USER MANUAL RASI 900 / RASI 901. EiUK Eurotron Instruments (UK) ltd

Size: px
Start display at page:

Download "USER MANUAL RASI 900 / RASI 901. EiUK Eurotron Instruments (UK) ltd"

Transcription

1

2 USER MANUAL RASI 900 / RASI 901 EiUK Eurotron Instruments (UK) ltd 2

3 We are grateful for your purchase of EiUK Eurotron Instruments (UK) ltd Gas Analyzer First read this instruction manual carefully until an adequate understanding is acquired, and then proceed to installation, operation and maintenance of the analyzer. Wrong handling may cause an accident or injury. The specifications of this analyzer are subject to change without prior notice for further product improvement. Modification of this analyzer is strictly prohibited unless a written approval is obtained from the manufacturer. EiUK Eurotron Instruments (UK) ltd will not bear any responsibility for a trouble caused by such a modification. This instruction manual shall be stored by the person who actually uses the analyzer. After reading the manual, be sure to store it at a place easier to access. This instruction manual should be delivered to the end user without fail. Request: It is prohibited to transfer part or all of this manual without EiUK Eurotron Instruments (UK) ltd permission in written format. Description in this manual is subject to change without prior notice for further improvement. The products described in this manual are subject to continuous development and improvement and it is therefore acknowledged that this manual may contain errors or omissions. Eurotron Instruments (UK) ltd encourage customer feedback and welcome any comments or suggestions relating to the product or documentation. These should be forwarded to the Customer Feedback Department at the address given below. Eurotron Instruments (UK) ltd Unit 13 Riley Close, Royal Oak Industrial Estate, Daventry, NN11 8QT, England Tel: +44 (0) Fax: +44 (0) sales@eurotronuk.com This manual is intended as a guide to use this product. Eurotron Instruments (UK) ltd shall not be liable for any loss or damage whatsoever arising from the wrong comment / interpretation of information s from this manual or any misuse come out of this manual. 3

4 Eventual problems after power-on 1. Display contrast problem After power-on the unit, if you have trouble reading the display, you can change display contrast by pressing the, keys: change of display contrast occur in 5 % steps! Analyzer display after power-on: 2. Software language problem After power-on the unit, if you have no installation of the language of the country of your origin, please press or for some 7 seconds. This will switch on the language change feature. By means of keys, you can search the language required Now press in order to accept the new language setting. The analyzer will restart with the selected language. 4

5 1 Content 1 Content Introduction Security advises Keypad Operating Menu overview Switch analyzer on/off and Timer Function Switch analyzer on/off in grid power supplied mode Timer function Switch analyzer on/off in battery powered mode (RASI 901) Self test Operation with gas cooler (RASI 901) Operation with heating line (RASI 901) Operation with heating probe (RASI 901/RASI 900)) Main menu measuring settings Measurement program selection Measuring program configuration CO-limit Preset of measurement display Preset for print-out Preset fuel type Configuration fuel type list User definable fuel types...19 Modify self definable fuel type Flow measurement (option) Automatic interval measurement Setup of auto-measurement Reset to zero Selection of temperature unit Selection of pressure unit Selection of interval auto-zero Main menu data Structure of a data block New entry of a site data block Delete all sites Site data from PC Memory occupation View stored data Delete all measuring data Send measuring data to PC Copy measuring data to SD card (option)

6 7.10 Format SD card (option) Main menu Xtras (settings) AUX-channel configuration Configuration AUX input type Terminal assignments Measuring name Measuring unit Lower / upper limit Resolution RS 232 settings Draft measurement PC Analyzer settings Set clock Setting display contrast Helping hints Country Negative gas readings Select T-gas sensor type Reference temperature Hold delay after purge Service values Adjustment Main adjustment menu Recommended calibration gases Adjustment electrochemical sensors Infrared bench adjustment Reset to zero Setup analog output (4 20 ma) Setup analog output During zeroing Setup analog output Output Channel Setup analog output Measuring value Setup analog output Minimal value (lower range end, correspond to 4 ma) Setup analog output Maximal value (upper range end, correspond to 20 ma) Main menu Info Unit status Device options General, functions Text edit Quick input of numbers Request user decision

7 2 Introduction The analyzer was developed for the semi-continuous and automatic measurement of gas component concentrations in flue gases (e.g. SO 2, CO, NO, NO 2 etc). 3 Security advises The following security advises have to be strictly observed. They are an essential and indispensable part of the user s documentation. Not observing the security advises can mean loss of your right to claim under guarantee. Qualified staff Certain works at the device and its accessories are only to be done by especially qualified staff. Such are persons which are authorized from the safety officer due to their education, knowledge and experience and which can recognize and avoid possible danger. Planning and installation Installation design, assembly, installation and maintenance work are carried out only by trained and qualified persons and are checked by responsible qualified staff. Certain jobs, such as electrical installation, require special knowledge. The qualification of persons performing these jobs must correspond to the degree of difficulty. Operation The device might cause danger, especially in case of not observing the security advises. Maintenance The analyzer system components are operating equipment for use in industrial high voltage installations. During installation, start-up and service wrong operational control or insufficient maintenance can cause severe injuries or damage to persons or equipment. To prevent equipment malfunctions, scheduled maintenance and inspections performed by qualified and servicing staffs are required. During maintenance work, make sure you switched off the power supply! During maintenance work, take care of hot surfaces! Attention, danger of skin burning! Wear safety gloves! Malfunctions Abnormal operating conditions must be regarded as serious indications for a functional degradation. Such indications are: heavy drift of measurement results increased power consumption or power failure increased temperatures inside the steel cabinet unusual noises or smells Quality Control Department Eurotron Instruments (UK) ltd 7

8 4 Keypad Position Symbol function 1 Save configuration key 2 Dependent on text/object Main menu: measurement display 3 Dependent on text/object Main menu: measurement settings 4 Dependent on text/object Main menu: Data menu 5 Dependent on text/object Main menu: Setting menu 6 Dependent on text/object Main menu : Information menu 7 Help function 8 Print function (optional) 9 ESC function 10 ENTER function 11 Change value/setting 12 Change value/setting 13 Move selection cursor 14 Move selection cursor 8

9 5 Operating 5.1 Menu overview The structure of this operating manual is based upon the structure of the main menus. This manual explains the analyzer inclusive all available options! Therefore the display contents could be insignificant deviate from your delivered analyzer. (IR-bench, electrochemical sensors, other sensors.) Flowchart main menus Switch on the mains switch in order to power-on analyzer switch on self test (automatically) set to zero (automatically) start measurement (automatically) press, keys to change display contrast press, for 7 seconds to change display language press : store actual measurement press, or : select main settings menu press : select main data menu press : select main Xtras menu press.: select main Info menu After power on, the analyzer performs automatically the self test and setting to zero. Then the measurement starts and the display shows the actual measuring values (main measuring menu). By means of the keys,,, and the menus of interest can be directly selected. 9

10 Main measuring window Store actual measuring values (Chapter 5.1) Measuring settings Measuring settings (Chapter 6) Gas flow velocity measurement (Chapter 6.5) Auto-measurement (Chapter 6.6) Reset to zero (Chapter 6.7) Measuring program (Chapter 6.2) Fuel type (Chapter 6.3) Temperature unit (Chapter 6.8) Pressure unit (Chapter 6.9) Interval auto-zero (Chapter 6.10) Main data menu Main data menu (Chapter 7) Select a site (Chapter 7.2) Delete all sites (Chapter 7.3) Sites from PC (Chapter 7.4) Memory occupation (Chapter 7.5) View stored data (Chapter 7.6) Delete all measured data (Chapter 7.7) Send measured data to PC (Chapter 7.8) Copy measured data to MMC/SD (Chapter 7.9) Format MMC/SD card (Chapter 7.10) Main Xtras menu Xtras (Chapter 8) RS232 settings (Chapter 8.2) Analyzer settings (Chapter 8.3) Service values (Chapter 8.4) Adjustment (Chapter 8.5) Analog output configuration (Chapter 8.6) Main information menu Main information menu (Chapter 9) Unit status (Chapter 9.1) Device options (Chapter 9.2) 10

11 5.2 Switch analyzer on/off and Timer Function After switch on, the analyzer shows on the display for a few seconds, a graphical logo (if installed) or the device name. Then the analyzer behaves differently in grid power or battery power supplied mode Switch analyzer on/off in grid power supplied mode When the user connects mains, the analyzer will switch on itself and start the battery charge function. It is not possible to switch off the analyzer as long as grid power supply stays connected. When grid power supply is disconnected while the battery is charging, the analyzer will switch off itself immediately. Pressing the - key will start the Self Test. Pressing the - key will start the Timer Function (chapter 5.2.2). In grid power supplied mode the Self Test will be left automatically, as soon as all components are OK. Pressing the - key will not switch off the analyzer, but the following window will appear: Select the desired function with Arrows and Enter keys. If there is an optional pressure sensor, this menu also offers to zero this sensor. This window will also be closed without any action by pressing the ESC key. Recharge battery means a restart, which will lead to the battery charge function. Save configuration will store eventually changed user settings in a non volatile memory. 11

12 5.2.2 Timer function The Timer Function allows scheduling a certain time, when the analyzer shall be perfectly ready for measurement, without stressing the measurement components in the meanwhile. In grid power supplied mode the Timer Function is accessed by pressing the function. - key during the battery charge Example: You want to measure in the morning at 8:00 am. In the evening before you install the analyzer at the sample point with connected mains, start the Timer Function and set it to 5:00 am. At 5:00 am the analyzer will leave the Timer window, will start Self Test and will leave the Self Test (when all components are OK) without a press on the ON - key. When you arrive at 8:00 am, the analyzer will already be running since 3 hours, therefore it is perfectly ready for measurement. You will just restart a zeroing when you arrive and after few minutes you can start the measurement with the best possible stability. When grid power supply is disconnected, the analyzer will switch off itself, but it will restart with the same timer settings, when grid power supply returns., : change the timer start in steps of 1 hour : go to Self Test => Zeroing => Measurement, Timer will be deactivated, : go to battery charge window, Timer will be deactivated On Power-ON, the timer is deactivated if is found out that the timer lies more than 1 week in the future or more than 2 days in the past (plausibility check). If the device was started by a timer-auto-start without the ON key was pressed, then press ON-key is faded in all 8 seconds in the middle of the LCD, until the user presses the required key Switch analyzer on/off in battery powered mode (RASI 901) The analyzer will be powered on by pressing the - key. It will go to Self Test. In battery powered mode the Self Test will not be left as long the ON key was not pressed. The analyzer will be powered off by pressing the - key. The following window will appear: Select the desired function with arrows and Enter keys. You can also switch off the analyzer by pressing the ON key again. If there is an optional pressure sensor, this menu also offers to zero this sensor. This window will also be closed without any action by pressing the ESC key. 12

