CLAMP-ON METERS THIS PUBLICATION SUPERSEDES NAVAIR 17-20AQ-32 DATED 1 OCTOBER 2009

Size: px
Start display at page:

Download "CLAMP-ON METERS THIS PUBLICATION SUPERSEDES NAVAIR 17-20AQ-32 DATED 1 OCTOBER 2009"

Transcription

1 TECHNICAL MANUAL AQ-32 INSTRUMENT CALIBRATION PROCEDURE CLAMP-ON METERS THIS PUBLICATION SUPERSEDES DATED 1 OCTOBER 2009 DISTRIBUTION STATEMENT C. DISTRIBUTION AUTHORIZED TO U.S. GOVERNMENT AGENCIES AND THEIR CONTRACTORS TO PROTECT PUBLICATIONS REQUIRED FOR OFFICIAL USE OR FOR ADMINISTRATIVE OR OPERATIONAL PURPOSES ONLY. OTHER REQUESTS FOR THIS DOCUMENT SHALL BE REFERRED TO MEASUREMENT SCIENCE DEPARTMENT, CORONA DIVISION, NAVAL SURFACE WARFARE CENTER, P.O. BOX 5000, CORONA, CA AQ-32 DESTRUCTION NOTICE - FOR UNCLASSIFIED, LIMITED DOCUMENTS, DESTROY BY ANY METHOD THAT WILL PREVENT DISCLOSURE OF CONTENTS OR RECONSTRUCTION OF THE DOCUMENT. PUBLISHED BY DIRECTION OF COMMANDER NAVAL AIR SYSTEMS COMMAND 1 MAY 2012

2 LIST OF EFFECTIVE PAGES The purpose of this list is to identify the pages in this document affected by any technical content changes made since the previous release of the document. NOTE: On a revised page, with the exception of the Title, the A, and the i pages, the technical changes are indicated by a vertical line in the outer margin of the page. Page No. ** Revision* Page No. ** Revision* Title A 0 A-1 to A-5 0 i 0 B-1 to B to to to to to to to Page No. ** Revision* 71 0 *Zero in this column indicates an original page. R in this column indicates a revised page. **All pages not listed are blank pages. A

3 TABLE OF CONTENTS Section Page 1 Introduction and Description Equipment Requirements Preliminary Operations Calibration Process Alternating Voltage Test Direct Voltage Tests Alternating Current Test Direct Current Test Ohmmeter Test... 9 Navy Calibration Checklist Appendix A: Representative Test Instruments... A-1 Appendix B: Current Coils and Test Points... B-1 ILLUSTRATIONS Figure Page 1 Current Test Configuration Using a Current Shunt Test Configuration Using an Alternating Current Source with Current Transformer... 7 TABLES Table Page 1 Calibration Description Equipment Requirements... 2 i

4 SECTION 1 INTRODUCTION AND DESCRIPTION 1.1 This procedure describes the calibration of the Clamp-On Meters. The Clamp-On Meters may also have voltages/currents and resistance measuring capabilities that are covered by this procedure. Instruments that can be calibrated by this procedure include, but are not limited to, the items listed in Appendix A. The instrument being calibrated is referred to herein as the TI (Test Instrument). 1.2 All comments concerning this procedure should be directed to the Measurement Science Department, Corona Division, Naval Surface Warfare Center, P.O. Box 5000, Corona, CA This procedure includes tests of essential performance parameters only. Any malfunction noticed during calibration, whether specifically tested for or not, should be corrected. NOTE For AC current frequencies >400 Hz, the Weston 327 Type 2 transformer and Ammeter should not be used. The current shunt and digital multimeter (frequency range from 50 Hz to 1 khz) are intended as an alternate to the current transformer and ammeter (frequency range from 25 to 500 Hz). The Altec Lansing 1570BN was retained as a listed standard since the 1570BN will tolerate loads as small as 0.8. If the 1590E functions correctly, it is also acceptable (see Appendix B). There are also compliance voltage limitations of the 5700AAN and 5725A (or substitutes). The AC current waveform is sine wave. It is acceptable to change the test points for the higher value currents at frequencies of 400 Hz or greater due to compliance voltage limitations. Refer to Appendix B for expanded information. TI Characteristics Alternating voltage Direct voltage Alternating current Direct current Resistance Table 1. Calibration Description Performance Specifications Range: 0 to 1000 V ac Tolerance: ±1% iv or greater Range: 0 to 1000 V dc Tolerance: ±0.8% iv or greater Range: 0 to Hz, 0 to Hz Tolerance: ±1% iv or greater Range: 0 to 500 A Tolerance: ±2% iv or greater Range: 0 to 2000 Tolerance: ±1% iv or greater Test Method Voltage is applied, adjusted to the desired TI indication, and measured. Voltage is applied, adjusted to the desired TI indication, and measured. Comparison with a metered current source, if required, and using a multiturn coil as a current multiplier. Comparison with a metered current source, if required, and using a multiturn coil as a current multiplier. Measurement of a decade resistor. 1

5 SECTION 2 EQUIPMENT REQUIREMENTS NOTES Minimum use specifications are the principal parameters required for performance of the calibration, and are included to assist in the selection of alternate equipment, which may be used at the discretion of the using laboratory. Satisfactory performance of alternate items shall be verified prior to use. All applicable equipment must bear evidence of current calibration. The instruments utilized in this procedure were selected from those known to be available at Navy calibration facilities, and the listing by make or model number carries no implication of preference, recommendation, or approval for use by other agencies. It is recognized that equivalent equipment produced by other manufacturers may be capable of equally satisfactory performance in this procedure. Item Table 2. Equipment Requirements Minimum Use Specifications 2.1 Alternating voltage source Range: 0 to 1000 V ac Uncertainty: ±0.25% Frequency: 60 and 400 Hz 2.2 Direct voltage source Range: 0 to 1000 V dc Uncertainty: ±0.2% iv 2.3 Alternating current source Range: 0 to 10 A minimum Uncertainty: ±0.25% iv Frequency: 50, 60, 400 Hz, 1 khz 2.4 Direct current source Range: 0 to 10 A minimum Uncertainty: ±0.5% iv 2.5 Resistance standard Range: 0 to 2000 Uncertainty: ±0.25% iv 2.6 Current multiplying coil (may be locally fabricated or purchased) To fabricate current-multiplying coil with local supply wire, possible sizes and current ratings are: 14 gauge wire to 20A 10 gauge wire to 35A 8 gauge wire to 50A 4 gauge wire to 100A 2.7 Clamp To connect cables together with minimum resistance 2.8 Digital multimeter (DMM) Voltage range: 0 to 2 V ac Uncertainty: 0.25% iv Frequency range: 50 Hz to 1 khz 2.9 Current shunt Current range: 0 to 10 A minimum Uncertainty: 0.25% iv Frequency range: DC, 50 Hz, 60 Hz, 1 khz Calibration Equipment Fluke 5700AAN Fluke 5700AAN Fluke 5700AAN with 5725A; or Altec 1570BN with Weston and Weston 327 Type 2; or Hewlett-Packard 34401A with Electrical Instrument Service SR15 Fluke 5700AAN with 5725A Fluke 5700AAN; or QuadTech ; or GenRad 1433W Local supply for wire or Fluke 5500A/COIL (50 turn coil) Local supply Hewlett-Packard 34401A Electrical Instrument Service SR15 2

6 SECTION 3 PRELIMINARY OPERATIONS 3.1 Select the appropriate equipment for the measurement system to be used. 3.2 Set all auxiliary equipment controls for minimum output to avoid damage to the equipment and so that dangerous voltages will not be present on the output terminal when the power switches are turned on. 3.3 Connect all auxiliary equipment to the appropriate power source. 3.4 Turn all power switches on, and allow the equipment to warm up. 3.5 If a checklist for the TI is not provided, select the calibration points as follows (exceptions are allowed for alternating current): Select graduations on suitable alternating and direct voltage ranges as calibration points. If the TI has current capability, select three graduations on suitable alternating and direct current ranges as calibration points, ensuring that all test values are less than or equal to the current limits listed in Section If the scales are linear for voltage, current or resistance, the cardinal graduations may be used. If the scales are non-linear, select graduations at deflections of approximately one-third and two-thirds of fullscale, and at fullscale, ensuring that test points for current values are less than or equal to the current limits listed in Section Ensure that the TI meter is mechanically zeroed. 3.7 Ensure that the TI jaws are clean and free of foreign substance and that they close properly. 3

7 SECTION 4 CALIBRATION PROCESS WARNINGS THIS PROCEDURE INVOLVES VOLTAGES THAT ARE HAZARDOUS TO PERSONNEL. USE EXTREME CAUTION. THIS PROCEDURE INVOLVES CURRENTS THAT ARE HAZARDOUS TO PERSONNEL. USE EXTREME CAUTION. NOTE Unless otherwise specified, verify the results of each test and take corrective action whenever the test requirement is not met, before proceeding. 4.1 ALTERNATING VOLTAGE TEST Set the TI selector switch for alternating voltage measurement Set the alternating voltage source (item 2.1) controls for 60 or 400 Hz alternating voltage output as required for the TI If applicable, perform the alternating voltage source meter electrical zero adjustment For each alternating voltage calibration point selected, perform a measurement as follows: Connect the TI to the alternating voltage source using the appropriate terminals If applicable, set the TI range switch to the appropriate voltage range Set the alternating voltage source controls to the voltage value selected as the calibration point Adjust the alternating voltage source output controls for the required TI nominal value selected as the calibration point. NOTE If a meter calibrator having a percent error meter is being used and the error exceeds its range, adjust the calibrator controls for the exact calibration point and verify that the TI indication is within the tolerance limits Verify that the alternating voltage source indicates within the tolerance limits for the selected calibration point. WARNING REDUCE THE ALTERNATING VOLTAGE SOURCE OUTPUT TO MINIMUM BEFORE CHANGING TI CONNECTIONS OR SWITCHING TO A LOWER RANGE Adjust the alternating voltage source controls for minimum output Unless other measurements are to be performed, skip to step

8 4.2 DIRECT VOLTAGE TESTS Set the TI selector switch for direct voltage measurements Set the direct voltage source (item 2.2) controls for direct voltage output If applicable, perform the direct voltage source meter electrical zero adjustment. If applicable, perform DC zero adjustments on the TI Perform a measurement as follows for each direct voltage calibration point selected: Connect the TI to the direct voltage source using the appropriate terminals If applicable, set the TI range switch to the appropriate voltage range Set the direct voltage source controls to the voltage value selected as the calibration point Adjust the direct voltage source output controls for the required TI nominal value selected as the calibration point. NOTE If a meter calibrator having a percent error meter is being used and the error exceeds its range, adjust the calibrator controls for the exact calibration point and verify that the TI indication is within the tolerance limits Verify that the direct voltage source indicates within the tolerance limits for the selected calibration point. WARNING REDUCE THE DIRECT VOLTAGE SOURCE OUTPUT TO MINIMUM BEFORE CHANGING TI CONNECTIONS, OR SWITCHING TO A LOWER RANGE Adjust the direct voltage source controls for minimum output Unless other measurements are to be performed, skip to step ALTERNATING CURRENT TEST Set the TI selector switch for alternating current measurements. NOTE The technician is given the choice of 3 methods. The first method uses a single wire. The second method uses the Fluke 5500A/COIL (50 turn current coil). The third method allows the technician to construct a multiturn coil. After the wire (or coil) is chosen, the technician is told to size the current for the test points (the 50 turn coil or the constructed coil are current multipliers, the single wire is not). The 3 methods are equivalent. The calibration technician is free to use whatever method is convenient for the test point desired. 5

