeons use WF-guided ablation. This technique consists on obtaining the ocular aberrations and to ablate the anterior surface of the cornea using the WF data. The advantage of refractive surgery is that customized treatment induces less high order aberrations (HOAs) than conventional treatments 4,5. The main disadvantage of this optical aberration correction is that the effect cannot be removed if patient s visual performance is not good. Additionally other factors tend to limit the success of WF-guided treatments and will need to be overcome to optimize clinical results (also considered for CLs and IOLs). These factors are: Age-related changes in HOAs, pupil size accommodation, ocular surface, retinal limits, neuroplasticity, postoperative issues: wound healing and biomechanical effects. Finally, we know that society is aging and the number of cataract surgeries is increasing. Currently, surgeons have different monofocal IOL 1 designs available in the market to be considered for replacing the crystalline lens, as an aspheric and a spherical IOL. The former design is the spherical IOL, but this one induces a residual spherical aberration (SA) and, theoretically, the retinal image quality deteriorates. The aspheric design tries to compensate the corneal SA to improve the retinal image quality. Theoretically it can be concluded that an aspheric monofocal IOL design induces better VA than spherical monofocal IOL design if exists balance between corneal and IOL aberrations. However, clinical studies did not find VA differences between aspheric and spherical designs Only two studies found 11,12 that an aspheric design has better VA than the spherical one. Discrepancies between theoretical and clinical studies can be: clinical studies do not consider the IOL position into the eye, these studies do not check if exist a balance between corneal and IOL aberrations, and there are different philosophies of IOL manufacturers exist to correct full or part of the corneal SA 10. On the other hand, we should considered that the potential benefits of aspheric IOLs are limited by inaccurate or absent preoperative measurement of the ocular parameters necessary for IOL power calculation, inaccurate manufacturing and inability to locate the IOL in the correct plane. However, in centered position, the optical and visual performance of aspheric IOLs is, even in the worst cases, equal to or better than that with spherical IOLs. But independently of the image quality that the IOL induces, there is an external problem which depends of the surgeon. This is an error placing the IOL in the capsular bag because surgeons can tilt or decenter the IOL. Consequently, surgeons should know what IOL design is most robust when it is placed inside the eye.

3 VISUAL SIMULATION IN MONOFOCAL IOLs ANALYSIS 145 Considering the differences reported above on this topic, we have done a pilot study to analyse two monofocal IOL design. One with an aspheric design, which partially compensates corneal SA, and the other one with a spherical design which does not compensate corneal SA. We used the visual simulator to induce the aberration pattern of each IOL + cornea, previously obtained, to measure the VA at different contrast under different conditions, when the lenses were centered, decentered and tilted. Depth-of-focus for the centered condition of both lenses was also measured. PATIENTS AND METHODS Subjects Seven individuals, aged between 21 and 30 years and experienced in psychophysical experiments were included in this study. Spherical refractive errors ranged between 2.00 and dioptres (D) with astigmatism < 0.50 D. They had clear intraocular media and no known ocular pathology. The tenets of the Declaration of Helsinki were followed. Informed consent was obtained from each participant after verbal and written explanation of the nature and possible consequences of the study. Adaptive Optics Visual Simulator A visual simulator 1,13-16 is a device which uses the principle of AO. This device can measure, simulate, correct and manipulate any aberration pattern. It has three basic components: WF sensor, control computer and WF corrector-simulator. On the following section we describe briefly these elements relating them with the components of the crx1 AO visual simulator (Imagine Eyes, Orsay, France) used during the experimental procedure. Figure 3 shows a scheme of this equipment. Wavefront sensor The WF sensor measures the optical aberration at the pupil plane of the eye. A recent review of Cerviño et al. 17 discussed about the advantages and disadvantages of different clinical WF sensors available in the market. Clinical aberrometers may be based on the Tscherning s aberroscope, the cross-cylinder technique, the spatially resolved refractometer, the laser ray tracing technique and the Shack-Hartman (S-H) sensor. We are going to focus on the S-H WF sensor because it is the WF sensor used in the visual simulator. The basic components of S-H WF sensor 1,18 are the lens let array, a monochromatic light source and a charge-coupled device (CCD) detector 1. The lens let array consists of two-dimensional array of 1024 lenses, Figure 3. Visual simulator scheme of the crx1 system. all with the same diameter and the same focal length. The typical diameter range of these lenses is about 100 to 600 µm. The typical focal length range is from 30 nm to a few millimetres. The light source that is used to measure aberrations uses a wavelength of 850 nm. The CCD detector is a device which is placed in the focal plane of the lens let array. Its function is to record the spot array pattern to calculate the WF. This WF sensor has some advantages and disadvantages 1. The advantages consider that the measurement is objective, automatic and quick; it can operate in real time and it has high spatial resolution. Disadvantages of this WF are that it has small dynamic range (it is limited by spacing and the focal length of the lens let) and that the spot pattern can be blurred. This defocus can cause an error in estimating the centroid of the spot. The measure procedure of the aberrations for the whole eye consists on sending into the eye a light from a laser. This is focused by the optics of the eye to a point on the retina. This light is reflected on the retina and emerges through the pupil as an aberrated WF. This aberration pattern travels from the eye through a lens let array, which is conjugate with the pupil of the eye. As the lens let array is in a pupil conjugate plane, the WF shape at the lens let plane is identical to the shape at the pupil of the eye. Finally, the lens let array forms an array of spots image on a CCD detector. This explanation is summarized on figure 4. This figure shows a schematic diagram of the AO system during the WF measurement. The red line shows the way that the light uses to do the measure. On the other hand, the same acquisition provides a calculation of the WF refraction which is compatible with optical corrections. The control computer This component 1 converts the raw output of the WF sensor into voltage commands that are sent to the WF corrector. Control systems are broadly classified in

4 146 VISUAL SIMULATION IN MONOFOCAL IOLs ANALYSIS Figure 4. Schematic diagram of the adaptive optics system during wavefront measurement. two group, open-loop systems and closed-loop systems. The closed-loop system is used in this visual simulator where the WF sensor measures the wave aberration after it is corrected by the WF corrector. Figure 5 shows a block diagram of closed-loop AO system for vision science. This control computer is faster and more accurate than open-loop control system. It increases the system accuracy and reduces the sensitivity of the ratio of the output to input due to variations in system characteristics. The disadvantages of closed-loop control are its increased complexity and cost. The wavefront corrector The WF corrector 1 compensates the aberration pattern generating a surface shape that is ideally conjugate to the aberration profile. In the visual simulator, the WF corrector has another application. This is inducing the aberration pattern that we need. This device is placed in the conjugated plane for both the pupil eye and the WF sensor. The WF correctors which are used in AO vision system to correct the ocular aberrations may be macroscopic discrete actuator deformable mirrors, bimorph mirrors and liquid crystal spatial light modulators. The present visual simulator uses the macroscopic discrete actuator deformable mirror 19. It consists of a continuous mirror surface whose profile is controlled Figure 6. Schematic diagram of the adaptive optic system during wavefront static correction/induction. by an underlying mini-magnet array. These mini-magnets electromagnetically push and pull on the surface, transforming it into a desired shape. Specially this WF corrector has 52 independent mini-magnets and can correct aberrations up to 5 th order. Figures 6 and 7 show a brief scheme about the working of this equipment. Figure 6 shows a diagram of the visual simulator during a correction of any aberration pattern and figure 7 shows a diagram of the visual simulator during a simulation of any aberration pattern. Other components of the visual simulator crx1 The crx1 has other components 19 such as Badal system, a micro-display and an artificial control pupil size. The Badal system controls the vergence of the stimulus. This device allows us to compensate the subject s refractive error or study the change of the ocular aberrations during accommodation (see figure 8 for an example). The micro-display has pixels and it works with white and black light. Consequently we only can show images which have these colours. Finally, the artificial control pupil size controls the aperture of the system. This device is optically conju- Figure 5. Closed-loop system for correcting wave aberration. Figure 7. Schematic diagram of the adaptive optic system during visual simulation.

