Millenium ML110 CONVERTER OPERATING & INSTALLATION MANUAL
|
|
- Bruno Berry
- 5 years ago
- Views:
Transcription
1 Millenium ML110 CONVERTER OPERATING & INSTALLATION MANUAL
2 FLOMOTION SYSTEMS, Inc. 586 N. French Rd., Suite 6 Buffalo, NY Toll Free: FLOW (3569) (USA & Canada only) Phone: Fax: info@flomotionsystems.com VER The software version is displayed when powering up the converter. 2
3 INDEX Introduction pag.4 Symbols Used on the manual pag.4 Technical characteristics pag.5 Electrical characteristics pag.5 Environmental conditions of use pag.5 Operative temperature pag.5 Overall dimensions pag.6 Electrical connections pag.7 Grounding instructions pag.7 Power supply converter pag.7 Internal view of the converter pag.8 Electrical connections sensor-converter pag.8 Outputs pag.9 On/Off output wiring (up to 1250 Hz) - low frequency pag.9 Analogical output - 0\4 20 ma pag.9 Start up and maintenance of the instruments pag.10 Access codes pag.10 How to access at the instrument functions pag.11 Converter visualization pages pag.11 Flags interpretation and led pag.12 Programming functions pag.13 Converter key board pag.15 Access to the configuration menu pag.16 Programming functions pag.18 Functions description pag.18 Alarm messages pag.24 Causes and action to be taken pag.24 Anomalies codes pag.24 3
4 INTRODUCTION This manual is integral part of the product. Read carefully the instructions contained since they give important indications for the safe use and maintenance. Technical information and relative products in this manual could undergo modifications without any previous notice. The flow meter must be used for what it has been built for. The improper use, possible tampering of the instrument or parts of it and substitutions of any not original components, make the warranty to decay automatically. The manufacturer is considered responsible only if the instrument it s used in his original configuration. Reproduction of the present manual and of any possible software supplied with the instrument is strictly forbidden. Symbols Used on the manual ATTENTION DANGER ELECTRIC SHOCK WARNING PRECAUTIONS 4
5 TECHNICAL CHARACTERISTICS ELECTRIC CHARACTERISTICS Classification of the instrument: class I, IP 65, category of installation II Power supply Power supply Power supply current Pmax versions voltage frequency max HV Vac Hz 3W/5VA 35 ma LV Vdc Vac Hz 3W/5VA 300 ma INPUT/OUTPUT ISOLATION Input/output are insulated up to 500V The output 4 20 ma and the output 24 Vdc are electrically connected ENVIRONMENTAL CONDITIONS OF USE The instrument can be installed inside or outside buildings Altitude: from 200 a 6000 m (from -656 to feet) Humidity range: 0 98% Line voltage range: (see table on technical characteristics) OPERATING TEMPERATURE CONVERTER Ambient Temp. Min. Max C F C F -10* -14* * For discontinuous use, the installation of a heating resistance is necessary 5
6 124 OVERALL DIMENSIONS COMPACT VERSION 127 SEPARATE VERSION
7 ELECTRICAL CONNECTIONS GROUNDING INSTRUCTIONS For the correct operation of the meter it s NECESSARY that sensor and liquid are equipotential, so ALWAYS connect sensor and converter to the ground CONVERTER POWER SUPPLY L (+) Wall version N (-) L N (+) (-) Panel version M3 M3 before connecting the power supply, verify that the mains voltage falls between the limits indicated on the tag plate ATTENTION: the converters on dc power supply line are not protected against the inversions of polarity. For the wiring use only approved conductors, with fire-proof properties. The power supply line must be equipped with an external protection for current overload (fuse or automatic line breaker with limiting capacity not greater than 10 A). Provide in the proximity of the instrument a circuit breaker that must be easily accessible from the operator and clearly identified. NOTE: characteristics of meter s power supply, see page 5 7
8 INTERNAL VIEW OF CONVERTER Power supply IF2 socket Signalling LED with light: interpretation at page Dangerous voltage on block 12-13: - 60 Vdc Max V Max on commutation coils Keyboard Power supply >HV< VAC >LV< VAC VDC ELECTRICAL CONNECTIONS SENSOR TO CONVERTER SEPARATE VERSION C018 CABLE 1 E1 2 E C SH B1 B2 24V - TERMINAL BLOCK OUT2 OUT1 24V E C E C COILS SC B2 B1 SC Sudden movements of the electrodes cable, can cause noises on measure Max length of cable: mt mA SC C E2 E1 ELECTRODES 8
9 OUTPUTS WIRING Output on/off 1250 Hz 43V 16 (out1) 18 (out2) 17 (out1) 19 (out2) Opto-insulated output with floating collector and emitter terminals freely connectable Maximum switching voltage: 40 Vdc Maximum switching current: 100mA Maximum saturation voltage between collector and 1,2V Maximum switching frequency (load on the collector or emitter, RL=470Ω, VOUT=24Vdc): 1250Hz Maximum reverse current bearable on the input during and accidental polarity reversion (VEC): 100mA Insulation from other secondary circuits: 500 Vdc Output 0/4 20mA +24V INTERNAL I ma 9 (+) 0/4..20mA OUT Opto-insulated output Maximum load 1000 ohm Maximum voltage without load 27 Vdc Refresh frequency is the same of the sample frequency of the connected sensor protected against persistent over voltages till 30 Vdc 10 (-) 24V COMMON The converter CHECK of load on the 4 20mA output; to disable this function set the value "ma Val. Fault" to 0 ( Pos. 4.5) 9
10 START UP AND MAINTENANCE OF THE INSTRUMENTS Before starting up the instrument please verify the following: Power supply voltage must correspond to that specified in the name plate Electric connections must be done as described at page 7-8 Ground connections must be done Verify periodically: The integrity of the power supply cables, wiring and other electrical parts connected The integrity of the instrument s housing (this must not have bruises or other damages that may compromises the hermetical sealing) The tightening of the sealing elements (cable glands, covers, etc.) The integrity of the front panel (display and keyboard), damages may compromise the sealing The mechanical fixing of the instrument on the pipe or on the wall stand ACCESS CODE FACTORY PRESETTINGS Some functions in the converter are enabled by the access codes. The information of this manual is related to all the functions available with L2 level. All the functions available through higher level are protected and reserved to the service. Description of the L2 access code (menu 11 Internal data pos. 11.1) with code L2 = (with this code only) you disable the request of code L2 NOTE: the availability of the functions is related to the selected block *with L2 customised (freely chosen by the user) you can proceed programming all the functions up to L2 security level, entering its code whenever you enter the Main menu *ATTENTION: take note very carefully of the customised code you have chosen, since there is no way for the user to retrieve it if it is forgotten The converter is delivered with access code L2: and with the "Quick start menu" enable. Press the key to access to the "Quick start menu" from one of the visualization pages The Quick start menu may be set without entering any access code (see example 1 on page 11). 10
11 VISUALIZATION PAGES The direct exposure of the converter to the solar rays, could damage the liquid crystals display. N.B. Contrast set see page 23 pos. 8.3 Active scale Alarm on * Direct/Reverse totalizer visualization Flow rate visualization Alternate visualization Push to change visualization CURRENCY FUNTION ENABLE * Direct totalizer Reverse totalizer Currency value Flow rate visualization Push to change visualization Flow rate visualization * Direct Reverse totalizer Flow speed Alarm description Push to visualize the alarm Sampling frequency * The maximum number shown from the totalizer is independently from the number of selected decimals. Beyond this value the totalise are reset. 11
12 Flags interpretation and LED FLAGS INTERPRETATION FLAGS FLAG DESCRIPTION M Alarm max activated m Alarm min activated - Interruption coils circuit! - Segnal error - Empty pipe C Calibration running S Simulation Pulse output saturation (reduce TIME PULSE ) LED PERMANENT LIGHT: initialisation LED INTERPRETATION FLASHING LIGHT ( 1 sec.): normal function FLASHING LIGHT (<1 SEC.): alarm on The LED signals the real alarm status only if the display visualizes one of the visualization pages suitable to page 17 12
