ATOMIC FORCE MICROSCOPY

Size: px
Start display at page:

Download "ATOMIC FORCE MICROSCOPY"

Transcription

1 B47 Physikalisches Praktikum für Fortgeschrittene Supervision: Prof. Dr. Sabine Maier ATOMIC FORCE MICROSCOPY Version: E1.4 first edit: 15/09/2015 last edit: 05/10/2018

2 2 CONTENT 1. Introduction Preparation The easyscan2 AFM Experiment Changing the Cantilever Measurements in Dynamic Mode Scan-parameter settings and tip-approach Measurement of the calibration grid: optimization of the scan parameter XY-Scanner calibration Calibration of the free vibration amplitude A mplitude-distance and resonance curves Contact mode measurements Exchanging the cantilever Measurement of an optical storage disc Measurement of Collagen Data Analysis Dynamic Mode Contact Mode References... 21

3 Experiment B47: Atomic Force Microscopy 3 1. INTRODUCTION The atomic force microscope (AFM) is a crucial analysis instrument in modern nanotechnology. Surface properties like topography, elasticity or adhesion can be determined by AFM measurements on microscopic scales. The AFM was invented in 1986 by Binning and Rohrer (IBM research lab Zürich and Stanford University). Like the scanning tunnelling microscope (STM) the AFM belongs to the family of the scanning probe methods. All members of this family have the same basic working principle: A microscopic probe scans over a surface and interacts with it. By recording a measurement parameter, which is related to the interaction between tip and surface, a local map of the surface properties is created. In the case of the AFM, the probe is a tip with a length of some micrometers and a tip radius of about 10 nm. The tip is mounted at a µm long cantilever that is used as force gauge (Fig. 1). Figure 1: Scanning electron microscope image of a rectangular AFM cantilever with the dimensions l = 445µm, w = 43µm, t = 4.5µm, h =14.75 µm (figure adapted from ref. [1]) Usually the whole system (cantilever plus tip) is produced as one part by etching it from a silicon piece. In Contact-Mode AFM the tip is approached to the surface until the (repulsive) interaction forces lead to a deflection of the cantilever, which can be detected and used as signal. To image the surface, the deflection of the cantilever is kept constant by a feedback control. In Dynamic-Mode AFM, the cantilever is oscillating and the change in the amplitude or the frequency of the oscillation is measured. The dynamic mode can be subdivided into different classes, especially into tapping-mode and non-contact AFM. In this experiment you will use contact mode and tapping mode to characterize different samples.

4 4 Preparation 2. PREPARATION To carry out this experiment successfully, an understanding of the following topics is necessary: The working principle and setup of an AFM Measuring modes of an AFM, especially the dynamic vs. static mode Properties of a damped harmonic oscillator under influence of an external force Forces between tip and sample You should be able to answer the following questions: 1. How does an AFM work? 2. What are the important measuring modes and their advantages and disadvantages? 3. How does the feedback control system (PI controller) work while imaging with an AFM? 4. Which forces exist between tip and sample? Which ones are short-ranged and which long-ranged? 5. How can the corresponding potential be approximated? How does the force look like? In which ranges of the potential are the different measurement modes settled? 6. What is the typical appearance of a force-distance-curve? What are the differences when measuring it under ambient conditions, in water, and in vacuum? 7. When do instabilities ("jump in", "jump out") take place in a force distance curve and how can they be explained? 8. In which order of magnitude are the forces measured by an AFM? 9. How does the resonance peak of a one-dimensional harmonic oscillator change under the influence of a constant external force or a constant external force gradient? The associated equation of motion is:.

5 Experiment B47: Atomic Force Microscopy 5 Literature for preparation (all literature chapters are also available in printed form from the supervisors on request): [1] B. Bhushan, Springer Handbook of Nanotechnology, Springer, Berlin 2007 As E-Book: Chapter 22: Principle of Operation, Instrumentation, and Probes Chapter 27: Dynamic modes of AFM [2] Nanosurf easyscan2 AFM Operating Instructions (see below, StudOn) [3] E. Meyer, H. Hug und R. Bennewitz, Scanning Probe Microscopy: The Lab on a tip, Springer Verlag [4] Further material for the experiment including software and manuals: On StudOn ( Online-Angebote» 4. Nat» 4.5 Physik» Physik der Kondensierten Materie» Professur für Experimentalphysik (Rastersondenmikroskopie)» Rasterkraftmikroskopie Password: B47AFM

6 6 The easyscan2 AFM 3. THE EASYSCAN2 AFM The easyscan2 AFM from Nanosurf is used in this experiment (Figure 2). It is easy to handle and manages many different measurement modes. In contrast to most other AFMs the easyscan2 uses an electromagnetic scanner instead of a piezoelectric one. The advantages of this system are a high linearity and the fact, that a high voltage supply is not necessary. (a) (b) Figure 2: (a) Overview of the easyscan2 (b) Top and bottom view of the scan head (Figures adapted from [2]). The force detection of the easyscan2 is based on a deflection sensor (see Figure 3). A laser spot is focused on the front part of the cantilever and its reflection is centered on a two-segment photodiode. The monitored signal is the difference of both photo currents. Before scanning, the system is normally adjusted to zero. Due to the bending of the cantilever during the scan, the reflection of the laser spot moves vertically on the diode and changes the difference between the photo currents. The sign of the difference is related to the direction of the bending and the value to the magnitude. (a) (b) Figure 3: (a) Schematic of an AFM with deflection sensor. (b) Assembly of the easyscan2 scan head (Figure adapted from [2]).

7 Experiment B47: Atomic Force Microscopy 7 Attention: The laser of the easyscan2 is a class 2M laser: DO NOT STARE INTO THE BEAM OR VIEW DIRECTLY WITH OPTICAL INSTRUMENTS. The easyscan2 scan head has to be placed during the entire experiment either on the scan table or the storage plate.

8 8 Experiment 4. EXPERIMENT 4.1 CHANGING THE CANTILEVER The first measurement is performed in the dynamic mode with a Budget Sensor Tap190Al-G cantilever (f R = 190 khz; k = 48 N/m). Exchange the cantilever ONLY under the supervision of the supervisor. Never touch the cantilever with bare hands. It can be easily damaged or destroyed. Always use tweezers when handling the cantilever. Always mount the Dropstop (see Figure 4a) before exchanging the cantilever. Otherwise the cantilever could drop into the scan head and cause damage. In addition the Dropstop shields the laser. The price of one cantilever is 40 Euro. How to exchange the cantilever: 1. Cantilever exchange: Remove the old cantilever (Figure 4) Switch off the scan electronic Turn the scan head upside down and mount the Dropstop (Figure 4a). Use the Push Rod to push down the cantilever fixing spring (Figure 4b). Remove the cantilever with tweezers and put it into the storage box (Figure 4c). Figure 4:: (a) Mounting the Dropstop. (b) Mounting the Push Rod. (c) Replacing the cantilever (Figure adapted from [2]).