5 VISUAL SIMULATION IN MONOFOCAL IOLs ANALYSIS 147 Table.1. Zernikes values of the intraocular lens centered plus cornea Zernike Coefficients Values (µm) SN60AT * SN60WF +CORNEA +CORNEA Figure 8. Variation of the wavefront pattern during accommodation. Left image shows the wavefront without accommodation and right image shows the wavefront with 5 dioptres of accommodation (5 mm pupil). gated with the eye-pupil plane and its aperture can be artificially adjusted selecting the diameter of an internal aperture. Then we can control the subject s pupil diameter if this is longer than the pupil diameter that we need for any experiment. Figure 9. Maltese cross. z (3, 3) z (3, 1) z (3, 1) z (3, 3) z (4, 4) z (4, 2) z (4, 0) z (4, 2) z (4, 4) * Spherical intraocular lens; Aspheric intraocular lens. Experimental Procedures Approximately 30 min before the experimental measurements, three drops of cyclopentolate hydrochloride 0.5% were instilled to paralyze the subject s accommodation and achieve pupil dilation up to 5 mm. While patient was dilating the mirror of the visual simulator was reshaped to flat shape using the commercially available program (HASO, Imagine Eyes, France). During all measurement procedure, we put 5 mm on the artificial control pupil size to ensure that all measures had the same pupil diameter. Once this process has been carried out, we used the commercially available program (Irx3, Imagine Eyes, France) program to measure subject s ocular aberrations while he was fixing to the Maltese cross (fig. 9). After that, we used the commercially available program (SVAO, Imagine Eyes, France) program to compensate the subject s aberration pattern and induce the aberration pattern which we needed. Before VA measurement we compensated the defocus that aberration pattern had induced using the Badal system. After that we measured the depth-of-focus when the IOL was centered. When these procedures were carried out, we measured the VA for 5 IOL positions (centered, decentered 0.2 mm and 0.4 mm and tilted 2 and 4 ). For each position we measured the VA for three contrast values: 100%, 50 % and 10%. All of these procedures must be done with all lenses positions. In addition, modulation transfer functions (MTFs) and point spread functions (PSFs) for both IOL + cornea were computed for all conditions (centered, decentered 0.2 mm and 0.4 mm and tilted 2 and 4 ). Corneal and Intraocular lenses Zernike s values The IOLs which we used were the monofocal AcrySof IQ SN60WF IOL (Alcon Laboratories, USA) which has an aspheric design; and the monofocal AcrySof SN60AT IOL, (Alcon Laboratories, USA) which has a spherical design. Figure 10 shows a frontal and a cross-section view of both lenses. For visual simulation we considered the Zernike s values of the IOLs 20 and the cornea which were obtained previously. Corneal Zernike s values used were obtained by Wang et al. 21 for a normal population. Table 1 shows the whole eye Zernike s values considering both corneal and IOL aberrations (defocus and astigmatism were cancelled). These values were obtained for a 5 mm pupil diameter. Visual Acuity Measurements To measure VA we have used the Freiburg Acuity Test (FrACT) 22, where we used the Landolt C at different contrasts. Subjects were instructed to recognize the orientation of the aperture of the C (forced-choice test). To do that, the subject had a numeric keypad and he had to press the number which corresponds with the orientation of the aperture.

6 148 VISUAL SIMULATION IN MONOFOCAL IOLs ANALYSIS Statistical analysis Data analysis was performed using SPSS for Windows software (version 14.0, SPSS, Inc.). Normality was checked using the Kolmogorov- Smirnov test, and outcomes were compared using t tests. Differences were considered statistically significant when the P value was less than RESULTS Tables 2 to 6 and figure 11 show the experimental results. Each table includes the VA values for the same IOL position at different contrasts (100% 50% and 10%). The contents of each table consider logmar VA values, its standard deviation (sd), the VA interval and the p-value. Figure 11 shows the depth-of-focus for both lenses. Table 2 includes the VA values obtained when both IOLs were centered. Here we can see that SN60WF IOL and SN60AT IOL provide good VA values at high contrast (about 20/20). However, at 50% and 10% of contrast VA values were worse (up to 20/40). When both lenses were decentered 0.2 mm (table 3), VA values were also good at 100% of contrast. Similarly to the centered situation, at 50% and 10% of contrast the VA values were reduced. At 0.4 mm of decentration (table 4) the VA values for the SN60WF IOL were considerably reduced (about 2 lines of VA) in relation Figure 11. Comparative of the depth-of-focus curve of the intraocular lenses used in this study. Y-label shows the visual acuity values in decimal scale. X-label shows the vergence of the stimulus in dioptres. Error bars are omitted for clarity. to the previous situation for all contrasts. In contrast, the SN60AT IOL showed better outcomes for all contrasts, which were significantly better in relation to the aspheric design (P<0.05). Table 5 shows the VA values when both IOLs were tilted 2. For the SN60AT IOL the VA values were good at 100% and 50% of contrasts. At 10% of contrast the VA value was highly deteriorated (about 20/50). For the SN60WF IOL at all contrasts the VA values were significantly worse than the spherical model (P<0.05). At 4 of tilt (table 6) both lenses provide good VA at 100%, being reduced at 50% and 10% of contrast. The spherical model showed statistically significant better VA in relation to the aspheric model (P<0.05). Table 2. Visual acuity values expressed in logmar scale at three different contrasts when the lenses are centered VA* 100 % VA 50 % VA 10 % SN60WF SN60AT SN60WF SN60AT SN60WF SN60AT Mean±sd 0.050± ± ± ± ± ±0.110 Range [ ] [ ] [ ] [ ] [ ] [ ] P value * Visual acuity; Aspheric intraocular lens; Spherical intraocular lens; Standard deviation. Table 3. Visual acuity values expressed in logmar scale at three different contrasts when the lenses are decentered 0.2 mm VA* 100 % VA 50 % VA 10 % SN60WF SN60AT SN60WF SN60AT SN60WF SN60AT mean±sd 0.030± ± ± ± ± ±0.100 Range [ ] [ ] [ ] [ ] [ ] [ ] P value * Visual acuity; Aspheric intraocular lens; Spherical intraocular lens; Standard deviation.

7 VISUAL SIMULATION IN MONOFOCAL IOLs ANALYSIS 149 Figure 11 shows the depth-of-focus of both IOLs. The SN60AT IOL showed the best VA value at 0.00 D of vergence. For ±0.50 D of vergence we can appreciate that this lens maintains good and comparable VA than the value found at 0.00 D of vergence. At ±1.00 D of vergence, the VA values decrease significantly (about 4 lines of VA). In relation to the aspheric model, the SN60WF IOL showed its best VA also at 0.00 D. When defocus was introduced the lens shows a similar pattern which was found for the spherical model. Differences statistically significant were not found between both IOLs in any of the defocus degrees studied (P>0.05). Figures show the MTFs and the PSFs for both IOLs. A descriptive analysis was done. In relation Figure 12. Variation of SN60AT intraocular lens (IOL) modulation transfer function (MTF) for all IOL positions studied. X label shows the frequencies in cycles per degree. Y label shows the MTF radial. Table 4. Visual acuity values expressed in logmar scale at three different contrasts when the lenses are decentered 0.4 mm VA* 100 % VA 50 % VA 10 % SN60WF SN60AT SN60WF SN60AT SN60WF SN60AT mean±sd 0.120± ± ± ± ± ±0.120 Range [ ] [ ] [ ] [ ] [ ] [ ] P value * Visual acuity; Aspheric intraocular lens; Spherical intraocular lens; Standard deviation; Differences statistically significant (P<0.05). Table 5. Visual acuity values expressed in logmar scale at three different contrasts when the lenses are tilted 2 VA* 100 % VA 50 % VA 10 % SN60WF SN60AT SN60WF SN60AT SN60WF SN60AT mean±sd 0.050± ± ± ± ± ±0.080 Range [ ] [ ] [ ] [ ] [ ] [ ] P value * Visual acuity; Aspheric intraocular lens; Spherical intraocular lens; Standard deviation; Differences statistically significant (P<0.05). Table 6. Visual acuity values expressed in logmar scale at three different contrasts when the lenses are tilted 4 VA* 100 % VA 50 % VA 10 % SN60WF SN60AT SN60WF SN60AT SN60WF SN60AT mean±sd 0.070± ± ± ± ± ±0.080 Range [ ] [ ] [ ] [ ] [ ] [ ] P value * Visual acuity; Aspheric intraocular lens; Spherical intraocular lens; Standard deviation; Differences statistically significant (P<0.05).

