Implementing Automated Oscilloscope Calibration Systems

Size: px
Start display at page:

Download "Implementing Automated Oscilloscope Calibration Systems"

Transcription

1 This paper was first presented at the National Conference of Standards Laboratories '97, Atlanta, Georgia, USA, on July 28, Implementing Automated Oscilloscope Calibration Systems Presenter: Richard Roddis Wavetek Ltd. Norwich NR6 6JB, UK Phone Fax Author: Richard Roddis, Wavetek Ltd. Abstract Automating oscilloscope calibration is not easy. Most oscilloscopes have multiple input channels and require a wide range of calibration waveforms. In the past, routing appropriate signals to these inputs has meant the use of a complex computer-controlled signal multiplexer and a large number of interconnecting leads. Both the multiplexer and the leads contribute additional uncertainties to the calibration system. These uncertainties must be evaluated and documented in order to maintain traceability. This paper describes an alternative approach to automated oscilloscope calibration, using fully characterized active heads that deliver all the necessary calibration waveforms directly to the oscilloscope's inputs. Because each head carries its own calibration data, all uncertainties are automatically included in the calibration system's uncertainty calculations. Introduction The increasing accuracy and complexity of modern oscilloscopes, coupled with the requirement of quality standards such as ISO9000 to have them routinely calibrated, is proving a major challenge for calibration laboratories. Although a high proportion of oscilloscopes with bandwidths below 100 MHz still remain uncalibrated, over 90% of the new breed of 400 MHz to 500 MHz scopes (now recommended as the standard workhorse oscilloscope for bench-top use), and virtually all oscilloscopes above 500 MHz, are calibrated. In addition, the vast majority of oscilloscopes sold today are digital-storage models. Because of their higher accuracy these digital storage scopes are increasingly being used to make critical measurements in real production processes, making the need for routine calibration even more important. With manual scope calibration requiring the skills of an experienced calibration engineer and typically taking several hours to complete, calibration laboratories face serious throughput problems. The length of time taken to calibrate modern high-performance oscilloscopes is not only the result of their functional complexity. It has also been a consequence of the lack of fully automated calibration systems. 1

2 Waveform Diversity Meets Scope Cal Needs Calibration of even basic oscilloscope functions, such as vertical and horizontal gain, vertical and horizontal bandwidth, timebase accuracy and trigger sensitivity, requires the use of a large number of calibration waveforms. Vertical and horizontal gain are normally calibrated using lowfrequency (typically 1 khz) square waves or DC levels. Square waves with amplitude accuracies of around 0.25% have traditionally been used for analog oscilloscopes because they allow DC offsets to be calibrated out. DC levels with accuracies better than 0.1% have been introduced to meet the higher accuracy needs of digital-storage oscilloscopes, many of which now interpolate to 14-bit vertical resolution. As oscilloscopes begin to offer higher resolutions, better accuracies will be required. To calibrate oscilloscope bandwidth a frequency response test using leveled sinewaves is normally used. This involves adjustment of the sinewave amplitude at a low reference frequency so that the trace occupies a given number of graticule divisions (typically 6 divisions), after which the frequency is increased until the displayed amplitude falls by 29% (typically from 6 to 4.2 divisions). The frequency at which this occurs is the oscilloscope's 3dB point (see Figure 1). Although it provides a direct measurement of bandwidth that can be checked against the scope manufacturer's published specifications, leveled sinewave testing provides no information about the phase response of the scope's input amplifiers. The observed 3dB point cannot therefore be used to predict the oscilloscope's risetime capabilities when it is subjected to a step (high-speed edge) input. Amplitude set to 6 divisions at reference frequency (typically 50 khz, 1 MHz or 6 MHz). 3-dB bandwidth indicated by upper frequency at which amplitude falls to 4.2 divisions. Figure 1 3dB bandwidth testing using leveled sinewaves For this reason, some oscilloscope manufacturers also include a separate pulse-response test in their calibration procedures. Because a 150 ps risetime pulse contains frequency components up 2

3 to and beyond a 1 GHz scope's 3 db point, the ability of the scope to faithfully display the pulse is a good indication of both its gain and phase response. Pulse performance is normally assessed by measuring the risetime, overshoot, undershoot and ringing that appears in the trace (see Figure 2.) Ringing Overshoot Undershoot 10% to 90% Risetime Figure 2 Typical pulse response waveform and aberrations Although leveled sinewave testing is not a good indication of pulse performance, pulse response testing does provide a good indication of bandwidth. The bandwidth can be deduced from the observed 10% to 90% risetime using the equation:- Bandwidth (MHz) = C / (t r1 2 - t r22 ) where t r1 is the observed trace risetime in nanoseconds, t r2 is the risetime of the calibrator output in nanoseconds, and C is a constant that depends on the high-frequency roll-off of the oscilloscope's input amplifier. This roll-off constant is typically 350 or 450, the appropriate value often being found in the oscilloscope's servicing manuals. If you use the above equation to determine bandwidth, it is important that the risetime of the calibrator output is accurately defined. For example, if a calibrator with a specified output risetime of /-100 ps is used to determine the bandwidth of a 1 GHz scope, the result could be in error by as much as 300 MHz. The calibrator output risetime needs to be much more tightly specified for this equation to be of use. Wavetek's Model 9500 calibrator, for example, has its actual risetime displayed on the calibrator screen. This risetime is specified to within a few picoseconds. 3

