Service Manual GMX X-Ray Generator

Size: px
Start display at page:

Download "Service Manual GMX X-Ray Generator"

Transcription

1 Service Manual GMX X-Ray Generator CMR X-Ray, LLC 2000 Pewaukee Road, Suite H Waukesha, Wisconsin Copyright CMR X-Ray, March 2015

2 GMX X-RAY GENERATOR SERVICE MANUAL CONTENTS 1) PRE- INSTALLATION Introduction Dimensions, Weight and Cable Access Environmental Requirements Packaging Electrical Requirements, 208 to 480 Volt Lines Electrical Requirements, Power Assist Units Safety Devices ) INSTALLATION Introduction Tools and Materials Measurement Equipment Pre- Installation Check Unpacking Power Cabinet Installation / power assist Operator Console Installation Interface Connections Generator Connections Left side Generator connections right side T1 Voltage Taps, Rotor and Filament X- Ray Tube Stator Connections DiGital Interface, 2TS Relay Board, Bucky Interface, 2TS System Connect, 2TS T1 Power Module AUX Power T1 Taps (Rotor & Filament) and Fuses TS- 1-1 Power assist transformer input i Copyright CMR X-Ray, 2015

3 3) GENERATOR CONFIGURATION Configuration procedure ) CALIBRATION TRUE- TRACK Calibration with Computer Typical ma and kv Wave Forms ) TROUBLESHOOTING Introduction Service RECOMMENDATIONs Error Codes ERROR Code Descriptions ) RENEWAL PARTS Power Module LINE Inverter ) ACCEPTANCE TESTING Introduction Preliminary Testing X- Ray Time Measurement kv/ma X- Ray Measurement Linearity Repeatability SUPPLEMENT B POWER ASSIST SUPPLEMENT C USEFUL TABLES SUPPLEMENT D HV CABLE MAINTENANCE ii

4 1) PRE-INSTALLATION 1.1 INTRODUCTION Planning and preparing the customer site. This list contains the main topics to consider during pre-installation planning: Equipment dimensions Installation and sizing of the electrical supply Installation of safety devices, Door Interlocks, X-Ray On-Lamps, Disconnect Circuit Breakers, etc. 1.2 DIMENSIONS, WEIGHT AND CABLE ACCESS POWER MODULE DIMENSIONS: FRONT 23 SIDE 24 HEIGTH 19.5 WEIGHT 178 lbs. (Illustrated W/O covers) 1

5 BASIC RAD CONSOLE DIMENSIONS: FRONT SIDE 9.75 HEIGTH 4.5 WEIGHT 4.75 lbs. POWER ASSIST DIMENSIONS: FRONT 23 SIDE 24 HEIGTH 10.0 WEIGHT 98 lbs. (Illustrated W/O covers) 1.1 GENERATOR COMPONENTS 2

6 1.3 ENVIRONMENTAL REQUIREMENTS Temperature: 10ºC to 40ºC (50 F to 104 F). Relative Humidity 0 to 90%. Maximum Altitude: 2440 meters (8000 feet) above sea level. Note: This X-Ray Generator will dissipate 544 btu/hr. 1.4 PACKAGING 1.7 SHIPPING CRATE 3

7 1.5 ELECTRICAL REQUIREMENTS, 208 TO 480 VOLT LINES Model Power Supply Voltage Frequency (Hz) GMX 325 RAD 30 kw 208/220, 1Ø 50/60 GMX 525 RAD 50 kw 220, 3Ø 50/60 GMX 625 RAD 60 kw 380, 3Ø 50/60 GMX 825 RAD 80 kw 480, 3Ø 50/ LINE VOLTAGES Power and Ground Wire Size (AWG): Model Power Ground Power Ground Power Ground Power Ground Power Ground Power Ground 0 to 30m 31 to 50m 51 to 70m 71 to 90m 91 to 110m 111 to 130m GMX 325 RAD 1/0 6 1/0 6 2/0 2 3/0 2 4/0 1/0 250 MCM 1/0 GMX 525 RAD 1/0 6 1/0 6 2/0 2 3/0 2 4/0 1/0 250 MCM 1/0 GMX 625 RAD 2 6 1/0 6 2/0 2 3/0 2 4/0 1/0 250 MCM 1/0 GMX 825 RAD 2 8 1/0 6 2/0 2 3/0 2 4/0 1/0 250 MCM 1/ POWER AND GROUND WIRE SIZING 4

8 electrical requirements continued. Line Disconnect to Power Module Model Power Ground 0 to 10m GMX 325 RAD 2AWG 2AWG GMX 525 RAD 2AWG 2AWG GMX 625 RAD 2AWG 6AWG GMX 825 RAD 2AWG 6AWG 1.4 WIRE SIZING, TO POWER MODULE Model Power Transformer Capacity (kva) Main Disconnect Switch GMX 325 RAD 40, 1Ø 100 Amp GMX 525 RAD 50, 3Ø 100 Amp GMX 625 RAD 65, 3Ø 100 Amp GMX 825 RAD 80, 3Ø 100 Amp 1.5 WIRE SIZING TOLINE TRANSFORMER Note : Line voltage variations must not exceed ± 10% from nominal voltage. Note : Line voltage drop during X-Ray exposures must not exceed 10%. 5

9 1.6 ELECTRICAL REQUIREMENTS, POWER ASSIST UNITS Power assist Generators require 115 VAC, 20 Amps, supplied from a standard wall outlet or 220VAC, 30 Amp. The Power assist Line Voltage is configured by taping its input transformer (located in the Power Assist Chassis) to the appropriate Line Voltage. See page 22, this manual. 1.7 SAFETY DEVICES Door safety switch: Prevents X-Ray exposures when the room door is open. Check local requirements. X-Ray On lamp: Installed outside the X-Ray suite, (typically above the room door). Check local requirements. Note: The GMX X-Ray Generator provides a switching function to activate the X-Ray On Lamp, the lamp voltage is supplied by the customer. Note: The GMX X-Ray Generator is not configured to supply any voltage or currents that may be required to interface other ancillary devices that may be installed with the X-Ray generator. Grounding and or load issues to the GMX circuits may result in faulty operation and function during X-Ray generation. 6

10 This Page left blank purposefully 7

11 2) INSTALLATION 2.1 INTRODUCTION This chapter describes procedures used to install the GMX X-Ray Generator. The installation process will depend on Generator type and ancillary equipment configuration: Check for transportation damage. Report shipping damage to the Freight Company or CMR X-Ray. Check that supplied components and cables match what was ordered system. NOTE: Power Assist Generator installation is covered in Appendix PA 2.2 TOOLS AND MATERIALS Required tools and materials: Hand tools. (Service Engineers Tool Kit) Terminal Crimp Tool (Wire sizes 12 through 22) Power Drill Motor, Drill Bit Index Rotary Hammer, Concrete Bits Extension cord. Soldering iron, 60 Watts. Silicone Insulating Grease DC4 Vapor Proofing Compound or Equivalent. Degreasing/Cleaning agent. Dielectric Oil. 8

12 2.3 MEASUREMENT EQUIPMENT Fluke 177 or equivalent digital multimeter. Tektronix TDS 2002 or equivalent oscilloscope. mas meter. KVP Meter Dose Meter 2.4 PRE-INSTALLATION CHECK Prior to installation, inspect the site and verify that the X-Ray room complies with Preinstallation requirements: Incoming Line, voltage and current capacity. Main Switch and Safety Devices. Conduit, size and location. Space Requirements. Note: Refer to Pre-Installation Chapter (1) 2.5 UNPACKING TWO PEOPLE ARE REQUIRED FOR MOVEMENT AND PLACEMENT OF THE HV TRANSFORMER AND POWER MODULE and/or POWER ASSIST. IT S HEAVY, DON T RISK AN INJURY! During the equipment unpacking, inspect all items for visible damage. After unpacking, verify the following: All items on the customer order are present. All hardware and internal wiring is secure. 9

13 2.6 POWER CABINET INSTALLATION / POWER ASSIST TWO PEOPLE ARE REQUIRED FOR MOVEMENT AND PLACEMENT OF THE HV TRANSFORMER AND POWER MODULE and/or POWER ASSIST. IT S HEAVY, DON T RISK AN INJURY! Power Cabinet Installation: 1. After unpacking, remove Power Module trim covers. 2. Remove the shipping screws that attach the Power Module to the pallet. Place the Power Module chassis near to its final position in the room. Two people are required for this operation. 3. Do not install HV Transformer wires, HV cables and covers until instructed to do so in this Manual. 4. Power Cabinet may be freestanding or located on a platform or shelf or bolted to the room floor. NOTE: Local requirements apply. 2.7 OPERATOR CONSOLE INSTALLATION 1. Place the Operator Console at its final location. 2. Connect provided interconnect cable, the other end will terminate at the Power Module. 10

14 2.8 INTERFACE CONNECTIONS The following drawings show the connections between the generator and peripherals, ie.: Table Bucky, Wall Bucky, Safety Lamp, Door Switch, etc GENERATOR CONNECTIONS LEFT SIDE 2TS1 Rotor AUX Voltages See pp 14 2TS2 Digital Interface 2TS3 Relay Board, Bucky Interface See pp 17 2TS4 System Connect Jumper 4 and 10 to enable Service Mode See pp 18 11