8 150 VISUAL SIMULATION IN MONOFOCAL IOLs ANALYSIS Figure 13. Variation of SN60WF intraocular lens (IOL) modulation transfer function (MTF) for all IOL positions studied. X label shows the frequencies in cycles per degree. Y label shows the MTF radial. Figure 14. Variation of the point spread function of the two lenses studied under 5 conditions studied. Left image (ideal) only affected by diffraction. to the spherical model, SN60AT IOL, the MTF is reduced in relation to the diffraction-limited and decreases when decentration and tilt were provoked. A similar pattern was found for the aspheric model, the SN60WF IOL. In relation to spherical design, SN60AT IOL, the PSFs are worse than diffraction-limited for all positions, centered, decentered and tilted. However, for SN60WF IOL, aspheric design, the PSF is similar to diffraction-limited in centered position. Though, for both decenters and tilts the PSFs are worse than centered position. DISCUSSION In relation to VA, we may observe that for centered and decentered 0.2 mm situations, both IOLs provide similar VA values. For other situations of the IOLs (high decenter and both tilts), the SN60AT IOL (spherical design) provides better VA than the SN60WF IOL (aspheric design). Considering these outcomes, it seems that the spherical design is more robust than the aspheric model for decentering and tilting. These values agree with some theoretical and clinical papers which compare both IOLs designs. On the theoretical paper, Dietze and Cox 23, with model-eye simulation, found that spherical IOLs perform more robust than aspheric IOL when it was displaced. This result agrees with the VA values obtained here. It can be observed from tables 2-6 that the VA values for the spherical design are similar at the same contrast and different IOL position. Figure 14 correlates with this too. Here, the PSFs of the spherical IOL are more constant than the PSFs of the aspheric IOLs at different situations. In a clinical paper, Rocha et al. 9,24 found that the aspheric design in a centered position induces lower HOAs than spherical IOL for a 5 mm pupil. Figures 12, 13 and 14 show an example of this outcome. We can appreciate, in centered position, that the PSF of the aspheric design is better than the PSF of the spherical one. Moreover, in terms of ocular aberrations, the aspheric design induces lower HOAs than spherical IOLs for 5 mm of pupil 9,24 because these lenses try to compensate ocular SA. Then, it can be concluded that, in centered position, aspheric IOL will induce better VA than spherical IOL. However, it must be taking into account that after performing theoretical simulations, Wang and Koch 25 concluded that there are two main reasons for customizing the asphericity of the IOL. First, there is a wide range of corneal SA within population 26,27. Second, other corneal HOAs interact with SA to increase or decrease optical performance. In fact, some residual defocus or astigmatism combined with SA may produce different outcomes depending on the balance between lower-order aberrations and HOAs. Although IOL asphericity can be customized according to corneal SA and desired postoperative ocular aberration, the following must be taken into account: the pupil in older patients is relatively small (senile miosis), HOAs are dominated by asymmetrical aberrations and thus cannot be compensated for by symmetrical IOL designs, and some tilting or displacement of the IOL and changes in corneal optical aberrations can occur after surgery (increasing asymmetrical aberrations). All these factors may contribute to unexpected visual performance outcomes after aspheric IOL implantation. There are two procedures to achieve good clinical results with an aspherical IOL. On the one hand, the asphericity of the lens must be customized because it produces better results compared with results in patient population which aspheric IOLs were not customized 28. On the other hand, the centering of the IOL into capsular bag must be very accurate because if it is not achieved, the IOL will induce more HOAs than spherical design. In terms of depth-of-focus, spherical IOL provides more depth-of-focus than aspheric IOLs because the tolerance to defocus is highest in the

9 VISUAL SIMULATION IN MONOFOCAL IOLs ANALYSIS 151 spherical design. On figure 11, it can be appreciate that aspheric IOL has about 0.25 D and the spherical has about 0.75 D of depth-of-focus (we have considered depth-of-focus the range of VA values that are equal to or higher than 1.00). Marcos et al. 29 found that the tolerance to defocus was significantly higher with spherical IOLs than with aspheric IOLs. This was explained by Rocha et al. 24 because they pointed out that residual SA can improve depth-offocus and that the tolerance to defocus seems to be higher in eyes with a spherical IOL than in eyes with an aspheric IOL. However, it must be kept in mind that this refers to a monofocal IOL that was not designed to provide a large depth of focus, as are pseudoaccommodating IOL. In summary, the most clinical advantage of the visual simulator is that patients can check, with objective evidences and with a non-invasive procedure, their visual performance before any clinical procedure is carried out (i.e refractive surgery or adapt a visual corrector device). Clinically, visual simulator can be used by ophthalmologists and optometrists. With the visual simulator, ophthalmologists can show to their patients their visual performance before the refractive procedure is carried out. On the other hand, optometrists can show to their patients their visual performance if they adapt some visual corrector device. In both cases, they use the same simulation procedure. Firstly, they must measure and correct patients ocular aberrations. After that, they must induce patients ocular aberrations that their eye will have after the clinical procedure. On the other hand, researchers may use the visual simulator for other activities. They use this device to study which ocular surgery, ocular compensator device or compensator design induces better visual performance after their adaption. For example in the pilot study, we studied the visual performance of two IOLs designs when they are centered, decentered or tilted. Another use that researchers do of this device is to study the effect of ocular aberrations in the visual performance. All this research activities have the same aim: to design more physiologic ocular compensation devices to obtain better visual performance. REFERENCES 1. Poter J, Queener H, Lin J, Thorn K, Awwal A. Adaptive Optics for vision science. Canada:Wiley Interscience Pérez-Vices C, Belda Salmerón L, García-Lázaro S, Madrid Costa D, Ferrer-Blasco T. Adaptive optics, wavefront and visual simulation. J Emmetropia 2011; 2: López-Gil N, Castejón-Mochón J, Fernández-Sánchez V. Limitations of the ocular wavefront correction with contact lenses. Vision Research. 2009; 49: Kaiserman I, Hazarbassanov R, Varssano D, Girnbaum A. Contrast sensitivity after wave-fron-guided LASIK. Opthalmology. 2004; 111: D Arcy F, Kirwan C, Qasem Q, O keefe M. Prospective contralateral eye study to compare conventional and wavefronguided laser in situ keratomileusis. Acta Opthalmol Kershner RM. Retinal image contrast and functional visual performance with aspheric, silicone, and acrylic intraocular lenses; prospective evaluation. J Cataract Refract Surg. 2003; 29: Packer M, Fine IH, Hoffman RS, Piers PA. Improvement functional vision with a modified prolate intraocular lens. J Cataract Refract Surg. 2004; 30: Muñoz G, Albarrán-Diego C, Montés-Micó R, Rodríguez- Galietero A, Alió JL. Spherical aberration and contrast sensitivity after cataract surgery with the Tecnis Z9000 intraocular lens. J Cataract Refract Surg. 2006; 32: Rocha KM, Soriano ES, Chalita MR, et al. Wavefront analysis and contrast sensitivity of aspheric and spherical intraocular lenses: a randomized prospective study. Am J Ophthalmol. 2006; 142: Montés-Micó R, Ferrer-Blasco T, Cerviño A. Analysis of the possible benefits of aspheric intraocular lenses: Review of the literature. J Cataract Refract Surg. 2009; 35: Mester U, Dillinger P, Anterist N. Impact of a modified optic design on visual function: clinical comparative study. J Cataract Refract Surg. 2003; 29: Bellucci R, Scialdone A, Buratto L, et al. Visual acuity and contrast sensitivity comparison between Tecnis and AcrySof SA60AT intraocular lenses: a multicentre randomized study. J Cataract Refract Surg. 2005; 31: ; errata Rocha KM, Vabre L, Harms F, Chateau N, Krueger RR. Effects of Zernike wavefront aberrations on visual acuity measured using electromagnetic adaptive optics Technology. J Refract Surg. 2007; 23: Rouger H, Benard Y, Legras R. Effect of monochromatic induced aberrations on visual performance measured by adaptive optics technology. J Refract Surg. 2010; 26: Legras R, Rouger H. Calculations and measurements of the visual benefit of correcting the higher-order aberrations using adaptive optics technology. J Optom. 2008; 1: Rocha KM, Vabre L, Chateau N, Krueger R. Expanding depth of focus by modifying higher-order aberrations induced by an adaptive optics visual simulator. J Cataract Refract Surg. 2009; 35: Cerviño A, Hosking SL, Montés-Micó R, Bates K. Clinical ocular wavefront analyzers. J. Refract. Surg. 2007; 23: Liang J, Grimm B, Goelz S, Bille J. Objective measurement of wave aberrations of the human eye with the use of a Hartman- Shack wave-front sensor. J. Opt. Soc. Am. A. 1994; 11: Chateau N. Instructions of adaptive optics visual simulator Crx1. France: Imagine Eyes, López-Gil N, Bonaque S, Montés-Micó R. Optical quality of rotated or translated IOLs after in vitro wavefront measurements. XXVII congress of the ESCRS. Barcelona Wang L, Dai E, Koch DD, Nathoo A. Optical aberrations of the human anterior cornea. J Cataract Refract Surg. 2003; 29: Bach M. The «Freiburg Visual Acuity Test»- automatic measurement of visual acuity. Optom Vis Sci 1996;20: Dietze HH, Cox MJ. Limitations of correcting spherical aberrations with aspheric intraocular lenses. J Refract. 2005; 21: S541-S Rocha KM, Soriano ES, Chamon W, Chalita MR, Nosé W. Spherical aberration and depth of focus in eyes implanted with aspheric and spherical intraocular lenses; a prospective randomized study. Ophthalmology. 2007; 114:

10 152 VISUAL SIMULATION IN MONOFOCAL IOLs ANALYSIS 25. Wang L, Koch DD. Custom optimization of intraocular lens asphericity. J Cataract Refract Surg. 2007; 33: Wang L, Dai E, Koch DD, Nathoo A. Optical aberrations of the human anterior cornea. J Cataract Refract Surg. 2003; 29: Beiko GHH, Haigis W, Steinmueller A. Distribution of corneal spherical aberration in a comprehensive ophthalmology practice and whether keratometry can predict aberration values. J Cataract Refract Surg. 2007; 33: Atchison DA, Charman WN, Woods RL. Subjective depth-offocus of the eye. Optom Vis Sci. 1997; 74: Marcos S, Barbero S, Jiménez-Alfaro I. Optical quality and depth-of-field of eyes implanted with spherical and aspheric intraocular lenses. J Refract Surg. 2005; 21: First author: Alberto Domínguez Vicent, OD Department of Optics University of Valencia, Spain

10/25/2017. Financial Disclosures. Do your patients complain of? Are you frustrated by remake after remake? What is wavefront error (WFE)?

10/25/2017. Financial Disclosures. Do your patients complain of? Are you frustrated by remake after remake? What is wavefront error (WFE)? Wavefront-Guided Optics in Clinic: Financial Disclosures The New Frontier November 4, 2017 Matthew J. Kauffman, OD, FAAO, FSLS STAPLE Program Soft Toric and Presbyopic Lens Education Gas Permeable Lens

More information

Comparison of higher order aberrations with spherical and aspheric IOLs compared to normal phakic eyes

Comparison of higher order aberrations with spherical and aspheric IOLs compared to normal phakic eyes European Journal of Ophthalmology / Vol. 18 no. 5, 2008 / pp. 728-732 Comparison of higher order aberrations with spherical and aspheric IOLs compared to normal phakic eyes M. RĘKAS, K. KRIX-JACHYM, B.

More information

4th International Congress of Wavefront Sensing and Aberration-free Refractive Correction ADAPTIVE OPTICS FOR VISION: THE EYE S ADAPTATION TO ITS

4th International Congress of Wavefront Sensing and Aberration-free Refractive Correction ADAPTIVE OPTICS FOR VISION: THE EYE S ADAPTATION TO ITS 4th International Congress of Wavefront Sensing and Aberration-free Refractive Correction (Supplement to the Journal of Refractive Surgery; June 2003) ADAPTIVE OPTICS FOR VISION: THE EYE S ADAPTATION TO

More information

Customized Correction of Wavefront Aberrations in Abnormal Human Eyes by Using a Phase Plate and a Customized Contact Lens

Customized Correction of Wavefront Aberrations in Abnormal Human Eyes by Using a Phase Plate and a Customized Contact Lens Journal of the Korean Physical Society, Vol. 49, No. 1, July 2006, pp. 121 125 Customized Correction of Wavefront Aberrations in Abnormal Human Eyes by Using a Phase Plate and a Customized Contact Lens

More information

Update on Aspheric IOL Technology

Update on Aspheric IOL Technology Peer-Reviewed Literature: Update on Aspheric IOL Technology Editor: Ming Wang, MD, PhD, Clinical Associate Professor of Ophthalmology at the University of Tennessee and Director of the Wang Vision Institute

More information

Corneal Asphericity and Retinal Image Quality: A Case Study and Simulations

Corneal Asphericity and Retinal Image Quality: A Case Study and Simulations Corneal Asphericity and Retinal Image Quality: A Case Study and Simulations Seema Somani PhD, Ashley Tuan OD, PhD, and Dimitri Chernyak PhD VISX Incorporated, 3400 Central Express Way, Santa Clara, CA

More information

In vitro optical performance of a new aberration-free intraocular lens

In vitro optical performance of a new aberration-free intraocular lens (2014) 28, 614 620 & 2014 Macmillan Publishers Limited All rights reserved 0950-222X/14 www.nature.com/eye LABORATORY STUDY Optometry Research Group, Optics Department, University of Valencia, Valencia,

More information

What is Wavefront Aberration? Custom Contact Lenses For Vision Improvement Are They Feasible In A Disposable World?

What is Wavefront Aberration? Custom Contact Lenses For Vision Improvement Are They Feasible In A Disposable World? Custom Contact Lenses For Vision Improvement Are They Feasible In A Disposable World? Ian Cox, BOptom, PhD, FAAO Distinguished Research Fellow Bausch & Lomb, Rochester, NY Acknowledgements Center for Visual

More information

Aberrations Before and After Implantation of an Aspheric IOL

Aberrations Before and After Implantation of an Aspheric IOL Ocular High Order Aberrations Before and After Implantation of an Aspheric IOL Fabrizio I. Camesasca, MD Massimo Vitali, Orthoptist Milan, Italy I have no financial interest to disclose Wavefront Measurement

More information

The Aberration-Free IOL:

The Aberration-Free IOL: The Aberration-Free IOL: Advanced Optical Performance Independent of Patient Profile Griffith E. Altmann, M.S., M.B.A.; Keith H. Edwards, BSc FCOptom Dip CLP FAAO, Bausch & Lomb Some of these results were

More information

NOW. Approved for NTIOL classification from CMS Available in Quar ter Diopter Powers. Accommodating. Aberration Free. Aspheric.

NOW. Approved for NTIOL classification from CMS Available in Quar ter Diopter Powers. Accommodating. Aberration Free. Aspheric. NOW Approved for NTIOL classification from CMS Available in Quar ter Diopter Powers Accommodating. Aberration Free. Aspheric. Accommodation Meets Asphericity in AO Merging Innovation & Proven Design The

More information

Analysis of the possible benefits of aspheric intraocular lenses: Review of the literature

Analysis of the possible benefits of aspheric intraocular lenses: Review of the literature REVIEW/UPDATE Analysis of the possible benefits of aspheric intraocular lenses: Review of the literature Robert Montés-Micó, OD, MPhil, PhD, Teresa Ferrer-Blasco, OD, MSc, PhD, Alejandro Cerviño, OD, PhD

More information

Visual Outcomes of Two Aspheric PCIOLs: Tecnis Z9000 versus Akreos AO

Visual Outcomes of Two Aspheric PCIOLs: Tecnis Z9000 versus Akreos AO Visual Outcomes of Two Aspheric PCIOLs: Tecnis Z9000 versus Akreos AO Ahmad-Reza Baghi, MD; Mohammad-Reza Jafarinasab, MD; Hossein Ziaei, MD; Zahra Rahmani, MD Shaheed Beheshti Medical University, Tehran,

More information

The Aberration Structure of the Keratoconic Eye

The Aberration Structure of the Keratoconic Eye The Aberration Structure of the Keratoconic Eye Geunyoung Yoon, Ph.D. Department of Ophthalmology Center for Visual Science Institute of Optics Department of Biomedical Engineering University of Rochester

More information

Ron Liu OPTI521-Introductory Optomechanical Engineering December 7, 2009

Ron Liu OPTI521-Introductory Optomechanical Engineering December 7, 2009 Synopsis of METHOD AND APPARATUS FOR IMPROVING VISION AND THE RESOLUTION OF RETINAL IMAGES by David R. Williams and Junzhong Liang from the US Patent Number: 5,777,719 issued in July 7, 1998 Ron Liu OPTI521-Introductory

More information

CLINICAL SCIENCES. Corneal Optical Aberrations and Retinal Image Quality in Patients in Whom Monofocal Intraocular Lenses Were Implanted

CLINICAL SCIENCES. Corneal Optical Aberrations and Retinal Image Quality in Patients in Whom Monofocal Intraocular Lenses Were Implanted CLINICAL SCIENCES Corneal Optical Aberrations and Retinal Image Quality in Patients in Whom Monofocal Intraocular Lenses Antonio Guirao, PhD; Manuel Redondo, PhD; Edward Geraghty; Patricia Piers; Sverker

More information

Role of Asphericity in Choice of IOLs for Cataract Surgery

Role of Asphericity in Choice of IOLs for Cataract Surgery Role of Asphericity in Choice of IOLs for Cataract Surgery Delhi J Ophthalmol 2015; 25 (3): 185-189 DOI: http://dx.doi.org/10.7869/djo.105 Aman Khanna, Rebika Dhiman, Rajinder Khanna, Yajuvendra Singh