13 5.3 Self test After power on, the displays shows the analyzer type and the analyzer status (actual voltage and temperature values). If one value is provided with an exclamation mark, the set to zero can not be started. One parameter can be ignored if no exclamation mark is being displayed and other measuring conditions are ready (all values are displayed with O.K.) The measurement can be started by means of pressing the key. As soon as ON-key for measurement will be displayed and all measurement conditions are satisfied, the measurement / set to zero can be started by pressing the - key (or automatically) The set to zero process will be started now, indicated by a progress bar in the display s right bottom. After a few minutes Esc will be displayed right of the progress bar, this means now the measurement is possible although the warm-up of the NDIR-bench is not finished. By pressing the Esc key the measurement can be started with reduced accuracy. Esc? With the key the sensors which are not ready for measurement with the best accuracy will be displayed. Set to zero is carried out in the background. For a sufficient accuracy according to analyzers specification, the infrared measuring technology requires a warm up period of at least 30 minutes, dependent on the outside ambient temperature! During this period of time, flue gas measurement cannot be started. Still, all other menu items can be selected. 13

14 After finishing set to zero, the analyzer starts automatically the measurement and the main measuring menu will be displayed. Reset to zero ESC is indicated beside the Reset to zero symbol, as soon as a measurement with diminished exactness of the infrared measuring technology is possible. Set to zero is carried out in the background. For a sufficient accuracy according to analyzers specification, the infrared measuring technology requires a warm up period of at least 30 minutes to get stables values. n the first zeroing after power-on, the analyser starts measurement as soon as the other sensors are ready (after about 10 minutes). Then the analyser starts the measurement, but the NDIR values (e.g. CO2, CH4) are hidden with "---". When 30 minutes are over, the analyser will show in the measurement window's bottom line "NDIR bench ready for zeroing". The user should restart zeroing then to get the measured NDIR values on the display. After finished reset to zero, the analyzer starts automatically the measurement and the main measuring window will be displayed. The multi-page (up to 5 windows) measuring display can be set individually and differently for each of the 4 measuring programs., : change the visible page, : toggle between 4 value indication (zoom) and 10 value indication : store momentary measuring values (assigned to currently selected vehicle) /. : change to menu settings.: change to menu data : change to menu Xtras : change to menu info 14

15 5.4 Operation with gas cooler (RASI ( 901) The built-in gas cooler cools the typical sample gas contains humidity of < 60 C dew point with line voltage supply on 5 C down, with battery operation for energy savings reasons on 10 C under device interior temperature. 5.5 Operation with heating line (RASI 901) A used heating line is heated only at line voltage supply. 5.6 Operation with heating probe (RASI ( 901/RASI 900)) A used heating probe is heated only at line voltage supply. 15

16 6 Main menu measuring settings, :Selection of the relevant parameter for measurement start flow measurement Chapter 6.5 setup auto measurement Chapter 6.6 start new reset to zero Chapter 6.7 configuration measuring programs Chapter 6.1 configuration fuel types Chapter 6.3 select measuring program, Chapter 6.1 select fuel type, Chapter 6.3 select temperature unit, Chapter 6.8 select pressure unit, Chapter Measurement program selection The following settings enable the fast measuring configuration by costumer specific settings in the Xtras menu. A measuring program contains the following facts for the measuring: 1. the fuel type 2. the CO-threshold 3. configuration of the measuring display 4. configuration of the print-out The device memorizes the fuel assigned with the last measurement and uses this setting again with the next measurement., : change the momentary measuring program (program 1 program 4) The program names are individually editable (Chapter 10.1) 16

17 6.2 Measuring program configuration, : select measuring program, : modify CO cut-off of the selected measuring program modify value in 200 ppm steps : modify the name (Chapter 10.1).: print setting of the selected measurement program : display setting of the selected meas. program : default setting (factory setting) set standard name of the measuring program, : return to setting menu CO-limit The CO-threshold is only available if the analyzer has installed an electrochemical CO sensor. The deposit ppm value determines the CO cut-off. If the measuring value reaches this threshold a second pump for fresh air purging switches on and the CO sensor is switches off by means of a solenoid valve of the gas path. If the ppm value falls short off 20 % of the threshold value the CO sensor will be admitted again with the measuring gas., : modify CO cut-off of the selected measuring program modify value in 200 ppm steps Preset of measurement display For every measuring program it is possible to adjust, which values should be visible in the measuring display. Each measuring program has 5 pages with 10 lines. They can be adjusted individually. During measurement, displays can be paged forward and backward., : select the line (in the first and the last line, the page will be turned), : change measuring value to be displayed at the selected position.: default setting (factory setting) : scroll page downwards : scroll page upwards, : return to setting menu Note: There is no problem in displaying one measuring value on several pages. If e.g. T-gas should be always visible, it can be displayed e.g. on every page. 17

18 Note: If a page only consists of blank lines, it will not be visible while paging. If e.g. only 2 pages are needed, the rest of them should be left blanc Preset for print-out It is possible to adjust for each measurement program, which values should be displayed on the print-out., : select the line (in the first and the last line, the page will be turned), : activate/deactivate print of the selected measuring value : activate/deactivate print of the selected measuring value : activate print of all measuring value : deactivate print of all measuring value (empty print-out), : return to setting menu Measuring values to be printed are marked with *. The list of measuring data can contain more than 40 lines (depending on the existing analyser options) and cannot be shown in one display, not even with basic equipment. 6.3 Preset fuel type, : change the actual fuel type (the analyzer stores the changed fuel type to the activated program. Note: Only the activated fuel types will be displayed (Chapter 6.3)! Activated fuel types are marked with *! 18

19 6.4 Configuration fuel type list Here you can carry out a fuel type pre-selection. The required fuel types are selected and the unused are left out. In the Measuring menu, section Fuel type, only the pre-selected fuel types will appear., : scroll selection cursor, or also scroll fuel type list (there are more fuel types available as displayed in one window), : switch on/off the selected fuel type (on= marked * ) : switch on/off the selected fuel type (on= marked * ).: to window user definable fuel types : change O2 value of the selected fuel type and fuel type information, : return to setting menu User definable fuel types Here, 4 fuels can be preset individually. The name as well as all combustion parameters (CO2 max, O2 reference value and the calculation parameters A2 and B) are adjustable. As the other fuel types, they can be pre-selected or left out., : select the fuel type, : activate selected individual fuel type (selected = *) : modify fuel types parameters (Chapter10.1) : activate selected individual fuel type (selected = *) : default setting (factory setting name of the selected fuel type), : return to fuel type selection Modify self definable fuel type, : scroll selection cursor, : modify name of fuel type or value : modify settings (Chapter 10.1), : return to setting menu 19

20 6.5 Flow measurement (option) This option enables the measurement of the gas flow velocity in the flue gas duct or stack. (option differential pressure is necessary!) The flow measurement is not interrupted. If the measurement display is configured corresponded to the flow measurement (Chapter 6.1) the flow rate [m/s], the flow volume [cfm] and the emission e.g.. CO [mg/s] will be displayed. Requires to measure: Dynamic pressure: measured with the Pitot tube Static pressure: measured with the Pitot tube Flue gas temperature measured with the probe Barometric pressure measured with a barometric pressure transmitter connected at the AUX input or preset the barometric pressure value with the keys Setup flow measurement, : move selection cursor, : change unit velocity unit [m/s] or [feet/s] cross-sectional area Quick input of numbers (Chapter 10.2 Cross-sectional area unit: [cm 2 ], [inch 2 ] or [feet 2 ] Volume flow unit: [l/s], [m 3 /h] or [cfm] Note: The input of the cross-sectional area is only for the volume flow necessary there the analyzer calculates the volume flow by means of the values of velocity and cross-sectional area. 20

21 Connection of Pitot tube for flow velocity measurement Analyzer connection: Pressure Pressure Delta p + Delta P If the measurement display is configured corresponded to the flow measurement the flow rate [m/s], the norm flow volume [cfm] (see Chapter 8.3.7) and the emission e.g. CO [mg/s] will be displayed., : change the visible page, : switch to big display ( Zoom function) : store momentary measuring values /. : change to menu settings.: change to menu data : change to menu Xtras : change to menu info 21

22 6.6 Automatic interval measurement By means of the option automatic interval measurement the analyzer is able to record the continuous measuring values itself. The user may vary the settings of the automatic program according to his needs. Once activated, this option generates measurements and other functions over a longer period of time. These functions can print and/or store the measurements. So the waiting phase is the only phase which is really being changed while changing the interval time. The user cannot change purge time, or zeroing time or gas measurement time. Due to a certain limit of the internal memory, there are also limits in the ratio between the duration of the total measurement and the selectable intervals. The analyzer automatically calculates and limits the total time allowance for the measurements respectively the minimal measuring interval depending on the memory space available. This means a shorter measuring interval will also result in a shorter total measuring time (duration). With an activated automatic print out the measuring interval can of course not be shorter than the printout needs to be completed. Hence the analyzer will not accept intervals shorter than 60 seconds. Please make sure that there is always plenty of paper in the printer when the automatic printout function is activated. Should the printer paper have to be changed during the on-going of an automatic measurement the interval should be long enough to allow the completion of this procedure. The analyzer does not monitor when it runs out of paper. Continuous transmission of the measured values over RS 232 is also ensured during activated automatic measurement. It is therefore easily possible to send data to a computer for storing and evaluation purposes, for example with the software Online View. When a measurement is started the automatic measurement is generally inactive. It has to be deliberately activated by the operator in order to avoid unintended print- and storing. We recommend calibrating the analyzer regularly. During long term measurements the sensors may offset drift a little bit which results in a loss of accuracy. The phase of sensor stability depends on the surrounding conditions and the gas concentration. A rule of thumb is that electrochemical sensors (O2, CO, NO) will not drift before 6 hours and the draft sensor not before 1 2 hours. Make sure that the probe and analyzer filters will not clog during long term measurements. 22

