TECHNICAL MANUAL INSTRUMENT CALIBRATION PROCEDURE PEAK POWER ANALYZER HEWLETT-PACKARD 8991AOPT003

Size: px
Start display at page:

Download "TECHNICAL MANUAL INSTRUMENT CALIBRATION PROCEDURE PEAK POWER ANALYZER HEWLETT-PACKARD 8991AOPT003"

Transcription

1 TECHNICAL MANUAL INSTRUMENT CALIBRAON PROCEDURE PEAK POWER ANALYZER HEWLETT-PACKARD 8991AOPT003 THIS PUBLICAON SUPERSEDES NAVAIR 17-20GP-48 DATED 1 MARCH 1998 DISTRIBUON STATEMENT C. DISTRIBUON AUTHORIZED TO U.S. GOVERNMENT AGENCIES AND THEIR CONTRACTORS TO PROTECT PUBLICAONS REQUIRED FOR OFFICIAL USE OR FOR ADMINISTRAVE OR OPERAONAL PURPOSES ONLY. OTHER REQUESTS FOR THIS DOCUMENT SHALL BE REFERRED TO NAVY MEASUREMENT SCIENCE DIRECTORATE, NAVAL WARFARE ASSESSMENT STAON, P.O. BOX 5000, CORONA, CA GP-48 DESTRUCON NOCE - FOR UNCLASSIFIED, LIMITED DOCUMENTS, DESTROY BY ANY METHOD THAT WILL PREVENT DISCLOSURE OF CONTENTS OR RECONSTRUCON OF THE DOCUMENT. PUBLISHED BY DIRECON OF COMMANDER NAVAL AIR SYSTEMS COMMAND 1 MAY 2000

2 LIST OF EFFECVE PAGES The purpose of this list is to identify the pages in this document affected by any technical content changes made since the previous release of the document. NOTE: On a revised page, with the exception of the Title, the A, and the i pages, the technical changes are indicated by a vertical line in the outer margin of the page. Page No. ** Revision* Title 0 Page No. ** Revision* Page No. ** Revision* A 0 i 0 1 to 14 0 *Zero in this column indicates an original page. R in this column indicates a revised page. ** All pages not listed are blank pages. A

3 TABLE OF CONTENTS Section Page 1 Introduction and Description Equipment Requirements Preliminary Operations Calibration Process Vertical Accuracy Tests Horizontal Accuracy Tests Rise Time Tests (Channels 2 and 3) Rise Time Tests (Channels 1 and 4) Trigger Sensitivity Tests (Channels 1 and 4) Instrumentation Uncertainty Tests Check Source Tests Navy Calibration Checklist TABLES Table Page 1 Calibration Description Equipment Requirements... 2 i

4 SECON 1 INTRODUCON AND DESCRIPON 1.1 This procedure describes the calibration of the Hewlett-Packard 8991AOPT003 Peak Power Analyzer. The instrument being calibrated is referred to herein as the (Test Instrument). NOTE This procedure does not calibrate the Hewlett-Packard 84812A or 84815A peak power sensors commonly used with the. To calibrate these peak power sensors, use NAVAIR 17-20GD This procedure was derived from a draft prepared by Naval Air Warfare Center, Aircraft Division (542800A), Patuxent River, MD. All comments concerning this procedure should be directed to Measurement Science Directorate, Naval Warfare Assessment Station, P.O. Box 5000, Corona, CA This procedure includes tests of essential performance parameters only. Any malfunction noticed during calibration, whether specifically tested for or not, should be corrected. Characteristics Type-N connector dimension Vertical accuracy (Ch. 2 and 3) Horizontal accuracy Rise time (Ch. 2 and 3) Rise time (Ch. 1 and 4) Trigger sensitivity (Ch. 1 and 4) Table 1. Calibration Description Performance Specifications In accordance with MIL-C composite: to in. for female connector Gain tolerance: +1.5% of display range Offset tolerance: +(2% of offset value + 20% of V/div setting) Timebase range: 2 ns to 5 s/div. Tolerance: +(0.005% iv + 2% of sec/div setting + 1 ns) Tolerance: <5 ns Tolerance <10 ns at high bandwidth; <1 µs at low bandwidth; <100 µs at CW; 84815A tolerance at High BW: <45 ns Minimum level: -15 dbm high bandwidth; -25 dbm low bandwidth Test Method Measured with a coaxial gage. Tested by having the measure DC voltage from an oscilloscope calibrator. Vertical accuracy = gain accuracy when offset = 0 and (gain accuracy + offset Accuracy) otherwise. The measures known horizontal markers from an oscilloscope calibrator. Tested by applying a fast rise pulse from an oscilloscope calibrator. For the high-bandwidth (<10 ns) test, a Pulse-modulated signal is Characterized with a crystal Detector and oscilloscope and then applied to. For all other tests, the signal itself has a rise time fast enough to be used as a standard. Verified with a signal generator. 1

5 Instrumentation uncertainty (Ch. 1 and 4) Characteristics IEEE-488 (HP-IB) data bus Check source Tolerance: ± Performance Specifications db at high bandwidth; ± db at low bandwidth, CW Conforms to IEEE-STD for talker and listener Frequency: 1.05 Hz Output level: 10 dbm Tolerance: +0.5 db signal in signal in 1 dbm dbm + 33 Test Method Verified with an automated procedure. Verified indirectly during the instrumentation uncertainty tests. Verified with the after its instrumentation uncertainty has been verified. SECON 2 EQUIPMENT REQUIREMENTS NOTE Minimum use specifications are the principal parameters required for performance of the calibration, and are included to assist in the selection of alternate equipment, which may be used at the discretion of the using laboratory. Satisfactory performance of alternate items shall be verified prior to use. All applicable equipment must bear evidence of current calibration. The instruments utilized in this procedure were selected from those known to be available at Navy calibration facilities, and the listing by make or model number carries no implication of preference, recommendation, or approval for use by other agencies. It is recognized that equivalent equipment produced by other manufacturers may be capable of equally satisfactory performance in this procedure. Item Table 2. Equipment Requirements Minimum Use Specifications Calibration Equipment 2.1 Coaxial gage Uncertainty: in. Maury Microwave A007A 2.2 Special BNC cable Delay: 5 ns +10% Length: Approx. 1.5 meters 2.3 Peak power sensors Peak power sensor # Peak power sensor #2 Compatible with Compatible with Hewlett-Packard 10503A Hewlett-Packard 84812A Hewlett-Packard 84815A 2

6 Item Table 2. Equipment Requirements Minimum Use Specifications Calibration Equipment 2.4 Oscilloscope Bandwidth: >200 MHz Tektronix TDS620A 2.5 Oscilloscope calibrator 2.6 Feedthrough termination Voltage: 0 to 20 V dc Uncertainty: % iv Marker spacing: 10 ns to 10 ms Uncertainty: +0.01% iv Fast Edge mode rise time: <2.5 ns Tektronix CG5011 with Programmable Pulse Head Impedance: 50 Ω Tektronix Signal generator Frequency range: 1 to 5 Hz Power range: 0 to -25 dbm Pulse modulation capability with rise time <30 ns Hewlett-Packard 83630AOPT001, 008, H Crystal detector Frequency range: 5 GHz Hewlett-Packard 8470BOPT Calibration software No substitute NAVAIR 17-20AXG D, CHG: Instrument controller No substitute Fluke 1722AAP 2.11 IEEE bus cable Meets IEEE STD and FCC regulation Local supply 2.12 Calibration cable No substitute NAWCAD Patuxent River Digital multimeter Used in Section 4.6 Hewlett-Packard 3458A 2.14 Cables, adapters, tees, etc. As required Local supply SECON 3 PRELIMINARY OPERAONS NOTE Throughout this procedure, hardkeys are those on the front panel, while softkeys are those blank buttons on the side of the display for which the display will indicate the softkey name according to the current menu selection. 3.1 Ensure that all power switches are set to off, and set all auxiliary equipment controls as necessary to avoid damage to the equipment and so that dangerous voltages will not be present on output terminals when the power switches are turned on. 3.2 Verify that the N-type connector of the sensor check source conforms to the MIL-C composite specification (0.187 to ) as follows: If the coaxial gage has been standardized with a MIL-C master gage block, verify that the gage dial indicates between and inches. 3

