Wide-Detector CT for TAVR Planning:

Size: px
Start display at page:

Download "Wide-Detector CT for TAVR Planning:"

Transcription

1 Wide-Detector CT for TAVR Planning: Impact on Iodine Dose, Radiation Dose, and Image Quality SCBTMR 2015 Annual Course Thursday, October 8 William P. Shuman MD FSCBTMR Department of Radiology University of Washington, Seattle

2 Disclosures WPS administers research grants from GE Healthcare which support clinical CT research

3 Wide Detector Technology 16 cm detector with focally aligned detector modules Post patient collimator

4 Detector Technology 16 cm Wide-Detector

5 Mitigating Wide Coverage Issues cone beam Focally aligned detector modules designed for cone beam geometry VHD reconstruction reduces Hounsfield unit uniformity variation across 160 mm z-axis coverage

6 Post-Patient Collimator

7

8 Wide Detector Technology Axial step-and-shoot technique with 16 cm slabs No helical artifacts Stitching reconstruction of anatomical structures to control or eliminate step-off artifacts

9

10 Wide Detector Technology Faster rotation time (0.28 sec) Better ACTM and auto-kvp Partial model-based iterative reconstruction (ASIR-V)

11

12

13 TAVR Transcutaneous Aortic Valve Replacement Patients with severe calcific AS and AI who are not operative candidates due to co-morbidities Femoral, transapical,or trans axillary approaches Implanted ring/valve overlies native valve, displacing leaflets

14 TAVR Transcutaneous Aortic Valve Replacement Patients with severe calcific AS and AI who are not operative candidates due to co-morbidities Femoral, transapical, or trans axillary approaches Implanted ring/valve overlies native valve, displacing leaflets

15

16

17

18

19

20

21

22

23

24

25 Background Wide-detector CT is faster, enabling scanning of chestabdomen-pelvis as a single acquisition with one bolus of contrast Wide Detector CT also has new radiation dose reduction features

26 Purpose To compare an older clinical standard TAVR planning protocl on a 4 cm detector CT with a new clinical standard protocol on a 16 cm wide-detector CT for: Iodine dose utilized Radiation dose Subjective and objective image quality

27 Methods IRB approved retrospective study December 2014 January 2015 Large urban academic medical center

28 Methods Patients Reviewed 36 consecutive patients scanned for TAVR planning on wide-detector Carefully match for gender, age, and BMI to 36 patients previously scanned for TAVR planning on 4 cm detector

29 Methods Wide-detector Protocol 1 Continuous scanning of ECG gate chest and non-gated abdomen-pelvis cm slabs in the chest, cm slabs in the abdomen-pelvis Scan duration: seconds

30 Methods Wide-detector Protocol 1 NI = 30 for chest, 18 for abd/pel, kvp = rotation 0.28 or 0.35 sec Contrast: cc at 3-5 cc/sec base on Z axis Injection duration 5 sec longer than scan duration Reconstructed with ASIR-V 70%

31 Methods Standard 4cm Detector Protocol 2 Separate acquisitions of the chest (helical ECG gated) and of the abdomen/pelvis (helical non-gated) 30 second pause between the two acquisitions Scan duration: 47 seconds

32 Methods Standard 4 cm detector Protocol 2 NI = 50 for chest, 30 for abd/pel, kvp = rotation time = 0.35 or 0.5 sec Contrast: cc at 4 cc/sec Reconstructed with 70% ASIR

33

34 Methods Images reconstructed: Axial 2.5 mm of the chest, abdomen, and pelvis mm double oblique of aortic valve mm magnified - with measurements of annulus, leaflet length, and coronary ostia distances.

35 Protocol 1 Protocol 2 A B C D

36 Methods Image Evaluation by 2 independent Reviewers (subjective 4 point Likert scales): aortic enhancement overall image quality for non-vascular structures image quality of aortic annulus

37 Methods Objective Measurements: HU vascular attenuation at 8 locations in aorta and iliofemorals Psoas HU and Image Noise in air Calculated SNR, CNR

38 Results Demographics No difference in age, gender, or BMI

39

40 Results Wide detector scanner acquisition time averaged 32 seconds less than 4 cm detector scanner

41 Results Wide detector scanner iodine dose averaged 33% less than standard 4 cm detector scanner

42 Results Wide detector scanner radiation dose in SSDE was 28 45% lower than standard 4cm detector scanner 53% reduction in total exam DLP with no statistical difference in scan range (mm)

43 Results Wide detector scanner mean vascular HU attenuation was statistically greater overall and was greater at each of the 8 measurement levels than standard 4cm detector scanner

44

45 Results Image noise, SNR, and CNR were the same for both scanners

46

47 Results Subjective scores for vascular enhancement and overall image quality were the same Subjective image quality scores for the magnified aortic root measurement images were statistically better for the wide detector scanner

48

49 Discussion Wide detector scanner features: Axial step-and-shoot technique No helical artifacts Whole heart in one beat Effective Stitching = no step-off artifacts

50 Discussion Wide detector scanner features: Focally aligned detector modules Less cone beam artifact Better dose efficiency

51 Discussion Wide detector scanner features: Three dimensional post-patient collimator Eliminates scatter radiation Less image noise

52 Discussion Wide detector scanner features: Faster rotation time 0.28 sec Better temporal resolution

53 Discussion Wide detector scanner features: Faster overall scan acquisition time Transition from ECG gated to ungated shortened from 30 secs. down to 5 secs. Enables single contrast bolus technique