4 In addition to the above analog waveforms, oscilloscope calibration also requires the use of precision timing markers to calibrate timebase accuracy. The triangular timing markers traditionally used to calibrate analog oscilloscopes are not, however, suitable for calibrating digital storage types. This is because a digital storage oscilloscope's asynchronous sampling technique may miss the peak of the triangle, making it difficult to align the timing markers accurately. Crystal controlled squarewaves or rectangular pulses are necessary for calibrating a digital storage scope's timebases, and can also be used to calibrate analog scopes. Additional waveforms are required to calibrate other scope functions such as trigger sensitivity and TV sync separation (for example, TV sync separation requires a composite-sync baseband video signal). To meet ISO9000 recommendations, which encourage the traceable verification of all instrument ranges and functions, calibration waveforms should be available at amplitudes that can exercise all an oscilloscope's VOLT/DIV ranges and multipliers. This typically means that the dynamic range of calibration waveforms should extend from a few mv to ±200V pk-pk. Problems with Connecting Cables Existing automated calibration systems typically employ many different pieces of equipment (DC calibrators, calibration waveform generators, RF signal generators, power meters, pulse generators, video waveform generators, etc.) to generate the necessary waveforms, and then route these via a signal multiplexer to the oscilloscope under test. This multiple-instrument arrangement not only increases the cost, complexity and size of the system (it is likely to occupy most of the space available in a 6-foot high equipment rack), it also introduces significant errors into the overall uncertainty equation, all of which must be traceably identified and taken into account. The multiplexer itself, and the cabling used between the calibration sources and the multiplexer and between the multiplexer and the oscilloscope, can cause significant errors. For example, as little as 1 metre of coaxial cabling between the output of a 150 ps risetime pulse generator and an oscilloscope's input connector will introduce waveform aberrations that will add to those caused by the oscilloscope you are trying to calibrate. Reflections caused by the inevitable mismatch at either end of the cable (i.e. between the cable and the oscilloscope's input and the cable and the pulse generator's output) will arrive back at the oscilloscope input some 10 ns to 15 ns after the original edge. As a result, the reflected signal aberration may be superimposed on the overshoot and ringing you are trying to measure at a point where the oscilloscope's aberration specifications are tight (see Figure 3a). In such circumstances, it is almost impossible to determine how much of the overshoot and ringing is due to the cable and how much is due to the scope. To make matters worse, cables are susceptible to damage and contact degradation (for example, where the cable terminates inside the connectors), which can further complicate the situation by causing additional mismatches and consequential reflections. Oscilloscope inputs never provide perfect 50 Ω loading over their entire bandwidth (their typical VSWR can be as high as 1.6), and it is therefore almost impossible to prevent signal reflections from occurring. The only way to overcome the potential problem they represent is to keep the signal path between the pulse generator and the oscilloscope input as short as possible, preferably no more than 60 to 70 mm in length for a 150 ps risetime pulse. In this way, reflections arrive 4

5 back at the oscilloscope input very close to the trace's initial overshoot and ringing, where the oscilloscope's aberration specifications are widest (see Figure 3b). As a result they represent a much smaller proportion of the overall accuracy specification, and test uncertainty ratios are maintained. The oscilloscope's input amplifier is now subjected to all the energy contained in the pulse edge during a single short burst, rather than receiving some of it in short bursts as reflections run up and down the cable. Figure 3a Long connecting cables result in reflections appearing where scope aberration specifications are tightest Figure 3b Short signal path results in reflections appearing where scope aberration specifications are widest, maintaining test uncertainty ratios. 5

6 The source and load mismatch at either end of connecting cables can also cause problems when delivering leveled sinewaves to an oscilloscope's inputs. One metre of coaxial cable between a sinewave generator and a scope input with reactive mismatch VSWRs of 1.2 at each end will result in the sinewave amplitude at the scope input changing by up to 0.9% every time the sinewave frequency hits a multiple of the cable's 1/4 wavelength. For this 1-metre cable, the sinewave amplitude would ripple up and down five times during a 0 to 500 MHz frequency scan (see Figure 4). When driving pulses or squarewaves into an oscilloscope input, connecting cables also cause an effect called 'dribble-up' a loss that increases the amount of time taken for the waveform to settle to its final amplitude. ERROR (%) 1.0 CABLE LENGTH = 1 METRE SOURCE OUTPUT VSWR = 1.2 SCOPE INPUT VSWR = FREQUENCY (MHz) Figure 4 Amplitude changes due to cable length and VSWR Active Heads Ease Reflection Problems All the requirements of high-speed pulse generation and delivery can be met by the use of active pulse-forming heads that plug directly into the oscilloscope's input connectors. These heads can generate high-speed edges within a few millimetres of the scope inputs and deliver them via a carefully controlled transmission line. Unlike a connecting cable, this transmission path is fully protected against accidental damage. The active head principle is not new, and has been used extensively by Tektronix in commercial high-performance pulse generators and by NIST in its reference pulse systems. The 150 ps return-to-zero pulses generated by the active heads of Wavetek's Model 9500 Oscilloscope Calibrator are illustrated in Figure 5. Figure 6 illustrates how the accurate 50-Ω source impedance of these heads absorbs reflected signals caused by an oscilloscope's input mismatch (the input mismatch used in this test was the same as that for Figures 3a and 3b). 6

7 3-volt falling edge (t f = 141 ps) Aberrations at 2%/div Figure 5 The 150ps return-to-zero pulses generated by the active heads of Wavetek's Model 9500 oscilloscope calibrator (measured over a 22 GHz bandwidth using a HP54750) 7

8 Figure 6 The effectiveness of the accurate 50-Ω source impedance of Model 9500 active heads in absorbing reflections from an oscilloscope input mismatch (cf. Figures 3a and 3b) If high-bandwidth amplitude sensing circuitry is also built into the head, the amplitude of leveled sinewave outputs can be accurately controlled very close to the oscilloscope input, reducing the amplitude ripple that occurs with interconnecting cables due to source and load mismatch. The amplitude of DC and squarewave outputs can also be 4-wire sensed at the oscilloscope input. The use of an active head has further advantages. For example, if the attenuators required to generate low-level calibration signals are incorporated into the head, the head itself can be driven with high-level signals. This dramatically reduces the effect of any interference (for example, common-mode noise, microphony or RFI) that is picked up by mainframe to head cabling. Active heads, especially when more than one can be driven at the same time, are also the key to fully automated oscilloscope calibration. Activation of individual heads to drive the oscilloscope with all the necessary calibration waveforms and trigger signals can be performed by the calibrator, eliminating the need for an external multiplexer. In addition, active heads can also include 50-Ω terminators that are switched in whenever the active heads need to drive leveled sinewaves or pulse waveforms into an oscilloscope that has high impedance inputs. The ability of Wavetek's Model 9500 Oscilloscope Calibrator to measure an oscilloscope's input resistance and capacitance means that high impedance inputs can be detected automatically. Heads Ease Support Problems The use of active heads can also improve the serviceability of an oscilloscope calibration system when it becomes necessary to recalibrate the calibrator against higher order metrology standards. 8