9 Experiment B47: Atomic Force Microscopy 9 2. Inserting a new cantilever Use tweezers to take a new cantilever from the storage box. Place the cantilever carefully on the alignment chip (Figure 5). Figure 5:: Left: Alignment chip on the microscope. Right: Bottom side of the cantilever chip (Figure adapted from [2]). Carefully move the cantilever-chip into the correct position by slightly tapping the cantilever chip on its top side. The position is correct, when the cantilever chip symmetrically fits to the alignment chip (Figure 6, left). Figure 6:: Left: Correct cantilever position. The cantilever chip and alignment-chip form small triangles in two corners which should be symmetric. The light reflection continuous. Middle/right: wrong cantilever positions. Remove carefully the Push Rod. If the cantilever chip still moves, it is not inserted correctly. Remove the Dropstop. Place the scan-head on the scan table. Connect all cables with the scan-head and activate the electronics. IMPORTANT: Always activate the electronic before starting the software since otherwise a proper communication between computer and electronics cannot be assured. In the case, there is no connection between computer and electronics, the software says "Simulation" in the bottom status bar, and both electronics and computer have to be restarted.

10 10 Experiment 4.2 MEASUREMENTS IN DYNAMIC MODE Scan-parameter settings and tip-approach Use the video-option in side view. Figure 1: Video image of the cantilever. Place the calibration grid on the scan table by using tweezers (NEVER touch the surface of the sample). Align the edges parallel to the scan table. Use the following parameter settings: Figure 2: Parameter settings for dynamic mode measurements. Check the "SPM Parameters" by clicking "More " and choosing the "Imaging" tab:

11 Experiment B47: Atomic Force Microscopy 11 Approach the cantilever to the surface by using the following procedure: 1. Use the adjustment screws to approach the cantilever to a cantilever-surface distance of 1-2 mm. Make sure the scan head stays parallel to the scan table. 2. Use "Advance" to further reduce the distance between cantilever and sample. Use the shadow of the cantilever on the surface as orientation: A gap between shadow and cantilever should still be visible. 3. Now activate the auto approach function of the microscope by clicking "Approach". 4. After short time a pop up window indicates that the approach was successful: 5. Check the LEDs at the control unit: The "probe-status-led" should be green. In case of a red LED the approach failed. Then retract the cantilever, check the placing of the cantilever on the alignment-chip and try it again. Important: Touching the scan-head or scan-table while the cantilever is engaged to the sample can distort the measurement, and most likely damages cantilever and sample.

12 12 Experiment Measurement of the calibration grid: optimization of the scan parameter Plane correction: The surface of the sample (i.e. measurement plane) and the plane of the xyscanner are ideally parallel to each other. In practical application this is rarely the case. A mismatch between both planes reduces the efficiency of the z-controller, and thereby also reduces the ability to resolve small details of the sample. A slope correction is able to correct this mismatch: Slope correction procedure: Start a scan and take a measurement without slope correction. Measure the angle between the horizontal and a scanned line in x-direction during scanning. Use the angle measurement tool from the Analysis-panel and apply it to a line scan in the cross-section window. Correct the slope in x-direction by inserting the measured angle in the "Slope X" field in the "Imaging" tab of the SPM parameters (click "More...", choose the "Imaging" tab). Rotate the scan angle by 90 to repeat the same procedure for the y-direction. Tasks: Measure the calibration grid before the slope correction. Do a slope correction and measure the calibration grid again. What is the difference between the "line-fit-filter" and the slope correction? Why are there stripes when the "line-fit-filter" is activated? Change the scan parameter (time per line, free vibration amplitude, P-/Igain) in a controlled way and find optimal settings. Describe your observations for too high and too low settings of the controller gain. Save one image with optimal parameters.

13 Experiment B47: Atomic Force Microscopy XY-Scanner calibration The scanner moves in x- and y-direction by applying a voltage to the scanner. In order to connect a distance to the applied voltage, the scanner has to be calibrated by using a calibration grid with known dimensions (Hint: A parallel alignment between scan direction and grid edges simplifies the calibration of the xy-scanner. If this is not the case, Retract, change the orientation of the grid and approach again). The xy-scanner is calibrated by using the following procedure: Use the "Measure Length" tool from the "Analysis" panel to determine the size of the calibration grid in your last measurement. Calculate the calibration factors for x- and y-direction by dividing the reference size of the grid by the measured size of the grid. Open the "Scan head calibration editor" ("Settings" -> "Calibration" -> "Edit") Insert the factors for x- and y-direction in the corresponding fields by clicking "Set". Confirm by clicking "Set". Close the "Scan head calibration editor" again by clicking "OK". The scanner should be calibrated now. Measure the calibration grid again. Do not forget to save the measurement. Check the size of the measured grid by using the "Measure Length" tool. Task: Calibrate the xy-scanner using the procedure described above. Write your calibration factors down?

14 14 Experiment Calibration of the free vibration amplitude The free vibration amplitude (deflection of the free oscillating cantilever) is by default given as a voltage. In order to determine its value in nanometer, a conversion factor needs to be derived from an amplitudedistance curve. The amplitude calibration: After each successful approach the easyscan2 software uses the current z-position as reference and sets it to zero. A negative value means a z-position above the sample surface and a positive value means a z-position below the surface. Always begin with "Start values" well above the surface (i.e. negative ones) and then decrease the "Start value" stepwise. Avoid pressing too strongly on the surface. Change to the spectroscopy mode by clicking "Spectroscopy" in the lower left panel. Use the following parameters: Start value -200 nm and End value -50 nm. Check that the "Amplitude - Spec forward"-option is activated. Measure amplitude-distance-curves by increasing the "Start value" and "End value" step by step (for example in 10 nm steps, and keep the range in constant ~150nm) until you observe a change in the amplitude (i.e. a change of the slope of the amplitude-distance-curve). Always confirm each change with the enter key, otherwise the change is not applied to the measurement. If the mean value of the amplitude is smaller than 50 mv, do not further increase the "End value, but ask your supervisor for help. Save one measurement and retract the cantilever from the sample by clicking "Withdraw".

15 Experiment B47: Atomic Force Microscopy 15 To calibrate the free vibration amplitude, measure the slope (= sensitivity) of the curve with the "Measure Length" tool from the "Analysis" panel. Open the "Scan head calibration editor" (Settings -> Calibration -> Edit) and go to the "I/O Signals" tab. Calculate from the measured slope the value for the deflection, which would be expected for an amplitude of 10 V and insert it in the corresponding field. Tasks: Determine the calibration factor for the vibration amplitude in Volt per Nanometer. How large is free amplitude of 200 mv (as used in the previous measurements) in nm?