23 6.6.1 Setup of auto-measurement, : change value, auto-measurement on/off, : move cursor, change the line : measuring values : minimum value : maximum value : modify values, Quick input of numbers (chap. 10.2) : end, back to settings Available settings: (Auto-measurement deactivated!) Duration: between 1 and 30 minutes Interval: between 1 und seconds Note: The duration time can never be lower as the interval time! Storage: switch on/off storage in the memory defines whether there is supposed to store measuring data after each interval. Print-out: switch on/off print-out defines whether there is supposed to print measuring data after each interval. Note: The lowest interval time is 30 seconds if print-out is activated. Mean values: calculation of the mean values of the past measuring interval. These mean values will be stored/printed-out instead of the actual values Note: only continuous measurement mode is available with mean values ON! Destinat. memory: selection of where data is going to be stored storage internal or storage on card (optional SD card) When mean values are printed, the printout shows e. g. average value 60s. If there is not enough available memory for the desired auto measurement, the analyzer automatically will shorten the duration and / or extends the interval time. 23

24 6.7 Reset to zero In case readings with ambient air are erroneous (for example negative reading drift), the operator can start manual reset to zero. For reset to zero, the analyzer switches automatically from gas sampling to ambient air sampling and after taking a new zero will switch back to sampling gas. 6.8 Selection of temperature unit, : select the temperature measuring unit between [ C] or [ F] 6.9 Selection of pressure unit, : select the pressure unit Selectable pressure units and the calculation factors: unit number of decimal places calculation factor (with reference to mbar) Pa 1 100,0 hpa 2 1,0 kpa 3 0,1 Mbar 2 1,0 mmh2o 1 10,197 cmh2o 2 1,0197 In.H2O 3 0,4015 mmhg 2 0,7501 In.Hg 3 0,02953 PSI 3 0,

25 6.10 Selection of interval auto-zero, : select the interval duration between two auto-zero cycles (Available settings minimal interval 1/2 hour up to maximal interval 24 hours) It is recommended to set an interval time for auto-zero cycle as small as possible: this will increase the accuracy of analyzer. Low auto-zero interval time will also increase the life time of electrochemical cells! Electrochemical cells require ambient air purge from time to time; the moisture of ambient air will recover the electrolyte from inside the cell. Do not expose electrochemical cells to very high concentration of gas! 25

26 7 Main menu data The analyzer has a storage organisation for measurements minimum. Every measurement has to be allocated a site. It is possible to allocate several measurements to one site. The measurements of one site will be stored directly side by side. If the memory is full occupied the analyzer it is not possible to save more measurements, without delete the whole or a part of the memory. A measurement data block consist: Date and clock time of the measurement Number of the site Measuring values Program and fuel Press key = data: 7.1 Structure of a data block A customer data block consists of: Field name Field size Note Input Site no. 15 definite no. necessary Pos.1 15 Description line 1 e.g. name of site not necessary Pos.2 15 Description line 2 not necessary Pos.3 15 Description line 3 not necessary Pos.4 15 Description line 4 not necessary Pos.5 15 Description line 5 not necessary An unlimited number of measuring data blocks (only limited by the size of the storage) can be saved under one site. When saving the measuring data, the corresponding site has to be selected. Hint: If generally, the customer administration is not used, but data should nevertheless be saved, a customer should be created (Chapter 7.2 ), inserting one blank in the corresponding customer number. In the future, the analyser saves all measuring data blocks without visible customer number and co-ordinates them to the pseudo customer. 7.2 New entry of a site data block Press key F1=new in the display Select a site, the analyser demands the input of the site number. The display Input of Text (see Chapter 10.1) will be opened. Please type the site number. If you cancel the input or do not insert digits, the new customer will not be entered. After input of the site number, the site will be stored in the data base. Now it is possible if requested to fill in the description of the site. 26

27 : new entry (Chapter 10.1) : modify data (Chapter 10.1) : delete data (Chapter 10.1) find data (Chapter 10.1) : end 7.3 Delete all sites All sites and all allocated measurements will be deleted 7.4 Site data from PC Via the RS 232 interface the analyser is able to receive customer numbers including addresses from a PC. For this, the necessary PC software (e.g. Online View) has to be installed. The necessary connection between the measuring analyser and the PC or handheld PC has to be carried out before transmission and shall not to be disconnected during transmission. The included transmission cable has to be connected at the RS 232 interface (COM) of the PC and at the RS 232- interface at the analyzer. Transmission software: 32-bit Data Logger or 32-bit data logger Now the analyser asks, if the already saved sites should be deleted or if the received site should be added to the site database: YES: all saved sites and all measurements will be deleted. NO: no erasing, the new data will be added to the site data. If the site number is not available the existing site will be updated with the new description and the potential measuring data remains allocated. The analyzer receives data from PC. If the transmission connection to PC is ok, the analyzer the started and the analyzer displays: waiting for data Transmission can be interrupted at any time with the If the transmission was in spite of correct connection not successful, the analyzer displays: ATTENTION Transmission was not successful! The data flow was interrupted. The transmission has to be started once again, after restarting the PC transmission program. 27 key

28 7.5 Memory occupation In the window Measurement windows occupation the number of all, the occupied and free measurement data storage blocks will be indicated. 7.6 View stored data, : scroll measurements backwards/forwards : find a measurement by means of the site number : measuring data block : delete the selected measuring data block : back to data menu 7.7 Delete all measuring data All stored measurements will be deleted 28

29 7.8 Send measuring data to PC Transmission of the analyzer s internal data memory (not on SD) to PC. Note: To transfer the data of the SD card to PC, you have to pull out the SD card of the RASI card reader and insert this card in the PC card reader. The software OnlineView can open the data files from the SD card in the PC card reader. (For further information see manual OnlineView or Online help of the program) 7.9 Copy measuring data to SD card (option) Copy data memory to SD card Note: Only the SD card # (1 GB) can be used! This card is marked with a sticker on the backside. Other types have no guaranty for usable ness with analyzers! The SD-card option allows the storage of continuous measurements as well as the copying of the data memory onto the SD-card. These dates keep on being able to be processed with the software "Online View" at the personal computer. e.g. option SD card with analyzer RASI

30 Example: Use of the SD card with Analyzer: Insert the card up to the lock. For unloading the SD-card press once again seen on the illustration below. The card will unlock and you may pull it out Format SD card (option) All data on SD card will be deleted. 30

31 8 Main menu Xtras (settings) At delivery, the analyzer has a standard pre-adjusted configuration (factory default), which should cover your needs in the most cases. However, this configuration is highly flexible and individually adjustable. If you would like to modify different configurations, we recommend doing this in a reasonable way. The better the configuration modification is planned, the lesser corrective action must be taken and the easier you are able to work with the analyser. Take advantage of the possibility to individually adjust measurement programs, measuring indications, printings and self-selection fuel in a way, that all of your needs in practice are covered. Then less correction regarding the configurations has to be carried out. The modification is at once useable. The new configuration will be stored after switch off the analyzer (battery operated) or by pressing the - key (power supply operated. Press : in all other main menus for selection of main settings menu 8.1 AUX-channel configuration The AUX channel is an input for recording external signals. These will be sent by a transducer (4 20 ma type or 0 10 V type). This transducer can be supplied direct from the AUX input (12 Vdc, max. 400 ma). If the analyzer is potable or mobile, the connection of a thermocouple type K is allowed. It is possible to edit the name of the transducer, to assign a measuring range and measuring unit. The readed signal will be displayed together with the other values, stored and transmitted to a PC., : selection of the different positions (Input type, name unit, : select AUX input, one of following configurations : default values (factory settings = OFF) : modify the values : store the modified settings, back to the analyser settings 31

32 8.1.1 Configuration AUX input type, /.: select the type of input. Available: OFF: Thermo: No signal of an AUX input will be read A connected thermo couple will be read The positions unit, lower limit and upper limit are pre-selected by the analyser. Now it is possible to measure a temperature from 40 C to C ma: A connected 4 20 ma signal sensor will be read 0 10 V A connected 0 10 V signal sensor will be read : default values (factory settings = OFF) : store the modified settings, back to the analyser settings Terminal assignments 32

33 8.1.3 Measuring name modifies the text (see Chapter 10.1) of the measuring name (in the measuring display and on the print-out) Measuring unit (only with 0 10 V and 4 to 20 ma) Lower / upper limit (only with 0 10 V and 4 to 20 ma) Set the low end of measuring range of signal (0 V or 4 ma) Set the high end of measuring range of signal (10 V or 20 ma) Resolution Resolution of measurement display, will be calculated by the analyzer. Ex. Min. Ex. Max. Resolution

34 8.2 RS 232 settings If the continuous data transmission is activated, the measuring values will be transmitting without request to the serial interface., : Select the continuous data transmission Possible settings: ON/OFF OFF: the receiver has to demand the data! Master Slave ON device PC OFF PC device, : select the transmission protocol I.P. : industry protocol for 32-bit data logger software O.V.: Online View for Online view software Modbus slave the analyzer is able to work as modbus slave using the RS232 or RS485 port (possibly with external RS232/RS485 adapter) supports RS485 interface with 2/4 wires (half/full duplex) supports only the binary Modbus protocol (RTU) the slave modbus address is user definable from 0 to 238 communication parameter are user definable as follows: baud baud - even parity and 1 stop bit - no parity and 2 stop bits Further information with defined register on demand! Draft measurement PC Use for draft measurements the differential pressure + connector. The draft values will displayed and/or - if configured - printed as differential pressure values. Transmission to PC Select RS232 settings/diff. Pressure as Draft ON: The values will be transmitted to PC as DRAFT. 34