54 Discussion Wide detector scanner features: ASIR-V reconstruction Partial model based iterative recon. 30% lower noise than statistical iterative recon Supports reduced radiation technique Similar reconstruction time

55 Conclusion Wide detector axial step-and-shoot TAVR planning CT resulted in: Greater vascular attenuation Similar subjective and objective image noise Lower radiation dose Less iodine dose

56

Pitfalls and Remedies of MDCT Scanners as Quantitative Instruments

Pitfalls and Remedies of MDCT Scanners as Quantitative Instruments intensity m(e) m (/cm) 000 00 0 0. 0 50 0 50 Pitfalls and Remedies of MDCT Scanners as Jiang Hsieh, PhD GE Healthcare Technology University of Wisconsin-Madison Root-Causes of CT Number Inaccuracies Nature

More information

12/21/2016. Siemens Medical Systems Research Agreement Philips Healthcare Research Agreement AAN and ASN Committees

12/21/2016. Siemens Medical Systems Research Agreement Philips Healthcare Research Agreement AAN and ASN Committees Joseph V. Fritz, PhD Nandor Pintor, MD Dent Neurologic Institute ASN 2017 Friday, January 20, 2017 Siemens Medical Systems Research Agreement Philips Healthcare Research Agreement AAN and ASN Committees

More information

Software and Hardware in CCTA. Elly Castellano PhD

Software and Hardware in CCTA. Elly Castellano PhD Software and Hardware in CCTA Elly Castellano PhD Outline technical requirements for coronary CTA the modern cardiac CT scanner ECG-gating technology image reconstruction algorithms 2 Technical requirements

More information

HISTORY. CT Physics with an Emphasis on Application in Thoracic and Cardiac Imaging SUNDAY. Shawn D. Teague, MD

HISTORY. CT Physics with an Emphasis on Application in Thoracic and Cardiac Imaging SUNDAY. Shawn D. Teague, MD CT Physics with an Emphasis on Application in Thoracic and Cardiac Imaging Shawn D. Teague, MD DISCLOSURES 3DR- advisory committee CT PHYSICS WITH AN EMPHASIS ON APPLICATION IN THORACIC AND CARDIAC IMAGING

More information

Research Support. Dual-Source CT: What is it and How Do I Test it? Cynthia H. McCollough, Ph.D.

Research Support. Dual-Source CT: What is it and How Do I Test it? Cynthia H. McCollough, Ph.D. Dual-Source CT: What is it and How Do I Test it? Cynthia H. McCollough, Ph.D. CT Clinical Innovation Center Department of Radiology Mayo Clinic College of Medicine Rochester, MN Research Support National

More information

Detector technology in simultaneous spectral imaging

Detector technology in simultaneous spectral imaging Computed tomography Detector technology in simultaneous spectral imaging Philips IQon Spectral CT Z. Romman, I. Uman, Y. Yagil, D. Finzi, N. Wainer, D. Milstein; Philips Healthcare While CT has become

More information

TOPICS: CT Protocol Optimization over the Range of Patient Age & Size and for Different CT Scanner Types: Recommendations & Misconceptions

TOPICS: CT Protocol Optimization over the Range of Patient Age & Size and for Different CT Scanner Types: Recommendations & Misconceptions CT Protocol Optimization over the Range of Patient Age & Size and for Different CT Scanner Types: Recommendations & Misconceptions TOPICS: Computed Tomography Quick Overview CT Dosimetry Effects of CT

More information

1. Patient size AEC. Large Patient High ma. Small Patient Low ma

1. Patient size AEC. Large Patient High ma. Small Patient Low ma Comparison of the function and performance of CT AEC systems CTUG meeting by Emily Field Trainee clinical scientist 14 th th Breakdown CT Automatic Exposure Control (AEC) Background Project Description

More information

Automated dose control in multi-slice CT. Nicholas Keat Formerly ImPACT, St George's Hospital, London

Automated dose control in multi-slice CT. Nicholas Keat Formerly ImPACT, St George's Hospital, London Automated dose control in multi-slice CT Nicholas Keat Formerly ImPACT, St George's Hospital, London Introduction to presentation CT contributes ~50+ % of all medical radiation dose Ideally all patients

More information

COMPUTED TOMOGRAPHY 1

COMPUTED TOMOGRAPHY 1 COMPUTED TOMOGRAPHY 1 Why CT? Conventional X ray picture of a chest 2 Introduction Why CT? In a normal X-ray picture, most soft tissue doesn't show up clearly. To focus in on organs, or to examine the

More information

Breast Tomosynthesis. Bob Liu, Ph.D. Department of Radiology Massachusetts General Hospital And Harvard Medical School

Breast Tomosynthesis. Bob Liu, Ph.D. Department of Radiology Massachusetts General Hospital And Harvard Medical School Breast Tomosynthesis Bob Liu, Ph.D. Department of Radiology Massachusetts General Hospital And Harvard Medical School Outline Physics aspects of breast tomosynthesis Quality control of breast tomosynthesis

More information

Iterative Reconstruction in Image Space. Answers for life.

Iterative Reconstruction in Image Space. Answers for life. Iterative Reconstruction in Image Space Answers for life. Iterative Reconstruction in Image Space * (IRIS) * Please note: IRIS is used as an abbreviation for Iterative Reconstruction in Image Space throughout

More information

160-slice CT SCANNER / New Standard for the Future

160-slice CT SCANNER / New Standard for the Future TECHNOLOGY HISTORY For over 130 years, Toshiba has been a world leader in developing technology to improve the quality of life. Our 50,000 global patents demonstrate a long, rich history of leading innovation.

