sw rev. 55B DEVICE FOR THE PERMANENT CONTROL OF INSULATION ON THE SUPPLY LINES IN MEDICAL PRACTICE PLACES INSTALLATIONS

Size: px
Start display at page:

Download "sw rev. 55B DEVICE FOR THE PERMANENT CONTROL OF INSULATION ON THE SUPPLY LINES IN MEDICAL PRACTICE PLACES INSTALLATIONS"

Transcription

1 INSTRUCTIONS MANUAL IM831-U v0.92 HRI-R40W sw rev. 55B DEVICE FOR THE PERMANENT CONTROL OF INSULATION ON THE SUPPLY LINES IN MEDICAL PRACTICE PLACES INSTALLATIONS INDEX: GENERAL TYPES ACESSORIES AND OPTIONS INSTALLATION WIRING DIAGRAMS CONNECTING TERMINALS DIAGRAMS WIRING DIAGRAM DESCRIPTION WORKING DESCRIPTION AND KEYBOARD WORKING DESCRIPTION AND PROGRAMMING KEY s FUNCTIONS PROGRAMMING INSTALLATION SETUP CONFIGURATION SETUP OPERATION DIMENSIONS TECHNICAL CHARACTERISTICS GENERAL HRI-R40W is a device for controlling the supply lines of medical practice places (installations with insulated neutral IT-M) which allows the insulation control towards earth of the supply network so as the thermal and electric overload of the transformer. In such a way it accomplishes what it is required by the specific standards regarding the present applications. The insulation resistance control is being made by applying a measure signal between the insulated line and the earth. By measuring the current flowing to earth it is possible to supervise the insulation level. The actual state of the art of measures allows the correct measure of the insulation value, even in presence of great disturbances with high harmonic components and direct current components. The HRI-R40W type uses a direct-current component control signal. In order to reduce the problems generated by the presence of direct current components in the line (presence of rectifiers, etc.,) this device is fitted with a digital filter capable to identify the direct current component present in the line. HRI-R40W has a wide range of programming with the front keys and the digital LED type display of 3 digits for visualising of measuring and programming parameters. The instrument is fitted with 2 temperature inputs for thermal s or PTC (DIN 44081) type, for controlling the thermal overload of the insulation transformer, and 1 measuring current input, by an external CT, for controlling the line s overload. The outputs can be used with the remote signalling panels type PR5. There is a relay output programmable by the user. There is one RS485 serial port output as an option for a bidirectional communication with a control system (PLC, PC, etc.). The communication protocol is the MODUBUS-RTU (see specific manual). The instrument complies with the following standards: EN / IEC / CEI 64.8/7-710 V2 / UNE TYPES: HRI-R40W Vaux: V 50-60Hz ACCESSORIES Accessories: - Remote flush-mounted signal panel (universal box E503) type PR5 HRI-R40W Instructions manual IM831-U v0.92 pag. 1 / 12

2 INSTALLATION The installation must be carried out in absence of voltage by qualified and authorized personnel only. Before installing, make it sure that the instrument has not suffered any damage due to transport. It has to be verified that the supply voltage is compatible with the product s defined supply voltage. Each independent line can use one control device only. The modular instrument (6 modules 17,50mm) is fitted of sealable transparent cover, to be mounted on DIN rail mounting 35mm. WIRING DIAGRAM Connecting terminals A Auxiliary power supply (rated voltage 115 / 230V 50-60Hz) B Link fail function input (permanent control of network connection) C Measure input for resistance and insulation impedance D Connection to remote signal panels type PR5 E Serial port RS485 F Thermal input 2 for or PTC G Thermal input 1 for or PTC H Current measuring input with external current transformer /5 A I Programmable auxiliary relay output (changeover contact 250V 5A) HRI-R40W Instructions manual IM831-U v0.92 pag. 2 / 12

3 Wiring diagram with PR4 panel for single phase line IT-M with insulation transformer and centre intake of the secondary: Vaux:* 115V 115V RS485 serial port (option) Vaux:* link fail 115V 115V control input 2 (or PTC) input 1 (or PTC) current input TA.../5A auxiliary relay output HRI-R40W insulation control PR4 remote control panel remote control panel PR power supply Vaux: 115Vca power supply Vaux: 230Vca PE S S current transformer.../5a PC L INSULATED NETWORK (IT-M) LOAD N INSULATION TRANSFORMER to other PR4 panels HRI-R40W Instructions manual IM831-U v0.92 pag. 3 / 12

4 Wiring diagram with PR5 panel for single phase line IT-M with insulation transformer and centre intake of the secondary: 19 A B C PT 2+A PT2+B PT2- PT1+A PT1+B PT1- CT-S2 CT-S1 R-NO R-C R-NC RS 485 serial port (option) Vaux:* link fail 115V 115V control input 2 ingput 1 current input TA.../5A insulation control remote control panel auxiliary relay output HRI-R40W AUX -L1 AUX-01 AUX -L2 AUX-02 LF-PE LF-L VC-L VC-PE TE ST+ TE ST- ACK ACUS OVER COM-P V-P FAULT remote control panel PR V-P ACUS ACK OVER FAULT COM-P TEST- TEST+ COM-S F-S TEST-S auxiliary power supply Vaux: 230Vca PE TO OTHERS PR5 PANELS S S PC L ISOLATED NETWORK (IT-M) LOAD N INSULATION TRANSFORMER HRI-R40W Instructions manual IM831-U v0.92 pag. 4 / 12

5 Three phase wiring diagram in IT-M line: RS485 serial port (option) input 2 input 1 current input TA.../5A auxiliary relay output HRI-R40 link fail control insulation control PR4 remote control panel to PR4 panel auxiliary power supply Vaux: 230Vca adder current transformer 5A A S1 S1 S1 LOAD S1 S1 P1 S1 P1 INSULATION TRANSFORMER INSULATED NETWORK (IT-M) P1 current transformers.../5a Wiring diagram thermal s type : Wiring diagram thermal s type PTC: PTC PTC R = 120 ohm R R PT2+A PT2+B PT2- PT1+A PT1+B PT input 2 input 1 PT2+A PT2+B PT2- PT1+A PT1+B PT input 2 input 1 HRI-R40W Instructions manual IM831-U v0.92 pag. 5 / 12

6 Wiring diagram measure section on single-phase phase IT-M line with insulation transformer without the centre tap of transformer s the secondary: Wiring connection measure section on threephase IT-M line three wires without neutral link fail control insulation control link fail control insulation control auxiliary power supply Vaux:230 Vca auxiliary power supply Vaux:230 Vca INSULATED NETWORK (IT-M) L1 L2 INSULATION TRANSFORMER L3 INSULATION TRANSFORMER INSULATED NETWORK max 230V L-L (IT-M) WIRING DIAGRAM DESCRIPTION HRI-R40 Auxiliary power supply Terminals The section of power supply is made with double input 115V of rated voltage. To supply the instrument with 230V, connect the two sections in series. To supply the instrument at 115V connect the two section in parallel. Link fail control Terminals 6-7 The two terminals must be connected between one phase of the insulated line and the equipotential node (PE). The maximum voltage is 250V (see insulation control). Insulation control Terminals 8-9 The two terminals must be connected between the centre tap of the transformer or to one phase of insulated line and the equipotential node (PE) The maximum voltage is 230Vac. Consequently the single phase network could be 230V, the three-phase 3 wires 230V phase-phase, whilst the three-phase 4 wires will be max. 230V phase-neutral. Connection to remote signal panel type PR5 Terminals The maximum voltage on these conductors is 24V. RS485 serial port (option) Terminals The serial bus is on the A-B terminals (19-20), the C terminal (21) is a ground signal that can be connected to the RS485 shield (if present). MODBUS-RTU is the standard used protocol (see the relative manual). HRI-R40W Instructions manual IM831-U v0.92 pag. 6 / 12