9 NOTES The technician may either construct a multi turn coil (see Appendix B) as item 2.6 or use the Fluke 5500A/COIL (a 50 turn coil listed as an accessory to the Fluke 5500A calibrator). It is recognized that there will be compliance voltage limitations when the 5700AAN and the 5725A are used as the current source. It is anticipated that the compliance voltage limitations will be more pronounced at the 400 Hz and higher frequencies for larger currents. For these reasons, the calibration technician is allowed to modify the current check points when compliance voltage is a problem. Refer to Appendix B for expanded instructions. The current shunt (item 2.9) and the digital multimeter (item 2.8) may be used in place of the 327 Type 2 current transformer and the ammeter. In Figure 2, the primary of the 327 Type 2 transformer is connected in series with the amplifier output and the coil. The shunt current terminals are to be connected in place of the primary terminals of the 327 Type 2 current transformer. The shunt output voltage terminals are to be connected to the voltage inputs of the multimeter. Set the multimeter to measure V ac. Figure 2 shows the configuration using the 327 Type 2 transformer and ammeter If applicable, perform the alternating current source (item 2.3) meter electrical zero adjustment. Connect the equipment, except the coil, as shown in Figure 1 or Figure For the TI, find the ac current ranges from Appendix A or the checklist, if it exists. NOTE If using Appendix A to find the test points, it is only necessary to use 3 test points for each current range at just one of the test frequencies used. For testing at the other frequencies, one test point on each range is sufficient. For testing at 50 Hz or 60 Hz, the 3 test points should be spread across each current range (suggested points are at 20 to 30% of range, 40 to 60% or range, and 90 to 100% of range). At 400 Hz or higher, the test point (or points) in each of the current ranges should be above 50% of the range if possible If compliance voltage problems occur (particularly with the 5700AAN and 5725A) when testing over 100 Hz, reduce the higher current test point magnitudes until the compliance voltage problem ceases. In this case, the SPECIAL CALIBRATION label is not necessary because the current ranges at the 60 Hz or 50 Hz frequencies were tested If the TI current test point (or range) is less than the output range of item 2.3 or item 2.4 standards, a single wire (through the center of the TI jaw area) may be used. If the TI current range is greater than the output range of item 2.3 or item 2.4 standards, a multiturn coil must be used. The multiturn coil may be either the Fluke 5500A/COIL or a coil constructed as described in Appendix B. 2.3 OR 2.4 CURRENT SOURCE 2.8 DMM 2.9 CURRENT SHUNT COIL TI JAWS MULTITURN COIL TI Figure 1. Current Test Configuration Using a Current Shunt 6

10 OSCILLATOR (200 CD) AMPLIFIER (1570 BN) CURRENT TRANSFORMER (327 TYPE 2) AC/DC AMMETER ( ) 2.3 ALTERNATING CURRENT SOURCE 100 A ± 50 A SEC 20 A 10 A NOTE THE WESTON 327 TYPE 2 CURRENT TRANSFORMER OUTPUT IS 5 A WITH MAXIMUM RATED INPUT AT ANY TAP. COIL TI Figure 2. Test Configuration Using an Alternating Current Source with Current Transformer Size the coil current, connect the coil or wire to the TI, and perform the TI calibration test points as follows: If a single wire is being used, then the wire current is the same as the current for the TI test points. Use the appropriate size of insulated wire (item 2.6) needed for the current used (through the single wire). Then connect the ends of the single wire as shown in Figure 1 or Figure 2. One end will be to the current source and the other end will be to the primary of the current transformer if Figure 2 is used. Test Configuration shown in Figure 1 is used if a current shunt is used as the standard. NOTE The current through the Fluke 5500A/COIL coil is not the same as the TI test value. Divide each TI test point current used by 50 to get each current through the 5500A/COIL If the Fluke 5500A/COIL is to be used, connect the primary of the Fluke 5500A/COIL to the current source If constructing a multiturn coil, read Appendix B. Connect the ends of the multiturn coil as shown in Figure 1 or Figure 2. The number of complete turns in the coil does not have to be the same for all the test points. Several coils with different numbers of turns may be used. Size the coil current for each test point For each TI alternating current calibration point (test point) selected, perform a measurement as follows: Set the TI range switch to the appropriate current range, and clamp the TI jaws around one side of the coil (or around the wire with the wire in the center area enclosed by the jaws) If the test configuration in Figure 2 is used, adjust the oscillator frequency controls to the frequency required for the TI. Set the oscillator amplitude controls for a 0.85 V indication on the amplifier panel meter (if the 1570BN is used). If the meter calibrator and amplifier are used, set the meter calibrator to the frequencies and amperages listed. 7

11 Adjust the current source or amplifier controls to obtain a TI indication at the calibration point Verify that the result is within the tolerance limits for the selected calibration point Adjust the current source controls for minimum output If the TI has AC current ranges that were not tested, indicate on an affixed SPECIAL CALIBRATION label, the current range not tested. For example, Not calibrated on 200 A ac range. 4.4 DIRECT CURRENT TEST CAUTION DO NOT APPLY DIRECT CURRENT TO THE WESTON 327 TYPE 2 TRANSFORMER Set the TI selector switch for direct current measurements If applicable, set the direct current source controls for direct current output If applicable, perform the direct current source meter electrical zero adjustment Select the equipment for the test method to be used For each direct current calibration point selected, perform a measurement as follows: If a direct current correction factor is marked on the TI, determine the corrected value for the current source by algebraically adding the direct current correction factor to the TI calibration point If the calibration point is within the range of the current source, use the appropriate size of insulated wire to make a single loop of wire, and connect the ends of the wire to the current source output terminals; if not within the range of the current source, connect the current multiplying coil to the current source output terminals. Observe the NOTE following NOTES To reduce hysteresis in the TI, press the TI trigger to open the jaws fully, and then release the trigger to allow the jaws to snap shut. Repeat this several times before each measurement. If the TI is produced by the Columbia Electric Mfg, demagnetize the jaws using an alternating current (10 amp max) with the equipment connected as shown in Figure 1 or Figure Set the TI range switch to the appropriate current range, and clamp the TI jaws around one side of the coil by allowing the TI jaws to snap shut Adjust the current source controls to obtain a TI indication at the calibration point Measure the output of the current source and multiply the value by the number of turns of wire in the coil. NOTE Some TIs are not capable of measuring direct current with the jaw position reversed. For TIs with this limitation, skip to step Without decreasing current, reverse the TI on the coil and allow the TI jaws to snap shut and repeat steps and Determine the average value for the two indications, and verify that the result is within the tolerance limits for the nominal or corrected nominal value. 8

12 Example: If the TI calibration point is 100 A dc and the TI correction factor is -2, the current source is adjusted for a TI indication of 100 A and the current source corrected nominal indication 98 A. The tolerance limits should be applied to the corrected nominal indication and not to the TI indication Adjust the current source controls for minimum output If the TI has ranges that were not tested at fullscale, indicate on an affixed SPECIAL CALIBRATION label, the highest current values tested. For example, Not calibrated above 600 A ac and/or 500 A dc If the TI was submitted with a resistance probe, perform Section 4.5. If no probe is supplied, skip to step and state in the SPECIAL CALIBRATION statement, The resistance function was not tested (NA the checklist steps). 4.5 OHMMETER TEST Perform a mid-scale measurement for each of the TI resistance ranges as follows: Set the TI controls to the appropriate resistance range Set the resistance standard (item 2.5) to the nominal value of the resistance to be measured Connect the TI test leads together, and zero the meter Connect the TI test leads to the applicable resistance standard and verify that the TI indication obtained is within tolerance Unless other instruments are to be tested, turn all power switches off, and disconnect the equipment If the TI has interchangeable scale range plug ins, or a resistance probe, affix a SPECIAL CALIBRATION label on the TI, and list the plug in units, or resistance probe by scale range and serial numbers, which can be used with the TI. Example: THIS INSTRUMENT TO BE USED ONLY WITH: SCALE RANGE 0-100, SN SCALE RANGE 0-200, SN SCALE RANGE 0-400, SN SCALE RANGE , SN RESISTANCE PROBE, SN NOTE Only those plug in units calibrated in the TI can be listed for use by the TI Affix a SPECIAL CALIBRATION label to each of the scale range plug in units calibrated with the TI. Example: THIS INSTRUMENT TO BE USED ONLY WITH: (mark TIs serial #) Unless other measurements are to be performed, turn all power switches to off or standby and disconnect the equipment from the TI. (10 blank) 9

13 TEST INST (S) Amprobe Instrument Co RS-3 Clamp-On Volt Ammeter 3.6 Adj mech zero -- ck ( ) NA 3.7 Jaws clean -- ck ( ) NA 4.1 Alternating Voltage Test (V) (V) V range ±3% or to ±3.6% or to ±4.5% or 95.5 to ±6% or 70.5 to ±9% or 45.5 to ±18% or 20.5 to V rang 300 ±3% or 291 to V rang 600 ±3% or 582 to Alternating Current Test (A) (A) A range 6.00 ±3% or 5.82 to ±3.6% or 4.82 to ±4.5% or 3.82 to ±6% or 2.82 to ±9% or 1.82 to ±18% or 0.82 to A range ±3% or to A range 40.0 ±3% or 38.8 to A range 100 ±3% or 97 to A range 300 ±3% or 291 to Ohmmeter Test Zero meter -- ck ( ) NA ±1 Div on Voltage Scale 32 Page 1 of 1 (12 blank) 11

14 TEST INST (S) Amprobe Instrument ACDC600A Digital Clamp-On Ammeter 3.6 Adj mech zero -- ck ( ) NA 3.7 Jaws clean -- ck ( ) NA 4.3 Alternating Current Test Test frequency 60 Hz (A) (A) A range to to to A range to Test frequency 400 Hz (A) (A) A range to to Direct Current Test A range (A) (A) to to to to to Page 1 of 1 (14 blank) 13

15 TEST INST (S) Amprobe Instrument ACDC620T True RMS Clamp-On Multimeter 4.1 Alternating Voltage Test V ac 60 Hz (V ac) (V ac) to to to to to to V ac 400 Hz (V ac) (V ac) to to to to to to Direct Voltage Voltage range 0 to 1000 V dc (V dc) (V dc) to to to to to to 1013 Page 1 of 3 15