15 2.8.2 GENERATOR CONNECTIONS RIGHT SIDE 1TS1 Pre Charge Functions & ON/OFF Functions MAKE NO CHANGES F6, F7 ON/OFF 1TS2 T1 Primary Tap 1TS3 Rotor Start & Run Taps Filament Inverter Volts Tap FUSE F8, 15A Pre Charge F9, 15A T1 F10, 15A RTR Start F11, 15A RTR Run F12, 15A Filament F14, 15A Filament 12

16 2.8.3 T1 VOLTAGE TAPS, ROTOR AND FILAMENT GMX X-RAY GENERATOR 2TS1-1 1F10* A , J1-3 1F11* A , J1-1 1F12* 1F13* A , J1-8 Fluoro Only 1F14* A , J1-1 1F15* Fluoro Only Stator Common Tap for Rotor Start Voltage Tap for Rotor Run Voltage Tap for Filament Inverter Voltage * All Fuses are 15A / 250 VAC 1TS V BLK 40V BRN 60V RED 100V ORN 130V YEL 220V BLU 310V WHT 270V GRY 0V VIO T1 220V RED 208V BRN 0V BLK 1TS

17 2.8.4 X-RAY TUBE STATOR CONNECTIONS 2TS1 X-Ray Tube 1 Common WHT 2 Phase Shift RED or GRN 3 Main BLK 4 Housing Ground 5 Thermo Switch t o 6 Thermo Switch RTN Tie Point (Open) 7 Tie Point (Open) 8 AUX +24, < 500mA 9 AUX +24 Common 10 Will not support more than 500 ma Load *2C1 Phase Shift Capacitor Value as/per X-Ray Tube Stator 2C1* 0 VAC 1TS3 ** Chassis Ground added by installer ** 2TS1 X-Ray Tube/ Stator Interface Low Speed Rotor Control A J3 1 3 Closes at Power UP NO Contact: UL Rating: 5 A at 30 VDC 5 A at 125 VAC Wall Bucky Suppl Cable. Generator Descriptio 14 Table

18 Varian Color Black Red White Orange-Brown Brown-Orange Green-Yellow Description Main Auxiliary Common Thermal switch Thermal switch Ground TOSHIBA Color Black Red White Yellow Blue Green-Yellow Description Main Auxiliary Common Thermal switch Thermal switch Ground KAILLONG Color Black Red White Blue Blue Green-Yellow Description Main Auxiliary Common Thermal switch Thermal switch Ground 15

19 2.8.5 DIGITAL INTERFACE, 2TS2 2TS2 Digital Interface Used for Software based interfaces 16

20 LED1 LED2 LED3 LED4 LED5 K1 D1 K2 D2 K3 D3 K4 D4 K5 D5 R1 R2 R3 R4 R5 K1C K1B K2C K2B K3C K3B K4C K4B K5C K5B P1 P3 P4 P6 P7 P9 P10 P12 P13 P15 P5 P8 P2 P11 P CMD Return Bucky 1 Start CMD Bucky 2 Start CMD Spare 1 Start CMD Spare 2 Start CMD Room Light Room Light Return EXP Start CMD EXP Start Return Tie Point (Open) A Ancillary Interface Board J V Bucky 1 Bucky 2 Bucky 3 Bucky 4 Room Light Bucky/Tomo EXP 2TS3 TS1 Tie Point (Open) GMX X-RAY GENERATOR RELAY BOARD, BUCKY INTERFACE, 2TS3 Tie Point (Open) 17

21 2.8.7 SYSTEM CONNECT, 2TS4 127 VAC and 2TS4 2 Descriptio 1 Tie Point (Open) Tie Point (Open) Tie Point (Open) *SERVICE MODE Spare Interlock **DOOR OPEN *** X-Ray Tube Thermo Switch Tie Point (Open) GND GND *Jumper to GND for Service Mode ** Must be LOW for EXP Through Door Switch or Jumper to GND *** Must be LOW for EXP Connects to Tube at 2TS1-5 2TS4 A System Connect 18

22 2.8.8T1 POWER MODULE AUX POWER 19

23 2.8.9 T1 TAPS (ROTOR & FILAMENT) AND FUSES GMX X-RAY GENERATOR 2TS1-1 1F10* A , J1-3 1F11* A , J1-1 1F12* A , J1-8 1F13* Fluoro Only 1F14* A , J1-1 1F15* Fluoro Only Stator Common Tap for Rotor Start Voltage Tap for Rotor Run Voltage Tap for Filament Inverter Voltage * All Fuses are 15A / 250 VAC 1TS V BLK 40V BRN 60V RED 100V ORN 130V YEL 220V BLU 310V WHT 270V GRY 0V VIO T1 220V RED 208V BRN 0V BLK 1TS

24 TS-1-1 POWER ASSIST TRANSFORMER INPUT GMX X-RAY GENERATOR Generator Descriptio 21

25 3) GENERATOR CONFIGURATION The following procedure describes the initial configuration for the GMX X-Ray Generator. Generator calibration follows in chapter 4. In order to successfully complete configuration and calibration you must observe the following: Required Hardware / Software: Computer, Laptop Windows 7 or higher Service Interface Box, CMR part Number G , with USB cable. Service Software (A4000-xx), CMR CD provided. Software Driver, USB to UART bridge (silicon labs CP 2102 USB to UART); available at Silicone Labs website : (also available on the CD provided with the CMR X-Ray Generator). 3.1 CONFIGURATION PROCEDURE 1. Verify that the Mains Power breaker is Off; Generators with Power Assist (PA) verify that the PA power cord is disconnected. 3. Connect the USB cable between your laptop computer and the Service Interface Box. 4. Connect the Generator to Operator Console cable (DB-15), from the Power Module front panel to the Service Interface Box. 5. Place a jumper between 2TS4-4 and 2TS4-10 on the A board located in the power module (see page 11 and page 18 in this manual).. 22

26 Turn the Mains wall breaker On, or connect the Power Assist unit to outlet power. Turn the Generator On with the ON switch located on the Service Interface Box. Close the toggle switch on the PA unit. 7. Insert the installation CD or flash drive into your PC. 8. Open Folder A v4, in that Folder find Folder Software, copy and paste Software to C Drive Open A v4 found on the installation disk or flash drive 11. Double click setup.exe, follow instructions, this will install the generator calibration program ion c:\program Files (x86)\a On your PC, Click Windows Start 13. Click on Control Panel 14. Click on Device Manager 12. Select Ports (COM & LPT), this will indicate which COM Port connects to the X-Ray Generator. Note and remember this number. If you don t see a comport that is connected to the generator via the service box then you don t have the UART bridge driver installed. Most of the time your computer will search for this driver and find it on the supplied CD. However if that does not work you must go online to the silicone labs website, this will allow your computer to download and install the correct USB two UART driver. This operation is extremely critical to the successful and functional application of the configuration and calibration features of this generator. 23

27 12 COM Port # 13. Run program A v4, accessible fron Windows Start or Program Files (x86)\a Or create a shortcut from Program Files 24

28 15. The next screen should look like this. Select the ON button. GMX X-RAY GENERATOR 14, ON 16. You will see this window displayed for approximately 10 seconds 25

29 17. Next, the Main screen will display. Note: Status box reflecting possible problems or errors. Note: that various selections located at the bottom of this window.. Select Configuration button. 11, Configuration 26

30 18. In configuration mode, the following screen is visible. System ID and System Ratings cannot be changed. 100 Amp Line Power is default it when 1 Tube Rad is selected Rather than defining and repeating the selections on this screen, you as the installer should be able to configure and select the various functions presented. If you have trouble recognizing these choices you may want to enlist a more experienced service engineer to help (call a friend). The selection of the correct X-Ray Tube Type is extremely important. 20, Close 18, SAVE 19. Once configuration selections have been made use the Save button to move these settings into Generator memory. 27

31 20. During Configuration Save you will see this screen 21. After saving the configuration settings, Select Close,, the software will return to the Main screen. NOTE: If you are not moving to Calibration Mode, remove the Service enable jumper on 2TS4 (see page 11 in this manual). If you are proceeding to the Calibration Mode, select ma Calibration,continue as follows. 20, ma Calibration 28

32 4) CALIBRATION 4.1 TRUE-TRACK CALIBRATION WITH COMPUTER TRUE-TRACK insures that X-Ray parameters calibrate to required specifications. If you have just finished Configuration, connect your scope and start at step # 6, this chapter. If you are checking Calibration only, install the software and setup as per Chapter 3.1 (pp 22 this manual), steps The following procedure will ensure compliance. Connect scope Channel 1 to TP12 (ma test point) on the PWM A board, Set Amplitude: 500mV per division Set.Time 5ms per division. Use channel 1 trigger in normal mode. Connect scope Channel 2 to TP3 (kv test point) on the PWM A board, Set Amplitude: 1V per division. PULSE WIDTH MODULATOR A TEST POINTS TP3, kv Test Point, Chanel 1 TP12, ma Test Point, Chanel 2 TP 23, kv and ma signal common 29