7 3.2.2 If the coaxial gage has been standardized with a MIL-C-71B master gage block, verify that the gage dial indicates between and inches. 3.3 Connect the and auxiliary equipment to appropriate power sources. Turn the auxiliary equipment power switches to on. Then, while holding any front panel key down, turn the power on. (This ensures that all parameters are reset.) 3.4 Allow sufficient time for the and auxiliary equipment to warm up (the requires one hour warm up time). 3.5 Press the UL hardkey, and then the more (if necessary) and selftest menu softkeys. Disconnect everything from the inputs, if necessary, and then press the test all softkey and ensure that the eventually displays PASSED for all 11 self tests. 3.6 Press the exit menu and instr cal menu softkeys. Set the CALIBRATOR switch on the rear panel to UNPROTECTED, removing any label covering the switch if necessary. 3.7 Press the cal select softkey as necessary to select VERCAL CAL. Press the channel softkey as necessary to select ALL. Press the start cal softkey and follow the instructions on the screen, using BNC cables and a BNC tee connector as necessary. 3.8 A full calibration lasts about 10 minutes. Verify that when the calibration is completed, the status: at the top of the display indicates passed cal. 3.9 Press the cal select softkey to select delay cal. Press the channel softkey as necessary to select channel 1. Press the start cal softkey and follow the instructions on the screen, making connections with the special BNC cable (item 2.2) and peak power sensor #1 (item ) as necessary After completion of the self tests, verify that the status: at the top of the display indicates passed cal Repeat steps 3.9 and 3.10 for channels 2, 3, and 4, except press the channel softkey as necessary to select the appropriate channel Disconnect the special BNC cable and peak power sensor #1 from the. Press the exit menu softkey and set the CALIBRATOR switch (rear panel) to PROTECTED. Push the DISPLAY hardkey Set each of the the softkey settings to the following: DISPLAY to avg # of avg to 16 off/frame/axes/grid to grid connect dots to 3.14 Press the DEFINE MEAS hardkey and then press the continuous softkey as necessary to select on Press the oscilloscope SETUP, Recall Factory Setup (on the bottom menu), and OK Confirm Factory Init keys. When the oscilloscope is ready, press the VERCAL MENU key and set the impedance to 50Ω. on 4

8 SECON 4 CALIBRAON PROCESS NOTE Unless otherwise specified, verify the results of each test and take corrective action whenever the test requirement is not met, before proceeding. 4.1 VERCAL ACCURACY TESTS Set the oscilloscope calibrator to output +DC voltage. Connect the oscilloscope calibrator pulse head output to the CH 2 input Press the CHAN/VERT hardkey, press the CHANNEL softkey as necessary to select CHANNEL 2, and then set CHANNEL 2 to on. Ensure that the offset is set to 0 V Press the Blue, AVG and 2 hardkeys to display the channel 2 average value. Set the oscilloscope calibrator output to ON Set the oscilloscope calibrator to output each of the following positive dc voltages/division and the corresponding number of divisions. For each value, set the scale values as indicated. Wait until the displays a flat line and the numerical value has settled. Then, verify that the displayed value is within the tolerance limits listed. NOTES The scale value is just above the offset value on the display. Pressing the CLEAR DISPLAY hardkey reduces settling time. Oscilloscope Calibrator (V/D) Oscilloscope Calibrator (DIV) Scale (mv/div.) Tolerance Limits (mv dc) to " to " to 560 " 2 " 940 to 1060 " 3 " 1.44 to 1.56 V Set the oscilloscope calibrator to output each of the following positive dc voltages/division and the corresponding number of divisions. For each value, set the offset and scale values as indicated. Wait until the displays a flat line and the numerical value has settled, and then verify that the displayed value is within the tolerance limits listed. 5

9 Oscilloscope Calibrator (V/D) Oscilloscope Calibrator (DIV) Offset (V) Scale (mv/div.) Tolerance Limits (mv dc) to " " 1.0 " to " " 1.3 " to " 4.7 " 4.87 to 5.13 " " 5.0 " 4.87 to 5.13 " " to " 9.4 " 9.75 to " " 10.0 " 9.74 to " " 10.6 " 9.72 to " to 20.5 " " 20.0 " 19.4 to Press the CHAN/VERT hardkey, and then set CHANNEL 2 to off. Press the Blue and CLR MEAS hardkeys Set the oscilloscope calibrator OUTPUT to off Connect the oscilloscope calibrator pulse head output to the CH 3 input. Press the CHAN/VERT hardkey, press the CHANNEL softkey as necessary to select CHANNEL 3, and then set CHANNEL 3 to on. Ensure that the offset is set to 0 V Press the Blue, AVG and 3 hardkeys to display the channel 3 average value. Set the oscilloscope calibrator output to ON Repeat steps through for channel Disconnect the equipment from the. 4.2 HORIZONTAL ACCURACY TESTS Connect the oscilloscope calibrator pulse head output to the CH 2 input. Press the oscilloscope calibrator MARKERS and OUTPUT ON buttons to output horizontal markers Set the oscilloscope calibrator to 10 ns/div and press the AUTOSCALE hardkey. Adjust the scale setting (CHAN/VERT hardkey menu) as necessary to display the entire signal on the screen Press the Blue, PRI (1/PRF) and 2 hardkeys to display the channel 2 period Set the oscilloscope calibrator to output each of the following marker values. For each value, press the MEBASE hardkey and set the MEBASE setting to the given value. Adjust the delay setting if necessary so that the displays only two markers in the middle of the screen. Wait until the numerical value has settled and then verify that the value is within the tolerance limits listed. 6

10 Oscilloscope Calibrator Marker Setting (Per Division) Timebase Setting Tolerance Limits 10 ns 2 ns to ns.1 µs 20 ns to ns 1 µs 200 ns to ns 10 µs 2 µs to µs.1 ms 20 µs to µs 1 ms 200 µs to µs 10 ms 2 ms to ms Set the oscilloscope calibrator OUTPUT to off Connect the oscilloscope calibrator pulse head output to the CH 3 input. Press the oscilloscope calibrator OUTPUT ON button to output horizontal markers Set the oscilloscope calibrator to 10 ns/div and press the AUTOSCALE hardkey. Adjust the scale setting (CHAN/VERT hardkey menu) as necessary to display the entire signal on the screen Press the Blue, PRI (1/PRF) and 3 hardkeys to display the channel 3 period Repeat steps and for channel Set the oscilloscope calibrator output to off and then disconnect the pulse head from the. 4.3 RISE ME TESTS (CHANNELS 2 AND 3) Connect the oscilloscope calibrator pulse head output through a feedthrough termination to the CH 2 input. Press the oscilloscope calibrator FAST EDGE and OUTPUT ON buttons and set it for positive polarity at a frequency of 100 khz Press the AUTOSCALE hardkey, followed by the Blue, RISEME and 2 hardkeys to measure the channel 2 rise time Press the MEBASE hardkey, and set the MEBASE setting to 2 ns/div Ensure that a single rising-edge pulse can be seen on the display. Press the delay softkey and then adjust the delay setting so that the center of the rising edge crosses the third vertical graticule from the left Allow the to settle for 30 seconds or until the < sign with the rise time value disappears. Verify that the rise time value is <5 ns Set the oscilloscope calibrator OUTPUT to off. NOTE In some instances, the < sign may not disappear, indicating that the rise time is too short for the timebase setting. If the < has not disappeared after 30 seconds, ensure that the most stable displayed value is <5 ns Connect the oscilloscope calibrator pulse head through the feedthrough termination to the CH 3 input. Press the oscilloscope calibrator OUTPUT ON. 7

11 4.3.8 Press the AUTOSCALE hardkey, followed by the Blue, RISEME and 3 hardkeys to measure the channel 3 rise time Repeat steps through for channel Set the oscilloscope calibrator output to off and then disconnect the pulse head from the. 4.4 RISE ME TESTS (CHANNELS 1 AND 4) Set the signal generator controls for a 0 dbm output at 5 GHz. Set the signal generator controls for internal pulse modulation, setting the internal pulse rate to 1 khz and pulse width to 500 µs Connect one end of the crystal detector to the signal generator RF output and the other end of the detector through a BNC cable to the oscilloscope CH 1 input. Ensure that the signal generator output is set to on. Press the oscilloscope AUTOSET key Press the oscilloscope SHIFT and ACQUIRE MENU keys, select Mode and Average and then set the average to 16. Press the oscilloscope TRIGGER MENU key and set the edge slope to negative Press the oscilloscope CH 1, MEASURE, and Select Measrment keys; then press more (as necessary) and select FALL ME. NOTE The crystal detector inverts the signal. Thus, the rise time of the signal will be measured as fall time on the oscilloscope Adjust the oscilloscope vertical scale and position as necessary to display full square waves on the screen. Ensure that the oscilloscope C1 Fall measurement does not display the words Low Signal Amplitude The oscilloscope will still display the words Low Resolution. Adjust the oscilloscope horizontal timebase to 50 ns/div and ensure that Low Resolution disappears from the oscilloscope screen. Note the C1 Fall value on the oscilloscope. Round this value to the nearest 0.5 ns and record it as value T (in ns) Calculate and record the maximum rise time from the formula: MAXIMUM RISE ME (ns) = 1 + T where, T (in ns) is the value recorded in step Disconnect the crystal detector and BNC cable from between the oscilloscope and signal generator Connect peak power sensor #1 to the CH 1 input and to the signal generator RF output. Press the AUTOSCALE hardkey Press the CARRIER FREQ hardkey, and then set the CARRIER FREQ softkey selection to CH 1=CH 4. Set the frequency to 5 GHz via the keypad Press the CHAN/VERT hardkey. Adjust the ref. Level softkey selection until the peak of the waveform is between the top two horizontal lines. Press the Blue, RISEME and 1 hardkeys to measure the channel 1 rise time Push the MEBASE hardkey, and then decrease the MEBASE setting until the < sign with the rise time value disappears. Press the delay softkey and then adjust the delay setting as necessary to keep the leading edge of the pulse centered on the display. 8