16 TEST INST (S) Amprobe Instrument ACDC620T True RMS Clamp-On Multimeter 4.3 Alternating Current Test Alternating current range 0 to 60 Hz (A ac) (A ac) to to to to Alternating current range 0 to 400 Hz (A ac) (A ac) to to to to Alternating current range 60 Hz (A ac) (A ac) to to to to 1030 Alternating current range 400 Hz (A ac) (A ac) to to to to Direct Current Test Direct current range 0 to 400 A dc (A dc) (A dc) to to to to Page 2 of 3 16

17 TEST INST (S) Amprobe Instrument ACDC620T True RMS Clamp-On Multimeter Range 400 to 1000 A dc (A dc) (A dc) to to to to Ohmmeter Test Zero meter check -- ck ( ) NA Resistance range 400 ( ) ( ) to to to Resistance range to to to 1013 Page 3 of 3 (18 blank) 17

18 TEST INST(S) General Electric AK-1 Clamp-On Volt Ammeter 3.6 Adj mech zero -- ck ( ) NA 3.7 Jaws clean -- ck ( ) NA 4.1 Alternating Voltage Test (V) V range ±3% or to ±3.6% or to ±4.5% or 95.5 to ±6% or 70.5 to ±9% or 45.5 to ±18% or 20.5 to V range 600 ±3% or 582 to Alternating Current Test (A) (A) A range ±3% or to ±3.8% or to ±5% or 8.55 to ±7.5% or 5.55 to ±15% or 2.55 to A range 60.0 ±3% or 58.2 to A range ±3% or to A 60 Hz 600 ±3% or 582 to A 400 Hz 300 ±3% or 282 to 318 Page 1 of 1 (20 blank) 19

19 TEST INST(S) General Electric AK-4 and AAAA1 Clamp-On Volt Ammeter 3.6 Adj mech zero -- ck ( ) NA 3.7 Jaws clean -- ck ( ) NA 4.1 Alternating Voltage Test (V) V range ±3% or to ±3.6% or to ±4.5% or 95.5 to ±6% or 70.5 to ±9% or 45.5 to ±18% or 20.5 to V range 300 ±3% or 291 to ±3% or 727 to Alternating Current Test (A) A range 10.0 ±3% or 9.7 to ±3.8% or 7.7 to ±5% or 5.7 to ±7.5% or 3.7 to ±15% or 1.7 to A range 30.0 ±3% or 29.1 to ±3% or 97 to ±3% or 291 to A 60 Hz 600 ±3% or 576 to A 400 Hz 300 ±3% or 276 to 324 Page 1 of 1 (22 blank) 21

20 TEST INST(S) Weston Instruments 633 Type VA-1 or TYPE A-1 Clamp-On Volt Ammeter 3.6 Adj mech zero -- ck ( ) NA 3.7 Jaws clean -- ck ( ) NA 4.1 Alternating Voltage Test (V) (V) V range ±3% or to ±3.6% or to ±4.5% or 95.5 to ±6% or 70.5 to ±9% or 45.5 to ±18% or 20.5 to V range 300 ±3% or 291 to ±3% or 582 to Alternating Current Test (A) (A) A range 10.0 ±3% or 9.7 to ±3.8% or 7.7 to ±5% or 5.7 to ±7.5% or 3.7 to ±15% or 1.7 to A range 25.0 ±3% or 24.2 to ±3% or 97 to ±3% or 242 to A 60 Hz 600 ±5% or 570 to A 400 Hz 300 ±10% or 270 to 330 Page 1 of 1 (24 blank) 23

21 TEST INST(S) Weston Instruments 749 Clamp-On Volt Ammeter 3.6 Adj mech zero -- ck ( ) NA 3.7 Jaws clean -- ck ( ) NA 4.1 Alternating Voltage Test 60 Hz tests (V) (V) V range ±3% or to ±3.6% or to ±4.5% or 95.5 to ±6% or 70.5 to ±9% or 45.5 to ±18% or 20.5 to V range 300 ±3% or 291 to ±3% or 582 to Alternating Current Test (A) (A) A range 6.00 ±3% or 5.82 to ±3.6% or 4.82 to ±4.5% or 3.82 to ±6% or 2.82 to ±9% or 1.82 to ±18% or 0.82 to A range ±3% or to ±3% or 29.1 to ±3% or 58.2 to ±3% or to ±3% or 291 to 309 Page 1 of 2 25

22 TEST INST(S) Weston Instruments 749 Clamp-On Volt Ammeter 4.1 Alternating Voltage Test 400 Hz tests (V) (V) V range 150 ±5% or to ±6% or to ±7.5% or 92.5 to ±10% or 67.5 to ±15% or 42.5 to ±30% or 17.5 to V range 300 ±5% or 285 to ±5% or 570 to Alternating Current Test (A) (A) A range 6.00 ±5% or 5.70 to ±6% or 4.70 to ±7.5% or 3.70 to ±10% or 2.70 to ±15% or 1.70 to ±30% or 0.70 to A range ±5% or to ±5% or 28.5 to ±5% or 57.0 to ±5% or to ±5% or 285 to 315 Page 2 of 2 26

23 TEST INST (S) Simpson Co 150, Amp-Clamp 3.6 Adj mech zero -- ck ( ) NA 3.7 Jaws clean -- ck ( ) NA 4.3 Alternating Current Test if 0 to 2.5 V ac range used (A) (A) A range 5.00 ±3% or 4.85 to ±3% or 9.7 to ±3% or to ±3% or 48.5 to ±3% or 97 to ±3% or to Alternating Current Test if 0 to 3 V ac range used (A) (A) A range 6.00 ±3% or 5.82 to ±3% or to ±3% or 29.1 to ±3% or 58.2 to ±3% or to ±3% or 291 to 309 NOTE This TI can be used with Simpson volt-ohmmeter having a 0 to 2.5 V or 0 to 3 V range. If used on a volt-ohmmeter with a 2.5 V scale, range goes to 250 A; if used on 3 V scale, range goes to 300 A. Page 1 of 1 (28 blank) 27

24 TEST INST(S) Columbia Electric AX, D, and C AC/DC Clamp-On Ammeters 3.7 Jaws clean -- ck ( ) NA NOTE: The 0-600/800/1000 ranges of Types C and D are calibrated AC or DC, not both. 4.3 Alternating Current Test Test frequency 60 Hz (A) (A) A range 600 ±8.3% or 550 to ±10% or 450 to A range 600 ±6.7% or 560 to ±8% or 460 to ±10% or 360 to A range 600 ±5% or 570 to ±6% or 470 to ±7.5% or 370 to ±10% or 270 to ±15% or 170 to A 400 Hz 300 ±12% or 264 to Direct Current Test A range (A) (A) 500 ±5% fs A range (A) (A) 500 ±5% fs A range (A) (A) 500 ±5% fs Page 1 of 7 29

25 TEST INST(S) Columbia Electric AX, D, and C AC/DC Clamp-On Ammeters 3.6 Adj mech zero -- ck ( ) NA 3.7 Jaws clean -- ck ( ) NA 4.3 Alternating Current Test Test frequency 60 Hz (A) (A) A range 500 ±2% or 490 to ±2.5% or 390 to ±3.33% or 290 to ±5.0% or 190 to 210 Test frequency 400 Hz (A) (A) A range 300 ±5% or 285 to ±7.5% or 185 to Direct Current Test A range (A) (A) 500 ±5% fs Alternating Current Test Test frequency 60 Hz (A) (A) A range 400 ±2% or 392 to ±2.67% or 292 to ±4% or 192 to 208 Test frequency 400 Hz (A) (A) A range 300 ±4% or 288 to ±6.0% or 188 to 212 Page 2 of 7 30

26 TEST INST(S) Columbia Electric AX, D, and C AC/DC Clamp-On Ammeters 4.4 Direct Current Test A range (A) (A) 400 ±5% fs Alternating Current Test Test frequency 60 Hz (A) (A) A range 300 ±2% or 294 to ±2.4% or 244 to ±3% or 194 to ±4% or 144 to ±6% or 94 to 106 Test frequency 400 Hz (A) (A) A range 300 ±3% or 291 to ±3.6% or 241 to ±4.5% or 191 to ±6.0% or 141 to ±9.0% or 91 to 109 Test frequency 60 Hz (A) (A) A range 150 ±2% or 147 to ±2.5% or 117 to ±3.3% or 87 to ±5.0% or 57 to 63 Test frequency 400 Hz (A) (A) A range ±3% or to ±3.8% or to ±5.0% or 85.5 to ±7.0% or 55.5 to 63.5 Page 3 of 7 31

27 TEST INST(S) Columbia Electric AX, D, and C AC/DC Clamp-On Ammeters 4.4 Direct Current Test A range (A) (A) 150 ±5% fs Alternating Current Test Test frequency 60 Hz (A) (A) A range 100 ±2% or 98 to ±2.5% or 78 to ±3.3% or 58 to ±5.0% or 38 to 42 Test frequency 400 Hz (A) (A) A range 100 ±3% or 97 to ±3.8% or 77 to ±5.0% or 57 to ±7.5% or 37 to Direct Current Test A range (A) (A) 300 ±5% fs Page 4 of 7 32

28 TEST INST(S) Columbia Electric AX, D, and C AC/DC Clamp-On Ammeters 3.7 Jaws clean -- ck ( ) NA 4.3 Alternating Current Test Test frequency 60 Hz (A) (A) A range 200 ±2% or 196 to ±2.67% or 146 to ±4.0% or 96 to 104 Test frequency 400 Hz (A) (A) A range 200 ±3% or 194 to ±4% or 144 to ±6.0% or 94 to Direct Current Test A range (A) (A) 200 ±5% fs A range (A) (A) 100 ±5% fs Page 5 of 7 33

29 TEST INST(S) Columbia Electric AX, D, and C AC/DC Clamp-On Ammeters 3.7 Jaws clean -- ck ( ) NA 4.3 Alternating Current Test Test frequency 60 Hz (A) (A) A range 75.0 ±2% or 73.5 to ±2.5% or 58.5 to ±3.75% or 38.5 to ±7.5% or 18.5 to 21.5 Test frequency 400 Hz (A) (A) A range 75.0 ±3% or 72.8 to ±3.8% or 57.8 to ±5.6% or 37.8 to ±11.3% or 17.8 to Direct Current Test A range (A) (A) 75 ±5% fs Page 6 of 7 34

30 TEST INST(S) Columbia Electric AX, D, and C AC/DC Clamp-On Ammeters 3.7 Jaws clean -- ck ( ) NA 4.3 Alternating Current Test Test frequency 60 Hz (A) (A) A range 50 ±2% or 49 to ±2.5 % or 39 to ±3.3 % or 29 to ±5.0 % or 19 to 21 Test frequency 400 Hz (A) (A) A range 50.0 ±3% or 48.5 to ±3.8% or 38.5 to ±5.0 % or 28.5 to ±7.5 % or 18.5 to Direct Current Test A range (A) (A) 50 ±5% fs Page 7 of 7 (36 blank) 35

31 TEST INST(S) Military MH or and Columbia Electric DP Clamp-On Volt Ammeters 3.7 Jaws clean -- ck ( ) NA 4.4 Direct Current Test (A) (A) ±5% of fullscale A ranges to to to to A range to to to to A range to to to A range to to to 300 Page 1 of 1 (38 blank) 37

