Clinical Importance on CT

Size: px
Start display at page:

Download "Clinical Importance on CT"

Transcription

1 183 Truncated-View Clinical Importance on CT..,..,.... Artifacts:.... James L. Lehr1 A truncated-view artifact in CT is produced whenever any part of the patient or imaged object is present in some but not all of the views obtained for a slice. The potential to create images with this artifact exists for any CT scanner in which the fan beam (or its equivalent) does not cover the entire gantry aperture. This includes most CT systems currently on the market. Although the artifact may not create a severe visual disturbance in the image, It can alter the CT numbers in a manner that will compromise the accuracy of quantitative analyses. This report describes the nature of the truncated-view artifact and presents simulated examples for both mathematical phantoms and clinical scans. The artifact can be eliminated by assuring that the entire patient and all foreign objects are included in the field of view, or it can be minimized by placing objects that cannot be entirely within the field of view as close to the edge of the gantry aperture as possible. CT systems reconstruct images from many x-ray projections. Intuitively one might expect that these projections should be obtained from all possible directions through all parts of the object to be imaged, and that if some part of the object were not viewed from all directions, errors would result. This is in fact the case, and the reconstruction methods used to produce CT images assume that the object lies entirely within a region that is viewed from all directions [1 ]. In real CT systems, this assumption may not be correct, and errors or artifacts in the image then will be produced. Figure 1 represents a typical third-generation or fan-beam CT system in which the x-ray tube rotates around a circle of radius D and is coupled to a curved detector array which intersects an angle a of the fan beam. As the x-ray tube rotates, the fan beams will intersect to form a circle (of radius r in fig. 1 ) centered at the center of rotation. An object inside the circle will lie within the fan beam for all positions of the tube, and its projection will be obtained from all directions. Therefore, an object lying entirely within this circular field of view can be reconstructed accurately, if nonlinear effects such as beam hardening [2] and scatter [3] are neglected. On the other hand, if the gantry aperture is larger than the field of view, it is possible for an object (such as that labeled A in fig. 1 ) to be in the gantry but outside the field of view. The x-ray projection of such an object will be detected for some but not all positions of the tube-detector apparatus. For example, the object A in figure 1 will lie in the beam when the tube is in position 1, but it will lie out of the beam when the tube has rotated to position 2. A reconstruction Received October ; accepted after re- made from projection data collected under these circumstances will produce an vision February 4, image that contains artifacts [4-7]. One approach to this situation 15 to regard Chi- the object as being too large for the field of view, and this perspective leads one to speak of the fat-man artifacts which are produced by the oversized object. AJR 141: , July X/83/ Another approach 5 to note that If the detector array were longer, so that the American Roentgen Ray Society field of view would be larger, the artifacts would not occur. This leads one to

2 184 LEHR AJR:141, July 1983 Fig. 1 -Geometry of third-generation CT system. X-ray tube moving along outer solid circle of radius D produces detected fan beam of angle a which defines field of view indicated by dashed circle of radius r. If gantry aperture (inner solid circle of radius G) is larger than field of view, object inside gantry such as that labeled A will be detected for some positions of tube such as that labeled 1, but it will be undetected for other tube positions, such as that labeled 2. This produces truncated-view artifact. speak of incomplete data, missing projections, limited field of view, or simply truncated-view artifacts. The artifact also can be generated by parallel-beam and by fourth-generation inverted-fan-beam scanners under similar circumstances. Various methods to correct at least partially for truncatedview artifacts have been described in the physics literature [4-7], but no thorough assessment of the potential effect of such artifacts in the interpretation and analysis of CT studies has appeared in the clinical literature. This report focuses on these aspects by presenting several simulated examples of truncated view artifacts for both simple mathematical objects and clinical scans. In addition, practical methods to detect, minimize, or avoid the effects of such artifacts in clinical work are suggested. Materials and Methods The approach taken to study truncated-view artifacts for this report begins by collecting a complete set of projection values, that is, a set in which the view from each tube position includes the entire object. A simple computer program then simulates a smaller field of view by setting the projection values associated with a certain number of the outer detectors to zero. By zeroing a larger number of projection values associated with a larger number of detectors at the ends of the array, the program simulates narrower fan beams and smaller fields of view. These truncated views are then used to create a reconstructed image that produces the artifacts that would be obtained if the outer detectors were not present. Fig. 2.-Clinical scans reconstructed from complete projections covering shoulders in pediatric patient displayed with wide window (window level 0, 53-cm field of view indicated by superimposed circle. A, Scan of abdomen in window width 2000). obese patient (window level 0, window width 51 2). B, Chest scan at level of

3 AJR:141, July 1983 Cl TRUNCATED-VIEW ARTIFACTS 185 C Fig. 3.-Images of mathematical 2-cm disk with CT number 91 2 H located cm from center of rotation in background of air. A, Reconstruction for 53-cm-diam field of view which corresponds to gantry aperture indicated by superimposed circle. Graph of CT numbers along dashed line passing through disk and center of rotation shows no truncated-view artifact (window level 0, window width 1 024). B, Reconstruction made from truncated views corresponding to 24-cm-diam field of view indicated by inner circle. (Outer circle again indicates gantry aperture.) Graph of CT numbers along dashed line shows that CT numbers in area of disk are too small (about -400 H) and that The objects for which truncated-view images have been produced include both mathematical phantoms and patient scans. Patient scans that were used included an abdominal scan in an obese patient(fig. 2A) and a chest scan atthe level ofthe shoulders in a small pediatric patient (fig. 2B). The mathematical phantoms comprised small and large circular disks, for which complete projection values were computed using appropriate formulae. The advantages of using mathematical phantoms are that the size, D CT numbers in air near disk are too high because of truncated-view artifact (window level , window width 32). C, Artifact image formed by subtracting A from B. Graph of CT numbers shows negative artifact in location of disk and positive artifact (about +600 H) near disk, but falls off nearly to zero at center of rotation (window level 0, window width 32). D, Same artifact image as C with graph of CT numbers along dashed line shown extending diagonally from center of rotation to just beyond field of view. Largest artifact of about 70 H occurs at periphery of field of view (window level 0. window width 32). location, and CT value of the object can be controlled relatively precisely, and that the simulated projection values do not include nonlinear effects such as beam hardening [2] or scatter [3]. The small disks had a diameter of 2 cm and a CT number of 912 Hounsfield units (H). They were placed in a background of air at various distances from the center of notation (fig. 3A). A small disk should approximate the effects of a foreign object such as a syringe containing dilute contrast material or an extremity in the gantry