12 Adjust the delay setting to put the center of the rising edge on the third vertical graticule from the left. Wait for the measurement to settle. Verify that the rise time measurement value displayed on the screen is less than or equal to the maximum rise time calculated in step Press the CHAN/VERT hardkey and then press the bandwidth softkey as necessary to set the bandwidth to low. Press the MEBASE hardkey, and then set the MEBASE setting to 500 ns/div. Verify that the rise time measurement value displayed is <1000 ns Press the CHAN/VERT hardkey and then press the bandwidth softkey as necessary to set the bandwidth to CW. Press the MEBASE hardkey, and then set the MEBASE setting to 50 µs/div. Verify that the rise time measurement value displayed is <100 µs Connect peak power sensor #1 to the CH 4 input and to the signal generator RF output. Press the AUTOSCALE hardkey Press the CHAN/VERT hardkey. Adjust the REF LEVEL softkey selection until the peak of the waveform is between the top two horizontal lines. Press the Blue, RISEME and 4 hardkeys to measure the channel 4 rise time Repeat steps through for channel 4, using peak power sensor # Connect peak power sensor #2 to the CH 1 input and to the signal generator RF output. Press the AUTOSCALE hardkey Press the CHAN/VERT hardkey. Adjust the REF LEVEL softkey selection until the peak of the waveform is between the top two horizontal lines. Press the Blue, RISE ME and 1 hardkeys to measure the channel 1 rise time Repeat steps and for channel 1, using peak power sensor #2, except verify that the rise time is <45 ns, and then repeat steps and Disconnect the peak power sensors (#1 and #2) from the and the signal generator, as applicable. 4.5 TRIGGER SENSIVITY TESTS (CHANNELS 1 AND 4) Set the signal generator controls for a -14 dbm output at 1 GHz. Set the signal generator controls for internal pulse modulation, setting the internal pulse rate to 1 khz and the width to 500 µsec Connect peak power sensor #1 to the CH 1 input and to the signal generator RF output Press the AUTOSCALE hardkey, and then press the CARRIER FREQ hardkey. Set the frequency to 1 GHz via the keypad Press the Blue, TOP and 1 hardkeys to measure the channel 1 top value. Slowly adjust the signal generator output level (if necessary) until the indicates <-15 dbm, allowing time for settling. Verify that the waveform can still be seen on the Press the CHAN/VERT hardkey and set the bandwidth to LOW Set the signal generator controls for a -24 dbm output. Adjust the trigger to about -27 dbm and adjust the reference level until the display is visible and the peak is near the top of the screen Slowly adjust the signal generator output level until the indicates <-25 dbm, allowing time for settling. Verify that the waveform can still be seen on the Connect peak power sensor #1 to the CH 4 input and to the signal generator RF output. Set the signal generator output level to -14 dbm. 9

13 4.5.9 Press the AUTOSCALE hardkey, and then press the CARRIER FREQ hardkey. Set the frequency to 1 GHz via the keypad Press the Blue, TOP and 4 hardkeys to measure the channel 4 top value. Slowly adjust the signal generator output level (if necessary) until the indicates <-15 dbm, allowing time for settling. Verify that the waveform can still be seen on the Press the CHAN/VERT hardkey and set the bandwidth to LOW Set the signal generator controls for a -24 dbm output. Adjust the trigger to about -27 dbm and adjust the reference level until the display is visible and the peak is near the top of the screen Slowly adjust the signal generator output level until the indicates <-25 dbm, allowing time for settling. Verify that the waveform can still be seen on the Disconnect the equipment from the. 4.6 INSTRUMENTAON UNCERTAINTY TESTS Insert the program disk into the instrument controller. Set the instrument controller as appropriate to run the disk Use the instructions provided with the calibration software to connect the to the instrument controller with an IEEE bus cable Follow the instructions in the calibration software to perform the indicated tests, using the calibration cable and the digital multimeter. Verify that all software tests pass Disconnect all equipment from the. 4.7 CHECK SOURCE TESTS Connect the SENSOR CHECK input to the CH 1 input, using peak power sensor # Press the UL hardkey and more softkey, and set the check source to pulse. Press the CARRIER FREQ hardkey and enter a frequency of 1.05 GHz Press the AUTOSCALE hardkey and ensure that a square wave appears on the display. Press the Blue, TOP and 1 hardkeys to measure the channel 1 top value. Verify that the top value is between 9.5 and 10.5 dbm Press the UL hardkey and set the check source to CW. Press the AUTOSCALE hardkey, followed by the Blue, AVG and 1 hardkeys to measure the channel 1 average value. Verify that the displayed average value is between 9.5 and 10.5 dbm Unless other measurements are to be performed, turn all power switches to off or standby and disconnect the equipment from the. 10

14 CALIBRAON CHECKLIST TEST INST (S) Hewlett-Packard 8991AOPT003 Peak Power Analyzer PROC. NO. NA 17-20GP-48 MFR. MODEL SER. NO. PROCEDURE MEASURED VALUES OUT STEP FUNCON TESTED NOMINAL FIRST RUN SECOND RUN OF CALIBRAON TOLERANCES NO. TOL (1) (2) (3) (4) (5) (6) (7) 3.2 (Connector Dimensions) Check Source: C ck( ) " to 0.000" Check Source: C-71B - - ck( ) " to " 3.8 Self test - - ck( ) passed cal 3.10 CH 1 self test - - ck( ) passed cal 3.11 CH 2 self test - - ck( ) passed cal " CH 3 self test - - ck( ) passed cal " CH 4 self test - - ck( ) passed cal 4.1 Vertical Accuracy Tests (Channel 2) mv/div scale 100 mv 88 to 112 mv " 200 " " 200 mv 176 to 224 mv " 500 " " 500 mv 440 to 560 mv " " " " 1000 mv 940 to 1060 mv " " " " 1.5 V 1.44 to 1.56 V Offset: 0.7 V 1.0 V to V " 1 Offset: 1.0 V 1.0 V to V " 1 Offset: 1.3 V 1.0 V to V " 5 Offset: 4.7 V 5.0 V 4.87 to 5.13 V " 5 Offset: 5.0 V 5.0 V 4.87 to 5.13 V " 5 Offset: 5.3 V 5.0 V 4.86 to 5.14 V " 10 Offset: 9.4 V 10.0 V 9.75 to V " 10 Offset: 10.0 V 10.0 V 9.74 to V " 10 Offset: 10.6 V 10.0 V 9.72 to V " 20 Offset: 18.5 V 20.0 V 19.5 to 20.5 V " 20 Offset: 20.0 V 20.0 V 19.4 to 20.6 V Page 1 of 4 11

15 CALIBRAON CHECKLIST TEST INST (S) Hewlett-Packard 8991AOPT003 Peak Power Analyzer PROC. NO. NA 17-20GP-48 MFR. MODEL SER. NO. PROCEDURE MEASURED VALUES OUT STEP FUNCON TESTED NOMINAL FIRST RUN SECOND RUN OF CALIBRAON TOLERANCES NO. TOL (1) (2) (3) (4) (5) (6) (7) (Channel 3) mv/div scale 100 mv 88 to 112 mv " 200 " " 200 mv 176 to 224 mv " 500 " " 500 mv 440 to 560 mv " " " " 1000 mv 940 to 1060 mv " " " " 1.5 V 1.44 to 1.56 V " 1 Offset: 0.7 V 1.0 V to V " 1 Offset: 1.0 V 1.0 V to V " 1 Offset: 1.3 V 1.0 V to V " 5 Offset: 4.7 V 5.0 V 4.87 to 5.13 V " 5 Offset: 5.0 V 5.0 V 4.87 to 5.13 V " 5 Offset: 5.3 V 5.0 V 4.86 to 5.14 V " 10 Offset: 9.4 V 10.0 V 9.75 to V " 10 Offset: 10.0 V 10.0 V 9.74 to V " 10 Offset: 10.6 V 10.0 V 9.72 to V " 20 Offset: 18.5 V 20.0 V 19.5 to 20.5 V " 20 Offset: 20.0 V 20.0 V 19.4 to 20.6 V 4.2 Horizontal Accuracy Tests (Channel 2) (ns) (ns) ns mkrs: 2 ns/div to ".1 µs mkrs: 20 ns/div to " 1 µs mkrs: 20 ns/div to " 10 µs mkrs: 2 µs/div to ".1 ms mkrs: 20 µs/div to " 1 ms mkrs: 20 µs/div to " 10 ms mkrs: 2 ms/div to Page 2 of 4 12