32 TEST INST(S) Amprobe Instrument RS-300 Clamp On Volt Ammeter 3.7 Jaws clean -- ck ( ) NA 4.1 Alternating Voltage Test (V) (V) V range ±3% or to ±3.6% or to ±4.5% or 95.5 to ±6% or 70.5 to ±9% or 45.5 to ±18% or 20.5 to V range 300 ±3% or 291 to ±3% or 582 to Alternating Current Test (A) (A) A range 6.00 ±3% or 5.82 to ±3.6% or 4.82 to ±4.5% or 3.92 to ±6% or 2.82 to ±9% or 1.82 to ±18% or 0.82 to A range ±3% or to ±3% or 38.8 to ±3% or 97 to ±3% or 291 to 309 Page 1 of 1 (40 blank) 39

33 TEST INST(S) Weston Instruments 633 Type A-1 Clamp-On Ammeter 3.7 Jaws clean -- ck ( ) NA 4.3 Alternating Current Test (A) (A) A range 10.0 ±3% or 9.7 to ±3.8% or 7.7 to ±5% or 5.7 to ±7.5% or 3.7 to ±15% or 1.7 to A range ±3% or to ±3% or 48.5 to ±3% or 97 to ±3% or to A 60 Hz 500 ±3% or 485 to A 400 Hz 300 ±5% or 285 to 315 Page 1 of 1 (42 blank) 41

34 TEST INST(S) Pyramid Instrument 300 Clamp-On Volt Ammeter 3.7 Jaws clean -- ck ( ) NA 4.1 Alternating Voltage Test (V) (V) V range ±3% or to ±3.6% or to ±4.5% or 95.5 to ±6% or 70.5 to ±9% or 45.5 to ±18% or 20.5 to V range 300 ±3% or 291 to ±3% or 582 to Alternating Current Test (A) (A) A range 6.00 ±3% or 5.82 to ±3.6% or 4.82 to ±4.5% or 3.82 to ±6% or 2.82 to ±9% or 1.82 to ±18% or 0.82 to A range ±3% or to ±3.6% or to ±4.5% or 9.55 to ±6% or 7.05 to ±9% or 4.55 to ±18% or 2.05 to A range 30.0 ±3% or 29.1 to ±3% or 58.2 to ±3% or to ±3% or 291 to 309 Page 1 of 1 (44 blank) 43

35 TEST INST(S) Weston Instrument 633 Type A-2, 633 Type A Jaws clean -- ck ( ) NA 4.3 Alternating Current Test (A) (A) A range 100 ±3% or 97 to ±3.8% or 77 to ±5% or 57 to ±7.5% or 37 to ±15% or 17 to A range ±3% or to ±3% or 48.5 to ±3% or 97 to ±3% or to ±3% or 485 to 515 Page 1 of 1 (46 blank) 45

36 TEST INST(S) A.W. Sperry Instruments SPR-300 plus AC Volt Ohm Ammeter 3.7 Jaws clean -- ck ( ) NA 4.1 Alternating Voltage Test Test frequency: 60 Hz -- ck ( ) NA Ranges (V ac) (V ac) V ac to to to to to to V ac to to Alternating Current Test Test frequency: 60 Hz Ranges (A ac) (A ac) A ac to to to to to to A ac to to to to Ohmmeter Test Zero meter -- ck ( ) NA mid scale 25 ±3% of arc length or ±1/16 Page 1 of 1 (48 blank) 47

37 TEST INST(S) A.W. Sperry Instruments SPR-300 AC Volt Ohm Ammeter 3.7 Jaws clean -- ck ( ) NA 4.1 Alternating Voltage Test Test frequency 60 Hz -- ck ( ) NA Ranges (V ac) (V ac) V ac to to to to to to V ac to to Alternating Current Test Test frequency 60 Hz Ranges (A ac) (A ac) A ac to to to to to to A ac to to to to Ohmmeter Test Zero meter -- ck ( ) NA mid scale 25 ±3% of arc length or ±1/16 Page 1 of 1 (50 blank) 49

38 TEST INST(S) A.W. Sperry Instruments DSA-2007 Clamp-on Volt Ohm Ammeter 3.7 Jaws clean -- ck ( ) NA 4.1 Alternating Voltage Test (V) (V) V range 60 Hz to to to V range 60 Hz to V range 400 Hz to to to V range 400 Hz to Alternating Current Test (A) (A) A range 60 Hz to to to A range 60 Hz to A range 400 Hz to to to A range 400 Hz to Ohmmeter Test Zero meter -- ck ( ) NA ( ) ( ) range to to 193 Page 1 of 1 (52 blank) 51

39 TEST INST (S) Sperry Instruments DSA-2009 Volt-Ohm-Ammeter 3.7 Jaws clean -- ck ( ) NA 4.1 Alternating Voltage Test (V) (V) V range 60 Hz to to to V range 60 Hz to V range 400 Hz to to to V range 400 Hz to Direct Voltage Voltage ranges (V dc) (V dc) Range 20 V dc to to Range 200 V dc to to to to Voltage range 200 to 1000 V dc (V dc) (V dc) to to to to Alternating Current Test (A) (A) A range 60 Hz to to to A range 60 Hz to to to 1018 Page 1 of 2 53

40 TEST INST (S) Sperry Instruments DSA-2009 Volt-Ohm-Ammeter (A) (A) A range 60 Hz to A range 400 Hz to to to A range 400 Hz to to to to to Direct Current Test Direct current range (A dc) (A dc) 0 to 200 A dc to to to to to 2000 A dc to to to to to Ohmmeter Test Zero meter check -- ck ( ) NA Resistance range 200 ( ) ( ) to to Resistance range to to to 1525 Page 2 of 2 54

41 TEST INST(S) Caterpillar 8T900 AC/DC Clamp-On Ammeter 4.3 Alternating Current Test 60 Hz (A ac) (A ac) A range to to to to to to to to Hz (A ac) (A ac) 500 A range to to to to to Direct Current Test Range (A dc) (A dc) A range to to to to to to 52 NOTE Add a SPECIAL CALIBRATION label stating, TI calibrated to Hz, Hz and 500 A dc. Page 1 of 1 (56 blank) 55

42 TEST INST(S) Universal Enterprises, DL90A or DL100 Digital Clamp-on Tester 3.7 Jaws clean -- ck ( ) NA 4.1 Alternating Voltage Test (V ac) (V ac) V range 60 Hz to to to to V range 400 Hz to Hz to Volt range 60 Hz to to to to Direct Voltage Test (mv dc) (mv dc) Millvolt range to to to to (V dc) (V dc) to to to to V range to to to to Page 1 of 2 57

43 TEST INST(S) Universal Enterprises, DL90A or DL100 Digital Clamp-on Tester 4.3 Alternating Current Test (A) (A) A range 60 Hz to to to to A range 400 Hz to Hz to Hz to to to to A range to to to to Ohmmeter Test ( ) ( ) range to to to to (k ) (k ) k range to to to to Page 2 of 2 58

44 TEST INST (S) Fluke, 36 Clamp-Meter 4.1 Alternating Voltage Test V ac 60 Hz (V ac) (V ac) to to to to V ac 400 Hz (V ac) (V ac) to to to to V ac 60 Hz (V ac) (V ac) to to to to V ac 400 Hz (V ac) (V ac) to to to to Direct Voltage Voltage range 0 to 200 V dc (V dc) (V dc) to to to to Voltage range 200 to 600 V dc (V dc) (V dc) to to to to 609 Page 1 of 3 59

45 TEST INST (S) Fluke, 36 Clamp-Meter 4.3 Alternating Current Test Alternating current range 0 to 60 Hz (A ac) (A ac) to to to to Alternating current range 0 to 400 Hz (A ac) (A ac) to to to to Alternating current range 0 to 60 Hz (A ac) (A ac) to to to to 615 Alternating current range 0 to 400 Hz (A ac) (A ac) to to to to Direct Current Test Direct current range 0 to 200 A dc (A dc) (A dc) to to to to Page 2 of 3 60

46 TEST INST (S) Fluke, 36 Clamp-Meter Range 0 to 600 A dc (A dc) (A dc) to to to to 616 Range 600 to 1000 A dc (A dc) (A dc) to to to Ohmmeter Test Zero meter check -- ck ( ) NA Resistance range 200 ( ) ( ) to to to Page 3 of 3 (62 blank) 61

47 TEST INST (S) Fluke 333 Clamp-On Ammeter PROCEDURE STEP FUNCTION TESTED NOMINAL MEASURED VALUES OUT FIRST RUN SECOND RUN OF CALIBRATION ERANCES 4.1 Alternating Voltage Test V ac 60 Hz (V ac) (V ac) to to to to to to to to Direct Voltage Voltage range 0 to 600 V dc (V dc) (V dc) to to to to to to to to Alternating Current Test Alternating current range 0 to 60 Hz TI Indication (A ac) (A ac) 25 A to A to A to A to Page 1 of 2 63

48 TEST INST (S) Fluke 333 Clamp-On Ammeter PROCEDURE STEP FUNCTION TESTED NOMINAL MEASURED VALUES OUT FIRST RUN SECOND RUN OF CALIBRATION ERANCES 4.3 Alternating Current Test Alternating current range 0 to 60 Hz TI Indication (A ac) (A ac) 300 A to A to Ohmmeter Test Zero meter check -- ck ( ) NA Resistance range 0 to 600 ( ) ( ) to to Page 2 of 2 64

49 TEST INST (S) Fluke 337 Clamp-On Ammeter 4.1 Alternating Voltage Test V ac 60 Hz (V ac) (V ac) to to to to to to to to V ac 400 Hz (V ac) (V ac) to to to to to to to to Direct Voltage Voltage range 0 to 600 V dc (V dc) (V dc) to to to to to to to to Page 1 of 3 65

50 TEST INST (S) Fluke 337 Clamp-On Ammeter 4.3 Alternating Current Test Alternating current range 0 to 60 Hz TI Indication (A ac) (A ac) 25 A to A to A to A to A to A to A to A to Alternating current range 0 to 400 Hz TI Indication (A ac) (A ac) 25 A to A to A to A to A to Direct Current Test Direct current range TI Indication (A dc) (A dc) 50 A to A to A to A to A to A to Page 2 of 3 66

51 TEST INST (S) Fluke 337 Clamp-On Ammeter 4.5 Ohmmeter Test Zero meter check -- ck ( ) NA Resistance range 0 to 600 ( ) ( ) to to Resistance range ( ) ( ) to to to to 6095 Page 3 of 3 (68 blank) 67

52 TEST INST (S) AEMC AC Current Probe Clamp On Ammeter 4.3 Alternating Current Test* to 1000 A 60 Hz (ma ac) (ma ac) 50 A 60 Hz to A 60 Hz to A 60 Hz to A 60 Hz to A 400 Hz to A 50 Hz to A 1000 Hz** to 51.0 * TI current output is to be used with item 2.8 (set to measure ac current). Current output from the TI is measured by item 2.8. ** The item 2.3 standards, the 327 Type 2 transformer, and the ammeter are not to be used for the 1 khz test, use either the shunts and DMM or the 5700AAN and 5725A instead. Page 1 of 1 (70 blank) 69