16 16 Experiment Amplitude-distance and resonance curves Acquisition of resonance curves: Approach the cantilever again to the surface. When pressing the approach button, a frequency sweep is done automatically by the software. Use the "Freq. Sweep" button in the "Acquisition" panel to display the resonance curve. Click "Capture" in order to save the resonance curve. How large is the excitation amplitude, which is needed to have a free vibration amplitude of 200 nm? Measure two amplitude-distance curves, one at a frequency above the resonance frequency and one with the frequency below the resonance frequency. You can change the excitation frequency in the "Vibration frequency search dialog" by shifting the marker in the "Amplitude-Frequency sweep" window (see Figure 9). Select the new frequency such, that the resulting amplitude is approximately the same as before. Figure 9: Resonance curve with chosen excitation frequency above (left) and below (right) the resonance frequency. The amplitude is in both cases approximately the same (blue horizontal line). Tasks: Examine the dependence of the amplitude-distance curves on the chosen excitation frequency. Measure therefore two amplitude-distance-curves, one with a frequency above the resonance frequency and one with the frequency below the resonance frequency (use the procedure explained above). Save at least one resonance curve to estimate the Q-factor in your report. On which variables does the Q- factor depend on? Which excitation amplitude of the cantilever is necessary to realize a free vibration amplitude of 200 mv?

17 Experiment B47: Atomic Force Microscopy 17

18 18 Experiment 4.3. CONTACT MODE MEASUREMENTS Exchanging the cantilever Exchange the cantilever in presence of the supervisor. Insert a contact mode cantilever (Cont-Al-G, f R = 25 khz, k = 0.2 N/m). Change the following parameters in the software: Figure 10: "Acquisition" und "Z-Controller" parameter for contact mode AFM Measurement of an optical storage disc We provide 3 samples of different optical storage discs (CD, DVD, BluRay). You can choose one of the three to determine which kind of optical storage disc it is. Choose a sample and put it into the microscope. Approach the cantilever carefully to the sample surface. IMPORTANT: The shadow of the cantilever might not be visible as clearly as on the calibration grid before. Tasks: Search a nice area to measure the structure of the optical disc. First perform an overview scan and apply a slope correction. Zoom stepwise into a good spot to determine the minimal bit size and the track pitch in your report (do not forget to save your measurements). Optimize the feedback parameters to get a good image. The optimal parameters vary strongly for each sample, cantilever,... Which kind of optical storage disc do you have? Conduct a force-distance curve with the same procedure as explained in chapter Save at least one successful force-distance curve and discuss it in your report. Are you able to calibrate the amplitude from mv to nanometer from this curve (analogous to chapter 4.2.4)?

19 Experiment B47: Atomic Force Microscopy Measurement of Collagen Insert the collagen sample into the easyscan2 AFM. Carefully approach the cantilever to the sample surface. You can switch the camera view in the videooption to identify a good place to approach on a collagen fibril beforehand. Tasks: Measure the collagen sample analogous to the measurement on the optical disc: Perform an overview scan and zoom stepwise to a reasonable spot until you can see the typical band pattern (see Figure 11 as an example). You may need to try different fibres to find a suitable one. Why is the band pattern of the fibres often better visible in the deflection signal than in the topography signal? Bonus: Conduct one force-distance curve measurement on the fibril and one on the silicon substrate. Figure 11: Topography and deflection signal of a Collagen sample. Check that you have really done all mentioned tasks and that you have saved all your files. For further analyzing your data at home you will need the easyscan2 software. A copy of the software is available on

20 20 Data Analysis StudOn, or ask your supervisor for a copy. Further analysis software can be downloaded without charge from the following homepages: GWYDDION: (looks similar to GIMP, easy to handle) WSxM: (more complex, more functions, usage of the latest version, not the "stable" one, is recommended) 5. DATA ANALYSIS Write a report which includes a description of the tasks from the experimental part. Use your measured AFMtopographies and curves for the documentation. In addition, answer the following questions in your report, and provide a rough determination of the measurement errors DYNAMIC MODE a) Describe what you see on the measurement of the calibration grid. What is the reason for the lines appearing while the "line-fit-filter" is activated? How can these stripes be removed? Describe your observations for too low and too high I- and P-Gain. Figure 12: Calibration grid with (left) and without (right) "line-fit-filter" (Figure adapted from [2]) b) Indicate your calibration factors for the xy-scanner (plus error). How large is the residual xy-error after the calibration? c) Indicate the calibration factor of the amplitude in mv/nm (plus error). What kind of sample do you need to get a meaningful calibration with this method? d) Provide an interpretation of the amplitude-distance curves above and below the resonance frequency: Compare their shape with the theoretical expected shape (as shown in the literature). What is the

21 Experiment B47: Atomic Force Microscopy 21 difference for an excitation below and above the resonance frequency of the cantilever? Estimate the size of the attractive interaction regime between tip and sample from the amplitude-distance curves. e) Determine the Q-factor of the resonance curve CONTACT MODE a) Discuss the shape of the force-distance curve. Explain the hysteresis between approaching and retracting the probe to/from the sample. Determine a suitable conversion factor for the force from the forcedistance curves. Calculate the maximal repulsive force and the maximal adhesion force (error analysis). b) Briefly describe the mean differences between CD, DVD, and BlueRay discs. Determine with the "Measure-Length" tool the minimal bit size and the pitch track. What kind of optical storage disc have you measured? c) Determine the characteristic periodicity of the collagen fibrils from your measurements. Use the crosssection tool to get a cross section across the collagen fibrils. d) Bonus: Force-distance curves: Which is softer, the Si substrate or the collagen fibrils? 6. REFERENCES [1] B. Bhushan, Springer Handbook of Nanotechnology (Springer, 2010). [2] Nanosurf, Nanosurf easyscan 2 AFM Operating Instructions 2011).

ATOMIC FORCE MICROSCOPY

ATOMIC FORCE MICROSCOPY B47 PhysikalischesPraktikumfür Fortgeschrittene Supervision: Prof.Dr.SabineMaier sabine.maier@physik.unierlangen.de ATOMICFORCEMICROSCOPY Version:E1.3 firstedit:15/09/2015 lastedit:16/12/2016 2 0BIntroduction

More information

UNIVERSITY OF WATERLOO Physics 360/460 Experiment #2 ATOMIC FORCE MICROSCOPY

UNIVERSITY OF WATERLOO Physics 360/460 Experiment #2 ATOMIC FORCE MICROSCOPY UNIVERSITY OF WATERLOO Physics 360/460 Experiment #2 ATOMIC FORCE MICROSCOPY References: http://virlab.virginia.edu/vl/home.htm (University of Virginia virtual lab. Click on the AFM link) An atomic force

More information

Standard Operating Procedure

Standard Operating Procedure Standard Operating Procedure Nanosurf Atomic Force Microscopy Operation Facility NCCRD Nanotechnology Center for Collaborative Research and Development Department of Chemistry and Engineering Physics The

More information

Basic methods in imaging of micro and nano structures with atomic force microscopy (AFM)

Basic methods in imaging of micro and nano structures with atomic force microscopy (AFM) Basic methods in imaging of micro and nano P2538000 AFM Theory The basic principle of AFM is very simple. The AFM detects the force interaction between a sample and a very tiny tip (

More information

Atomic Force Microscopy (Bruker MultiMode Nanoscope IIIA)

Atomic Force Microscopy (Bruker MultiMode Nanoscope IIIA) Atomic Force Microscopy (Bruker MultiMode Nanoscope IIIA) This operating procedure intends to provide guidance for general measurements with the AFM. For more advanced measurements or measurements with