More information

Customized intraocular lenses

Customized intraocular lenses Customized intraocular lenses Challenges and limitations Achim Langenbucher, Simon Schröder & Timo Eppig Customized IOL what does this mean? Aspherical IOL Diffractive multifocal IOL Spherical IOL Customized

More information

Choices and Vision. Jeffrey Koziol M.D. Thursday, December 6, 12

Choices and Vision. Jeffrey Koziol M.D. Thursday, December 6, 12 Choices and Vision Jeffrey Koziol M.D. How does the eye work? What is myopia? What is hyperopia? What is astigmatism? What is presbyopia? How the eye works How the Eye Works 3 How the eye works Light rays

More information

Clinical Update for Presbyopic Lens Options

Clinical Update for Presbyopic Lens Options Clinical Update for Presbyopic Lens Options Gregory D. Searcy, M.D. Erdey Searcy Eye Group Columbus, Ohio The Problem = Spherical Optics Marginal Rays Spherical IOL Light Rays Paraxial Rays Spherical Aberration

More information

Normal Wavefront Error as a Function of Age and Pupil Size

Normal Wavefront Error as a Function of Age and Pupil Size RAA Normal Wavefront Error as a Function of Age and Pupil Size Raymond A. Applegate, OD, PhD Borish Chair of Optometry Director of the Visual Optics Institute College of Optometry University of Houston

More information

Impact of scattering and spherical aberration in contrast sensitivity

Impact of scattering and spherical aberration in contrast sensitivity Journal of Vision (2009) 9(3):19, 1 10 http://journalofvision.org/9/3/19/ 1 Impact of scattering and spherical aberration in contrast sensitivity Guillermo M. Pérez Silvestre Manzanera Pablo Artal Laboratorio

More information

Crystalens AO: Accommodating, Aberration-Free, Aspheric Y. Ralph Chu, MD Chu Vision Institute Bloomington, MN

Crystalens AO: Accommodating, Aberration-Free, Aspheric Y. Ralph Chu, MD Chu Vision Institute Bloomington, MN Crystalens AO: Accommodating, Aberration-Free, Aspheric Y. Ralph Chu, MD Chu Vision Institute Bloomington, MN Financial Disclosure Advanced Medical Optics Allergan Bausch & Lomb PowerVision Revision Optics

More information

Principles and clinical applications of ray-tracing aberrometry (Part II)

Principles and clinical applications of ray-tracing aberrometry (Part II) UPDATE/REVIEW Principles and clinical applications of ray-tracing aberrometry (Part II) Alfredo Castillo Gómez, MD, PhD 1 ; Antonio Verdejo del Rey, OD 2 ; Carlos Palomino Bautista, MD 3 ; Ana Escalada

More information

In this issue of the Journal, Oliver and colleagues

In this issue of the Journal, Oliver and colleagues Special Article Refractive Surgery, Optical Aberrations, and Visual Performance Raymond A. Applegate, OD, PhD; Howard C. Howland,PhD In this issue of the Journal, Oliver and colleagues report that photorefractive

More information

*Simulated vision. **Individual results may vary and are not guaranteed. Visual Performance When It s Needed Most

*Simulated vision. **Individual results may vary and are not guaranteed. Visual Performance When It s Needed Most Simulated vision. Individual results may vary and are not guaranteed. Visual Performance When It s Needed Most The aspheric design of the AcrySof IQ IOL results in improved clarity and image quality. The

More information

Design of a Test Bench for Intraocular Lens Optical Characterization

Design of a Test Bench for Intraocular Lens Optical Characterization Journal of Physics: Conference Series Design of a Test Bench for Intraocular Lens Optical Characterization To cite this article: Francisco Alba-Bueno et al 20 J. Phys.: Conf. Ser. 274 0205 View the article

More information

Postoperative Wavefront Analysis and Contrast Sensitivity of a Multifocal Apodized Diffractive IOL (ReSTOR) and Three Monofocal IOLs

Postoperative Wavefront Analysis and Contrast Sensitivity of a Multifocal Apodized Diffractive IOL (ReSTOR) and Three Monofocal IOLs Postoperative Wavefront Analysis and Contrast Sensitivity of a Multifocal Apodized Diffractive IOL (ReSTOR) and Three Monofocal IOLs Karolinne Maia Rocha, MD; Maria Regina Chalita, MD; Carlos Eduardo B.

More information

Wavefront Aberrations in Eyes With Acrysof Monofocal Intraocular Lenses

Wavefront Aberrations in Eyes With Acrysof Monofocal Intraocular Lenses Wavefront Aberrations in Eyes With Acrysof Monofocal Intraocular Lenses Prema Padmanabhan, MS; Geunyoung Yoon, PhD; Jason Porter, PhD; Srinivas K. Rao, FRCSEd; Roy J, MSc; Mitalee Choudhury, BS ABSTRACT

More information

Comparison of contrast sensitivity and color discrimination after clear and yellow intraocular lens implantation

Comparison of contrast sensitivity and color discrimination after clear and yellow intraocular lens implantation J CATARACT REFRACT SURG - VOL 31, SEPTEMBER 2005 Comparison of contrast sensitivity and color discrimination after clear and yellow intraocular lens implantation Antonio Rodríguez-Galietero, MD, PhD, FEBO,

More information

Adaptive Optics. Adaptive optics for imaging. Adaptive optics to improve. Ocular High order Aberrations (HOA)

Adaptive Optics. Adaptive optics for imaging. Adaptive optics to improve. Ocular High order Aberrations (HOA) Effect of Adaptive Optics Correction on Visual Performance and Accommodation Adaptive optics for imaging Astromomy Retinal imaging Since 977, Hardy et al, JOSA A Since 989, Dreher et al. Appl Opt Susana

More information

Treatment of Presbyopia during Crystalline Lens Surgery A Review

Treatment of Presbyopia during Crystalline Lens Surgery A Review Treatment of Presbyopia during Crystalline Lens Surgery A Review Pierre Bouchut Bordeaux Ophthalmic surgeons should treat presbyopia during crystalline lens surgery. Thanks to the quality and advancements

More information

Choices and Vision. Jeffrey Koziol M.D. Friday, December 7, 12

Choices and Vision. Jeffrey Koziol M.D. Friday, December 7, 12 Choices and Vision Jeffrey Koziol M.D. How does the eye work? What is myopia? What is hyperopia? What is astigmatism? What is presbyopia? How the eye works Light rays enter the eye through the clear cornea,

More information

The design is distinctive. The outcomes are clear. Defocus tolerance 1 Glistening-free performance 1,2 Predictable outcomes 1

The design is distinctive. The outcomes are clear. Defocus tolerance 1 Glistening-free performance 1,2 Predictable outcomes 1 The design is distinctive. The outcomes are clear. Defocus tolerance 1 Glistening-free performance 1,2 Predictable outcomes 1 The clear choice for consistent visual excellence. For over 165 years Bausch

More information

HOYA aspherical IOL with ABC (Aspheric Balanced Curve) Design

HOYA aspherical IOL with ABC (Aspheric Balanced Curve) Design HOYA aspherical IOL with ABC (Aspheric Balanced Curve) Design Contents Basics of asphericity Visual quality and aspheric IOL Features of HOYA ABC Design 2 What is asphericity? Deviating from the spherical

More information

Pablo Artal. collaborators. Adaptive Optics for Vision: The Eye's Adaptation to its Point Spread Function

Pablo Artal. collaborators. Adaptive Optics for Vision: The Eye's Adaptation to its Point Spread Function contrast sensitivity Adaptive Optics for Vision: The Eye's Adaptation to its Point Spread Function (4 th International Congress on Wavefront Sensing, San Francisco, USA; February 23) Pablo Artal LABORATORIO

More information

WaveMaster IOL. Fast and accurate intraocular lens tester

WaveMaster IOL. Fast and accurate intraocular lens tester WaveMaster IOL Fast and accurate intraocular lens tester INTRAOCULAR LENS TESTER WaveMaster IOL Fast and accurate intraocular lens tester WaveMaster IOL is a new instrument providing real time analysis

More information

OPTOMETRY RESEARCH PAPER. Optical quality comparison among different Boston contact lens materials

OPTOMETRY RESEARCH PAPER. Optical quality comparison among different Boston contact lens materials C L I N I C A L A N D E X P E R I M E N T A L OPTOMETRY RESEARCH PAPER Optical quality comparison among different Boston contact lens materials Clin Exp Optom 2016; 99: 39 46 Alberto Domínguez-Vicent MSc