33 1. Connect the USB cable between your laptop and the SERVICE BOX, connect the Generator cable to the SERVICE BOX. 2. Place a jumper between 2TS4-4 and 2TS4-10 on the A board located in the power module (see page 11 in this manual) Install the.exe file found in the A folder. This software is compatible with Windows XP and Windows Turn the Mains breaker On 5. Turn the Generator On with the ON switch located on the OPERATOR CONSOLE SIMULATOR BOX 6. From the Main screen, access Calibration mode. 7, Calibration 30

34 7. Calibration screen. Note the Status window, if any of these are highlighted you will not be able to make an exposure. Repair or reconfigure as needed to satisfy all parameters in the Status window. 16, Status Window 8, Closed Loop. True-Track NOTE: When True Track is selected the software will default to the starting technique setting, 80 kv, 10 ma, at 40 ms 8. Insure that Closed Loop and True-Track are selected. 9. To start (10 ma at 80 kv) make sure that the Filament Demand set point is at least 2200 but not greater than 2400 (for most x-ray tubes). Note: Filament Demand Limit and starting Set Point are set relative to the X-Ray Tube selected during Configuration. 31

35 10. With a Hand switch connected to the service box make an exposure, the Filament Demand number will change. 11. Continue to hold exposure switch, X-Ray exposures will automatically continue, until a green highlight annotated with Calibration successful appears 12. Release the Expose and Prep switch, Note the WAIT indication in the STATUS Box, You must wait for the indicator to change to Ready to expose before continuing with the next ma Calibration. True-Track automatically saves the data and advances to the next station. 13. Continue to initiate the exposure switch until the next station is calibrated. 14. Repeat until all ma stations are calibrated. 15. A green indication of Cal complete will present when each station is calibrated. 12, WAIT Will Change to Ready to expose 16, OFF 10, Filament Demand 11, 15, Green 16. When all ma stations are Calibrated, Select OFF. 32

36 Note: to recalibrate a specific ma station select the MA value and select 80 kva, initiate the calibration process, after the 45 kva station is calibrated you must manually select the 125 kv station and then initiate the Cal process. 4.2 TYPICAL MA AND KV WAVE FORMS 80 kv 10 ma 80 kv 100 ma 80 kv 300 ma 33

37 You have just installed the Medical Imaging industry s most cost effective High Performance High Frequency X-Ray Generator. Thanks for your wise choice. 34

38 5) TROUBLESHOOTING 5.1 INTRODUCTION The generator includes self-diagnostic routines to facilitate troubleshooting. When in the unlikely event there is an error, the Operator console displays an error code. This chapter lists the errors, with corresponding corrective actions. 5.2 SERVICE RECOMMENDATIONS CAUTION: This X-Ray Generator has components that maybe damaged by electrostatic discharges (ESD). Use ESD precautions when working with the printed circuit boards. WARNING THE HIGH VOLTAGE INVERTER CAPACITORS RETAIN A LARGE PORTION OF CHARGE FOR APPROXIMATELY 2 MINUTES AFTER THE UNIT IS TURNED OFF. Auto-DISCHARGE CIRCUITS ARE IN PLACE BUT ALWAYS CHECK FOR BUS VOLTAGE WHEN WORKING IN THIS AREA. Do not produce Radiation without cause When making exposures during service activities, be mindful of the expected path of the primary X-ray beam. Keep all parts of your body outside of this path. Exposure to X ray radiation may be harmful, care must be taken to avoid direct exposure to the primary X Ray beam. Always Collimate the x-ray beam to as small an area as possible 35

39 Do not defeat or override any safety features on the GMX X-Ray generator. Note: In the unlikely event of a blown fuse: To insure the continued protection against fire replace uses only with the same type and ratings as supplied and specified by CMR X-Ray. After service, insure that all protective panels and covers are in place. 5.3 ERROR CODES Error codes, provided for troubleshooting. There are two categories (1) recoverable and (2) non-recoverable. A recoverable error prevents exposures until the conditions generating the Fault have been repaired, does not require a Generator restart. A non-recoverable error requires the Generator to be repaired and restarted. For the Basic RAD Console, the lamp of the selected function key will flash. When that key is pressed, the kv indicator displays the error code. To check if more than one error occurred, keep the function key pressed and simultaneously press one of the kv keys. For the Touch Screen Console, Error/Fault Codes display on the Operator Counsel using text, not numbers. 36

40 5.3.1 ERROR CODE DESCRIPTIONS Non-Recoverable Errors, Requires repair and Restart E02 E07 E10 E19 E27 E36 E38 E39 E44 E46 E47 E49 E52 E53 E56 E57 E58 Battery Fault (Battery Systems Only) X-Ray Tube Thermal Switch Fault. Bucky Signal Fault Rotor Signal Fault Communication Fault (Console to Power Module) Inverter Bus Voltage Low IPM Fault Filament Fault Console Pushbuttons activated during Power UP Microprocessor Battery Fault Tube Data Fault ma Station Not Calibrated kv Unbalanced Checksum Error, Flash Memory AEC No Calibration Data High Speed Rotor Fault Microprocessor Not Initialized Recoverable Error, Re-Start Not Required E18 E29 E32 E33 E51 E51 X-Ray Room Door Open Exposure Back Up Time Selected AEC Technique greater than 650 mas Selected Technique Exceeds Available Heat Units Pre-Charge Error Exposure Hand switch released before End of Exposure 37

41 6) RENEWAL PARTS 6.1 POWER MODULE Low Voltage Distribution A Low Voltage Power Supplies ,24,25 IPM Line Fuses Filament Driver A Contactors, Line & Pre Charge Low Speed Rotor Control A HV Transformer Inverter Module (CONT. Next Page) 38

42 Digital I/O A Micro Controller A Analog I/O A Pulse Width Modulator A

43 GMX X-RAY GENERATOR 6.2LINE INVERTER Assembly Number: A , Inverter Module, 1 Phase 240 VAC A , Inverter Module, 3 Phase 240 VAC A , Inverter Module, 3 Phase 480 VAC IPM kw kw kw Snubber Caps MF MF IPM Driver A Input Filter/Charge Monitor A (240V 1P) A (240V 3P) A (480V 3P) Line Rectifiers, (2 or 3 each) (Not Visible) Capacitor, 4 each MF (Not Visible) 240 VAC Fuses (480V 3P Only) Resistor, Pre- Charge E 30 kw E 80 kw Pre-Charge Contactor Line Contactor Line Fuses

44 7) ACCEPTANCE TESTING 7.1 INTRODUCTION This chapter defines the tests used to ensure that the calibrated operation of this equipment meets performance requirements set by the FDA in 21CFR, Sub Chapter J. Make these checks after installation calibration or any maintenance that would affect Generator calibration. 7.2 PRELIMINARY TESTING These procedures confirm the operation and calibration of the generator. 1. Turn the generator ON and note the operator console startup routine. The control console indicators will display the last ma, kv, and time 2. Change the technique factors and verify they are updating relative to operator input. 3. Press the prep pushbutton and make sure the X-Ray tube rotor is turning and the Ready Lamp illuminates. 4. Release the prep pushbutton and make sure the prep lamp turns off and the X-Ray Tube rotor is no longer powered. 41

45 7.3 X-RAY TIME MEASUREMENT Test Equipment Oscilloscope. Method Connect the oscilloscope at test point TP3 (KV test point), on the PWM board. Acceptance Limit Set time Acceptance limit 1ms at 5s +/- (1%+0.1ms) Procedure 1. Make an exposure at 80kV, 100mA, using the exposure times specified in table Measure the exposure time with oscilloscope. 3. Record the measurement information in form 6 Time Measurement (chapter 5, in this manual). The measured results must be within the indicated limits, see table

46 Exposure time (s) Limits (s) Table 6.1 Time Measurements. 43

47 7.8 KV/MA X-RAY MEASUREMENT Test Equipment Non-invasive kv meter. mas meter. Method Connect the kv meter as indicated by the manufacturer. Connect the mas meter to the red and black test terminals located on the high-voltage transformer. Acceptance Limits Factor Acceptance limit kv +/- (5% of kv + 1 kv) ma +/- (5% of ma + 1 ma) Table 6.2 kv and ma Measurements. 44

48 kv/ma Measurement (continued) Procedure 1. Select 50kV, 100ms and the lowest station in the set scale. 2. Make an exposure and record the measured values in Form 7 or Form 8 (Chapter 5, in this manual). 3. Select 80kV, 100ms 4. Make an exposure and record the readings for kv/ma in form 7 or 8, as applicable. 5. Select 120kV, 100ms 6. Make an exposure and record the readings for kv/ma in Form 7 or 8, as applicable. 7. Repeat until you are finished with the ma values and record in Form 7 or 8, as applicable. Note: Observe X-Ray Tube rating limitations during this procedure 45

49 7.8 LINEARITY Test Equipment Dosemeter. Handheld Calculator. mas meter. Aluminum Filter. Method Place the X-ray tube at a focus-film distance of 100cm (40 ). Place the dosemeter in the middle of the radiation field and a distance of 50cm (20 ) from tube focal point. Adjust the collimator as per the size of the radiation area. Place an aluminum filter over the image reception to minimize secondary radiation. Connect the mas meter in the red and black terminals aon the high-voltage transformer. Remove the shorting link located between the red and black terminals. 46