53 TEST INST (S) F. W. Bell CG-109A Clamp On Ammeter 4.3 Alternating Current Test * Alternating current range: 0 to 100 A 60 Hz (A ac) (V ac) (TI outputs V ac, fullscale is to V ac, use with DMM set to V ac) to to to 1000 A 60 Hz to (TI outputs V ac, fullscale is to V ac, use with DMM set to V ac) to to 100 A 400 Hz to (TI outputs V ac, fullscale is to V ac, use with DMM set to V ac) to to 1000 A 400 Hz to to 100 A 50 Hz to *Zero the TI before each measurement. Page 1 of 1 (72 blank) 71

DIAL AND VERNIER CALIPERS

DIAL AND VERNIER CALIPERS TECHNICAL MANUAL INSTRUMENT CALIBRATION PROCEDURE DIAL AND VERNIER CALIPERS THIS PUBLICATION SUPERSEDES NAVAIR 17 20MD 07 DATED 1 JUNE 1995 DISTRIBUTION STATEMENT C. DISTRIBUTION AUTHORIZED TO U.S. GOVERNMENT

More information

TECHNICAL MANUAL CALIBRATION PROCEDURE FOR MULTIMETERS

TECHNICAL MANUAL CALIBRATION PROCEDURE FOR MULTIMETERS TECHNICAL MANUAL CALIBRATION PROCEDURE FOR MULTIMETERS Distribution Statement C - Distribution authorized to U. S. Government agencies and their contractors for official use or for administrative or operational

More information

TABLE OF CONTENTS SECTION I INTRODUCTION AND DESCRIPTION

TABLE OF CONTENTS SECTION I INTRODUCTION AND DESCRIPTION Walter G. LEGGE Co., Inc. 444 Central Ave., Peekskill, NY 10566 tel # 800-345-3443 fax 800-332-2636 email address: info@leggesystems.com website: www.leggesystems.com TABLE OF CONTENTS SECTION PAGE 1 INTRODUCTION

More information

TECHNICAL MANUAL INSTRUMENT CALIBRATION PROCEDURE OPTICAL COMPARATORS THIS PUBLICATION SUPERSEDES NAVAIR 17 20MD 20 DATED 1 FEBRUARY 1988

TECHNICAL MANUAL INSTRUMENT CALIBRATION PROCEDURE OPTICAL COMPARATORS THIS PUBLICATION SUPERSEDES NAVAIR 17 20MD 20 DATED 1 FEBRUARY 1988 TECHNICAL MANUAL INSTRUMENT CALIBRATION PROCEDURE OPTICAL COMPARATORS THIS PUBLICATION SUPERSEDES NAVAIR 17 20MD 20 DATED 1 FEBRUARY 1988 DISTRIBUTION STATEMENT C. DISTRIBUTION AUTHORIZED TO U.S. GOVERNMENT

More information

TECHNICAL MANUAL CALIBRATION PROCEDURE FOR DIGITAL MULTIMETER 34401A (HEWLETT-PACKARD)

TECHNICAL MANUAL CALIBRATION PROCEDURE FOR DIGITAL MULTIMETER 34401A (HEWLETT-PACKARD) TECHNICAL MANUAL CALIBRATION PROCEDURE FOR DIGITAL MULTIMETER 34401A (HEWLETT-PACKARD) This publication replaces T.O. 33K8-4-1029-1 dated 30 May 2000. Distribution Statement - Distribution authorized to

More information

TECHNICAL MANUAL CALIBRATION PROCEDURE FOR HIGH VOLTAGE INSTRUMENTS

TECHNICAL MANUAL CALIBRATION PROCEDURE FOR HIGH VOLTAGE INSTRUMENTS TECHNICAL MANUAL CALIBRATION PROCEDURE FOR HIGH VOLTAGE INSTRUMENTS Distribution Statement - Distribution authorized to U. S. Government agencies and their contractors for official use or for administrative

More information

TECHNICAL MANUAL INSTRUMENT CALIBRATION PROCEDURE PEAK POWER ANALYZER HEWLETT-PACKARD 8991AOPT003

TECHNICAL MANUAL INSTRUMENT CALIBRATION PROCEDURE PEAK POWER ANALYZER HEWLETT-PACKARD 8991AOPT003 TECHNICAL MANUAL INSTRUMENT CALIBRAON PROCEDURE PEAK POWER ANALYZER HEWLETT-PACKARD 8991AOPT003 THIS PUBLICAON SUPERSEDES NAVAIR 17-20GP-48 DATED 1 MARCH 1998 DISTRIBUON STATEMENT C. DISTRIBUON AUTHORIZED

More information

CHANGE 1 DEPARTMENT OF THE ARMY TECHNICAL BULLETIN

CHANGE 1 DEPARTMENT OF THE ARMY TECHNICAL BULLETIN TB 9-6625-011-24 CHANGE 1 DEPARTMENT OF THE ARMY TECHNICAL BULLETIN CALIBRATION PROCEDURE FOR TRUE RMS VOLTMETER HEWLETT- PACKARD, MODEL 3400A Headquarters, Department of the Army, Washington, DC 18 September

More information

DEPARTMENT OF THE ARMY TECHNICAL BULLETIN

DEPARTMENT OF THE ARMY TECHNICAL BULLETIN *TB 9-6625-1356-24 DEPARTMENT OF THE ARMY TECHNICAL BULLETIN CALIBRATION PROCEDURE FOR TEST OSCILLATOR, HEWLETT-PACKARD MODELS 651A, 651B AND 652A (SG-763/U) Headquarters Department of the Army, Washington,

More information

DEPARTMENT OF THE ARMY TECHNICAL BULLETIN

DEPARTMENT OF THE ARMY TECHNICAL BULLETIN *TB 9-6625-2215-24 DEPARTMENT OF THE ARMY TECHNICAL BULLETIN CALIBRATION PROCEDURE FOR MEASURING SYSTEM, TD-1209/U (HEWLETT-PACKARD, MODEL 5300A); 50 MHz UNIVERSAL COUNTER, TD-1211/U (HEWLETT-PACKARD,

More information

TECHNICAL MANUAL CALIBRATION PROCEDURE FOR LCR METER 875A, 875B (B & K PRECISION) This publication replaces T.O. 33K dated 30 August 2011.

TECHNICAL MANUAL CALIBRATION PROCEDURE FOR LCR METER 875A, 875B (B & K PRECISION) This publication replaces T.O. 33K dated 30 August 2011. TECHNICAL MANUAL CALIBRATION PROCEDURE FOR LCR METER 875A, 875B (B & K PRECISION) This publication replaces T.O. 33K1-4-1260-1 dated 30 August 2011. Distribution Statement C - Distribution authorized to

More information

DEPARTMENT OF THE ARMY TECHNICAL BULLETIN

DEPARTMENT OF THE ARMY TECHNICAL BULLETIN *TB 9-6625-1213-24 DEPARTMENT OF THE ARMY TECHNICAL BULLETIN CALIBRATION PROCEDURE FOR AC AMPLIFIER AM-1881/U (HEWLETT- PACKARD MODEL 450A) AND HEWLETT- PACKARD MODEL 465A Headquarters, Department of the

More information

DEPARTMENT OF THE ARMY TECHNICAL BULLETIN

DEPARTMENT OF THE ARMY TECHNICAL BULLETIN *TB 9-6625-2107-24 DEPARTMENT OF THE ARMY TECHNICAL BULLETIN CALIBRATION PROCEDURE FOR AC VOLTMETER, ME-30F/U AND HEWLETT-PACKARD, MODELS 400F AND 400FL Headquarters, Department of the Army, Washington,

More information

DEPARTMENT OF THE ARMY TECHNICAL BULLETIN CALIBRATION PROCEDURE FOR DIGITAL MULTIMETER HEWLETT-PACKARD MODELS 3435A AND 3438A

DEPARTMENT OF THE ARMY TECHNICAL BULLETIN CALIBRATION PROCEDURE FOR DIGITAL MULTIMETER HEWLETT-PACKARD MODELS 3435A AND 3438A *TB 9-6625-2227-35 DEPARTMENT OF THE ARMY TECHNICAL BULLETIN CALIBRATION PROCEDURE FOR DIGITAL MULTIMETER HEWLETT-PACKARD MODELS 3435A AND 3438A Headquarters, Department of the Army, Washington, DC 18

More information

APPENDIX D DISCUSSION OF ELECTRONIC INSTRUMENTS

APPENDIX D DISCUSSION OF ELECTRONIC INSTRUMENTS APPENDIX D DISCUSSION OF ELECTRONIC INSTRUMENTS DC POWER SUPPLIES We will discuss these instruments one at a time, starting with the DC power supply. The simplest DC power supplies are batteries which

More information

Sept 13 Pre-lab due Sept 12; Lab memo due Sept 19 at the START of lab time, 1:10pm

Sept 13 Pre-lab due Sept 12; Lab memo due Sept 19 at the START of lab time, 1:10pm Sept 13 Pre-lab due Sept 12; Lab memo due Sept 19 at the START of lab time, 1:10pm EGR 220: Engineering Circuit Theory Lab 1: Introduction to Laboratory Equipment Pre-lab Read through the entire lab handout

More information

*This bulletin supersedes TB , dated 18 July 1988, including all changes.

