ELECTRICAL CURRENT FOR TISSUE HELALING
|
|
- Ethan Montgomery
- 6 years ago
- Views:
Transcription
1 ELECTRICAL CURRENT FOR TISSUE HELALING
2 Tissue healing 2 Chronic wound healing (ex. pressure ulcer) faster! Mechanism - attraction of appropriate cell type of area - activation of theses cells by altering cell membrane function - modification of endogenous electrical potential of the tissue in concert with healing potentials. - reduction of edema - enhancement of antimicrobial activity, and promotion of circulation Galvanotaxis ( 전기주성 ) - attraction to the negative pole : activated neutriphil, lymphocyte, platelet, mast cell, keratinocyte, fibroblast negative electrode : inflamed or infected wound healing 을촉진 - attraction to the positive pole: inactive neutriphil, macrophage, epidermal cell positive electrode : to promote healing of wound with no inflammation
3 Tissue healing 3 To enhance fibroblast replication and Increase the synthesis of DNA and collagen by fibroblast (voltage dependent) - proposed mechanism? Injured tissue : positive charge relative to the surrounding uninjured tissue returning to normal only after the wound closes to promote tissue healing through antimicrobial activity - monophasic current ex. Microampere level DC and HVPC - bacterial growth를억제하기위한 ES는 much higher voltage 또는 much longer times 적용 to promote tissue healing by increasing circulation during or after the stimulation
4 Parameter for ES for tissue healing 4 Electrode placement Waveform Polarity Pulse duration Frequency On:off time Current amplitude Treatment time
5 Electrode placement 5 Figure Electrode placement to promote tissue healing.
6 6 Waveform Monophasic waveform HVPC LIDC
7 7 High-voltage pulsed current (HVPC) DC dual peak pulse 장점맥동기간이짧고, 순간전압이높아통증이나조직손상을유발하지않고효과적으로심부조직을자극할수있다. 특징
8 Polarity 8 Polarity of the electrode on or nearest to the wound : wound healing의특정단계, infection or inflammation 유무에따라결정 - Negative polarity: early inflammation stage - Positive polarity : later to facilitate epithelial cell migration across the wound bed negative polarity positive polarity
9 9 Pulse duration : HVPC, 40~100 μs Frequency : 60~125 pps On:off time : off time X Current amplitude - sufficient to produce a comfortable sensation without a motor response - 치료부위의감각이감소되거나이상감각이있을경우에는정상감각부위에먼저 electrode 를적용하여적절한강도결정
10 Edema control 10 Edema formation associated with inflammation : negative polarity HVPC below the threshold for motor contraction microvessel permeability Edema formation caused by poor peripheral circulation due to lack of motion : motor level ES elevation, compression garment 함께사용하면효과적
11 Parameter for ES for edema control 11 ES for edema associated with inflammation ES for edema associated with lack of muscle contraction
12 Iontophoresis 12 transdermal drug delivery를촉진하기위해 low-amplitude DC 사용 : mechanism? Depth of drug delivery : uncertain, 3~20mm - salicyclic acid와 lidocain은 stratum corneum( 표피각질층 ) 투과가능 ma-min : table 8-5, 주로 40 ma-min권장 ionized drug의 continuous delivery를위해 DC 사용, akaline reaction: electrode 아래에서 undesirable chemical change 발생가능.! NaOH (- electrode), alkaline reaction, discomfort, skin irritation, chemical burn to reduce the risk : negative electrode larger, current amplitude decrease HCl (+ electrode) : acidic reaction, less uncomfortable than alkaline reaction Table 8-6
13 Parameter for the application of Iontophoresis 13 Electrode placement and size - drug delivery electrode, dispersive electrode - current density 가 0.5 ma(cathode), 1.0 ma(anode) 를초과하지않도록충분히커야한다. Polarity same polarity as the polarity of the active ion Current amplitude : patient comfort 에의해결정, 4mA 를넘어서는안됨. Treatment time - total treatment dose of 40 ma-min : table 7-7
14 14 Figure hour iontophoresis patch. Figure Hybrid iontophoresis device.
15 15 Figure The molecular structure of dex amethasone sodium phosphate. Figure Electrode placement for iontophoresis
16 Contraindication and precaution for the use of electrical current
17 Contraindication for the use of electrical currents 17 Demand pacemaker or unstable arrhythmias Over the carotid sinus Venous or arterial thrombophlebitis Pelvis, abdomen, trunk, and low back area during pregnancy
18 Precautions for the use of electrical currents 18 Cardiac disease Impairment mentation or impaired sensation Malignant tumor Skin irritation or open wounds Iontoporesis after another physical agent
19 Adverse effects of electrical current 19 Careful evaluation으로 adverse effect 최소화 Burn at DC and AC Skin irritation or inflammation Pain
20 20 Application technique Patient positioning Electrode type Electrode placement
21 Patient positioning Area to be treated, goal of treatment, the device used 에따라환자자세결정 ES for muscle strengthening midline end of the available ROM Positioning strap, cuff weight 활용하여고정
22 Electrode type 22 Electrode most commonly used today Self-adhesive gel coating that serves as the conduction medium disposable and flexible gel : decrease resistance between the electrode and the skin Gel coating start to dry out uneven current density burn Carbon rubber electrode more long-lasting electrode, not self-adhesive Should be cleaned with warm, soapy water and not alcohol Current density 와 electrode size 는반비례 larger electrode : more comfortable than smaller ones cannot target small area
23 Electrode placement 23 Electrode must lie smoothly against the skin without wrinkles or gap self adhesive electrode : maintain good contact other type : flexible bandage 필요, Should not be placed directly over bony prominence distance or spacing between electrodes Document electrode placement using diagram
24 24 General instruction for ES
25 25 Figure Treatment of upper back and neck pain with electrical stimulation.
