1 Trieu Viet Phuong Deputy head of Time and Frequency Laboratory (TFL) Vietnam Metrology Institute (VMI) N 0 8, Hoang Quoc Viet Road, Caugiay District, Hanoi, Vietnam
2 About Time & Frequency Laboratory (TFL) Time and Frequency Laboratory within Vietnam Metrology Institute (VMI). The Lab is a highest laboratory in the fields of Time and frequency in Vietnam testing measurement calibration scheme. + There are 7 staffs working at the TFL; + The laboratory has been accredited in accordance with ISO/IEC 17025:2005 by Vietnam Laboratory Accreditation Scheme (VILAS);
3 Main activities of Time & Frequency Laboratory (TFL) Currently, main activities of TFL are the maintenance of national time scale, UTC (VMI) and its dissemination through calibration services. Namely: + Maintenance of UTC (VMI) + Time comparison via satellites + Dissemination of standard time and frequency
4 1. Maintenance of UTC (VMI) In time and frequency laboratory of VMI, the 5 MHz frequency from Symmetricom 5071A (high-performance tube) is used as the external reference for AOG (Auxiliary Output Generator) and output of 1 PPS and 5 MHz from the AOG are defined as UTC (VMI) and Vietnam standard Frequency (VSF), respectively. UTC (VMI) is maintained with the help of highperformance 5071A Cesium clocks. It is generated by following BIPM tracking schedule for time comparisons via GPS commonview. For dissemination, it is maintained less than 100 ns of UTC. The Vietnam Standard Time is then defined as: UTC (VMI) + 7h
5 1. Maintenance of UTC (VMI) Fig 1.1 Clock comparison system traceable to BIPM
6 1. Maintenance of UTC (VMI) UTC (VMI) is maintained with the help of high-performance 5071A Cesium clocks. It is generated by following BIPM tracking schedule for time comparisons via GPS commonview. For dissemination, it is maintained less than 100 ns of UTC. The Vietnam Standard Time is then defined as: UTC (VMI) + 7h The three atomic clocks have been compared with one another by the TSC This comparison is performed daily at UTC 0 h (VST + 7 h). We send the collective data to the BIPM time section monthly. The traceability of VST and VSF based on the data of BIPM Circular-T
7 2. Time comparison via satellites There are two systems used to GPS common view time transfers at the TFL as Javad EU-80 and Septentrio PolarRx3 systems. To retain the traceability of VST and VSF, the timecomparison data using the GPS satellite are sent to BIPM every month. The reported data of (UTC UTC (VMI)) in Circular-T are checked regularly.
8 3. Dissemination of standard time and frequency 3.1 Dissemination of time in Vietnam - Standard time of UTC (VMI) is broadcasting though the radio the voice of Vietnam at 6:00, 12:00, 18:00, and 21:00 every day. - Time services though the internet: Standard time of UTC (VMI), we used the Sync Server S250i of Symmetricom (USA) to disseminate on the Website: The dissemination of time in Vietnam are shown in Fig 3.1
9 3. Dissemination of standard time and frequency 3.1 Dissemination of time in Vietnam Fig 3.1 Dissemination of time in Vietnam
10 3. Dissemination of standard time and frequency 3.2 Calibration services Officially, we have been providing frequency calibration services for universal counters, signal generators, local clocks, local frequency standards and the time interval from 0.01s, GPS receiver, Stop-watches, Tachometers In house computer programs have been written for the various types of counters and frequency measuring equipments used.
11 3. Dissemination of standard time and frequency 3.2 Calibration services Fig 3.2 Reference frequencies for calibration and measurements.
12 Salient point: + In 2008, we submitted CMC table; + On 19 July 2010 the CMC of the laboratory has been published in the KCDB of the BIPM.
13 Salient point
14 4. Other activities: + We have succeeded in the research, design, manufacture some equipments such as: a power cut/ on warrning via SMS system; tachometer; standard for verification of taximeters; + To develop calibration/ testing methods. + To develop, maintain the national measurement standards system
15 5. Future development plan - To develop remote calibration services; - To strengthen international and bilateral comparison with other NMIs in time and frequency field; - To develop dissemination of time in Vietnam through NTP server, speaking clock