7 Thermal input 2 Terminals It s possible to use (EN 60751) or PTC (DIN 44081) sensor type. If with 3 wires is used, it is required to connect a compensation conductor at the same terminal. With the PTC it s necessary to use an external resistor of 120 ohm min ¼ W. Thermal input 1 Terminals It s possible to use sensor (EN 60751) or PTC sensor (DIN 44081). If with 3 wires is used, it is required to connect a compensation conductor at the same terminal. With the PTC it s necessary to use an external resistor of 120 ohm min ¼ W. Measuring current input Terminals Connection to external current transformer with secondary 5 A (the transformation ratio is programmable). The current is measured in absolute value: A possible inversion of S1-S2 is not of great importance. Output auxiliary relay Terminals Voltage free changeover contact with programmable functions. Contact range: 250V 5A resistive load. FUNCTIONS DESCRIPTION AND KEYS R [k ] Z [k ] T1 [ C] T2 [ C] I [A] SET MIN / MAX + _ insulati on moni tor HRI-R40 ALARM OUTPUT RELAY RESET SET ERROR / LINK FAIL TEST ENTER DESCRIPTION: 1. Display to visualise the settings and the parameters under the control 2. Green SET LED to signal the instrument programming condition 3. Yellow ALARM LED to signal the alarm of parameter value beyond of threshold 4. Red LED OUTPUT RELAY to signal the condition of the auxiliary relay 5. Red LED ERROR / LINK FAIL to signal the alarm of internal failure, lack of connection of the line under control, thermal type open or in short circuit 6. Red LED R to signal the visualisation of the insulation resistance parameter (k ); flashing led if the parameter is beyond of threshold 7. Red LED Z to signal the visualisation of the insulation impedance C, flashing led if the parameter is beyond of threshold 8. Red LED T1 to signal the visualisation of the transformer s temperature parameter, flashing led if the parameter is beyond of threshold 9. Red LED T2 to signal the visualisation of the temperature parameter of the second, flashing led if the parameter is beyond of threshold 10. Red LED I to signal the visualisation of the line s current, flashing led if the parameter is beyond of threshold 11. +/- UP/DOWN keys to select the parameter to display, to configure the setting of the instrument and to visualise the maximum and minimum values stored. 12. RESET SET key to accede to the instrument programming, to silence the alarms and to delete the values stored 13. TEST ENTER key to test the instrument and the PR4 remote signal panel; to confirm the settings of the SETUP HRI-R40W Instructions manual IM831-U v0.92 pag. 7 / 12

8 FUNCTIONS AND PROGRAMMING 888 R [k ] Z [k ] T1 [ C] T2 [ C] I [A] SET MIN / MAX + _ insulation monitor HRI-R40 ALARM OUTPUT RELAY RESET SET ERROR / LINK FAIL TEST ENTER E D C B A Keys functions With normal operation the + and keys (D and C) allows to select the parameter to be visualised on the display and indicated by E led. ATTENTION: if a measure is not enabled (for example the T2 temperature input is not enabled), the relative LED will remain off. If only the resistance measure is enabled, the visualisation will be fixed. The B key allows silencing the acoustic signalling of the connected PR5 panels. The function is the same as the SILENCE function in the remote panels. In minimum and maximum visualisation mode this key allows the deleting of stored values. The A (TEST) key allows the functional test of the instrument simulating a low insulation: holding pressed the key it will appear on the display the simulated value of insulating resistance that it will go down (up to 0) causing the alarm trip for low resistance. The instrument comes back to the effective measure value when the key is released Visualisation of minimum and maximum memorised values By pressing at the same time the D C keys will accede to the visualisation of the minimum and maximum memorised values that will be displayed with a continuous flashing. With the C key it is possible to visualise the values of the various parameters. During the visualisation of the maximum and minimum values, it s possible to reset the memorised values, by pressing the B key (RESET): keep pressed the key during 3 seconds at least, until the message RES (RES) appears on the display in order to confirm the cancellation. To escape from the minimum and maximum visualisation it s necessary to wait 8 seconds without pressing any key. HRI-R40W Instructions manual IM831-U v0.92 pag. 8 / 12

9 PROGRAMMING The SETUP menu (programming) is subdivided in two sections: in the first one it s possible to set the condition of basic functioning (installation), whilst in the second it is possible to set the tripping threshold levels and the activation mode of the auxiliary relay. Installation SETUP: To accede to the installation SETUP press the C - B (SET / -) keys simultaneously during 3 seconds at least. The glowing of the LED SET confirms the entrance in the SETUP. It s possible to modify the parameters setting with the A-B keys. Press the A key (ENTER) to confirm. The setup is cyclic: after the last parameter, the setup comes back to the first. To escape from the SETUP it s necessary to wait 8 seconds without pressing any key. Press A or B keys to increase or decrease the values respectively (the variation is being made unit by unit). In order to speed up the operation, keep the B or C keys pressed, the variation will be made by tens and hundreds. To increase or decrease the value by unit, it is necessary to release and to press the key again. The OFF selection is selectable under the minimum of the scale of the parameter s setting. The indication of the parameter s setting blinks alternatively with a help message. Installation SETUP: trip delay Function measure enabling / trip delay measure enabling / type measure enabling / type measure enabling / trip delay transformation rate nominal frequency network under control not activated auxiliary relay status serial port setting (option) low insulation resistance low insulation impedance temperature input T1: temperature input T2: Setting Display indication Signalling (default) parameter flash led SET + led R 1 4 seconds DLY led SET + led Z led SET + led T1 led SET + led T2 off 1 4 seconds off PTC off PTC high current I led SET + led I off 1 60 seconds external current led SET + led I off transformer 1 40 Fail Safe function Off OFF ptc pt1 OFF ptc pt1 Off Off DLY MOD MOD DLY CT led SET 50 / 60 Hz 50_ / 60_ FRE led SET + led OUPUT RELAY blinking De-energized / excited (fail safe on) address network led SET Id address baud rate led SET baud parity, data bit, stop bit led SET none,8,1 none,8,2 even,8,1 odd,8,1 NOP / NCL SET ID N81 N82 E81 O81 The default parameters are underlined. The meaning of the delay for the resistance and the insulation impedance is an elaboration time for the internal filter. This time could be increased in presence of strong disturbances in the line under control. BDR PAR HRI-R40W Instructions manual IM831-U v0.92 pag. 9 / 12

10 Configuration SETUP: To accede to the configuration SETUP keep pressed the B key (SET) during 3 seconds at least. It s possible to modify the parameters setting with the A-B keys. Press the A key (ENTER) to confirm. The setup is cyclic: after the last parameter, the setup comes back to the first. To escape from the SETUP it s necessary to wait 8 seconds without pressing any key. Press A or B keys to increase or decrease the values respectively (the variation is being made unit by unit). In order to speed up the operation, keep the B or C keys pressed, the variation will be made by tens and hundreds. To increase or decrease the value by unit, it is necessary to release and to press the key again. Configuration SETUP Function (if activated) Configuration display indication Signalling (default) parameter flash led R kohm SET tripping threshold insulation resistance low auxiliary relay activation for low resistance tripping threshold low insulation impedance auxiliary relay activation for led Z + led OUTPUT RELAY off / on ON / Off REL low impedance flashing auxiliary relay activation for led OUTPUT RELAY + led off / on ON / Off REL error condition and Link Fail ERROR/LINK FAIL flashing tripping threshold high temperature T1: if selected led T1 off / C OFF / SET if PTC selected led T1 off / on ON / Off REL led R + led OUTPUT RELAY off / on ON / Off REL flashing led Z off / kohm OFF / SET auxiliary relay activation for led T1 + led OUTPUT RELAY off / on ON / Off REL high temperature T1 flashing tripping threshold high temperature T2: if selected led T2 off / C OFF / SET if PTC selected led T2 off / on ON / Off REL auxiliary relay activation for led T2 + led OUTPUT RELAY off / on ON / Off REL high temperature T2 flashing tripping threshold high current I led I off / 00,0 99,9 A OFF / SET auxiliary relay activation for high current I The default parameters are underlined. led I + led OUTPUT RELAY flashing off / on ON / Off REL Note: The resistance measure function is always enabled to the measure, the others functions can be set only if they are enabled in the installation SETUP. To come back to default value When the instrument is being switched on press the TEST and RESET keys simultaneously to come back to the default value for all SETUP. After 4 seconds it will appear the following messages: DEF, VAL, INI, YES (DEF, VAL, INI, YES). When the instrument is switched on again, the SETUP parameters will come back to default values. HRI-R40W Instructions manual IM831-U v0.92 pag. 10 / 12