13 ML210 Functions (for detail functions with symbol * see the manual from page 18) Attention: The function in grey colour are visualized on display only with other active functions or with optional modules 1.1 Insert ND of sensor ( ) 1.2 Calibration data of sensor visualized on sensor s label 1.3 Type of sensor: Enter the first two characters of the serial number of the sensor 1.4 Position for insertion sensors: 0=1/8DN, 1=1/2DN, 2=7/8DN 1.5 Factory parameter 1.6 Enables the empty pipe detection feature 1.7* Enables the automatic zero calibration system 1.8* Enables the automatic calibration procedure of the empty pipe detection 2.1*Full scale value set for range N.1 2.2*Full scale value set for range N.2 2.3* Unit of measure and number of decimal totalizes 2.4* Pulse value on channel 1 2.5* Pulse value on channel 2 2.6* Duration of the pulse generated on channel 1 2.7* Duration of the pulse generated on channel Full scale freq. for channel 1 (0.1Hz Hz) 2.9 Full scale freq. for channel 2 (0.1Hz Hz) 3.1* Time constant 3.3* Acceleration threshold 3.4* Anomalous signal pick cut off threshold 3.5 Low flow zero threshold: 0-25% of full scale value 3.2 Filter on the power supply: 0.1s= ready measure; 0.5s=filter of noise on the liquid 3.6 Enable every hour an internal cycle of calibration. The measure it's stopped for 8-15 sec. 3.7* Automatic change of scale 3.8* Energy saving 4.1 Maximum value alarm set for flow rate 4.2 Maximum value alarm set for flow rate 4.5 Hysteresis threshold set for the minimum and maximum flow rate alarms 4.6 Empty pipe detection threshold. It s automatically set by the function *Current output value in case of failure 4.8*Frequency output value in case of failure 13
14 6.1* Output 1 functions 6.2* Output 2 functions 6.5* Duty cycle value for pulses/frequency output 6.6* Choice of the function and the range of current output n.1 6.7* Choice of the function and the range of current output n Choice of the communication protocol for the IF2 device 8.1 Choice of the language: E= English, I=italian, F= French, S= Spanish 8.2 Updating frequency on the display: Hz 8.3 Display contrast 8.4 Quick start menu visualization 8.5 Enable the change value of the totalises (see function ) 8.6 Enable the page of net totalizer (difference between direct and reverse. see page 17) 8.7* Total direct (positive) flow totalise reset 8.8* Partial direct (positive) flow totalise reset 8.9* Total reverse (negative) flow totalise reset 8.10*Partial reverse (negative) flow totalise reset 8.11 Visualizes the values of the partial totalise in the unit of selected currency 8.12 Choice of the numbers of decimals for the visualization currency value: From 0 to *Value of conversion/currency for direct totalizer 8.14*Value of conversion/currency for reverse totalizer 10.1* Enable the calibration of the converter 10.2* Converter autotest 10.3* Flow rate simulation enabling 11.1 Level 2 access code enter 11.2 Load factory data pre-set 11.3 Load user data saved 11.4 Save user data 11.5 Visualisation of the total operation hours of the converter (function not editable) 11.6 Ignore the calibration error during the switch on test 11.7 Ks Coefficient 14
15 ACCESS TO THE INSTRUMENT KEYBOARD M3 The keyboard converter is accessible by opening the cover of the converter ; the three KEYS are set aside of the M1 terminal block KEYBOARD SHORT PRESSING (< 1 SECOND): It increases the numeric figure or the parameter selected by the cursor It goes to the previous subject on the menu batch start/stop (when enabled) LONG PRESSING (> 1 SECOND): It decreases the numeric figure or the parameter selected by the cursor It goes to the next subject on the menu SHORT PRESSING (< 1 SECOND): It moves the cursor rightwards on the input field It goes to the following subject of the menu It changes the display of the process data LONG PRESSING (> 1 SECOND): It moves the cursor leftwards on the input field It goes to the previous subject on the menu SHORT PRESSING (< 1 SECOND): It enters/leaves the selected function It enables the main menu for the instrument configuration It cancels the selected function under progress LONG PRESSING (> 1 SECOND): It leaves the current menu It enables the totalise reset request (when enabled) It confirms the selected function BLIND CONVERTERS For converter without keyboard (blind version), the programming of functions is made up by IF2 serial device: 15
16 ACCESS TO THE CONFIGURATION MENUES The access to the configuration menu can take place in two different modes: Through the Quick start menu where it is possible to access directly to some of the principal functions Through the Main menu where it is possible to access to all function with access code 2 We show below some examples relating to the change of the value in the Fs1 function EXAMPLE: modifying the full scale value from 4dm³/s to 5dm³/s. from Quick start menu Enter in the Quick start menu Access to the function Fs1 Push repeatedly Change the value Confirm the new value Long push Main page 16
17 EXAMPLE: modifying the full scale value from 4dm³/s to 5dm³/s. from Main Menu (quick start menu enable) Enter in the Quick start menu Access to the Main Menu X 5 TIMES Access to the Scale menu Access to the function Fs1 Push repeatedly Change the value Confirm the new value Long push Main page Long push 17
18 FUNCTIONS DESCRIPTION (description of the functions with access code< 3) Identification of the function (not visualized on display) MENU 1.SENSOR (POS. 1) Nominal diameter of sensor [ND= XXXX] Converter request Menu visualized on the converter ( from 1 to 11) Synthetic description of the function MENU 1.SENSOR (POS. 1.7) Autozero calibration [AUTOZERO CAL.] This function enables/disables the automatic zero calibration system. To perform the sensor it is absolutely necessary the sensor is full of liquid and that the liquid is perfectly staying still. Even very small movement of the liquid may affect the result of this function, and, consequently, the accuracy of the system. Once you are sure the a.m. conditions are fulfilled (and when the percentage flow rate value is stable) press for more than one second the key. Check the percentage flow rate value goes to zero, otherwise repeat the operation again. When the value is stable at zero, then press. (POS. 1.8) Empty pipe calibration [E.P. CALIBR.] this function enables/disables the automatic calibration procedure of the empty pipe detection function. Before enabling this function, the Empty Pipe test should be enabled first as above described. Before performing this function, the sensor has to be completely filled in with the liquid so that both the lining and the electrodes are wetted. The sensor has then to be emptied again and then you should press the key : the operation will have to be confirmed by pressing the key or cancelled by pressing the key. By this function the system sets the value of a parameter, which could also be manually changed (see function E.P.thr within the menu 4-ALARMS). MENU 2.SCALES (POS. 2.1) Full scale n 1-2 [FS1-2=dm³/SX.XXXX] Full scale value set for range N.1-2. There are four fields to fill in order to set this parameter, from left to right: 1) volume unit of measure, 2) type of unit, 3) time unit of measure and 4) numeric value. The selection is made by positioning the cursor on the field to modify. To change the type of unit of measure (metric, British or American, mass or volume) the cursor has to be positioned on the symbol / (field N. 2). When the nominal diameter is set to zero it is possible to modify only the numeric field, since the unit of measure stays at m/sec. The following tables show the units of measure available and the conversion factor by comparison with 1 dm 3 and 1 kg. The converter accepts any kind of combination of units of measure satisfying both the following conditions: Numeric field value / 25 fs max numeric field value fs max. where fs max is the maximum full scale value corresponding to the sensor, equal to a 10 m/sec liquid speed. The units of measure are shown as appear on the display. The British and American units are diversified by using capital and small characters. 18