16 Thank you for your attention!
REPORT OF TIME AND FREQUENCY LABORATORY (VIETNAM METROLOGY INSTITUTE)
Nguyen Bang Head of Time and Frequency Laboratory (TFL) Vietnam Metrology Institute (VMI) N 0 8, Hoang Quoc Viet Road, Caugiay District, Hanoi, Vietnam About Time & Frequency Laboratory (TFL) Time and
More informationSTATUS REPORT OF TIME AND FREQUENCY LAB. (VIETNAM METROLOGY INSTITUTE)
STATUS REPORT OF TIME AND FREQUENCY LAB. (VIETNAM METROLOGY INSTITUTE) Trieu Viet Phuong Head of Time and Frequency Laboratory, VMI Email: phuongtv@vmi.gov.vn DA NANG 11-2016 About TFL Laboratory of time
More informationMongolian Agency for Standardization and Metrology Time Frequency Lab. Unurbileg Darmaa Head, Length & Time and Frequency lab MASM
Mongolian Agency for Standardization and Metrology Time Frequency Lab Unurbileg Darmaa Head, Length & Time and Frequency lab MASM 2017-10-26 11/13/2017 Mongolian Agency for Standardization and Metrology
More informationHaruo Saito. National Institute of Information and Communications Technology
Calibration system at NICT Haruo Saito National Institute of Information and Communications Technology Organization of NICT Content Calibration system Calibration system Carried in system and remote system
More informationOutline. Introduction Laboratory Status Activities Future Plans
Laboratory Report Time and Frequency Metrology Sub Division Research Center for Metrology Indonesian Institute of Sciences RCM LIPI Metrology Enabling Developing Economies in Asia, MEDEA Project for GPS
More informationTime and Frequency Laboratory Measurement Units, Standards and Services Department (National Metrology Institute) MUSSD- Sri Lanka
Time and Frequency Laboratory Measurement Units, Standards and Services Department (National Metrology Institute) MUSSD- Sri Lanka Introduction Measurement Units, Standards and Services Department (MUSSD
More informationNPLI Report. for. Technical workshop and inter-laboratory comparison exercise for GPS time-transfer and calibration techniques under MEDEA
NPLI Report for Technical workshop and inter-laboratory comparison exercise for GPS time-transfer and calibration techniques under MEDEA Dr. V. N. Ojha, Dr. A. Agarwal, Mrs. D. Chaddha, Dr. S. Panja, Dr.
More informationCURRENT ACTIVITIES OF THE NATIONAL STANDARD TIME AND FREQUENCY LABORATORY OF THE TELECOMMUNICATION LABORATORIES, CHT TELECOM CO., LTD.
CURRENT ACTIVITIES OF THE NATIONAL STANDARD TIME AND FREQUENCY LABORATORY OF THE TELECOMMUNICATION LABORATORIES, CHT TELECOM CO., LTD., TAIWAN C. S. Liao, P. C. Chang, and S. S. Chen National Standard
More informationActivity Report of Time and Frequency Laboratory, NMIM (formally known as NML-SIRIM), Malaysia
Activity Report of Time and Frequency Laboratory, NMIM (formally known as NML-SIRIM), Malaysia Dr. Mohd. Nasir Senior Principal Metrologist Ahmad Sahar Senior Metrologist Mohd Rafiq Metrologist Mohd Izzulfitri
More informationDEC Workshop on Participation in Co-ordinated Universal Time
Report (Philippines) DEC Workshop on Participation in Co-ordinated Universal Time 6 7 November 2014 NIM, Beijing, China Sabino Paulo B. Leones, Jr. National Metrology Laboratory (NML) Industrial Technology
More informationTime and Frequency Laboratory Measurement Units, Standards. (National Metrology Institute) MUSSD-Sri Lanka
Time and Frequency Laboratory Measurement Units, Standards and Services Department (National Metrology Institute) MUSSD-Sri Lanka Introduction Measurement Units, Standards and Services Department (MUSSD
More informationTIME AND FREQUENCY ACTIVITIES AT THE CSIR NATIONAL METROLOGY LABORATORY
TIME AND FREQUENCY ACTIVITIES AT THE CSIR NATIONAL METROLOGY LABORATORY E. L. Marais and B. Theron CSIR National Metrology Laboratory PO Box 395, Pretoria, 0001, South Africa Tel: +27 12 841 3013; Fax:
More informationTraceability measurement results of accurate time and frequency in Bosnia and Herzegovina
INFOTEH-JAHORINA Vol. 11, March 2012. Traceability measurement results of accurate time and frequency in Bosnia and Herzegovina Osman Šibonjić, Vladimir Milojević, Fatima Spahić Institute of Metrology