26 26 Figure A, Electrical stimulation to increase hamstring strength
27 27 Figure B, Electrical stimulation to increase quadriceps strength.
28 28 General stimulations for ES
29 29 General term
30
31 31 Waveform
32 32 Alternating current Continuous bidirectional flow of charged particle equal ion flow in each direction, no pulse charge Frequency, cycle duration Relationship between frequency and cycle duration Frequency, cycle duration 1000~10,000 Hz
33 Medium frequency 33 AC with a frequency between 1000 and 10,000 Hz 2500 ~ 5000 Hz for clinical unit 2개의다른 medium frequency AC 의의해생성된 interferential current의형태로주로사용
34 Continuous current 34 Continuous flow of charged particles without interruption of break Continuous current Direct current(dc) : one direction Alternating current (AC) : two direction
35 35 Interferential current interferential current between two crossed pairs of electrodes ( 4 electrode) An alternating current with a frequency of 5000 Hz interfering with an alternating current with a frequency of 5100 Hz to produce an interferential current with a beat frequency of 100 Hz.
36 Interferential current 36 More comfortable than other waveform, Why? Low amplitude current on skin More total current than pulsed wave form Larger area than other waveform 많은 Animal study 에서 Inflammation 과 ischemia 와관련된 pain 을감소시킨다고하였으나, 특별하게다른효과없다고나타남.
37 37 Premodurated current AC with medium frequency Same form as an interferential current But single circuit and only 2 electrode No advantage of interferential current (lower current amplitude, large area)
38 Pulsed current(pulsatile current) 38 An interrupted flow of charged particle Type Monophasic pulsed current tissue healing and acute edema management application Commonly used High voltage pulsed current(hvpc) (= pulsed galvanic current) A, Monophasic; B, biphasic pulsed currents.
39 Pulsed current 39 Biphasic pulsed current Symmetrical : Asymmetrical : balanced, unbalanced Symmetrical, balanced asymmetrical biphasic current net chare of zero Often little difference between symmetrical and asymmetrical BPC Symmetrical : more comfortable for large muscle group Asymmetrical : for small muscle group A, Symmetrical; B, balanced asymmetrical; C, unbalanced asymmetrical biphasic pulsed currents.
40 Russian protocol 40 Waveform with specific parameter intended for quadriceps muscle strengthening Median frequency AC with 2500 Hz delivered in 50 bursts/second (= medium frequency burst AC)
41 41 Time-dependent parameter Frequency Interphase interval(intrapulse interval) Interpulse interval On:off time Phase duration Pulse duration Ramp up/ramp down Rise time/decay time Wave length
42 Frequency 42 Number of cycles or pulsed per second Measurement unit Hertz(Hz) for cycle Pulse per second(pps) for pulse Monophasic pulsed current with frequencies of 3 pps and 9 pps.
43 Interval 43 Interphase interval (intrapulse interval) Interpulse interval The time between phases of a pulse The time between pulse Interphase interval for a biphasic pulsed current
44 On: off time 44 On time: time during which a train of pulses occurs off time: time between trains of pulses when no current flows Usually only used when ES is used to muscle contraction On time, off time One time: muscle contraction off time : muscle relaxation, needed to reduce muscle fatigue during stimulation session Sequential on and off time to mimic the voluntary contraction and relaxation phases of normal physiological exercise On : off ratio 10sec on : 50 sec off 1:5 on off ratio On:off times for a biphasic current
45 Duration 45 Phase duration Pulse duration Duration of one phase of a pulse μs, ms Time from the beginning of the first phase of a pulse to the end of the last phase of a pulse μs
46 Time 46 Ramp up / ramp down time Rise time/decay time During the on time During any one phase
47 Wavelength 47 The duration of 1 cycles of AC The wavelength of AC is similar to the pulse duration of pulsed current
48 48 Other ES parameter Amplitude (intensity) Amplitude modulation Burst mode Frequency modulation Modulation Phase duration or pulse duration modulation Scan Sweep
49 Amplitude (intensity) The magnitude of current or voltage
50 50 Modulation
51 Burst mode 51 A current composed of series pulsed delivered in groups known as burst
52 Modulation 52 Used to limit neural adaptation to an electrical current Type Amplitude modulation Frequency modulation Phase duration /pulse duration modulation Scan : amplitude modulation of an interferential current Sweep: frequency modulation of an interferential current
53 53 Nerve and electrical current Absolute refractory period Accommodation Action potential Adaptation Chronaxie Depolarization Myelin Node of Ranvier Propagation Relative refractory current
54 54 Unn Figure Unn. Fig. 8-1.
