Laboratory 1 Generating and viewing signals rev. 20e. Oscilloscope Settings

Size: px
Start display at page:

Download "Laboratory 1 Generating and viewing signals rev. 20e. Oscilloscope Settings"

Transcription

1 1 2 Laboratory 1 Generating and viewing signals rev. 20e Purpose: Familiarization with the basic functions of an oscilloscope and of a signal generator. Adjusting and measuring specific parameters of signals. Summary of theory The oscilloscope (often abbreviated scope) is a device that allows viewing the instantaneous value of a voltage v(t) as a function of time, and quantitative measurements of voltage and time on the waveform, having broad applications in the analysis of electrical signals. In this lab we use the Tektronix TDS1001 digital scope. The description that follows corresponds to this model. The main parameters of a periodic signal - waveform (deterministic) in the time domain - period and frequency, f = 1/ T - maximum value U max - minimum value U min - peak-to-peak value, the signal range U PP = U max - U min ; - average (offset) value of the signal U avg or U DC - signal amplitude (if the signal is similar to a sine wave) U 0 = U max U DC = U DC - U min Remark: if U DC = 0, the amplitude, min and max values are equal in absolute values. - other parameters for certain waveforms, for example: - rectangular wave: duty cycle η = τ/t (τ represents the width of the high pulse, and T is the signal period), rise / fall time (t rise, t fall) - triangular wave: rising / falling slope Oscilloscope Settings Horizontally, the duration corresponding to the length of a division on the screen is adjustable from the C x knob (horizontal deflection coefficient). It is labeled in units of time per division. The following relation exists between the number of divisions N that an event occupies and its duration T x : x T x = N x C x Vertically, voltage U is applied from external input Y. Similar to the reading on the X axis, one can read the value of voltage U that occupies N y vertical divisions using the formula: U = N y C y C y is called vertical deflection coefficient and is labeled in units of volts per division. Example: A period of the sine wave in Fig. 2 occupies 7 horizontal divisions on the screen, and the time base is set at Cx=5ms/div. The signal s period is, thus, 35 ms. Vertically, the amplitude (peak value) occupies 3 divisions, which, for Cy= 2V/div, means 6V. U max τ U 0 U max t U 0 U DC U DC U min T U min T a. Sine wave b. Rectangular wave Figure 1. Periodic signals: sine, rectangular t Figure 2. Horizontal and vertical divisions

2 3 4 Values C x, C y are called calibrated and have standard form (1,2,5)*10 +/-K. for C y, and (1, 2.5, 5)*10 +/-K. Calibrated coefficient values for usual oscilloscopes are: Cy {5; 10; 20; 50; 100; 200; 500 mv/div; 1; 2; 5 V/div} Cx {5; 10; 25; 50; 100; 250; 500 ns/div; 1; 2,5; 5; 10; 25; 50; 100; 250; 500 μs/div; 1; 2,5; 5; 10, 25; 50; 100; 250; 500 ms/div, 1; 2,5; 5; 10 s/div} Triggering the oscilloscope: An image that is stable on the display of the oscilloscope is called triggered (synchronized). The physical meaning is the following: when 2 successive images of a periodic signal start at the same moment in time (relative to the signal period), the 2 images will overlap perfectly, and so will happen for subsequent images. Thus, the eye perceives a single stable image, although, in fact, we constantly have a new image superimposed on the previous one. An example in case of displaying a rising slope is given in Figure 3 (a). But if every display starts with some other moment of time, the images will differ, and the eye will perceive many different superposed images - Fig. 3 (b). In this case the image is called untriggered (unsynchronized) and is difficult or impossible to interpret. a) triggered image b) untriggered image Figure 3: oscilloscope triggering In order to obtain a triggered image, the operator must adjust the trigger settings. The most important are: trigger source, its level and a slope at which the displaying of the image should start. Usually, those adjustments are grouped in a trigger menu. Usually, in order for an image to be synchronized, the Trigger Level has to be between the [minimum, maximum] level of the signal. A smaller or larger level means that the Trigger Level does not intersect the signal, therefore it cannot trigger a display. In the measurements that will be performed next, the following trigger settings will especially be used: Source = CH1, slope = Rise, Mode = AUTO. For automatically adjusting the Trigger Level, the Set To 50% key has to be pushed. Generator settings The Rigol DG1022 function generator allows the generation of several waveforms (sine, rectangular, triangular etc) with different tuneable parameters. The waveform can be selected by pressing the corresponding buttons: The parameters of a certain waveform can be selected by pressing the function (unmarked) keys whose functions correspond to the indications on the display, above them: For example, in the case of pressing the Freq function key, that particular function (setting the frequency) becomes selected, and the value can be introduced either from the numerical keypad, or from the rotating knob. In the case of function keys which have a small arrow to the right side (like Freq above) successive presses of the key allow setting parameters in different variants. For instance, besides setting the amplitude (Ampl) you can also set the HiLevel value which corresponds with U max from figure 1a). After typing the numerical value, choose the desired measurement unit by pressing the function key beneath it. Pay attention! According to figure 1a), the amplitude U 0 and the value U max are equal only if U DC = 0 (null mean value). To eliminate possible confusions, this generator allows setting the amplitude in values of V PP (peak-topeak) which means that, actually, by choosing this measurement unit, the function key Ampl sets the peak-to-peak value, which is double the amplitude for signal with no DC level. For asymmetrical signals (with U DC nonzero), the peak-to-peak value stays the same, while U max and U min change (see again figure 1a).

3 5 6 Remark: there is no such thing as several types of volts, as the volt is uniquely defined. The optional index after the volt, like V PP, simply indicates the fact the the voltage is measured in volts between certain limits. For example, a symmetrical 2V amplitude signal is equivalent to a 4V PP signal, because the second way of expressing it explicitly signifies that it is measured between the extreme limits of the signal. The V RMS notation refers to the root mean square value which will be studied in the future lab work measuring voltages. The generator has 2 channels which can generate different waveforms. The above settings have effect on the channel that is selected using the key. Correspondingly, CH1 or CH2 will appear on the display. Moreover, the output of the respective channel is only active when the corresponding Output button is pressed and it lights up. On the scope, press CH1 MENU in order to display the settings of channel 1 (at repeated presses, channel 1 is alternatively turned on and off). Because a simple cable is used, not an attenuator probe, press the Probe soft-key until the indication is 1x (there are probes that contain a divisor that attenuates the signal times, in which case settings 10x, 100x would be used). Set the values for the oscilloscope C X1, C Y1 (written on the blackboard). Remark: review figure 2; the oscilloscope indicates the value of C Y of channel 1 with the notation CH1, and the value of C X as M (main time base). Notice that one period of the signal is displayed on the screen. Measure the period by counting the divisions and subdivisions N x of a period and use the formula Tmeas = Cx N x. Calculate the frequency f meas =1/T meas and verify that it corresponds with the value indicated by the generator. Repeat the previous steps for the vertical adjustment. By counting the vertical divisions and subdivisions N y and applying the formula U=N y C y, measure the amplitude of the signal (peak value U P ). In the same way, measure peak-to-peak value U PP. Calculate the ratio between U P and U PP (measured values). What is the theoretical value of the ratio? Remark: for identifying settings and indications of the oscilloscope you can use annex A. 2. Computing the display settings Measurements 1. Automatic settings of the oscilloscope Using the function generator, generate a sine wave with the following parameters: - channel 1 from the key (CH1 is displayed in the upper right corner) - Sine key sine wave - frequency Freq value f 1 written on the blackboard - amplitude AMPL value A 1 written on the blackboard (pay attention to the measurement unit! An amplitude of 2V is, like previously mentioned, equivalent to 4V PP! the measurement unit eliminates all doubt regarding the limits between which the voltage is specified). - no DC level (OFFSET =0V) Press the Output key next to the CH1 output so that it is illuminated; connect the CH1 output of the generator to channel 1 of the scope using a coaxial cable (or two crocodile clips connected together). Pay attention! This exercise is only for computing the settings. You do not have to display signals on the scope! a) Compute how many divisions, NX și NY, will the amplitude and period of the signal occupy on the scope display for the following signals and settings:: a1. sine wave with amplitude UP1=4V and frequency f1=20khz. Oscilloscope settings: CX1=10μs/div and CY1=1V/div. a2. sine wave with amplitude UP2=6V and frequency f2=8khz. Oscilloscope settings: CX2=25μs/div and CY2=2V/div. Indication 1: Use the following T = NX CX and UP = NY CY. You can compute T from the frequency value. Indication 2: Use as an example the solved exercises at the end of the laboratory work. Same goes for (b) and (c). b) Compute the deflection coefficients (CXcalc, CYcalc) that should be set for displaying a sine wave with the frequency of f3=1khz and amplitude UP3 =2V, such that the amplitude occupies two divisions and the period occupies four divisions.