11 OPERATION In the normal condition of operation the instrument shows the measure of the selected parameter. It is possible to select all of available measures (if enabled) with the + / - keys. By pressing + / - keys simultaneously it will accede to the maximum and minimum visualisation and the measure on display flashes. If the s used are PTC the value showed will be only TLO (low temperature), HOT (temperature near to the tripping threshold) or THI (high temperature). If a measure parameter exceeds the maximum end-scale, on the display it will appear the message HI_ (hi_ HIGH).The OUPUT RELAY led will glow only if the auxiliary relay is activated. The visualisation of the enabled measure parameters could be substituted by an error message (the ERROR /LINK FAIL LED glows). The signalling are the following: Measured parameter Visualisation Error condition Visualisation (if activated) led display (Led ERROR) led display insulation resistance led R / HI_ / HI_ Link Fail led R LF_ LF_ insulation impedance led Z / HI_ / HI_ led Z LF_ LF_ T1 temperature () led T / HI_ / HI_ in short circuit led T1 SHr (short) SHR T1 temperature () led T / HI_ / HI_ open circuit led T1 OPE (open) ope T1 temperature (PTC) led T1 TLO/HOT/THI TLO/HOT/THI T2 temperature () led T / HI_ / HI_ in short circuit led T2 SHr (short) SHR T2 temperature () led T / HI_ / HI_ open circuit led T2 OPE (open) ope T2 temperature (PTC) led T2 TLO/HOT/THI TLO/HOT/THI current I led I 0 99,9 / HI_ ,9 / HI_ The glowing of the LED ALARM indicates the condition of a parameter beyond threshold. The LED of the signalled parameter will flash if the visualisation is not selected. The signalling are the following: Measured parameter (if activated) Visualisation insulation resistance led R / HI_ (high) / HI_ insulation impedance led Z / HI_ (high) / HI_ T1 temperature led T / HI_ (high) () / HI_ T1 temperature led T1 TLO (low temp.) / HOT (PTC) TLO / HOT T2 temperature: led T / HI_ (high) () / HI_ T2 temperature: led T2 TLO (low temp.) / HOT (PTC) TLO / HOT current I led I 0 99,9 / Hi_ (high) ,9 / Hi_ Alarm conditions display flashing signalling / Hi_ /HI_ THI (high temp.) THI / Hi_ /HI_ THI (high temp.) THI 1 99,9 / Hi_ ,9 / Hi_ led R led Z led T1 led T1 led T2 led T2 NOTE Enabling measures. If the measures are not enabled in the installation SETUP, the insulation resistance R is the only measure parameter always enabled. In this condition no modification will be done by pressing the + and key. DIMENSIONS led I _ RESET TEST 45 DIN rail mounting 35mm according with DIN modules 17,5 mm HRI-R40W Instructions manual IM831-U v0.92 pag. 11 / 12

12 TECHNICAL FEATURES Auxiliary supply voltage V 50-60Hz ±20% Maximum consumption 5 VA Voltage of the line under control V 50-60Hz Maximum measuring current 1 ma Maximum measuring voltage 24V Control signal type Composite codify signal Internal impedance 200 kohm Insulation measure kohm / HIGH - resolution 1 k, accuracy 5% ± 1 digit Impedance measure kohm / HIGH - resolution 1 k, accuracy 5% ± 1 digit (test signal 2500 Hz) Temperature measures 2 inputs (one optional) from thermo type Rd at 2 or 3 wires C, - resolution 1 C, accuracy 2 % ± 1 digit or from PTC (DIN 44081) Current measure from external CT with secondary 5A, accuracy 5% ±1 digit, (transformation ratio 1 40) Tripping threshold low insulation resistance kohm - hysteresis 10% - delay settable, automatic reset typical tripping delay: 3 seconds, max 8 seconds low insulation impedance (possible to unable) kohm - hysteresis 10% - delay settable, automatic reset typical tripping time:3 seconds, max 8 seconds over temperature input 1 (possible to unable) C with - hysteresis 10%, automatic reset - (with PTC depending on the ) over load current ,9 A delay settable 1 60 seconds, automatic reset Diagnostic signalling Instrument out of line (link fail), thermal open or short circuit, internal error, interference error Outputs output to supply PR5 panel (up to 2 remote signal panel), max 24Vdc output for PR5signalling panel for low insulation, over temperature transformer, insulation and overload of the line auxiliary relay output with programmable function, NO-C-NC 5A 250Vac to enable for low insulation, low impedance, over temperature 1 and 2, over load, link fail or error, fail safe function Optional: serial port RS485, standard protocol MODBUS-RTU, baud rate 2,4-4,8-9,6-19,2 kbps / n,8,1 n,8,2 e,8,1 o,8,1 Visualisation insulation resistance value and signalling of value over maximum scale measured temperature value by measure channel 1 and 2 and signalling of value over maximum scale measured current value and signalling of value over maximum scale insulation impedance value and signalling of value over maximum scale programming parameter instrument off line (link fail), temperature error and internal error output status storing of minimum insulation, maximum temperature and maximum current Dimension, weight self-extinguishing thermoplastic enclosure - DIN 35mm mounting dimension 6 modules, protection by sealable transparent cover - 0,4 kg Connection screw terminals max cross section 2,5 mm 2 Protection degree IP50 frontal side, IP20 enclosure Environmental operating temperature C, storage temperature C, humidity < 95% Insulation 2,5 kv 60 sec. Standards safety CEI-EN of product CEI-EN / CEI 64.8/7-710 V2/ IEC / UNE electromagnetic compatibility CEI-EN Remark In consideration of the evolution of the products and standards, the company reserves the right to modify at any time the features of the product described in this literature, therefore we recommend to always verifying them beforehand. The manufacturer s liability for damages resulting from product defects may be reduced or deleted ( ) when the damage is attributable jointly to a product defect and to the negligence of the injured party or of a third party for whom the injured party is responsible (Article 8, 85/374/CEE) HRI-R40W Instructions manual IM831-U v0.92 pag. 12 / 12

INSTRUCTIONS MANUAL DEVICE FOR THE PERMANENT CONTROL OF INSULATION ON THE SUPPLY LINES IN MEDICAL PRACTICE PLACES INSTALLATIONS HRI-R40W

INSTRUCTIONS MANUAL DEVICE FOR THE PERMANENT CONTROL OF INSULATION ON THE SUPPLY LINES IN MEDICAL PRACTICE PLACES INSTALLATIONS HRI-R40W Page 2 / 16 INDEX: GENERAL TYPES ACESSORIES AND OPTIONS INSTALLATION WIRING DIAGRAMS CONNECTING TERMINALS DIAGRAMS WIRING DIAGRAM DESCRIPTION WORKING DESCRIPTION AND KEYBOARD WORKING DESCRIPTION AND PROGRAMMING