4 186 LEHR AJR:141, July 1983 aperture. The large disk had a CT value of 0 H and a diameter of 52 cm. It was centered at the system s center of rotation, again in a background of air. This disk should approximate the effects of scanning a very large patient. The scans on patients were performed on a Siemens Somatom II system in which the field of view corresponds to the gantry aperture. The abdominal scan used a slice thickness of 8 mm, a peak voltage of 1 25 kev, and 360 views of 51 2 projections, each collected in 5 sec. The chest scan used a thickness of 4 mm, a peak voltage of 1 25 kev, and 720 views of 51 2 projections, collected in 1 0 sec. The mathematical scans were computed for 720 views. For all objects, reconstructions have been made at 256 x 256 pixels for the entire field of view, a circle with a diameter of 53 cm. Truncated projections corresponding to smaller fields of view then were used to reconstruct images with the truncated-view artifacts. Subtracting the image produced with complete views from the image made with truncated views produces an image of the artifact. Images were photographed at selected settings, often with superimposed graphs of pixel values produced by analysis programs supplied with the CT system in order to demonstrate the magnitude of the artifact. All neconstructions were made using the filtered backpnojection software supplied with the Somatom II system and using the standard body-mode kennel. For some images comesponding to smaller fields of view, an extended scale option, which doubles the upper CT number range so that it reaches 2048 H, was used to avoid or minimize overshoot artifacts which occur when the computed CT number is larger than H. In addition, the average value of five pixels at the center of the artifact image was computed. This is referred to as the central error. Results The image reconstructed from complete views of a 2-cmdiam mathematical disk with its center located cm from the center of rotation is shown in figure 3A. The large circle superimposed on this figure shows the location of the gantry aperture, which is the same as the field of view. The superimposed graph shows the CT numbers along the dashed line passing through the center of the disk and the center of rotation. In this and similar graphs CT values that exceed the maximum or minimum values on the scale are shown as the maximum or minimum values, respectively. The image of this disk reconstructed from views that have been truncated to produce a limited field of view 24.0 cm in diameter is shown in figure 3B, along with a graph of CT values and larger and smaller circles representing the gantry aperture and field of view, respectively. Note that the window width has been narrowed and that the scale of the graph is different from the one in figure 3A. Because the projection of the disk is detected in some but not all of the truncated views, its density is not reconstructed accurately but is smeared out over a larger area. The artifact image formed by subtracting figure 3A from 3B is shown in figure 3C, again with superimposed circles and a graph of Cl values along the horizontal diameter of the image. It is apparent that most of the dense part of the artifact occurs in streaks extending from the disk tangent to the field of view. FIgure 3D shows the same artifact with a superimposed graph of the CT values along a line extending radially from the center of the field of view through the 0 Li 0 C a, r6cm 0 r:9cm 0 r:i2cm & A r:l8cm r:24cm. I I I I I Disk Distance, d (cm.) Fig. 4.-Central error generated from mathematical 2-cm disk with CT value of 91 2 H lying outside various-sized fields of view as function of distance between center of disk and center of rotation. For fixed field of view, central error decreases (or remains constant) as disk is moved further from center of rotation. For fixed distance between disk and center of rotation, central error increases as field of view becomes smaller. upper streak. The magnitude of the artifact along this line falls off rapidly from its maximum value of about 60 H near the edge of the field of view to a value of about 2 H near the center. Although the artifact outside the field of view is sometimes negative, the artifact within the field of view is always positive or zero. Because CT reconstruction is a linear operation, the magnitude of the truncated-view artifact should be a linear function of the CT number of the disk. Varying the CT value of a disk confirms that the magnitude of the artifact at any point is, as expected, directly proportional to the disk s CT number plus The additive term of is needed because in the Hounsfield scale, air, which has a linear attenuation of zero, is assigned the value of units. A CT value in attenuation units (AU) may be defined as the CT number in Hounsfield units plus In this scale air has a value of 0 AU, water has a value of AU, and the disk would have a value of AU. A disk with half the linear attenuation of that shown in figure 3A would then have a CT value of /2 or 856 AU, corresponding to a CT number of -44 H. If such a disk were scanned under the conditions for figure 3B, the magnitude of the artifact would be half as large, but the spatial distribution would be the same as that shown in figure 3C. In a series of experiments, the distance of the 2-cm disk from the center of rotation and the size of the limited field of view were varied (fig. 4). For a fixed size of the field of view, the total central error decreases (or remains constant) as the disk is moved away from the center of rotation. For a fixed disk location, the central error increases as the diameter of the field of view decreases. The large disk and clinical scans were reconstructed for limited fields of view of various sizes (figs. 5-7). Figure 5A I A

5 AJR:141, July 1983 Cl TRUNCATED-VIEW ARTIFACTS 187 shows the image of the 52-cm, water-density (0 H) disk reconstructed from truncated views corresponding to a limited field of view 48 cm in diameter. The only artifact that can be seen is an increase in the CT numbers near the edge of the field of view. The graph of CT numbers across the horizontal diameter of the image shows that the magnitude of the artifact near the edge of the limited field of view is about H, but that the central error is nearly 0 H (actually 1 2 H). Because the large disk has a CT number of 0 H, the limited-field-of-view image and the artifact image are identical inside the limited field. The image of the large disk for a limited field of view 24.0 cm in diameter (fig. SB) shows a more pronounced artifact. This image was made using the extended scale option. Even then, the artifact near the edge shows an overshoot effect, and one can state only that the edge artifact is greater than 2048 H. The central artifact is H. Limited-view images and superimposed graphs of the corresponding truncated-view artifacts for the abdominal scan in an obese patient (fig. 2A) are shown in figure 6. The image for a 48-cm-diam field of view (fig. 6A) shows a small edge artifact with a maximum value of about H and a central error of only 2.6 H. Indeed, if the patient had been more carefully centered in the aperture, the truncated-view artifact could have been eliminated. For the 24-cm diam field of view (fig. 6B) the artifact is much more pronounced, with an overshoot artifact indicating a maximum value of more than H near the edge and 243 H near the center. The 1 8-cm-diam field of view (fig. 6C) has a severe cupping artifact that again exceeds H near the edge of the field and is 509 H at the center. In all truncated-view images Fig. 5.-Limited-field-of-view reconstructions of mathematical 52-cmdiam disk with density corresponding to 0 H. A, Reconstruction made from truncated views corresponding to 48-cm-diam field of view indicated by outer margin of disk s image. Graph of CT values along horizontal diameter mdicated by dashed line shows that artifact is primarily evidenced by cupping near edge of field of view (window level 0, window width 51 2). B, Reconstruction made with extended CT scale option from truncated views corresponding to 24-cm-diam field of view indicated by outer margin of disk s image. Graph shows much more pronounced cupping artifact due to smaller field of view and to larger amount of disk lying outside field of view (window level 1400, window width 1024). produced from this scan, the artifact is slightly asymmetrical, being greater in the upper left part of the field. This is because the patient is not centered, so that slightly more of the out-of-field density lies beyond the upper left border of the field of view. Limited-view reconstructions of the pediatric chest scan (fig. 2B) are shown in figure 7. A 36-cm-diam field of view was used for figure 7A. The artifact shows high-density streaks near the edge of the field and resembles the sum of two small disk artifacts with the disk placed in the location of the patient s shoulders. The absence of out-of-field matenial above or below the patient leads to virtually no artifact along a vertical line through the center of rotation. The 18- cm-diam field of view (fig. 7B) is just large enough to include cross sections of both lungs, but as indicated by the graph of the artifact, if the limited-field scan were used to estimate the CT number of, say, a pulmonary nodule, the results would be incorrectly high by about 200 H for a nodule near the midline and by about H for a peripheral nodule. The 1 2-cm-diam field of view (fig. 7C) is associated with a severe artifact. Figure 8 shows x5 reconstructive magnifications of the central part of the abdominal scan (fig. 2A). It is important to distinguish between a limited field of view and a restricted area of reconstruction [8] which may be used to produce magnification images. No truncated-view artifact is produced as long as the magnified reconstruction is made from projections comprising complete views. Figure 8A shows a x 5 reconstructive magnification of the central part of the abdominal scan (fig. 2A) made from complete views and using the extended scale option. No artifact is present.