More information

Iterative Reconstruction

Iterative Reconstruction RECENT ADVANCES IN CT RADIATION DOSE REDUCTION TECHNIQUES Iterative Reconstruction Kalpana Kanal, PhD, FSCBTMR, FACR, FAAPM Professor and Director, Diagnostic Physics Section University of Washington Seattle,

More information

Authors: Cabral, Ricardo 1 ; Carvoeiras, Pedro 2 ; Fatana, João 2, ; Alves, Rita 1. 1 Centro Hospitalar Lisboa Norte - Hospital de Santa Maria; 2

Authors: Cabral, Ricardo 1 ; Carvoeiras, Pedro 2 ; Fatana, João 2, ; Alves, Rita 1. 1 Centro Hospitalar Lisboa Norte - Hospital de Santa Maria; 2 Authors: Cabral, Ricardo 1 ; Carvoeiras, Pedro 2 ; Fatana, João 2, ; Alves, Rita 1. 1 Centro Hospitalar Lisboa Norte - Hospital de Santa Maria; 2 Medical Consult, SA; Establish a method to correlate image

More information

PET: New Technologies & Applications, Including Oncology

PET: New Technologies & Applications, Including Oncology PET: New Technologies & Applications, Including Oncology, PhD, FIEEE Imaging Research Laboratory Department of Radiology University of Washington, Seattle, WA Disclosures Research Contract, GE Healthcare

More information

diagnostic examination

diagnostic examination RADIOLOGICAL PHYSICS 2011 Raphex diagnostic examination Adel A. Mustafa, Ph.D., Editor PUBLISHED FOR: RAMPS (Radiological and Medical Physics Society of New York) preface The RAPHEX Diagnostic exam 2011

More information

Redefining Ergonomics

Redefining Ergonomics Samsung Electronics Co., Ltd. inspires the world and shapes the future with transformative ideas and technologies, redefining the worlds of TVs, smartphones, wearable devices, tablets, cameras, digital

More information

1. Queries are issued to the image archive for information about computed tomographic (CT)

1. Queries are issued to the image archive for information about computed tomographic (CT) Appendix E1 Exposure Extraction Method examinations. 1. Queries are issued to the image archive for information about computed tomographic (CT) 2. Potential dose report screen captures (hereafter, dose

More information

TORNIER BLUEPRINT. 3D Planning + PSI SCAN PROTOCOL

TORNIER BLUEPRINT. 3D Planning + PSI SCAN PROTOCOL TORNIER BLUEPRINT 3D Planning + PSI SCAN PROTOCOL Contents 3 Introduction 3 Patient preparation 3 Scanning instructions 4 Image instructions 5 Scanning parameters 6 Technical instructions 2 BLUEPRINT 3D

More information

Computed Tomography. The Fundamentals of... THE FUNDAMENTALS OF... Jason H. Launders, MSc. Current Technology

Computed Tomography. The Fundamentals of... THE FUNDAMENTALS OF... Jason H. Launders, MSc. Current Technology The Fundamentals of... Computed Tomography Computed Tomography (CT) systems use x-rays to produce images of slices through a patient s anatomy. Despite having lower spatial resolution than other x-ray

More information

X-RAYS - NO UNAUTHORISED ENTRY

X-RAYS - NO UNAUTHORISED ENTRY Licencing of premises Premises Refer Guidelines A radiation warning sign and warning notice, X-RAYS - NO UNAUTHORISED ENTRY must be displayed at all entrances leading to the rooms where x-ray units are

More information

PET/CT Instrumentation Basics

PET/CT Instrumentation Basics / Instrumentation Basics 1. Motivations for / imaging 2. What is a / Scanner 3. Typical Protocols 4. Attenuation Correction 5. Problems and Challenges with / 6. Examples Motivations for / Imaging Desire

More information

Model Based Iterative Reconstructions represent a paradigm shift - Imaging with almost no noise

Model Based Iterative Reconstructions represent a paradigm shift - Imaging with almost no noise Model Based Iterative Reconstructions represent a paradigm shift - Imaging with almost no noise Jonas Rydberg, M.D. Professor of Radiology Indiana University School of Medicine Indianapolis, Indiana Medical

More information

Slide 1. Slide 2. Slide 3 ACR CT Accreditation. Multi-Slice CT Artifacts and Quality Control. What are the rules or recommendations for CT QC?

Slide 1. Slide 2. Slide 3 ACR CT Accreditation. Multi-Slice CT Artifacts and Quality Control. What are the rules or recommendations for CT QC? Slide 1 Multi-Slice CT Artifacts and Quality Control Dianna Cody, Ph.D. Chief, Radiologic Physics UT MD Anderson Cancer Center Houston, TX Slide 2 What are the rules or recommendations for CT QC? AAPM

More information

CT Basics: Image Quality Module 6

CT Basics: Image Quality Module 6 Module 6 For educational and institutional use. This transcript is licensed for noncommercial, educational inhouse or online educational course use only in educational and corporate institutions. Any broadcast,

More information

7/24/2014. Image Quality for the Radiation Oncology Physicist: Review of the Fundamentals and Implementation. Disclosures. Outline

7/24/2014. Image Quality for the Radiation Oncology Physicist: Review of the Fundamentals and Implementation. Disclosures. Outline Image Quality for the Radiation Oncology Physicist: Review of the Fundamentals and Implementation Image Quality Review I: Basics and Image Quality TH-A-16A-1 Thursday 7:30AM - 9:30AM Room: 16A J. Anthony

