EPR spectrometer & components

Size: px
Start display at page:

Download "EPR spectrometer & components"

Transcription

1 EPR spectrometer & components Water lines µ wave bridge VC41 Gas Flow Controller Temperature Controller Vacuum pump for cryostat on left hand side water chiller HASKRIS EMX EPR Spectrometer EMX Magnet Power Supply Diaphragm pump

2 Getting Started (turning the instrument on) 1. Open the two yellow handled water valves simultaneously. Check for leaks and report if any are found. 2. Turn on the Haskris water chiller. The power may require cycling a few times (Chuck states 6 times in his instructions) if it has been off for a while. Again, check for leaks. 3. Turn on the computer, monitor, and printer if they are not already on. 4. Turn on the bridge power supply with the big red switch on the back of the spectrometer console. The liquid crystal frequency display on the µ wave bridge should display something. 5. Check the water chiller to confirm the orange liquid level indicator is on and the green temperature light is flashing if not, the chiller requires water. 6. Allow ~15 minutes for the temperature to stabilize. 7. Turn on the magnet power supply at the bottom front of the console. Two steady green lights should indicate all is okay. The magnet and bridge require ½ to 1 hour to stabilize, during that time open the WINEPR acquisition software (see step 8), prepare samples, etc. 8. Launch software with Bruker EMX icon located on the computer desktop

3 Preparing to Tune Make sure the bridge is in stand by mode Disconnect the iris motor shaft from the white Teflon iris screw Using the down arrow, run the iris control all the way down until arrow greys, indicating maximum depth is reached. Manually screw the white Teflon iris screw clockwise to the maximum down position, then back it out ½ turn. Re insert the motor shaft, and reposition the motor so that the shaft is vertical. Back the iris out using the mouse and the up arrow 2 3 full turns. This is most easily done with a second person watching. Tuning the Bridge Go to tune mode Set the attenuation to 30 db. Move the frequency slider slowly until a dip is observed on the view screen. This may be hard to find if the bias, signal phase, and iris are far from optimal. Once observed, center the dip with the frequency control. Adjust the signal phase for a symmetrical dip, resulting in a signal like this Set the bias control approximately center scale (50%)

4 Fine Tuning the Bridge Change to operate mode and set the attenuation to 40 db Adjust the frequency to lock the AFC meter to center scale Adjust the Bias level: Set the attenuation to 50 db; if the AFC meter falls out of lock, readjust with the frequency control. Then, adjust the Bias slider until the Diode meter needle is centered. Match the cavity: Set the attenuation to 40 db and re center the diode meter with the iris control (up and down arrows). This may take several turns of the iris and the diode may go down before it comes back up. Keep the frequency centered with the frequency control. Then repeat the procedure at 30 db, 20 db, and 10 db. Adjust the signal phase control in a continuous single direction, until you reach a local maximum in the diode current. Leave the signal phase at this maximum setting and do not adjust it further. Re center the diode with the iris control (up and down arrows). Decrease the attenuation in small steps, keeping the diode meter centered with the iris control. If the AFC drops out of lock increase the attenuation and relock it with the frequency control. Then, increase the attenuation to 50 db in 10 db increments. The meters should remain centered with out adjusting anything.

5 Inserting the sample Make sure the instrument is in stand by mode. Carefully remove the compression fitting at the top of the cavity and locate one with the proper size tube opening (if the cryostat is in, use the appropriate sized top hat). Insert the fitting and lightly tighten. Wipe the sample tube clean and place it in the cavity. Adjust the sample depth by gently pushing up on the pedestal (white adjustment plug, at the bottom of the cavity) or down on the sample as necessary to center the sample in the cavity. (If the cryostat is in place, there is no pedestal / white adjustment plug.) After inserting a sample it may be necessary to retune.

6 Parameter table See subsequent pages for discussion of parameters to optimize for best sensitivity

7 Parameters: Optimization Candidates Conversion Time the amount of time the A to D converter spends integrating one field position before moving to the next value. Affects the dynamic range Time Constant filters out noise Beware, if you choose a time constant that is excessively high, relative to your sweep time, you may filter out your signal Resolution in X (Number of Data Points) Determines the appropriate sweep time; you should have at least 10 points within the narrowest line that you are trying to resolve to properly define it. Number of points available are 512, 1024, 2048, 4096 or 8192 As you double the number of data points, you should also double the time constant Modulation Amplitude Use a field modulation that is approximately the width of the narrowest EPR line you are trying to resolve Excessive modulation broadens the EPR and does not contribute to a more intense signal Shown below is the effect of using progressively higher field modulation on the EPR spectrum

8 Parameters: Optimization Candidates Receiver Gain Effect of using gain settings that are either optimal (a), too low (b), or too high (c) Microwave Bridge Power The intensity of the EPR signal increases with the square root of the power, in the absence of saturation effects Effect of using progressively higher powers on EPR spectrum: Number of X scans or signal averaging Signal averaging can achieve the same effect as a long scan time when in the absence of a stable laboratory environment

9 Instrument Shut Down Go to standby mode in the MW bridge control window of the EPR software Remove sample and cover the top of the cavity Shut off the power to bridge in back and magnet power supply at the bottom front. Tidy up while additional cool water circulates around the bridge control unit. Shut off the Haskris water chiller Take your stuff back to lab, but remember to come back. All the warm water in the chiller needs to flush out. Turn off the water with the two yellow handles