6 188 LEHR AJR:141, July 1983 Fig. 6.-Limited-field-of-view reconstructions for abdominal scan shown in fig. 2A. A. Reconstruction for 48-cm-diam field of view indicated by inner circle. Graph shows CT values of artifact image (not shown, but formed by subtracting fig. 2A from fig. 5A) along dashed line extending from upper left to lower right part of gantry aperture (outer circle). Artifact appears as increased density near upper left edge of field of view where patient contour extends beyond field (window level 0, window width 51 2). B, Reconstruction for 24-cm-diam field of view. Graph again shows magnitude of artifact along line extending diagonally across field. Artifact is more severe with central error of 243 H and value exceeding H at periphery. Window center has been raised to compensate for artifact (window level 300, window width 512). c, Reconstruction for 1 8-cm-diam field of view. Graph shows very severe artifact of about 509 H at center which again exceeds H near edge of field (window level 700, window width 51 2). Figure 8B shows a x 5 reconstructive magnification made from truncated views corresponding to a limited field of view 1 2 cm in diameter. The graph of artifact values shows a central error of about H and a relatively wide peripheral rim in which the artifact exceeds 2048 H. The image has been displayed at the same window width as figures 8A and 2A, in an attempt to maintain contrast, but the window level has been raised to H to compensate for the central error. However, the cupping artifact is so severe that only structures in the central part of the image are visible. Figure 8C shows the same image as figure 8B displayed at a wider window of 2000 H. Although more of the structures are visible, overall image contrast is greatly reduced. Discussion The largest error due to truncated-view artifact will occur near the edge of the field of view. Indeed, this may be the only part of the artifact seen under normal viewing conditions. Although it is possible to conceal this edge effect by computing reconstructed values over a region smaller than the field of view, the artifact will still be present in the central part of the image as a relatively constant value added to the CT numbers. One result of this additive constant will be to decrease image contrast. Although in theory this can be compensated for to a large extent by raising the window level and narrowing the window width when the image is

7 AJR:141, July 1983 CT TRUNCATED-VIEW ARTIFACTS 189 C displayed, this maneuver will also increase the visual impact of any cupping effect. Another result of the truncated-view artifact will be to frustrate attempts to perform quantitative analyses, because the CT number will depend not only on the absorption properties at the point represented, but also on the density, composition, and location of those parts of the patient or of other objects that lie outside the field of view. There are many benefits that can be achieved by using a limited field of view [9]. Certainly in a very obese patient, a Fig. 7.-Limited-field-of-view reconstructions for chest scan in fig. 2B. A, Reconstruction for 36-cm-diam field of view. Graph shows CT values along horizontal diameter indicated by dashed line in artifact image (not shown) formed by subtracting fig. 2B from fig. 7A. Artifact is primarily peripheral and in region of shoulders, which lie partially outside field of view (window level -300, window width 2000). B, Reconstruction for 18-cm-diam field of view. Graph shows more pronounced artifact with central error of about 200 H and value exceeding H at edge of field (window level , window width 2000). C, Reconstruction for 1 2-cm-diam field of view. Graph of artifact values shows central error greater than 500 H and 1 -cm-thick rim at periphery of field of view in which artifact exceeds H (window level 200, window width 2000). CT scan containing a truncated-view artifact would be more useful than no scan at all, and this may justify the use of a large gantry aperture. Furthermore, those parts of the patient that lie outside the x-ray beam in some of the views obviously will receive a lower dose. Finally, if the field of view is reduced by moving the x-ray tube closer to the patient while maintaining the same detected fan-beam angle (e.g., by moving the detector array an equal distance away from the center of rotation), the x-ray tube output can be used more efficiently, the sampling geometry will improve, and the ratio of scattered-to-primary photons will decrease [3]. Achieving these benefits without incurring artifacts is the goal of the correction techniques that have been proposed [4-7]. However, such techniques either are inexact or create additional technical problems in the construction of CT systems, and to my knowledge, no CT system that

8 190 LEHR AJR:141, July 1983 f &d*....:1 incorporates an accurate truncated-view correction procedure is available commercially. Because most CT systems currently on the market have a gantry aperture that is larger than the field of view, they are subject to truncated-view artifacts. Although this may be obvious from reviewing the system s technical specifications, it may be simpler to perform scans of reference objects (e.g., air or a water phantom) both with and without a syringe or a small bottle filled with diluted contrast material located at various distances from the center of the aperture. By subtracting the images formed without the object in the aperture from those with the object present, artifact images can be obtained and compared to those presented above (fig. 3). If a scanner is subject to truncated-view artifacts, the Fig. 8.-Reconstructive magnifications X 5 of central part of abdominal scan in fig. 2A. All reconstructions have been done using extended scale option which permits maximum CT value of A. Reconstruction from complete views. No artifact is present (window level 0, window width 51 2). B, x5 Reconstruction made from truncated views corresponding to 1 2-cm-diam field of view. Graph shows values of artifact along horizontal diameter indicated by dashed line (window level 1 200, window width 51 2). C, Same reconstruction and graph as B displayed at wider window setting to make more of structures visible. Note loss of image contrast (window level 1 600, window width 2048). artifacts can be avoided by carefully centering the patient in a large enough field of view to encompass both the patient and any other needed objects such as pads, blankets, restraining devices, drains, and intravenous tubing. If it is necessary to have some part of the patient, such as an extremity, or some foreign object inside the gantry aperture but outside the field of view, the magnitude of the artifact can be reduced by placing those objects as far from the center of rotation as possible (i.e., at the edge of the gantry aperture). When interpreting CT scans, the radiologist should watch carefully for increased Cl numbers near the outer edge of the reconstructed image, indicating the possibility that a truncated-view artifact is present. However, if the entire field of view is not reconstructed, the only indication of the

9 AJR:141, July 1983 Cl TRUNCATED-VIEW ARTIFACTS 191 artifact s presence may be a relatively uniform increase in the CT numbers which is difficult to detect visually. If a large enough area can be reconstructed to demonstrate that the entire patient is in the field of view, and if it is known that no other objects were in the gantry aperture during the scan, then the radiologist can be certain that no truncated-view artifact is present. Unless this is the case, radiologists should be quite cautious about basing any clinical opinion on quantitative analysis of CT numbers in the image. ACKNOWLEDGMENTS I thank Paul Davis who wrote the truncating program, and Jeffrey Crass, John Fennessy, Paul Frank, and Jonathan Rubin for review of the manuscript. REFERENCES 1. Herman GT. Image reconstructions from projections. New York: Academic, 1980:68 2. Stonestrom JP, Alvarez RE, Macovski A. A framework for spectral artifact connections in x-ray CT. IEEE Trans Biomed Eng 1981 ;28: Johns PC, Yaffe J. Scattered radiation in fan beam imaging systems. Med Phys I 982;9 : Herman GT, Lewitt AM. Evaluation of a preprocessing algonithm for truncated CT projections. J Comput Assist Tomogr 1981;5: 1 27-i Nalcioglu 0, Cho ZH, Lou RY. Limited file of view neconstnuction in computer tomography. IEEE Trans NucI Sci 1979;26 : Wagner W. Reconstructions from restricted region scan datanew means to reduce the patient dose. IEEE Trans NucI Sci 1 979;26 : Lewitt AM. Processing of incomplete measurement data in computed tomography. Med Phys 1979;6:4i Lehn JL, Davis P. CT magnification techniques. In: Proceedings of the seventh conference on computers in radiology. Chicago: American College of Radiology, 1982: , Zonneveld FW Jr. Computed tomography. Eindhoven, Netherlands: Philips Medical Systems, 1980:26

10 This article has been cited by: 1. J. Liebregts, T. Xi, R. Schreurs, B. van Loon, S. Bergé, T. Maal Three-dimensional virtual simulation of alar width changes following bimaxillary osteotomies. International Journal of Oral and Maxillofacial Surgery 45:10, [CrossRef] 2. Roya Zandparsa Latest Biomaterials and Technology in Dentistry. Dental Clinics of North America 58:1, [CrossRef] 3. Prashant P. Jaju, Mayuri Jain, Ajita Singh, Akanksha Gupta Artefacts in cone beam CT. Open Journal of Stomatology 03:05, [CrossRef] 4. AK Hunter, WD McDavid Characterization and correction of cupping effect artefacts in cone beam CT. Dentomaxillofacial Radiology 41:3, [CrossRef] 5. DAVID D. BARBEE, JENNIFER R. ALLEN, PATRICK R. GAVIN COMPUTED TOMOGRAPHY IN HORSES. Veterinary Radiology 28:5, [CrossRef]

diagnostic examination

diagnostic examination RADIOLOGICAL PHYSICS 2011 Raphex diagnostic examination Adel A. Mustafa, Ph.D., Editor PUBLISHED FOR: RAMPS (Radiological and Medical Physics Society of New York) preface The RAPHEX Diagnostic exam 2011

