1 Hollandse Signaalapparaten B.V. Research, Development & Technology Environmental Test Laboratory P.O. Box 42, 755 GD Hengelo The Netherlands Experimental Results Obtained in the Vibrating Intrinsic Reverberation Chamber Frank Leferink Jean-Claude Boudenot Wim van Etten University of Twente Thomson-CSF Comsys Faculty of Electrical Engineering Hardening, Safe Instumentation Department of and Systems Unit Telecommunication Engineering BP 82, 9274 Colombes P.O. Box 217, 75 AE Enschede France The Netherlands comsys.thomson-csf.com University of Twente Faculty of Electrical Engineering Department of Telecommunication Engineering P.O. Box 217, 75 AE Enschede The Netherlands Abstract: Measurements have been performed in a Vibrating Intrinsic Reverberation Chamber (VIRC). This chamber has varying angles between wall, floor and ceiling. Inside the VIRC a diffuse, statistically uniform electromagnetic field is created without the use of a mechanical rotating mode stirrer. In comparison to other reverberation chambers the VIRC displays an improved low frequency behavior, enabling faster, cost effective testing and allows for in-situ measurements. Several measurement results obtained in the VIRC are presented in this paper, such as the change in resonance frequenccy, the stirring ratio and the power density function. Introduction A reverberation chamber generally consists of a rectangular test room with metal walls and a mode stirrer, usually in the form of a large paddle, near the ceiling of the chamber. The EUT is placed in the chamber and exposed to an electromagnetic field while the stirrer slowly revolves. The average response of the EUT to the field is found by integrating the response over the time period of one revolution of the stirrer. The metal walls of the chamber allow a large field to be built up inside the chamber. The EUT is therefore exposed to a high field level consisting of several different polarizations [1,2,3,4]. The electromagnetic field inside a reverberation chamber is not time invariant as in most classic test methods utilizing free space set-ups, but provides a periodic electromagnetic environment, which is randomly polarized, i.e. the phase between all waves is random, spatialy uniform, i.e. the energy density in the chamber is uniform everywhere and isotropic, i.e. the energy flow in all directions is the same. It has been shown that in a Mode Stirred Chamber (MSC) at least 2 modes should be available [3] in order to fulfill these requirements. In [6] several mode modification techniques, such as irregular walls, a moving wall and the Vibrating Intrinsic Reverberation Chamber (VIRC) have been discussed. The VIRC has vibrating walls in order to change the boundary conditions with large amplitude and short time variation. The VIRC is decribed in the next paragraph. Several tests have been carried out using the VIRC. The test set-ups and results are discussed showing the advantages of the VIRC compared to other techniques. The Vibrating Intrinsic Reverberation Chamber The VIRC is a reverberation chamber with walls made of flexible conducting material. It is mounted in a rigid structure and connected to that structure via flexible rubber strings, as shown in Figure 1, and drawn schematically in Figure 2. By moving one or more ridges or one or more walls the modal behavior of the field inside the chamber is changed, and thus the resonance frequencies are changed. Because this frequency shift is much larger compared to what is possible with a conventional mode stirrer, the frequency range of the chamber is extended to lower frequencies compared to a reverberation chambers with equal dimensions. Note the natural corrugation of the flexible walls in Figure 1, which is beneficial for the spatial uniformity too. Another advantage is that the flexible chamber can be erected inside a standard anechoic chamber where the EUT has been installed for standard EMI tests, thus reducing measurement time and cost. Furthermore the VIRC does not need extra space inside the laboratory: it can be folded and put away fast. The most important advantage of the flexible structure of the VIRC is that it can be installed in-situ. Figure 1: The Vibrating Intrinsic Reverberation Chamber hanging in strings

2 z h y w Figure 2: The VIRC: a flexible tent with irregularly shaped walls and varied angles of the wall-floor-ceiling interface Experimental characterisation of a reverberation chamber Numerous experiments could be carried out to characterize a reverberation chamber. Some of them are: Resonance frequency variation By measuring the field strength as function of the frequency for several (fixed) boundary conditions, i.e. VIRC positions, the resonance frequencies and change in resonance frequencies can be observed. The change in resonance frequency provides an indication of the stirring efficiency of the VIRC Voltage Standing Wave Ratio By measuring the incident and reflected power as function of the frequency the voltage standing wave ratio (VSWR) can be obtained. The VSWR indicates at what frequency the maximum energy transfer can be achieved. Stirring Ratio By recording