55 55 Medium frequency Continuous current Direct current Alternating current Interferential premodulated current Russian protocole
ELECTRICAL CURRENT FOR PAIN CONTROL
ELECTRICAL CURRENT FOR PAIN CONTROL 2 Pain modulation ES for pain modulation (TENS) Parameter for ES for pain modulation Pain modulation 3 Gate control theory of pain modulation Pain perception nociceptor
More informationZ-STIM IF250 * Interferential / Russian Stimulator
Z-STIM IF250 * Interferential / Russian Stimulator User's Guide Amrex electrotherapy equipment a division of Amrex-Zetron, Inc. *Federal law (USA) restricts this device to sale by or on the order of a
More informationMICROWAVE DIATHERMY AND SURGICAL DIATHERMY DIATHERMICS
MICROWAVE DIATHERMY AND SURGICAL DIATHERMY 1 Microwave diathermy Microwave diathermy uses microwaves to generate heat in the body. It can be used to evenly warm deep tissues without heating the skin. Microwave
More informationUltima Neo Estim Programs for ND Clinicians
Ultima Neo Estim Programs for ND Clinicians 2.1.13 TENS MODE P-0 Frequency Auto Variable- range 6-20Hz- 250 pulse width Hands Asymmetrical Bi-Phasic Spiked P-1 Frequency 80Hz-150 pulse width Feet Asymmetrical
More informationUser Manual Operation & Installation Instructions for: Two Channel Transportable Electrotherapy Unit
Leading Therapy in a New Direction User Manual Operation & Installation Instructions for: 2783- Two Channel Transportable Electrotherapy Unit 2005 Encore Medical, L.P. TABLE OF CONTENTS FOREWORD........................................
More informationMULTISTIM MS100 Neuro-muscular Stimulator
MULTISTIM MS100 Neuro-muscular Stimulator Operating Instructions Metron Medical Australia Pty Ltd ACN 050 240 527 P.O. Box 2164, 57 Aster Avenue, Carrum Downs, Victoria, Australia 3201 Int Tel: (613) 9775
More informationUser Manual. Moving Rehabilitation Forward. Operation & Installation Instructions for: Two Channel Mobile Stimulation Unit
Moving Rehabilitation Forward User Manual Operation & Installation Instructions for: 2777- Two Channel Mobile Stimulation Unit DJO is an ISO 13485 Certified Company Electromagnetic Compatibility (EMC)
More informationInstruction Manual. CAUTION: Federal law requires a prescription from your physician before use of this product.
Instruction Manual Please read the Instruction Manual prior to use. CAUTION: Federal law requires a prescription from your physician before use of this product. 3504 Cragmont Dr. Suite #100 Tampa, FL 33619
More informationInTENSity Professional Series
InTENSity Professional Series InTENSity Professional Series USER MANUAL Operation Instructions For: DQ7000 InTENSity EX4 (Electrotherapy) DQ8000 InTENSity CX4 (Combination Electrotherapy & Ultrasound)
More informationPhysical Therapy. Methods in Physical Therapy. Non-electric heat therapy Therapeutique application of electric current
Physical Therapy Methods in Physical Therapy Hubbard Hydrotherapy Tank, Carlos Andreson, Watercolour, 943 Non-Electric heat therapy (heating or cooling) Electrotherapy Ultrasound therapy Magneto-therapy
More informationTranscutaneous Electrical Nerve Stimulator TENS 212. Instruction Manual. Read before using
Transcutaneous Electrical Nerve Stimulator TENS 212 Instruction Manual Read before using TABLE OF CONTENTS GENERAL DESCRIPTION 1 SYSTEM COMPONENTS 1 WARRANTY 1 INDICATIONS AND CONTRAINDICATIONS 2 WARNINGS
More informationLSI System IV. Interferential with Russian Stimulation Owner s Manual
LSI System IV Interferential with Russian Stimulation Owner s Manual KC Metro Area: 913-894-4493 Toll Free: 800-832-0053 Fax: 866-216-2541 lsiinternational.com Rev.B OM002 Updated 06/2012 LSI International,
More informationLab #9: Compound Action Potentials in the Toad Sciatic Nerve
Lab #9: Compound Action Potentials in the Toad Sciatic Nerve In this experiment, you will measure compound action potentials (CAPs) from an isolated toad sciatic nerve to illustrate the basic physiological
More informationI. Introduction to Animal Sensitivity and Response
Stray Voltage Field Guide Douglas J. Reinemann, Ph.D. Professor of Biological Systems Engineering University of Wisconsin Madison September 2007 Update I. Introduction to Animal Sensitivity and Response
More informationEmpi COMPLETE ELECTROTHERAPY SYSTEM USER S MANUAL. Read this manual carefully before operating the Phoenix. Visit us at
Empi COMPLETE ELECTROTHERAPY SYSTEM USER S MANUAL Read this manual carefully before operating the Phoenix Visit us at www.djoglobal.com Table of Contents 1. Foreword........................................
More informationMoving Rehabilitation Forward. User Manual Combination Ultrasound and Stimulation Unit. ISO Certified
Moving Rehabilitation Forward User Manual 2738- Combination Ultrasound and Stimulation Unit ISO 13485 Certified TABLE OF CONTENTS FOREWORD........................................ 1 PRODUCT DESCRIPTION..........................