4 7 8 c) Repeat (b) for a signal with f4=500khz și UP4=8V, if the amplitude occupies four divisions, and the period 10 divisions (CXcalc, CYcalc) 3. Generating a triangular wave a) Generate a triangular wave, (Ramp button) with no DC component (OFFSET 0V), symmetry 50%, with frequency f3 and amplitude UP3 (written on the blackboard). You have to set the oscilloscope such the display looks like the next figure (in order to have the start time like in the figure, press the SET TO 50% button below the Trigger level). with U P = 2V, u(t) varies within the limits [-2V, +2V], then for a signal with a DC component, written as: u(t) = U DC + U P sin ωt [V] [2] with U P = 2V și UDC= -1V, u(t) varies within the limits [-2-1V, +2-1V] = [-3V, +1V]. The two situations are illustrated in fig. 4: Fig. 4 Signal without DC (left) and with UDC= -1V (right) a1. What is the measured period of the signal (T 3 )? Compute the necessary C X to observe exactly 2 periods on the scope. Set this value C X on the oscilloscope. How many divisions does a period occupy on the scope (N X)? a2. Compute the necessary C Y s.t. the amplitude U V3 ocuppies exactly N Y = 2 div. Set this value C Y. b) Influence of C Y on the displayed image Set C Y to the value of CY =CY / 2. How many divisions N Y does the amplitude occupy now? Compute the amplitude based on the new image: U P = N Y C Y, and compare it with U P3. Explain the relationship between U P and U P3.. c) Influence of C X on the displayed image Modify C X to the value of CX =2Cx. How many divisions N X does the period occupy now? Compute the period based on the new image: T = N X C X, and compare it with T 3. Explain the relationship between T and T Generating and measuring a sine wave with a DC component a) Until now the signals did not have a DC component or DC level (also named offset), being symmetrical from 0 V. If for a signal without a DC offset written as: u(t) = U P sin ωt [V] [1] Remark: the DC component is also called the average value of the signal, because it is equal to the average of u(t) over a period. Notice that in eq. [1] and [2] above, the integration over a period (equivalent to averaging) gives us the values 0 and U DC respectively. Generate a sine wave (button Sine from the generator), with the frequency of f1=20khz, amplitude UP =2V and DC component UDC1= -1V. For setting the DC component at the generator, use the OFFSET = -1V function key. For the oscilloscope, use coefficients CX1=25μs/div and CY1=1V/div. If it isn t already, adjust the 0V (Ground) level in the middle of the screen, using Vertical Position. a1. Draw the image of the scope and specify the ground level (the arrow on the left). Use CH1 MENU->Coupling ->DC. This coupling mode means the signal is applied directly, without altering the possible DC component of the signal. Write down on the graph the position of the Ground level and the C x and C y values. a2. Draw the image when using AC coupling (CH1 MENU->Coupling - >AC). This coupling mode means that a capacitor is inserted in series between the signal and the oscilloscope signal path. As you should know, capacitors do not let continuous signals to pass through, but only alternative signals. Write down on the graph the position of the Ground level and the C x and C y values.

5 9 10 b) How can we measure, using the oscilloscope, the DC level of a given signal? - set the oscilloscope to AC coupling: (CH1 MENU->Coupling ->AC); the DC component is blocked; therefore, the signal is symmetrical on the screen (as if OFFSET were not set from the generator) - set CH1 MENU->Coupling ->DC. At this moment, the signal will rise or fall with a number of divisions N Y. Taking as reference a point on the waveform (typically, the maximum or minimum point), count how many divisions the signal rises or falls when swtiching from AC to DC. If it rises, then the offset is positive, if it falls, the offset is negative. Count how many divisions N Y, sign included, correspond to the value of U DC, and compute U DC = N Y C Y. This value should be the same as the one generated using the OFFSET from the generator. Remark: If needed, adjust the trigger (SET TO 50%) so that the image is synchronised (it s possible that when we move the signal up or down, because of the offset, it does no longer cross the trigger level). b1. Work in a team, as follows: - Set CH1 MENU->Coupling ->AC to hise the DC level. Keep the setting of CY = 1V/div. - One of the team members will set at the generator, from the function key Offset, a DC level of U DC1set different from zero, in the interval (-2V, 2V), without telling the chosen value to the colleague. The amplitude remains 2V. - The other team member will switch from CH1 MENU->Coupling ->AC, to DC, will count how many divisions NY1 the peak of the signal rises/falls, and will compute U DC1 (value and sign): U DC1 = N Y1 C Y - Draw the image with the setting CH1 MENU->Coupling ->DC (including the 0V level indicator). b2. Switch places, and repeat b1 for a different offset value. Explain why measuring the DC level is made by switching from AC to DC coupling, and not the other way around! 5. Generating a rectangular wave Generate a rectangular wave (Square button), no DC component (OFFSET 0V), with the amplitude of UP5, frequency of f5 (written on the blackboard). Compute C X5, C Y5 s.t. exactly one period is displayed, respectively the amplitude occupies 2 divisions. View the signal with CY5, CX5 and use coupling CH1 MENU->Coupling ->DC. Adjust the duty cycle of the signal (see fig. 1b) using the functional duty cycle button DtyCyc, then the measurement unit which is % implicitly successively, to the values of η1=50%, η2=20% - Measure (in divisions) the values of T and τ for both cases (see fig 1b). - Compute the ratio τ/t - Draw the two signals. Important remark: the duty cycle is a parameter that has meaning only for a rectangular wave, according to the definition in Fig. 1b. it has no meaning for a sine or triangular wave! (you can set the symmetry of the triangle). Solved exercises 1. A sine wave with a frequency of f=2khz and amplitude of UP=4V is displayed using an oscilloscope. The settings are: C Y=1V/div, C X=250 μs/div. Determine how many divisions the amplitude and period of the signal occupy on the display of the oscilloscope. Solution: The number of divisions that the amplitude occupies on the display can be determined using UP=NY CY NY = UP/CY = 4 [V] / 1[V/div] = 4 div The period of the signal is T = 1/f = 1/2000Hz = 500 μs The number of divisions that the period of the signal occupies on the display is T = NX CX NX = T/CX = [s] / [s/div] = 2 div 2. A sine wave with a frequency of f=2mhz and amplitude of UP=6V is displayed using an oscilloscope. Determine the values of C X and C Y such that the amplitude occupies 3 divisions (NY=3 div), and the period occupies two divisions (NX=2div). Solution: The equations in ex. 1 can be used: The period of the signal is T = 1/f =1/ s = 0,5 μs. CY = UP/NY = 6V / 3 div = 2 V/div CX = T/NX = 0,5 μs / 2 div = 0,25 μs/div 3. A sine wave is displayed using an oscilloscope. When the coupling is switched from AC to DC, the sine wave moves downwards by NY = 2 div. The vertical deflection coefficient is C Y=5V/div. Determine the DC level of the signal. Solution: The DC level value determines the downward movement of the signal image by a value that is equal to the value of the offset (for DC coupling). The direction of the movement determines the sign of the offset: upwards - positive; downwards - negative. Based on these observations, we can determine the DC component: UDC = - NY CY = -10V 4. A sine wave is displayed using an oscilloscope. When the coupling is switched from AC to DC, the sine wave moves downwards by NY = 4 div. The vertical deflection coefficient is CY=1V/div. Determine the DC level of the signal Solution:

6 11 12 The difference from ex 3 is that now we switch from DC to AC. Because, after the elimination of the DC component (AC coupling), the signal moves upwards, this means that it (the DC component) was pulling the signal downwards (in DC coupling). This means that the value of the offset is negative. UDC = - NY CY = -4V ANNEX 1. Tektronix TDS1001 Oscilloscope Exercises 1. An oscilloscope is set to Cy=0,5V/div. The amplitude of a signal measured on the oscilloscope screen is 3.8div. What is the signal amplitude in volts? 2. An oscilloscope is set on Cx=20ms/div. The measured period of a sine wave on the screen of the oscilloscope is 5 div. Determine the frequency of the sine wave. 3. Given a sinusoidal signal of frequency 10kHz and amplitude 4V, determine the values of the horizontal and vertical deflection coefficients so that the amplitude and the period of the signal can be precisely measured on the screen. 4. A sinusoidal signal is viewed with an oscilloscope. When the coupling button is set from AC to DC position the sinusoidal signal moves on the screen, downwards by 3 divisions. Cy=1V/div. Determine the DC level of the signal. 5. A symmetrical rectangular signal, having amplitude A=1V, no DC level and frequency of 1kHz, is applied at the input of an oscilloscope. The oscilloscope has Cy=0,5V/div, Cx=0,2ms/div, trigger level Utrig=0,5V and SLOPE=falling. Draw the image. Figure A1: front panel of the oscilloscope The front panel of the oscilloscope is represented in Fig. A1. The interface of the oscilloscope contains the following elements: 1. The screen of the oscilloscope Figure A2: Informations and symbols that are displayed on the screen of the oscilloscope

7 13 The screen contains an area for displaying the grid of the screen, the area of the control menu (right side of the grid area) and the area for displaying parameters (above and below the grid zone). The graticule area is formed from N x =10 units horizontally and N y =8 units vertically, and it is used for displaying the image. Beside the image there are several parameters of the oscilloscope or of the waveform that are displayed according to the selected working mode. On Fig. A2, the most relevant ones are: (1) acquisition time (normal, averaging, etc) (2) Trig d = triggered = synchronized (3) Trigger moment, it can be swept using the adjustment HORIZONTAL POSITION (5) Trigger Level, it is adjusted using TRIGGER LEVEL (6) identifier of traces 1 and 2, its position is swept with VERTICAL POSITION (8) C y values of both channels (9) BW= Bandwidth Limit (it limits the maximum frequency of the oscilloscope to 20MHz) (10,11) C x values for the main time base (Main) and the secondary time base (Window) (12,13) the source and the slope of the trigger (17) the measured frequency of the signal. 14 CH1 MENU the result of pressing the button is the display in the control area of fields that allow the control of the displaying on the vertical axis, for channel 1 (CH1). The following fields will be displayed: o Coupling - it selects the coupling type AC/DC/Ground (alternating current/continuous current/zero level) o BW Limit band limiting at 20MHz instead of 40 (option ON/OFF) o Volts/Div calibrated (Coarse) or noncalibrated (Fine) adjustment. For the calibrated adjustment the vertically deflection coefficient can only have values C y={1,2,5}x10 k V/div. o Probe- the type of probe used (x1/x10/x100/x1000). The value must be the same as that used at the probe of signal. o Invert reverses the image when ON. MATH MENU allows applying some mathematical functions on signals (add, subtract, Fourier Transform). 2. Control keys - they allow changing the control fields displayed on the screen of the oscilloscope. They are called soft keys because those fields change according to the selected menu/mode. 3. Channel Y adjustments (vertical adjustments) - there is a set of adjustments for each of the two channels of the oscilloscope. 4. Digital Functions Menu the effect of pressing a button from this area is displaying on the screen a menu that contains functions which are specific to digital oscilloscopes (saving, measurement, acquisition, cursors, utilities, display). For the laboratory the menu DISPLAY is the one that is interesting. It contains the control field Format, RUN/STOP and SINGLE SEQ buttons. Figure A4: Menus for digital functions a) adjustments of the Y channel b) adjustments of the X channel Figure A3 POSITION it allows moving the image vertically VOLTS/DIV knob used for changing the vertical deflection coefficient. Its value is displayed in the bottom side of the screen (the area for displaying parameters). DISPLAY o Format selecting operating mode y(t) (YT) or x(y) (XY) RUN/STOP in RUN mode the oscilloscope acquires signal continuously. In STOP mode the acquisition is stopped, the image which is displayed being the last image before pressing the STOP button. SINGLE SEQ the oscilloscope acquires a single image (that corresponds to a single course on the screen) and it waits for a new press of the button. Pressing the button is equivalent to a RESET. 5. Adjustments for X channel of the oscilloscope (horizontal axis adjustments) SEC/DIV the adjustment of the horizontal deflection coefficient C x. Its value is displayed in the bottom side of the screen as M 10ms, which is equivalent to C x =10ms/div.

8 15 HORIZ MENU displays the menu for control of the displaying on horizontal axis. o MAIN- selects the display of the image for main time basis (the usual working mode). o Window zone selects the display of the image for secondary time basis (in fact, secondary time basis is the name from analogical oscilloscope. In this context it is a portion of the image that is "dilated" on horizontal axis) o Window adjustment of the temporal window for the secondary time basis. o Trigger Knob it allows selecting the function of the LEVEL button from the TRIGGER buttons area: implicitly it represents the adjustment of the trigger level; when the holdoff adjustment is selected, the LED bellow LEVEL button is on (lighting). POSITION moves the image horizontally SET TO ZERO restores the image to its original position (horizontal displacement deleted) 6. Adjustments/Settings for the synchronizing circuit (TRIGGER) For the TDS 1001 oscilloscope, the trigger moment corresponds to the middle of the screen. LEVEL it allows to set the trigger level and the holdoff time. TRIG MENU enables the menu for control of synchronization (trigger). It contains the following control fields: o Type selects the type of the trigger: Edge triggering on the slope of the signal, Video - triggering on a video signal, Pulse - triggering on pulses. o Source the source of the triggering signal (CH1, CH2, EXT, EXT/5, AC signal from the mains) o Slope type of the front: positive or negative (Rising/Falling) o Mode synchronizing mode (Auto/Normal): AUTO: if trigger conditions are not satisfied, the oscilloscope automatically generates, after a certain time, a signal to trigger the display. This way, when the input signal is not present a horizontal line can be observed on the screen. It represents the zero level. This is the default mode if not specified otherwise! Normal This time, the display is triggered only if the trigger conditions are met. Otherwise, the oscilloscope does not display any image. The trigger level can also be set outside the limits of the signal. There exists the possibility that, although on the input of the oscilloscope a signal is present, the signal is not displayed because the trigger conditions are not met. o Coupling the coupling mode of the synchronization signal: AC eliminates the DC level from the trigger signal. DC the trigger signal 16 has nonzero DC level. Noise Reject the noise is eliminated from trigger signal. HF REJ (High Frequency Reject) it eliminates high frequencies from trigger signal. LF REJ it eliminates low frequencies from trigger signal.