More information

Dose Reduction in Helical CT: Dynamically Adjustable z-axis X-Ray Beam Collimation

Dose Reduction in Helical CT: Dynamically Adjustable z-axis X-Ray Beam Collimation Medical Physics and Informatics Original Research Christner et al. CT Dose Reduction Medical Physics and Informatics Original Research Downloaded from www.ajronline.org by 8.243.133.8 on 2/26/18 from IP

More information

make it easy, with Ray

make it easy, with Ray Lower dose - Quick scan times - Pulsed X-ray technology - Multiple scan modes 3 Dedicated detectors - Reliable performance - No damage - Long life span Easy upgrade - Ready to upgrade CBCT & Cephalometric

More information

Maximum Performance, Minimum Space

Maximum Performance, Minimum Space TECHNOLOGY HISTORY For over 130 years, Toshiba has been a world leader in developing technology to improve the quality of life. Our 50,000 global patents demonstrate a long, rich history of leading innovation.

More information

Diagnostic X-Ray Shielding

Diagnostic X-Ray Shielding Diagnostic X-Ray Shielding Multi-Slice CT Scanners Using NCRP 147 Methodology Melissa C. Martin, M.S., FAAPM, FACR Therapy Physics Inc., Bellflower, CA AAPM Annual Meeting, Orlando, FL FL Refresher Course

More information

Influence of different iteration levels in fourth generation iterative reconstruction technique on image noise in CT examinations of the neck

Influence of different iteration levels in fourth generation iterative reconstruction technique on image noise in CT examinations of the neck Influence of different iteration levels in fourth generation iterative reconstruction technique on image noise in CT examinations of the neck Poster No.: C-2205 Congress: ECR 2012 Type: Scientific Paper

More information

of sufficient quality and quantity

of sufficient quality and quantity of sufficient quality and quantity The patient s body attenuates the beam as it passes though the body More energy is deposited in organs located near the entry of the beam than near the exit of the beam

More information

Overview of Safety Code 35

Overview of Safety Code 35 Common Quality Control Procedures for All s Quality Control Procedures Film All s Daily Quality Control Tests Equipment Warm-up (D1) According to manufacturers instructions Can include auto calibration(d1)

More information

Aim. Images for this section: Page 2 of 13

Aim. Images for this section: Page 2 of 13 Changes in CT number of high atomic number materials with field of view when using an extended CT number to electron density curve and a metal artifact reduction reconstruction algorithm Poster No.: R-0094

More information

QC by the MPE in Belgium

QC by the MPE in Belgium Acceptance testing of state-of-the-art CT scanners using a new national protocol: first experience on a large number of scanners of different make and model the working group Radiology of the Belgian Hospital

More information

Digital Imaging started in the 1972 with Digital subtraction angiography Clinical digital imaging was employed from the 1980 ~ 37 years ago Amount of

Digital Imaging started in the 1972 with Digital subtraction angiography Clinical digital imaging was employed from the 1980 ~ 37 years ago Amount of Digital Imaging started in the 1972 with Digital subtraction angiography Clinical digital imaging was employed from the 1980 ~ 37 years ago Amount of radiation to the population due to Medical Imaging

More information

SAFIRE. Sinogram Affirmed Iterative Reconstruction. Answers for life.

SAFIRE. Sinogram Affirmed Iterative Reconstruction. Answers for life. Neuro Thoracic Abdominal Abdominal Cardiovascular Pediatric SAFIRE Sinogram Affirmed Iterative Reconstruction Answers for life. SAFIRE * (Sinogram Affirmed Iterative Reconstruction) * The information

More information

Maximizing clinical outcomes

Maximizing clinical outcomes Maximizing clinical outcomes Digital Tomosynthesis Dual Energy Subtraction Automated Long Length Imaging Improved image quality at a low dose Xray Xray Patented ISS capture technology promotes high sensitivity

More information

Image Quality and Dose. Image Quality and Dose. Image Quality and Dose Issues in MSCT. Scanner parameters affecting IQ and Dose

Image Quality and Dose. Image Quality and Dose. Image Quality and Dose Issues in MSCT. Scanner parameters affecting IQ and Dose Image Quality and Dose Issues in MSCT Image Quality and Dose Image quality Image noise Spatial resolution Contrast Artefacts Speckle and sharpness S. Edyvean St. George s Hospital London SW17 0QT Radiation

More information

CHAPTER 2 COMMISSIONING OF KILO-VOLTAGE CONE BEAM COMPUTED TOMOGRAPHY FOR IMAGE-GUIDED RADIOTHERAPY

CHAPTER 2 COMMISSIONING OF KILO-VOLTAGE CONE BEAM COMPUTED TOMOGRAPHY FOR IMAGE-GUIDED RADIOTHERAPY 14 CHAPTER 2 COMMISSIONING OF KILO-VOLTAGE CONE BEAM COMPUTED TOMOGRAPHY FOR IMAGE-GUIDED RADIOTHERAPY 2.1 INTRODUCTION kv-cbct integrated with linear accelerators as a tool for IGRT, was developed to

More information

Multislice CT: Current Technology and Future Developments

Multislice CT: Current Technology and Future Developments Multislice CT: Current Technology and Future Developments 1 Stefan Ulzheimer and Thomas Flohr Contents 1.1 Introduction 3 1.2 System Design 5 1.2.1 Gantry 6 1.2.2 X-Ray Tube and Generator 6 1.2.3 MDCT

More information

Introduction. Chapter 16 Diagnostic Radiology. Primary radiological image. Primary radiological image

Introduction. Chapter 16 Diagnostic Radiology. Primary radiological image. Primary radiological image Introduction Chapter 16 Diagnostic Radiology Radiation Dosimetry I Text: H.E Johns and J.R. Cunningham, The physics of radiology, 4 th ed. http://www.utoledo.edu/med/depts/radther In diagnostic radiology

More information

Exposure Indices and Target Values in Radiography: What Are They and How Can You Use Them?