50 Linearity (Continued) Specifications The linearity coefficient (CL) or difference between two radiation averages and the mas obtained for an two consecutive readings of the tube current, must not exceed 0.1 times the sum. That is: CL = ( X 1 X 2) 0.1( X 1 X 2) Where X and values. 1 X are the mr/mas averages obtained from two consecutive tube current 2 Procedure Select 80kV, 200mA, and 100ms. Perform a series of 5 exposures and record them in form 9 Linearity (chapter 5, in this manual). Select 80kV, 250mA y 100ms. Perform a series of 5 exposures and record them in form 9 Linearity (chapter 5, in this manual). Estimate the generator linearity with the data from the table and test its linear behavior. 47

51 7.8 REPEATABILITY Test Equipment Dosemeter. Calculator. Aluminum filter. Method Place the X-ray tube at focus-film distance of 100cm (40 ). Place the dosemeter in the middle of the radiation field and at a 50cm (20 ) distance from tube focal point. Adjust the collimator based on the size of the radiation area. Place an aluminum filter over the image reception to minimize secondary radiation. 48

52 Specifications The radiation coefficient of variability (CV) must not exceed 0.05 for any specific combination of technical factors. That is: CV = s mr Where: CV = Coefficient of variability. s = The simple standard deviation (σ (n-1)) or the square root of the differences and the square of the total of 5 exposures divided by 4(n-1). mr = Is the average doses for 5 exposures. Procedure Select 80kV, 100mA, and 200ms. Perform a series of 5 exposures and record the results in Form 10 Repeatability (Chapter 5, in this manual), for estimating the Repeatability. Repeatability must have a coefficient less than

53 SUPPLEMENT B POWER ASSIST The Optional CMR Power Assist allows CMR X-Ray Generators to operate from 115VAC, 20 Amp outlets. Once installed mimimal Service or Operator interaction is required. Power outages will necessitate the Operator to push the Reset Push Button when power is restored. POWER ASSIST MODULE POWER ASSIST INVERTER, LOCATED IN POWER MODULE 50

54 POWER ASSIST SERVICE SWITCH Located under the front cover of the Power Assist module. When Closed (Down Toggle) the Service Switch, will discharge the capacitor banks. Discharge may take up to 20 minutes, SWITCH, Push to reset PA Module Service Switch SWITCH, UP Position (ON), enables PA function SWITCH, UP Position (ON), enables power to Generator Power Module LAMP, Indicates a Charge on PA Bus LAMP, indicates Reset pushbutton must be activated POWER ASSIST CONTROL PANEL LAMP, Indicates PA Module is connected to power 51

55 SUPPLEMENT C USEFUL TABLES Selected ma Linear Scale Voltage at TP 12, PWM Linear ma Scale I Selected ma Renard Scale Voltage at TP 12, PWM Renard ma Scale Note: See page 11 for Test Point locations Selected KV Voltage at TP 12 PWM

56 SUPPLEMENT D HV CABLE MAINTENANCE If required, remove old insulating grease from cable ends and receptacles. Clean cable ends and receptacles with a hydrocarbon cleaning solvent. Avoid touching the cleaned areas with your hand to avoid leaving perspiration or skin oil on the insulating surfaces. Inspect cable ends and cables for evidence of arching, carbon tracking, burnt or pitted areas, and/cracked conditions. If any of these conditions cannot be cleaned it may be necessary to replace the cable or receptacle. Insure that the cable ends have a rubber sealing-ring and that the rubber ring is correctly positioned and clean. At the HV transformer, insulating oil (Shell Diala X or equivalent) may be used in the HV receptacles to provide insulation and vapor displacement. After cleaning, insert approximately ¼ inch of oil in the receptacle. Insert the cable(s) and tighten the spanner ring. At the X-Ray Tube, apply a fresh layer of insulating grease to the cable ends. Use a wooden or plastic spatula to apply the grease, avoid using fingers. On the side of the cable ends apply a 1 to 2mm layer of grease. At the cable end cover the cable pins with insulating grease. Insert the cable(s) and tighten the spanner ring. NOTE: after the X- Ray Tube has been heated recheck the tightness of the spanner ring; during the thermal cycling of the X-Ray tube the grease has melted and repositioned, causing the cable to loosen 53

57 END OF DOCUMENT 54

OPERATORS MANUAL AND INSTALLATION AND SERVICE MANUAL UNIVERSAL UNI-MATIC 325D GENERATOR MODEL NO. 3488/3550

OPERATORS MANUAL AND INSTALLATION AND SERVICE MANUAL UNIVERSAL UNI-MATIC 325D GENERATOR MODEL NO. 3488/3550 OPERATORS MANUAL AND INSTALLATION AND SERVICE MANUAL 8000-3488 UNIVERSAL UNI-MATIC 325D GENERATOR MODEL NO. 3488/3550 DEL MEDICAL IMAGING CORPORATION A Del Global Technologies Company 50B NORTH GARY AVENUE

More information

GE AMX 4+ Portable X-Ray

GE AMX 4+ Portable X-Ray GE AMX 4+ Portable X-Ray Typical Manufacturer s Picture GE Healthcare s AMX-4+ analog X-ray system provides high-performance in a compact, easy-to-maneuver package. The rotating arm and tube simplify positioning

More information

D. Gillespie Designs. SCA-35 Capacitor Board. Installation Manual. D. Gillespie Designs with EFB TM

D. Gillespie Designs. SCA-35 Capacitor Board. Installation Manual. D. Gillespie Designs  with EFB TM D. Gillespie Designs SCA-5 Capacitor Board with EFB TM Installation Manual D. Gillespie Designs www.tronola.com Thank you for choosing our SCA-5 Capacitor Board with *EFB. We feel it is the single most

More information

12V Victor 888 User Manual

12V Victor 888 User Manual The Victor speed controllers are specifically engineered for robotic applications. The high current capacity, low voltage drop, and peak surge capacity make the Victor ideal for drive systems while its

More information

CD-Pxx-00-0 Series Duct Mount CO 2 Transmitter

CD-Pxx-00-0 Series Duct Mount CO 2 Transmitter Installation Instructions CD-Pxx-00-0 Issue Date February 4, 006 CD-Pxx-00-0 Series Duct Mount CO Transmitter Application Requirements FCC Compliance Statement This equipment has been tested and found

More information

Contents 1. General instructions. 1.1 Precautions safety measures Protection mechanisms. 2. Description. 2.1 Instrument Familiarization. 2.

Contents 1. General instructions. 1.1 Precautions safety measures Protection mechanisms. 2. Description. 2.1 Instrument Familiarization. 2. Contents 1. General instructions. 1.1 Precautions safety measures... 1.2 Protection mechanisms. 2. Description. 2.1 Instrument Familiarization. 2.2 LCD Display 2.3 Keypad. 3. Function description. 3.1

More information

CMP 200 FUNCTIONAL DRAWINGS. DC Bus and Power Distribution

CMP 200 FUNCTIONAL DRAWINGS. DC Bus and Power Distribution CPI Canada Inc Functional Drawing Index CMP 00 FUNCTIONAL DRAWINGS DESCRIPTION DC Bus and Power Distribution System ON Room Interface XRay Exposure Radiographic kv Control and Feedback Filament Drive and

More information

NEO CAR AUDIO. Neo AUXiN AUX INPUT INTERFACE. Instruction Manual

NEO CAR AUDIO. Neo AUXiN AUX INPUT INTERFACE. Instruction Manual NEO CAR AUDIO Neo AUXiN AUX INPUT INTERFACE Instruction Manual IMPORTANT NOTE Neo AUXiN Dip switch positions MUST be set BEFORE any other step is taken. Otherwise, the kit will not operate properly. See

More information

MS8268 HANDHELD DIGITAL MULTIMETER OPERATOR S INSTRUCTION MANUAL

MS8268 HANDHELD DIGITAL MULTIMETER OPERATOR S INSTRUCTION MANUAL MS8268 HANDHELD DIGITAL MULTIMETER OPERATOR S INSTRUCTION MANUAL Table of Contents TITLE PAGE 1. GENERAL INSTRUCTIONS 1 1.1 Precaution safety measures 1 1.1.1 Preliminary 1 1.1.2 During use 2 1.1.3 Symbols

More information

Addendum to Instructions for Installation, Operation and Maintenance of Digitrip 3000 Protective Relays

Addendum to Instructions for Installation, Operation and Maintenance of Digitrip 3000 Protective Relays Dual-Source Power Supply Addendum to I.B. 17555 Addendum to Instructions for Installation, Operation and Maintenance of Digitrip 3000 Protective Relays Table of Contents Page 1.0 Introduction...1 2.0 General

More information

WRM-10 TM TRANSFORMER WINDING RESISTANCE METER

WRM-10 TM TRANSFORMER WINDING RESISTANCE METER WRM-10 TM TRANSFORMER WINDING RESISTANCE METER USER S MANUAL Vanguard Instruments Company, Inc. 1520 S. Hellman Ave. Ontario, California 91761, USA TEL: (909) 923-9390 FAX: (909) 923-9391 June 2009 Revision

More information

Electronically Commutated (EC) Motor Control with Solo, Select and Sync PWM Boards

Electronically Commutated (EC) Motor Control with Solo, Select and Sync PWM Boards Electronically Commutated (EC) Motor Control with Solo, Select and Sync PWM Boards The Solo, Select and Sync PWM boards provide a pulse-width modulated (PWM) signal to the EC motor to control fan speed.