10 Low Temperature Why? EPR frequencies (1 100 GHz) are in the microwave range o Aqueous solutions will warm up in the EPR cavity at RT The energy difference between the 2 energy levels due to the Zeeman splitting is very small, ~ 0.3 cm 1 for X band EPR. o Based on the Boltzmann distribution, n 1 = n 0 exp-(δe/kt), it can be shown that only at low temperatures there will be enough difference in the population of the S = ½ level (n0) and the S = + ½ level (n1) to create a signal. Spin Lattice Relaxation Rate decreases with decreasing temperature o The relaxation rate is anisotropic (i.e. is different for different parts of the spectrum). Due to the uncertainty principle the EPR spectra will broaden beyond detection at higher temperatures. At lower temperatures the spectra will sharpen up. o This sharpening up of the spectrum by cooling the sample is, however, limited by a temperature independent process: inhomogeneous broadening. For low temperature experiments, please see the facility manager about training and usage

Recording EPR Spectra using ER 4102ST Resonator

Recording EPR Spectra using ER 4102ST Resonator Recording EPR Spectra using ER 4102ST Resonator This protocol gives step-by-step instructions for recording an EPR spectrum using the high sensitivity Bruker SHQE cavity (assuming the SHQE is already in

More information

The University of Texas at Austin Electron Paramagnetic Resonance Facility. Bruker EMX+ CW EPR Spectrometer

The University of Texas at Austin Electron Paramagnetic Resonance Facility. Bruker EMX+ CW EPR Spectrometer The University of Texas at Austin Electron Paramagnetic Resonance Facility Bruker EMX+ CW EPR Spectrometer Contents: General Notes...2 Turning on the Instrument...2 Collecting a Spectrum...3 Turning off

More information

Recording EPR spectra using the Loop Gap Resonator (LGR)

Recording EPR spectra using the Loop Gap Resonator (LGR) Recording EPR spectra using the Loop Gap Resonator (LGR) This protocol gives step-by-step instructions for recording EPR spectra of spin labeled proteins (Nitroxide label like MTSSL) using the LGR assuming

More information

An EPR Primer 2. Basic EPR Theory 2.1. Introduction to Spectroscopy 2.1.1

An EPR Primer 2. Basic EPR Theory 2.1. Introduction to Spectroscopy 2.1.1 An EPR Primer 2 This chapter is an introduction to the basic theory and practice of EPR spectroscopy. It gives you sufficient background to understand the following chapters. In addition, we strongly encourage

More information

Magnettech GmbH Louis-Bleriot-Straße 5 D Berlin Germany

Magnettech GmbH Louis-Bleriot-Straße 5 D Berlin Germany MiniScope Control 6.51 IInstructiion Manuall Magnettech GmbH Louis-Bleriot-Straße 5 e-mail:info@magnettech.de MiniScope : ESR Spectrometer 2 1. TECHNICAL DATA...3 2. EXTENT OF DELIVERY...3 3. INSTALLATION...4

More information

Introduction 1. EPR Applications 1.1

Introduction 1. EPR Applications 1.1 Introduction 1 This document describes the basic operation of a Bruker Elexsys E 500 EPR (Electron Paramagnetic Resonance) spectrometer. No assumptions have been made about the background of the readers

More information

An Electron Spin Resonance Study Using a Toroidal Split Ring Resonator

An Electron Spin Resonance Study Using a Toroidal Split Ring Resonator An Electron Spin Resonance Study Using a Toroidal Split Ring Resonator AARON CLEMENTS SUPERVISED BY DR. JAKE BOBOWSKI Outline Background: The Zeeman effect the origin of Electron Spin Resonance (ESR) What

More information

Standard Operating Procedure for the Amray 1810 Scanning Electron Microscope Version: 29 NOVEMBER 2014

Standard Operating Procedure for the Amray 1810 Scanning Electron Microscope Version: 29 NOVEMBER 2014 Standard Operating Procedure for the Amray 1810 Scanning Electron Microscope Version: 29 NOVEMBER 2014 1. Utility Requirements a. System power is supplied by two 120 VAC/20 A circuits. When doing maintenance

More information

Operating Checklist for using the Scanning Electron Microscope, JEOL JSM 6400.

Operating Checklist for using the Scanning Electron Microscope, JEOL JSM 6400. Smith College August 2005 Operating Checklist for using the Scanning Electron Microscope, JEOL JSM 6400. CONTENT, page no. Pre-Check, 1 Specimen Insertion, 1 Startup, 2 Filament Saturation, 2 Beam Alignment,

More information

Operating Checklist for using the Scanning Electron. Microscope, JEOL JSM 6400.

Operating Checklist for using the Scanning Electron. Microscope, JEOL JSM 6400. Smith College August 2009 Operating Checklist for using the Scanning Electron Microscope, JEOL JSM 6400. CONTENT, page no. Pre-Check 1 Startup 1 Specimen Insertion 2 Filament Saturation 2 Beam Alignment

More information

OPERATION OF THE HITACHI S-450 SCANNING ELECTRON MICROSCOPE. by Doug Bray Department of Biological Sciences University of Lethbridge

OPERATION OF THE HITACHI S-450 SCANNING ELECTRON MICROSCOPE. by Doug Bray Department of Biological Sciences University of Lethbridge OPERATION OF THE HITACHI S-450 SCANNING ELECTRON MICROSCOPE by Doug Bray Department of Biological Sciences University of Lethbridge Revised September, 2000 Note: The terms in bold in this document represent

More information

The Discussion of this exercise covers the following points: Introduction How a tuning fork works Industrial applications. How a tuning fork works

The Discussion of this exercise covers the following points: Introduction How a tuning fork works Industrial applications. How a tuning fork works Exercise 3 Vibrating Level Switch EXERCISE OBJECTIVE Learn the working principle of vibrating level switches and learn how to use the vibrating level switch, Model 46933. DISCUSSION OUTLINE The Discussion

More information

Micro Automation- Model 1006 Dicing Saw Instructions. Serial # Rev 2 ( R.DeVito) Location Chase 1

Micro Automation- Model 1006 Dicing Saw Instructions. Serial # Rev 2 ( R.DeVito) Location Chase 1 Micro Automation- Model 1006 Dicing Saw Instructions Serial # Rev 2 (12-23-05 R.DeVito) Location Chase 1 Dicing Saw Instructions (Revised 8/9/03 - K.J) 1. On the Log Sheet sign in, including Name and Date.