EXPERIMENT NUMBER 2 BASIC OSCILLOSCOPE OPERATIONS

EXPERIMENT NUMBER 2 BASIC OSCILLOSCOPE OPERATIONS 1 EXPERIMENT NUMBER 2 BASIC OSCILLOSCOPE OPERATIONS The oscilloscope is the most versatile and most important tool in this lab and is probably the best tool an electrical engineer uses. This outline guides

More information

The Oscilloscope. Vision is the art of seeing things invisible. J. Swift ( ) OBJECTIVE To learn to operate a digital oscilloscope.

The Oscilloscope. Vision is the art of seeing things invisible. J. Swift ( ) OBJECTIVE To learn to operate a digital oscilloscope. The Oscilloscope Vision is the art of seeing things invisible. J. Swift (1667-1745) OBJECTIVE To learn to operate a digital oscilloscope. THEORY The oscilloscope, or scope for short, is a device for drawing

More information

EENG-201 Experiment # 4: Function Generator, Oscilloscope

EENG-201 Experiment # 4: Function Generator, Oscilloscope EENG-201 Experiment # 4: Function Generator, Oscilloscope I. Objectives Upon completion of this experiment, the student should be able to 1. To become familiar with the use of a function generator. 2.

More information

Sonoma State University Department of Engineering Science Spring 2017

Sonoma State University Department of Engineering Science Spring 2017 EE 110 Introduction to Engineering & Laboratory Experience Saeid Rahimi, Ph.D. Lab 4 Introduction to AC Measurements (I) AC signals, Function Generators and Oscilloscopes Function Generator (AC) Battery

More information

PHYSICS 171 UNIVERSITY PHYSICS LAB II. Experiment 4. Alternating Current Measurement

PHYSICS 171 UNIVERSITY PHYSICS LAB II. Experiment 4. Alternating Current Measurement PHYSICS 171 UNIVERSITY PHYSICS LAB II Experiment 4 Alternating Current Measurement Equipment: Supplies: Oscilloscope, Function Generator. Filament Transformer. A sine wave A.C. signal has three basic properties:

More information

EE 201 Function / Arbitrary Waveform Generator and Oscilloscope Tutorial

EE 201 Function / Arbitrary Waveform Generator and Oscilloscope Tutorial EE 201 Function / Arbitrary Waveform Generator and Oscilloscope Tutorial 1 This is a programmed learning instruction manual. It is written for the Agilent DSO3202A Digital Storage Oscilloscope. The prerequisite

More information

Exercise 4 - THE OSCILLOSCOPE

Exercise 4 - THE OSCILLOSCOPE Exercise 4 - THE OSCILLOSCOPE INTRODUCTION You have been exposed to analogue oscilloscopes in the first year lab. As you are probably aware, the complexity of the instruments, along with their importance

More information

Notes on Experiment #1

Notes on Experiment #1 Notes on Experiment #1 Bring graph paper (cm cm is best) From this week on, be sure to print a copy of each experiment and bring it with you to lab. There will not be any experiment copies available in

More information

Agilent 33522A Function Arbitrary Waveform Generator. Tektronix TDS 3012B Oscilloscope

Agilent 33522A Function Arbitrary Waveform Generator. Tektronix TDS 3012B Oscilloscope Agilent 33522A Function/Arbitrary Waveform Generator and Tektronix TDS 3012B Oscilloscope Agilent 33522A Function Arbitrary Waveform Generator The signal source for this lab is the Agilent 33522A Function

More information

2 Oscilloscope Familiarization

2 Oscilloscope Familiarization Lab 2 Oscilloscope Familiarization What You Need To Know: Voltages and currents in an electronic circuit as in a CD player, mobile phone or TV set vary in time. Throughout the course you will investigate

More information

ECE65 Introduction to the Function Generator and the Oscilloscope Created by: Eldridge Alcantara (Spring 2007)

ECE65 Introduction to the Function Generator and the Oscilloscope Created by: Eldridge Alcantara (Spring 2007) ECE65 Introduction to the Function Generator and the Oscilloscope Created by: Eldridge Alcantara (Spring 2007) I. Getting Started with the Function Generator OUTPUT Red Clip Small Black Clip 1) Turn on

More information

AE Agricultural Customer Services Play-by-Play Tekscope Manual

AE Agricultural Customer Services Play-by-Play Tekscope Manual 1 2012 AE Agricultural Customer Services Play-by-Play Tekscope Manual TABLE OF CONTENTS I. Definitions II. Waveform Properties 1 III. Scientific Notation... 2 IV. Transient Levels of Concern a. ASAE Paper

More information

Combinational logic: Breadboard adders

Combinational logic: Breadboard adders ! ENEE 245: Digital Circuits & Systems Lab Lab 1 Combinational logic: Breadboard adders ENEE 245: Digital Circuits and Systems Laboratory Lab 1 Objectives The objectives of this laboratory are the following:

More information

EE EXPERIMENT 1 (2 DAYS) BASIC OSCILLOSCOPE OPERATIONS INTRODUCTION DAY 1

EE EXPERIMENT 1 (2 DAYS) BASIC OSCILLOSCOPE OPERATIONS INTRODUCTION DAY 1 EE 2101 - EXPERIMENT 1 (2 DAYS) BASIC OSCILLOSCOPE OPERATIONS INTRODUCTION The oscilloscope is the most versatile and most important tool in this lab and is probably the best tool an electrical engineer

More information

Faculty of Engineering, Thammasat University

Faculty of Engineering, Thammasat University Faculty of Engineering, Thammasat University Experiment 6: Oscilloscope (For room 506) Objectives: 1. To familiarize you with the Oscilloscope and Function Generator User Manual: Oscilloscope 1 5 9 4 7

More information

Oscilloscope and Function Generators

Oscilloscope and Function Generators MEHRAN UNIVERSITY OF ENGINEERING AND TECHNOLOGY, JAMSHORO DEPARTMENT OF ELECTRONIC ENGINEERING ELECTRONIC WORKSHOP # 02 Oscilloscope and Function Generators Roll. No: Checked by: Date: Grade: Object: To

More information

Appendix A: Specifications

Appendix A: Specifications All specifications apply to the TDS 200-Series Digital Oscilloscopes and a P2100 probe with the Attenuation switch set to 10X unless noted otherwise. To meet specifications, two conditions must first be

More information

University of California, San Diego Department of Electrical and Computer Engineering

University of California, San Diego Department of Electrical and Computer Engineering University of California, San Diego Department of Electrical and Computer Engineering Part One: Introduction of Lab TAs ECE65, Spring 2007 Lab 0, ECE 65 Lab Orientation 1) James Liao, geniojames@yahoo.com

More information

Experiment 5 The Oscilloscope

Experiment 5 The Oscilloscope Experiment 5 The Oscilloscope Vision is the art of seeing things invisible. J. Swift (1667-1745) OBJECTIVE To learn to operate a cathode ray oscilloscope. THEORY The oscilloscope, or scope for short, is

More information

Sirindhorn International Institute of Technology Thammasat University at Rangsit

Sirindhorn International Institute of Technology Thammasat University at Rangsit Sirindhorn International Institute of Technology Thammasat University at Rangsit School of Information, Computer and Communication Technology COURSE : ECS 210 Basic Electrical Engineering Lab INSTRUCTOR

More information

Getting Started. MSO/DPO Series Oscilloscopes. Basic Concepts

Getting Started. MSO/DPO Series Oscilloscopes. Basic Concepts Getting Started MSO/DPO Series Oscilloscopes Basic Concepts 001-1523-00 Getting Started 1.1 Getting Started What is an oscilloscope? An oscilloscope is a device that draws a graph of an electrical signal.