9 Traceably calibrating 1.1 GHz leveled sinewaves or 150 picosecond risetime edges requires sophisticated equipment which is beyond the reach of most users. Calibrators that generate these waveforms entirely within their mainframe unit will therefore need to be returned to a suitably equipped calibration laboratory for recalibration, resulting in considerable system downtime. However, for calibrators that generate these signals within active heads, only the heads themselves will require sending away for recalibration. DC and low-frequency functions contained in the mainframe unit can be calibrated to sufficient accuracy on-site, often only requiring the use of a long-scale length DMM. If a spare head is substituted for the one that is away being calibrated, system downtime can be eliminated. In Wavetek's Model 9500 Oscilloscope Calibrator, each active head carries with it its own calibration data. This data is interrogated by the Model 9500 each time a new active head is plugged into it so that the head's calibration uncertainties can automatically be included into the overall uncertainty equation. Conclusion By choosing the right equipment, fully automated oscilloscope calibration is not only achievable, it can also be made simple. The availability of fully integrated calibrators that generate all the required waveforms simplifies system design and reduces support costs. Accurate delivery of these waveforms to the oscilloscope's input connectors via active heads reduces systematic errors and conforms to best-practice metrology techniques. References Bargroff, Keith P, 'New Techniques in Automated, High Performance, Oscilloscope Calibration', Cal Lab, Volume 3, No. 3, May/June 1996, p17 to p22. Roddis, Richard, 'Innovative Oscilloscope Calibrator Architecture', NCSL Proceedings, 1996, p467 to p476. Crisp, Peter, 'Significant Advances in the Provision of Traceability to Oscilloscope Calibrators', NCSL Proceedings, 1996, p391 to p395. Johnson, Mitch, 'Simple Tests for Evaluating Digital Oscilloscopes', Cal Lab, Volume 4, No. 2, March/April 1997, p28 to p32. 9

Cost-Effective Traceability for Oscilloscope Calibration. Author: Peter B. Crisp Head of Metrology Fluke Precision Instruments, Norwich, UK

Cost-Effective Traceability for Oscilloscope Calibration. Author: Peter B. Crisp Head of Metrology Fluke Precision Instruments, Norwich, UK Cost-Effective Traceability for Oscilloscope Calibration Author: Peter B. Crisp Head of Metrology Fluke Precision Instruments, Norwich, UK Abstract The widespread adoption of ISO 9000 has brought an increased

More information

Oscilloscope Calibrator

Oscilloscope Calibrator Header The Fluke 9500B Oscilloscope Calibrator The world s highest performance, fully automated, upgradeable oscilloscope calibration workstation * 9500B Calibrator The world s highest performance, fully

More information

PXIe Contents. Required Software CALIBRATION PROCEDURE

PXIe Contents. Required Software CALIBRATION PROCEDURE CALIBRATION PROCEDURE PXIe-5160 This document contains the verification and adjustment procedures for the PXIe-5160. Refer to ni.com/calibration for more information about calibration solutions. Contents

More information

SCOPE OF ACCREDITATION TO ISO/IEC 17025:2005 & ANSI/NCSLI Z

SCOPE OF ACCREDITATION TO ISO/IEC 17025:2005 & ANSI/NCSLI Z SCOPE OF ACCREDITATION TO ISO/IEC 17025:2005 & ANSI/NCSLI Z540-1-1994 TEKTRONIX, INC. 13725 SW Karl Braun Drive Beaverton, OR 97077 Andrew Mumford Phone: 503 627 2476 CALIBRATION Valid to: August 31, 2018

More information

Traceability for Oscilloscopes and Oscilloscope Calibrators

Traceability for Oscilloscopes and Oscilloscope Calibrators Traceability for Oscilloscopes and Oscilloscope Calibrators in relation to RF Voltage measurements Paul C. A. Roberts Fluke Precision Measurement PCAR Traceability for Scope Cal Mar 2006 1 Introduction

More information

Experiment 1: Instrument Familiarization (8/28/06)

Experiment 1: Instrument Familiarization (8/28/06) Electrical Measurement Issues Experiment 1: Instrument Familiarization (8/28/06) Electrical measurements are only as meaningful as the quality of the measurement techniques and the instrumentation applied

More information

Experiment 1: Instrument Familiarization

Experiment 1: Instrument Familiarization Electrical Measurement Issues Experiment 1: Instrument Familiarization Electrical measurements are only as meaningful as the quality of the measurement techniques and the instrumentation applied to the

More information

Improving TDR/TDT Measurements Using Normalization Application Note

Improving TDR/TDT Measurements Using Normalization Application Note Improving TDR/TDT Measurements Using Normalization Application Note 1304-5 2 TDR/TDT and Normalization Normalization, an error-correction process, helps ensure that time domain reflectometer (TDR) and

More information

Introduction to Basic Laboratory Instruments

Introduction to Basic Laboratory Instruments Introduction to Contents: 1. Objectives... 2 2. Laboratory Safety... 2 3.... 2 4. Using a DC Power Supply... 2 5. Using a Function Generator... 3 5.1 Turn on the Instrument... 3 5.2 Setting Signal Type...

More information

Oscilloscope Calibration Options for Fluke 5500A/5520A Multi-Product Calibrators Extended Specifications

Oscilloscope Calibration Options for Fluke 5500A/5520A Multi-Product Calibrators Extended Specifications Oscilloscope Calibration Options for Fluke 5500A/5520A Multi-Product Calibrators Extended Specifications These specifications apply to the 5520A-SC1100, 5500A-SC600 and 5500A-SC300 Oscilloscope Calibration

More information

RF and Microwave Power Sensors/Meters Tektronix PSM3000, PSM4000, and PSM5000 Series Data Sheet

RF and Microwave Power Sensors/Meters Tektronix PSM3000, PSM4000, and PSM5000 Series Data Sheet 99 Washington Street Melrose, MA 02176 Phone 781-665-1400 Toll Free 1-800-517-8431 Visit us at www.testequipmentdepot.com RF and Microwave Sensors/Meters Tektronix PSM3000, PSM4000, and PSM5000 Series

More information

Oscilloscope Calibration Options for 55XX Series Multi-Product Calibrators

Oscilloscope Calibration Options for 55XX Series Multi-Product Calibrators Oscilloscope Calibration Options for 55XX Series Multi-Product Calibrators Extended Specifications These specifications apply to the 5520A-SC1100, 5500A- SC600 and 5500A-SC300 Oscilloscope Calibration

More information

Oscilloscope Calibration Options for Fluke 5500A/5520A Multi-Product Calibrators

Oscilloscope Calibration Options for Fluke 5500A/5520A Multi-Product Calibrators Oscilloscope Calibration Options for Fluke 5500A/5520A Multi-Product Calibrators Extended Specifications November 1999 General Specifications These specifications apply to the 5520A-SC1100, 5500A-SC600

More information

TG550 AIM & THURLBY THANDAR INSTRUMENTS. 5MHz function generator with sweep

TG550 AIM & THURLBY THANDAR INSTRUMENTS. 5MHz function generator with sweep AIM & THURLBY THANDAR INSTRUMENTS TG550 5MHz function generator with sweep simultaneous display of frequency & amplitude crystal controlled digital frequency locking seven digit external frequency counter

More information

CHAPTER 6. Motor Driver

CHAPTER 6. Motor Driver CHAPTER 6 Motor Driver In this lab, we will construct the circuitry that your robot uses to drive its motors. However, before testing the motor circuit we will begin by making sure that you are able to