*This bulletin supersedes TB , dated 18 July 1988, including all changes. * SUPERSEDED COPY DATED 18 JULY 1988 DEPARTMENT OF THE ARMY TECHNICAL BULLETIN CALIBRATION PROCEDURE FOR DIGITAL MULTIMETER, TEKTRONIX TYPES DM 501 AND DM 501A WITH DC HIGH VOLTAGE PROBE, BALLANTINE, MODEL

More information

SOLICITUD XXVI CONGRESO NACIONAL DE METROLOGIA

SOLICITUD XXVI CONGRESO NACIONAL DE METROLOGIA SOLICITUD XXVI CONGRESO NACIONAL DE METROLOGIA Técnicas para la calibración de medidores analógicos Bill Gaviria Regional Product Manager Direct: +1.425.446.5335 Cell: +1.321.626.7845 Email: bill.gaviria@flukecal.com

More information

34134A AC/DC DMM Current Probe. User s Guide. Publication number April 2009

34134A AC/DC DMM Current Probe. User s Guide. Publication number April 2009 User s Guide Publication number 34134-90001 April 2009 For Safety information, Warranties, Regulatory information, and publishing information, see the pages at the back of this book. Copyright Agilent

More information

DEPARTMENT OF THE ARMY TECHNICAL BULLETIN

DEPARTMENT OF THE ARMY TECHNICAL BULLETIN *TB 9-6625-2213-24 DEPARTMENT OF THE ARMY TECHNICAL BULLETIN CALIBRATION PROCEDURE FOR VECTOR VOLTMETER, ME-512/U AND HEWLETT-PACKARD, MODELS 8405A AND 8405A-H16 Headquarters, Department of the Army, Washington,

More information

80i-110s AC/DC CURRENT PROBE

80i-110s AC/DC CURRENT PROBE Table of Contents 3 80i-110s AC/DC CURRENT PROBE Users Manual November 1996 1994, 1996 Fluke Corporation. All rights reserved. All product names are trademarks of their respective companies. Printed in

More information

DEPARTMENT OF THE ARMY TECHNICAL BULLETIN

DEPARTMENT OF THE ARMY TECHNICAL BULLETIN *TB 9-6625-1947-24 DEPARTMENT OF THE ARMY TECHNICAL BULLETIN CALIBRATION PROCEDURE FOR TEST OSCILLATOR HEWLETT-PACKARD, MODEL 654A Headquarters, Department of the Army, Washington, DC 11 March 2008 Distribution

More information

Change 4 DEPARTMENT OF THE ARMY TECHNICAL BULLETIN CALIBRATION PROCEDURE FOR VOLTAGE STANDARAD, JOHN FLUKE, MODELS 332A, 332B, AND 332B/AF

Change 4 DEPARTMENT OF THE ARMY TECHNICAL BULLETIN CALIBRATION PROCEDURE FOR VOLTAGE STANDARAD, JOHN FLUKE, MODELS 332A, 332B, AND 332B/AF Change 4 DEPARTMENT OF THE ARMY TECHNICAL BULLETIN CALIBRATION PROCEDURE FOR VOLTAGE STANDARAD, JOHN FLUKE, MODELS 332A, 332B, AND 332B/AF Headquarters, Department of the Army, Washington, DC 9 March 1988,

More information

DEPARTMENT OF THE ARMY TECHNICAL BULLETIN

DEPARTMENT OF THE ARMY TECHNICAL BULLETIN *TB 9-6625-2072-24 DEPARTMENT OF THE ARMY TECHNICAL BULLETIN CALIBRATION PROCEDURE FOR OSCILLOSCOPE, OS-189A(P), AND HEWLETT- PACKARD, MODELS 180A, 180C AND 180D; DUAL CHANNEL VERTICAL AMPLIFIER, PL-1186/USM-281

More information

DEPARTMENT OF THE ARMY TECHNICAL BULLETIN

DEPARTMENT OF THE ARMY TECHNICAL BULLETIN DEPARTMENT OF THE ARMY TECHNICAL BULLETIN CALIBRATION PROCEDURE FOR FREQUENCY COUNTER HEWLETT-PACKARD, MODEL 5340A AND MICROWAVE FREQUENCY COUNTER TD-1225A(V)1/U (HEWLETT-PACKARD, MODEL 5342A/H14), TD-1225A(V)2/U

More information

Electronic Instrument Disadvantage of moving coil meter Low input impedance High loading error for low-voltage range voltmeter

Electronic Instrument Disadvantage of moving coil meter Low input impedance High loading error for low-voltage range voltmeter EIE 240 Electrical and Electronic Measurement Class 6, February 20, 2015 1 Electronic Instrument Disadvantage of moving coil meter Low input impedance High loading error for low-voltage range voltmeter

More information

EE Chapter 7 Measuring Instruments

EE Chapter 7 Measuring Instruments EE 2145230 Chapter 7 Measuring Instruments 7.1 Meter Movements The basic principle of many electric instruments is that of the galvanometer. This is a device which reacts to minute electromagnetic influences

More information

MEASUREMENTS & INSTRUMENTATION ANALOG AND DIGITAL METERS

MEASUREMENTS & INSTRUMENTATION ANALOG AND DIGITAL METERS MEASUREMENTS & INSTRUMENTATION ANALOG AND DIGITAL METERS ANALOG Metering devices Provides monotonous (continuous) movement. ELECTRICAL MEASURING INSTRUMENTS ANALOG METERS A d Arsonval galvanometer (Moving

More information

DEPARTMENT OF THE ARMY TECHNICAL BULLETIN

DEPARTMENT OF THE ARMY TECHNICAL BULLETIN *TB 9-6625-2057-24 DEPARTMENT OF THE ARMY TECHNICAL BULLETIN CALIBRATION PROCEDURE FOR SIGNAL GENERATORS SG-543/U, SG-543B/U, AND SG-543C/U; AND HEWLETT-PACKARD, MODELS 204B AND 204C Headquarters, Department

More information

DEPARTMENT OF DEFENSE TEST METHOD STANDARD METHOD 213, SHOCK (SPECIFIED PULSE)

DEPARTMENT OF DEFENSE TEST METHOD STANDARD METHOD 213, SHOCK (SPECIFIED PULSE) INCH-POUND MIL-STD-202-213 18 April 2015 SUPERSEDING MIL-STD-202G w/change 2 (IN PART) 28 June 2013 (see 6.1) DEPARTMENT OF DEFENSE TEST METHOD STANDARD METHOD 213, SHOCK (SPECIFIED PULSE) AMSC N/A FSC

More information

DEPARTMENT OF THE ARMY TECHNICAL BULLETIN

DEPARTMENT OF THE ARMY TECHNICAL BULLETIN *TB 9-6625-2353-24 DEPARTMENT OF THE ARMY TECHNICAL BULLETIN CALIBRATION PROCEDURE FOR DIGITAL MULTIMETER, AN/PSM-45 AND (SIMPSON, MODEL 467) AND SIMPSON, MODEL 467E Headquarters, Department of the Army,

More information

AC/DC DIGITAL CLAMP METER OPERATION MANUAL

AC/DC DIGITAL CLAMP METER OPERATION MANUAL AC/DC DIGITAL CLAMP METER OPERATION MANUAL HYS005661 A0 ACCESSORIES 6. ACCESSORIES 1) Test Leads: Electric Ratings 1000V 10A 1 pair (set) 2) Operating Manual 1 copy 3) 1.5V AAA Battery 3 piece - - 55 -

More information

TECHNICAL MANUAL CALIBRATION PROCEDURE FOR THREADED PLUG GAGES. This publication replaces T.O. 33K dated 30 January 2013.

TECHNICAL MANUAL CALIBRATION PROCEDURE FOR THREADED PLUG GAGES. This publication replaces T.O. 33K dated 30 January 2013. TECHNICAL MANUAL CALIBRATION PROCEDURE FOR THREADED PLUG GAGES This publication replaces T.O. 33K6-4-203-1 dated 30 January 2013. Distribution Statement C - Distribution authorized to U. S. Government

More information

6100A/6101A - Alternative verification methods

6100A/6101A - Alternative verification methods 6100A/6101A - Alternative verification methods Alternative verification of 6100A/6101A 6100A/6101A - Alternative verification methods Title Page Alternative verification of 6100A/6101A... 2 Recommendation...

More information

ECE 53A: Fundamentals of Electrical Engineering I

ECE 53A: Fundamentals of Electrical Engineering I ECE 53A: Fundamentals of Electrical Engineering I Laboratory Assignment #1: Instrument Operation, Basic Resistor Measurements and Kirchhoff s Laws Fall 2007 General Guidelines: - Record data and observations

More information

Lab 1: Basic Lab Equipment and Measurements

Lab 1: Basic Lab Equipment and Measurements Abstract: Lab 1: Basic Lab Equipment and Measurements This lab exercise introduces the basic measurement instruments that will be used throughout the course. These instruments include multimeters, oscilloscopes,

More information

SUPERSEDED COPY DATED 15 OCTOBER 1990 DEPARTMENT OF THE ARMY TECHNICAL BULLETIN CALIBRATION PROCEDURE FOR DISTORTION ANALYZER, TS-4084/G

SUPERSEDED COPY DATED 15 OCTOBER 1990 DEPARTMENT OF THE ARMY TECHNICAL BULLETIN CALIBRATION PROCEDURE FOR DISTORTION ANALYZER, TS-4084/G *TB 9-6625-2170-35 SUPERSEDED COPY DATED 15 OCTOBER 1990 DEPARTMENT OF THE ARMY TECHNICAL BULLETIN CALIBRATION PROCEDURE FOR DISTORTION ANALYZER, TS-4084/G Headquarters, Department of the Army, Washington,

More information

201AP Charge Amplifier User Manual

201AP Charge Amplifier User Manual Trig-Tek 201AP Charge Amplifier User Manual Publication No. 980996 Rev. A Astronics Test Systems Inc. 4 Goodyear, Irvine, CA 92618 Tel: (800) 722-2528, (949) 859-8999; Fax: (949) 859-7139 atsinfo@astronics.com

More information

ABCs of DMMs Multimeter features and functions explained Application Note

ABCs of DMMs Multimeter features and functions explained Application Note ABCs of DMMs Multimeter features and functions explained Application Note Digital multimeters offer a wide selection of features. Choosing the right meter for the job can be challenging unless you know

More information

AP034-OM-E Rev D ISSUED: January 2000 ²

AP034-OM-E Rev D ISSUED: January 2000 ² 3HUIRUPDQFH9HULILFDWLRQ 3HUIRUPDQFH9HULILFDWLRQ This procedure can be used to verify the warranted characteristics of the AP034 Active Differential Probe. The recommended calibration interval for the model

More information

ABCs of DMMs. Multimeter features and functions explained. Application Note. Introduction. Choosing your DMM. Some basics

ABCs of DMMs. Multimeter features and functions explained. Application Note. Introduction. Choosing your DMM. Some basics ABCs of DMMs Multimeter features and functions explained Application Note Introduction Multimeters. They ve been described as the tape measure of the new millennium. But what exactly is a digital multimeter

More information

Manual Supplement. This supplement contains information necessary to ensure the accuracy of the above manual.

Manual Supplement. This supplement contains information necessary to ensure the accuracy of the above manual. Manual Title: 550A Getting Started Supplement Issue: Part Number: 415509 Issue Date: 9/18 Print Date: November 01 Page Count: 19 Revision/Date: This supplement contains information necessary to ensure

More information

CHANGE 1 DEPARTMENT OF THE ARMY TECHNICAL BULLETIN

CHANGE 1 DEPARTMENT OF THE ARMY TECHNICAL BULLETIN CHANGE 1 DEPARTMENT OF THE ARMY TECHNICAL BULLETIN CALIBRATION PROCEDURE FOR DIGITAL MULTIMETER, AN/PSM-45 AND (SIMPSON, MODEL 467) AND SIMPSON, MODEL 467E Headquarters, Department of the Army, Washington,

More information

STATEMENTS MADE AND DATA SHOWN, ARE TO THE BEST OF OUR KNOWLEDGE AND BELIEF, CORRECT AND WITHIN THE USUAL LIMITS OF COMMERCIAL TESTING PRACTICES.