the field strength as a function of time during a periodic or sufficiently random change in boundary conditions the so-called stirring ratio (SR) is obtained. This figure is defined as the ratio between the maximum and minimum field strength at a specific position and frequency: Emax SR( f ) = E The stirring ratio indicates the effectiveness of a stirring mechanism. If the stirring ratio would be small then the boundary conditions are changed only slightly, and, as a result, the spatial uniformity could be low. In [8] a stirring ratio of at least 2 db is considered as sufficient. The SR is an assess of the stirring efficiency in terms of field strength amplitude, which is an amplitude modulation. The change in resonance frequency due to a change in boundary condition is also an assess of the stirring efficiency, but now in terms of frequency modulation. Power Density Function The experimental power density function (PDF) is obtained by normalizing the measured absolute values of the electric field strength (in three directions) by their mean, and sorting the samples of total field strength from min l x the minimum field strength to the maximum field strength. By comparison of the experimental PDF with the theoretical PDF the quality of the chamber can be found: if the experimental PDF is equal to the theoretical PDF then the accuracy of a prediction is very high. Cumulative Density Function The cumulative probability density function (CDF) is obtained by integrating the PDF. Via the CDF a comparison of PDFs for different frequencies, or a comparison of chambers is facilitated. Spatial Field Uniformity Another important parameter is the spatial field uniformity (SFU). The SFU gives the ability to generate an isotropic, randomly polarised field which is stochastic equal in the whole volume of the chamber, exept near the walls. The SFU has not been measured yet. Quality Factor The chamber quality factor Q determines the efficiency of the chamber in relation to the power needed for generating a specific field strength or power density. The quality factor Q is an important parameter. It should be very high so that the generated field strength at a specific net input power is high, and thus a high efficiency. On the other hand, a low quality factor results in a higher mode overlap. The quality factor has not been measured yet. Measurement results The measurements were carried out using an experimental VIRC, as shown in Figure 1. The size of the VIRC was approximately 1.5 x 1.7 x 1.9 m. The VIRC was not vibrating for the measurement of the resonance frequency variation and the VSWR. For the other measurements (SR, PDF and CDF) the VIRC was randomly excited by three independent persons creating an amplitude variation of at least % of the length, width and heigth of the VIRC. Because the field strength inside the VIRC changes rapidly compared to the field inside the mode stirred chamber the standard field strength sensors, as used for EMI testing, are too slow. Therefore a very fast 3-axis sensor was used for the field strength measurements: the ET23 [7]. The advantages of this sensor are the low loading of the chamber, thus not decreasing the quality factor of the chamber, small dimensions (6x6x6 mm), large bandwidth with flat frequency response ( khz GHz), large dynamic range (nearly 8 db) and short risetime (2 ps) making this sensor very useful for reverberation chamber testing. Resonance frequency variation The field strength in the three orthogonal directions was measured inside the VIRC as function of the frequency using the set-up as drawn in Figure 3.

3 MHz and 2 MHz has been drawn in Figure 6. The resonance frequency variation in the VIRC is quite large, upto MHz for the f 11 mode, while the resonance frequency variation in the MSC is only a few MHz. 2 Figure 3: Measurement set-up for field modification An electromagnetic field was generated inside the VIRC by a log-spiral antenna (15 MHz-1 GHz). The antenna was activated by the output of the tracking generator of the Rohde&Schwarz ESS measuring receiver. The field strength inside the VIRC has been measured using the Thomson-CSF ET23 three-dimensional E-field sensor. The output signal of the sensor was fed to the optical receiver via an optical fiber. The output voltage level of the optical receiver was measured by the ESS measuring receiver for the three axes of the field. The total measured electric field strength between 9 MHz and 1 GHz has been drawn in Figure 4. During the measurements the VIRC was at two fixed positions (position 3 and 4), i.e. not vibrating. In this frequency range, 9 MHz 1 GHz, numerous resonance frequencies are available and a distinctive resonance frequency can thus not be observed. Field strength [dbv/m/dbm] coaxial - ESS measuring receiver optical BR2-3 optical receiver Etotal 3 Etotal Figure 4: Total electric field strength for VIRC position 3 and 4 between 9 MHz and 1 GHz In Figure 5 the total electric field strength between 5 MHz and 2 MHz has been drawn for the situation that the VIRC was at three fixed positions (position 2, 3 and 4). The resonance frequencies are evident. The