More information

Laboratory Exercise 6 THE OSCILLOSCOPE

Laboratory Exercise 6 THE OSCILLOSCOPE Introduction Laboratory Exercise 6 THE OSCILLOSCOPE The aim of this exercise is to introduce you to the oscilloscope (often just called a scope), the most versatile and ubiquitous laboratory measuring

More information

APPENDIX D DISCUSSION OF ELECTRONIC INSTRUMENTS

APPENDIX D DISCUSSION OF ELECTRONIC INSTRUMENTS APPENDIX D DISCUSSION OF ELECTRONIC INSTRUMENTS DC POWER SUPPLIES We will discuss these instruments one at a time, starting with the DC power supply. The simplest DC power supplies are batteries which

More information

Data Sheet. Peak, CW & Average. Power Sensors. Taking performance to a new peak

Data Sheet. Peak, CW & Average. Power Sensors. Taking performance to a new peak Data Sheet Peak, CW & Average Power Sensors Taking performance to a new peak Peak, CW & Average Power Sensors The overall performance of a power meter dependents on the power sensor employed. Boonton has

More information

Using the 55XX Series Oscilloscope Calibration Options

Using the 55XX Series Oscilloscope Calibration Options Using the 55XX Series Oscilloscope Calibration Options Application Note Modern technology, government regulations, and business trends are driving a demand for increased oscilloscope calibration. Many

More information

Enhanced Sample Rate Mode Measurement Precision

Enhanced Sample Rate Mode Measurement Precision Enhanced Sample Rate Mode Measurement Precision Summary Enhanced Sample Rate, combined with the low-noise system architecture and the tailored brick-wall frequency response in the HDO4000A, HDO6000A, HDO8000A

More information

Frequency and Time Domain Representation of Sinusoidal Signals

Frequency and Time Domain Representation of Sinusoidal Signals Frequency and Time Domain Representation of Sinusoidal Signals By: Larry Dunleavy Wireless and Microwave Instruments University of South Florida Objectives 1. To review representations of sinusoidal signals

More information

B. Equipment. Advanced Lab

B. Equipment. Advanced Lab Advanced Lab Measuring Periodic Signals Using a Digital Oscilloscope A. Introduction and Background We will use a digital oscilloscope to characterize several different periodic voltage signals. We will

More information

AIM & THURLBY THANDAR INSTRUMENTS

AIM & THURLBY THANDAR INSTRUMENTS AIM & THURLBY THANDAR INSTRUMENTS I-prober 520 positional current probe Unique technology enabling current measurement in PCB tracks bandwidth of DC to 5MHz, dynamic range of 10mA to 20A pk-pk useable

More information

Equipment: You will use the bench power supply, function generator and oscilloscope.

Equipment: You will use the bench power supply, function generator and oscilloscope. EE203 Lab #0 Laboratory Equipment and Measurement Techniques Purpose Your objective in this lab is to gain familiarity with the properties and effective use of the lab power supply, function generator

More information

Oscilloscope calibration Application Note

Oscilloscope calibration Application Note Oscilloscope calibration Application Note A guide to oscilloscope calibration using dedicated or multiproduct calibrators User requirements Oscilloscopes are very complex instruments, mainly because of

More information

Fluke 5820A Oscilloscope Calibrator Specifications

Fluke 5820A Oscilloscope Calibrator Specifications Fluke 5820A Oscilloscope Calibrator Specifications Voltage Function Volt Function DC Signal Square Wave Signal 1 Load into 50Ω into 1 MΩ into 50Ω into 1 MΩ Amplitude Range 0V to ±6.6V 0V to ±130V ±1 mv

More information

UNIT 2. Q.1) Describe the functioning of standard signal generator. Ans. Electronic Measurements & Instrumentation

UNIT 2. Q.1) Describe the functioning of standard signal generator. Ans.   Electronic Measurements & Instrumentation UNIT 2 Q.1) Describe the functioning of standard signal generator Ans. STANDARD SIGNAL GENERATOR A standard signal generator produces known and controllable voltages. It is used as power source for the

More information

Keysight Measuring High Impedance Sources Using the U8903B Audio Analyzer. Application Note

Keysight Measuring High Impedance Sources Using the U8903B Audio Analyzer. Application Note Keysight Measuring High Impedance Sources Using the U8903B Audio Analyzer Application Note Introduction This note details the input impedance of the U8903B Audio Analyzer, and shows that this needs to

More information

EENG-201 Experiment # 4: Function Generator, Oscilloscope

EENG-201 Experiment # 4: Function Generator, Oscilloscope EENG-201 Experiment # 4: Function Generator, Oscilloscope I. Objectives Upon completion of this experiment, the student should be able to 1. To become familiar with the use of a function generator. 2.

More information

Introduction to basic laboratory instruments

Introduction to basic laboratory instruments Introduction to basic laboratory instruments 1. OBJECTIVES... 2 2. LABORATORY SAFETY... 2 3. BASIC LABORATORY INSTRUMENTS... 2 4. USING A DC POWER SUPPLY... 2 5. USING A FUNCTION GENERATOR... 3 5.1 TURN

More information

AC Current Probes CT1 CT2 CT6 Data Sheet

AC Current Probes CT1 CT2 CT6 Data Sheet View at www.testequipmentdepot.com AC Current Probes CT1 CT2 CT6 Data Sheet Features & Benefits High Bandwidth Ultra-low Inductance Very Small Form Factor Characterize Current Waveforms up to

More information

Improving CDM Measurements With Frequency Domain Specifications

Improving CDM Measurements With Frequency Domain Specifications Improving CDM Measurements With Frequency Domain Specifications Jon Barth (1), Leo G. Henry Ph.D (2), John Richner (1) (1) Barth Electronics, Inc, 1589 Foothill Drive, Boulder City, NV 89005 USA tel.:

More information

Combinational logic: Breadboard adders

Combinational logic: Breadboard adders ! ENEE 245: Digital Circuits & Systems Lab Lab 1 Combinational logic: Breadboard adders ENEE 245: Digital Circuits and Systems Laboratory Lab 1 Objectives The objectives of this laboratory are the following:

More information

Appendix A: Specifications

Appendix A: Specifications All specifications apply to the TDS 200-Series Digital Oscilloscopes and a P2100 probe with the Attenuation switch set to 10X unless noted otherwise. To meet specifications, two conditions must first be

More information

FISCHER CUSTOM COMMUNICATIONS, INC.