16 CALIBRAON CHECKLIST TEST INST (S) Hewlett-Packard 8991AOPT003 Peak Power Analyzer PROC. NO. NA 17-20GP-48 MFR. MODEL SER. NO. PROCEDURE MEASURED VALUES OUT STEP FUNCON TESTED NOMINAL FIRST RUN SECOND RUN OF CALIBRAON TOLERANCES NO. TOL (1) (2) (3) (4) (5) (6) (7) (Channel 3) (ns) (ns) ns mkrs: 2 ns/div to ".1 µs mkrs: 20ns/div to " 1 µs mkrs: 200 s/div to " 10 µs mkrs: 2 µs/div to ".1 ms mkrs: 20µs/div to " 1 ms mkrs: 200µs/div to " 10 ms mkrs: 2 ms/div to Rise Time Tests (Channels 2 and 3) Channel ck ( ) <5 ns Channel ck ( ) <5 ns 4.4 Rise Time Tests (Channels 1 and 4) (Sensor #1) C1 Fall (T) ck ( ) NA Record fall time meas Max. Rise Time (ns) = " ck ( ) NA Calculate and record 1 + T ck ( ) Ch 1 rise time: BW-high - - ck ( ) < (value of step 4.4.7) Ch 1 rise time: BW-low - - ck ( ) <1000 ns Ch 1 rise time: BW-CW - - ck ( ) <100 µs Ch 4 rise time: BW-high - - ck ( ) < (value of step 4.4.7) " Ch 4 rise time: BW-low - - ck ( ) <1000 ns " Ch 4 rise time: BW-CW - - ck ( ) <100 µs (Sensor #2) Ch 1 rise time: BW-high - - ck ( ) <45 ns " Ch 1 rise time: BW-low - - ck ( ) <1000 ns " Ch 1 rise time: BW-CW - - ck ( ) <100 µs Page 3 of 4 13

17 CALIBRAON CHECKLIST TEST INST (S) Hewlett-Packard 8991AOPT003 Peak Power Analyzer PROC. NO. NA 17-20GP-48 MFR. MODEL SER. NO. PROCEDURE MEASURED VALUES OUT STEP FUNCON TESTED NOMINAL FIRST RUN SECOND RUN OF CALIBRAON TOLERANCES NO. TOL (1) (2) (3) (4) (5) (6) (7) 4.5 Trigger Sensitivity Tests Ch -15 dbm: BW high - - ck( ) Waveform visible Ch -25 dbm: BW low - - ck( ) " Ch -15 dbm: BW high - - ck( ) " Ch -25 dbm: BW low - - ck( ) " 4.6 Instrumentation Uncertainty Tests Software calibration - - ck( ) All tests pass 4.7 Check Source Tests (dbm) (dbm) Pulse: TOP to CW: AVG " " Page 4 of 4 14

CLAMP-ON METERS THIS PUBLICATION SUPERSEDES NAVAIR 17-20AQ-32 DATED 1 OCTOBER 2009

CLAMP-ON METERS THIS PUBLICATION SUPERSEDES NAVAIR 17-20AQ-32 DATED 1 OCTOBER 2009 TECHNICAL MANUAL AQ-32 INSTRUMENT CALIBRATION PROCEDURE CLAMP-ON METERS THIS PUBLICATION SUPERSEDES DATED 1 OCTOBER 2009 DISTRIBUTION STATEMENT C. DISTRIBUTION AUTHORIZED TO U.S. GOVERNMENT AGENCIES AND

More information

DIAL AND VERNIER CALIPERS

DIAL AND VERNIER CALIPERS TECHNICAL MANUAL INSTRUMENT CALIBRATION PROCEDURE DIAL AND VERNIER CALIPERS THIS PUBLICATION SUPERSEDES NAVAIR 17 20MD 07 DATED 1 JUNE 1995 DISTRIBUTION STATEMENT C. DISTRIBUTION AUTHORIZED TO U.S. GOVERNMENT

More information

TECHNICAL MANUAL CALIBRATION PROCEDURE FOR DIGITAL MULTIMETER 34401A (HEWLETT-PACKARD)

TECHNICAL MANUAL CALIBRATION PROCEDURE FOR DIGITAL MULTIMETER 34401A (HEWLETT-PACKARD) TECHNICAL MANUAL CALIBRATION PROCEDURE FOR DIGITAL MULTIMETER 34401A (HEWLETT-PACKARD) This publication replaces T.O. 33K8-4-1029-1 dated 30 May 2000. Distribution Statement - Distribution authorized to

More information

TECHNICAL MANUAL INSTRUMENT CALIBRATION PROCEDURE OPTICAL COMPARATORS THIS PUBLICATION SUPERSEDES NAVAIR 17 20MD 20 DATED 1 FEBRUARY 1988

TECHNICAL MANUAL INSTRUMENT CALIBRATION PROCEDURE OPTICAL COMPARATORS THIS PUBLICATION SUPERSEDES NAVAIR 17 20MD 20 DATED 1 FEBRUARY 1988 TECHNICAL MANUAL INSTRUMENT CALIBRATION PROCEDURE OPTICAL COMPARATORS THIS PUBLICATION SUPERSEDES NAVAIR 17 20MD 20 DATED 1 FEBRUARY 1988 DISTRIBUTION STATEMENT C. DISTRIBUTION AUTHORIZED TO U.S. GOVERNMENT

More information

TECHNICAL MANUAL CALIBRATION PROCEDURE FOR SYNTHESIZED SIGNAL GENERATOR 7200() (GIGA-TRONICS)

TECHNICAL MANUAL CALIBRATION PROCEDURE FOR SYNTHESIZED SIGNAL GENERATOR 7200() (GIGA-TRONICS) T.O. 33K3-4-3051-1 TECHNICAL MANUAL CALIBRATION PROCEDURE FOR SYNTHESIZED SIGNAL GENERATOR 7200() (GIGA-TRONICS) This publication replaces TO. 33K3-4-3051-1 dated 30 July 1997 and Change 1 30 December

More information

DEPARTMENT OF THE ARMY TECHNICAL BULLETIN

DEPARTMENT OF THE ARMY TECHNICAL BULLETIN *TB 9-6625-2333-24 DEPARTMENT OF THE ARMY TECHNICAL BULLETIN CALIBRATION PROCEDURE FOR SPECTRUM ANALYZER AGILENT MODELS 8562EC AND 8562EC-104 Headquarters, Department of the Army, Washington, DC 17 June

More information

Supplement. TDS5032 and TDS5034 Digital Phosphor Oscilloscopes

Supplement. TDS5032 and TDS5034 Digital Phosphor Oscilloscopes TDS5032 and TDS5034 Digital Phosphor Oscilloscopes 071-1316-00 www.tektronix.com 071131600 Copyright Tektronix, Inc. All rights reserved. Tektronix products are covered by U.S. and foreign patents, issued

More information

Frequency and Time Domain Representation of Sinusoidal Signals

Frequency and Time Domain Representation of Sinusoidal Signals Frequency and Time Domain Representation of Sinusoidal Signals By: Larry Dunleavy Wireless and Microwave Instruments University of South Florida Objectives 1. To review representations of sinusoidal signals

More information

Agilent 81133A/81134A

Agilent 81133A/81134A Agilent 81133A/81134A Performance Verification Rev. 2.3, Dec. 2009 Agilent Technologies Introduction Use these tests if you want to check that the Agilent 81133A / 81134A Pulse / Pattern Generator is

More information

DEPARTMENT OF THE ARMY TECHNICAL BULLETIN

DEPARTMENT OF THE ARMY TECHNICAL BULLETIN *TB 9-6625-1914-24 DEPARTMENT OF THE ARMY TECHNICAL BULLETIN CALIBRATION PROCEDURE FOR SPECTRUM ANALYZER, IF, LF, AND RF PLUG-IN UNITS, HEWLETT-PACKARD, MODELS 8552( ), 8553( ), 8554( ), 8555( ), AND 8556(

More information

TABLE OF CONTENTS SECTION I INTRODUCTION AND DESCRIPTION

TABLE OF CONTENTS SECTION I INTRODUCTION AND DESCRIPTION Walter G. LEGGE Co., Inc. 444 Central Ave., Peekskill, NY 10566 tel # 800-345-3443 fax 800-332-2636 email address: info@leggesystems.com website: www.leggesystems.com TABLE OF CONTENTS SECTION PAGE 1 INTRODUCTION

More information

DEPARTMENT OF THE ARMY TECHNICAL BULLETIN

DEPARTMENT OF THE ARMY TECHNICAL BULLETIN *TB 9-6625-2050-24 DEPARTMENT OF THE ARMY TECHNICAL BULLETIN CALIBRATION PROCEDURE FOR POWER METERS, TS-3793 (HEWLETT-PACKARD 436A AND 436AOPT9/22) HEWLETT-PACKARD MODELS 435A, 435B, 435BOPT001, AND 435BE21