More information

Electronic Noise in CT Detectors: Impact on Image Noise and Artifacts

Electronic Noise in CT Detectors: Impact on Image Noise and Artifacts Medical Physics and Informatics Original Research Duan et al. Electronic Noise in CT Detectors Medical Physics and Informatics Original Research Xinhui Duan 1 Jia Wang 1,2 Shuai Leng 1 ernhard Schmidt

More information

HIGH RESOLUTION COMPUTERIZED TOMOGRAPHY SYSTEM USING AN IMAGING PLATE

HIGH RESOLUTION COMPUTERIZED TOMOGRAPHY SYSTEM USING AN IMAGING PLATE HIGH RESOLUTION COMPUTERIZED TOMOGRAPHY SYSTEM USING AN IMAGING PLATE Takeyuki Hashimoto 1), Morio Onoe 2), Hiroshi Nakamura 3), Tamon Inouye 4), Hiromichi Jumonji 5), Iwao Takahashi 6); 1)Yokohama Soei

More information

Detector technology in simultaneous spectral imaging

Detector technology in simultaneous spectral imaging Computed tomography Detector technology in simultaneous spectral imaging Philips IQon Spectral CT Z. Romman, I. Uman, Y. Yagil, D. Finzi, N. Wainer, D. Milstein; Philips Healthcare While CT has become

More information

HISTORY. CT Physics with an Emphasis on Application in Thoracic and Cardiac Imaging SUNDAY. Shawn D. Teague, MD

HISTORY. CT Physics with an Emphasis on Application in Thoracic and Cardiac Imaging SUNDAY. Shawn D. Teague, MD CT Physics with an Emphasis on Application in Thoracic and Cardiac Imaging Shawn D. Teague, MD DISCLOSURES 3DR- advisory committee CT PHYSICS WITH AN EMPHASIS ON APPLICATION IN THORACIC AND CARDIAC IMAGING

More information

Automated dose control in multi-slice CT. Nicholas Keat Formerly ImPACT, St George's Hospital, London

Automated dose control in multi-slice CT. Nicholas Keat Formerly ImPACT, St George's Hospital, London Automated dose control in multi-slice CT Nicholas Keat Formerly ImPACT, St George's Hospital, London Introduction to presentation CT contributes ~50+ % of all medical radiation dose Ideally all patients

More information

OPTIMIZING SPATIAL RESOLUTION WITH THE MECHANICAL DESIGN OF AN X-RAY COMPU1ED TOMOGRAPHY SCANNER

OPTIMIZING SPATIAL RESOLUTION WITH THE MECHANICAL DESIGN OF AN X-RAY COMPU1ED TOMOGRAPHY SCANNER OPTIMIZING SPATIAL RESOLUTION WITH THE MECHANICAL DESIGN OF AN X-RAY COMPU1ED TOMOGRAPHY SCANNER Lowell D. Harris, RichardT. Bernardi, Simon H. C. Hughes, and Robert E. Slocum Bio-Imaging Research, Inc.

More information

Pitfalls and Remedies of MDCT Scanners as Quantitative Instruments

Pitfalls and Remedies of MDCT Scanners as Quantitative Instruments intensity m(e) m (/cm) 000 00 0 0. 0 50 0 50 Pitfalls and Remedies of MDCT Scanners as Jiang Hsieh, PhD GE Healthcare Technology University of Wisconsin-Madison Root-Causes of CT Number Inaccuracies Nature

More information

COMPUTED TOMOGRAPHY 1

COMPUTED TOMOGRAPHY 1 COMPUTED TOMOGRAPHY 1 Why CT? Conventional X ray picture of a chest 2 Introduction Why CT? In a normal X-ray picture, most soft tissue doesn't show up clearly. To focus in on organs, or to examine the

More information

Radionuclide Imaging MII Single Photon Emission Computed Tomography (SPECT)

Radionuclide Imaging MII Single Photon Emission Computed Tomography (SPECT) Radionuclide Imaging MII 3073 Single Photon Emission Computed Tomography (SPECT) Single Photon Emission Computed Tomography (SPECT) The successful application of computer algorithms to x-ray imaging in

More information

Suppression of metal artifacts using image-based monoenergetic DECT imaging

Suppression of metal artifacts using image-based monoenergetic DECT imaging Suppression of metal artifacts using image-based monoenergetic DECT imaging Poster No.: C-0519 Congress: ECR 2011 Type: Scientific Paper Authors: B. Krauss, B. Schmidt, M. Sedlmair, T. Flohr; Forchheim/DE

More information

Related topics Beam hardening, cupping effect, Beam hardening correction, metal artefacts, photon starvation

Related topics Beam hardening, cupping effect, Beam hardening correction, metal artefacts, photon starvation Beam hardening and metal artefacts TEP Related topics Beam hardening, cupping effect, Beam hardening correction, metal artefacts, photon starvation Principle X-ray sources produce a polychromatic spectrum

More information

The effect of compensating filter on image quality in lateral projection of thoraco lumbar radiography

The effect of compensating filter on image quality in lateral projection of thoraco lumbar radiography Journal of Physics: Conference Series OPEN ACCESS The effect of compensating filter on image quality in lateral projection of thoraco lumbar radiography To cite this article: N A A Daud et al 2014 J. Phys.:

More information

QC Testing for Computed Tomography (CT) Scanner

QC Testing for Computed Tomography (CT) Scanner QC Testing for Computed Tomography (CT) Scanner QA - Quality Assurance All planned and systematic actions needed to provide confidence on a structure, system or component. all-encompassing program, including

More information

An Introduction to TG-142 Imaging QA Using Standard Imaging Products. Mark Wiesmeyer, PhD, DABR Technical Product Manager Standard Imaging, Inc.

An Introduction to TG-142 Imaging QA Using Standard Imaging Products. Mark Wiesmeyer, PhD, DABR Technical Product Manager Standard Imaging, Inc. An Introduction to TG-142 Imaging QA Using Standard Imaging Products Mark Wiesmeyer, PhD, DABR Technical Product Manager Standard Imaging, Inc. Goals Understand the nature and intent of TG 142 imaging

More information

Wide-Detector CT for TAVR Planning:

Wide-Detector CT for TAVR Planning: Wide-Detector CT for TAVR Planning: Impact on Iodine Dose, Radiation Dose, and Image Quality SCBTMR 2015 Annual Course Thursday, October 8 William P. Shuman MD FSCBTMR Department of Radiology University

More information

Image Quality and Dose. Image Quality and Dose. Image Quality and Dose Issues in MSCT. Scanner parameters affecting IQ and Dose

Image Quality and Dose. Image Quality and Dose. Image Quality and Dose Issues in MSCT. Scanner parameters affecting IQ and Dose Image Quality and Dose Issues in MSCT Image Quality and Dose Image quality Image noise Spatial resolution Contrast Artefacts Speckle and sharpness S. Edyvean St. George s Hospital London SW17 0QT Radiation

More information

Introduction. Chapter 16 Diagnostic Radiology. Primary radiological image. Primary radiological image

Introduction. Chapter 16 Diagnostic Radiology. Primary radiological image. Primary radiological image Introduction Chapter 16 Diagnostic Radiology Radiation Dosimetry I Text: H.E Johns and J.R. Cunningham, The physics of radiology, 4 th ed. http://www.utoledo.edu/med/depts/radther In diagnostic radiology

More information

I. PERFORMANCE OF X-RAY PRODUCTION COMPONENTS FLUOROSCOPIC ACCEPTANCE TESTING: TEST PROCEDURES & PERFORMANCE CRITERIA

I. PERFORMANCE OF X-RAY PRODUCTION COMPONENTS FLUOROSCOPIC ACCEPTANCE TESTING: TEST PROCEDURES & PERFORMANCE CRITERIA FLUOROSCOPIC ACCEPTANCE TESTING: TEST PROCEDURES & PERFORMANCE CRITERIA EDWARD L. NICKOLOFF DEPARTMENT OF RADIOLOGY COLUMBIA UNIVERSITY NEW YORK, NY ACCEPTANCE TESTING GOALS PRIOR TO 1st CLINICAL USAGE