More informationStatus Report on Time and Frequency Activities at National Physical Laboratory India
Status Report on Time and Frequency Activities at National Physical Laboratory India (TCTF 2015) Ashish Agarwal *, S. Panja. P. Arora, P. Thorat, S. De, S. Yadav, P. Kandpal, M. P. Olaniya, S S Rajput,
More informationCRITERIA FOR LABORATORY ACCREDITATION IN THE FIELD OF TIME AND FREQUENCY METROLOGY
CRITERIA FOR LABORATORY ACCREDITATION IN THE FIELD OF TIME AND FREQUENCY METROLOGY Approved By: Chief Executive Officer: Ron Josias Senior Manager: Mpho Phaloane Revised By: Specialist Technical Committee
More informationTIME & FREQUENCY SERVICES IN MALAYSIA
TIME & FREQUENCY SERVICES IN MALAYSIA Presented by : Mohd Rafiq Bin Abdul Kamal Metrologist National Metrology Laboratory SIRIM Berhad, Malaysia Time & Frequency Laboratory National Metrology Laboratory
More informationLITHUANIAN NATIONAL TIME AND FREQUENCY STANDARD
LITHUANIAN NATIONAL TIME AND FREQUENCY STANDARD Rimantas Miškinis Semiconductor Physics Institute A. Goštauto 11, Vilnius 01108, Lithuania Tel/Fax: +370 5 2620194; E-mail: miskinis@pfi.lt Abstract The
More informationTime and Frequency Research Activity in NIM
Time and Frequency Research Activity in NIM Gao Xiaoxun National Institute of Metrology Bei San Huan Dong Lu No.18 Beijing P.R.China Abstract This paper will introduce scientific research activities in
More informationStatus Report on Time and Frequency Activities at CSIR-NPL India
Status Report on Time and Frequency Activities at CSIR-NPL India (APMP -TCTF 2016) S. Panja, A. Agarwal, D. Chadha, P. Arora, P. Thorat, S. De, S. Yadav, P. Kandpal, M. P. Olaniya and V. N. Ojha (Da Nang,
More informationNov.6-7,2014 DEC Workshop on Participation in Coordinated Universal Time. Aimin Zhang National Institute of Metrology (NIM)
Nov.6-7,2014 DEC Workshop on Participation in Coordinated Universal Time Aimin Zhang National Institute of Metrology (NIM) Introduction UTC(NIM) at old campus Setup of new UTC(NIM) Algorithm of UTC(NIM)
More informationMEASUREMENT UNITS STANDARDS AND SERVICES DEPARTMENT(MUSSD) SRI LANKA PRESENTED BY: R.G.S.A. PERERA
MEASUREMENT UNITS STANDARDS AND SERVICES DEPARTMENT(MUSSD) SRI LANKA PRESENTED BY: R.G.S.A. PERERA SRI LANKA Government type: Democratic Socialist Republic Capital city: Sri Jayawardenapura Location: South
More informationINRIM TIME AND FREQUENCY LABORATORY: STATUS AND ONGOING ENHANCEMENT ACTIVITIES
INRIM TIME AND FREQUENCY LABORATORY: STATUS AND ONGOING ENHANCEMENT ACTIVITIES Valerio Pettiti, Roberto Costa, Giancarlo Cerretto, Cedric Plantard, and Franco Cordara Istituto Nazionale di Ricerca Metrologica
More informationMETAS TIME & FREQUENCY METROLOGY REPORT
METAS TIME & FREQUENCY METROLOGY REPORT Laurent-Guy Bernier METAS Federal Office of Metrology Lindenweg 50, Bern-Wabern, Switzerland, CH-3003 E-mail: laurent-guy.bernier@metas.ch, Fax: +41 31 323 3210
More informationExpert Site Visit Report
Expert Site Visit Report Time and Frequency Metrology Sub Division Research Center for Metrology Indonesian Institute of Sciences RCM LIPI Metrology Enabling Developing Economies in Asia, MEDEA Project
More informationSTABILITY AND ACCURACY OF THE REALIZATION OF TIME SCALE IN SINGAPORE
90th Annual Precise Time and Time Interval (PTTI) Meeting STABILITY AND ACCURACY OF THE REALIZATION OF TIME SCALE IN SINGAPORE Dai Zhongning, Chua Hock Ann, and Neo Hoon Singapore Productivity and Standards
More informationUpgradation and Strengthening of National Time Scale of India
Upgradation and Strengthening of National Time Scale of India (ATF 2017) Ashish Agarwal, P. Thorat, M. P. Olaniya, S. Yadav, P. Kandpal, P. Arora, S. Panja, S. De, T. Bharadwaj, N. Sharma, S. Kazim, B.
More informationTHE TIMING ACTIVITIES OF THE NATIONAL TIME AND FREQUENCY STANDARD LABORATORY OF THE TELECOMMUNICATION LABORATORIES, CHT CO. LTD.
THE TIMING ACTIVITIES OF THE NATIONAL TIME AND FREQUENCY STANDARD LABORATORY OF THE TELECOMMUNICATION LABORATORIES, CHT CO. LTD., TAIWAN P. C. Chang, J. L. Wang, H. T. Lin, S. Y. Lin, W. H. Tseng, C. C.