More information

BU: EPBP GPG: DIN Rail Products Devices for the permanent control of insulation on supply lines for medical locations ISOLTESTER-DIG-RZ/RS/PLUS

BU: EPBP GPG: DIN Rail Products Devices for the permanent control of insulation on supply lines for medical locations ISOLTESTER-DIG-RZ/RS/PLUS INSTRUCTION MANUAL BU: EPBP GPG: DIN Rail Products Devices for the permanent control of insulation on supply lines for medical locations ISOLTESTER-DIG-RZ/RS/PLUS 1/23 More information than that reported

More information

DMTME-96 2CSG133030R4022 M DMTME-I Operation and assembly instructions 2CSG163030R4022 M CSG445001D0202

DMTME-96 2CSG133030R4022 M DMTME-I Operation and assembly instructions 2CSG163030R4022 M CSG445001D0202 DMTME-96 2CSG133030R4022 M204675 DMTME-I-485-96 2CSG163030R4022 M204685 2CSG445001D0202 GB Operation and assembly instructions DMTME-96: three-phase multimeter for panel mounting, used also in singlephase

More information

RI-F200 Series. Single and Three Phase Multifunction Energy Meter. Telephone : +44 (0) Displayed Parameters

RI-F200 Series. Single and Three Phase Multifunction Energy Meter. Telephone : +44 (0) Displayed Parameters RI-F200 Series Single and Three Phase Multifunction Energy Meter DIN 96 panel mounted -/1A or -/5A current transformer input Single phase or three phase network compatible Programmable voltage and current

More information

PRELIMINARY DATA SHEET RI-D140. Three Phase Multifunction DIN Rail Energy Meter (MID Certified) MID

PRELIMINARY DATA SHEET RI-D140. Three Phase Multifunction DIN Rail Energy Meter (MID Certified) MID RI-D140 Three Phase Multifunction DIN Rail Energy Meter (MID Certified) Four module DIN rail mounted Energy pulse LED True RMS measurement Cost effective and accurate Modbus communication -/1A or -/5A

More information

Analogue temperature controllers

Analogue temperature controllers Analogue temperature controllers CT8A Input by J-K thermo-couple or by thermo-resistance Pt 00 (-wire) regulation modes : ON/OFF or proportional derivative selected by wiring Relay output Specifications

More information

Earth Leakage Monitoring System IsoBase IsoHub IsoOut

Earth Leakage Monitoring System IsoBase IsoHub IsoOut 2014 11 21 Earth Leakage Monitoring System IsoBase IsoHub IsoOut User Manual Due to our policy of continual improvement, specifications may change without prior notice Page 2 (34) Contents Earth Leakage

More information

ISOMETER IRDH575. Approvals

ISOMETER IRDH575. Approvals Insulation monitoring device for unearthed AC, DC and AC/DC systems (IT systems) with control and display function for EDS insulation fault location systems IRDH575_D00089_02_D_XXEN/09.2018 Insulation

More information

SPAE 010, 011 High Impedance Protection Relay

SPAE 010, 011 High Impedance Protection Relay SPAE 010, 011 High Impedance Protection Relay User s manual and Technical description f n = 50/60 Hz U n = 50 / 100 / 200 V 2 5 U REF > SPAE 010 0.8 U aux 0.6 1.0 RESET OK x 80... 265 V ~ _ 0.4 U > U n

More information

B850 Boiler House Energy Monitor

B850 Boiler House Energy Monitor Local regulations may restrict the use of this product to below the conditions quoted. In the interests of development and improvement of the product, we reserve the right to change the specification without

More information

PROMUX Distributed MODBUS I/O Modules Catalog and Design Guide

PROMUX Distributed MODBUS I/O Modules Catalog and Design Guide PROMUX Distributed MODBUS I/O Modules Catalog and Design Guide 12/03/2012 V11.0 P.O.Box 164 Seven Hills 1730 NSW AUSTRALIA Tel: +61 2 96248376 Fax: +61 2 9620 8709 Email: proconel@proconel.com Web: www.proconel.com

More information

Multifuncion Measuring Device, 4 mod. DIN, connection via CT, with RS485 port

Multifuncion Measuring Device, 4 mod. DIN, connection via CT, with RS485 port Viale Borri 231, 21100 Varese Italy Multifuncion Measuring Device, 4 mod. Contents Pages 1. Description Use... 2 2. Range... 2 3. Overall dimensions... 2 4. Preparation Connection... 2 5. General characteristics...

More information

Centrale de mesure Power Meter PM500 Merlin Gerin

Centrale de mesure Power Meter PM500 Merlin Gerin Notice d'installation et d'utilisation Installation and user manual Centrale de mesure Power Meter PM500 Merlin Gerin 059473_D Introduction and description Package contents c one PM500 power meter with

More information

Electronic relays. Monitoring. Monitoring

Electronic relays. Monitoring. Monitoring Electronic relays Neither overload nor underload are particularly good for machines and installations in industrial applications. An unsteady power supply may lead to machine failure and, thus, expensive

More information

Data Acquisition Modules/ Distributed IO Modules

Data Acquisition Modules/ Distributed IO Modules User Manual Data Acquisition Modules/ Distributed IO Modules Future Design Controls, Inc. 7524 West 98 th Place / P.O. Box 1196 Bridgeview, IL 60455 888.751.5444 - Office: 888.307.8014 - Fax 866.342.5332

More information

Modbus communication module for TCX2: AEX-MOD

Modbus communication module for TCX2: AEX-MOD Modbus communication module for TCX2: Communication Specification TCX2 is factory installed in TCX2 series controllers with -MOD suffix, and is also available separately upon request for customer installation

More information

Energy Management Three-phase energy meter with output modules Type EM4-DIN

Energy Management Three-phase energy meter with output modules Type EM4-DIN Energy Management Three-phase energy meter with output modules Type EM4-DIN Class 1 (active energy) Class 2 (reactive energy) Three-phase multi-function energy meter Back-lighted LCD display 3 1 / 2 DGT

More information

RI-D440. Three Phase easywire Multifunction DIN Rail Energy Meter. Telephone : +44 (0) Displayed Parameters

RI-D440. Three Phase easywire Multifunction DIN Rail Energy Meter. Telephone : +44 (0) Displayed Parameters RI-D440 Three Phase easywire Multifunction DIN Rail Energy Meter Four module DIN rail mounted Energy pulse LED 330mV Input from easywire CTs (or -/1A and -/5A current transformer input with TAS-SCTEWA

More information

Technical Datasheet. True RMS Digital Protection Relay. Line Monitoring Relay

Technical Datasheet. True RMS Digital Protection Relay. Line Monitoring Relay Technical Datasheet RELAY-1 RELAY-2 k V Hz RESET ABH TEST True RMS Measurement True RMS Digital Protection V/Hz is used to protect against Over Voltage, Under Voltage, Phase Unbalance, Phase Sequence detection,

More information

Energy Management Energy Meter with plug-in Output Modules Type EM4-DIN

Energy Management Energy Meter with plug-in Output Modules Type EM4-DIN Energy Management Energy Meter with plug-in Output Modules Type EM4-DIN Degree of protection (front): IP 40 Front dimensions: 9 DIN modules RS 422/485 Serial port by means of optional module Dual pulse

More information

METERSUK. SmartLink D Y N A M i T E. EM415 MOD Series. User Manual KEY FEATURES INDEX 3. TECHNICAL DESCRIPTION. Single Phase Metering

METERSUK. SmartLink D Y N A M i T E. EM415 MOD Series. User Manual KEY FEATURES INDEX 3. TECHNICAL DESCRIPTION. Single Phase Metering METERSUK LTD s a l e s @ m e t e r s. c o. u k 0 1 5 2 4 5 5 5 9 2 9 Series Smart Energy Meter KEY FEATURES Single Phase Metering Single phase metering 4 din modules, Standard DIN rail Format (DIN43880)