More information

PXIe Contents. Required Software CALIBRATION PROCEDURE

PXIe Contents. Required Software CALIBRATION PROCEDURE CALIBRATION PROCEDURE PXIe-5160 This document contains the verification and adjustment procedures for the PXIe-5160. Refer to ni.com/calibration for more information about calibration solutions. Contents

More information

DEPARTMENT OF THE ARMY TECHNICAL BULLETIN

DEPARTMENT OF THE ARMY TECHNICAL BULLETIN *TB 9-6625-2240-35 DEPARTMENT OF THE ARMY TECHNICAL BULLETIN CALIBRATION PROCEDURE FOR OSCILLOSCOPE OS261U (TEKTRONIX, TYPE 475), OS261A(V)1U (TEKTRONIX, TYPE 475 OPTION 7), OS261B(V)1U (TEKTRONIX, TYPE

More information

AP034-OM-E Rev D ISSUED: January 2000 ²

AP034-OM-E Rev D ISSUED: January 2000 ² 3HUIRUPDQFH9HULILFDWLRQ 3HUIRUPDQFH9HULILFDWLRQ This procedure can be used to verify the warranted characteristics of the AP034 Active Differential Probe. The recommended calibration interval for the model

More information

100 Hz to 22. HP 8566B Spectrum Analyzer. Discontinued Product Support Information Only. Outstanding Precision and Capability

100 Hz to 22. HP 8566B Spectrum Analyzer. Discontinued Product Support Information Only. Outstanding Precision and Capability Discontinued Product Support Information Only This literature was published years prior to the establishment of Agilent Technologies as a company independent from Hewlett-Packard and describes products

More information

Calibration Guide. 8590L Spectrum Analyzer

Calibration Guide. 8590L Spectrum Analyzer Calibration Guide 8590L Spectrum Analyzer Manufacturing Part Number: 08590-90315 Supersedes: 08590-90269 Printed in USA April 2001 Copyright 1994-1995, 2000-2001 Agilent Technologies, Inc. The information

More information

PXIe Contents. Required Software CALIBRATION PROCEDURE

PXIe Contents. Required Software CALIBRATION PROCEDURE CALIBRATION PROCEDURE PXIe-5113 This document contains the verification and adjustment procedures for the PXIe-5113. Refer to ni.com/calibration for more information about calibration solutions. Contents

More information

Laboratory Equipment Instruction Manual 2011

Laboratory Equipment Instruction Manual 2011 University of Toronto Department of Electrical and Computer Engineering Instrumentation Laboratory GB341 Laboratory Equipment Instruction Manual 2011 Page 1. Wires and Cables A-2 2. Protoboard A-3 3. DC

More information

Specifications for DS1000CA Series

Specifications for DS1000CA Series Revised December, 2009 RIGOL Specifications for DS1000CA Series All specifications apply to the DS1000CA Series Oscilloscopes unless noted otherwise. To meet these specifications, two conditions must first

More information

SUPERSEDED COPY DATED 15 OCTOBER 1990 DEPARTMENT OF THE ARMY TECHNICAL BULLETIN CALIBRATION PROCEDURE FOR DISTORTION ANALYZER, TS-4084/G

SUPERSEDED COPY DATED 15 OCTOBER 1990 DEPARTMENT OF THE ARMY TECHNICAL BULLETIN CALIBRATION PROCEDURE FOR DISTORTION ANALYZER, TS-4084/G *TB 9-6625-2170-35 SUPERSEDED COPY DATED 15 OCTOBER 1990 DEPARTMENT OF THE ARMY TECHNICAL BULLETIN CALIBRATION PROCEDURE FOR DISTORTION ANALYZER, TS-4084/G Headquarters, Department of the Army, Washington,

More information

Sept 13 Pre-lab due Sept 12; Lab memo due Sept 19 at the START of lab time, 1:10pm

Sept 13 Pre-lab due Sept 12; Lab memo due Sept 19 at the START of lab time, 1:10pm Sept 13 Pre-lab due Sept 12; Lab memo due Sept 19 at the START of lab time, 1:10pm EGR 220: Engineering Circuit Theory Lab 1: Introduction to Laboratory Equipment Pre-lab Read through the entire lab handout

More information

HP 16533A 1-GSa/s and HP 16534A 2-GSa/s Digitizing Oscilloscope

HP 16533A 1-GSa/s and HP 16534A 2-GSa/s Digitizing Oscilloscope User s Reference Publication Number 16534-97009 February 1999 For Safety Information, Warranties, and Regulatory Information, see the pages behind the Index Copyright Hewlett-Packard Company 1991 1999

More information

OSCILLOSCOPES. Oscilloscopes CS-5300 SERIES RS-232C OPTION OPTION CS-5370P/5370/5350 FEATURES OUTLINE. Photo: CS-5370P

OSCILLOSCOPES. Oscilloscopes CS-5300 SERIES RS-232C OPTION OPTION CS-5370P/5370/5350 FEATURES OUTLINE. Photo: CS-5370P Oscilloscopes 100MHz 2-Channel Programmable Oscilloscope ( With Digital Readout / Cursor) CS-5370P CS-5370 100MHz 3-Channel Oscilloscope ( With Digital Readout / Cursor) 50MHz 3-Channel Oscilloscope (

More information

Introduction to Oscilloscopes Instructor s Guide

Introduction to Oscilloscopes Instructor s Guide Introduction to Oscilloscopes A collection of lab exercises to introduce you to the basic controls of a digital oscilloscope in order to make common electronic measurements. Revision 1.0 Page 1 of 25 Copyright

More information

PXIe Contents CALIBRATION PROCEDURE

PXIe Contents CALIBRATION PROCEDURE CALIBRATION PROCEDURE PXIe-5632 This document contains the verification and adjustment procedures for the PXIe-5632 Vector Network Analyzer. Refer to ni.com/calibration for more information about calibration

More information

DIGITAL STORAGE OSCILLOSCOPES

DIGITAL STORAGE OSCILLOSCOPES 99 Washington Street Melrose, MA 02176 Fax 781-665-0780 TestEquipmentDepot.com DIGITAL STORAGE OSCILLOSCOPES Digital Storage Oscilloscope 100MS/s Acquisition (40MS/s 2 Acquisition) 100MHz 2 channel. OUTLINE

More information

ScopeMeter 190 Series II

ScopeMeter 190 Series II ScopeMeter 190 Series II Fluke 190-062, -102, -104, -202, - 204, -502 Service Manual PN 4822 872 05405 July 2011, Rev. 2, March 2012 2011, 2011, 2012 Fluke Corporation, All rights reserved. Printed in

More information

Performance Characteristics

Performance Characteristics The performance characteristics describe the typical performance of the oscilloscope. You will notice that some of the characteristics are marked as tested, these are values that you can verify with the

More information

The University of Jordan Mechatronics Engineering Department Electronics Lab.( ) Experiment 1: Lab Equipment Familiarization

The University of Jordan Mechatronics Engineering Department Electronics Lab.( ) Experiment 1: Lab Equipment Familiarization The University of Jordan Mechatronics Engineering Department Electronics Lab.(0908322) Experiment 1: Lab Equipment Familiarization Objectives To be familiar with the main blocks of the oscilloscope and

More information

DEPARTMENT OF THE ARMY TECHNICAL BULLETIN

DEPARTMENT OF THE ARMY TECHNICAL BULLETIN *TB 9-6625-2072-24 DEPARTMENT OF THE ARMY TECHNICAL BULLETIN CALIBRATION PROCEDURE FOR OSCILLOSCOPE, OS-189A(P), AND HEWLETT- PACKARD, MODELS 180A, 180C AND 180D; DUAL CHANNEL VERTICAL AMPLIFIER, PL-1186/USM-281

More information

Agilent InfiniiVision 4000 X-Series Oscilloscopes. Service Guide

Agilent InfiniiVision 4000 X-Series Oscilloscopes. Service Guide Agilent InfiniiVision 4000 X-Series Oscilloscopes Service Guide s1 Notices Agilent Technologies, Inc. 2008-2013 No part of this manual may be reproduced in any form or by any means (including electronic

More information

Lab 0: Introduction to basic laboratory instruments. Revised by Dan Hoang & Tai-Chang Chen 03/30/2009

Lab 0: Introduction to basic laboratory instruments. Revised by Dan Hoang & Tai-Chang Chen 03/30/2009 Lab 0: Introduction to basic laboratory instruments Revised by Dan Hoang & Tai-Chang Chen 03/30/2009 1. Objectives 1. To learn safety procedures in the laboratory. 2. To learn how to use basic laboratory

More information

Cost-Effective Traceability for Oscilloscope Calibration. Author: Peter B. Crisp Head of Metrology Fluke Precision Instruments, Norwich, UK

Cost-Effective Traceability for Oscilloscope Calibration. Author: Peter B. Crisp Head of Metrology Fluke Precision Instruments, Norwich, UK Cost-Effective Traceability for Oscilloscope Calibration Author: Peter B. Crisp Head of Metrology Fluke Precision Instruments, Norwich, UK Abstract The widespread adoption of ISO 9000 has brought an increased