More informationTime and Frequency Activities at NICT, Japan
Time and Frequency Activities at NICT, Japan Yasuhiro Koyama, Kuniyasu Imamura, Tsukasa Iwama, Shin'ichi Hama, Jun Amagai, Ryuichi Ichikawa, and Mizuhiko Hosokawa National Institute of Information and
More informationReport of the TC Time and Frequency. Ramiz Hamid TC-TF Chair, TÜBİTAK UME, Turkey
Report of the TC Time and Frequency Ramiz Hamid TC-TF Chair, TÜBİTAK UME, Turkey Contents TC-TF meeting and T&F strategy EMRP Projects and future optical redefinition of the second Time scale generation
More informationEstablishing Traceability to UTC
White Paper W H I T E P A P E R Establishing Traceability to UTC "Smarter Timing Solutions" This paper will show that the NTP and PTP timestamps from EndRun Technologies Network Time Servers are traceable
More informationRECENT ACTIVITIES IN THE FIELD OF TIME AND FREQUENCY IN POLAND
RECENT ACTIVITIES IN THE FIELD OF TIME AND FREQUENCY IN POLAND Jerzy Nawrocki Astrogeodynamical Observatory, Borowiec near Poznań, and Central Office of Measures, Warsaw, Poland Abstract The work of main
More informationMEASUREMENT UNITS STANDARDS AND SERVICES DEPARTMENT(MUSSD) SRI LANKA
COUNTRY REPORT: MEASUREMENT UNITS STANDARDS AND SERVICES DEPARTMENT(MUSSD) SRI LANKA PRESENTED BY: R.G.S.ANURADHA. PERERA SRI LANKA Government type: Democratic Socialist Republic Location: South Asia,
More informationCertificate of Calibration No
Federal Department of Justice olice FDJP Federal Office of Metrology METAS Certificate of Calibration No 7-006 Object GPS rcvr type Septentrio PolaRx4TR PRO serial 005 Antenna type Aero AT-675 serial 500
More informationRapid UTC: a step forward for enhancing GNSS system times Elisa Felicitas Arias
Rapid UTC: a step forward for enhancing GNSS system times Elisa Felicitas Arias Eighth Meeting of the International Committee on Global Navigation Satellite Systems (ICG) Dubai, United Arab Emirates 9-14
More informationFederal Department of Justice and Police FDJP Federal Office of Metrology METAS. Measurement Report No
Federal epartment of Justice olice FJP Federal Office of Metrology METAS Measurement Report No 9-0009 Object GPS receiver type Septentrio PolaRxeTR serial 05 Antenna type Aero AT-775 serial 5577 Cable
More informationDesignated Institutes participating in the CIPM MRA
Designated Institutes participating in the CIPM MRA Expectations and nomination form Version 1.2 Contents 1. Introduction... 1 2. Designated Institute in Appendix A of the KCDB... 2 2.1. Designation of
More informationJoint ILAC CIPM Communication regarding the. Accreditation of Calibration and Measurement Services. of National Metrology Institutes.
Joint ILAC CIPM Communication regarding the Accreditation of Calibration and Measurement Services of National Metrology Institutes 7 March 2012 Authorship This document was prepared by the International
More informationPRECISE RECEIVER CLOCK OFFSET ESTIMATIONS ACCORDING TO EACH GLOBAL NAVIGATION SATELLITE SYSTEMS (GNSS) TIMESCALES
ARTIFICIAL SATELLITES, Vol. 52, No. 4 DOI: 10.1515/arsa-2017-0009 PRECISE RECEIVER CLOCK OFFSET ESTIMATIONS ACCORDING TO EACH GLOBAL NAVIGATION SATELLITE SYSTEMS (GNSS) TIMESCALES Thayathip Thongtan National
More information5-2 Generating and Measurement System for Japan Standard Time
5-2 Generating and Measurement System for Japan Standard Time HANADO Yuko, IMAE Michito, KURIHARA Noriyuki, HOSOKAWA Mizuhiko, AIDA Masanori, IMAMURA Kuniyasu, KOTAKE Noboru, ITO Hiroyuki, SUZUYAMA Tomonari,
More information5-3 JJY, The National Standard on Time and Frequency in Japan
5-3 JJY, The National Standard on Time and Frequency in Japan The Communications Research Laboratory (CRL) determines the national standard time and frequency in Japan, which is disseminated throughout
More informationProgramme of work and budget for Plans for Time Department
Programme of work and budget for 2013-2015 Plans for 2016-2019 Time Department Elisa Felicitas Arias 101 th Meeting of the CIPM, Session 1 BIPM, Sèvres, 8 June 2012 Programme of work 2013-2015 Continues
More informationRemote Time Calibrations via the NIST Time Measurement and Analysis Service
Remote Time Calibrations via the NIST Time Measurement and Analysis Service Michael A. Lombardi and Andrew N. Novick Abstract: The National Institute of Standards and Technology (NIST) now offers a new