19 19 cm 3 Cubic centimeter in 3 Cubic inch Oz Ounce ml Milliliter Gal American gallon Lb Pound l Liter GAL British gallon Ton short tons dm 3 Cubic decimeter ft 3 Cubic foot dal Decalitre Bbl Standard barrel hl Hectolitre BBL Oil barrel G Gram m 3 Cubic metre yd 3 Cubic yard Kg Kilogram kgl KAmerican gallon T Ton KGL KBritish gallon When a mass unit of measure is set, the specific gravity function is automatically enabled by the system. Please, note that the mass measure is heavily affected by the temperature and therefore with certain liquids this may cause significant measure errors. The units of measure of time may be chosen among the following values: s = second, m = minute, h = hour, d = day. (POS. 2.3) Unit of measure and number of decimal totalizes [UM.tot:dm³ X.XXX] Setting the unit of measure and number of decimals for visualized the totalizes or the volumes to batch. For set the unit of measure, position the cursor on field of the actual unit of measure; For set the type of unit, position the cursor on the blank space between the unit of measure and the numeric value; For set the number of decimal totalizes position the cursor on numeric field and choose one of the possible combinations: (POS ) Pulse value channel 1-2 and unit of measure of the total. [IMP1-2=dm³X.XXXXX] Setting of the pulse volume corresponding to channel 1-2 and of the totalizers measure units. There are three fields to fill in to set this parameter, from left to right: 1) measure unit, 2) unit type and 3) numeric value. The selection is performed by positioning the cursor on the field to be modified. To change the unit type (metric, British or American, mass or volume) just position the cursor on the blank space between the measure unit and the numeric value. When the nominal diameter is set to zero it is possible to modify only the numeric field since the measure unit stays at meter (m) or feet (ft). The possible measure units are those above described. This function is active only if the pulse emission on channel 1 has been set as enabled. (POS ) Pulse duration channel 1-2 [TPUL1-2=ms XXXX.XX] Setting of the duration of the pulse generated on channel 1-2. Its value is expressed in milliseconds and has to be between 0,4 and 9999,99. When the high frequency output is present, then the minimum value can go down to 0,04 milliseconds. ATTENTION: since the instrument cannot detect which type of device it is connected to, it is up to the user to verify the set pulse duration is compatible with the external device processing such pulses. If, for example, an electro-mechanical pulse counter is connected, then two kind of problems may occur: if the pulse is too long than the coil may burn or, if it is too short, the counter may not be able to count and eventually even cause the damaging of the output itself. MENU 3.MEASURE (POS. 3.1) Time constant [TCONST=sXXXX.X] Time constant set. This parameter affects the integrating filter making the instrument response quicker or slower, according to the set value. A higher value corresponds to a more stable but slower measure, a smaller value the opposite. The most common values are from 1 to 5 seconds. The value of this parameter has to be within the range from 0 (integral filter disabled) to seconds. The following diagram shows the response of the instrument for a flow rate variation from 0 to 100% within the T time constant period. (POS. 3.2) Acceleration threshold [SKIP THR=% XXX] Acceleration threshold set. The acceleration threshold stands for the limit beyond which a flow rate variation determines an immediate response at the output, without being filtered by the time constant. This system allows the instrument to have an immediate response in case of big variations of the flow rate, filtering (and delaying) the response to small variations. The result of that is a very stable measure, ready to follow the process. The value is set as percentage of the full scale value from 0 to 125%. If such a value is set to zero any flow rate variation bigger than 0.5% of the full scale value will immediately affect the outputs. The following diagram shows the instrument response in two cases: a flow rate variation from 0 to 10% completely absorbed by the time constant effect
20 and a variation form 10% to 100% exceeding the acceleration threshold and then immediately sent to the output. In actual fact there is always a minimum time between the measure acquisition and the outputs update. (POS. 3.3) Peak cut off threshold [PEAK THR=% XXX] Anomalous signal pick cut off threshold set. This parameter allows setting the maximum value of deviation of the actual measure sample by comparison with the average one. If the new value is higher than the set limit, than such a value is cut to the limit value. This function is used to make the meter less sensitive to big perturbations on the flow rate measure, as it may happen when there are solids in suspension in the liquid hitting against the electrodes determining a high electrical noise. The permitted values of this function range from 0 to 125 % and are referred to the full scale value. If this parameter is set to zero the peak detection function is disabled and any new measure ample will be accepted and processed as it is by the converter. (POS. 3.7) Automatic scale change enable [AUTORANGE=ON/OFF] Enables/disables the automatic change of scale. The meter may have two different working ranges in order to suit to the variable process conditions. In order to get the best results out of this function it is important that range N.2 is bigger than N.1. When the flow rate increases and reaches the 100% of the full scale 1, then the meter automatically switches to scale N.2. When the flow rate decreases again reaching a value on scale 2 equal to the 90% of full scale N.1, then the active scale is N.1 again. Allowed values for this parameter: ON / OFF. (POS. 3.8) Energy saving enable [E.SAVING=ON/OFF] Automatic energy saving function enables. This function is used when the instrument is powered by a battery or solar cells, allowing an energy saving up the The energy consumption is controlled by the ratio between the measuring cycles powering the coils and the cycles without powering the coils. When the flow rate is stable the number of off cycles is higher than the on ones, so that the average consumption is strongly reduced. If the flow rate suddenly changes, then the meter switches on a higher number of measuring cycles, in order to get a higher response time, switching off the cycles as soon as the flow rate gets back to be stable. If the flow rate varies below the acceleration threshold percentage value, then the meter goes on with off cycles, but as soon as the flow rate value exceeds such a threshold, the meter switches on many measuring cycles again. The speed the meter switches on and off the excitation cycles is different: from a constant flow rate to a variable one it is a very fast process, while from variable flow rate to a stable one such a process is much slower. Allowed values for this parameter: ON/OFF. N.B.: to optimise this function it is recommended choose a value for the acceleration threshold within MENU 4.ALARMS (POS. 4.5) Current output value in case of failure [ma VAL.FAULT =% XXX] Setting of the value of the 0/ ma current output in one or more of the following cases: empty pipe ; ADC error ; coils interrupted The allowed range is from 0 to 120% of the ma scales, 120% corresponds to 24 ma and does not depend on the selected range (0 20 / 4 20 ma). The NAMUR NE43 recommendations asks for an alarms signalling value for the current output lower than 3.6 ma (<18%) or bigger than 21 ma (>105%). It would then be preferable to set the value of 20