More information

Introduction to Basic Laboratory Instruments

Introduction to Basic Laboratory Instruments Introduction to Contents: 1. Objectives... 2 2. Laboratory Safety... 2 3.... 2 4. Using a DC Power Supply... 2 5. Using a Function Generator... 3 5.1 Turn on the Instrument... 3 5.2 Setting Signal Type...

More information

The University of Jordan Mechatronics Engineering Department Electronics Lab.( ) Experiment 1: Lab Equipment Familiarization

The University of Jordan Mechatronics Engineering Department Electronics Lab.( ) Experiment 1: Lab Equipment Familiarization The University of Jordan Mechatronics Engineering Department Electronics Lab.(0908322) Experiment 1: Lab Equipment Familiarization Objectives To be familiar with the main blocks of the oscilloscope and

More information

Laboratory 3 (drawn from lab text by Alciatore)

Laboratory 3 (drawn from lab text by Alciatore) Laboratory 3 (drawn from lab text by Alciatore) The Oscilloscope Required Components: 1 10 resistor 2 100 resistors 2 lk resistors 1 2k resistor 2 4.7M resistors 1 0.F capacitor 1 0.1 F capacitor 1 1.0uF

More information

332:223 Principles of Electrical Engineering I Laboratory Experiment #2 Title: Function Generators and Oscilloscopes Suggested Equipment:

332:223 Principles of Electrical Engineering I Laboratory Experiment #2 Title: Function Generators and Oscilloscopes Suggested Equipment: RUTGERS UNIVERSITY The State University of New Jersey School of Engineering Department Of Electrical and Computer Engineering 332:223 Principles of Electrical Engineering I Laboratory Experiment #2 Title:

More information

EC310 Security Exercise 20

EC310 Security Exercise 20 EC310 Security Exercise 20 Introduction to Sinusoidal Signals This lab demonstrates a sinusoidal signal as described in class. In this lab you will identify the different waveform parameters for a pure

More information

Specifications for DS1000CA Series

Specifications for DS1000CA Series Revised December, 2009 RIGOL Specifications for DS1000CA Series All specifications apply to the DS1000CA Series Oscilloscopes unless noted otherwise. To meet these specifications, two conditions must first

More information

UCE-DSO210 DIGITAL OSCILLOSCOPE USER MANUAL. FATIH GENÇ UCORE ELECTRONICS REV1

UCE-DSO210 DIGITAL OSCILLOSCOPE USER MANUAL. FATIH GENÇ UCORE ELECTRONICS REV1 UCE-DSO210 DIGITAL OSCILLOSCOPE USER MANUAL FATIH GENÇ UCORE ELECTRONICS www.ucore-electronics.com 2017 - REV1 Contents 1. Introduction... 2 2. Turn on or turn off... 3 3. Oscilloscope Mode... 3 3.1. Display

More information

Introduction to Oscilloscopes Instructor s Guide

Introduction to Oscilloscopes Instructor s Guide Introduction to Oscilloscopes A collection of lab exercises to introduce you to the basic controls of a digital oscilloscope in order to make common electronic measurements. Revision 1.0 Page 1 of 25 Copyright

More information

N acquisitions, all channels simultaneously, N is selectable from 4, 16, 64, and 128 Inputs

N acquisitions, all channels simultaneously, N is selectable from 4, 16, 64, and 128 Inputs With compliments All specifications apply to the TDS 200-Series Digital Real-Time Oscilloscope with a P2100 probe with the Attenuation switch set to 10X unless noted otherwise. To meet specifications,

More information

Experiment 9 The Oscilloscope and Function Generator

Experiment 9 The Oscilloscope and Function Generator Experiment 9 The Oscilloscope and Function Generator Introduction The oscilloscope is one of the most important electronic instruments available for making circuit measurements. It displays a curve plot

More information

Experiment #2: Introduction to Lab Equipment: Function Generator, Oscilloscope, and Multisim

Experiment #2: Introduction to Lab Equipment: Function Generator, Oscilloscope, and Multisim SCHOOL OF ENGINEERING AND APPLIED SCIENCE DEPARTMENT OF ELECTRICAL AND COMPUTER ENGINEERING ECE 2110: CIRCUIT THEORY LABORATORY Experiment #2: Introduction to Lab Equipment: Function Generator, Oscilloscope,

More information

UNIVERSITY OF CALIFORNIA, SANTA BARBARA Department of Electrical and Computer Engineering. ECE 2A & 2B Laboratory Equipment Information

UNIVERSITY OF CALIFORNIA, SANTA BARBARA Department of Electrical and Computer Engineering. ECE 2A & 2B Laboratory Equipment Information UNIVERSITY OF CALIFORNIA, SANTA BARBARA Department of Electrical and Computer Engineering ECE 2A & 2B Laboratory Equipment Information Table of Contents Digital Multi-Meter (DMM)... 1 Features... 1 Using

More information

LAB I. INTRODUCTION TO LAB EQUIPMENT

LAB I. INTRODUCTION TO LAB EQUIPMENT LAB I. INTRODUCTION TO LAB EQUIPMENT 1. OBJECTIVE In this lab you will learn how to properly operate the basic bench equipment used for characterizing active devices: 1. Oscilloscope (Keysight DSOX 1102A),

More information

Introduction to basic laboratory instruments

Introduction to basic laboratory instruments Introduction to basic laboratory instruments 1. OBJECTIVES... 2 2. LABORATORY SAFETY... 2 3. BASIC LABORATORY INSTRUMENTS... 2 4. USING A DC POWER SUPPLY... 2 5. USING A FUNCTION GENERATOR... 3 5.1 TURN

More information

The oscilloscope and RC filters

The oscilloscope and RC filters (ta initials) first name (print) last name (print) brock id (ab17cd) (lab date) Experiment 4 The oscilloscope and C filters The objective of this experiment is to familiarize the student with the workstation

More information

B. Equipment. Advanced Lab

B. Equipment. Advanced Lab Advanced Lab Measuring Periodic Signals Using a Digital Oscilloscope A. Introduction and Background We will use a digital oscilloscope to characterize several different periodic voltage signals. We will

More information

ME 365 EXPERIMENT 1 FAMILIARIZATION WITH COMMONLY USED INSTRUMENTATION

ME 365 EXPERIMENT 1 FAMILIARIZATION WITH COMMONLY USED INSTRUMENTATION Objectives: ME 365 EXPERIMENT 1 FAMILIARIZATION WITH COMMONLY USED INSTRUMENTATION The primary goal of this laboratory is to study the operation and limitations of several commonly used pieces of instrumentation:

More information

Introduction to Lab Instruments

Introduction to Lab Instruments ECE316, Experiment 00, 2017 Communications Lab, University of Toronto Introduction to Lab Instruments Bruno Korst - bkf@comm.utoronto.ca Abstract This experiment will review the use of three lab instruments

More information

Lab #1 Lab Introduction

Lab #1 Lab Introduction Cir cuit s 212 Lab Lab #1 Lab Introduction Special Information for this Lab s Report Because this is a one-week lab, please hand in your lab report for this lab at the beginning of next week s lab. The

More information

Name EET 1131 Lab #2 Oscilloscope and Multisim

Name EET 1131 Lab #2 Oscilloscope and Multisim Name EET 1131 Lab #2 Oscilloscope and Multisim Section 1. Oscilloscope Introduction Equipment and Components Safety glasses Logic probe ETS-7000 Digital-Analog Training System Fluke 45 Digital Multimeter

More information

UCE-DSO212 DIGITAL OSCILLOSCOPE USER MANUAL. UCORE ELECTRONICS

UCE-DSO212 DIGITAL OSCILLOSCOPE USER MANUAL. UCORE ELECTRONICS UCE-DSO212 DIGITAL OSCILLOSCOPE USER MANUAL UCORE ELECTRONICS www.ucore-electronics.com 2017 Contents 1. Introduction... 2 2. Turn on or turn off... 3 3. Oscilloscope Mode... 4 3.1. Display Description...