More information2-2 Summary and Improvement of Japan Standard Time Generation System
2-2 Summary and Improvement of Japan Standard Time Generation System NAKAGAWA Fumimaru, HANADO Yuko, ITO Hiroyuki, KOTAKE Noboru, KUMAGAI Motohiro, IMAMURA Kuniyasu, and KOYAMA Yasuhiro Japan Standard
More informationSpecific Accreditation Criteria Calibration ISO/IEC Annex. Electrical metrology
Specific Accreditation Criteria Calibration ISO/IEC 17025 Annex Electrical metrology January 2018 Copyright National Association of Testing Authorities, Australia 2014 This publication is protected by
More informationTime Traceability for the Finance Sector Fact Sheet
Time Traceability for the Finance Sector Fact Sheet Version 1.4 14 March 2016 NPL Management Ltd is a company registered in England and Wales No. 2937881 Registered Office: NPL Management Ltd, Hampton
More informationVersion 1.1 (01/2015) EURAMET e.v. Bundesallee 100 D Braunschweig Germany. Phone:
Authorship and Imprint This document was developed by the EURAMET e.v. Authors: W. Schmid (EURAMET), M. Sega (INRIM, Italy, EURAMET TC-MC Chair), J. Drnovsek (MIRS/UL-FE/LMK, Slovenia, EURAMET Vice-Chairperson
More informationESTIMATING THE RECEIVER DELAY FOR IONOSPHERE-FREE CODE (P3) GPS TIME TRANSFER
ESTIMATING THE RECEIVER DELAY FOR IONOSPHERE-FREE CODE (P3) GPS TIME TRANSFER Victor Zhang Time and Frequency Division National Institute of Standards and Technology Boulder, CO 80305, USA E-mail: vzhang@boulder.nist.gov
More informationCCTF 2012 Report on Time & Frequency activities at National Physical Laboratory, India (NPLI)
CCTF 2012 Report on Time & Frequency activities at National Physical Laboratory, India (NPLI) Major activities of the Time & Frequency division of NPLI in the last three years have been: 1. Maintenance
More informationStatus Report on Time and Frequency Activities at NPL India
Status Report on Time and Frequency Activities at NPL India (APMP TCTF 2013) A. Sen Gupta, A. Chatterjee, A. K. Suri, A. Agarwal, S. Panja P. Arora, S. De, P. Thorat, S. Yadav, P. Kandpal, M. P. Olaniya
More informationTIME COORDINATION THROUGHOUT THE AMERICAS VIA THE SIM COMMON-VIEW GPS NETWORK
TIME COORDINATION THROUGHOUT THE AMERICAS VIA THE SIM COMMON-VIEW GPS NETWORK Michael A. Lombardi a, Andrew N. Novick a, J. Mauricio Lopez R. b, Jean-Simon Boulanger c, Raymond Pelletier c, and Carlos
More informationSTEERING UTC (AOS) AND UTC (PL) BY TA (PL)
STEERING UTC (AOS) AND UTC (PL) BY TA (PL) J. Nawrocki 1, Z. Rau 2, W. Lewandowski 3, M. Małkowski 1, M. Marszalec 2, and D. Nerkowski 2 1 Astrogeodynamical Observatory (AOS), Borowiec, Poland, nawrocki@cbk.poznan.pl
More informationBIPM TIME ACTIVITIES UPDATE
BIPM TIME ACTIVITIES UPDATE A. Harmegnies, G. Panfilo, and E. F. Arias 1 International Bureau of Weights and Measures (BIPM) Pavillon de Breteuil F-92312 Sèvres Cedex, France 1 Associated astronomer at
More informationEvaluation of timing GPS receivers for industrial applications
12th IMEKO TC1 Workshop on Technical Diagnostics June 6-7, 213, Florence, Italy Evaluation of timing GPS receivers for industrial applications Vojt ch Vigner 1, Jaroslav Rozto il 2, Blanka emusová 3 1,
More informationTraceability in Time and Frequency Metrology
Traceability in Time and Frequency Metrology Michael A. Lombardi National Institute of Standards and Technology Time and Frequency Division 325 Broadway Boulder, CO 80303 United States of America (303)
More informationTime and frequency transfer methods based on GNSS. LIANG Kun, National Institute of Metrology(NIM), China
Time and frequency transfer methods based on GNSS LIANG Kun, National Institute of Metrology(NIM), China Outline Remote time and frequency transfer GNSS time and frequency transfer methods Data and results
More informationTime Comparisons by GPS C/A, GPS P3, GPS L3 and TWSTFT at KRISS
Time Comparisons by GPS C/A, GPS, GPS L3 and at KRISS Sung Hoon Yang, Chang Bok Lee, Young Kyu Lee Division of Optical Metrology Korea Research Institute of Standards and Science Daejeon, Republic of Korea
More informationGet in Sync and Stay that Way