STATEMENTS MADE AND DATA SHOWN, ARE TO THE BEST OF OUR KNOWLEDGE AND BELIEF, CORRECT AND WITHIN THE USUAL LIMITS OF COMMERCIAL TESTING PRACTICES. STATEMENTS MADE AND DATA SHOWN, ARE TO THE BEST OF OUR KNOWLEDGE AND BELIEF, CORRECT AND WITHIN THE USUAL LIMITS OF COMMERCIAL TESTING PRACTICES. ADMINISTRATIVE DATA 1.0 PURPOSE OF TEST For compliance

More information

5080A Multi-Product Calibrator

5080A Multi-Product Calibrator 5080A Multi-Product Calibrator Calibration solutions for your analog and digital workload 5080A Multi-Product Calibrator: expanded workload coverage Calibrate analog and digital meters, and much more The

More information

BENNING. CM7 Electrical Tester. Service Information

BENNING. CM7 Electrical Tester. Service Information BENNING CM7 Electrical Tester Service Information TABLE OF CONTENTS Page Introduction 1 Precautions and Safety Information 1 Symbols 1 Safety 2 Specifications 3 General Specification 3 Voltage Specification

More information

Using Circuits, Signals and Instruments

Using Circuits, Signals and Instruments Using Circuits, Signals and Instruments To be ignorant of one s ignorance is the malady of the ignorant. A. B. Alcott (1799-1888) Some knowledge of electrical and electronic technology is essential for

More information

*TB DEPARTMENT OF THE ARMY TECHNICAL BULLETIN

*TB DEPARTMENT OF THE ARMY TECHNICAL BULLETIN * DEPARTMENT OF THE ARMY TECHNICAL BULLETIN CALIBRATION PROCEDURE FOR SYNTHESIZER/LEVEL GENERATOR, HEWLETT-PACKARD, MODEL 3335A AND 3335A OPT 001 Headquarters Department of the Army, Washington, DC 18

More information

Introduction to Electronic Equipment

Introduction to Electronic Equipment Introduction to Electronic Equipment INTRODUCTION This semester you will be exploring electricity and magnetism. In order to make your time in here more instructive we ve designed this laboratory exercise

More information

5700A/5720A Series II Multi-Function Calibrator

5700A/5720A Series II Multi-Function Calibrator 5700A/5720A Series II Multi-Function Calibrator Operator Guide PN 601648 May 1996 1996 Fluke Corporation, Inc. All rights reserved. Printed in U.S.A. Contents What is in this Guide?... 2 Safety Summary...

More information

DIGITAL DUAL DISPLAY AC/DC CLAMP METER MODEL-860A OPERATION MANUAL

DIGITAL DUAL DISPLAY AC/DC CLAMP METER MODEL-860A OPERATION MANUAL DIGITAL DUAL DISPLAY AC/DC CLAMP METER MODEL-860A OPERATION MANUAL DIGITAL DUAL DISPLAY AC/DC CLAMP METER MODEL-860A TABLE OF CONTENTS TITLE PAGE Safety Information Safety Symbols... 1 Meter Description...

More information

DEPARTMENT OF THE ARMY TECHNICAL BULLETIN CALIBRATION PROCEDURE FOR SHF SIGNAL GENERATOR AN/USM-47 (HEWLETT-PACKARD MODEL 626A) (NSN )

DEPARTMENT OF THE ARMY TECHNICAL BULLETIN CALIBRATION PROCEDURE FOR SHF SIGNAL GENERATOR AN/USM-47 (HEWLETT-PACKARD MODEL 626A) (NSN ) DEPARTMENT OF THE ARMY TECHNICAL BULLETIN CALIBRATION PROCEDURE FOR SHF SIGNAL GENERATOR AN/USM-47 (HEWLETT-PACKARD MODEL 626A) (NSN 6625-00-455-6917) Headquarters, Department of the Army, Washington,

More information

IDEAL INDUSTRIES, INC. TECHNICAL MANUAL MODEL:

IDEAL INDUSTRIES, INC. TECHNICAL MANUAL MODEL: IDEAL INDUSTRIES, INC. TECHNICAL MANUAL MODEL: 61-352 The Service Information provides the following information: Precautions and safety information Specifications Basic maintenance (cleaning, replacing

More information

Meters and Test Equipment

Meters and Test Equipment Installation Knowledge and Techniques Meters and Test Equipment OBJECTIVES Meters and Test Equipment DMM s and VOM s Describe the difference between a DMM and a VOM. Describe the methods for measuring

More information

Agilent 1146A AC/DC Oscilloscope Current Probe

Agilent 1146A AC/DC Oscilloscope Current Probe A User s Guide Publication number 01146-92003 May 2005 Copyright Agilent Technologies 1996-2005 All Rights Reserved Contents Receiving Your Shipment -2 Packaging -2 Description -3 Compatibility -3 Control

More information

AC/DC Current Oscilloscope Probe Model SL261

AC/DC Current Oscilloscope Probe Model SL261 AC/DC Current Oscilloscope Probe Model SL261 USER MANUAL I ZERO 100 mv/a 10 mv/a OFF Statement of Compliance Chauvin Arnoux, Inc. d.b.a. AEMC Instruments certifies that this instrument has been calibrated

More information

M 140i. Multifunction Calibrator. Operation Manual

M 140i. Multifunction Calibrator. Operation Manual M 140i Multifunction Calibrator Operation Manual M-140i Multifunction Calibrator MEATEST, s.r.o. 2 Operation Manual v35 MEATEST, s.r.o. M-140i Multifunction Calibrator Content Operation Manual... 1 Basic

More information

Physics 120 Lab 1 (2018) - Instruments and DC Circuits

Physics 120 Lab 1 (2018) - Instruments and DC Circuits Physics 120 Lab 1 (2018) - Instruments and DC Circuits Welcome to the first laboratory exercise in Physics 120. Your state-of-the art equipment includes: Digital oscilloscope w/usb output for SCREENSHOTS.

More information

Guide for Conducting a Stray Voltage Evaluation Under the MPSC Stray Voltage Rule 1

Guide for Conducting a Stray Voltage Evaluation Under the MPSC Stray Voltage Rule 1 Guide for Conducting a Stray Voltage Evaluation Under the MPSC Stray Voltage Rule 1 Introduction and Overall Purpose: This guide is intended to be used in conjunction with the administrative rules addressing

More information

DEPARTMENT OF THE ARMY TECHNICAL BULLETIN

DEPARTMENT OF THE ARMY TECHNICAL BULLETIN *TB 9-6625-1914-24 DEPARTMENT OF THE ARMY TECHNICAL BULLETIN CALIBRATION PROCEDURE FOR SPECTRUM ANALYZER, IF, LF, AND RF PLUG-IN UNITS, HEWLETT-PACKARD, MODELS 8552( ), 8553( ), 8554( ), 8555( ), AND 8556(

More information

Introduction to the Laboratory

Introduction to the Laboratory Memorial University of Newfoundland Department of Physics and Physical Oceanography Physics 2055 Laboratory Introduction to the Laboratory The purpose of this lab is to introduce you to some of the equipment

More information

Agilent N2780A, N2781A, N2782A, and N2783A Current Probes

Agilent N2780A, N2781A, N2782A, and N2783A Current Probes Agilent N2780A, N2781A, N2782A, and N2783A Current Probes User s and Service Guide Agilent Technologies Notices Agilent Technologies, Inc. 2007 No part of this manual may be reproduced in any form or by

More information

5502A Multi-Product Calibrator. Extended specifications

5502A Multi-Product Calibrator. Extended specifications 550A Multi-Product Calibrator Extended specifications General Specifications The following tables list the 550A specifications. All specifications are valid after allowing a warm-up period of 30 minutes,

More information

PERFORMANCE SPECIFICATION SHEET ELECTRON TUBE, POWER TYPE 8660

PERFORMANCE SPECIFICATION SHEET ELECTRON TUBE, POWER TYPE 8660 INCH-POUND MIL-PRF-1/1648D 23 September 2014 SUPERSEDING MIL-PRF-1/1648C w/amendment 1 10 April 2008 DESCRIPTION: Tetrode, ceramic-metal. See figure 2. Mounting position: Any. PERFORMANCE SPECIFICATION

More information

TR CRITERIA FOR LABORATORY ACCREDITATION IN THE FIELD OF ELECTRICAL DCLF METROLOGY. Approved By: Senior Manager: Mpho Phaloane Revised By:

TR CRITERIA FOR LABORATORY ACCREDITATION IN THE FIELD OF ELECTRICAL DCLF METROLOGY. Approved By: Senior Manager: Mpho Phaloane Revised By: CRITERIA FOR LABORATORY ACCREDITATION IN THE FIELD OF ELECTRICAL LF METROLOGY Approved By: Senior Manager: Mpho Phaloane Revised By: Specialist Technical Committee Date of Approval: 2015-08-26 Date of

More information

DEPARTMENT OF THE ARMY TECHNICAL BULLETIN

DEPARTMENT OF THE ARMY TECHNICAL BULLETIN *TB 9-6625-2050-24 DEPARTMENT OF THE ARMY TECHNICAL BULLETIN CALIBRATION PROCEDURE FOR POWER METERS, TS-3793 (HEWLETT-PACKARD 436A AND 436AOPT9/22) HEWLETT-PACKARD MODELS 435A, 435B, 435BOPT001, AND 435BE21

More information

5520A. Service Manual. PQ Option

5520A. Service Manual. PQ Option 5520A PQ Option Service Manual PN 1629170 October 2001 2001 Fluke Corporation, All rights reserved. Printed in USA All product names are trademarks of their respective companies. LIMITED WARRANTY AND LIMITATION

More information

LAB 1: Familiarity with Laboratory Equipment (_/10)

LAB 1: Familiarity with Laboratory Equipment (_/10) LAB 1: Familiarity with Laboratory Equipment (_/10) PURPOSE o gain familiarity with basic laboratory equipment oscilloscope, oscillator, multimeter and electronic components. EQUIPMEN (i) Oscilloscope

More information

AME140 Lab #2 INTRODUCTION TO ELECTRONIC TEST EQUIPMENT AND BASIC ELECTRONICS MEASUREMENTS

AME140 Lab #2 INTRODUCTION TO ELECTRONIC TEST EQUIPMENT AND BASIC ELECTRONICS MEASUREMENTS INTRODUCTION TO ELECTRONIC TEST EQUIPMENT AND BASIC ELECTRONICS MEASUREMENTS The purpose of this document is to guide students through a few simple activities to increase familiarity with basic electronics

More information

MS8268 HANDHELD DIGITAL MULTIMETER OPERATOR S INSTRUCTION MANUAL

MS8268 HANDHELD DIGITAL MULTIMETER OPERATOR S INSTRUCTION MANUAL MS8268 HANDHELD DIGITAL MULTIMETER OPERATOR S INSTRUCTION MANUAL Table of Contents TITLE PAGE 1. GENERAL INSTRUCTIONS 1 1.1 Precaution safety measures 1 1.1.1 Preliminary 1 1.1.2 During use 2 1.1.3 Symbols

More information

CA60 ma Current Clamp

CA60 ma Current Clamp Operating Instructions CA60 AC/DC ma CA6O Current Clamp 1mV / 1OOmA 1mV / 1OmA OFF POWER LO BAT ZERO INPUT: 6OA MAX OUTPUT: 1OmV/A, 1OOmV/A MAX _ VOLTAGE: 24OVAC/3OOVDC _ 3OOV. CATII 6OA ~ ~ Essential

More information

Current Probes. User Manual

Current Probes. User Manual Current Probes User Manual ETS-Lindgren Inc. reserves the right to make changes to any product described herein in order to improve function, design, or for any other reason. Nothing contained herein shall

More information

PARALLEL MULTI-AMP KIT for 7200 Series AMPLIFIERS INSTRUCTION SHEET

PARALLEL MULTI-AMP KIT for 7200 Series AMPLIFIERS INSTRUCTION SHEET 2 5 0 7 W a r r e n S t r e e t, E l k h a r t, I N 4 6 5 1 6 U S A 5 7 4. 2 9 5. 9 4 9 5 w w w. A E T e c h r o n. c o m PARALLEL MULTI-AMP KIT for 7200 Series AMPLIFIERS INSTRUCTION SHEET Kit Contents:

More information

8000 SERIES PRECISION MULTIMETER VERIFICATION AND ADJUSTMENT GUIDE

8000 SERIES PRECISION MULTIMETER VERIFICATION AND ADJUSTMENT GUIDE 8000 SERIES PRECISION MULTIMETER VERIFICATION AND ADJUSTMENT GUIDE TRANSMILLE LTD. Version 1.1 : Apr 2015 TABLE OF CONTENTS PREPARING FOR CALIBRATION... 4 INTRODUCTION... 4 CALIBRATION INTERVAL SELECTION...