More information

Time-Varying Signals

Time-Varying Signals Time-Varying Signals Objective This lab gives a practical introduction to signals that varies with time using the components such as: 1. Arbitrary Function Generator 2. Oscilloscopes The grounding issues

More information

Oscilloscope Operation. Visualizing Signals and Making Measurements

Oscilloscope Operation. Visualizing Signals and Making Measurements Oscilloscope Operation Visualizing Signals and Making Measurements Set Up Oscilloscope Start with the oscilloscope off, with the input plugged into channel one. Press the power button to turn the scope

More information

Lab: INTRODUCTION TO THE WAVEFORM GENERATOR AND THE OSCILLOSCOPE

Lab: INTRODUCTION TO THE WAVEFORM GENERATOR AND THE OSCILLOSCOPE Name EET101/Lab#5; EET121/Lab#5; EGR104/Lab#3 Sec / Night Date Lab Partner(s) Name(s) Lab: INTRODUCTION TO THE WAVEFORM GENERATOR AND THE OSCILLOSCOPE Objectives: Each student will: 1. Know the function

More information

DIGITAL STORAGE OSCILLOSCOPES

DIGITAL STORAGE OSCILLOSCOPES DIGITAL STORAGE OSCILLOSCOPES Electronic Measurements Lab Massimo Ortolano 2016 POLITECNICO DI TORINO c 2011 2016 Massimo Ortolano Dipartimento di Elettronica e Telecomunicazioni (DET) Politecnico di Torino

More information

Laboratory equipments. Parameters of digital signals.

Laboratory equipments. Parameters of digital signals. Laboratory 1 Laboratory equipments. Parameters of digital signals. 1.1 Objectives This laboratory presents detailed description of the equipments used during the lab and measurement techniques specifically

More information

LAB I. INTRODUCTION TO LAB EQUIPMENT

LAB I. INTRODUCTION TO LAB EQUIPMENT 1. OBJECTIVE LAB I. INTRODUCTION TO LAB EQUIPMENT In this lab you will learn how to properly operate the oscilloscope Agilent MSO6032A, the Keithley Source Measure Unit (SMU) 2430, the function generator

More information

This section describes the basic operation including start-up. See Section 2, Explanation of Functions for details.

This section describes the basic operation including start-up. See Section 2, Explanation of Functions for details. 1 Section 1 Basic Operation This section describes the basic operation including start-up. See Section 2, Explanation of Functions for details. 2 Appearance ƒ Front View Power switch Input terminal ƒ External

More information

Data Sheet. Digital Storage Oscilloscope. Features & Benefits. Applications. Ease-of-Use Feature DSO5202BMT DSO5102BMT DSO5062BMT

Data Sheet. Digital Storage Oscilloscope. Features & Benefits. Applications. Ease-of-Use Feature DSO5202BMT DSO5102BMT DSO5062BMT Data Sheet Digital Storage Oscilloscope DSO5202BMT DSO5102BMT DSO5062BMT Features & Benefits 200/100/60MHz Bandwidths 1GSa/s Real Time Sample Rate 2M Memory Depth Trigger mode: Edge, Pulse Width, Video,

More information

Department of Electrical and Computer Engineering. Laboratory Experiment 1. Function Generator and Oscilloscope

Department of Electrical and Computer Engineering. Laboratory Experiment 1. Function Generator and Oscilloscope Department of Electrical and Computer Engineering Laboratory Experiment 1 Function Generator and Oscilloscope The purpose of this first laboratory assignment is to acquaint you with the function generator

More information

Chapter 5 : Specifications

Chapter 5 : Specifications Chapter 5 : Specifications All specifications apply to the DS1000B Series Oscilloscopes and a probe with the Attenuation switch set to 10X unless noted otherwise. To meet these specifications, two conditions

More information

Oscilloscope. 1 Introduction

Oscilloscope. 1 Introduction Oscilloscope Equipment: Capstone, BK Precision model 2120B oscilloscope, Wavetek FG3C function generator, 2-3 foot coax cable with male BNC connectors, 2 voltage sensors, 2 BNC banana female adapters,

More information

User s Guide RIGOL. DS1000CA Series Digital Oscilloscopes DS1302CA, DS1202CA, DS1102CA, DS1062CA. Publication number DS1000CA April 2008

User s Guide RIGOL. DS1000CA Series Digital Oscilloscopes DS1302CA, DS1202CA, DS1102CA, DS1062CA. Publication number DS1000CA April 2008 User s Guide RIGOL Publication number DS1000CA-080512 April 2008 DS1000CA Series Digital Oscilloscopes DS1302CA, DS1202CA, DS1102CA, DS1062CA Copyright 1998-2008 RIGOL Technologies, Inc. All Rights Reserved

More information

Laboratory Equipment Instruction Manual 2011

Laboratory Equipment Instruction Manual 2011 University of Toronto Department of Electrical and Computer Engineering Instrumentation Laboratory GB341 Laboratory Equipment Instruction Manual 2011 Page 1. Wires and Cables A-2 2. Protoboard A-3 3. DC

More information

Oscilloscope Fundamentals. For Electrical Engineering and Physics Undergraduate Students

Oscilloscope Fundamentals. For Electrical Engineering and Physics Undergraduate Students Oscilloscope Fundamentals For Electrical Engineering and Physics Undergraduate Students Agenda What is an oscilloscope? Probing basics (low-frequency model) Making voltage and timing measurements Properly

More information

LABORATORY 4. Palomar College ENGR210 Spring 2017 ASSIGNED: 3/21/17

LABORATORY 4. Palomar College ENGR210 Spring 2017 ASSIGNED: 3/21/17 LABORATORY 4 ASSIGNED: 3/21/17 OBJECTIVE: The purpose of this lab is to evaluate the transient and steady-state circuit response of first order and second order circuits. MINIMUM EQUIPMENT LIST: You will

More information

CHAPTER 6. Motor Driver

CHAPTER 6. Motor Driver CHAPTER 6 Motor Driver In this lab, we will construct the circuitry that your robot uses to drive its motors. However, before testing the motor circuit we will begin by making sure that you are able to

More information

54645D. Mixed Signal Oscilloscope

54645D. Mixed Signal Oscilloscope 54645D Mixed Signal Oscilloscope Page 1 of 42 Instructions for the use of the 54645D Mixed Signal Oscilloscope This pamphlet is intended to give you (the student) an overview on the use of the 54645D Mixed

More information

DSO5000P Series Digital Storage Oscilloscope User Manual. (Version 1.1)

DSO5000P Series Digital Storage Oscilloscope User Manual. (Version 1.1) DSO5000P Series Digital Storage Oscilloscope User Manual (Version 1.1) Contents Contents Contents... i Chapter 1 Safety Tips... 1 1.1 General Safety Summary... 1 1.2 Safety Terms and Symbols... 2 1.3 Terms

More information

AP034-OM-E Rev D ISSUED: January 2000 ²

AP034-OM-E Rev D ISSUED: January 2000 ² 3HUIRUPDQFH9HULILFDWLRQ 3HUIRUPDQFH9HULILFDWLRQ This procedure can be used to verify the warranted characteristics of the AP034 Active Differential Probe. The recommended calibration interval for the model

More information

MSO-5000B Mixed Storage Oscilloscope User Manual

MSO-5000B Mixed Storage Oscilloscope User Manual MSO-5000B Mixed Storage Oscilloscope User Manual Contents Contents CONTENTS... I COPYRIGHT DECLARATION... IV CHAPTER 1 SAFETY TIPS... 1 1.1 GENERAL SAFETY SUMMARY... 1 1.2 SAFETY TERMS AND SYMBOLS... 2