Exposure Indices and Target Values in Radiography: What Are They and How Can You Use Them? Exposure Indices and Target Values in Radiography: What Are They and How Can You Use Them? Definition and Validation of Exposure Indices Ingrid Reiser, PhD DABR Department of Radiology University of Chicago

More information

Wide beam CT dosimetry. Elly Castellano

Wide beam CT dosimetry. Elly Castellano Wide beam CT dosimetry Elly Castellano Outline revision: CT dose indices wide-beam CT: the end of the road for CTDI? the IEC rescue plan for CTDI 100 the american way AAPM report 111 better estimates of

More information

An Activity in Computed Tomography

An Activity in Computed Tomography Pre-lab Discussion An Activity in Computed Tomography X-rays X-rays are high energy electromagnetic radiation with wavelengths smaller than those in the visible spectrum (0.01-10nm and 4000-800nm respectively).

More information

QC Testing for Computed Tomography (CT) Scanner

QC Testing for Computed Tomography (CT) Scanner QC Testing for Computed Tomography (CT) Scanner QA - Quality Assurance All planned and systematic actions needed to provide confidence on a structure, system or component. all-encompassing program, including

More information

T h e P h a n t o m L a b o r a t o r y

T h e P h a n t o m L a b o r a t o r y T h e P h a n t o m L a b o r a t o r y 1 CCT228 ATCM Phantom Manual Copyright 2017 WARRANTY THE PHANTOM LABORATORY INCORPORATED ( Seller ) warrants that this product shall remain in good working order

More information

Abdominal Radiology ISSN X. Abdom Radiol DOI /s

Abdominal Radiology ISSN X. Abdom Radiol DOI /s Assessment of 70-keV virtual monoenergetic spectral images in abdominal CT imaging: A comparison study to conventional polychromatic 120- kvp images Negin Rassouli, Hamid Chalian, Prabhakar Rajiah, Amar

More information

Test Equipment for Radiology and CT Quality Control Contents

Test Equipment for Radiology and CT Quality Control Contents Test Equipment for Radiology and CT Quality Control Contents Quality Control Testing...2 Photometers for Digital Clinical Display QC...3 Primary Workstations...3 Secondary Workstations...3 Testing of workstations...3

More information

Brilliance in everything Philips CT products and services

Brilliance in everything Philips CT products and services Brilliance in everything Philips CT products and services Ready for anything No one does more than Philips to help you gain the productivity you need with a comprehensive approach to CT that marries significant

More information

The future of nuclear imaging is clear

The future of nuclear imaging is clear Cardius X-ACT The future of nuclear imaging is clear Increased regulations, growing competition, and concerns about radiation exposure are just a sampling of the current challenges facing the nuclear medicine

More information

I. PERFORMANCE OF X-RAY PRODUCTION COMPONENTS FLUOROSCOPIC ACCEPTANCE TESTING: TEST PROCEDURES & PERFORMANCE CRITERIA

I. PERFORMANCE OF X-RAY PRODUCTION COMPONENTS FLUOROSCOPIC ACCEPTANCE TESTING: TEST PROCEDURES & PERFORMANCE CRITERIA FLUOROSCOPIC ACCEPTANCE TESTING: TEST PROCEDURES & PERFORMANCE CRITERIA EDWARD L. NICKOLOFF DEPARTMENT OF RADIOLOGY COLUMBIA UNIVERSITY NEW YORK, NY ACCEPTANCE TESTING GOALS PRIOR TO 1st CLINICAL USAGE

More information

CT Basics: Data Acquisition Module 3

CT Basics: Data Acquisition Module 3 Module 3 Transcript For educational and institutional use. This transcript is licensed for noncommercial, educational inhouse or online educational course use only in educational and corporate institutions.

More information

An Activity in Computed Tomography

An Activity in Computed Tomography Pre-lab Discussion An Activity in Computed Tomography X-rays X-rays are high energy electromagnetic radiation with wavelengths smaller than those in the visible spectrum (0.01-10nm and 4000-800nm respectively).

More information

Radionuclide Imaging MII Single Photon Emission Computed Tomography (SPECT)

Radionuclide Imaging MII Single Photon Emission Computed Tomography (SPECT) Radionuclide Imaging MII 3073 Single Photon Emission Computed Tomography (SPECT) Single Photon Emission Computed Tomography (SPECT) The successful application of computer algorithms to x-ray imaging in

More information

Multiple Choice Identify the letter of the choice that best completes the statement or answers the question.