35 8.3 Analyzer settings, : select item of analyser to be modified (highlighted item), : modify selected item Set clock Allows operator to set time and date., : modify selected item (hours, minutes, seconds, day, month or year), : move selection cursor Setting display contrast Helping hints The display contrast is depending on the ambient temperature of analyzer (default = 0 % / 20 C). The analyzer compensates the temperature s influence, but there might remain a slight affect. Meaning: Display contrast, 0 % is normal with a temperature of about 20 C. Possible setting: -50 % to + 50 %, (in steps of 5 %) switch ON / OFF Possible settings: ON/OFF There are a couple of messages which are generated automatically by the analyzer and which help an initial user, but which may annoy an experienced user. Examples: 35

36 8.3.4 Country This function allows the change of display language. Following languages are pre-programmed in the firmware of the analyzer: Germany, France, England, Italia, Netherlands, Polska, Russia, USA, Japan, Romania, Turkey, Österreich, Belgium, Ceska,.: change country/language ATTENTION! A change of language/country also changes the list of fuel types and there parameters. As losses and efficiency are calculated with fuel parameters, previously stored measurements will be recalculated most probably with other parameters and will therefore result in other/wrong values. In this case you should change the language/country back in order to see correct storage contents Negative gas readings Select T-gas sensor type Switch on / off the negative gas readings with, keys. If the analyzer is operating for long time without a new reset to zero, the measuring values could drift from zero in both directions, negative and positive. HINT: If this function is switched on, it is a good chance, to see that the analyzer requires a new reset to zero! To see negative drift on the display, set Negative gas readings ON To hide negative drift on the display, set Negative gas readings OFF Available settings between thermocouple type K and type S selectable with, keys. T gas sensor type K: NiCrNi thermocouple T-gas-sensor type S: PtRhPt thermocouple Thermocouple type S is necessary for probes with temperature range up to C Reference temperature, : select 0 C, 15 C or 25 C In different countries, there are different local regulatory of standard gas condition. For example in Europe the norm cubic meter of gas is determined at 0 C, other countries require higher temperature for norm cubic meter (or feet)! This setting has influence on the calculation of gas concentrations in mg/m3, e.g. CO [mg/m3] and the norm flow volume [cfm]. Standard setting: 0 C, other settings: 15 C or 25 C Hold delay after purge, : select min. 0 sec. to max. 225 sec in 15 sec. steps In case of automatic reset to zero (or back purge of probe for stationary analyzers), it is possible to hold the last value of analog outputs 4 20 ma and in the measurement menu previous to these functions (see analog output configuration Chapter 8.6?) In addition to hold time during auto-zero (back 36 purge), analyzer can hold these values with a user selectable delay time up to max. 255 seconds.

37 8.4 Service values If an analyser malfunction should occur (e. g. message during calibration: " O 2 sensor not OK"), the malfunction can mostly be localized in the Service menu. The analogous values (= not calculated rough values) of all sensors will be displayed. Please contact our after-sales service in case of an error (see address / phone number on page 3). In order to localize the defective part, our technicians may ask you same service values or ask Eurotron Instruments (UK) ltd., : page turning, : move lines (in the first and the last line, the list will be moved = scroll) : switch between service values analog and infrared module.: tests: only for service department : trends.: only for service department : Internal heating: only for service department : back to the menu Xtras 37

38 8.5 Adjustment The adjustment menu is protected against use of non-authorized persons by a PIN code. For PIN code ask the service department. At wrong input of PIN code, the analyser will return to main settings menu Main adjustment menu Note: The adjustments temperature, pressure, AUX and analog output are factory made, it is generally not necessary to repeat them! Before starting the adjustment of gas measurement, it is advised to carry out a zeroing of this measurement. Connect analyser s inlet port to the gas calibration facility as in the picture below. Observe the flow of bypass calibration gas in the flow meter (4) and adjust regulator (2) for some 3 7 l/h. Cal gas will be sucked by the gas pump and has no pressure! A small excess of cal gas will be vent by the flow meter, insuring that no false air will be sucked by the gas pump. Pressurizing the electrochemical sensors will be damage them! 1 test gas cylinder 2 pressure reducer 3 T-connector 4 Flow-meter 5 Calibration gas inlet of analyser 38

39 8.5.2 Recommended calibration gases For adjustment, the following (analysis certified) calibration gas cylinders are recommended: range recommended cal gas manufacturer leak proof test Displayed O2 value = 0 0,1 % N2 = 100 % electrochem. sensors O ,0 Vol-% O2 = 10 % (for linerity error)) 10% CO (H 2 komp.) ppm, overload up to ppm CO = ppm rest N2 500 ppm Mixture (10 % O2), (800 ppm CO) and 850 ppm H2) rest H2 Gas mixer NO ppm, overload up to ppm NO = ppm rest N2 80 / 500 ppm NO ppm, overload up to ppm NO2 = ppm rest N2 50 ppm SO ppm, overload up to ppm SO2 = ppm rest N2 500 ppm H2S 0-50 ppm, overload up to 500 ppm H2S = 50 ppm rest N2 50 ppm H2 0-1 %, overload up to 2 % H2 = % rest N2 2% CO very high 0-4,00 %, overload up to 10,00 % CO = % rest N2 4% multi gas IR-bench CO CO 2 C X H Y equiv CH4 # 61447, 103 mm, CO, CO2, CH4 min ppm up to max ppm min. 0-3 % up to max % min ppm up to max ppm equiv. CH 4 CO approx % reading rest N2 CO2 approx % reading rest N2 CH4 approx % reading rest N2 Gas mixer Gas mixer Gas mixer multi gas IR-bench CO CO 2 C X H Y equiv CH4 # 58118, 65 mm, CO, CO2, CH4 min. 0-3 % up to max % min. 0-3 % up to max % min. 0-1 % up to max. 0-3 % equiv. CH 4 CO approx % reading rest N2 CO2 approx % reading rest N2 CH4 approx % reading rest N2 Gas mixer Gas mixer Gas mixer multi gas IR-bench CO CO 2 C X H Y equiv C3H8 # 58986, 65 mm, CO, CO2, C3H8 min. 0-3 % up to max % min. 0-3 % up to max % min ppm up to max ppm equiv. C 3 H 8 CO approx % reading rest N2 CO2 approx %readimg rest N2 C3H8 approx % reading rest N2 Gas mixer Gas mixer Gas mixer NOTE: Within the given gas concentration one chooses this cal gas with values like the real gas values! 39

40 8.5.3 Adjustment electrochemical sensors 1) Device with only O 2 measurement 2) Device with several electrochemical cells., : select the electrochemical cell / O2 sensor menu by moving the cursor : enter selected menu (highlighted line) Adjustment menu electrochemical sensors Manual calibration:, : select the gas to be calibrated, : change the factor with small steps ( 1digit steps) F1: CO off: not active F2: O2 offset: zero point offset correction of oxygen cell by means of N 2 calibration gas (sensor signal and display is decreasing to zero). Wait until the oxygen value is stable, then press F3: modify: change factor F4: IR: not active to adjust offset F5 or ENTER: end: end of adjustment with saving Adjustment CO (H2 compensated) - Connect the test gas cylinder and wait for measurement to stabilize - Select line for CO with keys, : - Adjust the span factor with keys, until the cylinder value and the display value are the same. H2 compensation: - Connect the test gas cylinder and wait for measurement to stabilize. - Select line for H 2 with keys, : - Adjust the correction factor for H 2 with keys and / or, until the CO value of cylinder and the display value for CO are the same. 40

41 8.5.4 Infrared bench adjustment Before starting the adjustment of gas measurement, it is advised to carry out a zeroing of this measurement (see chapter 8.5.5). Adjustment menu Infrared bench : activate NDIR bench adjustment : end: return to adjustment menu, : select the gas (in this example CH4) ENTER: F1 : confirm adjust: adjustment of the system to the values of the cylinder values which are set. The gases which cylinder values are zero will not be affected. The reaction time of the system is about 20 seconds. The adjustment of a gas only accomplished if between current value and cylinder value none to large difference (factor 3) is e.g. current 10ppm cylinder 1000 ppm => no adjustment F3: modify the values (value calibration gas cylinder) F5 or ENTER:: end of adjustment with saving. ATTENTION: The adjustment is not returnable by leaving the adjustment menu by means of the F5 -key! With the key F4 the display can be switch between electrochemical and NDIR adjustment! 41

42 Example: - Select the 3 gases CH4, CO und CO2 by means the keys, and set with F3 the cylinder values (CH4 = 1520, CO = 3342 und CO2 = 13,71) - Connect the test gas cylinder and wait for measurement to stabilize (min. 60 sec.) - then press F1 to adjust - the cylinder value (cylind) and the display value (current) are the same Reset to zero Before starting the adjustment of gas measurement, it is advised to carry out a zeroing of this measurement., : select line Reset to zero Adjustment menu Reset to zero : activate the function «Reset to zero» : continue: reset to zero and then return to adjustment menu.: abort: return to adjustment menu 42

43 8.6 Setup analog output (4 20 ma) Setup analog output allows configuring all 8 analog output channels 4 20 ma. To each of the analog output channels you can assign a measuring value and a measuring range., : move selection cursor, : print-out the current analog-output configuration (for usage at another place),, : change the setting selected by the cursor,, : back to the Xtras menu The example illustration shows the following setting: O2 [%] Channel 1 [ma] Setup analog output During zeroing This setting determines the analog output s behavior during zeroing: hold: The outputs will keep the last measured value before the zeroing. 2mA: The outputs will change to 2mA in order to indicate the zeroing state. This setting is common for all analog outputs; they will behave in the same way Setup analog output Output Channel Select the channel 1 through 8, which you want to configure. Below this line you can check or change this channel s assigned measuring value and range (min and max) Setup analog output Measuring value Here you can assign a measuring value to the selected Output Channel. All values are assignable, directly measured values e.g. CO [ppm] as well as calculated values e.g. Exc.Air Setup analog output Minimal value (lower range end, correspond to 4 ma) This setting determines the lower output range s end value, which corresponds to 4mA. If the assigned measuring value falls below this end value, the output will stay on 4mA Setup analog output Maximal value (upper range end, correspond to 20 ma) This setting determines the upper output range s end value, which corresponds to 20mA. If the assigned measuring value exceeds this end value, the output will stay on 20mA. 43