More information

MODERN CATARACT SURGERY AND LENS REplacement

MODERN CATARACT SURGERY AND LENS REplacement Wavefront Analysis and Contrast Sensitivity of Aspheric and Spherical Intraocular Lenses: A Randomized Prospective Study KAROLINNE MAIA ROCHA, MD, EDUARDO S. SORIANO, MD, MARIA REGINA CHALITA, MD, ANA

More information

Objective and subjective outcomes in comparing three different aspheric intraocular lens implants with their spherical counterparts

Objective and subjective outcomes in comparing three different aspheric intraocular lens implants with their spherical counterparts (2009) 23, 877 883 & 2009 Macmillan Publishers Limited All rights reserved 0950-222X/09 $32.00 www.nature.com/eye Objective and subjective outcomes in comparing three different aspheric intraocular lens

More information

Evaluation of the Impact of Intraocular Lens Tecnis Z9000 Misalignment on the Visual Quality Using the Optical Eye Modeling

Evaluation of the Impact of Intraocular Lens Tecnis Z9000 Misalignment on the Visual Quality Using the Optical Eye Modeling Evaluation of the Impact of Intraocular Lens Tecnis Z9000 Misalignment on the Visual Quality Using the Optical Eye Modeling Azam Asgari 1 Ali Asghar Parach 1 Keykhosro Keshavarzi 2 Abstract Purpose: The

More information

Long-term quality of vision is what every patient expects

Long-term quality of vision is what every patient expects Long-term quality of vision is what every patient expects Innovative combination of HOYA technologies provides: 1-piece aspheric lens with Vivinex hydrophobic acrylic material Unique surface treatment

More information

Assessing Visual Quality With the Point Spread Function Using the NIDEK OPD-Scan II

Assessing Visual Quality With the Point Spread Function Using the NIDEK OPD-Scan II Assessing Visual Quality With the Point Spread Function Using the NIDEK OPD-Scan II Edoardo A. Ligabue, MD; Cristina Giordano, OD ABSTRACT PURPOSE: To present the use of the point spread function (PSF)

More information

University of Groningen. Young eyes for elderly people van Gaalen, Kim

University of Groningen. Young eyes for elderly people van Gaalen, Kim University of Groningen Young eyes for elderly people van Gaalen, Kim IMPORTANT NOTE: You are advised to consult the publisher's version (publisher's PDF) if you wish to cite from it. Please check the

More information

This is the author s version of a work that was submitted/accepted for publication in the following source:

This is the author s version of a work that was submitted/accepted for publication in the following source: This is the author s version of a work that was submitted/accepted for publication in the following source: Atchison, David A. & Mathur, Ankit (2014) Effects of pupil center shift on ocular aberrations.

More information

Explanation of Aberration and Wavefront

Explanation of Aberration and Wavefront Explanation of Aberration and Wavefront 1. What Causes Blur? 2. What is? 4. What is wavefront? 5. Hartmann-Shack Aberrometer 6. Adoption of wavefront technology David Oh 1. What Causes Blur? 2. What is?

More information

Roadmap to presbyopic success

Roadmap to presbyopic success Roadmap to presbyopic success Miltos O Balidis MD, PhD, FEBOphth, ICOphth Early experience with Presbyopic correction 2003 Binocular Distance-Corrected Intermediate and Near Vision Binocular Distance-Corrected

More information

Effects of Pupil Center Shift on Ocular Aberrations

Effects of Pupil Center Shift on Ocular Aberrations Visual Psychophysics and Physiological Optics Effects of Pupil Center Shift on Ocular Aberrations David A. Atchison and Ankit Mathur School of Optometry & Vision Science and Institute of Health & Biomedical

More information

Comparison of Visual Acuity, Contrast Sensitivity and Spherical Aberration after Implantation of Aspheric and Spheric Intraocular Lenses

Comparison of Visual Acuity, Contrast Sensitivity and Spherical Aberration after Implantation of Aspheric and Spheric Intraocular Lenses Comparison of Visual Acuity, Contrast Sensitivity and Spherical Aberration after Implantation of Aspheric and Spheric Intraocular Lenses Mohammad Nasser Hashemian, MD 1 Morteza Movassat, MD 2 Abdolreza

More information

Raise your expectations. Deliver theirs.

Raise your expectations. Deliver theirs. 66 EXTENDED RANGE OF VISION MONOFOCAL-LIKE DISTANCE Raise your expectations. Deliver theirs. Now you can give your patients the best of both worlds with the first and only hybrid designed monofocal-multifocal

More information

Abetter understanding of the distribution of aberrations in

Abetter understanding of the distribution of aberrations in Predicting the Optical Performance of Eyes Implanted with IOLs to Correct Spherical Aberration Juan Tabernero, 1 Patricia Piers, 2 Antonio Benito, 1 Manuel Redondo, 3 and Pablo Artal 1 PURPOSE. To use

More information

SEE BEYOND WITH FULLRANGE OPTICS. Developed by Hanita Lenses

SEE BEYOND WITH FULLRANGE OPTICS. Developed by Hanita Lenses SEE BEYOND WITH FULLRANGE OPTICS Developed by Hanita Lenses SEE beyond with FullRange optics FullRange optic lenses are proven, highlyreliable and safe intraocular lenses designed to provide a solution

More information

WaveMaster IOL. Fast and Accurate Intraocular Lens Tester

WaveMaster IOL. Fast and Accurate Intraocular Lens Tester WaveMaster IOL Fast and Accurate Intraocular Lens Tester INTRAOCULAR LENS TESTER WaveMaster IOL Fast and accurate intraocular lens tester WaveMaster IOL is an instrument providing real time analysis of

More information

Clinical Evaluation 3-month Follow-up Report

Clinical Evaluation 3-month Follow-up Report Clinical Evaluation 3-month Follow-up Report Of SeeLens HP Intraocular Lens 27 December 2010 version 1.1 1of 16 Table of Contents TABLE OF CONTENTS... 1 OBJECTIVES... 2 EFFICACY AND SAFETY ASSESSMENTS...

More information

Correcting Highly Aberrated Eyes Using Large-stroke Adaptive Optics

Correcting Highly Aberrated Eyes Using Large-stroke Adaptive Optics Correcting Highly Aberrated Eyes Using Large-stroke Adaptive Optics Ramkumar Sabesan, BTech; Kamran Ahmad, MS; Geunyoung Yoon, PhD ABSTRACT PURPOSE: To investigate the optical performance of a large-stroke

More information

Comparison of Contrast Sensitivity in Eyes with Toric and Aspheric Intraocular Lenses

Comparison of Contrast Sensitivity in Eyes with Toric and Aspheric Intraocular Lenses ARTICLE Comparison of Contrast Sensitivity in Eyes with Toric and Aspheric Intraocular Lenses Francisco Pastor-Pascual, PhD 1 ; Aitor Lanzagorta-Aresti, MD 1 ; Pablo Alcocer-Yuste, MD 1 ; Cristina Puchades-Lladró

More information

Correlation of pupil size with visual acuity and contrast sensitivity after implantation of an apodized diffractive intraocular lens

Correlation of pupil size with visual acuity and contrast sensitivity after implantation of an apodized diffractive intraocular lens ARTICLE Correlation of pupil size with visual acuity and contrast sensitivity after implantation of an apodized diffractive intraocular lens José F. Alfonso, MD, PhD, Luis Fernández-Vega, MD, PhD, M. Begoña

More information

Optical Connection, Inc. and Ophthonix, Inc.