FISCHER CUSTOM COMMUNICATIONS, INC. FISCHER CUSTOM COMMUNICATIONS, INC. Current Probe Catalog FISCHER CUSTOM COMMUNICATIONS, INC. Fischer Custom Communications, Inc., is a manufacturer of custom electric and magnetic field sensors for military

More information

PXIe Contents. Required Software CALIBRATION PROCEDURE

PXIe Contents. Required Software CALIBRATION PROCEDURE CALIBRATION PROCEDURE PXIe-5113 This document contains the verification and adjustment procedures for the PXIe-5113. Refer to ni.com/calibration for more information about calibration solutions. Contents

More information

MODEL GB/S BROADBAND AMPLIFIER

MODEL GB/S BROADBAND AMPLIFIER Electro-Absorption Modulator driver or optical receiver amplifier khz - 43 GHz bandwidth 8 ps risetime.7 V amp eye amplitude 8.5 db gain MODEL 5881 4 GB/S BROADBAND AMPLIFIER The 5881 is extremely broadband,

More information

CERTIFICATE OF CALIBRATION

CERTIFICATE OF CALIBRATION CERTIFICATE OF CALIBRATION Issued by: Pico Technology Ltd. Certificate Number: 9999 of: James House, Colmworth Business Park, St. Neots, Cambridgeshire, Signature: PE19 8YP UNITED KINGDOM Tel: +44 (0)

More information

High Speed Competitive Comparison Report. Samtec MMCX-J-P-H-ST-TH1 Mated With MMCX-P-P-H-ST-TH1 Competitor A (Mated Set) Competitor B (Mated Set)

High Speed Competitive Comparison Report. Samtec MMCX-J-P-H-ST-TH1 Mated With MMCX-P-P-H-ST-TH1 Competitor A (Mated Set) Competitor B (Mated Set) High Speed Competitive Comparison Report Samtec MMCX-J-P-H-ST-TH1 Mated With MMCX-P-P-H-ST-TH1 Competitor A (Mated Set) Competitor B (Mated Set) REVISION DATE: January 6, 2005 TABLE OF CONTENTS Introduction...

More information

Models 296 and 295 combine sophisticated

Models 296 and 295 combine sophisticated Established 1981 Advanced Test Equipment Rentals www.atecorp.com 800-404-ATEC (2832) Models 296 and 295 50 MS/s Synthesized Multichannel Arbitrary Waveform Generators Up to 4 Independent Channels 10 Standard

More information

The Oscilloscope. Vision is the art of seeing things invisible. J. Swift ( ) OBJECTIVE To learn to operate a digital oscilloscope.

The Oscilloscope. Vision is the art of seeing things invisible. J. Swift ( ) OBJECTIVE To learn to operate a digital oscilloscope. The Oscilloscope Vision is the art of seeing things invisible. J. Swift (1667-1745) OBJECTIVE To learn to operate a digital oscilloscope. THEORY The oscilloscope, or scope for short, is a device for drawing

More information

LB480A Pulse Profiling USB PowerSensor+ Data Sheet

LB480A Pulse Profiling USB PowerSensor+ Data Sheet Key PowerSensor+ Specifications 50 MHz to 8 GHz (functional to 10 GHz) - 60 dbm to +20 dbm 1.95% Total Error* 1.09:1 VSWR (-27 db Return Loss) * Measuring a well matched DUT (-20 dbm @ 1 GHz) No Zero No

More information

Lab: INTRODUCTION TO THE WAVEFORM GENERATOR AND THE OSCILLOSCOPE

Lab: INTRODUCTION TO THE WAVEFORM GENERATOR AND THE OSCILLOSCOPE Name EET101/Lab#5; EET121/Lab#5; EGR104/Lab#3 Sec / Night Date Lab Partner(s) Name(s) Lab: INTRODUCTION TO THE WAVEFORM GENERATOR AND THE OSCILLOSCOPE Objectives: Each student will: 1. Know the function

More information

Application Note AN-23 Copyright September, 2009

Application Note AN-23 Copyright September, 2009 Removing Jitter From Picosecond Pulse Measurements James R. Andrews, Ph.D, IEEE Fellow PSPL Founder and former President (retired) INTRODUCTION: Uncertainty is always present in every measurement. Uncertainties

More information

N acquisitions, all channels simultaneously, N is selectable from 4, 16, 64, and 128 Inputs

N acquisitions, all channels simultaneously, N is selectable from 4, 16, 64, and 128 Inputs With compliments All specifications apply to the TDS 200-Series Digital Real-Time Oscilloscope with a P2100 probe with the Attenuation switch set to 10X unless noted otherwise. To meet specifications,

More information

SCOPE OF ACCREDITATION TO ISO/IEC 17025:2005

SCOPE OF ACCREDITATION TO ISO/IEC 17025:2005 SCOPE OF ACCREDITATION TO ISO/IEC 17025:2005 JEL LIMITED 2971 Nakabyo, Abiko-City Chiba-Prefecture, 270-1121, JAPAN Keiichiro Murata Phone: +81 4 7188 5333 Email: murata@jel.co.jp CALIBRATION Valid To:

More information

For the National Voluntary Laboratory Accreditation Program

For the National Voluntary Laboratory Accreditation Program SCOPE OF ACCREDITATION TO ISO/IEC 17025:2005 Intertek Japan K.K. Calibration Laboratory 3-2 Sunayama Kamisu Ibaraki 314-0255 JAPAN Ms. Masako Oyamada Phone: 81-465 89 2316 Fax: 81-465 89 2160 E-mail: masako.oyamada@intertek.com

More information

Model 310H Fast 800V Pulse Generator

Model 310H Fast 800V Pulse Generator KEY FEATURES Temperature Stability +/-5ppm 100 V to 800 V into 50 Ω

More information

High Speed Characterization Report

High Speed Characterization Report High Speed Characterization Report MMCX-P-P-H-ST-TH1 mated with MMCX-J-P-H-ST-TH1 MMCX-P-P-H-ST-MT1 mated with MMCX-J-P-H-ST-MT1 MMCX-P-P-H-ST-SM1 mated with MMCX-J-P-H-ST-SM1 MMCX-P-P-H-ST-EM1 mated with

More information

Radio Frequency Power Meter Design Project

Radio Frequency Power Meter Design Project Radio Frequency Power Meter Design Project Timothy Holt and Andrew Milks University of Akron, Akron Ohio Abstract This student paper discusses a radio frequency power meter developed and prototyped as

More information

UNIVERSITY OF CALIFORNIA, SANTA BARBARA Department of Electrical and Computer Engineering. ECE 2A & 2B Laboratory Equipment Information

UNIVERSITY OF CALIFORNIA, SANTA BARBARA Department of Electrical and Computer Engineering. ECE 2A & 2B Laboratory Equipment Information UNIVERSITY OF CALIFORNIA, SANTA BARBARA Department of Electrical and Computer Engineering ECE 2A & 2B Laboratory Equipment Information Table of Contents Digital Multi-Meter (DMM)... 1 Features... 1 Using

More information

P a g e 1 ST985. TDR Cable Analyzer Instruction Manual. Analog Arts Inc.