More information

High Frequency SineWave Guardian TM Filter

High Frequency SineWave Guardian TM Filter High Frequency SineWave Guardian TM Filter 380V 480V TECHNICAL REFERENCE MANUAL WARNING High Voltage! Only a qualified electrician can carry out the electrical installation of this filter. Quick Reference

More information

RELIANT Basic High Frequency X-ray Generator. Technical Manual

RELIANT Basic High Frequency X-ray Generator. Technical Manual RELIANT Basic High Frequency X-ray Generator Technical Manual 2 RELIANT TECHNICAL MANUAL Table of Contents Preface 3 Warranty 3 Presentation 3 Notation 3 Certifications 5 Applicability 6 Safety 6 Environment

More information

PATRIARCH High Frequency X-ray Generator. Service Manual Models: P200, P300, P400, P500, P600, P700

PATRIARCH High Frequency X-ray Generator. Service Manual Models: P200, P300, P400, P500, P600, P700 PATRIARCH High Frequency X-ray Generator Service Manual Models: P200, P300, P400, P500, P600, P700 2 PATRIARCH SERVICE MANUAL Table of Contents Preface 3 Warranty 3 Presentation 3 Notation 3 Certifications

More information

TECHNICAL PRODUCT DATASHEET

TECHNICAL PRODUCT DATASHEET FORM-ENG-0018 REV A 06-02-03 ISO 9001 CERTIFIED Phone: (352) 629-5020 or 800-533-3569 Fax: (352)-629-2902 SUITABLE FOR OEM DISTRIBUTION ONLY TECHNICAL PRODUCT DATASHEET High Density PDM 21 Output / 10

More information

OPERATION & SERVICE MANUAL FOR FC 110 AC POWER SOURCE

OPERATION & SERVICE MANUAL FOR FC 110 AC POWER SOURCE OPERATION & SERVICE MANUAL FOR FC 100 SERIES AC POWER SOURCE FC 110 AC POWER SOURCE VERSION 1.3, April 2001. copyright reserved. DWG No. FC00001 TABLE OF CONTENTS CHAPTER 1 INTRODUCTION... 1 1.1 GENERAL...

More information

IDEAL INDUSTRIES, INC. TECHNICAL MANUAL MODEL: MODEL: Multimeter Service Information

IDEAL INDUSTRIES, INC. TECHNICAL MANUAL MODEL: MODEL: Multimeter Service Information IDEAL INDUSTRIES, INC. TECHNICAL MANUAL MODEL: 61-340 MODEL: 61-342 Multimeter Service Information The Service Information provides the following information: Precautions and safety information Specifications

More information

PATRIARCH High Frequency X-ray Generator. Technical Manual Models: P200, P300, P400, P500, P600

PATRIARCH High Frequency X-ray Generator. Technical Manual Models: P200, P300, P400, P500, P600 PATRIARCH High Frequency X-ray Generator Technical Manual Models: P200, P300, P400, P500, P600 2 PATRIARCH TECHNICAL MANUAL Table of Contents Preface 3 Warranty 3 Presentation 3 Notation 3 Certifications

More information

15B+/17B+/18B+ Calibration Manual. Digital Multimeter

15B+/17B+/18B+ Calibration Manual. Digital Multimeter 5B+/7B+/8B+ Digital Multimeter Calibration Manual April 206 206 Fluke Corporation. All rights reserved. Specifications are subject to change without notice. All product names are trademarks of their respective

More information

High Frequency Sinewave Guardian TM Filter

High Frequency Sinewave Guardian TM Filter High Frequency Sinewave Guardian TM Filter 380V 480V TECHNICAL REFERENCE MANUAL FORM: SHF-TRM-E REL. April 2015 REV. 001 2015 MTE Corporation Caution Prior to start up; confirm the drive operation mode

More information

8000 SERIES PRECISION MULTIMETER VERIFICATION AND ADJUSTMENT GUIDE

8000 SERIES PRECISION MULTIMETER VERIFICATION AND ADJUSTMENT GUIDE 8000 SERIES PRECISION MULTIMETER VERIFICATION AND ADJUSTMENT GUIDE TRANSMILLE LTD. Version 1.1 : Apr 2015 TABLE OF CONTENTS PREPARING FOR CALIBRATION... 4 INTRODUCTION... 4 CALIBRATION INTERVAL SELECTION...

More information

PATRIARCH High Frequency X-ray Generator. Stored Energy Service Manual Models: PSE-30V, PSE-30H

PATRIARCH High Frequency X-ray Generator. Stored Energy Service Manual Models: PSE-30V, PSE-30H PATRIARCH High Frequency X-ray Generator Stored Energy Service Manual Models: PSE-30V, PSE-30H 2 PATRIARCH STORED ENERGY SERVICE MANUAL Table of Contents Preface 3 Warranty 3 Presentation 3 Notation 3

More information

TOSHIBA International Corp

TOSHIBA International Corp TOSHIBA International Corp GUIDE SPECIFICATIONS THREE PHASE UNINTERRUPTIBLE POWER SYSTEM TOSHIBA 4200FA 30 kva CT Internal Battery UPS GUIDE SPECIFICATIONS 1 (30 kva CT) 1.0 SCOPE 1.1 System This specification

More information

Operation Manual January Opticom Infrared System. 700 Series Emitters Detectors Phase Selectors Discriminators Accessories

Operation Manual January Opticom Infrared System. 700 Series Emitters Detectors Phase Selectors Discriminators Accessories Operation Manual January 2009 Opticom Infrared System 700 Series Emitters Detectors Phase Selectors Discriminators Accessories Table of Contents 1. Introduction... 1-1 2. System Theory of Operation...

More information

INTELLIGEN. TECHNICAL MANUAL i-325/425/425t

INTELLIGEN. TECHNICAL MANUAL i-325/425/425t INTELLIGEN TECHNICAL MANUAL i-325/425/425t 2 INTELLIGEN TECHNICAL MANUAL Table of Contents Preface 3 Warranty 3 Presentation 3 Notation 3 Applicability 4 Safety 4 Environment 4 Radiation Warning 4 Chapter

More information

DUAL OUTPUT AC CURRENT/VOLTAGE TRANSDUCER

DUAL OUTPUT AC CURRENT/VOLTAGE TRANSDUCER OPERATOR S MANUAL DUAL OUTPUT AC CURRENT/VOLTAGE TRANSDUCER Masibus Automation & Instrumentation Pvt. Ltd. B/30, GIDC Electronics Estate, Sector-25, Gandhinagar-382044, Gujarat, India Web Site: www..com

More information

Current Loop Interface Module M 2025D

Current Loop Interface Module M 2025D CONTROLS Current Loop Interface Module M 2025D Current-to-Voltage Analog Converter used with the M 2001 Series Digital Tapchanger Control and M-2948 Tap Position Sensors to provide positive tap position

More information

OPERATOR S INSTRUCTION MANUAL M-2625 AUTO RANGING DIGITAL MULTIMETER

OPERATOR S INSTRUCTION MANUAL M-2625 AUTO RANGING DIGITAL MULTIMETER OPERATOR S INSTRUCTION MANUAL M-2625 AUTO RANGING DIGITAL MULTIMETER with Temperature Probe Copyright 2007 Elenco Electronics, Inc. Contents 1. Safety Information 3,4 2. Safety Symbols 5 3. Front Plate

More information

External RF Driver. Electrical Driver for AOTF-crystal. Instruction Manual

External RF Driver. Electrical Driver for AOTF-crystal. Instruction Manual External RF Driver Electrical Driver for AOTF-crystal Instruction Manual Table of Contents 1 General... 3 2 Compliance... 3 3 Labeling... 4 4 Interface... 5 4.1 Front Panel... 5 4.2 Back Panel... 6 4.2.1

More information

3050 Stereo Power Amplifier

3050 Stereo Power Amplifier 3050 Stereo Power Amplifier Owners Manual 10/26/2016 Boulder Amplifiers, Inc. 255 Taylor Ave. Louisville, CO 80027 (303) 449-8220 www.boulderamp.com Fault Conditions Boulderlink Appendix Remote Control

More information

PEN TYPE DIGITAL MULTIMETER OPERATION MANUAL T8211D

PEN TYPE DIGITAL MULTIMETER OPERATION MANUAL T8211D PEN TYPE DIGITAL MULTIMETER OPERATION MANUAL T8211D T8211D 1 1. SAFETY INFORMATION BE EXTREMELY CAREFUL IN THE USE OF THIS METER. Improper use of this device can result in electric shock or destroy of

More information

Commercial Microwave Technical Information. CAUTION All safety information must be followed as provided in Service Manual

Commercial Microwave Technical Information. CAUTION All safety information must be followed as provided in Service Manual Commercial Microwave Technical Information 0 V, 0 Hz Models RCS0MPA P00M RCS0A P00M RCS0PBDA P00M RCS0MPSA P00M RFSMPSA P00M RFSSWA P00M RCS0SWA P0M RCS0MPA P0M Due to possibility of personal injury or

More information

MW3105 DIGITAL CLAMP MULTIMETER

MW3105 DIGITAL CLAMP MULTIMETER MW3105 DIGITAL CLAMP MULTIMETER 2 M MW3105 A 01 INTRODUCTION 1.1 - Unpacking and inspection Upon removing your new Digital Clamp Meter from its packing, you should have the following items: 1. Digital