More information

Outline: Introduction: What is SPM, history STM AFM Image treatment Advanced SPM techniques Applications in semiconductor research and industry

Outline: Introduction: What is SPM, history STM AFM Image treatment Advanced SPM techniques Applications in semiconductor research and industry 1 Outline: Introduction: What is SPM, history STM AFM Image treatment Advanced SPM techniques Applications in semiconductor research and industry 2 Back to our solutions: The main problem: How to get nm

More information

Investigate in magnetic micro and nano structures by Magnetic Force Microscopy (MFM)

Investigate in magnetic micro and nano structures by Magnetic Force Microscopy (MFM) Investigate in magnetic micro and nano 5.3.85- Related Topics Magnetic Forces, Magnetic Force Microscopy (MFM), phase contrast imaging, vibration amplitude, resonance shift, force Principle Caution! -

More information

Prepare Sample 3.1. Place Sample in Stage. Replace Probe (optional) Align Laser 3.2. Probe Approach 3.3. Optimize Feedback 3.4. Scan Sample 3.

Prepare Sample 3.1. Place Sample in Stage. Replace Probe (optional) Align Laser 3.2. Probe Approach 3.3. Optimize Feedback 3.4. Scan Sample 3. CHAPTER 3 Measuring AFM Images Learning to operate an AFM well enough to get an image usually takes a few hours of instruction and practice. It takes 5 to 10 minutes to measure an image if the sample is

More information

Akiyama-Probe (A-Probe) guide

Akiyama-Probe (A-Probe) guide Akiyama-Probe (A-Probe) guide This guide presents: what is Akiyama-Probe, how it works, and what you can do Dynamic mode AFM Version: 2.0 Introduction NANOSENSORS Akiyama-Probe (A-Probe) is a self-sensing

More information

Akiyama-Probe (A-Probe) guide

Akiyama-Probe (A-Probe) guide Akiyama-Probe (A-Probe) guide This guide presents: what is Akiyama-Probe, how it works, and its performance. Akiyama-Probe is a patented technology. Version: 2009-03-23 Introduction NANOSENSORS Akiyama-Probe

More information

Instructions for easyscan Atomic Force Microscope

Instructions for easyscan Atomic Force Microscope UVA's Hands-on Introduction to Nanoscience Instructions for easyscan Atomic Force Microscope (revision 8 November 2012) NOTE: Instructions assume software is pre-configured per "UVA Instructor Guide for

More information

Electric polarization properties of single bacteria measured with electrostatic force microscopy

Electric polarization properties of single bacteria measured with electrostatic force microscopy Electric polarization properties of single bacteria measured with electrostatic force microscopy Theoretical and practical studies of Dielectric constant of single bacteria and smaller elements Daniel

More information

ezafm OPERATING MANUAL

ezafm OPERATING MANUAL ezafm OPERATING MANUAL 2013rev 2.0 1 Table of Contents CHAPTER 1:ezAFM... 4 1.1. Introduction... 5 1.2. System Components... 5 1.3. Unpacking and Packing the Instrument... 6 1.3.1. Before Installation...

More information

Measurement of Microscopic Three-dimensional Profiles with High Accuracy and Simple Operation

Measurement of Microscopic Three-dimensional Profiles with High Accuracy and Simple Operation 238 Hitachi Review Vol. 65 (2016), No. 7 Featured Articles Measurement of Microscopic Three-dimensional Profiles with High Accuracy and Simple Operation AFM5500M Scanning Probe Microscope Satoshi Hasumura

More information

Unit-25 Scanning Tunneling Microscope (STM)

Unit-25 Scanning Tunneling Microscope (STM) Unit-5 Scanning Tunneling Microscope (STM) Objective: Imaging formation of scanning tunneling microscope (STM) is due to tunneling effect of quantum physics, which is in nano scale. This experiment shows

More information

Park NX-Hivac: Phase-lock Loop for Frequency Modulation Non-Contact AFM

Park NX-Hivac: Phase-lock Loop for Frequency Modulation Non-Contact AFM Park Atomic Force Microscopy Application note #21 www.parkafm.com Hosung Seo, Dan Goo and Gordon Jung, Park Systems Corporation Romain Stomp and James Wei Zurich Instruments Park NX-Hivac: Phase-lock Loop

More information

Study of shear force as a distance regulation mechanism for scanning near-field optical microscopy

Study of shear force as a distance regulation mechanism for scanning near-field optical microscopy Study of shear force as a distance regulation mechanism for scanning near-field optical microscopy C. Durkan a) and I. V. Shvets Department of Physics, Trinity College Dublin, Ireland Received 31 May 1995;

More information

Comparison of resolution specifications for micro- and nanometer measurement techniques

Comparison of resolution specifications for micro- and nanometer measurement techniques P4.5 Comparison of resolution specifications for micro- and nanometer measurement techniques Weckenmann/Albert, Tan/Özgür, Shaw/Laura, Zschiegner/Nils Chair Quality Management and Manufacturing Metrology

More information

Lecture 20: Optical Tools for MEMS Imaging

Lecture 20: Optical Tools for MEMS Imaging MECH 466 Microelectromechanical Systems University of Victoria Dept. of Mechanical Engineering Lecture 20: Optical Tools for MEMS Imaging 1 Overview Optical Microscopes Video Microscopes Scanning Electron

More information

Microscopic Structures

Microscopic Structures Microscopic Structures Image Analysis Metal, 3D Image (Red-Green) The microscopic methods range from dark field / bright field microscopy through polarisation- and inverse microscopy to techniques like

More information

Advanced Nanoscale Metrology with AFM

Advanced Nanoscale Metrology with AFM Advanced Nanoscale Metrology with AFM Sang-il Park Corp. SPM: the Key to the Nano World Initiated by the invention of STM in 1982. By G. Binnig, H. Rohrer, Ch. Gerber at IBM Zürich. Expanded by the invention

More information

Nanosurf easyscan 2 FlexAFM

Nanosurf easyscan 2 FlexAFM Nanosurf easyscan 2 FlexAFM Your Versatile AFM System for Materials and Life Science www.nanosurf.com The new Nanosurf easyscan 2 FlexAFM scan head makes measurements in liquid as simple as measuring in

More information

- Near Field Scanning Optical Microscopy - Electrostatic Force Microscopy - Magnetic Force Microscopy

- Near Field Scanning Optical Microscopy - Electrostatic Force Microscopy - Magnetic Force Microscopy - Near Field Scanning Optical Microscopy - Electrostatic Force Microscopy - Magnetic Force Microscopy Yongho Seo Near-field Photonics Group Leader Wonho Jhe Director School of Physics and Center for Near-field

More information

Standard Operating Procedure of Atomic Force Microscope (Anasys afm+)

Standard Operating Procedure of Atomic Force Microscope (Anasys afm+) Standard Operating Procedure of Atomic Force Microscope (Anasys afm+) The Anasys Instruments afm+ system incorporates an Atomic Force Microscope which can scan the sample in the contact mode and generate