More informationTrio*Stim. Instruction Manual
Trio*Stim Instruction Manual Please read this manual carefully before using the Trio*Stim. This manual is comprised as an essential part of the Trio*Stim. Save this manual in a designated place for your
More informationC14 M 3.3 Diathermy Role Name Affiliation Principal Investigator Dr. Asis Goswami Ramakrishna Mission Vivekananda University Co-Principal Investigator
C14 M 3.3 Diathermy Role Name Affiliation Principal Investigator Dr. Asis Goswami Ramakrishna Mission Vivekananda University Co-Principal Investigator Dr. P.K. Nag Ramakrishna Mission Vivekananda University
More informationSERVICE MANUAL. Dynatron 950plus & Dynatron 650plus
SERVICE MANUAL Dynatron 950plus & Dynatron 650plus CAUTION: Federal law restricts this device to sale by or on the order of a physician, chiropractor, physical therapist, or dentist licensed by the law
More informationSonicator Plus 994 Specifications
Sonicator Plus 994 Specifications General Specifications: Input: Certification: Domestic model ETL and C-ETL Listed: Domestic model Classification: CE model Year 2000 Compliant Weight: Dimensions: Operating
More informationMH8001 Four Channels Pre-Programmed TENS and EMS combo device
MH8001 Four Channels Pre-Programmed TENS and EMS combo device The MH8001 is provided with 4 controllable output channels, each independent of each other. The unit integrates EMS and TENS functions with
More informationTENSPros. Twin Stim Plus 3 rd Edition INSTRUCTION MANUAL. Combo Stimulator
Twin Stim Plus 3 rd Edition Combo Stimulator INSTRUCTION MANUAL This manual is valid for the Twin Stim Plus 3 rd Edition Combo Stimulator This user manual is published by Roscoe Medical, Inc. Roscoe Medical,
More informationRICH-MAR Theratouch 7.7-S
RICH-MAR Theratouch 7.7-S OPERATION HANDBOOK AND MANUAL P/N: 400-120 Rev. 2 1 CAUTION This device is not designed to be connected with any electrical equipment unless manufactured and approved by Rich-Mar.
More informationDIATHERMY UNITS HIGH FREQUENCY HEAT THERAPY:
DIATHERMY UNITS HIGH FREQUENCY HEAT THERAPY: The dipole molecules of the body are normally placed randomly. Under electric field, they rotate in the direction of the field lines so that the positively
More informationAS SUPER 4 digital. Elektrischer Nadelstimulator / Electrical needle stimulator. Art.-Nr
AS SUPER 4 digital Elektrischer Nadelstimulator Electrical needle stimulator Art.-Nr. 200510 Gebrauchsanweisung Instruction Manual - Art.-Nr. 101477 20 Contents Foreword...21 Purpose for use...21 Safety
More informationOPERATING INSTRUCTIONS. PHYSIOMED-IF-Expert GB
OPERATING INSTRUCTIONS PHYSIOMED-IF-Expert 00904 GB The technical data in this manual is as at the time of printing and subject to alteration. Copyright 2001-2013 by PHYSIOMED ELEKTROMEDIZIN AG All rights,
More informationINTELECT. User Manual LEGEND SERIES COMBO. Models 2C & 4C ISO CERTIFIED
INTELECT LEGEND SERIES User Manual COMBO Models 2C & 4C ISO 13485 CERTIFIED Table of Contents Foreword................................................................................. 3 Product Description.........................................................................
More informationI. Introduction to Animal Sensitivity and Response
I. Introduction to Animal Sensitivity and Response The term stray voltage has been used to describe a special case of voltage developed on the grounded neutral system of a farm. If this voltage reaches
More informationCAUTION: United States Federal Law restricts this device to sale by or on the order of a physician
INSTRUCTION MANUAL CAUTION: United States Federal Law restricts this device to sale by or on the order of a physician www.roscoemedical.com TM This manual is valid for the Quattro 2.5 This user manual
More informationEDL Group #3 Final Report - Surface Electromyograph System
EDL Group #3 Final Report - Surface Electromyograph System Group Members: Aakash Patil (07D07021), Jay Parikh (07D07019) INTRODUCTION The EMG signal measures electrical currents generated in muscles during
More informationINTRODUCTION. It is also intended for symptomatic relief and management of chronic, intractable pain and relief of pain associated with arthritis.
TENS / HEAT 1 2 TABLE OF CONTENTS Introduction...4 Indications for Use...4 Safety Warning...5 Contraindications...5 Warnings...5 Precautions...6 Adverse Reactions...8 Symbol and Title...8 Environmental
More informationSys * Stim 294. Instruction Manual
Sys * Stim 294 Instruction Manual 1333 South Claudina Street Anaheim, CA 92805, U. S. A. Toll Free: (800) 854 9305 Telephone: (714) 533 2221 FAX: (714) 635 7539 Web Site: http://www.mettlerelectronics.com
More informationAmrex. SynchroSonic U/HVG50 * User's Guide. electrotherapy equipment a division of Amrex-Zetron, Inc.