More information

Agilent 33522A Function Arbitrary Waveform Generator. Tektronix TDS 3012B Oscilloscope

Agilent 33522A Function Arbitrary Waveform Generator. Tektronix TDS 3012B Oscilloscope Agilent 33522A Function/Arbitrary Waveform Generator and Tektronix TDS 3012B Oscilloscope Agilent 33522A Function Arbitrary Waveform Generator The signal source for this lab is the Agilent 33522A Function

More information

3.003 Lab 3 Part A. Measurement of Speed of Light

3.003 Lab 3 Part A. Measurement of Speed of Light 3.003 Lab 3 Part A. Measurement of Speed of Light Objective: To measure the speed of light in free space Experimental Apparatus: Feb. 18, 2010 Due Mar. 2, 2010 Components: 1 Laser, 4 mirrors, 1 beam splitter

More information

Appendix A: Specifications

Appendix A: Specifications All specifications apply to the TDS 200-Series Digital Oscilloscopes and a P2100 probe with the Attenuation switch set to 10X unless noted otherwise. To meet specifications, two conditions must first be

More information

HP 8901B Modulation Analyzer. HP 11722A Sensor Module. 150 khz MHz. 100 khz MHz. Technical Specifications. Four Instruments In One

HP 8901B Modulation Analyzer. HP 11722A Sensor Module. 150 khz MHz. 100 khz MHz. Technical Specifications. Four Instruments In One HP 8901B Modulation Analyzer 150 khz - 1300 MHz HP 11722A Sensor Module 100 khz - 2600 MHz Technical Specifications Four Instruments In One RF Power: ±0.02 db instrumentation accuracy RF Frequency: 10

More information

DEPARTMENT OF THE ARMY TECHNICAL BULLETIN CALIBRATION PROCEDURE FOR SIGNAL GENERATOR SG-1207/U (HEWLETT-PACKARD, MODEL 8642M)

DEPARTMENT OF THE ARMY TECHNICAL BULLETIN CALIBRATION PROCEDURE FOR SIGNAL GENERATOR SG-1207/U (HEWLETT-PACKARD, MODEL 8642M) *TB 9-6625-2182-24 DEPARTMENT OF THE ARMY TECHNICAL BULLETIN CALIBRATION PROCEDURE FOR SIGNAL GENERATOR SG-1207/U (HEWLETT-PACKARD, MODEL 8642M) Headquarters, Department of the Army, Washington, DC 2 July

More information

Model 310H Fast 800V Pulse Generator

Model 310H Fast 800V Pulse Generator KEY FEATURES Temperature Stability +/-5ppm 100 V to 800 V into 50 Ω

More information

User Manual. TDS 420A, TDS 430A, TDS 460A & TDS 510A Digitizing Oscilloscopes

User Manual. TDS 420A, TDS 430A, TDS 460A & TDS 510A Digitizing Oscilloscopes User Manual TDS 420A, TDS 430A, TDS 460A & TDS 510A Digitizing Oscilloscopes 070-9701-03 www.tektronix.com Copyright Tektronix, Inc. All rights reserved. Tektronix products are covered by U.S. and foreign

More information

Implementing Automated Oscilloscope Calibration Systems

Implementing Automated Oscilloscope Calibration Systems This paper was first presented at the National Conference of Standards Laboratories '97, Atlanta, Georgia, USA, on July 28, 1997. Implementing Automated Oscilloscope Calibration Systems Presenter: Richard

More information

User Manual. TDS 420A, TDS 430A, TDS 460A & TDS 510A Digitizing Oscilloscopes

User Manual. TDS 420A, TDS 430A, TDS 460A & TDS 510A Digitizing Oscilloscopes User Manual TDS 420A, TDS 430A, TDS 460A & TDS 510A Digitizing Oscilloscopes 070-9701-02 Copyright Tektronix, Inc. All rights reserved. Tektronix products are covered by U.S. and foreign patents, issued

More information

EENG-201 Experiment # 4: Function Generator, Oscilloscope

EENG-201 Experiment # 4: Function Generator, Oscilloscope EENG-201 Experiment # 4: Function Generator, Oscilloscope I. Objectives Upon completion of this experiment, the student should be able to 1. To become familiar with the use of a function generator. 2.

More information

N acquisitions, all channels simultaneously, N is selectable from 4, 16, 64, and 128 Inputs

N acquisitions, all channels simultaneously, N is selectable from 4, 16, 64, and 128 Inputs With compliments All specifications apply to the TDS 200-Series Digital Real-Time Oscilloscope with a P2100 probe with the Attenuation switch set to 10X unless noted otherwise. To meet specifications,

More information

DEPARTMENT OF THE ARMY TECHNICAL BULLETIN

DEPARTMENT OF THE ARMY TECHNICAL BULLETIN *TB 9-6625-2213-24 DEPARTMENT OF THE ARMY TECHNICAL BULLETIN CALIBRATION PROCEDURE FOR VECTOR VOLTMETER, ME-512/U AND HEWLETT-PACKARD, MODELS 8405A AND 8405A-H16 Headquarters, Department of the Army, Washington,

More information

CERTIFICATE OF CALIBRATION

CERTIFICATE OF CALIBRATION CERTIFICATE OF CALIBRATION Issued by: Pico Technology Ltd. Certificate Number: 9999 of: James House, Colmworth Business Park, St. Neots, Cambridgeshire, Signature: PE19 8YP UNITED KINGDOM Tel: +44 (0)

More information

ECE65 Introduction to the Function Generator and the Oscilloscope Created by: Eldridge Alcantara (Spring 2007)

ECE65 Introduction to the Function Generator and the Oscilloscope Created by: Eldridge Alcantara (Spring 2007) ECE65 Introduction to the Function Generator and the Oscilloscope Created by: Eldridge Alcantara (Spring 2007) I. Getting Started with the Function Generator OUTPUT Red Clip Small Black Clip 1) Turn on

More information

Name EET 1131 Lab #2 Oscilloscope and Multisim

Name EET 1131 Lab #2 Oscilloscope and Multisim Name EET 1131 Lab #2 Oscilloscope and Multisim Section 1. Oscilloscope Introduction Equipment and Components Safety glasses Logic probe ETS-7000 Digital-Analog Training System Fluke 45 Digital Multimeter

More information

DEPARTMENT OF THE ARMY TECHNICAL BULLETIN

DEPARTMENT OF THE ARMY TECHNICAL BULLETIN DEPARTMENT OF THE ARMY TECHNICAL BULLETIN CALIBRATION PROCEDURE FOR FREQUENCY COUNTER HEWLETT-PACKARD, MODEL 5340A AND MICROWAVE FREQUENCY COUNTER TD-1225A(V)1/U (HEWLETT-PACKARD, MODEL 5342A/H14), TD-1225A(V)2/U

More information

Introduction to Basic Laboratory Instruments

Introduction to Basic Laboratory Instruments Introduction to Contents: 1. Objectives... 2 2. Laboratory Safety... 2 3.... 2 4. Using a DC Power Supply... 2 5. Using a Function Generator... 3 5.1 Turn on the Instrument... 3 5.2 Setting Signal Type...

More information

Oscilloscope Calibration Options for Fluke 5500A/5520A Multi-Product Calibrators Extended Specifications

Oscilloscope Calibration Options for Fluke 5500A/5520A Multi-Product Calibrators Extended Specifications Oscilloscope Calibration Options for Fluke 5500A/5520A Multi-Product Calibrators Extended Specifications These specifications apply to the 5520A-SC1100, 5500A-SC600 and 5500A-SC300 Oscilloscope Calibration

More information

MULT SWP X1K K VERN START FREQ DURATION AMPLITUDE 0 TTL OUT RAMP

MULT SWP X1K K VERN START FREQ DURATION AMPLITUDE 0 TTL OUT RAMP Signal Generators This document is a quick reference guide to the operation of the signal generators available in the laboratories. Major functions will be covered, but some features such as their sweep

More information

NI PXIe-5171R. Contents. Required Software CALIBRATION PROCEDURE

NI PXIe-5171R. Contents. Required Software CALIBRATION PROCEDURE CALIBRATION PROCEDURE NI PXIe-5171R This document contains the verification and adjustment procedures for the NI PXIe-5171R (NI 5171R). Refer to ni.com/calibration for more information about calibration

More information

2400C Series Microwave Signal Generators 10 MHz to 40 GHz. Preliminary Technical Datasheet. Low Phase Noise and Fast-Switching Speed in a Single Unit

2400C Series Microwave Signal Generators 10 MHz to 40 GHz. Preliminary Technical Datasheet. Low Phase Noise and Fast-Switching Speed in a Single Unit Preliminary Technical Datasheet 2400C Series Microwave Signal Generators 10 MHz to 40 GHz Low Phase Noise and Fast-Switching Speed in a Single Unit 2400C Series Microwave Signal Generator Signal Generator

More information

Model 855 RF / Microwave Signal Generator

Model 855 RF / Microwave Signal Generator Features Very low phase noise Fast switching Phase coherent switching option 2 to 8 phase coherent outputs USB, LAN, GPIB interfaces Applications Radar simulation Quantum computing High volume automated

More information

8000 SERIES PRECISION MULTIMETER VERIFICATION AND ADJUSTMENT GUIDE

8000 SERIES PRECISION MULTIMETER VERIFICATION AND ADJUSTMENT GUIDE 8000 SERIES PRECISION MULTIMETER VERIFICATION AND ADJUSTMENT GUIDE TRANSMILLE LTD. Version 1.1 : Apr 2015 TABLE OF CONTENTS PREPARING FOR CALIBRATION... 4 INTRODUCTION... 4 CALIBRATION INTERVAL SELECTION...