Optical Connection, Inc. and Ophthonix, Inc. Optical Connection, Inc. and Ophthonix, Inc. Partners in the delivery of nonsurgical vision optimization www.opticonnection.com www.ophthonix.com The human eye has optical imperfections that can not be

More information

Pablo Artal. Adaptive Optics visual simulator ( and depth of focus) LABORATORIO DE OPTICA UNIVERSIDAD DE MURCIA, SPAIN

Pablo Artal. Adaptive Optics visual simulator ( and depth of focus) LABORATORIO DE OPTICA UNIVERSIDAD DE MURCIA, SPAIN Adaptive Optics visual simulator ( and depth of focus) Pablo Artal LABORATORIO DE OPTICA UNIVERSIDAD DE MURCIA, SPAIN 8th International Wavefront Congress, Santa Fe, USA, February New LO UM building! Diego

More information

Research Article In Vitro Aberrometric Assessment of a Multifocal Intraocular Lens and Two Extended Depth of Focus IOLs

Research Article In Vitro Aberrometric Assessment of a Multifocal Intraocular Lens and Two Extended Depth of Focus IOLs Hindawi Ophthalmology Volume 2017, Article ID 7095734, 7 pages https://doi.org/10.1155/2017/7095734 Research Article In Vitro Aberrometric Assessment of a Multifocal Intraocular Lens and Two Extended Depth

More information

Surgical data reveals that Q-Factor is important for good surgical outcome

Surgical data reveals that Q-Factor is important for good surgical outcome Surgical data reveals that Q-Factor is important for good surgical outcome Michael Mrochen, PhD Michael Bueeler, PhD Tobias Koller, MD Theo Seiler, MD, PhD IROC AG Institut für Refraktive und Ophthalmo-Chirurgie

More information

Prospective sual evaluation of apodized diffractive intraocular lenses

Prospective sual evaluation of apodized diffractive intraocular lenses See discussions, stats, and author profiles for this publication at: https://www.researchgate.net/publication/6251759 Prospective sual evaluation of apodized diffractive intraocular lenses ARTICLE in JOURNAL

More information

Improving Lifestyle Vision. with Small Aperture Optics

Improving Lifestyle Vision. with Small Aperture Optics Improving Lifestyle Vision with Small Aperture Optics The Small Aperture Premium Lens Solution The IC-8 small aperture intraocular lens (IOL) is a revolutionary lens that extends depth of focus by combining

More information

Optical aberrations and the eye Part 3

Optical aberrations and the eye Part 3 clinical 22 Optical aberrations and the eye Part 3 In the final part of our series, Alejandro Cerviño and Dr Shehzad Naroo discuss the methods of correction required for low and high order wavefront aberrations

More information

Although, during the last decade, peripheral optics research

Although, during the last decade, peripheral optics research Visual Psychophysics and Physiological Optics Comparison of the Optical Image Quality in the Periphery of Phakic and Pseudophakic Eyes Bart Jaeken, 1 Sandra Mirabet, 2 José María Marín, 2 and Pablo Artal

More information

ORIGINAL ARTICLE. ESTHER MORENO-BARRIUSO, PhD, SUSANA MARCOS, PhD, RAFAEL NAVARRO, PhD, and STEPHEN A. BURNS, PhD

ORIGINAL ARTICLE. ESTHER MORENO-BARRIUSO, PhD, SUSANA MARCOS, PhD, RAFAEL NAVARRO, PhD, and STEPHEN A. BURNS, PhD 1040-5488/01/7803-0152/0 VOL. 78, NO. 3, PP. 152 156 OPTOMETRY AND VISION SCIENCE Copyright 2001 American Academy of Optometry ORIGINAL ARTICLE Comparing Laser Ray Tracing, the Spatially Resolved Refractometer,

More information

The Blueprint for Improved Image Quality

The Blueprint for Improved Image Quality Insert to January 2009 The Blueprint for Improved Image Quality The function and design of the AcrySof IQ IOL. Supported by an educational grant from Alcon Laboratories, Inc. The Function and Design of

More information

UNIVERSIDAD COMPLUTENSE DE MADRID

UNIVERSIDAD COMPLUTENSE DE MADRID UNIVERSIDAD COMPLUTENSE DE MADRID FACULTAD DE OPTICA Y OPTOMETRÍA Departamento de Óptica TESIS DOCTORAL Vision under manipulated aberrations : towards improved multifocal corrections MEMORIA PARA OPTAR

More information

TRANSLATIONAL SCIENCE. Effect of Crystalline Lens Aberrations on Adaptive Optics Simulation of Intraocular Lenses

TRANSLATIONAL SCIENCE. Effect of Crystalline Lens Aberrations on Adaptive Optics Simulation of Intraocular Lenses TRANSLATIONAL SCIENCE Effect of Crystalline Lens Aberrations on Adaptive Optics Simulation of Intraocular Lenses Eloy A. Villegas, PhD; Silvestre Manzanera, PhD; Carmen M. Lago, MSc; Lucía Hervella, MSc;

More information

Development of a Calibration Standard for Spherical Aberration

Development of a Calibration Standard for Spherical Aberration Development of a Calibration Standard for David C. Compertore, Filipp V. Ignatovich, Matthew E. Herbrand, Michael A. Marcus, Lumetrics, Inc. 1565 Jefferson Road, Rochester, NY (United States) ABSTRACT

More information

Effect of an aspheric intraocular lens on the ocular wave-front adjusted for pupil size and capsulorhexis size

Effect of an aspheric intraocular lens on the ocular wave-front adjusted for pupil size and capsulorhexis size Effect of an aspheric intraocular lens on the ocular wave-front adjusted for pupil size and capsulorhexis size Alja Crnej, 1 Wolf Buehl, 2 Roman Greslechner, 2 Nino Hirnschall 1,3 and Oliver Findl 1,3

More information

Accommodation with higher-order monochromatic aberrations corrected with adaptive optics

Accommodation with higher-order monochromatic aberrations corrected with adaptive optics Chen et al. Vol. 23, No. 1/ January 2006/ J. Opt. Soc. Am. A 1 Accommodation with higher-order monochromatic aberrations corrected with adaptive optics Li Chen Center for Visual Science, University of

More information

Adaptive Optics for Vision Science. Principles, Practices, Design, and Applications

Adaptive Optics for Vision Science. Principles, Practices, Design, and Applications Adaptive Optics for Vision Science Principles, Practices, Design, and Applications Edited by JASON PORTER, HOPE M. QUEENER, JULIANNA E. LIN, KAREN THORN, AND ABDUL AWWAL m WILEY- INTERSCIENCE A JOHN WILEY

More information

Transferring wavefront measurements to ablation profiles. Michael Mrochen PhD Swiss Federal Institut of Technology, Zurich IROC Zurich

Transferring wavefront measurements to ablation profiles. Michael Mrochen PhD Swiss Federal Institut of Technology, Zurich IROC Zurich Transferring wavefront measurements to ablation profiles Michael Mrochen PhD Swiss Federal Institut of Technology, Zurich IROC Zurich corneal ablation Calculation laser spot positions Centration Calculation

More information

ORIGINAL ARTICLE. Correlation between Optical and Psychophysical Parameters as a Function of Defocus

ORIGINAL ARTICLE. Correlation between Optical and Psychophysical Parameters as a Function of Defocus 1040-5488/02/7901-0001/0 VOL. 79, NO. 1, PP. 60-67 OPTOMETRY AND VISION SCIENCE Copyright 2002 American Academy of Optometry A schematic view of the apparatus used is shown in Fig. 1. It is a double-pass

More information

Aberration Interaction In Wavefront Guided Custom Ablation

Aberration Interaction In Wavefront Guided Custom Ablation Aberration Interaction In Wavefront Guided Custom Ablation Scott M. MacRae MD Professor of Ophthalmology Professor of Visual Science University of Rochester Collaborators and Disclosures: Manoj Subbaram

More information

Unique Aberration-Free IOL: A Vision that Patients

Unique Aberration-Free IOL: A Vision that Patients Unique Aberration-Free IOL: A Vision that Patients Can Appreciate An Aspheric Optic for Improved Quality of Vision n Traditional spherical IOLs create Bilateral implantation study spherical aberration

More information

Causes of refractive error post premium IOL s 3/17/2015. Instruction course: Refining the Refractive Error After Premium IOL s.

Causes of refractive error post premium IOL s 3/17/2015. Instruction course: Refining the Refractive Error After Premium IOL s. Instruction course: Refining the Refractive Error After Premium IOL s. Senior Instructor: Mounir Khalifa, MD Instructors: David Hardten,MD Scott MacRea,MD Matteo Piovella,MD Dr. Khalifa: Causes of refractive

More information

In recent years there has been an explosion of

In recent years there has been an explosion of Line of Sight and Alternative Representations of Aberrations of the Eye Stanley A. Klein, PhD; Daniel D. Garcia, PhD ABSTRACT Several methods for representing pupil plane aberrations based on wavefront

More information

phone extn.3662, fax: , nitt.edu ABSTRACT

phone extn.3662, fax: , nitt.edu ABSTRACT Analysis of Refractive errors in the human eye using Shack Hartmann Aberrometry M. Jesson, P. Arulmozhivarman, and A.R. Ganesan* Department of Physics, National Institute of Technology, Tiruchirappalli

More information

Contrast Sensitivity after Refractive Lens Exchange with A Multifocal Diffractive Aspheric Intraocular Lens