SynchroSonic U/HVG50 * User's Guide Amrex electrotherapy equipment a division of Amrex-Zetron, Inc. *Federal law (USA) restricts this device to sale by or on the order of a licensed practitioner licensed
More informationINSTRUCTION MANUAL. TENSPros. CAUTION: United States Federal Law restricts this device to sale by or on the order of a physician
INSTRUCTION MANUAL CAUTION: United States Federal Law restricts this device to sale by or on the order of a physician www.currentsolutionsnow.com This manual is valid for the ComboCare This user manual
More informationstock Microstim2(v2) Instruction Manual
stock Microstim2(v2) Instruction Manual SALISBURY DISTRICT HOSPITAL DEPARTMENT OF MEDICAL PHYSICS & BIOMEDICAL ENGINEERING www.salisburyfes.com WARNING: Output has a physiological effect! Read instructions
More informationPINXEL. Micro Needle Rf & Fractional Rf
Micro Needle Rf & Fractional Rf Introduction of Pinxel Pinxel is designed specially for delivering RF energy directly to epidermis or dermis layer in order to stimulate and re-model collagen effectively
More informationCOMBI 200. User Manual. Pasweg 6A B-3740 Bilzen. Tel.: (+32) (0) 89/ Fax: (+32) (0) 89/
User Manual Pasweg 6A B-3740 Bilzen Tel.: (+32) (0) 89/510.510 Fax: (+32) (0) 89/510.511 www.gymna-uniphy.com E-mail: info@gymna-uniphy.com Your dealer: COMBI 200 User Manual Combi 200 Device for electrotherapy,
More informationAmrex. MS324A * Low Volt AC Stimulator. User's Guide. electrotherapy equipment a division of Amrex-Zetron, Inc.
MS324A * Low Volt AC Stimulator User's Guide Amrex electrotherapy equipment a division of Amrex-Zetron, Inc. *Federal law (USA) restricts this device to sale by or on the order of a licensed practitioner
More informationSANO. Pinxel Introduction
Pinxel Introduction SANO Our Pinxel model is specially designed to deliver RF energy directly to the epidermis or dermis skin layer. This stimulates and remodels collagen effectively by either RF microneedling
More informationSonicator Plus 940. Instruction Manual
Sonicator Plus 940 Instruction Manual 1333 South Claudina Street Anaheim, CA 92805, U. S. A. Toll Free: (800) 854 9305 Telephone: (714) 533 2221 FAX: (714) 635 7539 Web Site: http://www.mettlerelectronics.com
More informationCHAPTER 7 INTERFERENCE CANCELLATION IN EMG SIGNAL
131 CHAPTER 7 INTERFERENCE CANCELLATION IN EMG SIGNAL 7.1 INTRODUCTION Electromyogram (EMG) is the electrical activity of the activated motor units in muscle. The EMG signal resembles a zero mean random
More informationUser Manual. Before Using Your WiTouch Pro Device
User Manual Before Using Your WiTouch Pro Device Sync the Remote Control and the WiTouch Pro Device. Using the provided screwdriver, remove the back cover from the WiTouch Pro device. 2. Remove the clear
More informationTENS eco 2. schwa-medico. Transcutaneous nerve and muscle stimulator. Art.-No Operating manual Art.-No V03.
TENS eco 2 Transcutaneous nerve and muscle stimulator Art.-No. 104062 Operating manual Art.-No. 101446-V03 schwa-medico holistic health 2 Content Before using TENS eco 2...3 Intended use...3 Safety instructions...3
More informationFundamentals of Pacing Therapy
1 4 7 2 5 8 3 6 9 Fundamentals of Pacing Therapy This presentation is provided with the understanding that the slide content must not be altered in any manner as the content is subject to FDA regulations.
More informationSys * Stim 228. Instruction Manual
Sys * Stim 228 Instruction Manual 1333 South Claudina Street Anaheim, CA 92805, U. S. A. Toll Free: (800) 854 9305 Telephone: (714) 533 2221 FAX: (714) 635 7539 Web site: www.mettlerelectronics.com email:
More informationBiomedical Engineering Evoked Responses
Biomedical Engineering Evoked Responses Dr. rer. nat. Andreas Neubauer andreas.neubauer@medma.uni-heidelberg.de Tel.: 0621 383 5126 Stimulation of biological systems and data acquisition 1. How can biological
More informationINSTRUCTIONS FOR YOUR NEW PLASMAFLOW. Vascular Therapy System (Compressible Limb Sleeve Device)
INSTRUCTIONS FOR YOUR NEW PLASMAFLOW Vascular Therapy System (Compressible Limb Sleeve Device) Customer Service Toll Free: 888-508-0712 Email: CustomerService@manamed.net Web: www.manamed.net 1511 W. Alton
More informationBiomedical. Measurement and Design ELEC4623/ELEC9734. Electrical Safety and Performance Standards
Biomedical Instrumentation, Measurement and Design ELEC4623/ELEC9734 Electrical Safety and Performance Standards Contents Physiological Effects of Electrical Currents Safety Standards for Medical Instrumentation
More informationSCENAR: the secrets of effectiveness. RITM OKB ZAO, M. Unakafov
SCENAR: the secrets of effectiveness RITM OKB ZAO, M. Unakafov A few words about electrotreatment Galvanization is low-strength (under 50mA) and lowvoltage (30-80V) direct current. Galvanization exists
More informationLIGHT THERAPY. Long wavelength. Low frequency. High frequency. Short wavelength. Part 2. General Sciences. application. application. wavelengths.