21 this function at the 10%, so that the current value in case of the a.m. cases would be 2 ma, allowing the following diagnostics: current < 2 ma - 5%: line interrupted, power supply failure or faulty converter; 2 ma -5% current 2 ma + 5%: hardware alarm; 4 ma current 20 ma: normal working range; 20 ma < current 22 ma: out of range, measure above 100% f.s. (POS. 4.6) Frequency output value in case of failure [Hz VAL.FAULT=% XXX] Setting of the frequency value to assign to the on/off output in one or more of the following cases: Empty pipe ; Coils interrupted ; ADC error The allowed range is from 0 to 125% of the frequency full scale value. Although there are not specific rules regulating cases like this one, it would be convenient to use the failure information as follows: 0% Hz frequency 100% f.s.: normal working range; 100% f.s. < frequency 110% f.s.: overflow, measure above the 100% of the f.s.; 115% f.s. frequency 125% f.s.: hardware alarm condition. MENU 6.OUTPUTS (POS. 6.1) Choice of the function corresponding to on/off output 1 [OUT1=XXXXXX] Choice of the function corresponding to digital Output 1. The functions are listed in the table below. (POS. 6.2) Choice of the function corresponding to on/off output 2 [OUT2=XXXXXX] Choice of the function corresponding to digital Output 2. The functions are listed in the table below. FUNCTION FOR OUTPUT 1, 2. OFF: DISABLED #1PULS+: PULSE ON CHANNEL 1 FOR POSITIVE FLOW RATE #1PULS-: PULSE ON CHANNEL 1 FOR NEGATIVE FLOW RATE #1PULS±: PULSE ON CHANNEL 1 FOR POSITIVE AND NEGATIVE FLOW RATE #2PULS+: PULSE ON CHANNEL 2 FOR POSITIVE FLOW RATE #2PULS-: PULSE ON CHANNEL 2 FOR NEGATIVE FLOW RATE #2PULS±: PULSE ON CHANNEL 2 FOR POSITIVE AND NEGATIVE FLOW RATE #1FREQ+: FREQUENCY CHANNEL 1 FOR POSITIVE FLOW RATE #1FREQ-: FREQUENCY CHANNEL 1 FOR NEGATIVE FLOW RATE #1FREQ±: FREQUENCY CHANNEL 1 FOR POSITIVE AND NEGATIVE FLOW RATE #2FREQ+: FREQUENCY CHANNEL 2 FOR POSITIVE FLOW RATE #2FREQ-: FREQUENCY CHANNEL 2 FOR NEGATIVE FLOW RATE #2FREQ±: FREQUENCY CHANNEL 2 FOR POSITIVE AND NEGATIVE FLOW RATE SIGN: FLOW DIRECTION OUTPUT (ENERGISED = -) RANGE: RANGE INDICATION OUTPUT (ENERGISED = SCALE 2) MAX AL: MAX FLOW RATE OUTPUT(ENERGISED = AL. OFF) MIN AL: MIN FLOW RATE OUTPUT(ENERGISED = AL. OFF) MAX+MIN: MAX AND MIN FLOW RATE ALARM OUTPUT (ENERGISED = AL. OFF) EMPTY PIPE: EMPTY PIPE ALARM OUTPUT (ENERGISED = FULL PIPE) OVERFLOW.: OUT OF RANGE ALARM OUTPUT (ENERGISED = FLOW RATE OK) HW ALARM: CUMULATIVE ALARM OUT interrupt coils, empty pipe, measure error (ENERGISED=NO ALARMS) (POS. 6.3) duty cycle value for pulses/frequency output The duty cycle function defines the time ratio between ON and OFF state when frequency output are used: 50% means that the ON phase will last as much as the OFF phase, 60% means that the ON phase will be 60 % and the OFF phase will be 40% of the total cycle time. When the pulses outputs are used, the duty cycle defines the OFF phase because the ON phase is already set with the function "PULSE DURATION"( see menu SCALE ). In this case if, for example, the duty cycle is set at 50% and the pulse duration at 50ms, the OFF phase will be the same of the ON phase. The formula to calculate the minimum time of the OFF phase and the time of total cycle is the following: T. total cycle= 100 x (pulse duration in ms)/ (duty cycle) T. OFF phase = T. total cycle - pulse duration N.B.: If the value of the function is set to 0 the emission of the pulses happens in synchronous mode with the flow rate, therefore when using the function in frequency mode, DO NOT set the duty cycle to 0. The function is active only if one of the outputs is set on pulse and/or frequency function 21
22 (POS. 6.4) Choice of the function and the range of current output [OUT.mA=X XX±] Choice of the function and the range of current output N.1. The current output N.1 is optional and it is mounted on the main board. There are three fields to modify for this function: Scale zero: 4 or 0 ma Full scale: 20 or 22 ma Field: + = positive, - = negative, ± = both, -0+ = central zero scale The values corresponding to the scale points are shown in the following chart: CURRENT VALUES IN ma ASSOCIATE TO THE % VALUE OF FULL SCALE REVERSE FLOW VALUE ZERO DIRECT FLOW VALUE POSSIBLE FIELD -110% -100% 0% +100% +110% OutmA = OutmA = OutmA = * OutmA = OutmA = OutmA = OutmA = OutmA = OutmA = 0 20 ± OutmA = 0 22 ± OutmA = 4 20 ± OutmA = 4 22 ± OutmA = OutmA = ** OutmA = OutmA = In hardware alarm conditions HW ALARM (interrupted coils, empty pipe, measure error) the current value is programmed by the function ma VALL. FAULT (pos. 4.7) and it is expressed as percentage of a fixed current range, where: 0% = 0 ma e 110% = 22 ma. * Example 1: out I (ma) 22 ma 20 ma 4 ma -110% F.S. -100% F.S. +100% F.S. +110% F.S. zero ** Example 2: out I (ma) 20 ma 12 ma 4 ma -110% F.S. -100% F.S. +100% F.S. +110% F.S. zero 22
23 MENU 8.DISPLAY (POS ) Total/Partial totalizer reset enable [RESET T/P/-/+] Activates the reset of Total and partial flow totalizer Functions are activated by pressing the key during the visualisation of the function itself. When " EXECUTE?" is required, press the key to proceed. Press any other key to delete the operation N.B.: The PARTIAL totalizers reset can be done also in the following way: From one visualization page press the key. Set the L2 code and press the key. The following question will be visualized: "EXECUTE?" Press the key to proceed. Press any other key to delete the operation (POS ) Conversion factor for direct flow rate totalizers [EUR/dm³+ =X] Setting of the conversion/currency value for direct flow totalizers (positive). There are three setting fields for this parameter, from left to right: 1) monetary symbol, 2 ) default/personalized monetary symbol, 3 ) conversion coefficient. For the selection set the cursor over the field to be modified. The monetary symbol setting can be done in two ways: 1. choosing one of the 7 predetermined monetary symbols (standard ISO 4217-REV81): EUR=Eur; USD=USA dollar; CAD=Canadian dollar; AUD=Australian dollar; GPB=English pound; CHF=Swissfranc; JPY=Japanese yen. 2. Entering a three characters (numbers or letters) personalised password. To change the characters, the cursor has to be set on the symbol / (field N. 2) MENU 10. DIAGNOSTIC (POS. 10.1) Meter calibration [CALIBRATION] Enables the calibration function of the meter. The activation of this function happens by pressing the key during the visualization of the function. The following question will be visualized: " EXECUTE?". Press the key to proceed. Press any other key to delete the operation. (POS. 10.2) Autotest function [SELF TEST] Meter autotest function. This function stops the normal functions of the meter and performs a complete test cycle on the measure input circuits and on the excitation generator. To activate this function, after selecting it, press the key, at the question: EXECUTE? press the key to start the autotest, or any other key to delete the operation. The result of the test is shown on the display. At the end of the operation the display will go back to one of the visualisation pages. This function is automatically performed when switching on the device. (POS. 10.3) Flow rate simulation [SIMULATION] Flow rate simulation enabling. With this function it is possible to generate an internal signal that simulates the flow rate, allowing the outputs and all the connected instruments test. After enabling it, the flow rate simulation can be: - set: by pushing the key from one of the four visualisation pages - started: by pushing the key after setting it - finished: by pushing the key from the visualisation pages and then by pushing the key. 23
24 Alarm messages, causes and actions to be taken Messages ANOMALIES ACTION TO TAKE NO ALARMS All works regularly MAX ALARM The flow rate is higher than the maximum threshold set Check the maximum flow rate threshold set and the process conditions MIN ALARM The flow rate is lower than the minimum threshold set Check the minimum flow rate threshold set and the process conditions FLOW RATE >FS PULSE/FREQ>FS EMPTY PIPE INPUT NOISY The flow rate is higher than the full scale value set on the instrument The pulse generation output of the device is saturated and cannot generate the sufficient number of impulses The measuring pipe is empty or the detection system has not been properly calibrated The measure is strongly effected by external noise or the cable connecting the converter to the sensor is broken Check the full scale value set on the instrument and the process conditions Set a bigger unit of volume or, if the connected counting device allows it, reduce the pulse duration value Check whether the pipe is empty or perform again the empty pipe calibration procedure Check the status of the cables connecting the sensor, the grounding connections of the devices or the possible presence of noise sources EXCITATION FAIL The coils or the cable connecting the sensor are interrupted Check the connecting cables to the sensor CURR. LOOP OPEN P.SUPPLY FAIL Verify the load is applied to the output (max 1000 The 0/4...20mA output on board or the optional one are not ohm). correctly closed on a valid load To disable the alarm,set the ma VAL.FAULT value ( menu alarm ) to 0. Verify that the power supply is that indicated on the Power supply different from that indicated on the label. label Anomalies codes CODES ANOMALIE DESCRIPTIONS ACTION TO TAKE 0001 problem with watch-dog circuit 0002 wrong configuration work data in eeprom 0004 wrong configuration safety data in eeprom 0008 defective eeprom 0010 defective keyboard (one or more key are pushed during the test) ADDRESS TO SERVICE 0020 Power supply voltage (+3.3) is out of range 0040 Power supply voltage (+13) is too low (<10V) 0080 Power supply voltage (+13) it s too high (>14V) 0200 timeout calibration input (input circuit is broken) Check the status of the cables connecting the sensor to 0400 Input stage gaining is out of range the converter, the grounding connections of the devices or the possible presence of strong and anomalous noise sources Check the status of the cables connecting the sensor to 0800 Interruption on the coils circuit the converter 0C00 Cumulative alarm see single code 24