Multiple Choice Identify the letter of the choice that best completes the statement or answers the question. RA110 test 3 Multiple Choice Identify the letter of the choice that best completes the statement or answers the question. 1. An object 35 cm in width is radiographed at 100 cm SID and at a 50 cm SOD. What

More information

Multi-Access Biplane Lab

Multi-Access Biplane Lab Multi-Access Biplane Lab Advanced technolo gies deliver optimized biplane imaging Designed in concert with leading physicians, the Infinix VF-i/BP provides advanced, versatile patient access to meet the

More information

Features and Weaknesses of Phantoms for CR/DR System Testing

Features and Weaknesses of Phantoms for CR/DR System Testing Physics testing of image detectors Parameters to test Features and Weaknesses of Phantoms for CR/DR System Testing Spatial resolution Contrast resolution Uniformity/geometric distortion Dose response/signal

More information

Acceptance Testing of a Digital Breast Tomosynthesis Unit

Acceptance Testing of a Digital Breast Tomosynthesis Unit Acceptance Testing of a Digital Breast Tomosynthesis Unit 2012 AAPM Spring Clinical Meeting Jessica Clements, M.S., DABR Objectives Review of technology and clinical advantages Acceptance Testing Procedures

More information

Dose Reduction and Image Preservation After the Introduction of a 0.1 mm Cu Filter into the LODOX Statscan unit above 110 kvp

Dose Reduction and Image Preservation After the Introduction of a 0.1 mm Cu Filter into the LODOX Statscan unit above 110 kvp Dose Reduction and Image Preservation After the Introduction of a into the LODOX Statscan unit above 110 kvp Abstract: CJ Trauernicht 1, C Rall 1, T Perks 2, G Maree 1, E Hering 1, S Steiner 3 1) Division

More information

FOREWORD. Acknowledgements

FOREWORD. Acknowledgements ΠΑΝΕΠΙΣΗΜΙΟ ΠΑΣΡΩΝ Διαημημαηικό Πρόγραμμα Μεηαπηστιακών ποσδών ζηην Ιαηρική Φσζική Διπλωμαηική εργαζία «ΔΟΙΜΕΣΡΙΑ ΑΘΕΝΩΝ Ε ΕΞΕΣΑΕΙ ΤΠΟΛΟΓΙΣΙΚΗ ΣΟΜΟΓΡΑΦΙΑ ΠΟΛΛΑΠΛΩΝ ΣΟΜΩΝ» ηέλλα Γ. Θαλαζζινού Α.Μ : 1575

More information

abc MHRA Philips Mx8000 IDT CT scanner technical evaluation September 2004 Best choice best practice nww.medical-devices.nhs.

abc MHRA Philips Mx8000 IDT CT scanner technical evaluation September 2004 Best choice best practice   nww.medical-devices.nhs. abc September 2004 MHRA 04099 Philips Mx8000 IDT CT scanner technical evaluation Best choice best practice www.mhra.gov.uk nww.medical-devices.nhs.uk About MHRA evaluation reports. What you can expect.

More information

Philips Astonish. Key advantages. including improved image quality and

Philips Astonish. Key advantages. including improved image quality and Philips Key advantages including improved image quality and with AC offers improved image quality, interpretative certainty, diagnostic Alternatively, simplify patient care by exposing patients to reduced

More information

RADspeed Pro. EDGEpackage C501-E041C

RADspeed Pro. EDGEpackage C501-E041C RADspeed Pro EDGEpackage C501-E041C 2 Some of the FPDs may be not available in your country. Please contact us to check the availability in your country. 3 Tomosynthesis in the Standing Position Tomosynthesis

More information

3/31/2011. Objectives. Emory University. Historical Development. Historical Development. Historical Development

3/31/2011. Objectives. Emory University. Historical Development. Historical Development. Historical Development Teaching Radiographic Technique in a Digital Imaging Paradigm Objectives 1. Discuss the historical development of digital imaging. Dawn Couch Moore, M.M.Sc., RT(R) Assistant Professor and Director Emory

More information

Quantitation of clinical feedback on image quality differences between two CT scanner models

Quantitation of clinical feedback on image quality differences between two CT scanner models Received: 4 August 2016 Revised: 4 November 2016 Accepted: 12 December 2016 DOI: 10.1002/acm2.12050 MEDICAL IMAGING Quantitation of clinical feedback on image quality differences between two CT scanner

More information

NeuViz 16 Computed Tomography. Elevating routine imaging for exceptional results

NeuViz 16 Computed Tomography. Elevating routine imaging for exceptional results NeuViz 16 Computed Tomography Elevating routine imaging for exceptional results Essence NeuViz 16 Raising the bar on clinical utility in routine imaging. Get more. More clinical information for patients.

More information

Pure confidence in action. Philips Ingenuity CT specifications

Pure confidence in action. Philips Ingenuity CT specifications Pure confidence in action Philips Ingenuity CT specifications Table of contents 1 Introduction 3 2 Imaging 2.0 4 3 Results Driven Scanning 5 3.1 ExamCards 5 3.2 ScanRuler 5 3.3 SyncRight 5 4 Dose management

More information

Initial Certification

Initial Certification Initial Certification Nuclear Medical Physics (NMP) Study Guide Part 2 Content Guide and Sample Questions The content of all ABR exams is determined by a panel of experts who select the items based on

More information

Changing the Shape of Nuclear Medicine

Changing the Shape of Nuclear Medicine TRUTH IN IMAGING Changing the Shape of Nuclear Medicine Multi-Purpose SPECT Scanner Nothing Gets Closer Introducing 360 Body Contour Scanning With 360 degree detector coverage, and unique proximity sensors

More information

TISSUE EQUIVALENT PHANTOMS FOR EVALUATING IN-PLANE TUBE CURRENT MODULATED CT DOSE AND IMAGE QUALITY

TISSUE EQUIVALENT PHANTOMS FOR EVALUATING IN-PLANE TUBE CURRENT MODULATED CT DOSE AND IMAGE QUALITY TISSUE EQUIVALENT PHANTOMS FOR EVALUATING IN-PLANE TUBE CURRENT MODULATED CT DOSE AND IMAGE QUALITY By RYAN F. FISHER A THESIS PRESENTED TO THE GRADUATE SCHOOL OF THE UNIVERSITY OF FLORIDA IN PARTIAL FULFILLMENT