More information

Technical Datasheet. True RMS Digital Protection Relay. Voltage Protection Relay

Technical Datasheet. True RMS Digital Protection Relay. Voltage Protection Relay Technical Datasheet RELAY-1 RISHABH k V RESET True RMS Digital Protection Relay RELAY-2 TEST True RMS Measurement RISH Relay V is used to protect against Over Voltage, Under Voltage, Phase Sequence detection,

More information

Temperature Controller model MFC-301/T-Dry. Version for Dry Transformers and Motors. Technical Manual. Licht

Temperature Controller model MFC-301/T-Dry. Version for Dry Transformers and Motors. Technical Manual. Licht Temperature Controller model MFC-301/T-Dry Version for Dry Transformers and Motors Technical Manual Licht Contents 1 Introduction 2 2 Operating principle 3 2.1 General principle 3 2.2 RTD operation 3 3

More information

VFSC9 ELECTRONIC SPEED CONTROLLER. Mounting and operating instructions

VFSC9 ELECTRONIC SPEED CONTROLLER. Mounting and operating instructions ELECTRONIC SPEED CONTROLLER Mounting and operating instructions Table of contents SAFETY AND PRECAUTIONS 3 PRODUCT DESCRIPTION 4 ARTICLE CODES 4 INTENDED AREA OF USE 4 TECHNICAL DATA 4 STANDARDS 5 WIRING

More information

ALPTEC POWER FACTOR CONTROLLER

ALPTEC POWER FACTOR CONTROLLER ALPTEC POWER FACTOR CONTROLLER ALPTEC3 ALPTEC5 ALPTEC7 ALPTEC12 0 REF : 2008-ALPTEC3.5.7.12-01-ANG CONTENTS I GENERAL INFORMATION. 2 II WAYS TO SET UP THE CONTROLLER.. 4 III OPERATING MODE..6 IV ADVANCED

More information

ioselect Z-NET Z-4RTD 4 Channel RTD Input Module

ioselect Z-NET Z-4RTD 4 Channel RTD Input Module 4 Channel nput Module : PT00, PT000, PT500 & N00: 2, 3 & 4 wire 500 V solation, 6-way Modbus Communications Adjustable Sampling Time Accuracy to 0.2 C Configuration and Setup Software EC 63 Flexible Power

More information

T/3000 T/3000. Substation Maintenance and Commissioning Test Equipment

T/3000 T/3000. Substation Maintenance and Commissioning Test Equipment T/3000 Substation Maintenance and Commissioning Test Equipment MULTI FUNCTION SYSTEM FOR TESTING SUBSTATION EQUIPMENT SUCH AS: CURRENT, VOLTAGE AND POWER TRANSFORMERS, ALL TYPE OF PROTECTION RELAYS, ENERGY

More information

Relay. Technical Datasheet. True RMS Digital Protection Relay. Line Monitoring Relay

Relay. Technical Datasheet. True RMS Digital Protection Relay. Line Monitoring Relay Line Monitoring Technical Datasheet RISH RELAY-1 RELAY-2 RISHABH k V Hz True RMS Digital Protection RISH V/Hz is used to protect against Over Voltage, Under Voltage, Phase Unbalance, Phase Sequence detection,

More information

Fault indicator. Application. SPEF 3A2 C 1MRS MBG Issued: April 1999 Status: Updated Version: B/ Data subject to change without notice

Fault indicator. Application. SPEF 3A2 C 1MRS MBG Issued: April 1999 Status: Updated Version: B/ Data subject to change without notice Issued: April 1999 Status: Updated Version: B/08.11.200 Data subject to change without notice Features Versatile, multifunction line fault indicator for distribution networks Overcurrent, earth-fault and

More information

Position Indicator model MFC-300/IP. Technical Manual. Licht

Position Indicator model MFC-300/IP. Technical Manual. Licht Position Indicator model MFC-300/IP Technical Manual Licht Contents 1 Introduction 2 2 Front panel indication 3 3 Error indication 4 4 Manual commands 5 5 Configuration 6 5.1 Parameter reset 6 6 Programmable

More information

Multimeter 500CVD21 RTU500 series

Multimeter 500CVD21 RTU500 series Remote Terminal Units - Data sheet Multimeter 500CVD21 RTU500 series CT/VT interface with 4 voltage and 24 current inputs for direct monitoring of 3/4 wire 0 300 V AC (line to earth), 0...500 V AC (phase

More information

10/2 Product overview. 10/3 4AC3 0, 4AC3 1 bell transformers. 10/5 4AC AC3 6 transformers for permanent loads. 10/8 4AC2 4 power supply units

10/2 Product overview. 10/3 4AC3 0, 4AC3 1 bell transformers. 10/5 4AC AC3 6 transformers for permanent loads. 10/8 4AC2 4 power supply units BETA Switching Transformers, Bells and Socket Outlets /2 Product overview /3 4AC3 0, 4AC3 1 bell transformers /5 4AC3 4... 4AC3 transformers for permanent loads /8 4AC2 4 power supply units / 7LQ2 2 bells

More information

Data Sheet. RISH Master Record %THD

Data Sheet. RISH Master Record %THD Data Sheet %THD Application : measures important electrical parameters & replaces the multiple analog panel meters. It measures electrical parameters like AC current, Voltage, frequency, active energy

More information

269 Troubleshooting Guide 11/4/98 1

269 Troubleshooting Guide 11/4/98 1 269 Troubleshooting Guide 11/4/98 1 Table of Contents Phase currents 3 Communications. 6 RTDs.. 9 Ground Fault Currents... 13 Output Relays. 15 Analog Output 17 Switch Inputs...18 11/4/98 2 Phase Currents

More information

Digital Multifunction Instrument - Rish Master 3440

Digital Multifunction Instrument - Rish Master 3440 Application Rish Master 3440 measures important electrical parameters in 3 phase and single phase etwork & replaces the multiple analog panel meters. It measures electrical parameters like AC current,

More information

NetBiter I/O Extender 4RO 6RTD 8DIO - DAIO User Manual Revision 1.00

NetBiter I/O Extender 4RO 6RTD 8DIO - DAIO User Manual Revision 1.00 NetBiter I/O Extender 4RO 6RTD 8DIO DAIO User Manual Revision 1.00 IntelliCom Innovation AB Linjegatan 3D SE302 50 Halmstad SWEDEN Phone +46 35 18 21 70 Fax +46 35 18 21 99 email info@intellicom.se www

More information

PROMUX Distributed MODBUS I/O Modules Catalog and Design Guide

PROMUX Distributed MODBUS I/O Modules Catalog and Design Guide PROMUX Distributed MODBUS I/O Modules Catalog and Design Guide 14/11/2006 V10 P.O.Box 24 Stanfield 3613 SOUTH AFRICA Tel: +27 (031) 7028033 Fax: +27 (031) 7028041 Email: proconel@proconel.com Web: www.proconel.com

More information

XR1 Rotor Earth Fault Relay. Manual XR1 (Revision C)

XR1 Rotor Earth Fault Relay. Manual XR1 (Revision C) XR1 Rotor Earth Fault Relay Manual XR1 (Revision C) Woodward Manual XR1 (EN) Woodward Governor Company reserves the right to update any portion of this publication at any time. Information provided by

More information

SUPPLY NETWORK ANALYZER CVM-96 SERIES

SUPPLY NETWORK ANALYZER CVM-96 SERIES SUPPLY NETWORK ANALYZER CVM-96 SERIES (Power Demand) INSTRUCTION MANUAL ( M 981 326 / 00B - GB) (c) CIRCUTOR S.A. ----- Supply network analyzer CVM-96 ------ User's manual --- Page No. 1 CVM-96 SUPPLY

More information

Monitoring Electric Network Via Internet

Monitoring Electric Network Via Internet Monitoring Electric Network Via Internet NETWORK ANALYSER let "it" analyse! MPR63 Real time monitoring via internet / intranet Measurement of all electrical parameters Including Harmonics Automatic recording

More information

3-phase meter with direct connection and pulse or Modbus RS485 output

3-phase meter with direct connection and pulse or Modbus RS485 output 87045 LIMOGES Cedex Telephone: +33 (0)5 55 06 87 87 Fax: +33 (0)5 55 06 88 88 3-phase meter with direct connection and CONTENTS PAGES 1. Description, usage... 1 2. Range... 1 3. Dimensions... 1 4. Positioning...