More information

Alumajust Installation Instructions

Alumajust Installation Instructions Alumajust Installation Instructions Parts Index.Part No 105 106 107 108 109 110 111 112 150 A 150 160 A 160 B 160 160C Name & description Louver beam bracket beam to rafter connection Louver beam Drive

More information

Exercise 6. Range and Angle Tracking Performance (Radar-Dependent Errors) EXERCISE OBJECTIVE

Exercise 6. Range and Angle Tracking Performance (Radar-Dependent Errors) EXERCISE OBJECTIVE Exercise 6 Range and Angle Tracking Performance EXERCISE OBJECTIVE When you have completed this exercise, you will be familiar with the radardependent sources of error which limit range and angle tracking

More information

CNC Using the FlexiCam CNC and HMI Software. Guldbergsgade 29N, P0 E: T:

CNC Using the FlexiCam CNC and HMI Software. Guldbergsgade 29N, P0 E: T: CNC Using the FlexiCam CNC and HMI Software Guldbergsgade 29N, P0 E: makerlab@kea.dk T: +46 46 03 90 This grey box is the NC controller. Let s start by turning the red switch to the ON position, then press

More information

CRL MASS SPECTROMETRY FACILITY INSTRUMENT USER MANUAL MALDI MICRO. Operating Instructions. Basement Spec Lab:

CRL MASS SPECTROMETRY FACILITY INSTRUMENT USER MANUAL MALDI MICRO. Operating Instructions. Basement Spec Lab: CRL MASS SPECTROMETRY FACILITY INSTRUMENT USER MANUAL MALDI MICRO Operating Instructions Basement Spec Lab: 00.097 crs Page of 2 This is a guide to using the MALDI Micro for those who have received training.

More information

FPPO 1000 Fiber Laser Pumped Optical Parametric Oscillator: FPPO 1000 Product Manual

FPPO 1000 Fiber Laser Pumped Optical Parametric Oscillator: FPPO 1000 Product Manual Fiber Laser Pumped Optical Parametric Oscillator: FPPO 1000 Product Manual 2012 858 West Park Street, Eugene, OR 97401 www.mtinstruments.com Table of Contents Specifications and Overview... 1 General Layout...

More information

GC-1032 Metal Detector OWNER S MANUAL

GC-1032 Metal Detector OWNER S MANUAL GC-1032 Metal Detector OWNER S MANUAL 1 With your GC-1032 metal detector, you can hunt for coins, relics, jewelry, gold, and silver just about anywhere. The detector comes with high sensitivity and strong

More information

2. Refraction and Reflection

2. Refraction and Reflection 2. Refraction and Reflection In this lab we will observe the displacement of a light beam by a parallel plate due to refraction. We will determine the refractive index of some liquids from the incident

More information

Exercise 4. Angle Tracking Techniques EXERCISE OBJECTIVE

Exercise 4. Angle Tracking Techniques EXERCISE OBJECTIVE Exercise 4 Angle Tracking Techniques EXERCISE OBJECTIVE When you have completed this exercise, you will be familiar with the principles of the following angle tracking techniques: lobe switching, conical

More information

impact VC-500LR Monolight INSTRUCTIONS

impact VC-500LR Monolight INSTRUCTIONS impact lighting equipment and accessories VC-500LR Monolight INSTRUCTIONS Congratulations on your purchase of the Impact VC-500LR Monolight. We feel that it will contribute much to your photographic skill

More information

Introduction to Soldering

Introduction to Soldering Introduction to Soldering The Soldering Iron American Standard Wire Gage (AWG) Solder The Soldering Process Stripping & Tinning Wire Connecting/Soldering Wire Component Soldering De-Soldering Lab Exercise

More information

Basic Operating Instructions for Strata Dual Beam 235 FIB/SEM

Basic Operating Instructions for Strata Dual Beam 235 FIB/SEM Basic Operating Instructions for Strata Dual Beam 235 FIB/SEM Warning Always adjust your specimen height before closing the chamber door to make sure your specimen will not hit the bottom of the lens;

More information

Caution: Installation and Operation Instructions. Locking Mechanism (if supplied) Set Screws Backed Out. Mounting Orientation

Caution: Installation and Operation Instructions. Locking Mechanism (if supplied) Set Screws Backed Out. Mounting Orientation Genie Probe Regulator (GPR)/ Genie Probe (GP2)/ Genie Probe for Composite Sampler (CSA)/ And (Model H) Installation and Operation Instructions Caution: Not designed for external fire. Prior to use in a

More information

QUICKSILVER MX-190 OPERATING INSTRUCTIONS ,,-

QUICKSILVER MX-190 OPERATING INSTRUCTIONS ,,- QUICKSILVER MX-190 OPERATING INSTRUCTIONS -------..,,- INPUT CONNECTIONS To maintain a short and concise signal path, the input connectors are mounted directly on the plug-in front-end circuit boards.