More information

TOPICS: CT Protocol Optimization over the Range of Patient Age & Size and for Different CT Scanner Types: Recommendations & Misconceptions

TOPICS: CT Protocol Optimization over the Range of Patient Age & Size and for Different CT Scanner Types: Recommendations & Misconceptions CT Protocol Optimization over the Range of Patient Age & Size and for Different CT Scanner Types: Recommendations & Misconceptions TOPICS: Computed Tomography Quick Overview CT Dosimetry Effects of CT

More information

A Skew-Slit Collimator for Small-Animal SPECT

A Skew-Slit Collimator for Small-Animal SPECT A Skew-Slit Collimator for Small-Animal SPECT Gengsheng L. Zeng Department of Radiology, Utah Center for Advanced Imaging Research (UCAIR), University of Utah, Salt Lake City, Utah The main objective of

More information

An Activity in Computed Tomography

An Activity in Computed Tomography Pre-lab Discussion An Activity in Computed Tomography X-rays X-rays are high energy electromagnetic radiation with wavelengths smaller than those in the visible spectrum (0.01-10nm and 4000-800nm respectively).

More information

QC by the MPE in Belgium

QC by the MPE in Belgium Acceptance testing of state-of-the-art CT scanners using a new national protocol: first experience on a large number of scanners of different make and model the working group Radiology of the Belgian Hospital

More information

CHAPTER 2 COMMISSIONING OF KILO-VOLTAGE CONE BEAM COMPUTED TOMOGRAPHY FOR IMAGE-GUIDED RADIOTHERAPY

CHAPTER 2 COMMISSIONING OF KILO-VOLTAGE CONE BEAM COMPUTED TOMOGRAPHY FOR IMAGE-GUIDED RADIOTHERAPY 14 CHAPTER 2 COMMISSIONING OF KILO-VOLTAGE CONE BEAM COMPUTED TOMOGRAPHY FOR IMAGE-GUIDED RADIOTHERAPY 2.1 INTRODUCTION kv-cbct integrated with linear accelerators as a tool for IGRT, was developed to

More information

Dose Reduction and Image Preservation After the Introduction of a 0.1 mm Cu Filter into the LODOX Statscan unit above 110 kvp

Dose Reduction and Image Preservation After the Introduction of a 0.1 mm Cu Filter into the LODOX Statscan unit above 110 kvp Dose Reduction and Image Preservation After the Introduction of a into the LODOX Statscan unit above 110 kvp Abstract: CJ Trauernicht 1, C Rall 1, T Perks 2, G Maree 1, E Hering 1, S Steiner 3 1) Division

More information

Computed Tomography. The Fundamentals of... THE FUNDAMENTALS OF... Jason H. Launders, MSc. Current Technology

Computed Tomography. The Fundamentals of... THE FUNDAMENTALS OF... Jason H. Launders, MSc. Current Technology The Fundamentals of... Computed Tomography Computed Tomography (CT) systems use x-rays to produce images of slices through a patient s anatomy. Despite having lower spatial resolution than other x-ray

More information

An Activity in Computed Tomography

An Activity in Computed Tomography Pre-lab Discussion An Activity in Computed Tomography X-rays X-rays are high energy electromagnetic radiation with wavelengths smaller than those in the visible spectrum (0.01-10nm and 4000-800nm respectively).

More information

IBEX TECHNOLOGY APPLIED TO DIGITAL RADIOGRAPHY

IBEX TECHNOLOGY APPLIED TO DIGITAL RADIOGRAPHY WHITE PAPER: IBEX TECHNOLOGY APPLIED TO DIGITAL RADIOGRAPHY IBEX Innovations Ltd. Registered in England and Wales: 07208355 Address: Discovery 2, NETPark, William Armstrong Way, Sedgefield, UK Patents:

More information

Studies on reduction of exposure dose using digital scattered X-ray removal processing

Studies on reduction of exposure dose using digital scattered X-ray removal processing Studies on reduction of exposure dose using digital scattered X-ray removal processing Poster No.: C-1834 Congress: ECR 2015 Type: Scientific Exhibit Authors: K. Kashiyama, M. Funahashi, T. Nakaoka, T.

More information

Studies on reduction of exposure dose using digital scattered X-ray removal processing

Studies on reduction of exposure dose using digital scattered X-ray removal processing Studies on reduction of exposure dose using digital scattered X-ray removal processing Poster No.: C-1834 Congress: ECR 2015 Type: Scientific Exhibit Authors: K. Kashiyama, M. Funahashi, T. Nakaoka, T.

More information

SECTION I - CHAPTER 2 DIGITAL IMAGING PROCESSING CONCEPTS

SECTION I - CHAPTER 2 DIGITAL IMAGING PROCESSING CONCEPTS RADT 3463 - COMPUTERIZED IMAGING Section I: Chapter 2 RADT 3463 Computerized Imaging 1 SECTION I - CHAPTER 2 DIGITAL IMAGING PROCESSING CONCEPTS RADT 3463 COMPUTERIZED IMAGING Section I: Chapter 2 RADT

More information

PD233: Design of Biomedical Devices and Systems

PD233: Design of Biomedical Devices and Systems PD233: Design of Biomedical Devices and Systems (Lecture-8 Medical Imaging Systems) (Imaging Systems Basics, X-ray and CT) Dr. Manish Arora CPDM, IISc Course Website: http://cpdm.iisc.ac.in/utsaah/courses/

More information

Projection Data and Scanograms in X-ray Computed Tomography

Projection Data and Scanograms in X-ray Computed Tomography DICTA2002: Digital Image Computing Techniques and Applications, 21 22 January 2002, Melbourne, Australia 1 Projection Data and Scanograms in X-ray Computed Tomography Russell Horney Imants Svalbe Peter

More information

1. Patient size AEC. Large Patient High ma. Small Patient Low ma

1. Patient size AEC. Large Patient High ma. Small Patient Low ma Comparison of the function and performance of CT AEC systems CTUG meeting by Emily Field Trainee clinical scientist 14 th th Breakdown CT Automatic Exposure Control (AEC) Background Project Description

More information

Scatter Correction by Modulation of Primary Radiation in Industrial X-ray CT: Beam-hardening Effects and their Correction

Scatter Correction by Modulation of Primary Radiation in Industrial X-ray CT: Beam-hardening Effects and their Correction International Symposium on Digital Industrial Radiology and Computed Tomography - Mo.3.2 Scatter Correction by Modulation of Primary Radiation in Industrial X-ray CT: Beam-hardening Effects and their Correction

More information

Y11-DR Digital Radiography (DR) Image Quality

Y11-DR Digital Radiography (DR) Image Quality Y11-DR Digital Radiography (DR) Image Quality Image quality is stressed for all systems in Safety Code 35. In the relevant sections Health Canada s advice is the manufacturer s recommended test procedures

More information

12/21/2016. Siemens Medical Systems Research Agreement Philips Healthcare Research Agreement AAN and ASN Committees

12/21/2016. Siemens Medical Systems Research Agreement Philips Healthcare Research Agreement AAN and ASN Committees Joseph V. Fritz, PhD Nandor Pintor, MD Dent Neurologic Institute ASN 2017 Friday, January 20, 2017 Siemens Medical Systems Research Agreement Philips Healthcare Research Agreement AAN and ASN Committees

More information

NON-UNIFORM ATTENUATION CORRECTION USING SIMULTANEOUS TRANSMISSION AND EMISSION CONVERGING TOMOGRAPHY

NON-UNIFORM ATTENUATION CORRECTION USING SIMULTANEOUS TRANSMISSION AND EMISSION CONVERGING TOMOGRAPHY 1134 IEEE TRANSACTIONS ON NUCLEAR SCIENCE, VOL. 39, NO. 4,1992 NON-UNIFORM ATTENUATION CORRECTION USING SIMULTANEOUS TRANSMISSION AND EMISSION CONVERGING TOMOGRAPHY C-H Tung, G. T. Gullberg, G. L. Zeng,