25 DECLARATION OF CONFORMITY According to ISO / IEC Guide 22 and EN Product s name: Electromagnetic flow meter series Converter model: ML110 Option: all applicable Sensors model: MS 500 MS 501 MS 600 MS 1000 MS 2410 MS 2500 MS 3700 MS 3770 MS 5000 ISOIL INDUSTRIA S.p.A. declares that the above mentioned products satisfy the following requirements: Safety EN61010, dielectric strength = 4 kv, installation category II, IP65 EMC EMC reference : Immunity: EN Emission: EN Test : EN55011 (150 khz 30 MHz): Group 1, class B EN55011 (30 MHz 1GHz): Group 1, class B IEC : 4 kv CD, 8 kv AD IEC (f = 80 MHz 1 GHz, antenna at 3 m, AM modulation 1kHz 80%): 10 V/m IEC : 4 kv on all ports IEC (2kV diff/2kv common mode) IEC (f = 150 khz 80 MHz, AM modulation 1 khz 80%): 10 V IEC
26 NOTE 26
27 NOTE 27
28 110_EN_IS_4_3_5X.doc The last three character of file name, identify the sw version, which the manual is refer. The sw version is visualized during switch on of converter
OPERATING AND INSTALLATION MANUAL CONVERTER
OPERATING AND INSTALLATION MANUAL CONVERTER ML 110 CONTENTS INTRODUCTION...3 SYMBOLS USED IN THIS MANUAL...3 TECHNICAL CHARACTERISTICS...4 INPUT/OUTPUT ISOLATION...4 ENVIRONMENTAL CONDITIONS OF USE...4
More informationFAST SAMPLING CONVERTER
FAST SAMPLING CONVERTER ML4-F1 HIGH SAMPLING RATE CONVERTER (UP TO 400 SAMPLES/SECOND) Warranty conditions are available on this website: www.isomag.eu only in English version INDEX TECHNICAL DATA... 3
More information( F L O W I Z F A M I L Y )
THE MOST ACCURATE BATTERY POWERED SYSTEM ( F L O W I Z F A M I L Y ) E l e c t r o m a g n e t i c c o n v e r t e r p o w e r e d b y b a t t e r i e s, s o l a r p a n e l o r D C p o w e r w i t h 4
More informationWarranty conditions are available on this website: only in English version
THE MOST ACCURATE BATTERY POWERED SYSTEM FLOWIZ ( M L 2 5 2 ) C O N V E R T E R W I T H B A T T E R Y S U P P L Y Warranty conditions are available on this website: www.isomag.eu only in English version
More informationISOMAG UNIQUE PID REGULATOR + MAG CONVERTER THE FRIENDLY MAG METER
UNIQUE PID REGULATOR MAG CONVERTER ISOMAG THE FRIENDLY MAG METER CONVERTER ML 212 Integrate PID Regulator N 1 output (0/4 20mA; frequency; pulses variable duty cycle ) for the actuator N 2 inputs (0 5/10V
More informationISOIL ML210 CONVERTER For use with ELECTROMAGNETIC FLOWMETERS
ISOIL ML210 CONVERTER For use with ELECTROMAGNETIC FLOWMETERS 1 2 ML210 CONVERTER MANUAL Including Installation Instructions for M/MS SERIES ELECTROMAGNETIC FLOWMETERS April 2005 VER 3.0x ML 210 - Version
More informationMC 308 CONVERTER MUT SERIES FLOWMETERS
MC 308 CONVERTER MUT SERIES FLOWMETERS Page 2 FLOMOTION SYSTEMS, Inc. www.flomotionsystems.com 716.691.3941 info@flomotionsystems.com FEBRUARY 2003 MC 308 CONVERTER MANUAL Including Installation Instructions
More informationFLOMID-MX Control unit for electromagnetic flowmeters USERS MANUAL. instrumentación para fluidos. Realitzat: Revisat R+D Electrònic: Aprovat: Data:
instrumentación para fluidos FLOMID-MX Control unit for electromagnetic flowmeters USERS MANUAL Realitzat: Data: Revisat R+D Electrònic: Data: Revisat Of. Comercial: Data: Aprovat: Data: R-MI-Flomid-MX
More informationCBCT TM CIRCUIT BREAKER COIL TESTER
CBCT TM CIRCUIT BREAKER COIL TESTER USER S MANUAL Vanguard Instruments Company, Inc. 1520 S. Hellman Ave. Ontario, California 91761, USA TEL: (909) 923-9390 FAX: (909) 923-9391 August 2010 Revision 1 SAFETY
More informationUSER MANUAL. EPP Intelligent Positioner Control Unit 1/22.
USER MANUAL - Intelligent Positioner Control Unit 1/22 Table of contents: 1 General... 3 1.1 Safety instructions... 3 2 Application... 4 3 Electrical specifications and terminals... 5 3.1 Control loop...
More informationCurrent Loop Interface Module M 2025D
CONTROLS Current Loop Interface Module M 2025D Current-to-Voltage Analog Converter used with the M 2001 Series Digital Tapchanger Control and M-2948 Tap Position Sensors to provide positive tap position
More informationD.C. BRUSHLESS MOTORS DRIVE. BLD07-IT Service Manual
D.C. BRUSHLESS MOTORS DRIVE BLD07-IT Service Manual INTECNO s.r.l. Via Caduti di Sabbiuno n. 9/E 40011 Anzola Emilia (BO) - Italy tel. 051.19985350 fax 051.19985360 www.intecno-srl.com INDEX Description
More informationModel Sono-Trak Transit Time
Model Sono-Trak Transit Time Ultrasonic Flow Meter Engineered for performance excellence, the Sono-Trak Transit Time ultrasonic meter combines non-invasive, bi-directional measurement with advanced ultrasonic
More informationEfcon Water. Installation & User Manual Electro Magnetic Flowmeter (EMF) Efcon Water compact Electro Magnetic Flowmeter (EMF) Manual Version UK
Efcon Water Installation & User Manual Electro Magnetic Flowmeter (EMF) Page 1 Table of content Page 1.1 General 3 1.2 Application area 4 1.3 Transportation 4 1.4 Principle of Operation 4 1.5 Product Description
More informationsw rev. 55B DEVICE FOR THE PERMANENT CONTROL OF INSULATION ON THE SUPPLY LINES IN MEDICAL PRACTICE PLACES INSTALLATIONS
INSTRUCTIONS MANUAL IM831-U v0.92 HRI-R40W sw rev. 55B DEVICE FOR THE PERMANENT CONTROL OF INSULATION ON THE SUPPLY LINES IN MEDICAL PRACTICE PLACES INSTALLATIONS INDEX: GENERAL TYPES ACESSORIES AND OPTIONS
More informationSeries BDF for Process
Series DF for Process LARGE DISPLAYS for Process signals DF-xx-32 for ma signals DF-xx-36 for Vdc signals IDEAL SOLUTION for reading process signals at long distances, with scaled values on display, coming
More informationCustomer: Standard type Page 1 of 7 Description Fluxgate current sensor with toroidal core PCB mounting
Customer: Standard type Page 1 of 7 Description Fluxgate current sensor with toroidal core PCB mounting Electrical data Ratings min. typ. max. Unit I P Primary nominal RMS current (1phase / 3phase) 80
More informationFLOWIZ ( M L ) Warranty conditions are available on this website: only in English version
THE MOST ACCURATE BATTERY POWERED SYSTEM FLOWIZ ( M L 2 5 5 ) E l e c t r o m a g n e t i c c o n v e r t e r p o w e r e d b y b a t t e r i e s a n d u n i v e r s a l p o w e r s u p p l y. O p t i
More informationMultifunctional counter tico 772/773/774
Universally applicable as a tachometer, timer, shift counter, batch counter Easy to operate Four different backlight colours Intensity of backlighting can now be adjusted Large, clearly readable 48 x 48
More informationMW3105 DIGITAL CLAMP MULTIMETER
MW3105 DIGITAL CLAMP MULTIMETER 2 M MW3105 A 01 INTRODUCTION 1.1 - Unpacking and inspection Upon removing your new Digital Clamp Meter from its packing, you should have the following items: 1. Digital
More informationSHI Series Pure Sine Wave Inverter (SHI2000 AND SHI3000) User s Manual
SHI Series Pure Sine Wave Inverter (SHI2000 AND SHI3000) User s Manual The information presented in this document does not form part of any quotation or contract, is believed to be accurate and reliable
More informationInstallation, Operation and Maintenance Manual
M518 Rev. B Model 1030F Indicator-Totalizer Model 1030F Indicator-Totalizer Installation, Operation and Maintenance Manual Table of Contents General... 1 Installation... 1 Operation... 7 Programming...
More informationIDEAL INDUSTRIES, INC. TECHNICAL MANUAL MODEL: MODEL: Multimeter Service Information
IDEAL INDUSTRIES, INC. TECHNICAL MANUAL MODEL: 61-340 MODEL: 61-342 Multimeter Service Information The Service Information provides the following information: Precautions and safety information Specifications
More informationDSA150 Series. xxx Series. 150 Watts. AC-DC Power Supplies. Models & Ratings. Mechanical Details 3.92 (99.8) 2.18 (55.5) 4.92 (125.