More information

University of Utah Electrical & Computer Engineering Department ECE 2210/2200 Lab 4 Oscilloscope

University of Utah Electrical & Computer Engineering Department ECE 2210/2200 Lab 4 Oscilloscope University of Utah Electrical & Computer Engineering Department ECE 2210/2200 Lab 4 Oscilloscope Objectives 1 Introduce the Oscilloscope and learn some uses. 2 Observe Audio signals. 3 Introduce the Signal

More information

Lab #5 Steady State Power Analysis

Lab #5 Steady State Power Analysis Lab #5 Steady State Power Analysis Steady state power analysis refers to the power analysis of circuits that have one or more sinusoid stimuli. This lab covers the concepts of RMS voltage, maximum power

More information

User Manual Series. Digital Storage Oscilloscope 6810, 6806, March Copyright Protek Test & Measurement 2005 All Rights Reserved

User Manual Series. Digital Storage Oscilloscope 6810, 6806, March Copyright Protek Test & Measurement 2005 All Rights Reserved User Manual March 2005 6800 Series Digital Storage Oscilloscope 6810, 6806, 6804 Copyright Protek Test & Measurement 2005 All Rights Reserved Copyright Protek Test & Measurement 2005 All Rights Reserved.

More information

User s Guide RIGOL. DS1000E, DS1000D Series Digital Oscilloscopes DS1102E, DS1052E, DS1102D, DS1052D. Publication number UGA July 2009

User s Guide RIGOL. DS1000E, DS1000D Series Digital Oscilloscopes DS1102E, DS1052E, DS1102D, DS1052D. Publication number UGA July 2009 User s Guide RIGOL Publication number UGA07111-1110 July 2009 DS1000E, DS1000D Series Digital Oscilloscopes DS1102E, DS1052E, DS1102D, DS1052D All Rights Reserved 1. All Rights Reserved 2. RIGOL products

More information

Introduction to basic laboratory instruments

Introduction to basic laboratory instruments BEE 233 Laboratory-1 Introduction to basic laboratory instruments 1. Objectives To learn safety procedures in the laboratory. To learn how to use basic laboratory instruments: power supply, function generator,

More information

Electric Circuit II Lab Manual Session #1

Electric Circuit II Lab Manual Session #1 Department of Electrical Engineering Electric Circuit II Lab Manual Session #1 Subject Lecturer Dr. Yasser Hegazy Name:-------------------------------------------------- Group:--------------------------------------------------

More information

DS1000B Series Digital Oscilloscopes

DS1000B Series Digital Oscilloscopes Product Overview DS1000B series oscilloscopes are designed with four analog channels and 1 external trigger channel, which can capture multi-channel signal simultaneously and meet industrial needs. The

More information

University of Saskatchewan Electrical Engineering Laboratory Equipment Manual. Analog Meters

University of Saskatchewan Electrical Engineering Laboratory Equipment Manual. Analog Meters Analog Meters There are a wide variety of analog meters available in the labs. There are ac & dc voltmeters, ac & dc ammeters, ac & dc wattmeters, and several true rms voltmeters. They are all designed

More information

LAB II. INTRODUCTION TO LAB EQUIPMENT

LAB II. INTRODUCTION TO LAB EQUIPMENT 1. OBJECTIVE LAB II. INTRODUCTION TO LAB EQUIPMENT In this lab you will learn how to properly operate the oscilloscope Keysight DSOX1102A, the Keithley Source Measure Unit (SMU) 2430, the function generator

More information

UTD1000 User Manual. Table of Contents

UTD1000 User Manual. Table of Contents Table of Contents Introduction : UTD1000 Series Digital Storage Oscilloscope Chapter 1 User Guide 1. Getting to know your UTD1000 Digital Storage Oscilloscope 2. General Inspection 3. Functional Check

More information

Measurement Bench. Accessories. Power supply. Wave form generator. Multimetre. Oscilloscope. Dr. L.Scucchia

Measurement Bench. Accessories. Power supply. Wave form generator. Multimetre. Oscilloscope. Dr. L.Scucchia Measurement Bench Accessories Power supply Wave form generator Multimetre Oscilloscope OSCILLOSCOPE Oscilloscope (1) The oscilloscope allows to display a voltage (vertical axis - Y axis) versus time (horizontal

More information

DST Series B Type Digital Storage Oscilloscope User Manual

DST Series B Type Digital Storage Oscilloscope User Manual DST Series B Type Digital Storage Oscilloscope User Manual Contents Contents Contents... i Copyright Declaration... iv Chapter 1 Safety Tips... 1 1.1 General Safety Summary... 1 1.2 Safety Terms and Symbols...

More information

ENGR 210 Lab 6 Use of the Function Generator & Oscilloscope

ENGR 210 Lab 6 Use of the Function Generator & Oscilloscope ENGR 210 Lab 6 Use of the Function Generator & Oscilloscope In this laboratory you will learn to use two additional instruments in the laboratory, namely the function/arbitrary waveform generator, which

More information

PHY152 Experiment 4: Oscillations in the RC-Circuits (Measurements with an oscilloscope)

PHY152 Experiment 4: Oscillations in the RC-Circuits (Measurements with an oscilloscope) PHY152 Experiment 4: Oscillations in the RC-Circuits (Measurements with an oscilloscope) If you have not used an oscilloscope before, the web site http://www.upscale.utoronto.ca/generalinterest/harrison/oscilloscope/oscilloscope.html

More information

User s Guide RIGOL. DS1000E, DS1000D Series Digital Oscilloscopes DS1102E, DS1052E, DS1102D, DS1052D. Publication number UGA July 2008

User s Guide RIGOL. DS1000E, DS1000D Series Digital Oscilloscopes DS1102E, DS1052E, DS1102D, DS1052D. Publication number UGA July 2008 User s Guide RIGOL Publication number UGA07107-1110 July 2008 DS1000E, DS1000D Series Digital Oscilloscopes DS1102E, DS1052E, DS1102D, DS1052D All Rights Reserved All Rights Reserved RIGOL products are

More information

User Manual RIGOL. DS1000KCA Series Digital Oscilloscopes DS1302CA, DS1202CA, DS1102CA, DS1062CA. Publication number DS1KCA June 2007

User Manual RIGOL. DS1000KCA Series Digital Oscilloscopes DS1302CA, DS1202CA, DS1102CA, DS1062CA. Publication number DS1KCA June 2007 User Manual RIGOL Publication number DS1KCA-070712 June 2007 DS1000KCA Series Digital Oscilloscopes DS1302CA, DS1202CA, DS1102CA, DS1062CA Copyright RIGOL Technologies, Inc. 2007 All Rights Reserved Copyright

More information

Get started with DSO-X 2014A

Get started with DSO-X 2014A Get started with DSO-X 2014A The oscilloscope has some built-in training signals! Connect two measurement cables with probes to the demo-connectors. Turn the oscilloscope on. Press Default Setup settings

More information

LAB #7: Digital Signal Processing

LAB #7: Digital Signal Processing LAB #7: Digital Signal Processing Equipment: Pentium PC with NI PCI-MIO-16E-4 data-acquisition board NI BNC 2120 Accessory Box VirtualBench Instrument Library version 2.6 Function Generator (Tektronix

More information

MULT SWP X1K K VERN START FREQ DURATION AMPLITUDE 0 TTL OUT RAMP

MULT SWP X1K K VERN START FREQ DURATION AMPLITUDE 0 TTL OUT RAMP Signal Generators This document is a quick reference guide to the operation of the signal generators available in the laboratories. Major functions will be covered, but some features such as their sweep

More information

EET 223 RF COMMUNICATIONS LABORATORY EXPERIMENTS

EET 223 RF COMMUNICATIONS LABORATORY EXPERIMENTS EET 223 RF COMMUNICATIONS LABORATORY EXPERIMENTS Experimental Goals A good technician needs to make accurate measurements, keep good records and know the proper usage and limitations of the instruments

More information

Experiment # 1 Introduction to Lab Equipment

Experiment # 1 Introduction to Lab Equipment Experiment # 1 Introduction to Lab Equipment 1. Synopsis: In this introductory lab, we will review the basic concepts of digital logic design and learn how to use the equipment available in the laboratory.