P a g e 1 ST985. TDR Cable Analyzer Instruction Manual. Analog Arts Inc. P a g e 1 ST985 TDR Cable Analyzer Instruction Manual Analog Arts Inc. www.analogarts.com P a g e 2 Contents Software Installation... 4 Specifications... 4 Handling Precautions... 4 Operation Instruction...

More information

LB680A Pulse Profiling USB PowerSensor+ Data Sheet

LB680A Pulse Profiling USB PowerSensor+ Data Sheet Key PowerSensor+ Specifications 50 MHz to 20 GHz - 40 dbm to +20 dbm 2.8% Total Error* 1.20:1 VSWR (-21 db Return Loss) * Measuring a well matched DUT (-20 dbm @ 2 GHz) Measurement Capability Time Gated

More information

Precision in Practice Achieving the best results with precision Digital Multimeter measurements

Precision in Practice Achieving the best results with precision Digital Multimeter measurements Precision in Practice Achieving the best results with precision Digital Multimeter measurements Paul Roberts Fluke Precision Measurement Ltd. Abstract Digital multimeters are one of the most common measurement

More information

TAKE THE MYSTERY OUT OF PROBING. 7 Common Oscilloscope Probing Pitfalls to Avoid

TAKE THE MYSTERY OUT OF PROBING. 7 Common Oscilloscope Probing Pitfalls to Avoid TAKE THE MYSTERY OUT OF PROBING 7 Common Oscilloscope Probing Pitfalls to Avoid Introduction Understanding common probing pitfalls and how to avoid them is crucial in making better measurements. In an

More information

Amplifier Test Bench Taking performance to a new peak

Amplifier Test Bench Taking performance to a new peak Data Sheet Amplifier Test Bench Taking performance to a new peak Amplifier Test Bench Boonton s Amplifier Test Bench is a powerful software tool especially designed for efficient and accurate, test verification

More information

NI Contents CALIBRATION PROCEDURE

NI Contents CALIBRATION PROCEDURE CALIBRATION PROCEDURE NI 5450 Contents This document describes processes to calibrate the National Instruments PXIe-5450 (NI 5450) differential I/Q signal generator. This document provides performance

More information

Department of Electrical and Computer Engineering. Laboratory Experiment 1. Function Generator and Oscilloscope

Department of Electrical and Computer Engineering. Laboratory Experiment 1. Function Generator and Oscilloscope Department of Electrical and Computer Engineering Laboratory Experiment 1 Function Generator and Oscilloscope The purpose of this first laboratory assignment is to acquaint you with the function generator

More information

ECE2019 Sensors, Signals, and Systems A Lab #6: Electromagnetic Field Sensing

ECE2019 Sensors, Signals, and Systems A Lab #6: Electromagnetic Field Sensing ECE2019 Sensors, Signals, and Systems A 2012 Lab #6: Electromagnetic Field Sensing Introduction This lab involves construction of circuits which demonstrate electromagnetic properties of cables used in

More information

CONNECTING THE PROBE TO THE TEST INSTRUMENT

CONNECTING THE PROBE TO THE TEST INSTRUMENT 2SHUDWLRQ 2SHUDWLRQ Caution The input circuits in the AP034 Active Differential Probe incorporate components that protect the probe from damage resulting from electrostatic discharge (ESD). Keep in mind

More information

5520A. Multi-Product Calibrator. Extended Specifications 2005

5520A. Multi-Product Calibrator. Extended Specifications 2005 5520A Multi-Product Calibrator Extended Specifications 2005 5520A Specifications The following tables list the 5520A specifications. All specifications are valid after allowing a warm-up period of 30 minutes,

More information

5500A. Multi-Product Calibrator. Extended Specifications 2005

5500A. Multi-Product Calibrator. Extended Specifications 2005 5500A Multi-Product Calibrator Extended Specifications 2005 5500A Specifications The following paragraphs detail specifications for the 5500A Calibrator. The specifications are valid after allowing a warm-up

More information

FREQUENCY SYNTHESIZERS, SIGNAL GENERATORS

FREQUENCY SYNTHESIZERS, SIGNAL GENERATORS SYNTHESIZED SIGNAL GENERATOR MG3641A/MG3642A 12 khz to 1040/2080 MHz NEW New Anritsu synthesizer technology permits frequency to be set with a resolution of 0.01 Hz across the full frequency range. And

More information

Laboratory 3 (drawn from lab text by Alciatore)

Laboratory 3 (drawn from lab text by Alciatore) Laboratory 3 (drawn from lab text by Alciatore) The Oscilloscope Required Components: 1 10 resistor 2 100 resistors 2 lk resistors 1 2k resistor 2 4.7M resistors 1 0.F capacitor 1 0.1 F capacitor 1 1.0uF

More information

TG1010A AIM & THURLBY THANDAR INSTRUMENTS. 10MHz programmable DDS function generator. Direct Digital Synthesis

TG1010A AIM & THURLBY THANDAR INSTRUMENTS. 10MHz programmable DDS function generator. Direct Digital Synthesis AIM & THURLBY THANDAR INSTRUMENTS TG1010A 10MHz programmable DDS function generator Arbitrary Waveform Capability, Extensive Modulation Modes Direct Digital Synthesis All the versatility of a function

More information

Rigol DG1022A Function / Arbitrary Waveform Generator

Rigol DG1022A Function / Arbitrary Waveform Generator Rigol DG1022A Function / Arbitrary Waveform Generator The Rigol DG1000 series Dual-Channel Function/Arbitrary Waveform Generator adopts DDS (Direct Digital Synthesis) technology to provide stable, high-precision,

More information

2520 Pulsed Laser Diode Test System

2520 Pulsed Laser Diode Test System Complete pulse test of laser diode bars and chips with dual photocurrent measurement channels 0 Pulsed Laser Diode Test System Simplifies laser diode L-I-V testing prior to packaging or active temperature

More information

NI PXIe-5171R. Contents. Required Software CALIBRATION PROCEDURE

NI PXIe-5171R. Contents. Required Software CALIBRATION PROCEDURE CALIBRATION PROCEDURE NI PXIe-5171R This document contains the verification and adjustment procedures for the NI PXIe-5171R (NI 5171R). Refer to ni.com/calibration for more information about calibration