44 9 Main menu Info This menu contains important data of the analyzer for the service department. 9.1 Unit status Information about the actual temperature components measured with the analyzer. 9.2 Device options The information about the actual device options will be displayed. example: EC (1%) = electrochemical sensor resolution 1 % IR = Infrared bench 44

45 10 General, functions 10.1 Text edit A number of text and denominations can be edited in the analyser (e.g. the names of the self definable fuel types, customer numbers, and names of pressure). When starting the edit function, the following display appears: Input cursor Selection cursor, : select the character, : select the character / : cancel, character will not be taken.: character above or left of the input cursor will be deleted : character selected by the selection cursor will be inserted or : character selected by the selection cursor will be overwritten : end of text edit with saving 10.2 Quick input of numbers, : select the decimal place, : select the number : abort the function : set the number in the frame to zero : set the sign character (not used) : end of input, input will be saved 10.3 Request user decision At several functions the analyzer requests for a user decision., : select the line / : abort the function :: end of input, input will be saved 45

46 46

Configuration of C310 multifunction transmitters by keypad

Configuration of C310 multifunction transmitters by keypad Configuration of C310 multifunction transmitters by keypad Table of contents 1. Introduction...5 1.1. Description of the transmitter...5 1.2. Description of the keys...5 1.3. Protection tips of the sensor...6

More information

Configuration of class 210 transmitters by keypad

Configuration of class 210 transmitters by keypad Configuration of class 210 transmitters by keypad Table of contents 1. Introduction...5 1.1. Description of the transmitter...5 1.2. Description of the keys...5 1.3. Selection of the output signal...6

More information

Configuration of CPE 310-S and CPE 311-S transmitters by keypad

Configuration of CPE 310-S and CPE 311-S transmitters by keypad Configuration of CPE 310-S and CPE 311-S transmitters by keypad Table of contents 1. Introduction...5 1.1. Description of the transmitter...5 1.2. Description of the keys...5 1.3. Protection tips of the

More information

User manual. paper moisture meter RH5.1 with sword-sensor

User manual. paper moisture meter RH5.1 with sword-sensor User manual paper moisture meter RH5.1 with sword-sensor version 2.0_en Schaller GmbH 2012 User manual Positioning the instrument Insert the sword-sensor into the stack for only approx. 10 cm, and push

More information

Operating Manual. Differential Pressure Transmitter DPS 100

Operating Manual. Differential Pressure Transmitter DPS 100 Operating Manual Differential Pressure Transmitter DPS 100 Important notes: Please read this operating manual carefully before installing and starting up the pressure transmitter. This operating manual

More information

Senseair S8 Commercial

Senseair S8 Commercial Product Specification Senseair S8 Commercial Miniature infrared CO 2 sensor module 1 (8) Item Senseair S8 Commercial Article No. 004-0-0010, 004-0-0075 Target gas CO 2 Operating Principle Non-dispersive

More information

Product Specification SenseAir S8

Product Specification SenseAir S8 Product Specification SenseAir S8 Miniature CO 2 Sensor module Driesen + Kern GmbH SenseAir S8 Miniature infrared CO 2 sensor module Warning! ESD sensitive device! Figure 1. SenseAir S8 Article no. 004-0-005

More information

ADI-100 Interrupter. Operator s Manual. 526 S. Seminole Bartlesville, OK /

ADI-100 Interrupter. Operator s Manual. 526 S. Seminole Bartlesville, OK / ADI-100 Interrupter Operator s Manual 526 S. Seminole Bartlesville, OK 74003 918/336-1221 www.sescocp.com ADI - 100 Interrupter FEATURES Clock Accuracy 100% during GPS lock Clock Drift 30 µsec per degree

More information

User Manual. PolyGard AT-2000 V3. Infrared Refrigerant Gas Transmitter Serial No. AT March, 2009

User Manual. PolyGard AT-2000 V3. Infrared Refrigerant Gas Transmitter Serial No. AT March, 2009 PolyGard AT-2000 V3 Infrared Refrigerant Gas Transmitter Serial No. AT-2000 User Manual March, 2009 February 16, 2016 Revision Specifications subject to change without notice. USA 160216 Page 2 of 13 1

More information

TABLE OF CONTENTS. Electromatic Equipment Co., Inc. -1-

TABLE OF CONTENTS. Electromatic Equipment Co., Inc. -1- TABLE OF CONTENTS 1.0 Introduction 2 1.1 Calibration curves 1.2 Explanation of calibration curves 2.0 Operating the instrument. 3 3.0 View of the instrument. 4 4.0 Menu.. 5 5.0 Other symbols and key functions..

More information

USER MANUAL ENGLISH 1450 COIN COUNTER & SORTER

USER MANUAL ENGLISH 1450 COIN COUNTER & SORTER USER MANUAL ENGLISH 1450 COIN COUNTER & SORTER INTRODUCTION ENGLISH Thank you for purchasing the Safescan 1450 coin counter and sorter. For proper use and maintenance, we advise to read this user manual

More information

CDG. Carbon Dioxide - User Manual Gas Detector. 1 Intended Use. 2 Functional Description. 3 Installation. 4 Electrical Connection.

CDG. Carbon Dioxide - User Manual Gas Detector. 1 Intended Use. 2 Functional Description. 3 Installation. 4 Electrical Connection. Carbon Dioxide - User Manual 1 Intended Use 2 Functional Description 2.1 Control Mode 2.2 Sensor 3 Installation 3.1 Mounting instructions 3.2 Installation 4 Electrical Connection 4.1 Instruction 5 Commissioning

More information

Product Specification. SenseAir S8 LP. Miniature infrared CO 2 sensor module. Document PSP 126. Rev 8. Page 1 (8)

Product Specification. SenseAir S8 LP. Miniature infrared CO 2 sensor module. Document PSP 126. Rev 8. Page 1 (8) Product Specification SenseAir S LP Miniature infrared CO 2 sensor module 1 () Item SenseAir S LP Article no. 004-0-0053 Target gas CO 2 Operating Principle Non-dispersive infrared (NDIR) Measurement range

More information

PWM regulator for electrolysis cell Η 2

PWM regulator for electrolysis cell Η 2 PWM regulator for electrolysis cell Η 2 Ο Red or Gray Battery positive end (+) via fuse 25Α (30Α max) Blue Positive PWM output towards solution 1 Brown Negative PWM output towards solution Solution state

More information

User Manual TH300 > CTV310 > TH300 > < TT300 TT300 > Air velocity. Temperature Humidity. Temperature. Pressure CP300 > Standard probe.

User Manual TH300 > CTV310 > TH300 > < TT300 TT300 > Air velocity. Temperature Humidity. Temperature. Pressure CP300 > Standard probe. User Manual Configuration of Class 00 Transmitters New New Keypad Remote control network Temperature Humidity Air velocity TH00 Remote probe Pressure TH00 Standard probe CTV0 Standard probe Temperature

More information

Product Specification. SenseAir S8 LP. Miniature infrared CO 2 sensor module. Document PSP 126. Rev 4. Page 1 (10)

Product Specification. SenseAir S8 LP. Miniature infrared CO 2 sensor module. Document PSP 126. Rev 4. Page 1 (10) Product Specification SenseAir S8 LP Miniature infrared CO 2 sensor module Document PSP 126 Rev 4 Page 1 (10) SenseAir S8 LP Miniature infrared CO 2 sensor module Warning! ESD sensitive device! Figure

More information

Warning! ESD sensitive device!

Warning! ESD sensitive device! Datasheet & Manual SenseAir S8 Miniature CO2 sensor Model SE-0031 3.2% CO2 Warning! ESD sensitive device! Document PSP 108 Key technical specifications Item Target gas Operating Principle SenseAir S8 CO2

More information

Gas and Air Sensors. Product Specification. SenseAir. Miniature CO2 sensor

Gas and Air Sensors. Product Specification. SenseAir. Miniature CO2 sensor Gas and Air Sensors Product Specification SenseAir S8 Miniature CO2 sensor SenseAir S8 Miniature infrared CO 2 sensor module Warning! ESD sensitive device! Figure 1. SenseAir S8 Article no. 004-0-0050

More information

S8 Residential pin header

S8 Residential pin header Product Specification S8 Residential pin header Miniature infrared CO sensor module Item S8 Residential pin header Article no. 004-0-0056 Target gas CO Operating Principle Measurement range Measurement

More information

Instruction Manual HAS3xE-IM-SW39 10/2005

Instruction Manual HAS3xE-IM-SW39 10/2005 Instruction Manual HAS3xE-IM-SW39 10/2005 Software Version 3.9.x NGA 2000 Software Manual for MLT or CAT 200 Analyzer and MLT or CAT 200 Analyzer Module (combined with NGA 2000 Platform, MLT, CAT 200 or

More information

Instruction also available on

Instruction also available on TERA Radon Program EN TCR3 Central Unit Technical Specifications & Operation Manual v.2 2016 Table of Contents 1 Introduction...2 2 Description and Utilization...2 3 Scope of Delivery...4 4 Product Specification...5

More information

Air flow meter For ventilation and air-conditioning Model A2G-25

Air flow meter For ventilation and air-conditioning Model A2G-25 Flow ir flow meter For ventilation and air-conditioning Model 2G-25 WIK data sheet SP 69.04 for further approvals see page 5 pplications For measuring the air flows of radial ventilators For measuring

More information

PhenoMaster / LabMaster Phenotyping Research Platform

PhenoMaster / LabMaster Phenotyping Research Platform Operating Instructions PhenoMaster / LabMaster Phenotyping Research Platform SOFTWARE Operating Instructions Calibration Description of Standard Components 2 PhenoMaster/LabMaster Software Operating Instructions

More information

Driesen + Kern GmbH. Product Specification. SenseAir S8. Commercial. Miniature infrared CO 2 sensor module. Document Rev Page 1 (10)

Driesen + Kern GmbH. Product Specification. SenseAir S8. Commercial. Miniature infrared CO 2 sensor module. Document Rev Page 1 (10) Driesen + Kern GmbH Product Specification SenseAir S8 Commercial Miniature infrared CO 2 sensor module 1 (10) SenseAir S8 Commercial Miniature infrared CO 2 sensor module Warning! ESD sensitive device!