142 Figure 11 application. Direct high-frequency There are two methods for applying high-frequency current. 1. Direct surface application. Do not apply any product to the client s face. If the glass electrode
More informationCyMedica Orthopedics. e-vive USER S MANUAL. Read this manual carefully before operating the e-vive System. Customer Service CYM-2014
CyMedica Orthopedics e-vive USER S MANUAL Read this manual carefully before operating the e-vive System Customer Service 1-844-CYM-2014 www.cymedicaortho.com Please log onto www.cymedicaortho.com for instructional
More informationTENS / EMS Dual Channel Electrical Stimulator Model GF-DF5
TENS / EMS Dual Channel Electrical Stimulator Model GF-DF5 Operation Manual Read this manual before operating your GF-DF5 Dual Channel Electrical Stimulator. Save this manual for future use. GF-DF5-INS-LAB-RevA16
More informationEE 230 Experiment 10 ECG Measurements Spring 2010
EE 230 Experiment 10 ECG Measurements Spring 2010 Note: If for any reason the students are uncomfortable with doing this experiment, please talk to the instructor for the course and an alternative experiment
More informationTable of Contents. 6.6 Operation in the Parameter Setting mode Memory checking interface In all operation interfaces...
Table of Contents 1. Getting to know your device... 1 1.1 What is Electrical Stimulator and what can it do?..1 1.2 Important information... 2 1.3 WARNINGS... 2 1.4 CAUTIONS... 3 1.5 Adverse Reactions...
More informationiworx Sample Lab Experiment AN-2: Compound Action Potentials
Experiment AN-2: Compound Action Potentials Exercise 1: The Compound Action Potential Aim: To apply a brief stimulus at the proximal end of the nerve and record a compound action potential from the distal
More informationEXPERIMENT 8 Bio-Electric Measurements
EXPERIMENT 8 Bio-Electric Measurements Objectives 1) Determine the amplitude of some electrical signals in the body. 2) Observe and measure the characteristics and amplitudes of muscle potentials due to
More informationLecture 4. Medical Textiles Total Consumption. cancer. Properties of Medical Textiles
Consumption Growth Lecture 4 Transport Home textiles Industrial Medical Building Agriculture Technical components of footwear and clothing Packaging Geotextiles Sports Personal and property protection
More informationDUO 500 ELECTROTHERAPY UNIT
DUO 500 ELECTROTHERAPY UNIT OPERATING INSTRUCTIONS Your supplier is: The DUO 500 is made in Belgium by GymnaUniphy At the beginning of these Operating Instructions you will find an exploded view of the
More informationInstruction Manual for the T.I.M.E. Multitherapist
Instruction Manual for the T.I.M.E. Multitherapist Model MT9000 www.tenscare.co.uk 1 This manual is valid for the MT9000 Combo TENS/EMS/IF/MIC Stimulator This user manual is published by Shenzhen Dongdixin
More informationSys*Stim 226. Maintenance Manual
Sys*Stim 226 Maintenance Manual IR7 82 1333 S. Claudina Street Anaheim, CA 92805 U.S.A. Call toll free: (800) 854 9305 (U.S.A.) Tel: 1 (714) 533 2221 FAX: 1 (714) 635 7539 Web site: www.mettlerelectronics.com
More informationThis manual is valid for the TM. In TENSity 5000 TENS Stimulator. This user manual is published by Current Solutions, LLC
INSTRUCTION MANUAL This manual is valid for the TM In TENSity 5000 TENS Stimulator This user manual is published by Current Solutions, LLC Current Solutions, LLC does not guarantee its contents and reserves
More informationTABLE OF CONTENTS CONVENTIONS AND GRAPHICAL SYMBOLS 1. INTRODUCTION 2 2. WARNINGS 3 3. CAUTIONS 4 4. ADDITIONAL BENEFITS 5 5. PACKAGE CONTENTS 6
USER S MANUAL TABLE OF CONTENTS 1. INTRODUCTION 2 2. WARNINGS 3 3. CAUTIONS 4 4. ADDITIONAL BENEFITS 5 5. PACKAGE CONTENTS 6 6. USING LIVIA 6 7. CHARGING LIVIA 11 8. SPECIFICATIONS 11 9. SYSTEM CONTROLS
More informationitouch Easy and itouch Plus User s Guide
user_manual_col_v2 13/1/08 14:32 Page 1 itouch Easy and itouch Plus User s Guide Includes Setup and troubleshooting information for your itouch Easy and Plus Intelligent TENS TensCare Ltd, 9 Blenheim Road,
More informationComponent Technical Specifications
Component Technical Specifications 11 Control Unit Specifications Classification Operation Modes Battery Type Controls Indications Carrying Options Dimensions Environmental Ranges Internally powered, continuous
More informationThis manual is valid for the InTENSity TM Micro Combo TENS/MIC Stimulator
INSTRUCTION MANUAL This manual is valid for the InTENSity TM Micro Combo TENS/MIC Stimulator This user manual is published by Roscoe Medical, Inc. Roscoe Medical does not guarantee its contents and reserves
More informationDual Channel TENS. User Manual. Prescription Strength Pain Relief
User Manual Dual Channel TENS Prescription Strength Pain Relief Quick and Easy Set-Up 20 Minute Treatment Timer 12 Preset Modes 100 Intensity Levels Treatment Record Log Auto Shut-Off This manual is valid
More informationINSTRUCTION MANUAL MULTI MODALITY ELECTRICAL STIMULATOR INTERFERENTIAL TENS EMS MICROCURRENT
INSTRUCTION MANUAL MULTI MODALITY ELECTRICAL STIMULATOR INTERFERENTIAL TENS EMS MICROCURRENT Table of Contents 1.Safety information..... 4 1.1 General description 1.2 Medical background 1.3 Indication
More informationAmrex. MS324C * Low Volt AC Stimulator. User's Guide. electrotherapy equipment a division of Amrex-Zetron, Inc.