More information

Nuclear Associates , , CT Head and Body Dose Phantom

Nuclear Associates , , CT Head and Body Dose Phantom Nuclear Associates 76-414,76-414-4150,76-415 CT Head and Body Dose Phantom Users Manual March 2005 Manual No. 76-414-1 Rev. 2 2004, 2005 Fluke Corporation, All rights reserved. Printed in U.S.A. All product

More information

PET/CT Instrumentation Basics

PET/CT Instrumentation Basics / Instrumentation Basics 1. Motivations for / imaging 2. What is a / Scanner 3. Typical Protocols 4. Attenuation Correction 5. Problems and Challenges with / 6. Examples Motivations for / Imaging Desire

More information

Your worldwide source of grids. What a difference a grid makes

Your worldwide source of grids. What a difference a grid makes Your worldwide source of grids What a difference a grid makes The right assembly to meet your every need Trust the experts A high degree of expertise is required to design and manufacture quality X-ray

More information

MC SIMULATION OF SCATTER INTENSITIES IN A CONE-BEAM CT SYSTEM EMPLOYING A 450 kv X-RAY TUBE

MC SIMULATION OF SCATTER INTENSITIES IN A CONE-BEAM CT SYSTEM EMPLOYING A 450 kv X-RAY TUBE MC SIMULATION OF SCATTER INTENSITIES IN A CONE-BEAM CT SYSTEM EMPLOYING A 450 kv X-RAY TUBE A. Miceli ab, R. Thierry a, A. Flisch a, U. Sennhauser a, F. Casali b a Empa - Swiss Federal Laboratories for

More information

RaySafe X2. Effortless measurements of X-ray

RaySafe X2. Effortless measurements of X-ray RaySafe X2 Effortless measurements of X-ray At your fingertips We ve grown accustomed to intuitive interactions with our devices. After all, it s not the device that s most important, but what you can

More information

TISSUE EQUIVALENT PHANTOMS FOR EVALUATING IN-PLANE TUBE CURRENT MODULATED CT DOSE AND IMAGE QUALITY

TISSUE EQUIVALENT PHANTOMS FOR EVALUATING IN-PLANE TUBE CURRENT MODULATED CT DOSE AND IMAGE QUALITY TISSUE EQUIVALENT PHANTOMS FOR EVALUATING IN-PLANE TUBE CURRENT MODULATED CT DOSE AND IMAGE QUALITY By RYAN F. FISHER A THESIS PRESENTED TO THE GRADUATE SCHOOL OF THE UNIVERSITY OF FLORIDA IN PARTIAL FULFILLMENT

More information

Maximum Performance, Minimum Space

Maximum Performance, Minimum Space TECHNOLOGY HISTORY For over 130 years, Toshiba has been a world leader in developing technology to improve the quality of life. Our 50,000 global patents demonstrate a long, rich history of leading innovation.

More information

Diamond X-ray Rocking Curve and Topograph Measurements at CHESS

Diamond X-ray Rocking Curve and Topograph Measurements at CHESS Diamond X-ray Rocking Curve and Topograph Measurements at CHESS G. Yang 1, R.T. Jones 2, F. Klein 3 1 Department of Physics and Astronomy, University of Glasgow, Glasgow, UK G12 8QQ. 2 University of Connecticut

More information

INTRODUCTION THIN LENSES. Introduction. given by the paraxial refraction equation derived last lecture: Thin lenses (19.1) = 1. Double-lens systems

INTRODUCTION THIN LENSES. Introduction. given by the paraxial refraction equation derived last lecture: Thin lenses (19.1) = 1. Double-lens systems Chapter 9 OPTICAL INSTRUMENTS Introduction Thin lenses Double-lens systems Aberrations Camera Human eye Compound microscope Summary INTRODUCTION Knowledge of geometrical optics, diffraction and interference,

More information

Nuclear Associates

Nuclear Associates Nuclear Associates 76-700 Digital Subtraction Angiography Phantom Users Manual March 2005 Manual No. 76-700-1 Rev. 2 2004, 2005 Fluke Corporation, All rights reserved. Printed in U.S.A. All product names

More information

CT Basics: Image Quality Module 6

CT Basics: Image Quality Module 6 Module 6 For educational and institutional use. This transcript is licensed for noncommercial, educational inhouse or online educational course use only in educational and corporate institutions. Any broadcast,

More information

Research Support. Dual-Source CT: What is it and How Do I Test it? Cynthia H. McCollough, Ph.D.

Research Support. Dual-Source CT: What is it and How Do I Test it? Cynthia H. McCollough, Ph.D. Dual-Source CT: What is it and How Do I Test it? Cynthia H. McCollough, Ph.D. CT Clinical Innovation Center Department of Radiology Mayo Clinic College of Medicine Rochester, MN Research Support National

More information

Ludlum Medical Physics

Ludlum Medical Physics Ludlum Medical Physics Medical Imaging Radiology QA Test Tools NEW LUDLUM PRODUCT LINE Medical Physics Products Medical Physics Products What are they? Products used to measure radiation output and to

More information

Distortion Correction in LODOX StatScan X-Ray Images

Distortion Correction in LODOX StatScan X-Ray Images Distortion Correction in LODOX StatScan X-Ray Images Matthew Paul Beets November 27, 2007 X-Ray images produced by the LODOX StatScan machine contain a non-linear distortion in the direction of the beam

More information

Acceptance Testing of a Digital Breast Tomosynthesis Unit

Acceptance Testing of a Digital Breast Tomosynthesis Unit Acceptance Testing of a Digital Breast Tomosynthesis Unit 2012 AAPM Spring Clinical Meeting Jessica Clements, M.S., DABR Objectives Review of technology and clinical advantages Acceptance Testing Procedures

More information

Geometric image distortion in flat-panel X-ray detectors and its influence on the accuracy of CT-based dimensional measurements

Geometric image distortion in flat-panel X-ray detectors and its influence on the accuracy of CT-based dimensional measurements Geometric image distortion in flat-panel X-ray detectors and its influence on the accuracy of CT-based dimensional measurements Daniel Weiß, Ronald Lonardoni, Andreas Deffner, Christoph Kuhn Carl Zeiss

More information

Breast Tomosynthesis. Bob Liu, Ph.D. Department of Radiology Massachusetts General Hospital And Harvard Medical School

Breast Tomosynthesis. Bob Liu, Ph.D. Department of Radiology Massachusetts General Hospital And Harvard Medical School Breast Tomosynthesis Bob Liu, Ph.D. Department of Radiology Massachusetts General Hospital And Harvard Medical School Outline Physics aspects of breast tomosynthesis Quality control of breast tomosynthesis

More information

160-slice CT SCANNER / New Standard for the Future

160-slice CT SCANNER / New Standard for the Future TECHNOLOGY HISTORY For over 130 years, Toshiba has been a world leader in developing technology to improve the quality of life. Our 50,000 global patents demonstrate a long, rich history of leading innovation.