More information

Data sheet and manual. SenseAir S8. Alarm 5% Miniature infrared CO 2 sensor module

Data sheet and manual. SenseAir S8. Alarm 5% Miniature infrared CO 2 sensor module Data sheet and manual SenseAir S8 Alarm 5% Miniature infrared CO 2 sensor module Key technical specification Item SenseAir S8 Alarm 5% Article no. 004-0-001 Target gas CO 2 Operating Principle Non-dispersive

More information

Temp / RH / CO 2 Hand-Held Meter

Temp / RH / CO 2 Hand-Held Meter Temp / RH / CO 2 Hand-Held Meter PRODUCT MANUAL Item # 3440 CONTENTS Specifications 2 Battery 3 LCD Display 3 Keypad 3 Operation 4 Troubleshooting 7 Warranty 8 CE Declaration of Conformity 8 Thank you

More information

Temp. & humidity indicator

Temp. & humidity indicator Temp. & humidity indicator AH8008 Product Manual www.aosong.com 1 Product Overview AH8008 handheld multi-function temperature and humidity instrumentation consists of two parts: the AH8008 instrument and

More information

DP2500 DP0100 DP0250 Differential Pressure Transmitter

DP2500 DP0100 DP0250 Differential Pressure Transmitter PB_DP_03 2011 DP2500 DP0100 DP0250 Differential Pressure Transmitter The DP Low Differential Pressure Transmitter series is an accurate and cost competitive solution for measuring low pressures of air

More information

Global Water Instrumentation, Inc.

Global Water Instrumentation, Inc. Global Water Instrumentation, Inc. 151 Graham Road P.O. Box 9010 College Station, TX 77842-9010 T: 800-876-1172 Int l: (979) 690-5560, Fax: (979) 690-0440 E-mail : globalw@globalw.com PC320 Process Controller

More information

Temperature Humidity < THA 300

Temperature Humidity < THA 300 User Manual Presure Temperature Humidity Air Velocity Air Flow USER MANUAL Configuration of CPA00 and THA00 New New Remote control MODBUS System Temperature Humidity < THA 00 Pressure CPA 00 > Summary.

More information

Gas and Air Sensors. Product Specification. SenseAir S8. Miniature CO 2 sensor module

Gas and Air Sensors. Product Specification. SenseAir S8. Miniature CO 2 sensor module Gas and Air Sensors Product Specification SenseAir S8 Miniature CO 2 sensor module SenseAir S8 Miniature infrared CO 2 sensor module Warning! ESD sensitive device! Figure 1a. SenseAir S8 Article no. 004-0-0013

More information

Gas and Air Sensors. Product Specification. SenseAir S8. Miniature CO2 sensor. Document PSP 107. Rev 8. Page 1 (10)

Gas and Air Sensors. Product Specification. SenseAir S8. Miniature CO2 sensor. Document PSP 107. Rev 8. Page 1 (10) Gas and Air Sensors Product Specification SenseAir S8 Miniature CO2 sensor 8 1 (10) SenseAir S8 Miniature infrared CO 2 sensor module Warning! ESD sensitive device! Figure 1. SenseAir S8 Article no. 004-0-0013

More information

Driesen + Kern GmbH. Product Specification. SenseAir S8. Miniature CO 2 sensor. Document PSP 113. Rev 3. Page 1 (12)

Driesen + Kern GmbH. Product Specification. SenseAir S8. Miniature CO 2 sensor. Document PSP 113. Rev 3. Page 1 (12) Driesen + Kern GmbH Product Specification SenseAir S8 Miniature CO 2 sensor 1 (12) SenseAir S8 Miniature infrared CO 2 sensor module Warning! ESD sensitive device! Figure 1. SenseAir S8 Article no. 004-0-0017

More information

Multi-Channel In-Out Thermometer with Cable Free Sensor and RF Clock

Multi-Channel In-Out Thermometer with Cable Free Sensor and RF Clock Multi-Channel In-Out Thermometer with Cable Free Sensor and RF Clock MAIN FEATURES: MAIN UNIT GB MODEL: RMR182 USER'S MANUAL INTRODUCTION Congratulations on your purchase of the RMR182 Multi- Channel In-Out

More information

GE Sensing. Introduction. Wiring Diagrams (Typical) Field Calibration. Installing the Sensor

GE Sensing. Introduction. Wiring Diagrams (Typical) Field Calibration. Installing the Sensor clear mode enter clear mode enter GE Sensing Introduction The GE Telaire Vaporstat 900 sensor measures in applications in the range of 0 to 0 F dew point. The sensor package is designed for wall mounting.

More information

Model OI-6940 Notis Quad 4-Gas Sensor Assembly

Model OI-6940 Notis Quad 4-Gas Sensor Assembly Model OI-6940 Notis Quad 4-Gas Sensor Assembly Operation Manual Revision 2.3w Product Overview The Otis Instruments, Inc. Gen II WireFree OI-6940 Notis Quad is a battery-powered explosion-proof 4-gas sensor

More information

PROCESS ANALYSERS. SERVOFLEX MiniFoodPack Gas Analyser. Operator Manual. Part Number: A Revision: 4 Language: UK English

PROCESS ANALYSERS. SERVOFLEX MiniFoodPack Gas Analyser. Operator Manual. Part Number: A Revision: 4 Language: UK English PROCESS ANALYSERS SERVOFLEX MiniFoodPack Gas Analyser Operator Manual Part Number: 05240001A Revision: 4 Language: UK English This page intentionally blank This page intentionally blank SERVOMEX - SPECIALISTS

More information

0.01kV (0 to ±0.99 kv)

0.01kV (0 to ±0.99 kv) Handheld Electrostatic Meter Series IZH10 Easy-to-use handheld electrostatic meter Rated charge amount range: ±20.0 kv Minimum display unit: 0.1kV (±1.0 to ±20.0 kv) Check the current situation before

More information

Driesen + Kern GmbH. SenseAir S8

Driesen + Kern GmbH. SenseAir S8 SenseAir S8 SenseAir S8 Miniature infrared CO 2 sensor module Warning! ESD sensitive device! SenseAir S8 Article no. 004-0-0050 SenseAir S8 Article no. 004-0-0051 2 (10) Key technical specification Item

More information

Flowmetering System Controller KMA4000 / PLU4000 Operating instructions

Flowmetering System Controller KMA4000 / PLU4000 Operating instructions System Controller KMA4000 / PLU4000 Operating instructions AVL Pierburg Flow Technology GmbH D-41460 Neuss Leuschstr. 11 Phone:+49 2131 40256-0 Fax:+49 2131 40256-99 Notes Before connecting and operating

More information

EdvoCycler & MegaCycler

EdvoCycler & MegaCycler Operation Manual: EdvoCycler & MegaCycler Cat. #541 and #542 Features: 25 x 0.2 ml Tube Block (EdvoCycler) 49 x 0.2 ml Tube Block (MegaCycler) Heated Oil-Free Lid with Magnetic Latch Vivid 7 Line LCD Display

More information

M O D U L E - 7 D Model CS-CAL/DLIU V6.0 Operator s Manual

M O D U L E - 7 D Model CS-CAL/DLIU V6.0 Operator s Manual O P E R A T I O N S A N D P R O C E D U R E S F O R C S - C A L / D L I U V 6. 0 A N D A C C E S S O R I E S Model CS-CAL/DLIU V6.0 Visual screen display allows easy step-by-step operation Rechargeable

More information

GREISINGER electronic GmbH D Regenstauf, Hans-Sachs-Straße 26 phone: / , fax: / ,

GREISINGER electronic GmbH D Regenstauf, Hans-Sachs-Straße 26 phone: / , fax: / , H66.0.01.6C-05 Operating Manual Conductivity measuring instrument from Version 1.2 GLF 100 GREISINGER electronic GmbH D - 93128 Regenstauf, Hans-Sachs-Straße 26 phone: +49 9402 / 9383-0, fax: +49 9402

More information

Data Acquisition Modules/ Distributed IO Modules

Data Acquisition Modules/ Distributed IO Modules User Manual Data Acquisition Modules/ Distributed IO Modules Future Design Controls, Inc. 7524 West 98 th Place / P.O. Box 1196 Bridgeview, IL 60455 888.751.5444 - Office: 888.307.8014 - Fax 866.342.5332

More information

Table of Contents. Instruction Manual - TORBAL DRX 5 Series

Table of Contents. Instruction Manual - TORBAL DRX 5 Series Table of Contents Chapter 1.CAUTIONARY NOTES AND PRECAUTIONS... 3 Chapter 2.SPECIFICATIONS... 4 Chapter 3.Keys and display indicators... 5 Chapter 4.Parts Description... 6 Chapter 5.Unpacking the balance

More information

SenseAir S8 Miniature infrared CO 2 sensor module

SenseAir S8 Miniature infrared CO 2 sensor module SenseAir S SenseAir S Miniature infrared CO 2 sensor module Warning! ESD sensitive device! SenseAir S Article no. 004-0-0010 SenseAir S Article no. 004-0-0011 2 (10) Item Target gas Operating Principle

More information

High-precision process calibrator Model CED7000

High-precision process calibrator Model CED7000 Calibration technology High-precision process calibrator Model CED7000 WIKA data sheet CT 85.51 Applications Research and development laboratories Calibration service companies and service industry Industry

More information

Please enter the identity code of your device here!