More information

CHAPTER 6. Motor Driver

CHAPTER 6. Motor Driver CHAPTER 6 Motor Driver In this lab, we will construct the circuitry that your robot uses to drive its motors. However, before testing the motor circuit we will begin by making sure that you are able to

More information

WITec Alpha 300R Quick Operation Summary October 2018

WITec Alpha 300R Quick Operation Summary October 2018 WITec Alpha 300R Quick Operation Summary October 2018 This document is frequently updated if you feel information should be added, please indicate that to the facility manager (currently Philip Carubia,

More information

MICROWAVE AND RADAR LAB (EE-322-F) LAB MANUAL VI SEMESTER

MICROWAVE AND RADAR LAB (EE-322-F) LAB MANUAL VI SEMESTER 1 MICROWAVE AND RADAR LAB (EE-322-F) MICROWAVE AND RADAR LAB (EE-322-F) LAB MANUAL VI SEMESTER RAO PAHALD SINGH GROUP OF INSTITUTIONS BALANA(MOHINDERGARH)123029 Department Of Electronics and Communication

More information

ProLogic Xtreme L22 - Electromechanical Redundant Safe Lock System - Instructions

ProLogic Xtreme L22 - Electromechanical Redundant Safe Lock System - Instructions ProLogic Xtreme L22 - Electromechanical Redundant Safe Lock System - Instructions Please visit the website below by scanning the QR code with your smartphone or by typing in the address below for video

More information

MICROWAVE MICROWAVE TRAINING BENCH COMPONENT SPECIFICATIONS:

MICROWAVE MICROWAVE TRAINING BENCH COMPONENT SPECIFICATIONS: Microwave section consists of Basic Microwave Training Bench, Advance Microwave Training Bench and Microwave Communication Training System. Microwave Training System is used to study all the concepts of

More information

MASSACHUSETTS INSTITUTE OF TECHNOLOGY Department of Electrical Engineering and Computer Science

MASSACHUSETTS INSTITUTE OF TECHNOLOGY Department of Electrical Engineering and Computer Science Student Name Date MASSACHUSETTS INSTITUTE OF TECHNOLOGY Department of Electrical Engineering and Computer Science 6.161 Modern Optics Project Laboratory Laboratory Exercise No. 6 Fall 2010 Solid-State

More information

EECE 2413 Electronics Laboratory

EECE 2413 Electronics Laboratory EECE 2413 Electronics Laboratory Lab #2: Diode Circuits Goals In this lab you will become familiar with several different types of pn-junction diodes. These include silicon and germanium junction diodes,

More information

AVN Training HartRAO 2016

AVN Training HartRAO 2016 AVN Training HartRAO 2016 Microwave 1 Overview Introduction to basic components used in microwave receivers. Performance characteristics of these components. Assembly of components into a complete microwave

More information

Dickinson College Department of Geology

Dickinson College Department of Geology Dickinson College Department of Geology Title: Equipment: BASIC OPERATION OF THE SCANNING ELECTRON MICROSCOPE (SEM) JEOL JSM-5900 SCANNING ELECTRON MICROSCOPE Revision: 2.2 Effective Date: 1/29/2003 Author(s):

More information

Overview of Multi-Frequency EPR

Overview of Multi-Frequency EPR Overview of Multi-Frequency EPR Most EPR (Electron Paramagnetic Resonance) experiments are performed at X-band (9-10 GHz). Many factors contribute to the choice of this frequency. One factor is the availability

More information

MAHAVEER INSTITUTE OF SCIENCE & TECHNOLOGY. Microwave and Digital Communications Lab. Department Of Electronics and Communication Engineering

MAHAVEER INSTITUTE OF SCIENCE & TECHNOLOGY. Microwave and Digital Communications Lab. Department Of Electronics and Communication Engineering MAHAVEER INSTITUTE OF SCIENCE & TECHNOLOGY Microwave and Digital Communications Lab Department Of Electronics and Communication Engineering MICROWAVE ENGINEERING LAB List of Experiments: 1.Reflex Klystron

More information

Lab Equipment EECS 311 Fall 2009

Lab Equipment EECS 311 Fall 2009 Lab Equipment EECS 311 Fall 2009 Contents Lab Equipment Overview pg. 1 Lab Components.. pg. 4 Probe Compensation... pg. 8 Finite Instrumentation Impedance. pg.10 Simulation Tools..... pg. 10 1 - Laboratory

More information

DIODE LASER SPECTROSCOPY (160309)

DIODE LASER SPECTROSCOPY (160309) DIODE LASER SPECTROSCOPY (160309) Introduction The purpose of this laboratory exercise is to illustrate how we may investigate tiny energy splittings in an atomic system using laser spectroscopy. As an

More information

ARROW SAW PRECISE CUT 8000 RPM WITH DUST COLLECTING ATTACHMENT INSTRUCTION BOOK MODEL NO

ARROW SAW PRECISE CUT 8000 RPM WITH DUST COLLECTING ATTACHMENT INSTRUCTION BOOK MODEL NO ATTENTION If any components of this unit are broken or the unit does not operate properly, please contact Cabela s Customer Service. Retail Store Purchases: 1-800-905-2731 (U.S. & Canada) Catalog and Internet

More information

Scratch Assay Device for Wound Healing Studies

Scratch Assay Device for Wound Healing Studies Scratch Assay Device for Wound Healing Studies By Team Cell Scratcher: Hannah Morgan, Charles Merchant, and Zachariah Cribbin 1 Quick Start Guide 2 Comments: _ Table of Contents Cover Page 1 1.0 Quick

More information

Precalculations Individual Portion Introductory Lab: Basic Operation of Common Laboratory Instruments