Contrast Sensitivity after Refractive Lens Exchange with A Multifocal Diffractive Aspheric Intraocular Lens Contrast Sensitivity after Refractive Lens Exchange with A Multifocal Diffractive Aspheric Intraocular Lens Teresa Ferrer-Blasco, PhD Santiago García-Lázaro, PhD César Albarrán-Diego, MSc 3 Lurdes Belda-Salmerón,

More information

ORIGINAL ARTICLE. Predicting and Assessing Visual Performance with Multizone Bifocal Contact Lenses. JOY A. MARTIN, OD and AUSTIN ROORDA, PhD

ORIGINAL ARTICLE. Predicting and Assessing Visual Performance with Multizone Bifocal Contact Lenses. JOY A. MARTIN, OD and AUSTIN ROORDA, PhD 1040-5488/03/8012-0812/0 VOL. 80, NO. 12, PP. 812 819 OPTOMETRY AND VISION SCIENCE Copyright 2003 American Academy of Optometry ORIGINAL ARTICLE Predicting and Assessing Visual Performance with Multizone

More information

Generation of third-order spherical and coma aberrations by use of radially symmetrical fourth-order lenses

Generation of third-order spherical and coma aberrations by use of radially symmetrical fourth-order lenses López-Gil et al. Vol. 15, No. 9/September 1998/J. Opt. Soc. Am. A 2563 Generation of third-order spherical and coma aberrations by use of radially symmetrical fourth-order lenses N. López-Gil Section of

More information

Corporate Perspective Alcon Unanswered Technical Challenges that Still Need to be Overcome

Corporate Perspective Alcon Unanswered Technical Challenges that Still Need to be Overcome Corporate Perspective Alcon Unanswered Technical Challenges that Still Need to be Overcome Ronald Krueger, MD Refractive Industry Challenges Diagnostic Improvement Optimal Laser Performance Corneal Factors

More information

IOL Types. Hazem Elbedewy. M.D., FRCS (Glasg.) Lecturer of Ophthalmology Tanta university

IOL Types. Hazem Elbedewy. M.D., FRCS (Glasg.) Lecturer of Ophthalmology Tanta university IOL Types Hazem Elbedewy M.D., FRCS (Glasg.) Lecturer of Ophthalmology Tanta university Artificial intraocular lenses are used to replace the eye natural lens when it has been removed during cataract surgery.

More information

In Vitro Strehl Ratios with Spherical, Aberration-Free, and customized spherical aberration-correcting METHODS

In Vitro Strehl Ratios with Spherical, Aberration-Free, and customized spherical aberration-correcting METHODS In Vitro Strehl Ratios with Spherical, Aberration-Free, Average, and Customized Spherical Aberration-Correcting Intraocular Lenses Stefan Pieh, 1 Werner Fiala, 1 Andre Malz, 2 and Wilhelm Stork 2 PURPOSE.

More information

ROTATIONAL STABILITY MAKES THE DIFFERENCE

ROTATIONAL STABILITY MAKES THE DIFFERENCE The Bi-Flex platform the proven platform of Excellence 01 Proven Stability less than 02 Optimal biomaterials 2 degrees long term rotation 03 Posterior Toric Lens surface with marks indicating the flat

More information

Subjective Image Quality Metrics from The Wave Aberration

Subjective Image Quality Metrics from The Wave Aberration Subjective Image Quality Metrics from The Wave Aberration David R. Williams William G. Allyn Professor of Medical Optics Center For Visual Science University of Rochester Commercial Relationship: Bausch

More information

CLINICAL SCIENCE INTRODUCTION

CLINICAL SCIENCE INTRODUCTION CLINICS 2009;64(10):953-60 CLINICAL SCIENCE A CONTRALATERAL EYE STUDY COMPARING APODIZED DIFFRATIVE AND FULL DIFFRATIVE LENSES: WAVEFRONT ANALYSIS AND DISTANCE AND NEAR UNCORRECTED VISUAL ACUITY Marcony

More information

Energy balance in apodized diffractive multifocal intraocular lenses

Energy balance in apodized diffractive multifocal intraocular lenses Energy balance in apodized diffractive multifocal intraocular lenses Francisco Alba-Bueno *a, Fidel Vega a, María S. Millán a a Optics and Optometry Department, Polytechnic University of Catalonia. ABSTRACT

More information

Multifocal Intraocular Lenses for the Treatment of Presbyopia: Benefits and Side-effects

Multifocal Intraocular Lenses for the Treatment of Presbyopia: Benefits and Side-effects Published on Points de Vue International Review of Ophthalmic Optics () Home > Multifocal Intraocular Lenses for the Treatment of Presbyopia: Benefits and Side-effects Multifocal Intraocular Lenses for

More information

Effect of rotation and translation on the expected benefit of an ideal method to correct the eye s higher-order aberrations

Effect of rotation and translation on the expected benefit of an ideal method to correct the eye s higher-order aberrations Guirao et al. Vol. 18, No. 5/May 2001/J. Opt. Soc. Am. A 1003 Effect of rotation and translation on the expected benefit of an ideal method to correct the eye s higher-order aberrations Antonio Guirao

More information

7 DO IT. A SAFER WAY TO TRIFOCALITY * : ELEVATED PHASE SHIFT (EPS) ** 7 DIFFRACTIVE RINGS FOR OPTIMAL LIGHT DISTRIBUTION AND LESS DISTURBANCE

7 DO IT. A SAFER WAY TO TRIFOCALITY * : ELEVATED PHASE SHIFT (EPS) ** 7 DIFFRACTIVE RINGS FOR OPTIMAL LIGHT DISTRIBUTION AND LESS DISTURBANCE A SAFER WAY TO TRIFOCALITY * : ELEVATED PHASE SHIFT (EPS) ** 7 DIFFRACTIVE RINGS FOR OPTIMAL LIGHT DISTRIBUTION AND LESS DISTURBANCE 7 DO IT. TRIFOCAL PERFORMANCE & GLASS INDEPENDENCY UNCOMPROMISED CONTRAST

More information

Repeatability of measurements with a double-pass system

Repeatability of measurements with a double-pass system ARTICLE Repeatability of measurements with a double-pass system Alain Saad, MD, Marc Saab, MD, Damien Gatinel, MD, PhD PURPOSE: To evaluate the repeatability of measurements with a double-pass system.

More information

Evolution of Diffractive Multifocal Intraocular Lenses

Evolution of Diffractive Multifocal Intraocular Lenses Evolution of Diffractive Multifocal Intraocular Lenses Wavefront Congress February 24, 2007 Michael J. Simpson, Ph.D. Alcon Research, Ltd., Fort Worth, Texas Presentation Overview Multifocal IOLs two lens

More information

VISUAL ACUITY AND IMAGE QUALITY IN FIVE DIFFRACTIVE. Short title: Visual acuity and image quality in five diffractive intraocular lenses

VISUAL ACUITY AND IMAGE QUALITY IN FIVE DIFFRACTIVE. Short title: Visual acuity and image quality in five diffractive intraocular lenses VISUAL ACUITY AND IMAGE QUALITY IN FIVE DIFFRACTIVE INTRAOCULAR LENSES Short title: Visual acuity and image quality in five diffractive intraocular lenses Genís Cardona a, Fidel Vega a, Miguel A. Gil b,

More information

Through-focus response of multifocal intraocular lenses evaluated with a spatial light modulator

Through-focus response of multifocal intraocular lenses evaluated with a spatial light modulator Through-focus response of multifocal intraocular lenses evaluated with a spatial light modulator Laura Remón, 1 Augusto Arias, 2,3 Arnau Calatayud, 1 Walter D. Furlan, 3 and Juan A. Monsoriu 1, * 1 Centro

More information

Refractive Power / Corneal Analyzer. OPD-Scan III

Refractive Power / Corneal Analyzer. OPD-Scan III Refractive Power / Corneal Analyzer OPD-Scan III Comprehensive Vision Analysis and NIDEK, a global leader in ophthalmic and optometric equipment, has created the OPD-Scan III, the third generation aberrometer

More information

Calculated impact of higher-order monochromatic aberrations on retinal image quality in a population of human eyes: erratum

Calculated impact of higher-order monochromatic aberrations on retinal image quality in a population of human eyes: erratum ERRATA Calculated impact of higher-order monochromatic aberrations on retinal image quality in a population of human eyes: erratum Antonio Guirao* Laboratorio de Optica, Departamento de Física, Universidad

More information

Is Aberration-Free Correction the Best Goal

Is Aberration-Free Correction the Best Goal Is Aberration-Free Correction the Best Goal Stephen Burns, PhD, Jamie McLellan, Ph.D., Susana Marcos, Ph.D. The Schepens Eye Research Institute. Schepens Eye Research Institute, an affiliate of Harvard

More information