Get in Sync and Stay that Way CHOOSING THE RIGHT FREQUENCY FOR YOUR WIRELESS TIMEKEEPING SOLUTION Prepared by Primex Wireless 965 Wells Street Lake Geneva, WI 53147 U.S. 800-537-0464 Canada 800-330-1459
More informationILAC input to CIPM MRA Review Workshop October 2015
ILAC input to CIPM MRA Review Workshop 13-14 October 2015 BIPM Sevres 13 October 2015 By Erik Oehlenschlaeger ILAC (DANAK) ILAC History ILAC first started as a conference in 1977 (Copenhagen) with the
More informationCIPM and CCPR What are these organizations and how do they affect my testing results. Maria Nadal Photometry, Surface Color and Appearance NIST
CIPM and CCPR What are these organizations and how do they affect my testing results Maria Nadal Photometry, Surface Color and Appearance NIST CIE USA Annual Meeting October 6-7, 2014 Calibration Laboratory
More information2-5 Frequency Calibration
2-5 Frequency SAITO Haruo, IWAMA Tsukasa, TSUCHIYA Shigeru, and KOYAMA Yasuhiro The Japan Standard Time (JST) and the Coordinated Universal Time (UTC(NICT)), which are constructed by National Institute
More informationHOW TO RECEIVE UTC AND HOW TO PROVE ACCURACY
HOW TO RECEIVE UTC AND HOW TO PROVE ACCURACY Marc Weiss, Ph.D. Independent Consultant to Booz Allen Hamilton Weiss_Marc@ne.bah.com Innovation center, Washington, D.C. JANUARY 23, 2018 HOW DO YOU GET UTC
More informationBureau International des Poids et Mesures. International Recognition of NMI Calibration and Measurement Capabilities: The CIPM MRA
Bureau International des Poids et Mesures International Recognition of NMI Calibration and Measurement Capabilities: The CIPM MRA Prof. Michael Kühne International School of Physics Enrico Fermi Metrology
More informationTHE FIRST TWO-WAY TIME TRANSFER LINK BETWEEN ASIA AND EUROPE
35 th Annual Precise Time and Time Interval (PTTI) Meeting THE FIRST TWO-WAY TIME TRANSFER LINK BETWEEN ASIA AND EUROPE H. T. Lin, W. H. Tseng, S. Y. Lin, H. M. Peng, C. S. Liao Telecommunication Laboratories,
More informationRemote Frequency Calibration Service at NMIJ
Remote Frequency Calibration Service at NMIJ Case Example for APMP-WS 2014 Frequency Measurement Systems Section National Metrology Institute of Japan (NMIJ), AIST, JAPAN Content Introduction Overview
More informationCOMPARISON OF THE ONE-WAY AND COMMON- VIEW GPS MEASUREMENT TECHNIQUES USING A KNOWN FREQUENCY OFFSET*
COMPARISON OF THE ONE-WAY AND COMMON- VIEW GPS MEASUREMENT TECHNIQUES USING A KNOWN FREQUENCY OFFSET* Michael A. Lombardi and Andrew N. Novick Time and Frequency Division National Institute of Standards
More informationReport. Bilateral Comparison on Time Differences Between Two Pulses Between TÜBİTAK UME and SASO NMCC GULFMET.TF-S1
Report Bilateral Comparison on Time Differences Between Two Pulses Between TÜBİTAK UME and SASO NMCC GULFMET.TF-S1 (Rev. 1) April 21, 2017 Contents Contents... 2 1. Introduction... 3 2. Travelling Standard...
More informationEnhanced Primary Clocks and Time Transfer
Deutsche Telekom Enhanced Primary Clocks and Time Transfer Helmut Imlau ITSF 2017, November 8 th ITSF 2017: Enhanced Primary Clocks and Time Transfer, Deutsche Telekom, Helmut Imlau 1 Agenda (a) Enhanced
More informationComparison of Cesium Fountain Clocks in Europe and Asia
APMP/TCTF workshop 214,Daejeon, Korea Comparison of Cesium Fountain Clocks in Europe and Asia Aimin Zhang National Institute of Metrology(NIM) Sep.2,214 Outline Introduction Setup of PFS comparison Comparison
More informationF6052 Universal Time Synchronizer
F6052 Universal Time Synchronizer Doble Engineering Company March 2014 2013 Doble Engineering Company. All Rights Reserved 1 2013 Doble Engineering Company. All Rights Reserved History of Portable Time
More informationSchedule of Accreditation issued by United Kingdom Accreditation Service 2 Pine Trees, Chertsey Lane, Staines-upon-Thames, TW18 3HR, UK
72 Manchester Road Kearsley Bolton BL4 8NZ Contact: Mr A Leonard Tel: +44 (0) 1204 571499 Fax: +44 (0) 1204 571734 E-Mail: enquiries@airflowmeasurements.com Website: www.airflowmeasurements.com AIR VELOCITY
More informationTiming-oriented Processing of Geodetic GPS Data using a Precise Point Positioning (PPP) Approach
6 th Meeting of Representatives of Laboratories Contributing to TAI BIPM, 31 March 2004 Timing-oriented Processing of Geodetic GPS Data using a Precise Point Positioning (PPP) Approach Patrizia TAVELLA,