More information

LB480A Pulse Profiling USB PowerSensor+ Data Sheet

LB480A Pulse Profiling USB PowerSensor+ Data Sheet Key PowerSensor+ Specifications 100 MHz to 8 GHz (functional to 10 GHz) -60 dbm to +20 dbm 1.95% Total Error* 1.09:1 VSWR (-27 db Return Loss) * Measuring a well matched DUT (-20 dbm @ 1 GHz) Measurement

More information

SCOPE OF ACCREDITATION TO ISO/IEC 17025:2005 AMETEK CO., LTD. NAGOYA OFFICE. (Main Laboratory) Onna, Atsugi-shi, Kanagawa-ken, Japan

SCOPE OF ACCREDITATION TO ISO/IEC 17025:2005 AMETEK CO., LTD. NAGOYA OFFICE. (Main Laboratory) Onna, Atsugi-shi, Kanagawa-ken, Japan SCOPE OF ACCREDITATION TO ISO/IEC 17025:2005 AMETEK CO., LTD. NAGOYA OFFICE (Main Laboratory) 5-1-1 Onna, Atsugi-shi, Kanagawa-ken, 243-0032 Japan (Commercial Office) 1-329 Kifune, Meito-ku Nagoya-shi,

More information

Lab 0: Introduction to basic laboratory instruments. Revised by Dan Hoang & Tai-Chang Chen 03/30/2009

Lab 0: Introduction to basic laboratory instruments. Revised by Dan Hoang & Tai-Chang Chen 03/30/2009 Lab 0: Introduction to basic laboratory instruments Revised by Dan Hoang & Tai-Chang Chen 03/30/2009 1. Objectives 1. To learn safety procedures in the laboratory. 2. To learn how to use basic laboratory

More information

Test & Calibration Benefits from a New Precision RF/Microwave Calibrator

Test & Calibration Benefits from a New Precision RF/Microwave Calibrator Test & Calibration Benefits from a New Precision RF/Microwave Calibrator Topics: RF & Microwave calibration signal requirements Design philosophy and architecture of the new RF Calibrator. Spectrum analyzer

More information

LAB I. INTRODUCTION TO LAB EQUIPMENT

LAB I. INTRODUCTION TO LAB EQUIPMENT LAB I. INTRODUCTION TO LAB EQUIPMENT 1. OBJECTIVE In this lab you will learn how to properly operate the basic bench equipment used for characterizing active devices: 1. Oscilloscope (Keysight DSOX 1102A),

More information

LFR: flexible, clip-around current probe for use in power measurements

LFR: flexible, clip-around current probe for use in power measurements LFR: flexible, clip-around current probe for use in power measurements These technical notes should be read in conjunction with the LFR short-form datasheet. Power Electronic Measurements Ltd Nottingham

More information

Application Note #5 Direct Digital Synthesis Impact on Function Generator Design

Application Note #5 Direct Digital Synthesis Impact on Function Generator Design Impact on Function Generator Design Introduction Function generators have been around for a long while. Over time, these instruments have accumulated a long list of features. Starting with just a few knobs

More information

2026A/B 10 khz to 2.05/2.51 GHz MultiSource Generator

2026A/B 10 khz to 2.05/2.51 GHz MultiSource Generator Signal Sources 2026A/B 10 khz to 2.05/2.51 GHz MultiSource Generator Up to three fully functional signal generators in one unit offering a unique solution for complex tests on receivers, components and

More information

Probe Considerations for Low Voltage Measurements such as Ripple

Probe Considerations for Low Voltage Measurements such as Ripple Probe Considerations for Low Voltage Measurements such as Ripple Our thanks to Tektronix for allowing us to reprint the following article. Figure 1. 2X Probe (CH1) and 10X Probe (CH2) Lowest System Vertical

More information

Why Modern Servicing Requires Complete Waveform & Circuit Analyzing!

Why Modern Servicing Requires Complete Waveform & Circuit Analyzing! Why Modern Servicing Requires Complete Waveform & Circuit Analyzing! DC Bias Voltages DC Currents Resistance AC Signals Of Various Waveshapes & Amplitudes Continuity Of Circuit Paths & Components If you

More information

EMC Pulse Measurements

EMC Pulse Measurements EMC Pulse Measurements and Custom Thresholding Presented to the Long Island/NY IEEE Electromagnetic Compatibility and Instrumentation & Measurement Societies - May 13, 2008 Surge ESD EFT Contents EMC measurement

More information

Basic Communication Laboratory Manual. Shimshon Levy&Harael Mualem

Basic Communication Laboratory Manual. Shimshon Levy&Harael Mualem Basic Communication Laboratory Manual Shimshon Levy&Harael Mualem September 2006 CONTENTS 1 The oscilloscope 2 1.1 Objectives... 2 1.2 Prelab... 2 1.3 Background Theory- Analog Oscilloscope...... 3 1.4

More information

G5100A: 50 MHz Arbitrary Function Generator

G5100A: 50 MHz Arbitrary Function Generator G5100A: 50 MHz Arbitrary Function Generator Key Features 50 MHz Sine Wave 25 MHz Square Wave Pulse, Ramp, Triangle, Noise, and DC waveforms AM, FM, PM, FSK, and PWM modulation types Linear & logarithmic

More information

Introduction to basic laboratory instruments

Introduction to basic laboratory instruments BEE 233 Laboratory-1 Introduction to basic laboratory instruments 1. Objectives To learn safety procedures in the laboratory. To learn how to use basic laboratory instruments: power supply, function generator,

More information

Traceability and Modulated-Signal Measurements

Traceability and Modulated-Signal Measurements Traceability and Modulated-Signal Measurements Kate A. Remley 1, Dylan F. Williams 1, Paul D. Hale 2 and Dominique Schreurs 3 1. NIST Electromagnetics Division 2. NIST Optoelectronics Division 3. K.U.