More information

Protection and safety SQZ3 phase and sequence relay

Protection and safety SQZ3 phase and sequence relay SQZ3 phase and sequence relay Supply voltage [Vn] 400 V a.c. Frequency [Hz] 0/60 SQZ3 2CSC4001F0201 Contact type [A] 1 CO, 20 V, 10 A (cos =1) safety switching Minimum voltage adjustment trimmer [%] 100

More information

PM-311x Quick Start Ver PM-311x introduction 1.1. Caution & Warning 1.2. Product Warranty & Customer Support

PM-311x Quick Start Ver PM-311x introduction 1.1. Caution & Warning 1.2. Product Warranty & Customer Support PM-311x Quick Start Ver. 1.1 1. PM-311x introduction ICP DAS brings the most powerful, cost-effective, advanced Smart Power Meters PM-3000 series that gives you access to real-time electric usage for single-phase

More information

Technical Data Sheet

Technical Data Sheet Technical Data Sheet parameters in 3 phase 4 Wire and 3 phase 3 Wire Network & replaces the multiple analog panel meters. Special Features Pulse/Limit Switch output (optional) Number of interruption Application

More information

B850 Boiler House Energy Monitor

B850 Boiler House Energy Monitor Local regulations may restrict the use of this product to below the conditions quoted. In the interests of development and improvement of the product, we reserve the right to change the specification without

More information

Energy Management Energy Meter Type EM111

Energy Management Energy Meter Type EM111 Energy Management Energy Meter Type EM111 Easy connection or wrong current direction detection Other versions available (not certified, option X: see how to order on the next page Single phase energy meter

More information

Energy Management Energy Meter Type EM112

Energy Management Energy Meter Type EM112 Energy Management Energy Meter Type EM112 Easy connection or wrong current direction detection Other versions available (not certified, option X: see how to order on the next page Single phase energy meter

More information

XR1 Rotor Earth Fault Relay. (May 2007) Manual XR1 (Revision New)

XR1 Rotor Earth Fault Relay. (May 2007) Manual XR1 (Revision New) XR1 Rotor Earth Fault Relay (May 2007) Manual XR1 (Revision New) Woodward Manual XR1 GB Woodward Governor Company reserves the right to update any portion of this publication at any time. Information provided

More information

Technical Data Sheet. x1000. kwh R Y B

Technical Data Sheet. x1000. kwh R Y B Technical Data Sheet x1000 Von kwh IRev 3600 imp /kwh R Y B is a compact multifunction instrument which is a 96mm x 96mm panel mounted kilowatt hour meter it measures active energy with class 1.0 accuracy

More information

Single Channel Loop Detector

Single Channel Loop Detector Single Channel Loop Detector Model LD120T Series The LD120T is a series of single channel inductive loop detectors. The use of microprocessor and surface mount technology enables a large number of functions

More information

PROCESS & TEMPERATURE CONTROLLERS

PROCESS & TEMPERATURE CONTROLLERS PROCESS & TEMPERATURE CONTROLLERS NOVA PD54 Series Thermocouple, RTD, & Process Inputs High Accuracy Auto-Tuning PID Heating & Cooling Models Universal Power Supply 1-24 VAC Up to 3 Relays & 2 Analog Outputs

More information

XP2-R Power and Reverse Power Relay

XP2-R Power and Reverse Power Relay XP2-R Power and Reverse Power Relay Manual XP2-R (Revision C) Woodward Manual XP2-RE Woodward reserves the right to update any portion of this publication at any time. Information provided by Woodward

More information

EARTH LEAKAGE RELAY RELAY FOR PERMANENT CONTROL OF THE MCCB S TRIPPING CIRCUIT AND ACTUATOR FOR SAFETY CIRCUITS

EARTH LEAKAGE RELAY RELAY FOR PERMANENT CONTROL OF THE MCCB S TRIPPING CIRCUIT AND ACTUATOR FOR SAFETY CIRCUITS April 009 RELAY FOR PERMANENT CONTROL OF THE MCCB S PING CIRCUIT AND ACTUATOR FOR SAFETY CIRCUITS STATIC RELAY FOR MOTOR RE-START AND REACCELERATION www. microener.com ELR Multi-range Earth Leakage Relays

More information

MT 560/UMT TRANSDUCER & ANALYZER

MT 560/UMT TRANSDUCER & ANALYZER MT 560/UMT 560 - TRANSDUCER & ANALYZER USE The MT 560/UMT 560 multi transducer and analyzer is used for a permanent analysis of electricity supply quality in compliance with the SIST EN 50160 standard.

More information

ALPHA 50 MULTIFUNCTION METER (ALPHA SERIES)

ALPHA 50 MULTIFUNCTION METER (ALPHA SERIES) Alpha Series www.sifamtinsley.co.uk Multifunction Meters Transducers & Isolators Temperature Controllers Converters & Recorders Digital Panel Meters Current Transformers Analogue Panel Meters Shunts ALPHA

More information

DPF-MAC Auto power factor controller

DPF-MAC Auto power factor controller Digital Electric& Electronics System DPF-MAC Auto power factor controller MULTI POWER FACTOR AUTOMATIC MULTI POWER FACTOR AUTOMATIC Introduction DPF-MAC increases usage efficiency by controlling the power

More information

ENA33LCD. Power line analyzer. User and service manual. Obrezija 5 SI-1411 Izlake

ENA33LCD. Power line analyzer. User and service manual.   Obrezija 5 SI-1411 Izlake ENA33LCD Power line analyzer User and service manual version 2.9 (FW version 6.8 and newer) ETI, d.o.o. Obrezija 5 SI-1411 Izlake www.etigroup.eu/products-services 1. Front control panel and terminal plate

More information

Selection guide. Communication accessories. Communication accessory selection guide. Characteristics

Selection guide. Communication accessories. Communication accessory selection guide. Characteristics Communication accessories Selection guide There are 2 types of Sepam communication accessories: b communication interfaces, which are essential for connecting Sepam to the communication network b converters

More information

Single phase meter with direct connection and pulse or Modbus RS485 output

Single phase meter with direct connection and pulse or Modbus RS485 output 87045 LIMOGES Cedex Telephone: +33(0)5 55 06 87 87 Fax: +33(0)5 55 06 88 88 Single phase meter with direct connection CONTENTS PAGES 1. Description, usage... 1 2. Range... 1 3. Dimensions... 1 4. Positioning...

More information

2.7. IT system floor-standing distribution cabinet Series -IPS-F/EDS

2.7. IT system floor-standing distribution cabinet Series -IPS-F/EDS Dipl.-Ing. W. Bender GmbH & Co. KG Londorfer Str. 65 35305 Grünberg Tel.: 0640 807-0 Fax: 0640 807-259 IT system floor-standing distribution cabinet Series -IPS-F/EDS for supplying power to medical locations

More information

ZHUHAI PILOT TECHNOLOGY CO., LTD.