MS324C * Low Volt AC Stimulator User's Guide Amrex electrotherapy equipment a division of Amrex-Zetron, Inc. *Federal law (USA) restricts this device to sale by or on the order of a licensed practitioner
More informationEmpi Elite COMPLETE ELECTROTHERAPY SYSTEM. USER S MANUAL Read this manual carefully before operating the Elite Visit us at
Empi Elite COMPLETE ELECTROTHERAPY SYSTEM USER S MANUAL Read this manual carefully before operating the Elite Visit us at www.djoglobal.com Table of Contents Foreword...........................................
More informationFront View 3 2. Selection and Operating Elements
Front View 3 2 4 5 6 7 1 Selection and Operating Elements 1 On and Off switch 2 Navigation keys 3 Title bar 4 Keys on screen 5 Channel I intensity regulator 6 Status bar 7 Channel II intensity regulator
More informationThis manual is valid for the InTENSity TM Twin Stim III TENS/EMS Combo Stimulator
INSTRUCTION MANUAL This manual is valid for the InTENSity TM Twin Stim III TENS/EMS Combo Stimulator This user manual is published by Compass Health Brands Corp. Compass Health Brands does not guarantee
More informationisoothe User Manual WIRELESS RECHARGEABLE TENS INV-942R
isoothe WIRELESS RECHARGEABLE TENS Rechargeable Battery 20 Minute Treatment Timer 5 Preset Modes 15 Intensity Levels Auto Shut-Off Clinically Proven User Manual INV-942R This manual is applicable to the
More informationANC: Section 2. Unidirectional Propagation - 1 J Thomas Mortimer & Narendra Bhadra
ANC: Section 2. Unidirectional Propagation - 1 J Thomas Mortimer & Narendra Bhadra Under physiological conditions, a nerve action potential (AP) is generated at one end of an axon and proceeds towards
More informationBiomechatronic Systems
Biomechatronic Systems Unit 4: Control Mehdi Delrobaei Spring 2018 Open-Loop, Closed-Loop, Feed-Forward Control Open-Loop - Walking with closed eyes - Changing sitting position Feed-Forward - Visual balance
More informationBiomechatronic Systems
Biomechatronic Systems Unit 4: Control Mehdi Delrobaei Spring 2018 Open-Loop, Closed-Loop, Feed-Forward Control Open-Loop - Walking with closed eyes - Changing sitting position Feed-Forward - Visual balance
More informationSensor Array. Multi-Electrode Mapping Vest. Technical Manual
Sensor Array Multi-Electrode Mapping Vest Technical Manual Explanation of packaging and product symbols Refer to the package label and product to see which symbols apply to this product. Item Description
More informationEMP2 PRO 2-CHANNEL MUSCLE STIMULATOR. Instruction Manual Art. no Item no
EMP2 PRO 2-CHANNEL MUSCLE STIMULATOR Item no. 104068 Instruction Manual Art. no. 101674 2 EMP2 PRO Instruction Manual Contents 1 Intended use... 4 2 Safety instructions... 4 Important safety hints... 4
More informationINTELECT SHORTWAVE DIATHERMY MOVING REHABILITATION FORWARD
INTELECT SHORTWAVE DIATHERMY MOVING REHABILITATION FORWARD Chattanooga Group is the world s largest manufacturer of rehabilitation equipment for treating musculoskeletal, neurological and soft tissue disorders.
More informationINDEX. Chapter Contents
INDEX Chapter Contents Page Index....1 1. Introduction. 2 2. Cautions.. 2 3. Warnings.. 3 4. General Description... 4 5. Construction......5 6. Technical Specifications...6 7. Replaceable Parts......7
More informationExperiment HN-12: Nerve Conduction Velocity & Hand Dominance
Experiment HN-12: Nerve Conduction Velocity & Hand Dominance This lab written with assistance from: Nathan Heller, Undergraduate research assistant; Kathryn Forti, Undergraduate research assistant; Keith
More informationCompound Action Potentials
Compound Action Potentials Background As last week s lab showed, even a piece of thread can conduct electricity from stimulating electrodes to recording electrodes in the nerve chamber. You know from lecture
More informationINDEX. Chapter Contents
CMYK INDEX Chapter Contents Page Index....1 1. Introduction... 2 2. Cautions... 3 3. Warnings... 4 4. General Description... 4 5. Construction.....5 6. Technical Specifications...... 6 7. Replaceable Parts....