More information

GE Healthcare. Senographe 2000D Full-field digital mammography system

GE Healthcare. Senographe 2000D Full-field digital mammography system GE Healthcare Senographe 2000D Full-field digital mammography system Digital has arrived. The Senographe 2000D Full-Field Digital Mammography (FFDM) system gives you a unique competitive advantage. That

More information

SOMATOM Esprit A Bundle of Energy

SOMATOM Esprit A Bundle of Energy SOMATOM Esprit A Bundle of Energy DATA SOMATOM Esprit An economical CT scanner designed for...... Excellent spiral image quality... A wide range of clinical applications... Value performance and reliabilty

More information

Y11-DR Digital Radiography (DR) Image Quality

Y11-DR Digital Radiography (DR) Image Quality Y11-DR Digital Radiography (DR) Image Quality Image quality is stressed for all systems in Safety Code 35. In the relevant sections Health Canada s advice is the manufacturer s recommended test procedures

More information

Lunar Technology Advantages

Lunar Technology Advantages Lunar Technology Advantages DXA stands for Dual-Energy X-ray Absorptiometry. It is a measurement method that uses the differences in the absorption of high energy and low energy X-ray photons by different

More information

CT Basics: Computed Tomography Fundamentals Module 1

CT Basics: Computed Tomography Fundamentals Module 1 Module 1 Transcript For educational and institutional use. This transcript is licensed for noncommercial, educational inhouse or online educational course use only in educational and corporate institutions.

More information

Beam-Restricting Devices

Beam-Restricting Devices Beam-Restricting Devices Three factors contribute to an increase in scatter radiation: Increased kvp Increased Field Size Increased Patient or Body Part Size. X-ray Interactions a some interact with the

More information

Suppression of metal artifacts using image-based monoenergetic DECT imaging

Suppression of metal artifacts using image-based monoenergetic DECT imaging Suppression of metal artifacts using image-based monoenergetic DECT imaging Poster No.: C-0519 Congress: ECR 2011 Type: Scientific Paper Authors: B. Krauss, B. Schmidt, M. Sedlmair, T. Flohr; Forchheim/DE

More information

COCIR SELF-REGULATORY INITIATIVE FOR MEDICAL IMAGING EQUIPMENT COMPUTED TOMOGRAPHY MEASUREMENT OF ENERGY CONSUMPTION

COCIR SELF-REGULATORY INITIATIVE FOR MEDICAL IMAGING EQUIPMENT COMPUTED TOMOGRAPHY MEASUREMENT OF ENERGY CONSUMPTION COCIR SELF-REGULATORY INITIATIVE FOR MEDICAL IMAGING EQUIPMENT COMPUTED TOMOGRAPHY MEASUREMENT OF ENERGY CONSUMPTION Revision: 1 Date: June 2015 Approved: June 2015 TABLE OF CONTENT 1. INTRODUCTION...

More information

Advanced Noise Reduction Processing for X-ray CT System with Iterative Processing. Koichi Hirokawa MEDIX VOL. 56 P.43 P.46

Advanced Noise Reduction Processing for X-ray CT System with Iterative Processing. Koichi Hirokawa MEDIX VOL. 56 P.43 P.46 Advanced Noise Reduction Processing for X-ray CT System with Iterative Processing Taiga Goto Koichi Hirokawa Hisashi Takahashi MEDIX VOL. 56 P.43 P.46 Advanced Noise Reduction Processing for X-ray CT System

More information

Four-dimensional Computed Tomography (4D CT) Concepts and Preliminary Development

Four-dimensional Computed Tomography (4D CT) Concepts and Preliminary Development ORIGINAL ARTICLE ORIGINAL ARTICLE Radiation Medicine: Vol. 21 No. 1, 17 22 p.p., 2003 Four-dimensional Computed Tomography (4D CT) Concepts and Preliminary Development Masahiro Endo,* Takanori Tsunoo,*

More information

Digital Imaging and Communications in Medicine (DICOM) Supplement 188: Multi-energy CT Images

Digital Imaging and Communications in Medicine (DICOM) Supplement 188: Multi-energy CT Images Supplement 188: Multi-energy CT Images Page 1 2 4 6 Digital Imaging and Communications in Medicine (DICOM) 8 Supplement 188: Multi-energy CT Images 10 12 14 16 18 20 Prepared by: 22 DICOM Standards Committee,

More information

SIGNA Explorer Lift revives our MR

SIGNA Explorer Lift revives our MR Seiji Shiotani, MD, PhD Seirei Fuji Hospital in Fuji City, Shizuoka, Japan Masayoshi Sugimura Seirei Fuji Hospital in Fuji City, Shizuoka, Japan SIGN Explorer Lift revives our MR The clinical usefulness

More information

Clinical Experiences with a Patient Skin Dose Monitoring and Tracking Program

Clinical Experiences with a Patient Skin Dose Monitoring and Tracking Program Clinical Experiences with a Patient Skin Dose Monitoring and Tracking Program Allen R. Goode, MS, DABR Chief Diagnostic Medical Physicist Department of Radiology & Medical Imaging University of Virginia

More information

Clinical Experience Using the Open Bore Multislice CT System Supria (16 slice CT) MEDIX VOL. 61 P.8 P.11

Clinical Experience Using the Open Bore Multislice CT System Supria (16 slice CT) MEDIX VOL. 61 P.8 P.11 Clinical Experience Using the Open Bore Multislice CT System Supria (16 slice CT) Hiroki Kadoya Yukiko Kitagawa MEDIX VOL. 61 P.8 P.11 Clinical Experience Using the Open Bore Multislice CT System Supria

More information

Patient-Assisted Compression Impact on Image Quality and Workflow

Patient-Assisted Compression Impact on Image Quality and Workflow Patient-Assisted Compression Impact on Image Quality and Workflow Senographe Pristina In 2017, GE Healthcare s Senographe Pristina ( Pristina ) was approved by the FDA using the standard technologist-controlled

More information

Chiara Secco. PET Performance measurements of the new LSO-Based Whole Body PET/CT. Scanner biograph 16 HI-REZ using the NEMA NU Standard.