More information

USER'S MANUAL DMR-6700

USER'S MANUAL DMR-6700 USER'S MANUAL Multimeter True RMS DMR-6700 CIRCUIT-TEST ELECTRONICS www.circuittest.com Introduction This meter measures AC/DC Voltage, AC/DC Current, Resistance, Capacitance, Frequency (electrical & electronic),

More information

DVM645BI BENCH MULTIMETER TAFELMULTIMETER MULTIMETRE DE TABLE BANCO MULTÍMETRO TISCHMULTIMETER. User Manual. Gebruikershandleiding

DVM645BI BENCH MULTIMETER TAFELMULTIMETER MULTIMETRE DE TABLE BANCO MULTÍMETRO TISCHMULTIMETER. User Manual. Gebruikershandleiding BENCH MULTIMETER TAFELMULTIMETER MULTIMETRE DE TABLE BANCO MULTÍMETRO TISCHMULTIMETER User Manual Gebruikershandleiding Manuel d'utilisation Gebrauchsanleitung Introduction BENCH MULTIMETER This manual

More information

A53106 SERIES DC-TO-DC CONVERTER

A53106 SERIES DC-TO-DC CONVERTER INSTALLATION & MAINTENANCE A53106 SERIES DC-TO-DC CONVERTER AUGUST 2011, REVISED AUGUST 2014 DOCUMENT NO. COM-00-04-20 VERSION C.1 Siemens Industry, Inc., Rail Automation 9568 Archibald Ave., Suite 100,

More information

INSTALLATION AND MAINTENANCE MANUAL FOR GROUND MONITOR GM-250 COPYRIGHT 1983 AMERICAN MINE RESEARCH, INC.

INSTALLATION AND MAINTENANCE MANUAL FOR GROUND MONITOR GM-250 COPYRIGHT 1983 AMERICAN MINE RESEARCH, INC. INSTALLATION AND MAINTENANCE MANUAL FOR GROUND MONITOR GM-250 COPYRIGHT 1983 AMERICAN MINE RESEARCH, INC. MANUAL PART NUMBER 180-0036 ORIGINAL: 1-17-83 REVISION: B (8-26-86) NOT TO BE CHANGED WITHOUT MSHA

More information

IDEAL INDUSTRIES, INC. TECHNICAL MANUAL MODEL:

IDEAL INDUSTRIES, INC. TECHNICAL MANUAL MODEL: IDEAL INDUSTRIES, INC. TECHNICAL MANUAL MODEL: 61-352 The Service Information provides the following information: Precautions and safety information Specifications Basic maintenance (cleaning, replacing

More information

HE425 HE425. X-Ray Generator Installation, Operation & Service Manual. P/N 8000-HE425 Revision: D, November 18, 2005

HE425 HE425. X-Ray Generator Installation, Operation & Service Manual. P/N 8000-HE425 Revision: D, November 18, 2005 P/N 8000-HE425 Revision: D, November 18, 2005 HE425 HE425 X-Ray Generator Installation, Operation & Service Manual HE425 X-Ray Generator Revision D Periodic Maintenance Revision History Revision Date Description

More information

Rotating Anode X-Ray Tube Housing Assembly. General Data

Rotating Anode X-Ray Tube Housing Assembly. General Data Display Devices & Components Company TECHNICAL DATA ROTANODE TM E7252X Rotating Anode X-Ray Tube Housing Assembly! High speed rotating anode X-ray tube housing assembly for high energy radiographic and

More information

AC/DC Clamp Meter. Owner's Manual. Model No Safety Operation Maintenance Español

AC/DC Clamp Meter. Owner's Manual. Model No Safety Operation Maintenance Español Owner's Manual AC/DC Clamp Meter Model No. 82369 CAUTION: Read, understand and follow Safety Rules and Operating Instructions in this manual before using this product. Safety Operation Maintenance Español

More information

CP /240-MC4 User Manual

CP /240-MC4 User Manual CP-250-60-208/240-MC4 User Manual Chilicon Power LLC Jan 2014 1 CONTENTS Important Safety Instructions... 3 Safety Instructions... 3 CP-250 Microinverter System Introduction... 4 Inverter Label Information...

More information

AM-510 Commercial / Residential Multimeter. AM-510-EUR Digital Multimeter. Users Manual

AM-510 Commercial / Residential Multimeter. AM-510-EUR Digital Multimeter. Users Manual AM-510 Commercial / Residential Multimeter AM-510-EUR Digital Multimeter Users Manual AM-510 Commercial / Residential Multimeter AM-510-EUR Digital Multimeter English Users Manual Limited Warranty and

More information

CHAPTER 2 INSTALLATION

CHAPTER 2 INSTALLATION CPI Canada Inc. Installation 2 CHAPTER 2 INSTALLATION CONTENTS: 2.1.0 INTRODUCTION... 2-2 2.2.0 UNPACKING... 2-2 2.3.0 REMOVING THE GENERATOR COVER... 2-3 2.4.0 MAJOR COMPONENT LAYOUT... 2-3 2.5.0 EQUIPMENT

More information

LPB OMNI Broadcast Transmitter Installation Procedure Checklist

LPB OMNI Broadcast Transmitter Installation Procedure Checklist Installation Procedure LPB OMNI Broadcast Transmitter Installation Procedure Checklist LPB Communications, Inc. 960 Brook Road, Unit 5 Conshohocken, PA 19428 610-825-4100 phone 610-825-4047 fax www.lpbinc.com!

More information

OPERATION MANUAL. Model 1010 TE Survey Meter. August Health Physics Instruments 330 South Kellogg Ave, Suite D Goleta, CA 93117

OPERATION MANUAL. Model 1010 TE Survey Meter. August Health Physics Instruments 330 South Kellogg Ave, Suite D Goleta, CA 93117 OPERATION MANUAL Model 1010 TE Survey Meter August 1998 Health Physics Instruments 330 South Kellogg Ave, Suite D Goleta, CA 93117 GENERAL INFORMATION This instrument is manufactured in the United States

More information

OVA-50 Optical Variable Attenuator User s Manual

OVA-50 Optical Variable Attenuator User s Manual OVA-50 Optical Variable Attenuator User s Manual Shineway Technologies, Inc. All rights reserved. Safety Instructions The WARNING sign denotes a hazard. It calls attention to a procedure, practice, or

More information

Central Electronics Model 600L Linear Amplifier

Central Electronics Model 600L Linear Amplifier INTRODUCTION This manual has been reproduced by James Lawrence, NA5RC, a 600L owner. Text no longer applicable such as insurance claim with the carrier has been deleted. Some capitalization and grammar

More information

Tel: Fax:

Tel: Fax: Grid Tie Solar Inverter Operator's Manual SG-4000 1 IMPORTANT SAFETY INSTRUCTIONS SAVE THESE INSTRUCTIONS To reduce the risk of electrical shock and to ensure safe installation and operation of Grid Tie

More information

TECHNICAL SPECIFICATIONS for 300KV CONSTANT POTENTIAL X RAY EQUIPMENT (for industrial applications) S. No. PARTICULARS BHEL SPECIFICATIONS

TECHNICAL SPECIFICATIONS for 300KV CONSTANT POTENTIAL X RAY EQUIPMENT (for industrial applications) S. No. PARTICULARS BHEL SPECIFICATIONS TECHNICAL SPECIFICATIONS for 300KV CONSTANT POTENTIAL X RAY EQUIPMENT (for industrial applications) BIDDER s OFFER S. No. PARTICULARS BHEL SPECIFICATIONS 1.0 Area of Application The 300kV Constant potential

More information

MODEL , MODEL 310SAO, AND MODEL 310-ALT

MODEL , MODEL 310SAO, AND MODEL 310-ALT MODEL -0, MODEL SAO, AND MODEL -ALT VOLUME CALL AudioMaster MODEL -0 VOLTS: / VDC FEDERAL SIGNAL CORPORATION UNIVERSITY PARK, IL. U.S.A. WARNING: DISCONNECT POWER BEFORE REMOVING COVER TALK TM LISTEN 0A0

More information

User s Manual. MiniTec TM Series. Model MN26 (Model MN26T includes temperature probe) Mini Autoranging MultiMeter

User s Manual. MiniTec TM Series. Model MN26 (Model MN26T includes temperature probe) Mini Autoranging MultiMeter User s Manual MiniTec TM Series Model MN26 (Model MN26T includes temperature probe) Mini Autoranging MultiMeter Introduction Congratulations on your purchase of Extech s MN26 Autoranging Multimeter. This

More information

English KS-DR3005D POWER AMPLIFIER: INSTRUCTION MANUAL

English KS-DR3005D POWER AMPLIFIER: INSTRUCTION MANUAL KS-DR005D POWER AMPLIFIER: INSTRUCTION MANUAL B5E-009-00/00 [W] WARNING If the fuse blows, first make sure the wires aren t touching to cause a short circuit then replace the old fuse with one with the

More information

USER'S MANUAL DMR-4350

USER'S MANUAL DMR-4350 USER'S MANUAL DIGITAL MULTIMETER DMR-4350 CIRCUIT-TEST ELECTRONICS www.circuittest.com TABLE OF CONTENTS SAFETY Safety Information...................................... 2 Safety Symbols........................................