Please enter the identity code of your device here! Operating Instructions DULCOMETER D1C Part 2: Adjustment and Operation, Measured Variable Conductive Conductivity ProMinent D1C2-Leit.-001-GB Conductivity 100 µs/cm DULCOMETER STOP START Conductivity 100

More information

FW100 Dust Measurement Device

FW100 Dust Measurement Device product InForMAtIon FW100 Dust Measurement Device Preventing fine dust polluting the air by using scattered-light technology Analyzers and Process Instrumentation FW100 using Scattered Light Technology

More information

Alemite FCS Fluid Control System

Alemite FCS Fluid Control System 343000 343000-1 Alemite FCS Fluid Control System Installation and Operation Guide 670996 TABLE OF CONTENTS DISCLAIMER... 3 BEFORE YOU BEGIN........................................................ 4 1.0

More information

LCC-100 software User manual

LCC-100 software User manual User Manual Pressure Temperature Humidity Air Velocity Airflow Sound level New LCC-100 software User manual Configuration software for Monostats - Class 50 and 100 units - ranges - relays - set points

More information

200 us/2 ms/20 ms/200 ms, TDS ( PPM ) Real time data logger, Data logger no., RS232 CONDUCTIVITY METER Model : YK-2005CD

200 us/2 ms/20 ms/200 ms, TDS ( PPM ) Real time data logger, Data logger no., RS232 CONDUCTIVITY METER Model : YK-2005CD 200 us/2 ms/20 ms/200 ms, TDS ( PPM ) Real time data logger, 16000 Data logger no., RS232 CONDUCTIVITY METER Model : YK-2005CD TABLE OF CONTENTS 1. FEATURES...1 2. SPECIFICATIONS... 2 2-1 General Specifications...2

More information

Characteristics and functioning

Characteristics and functioning Characteristics and functioning 1/36 enod4-d Characteristics and functioning NU-eNod4D-ETH-E-1014_216710-A 1 ENOD4 PRODUCT RANGE... 5 1.1 General presentation... 5 1.2 Versions and options... 5 1.2.1 Versions...

More information

Power Meter. Measurement Guide. for Anritsu RF and Microwave Handheld Instruments BTS Master Site Master Spectrum Master Cell Master

Power Meter. Measurement Guide. for Anritsu RF and Microwave Handheld Instruments BTS Master Site Master Spectrum Master Cell Master Measurement Guide Power Meter for Anritsu RF and Microwave Handheld Instruments BTS Master Site Master Spectrum Master Cell Master Power Meter Option 29 High Accuracy Power Meter Option 19 Inline Peak

More information

Technical Specification Temperature-Humidity-Pressure Sensor

Technical Specification Temperature-Humidity-Pressure Sensor Technical Specification ITEM CODE RELEASE DATE AUTHOR B80951-01 24 th October 2017 T. Thiesing REVISION REVISION DATE VERSION AUTHOR B-COMMAND GmbH Gruetzmuehlenweg 46 22339 Hamburg Germany FON: +49 40

More information

Bedienungsanleitung Operating manual Mode d emploi Instrucciones de operación

Bedienungsanleitung Operating manual Mode d emploi Instrucciones de operación Bedienungsanleitung Operating manual Mode d emploi Instrucciones de operación ba7542defs M PRT CAL STO O ESC Universal-Taschenmessgerät Seite 3 Universal Pocket Meter Page 7 Instrument de poche universel

More information

( F L O W I Z F A M I L Y )

( F L O W I Z F A M I L Y ) THE MOST ACCURATE BATTERY POWERED SYSTEM ( F L O W I Z F A M I L Y ) E l e c t r o m a g n e t i c c o n v e r t e r p o w e r e d b y b a t t e r i e s, s o l a r p a n e l o r D C p o w e r w i t h 4

More information

MODBUS for mamos analyser manual v

MODBUS for mamos analyser manual v sales@.com MODBUS for mamos analyser v.1.7.5 2017-05 1 / 27 sales@.com 1 Basic connection data... 2 2 General information about modbus... 3 3 Modbus connection... 4 4 Types of data available via modbus:...

More information

JUMO Wtrans E01. Measuring probe for humidity, temperature, and CO 2 with wireless data transmission. Brief description. Universal Wtrans receiver

JUMO Wtrans E01. Measuring probe for humidity, temperature, and CO 2 with wireless data transmission. Brief description. Universal Wtrans receiver Page 1/13 JUMO Wtrans E01 Measuring probe for humidity, temperature, and CO 2 with wireless data transmission Humidity from 0 to 100 % RH (incl. -40 to +80 C) or CO 2 from 0 to 2000/5000/10000 ppm or Temperature

More information

BS-3520 Digital Indicator

BS-3520 Digital Indicator BS-3520 Digital Indicator The Better Way for Weighing & easurements Table of Contents. Introduction...3 - Trait... 3-2 Warning... 4 2. Specification...5 3. External Size...6 4. Description on Front Panel...7

More information

PalmGauss SC PGSC-5G. Instruction Manual

PalmGauss SC PGSC-5G. Instruction Manual PalmGauss SC PGSC-5G Instruction Manual PalmGauss SC PGSC 5G Instruction Manual Thank you very much for purchasing our products. Please, read this instruction manual in order to use our product in safety

More information

OVEN INDUSTRIES, INC. Model 5C7-362

OVEN INDUSTRIES, INC. Model 5C7-362 OVEN INDUSTRIES, INC. OPERATING MANUAL Model 5C7-362 THERMOELECTRIC MODULE TEMPERATURE CONTROLLER TABLE OF CONTENTS Features... 1 Description... 2 Block Diagram... 3 RS232 Communications Connections...

More information

DP2500 DP0100 DP0250 Differential Pressure Transmitter

DP2500 DP0100 DP0250 Differential Pressure Transmitter PB_DPT_03 2010 DP2500 DP0100 DP0250 Differential Pressure Transmitter The DP Low Differential Pressure Transmitter series is an accurate and cost competitive solution for measuring low pressures of air

More information

Kurz Instruments Inc. 19 August 2010

Kurz Instruments Inc. 19 August 2010 BUILT-IN ZERO-MIDSPAN-SPAN DRIFT CHECK/CALIBRATOR Introduction Continuous Emissions Monitoring Systems (CEMS) have a long term stability verification requirement. Most of the equipment is verified using

More information

MD-5060 DIGITAL MEGOHMMETER

MD-5060 DIGITAL MEGOHMMETER DIGITAL MEGOHMMETER ( M98126501-03 / 04A ) TECHNICAL SPECIFICATIONS OPERATING INSTRUCTIONS DESCRIPTION The CIRCUTOR model MD-5060 digital electronic ohmmeter is a versatile, robust and easy-to-use equipment.

More information

INSTRUCTION MANUAL INF Fax: (503)

INSTRUCTION MANUAL INF Fax: (503) INSTRUCTION MANUAL INF151 1-800-547-5740 Fax: (503) 643-6322 www.ueiautomotive.com email: info@ueitest.com Introduction Congratulations on your purchase of the INF151 infrared thermometer. Like all UEi

More information

Moving money forward. CASSIDA TillTally + TillTally Elite Money Counting Scales

Moving money forward. CASSIDA TillTally + TillTally Elite Money Counting Scales Moving money forward CASSIDA TillTally + TillTally Elite Money Counting Scales Table of contents: 1. INTRODUCTION 1.1 About the Cassida TillTally 2 1.2 Box contents 2 1.3 Front and rear views 3 1.4 Display

More information

BusWorks 900EN Series Modbus TCP/IP 10/100M Industrial Ethernet I/O Modules

BusWorks 900EN Series Modbus TCP/IP 10/100M Industrial Ethernet I/O Modules BusWorks 900EN Series Modbus TCP/IP 10/100M Industrial Ethernet I/O Modules Six Differential Current Inputs Six Differential Voltage Inputs USER S MANUAL ACROMAG INCORPORATED Tel: (248) 295-0880 30765

More information

Atlas 5TM. Instruction Manual. Handheld Indoor Air Quality Monitor. Measures CO 2, Temperature and Humidity

Atlas 5TM. Instruction Manual. Handheld Indoor Air Quality Monitor. Measures CO 2, Temperature and Humidity Atlas 5TM Handheld Indoor Air Quality Monitor Instruction Manual Measures C 2, Temperature and Humidity ITRDUCTI Thank you for purchasing this portable meter. This device measures ppm levels, air temperature

More information

Please enter the identity code of your device here!

Please enter the identity code of your device here! Operating Instructions DULCOMETER D1C Part 2: Adjustment and Operation, Measured Variable Ozone ProMinent D1C2-03-001-GB O 3 DULCOMETER STOP START O 3 DULCOMETER STOP START Type D Type W D1C A Please enter

More information

R PROFLAME Instruction Book Collection

R PROFLAME Instruction Book Collection 9.956.028 R00 584 PROFLAME Instruction Book Collection 4-17 18-29 584 PROFLAME System 30-39 Appendix: DIP SWITCH NUMBER (0=ON 1=OFF) 40-41 4-17 Fig. 1 The SIT is a device that allows, in conjunction with

More information

JUMO Wtrans Receiver with Wireless Data Transmission

JUMO Wtrans Receiver with Wireless Data Transmission Data Sheet 902931 Page 1/10 JUMO Wtrans Receiver with Wireless Data Transmission k For measuring temperature, pressure, potentiometer, and voltage. k Interface RS485 with Modbus protocol k Wireless measured

More information

EIB/KNX Switch Actuators. User manual

EIB/KNX Switch Actuators. User manual EIB/KNX Switch Actuators User manual IT KNT 004 IT KNT 012 Tel.: +34943627988 E-mail: knx@dinuy.com Web: www.dinuy.com Contents 1. Introduction --------------------------------------------------------------------------------------------------------------

More information

Submersible pressure sensor For water and wastewater Model LW-1

Submersible pressure sensor For water and wastewater Model LW-1 Replacement product: Model LF-1 Level measurement Submersible pressure sensor For water and wastewater Model LW-1 WIKA data sheet LM 40.03 Applications Level measurement in rivers and lakes Deep well and

More information

KM-4800w. Copy/Scan Operation Manual

KM-4800w. Copy/Scan Operation Manual KM-4800w Copy/Scan Operation Manual NOTE: This Operation Manual contains information that corresponds to using both the metric and inch versions of these machines. The metric versions of these machines

More information

Portable Appliance Testers. OmegaPAT MI 2140 BetaPAT MI 2141 User Manual Ver Code No

Portable Appliance Testers. OmegaPAT MI 2140 BetaPAT MI 2141 User Manual Ver Code No Portable Appliance Testers OmegaPAT MI 2140 BetaPAT MI 2141 User Manual Ver. 1.2. Code No. 20 750 684 Distributor: Producer: METREL d.d. Ljubljanska 77 SI-1354 Horjul E-mail: metrel@metrel.si http://www.metrel.si