Precalculations Individual Portion Introductory Lab: Basic Operation of Common Laboratory Instruments Name: Date of lab: Section number: M E 345. Lab 1 Precalculations Individual Portion Introductory Lab: Basic Operation of Common Laboratory Instruments Precalculations Score (for instructor or TA use only):

More information

EMO. Service Instruction. created by Frank Weithöner. Table of contents: Special Tools Assembling Mixing Chamber

EMO. Service Instruction. created by Frank Weithöner. Table of contents: Special Tools Assembling Mixing Chamber EMO Service Instruction created by Frank Weithöner Table of contents: - Special Tools Disassembling Mixing Chamber Assembling Mixing Chamber Adjustment Rotor / Level Indicator Unit Temperature Compensating

More information

GP-SD GENIE PROBE SMALL DIAMETER

GP-SD GENIE PROBE SMALL DIAMETER For GENIE PROBE SMALL DIAMETER Installation and Operation Instructions 2 for s Locking Mechanism (if supplied) Set Screws Backed Out To ensure that the is Locking never accidentally loosened or Mechanism

More information

Removing and Replacing the Y-truck

Removing and Replacing the Y-truck Service Documentation Removing and Replacing the Y-truck To remove and replace the Y-truck you will need the following tools: 4mm Allen wrench 12mm stamped flat wrench #2 Phillips screwdriver (magnetic

More information

Kit for building your own THz Time-Domain Spectrometer

Kit for building your own THz Time-Domain Spectrometer Kit for building your own THz Time-Domain Spectrometer 16/06/2016 1 Table of contents 0. Parts for the THz Kit... 3 1. Delay line... 4 2. Pulse generator and lock-in detector... 5 3. THz antennas... 6

More information

(N)MR Imaging. Lab Course Script. FMP PhD Autumn School. Location: C81, MRI Lab B0.03 (basement) Instructor: Leif Schröder. Date: November 3rd, 2010

(N)MR Imaging. Lab Course Script. FMP PhD Autumn School. Location: C81, MRI Lab B0.03 (basement) Instructor: Leif Schröder. Date: November 3rd, 2010 (N)MR Imaging Lab Course Script FMP PhD Autumn School Location: C81, MRI Lab B0.03 (basement) Instructor: Leif Schröder Date: November 3rd, 2010 1 Purpose: Understanding the basic principles of MR imaging

More information

Operation Hints. for the. MiniScope MS100-MS300

Operation Hints. for the. MiniScope MS100-MS300 Operation Hints for the MiniScope MS100-MS300 Magnettech GmbH Louis Bleriot Straße 5 D-12487 Berlin phone: +49 / 30 / 6780 2526 fax: +49 / 30 / 6322 4101 e-mail: info@magnettech.de homepage: http://www.magnettech.de

More information

FREQUENCY AGILE FM MODULATOR INSTRUCTION BOOK IB

FREQUENCY AGILE FM MODULATOR INSTRUCTION BOOK IB FMT615C FREQUENCY AGILE FM MODULATOR INSTRUCTION BOOK IB1215-02 TABLE OF CONTENTS SECTION SUBJECT 1.0 Introduction 2.0 Installation & Operating Instructions 3.0 Specification 4.0 Functional Description

More information

Lab 1: Basic Lab Equipment and Measurements

Lab 1: Basic Lab Equipment and Measurements Abstract: Lab 1: Basic Lab Equipment and Measurements This lab exercise introduces the basic measurement instruments that will be used throughout the course. These instruments include multimeters, oscilloscopes,

More information

R. J. Jones Optical Sciences OPTI 511L Fall 2017

R. J. Jones Optical Sciences OPTI 511L Fall 2017 R. J. Jones Optical Sciences OPTI 511L Fall 2017 Semiconductor Lasers (2 weeks) Semiconductor (diode) lasers are by far the most widely used lasers today. Their small size and properties of the light output

More information

Standard Kit #1 (3-way switch)

Standard Kit #1 (3-way switch) Standard Kit #1 (3-way switch) Please Read All Instructions Before Beginning. Tools you will need: Soldering Iron (35 watt preferably) Solder Wet Sponge Wire Clippers 3/8 Drill Bit 1/4 Drill Bit Variable

More information

NUMBER OF COPIES: 60 Rerun November 1963: 100

NUMBER OF COPIES: 60 Rerun November 1963: 100 NATIONAL RADIO ASTRONOMY OBSERVATORY Green Bank, West Virginia Electronics Division Internal Report No 9 PARAMETRIC AMPLIFIER AIRBORNE INSTRUMENTS LABORATORY 13y Dewey Ross June 1963 NUMBER OF COPIES:

More information

Repairing Microsoft Wedge Touch Mouse Battery Cover Retaining Clip

Repairing Microsoft Wedge Touch Mouse Battery Cover Retaining Clip Repairing Microsoft Wedge Touch Mouse Battery Cover Retaining Clip Disassembly, repair and reassembly of Wedge Touch mouse when the battery cover will not stay closed. Also is a good guide to repair other

More information

PM24 Installation Instructions

PM24 Installation Instructions Marchand Electronics Inc. PO Box 473, Webster, NY 14580 Tel:(716) 872-0980 Fax:(716) 872-1960 info@marchandelec.com http://www.marchandelec.com (c)1997 Marchand Electronics Inc. PM24 Installation Instructions

More information

Product End-of-Life Disassembly Instructions Product Category: Printers

Product End-of-Life Disassembly Instructions Product Category: Printers Product End-of-Life Disassembly Instructions Product Category: Printers Marketing Name / Model [List multiple models if applicable.] HP Officejet Pro X45dn/ CN459A HP Officejet Pro X45dw/CN463A HP Officejet

More information

AA-35 ZOOM. RigExpert. User s manual. Antenna and cable analyzer

AA-35 ZOOM. RigExpert. User s manual. Antenna and cable analyzer AA-35 ZOOM Antenna and cable analyzer RigExpert User s manual . Table of contents Introduction Operating the AA-35 ZOOM First time use Main menu Multifunctional keys Connecting to your antenna SWR chart

More information

EPR2010 Puerto Rico. Rapid Scan EPR. Mark Tseitlin, Deborah G. Mitchell, Joshua A. Biller, Richard W. Quine, George A. Rinard, Sandra S.