Chiara Secco. PET Performance measurements of the new LSO-Based Whole Body PET/CT. Scanner biograph 16 HI-REZ using the NEMA NU Standard. Chiara Secco PET Performance measurements of the new LSO-Based Whole Body PET/CT Scanner biograph 16 HI-REZ using the NEMA NU 2-2001 Standard. INTRODUCTION Since its introduction, CT has become a fundamental

More information

CT Basics: Equipment and Instrumentation Module 2

CT Basics: Equipment and Instrumentation Module 2 Module 2 Transcript For educational and institutional use. This transcript is licensed for noncommercial, educational in-house or online educational course use only in educational and corporate institutions.

More information

2D, 3D CT Intervention, and CT Fluoroscopy

2D, 3D CT Intervention, and CT Fluoroscopy 2D, 3D CT Intervention, and CT Fluoroscopy SOMATOM Definition, Definition AS, Definition Flash Answers for life. Siemens CT Vision Siemens CT Vision The justification for the existence of the entire medical

More information

Ludlum Medical Physics

Ludlum Medical Physics Ludlum Medical Physics Medical Imaging Radiology QA Test Tools NEW LUDLUM PRODUCT LINE Medical Physics Products Medical Physics Products What are they? Products used to measure radiation output and to

More information

Multisection CT: Scanning Techniques and Clinical Applications 1

Multisection CT: Scanning Techniques and Clinical Applications 1 IMAGING & THERAPEUTIC TECHNOLOGY Multisection CT: Scanning Techniques and Clinical Applications 1 1787 Jonas Rydberg, MD Kenneth A. Buckwalter, MD Karen S. Caldemeyer, MD Michael D. Phillips, MD Dewey

More information

Industry Breakthrough

Industry Breakthrough Industry Breakthrough Dynamic SPECT Acquisition Quantifying Myocardial Blood Flow Nuclear Cardiology in the 21st Century In the 21st century, most nuclear cameras are still relying on a technology invented

More information

A comparison of two methods for the determination of freein-air geometric efficiency in MDCT

A comparison of two methods for the determination of freein-air geometric efficiency in MDCT A comparison of two methods for the determination of freein-air geometric efficiency in MDCT Theocharis Berris *1, Kostas Perisinakis 1,, Antonios E. Papadakis and John Damilakis 1, 1 Department of Medical

More information

Overview. Professor Roentgen was a Physicist!!! The Physics of Radiation Oncology X-ray Imaging

Overview. Professor Roentgen was a Physicist!!! The Physics of Radiation Oncology X-ray Imaging The Physics of Radiation Oncology X-ray Imaging Charles E. Willis, Ph.D. DABR Associate Professor Department of Imaging Physics The University of Texas M.D. Anderson Cancer Center Houston, Texas Overview

More information

MUSICA Nerve Center. Artificial Intelligence. Intelligent tools for your Digital Radiography workflow. Fluoroscopy. Workflow Optimization

MUSICA Nerve Center. Artificial Intelligence. Intelligent tools for your Digital Radiography workflow. Fluoroscopy. Workflow Optimization Image Quality Bariatric Abdomen Pediatric Imaging Diagnostic Confidence Fluoroscopy Neonatal Imaging Scatter Suppression Dental Full Leg Full Spine Exposure Control Index Artificial Intelligence General

More information

Digital Imaging and Communications in Medicine (DICOM) Supplement 188: Multi-energy CT Images

Digital Imaging and Communications in Medicine (DICOM) Supplement 188: Multi-energy CT Images Supplement 188: Multi-energy CT Images Page 1 2 4 6 Digital Imaging and Communications in Medicine (DICOM) 8 Supplement 188: Multi-energy CT Images 10 12 14 16 18 20 Prepared by: 22 DICOM Standards Committee,

More information

Truly flexible to meet your clinical needs

Truly flexible to meet your clinical needs Truly flexible to meet your clinical needs 2 Adapting to meet your needs Flexible Fast and responsive Excellent image quality Designed with ergonomic efficiency Equipped with dose management tools 3 Three

More information

Measurement of table feed speed in modern CT

Measurement of table feed speed in modern CT JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, VOLUME 15, NUMBER 3, 2014 Measurement of table feed speed in modern CT Atsushi Fukuda, 1,2a Pei-Jan P. Lin, 3 Kosuke Matsubara, 2 Tosiaki Miyati 2 Department

More information

Focal Spot Blooming in CT: We Didn t Know We Had a Problem Until We Had a Solution

Focal Spot Blooming in CT: We Didn t Know We Had a Problem Until We Had a Solution Focal Spot Blooming in CT: We Didn t Know We Had a Problem Until We Had a Solution Cynthia H. McCollough, PhD, DABR, FAAPM, FACR Director, CT Clinical Innovation Center Professor of Medical Physics and

More information