More information

X-ray Imaging. PHYS Lecture. Carlos Vinhais. Departamento de Física Instituto Superior de Engenharia do Porto

X-ray Imaging. PHYS Lecture. Carlos Vinhais. Departamento de Física Instituto Superior de Engenharia do Porto X-ray Imaging PHYS Lecture Carlos Vinhais Departamento de Física Instituto Superior de Engenharia do Porto cav@isep.ipp.pt Overview Projection Radiography Anode Angle Focal Spot Magnification Blurring

More information

LSO PET/CT Pico Performance Improvements with Ultra Hi-Rez Option

LSO PET/CT Pico Performance Improvements with Ultra Hi-Rez Option LSO PET/CT Pico Performance Improvements with Ultra Hi-Rez Option Y. Bercier, Member, IEEE, M. Casey, Member, IEEE, J. Young, Member, IEEE, T. Wheelock, Member, IEEE, T. Gremillion Abstract-- Factors which

More information

X-RAY BACKSCATTER IMAGING: PHOTOGRAPHY THROUGH BARRIERS

X-RAY BACKSCATTER IMAGING: PHOTOGRAPHY THROUGH BARRIERS Copyright JCPDS-International Centre for Diffraction Data 2006 ISSN 1097-0002 X-RAY BACKSCATTER IMAGING: PHOTOGRAPHY THROUGH BARRIERS 13 Joseph Callerame American Science & Engineering, Inc. 829 Middlesex

More information

CT Basics: Data Acquisition Module 3

CT Basics: Data Acquisition Module 3 Module 3 Transcript For educational and institutional use. This transcript is licensed for noncommercial, educational inhouse or online educational course use only in educational and corporate institutions.

More information

Development of new dosimeter for measuring dose distribution in CT

Development of new dosimeter for measuring dose distribution in CT Development of new dosimeter for measuring dose distribution in CT Poster No.: C-2925 Congress: ECR 2010 Type: Scientific Exhibit Topic: Physics in Radiology - Without Subtopic Authors: Y. Muramatsu, K.

More information

BASICS OF FLUOROSCOPY

BASICS OF FLUOROSCOPY Medical Physics Residents Training Program BASICS OF FLUOROSCOPY Dr. Khalid Alyousef, PhD Department of Medical Imaging King Abdulaziz Medical City- Riyadh Edison examining the hand of Clarence Dally with

More information

Application Note (A11)

Application Note (A11) Application Note (A11) Slit and Aperture Selection in Spectroradiometry REVISION: C August 2013 Gooch & Housego 4632 36 th Street, Orlando, FL 32811 Tel: 1 407 422 3171 Fax: 1 407 648 5412 Email: sales@goochandhousego.com

More information

I. Introduction.

I. Introduction. JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, VOLUME 15, NUMBER 1, 2014 Accuracy of measuring half- and quarter-value layers and appropriate aperture width of a convenient method using a lead-covered case

More information

Automated Detection of Early Lung Cancer and Tuberculosis Based on X- Ray Image Analysis

Automated Detection of Early Lung Cancer and Tuberculosis Based on X- Ray Image Analysis Proceedings of the 6th WSEAS International Conference on Signal, Speech and Image Processing, Lisbon, Portugal, September 22-24, 2006 110 Automated Detection of Early Lung Cancer and Tuberculosis Based

More information

Investigation of the line-pair pattern method for evaluating mammographic focal spot performance

Investigation of the line-pair pattern method for evaluating mammographic focal spot performance Investigation of the line-pair pattern method for evaluating mammographic focal spot performance Mitchell M. Goodsitt, a) Heang-Ping Chan, and Bob Liu Department of Radiology, University of Michigan, Ann

More information

Nuclear Associates , &

Nuclear Associates , & Nuclear Associates 76-823, 76-824 & 76-825 PET/SPECT Phantom Source Tank, Phantom Inserts and Cardiac Insert Users Manual March 2005 Manual No. 76-823-1 Rev. 2 2004, 2005 Fluke Corporation, All rights

More information

QUANTITATIVE COMPUTERIZED LAMINOGRAPHY. Suzanne Fox Buchele and Hunter Ellinger

QUANTITATIVE COMPUTERIZED LAMINOGRAPHY. Suzanne Fox Buchele and Hunter Ellinger QUANTITATIVE COMPUTERIZED LAMINOGRAPHY Suzanne Fox Buchele and Hunter Ellinger Scientific Measurement Systems, Inc. 2201 Donley Drive Austin, Texas 78758 INTRODUCTION Industrial computerized-tomography

More information

Exposure Indices and Target Values in Radiography: What Are They and How Can You Use Them?

Exposure Indices and Target Values in Radiography: What Are They and How Can You Use Them? Exposure Indices and Target Values in Radiography: What Are They and How Can You Use Them? Definition and Validation of Exposure Indices Ingrid Reiser, PhD DABR Department of Radiology University of Chicago

More information

C a t p h a n. T h e P h a n t o m L a b o r a t o r y. Ordering Information

C a t p h a n. T h e P h a n t o m L a b o r a t o r y. Ordering Information Ordering Information Please contact us if you have any questions or if you would like a quote or delivery schedule regarding the Catphan phantom. phone 800-525-1190, or 518-692-1190 fax 518-692-3329 mail

More information

Computerized Medical Imaging and Graphics

Computerized Medical Imaging and Graphics Computerized Medical Imaging and Graphics 36 (2012) 387 395 Contents lists available at SciVerse ScienceDirect Computerized Medical Imaging and Graphics jo ur n al homep age : www.elsevier.com/locate/compmedimag

More information

Medical Imaging. X-rays, CT/CAT scans, Ultrasound, Magnetic Resonance Imaging

Medical Imaging. X-rays, CT/CAT scans, Ultrasound, Magnetic Resonance Imaging Medical Imaging X-rays, CT/CAT scans, Ultrasound, Magnetic Resonance Imaging From: Physics for the IB Diploma Coursebook 6th Edition by Tsokos, Hoeben and Headlee And Higher Level Physics 2 nd Edition

More information

X-rays in medical diagnostics

X-rays in medical diagnostics X-rays in medical diagnostics S.Dolanski Babić 2017/18. History W.C.Röntgen (1845-1923) discovered a new type of radiation Nature, Jan. 23. 1896.; Science, Feb.14. 1896. X- rays: Induced the ionization

More information

Philip Sperling. Sales Science and New Materials, YXLON International GmbH, Essener Bogen 15, Hamburg, Germany.

Philip Sperling. Sales Science and New Materials, YXLON International GmbH, Essener Bogen 15, Hamburg, Germany. A new generation of x-ray computed tomography devices for quality inspection and metrology inspection in the field of additive manufacturing and other sciences Philip Sperling Sales Science and New Materials,

More information

Towards accurate measurements with synchrotron tomography Problems and pitfalls. Robert C. Atwood. Nghia T. Vo, Michael Drakopoulos, Thomas Connolley

Towards accurate measurements with synchrotron tomography Problems and pitfalls. Robert C. Atwood. Nghia T. Vo, Michael Drakopoulos, Thomas Connolley Towards accurate measurements with synchrotron tomography Problems and pitfalls Robert C. Atwood Nghia T. Vo, Michael Drakopoulos, Thomas Connolley Artefacts in Synchrotron X-ray Tomography Rings Rings

More information

10.2 Images Formed by Lenses SUMMARY. Refraction in Lenses. Section 10.1 Questions

10.2 Images Formed by Lenses SUMMARY. Refraction in Lenses. Section 10.1 Questions 10.2 SUMMARY Refraction in Lenses Converging lenses bring parallel rays together after they are refracted. Diverging lenses cause parallel rays to move apart after they are refracted. Rays are refracted

More information

Test Equipment for Radiology and CT Quality Control Contents

Test Equipment for Radiology and CT Quality Control Contents Test Equipment for Radiology and CT Quality Control Contents Quality Control Testing...2 Photometers for Digital Clinical Display QC...3 Primary Workstations...3 Secondary Workstations...3 Testing of workstations...3

More information

RECOMMENDATION ITU-R F *

RECOMMENDATION ITU-R F * Rec. ITU-R F.699-6 1 RECOMMENATION ITU-R F.699-6 * Reference radiation patterns for fixed wireless system antennas for use in coordination studies and interference assessment in the frequency range from

More information

Beam-Restricting Devices

Beam-Restricting Devices Beam-Restricting Devices Three factors contribute to an increase in scatter radiation: Increased kvp Increased Field Size Increased Patient or Body Part Size. X-ray Interactions a some interact with the

More information

Radiographic sensitivity improved by optimized high resolution X -ray detector design.