ZHUHAI PILOT TECHNOLOGY CO., LTD. SPM32 Multifunctional Power Meter SPM32 Multifunction Power Meter Installation & Operation Manual V1.0 ZHUHAI PILOT TECHNOLOGY CO., LTD. Danger and warning! This device can be installed only by professionals.

More information

XP2-R Power and reverse power relay. (January 2006) Manual XP2-R (Revision New)

XP2-R Power and reverse power relay. (January 2006) Manual XP2-R (Revision New) XP2-R Power and reverse power relay (January 2006) Manual XP2-R (Revision New) Woodward Manual XP2-R GB Woodward Governor Company reserves the right to update any portion of this publication at any time.

More information

RISH EM 3490 SS Kilowatt Hour Energy Meter With Rs485 RISH EM 3490 SS. Application : Product Features: Indication: Pulse Indication:

RISH EM 3490 SS Kilowatt Hour Energy Meter With Rs485 RISH EM 3490 SS. Application : Product Features: Indication: Pulse Indication: Application : RISH Master 3490 SS is a 96mm x 96mm panel mounted kilowatt hour meter it measures active energy with class 1.0 accuracy having auto-resetting 8 digit seven segment LED counter. The unit

More information

High-set undervoltage stage with definitetime. or inverse definite minimum time (IDMT) characteristic. Low-set undervoltage stage with definitetime

High-set undervoltage stage with definitetime. or inverse definite minimum time (IDMT) characteristic. Low-set undervoltage stage with definitetime Issued: 5.06.999 Status: 5.06.999 Version: B/09..00 Data subject to change without notice Features Overvoltage and undervoltage protection Single- or three-phase operation High-set overvoltage stage with

More information

UNIVERSAL MEASURING INSTRUMENTS. TNM 34xx. TNM 3410 / 3420 / 3430 / 3440 Universal measuring instrument 1/49

UNIVERSAL MEASURING INSTRUMENTS. TNM 34xx. TNM 3410 / 3420 / 3430 / 3440 Universal measuring instrument 1/49 TNM 3410 / 3420 / 3430 / 3440 Universal measuring instrument True RMS measurement User-friendly programing Four different types for optimal specification RS485 interface For 3- or 4-wire connection Compact

More information

Current monitor Type C 551

Current monitor Type C 551 Current monitor Type C 551 Relay for current monitoring, single-phase Monitoring AC and DC for overcurrent or undercurrent 1 yellow LED for indication of the relay state 1 green LED for indication of the

More information

CROSS Chassis from 160 A to 450 A

CROSS Chassis from 160 A to 450 A CROSS Chassis from 160 A to 450 A STS CATALOGUE Important note! The technical data enclosed is for general information. Please note that the operating instructions and references indicated on the products

More information

INTELLIGENT DIGITAL MULTI POWER METER [DIC-MPM]

INTELLIGENT DIGITAL MULTI POWER METER [DIC-MPM] Feature Suit for LV/ HV voltage system Compact design with all real-time measurement Consist of basic unit and optional modules 10 years back-up of integrated energy data Large LCD, high brightness High

More information

PLA 33. Power line analyzer. User and service manual. version 2.4

PLA 33. Power line analyzer. User and service manual. version 2.4 PLA 33 Power line analyzer User and service manual version 2.4 Content. Front control panel and terminal plate...3 7.2.2. System frequency setting...0 2. Device description...4 7.2.3. Password protection...0

More information

WIRELESS MODBUS GATEWAY WGW410

WIRELESS MODBUS GATEWAY WGW410 WIRELESS MODBUS GATEWAY WGW410 The Tekon Wireless Modbus Gateway WGW410 is specifically designed to meet the most rigorous requirements of operation in the industrial process environments. Due to their

More information

FCR 06, FCR 12. Power factor correction controller. User and service manual

FCR 06, FCR 12. Power factor correction controller. User and service manual FCR 06, FCR 12 Power factor correction controller User and service manual version 2.3 Czech Republic Czech Republic 1 Content 1. Control and signal elements... 3 2. Device description... 4 3. Instruction

More information

Contents. Introduction and description Package contents Device identification PM Options... 80

Contents. Introduction and description Package contents Device identification PM Options... 80 Contents 1 Introduction and description Package contents... 77 Device identification... 77 2 Characteristics PM500... 78 Options... 80 3 Installation Front-panel cut-out... 82 Mounting... 82 4 Connections

More information

These parameters are available only on EMC-D3b-485 model and readable only by RS485 interface.

These parameters are available only on EMC-D3b-485 model and readable only by RS485 interface. ISTRUCTIO MAUAL IM450-U v3.1 -D3b -D3b-485 EERGY MULTIFUCTIO COUTER AVAILABLE MODELS -D3b: 3-phase energy counter with 7+1 digit display. 1 digital input and two digital pulse output for kwhr and kvarhr.

More information

1 INTRODUCTION ORDER CODE / INFORMATION

1 INTRODUCTION ORDER CODE / INFORMATION INTRODUCTION ORDER CODE / INFORMATION 269/269Plus * * * * * * 269/269Plus SV D/O.4 ORDER CODE / INFORMATION Motor management relay Standard version Drawout version Phase CT Ground CT (required for D/O

More information

Protection relays and transformers

Protection relays and transformers NEW ZEALAND WWW.LPINZ.CO.NZ Unit 6/22 Moselle Ave, Henderson, Auckland, New Zealand PO Box 21-872, Henderson, Auckland, New Zealand Phone:+64 9 833 5749 Email:info@LPINZ.co.nz Web: www.lpinz.co.nz Protection

More information

RISH EM 3490 DS Dual Source Energy Meter (With All Display Parameters) RISH EM 3490 DS. Application : Product Features:

RISH EM 3490 DS Dual Source Energy Meter (With All Display Parameters) RISH EM 3490 DS. Application : Product Features: (With All Display Parameters) Application : RISH EM 3490 DS measures important electrical parameters of Utility (in normal mode) & Generators (in Power back up) in three phase and single phase Network

More information

Power Meter Series PM3200 Functions and characteristics

Power Meter Series PM3200 Functions and characteristics Functions and characteristics PB108433 This PowerLogic Power meter offers basic to advanced measurement capabilities. With compact size and DIN rail mounting, the PM3200 allows mains and feeders monitoring

More information

AC/DC. Monitoring Technique. VARIMETER IMD Insulation monitor RN 5897/300

AC/DC. Monitoring Technique. VARIMETER IMD Insulation monitor RN 5897/300 Monitoring Technique VARIMETER IMD Insulation monitor RN 5897/300 0274214 Your Advantages For mobile generator sets according to DIN VDE 0100-551 Preventive fire and system protection Detection of symmetric

More information

Theta 20A / 20V. Technical Data Sheet. Special Features

Theta 20A / 20V. Technical Data Sheet. Special Features Technical Data Sheet Theta 20A / 20V Theta 20A / 20V is used to measure and convert AC Voltage or Current input into a load independent DC current or voltage output signal. Special Features Available in

More information

Energy Management Energy Analyzer Type EM112

Energy Management Energy Analyzer Type EM112 Energy Management Energy Analyzer Type EM112 Single phase energy analyzer Class 1 (kwh according to EN62053-21 Class B (kwh according to EN50470-3 Accuracy ±0.5% RDG (current/voltage Direct current measurement

More information

DIN rail mounting temperature controller with current transformer input deltadue series D1 line

DIN rail mounting temperature controller with current transformer input deltadue series D1 line BT.. D/E S E R I E S DIN rail mounting temperature controller with current transformer input deltadue series D line The controller with load control The deltadue series includes a powerful DIN rail mounting

More information

PowerLogic power-monitoring units. PM700 series power meter. Technical data sheet

PowerLogic power-monitoring units. PM700 series power meter. Technical data sheet PowerLogic power-monitoring units PM700 series power meter Technical data sheet 2009 Functions and characteristics The PowerLogic PM700 series power meter offers all the measurement capabilities required