More informationPain Management System
TM Pain Management System Model: ireliev Model #: ET-1313 Operating & Instruction Manual Read Before Using ireliev Pain Management System Intended Use The ireliev Pain Management System (Model # ET-1313)
More informationLG QUAD Combo INSTRUCTION MANUAL
LG QUAD Combo INSTRUCTION MANUAL This manual is valid for the LG QUAD Combo TENS/IF/MIC/EMS Stim device This user manual is published by Current Solutions, LLC Current Solutions, LLC does not guarantee
More informationSonicator Plus 994. Instruction Manual
Sonicator Plus 994 Instruction Manual 1333 South Claudina Street Anaheim, CA 92805, U. S. A. Toll Free: (800) 854 9305 Telephone: (714) 533 2221 FAX: (714) 635 7539 Web Site: http://www.mettlerelectronics.com
More informationOPERATING MANUAL MINIDOP ES-100VX POCKET DOPPLER
OPERATING MANUAL MINIDOP ES-100VX POCKET DOPPLER CONTENTS * Features.......................... 1 * Cautions.......................... 2 * Clinical applications................. 4 * Operating controls..................
More informationINDEX. Chapter Contents Page
INDEX Chapter Contents Page 1. Introduction... 2 2. Cautions... 3 3. Warnings... 4 4. Contraindications... 5 5. Adverse Reactions... 5 6. General Description... 5 7. Construction... 6 8. Technical Specifications...
More informationPRODUCT CODE: XT EN
PRODUCT CODE: XT-46006-EN XM400-31A04-07 i Manufacturer Xavant Technology Pty (Ltd) Unit 102, The Tannery Industrial Park, 309 Derdepoort Rd Silverton, Pretoria, South Africa, 0184 Tel: +27 (0) 12 743
More informationINSTRUCTION MANUAL FOR THE
INSTRUCTION MANUAL FOR THE PAIN-CARE TENS Distributed by: V1.0 0434 INDEX Chapter Contents Page 1. General Description...2 2. Introduction...2 3. Cautions...3 4. Warnings...4 5. Contraindications...5 6.
More informationDE / EN. Operating Manual Art.-Nr Amplimove medical NEURO MUSCULAR ELECTRO-STIMULATOR for amplitude modulated. schwa-medico.
DE / EN Operating Manual Art.-Nr. 101799 schwa-medico holistic health Amplimove medical NEURO MUSCULAR ELECTRO-STIMULATOR for amplitude modulated medium frequency electro-therapy abbreviatedinstructions
More informationUsed to overcome ventricular fibrillation may be due to coronary occlusion, shock, or abnormalities in blood chemistry
Used to overcome ventricular fibrillation may be due to coronary occlusion, shock, or abnormalities in blood chemistry Main problem: heart muscle fibers are continuously stimulated by adjacent muscles
More informationTranscutaneous Electrical Nerve Stimulator
Transcutaneous Electrical Nerve Stimulator Simple, effective pain relief *For the temporary relief of pain as part of current medication programme. Distributor Medi-Direct International Ltd. Unit 24, Wilford
More informationQuick Guide - Some hints to improve ABR / ABRIS / ASSR recordings
Quick Guide - Some hints to improve ABR / ABRIS / ASSR recordings Several things can influence the results obtained during ABR / ABRIS / ASSR testing. In this guide, some hints for improved recordings
More informationModel BSI-1A BIPHASIC STIMULUS ISOLATOR
MicroProbes for Life Science Model BSI-1A BIPHASIC STIMULUS ISOLATOR Owner Manual OUTPUT WAVEFORM FOLLOWS INPUT (LINEAR) PLUS AND MINUS 125 VOLTS COMPLIANCE CONSTANT CURRENT AND CONSTANT VOLTAGE MODES
More informationMASIMO RADICAL 7 Signal Extraction Pulse CO-Oximeter
MASIMO RADICAL 7 Signal Extraction Pulse CO-Oximeter Women s Health Manual MCH Only Policy Group: Cardiovascular Approved by: Heather Crosland Director, Women s Health, Covenant Health, GNH/MCH Site Lead
More informationMini TENS. User Manual. Natural, drug-free pain relief. 20 Minute Treatment Single Patient Use. Model ACRL Clinically Proven
Mini TENS Natural, drug-free pain relief 20 Minute Treatment Single Patient Use Clinically Proven User Manual Model ACRL-1000 This manual is valid for the AccuRelief Mini TENS ACRL-1000 This instruction
More informationSonicator Plus 930. Instruction Manual
Sonicator Plus 930 Instruction Manual 1333 South Claudina Street Anaheim, CA 92805, U. S. A. Toll Free: (800) 854 9305 Telephone: (714) 533 2221 FAX: (714) 635 7539 Web Site: http://www.mettlerelectronics.com
More informationManual of Electronic Pulse Stimulator
Manual of Electronic Pulse Stimulator PL-029K8B Operation Manual PL-029K8B Edition V1.0 1 Table of Contents Introduction...3 Indications for Use...3 Safety Warning... 4 Contraindications...4 Warnings...4
More informationOutput Circuits for Cutaneous Muscle Stimulators A THESIS SUBMITTED TO THE FACULTY OF THE GRADUATE SCHOOL OF THE UNIVERSITY OF MINNESOTA BY
Output Circuits for Cutaneous Muscle Stimulators A THESIS SUBMITTED TO THE FACULTY OF THE GRADUATE SCHOOL OF THE UNIVERSITY OF MINNESOTA BY Joseph Robert Young IN PARTIAL FULFILLMENT OF THE REQUIREMENTS
More informationHuman-to-Human Interface
iworx Physiology Lab Experiment Experiment HN-8 Human-to-Human Interface Introduction to Neuroprosthetics and Human-to-Human Muscle Control Background Set-up Lab Note: The lab presented here is intended
More information