More informationInternationally accepted framework for metrology
Internationally accepted framework for metrology Andy Henson BIPM Working Groups The BIPM Bureau International des Poids et Measures the intergovernmental organization through which Member States act together
More informationTHE NATIONAL TIME AND FREQUENCY SERVICE OF THE RUSSIAN FEDERATION
THE NATIONAL TIME AND FREQUENCY SERVICE OF THE RUSSIAN FEDERATION V. Krutikov Gosstandard of Russia, Moscow 119991, Russia V. Kostromin and N. Koshelyaevsky Institute of Metrology for Time and Space FGUP
More informationRecommendation 16-A for Committee Decision
Recommendation 16-A for Committee Decision Information on the works related to the proposed redefinition of UTC (revision of Recommendation 16 (2012) Considering that: the navigation systems have unique
More informationLaboratory Report National Institute of Information and Communications Technology (NICT), Japan. Section 1: Laboratory Related Matters
32 nd Asia Pacific Metrology Programme General Assembly 17-18 November 2016 Da Nang, Vietnam Laboratory Report National Institute of Information and Communications Technology (NICT), Japan Section 1: Laboratory
More informationDEMETRA A Time Service Demonstrator Patrizia Tavella, INRIM Torino Italy on behalf of DEMETRA consortium
This project has received funding from the European GNSS Agency under the European Union s Horizon 2020 research and innovation programme under grant agreement No 640658. DEMETRA A Time Service Demonstrator
More informationPendulum Instruments AB Sorterargatan 26 SE VÄLLINGBY SWEDEN
à Pendulum Instruments AB Sorterargatan 26 SE-162 15 VÄLLINGBY SWEDEN Handläggare, enhet / +DQGOHGÃE\ÃGHSDUWPHQW Datum / 'DWH Beteckning / 5HIHUHQFH Sida / 3DJH Kenneth Jaldehag, Fysik och Elteknik 2000-09-04
More informationReport of the GPS calibration trip among NIMT, NMIM, and VMI in 2017
Report of the GPS calibration trip among NIMT, NMIM, and VMI in 2017 Y. Huang, T. Chiu, H. Lin and C. Liao 2017 ATF, CSIR-NPL, India November 25, 2017 Calibration for time transfer UTC(k) UTC(j) GPS Receiver
More informationRecent Calibrations of UTC(NIST) - UTC(USNO)
Recent Calibrations of UTC(NIST) - UTC(USNO) Victor Zhang 1, Thomas E. Parker 1, Russell Bumgarner 2, Jonathan Hirschauer 2, Angela McKinley 2, Stephen Mitchell 2, Ed Powers 2, Jim Skinner 2, and Demetrios
More informationMetrological and legal traceability of time signals
Metrological and legal traceability of time signals Demetrios Matsakis 1, Judah Levine 2, and Michael A. Lombardi 2 1 United States Naval Observatory, Washington, DC, USA 2 Time and Frequency Division,
More informationNMI's Role and Expertise in Synchronization Applications
NMI's Role and Expertise in Synchronization Applications Wen-Hung Tseng National Time and Frequency standard Lab, Telecommunication Laboratories, Chunghwa Telecom Co., Ltd., Taiwan APMP 2014 Time-transfer
More informationRECENT TIMING ACTIVITIES AT THE U.S. NAVAL RESEARCH LABORATORY
RECENT TIMING ACTIVITIES AT THE U.S. NAVAL RESEARCH LABORATORY Ronald Beard, Jay Oaks, Ken Senior, and Joe White U.S. Naval Research Laboratory 4555 Overlook Ave. SW, Washington DC 20375-5320, USA Abstract
More informationThe Inter-American Metrology System (SIM) Common-View GPS Comparison Network
The Inter-American Metrology System (SIM) Common-View GPS Comparison Network Michael A. Lombardi and Andrew N. Novick National Institute of Standards and Technology (NIST) * Boulder, Colorado, United States
More informationLONG-BASELINE COMPARISONS OF THE BRAZILIAN NATIONAL TIME SCALE TO UTC (NIST) USING NEAR REAL-TIME AND POSTPROCESSED SOLUTIONS
LONG-BASELINE COMPARISONS OF THE BRAZILIAN NATIONAL TIME SCALE TO UTC (NIST) USING NEAR REAL-TIME AND POSTPROCESSED SOLUTIONS Michael A. Lombardi and Victor S. Zhang Time and Frequency Division National
More informationBorderline between CIPM MRA and testing activities
Borderline between CIPM MRA and testing activities Michela Sega, EURAMET TC-MC Chair INRIM-Italy 7 th Meeting of the Focus Group on Facilitating National Metrology Infrastructure Development Zagreb, Croatia,
More informationA NEW APPROACH TO COMMON-VIEW TIME TRANSFER USING ALL-IN-VIEW MULTI-CHANNEL GPS AND GLONASS OBSERVATIONS