More information

GT 9000 GT 9000S MICROWAVE

GT 9000 GT 9000S MICROWAVE Page 1 of 6 GT 9000 GT 9000S MICROWAVE Now you can get the performance you need and the capability you want, at a price you can afford. Both the Giga-tronics GT9000 Microwave Synthe- techniques.together,

More information

Instruction Manual. TDS 420A, TDS 430A & TDS 460A Digitizing Oscilloscope Performance Verification and Specifications

Instruction Manual. TDS 420A, TDS 430A & TDS 460A Digitizing Oscilloscope Performance Verification and Specifications Instruction Manual TDS 420A, TDS 430A & TDS 460A Digitizing Oscilloscope Performance Verification and Specifications 070-9705-02 Copyright Tektronix, Inc. All rights reserved. Tektronix products are covered

More information

Agilent X-Series Signal Analyzer This manual provides documentation for the following X-Series Analyzer: CXA Signal Analyzer N9000A

Agilent X-Series Signal Analyzer This manual provides documentation for the following X-Series Analyzer: CXA Signal Analyzer N9000A Agilent X-Series Signal Analyzer This manual provides documentation for the following X-Series Analyzer: CXA Signal Analyzer N9000A N9000A CXA Functional Tests Notices Agilent Technologies, Inc. 2006-2008

More information

*TB DEPARTMENT OF THE ARMY TECHNICAL BULLETIN

*TB DEPARTMENT OF THE ARMY TECHNICAL BULLETIN * DEPARTMENT OF THE ARMY TECHNICAL BULLETIN CALIBRATION PROCEDURE FOR SYNTHESIZER/LEVEL GENERATOR, HEWLETT-PACKARD, MODEL 3335A AND 3335A OPT 001 Headquarters Department of the Army, Washington, DC 18

More information

Test ID 5-15 Utility Line Impedance Test Procedures Guide

Test ID 5-15 Utility Line Impedance Test Procedures Guide Test ID 5-15 Utility Line Impedance Test Procedures Guide Revision 1.1 Tektronix October 13, 2010 Page 2 of 18 Equipment Required Table 1 lists the equipment required to perform the Utility Line Impedance

More information

332:223 Principles of Electrical Engineering I Laboratory Experiment #2 Title: Function Generators and Oscilloscopes Suggested Equipment:

332:223 Principles of Electrical Engineering I Laboratory Experiment #2 Title: Function Generators and Oscilloscopes Suggested Equipment: RUTGERS UNIVERSITY The State University of New Jersey School of Engineering Department Of Electrical and Computer Engineering 332:223 Principles of Electrical Engineering I Laboratory Experiment #2 Title:

More information

Instruction Manual CT-6 High Frequency AC Current Probe

Instruction Manual CT-6 High Frequency AC Current Probe Instruction Manual CT-6 High Frequency AC Current Probe 071-0453-00 Revision A Copyright Tektronix, Inc. All rights reserved. Tektronix products are covered by U.S. and foreign patents, issued and pending.

More information

Agilent 83711B and 83712B Synthesized CW Generators

Agilent 83711B and 83712B Synthesized CW Generators View at www.testequipmentdepot.com Agilent 83711B and 83712B Synthesized CW Generators Agilent 83731B and 83732B Synthesized Signal Generators Data Sheet 10 MHz to 20 GHz 1 to 20 GHz Specifications describe

More information

EXPERIMENT 1 PRELIMINARY MATERIAL

EXPERIMENT 1 PRELIMINARY MATERIAL EXPERIMENT 1 PRELIMINARY MATERIAL BREADBOARD A solderless breadboard, like the basic model in Figure 1, consists of a series of square holes, and those columns of holes are connected to each other via

More information

DEPARTMENT OF THE ARMY TECHNICAL BULLETIN

DEPARTMENT OF THE ARMY TECHNICAL BULLETIN *TB 9-6625-1356-24 DEPARTMENT OF THE ARMY TECHNICAL BULLETIN CALIBRATION PROCEDURE FOR TEST OSCILLATOR, HEWLETT-PACKARD MODELS 651A, 651B AND 652A (SG-763/U) Headquarters Department of the Army, Washington,

More information

P a g e 1 ST985. TDR Cable Analyzer Instruction Manual. Analog Arts Inc.

P a g e 1 ST985. TDR Cable Analyzer Instruction Manual. Analog Arts Inc. P a g e 1 ST985 TDR Cable Analyzer Instruction Manual Analog Arts Inc. www.analogarts.com P a g e 2 Contents Software Installation... 4 Specifications... 4 Handling Precautions... 4 Operation Instruction...

More information

Oscilloscope Calibration Options for 55XX Series Multi-Product Calibrators

Oscilloscope Calibration Options for 55XX Series Multi-Product Calibrators Oscilloscope Calibration Options for 55XX Series Multi-Product Calibrators Extended Specifications These specifications apply to the 5520A-SC1100, 5500A- SC600 and 5500A-SC300 Oscilloscope Calibration

More information

Circuit Characterization with the Agilent 8714 VNA

Circuit Characterization with the Agilent 8714 VNA Circuit Characterization with the Agilent 8714 VNA By: Larry Dunleavy Wireless and Microwave Instruments University of South Florida Objectives 1) To examine the concepts of reflection, phase shift, attenuation,

More information

Signal Generators for Anritsu RF and Microwave Handheld Instruments

Signal Generators for Anritsu RF and Microwave Handheld Instruments Measurement Guide Signal Generators for Anritsu RF and Microwave Handheld Instruments BTS Master Spectrum Master Tracking Generator Option 20 Vector signal Generator Option 23 Anritsu Company 490 Jarvis

More information

Help Volume Agilent Technologies. All rights reserved. Instrument: Agilent Technologies 16533/34A Digitizing Oscilloscope

Help Volume Agilent Technologies. All rights reserved. Instrument: Agilent Technologies 16533/34A Digitizing Oscilloscope Help Volume 1992-2002 Agilent Technologies. All rights reserved. Instrument: Agilent Technologies 16533/34A Digitizing Oscilloscope Agilent Technologies 16533/34A Digitizing Oscilloscope The Agilent Technologies

More information

Advanced Test Equipment Rentals ATEC (2832)

Advanced Test Equipment Rentals ATEC (2832) Established 1981 Advanced Test Equipment Rentals www.atecorp.com 800-404-ATEC (2832) INSTRUCTION MANUAL MODEL 960 SERIES MICRO SWEEP 1 TO 18.0 GHz MICROWAVE GENERATORS 1985 Wavetek THIS DOCUMENT CONTAINS

More information

Data Sheet. Digital Storage Oscilloscope. Features & Benefits. Applications. Ease-of-Use Feature DSO5202BMT DSO5102BMT DSO5062BMT

Data Sheet. Digital Storage Oscilloscope. Features & Benefits. Applications. Ease-of-Use Feature DSO5202BMT DSO5102BMT DSO5062BMT Data Sheet Digital Storage Oscilloscope DSO5202BMT DSO5102BMT DSO5062BMT Features & Benefits 200/100/60MHz Bandwidths 1GSa/s Real Time Sample Rate 2M Memory Depth Trigger mode: Edge, Pulse Width, Video,

More information

OSCILLOSCOPES. Oscilloscopes CS-5400 SERIES CS-5400/5450 FEATURES OUTLINE CS-5400

OSCILLOSCOPES. Oscilloscopes CS-5400 SERIES CS-5400/5450 FEATURES OUTLINE CS-5400 99 Washington Street Melrose, MA 02176 Fax 781-665-0780 TestEquipmentDepot.com Oscilloscopes 100MHz 3-Channel Oscilloscope (With Digital Readout / Cursor) CS-5400 100MHz 3-Channel Oscilloscope CS-5405

More information

DEPARTMENT OF THE ARMY TECHNICAL BULLETIN

DEPARTMENT OF THE ARMY TECHNICAL BULLETIN *TB 9-6625-2215-24 DEPARTMENT OF THE ARMY TECHNICAL BULLETIN CALIBRATION PROCEDURE FOR MEASURING SYSTEM, TD-1209/U (HEWLETT-PACKARD, MODEL 5300A); 50 MHz UNIVERSAL COUNTER, TD-1211/U (HEWLETT-PACKARD,