EPR2010 Puerto Rico. Rapid Scan EPR. Mark Tseitlin, Deborah G. Mitchell, Joshua A. Biller, Richard W. Quine, George A. Rinard, Sandra S. EPR2010 Puerto Rico Rapid Scan EPR Mark Tseitlin, Deborah G. Mitchell, Joshua A. Biller, Richard W. Quine, George A. Rinard, Sandra S. Eaton, Gareth R. Eaton, and Ralph T. Weber University of Denver and

More information

Please read carefuly before using. Aerobic Lateral Trainer ASSEMBLY MANUAL H901

Please read carefuly before using. Aerobic Lateral Trainer ASSEMBLY MANUAL H901 Please read carefuly before using. Aerobic Lateral Trainer ASSEMBLY MANUAL H901 TABLE OF CONTENTS 1 IMPORTANT SAFETY INFORMATION 2 Important Safety Precautions 2 Weight Limit Capacities 2 Warning 2 2

More information

LBI-4938C. Mobile Communications MASTR II POWER AMPLIFIER MODELS 4EF4A1,2,3. Printed in U.S.A. Maintenance Manual

LBI-4938C. Mobile Communications MASTR II POWER AMPLIFIER MODELS 4EF4A1,2,3. Printed in U.S.A. Maintenance Manual C Mobile Communications MASTR II POWER AMPLIFIER MODELS 4EF4A1,2,3 Printed in U.S.A. Maintenance Manual TABLE OF CONTENTS DESCRIPTION.................................................... 1 CIRCUIT ANALYSIS.................................................

More information

Quick Start MMRC Nicolet 6700 Revised 10/24/2018

Quick Start MMRC Nicolet 6700 Revised 10/24/2018 Do Not Remove Quick Start MMRC Nicolet 6700 Revised 10/24/2018 1. Fill out the logbook before you begin. 2. Before you start the Nicolet should be on with purge gas flowing. 3. Check that the purge gas

More information

Repair Instructions grand MA

Repair Instructions grand MA Repair Instructions grand MA Grand MA 1 Repair Instructions grand MA The following tasks may be performed by qualified personnel only! By disregarding the instructions or the prescribed sequence of tasks

More information

7. Experiment K: Wave Propagation

7. Experiment K: Wave Propagation 7. Experiment K: Wave Propagation This laboratory will be based upon observing standing waves in three different ways, through coaxial cables, in free space and in a waveguide. You will also observe some

More information

Lab 12 Microwave Optics.

Lab 12 Microwave Optics. b Lab 12 Microwave Optics. CAUTION: The output power of the microwave transmitter is well below standard safety levels. Nevertheless, do not look directly into the microwave horn at close range when the

More information

Micro-manipulated Cryogenic & Vacuum Probe Systems

Micro-manipulated Cryogenic & Vacuum Probe Systems Janis micro-manipulated probe stations are designed for non-destructive electrical testing using DC, RF, and fiber-optic probes. They are useful in a variety of fields including semiconductors, MEMS, superconductivity,

More information

Energate Foundation Meter Data Collector Installation Guide

Energate Foundation Meter Data Collector Installation Guide Energate Foundation Meter Data Collector Installation Guide The Meter Data Collector works with Foundation s built-in Meter Data Receiver. The collector attaches to the meter provided by your electricity

More information

Strata DB235 FESEM FIB

Strata DB235 FESEM FIB Strata DB235 FESEM FIB Standard Operating Procedure Revision: 5.0 Last Updated: August 16/2016, revised by Li Yang Overview This document will provide a detailed operation procedure of the Focused Ion

More information

SOP-P051. Scanning of Optical Filters With USB2000. Objective: To determine the spectral transmittance properties of an optical filter.

SOP-P051. Scanning of Optical Filters With USB2000. Objective: To determine the spectral transmittance properties of an optical filter. Purdue University Cytometry Laboratories SOP-P051 Scanning of Optical Filters With USB2000 Objective: To determine the spectral transmittance properties of an optical filter. Procedure: 1. Ensure that

More information

Physical Properties Measurement System (PPMS): Detailed specifications: Basic unit cryogen- free

Physical Properties Measurement System (PPMS): Detailed specifications: Basic unit cryogen- free Physical Properties Measurement System (PPMS): A Cryogen-free Physical Properties Measurement system that operates over a wider range of temperature and magnetic fields: fully automated/computer controlled

More information

Glenco Gas/Liquid Syringes from Spectrum Chromatography

Glenco Gas/Liquid Syringes from Spectrum Chromatography Glenco Gas/Liquid Syringes from Spectrum Chromatography D10122 180296 Copyright 1996 by Spectrum Chromatography PO Box 672026 Houston, TX 77267-2026 Phone: (281) 443-2900 (800) 459-9700 (US & Canada) Fax:

More information

Student Name: Date Completed: Supervisor:

Student Name: Date Completed: Supervisor: 2 NMR Training for the 600 MHz NMR with Chempack INOVA 600 Tests and Assignment Certification Student Name: 600-Test #1: The student will be given a written test administered by Dr. Lee. This test will