Radiographic sensitivity improved by optimized high resolution X -ray detector design. DIR 2007 - International Symposium on Digital industrial Radiology and Computed Tomography, June 25-27, 2007, Lyon, France Radiographic sensitivity improved by optimized high resolution X -ray detector

More information

Lunar Technology Advantages

Lunar Technology Advantages Lunar Technology Advantages DXA stands for Dual-Energy X-ray Absorptiometry. It is a measurement method that uses the differences in the absorption of high energy and low energy X-ray photons by different

More information

Phase Imaging Using Focused Polycapillary Optics

Phase Imaging Using Focused Polycapillary Optics Phase Imaging Using Focused Polycapillary Optics Sajid Bashir, Sajjad Tahir, Jonathan C. Petruccelli, C.A. MacDonald Dept. of Physics, University at Albany, Albany, New York Abstract Contrast in conventional

More information

Geometric Dimensioning and Tolerancing

Geometric Dimensioning and Tolerancing Geometric Dimensioning and Tolerancing (Known as GDT) What is GDT Helps ensure interchangeability of parts. Use is dictated by function and relationship of the part feature. It does not take the place

More information

CT Data Storage Reduction by Means of Compressing Projection Data Instead of Images: Feasibility Study 1

CT Data Storage Reduction by Means of Compressing Projection Data Instead of Images: Feasibility Study 1 Kyongtae T. Bae, MD, PhD Bruce R. Whiting, PhD Index terms: Computed tomography (CT), image display and recording Computed tomography (CT), image processing Computed tomography (CT), technology Data compression

More information

MULTIPLE SENSORS LENSLETS FOR SECURE DOCUMENT SCANNERS

MULTIPLE SENSORS LENSLETS FOR SECURE DOCUMENT SCANNERS INFOTEH-JAHORINA Vol. 10, Ref. E-VI-11, p. 892-896, March 2011. MULTIPLE SENSORS LENSLETS FOR SECURE DOCUMENT SCANNERS Jelena Cvetković, Aleksej Makarov, Sasa Vujić, Vlatacom d.o.o. Beograd Abstract -

More information

Reducing Radiation Exposure from Survey CT Scans

Reducing Radiation Exposure from Survey CT Scans Reducing Survey CT Scan Exposure Pediatric Imaging Original Research Jennifer C. O Daniel 1 Donna M. Stevens 2 Dianna D. Cody 2 O Daniel JC, Stevens DM, Cody DD Received July 28, 2004; accepted after revision

More information

2 nd generation TOMOSYNTHESIS

2 nd generation TOMOSYNTHESIS 2 nd generation TOMOSYNTHESIS 2 nd generation DBT true innovation in breast imaging synthesis graphy Combo mode Stereotactic Biopsy Works in progress: Advanced Technology, simplicity and ergonomics Raffaello

More information

Real-Time Scanning Goniometric Radiometer for Rapid Characterization of Laser Diodes and VCSELs

Real-Time Scanning Goniometric Radiometer for Rapid Characterization of Laser Diodes and VCSELs Real-Time Scanning Goniometric Radiometer for Rapid Characterization of Laser Diodes and VCSELs Jeffrey L. Guttman, John M. Fleischer, and Allen M. Cary Photon, Inc. 6860 Santa Teresa Blvd., San Jose,

More information

Nuclear Associates and

Nuclear Associates and Nuclear Associates 76-410-4130 and 76-411 AAPM CT Performance Phantom Users Manual March 2005 Manual No. 76-410-4130-1 Rev. 2 2004, 2005 Fluke Corporation, All rights reserved. Printed in U.S.A. All product

More information

Master of Science Thesis. SIMIND Based Pinhole Imaging

Master of Science Thesis. SIMIND Based Pinhole Imaging Master of Science Thesis SIMIND Based Pinhole Imaging * Development and Validation Kurt Sundin Supervisor: Michael Ljungberg, PhD Medical Radiation Physics Clinical Sciences, Lund Lund University, 2006

More information

CHAPTER 8: EXTENDED TETRACHORD CLASSIFICATION

CHAPTER 8: EXTENDED TETRACHORD CLASSIFICATION CHAPTER 8: EXTENDED TETRACHORD CLASSIFICATION Chapter 7 introduced the notion of strange circles: using various circles of musical intervals as equivalence classes to which input pitch-classes are assigned.

More information

Opto Engineering S.r.l.

Opto Engineering S.r.l. TUTORIAL #1 Telecentric Lenses: basic information and working principles On line dimensional control is one of the most challenging and difficult applications of vision systems. On the other hand, besides

More information

Virtual monochromatic imaging in dual-source and dual-energy CT for visualization of acute ischemic stroke

Virtual monochromatic imaging in dual-source and dual-energy CT for visualization of acute ischemic stroke Virtual monochromatic imaging in dual-source and dual-energy CT for visualization of acute ischemic stroke Hidetake Hara and Hiroshi Muraishi School of Allied Health Sciences, Kitasato University, Japan

More information

NeuViz 16 Computed Tomography. Elevating routine imaging for exceptional results

NeuViz 16 Computed Tomography. Elevating routine imaging for exceptional results NeuViz 16 Computed Tomography Elevating routine imaging for exceptional results Essence NeuViz 16 Raising the bar on clinical utility in routine imaging. Get more. More clinical information for patients.

More information

Iterative Reconstruction in Image Space. Answers for life.

Iterative Reconstruction in Image Space. Answers for life. Iterative Reconstruction in Image Space Answers for life. Iterative Reconstruction in Image Space * (IRIS) * Please note: IRIS is used as an abbreviation for Iterative Reconstruction in Image Space throughout

More information

Thermionic x-ray. Alternative technologies. Electron Field Emission. CNT Based Field Emission X-Ray Source

Thermionic x-ray. Alternative technologies. Electron Field Emission. CNT Based Field Emission X-Ray Source Energy Level (ev) Multi-beam x-ray source array based on carbon nanotube field emission O. Zhou, JP Lu, X. Calderon-Colon, X. Qian, G. Yang, G. Cao, E. Gidcumb, A. Tucker, J. Shan University of North Carolina

More information

SYLLABUS. 1. Identification of Subject:

SYLLABUS. 1. Identification of Subject: SYLLABUS Date/ Revision : 30 January 2017/1 Faculty : Life Sciences Approval : Dean, Faculty of Life Sciences SUBJECT : Biophysics 1. Identification of Subject: Name of Subject : Biophysics Code of Subject

More information

Optimization of Energy Modulation Filter for Dual Energy CBCT Using Geant4 Monte-Carlo Simulation

Optimization of Energy Modulation Filter for Dual Energy CBCT Using Geant4 Monte-Carlo Simulation Original Article PROGRESS in MEDICAL PHYSICS 27(3), Sept. 2016 http://dx.doi.org/10.14316/pmp.2016.27.3.125 pissn 2508-4445, eissn 2508-4453 Optimization of Energy Modulation Filter for Dual Energy CBCT

More information

Four-dimensional Computed Tomography (4D CT) Concepts and Preliminary Development

Four-dimensional Computed Tomography (4D CT) Concepts and Preliminary Development ORIGINAL ARTICLE ORIGINAL ARTICLE Radiation Medicine: Vol. 21 No. 1, 17 22 p.p., 2003 Four-dimensional Computed Tomography (4D CT) Concepts and Preliminary Development Masahiro Endo,* Takanori Tsunoo,*

More information

A comparison of two methods for the determination of freein-air geometric efficiency in MDCT

A comparison of two methods for the determination of freein-air geometric efficiency in MDCT A comparison of two methods for the determination of freein-air geometric efficiency in MDCT Theocharis Berris *1, Kostas Perisinakis 1,, Antonios E. Papadakis and John Damilakis 1, 1 Department of Medical

More information