More information

Bruker Dimension Icon AFM Quick User s Guide

Bruker Dimension Icon AFM Quick User s Guide Bruker Dimension Icon AFM Quick User s Guide August 8 2014 GLA Contacts Jingjing Jiang (jjiang2@caltech.edu 626-616-6357) Xinghao Zhou (xzzhou@caltech.edu 626-375-0855) Bruker Tech Support (AFMSupport@bruker-nano.com

More information

Scanning Tunneling Microscopy

Scanning Tunneling Microscopy EMSE-515 02 Scanning Tunneling Microscopy EMSE-515 F. Ernst 1 Scanning Tunneling Microscope: Working Principle 2 Scanning Tunneling Microscope: Construction Principle 1 sample 2 sample holder 3 clamps

More information

Lateral Force: F L = k L * x

Lateral Force: F L = k L * x Scanning Force Microscopy (SFM): Conventional SFM Application: Topography measurements Force: F N = k N * k N Ppring constant: Spring deflection: Pieo Scanner Interaction or force dampening field Contact

More information

; A=4π(2m) 1/2 /h. exp (Fowler Nordheim Eq.) 2 const

; A=4π(2m) 1/2 /h. exp (Fowler Nordheim Eq.) 2 const Scanning Tunneling Microscopy (STM) Brief background: In 1981, G. Binnig, H. Rohrer, Ch. Gerber and J. Weibel observed vacuum tunneling of electrons between a sharp tip and a platinum surface. The tunnel

More information

Bruker Dimension Icon AFM Quick User s Guide

Bruker Dimension Icon AFM Quick User s Guide Bruker Dimension Icon AFM Quick User s Guide March 3, 2015 GLA Contacts Jingjing Jiang (jjiang2@caltech.edu 626-616-6357) Xinghao Zhou (xzzhou@caltech.edu 626-375-0855) Bruker Tech Support (AFMSupport@bruker-nano.com

More information

Constant Frequency / Lock-In (AM-AFM) Constant Excitation (FM-AFM) Constant Amplitude (FM-AFM)

Constant Frequency / Lock-In (AM-AFM) Constant Excitation (FM-AFM) Constant Amplitude (FM-AFM) HF2PLL Phase-locked Loop Connecting an HF2PLL to a Bruker Icon AFM / Nanoscope V Controller Zurich Instruments Technical Note Keywords: AM-AFM, FM-AFM, AFM control Release date: February 2012 Introduction

More information

SENSOR+TEST Conference SENSOR 2009 Proceedings II

SENSOR+TEST Conference SENSOR 2009 Proceedings II B8.4 Optical 3D Measurement of Micro Structures Ettemeyer, Andreas; Marxer, Michael; Keferstein, Claus NTB Interstaatliche Hochschule für Technik Buchs Werdenbergstr. 4, 8471 Buchs, Switzerland Introduction

More information

University of MN, Minnesota Nano Center Standard Operating Procedure

University of MN, Minnesota Nano Center Standard Operating Procedure Equipment Name: Atomic Force Microscope Badger name: afm DI5000 PAN Revisionist Paul Kimani Model: Dimension 5000 Date: October 6, 2017 Location: Bay 1 PAN Revision: 1 A. Description i. Enhanced Motorized

More information

Near-field Optical Microscopy

Near-field Optical Microscopy Near-field Optical Microscopy R. Fernandez, X. Wang, N. Li, K. Parker, and A. La Rosa Physics Department Portland State University Portland, Oregon Near-Field SPIE Optics Microscopy East 2005 Group PSU

More information

Basic methods in imaging of micro and nano structures with atomic force microscopy (AFM) Related Topics. Principle. Equipment

Basic methods in imaging of micro and nano structures with atomic force microscopy (AFM) Related Topics. Principle. Equipment Basic methods in imaging of micro and nano P2538000 Related Topics Atomic Force Microscopy (AFM), Lennard Jones potential, static force mode, dynamic force mode, feedback (PID), setpoint, Vibration amplitude,

More information

Nanosurf Nanite. Automated AFM for Industry & Research.

Nanosurf Nanite. Automated AFM for Industry & Research. Nanosurf Nanite Automated AFM for Industry & Research www.nanosurf.com Multiple Measurements Automated Got work? Nanosurf has the solution! The Swiss-based innovator and manufacturer of the most compact

More information

A Project Report Submitted to the Faculty of the Graduate School of the University of Minnesota By

A Project Report Submitted to the Faculty of the Graduate School of the University of Minnesota By Observation and Manipulation of Gold Clusters with Scanning Tunneling Microscopy A Project Report Submitted to the Faculty of the Graduate School of the University of Minnesota By Dogukan Deniz In Partial

More information

Imaging Carbon Nanotubes Magdalena Preciado López, David Zahora, Monica Plisch

Imaging Carbon Nanotubes Magdalena Preciado López, David Zahora, Monica Plisch Imaging Carbon Nanotubes Magdalena Preciado López, David Zahora, Monica Plisch I. Introduction In this lab you will image your carbon nanotube sample from last week with an atomic force microscope. You

More information

attosnom I: Topography and Force Images NANOSCOPY APPLICATION NOTE M06 RELATED PRODUCTS G

attosnom I: Topography and Force Images NANOSCOPY APPLICATION NOTE M06 RELATED PRODUCTS G APPLICATION NOTE M06 attosnom I: Topography and Force Images Scanning near-field optical microscopy is the outstanding technique to simultaneously measure the topography and the optical contrast of a sample.

More information

Akiyama-Probe (A-Probe) technical guide This technical guide presents: how to make a proper setup for operation of Akiyama-Probe.

Akiyama-Probe (A-Probe) technical guide This technical guide presents: how to make a proper setup for operation of Akiyama-Probe. Akiyama-Probe (A-Probe) technical guide This technical guide presents: how to make a proper setup for operation of Akiyama-Probe. Version: 2.0 Introduction To benefit from the advantages of Akiyama-Probe,

More information

INDIAN INSTITUTE OF TECHNOLOGY BOMBAY

INDIAN INSTITUTE OF TECHNOLOGY BOMBAY IIT Bombay requests quotations for a high frequency conducting-atomic Force Microscope (c-afm) instrument to be set up as a Central Facility for a wide range of experimental requirements. The instrument

More information

Figure for the aim4np Report

Figure for the aim4np Report Figure for the aim4np Report This file contains the figures to which reference is made in the text submitted to SESAM. There is one page per figure. At the beginning of the document, there is the front-page

More information

Cutting-edge Atomic Force Microscopy techniques for large and multiple samples

Cutting-edge Atomic Force Microscopy techniques for large and multiple samples Cutting-edge Atomic Force Microscopy techniques for large and multiple samples Study of up to 200 mm samples using the widest set of AFM modes Industrial standards of automation A unique combination of

More information

LAB UNIT 1: Introduction Scanning Force Microscopy

LAB UNIT 1: Introduction Scanning Force Microscopy LAB UNIT 1: Introduction Specific Assignment: Setup of scanning force microscopy experiment and first contact measurements Objective Outcome Synopsis The student will become familiar with contact mode

More information

Keysight Technologies Using Non-Contact AFM to Image Liquid Topographies. Application Note

Keysight Technologies Using Non-Contact AFM to Image Liquid Topographies. Application Note Keysight Technologies Using Non-Contact AFM to Image Liquid Topographies Application Note Introduction High resolution images of patterned liquid surfaces have been acquired without inducing either capillary

More information

Park NX-Hivac The world s most accurate and easy to use high vacuum AFM for failure analysis.