More information

Standard Kit #1 (5-way switch)

Standard Kit #1 (5-way switch) Standard Kit #1 (5-way switch) Please Read All Instructions Before Beginning. Tools you will need: Soldering Iron (35 watt preferably) Solder Wet Sponge Wire Clippers 3/8 Drill Bit 1/4 Drill Bit Variable

More information

KN-8828B Upgrade Directions

KN-8828B Upgrade Directions KN-8828B Upgrade Directions This document outlines the steps to take to update earlier Hottop Bean Roasters to the KN-8828B 2007 by Chang Yue and Hottop USA - All Rights Reserved No part of this document

More information

MetalliScanner 6.0 Components

MetalliScanner 6.0 Components MetalliScanner 6.0 Components 1. Mode Switch 2. Calibration Switch 3. Crosshairs 4. Liquid Crystal Display 5. Battery Compartment LCD Components 1. Depth Bars 2. Depth Numbers 3. Magnetic Icon 4. Low Battery

More information

User manual Bruker DPX200 NMR spectrometer

User manual Bruker DPX200 NMR spectrometer User manual Bruker DPX200 NMR spectrometer Insert the NMR tube in the spinner in such a way that the bottom of the tube reaches the grey disc at the bottom of the spinnerholder. Make sure that the NMR

More information

Rabbit Series Laser Engraving Machine HX3040. User s Manual. 1 Jinan King Rabbit Technology Development Co. Ltd.

Rabbit Series Laser Engraving Machine HX3040. User s Manual. 1 Jinan King Rabbit Technology Development Co. Ltd. Rabbit Laser Engraving Machine HX3040 User s Manual 1 Preface Thank you for buying our Rabbit Series HX3040 Laser Engraving Machine. You will find it has many nice features including easy operation, high

More information

Cartridge Machine USER MANUAL

Cartridge Machine USER MANUAL Cartridge Machine USER MANUAL 2018 USER MANUAL MACHINE OVERVIEW PACKAGE CONTENTS User Manual, Spektra Edge Body Hex Drive MotorBolt Oil Lubricant 0.05 Hex Screwdriver Spare 4-40 3/16 Stroke Set Screw Spare

More information

SoftRock v6.0 Builder s Notes. May 22, 2006

SoftRock v6.0 Builder s Notes. May 22, 2006 SoftRock v6.0 Builder s Notes May 22, 2006 Be sure to use a grounded tip soldering iron in building the v6.0 SoftRock circuit board. The soldering iron needs to have a small tip, (0.05-0.1 inch diameter),

More information

Instruction for Operating the Bruker Avance III 800 MHz NMR Spectrometers in UTMB

Instruction for Operating the Bruker Avance III 800 MHz NMR Spectrometers in UTMB Instruction for Operating the Bruker Avance III 800 MHz NMR Spectrometers in UTMB Written by Tianzhi Wang, date: February 8, 2013. No food, no drink in NMR room and no internet in NMR host computer except

More information

Pre-Laboratory Assignment

Pre-Laboratory Assignment Measurement of Electrical Resistance and Ohm's Law PreLaboratory Assignment Read carefully the entire description of the laboratory and answer the following questions based upon the material contained

More information

Maintenance Manual for Auto Lab

Maintenance Manual for Auto Lab Version 1.1 9/30/2012 Maintenance Manual for Auto Lab Lubricate X axis There are two locations for lubrication at the back side of the gantry as shown in the below picture. The left one (from the backside

More information

P H Y S I C A L P R O P E R T Y M E A S U R E M E N T S Y S T E M. Quantum Design

P H Y S I C A L P R O P E R T Y M E A S U R E M E N T S Y S T E M. Quantum Design P H Y S I C A L P R O P E R T Y M E A S U R E M E N T S Y S T E M Quantum Design S Y S T E M F E A T U R E S THE QUANTUM DESIGN PHYSICAL PROPERTY EASE OF USE MEASUREMENT SYSTEM (PPMS) REPRESENTS A UNIQUE

More information

RZ 200 Parts Diagram. To avoid confusion during installation remove this page and router being installed pages, return others to box.

RZ 200 Parts Diagram. To avoid confusion during installation remove this page and router being installed pages, return others to box. Collar Page 1 To avoid confusion during installation remove this page and router being installed pages, return others to box. Caution: Before and during installation of Router Raizer make sure power switch

More information

MaxiMist Pro Series Spray Gun

MaxiMist Pro Series Spray Gun MaxiMist Pro Series Spray Gun Your purchase includes the MaxiMist Pro Series Spray Gun Please read these instructions carefully and keep for your reference. Refer to the included instruction manual as

More information

ECE 203 LAB 6: INVERTED PENDULUM

ECE 203 LAB 6: INVERTED PENDULUM Version 1.1 1 of 15 BEFORE YOU BEGIN EXPECTED KNOWLEDGE Basic Circuit Analysis EQUIPMENT AFG Oscilloscope Programmable Power Supply MATERIALS Three 741 Opamps TIP41 NPN power transistor TIP42 PNP power

More information

Using the AmScope Microscope Cameras

Using the AmScope Microscope Cameras Using the AmScope Microscope Cameras Part 1 Setup. In order to use the camera, you will need: a) the camera system; b) a computer running the camera software. The camera system is contained in a Pelican

More information

PHYS 3153 Methods of Experimental Physics II O2. Applications of Interferometry

PHYS 3153 Methods of Experimental Physics II O2. Applications of Interferometry Purpose PHYS 3153 Methods of Experimental Physics II O2. Applications of Interferometry In this experiment, you will study the principles and applications of interferometry. Equipment and components PASCO

More information

Orbital Test Stand. Operations and Assembly Manual. Department of Mechanical Engineering Northern Arizona University Flagstaff, AZ 86011

Orbital Test Stand. Operations and Assembly Manual. Department of Mechanical Engineering Northern Arizona University Flagstaff, AZ 86011 Orbital Test Stand Operations and Assembly Manual Department of Mechanical Engineering Northern Arizona University Flagstaff, AZ 86011 TABLE OF CONTENTS 1.0 Components List... 2 2.0 Assembly Instructions...