29th Annual Preciae Time and Time Interval (PTTI) Meeting A NEW APPROACH TO COMMONVIEW TIME TRANSFER USING ALLINVIEW MULTICHANNEL GPS AND GLONASS OBSERVATIONS J. Azoubib, G, de Jon2, J. Danahe?, W. Lewandowski
More informationCalibration Specialists Ltd
The National Technological Park, Castletroy, Limerick Calibration Laboratory Registration number: 001C is accredited by the Irish National Board (INAB) to undertake calibration as detailed in the Schedule
More informationLength section: New calibration and research services
Length section: New calibration and research services O Kruger October 2015 T026 Overview Traditional traceability chart Traceability chart with length R&D projects Overview of Various R&D projects Conclusion
More informationReport to the 20th Meeting of CCTF Research Activities on Time and Frequency National Metrology Institute of Japan (NMIJ)/AIST
Report to the 20th Meeting of CCTF Research Activities on Time and Frequency National Metrology Institute of Japan (NMIJ)/AIST The National Metrology Institute of Japan (NMIJ) is responsible for almost
More informationCALIBRATION OF THE BEV GPS RECEIVER BY USING TWSTFT
CALIBRATION OF THE BEV GPS RECEIVER BY USING TWSTFT A. Niessner 1, W. Mache 1, B. Blanzano, O. Koudelka, J. Becker 3, D. Piester 3, Z. Jiang 4, and F. Arias 4 1 Bundesamt für Eich- und Vermessungswesen,
More informationAn Experiment Study for Time Synchronization Utilizing USRP and GNU Radio
GNU Radio Conference 2017, September 11-15th, San Diego, USA An Experiment Study for Time Synchronization Utilizing USRP and GNU Radio Won Jae Yoo, Kwang Ho Choi, JoonHoo Lim, La Woo Kim, Hyoungmin So
More informationReport of Time and Frequency Activities at NICT
Report of Time and Frequency Activities at NICT National Institute of Information and Communications Technology Koganei, Tokyo, Japan 1. Introduction At National Institute of Information and Communications
More informationMeasurements and Distribution: PNT and Other Applications
NIST Time and Frequency Metrology Precision Time and Frequency Measurements and Distribution: PNT and Other Applications Tom O Brian Chief, NIST Time and Frequency Division and NIST Quantum Physics Division
More informationActivity report from NICT
Activity report from NICT APMP 2013 / TCTF meeting 25-26 November, 2013 National Institute of Information and Communications Technology (NICT) Japan 1 1 Activities of our laboratory Atomic Frequency Standards
More informationTime and Frequency Activities at NICT, Japan
Time and Frequency Activities at NICT, Japan Yasuhiro Koyama, Kuniyasu Imamura, Tsukasa Iwama, Shin'ichi Hama, Jun Amagai, Ryuichi Ichikawa, Yuko Hanado, and Mizuhiko Hosokawa National Institute of Information
More informationTime and Frequency Activities at KRISS
Time and Frequency Activities at KRISS Dai-Hyuk Yu Center for Time and Frequency Metrology, Division of Physical Metrology Korea Research Institute of Standards and Science (KRISS) dhyu@kriss.re.kr Time
More informationANALYSIS OF ONE YEAR OF ZERO-BASELINE GPS COMMON-VIEW TIME TRANSFER AND DIRECT MEASUREMENT USING TWO CO-LOCATED CLOCKS
ANALYSIS OF ONE YEAR OF ZERO-BASELINE GPS COMMON-VIEW TIME TRANSFER AND DIRECT MEASUREMENT USING TWO CO-LOCATED CLOCKS Gerrit de Jong and Erik Kroon NMi Van Swinden Laboratorium P.O. Box 654, 2600 AR Delft,
More informationActivity report from NICT
Activity report from NICT APMP 2017 / TCTF meeting 27-28 November, 2017 National Institute of Information and Communications Technology (NICT) Japan 1 1 Space time standards laboratory Atomic Frequency
More informationRadiocommunication Study Group 7 DRAFT REVISION OF RECOMMENDATION ITU-R TF Standard-frequency and time-signal emissions
Radiocommunication Assembly (RA-12) Geneva, 16-20 January 2012 Subject: Question ITU-R 236/7 Document 7/1005-E 20 October 2011 Radiocommunication Study Group 7 DRAFT REVISION OF RECOMMENDATION ITU-R TF.460-6
More informationSchedule of Accreditation issued by United Kingdom Accreditation Service 2 Pine Trees, Chertsey Lane, Staines-upon-Thames, TW18 3HR, UK
DC RESISTANCE Unit 4 Riverside Business Centre Walnut Tree Close Guildford Surrey GU1 4UG Contact: Ian Norman Tel: +44 (0)1483 302700 Fax: +44 (0)1483 300562 E-Mail: inorman@caltest.co.uk Website: www.caltest.co.uk
More information