Park NX-Hivac The world s most accurate and easy to use high vacuum AFM for failure analysis. Park NX-Hivac The world s most accurate and easy to use high vacuum AFM for failure analysis www.parkafm.com Park NX-Hivac High vacuum scanning for failure analysis applications 4 x 07 / Cm3 Current (µa)

More information

MEMS for RF, Micro Optics and Scanning Probe Nanotechnology Applications

MEMS for RF, Micro Optics and Scanning Probe Nanotechnology Applications MEMS for RF, Micro Optics and Scanning Probe Nanotechnology Applications Part I: RF Applications Introductions and Motivations What are RF MEMS? Example Devices RFIC RFIC consists of Active components

More information

Optical Microscope. Active anti-vibration table. Mechanical Head. Computer and Software. Acoustic/Electrical Shield Enclosure

Optical Microscope. Active anti-vibration table. Mechanical Head. Computer and Software. Acoustic/Electrical Shield Enclosure Optical Microscope On-axis optical view with max. X magnification Motorized zoom and focus Max Field of view: mm x mm (depends on zoom) Resolution : um Working Distance : mm Magnification : max. X Zoom

More information

Instructions for the Experiment

Instructions for the Experiment Instructions for the Experiment Excitonic States in Atomically Thin Semiconductors 1. Introduction Alongside with electrical measurements, optical measurements are an indispensable tool for the study of

More information

Nanosurf easyscan 2 Your Modular Scanning Probe Microscopy System

Nanosurf easyscan 2 Your Modular Scanning Probe Microscopy System Nanosurf Your Modular Scanning Probe Microscopy System www.nanosurf.com Perfect Modularity Nanosurf s easyscan series has gained worldwide popularity through its affordability, portability, and ease of

More information

Nanosurf Easyscan 2 STM

Nanosurf Easyscan 2 STM Nanosurf Easyscan 2 STM Operating Instructions for SPM Control Software Version 3.1 NANOSURF AND THE NANOSURF LOGO ARE TRADEMARKS OF NANOSURF AG, REGISTERED AND/OR OTHERWISE PROTECTED IN VARIOUS COUNTRIES.

More information

SOLVER NANO Scanning Probe Microscope

SOLVER NANO Scanning Probe Microscope SOLVER NANO Scanning Probe Microscope Instruction Manual 2012 Copyright «NT-MDT» Web Page: http://www.ntmdt.com General Information: spm@ntmdt.ru Technical Support: support@ntmdt.ru NT-MDT Co., building

More information

NSOM (SNOM) Overview

NSOM (SNOM) Overview NSOM (SNOM) Overview The limits of far field imaging In the early 1870s, Ernst Abbe formulated a rigorous criterion for being able to resolve two objects in a light microscope: d > ë / (2sinè) where d

More information

Kit for building your own THz Time-Domain Spectrometer

Kit for building your own THz Time-Domain Spectrometer Kit for building your own THz Time-Domain Spectrometer 16/06/2016 1 Table of contents 0. Parts for the THz Kit... 3 1. Delay line... 4 2. Pulse generator and lock-in detector... 5 3. THz antennas... 6

More information

Distinguishing Between Mechanical and Electrostatic. Interaction in Single-Pass Multifrequency Electrostatic Force

Distinguishing Between Mechanical and Electrostatic. Interaction in Single-Pass Multifrequency Electrostatic Force SUPPORTING INFORMATION Distinguishing Between Mechanical and Electrostatic Interaction in Single-Pass Multifrequency Electrostatic Force Microscopy on a Molecular Material Marta Riba-Moliner, Narcis Avarvari,

More information

attocfm I for Surface Quality Inspection NANOSCOPY APPLICATION NOTE M01 RELATED PRODUCTS G

attocfm I for Surface Quality Inspection NANOSCOPY APPLICATION NOTE M01 RELATED PRODUCTS G APPLICATION NOTE M01 attocfm I for Surface Quality Inspection Confocal microscopes work by scanning a tiny light spot on a sample and by measuring the scattered light in the illuminated volume. First,

More information

Operating Instructions. easyscan 2 STM. Version 2.1

Operating Instructions. easyscan 2 STM. Version 2.1 Operating Instructions easyscan 2 STM Version 2.1 NANOSURF AND THE NANOSURF LOGO ARE TRADEMARKS OF NANOSURF AG, REGISTERED AND/OR OTHERWISE PROTECTED IN VARIOUS COUNTRIES. COPYRIGHT APRIL 2009, NANOSURF

More information

Scanning Electron Microscopy

Scanning Electron Microscopy Scanning Electron Microscopy For the semiconductor industry A tutorial Titel Vorname Nachname Titel Jobtitle, Bereich/Abteilung Overview Scanning Electron microscopy Scanning Electron Microscopy (SEM)

More information

ADALAM Sensor based adaptive laser micromachining using ultrashort pulse lasers for zero-failure manufacturing D2.2. Ger Folkersma (Demcon)

ADALAM Sensor based adaptive laser micromachining using ultrashort pulse lasers for zero-failure manufacturing D2.2. Ger Folkersma (Demcon) D2.2 Automatic adjustable reference path system Document Coordinator: Contributors: Dissemination: Keywords: Ger Folkersma (Demcon) Ger Folkersma, Kevin Voss, Marvin Klein (Demcon) Public Reference path,

More information

Operating Instructions. easyscan E-STM Version 2.0

Operating Instructions. easyscan E-STM Version 2.0 Operating Instructions easyscan E-STM Version 2.0 1 TEXT & LAYOUT: KARIN HOOL, R. SUM, PIETER VAN SCHENDEL ENGLISH: VICKY CONNOLLY NANOSURF AND THE NANOSURF LOGO ARE TRADEMARKS OF NANOSURF AG, REGISTERED

More information

HybridStage - Automated, large sample-area mapping made easy

HybridStage - Automated, large sample-area mapping made easy HybridStage - Automated, large sample-area mapping made easy Motivation Crucial parameters that affect cell adhesion, morphogenesis, cell differentiation and cancer invasion include the molecular interactions

More information

Nanoscale Material Characterization with Differential Interferometric Atomic Force Microscopy

Nanoscale Material Characterization with Differential Interferometric Atomic Force Microscopy Nanoscale Material Characterization with Differential Interferometric Atomic Force Microscopy F. Sarioglu, M. Liu, K. Vijayraghavan, A. Gellineau, O. Solgaard E. L. Ginzton Laboratory University Tip-sample