15 Watts xxx Series Ultra Slim Design 15% Peak Load for 3 seconds Ambient Operation from -1 C to +7 C High Efficiency Selectable Overload Characteristic Selectable Remote Inhibit or Enable 3 Year Warranty
More informationOperating instructions Electronic level sensor LK / / 2008
Operating instructions Electronic level sensor LK31 UK 704046 / 00 01 / 2008 Contents Safety instructions...2 Menu structure...3 Controls and indicating elements...4 Function and features...5 Functional
More informationCustomer: Standard type Page 1 of 7 Description Fluxgate current sensor with toroidal core PCB mounting
Customer: Standard type Page 1 of 7 Description Fluxgate current sensor with toroidal core PCB mounting Characteristics Excellent accuracy AEC-Q qualified components Switching open-collector outputs Compact
More informationSolid-State Digital Counter H7AN
Solid-State Digital Counter Preset and Totalizing Counters with Up to 8-Digit LED Displays Draw-out construction allows setting, servicing without disconnecting wiring Choose from selectable UP/DOWN or
More informationKNC-PLC-K506 Series FEATURES DESCRIPTION FEATURES
FEATURES Two High Speed Counters Two Pulse Train Outputs Two Pulse Width Modulation Outputs Inputs 10 Outputs 1 RS232 Port 2 RS485 Ports Supports Modbus RTU Protocol Communicate with up to 32 devices DESCRIPTION
More informationSITRANS F flowmeters SITRANS F M. Transmitter Transmag 2 with sensor 911/E 4/77. Overview
Overview Design The complete flowmeter consists of a flow sensor and an associated transmitter from the Transmag 2 for pulsed alternating field. These are available as remote and compact versions ( 911/E
More informationAA-35 ZOOM. RigExpert. User s manual. Antenna and cable analyzer
AA-35 ZOOM Antenna and cable analyzer RigExpert User s manual . Table of contents Introduction Operating the AA-35 ZOOM First time use Main menu Multifunctional keys Connecting to your antenna SWR chart
More informationPLC-K506 Series FEATURES DESCRIPTION FEATURES
FEATURES Two High Speed Counters Two Pulse Train Outputs Two Pulse Width Modulation Outputs Inputs 10 Outputs 1 RS232 Port 2 RS485 Ports Supports Modbus RTU Protocol Communicate With Up to 32 Devices DESCRIPTION
More information1000TR. Instructions
1000TR ph Instructions CONTENTS 1. INTRODUCTION... 2 1.1 COMMON INTRODUCTION... 2 1.2 PARTS & ACCESSORIES... 2 2. INSTALLATION... 3 2.1 CASING... 3 2.2 MOUNTING... 3 2.3 ELECTRICAL INSTALLATION... 3 2.3.1
More informationTECHNICAL PRODUCT DATASHEET
FORM-ENG-0018 REV A 06-02-03 ISO 9001 CERTIFIED Phone: (352) 629-5020 or 800-533-3569 Fax: (352)-629-2902 SUITABLE FOR OEM DISTRIBUTION ONLY TECHNICAL PRODUCT DATASHEET High Density PDM 21 Output / 10
More informationUser s Manual. F2 Calculator. A metering component for thermal energy applications
User s Manual F2 Calculator A metering component for thermal energy applications Rev 08.24.2007 1 Installation 1.1 At delivery F2 calculators are delivered in the Transport Mode. Only the clock is active
More informationBATCHMATE 1500 Batch Control Computer Technical Bulletin
TS-5(C) BATCHMATE 5 Batch Control Computer Technical Bulletin DESCRIPTION The BATCHMATE features an 8 digit.55-in. alphanumeric LED display. The pulse input model will accept up to 2, pulses per second
More informationGE Infrastructure Sensing. Druck DPI 841/842. Frequency calibrator and Frequency loop calibrator User manual - K395
GE Infrastructure Sensing Druck DPI 841/842 Frequency calibrator and Frequency loop calibrator User manual - K395 A1 B1 10 1 A 2 9 A 3 8 2 3 7 6 11 4 5 B1 12 A2 DPI 842 13 15 14 A3 19 18 17 16 27 20 21
More informationDifferential Current Sensor acc. to the standard UL2231 Ed.2.0. Customer: Standard type Page 1 of 7. Electrical data Ratings min. typ. max.
Customer: Standard type Page 1 of 7 Electrical data Ratings min. typ. max. Unit I P Primary nominal RMS current (1phase / 3phase) 32 40 A I N1 Rated residual operating current 1 5 ma rms I N2 Rated residual
More informationCROSS Chassis from 160 A to 450 A
CROSS Chassis from 160 A to 450 A STS CATALOGUE Important note! The technical data enclosed is for general information. Please note that the operating instructions and references indicated on the products
More informationOperating manual IMPORTANT! READ CAREFULLY BEFORE USE. KEEP FOR THE ENTIRE LIFE OF THE PRODUCT. Heat meters. Deltamess TKS-WM compact heat meter
Operating manual IMPORTANT! READ CAREFULLY BEFORE USE. KEEP FOR THE ENTIRE LIFE OF THE PRODUCT. Heat meters Deltamess TKS-WM compact heat meter 1 Contents Safety and warranty... 3 Radio system... 3 Lithium
More informationT/3000 T/3000. Substation Maintenance and Commissioning Test Equipment
T/3000 Substation Maintenance and Commissioning Test Equipment MULTI FUNCTION SYSTEM FOR TESTING SUBSTATION EQUIPMENT SUCH AS: CURRENT, VOLTAGE AND POWER TRANSFORMERS, ALL TYPE OF PROTECTION RELAYS, ENERGY
More informationInstallation and Operational Instructions for ROBA -switch Type 017._00.2
OBA -switch Type 017._00.2 Guidelines on the Declaration of Conformity A conformity evaluation has been carried out for the product in terms of the EC Low Voltage Directive 2014/35/ EC and the EMC Directive
More informationLOW VOLTAGE CONVERTER WITH THREE POINT GALVANIC INSULATION
EN K09LV LOW VOLTAGE CONVERTER WITH THREE POINT GALVANIC INSULATION General Description The K09LV instrument is a three points galvanic insulator, with input on high sensitivity voltage, and active output.
More informationI P. /dt. di p V S Applications. Standards 1) IEC : 2007; IEC : ) IEC : 2016; IEC : 2017
Ref: ART-B22-D70, ART-B22-D125, ART-B22-D175, ART-B22-D300 Flexible clip-around Rogowski coil for the electronic measurement of AC current with galvanic separation between the primary circuit (power) and
More informationEGC100W MAGMAX. Converter for Electromagnetic Flowmeter GENERAL FEATURES STANDARD SPECIFICATION. General Specification
EGCW Converter for Electromagnetic Flowmeter GENERAL MAGMAX EGCW is the high-performance separate type converter for electromagnetic flowmeters with the compact housing. Improved self-diagnostic functions
More informationEarth 2/3 MI 2126 Instruction Manual Version 1.0, Code No
Earth 2/3 MI 2126 Instruction Manual Version 1.0, Code No. 20 750 190 Distributor: Manufacturer: METREL d.d. Ljubljanska cesta 77 1354 Horjul Slovenia web site: http://www.metrel.si e-mail: metrel@metrel.si
More informationDUAL OUTPUT AC CURRENT/VOLTAGE TRANSDUCER
OPERATOR S MANUAL DUAL OUTPUT AC CURRENT/VOLTAGE TRANSDUCER Masibus Automation & Instrumentation Pvt. Ltd. B/30, GIDC Electronics Estate, Sector-25, Gandhinagar-382044, Gujarat, India Web Site: www..com
More informationNPRG860 NPRG870 REGULATION. Automatic Synchronizer for Generator
REGULATION Automatic Synchronizer for Generator NPRG860 & NPRG870 perform synchronization and paralleling of generators with electrical network. NPRG860 features a speed adjustment function. NPRG870 adds
More informationWATERFLUX 3000 Quick Start
WATERFLUX 3000 Quick Start Electromagnetic flow sensor The documentation is only complete when used in combination with the relevant documentation for the signal converter. KROHNE CONTENTS WATERFLUX 3000
More informationTemperature monitoring relays CM-TCS Monitoring relays for monitoring temperatures with a PT100 sensor (2- or 3-wire connection)
Data sheet Temperature monitoring relays CM-TCS Monitoring relays for monitoring temperatures with a PT100 sensor (2- or 3-wire connection) The temperature monitoring relays CM-TCS monitor overtemperature,
More informationD7F. Detects Changes in Machine. Vibration. Vibration Sensor with Linear Output or Manual Setting. Ordering Information
with Linear Output or Manual Setting Detects Changes in Machine ON/OFF output can be set and checked quickly and easily from the vibration level meter. waveforms can be checked using the AC monitor output.