More information

Physics 4C Chabot College Scott Hildreth

Physics 4C Chabot College Scott Hildreth Physics 4C Chabot College Scott Hildreth The Inverse Square Law for Light Intensity vs. Distance Using Microwaves Experiment Goals: Experimentally test the inverse square law for light using Microwaves.

More information

SPECIFICATION OF SOLAR CELL TESTER

SPECIFICATION OF SOLAR CELL TESTER SPECIFICATION OF SOLAR CELL TESTER A. Purpose: This equipment is to be used for testing of mono and multi crystalline silicon solar cells under standard testing conditions. The equipment shall be interfaced

More information

PN9000 PULSED CARRIER MEASUREMENTS

PN9000 PULSED CARRIER MEASUREMENTS The specialist of Phase noise Measurements PN9000 PULSED CARRIER MEASUREMENTS Carrier frequency: 2.7 GHz - PRF: 5 khz Duty cycle: 1% Page 1 / 12 Introduction When measuring a pulse modulated signal the

More information

I specifically did this for my 1980 Suzuki GS450L which uses the Mikuni BS34SS.

I specifically did this for my 1980 Suzuki GS450L which uses the Mikuni BS34SS. This is a carburetor cleaning guide for the Mikuni BS34SS which is used on the Suzuki GS450 E, L, and S models. Confirm that you have the same carburetors by looking at the pictures provided on the next

More information

FFP-C Fiber Fabry-Perot Controller OPERATING INSTRUCTIONS. Version 1.0 MICRON OPTICS, INC.

FFP-C Fiber Fabry-Perot Controller OPERATING INSTRUCTIONS. Version 1.0 MICRON OPTICS, INC. FFP-C Fiber Fabry-Perot Controller OPERATING INSTRUCTIONS Version 1.0 MICRON OPTICS, INC. 1852 Century Place NE Atlanta, GA 30345 USA Tel (404) 325-0005 Fax (404) 325-4082 www.micronoptics.com Page 2 Table

More information

Orion m Series Welding System User Manual

Orion m Series Welding System User Manual Orion m Series Welding System User Manual Table of Contents Chapter 1: Setup and Assembly... p.5 What is in the Box... p.5 Darkening Lens & Power Supply Setup... p.5 Microscope Arm Setup... p.6 Microscope

More information

2009 ODU Point and Peak Job Aid

2009 ODU Point and Peak Job Aid Summary This Job Aid covers: Preparing the Antenna for Pointing and Peaking Point Elevation Set the Skew Point Azimuth Peak Azimuth Peak Elevation Push/Pull Test This Job Aid supports all Technician audiences.

More information

1 FUNCTIONAL DESCRIPTION WAY SPLITTER/INPUT BOARD FET RF AMPLIFIERS WAY POWER COMBINER VSWR CONTROL BOARD...

1 FUNCTIONAL DESCRIPTION WAY SPLITTER/INPUT BOARD FET RF AMPLIFIERS WAY POWER COMBINER VSWR CONTROL BOARD... CONTENTS 1 FUNCTIONAL DESCRIPTION...1 2 4-WAY SPLITTER/INPUT BOARD...2 3 FET RF AMPLIFIERS...3 4 4-WAY POWER COMBINER...4 5 VSWR CONTROL BOARD...5 6 ADJUSTMENT OF BIAS VOLTAGE TO ESTABLISH PROPER QUIESCENT

More information

PM124 Installation Instructions. See important note about revisions of this board on the last page.

PM124 Installation Instructions. See important note about revisions of this board on the last page. Marchand Electronics Inc. PO Box 473, Webster, NY 14580 Tel:(716) 872-0980 Fax:(716) 872-1960 info@marchandelec.com http://www.marchandelec.com (c)1997 Marchand Electronics Inc. PM124 Installation Instructions

More information

Renishaw InVia Raman microscope

Renishaw InVia Raman microscope Laser Spectroscopy Labs Renishaw InVia Raman microscope Operation instructions 1. Turn On the power switch, system power switch is located towards the back of the system on the right hand side. Wait ~10

More information

Title: Amray 1830 SEM#2 Semiconductor & Microsystems Fabrication Laboratory Revision: D Rev Date: 03/18/2016

Title: Amray 1830 SEM#2 Semiconductor & Microsystems Fabrication Laboratory Revision: D Rev Date: 03/18/2016 Approved by: Process Engineer / / / / Equipment Engineer 1 SCOPE The purpose of this document is to detail the use of the Amray 1830 SEM. All users are expected to have read and understood this document.

More information

CUTTING SAMPLES WITH THE BUEHLER PETROTHIN: USER GUIDE

CUTTING SAMPLES WITH THE BUEHLER PETROTHIN: USER GUIDE Cutting Sample with the Buehler PetroThin CUTTING SAMPLES WITH THE BUEHLER PETROTHIN: USER GUIDE Author: Reviewer(s): Management Approval (Name, Title, Date): Audience: E. Fisher Origination date: 4/29/2013

More information