More information

Microprocessor Control Board Set Up Procedures (OR PLC)

Microprocessor Control Board Set Up Procedures (OR PLC) Microprocessor Control Board Set Up Procedures (OR-00 PLC) SWITCHES/PUSHBUTTONS Push Buttons at display SW Enter button SW Back button SW Down SW UP Back light on/off switch Rotary switches on main board

More information

Analog Technologies. Multimeter 15B and17b

Analog Technologies. Multimeter 15B and17b Figure 1. The Photo of Actual 17B Figure 2. The Photo of Actual 15B FEATURES Maximum Voltage between Any Terminal and Earth Ground: 1000V Temperature Coefficient: 0.1 (specified accuracy)/ (28

More information

Compact Autoranging Clamp Meters. Models (400 A AC), (400 A AC/DC)

Compact Autoranging Clamp Meters. Models (400 A AC), (400 A AC/DC) User Manual Compact Autoranging Clamp Meters with NIST-Traceable Calibration Models 20250-55 (400 A AC), 20250-56 (400 A AC/DC) THE STANDARD IN PRECISION MEASUREMENT 1065DGMAN_20250-55,-56 DS Clamp Meter

More information

Model 7000 Low Noise Differential Preamplifier

Model 7000 Low Noise Differential Preamplifier Model 7000 Low Noise Differential Preamplifier Operating Manual Service and Warranty Krohn-Hite Instruments are designed and manufactured in accordance with sound engineering practices and should give

More information

400Amp True RMS AC Clamp Meter + NCV

400Amp True RMS AC Clamp Meter + NCV User Guide 400Amp True RMS AC Clamp Meter + NCV Model MA410T Introduction Thank you for selecting the Extech MA410 Clamp Meter. This meter measures AC Current, AC/DC Voltage, Resistance, Capacitance, Frequency,

More information

Rotating Anode X-Ray Tube Housing Assembly. General Data. IEC Classification... Class I

Rotating Anode X-Ray Tube Housing Assembly. General Data. IEC Classification... Class I Technical Data TD ROTANODE E7239X E7239FX E7239GX 197 Rotating Anode X-Ray Tube Housing Assembly Rotating anode X-ray tube housing assembly for the purpose of general diagnostic X-ray procedures. Specially

More information

200Amp AC Clamp Meter + NCV Model MA250

200Amp AC Clamp Meter + NCV Model MA250 User's Guide 200Amp AC Clamp Meter + NCV Model MA250 Introduction Congratulations on your purchase of this Extech MA250 Clamp Meter. This meter measures AC Current, AC/DC Voltage, Resistance, Capacitance,

More information

Rotating Anode X-Ray Tube Housing Assembly. General Data. IEC Classification... Class I Type B

Rotating Anode X-Ray Tube Housing Assembly. General Data. IEC Classification... Class I Type B Technical Data TD ROTANODE E7239X E7239FX E7239GX 0197 Rotating Anode X-Ray Tube Housing Assembly Rotating anode X-ray tube housing assembly for the purpose of general diagnostic X-ray procedures. Specially

More information

MOXTEK. 50kV 10 Watt MAGNUM X-ray Source. X-ray Sources. Contents

MOXTEK. 50kV 10 Watt MAGNUM X-ray Source. X-ray Sources. Contents X-ray Sources 0kV 10 Watt MAGNUM X-ray Source Manual Contents MAGNUM X-ray Source Characteristics Initial Inspection and Handling Tube Setup Operating Conditions Operating Precautions and Warnings Operating

More information

ASX SERIES PACIFIC OPERATION MANUAL AC POWER SOURCE POWER SOURCE

ASX SERIES PACIFIC OPERATION MANUAL AC POWER SOURCE POWER SOURCE ASX SERIES AC POWER SOURCE OPERATION MANUAL PACIFIC POWER SOURCE ASX-SERIES OPERATION MANUAL FOR THE MODELS 115-ASX, 120-ASX 140-ASX, 160-ASX 315-ASX, 320-ASX 345-ASX, 360-ASX 390-ASX, 3120-ASX PPS PART

More information

WESTREX RA-1712 PHOTOGRAPHIC SOUND RECORD ELECTRONICS

WESTREX RA-1712 PHOTOGRAPHIC SOUND RECORD ELECTRONICS INTRODUCTION The RA-1712 solid state Record Electronics is an integrated system for recording photographic sound tracks on a Westrex photographic sound recorder. It accepts a 600Ω input signal level from

More information

CL900. True RMS 1000V 2000A 60MΩ ENGLISH. INSTRUCTION MANUAL 2000A Digital Clamp Meter. Measurement Technology

CL900. True RMS 1000V 2000A 60MΩ ENGLISH. INSTRUCTION MANUAL 2000A Digital Clamp Meter. Measurement Technology ENGLISH INSTRUCTION MANUAL 2000A Digital Clamp Meter True RMS Measurement Technology NON-CONTACT VOLTAGE TESTING INRUSH CURRENT LOW IMPEDANCE DATA HOLD RANGE HOLD AUDIBLE CONTINUITY DIODE TEST CAPACITANCE

More information

CM605. User Manual AC/DC LOW CURRENT CLAMP-ON METER ENGLISH

CM605. User Manual AC/DC LOW CURRENT CLAMP-ON METER ENGLISH AC/DC LOW CURRENT CLAMP-ON METER CM605 ENGLISH User Manual Statement of Compliance Chauvin Arnoux, Inc. d.b.a. AEMC Instruments certifies that this instrument has been calibrated using standards and instruments

More information

LL-105-R LED Dimmer & 0-10V Bridge Setup Guide

LL-105-R LED Dimmer & 0-10V Bridge Setup Guide www.i2systems.com i2systems is a registered trademark of Integrated Illumination Systems, Inc. LightLink is a trademark of Integrated Illumination Systems, Inc. 2018 Integrated Illumination Systems, Inc.

More information

MS8250D DUAL DISPLAY DIGITAL MULTIMETER User s Manual

MS8250D DUAL DISPLAY DIGITAL MULTIMETER User s Manual DUAL DISPLAY DIGITAL MULTIMETER User s Manual MS8250D 1. Safety Information Warning Use caution and follow all safety guidelines to prevent electric shock or damage to the meter. Please ready carefully

More information

AX-C Introduction. 2. Safety Information

AX-C Introduction. 2. Safety Information AX-C708 1. Introduction Read Safety Information before using the meter. ProcessMeter ( referred to as the meter )is a handheld, battery-operated tool for measuring electrical parameters. It has all the

More information

Distribution/Substation Transformer

Distribution/Substation Transformer Distribution/Substation Transformer Type VFI, Vacuum Fault Interrupter Transformer Option Functional Specification Guide Functional specification for 15 kv, 25 kv, or 35 kv vacuum fault interrupter distribution/substation

More information

INSTRUCTION MANUAL. Type SWF. Sine Wave Filters 690 Volts, 50Hz

INSTRUCTION MANUAL. Type SWF. Sine Wave Filters 690 Volts, 50Hz POWER QUALITY INSTRUCTION MANUAL Type SWF Sine Wave Filters 690 Volts, 50Hz Page 2 of 24 Contents 1. Introduction & SAFETY 2. Theory of operation 3. Typical results of the Sine Wave Filters 4. Transportation,

More information

IS7705. Installation & Operation Manual AUDIO INTEGRATION KIT. TranzIt LINK

IS7705. Installation & Operation Manual AUDIO INTEGRATION KIT. TranzIt LINK GET CONNECTED Installation & Operation Manual AUDIO INTEGRATION KIT IS7705 Note to Readers, The information contained within the following documentation is subject to change without notice. Features discussed

More information

Ambient Level Controller

Ambient Level Controller Ambient Level Controller Installation and Use Manual Issue 1, October 1999 1999 Bogen Communications, Inc. All rights reserved. 54-2028-01 9910 Model: LUALC PEC Code: 5335-621 COM Code: 408184273 Select

More information

CARSON MANUFACTURING CO., INC NORTH RURAL STREET INDIANAPOLIS, IN (888) TECHNICAL BULLETIN

CARSON MANUFACTURING CO., INC NORTH RURAL STREET INDIANAPOLIS, IN (888) TECHNICAL BULLETIN Carson MANUFACTURING COMPANY, INC. CARSON MANUFACTURING CO., INC. 5451 NORTH RURAL STREET INDIANAPOLIS, IN 46220 (888) 577 6877 www.carsonsirens.com TECHNICAL BULLETIN RADIO MAN SIREN PA VOL WAIL YELP

More information

NUCLEARENVIRONMENT VELOCITY - TO-DISPLACEMENTCONVERTER MOD~L BENTL NEVADA

NUCLEARENVIRONMENT VELOCITY - TO-DISPLACEMENTCONVERTER MOD~L BENTL NEVADA Manual No. TW8029360' NUCLEARENVIRONMENT VELOCITY - TO-DISPLACEMENTCONVERTER MOD~L 232-691 BENTL NEVADA Y @ 8ENTLY NEVADA r'\ V..WoridWider~serVice... Copyright

More information

Power Meter. Measurement Guide. for Anritsu RF and Microwave Handheld Instruments BTS Master Site Master Spectrum Master Cell Master

Power Meter. Measurement Guide. for Anritsu RF and Microwave Handheld Instruments BTS Master Site Master Spectrum Master Cell Master Measurement Guide Power Meter for Anritsu RF and Microwave Handheld Instruments BTS Master Site Master Spectrum Master Cell Master Power Meter Option 29 High Accuracy Power Meter Option 19 Inline Peak

More information

ECI 6D. Balanced Integrated Amplifier. Owner's Manual. (with a built-in DAC) ENGLISH

ECI 6D. Balanced Integrated Amplifier. Owner's Manual. (with a built-in DAC) ENGLISH ECI 6D Balanced Integrated Amplifier (with a built-in DAC) Owner's Manual EN ENGLISH Unpacking the ECI 6D Immediately upon receipt of the ECI 6D, inspect the carton for possible damage during shipment.