More informationInstallation and Operational Instructions for ROBA -multiswitch Type 019._00.2
Guidelines on the Declaration of Conformity A conformity evaluation has been carried out for the product in terms of the EU Low Voltage Directive 2014/35/ EU and the Electromagnetic Compatibility (EMC)
More informationType Nominal voltage (V) Code HSV V
Speed control Speed control relay - mm Control of overspeed, underspeed, operating rate, stopping Measurement via discrete sensors - -wire PNP or NPN, Namur, voltage 0-0V or volt-free contact type Works
More informationADC5000 SERIES. AC/DC Switch Mode Power Supplies and Rectifiers for Industrial and Telecom Applications. 60W, 125W and 250 W
ADC5000 SERIES AC/DC Switch Mode Power Supplies and Rectifiers for Industrial and Telecom Applications 60W, 125W and 250 W Input voltage 230/115 VAC voltages 12, 24, 36 or 48 VDC Statistical MTBF >3 000
More informationHigh precision measurement system for current and voltage IHC-A/B-RM01/03
Measurement Offset-free and low-noise 16 bit data acquisition system ISA-ASIC Internal sample rate 3,500 Hz Communication Standard RS232- or RS485 interface Advantages Direct measurement on the bus bar
More informationBATCHFLUX 5500 C Technical Datasheet
BATCHFLUX 5500 C Technical Datasheet Electromagnetic flowmeter for volumetric filling machines High-stability ceramic measuring tube, therefore extremely long-term stability Optimum hygienic design Only
More informationBrunata Optuna H Ultrasonic energy meter Type 775 Installation Guide Edition 1.2
Ultrasonic energy meter Type 775 Installation Guide Edition 1.2 UK-QB101575 / 29.05.2012 Brunata a/s is a Danish owned company. We have more than 90 years of experience within developing and producing
More informationM1 Series. Humidity - Temperature Transmitter INSTRUCTION MANUAL
M1 Series Humidity - Temperature Transmitter INSTRUCTION MANUAL 20031110 -2- CONTENTS Overview... 3 Operation... 4 Power supply... 4 Operating range and limits... 4 Temperature compensation of the humidity
More informationI P. /dt. di p V S+ Applications. Standards. 1) IEC ed1.0: 2007; IEC : ed1.0: 2012
Ref: ART-B22-D70, ART-B22-D125, ART-B22-D175 Flexible clip-around Rogowski coil for the electronic measurement of AC current with galvanic separation between the primary circuit (power) and the secondary
More informationUNICONT. PMG-400 Universal controller and display unit USER'S AND PROGRAMMING MANUAL 1. pmg4111a0600p_01 1 / 24. ST edition
UNICONT PMG-400 Universal controller and display unit USER'S AND PROGRAMMING MANUAL 1 ST edition pmg4111a0600p_01 1 / 24 TABLE OF CONTENTS 1. GENERAL DESCRIPTION... 3 2. ORDER CODE... 3 3. TECHNICAL DATA...
More informationWATERFLUX 3000 Quick Start
WATERFLUX 3000 Quick Start Electromagnetic flowmeter The documentation is only complete when used in combination with the relevant documentation for the signal converter. KROHNE CONTENTS WATERFLUX 3000
More informationPortable Appliance Testers. OmegaPAT MI 2140 BetaPAT MI 2141 User Manual Ver Code No
Portable Appliance Testers OmegaPAT MI 2140 BetaPAT MI 2141 User Manual Ver. 1.2. Code No. 20 750 684 Distributor: Producer: METREL d.d. Ljubljanska 77 SI-1354 Horjul E-mail: metrel@metrel.si http://www.metrel.si
More informationDIGEM f 96 x 48 EK. Applications. Description
3-349-002-03 1/7.98 Front panel dimensions: 96 x 48 mm LED display: red or green, 14 mm high Max. display range: 19 999 to + 32 765 Modular connectors for flexible use Simple adjustments at front panel
More informationAn RS485 bus is used for command, monitoring and diagnostic information that can be supplied to a system controller.
The TCP3500 series is an AC-DC converter with adjustable DC output and universal 3-phase AC input. Conduction cooling (No Fans) makes this power supply series suitable for a wide variety of Industrial
More information3 AXIS RF ELECTROMAGNETIC FIELD METER
50 MHz to 3 GHz Radio Frequency Radiation Meters Electromagnetic Field strength measurement 3 AXIS RF ELECTROMAGNETIC FIELD METER Model : EMF-819 Your purchase of this RF EMF METER marks a step forward
More informationMODEL 3810/2 Line Impedance Stabilization Network
EMC TEST SYSTEMS FEBRUARY 1996 REV C PN 399197 MODEL 3810/2 Line Impedance Stabilization Network OPERATION MANUAL USA P.O. Box 80589 Austin, Texas 78708-0589 2205 Kramer Lane, Austin, Texas 78758-4047
More informationHigh-set undervoltage stage with definitetime. or inverse definite minimum time (IDMT) characteristic. Low-set undervoltage stage with definitetime
Issued: 5.06.999 Status: 5.06.999 Version: B/09..00 Data subject to change without notice Features Overvoltage and undervoltage protection Single- or three-phase operation High-set overvoltage stage with
More informationOPTITEMP TT 20 Technical Datasheet
OPTITEMP TT 20 Technical Datasheet Analog PC-programmable two-wire transmitters for Pt100 Efficient PC-configuration without recalibration Very stable output Very fast response time The documentation is
More informationModel Number Structure. Ordering Information. Solid-state Power OFF-delay Timer H3DE-H. Model Number Legend. List of Models
Solid-state Power OFF-delay Timer H3DE-H Timers Two delay-time models available. 0.1 to 12 seconds (S Series) 1 to 120 seconds (L Series) Covers wide range of supply voltage. Model Number Structure Model
More informationNPRG860 & NPRG870 perform synchronization and paralleling of generators with electrical network. NPRG860 features a speed adjustment function.
REGULATION AUTOMATIC SYNCHRONIZER for GENERATOR NPRG860 & NPRG870 perform synchronization and paralleling of generators with electrical network. NPRG860 features a speed adjustment function. NPRG870 adds
More informationLaboratory Grade Instruments Series & 4021 Power Meter SERIES Power Sensor
Laboratory Grade Instruments 4020 Series & 4021 Power Meter Semiconductor 4020 SERIES Power Sensor 4021 4022 4024 4025 Power Input 300 mw to 1 kw 300 mw to 1 kw 3 W to 10 kw 3 W to 10 kw (1.2 kw max.)
More informationINSTALLATION MANUAL. Model: Smart Analyzer Manufacturer: Smart Impulse. Power meter with consumption breakdown by use 03/12/13
INSTALLATION MANUAL Model: Smart Analyzer Manufacturer: Smart Impulse Power meter with consumption breakdown by use 03/12/13 Table of contents Table of contents... 2 1. Introduction... 3 2. Installation
More informationWRM-10 TM TRANSFORMER WINDING RESISTANCE METER
WRM-10 TM TRANSFORMER WINDING RESISTANCE METER USER S MANUAL Vanguard Instruments Company, Inc. 1520 S. Hellman Ave. Ontario, California 91761, USA TEL: (909) 923-9390 FAX: (909) 923-9391 June 2009 Revision
More informationLiquid level monitoring relay CM-ENS.2x
Data sheet Liquid level monitoring relay CM-ENS.2x The CM-ENS.2x is served to regulate and control liquid levels and ratios of mixtures of conductive fluids. It can be used for overflow protection, dry
More informationApplications: Power generation, Cogeneration, Stationary power Large engines
Features: 8 inputs are user selectable from the following. 0-5V, 0-10V, 4-20 ma or 0-20 ma Digital inputs for interface to switches, etc. PWM signal, pulse or 16-bit counter inputs from sensors or diesel
More informationAPPENDIX APPENDIX A 1
A 1 SPECIFICATIONS Ratings Supply voltage 100 to 240 VAC, 50/60 Hz 24 VAC, 50/60 Hz/24 VDC Operating voltage range 85 to 110% of rated supply voltage Power consumption 7VA 4VA/2.5W Sensor input Thermocouple
More informationDIGITAL SYNCHRONIZING UNIT KS3.2 SERVICE MANUAL. 144 x 144 x 77 mm C E R T I F I E D
DIGITAL SYNCHRONIZING UNIT KS3.2 C E R T I F I E D 144 x 144 x 77 mm SERVICE MANUAL 1 CONTENTS 1. APPLICATION... 3 2. DELIVERY SPECIFICATION... 3 3. BASIC SAFETY REQUIREMENTS AND USER S SAFETY... 4 4.