More information

GE Infrastructure Sensing. Druck DPI 841/842. Frequency calibrator and Frequency loop calibrator User manual - K395

GE Infrastructure Sensing. Druck DPI 841/842. Frequency calibrator and Frequency loop calibrator User manual - K395 GE Infrastructure Sensing Druck DPI 841/842 Frequency calibrator and Frequency loop calibrator User manual - K395 A1 B1 10 1 A 2 9 A 3 8 2 3 7 6 11 4 5 B1 12 A2 DPI 842 13 15 14 A3 19 18 17 16 27 20 21

More information

UNIVERSAL MOISTURE TESTER HB-300. Operating Manual

UNIVERSAL MOISTURE TESTER HB-300. Operating Manual UNIVERSAL MOISTURE TESTER HB-300 Operating Manual Contents 1. UNIVERSAL MOISTURE TESTER HB-300...4 2. FEATURES... 5 3. SPECIFICATIONS...6 4. NOMENCLATURE... 7 5. DISPLAY SCREEN...8 6. OPERATION PANEL

More information

AUTOMATION. Operator s Manual. IRU-2xx4/3xx4 Series. Rev. A2, 11/08 Doc

AUTOMATION. Operator s Manual. IRU-2xx4/3xx4 Series. Rev. A2, 11/08 Doc AUTOMATION P R O D U C T S G R O U P, I N C. Operator s Manual IRU-2xx4/3xx4 Series Rev. A2, 11/08 Doc. 9002673 Tel: 1/888/525-7300 Fax: 1/435/753-7490 www.apgsensors.com E-mail: sales@apgsensors.com IRU-2xx4/3xx4

More information

2F. No.25, Industry E. 9 th Rd., Science-Based Industrial Park, Hsinchu, Taiwan Application Note of OGM220, AN001 V1.8

2F. No.25, Industry E. 9 th Rd., Science-Based Industrial Park, Hsinchu, Taiwan Application Note of OGM220, AN001 V1.8 Application Note of OGM220, AN001 V1.8 1.0 Introduction OGM220 series is a dual channels NDIR module having a digital output directly proportional to CO2 concentration. OGM220 is designed for multi-dropped

More information

SERVICE-MANUAL ROOM TEMPERATURE CONTROLLER NEA 230V / 24V. Construction Automotive Industry

SERVICE-MANUAL ROOM TEMPERATURE CONTROLLER NEA 230V / 24V. Construction Automotive Industry ROOM TEMPERATURE CONTROLLER NEA 230V / 24V Construction Automotive Industry ROOM TEMPERATURE CONTROLLER NEA 230V / 24V TABLE OF CONTENTS 1....... Information and safety guidelines......................................................

More information

User Manual. PolyGard ADTD Infrared Carbon Dioxide Transmitter Serial No. SM03_00X. October, 2011

User Manual. PolyGard ADTD Infrared Carbon Dioxide Transmitter Serial No. SM03_00X. October, 2011 PolyGard Infrared Carbon Dioxide Transmitter Serial No. SM0_00X User Manual October, 0 GAADTD6_E_0 Page Intended Use... Functional Description.... Control Mode.... Sensor... Installation.... Mounting Instructions....

More information

IP 251 Universal Signal Converter SSI parallel RS232 parallel SSI RS232

IP 251 Universal Signal Converter SSI parallel RS232 parallel SSI RS232 control motion interface ELEKTRO-TRADING sp. z o.o Tel. +48 (0-32) 734-55-72 Tel/Fax +48 (0-32) 734-55-70 E-Mail et@elektro-trading.com.pl http://www.elektro-trading.com.pl IP 251 Universal Signal Converter

More information

RSMFX-2R MULTIFUNCTIONAL

RSMFX-2R MULTIFUNCTIONAL Mounting and operating instructions Table of contents SAFETY AND PRECAUTIONS 3 PRODUCT DESCRIPTION 4 ARTICLE CODES 4 INTENDED AREA OF USE 4 TECHNICAL DATA 4 STANDARDS 4 OPERATIONAL DIAGRAMS 5 WIRING AND

More information

CS105 BAROMETRIC PRESSURE SENSOR

CS105 BAROMETRIC PRESSURE SENSOR REVISION: 3/97 COPYRIGHT (c) 1995-1997 CAMPBELL SCIENTIFIC, INC. WARRANTY AND ASSISTANCE The CS105 BAROMETRIC PRESSURE SENSOR is warranted by CAMPBELL SCIENTIFIC, INC. to be free from defects in materials

More information

Operation. 1 of 14. Serial number label

Operation. 1 of 14. Serial number label Telecody Keypad Transmitter (US) WARNING There is a possible risk of injury or death if safety instructions will not be observed. Please read and follow all instructions in this manual as well as the information

More information

HC500 Heater Controller

HC500 Heater Controller HC500 Heater Controller Function description - basics HETRONIK GmbH Heisinger Str. 12 D-87437 Kempten / Germany phone: +49 / (0)831-56 58 59-34 fax: +49 / (0)831-56 58 59-39 e-mail: contact@hetronik.de

More information

Fluke 125. Getting Started. Industrial ScopeMeter

Fluke 125. Getting Started. Industrial ScopeMeter Fluke 125 Industrial ScopeMeter Getting Started GB Dec 2006, Rev. 1, 09/2009 2006, 2009 Fluke Corporation, All rights reserved. Printed in The Netherlands All product names are trademarks of their respective

More information

Carbon Monoxide (CO) and Carbon Dioxide (CO 2 ) Meter With Air Temp/RH Measurements and Dew Point/Wet Bulb calculations

Carbon Monoxide (CO) and Carbon Dioxide (CO 2 ) Meter With Air Temp/RH Measurements and Dew Point/Wet Bulb calculations User Manual Carbon Monoxide (CO) and Carbon Dioxide (CO 2 ) Meter With Air Temp/RH Measurements and Dew Point/Wet Bulb calculations Model CO260 Additional User Manual Translations available at www.extech.com

More information

Please enter the identity code of your device here!

Please enter the identity code of your device here! Operating Instructions DULCOMETER D2C Part 2: Adjustment and Operation, Measured Variables chlorine/chlorine ProMinent D2C2-001-chlorine/chlorine-GB CI 2 /Cl 2 0.50 ppm CLE 0.20 ppm CI DULCOMETER STOP

More information

MT3500 Hand-Held Engine Analyzer Safety Operation Regulations

MT3500 Hand-Held Engine Analyzer Safety Operation Regulations MT3500 Hand-Held Engine Analyzer Safety Operation Regulations - 1 - Chapter One Introduction A. Notice of Usage: MT3500 Hand-Held Engine Analyzer must be operated by trained professionals, who must know

More information

ProfiScale MULTI Multimeter

ProfiScale MULTI Multimeter 1,5 V 9V 200 mv 600 V 200 ma 1/10 A ProfiScale MULTI Multimeter en Operating instructions BURG-WÄCHTER KG Altenhofer Weg 15 58300 Wetter Germany Introduction Want the reassurance of knowing whether current

More information

Table of Contents. Chapter 3. The 843-R Power Meter Unit.. 17

Table of Contents. Chapter 3. The 843-R Power Meter Unit.. 17 Table of Contents Chapter 1. Introduction: How to Use This Manual... 3 Chapter 2. Quick Reference... 4 2.1 Getting Started... 4 2.2 Functions with No Sensor Connected... 5 2.3 Thermal Sensors... 8 2.4

More information

K-BUS Switch Actuator

K-BUS Switch Actuator K-BUS Switch Actuator User manual-ver. 2 KA/R0416.1 KA/R0816.1 KA/R1216.1 Contents Contents... 2 1. Introduction... 3 1.1 Product and function overview... 3 2. Technical Properties... 3 3. Commissioning...

More information

Safety Information (Attach ed at the upper cover of t his product) Safety Warning (Attached on the output fiber isolato r & collimator) Figure 1 Laser

Safety Information (Attach ed at the upper cover of t his product) Safety Warning (Attached on the output fiber isolato r & collimator) Figure 1 Laser Safety Please read this user manual carefully before operating the YDFLP fiber laser. It provides essential information regarding safety, product operation, and other necessary reference information. For

More information

Mars-Energo. Volt/mA Calibrator CALMAR-S. User s manual Edition 3.1

Mars-Energo. Volt/mA Calibrator CALMAR-S. User s manual Edition 3.1 Mars-Energo Volt/mA Calibrator CALMAR-S User s manual Edition 3.1 Manufacturer: OOO NPP Mars-Energo Address: V.O. 13 Line, 6-8, office 41H, Saint-Petersburg, Russia, 199034 2017 MARS-ENERGO Table of contents

More information

OVA-50 Optical Variable Attenuator User s Manual

OVA-50 Optical Variable Attenuator User s Manual OVA-50 Optical Variable Attenuator User s Manual Shineway Technologies, Inc. All rights reserved. Safety Instructions The WARNING sign denotes a hazard. It calls attention to a procedure, practice, or

More information

The CO2 Sensor Calibration Kit

The CO2 Sensor Calibration Kit The CO2 Sensor Kit For use with all BAPI CO 2 Sensors Instruction Manual CO 2 Kit Product Identification and Overview BAPI s CO 2 Sensor Kit is designed to calibrate and verify the operation of all BAPI

More information

TX CONTROLLER Model EM-IP Quick Start Guide

TX CONTROLLER Model EM-IP Quick Start Guide TX CONTROLLER Model EM-IP Quick Start Guide 860 boul. de la Chaudière, suite 200 Québec (Qc), Canada, G1X 4B7 Tel.: +1 (418) 877-4249 Fax: +1 (418) 877-4054 E-Mail: gdd@gdd.ca Web site: www.gdd.ca Visit

More information

HG G B. Gyroscope. Gyro for AGV. Device Description HG G B. Innovation through Guidance. Autonomous Vehicles

HG G B. Gyroscope. Gyro for AGV. Device Description HG G B.   Innovation through Guidance. Autonomous Vehicles Device Description HG G-84300-B Autonomous Vehicles Gyroscope HG G-84300-B Gyro for AGV English, Revision 06 Date: 24.05.2017 Dev. by: MG/WM/Bo Author(s): RAD Innovation through Guidance www.goetting-agv.com

More information