More information

nanovea.com PROFILOMETERS 3D Non Contact Metrology

nanovea.com PROFILOMETERS 3D Non Contact Metrology PROFILOMETERS 3D Non Contact Metrology nanovea.com PROFILOMETER INTRO Nanovea 3D Non-Contact Profilometers are designed with leading edge optical pens using superior white light axial chromatism. Nano

More information

3B SCIENTIFIC PHYSICS

3B SCIENTIFIC PHYSICS B SCIENTIFIC PHYSICS Cavendish Torsion Balance 007 Operating instructions 06/8 ALF. Description The Cavendish torsion balance is for demonstrating the gravitational attraction between two masses and determining

More information

SUPPLEMENTARY INFORMATION

SUPPLEMENTARY INFORMATION Figure S. Experimental set-up www.nature.com/nature Figure S2. Dependence of ESR frequencies (GHz) on a magnetic field (G) applied in different directions with respect to NV axis ( θ 2π). The angle with

More information

Atomic Force Microscopy (I)

Atomic Force Microscopy (I) Atomic Force Microscopy (I) - Optical Grating AFM and the thermal noise measurement 2.674 Lab 10 Spring 2016 Pappalardo II Micro/Nano Laboratories AFM Imaging (with home-made AFMs) I. Safety Notes This

More information

Magnetic Levitation System

Magnetic Levitation System Introduction Magnetic Levitation System There are two experiments in this lab. The first experiment studies system nonlinear characteristics, and the second experiment studies system dynamic characteristics

More information

P a g e 1 ST985. TDR Cable Analyzer Instruction Manual. Analog Arts Inc.

P a g e 1 ST985. TDR Cable Analyzer Instruction Manual. Analog Arts Inc. P a g e 1 ST985 TDR Cable Analyzer Instruction Manual Analog Arts Inc. www.analogarts.com P a g e 2 Contents Software Installation... 4 Specifications... 4 Handling Precautions... 4 Operation Instruction...

More information

Fast Quality Control of Suspension Parts AN 53

Fast Quality Control of Suspension Parts AN 53 Application Note for the KLIPPEL QC SYSTEM The performance and quality of loudspeaker drivers and complete audio systems is mainly determined by the quality of the single components. To ensure a consistent

More information

A scanning tunneling microscopy based potentiometry technique and its application to the local sensing of the spin Hall effect

A scanning tunneling microscopy based potentiometry technique and its application to the local sensing of the spin Hall effect A scanning tunneling microscopy based potentiometry technique and its application to the local sensing of the spin Hall effect Ting Xie 1, a), Michael Dreyer 2, David Bowen 3, Dan Hinkel 3, R. E. Butera

More information

Experiment 3 Topic: Dynamic System Response Week A Procedure

Experiment 3 Topic: Dynamic System Response Week A Procedure Experiment 3 Topic: Dynamic System Response Week A Procedure Laboratory Assistant: Email: Office Hours: LEX-3 Website: Brock Hedlund bhedlund@nd.edu 11/05 11/08 5 pm to 6 pm in B14 http://www.nd.edu/~jott/measurements/measurements_lab/e3

More information

NTEGRA Spectra Probe NanoLaboratory. SNOM Measuring Head

NTEGRA Spectra Probe NanoLaboratory. SNOM Measuring Head NTEGRA Spectra Probe NanoLaboratory SNOM Measuring Head Instruction Manual 16 April 2010 Copyright NT-MDT Web Page: http://www.ntmdt.com General Information: spm@ntmdt.ru Technical Support: support@ntmdt.ru

More information

Surface Finish Measurement Methods and Instrumentation

Surface Finish Measurement Methods and Instrumentation 125 years of innovation Surface Finish Measurement Methods and Instrumentation Contents Visual Inspection Surface Finish Comparison Plates Contact Gauges Inductive / Variable Reluctance (INTRA) Piezo Electric

More information

2 Oscilloscope Familiarization

2 Oscilloscope Familiarization Lab 2 Oscilloscope Familiarization What You Need To Know: Voltages and currents in an electronic circuit as in a CD player, mobile phone or TV set vary in time. Throughout the course you will investigate

More information

Indentation Cantilevers

Indentation Cantilevers curve is recorded utilizing the DC displacement of the cantilever versus the extension of the scanner. Many indentations may be made using various forces, rates, etc. Upon exiting indentation mode, TappingMode

More information

Radial Polarization Converter With LC Driver USER MANUAL

Radial Polarization Converter With LC Driver USER MANUAL ARCoptix Radial Polarization Converter With LC Driver USER MANUAL Arcoptix S.A Ch. Trois-portes 18 2000 Neuchâtel Switzerland Mail: info@arcoptix.com Tel: ++41 32 731 04 66 Principle of the radial polarization

More information

Exercise 4. Angle Tracking Techniques EXERCISE OBJECTIVE

Exercise 4. Angle Tracking Techniques EXERCISE OBJECTIVE Exercise 4 Angle Tracking Techniques EXERCISE OBJECTIVE When you have completed this exercise, you will be familiar with the principles of the following angle tracking techniques: lobe switching, conical

More information

Ph 3455 The Photoelectric Effect

Ph 3455 The Photoelectric Effect Ph 3455 The Photoelectric Effect Required background reading Tipler, Llewellyn, section 3-3 Prelab Questions 1. In this experiment you will be using a mercury lamp as the source of photons. At the yellow

More information

LAB I. INTRODUCTION TO LAB EQUIPMENT

LAB I. INTRODUCTION TO LAB EQUIPMENT 1. OBJECTIVE LAB I. INTRODUCTION TO LAB EQUIPMENT In this lab you will learn how to properly operate the oscilloscope Agilent MSO6032A, the Keithley Source Measure Unit (SMU) 2430, the function generator

More information

Fluorescence Imaging of Single Spins in Nitrogen-Vacancy centers using a Confocal Microscope. Advanced Lab Course University of Basel

Fluorescence Imaging of Single Spins in Nitrogen-Vacancy centers using a Confocal Microscope. Advanced Lab Course University of Basel Fluorescence Imaging of Single Spins in Nitrogen-Vacancy centers using a Confocal Microscope Advanced Lab Course University of Basel October 6, 2015 Contents 1 Introduction 2 2 The Confocal Microscope

More information

2. Operating modes in scanning probe microscopy

2. Operating modes in scanning probe microscopy . Operating modes in scanning probe microscopy.1. Scanning tunneling microscopy Historically, the first microscope in the family of probe microscopes is the scanning tunneling microscope. The working principle

More information

Development of a Package for a Triaxial High-G Accelerometer Optimized for High Signal Fidelity

Development of a Package for a Triaxial High-G Accelerometer Optimized for High Signal Fidelity Development of a Package for a Triaxial High-G Accelerometer Optimized for High Signal Fidelity R. Langkemper* 1, R. Külls 1, J. Wilde 2, S. Schopferer 1 and S. Nau 1 1 Fraunhofer Institute for High-Speed