More informationIFC 010 Signal Converter
KROHNE 04/2003 7.02347.22.00 GR Signal Converter for electromagnetic flowmeters Easy to install Low-cost signal converter, measuring error < 0.5% of the measured value Easy user-friendly operation Self-monitoring
More informationInstruction manual SMART 96-2 / WG-2 ph
Instruction manual SMART 96-2 / WG-2 ph Stand: 06.09.2015 SMART 96-2 / WG-2 ph / mv Instructions for Installation Configuration PH-mV / Temp. 0...20 ma 4...20mA 20 ma PH-mV / Temp. 0...20 ma 4...20mA 20
More informationENGLISH ENGLISH ENGLISH ENGLISH. Installation and User Manual version 1.03 TLB4 2014/30/UE EN55022:2010 EN :2005 EN :2007
ENGLISH ENGLISH ENGLISH ENGLISH Installation and User Manual version 1.03 TLB4 2014/30/UE EN55022:2010 EN61000-6-2:2005 EN61000-6-4:2007 SYSTEM IDENTIFICATION KEY TO SYMBOLS Below are the symbols used
More informationSimplified Operations for TM-3100 Series Digital Tachometer
Simplified Operations for TM-3100 Series Digital Tachometer The TM-3100 series Digital Tachometers has the most fundamental function in rotational measurement of four standard models (rotation speed display,
More information1) What meter do I need in this application?
ELECTRONIC ELECTRONIC Choices 1) What meter do I need in this application? Meter choice is determined by: Level of accuracy required, flowrate, line size, viscosity, fitting type, pressure rating, temperature,
More informationInstruction Manual ABM HART Gateway Software. Instruction Manual Revision A.1
Instruction Manual ABM HART Gateway Software Instruction Manual Revision A.1 Table of Contents Section 1: Getting Started... 3 1.1 Setup Procedure... 3 1.2 Quick Setup Guide for Ultrasonic Sensors... 11
More informationOperator Manual. Transmitters Series KAT-... Analog transmitters in 2-wire technique
Operator Manual Transmitters Series KAT-... Analog transmitters in 2-wire technique Important Safety Notes! In the case of non-observance of the installation and safety notes or improper use of the transmitter
More informationModel FLSC-C1-LIQ. Microprocessor Based Loop Powered Transmitter
Model FLSC-C1-LIQ Microprocessor Based Loop Powered Transmitter CONTENTS 1. Introduction ------------------------------------------------------------ 2 2. Specifications-----------------------------------------------------------
More informationKMS101W. ManuFlo Page 1 MANU ELECTRONICS PTY LTD Flow Measurement Products. Electromagnetic Wafer Flowmeters with S100 display
KMS101W Electromagnetic Wafer Flowmeters with S100 display ideal as resettable/batching meter (sizes: 15mm to 150mm) 0BFEATURES For ADMIXTURE Batching, Shotcrete, Mild-Recycled Water & Selected Chemical
More informationSPECIFICATION FOR OVERCURRENT RELAYS
SPECIFICATION FOR OVERCURRENT RELAYS 1/9 1. Procedure A manufacturer willing to classify an overcurrent relay according to this specification should provide: A complete file providing a clear, unambiguous
More informationEnergy Management Energy Meter Type EM111
Energy Management Energy Meter Type EM111 Easy connection or wrong current direction detection Other versions available (not certified, option X: see how to order on the next page Single phase energy meter
More informationGEMINI PRESETTABLE DUAL RATE INDICATOR WITH RATIO (A/B)
Bulletin No. GEM5-C Drawing No. LP0321 Released 1/05 GEMINI 5200 - PRESETTABLE DUAL RATE INDICATOR WITH RATIO (A/B)! DIFFERENCE (A-B), OR DRAW [(A-B)/B] INDICATION! 6-DIGIT, 0.56" (14.2 mm) HIGH LED DISPLAY
More informationEnergy Management Energy Meter Type EM112
Energy Management Energy Meter Type EM112 Easy connection or wrong current direction detection Other versions available (not certified, option X: see how to order on the next page Single phase energy meter
More informationPD30CNB25xxPS. Photoelectrics, Background suppression reflective - PointSpot. Main features. Main functions. Description
Photoelectrics, Background suppression reflective - PointSpot Main features Miniature sensor range The visible red PointSpot light makes alignment very easy Range: 250 mm Sensitivity adjustment by potentiometer
More informationThe keypad is used to set the flowmeter. The function of the keys are as follows: This key selects the figures to be changed.
7. 7.1 Keypad and display layout Keypad The keypad is used to set the flowmeter. The function of the keys are as follows: TOP UP KEY FORWARD KEY BACKWARD KEY CHANGE KEY SELECT KEY LOCK/UNLOCK KEY This
More informationHeart Rate Noise Detector REV. 1.1
Heart Rate Noise Detector REV. 1.1 HR Noise Detector: quick reference rev. 1.1 Contents 1. INTRODUCTION... 1.1 2. TECHNICAL NOTES ON CARDIO RECEIVERS... 2.1 2.1. GENERAL INFORMATION... 2.1 2.2. ELECTROMAGNETIC
More informationT6+ Analog I/O Section. Installation booklet for part numbers: 5/4-80A-115 5/4-90A-115 5/4-80A /4-90A-1224
T and T+ are trade names of Trol Systems Inc. TSI reserves the right to make changes to the information contained in this manual without notice. publication /4A115MAN- rev:1 2001 TSI All rights reserved
More informationFuzzy Temperature Controllers E5AF
Fuzzy Temperature Controllers 1/4 DIN Controller Combines Fuzzy and PID Control For Fast Response to Process Disturbances Advanced PID control provides optimal response during start-up and steadystate
More informationSignet Flow Transmitter
Signet 80- Flow *-80.090-* English -80.090- Rev. L 0/ English WARNING! Remove power to unit before wiring input and output connections. Follow instructions carefully to avoid personal injury. Contents.
More informationN24, N25 DIGITAL PANEL METER
DIGITAL PANEL METER www.sifamtinsley.co.uk USER S MANUAL Multifunction Meters Transducers & Isolators Temperature Controllers Converters & Recorders Digital Panel Meters DIGITAL PANEL METER Current Transformers
More informationPlease enter the identity code of your device here!
Operating Instructions DULCOMETER D1C Part 2: Adjustment and Operation, Measured Variable Conductive Conductivity ProMinent D1C2-Leit.-001-GB Conductivity 100 µs/cm DULCOMETER STOP START Conductivity 100
More informationPULSE/FREQUENCY/CONDITIONER USER GUIDE
PULSE/FREQUENCY/CONDITIONER USER GUIDE D2560-02-02 CN5220 1600F USER GUIDE Page 1 of 13 PULSE/FREQUENCY/CONDITIONER USER GUIDE FREQUENCY INPUT PSU 1 2 + - SUPPLY (to4) Vdc (to32) Vac UNIVERSL Excitation
More informationInstallation and Operational Instructions for ROBA -switch Type 017._00.2
Manufacturer s Declaration This product is intended for installation in a machine or system, based on the machine directive 2006/42/EC. It is forbidden to start use of the product until the machine or
More informationTemposonics. E-Series Model ER. Magnetostrictive, Absolute, Non-contact Linear-Position Sensors. Analog and Start/Stop Outputs.
Temposonics Magnetostrictive, Absolute, Non-contact Linear-Position Sensors Analog and Start/Stop Outputs SENSORS Document Part Number: 550996 Revision E Data Sheet FEATURES Linear, Absolute Measurement
More information