More information

Arbitrary/Function Waveform Generators 4075B Series

Arbitrary/Function Waveform Generators 4075B Series Data Sheet Arbitrary/Function Waveform Generators Point-by-Point Signal Integrity The Arbitrary/Function Waveform Generators are versatile high-performance single- and dual-channel arbitrary waveform generators

More information

Key Reference. Agilent Technologies E8663B Analog Signal Generator

Key Reference. Agilent Technologies E8663B Analog Signal Generator Agilent Technologies E8663B Analog Signal Generator For the latest revision of this guide, go to http://www.agilent.com/find/e8663b and click Product Library. Manufacturing Part Number: E8663-90004 Printed

More information

Introduction to basic laboratory instruments

Introduction to basic laboratory instruments Introduction to basic laboratory instruments 1. OBJECTIVES... 2 2. LABORATORY SAFETY... 2 3. BASIC LABORATORY INSTRUMENTS... 2 4. USING A DC POWER SUPPLY... 2 5. USING A FUNCTION GENERATOR... 3 5.1 TURN

More information

Agilent 8360B Series Synthesized Swept Signal Generators 8360L Series Synthesized Swept CW Generators Data Sheet

Agilent 8360B Series Synthesized Swept Signal Generators 8360L Series Synthesized Swept CW Generators Data Sheet Agilent 8360B Series Synthesized Swept Signal Generators 8360L Series Synthesized Swept CW Generators Data Sheet 10 MHz to 110 GHz Specifications apply after full user calibration, and in coupled attenuator

More information

Combinational logic: Breadboard adders

Combinational logic: Breadboard adders ! ENEE 245: Digital Circuits & Systems Lab Lab 1 Combinational logic: Breadboard adders ENEE 245: Digital Circuits and Systems Laboratory Lab 1 Objectives The objectives of this laboratory are the following:

More information

Traceability for Oscilloscopes and Oscilloscope Calibrators

Traceability for Oscilloscopes and Oscilloscope Calibrators Traceability for Oscilloscopes and Oscilloscope Calibrators in relation to RF Voltage measurements Paul C. A. Roberts Fluke Precision Measurement PCAR Traceability for Scope Cal Mar 2006 1 Introduction

More information

Specifications. Specifications and Characteristics Specifications

Specifications. Specifications and Characteristics Specifications Specifications and Specifications Specifications All specifications are warranted. Specifications are valid after a 30-minute warm-up period and ±5 C from last calibration temperature. Bandwidth (-3dB)

More information

Project Description and Guidelines

Project Description and Guidelines EE 351 Project Due Friday, Apr. 30, 2010 Project Description and Guidelines For this project your team is required to build and characterize an antenna (half-wavelength, waveguide, etc.) that will operate

More information

DS1000B Series Digital Oscilloscopes

DS1000B Series Digital Oscilloscopes Product Overview DS1000B series oscilloscopes are designed with four analog channels and 1 external trigger channel, which can capture multi-channel signal simultaneously and meet industrial needs. The

More information

Laboratory 1 Generating and viewing signals rev. 20e. Oscilloscope Settings

Laboratory 1 Generating and viewing signals rev. 20e. Oscilloscope Settings 1 2 Laboratory 1 Generating and viewing signals rev. 20e Purpose: Familiarization with the basic functions of an oscilloscope and of a signal generator. Adjusting and measuring specific parameters of signals.

More information

Spectrum Analyzer R&S FS300

Spectrum Analyzer R&S FS300 Spectrum Analyzer R&S FS300 9 khz to 3 GHz The new product family from Rohde & Schwarz Professional test equipment for laboratory, service and production The R&S FS300 is a highly accurate spectrum analyzer

More information

Exercise 6. Range and Angle Tracking Performance (Radar-Dependent Errors) EXERCISE OBJECTIVE

Exercise 6. Range and Angle Tracking Performance (Radar-Dependent Errors) EXERCISE OBJECTIVE Exercise 6 Range and Angle Tracking Performance EXERCISE OBJECTIVE When you have completed this exercise, you will be familiar with the radardependent sources of error which limit range and angle tracking

More information

Oscilloscope Calibration Options for Fluke 5500A/5520A Multi-Product Calibrators

Oscilloscope Calibration Options for Fluke 5500A/5520A Multi-Product Calibrators Oscilloscope Calibration Options for Fluke 5500A/5520A Multi-Product Calibrators Extended Specifications November 1999 General Specifications These specifications apply to the 5520A-SC1100, 5500A-SC600

More information

DSM303-V4 3.0 GHz Arbitrary Frequency Chirping Module

DSM303-V4 3.0 GHz Arbitrary Frequency Chirping Module DSM303-V4 3.0 GHz Arbitrary Frequency Chirping Module PRODUCT DESCRIPTION The DSM303-V4 module generates arbitrary frequency chirping CW with frequency update rates up to 312.5 updates/microsecond (1/8

More information

Agilent N9923A FieldFox RF Vector Network Analyzer 2 MHz to 4/6 GHz. Data Sheet

Agilent N9923A FieldFox RF Vector Network Analyzer 2 MHz to 4/6 GHz. Data Sheet Agilent N9923A FieldFox RF Vector Network Analyzer 2 MHz to 4/6 GHz Data Sheet Table of Contents Definitions... 2 FieldFox RF Vector Network Analyzer... 3 Cable and Antenna Analyzer (Option 305)... External

More information

Tektronix: Products > AWG 2040 Arbitrary Waveform Generator

Tektronix: Products > AWG 2040 Arbitrary Waveform Generator Page 1 of 7 Arbitrary Waveform Generator AWG 2040 This product is no longer carried in our catalog. Features 1.024 GS/sec Clock Rate Provides up to 500 MHz Waveforms 1 MB Record Length (4 MB with Opt.

More information

HP 33120A Function Generator / Arbitrary Waveform Generator

HP 33120A Function Generator / Arbitrary Waveform Generator Note: Unless otherwise indicated, this manual applies to all Serial Numbers. The HP 33120A is a high-performance 15 MHz synthesized function generator with built-in arbitrary waveform capability. Its combination

More information

Faculty of Engineering, Thammasat University

Faculty of Engineering, Thammasat University Faculty of Engineering, Thammasat University Experiment 6: Oscilloscope (For room 506) Objectives: 1. To familiarize you with the Oscilloscope and Function Generator User Manual: Oscilloscope 1 5 9 4 7

More information

Advanced Test Equipment Rentals ATEC (2832)

Advanced Test Equipment Rentals ATEC (2832) Established 1981 Advanced Test Equipment Rentals www.atecorp.com 800-404-ATEC (2832) Agilent 2-Port and 4-Port PNA-X Network Analyzer N5249A - 10 MHz to 8.5 GHz N5241A - 10 MHz to 13.5 GHz N5242A - 10

More information

LAB I. INTRODUCTION TO LAB EQUIPMENT

LAB I. INTRODUCTION TO LAB EQUIPMENT 1. OBJECTIVE LAB I. INTRODUCTION TO LAB EQUIPMENT In this lab you will learn how to properly operate the oscilloscope Agilent MSO6032A, the Keithley Source Measure Unit (SMU) 2430, the function generator

More information

Agilent Technologies PSA Series Spectrum Analyzers Test and Adjustment Software

Agilent Technologies PSA Series Spectrum Analyzers Test and Adjustment Software Test System Overview Agilent Technologies PSA Series Spectrum Analyzers Test and Adjustment Software Test System Overview The Agilent Technologies test system is designed to verify the performance of the

More information

WaveStation Function/Arbitrary Waveform Generators

WaveStation Function/Arbitrary Waveform Generators Function/Arbitrary Waveform Generators Key Features High performance with 14-bit waveform generation, up to 500 MS/s sample rate and up to 512 kpts memory 2 channels on all models Large color display for

More information

CHANGE 1 DEPARTMENT OF THE ARMY TECHNICAL BULLETIN

CHANGE 1 DEPARTMENT OF THE ARMY TECHNICAL BULLETIN TB 9-6625-011-24 CHANGE 1 DEPARTMENT OF THE ARMY TECHNICAL BULLETIN CALIBRATION PROCEDURE FOR TRUE RMS VOLTMETER HEWLETT- PACKARD, MODEL 3400A Headquarters, Department of the Army, Washington, DC 18 September

More information

LeCroy 9304A, 9304AM Digital Oscilloscopes 200 MHz Bandwidth, 100 MS/s. Main Features

LeCroy 9304A, 9304AM Digital Oscilloscopes 200 MHz Bandwidth, 100 MS/s. Main Features LeCroy 9304A, 9304AM Digital Oscilloscopes 200 MHz Bandwidth, 100 MS/s Main Features Four Channels 50k and 200k Point Records DOS Compatible Floppy Disk, PCMCIA portable hard drive and Memory Card Options

More information