More information

ROTANODE TM E7252X E7252FX E7252GX Rotating Anode X-Ray Tube Housing Assembly. General Data. IEC Classification... Class I Electrical :

ROTANODE TM E7252X E7252FX E7252GX Rotating Anode X-Ray Tube Housing Assembly. General Data. IEC Classification... Class I Electrical : Technical Data TD ROTANODE TM E7252X E7252FX E7252GX Rotating Anode X-Ray Tube Housing Assembly High speed rotating anode X-ray tube housing assembly for high energy radiographic and cine-fluoroscopic

More information

Installing the IPS 4240 and the IPS 4255

Installing the IPS 4240 and the IPS 4255 CHAPTER 2 Installing the IPS 4240 and the IPS 4255 All IPS platforms allow ten concurrent CLI sessions. This chapter describes the IPS 4240 and the IPS 4255 and how to install them. It also describes the

More information

Model Hz to 10MHz Precision Phasemeter. Operating Manual

Model Hz to 10MHz Precision Phasemeter. Operating Manual Model 6610 1Hz to 10MHz Precision Phasemeter Operating Manual Service and Warranty Krohn-Hite Instruments are designed and manufactured in accordance with sound engineering practices and should give long

More information

Dawson DDM190. Digital Multimeter User s Manual

Dawson DDM190. Digital Multimeter User s Manual Dawson DDM190 Digital Multimeter User s Manual TABLE OF CONTENTS LIMITED WARRANTY AND LIMITATION OF LIABILITY... 3 Out of the Box... 3 Accessories.. Error! Bookmark not defined. Safety Information... 7

More information

B1793 AC Blower Innovation in Motion

B1793 AC Blower Innovation in Motion www.pelonistechnologies.com B1793 AC Blower Innovation in Motion 2 OVERVIEW Pelonis Technologies innovative B1793 Titan Series AC blower is uniquely engineered for rugged performance to meet the needs

More information

SPECIFICATION, CONTROLS AND ACCESSORIES

SPECIFICATION, CONTROLS AND ACCESSORIES AS440 Automatic Voltage Regulator (AVR) SPECIFICATION, CONTROLS AND ACCESSORIES English Original Instructions A043Y697 (Issue 2) Table of Contents 1. DESCRIPTION... 1 2. SPECIFICATION... 3 3. CONTROLS...

More information

I n s t a l l a t i o n M a n u a l. T E D P r o L i t e A B C. f o r. Shop for The Energy Detective products online at: Rev 3.

I n s t a l l a t i o n M a n u a l. T E D P r o L i t e A B C. f o r. Shop for The Energy Detective products online at: Rev 3. Rev 3.5 I n s t a l l a t i o n M a n u a l f o r T E D P r o H o m e T E D P r o L i t e A B C Shop for The Energy Detective products online at: 1.877.766.5412 IMPORTANT: The installation of your TED

More information

DM-45 Digital Multimeter

DM-45 Digital Multimeter INSTRUCTION MANUAL DM-45 Digital Multimeter Read and understand all of the instructions and safety information in this manual before operating or servicing this tool. Description The Greenlee DM-45 Digital

More information

Ambient Conditions Storage Conditions Installation Minimum Clearances and Air Flow...2 3

Ambient Conditions Storage Conditions Installation Minimum Clearances and Air Flow...2 3 CHAPTER INSTALLATION 2 AND WIRING Contents of this Chapter... Ambient Conditions..............................2 2 Storage Conditions...............................2 2 Installation.....................................2

More information

Nuclear Associates &

Nuclear Associates & Nuclear Associates 06-526 & 06-526-2200 RAD-CHECK PLUS Operators Manual March 2005 Manual No. 136201 Rev. 5 2003, 2005 Fluke Corporation, All rights reserved. Printed in U.S.A. All product names are trademarks

More information

Com-Trol ADV-6000 Trouble Shooting Guide Click on red text to go to that page in guide

Com-Trol ADV-6000 Trouble Shooting Guide Click on red text to go to that page in guide Com-Trol ADV-6000 Trouble Shooting Guide Click on red text to go to that page in guide Topic Introduction 1 Tool Requirements 1 Trouble Shooting Check List 1 Page(s) Lost communications to controller(s)

More information

DIODE / TRANSISTOR TESTER KIT

DIODE / TRANSISTOR TESTER KIT DIODE / TRANSISTOR TESTER KIT MODEL DT-100K Assembly and Instruction Manual Elenco Electronics, Inc. Copyright 1988 Elenco Electronics, Inc. Revised 2002 REV-K 753110 DT-100 PARTS LIST If you are a student,

More information

Nuclear Associates

Nuclear Associates Nuclear Associates 07-424 Digital Densitometer II Operators Manual March 2005 Manual No. 112111 Rev. 4 2003, 2005 Fluke Corporation, All rights reserved. Printed U.S.A. All product names are trademarks

More information

OPERATING MANUAL SERIES R B F BRUSHLESS RACK SYSTEM

OPERATING MANUAL SERIES R B F BRUSHLESS RACK SYSTEM OPERATING MANUAL SERIES R B F BRUSHLESS RACK SYSTEM Version 3.0 (European version 1.3) This is a general manual describing a series of racks receiving Servo Amplifiers having output capability suitable

More information

MM V 10A ENGLISH. INSTRUCTION MANUAL Auto-Ranging DATA HOLD AUDIBLE CONTINUITY MIN / MAX TEMPERATURE DIODE TEST CAPACITANCE

MM V 10A ENGLISH. INSTRUCTION MANUAL Auto-Ranging DATA HOLD AUDIBLE CONTINUITY MIN / MAX TEMPERATURE DIODE TEST CAPACITANCE INSTRUCTION MANUAL Auto-Ranging Digital Multimeter MM400 DATA HOLD AUDIBLE CONTINUITY MIN / MAX TEMPERATURE DIODE TEST CAPACITANCE 600V 10A 40MΩ 2 GENERAL SPECIFICATIONS Klein Tools MM400 is an auto-ranging

More information

Agilent U1253B True RMS OLED Multimeter. Quick Start Guide

Agilent U1253B True RMS OLED Multimeter. Quick Start Guide Agilent U1253B True RMS OLED Multimeter Quick Start Guide The following items are included with your multimeter: Silicone test leads 4 mm probes Alligator clips Printed Quick Start Guide Rechargeable 8.4

More information

AMM-1022 Digital Multimeter USER`S MANUAL

AMM-1022 Digital Multimeter USER`S MANUAL Digital Multimeter USER`S MANUAL www.tmatlantic.com CONTENTS 1. SAFETY INFORMATION.3 2. DESCRIPTION..6 3. SPECIFICATIONS.8 4. OPERATING INSTRUCTION..11 4.1 Voltage measurement...11 4.2 Current measurement

More information

XR12. Frequency Change Procedure IS Issue August 2007

XR12. Frequency Change Procedure IS Issue August 2007 XR12 Frequency Change Procedure IS07013 Issue 1.0... 31 August 2007 Nautel Limited 10089 Peggy's Cove Road, Hackett's Cove, NS, Canada B3Z 3J4 T.877 6 nautel (628835) or +1.902.823.2233 F.+1.902.823.3183

More information

UNITY/I TM. Installation Manual. UT3K, UT4K, UT5K and UT8K. Single-Phase Uninterruptible Power Systems

UNITY/I TM. Installation Manual. UT3K, UT4K, UT5K and UT8K. Single-Phase Uninterruptible Power Systems UNITY/I TM UT3K, UT4K, UT5K and UT8K Single-Phase Uninterruptible Power Systems Installation Manual MLS-0351C-OL Copyright 1994-1997 Best Power. All rights reserved. å IMPORTANT SAFETY INSTRUCTIONS! SAVE

More information

815-BR SERVO AMPLIFIER FOR BRUSH SERVOMOTORS

815-BR SERVO AMPLIFIER FOR BRUSH SERVOMOTORS 815-BR SERVO AMPLIFIER FOR BRUSH SERVOMOTORS USER GUIDE September 2004 Important Notice This document is subject to the following conditions and restrictions: This document contains proprietary information

More information

1507/1503. Insulation Testers. Users Manual

1507/1503. Insulation Testers. Users Manual 1507/1503 Insulation Testers 1507/1503 Insulation Testers Introduction The Fluke model 1507 and model 1503 are battery-powered insulation testers (hereafter, "the Tester"). Although this manual describes

More information