More information

SAMPULSE50GHz and SAMPULSE70GHzANT PULSER/SAMPLER DEMO BOARD

SAMPULSE50GHz and SAMPULSE70GHzANT PULSER/SAMPLER DEMO BOARD SAMPULSE50GHz and SAMPULSE70GHzANT PULSER/SAMPLER DEMO BOARD REV B. PRELIMINARY USER MANUAL 2/25/15 Figure 1. Connections to SAMPULSE50GHz (left) or SAMPULSE70GHzANT (right). INTRODUCTION: The SAMPULSE50GHz

More information

Phase Matrix, Inc. 545B 548B. Phase Matrix, Inc. EIP 545B and 548B CW Frequency Counters. Instruments You Can Count On

Phase Matrix, Inc. 545B 548B. Phase Matrix, Inc. EIP 545B and 548B CW Frequency Counters. Instruments You Can Count On Phase Matrix, Inc. Instruments You Can Count On 545B 548B Phase Matrix, Inc. EIP 545B and 548B CW Frequency Counters Full Function CW Microwave Frequency Counters with Selective Power Measurement Keyboard

More information

Experiment 2: Transients and Oscillations in RLC Circuits

Experiment 2: Transients and Oscillations in RLC Circuits Experiment 2: Transients and Oscillations in RLC Circuits Will Chemelewski Partner: Brian Enders TA: Nielsen See laboratory book #1 pages 5-7, data taken September 1, 2009 September 7, 2009 Abstract Transient

More information

LIMITATIONS IN MAKING AUDIO BANDWIDTH MEASUREMENTS IN THE PRESENCE OF SIGNIFICANT OUT-OF-BAND NOISE

LIMITATIONS IN MAKING AUDIO BANDWIDTH MEASUREMENTS IN THE PRESENCE OF SIGNIFICANT OUT-OF-BAND NOISE LIMITATIONS IN MAKING AUDIO BANDWIDTH MEASUREMENTS IN THE PRESENCE OF SIGNIFICANT OUT-OF-BAND NOISE Bruce E. Hofer AUDIO PRECISION, INC. August 2005 Introduction There once was a time (before the 1980s)

More information

When you have completed this exercise, you will be able to relate the gain and bandwidth of an op amp

When you have completed this exercise, you will be able to relate the gain and bandwidth of an op amp Op Amp Fundamentals When you have completed this exercise, you will be able to relate the gain and bandwidth of an op amp In general, the parameters are interactive. However, in this unit, circuit input

More information

DC to 3.5-GHz Amplified Photoreceivers Models 1591 & 1592

DC to 3.5-GHz Amplified Photoreceivers Models 1591 & 1592 USER S GUIDE DC to 3.5-GHz Amplified Photoreceivers Models 1591 & 1592 These photoreceivers are sensitive to electrostatic discharges and could be permanently damaged if subjected even to small discharges.

More information

University of New Hampshire InterOperability Laboratory Gigabit Ethernet Consortium

University of New Hampshire InterOperability Laboratory Gigabit Ethernet Consortium University of New Hampshire InterOperability Laboratory Gigabit Ethernet Consortium As of June 18 th, 2003 the Gigabit Ethernet Consortium Clause 40 Physical Medium Attachment Conformance Test Suite Version

More information

HP 8901B Modulation Analyzer. HP 11722A Sensor Module. 150 khz MHz. 100 khz MHz. Technical Specifications. Four Instruments In One

HP 8901B Modulation Analyzer. HP 11722A Sensor Module. 150 khz MHz. 100 khz MHz. Technical Specifications. Four Instruments In One HP 8901B Modulation Analyzer 150 khz - 1300 MHz HP 11722A Sensor Module 100 khz - 2600 MHz Technical Specifications Four Instruments In One RF Power: ±0.02 db instrumentation accuracy RF Frequency: 10

More information

3.003 Lab 3 Part A. Measurement of Speed of Light

3.003 Lab 3 Part A. Measurement of Speed of Light 3.003 Lab 3 Part A. Measurement of Speed of Light Objective: To measure the speed of light in free space Experimental Apparatus: Feb. 18, 2010 Due Mar. 2, 2010 Components: 1 Laser, 4 mirrors, 1 beam splitter

More information

SPECTRUM ANALYZERS. MS710C/D/E/F 10 khz to 23 GHz (18 to 140 GHz) GPIB SPECTRUM ANALYZER

SPECTRUM ANALYZERS. MS710C/D/E/F 10 khz to 23 GHz (18 to 140 GHz) GPIB SPECTRUM ANALYZER SPECTRUM ANALYZER MS710C/D/E/F 10 khz to 23 GHz (18 to 140 GHz) 2 GPIB The MS710C/D/E/F has been designed as a high-performance microwave spectrum analyzer with wide user applications. The MS710C/D/ E/F

More information

Analog Arts SG985 SG884 SG834 SG814 Product Specifications [1]

Analog Arts SG985 SG884 SG834 SG814 Product Specifications [1] www.analogarts.com Analog Arts SG985 SG884 SG834 SG814 Product Specifications [1] 1. These models include: an oscilloscope, a spectrum analyzer, a data recorder, a frequency & phase meter, and an arbitrary

More information

Frequency Agility and Barrage Noise Jamming

Frequency Agility and Barrage Noise Jamming Exercise 1-3 Frequency Agility and Barrage Noise Jamming EXERCISE OBJECTIVE To demonstrate frequency agility, a radar electronic protection is used against spot noise jamming. To justify the use of barrage

More information

SCOPE OF ACCREDITATION TO ISO/IEC 17025:2005 & ANSI/NCSL Z

SCOPE OF ACCREDITATION TO ISO/IEC 17025:2005 & ANSI/NCSL Z SCOPE OF ACCREDITATION TO ISO/IEC 17025:2005 & ANSI/NCSL Z540.3-2006 ELECTRO RENT CORPORATION 15385 Oxnard Street Van Nuys, CA 91411 Karl Haynes Phone: 770 310 5228 CALIBRATION Valid To: June 30, 2019

More information

Laboratory Experience #5: Digital Spectrum Analyzer Basic use

Laboratory Experience #5: Digital Spectrum Analyzer Basic use TELECOMMUNICATION ENGINEERING TECHNOLOGY PROGRAM TLCM 242: INTRODUCTION TO TELECOMMUNICATIONS LABORATORY Laboratory Experience #5: Digital Spectrum Analyzer Basic use 1.- INTRODUCTION Our normal frame

More information

Supplement. TDS5032 and TDS5034 Digital Phosphor Oscilloscopes

Supplement. TDS5032 and TDS5034 Digital Phosphor Oscilloscopes TDS5032 and TDS5034 Digital Phosphor Oscilloscopes 071-1316-00 www.tektronix.com 071131600 Copyright Tektronix, Inc. All rights reserved. Tektronix products are covered by U.S. and foreign patents, issued

More information

Model 7000 Series Phase Noise Test System

Model 7000 Series Phase Noise Test System Established 1981 Advanced Test Equipment Rentals www.atecorp.com 800-404-ATEC (2832) Model 7000 Series Phase Noise Test System Fully Integrated System Cross-Correlation Signal Analysis to 26.5 GHz Additive

More information

Manual Supplement. This supplement contains information necessary to ensure the accuracy of the above manual.

Manual Supplement. This supplement contains information necessary to ensure the accuracy of the above manual. Manual Title: 5502E Getting Started Supplement Issue: 3 Part Number: 4155211 Issue Date: 9/18 Print Date: November 2012 Page Count: 12 Revision/Date: This supplement contains information necessary to ensure

More information