More information

Manual Supplement. This supplement contains information necessary to ensure the accuracy of the above manual.

Manual Supplement. This supplement contains information necessary to ensure the accuracy of the above manual. Manual Title: 5502E Getting Started Supplement Issue: 3 Part Number: 4155211 Issue Date: 9/18 Print Date: November 2012 Page Count: 12 Revision/Date: This supplement contains information necessary to ensure

More information

User s Manual Current Probe. IM E 1st Edition. Yokogawa Electric Corporation

User s Manual Current Probe. IM E 1st Edition. Yokogawa Electric Corporation User s Manual 701932 Current Probe Yokogawa Electric Corporation 1st Edition Foreword Revisions Thank you for purchasing the 701932 Current Probe. This user's manual contains useful information about the

More information

Experiment 9 AC Circuits

Experiment 9 AC Circuits Experiment 9 AC Circuits "Look for knowledge not in books but in things themselves." W. Gilbert (1540-1603) OBJECTIVES To study some circuit elements and a simple AC circuit. THEORY All useful circuits

More information

5522A Multi-Product Calibrator Extended specifications

5522A Multi-Product Calibrator Extended specifications 5522A Multi-Product Calibrator Extended specifications General Specifications The following tables list the 5522A specifications. All specifications are valid after allowing a warm-up period of 30 minutes,

More information

INSTALATION, OPERATION & MAINTENANCE MANUAL. PA-1001A Series SIGNAL CONDITIONER & CONVERTORS

INSTALATION, OPERATION & MAINTENANCE MANUAL. PA-1001A Series SIGNAL CONDITIONER & CONVERTORS INSTALATION, OPERATION & MAINTENANCE MANUAL FOR PA-1001A Series SIGNAL CONDITIONER & CONVERTORS PA1001A 7/02 Page 1 of 11 SIGNAL CONDITIONER & CONVERTERS PA1001A Series INTRODUCTION: The PA1001A series

More information

Group: Names: Resistor Band Colors Measured Value ( ) R 1 : 1k R 2 : 1k R 3 : 2k R 4 : 1M R 5 : 1M

Group: Names: Resistor Band Colors Measured Value ( ) R 1 : 1k R 2 : 1k R 3 : 2k R 4 : 1M R 5 : 1M 2.4 Laboratory Procedure / Summary Sheet Group: Names: (1) Select five separate resistors whose nominal values are listed below. Record the band colors for each resistor in the table below. Then connect

More information

5320A. Multifunction Electrical Tester Calibrator. Extended Specifications

5320A. Multifunction Electrical Tester Calibrator. Extended Specifications 5320A Multifunction Electrical Tester Calibrator Extended Specifications General Specifications Warm-Up Time...30 minutes Specifications Confidence Level...99 % Specifications Interval...1 year Temperature

More information

DEPARTMENT OF THE ARMY TECHNICAL BULLETIN

DEPARTMENT OF THE ARMY TECHNICAL BULLETIN *TB 9-6625-2224-24 DEPARTMENT OF THE ARMY TECHNICAL BULLETIN CALIBRATION PROCEDURE FOR SIGNAL GENERATOR, SG-644/U (HEWLETT-PACKARD, MODEL 8614A) AND HEWLETT-PACKARD, 8616A Headquarters, Department of the

More information

80i-110s. AC/DC Current Probe Instructions. Introduction

80i-110s. AC/DC Current Probe Instructions. Introduction 80i-110s.book Page 1 Monday, December 5, 2016 12:12 PM 80i-110s AC/DC Current Probe Instructions Introduction The Fluke 80i-110s (the Probe or Product) is a clamp-on AC/DC Current Probe that reproduces

More information

5700A/5720A. Operator Guide. Series II Multi-Function Calibrator

5700A/5720A. Operator Guide. Series II Multi-Function Calibrator 5700A/5720A Series II Multi-Function Calibrator Operator Guide PN 3474006 May 1996, Rev. 1, 5/09 2009 Fluke Corporation. All rights reserved. Printed in USA. All product names are trademarks of their respective

More information

DISCUSSION The best way to test a transistor is to connect it in a circuit that uses the transistor.

DISCUSSION The best way to test a transistor is to connect it in a circuit that uses the transistor. Exercise 1: EXERCISE OBJECTIVE When you have completed this exercise, you will be able to test a transistor by forward biasing and reverse biasing the junctions. You will verify your results with an ohmmeter.

More information

Agilent U1253B True RMS OLED Multimeter. Quick Start Guide

Agilent U1253B True RMS OLED Multimeter. Quick Start Guide Agilent U1253B True RMS OLED Multimeter Quick Start Guide The following items are included with your multimeter: Silicone test leads 4 mm probes Alligator clips Printed Quick Start Guide Rechargeable 8.4

More information

DEPARTMENT OF THE ARMY TECHNICAL BULLETIN

DEPARTMENT OF THE ARMY TECHNICAL BULLETIN *TB 9-6625-2278-24 DEPARTMENT OF THE ARMY TECHNICAL BULLETIN CALIBRATION PROCEDURE FOR POWER METER HEWLETT-PACKARD, MODEL 438A AND 438AOPT002 Headquarters, Department of the Army, Washington, DC 17 June

More information

Laboratory Equipment Instruction Manual 2011

Laboratory Equipment Instruction Manual 2011 University of Toronto Department of Electrical and Computer Engineering Instrumentation Laboratory GB341 Laboratory Equipment Instruction Manual 2011 Page 1. Wires and Cables A-2 2. Protoboard A-3 3. DC

More information

AC Clamp Meter + Phase Rotation Tester

AC Clamp Meter + Phase Rotation Tester User's Manual AC Clamp Meter + Phase Rotation Tester Model 380974 Hz Introduction Congratulations on your purchase of the Extech 380974 AC Clamp Meter + Phase Rotation Tester. This professional 1000A Clamp

More information

Exercise 1: AC Waveform Generator Familiarization

Exercise 1: AC Waveform Generator Familiarization Exercise 1: AC Waveform Generator Familiarization EXERCISE OBJECTIVE When you have completed this exercise, you will be able to operate an ac waveform generator by using equipment provided. You will verify

More information

INSTALLATION & OPERATION MANUAL

INSTALLATION & OPERATION MANUAL INSTALLATION & OPERATION MANUAL AN25 Totalizer/ Rate Indicator DOC#: MN-AN25.doc Sponsler Co., Inc. AN25 Totalizer/Rate Indicator pg. 2 DOC#: MN-AN25 SPECIFICATIONS Temperature: Operating 0 to 70 C Storage

More information

RESISTANCE & OHM S LAW (PART I

RESISTANCE & OHM S LAW (PART I RESISTANCE & OHM S LAW (PART I and II) Objectives: To understand the relationship between potential and current in a resistor and to verify Ohm s Law. To understand the relationship between potential and

More information

Accuracy : ± (0.8% rdg + 1dgt) on all ranges except Ranges : 200mV, 2V, 20V, 200V, 1000V

Accuracy : ± (0.8% rdg + 1dgt) on all ranges except Ranges : 200mV, 2V, 20V, 200V, 1000V . SPECIFICATIONS Digital Multimeter Model : 0 Instruction Manual. General Specifications Display : ½ digit liquid crystal display (LCD) with a maximum reading of 999. Polarity : Automatic, (-)negative

More information

EXPERIMENT 1 PRELIMINARY MATERIAL

EXPERIMENT 1 PRELIMINARY MATERIAL EXPERIMENT 1 PRELIMINARY MATERIAL BREADBOARD A solderless breadboard, like the basic model in Figure 1, consists of a series of square holes, and those columns of holes are connected to each other via

More information

MODEL: D03128 CLAMP METER

MODEL: D03128 CLAMP METER MODEL: D03128 CLAMP METER 1 CONTENTS Page Number Details 3 Important Safety Information 3 Features 4 Product Overview 5 Switches, Buttons & Input Jacks 5 LCD 6 Specifications 6 Electrical Specifications

More information

User s Manual. Ground Resistance Clamp On Tester MODEL

User s Manual. Ground Resistance Clamp On Tester MODEL User s Manual Ground Resistance Clamp On Tester MODEL 382357 Warranty EXTECH INSTRUMENTS CORPORATION warrants the basic instrument to be free of defects in parts and workmanship for one year from date

More information

DIGITAL CLAMP METER Use s Manual

DIGITAL CLAMP METER Use s Manual MS2026 MS2026R DIGITAL CLAMP METER Use s Manual MS2026 FUNC. RANGE MAX/MIN Hz/% HOLD AC CLAMP METER Auto Range AUTO DC AC MAX MIN REL μnf %Hz kmω mva CAT III 600 V CONTENTS 1. Safety Information...1 1.1

More information

UNIVERSITY OF TECHNOLOGY, JAMAICA SCHOOL OF ENGENEERING. Electrical Engineering Science. Laboratory Manual

UNIVERSITY OF TECHNOLOGY, JAMAICA SCHOOL OF ENGENEERING. Electrical Engineering Science. Laboratory Manual UNIVERSITY OF TECHNOLOGY, JAMAICA SCHOOL OF ENGENEERING Electrical Engineering Science Laboratory Manual Table of Contents Experiment #1 OHM S LAW... 3 Experiment # 2 SERIES AND PARALLEL CIRCUITS... 8

More information

Loop calibration and maintenance

Loop calibration and maintenance Loop calibration and maintenance pplication Note Introduction Process instrumentation requires periodic calibration and maintenance to ensure that it is operating correctly. This application note contains

More information

The Art of Electrical Measurements

The Art of Electrical Measurements The Art of Electrical Measurements Purpose: Introduce fundamental electrical test and measurement tools and the art of making electrical measurements. Equipment Required Prelab 1 Digital Multimeter 1 -

More information