More information

Nanoscience Instruments

Nanoscience Instruments Nanoscience Instruments Operating Instructions for traxstm Control Software Version 3.3 NANOSCIENCE INSTRUMENTS AND THE NANOSCIENCE INSTRUMENTS LOGO ARE TRADEMARKS OF NANOSCIENCE INSTRUMENTS INC, REGISTERED

More information

Controller Design for Z Axis Movement of STM Using SPM Control Software

Controller Design for Z Axis Movement of STM Using SPM Control Software Controller Design for Z Axis Movement of STM Using SPM Control Software Neena Tom, Rini Jones S. B Abstract Scanning probe microscopy is a branch of microscopy that forms images of surfaces using a physical

More information

ScanArray Overview. Principle of Operation. Instrument Components

ScanArray Overview. Principle of Operation. Instrument Components ScanArray Overview The GSI Lumonics ScanArrayÒ Microarray Analysis System is a scanning laser confocal fluorescence microscope that is used to determine the fluorescence intensity of a two-dimensional

More information

Fire CR Calibration Guide

Fire CR Calibration Guide 1 Fire CR Calibration Guide This reference guide will guide you through the steps to complete the calibration for the Fire CR.. Getting Started: 1. Click on the Opal Icon on the Desktop. Figure 1 2. Once

More information

Improving the Collection Efficiency of Raman Scattering

Improving the Collection Efficiency of Raman Scattering PERFORMANCE Unparalleled signal-to-noise ratio with diffraction-limited spectral and imaging resolution Deep-cooled CCD with excelon sensor technology Aberration-free optical design for uniform high resolution

More information

Instruction Manual for HyperScan Spectrometer

Instruction Manual for HyperScan Spectrometer August 2006 Version 1.1 Table of Contents Section Page 1 Hardware... 1 2 Mounting Procedure... 2 3 CCD Alignment... 6 4 Software... 7 5 Wiring Diagram... 19 1 HARDWARE While it is not necessary to have

More information

Profile Measurement of Resist Surface Using Multi-Array-Probe System

Profile Measurement of Resist Surface Using Multi-Array-Probe System Sensors & Transducers 2014 by IFSA Publishing, S. L. http://www.sensorsportal.com Profile Measurement of Resist Surface Using Multi-Array-Probe System Shujie LIU, Yuanliang ZHANG and Zuolan YUAN School

More information

Manufacturing Metrology Team

Manufacturing Metrology Team The Team has a range of state-of-the-art equipment for the measurement of surface texture and form. We are happy to discuss potential measurement issues and collaborative research Manufacturing Metrology

More information

SUPPORTING INFORMATION FOR

SUPPORTING INFORMATION FOR SUPPORTING INFORMATION FOR Internal Hydration Properties of Single Bacterial Endospores Probed by Electrostatic Force Microscopy Marc Van Der Hofstadt 1,2, Rene Fabregas 1,2, Ruben Millan-Solsona 1, Antonio

More information

Operating the Hitachi 7100 Transmission Electron Microscope Electron Microscopy Core, University of Utah

Operating the Hitachi 7100 Transmission Electron Microscope Electron Microscopy Core, University of Utah Operating the Hitachi 7100 Transmission Electron Microscope Electron Microscopy Core, University of Utah Follow the procedures below when you use the Hitachi 7100 TEM. Starting Session 1. Turn on the cold

More information

IMAGING P-N JUNCTIONS BY SCANNING NEAR-FIELD OPTICAL, ATOMIC FORCE AND ELECTRICAL CONTRAST MICROSCOPY. G. Tallarida Laboratorio MDM-INFM

IMAGING P-N JUNCTIONS BY SCANNING NEAR-FIELD OPTICAL, ATOMIC FORCE AND ELECTRICAL CONTRAST MICROSCOPY. G. Tallarida Laboratorio MDM-INFM Laboratorio MDM - INFM Via C.Olivetti 2, I-20041 Agrate Brianza (MI) M D M Materiali e Dispositivi per la Microelettronica IMAGING P-N JUNCTIONS BY SCANNING NEAR-FIELD OPTICAL, ATOMIC FORCE AND ELECTRICAL

More information

XYZ Stage. Surface Profile Image. Generator. Servo System. Driving Signal. Scanning Data. Contact Signal. Probe. Workpiece.

XYZ Stage. Surface Profile Image. Generator. Servo System. Driving Signal. Scanning Data. Contact Signal. Probe. Workpiece. Jpn. J. Appl. Phys. Vol. 40 (2001) pp. 3646 3651 Part 1, No. 5B, May 2001 c 2001 The Japan Society of Applied Physics Estimation of Resolution and Contact Force of a Longitudinally Vibrating Touch Probe

More information

Vibration Isolation for Scanning Tunneling Microscopy

Vibration Isolation for Scanning Tunneling Microscopy Vibration Isolation for Scanning Tunneling Microscopy Catherine T. Truett Department of Physics, Michigan State University East Lansing, Michigan 48824 ABSTRACT Scanning Tunneling Microscopy measures tunneling

More information

Eddy current flaw detector «Eddycon C»

Eddy current flaw detector «Eddycon C» ULTRACON-SERVICE LLC Eddy current flaw detector «Eddycon C» Quick start guide CONTENTS P. 1 CONTROLLERS OF EDDYCON C FLAW DETECTOR... 3 2 SETTINGS OF «TEST» MENU... 5 3 INSTRUCTIONS FOR USE... 8 3.1 THRESHOLD

More information

Multi-Probe Atomic Force Microscopy Using Piezo-Resistive Cantilevers and Interaction between Probes

Multi-Probe Atomic Force Microscopy Using Piezo-Resistive Cantilevers and Interaction between Probes e-journal of Surface Science and Nanotechnology 26 January 2013 e-j. Surf. Sci. Nanotech. Vol. 11 (2013) 13-17 Regular Paper Multi-Probe Atomic Force Microscopy Using Piezo-Resistive Cantilevers and Interaction

More information

2 How to operate the microscope/obtain an image

2 How to operate the microscope/obtain an image Morgagni Operating Instructions 50079 010912 2-1 2 ow to operate the microscope/obtain an image 2.1 Starting the microscope 2.1.1 Starting the microscope with several manually-operated steps 1. Turn on

More information

Akiyama-Probe (A-Probe) simple DIY controller This technical guide presents: simple and low-budget DIY controller

Akiyama-Probe (A-Probe) simple DIY controller This technical guide presents: simple and low-budget DIY controller Akiyama-Probe (A-Probe) simple DIY controller This technical guide presents: simple and low-budget DIY controller Version: 2.0 Introduction NANOSENSORS has developed a simple and low-budget controller

More information

Confocal Application Notes Vol. 5 July 2010

Confocal Application Notes Vol. 5 July 2010 Tile Scan Prepared by Myriam Gastard, PhD Application and Technical Support Group, Leica Microsystems, Inc. In this issue of our Confocal Application Notes, proper set up of the Tile function enables you

More information

Part 2: Second order systems: cantilever response

Part 2: Second order systems: cantilever response - cantilever response slide 1 Part 2: Second order systems: cantilever response Goals: Understand the behavior and how to characterize second order measurement systems Learn how to operate: function generator,

More information