More information

DEPARTMENT OF THE ARMY TECHNICAL BULLETIN

DEPARTMENT OF THE ARMY TECHNICAL BULLETIN *TB 9-6625-2240-35 DEPARTMENT OF THE ARMY TECHNICAL BULLETIN CALIBRATION PROCEDURE FOR OSCILLOSCOPE OS261U (TEKTRONIX, TYPE 475), OS261A(V)1U (TEKTRONIX, TYPE 475 OPTION 7), OS261B(V)1U (TEKTRONIX, TYPE

More information

Lab 0: Introduction to basic laboratory instruments. Revised by Dan Hoang & Tai-Chang Chen 03/30/2009

Lab 0: Introduction to basic laboratory instruments. Revised by Dan Hoang & Tai-Chang Chen 03/30/2009 Lab 0: Introduction to basic laboratory instruments Revised by Dan Hoang & Tai-Chang Chen 03/30/2009 1. Objectives 1. To learn safety procedures in the laboratory. 2. To learn how to use basic laboratory

More information

Performance Characteristics

Performance Characteristics The performance characteristics describe the typical performance of the oscilloscope. You will notice that some of the characteristics are marked as tested, these are values that you can verify with the

More information

Exercise 8 Measurements of time and frequency.

Exercise 8 Measurements of time and frequency. Exercise 8 Measurements of time and frequency. 1. Aim of the exercise The aim of the exercise is to familiarize students with methods of measuring time parameters of electrical signals such as frequency,

More information

OSCILLOSCOPES. Oscilloscopes CS-5400 SERIES CS-5400/5450 FEATURES OUTLINE CS-5400

OSCILLOSCOPES. Oscilloscopes CS-5400 SERIES CS-5400/5450 FEATURES OUTLINE CS-5400 99 Washington Street Melrose, MA 02176 Fax 781-665-0780 TestEquipmentDepot.com Oscilloscopes 100MHz 3-Channel Oscilloscope (With Digital Readout / Cursor) CS-5400 100MHz 3-Channel Oscilloscope CS-5405

More information

User s Guide RIGOL. DS1000CA Series Digital Oscilloscopes DS1302CA, DS1202CA, DS1102CA, DS1062CA. Publication number UGA Oct.

User s Guide RIGOL. DS1000CA Series Digital Oscilloscopes DS1302CA, DS1202CA, DS1102CA, DS1062CA. Publication number UGA Oct. User s Guide RIGOL Publication number UGA03106-1110 Oct. 2008 DS1000CA Series Digital Oscilloscopes DS1302CA, DS1202CA, DS1102CA, DS1062CA 2007 RIGOL Technologies, Inc. All Rights Reserved. 2007 RIGOL

More information

2 AC and RMS. To pass this lab you must solve tasks 1-2. Tasks 3 and 4 are included in the grading of the course.

2 AC and RMS. To pass this lab you must solve tasks 1-2. Tasks 3 and 4 are included in the grading of the course. 2 AC and RMS Purpose of the lab: to familiarize yourself with the oscilloscope to familiarize yourself with AC voltages and different waveforms to study RMS and average values In this lab, you have the

More information

LAB INSTRUMENTATION. RC CIRCUITS.

LAB INSTRUMENTATION. RC CIRCUITS. LAB INSTRUMENTATION. RC CIRCUITS. I. OBJECTIVE a) Becoming accustomed to using the lab instrumentation (voltage supply, digital multimeter, signal generator, oscilloscope) necessary to the experimental

More information

EXPERIMENT 1 PRELIMINARY MATERIAL

EXPERIMENT 1 PRELIMINARY MATERIAL EXPERIMENT 1 PRELIMINARY MATERIAL BREADBOARD A solderless breadboard, like the basic model in Figure 1, consists of a series of square holes, and those columns of holes are connected to each other via

More information

Function Generator Guide Tektronix AFG3102

Function Generator Guide Tektronix AFG3102 Tektronix AFG3102 ersion 2008-Jan-1 Dept. of Electrical & Computer Engineering Portland State University Copyright 2008 Portland State University 1 Basic Information This guide provides basic instructions

More information

DEPARTMENT OF INFORMATION ENGINEERING. Test No. 1. Introduction to Scope Measurements. 1. Correction. Term Correction. Term...

DEPARTMENT OF INFORMATION ENGINEERING. Test No. 1. Introduction to Scope Measurements. 1. Correction. Term Correction. Term... 2. Correction. Correction Report University of Applied Sciences Hamburg Group No : DEPARTMENT OF INFORMATION ENGINEERING Laboratory for Instrumentation and Measurement L: in charge of the report Test No.

More information

MASSACHUSETTS INSTITUTE OF TECHNOLOGY /6.071 Introduction to Electronics, Signals and Measurement Spring 2006

MASSACHUSETTS INSTITUTE OF TECHNOLOGY /6.071 Introduction to Electronics, Signals and Measurement Spring 2006 MASSACHUSETTS INSTITUTE OF TECHNOLOGY.071/6.071 Introduction to Electronics, Signals and Measurement Spring 006 Lab. Introduction to signals. Goals for this Lab: Further explore the lab hardware. The oscilloscope

More information

Tektronix digital oscilloscope, BK Precision Function Generator, coaxial cables, breadboard, the crystal earpiece from your AM radio kit.

Tektronix digital oscilloscope, BK Precision Function Generator, coaxial cables, breadboard, the crystal earpiece from your AM radio kit. Experiment 0: Review I. References The 174 and 275 Lab Manuals Any standard text on error analysis (for example, Introduction to Error Analysis, J. Taylor, University Science Books, 1997) The manual for

More information

Agilent 33220A Function Generator Tutorial

Agilent 33220A Function Generator Tutorial Contents UNIVERSITY OF CALIFORNIA AT BERKELEY College of Engineering Department of Electrical Engineering and Computer Sciences EE105 Lab Experiments Agilent 33220A Function Generator Tutorial 1 Introduction

More information

HP 16533A 1-GSa/s and HP 16534A 2-GSa/s Digitizing Oscilloscope

HP 16533A 1-GSa/s and HP 16534A 2-GSa/s Digitizing Oscilloscope User s Reference Publication Number 16534-97009 February 1999 For Safety Information, Warranties, and Regulatory Information, see the pages behind the Index Copyright Hewlett-Packard Company 1991 1999

More information

RIGOL Data Sheet. DS1000E, DS1000D Series Digital Oscilloscopes DS1102E, DS1052E, DS1102D, DS1052D. Product Overview. Easy to Use Design.

RIGOL Data Sheet. DS1000E, DS1000D Series Digital Oscilloscopes DS1102E, DS1052E, DS1102D, DS1052D. Product Overview. Easy to Use Design. RIGOL Data Sheet DS1000E, DS1000D Series Digital Oscilloscopes DS1102E, DS1052E, DS1102D, DS1052D Product Overview The DS1000E, DS1000D series instruments are economical, high-performance digital oscilloscopes.

More information

EE-4022 Experiment 2 Amplitude Modulation (AM)

EE-4022 Experiment 2 Amplitude Modulation (AM) EE-4022 MILWAUKEE SCHOOL OF ENGINEERING 2015 Page 2-1 Student objectives: EE-4022 Experiment 2 Amplitude Modulation (AM) In this experiment the student will use laboratory modules to implement operations

More information