More information

- SMART TUNE- PID CONTROL - UNIVERSAL, 3 WIRE- TC, RTD AND LINEAR INPUT - AUX- REMOTE SET POINT/ TRIM INPUT - 2x ISOLATED CONTROL AND RETRANSMISSION

- SMART TUNE- PID CONTROL - UNIVERSAL, 3 WIRE- TC, RTD AND LINEAR INPUT - AUX- REMOTE SET POINT/ TRIM INPUT - 2x ISOLATED CONTROL AND RETRANSMISSION ADVANCED - CONTROLLERS - SMART TUNE- PID CONTROL - UNIVERSAL, 3 WIRE- TC, RTD AND INPUT - AUX- REMOTE SET POINT/ TRIM INPUT - 2x ISOLATED CONTROL AND RETRANSMISSION (ma) OUTPUTS - 4x EVENT OR ALARM OUTPUTS

More information

PD300. Transformer, generator and motor protection Data sheet

PD300. Transformer, generator and motor protection Data sheet PD300 Transformer, generator and motor protection Data sheet DSE_PD300_eng_AO No part of this publication may be reproduced by whatever means without the prior written permission of Ingeteam T&D. One of

More information

Digital Panel Meters DC/AC Current and Voltage Indicator/Controller Type LDM40

Digital Panel Meters DC/AC Current and Voltage Indicator/Controller Type LDM40 Digital Panel Meters Current and Voltage Indicator/Controller Type LDM40 Multi-input instrument 4-DGT LED 0.1% RDG basic accuracy TRMS AC current and voltage measurements AC/DC current measurements: selectable

More information

Model Number Structure

Model Number Structure Current Sensor CSM DS_E Solid-state, Plug-in Current Sensor Applicable to motor overcurrent protection and 3-phase AC current detection. Inverse-type, start-up lock type, and instantaneous type overcurrent

More information

Type Nominal voltage (V) Code HSV V

Type Nominal voltage (V) Code HSV V Speed control Speed control relay - mm Control of overspeed, underspeed, operating rate, stopping Measurement via discrete sensors - -wire PNP or NPN, Namur, voltage 0-0V or volt-free contact type Works

More information

Transducing electrical quantities Iskra MIS d.d. Measuring Transducers. Features

Transducing electrical quantities Iskra MIS d.d. Measuring Transducers. Features Measuring Transducers Transducing electrical quantities Iskra MIS d.d. Measuring Transducers Features AC, DC POWER SYSTEM MEASUREMENT MICROPROCESSOR CONTROL HOUSING FOR DIN RAIL MOUNTING IN COMPLIANCE

More information

ADC5000 SERIES. AC/DC Switch Mode Power Supplies and Rectifiers for Industrial and Telecom Applications. 60W, 125W and 250 W

ADC5000 SERIES. AC/DC Switch Mode Power Supplies and Rectifiers for Industrial and Telecom Applications. 60W, 125W and 250 W ADC5000 SERIES AC/DC Switch Mode Power Supplies and Rectifiers for Industrial and Telecom Applications 60W, 125W and 250 W Input voltage 230/115 VAC voltages 12, 24, 36 or 48 VDC Statistical MTBF >3 000

More information

Measuring and monitoring relays CM-SRS.1 Current monitoring relays, single-phase AC/DC

Measuring and monitoring relays CM-SRS.1 Current monitoring relays, single-phase AC/DC 2CDC 251 244 F0t05 Characteristics Monitoring of DC and AC currents: RMS measuring principle One device includes 3 measuring ranges Over- or undercurrent monitoring configurable Hysteresis adjustable from

More information

Measuring Transducer MT4xx series Programmable AC Current Transducer MT418

Measuring Transducer MT4xx series Programmable AC Current Transducer MT418 Measuring Transducer MT4xx series Programmable AC Current Transducer MT418 o RMS AC current or frequency measurements o Current auto range measurements up to 12 A o Frequency measurement range 16 400 Hz

More information

Multifuncion Measuring Device, 4 mod. DIN, connection via CT, with RS485 port

Multifuncion Measuring Device, 4 mod. DIN, connection via CT, with RS485 port 87045 LIMOGES Cedex Telephone : 05 55 06 87 87 Fax: 05 55 06 88 88 Multifuncion Measuring Device, 4 mod. Contents Pages 1. Description Use... 1 2. Range... 1 3. Overall dimensions... 1 4. Preparation Connection...

More information

Temperature monitoring relays CM-TCS Monitoring relays for monitoring temperatures with a PT100 sensor (2- or 3-wire connection)

Temperature monitoring relays CM-TCS Monitoring relays for monitoring temperatures with a PT100 sensor (2- or 3-wire connection) Data sheet Temperature monitoring relays CM-TCS Monitoring relays for monitoring temperatures with a PT100 sensor (2- or 3-wire connection) The temperature monitoring relays CM-TCS monitor overtemperature,

More information

VOLTAGE DETECTING SYSTEMS

VOLTAGE DETECTING SYSTEMS INTEGRATED OPTOELECTRIC SYSTEM OF VOLTAGE DETECTION WITH TOTAL SAFETY GUARANTEED BY GALVANIC INSULATION AND INTERNAL FAULT SUPERVISION pag.1 di 13 INDE 1.PREVIEW.. 3 2.GENERAL.. 3 3.STANDARD REFERENCES..

More information

Thermistor motor protection relays

Thermistor motor protection relays Thermistor motor protection relays Content Benefits and advantages... / 68 Selection table... / 68 Ordering details CM-MSE... / 69 CM-MSS... / 69 CM-MSN... / 7 PTC sensor C0... / 7 Technical data... /

More information

Energy Management Energy Analyzer Type EM111

Energy Management Energy Analyzer Type EM111 Energy Management Energy Analyzer Type EM111 Single phase energy analyzer Class 1 (kwh according to EN62053-21 Class B (kwh according to EN50470-3 Accuracy ±0.5% RDG (current/voltage Direct current measurement

More information

Dipl.-Ing. W. Bender GmbH & Co. KG Londorfer Str Grünberg Tel.: Fax:

Dipl.-Ing. W. Bender GmbH & Co. KG Londorfer Str Grünberg Tel.: Fax: Dipl.-Ing. W. Bender GmbH & Co. KG Londorfer Str. 65 35305 Grünberg Tel.: 06401 807-0 Fax: 06401 807-259 Insulation fault locator Insulation fault locator for DC IT systems with high system leakage capacitances

More information

UMG 103. UMG 103 Universal measurement device for DIN rails. GridVis Analysis software. Harmonics. Measurement accuracy 0.5.

UMG 103. UMG 103 Universal measurement device for DIN rails. GridVis Analysis software. Harmonics. Measurement accuracy 0.5. Harmonics GridVis Analysis software Modbus interface Measurement accuracy 0.5 Universal measurement device for DIN rails Communication Protocols: Modbus RTU / Slave Interface RS485 Power quality Harmonics

More information

R&D TECHNICAL SPECIFI

R&D TECHNICAL SPECIFI INTRODUCTION Present document defines technical specs of a three phase UPS covering the power range from 20 to 30 kva ON-LINE DOUBLE CONVERSION Uninterruptible Power Supply Systems with forced ventilation

More information

Application Alpha 20 Measures important electrical parameters in 3 phase 4 Wire and 3 phase 3 Wire Network & replaces the multiple analog panel meters

Application Alpha 20 Measures important electrical parameters in 3 phase 4 Wire and 3 phase 3 Wire Network & replaces the multiple analog panel meters Technical Data Sheet Alpha 20 Alpha 20 is a compact multifunction instrument which Measures important electrical parameters in 3 phase 4 Wire and 3 phase 3 Wire Network & replaces the multiple analog panel

More information