resonance frequency variation due to the changed VIRC positions is also noticeable. A similar measurement has been carried out in a MSC with dimensions of approximately 2x2x2 m and for three stirrer positions. The total electric field strength between 5 Field strength [dbv/m/dbm] Field strength [dbuv/m] - -2 Etotal 2 Etotal 3 Etotal Figure 5: Total electric field strength as function of the frequency for VIRC position 2, 3 and f 1 f 1 f 11 MSC, Stirrer pos. 1 MSC, Stirrer pos. 2 MSC, Stirrer pos. 3 (stirrer moved) Figure 6: Total electric field strength as function of the frequency for a MSC with 3 stirrer positions Voltage standing wave ratio The standing wave ratio (SWR) is defined as the ratio between the incident and reflected power at a specific position and frequency. If the measurement impedances are equal (5 Ω) then the voltage SWR (VSWR) in terms of incident U i and reflected voltage U r is: Ur VSWR = U If the VSWR equals 1 then all incident energy is absorbed in the reverberation chamber. If the VSWR is high then a large portion of the energy is reflected. The VSWR for VIRC position 3 has been drawn in Figure 7, using the measurement set-up as drawn in Figure 3. From the VSWR we can calculate the net power injected into the VIRC. This net power has been drawn in Figure 8. The electric field strength for VIRC position 3 has been drawn in Figure 9. By comparison of Figure 8 and 9 the resemblance between net power and resonance frequencies is high, as could be expected. i

4 VSWR , 12 and 13 respectively. The measurement period was 4 s, with 8 samples and thus.5 ms/sample. It can be observed in the figures that the SR for 15 MHz was less than 2dB. For 4MHz and 1 GHz the SR satisfies the rule SR>2dB, as stated in [8]. Net Power [dbm] Figure 7: VSWR for VIRC position Figure 8: Net power for VIRC position 3 8 Field strength [dbv/m] - -2 Ex Ey Ez Etot Sample Figure 11: Field strength as function of time or stirring ratio, for 15 MHz Field strength [dbuv/m] VIRC Ex 3 VIRC Ey 3 VIRC Ez 3 Etotal Figure 9: Electric field strength for VIRC position 3 Stirring ratio The stirring ratio has been obtained by measuring the field strength as a function of the time at a fixed frequency. In the measurement period the VIRC was randomly excited by three independent persons creating an amplitude variation of at least % in width, heigth and length of the VIRC. The test setup has been drawn in Figure. The electromagnetic field inside the VIRC was generated by a R&S SMS-2 generator, amplified by a Kalmus amplifier and injected by the log-spiral antenna. The power level was adjusted to 3 dbm incident power. The field strength inside the VIRC has been measured using the Thomson-CSF ET23 three-dimensional E-field sensor. The output signal of the sensor was fed via an optical fiber to the optical receiver. The output voltage levels, for every axis (x,y,z) one channel, was detected by a logarithmic detector. This detector makes use of an AD 8313 integrated circuit, which has a bandwidth of 2.5 GHz, a dynamic range of 7 db, an amplitude flatness of ±1 db, a noise level of less than 73 dbm and a response time of 4 ns. The stirring ratio, or the field strength as function of the time, for 15 MHz, 4 MHz and 1 GHz has been drawn in Figures Field strength [dbv/m] Field strength [dbv/m] - -2 Ex Ey Ez Etot Sample Figure 12: Field strength as function of time, 4 MHz - -2 Ex Ey Ez Etot Sample Figure 13: Field strength as function of time, 1 GHz Probability density function The electric field strength has been measured in the three ortogonal directions while the boundary conditions were changing; i.e. the VIRC was vibrating. The movements were random: three persons were independently moving the VIRC. The test set-up was as drawn in Figure. Measurements have been performed using a sample time of 1 ms for a signal frequency of 15 MHz decreasing to.1 ms for a signal frequency of 1 GHz. For every signal frequency (15 MHz,

5 SMS-2 amplifier splitter Figure : Test set-up for measuring the field strength inside the VIRC as function of change in boundary conditions 2 MHz, 4 MHz, 6 MHz, 8 MHz and 1 GHz) 4 samples have been recorded. The samples were recorded simultaneously for the incident and reflected power and the field strength for every axis, giving a total of 5 channels with 4 samples. To obtain the probability density function the measured absolute value of the electric field strengths in the three directions were normalized. The samples of the total electric field strength have been sorted from the minimum field strength to the maximum field strength. This yields the experimental probability density function (PDF). The PDFs are shown for the frequencies 15 MHz, 2 MHz, 3 MHz, 4 MHz, 6 MHz, 8 MHz and 1 GHz in Figures 14 through 2. 3 power meter dir. coupler 5-channel logarithmic detector AR1 5-ch. analysing recorder coaxial optical BR2-3 optical receiver 5 45 Measured, VIRC, MP1, 3MHz Figure 16: PDF for 3 MHz 4 35 Measured, VIRC, MP1, 4MHz Figure 17: PDF for 4 MHz 35 3 Measured, VIRC, MP1, 6MHz Figure 18: PDF for 6 MHz Measured, VIRC, MP1, 15MHz 2 15 Measured, 8MHz, VIRC Figure 14: PDF for 15 MHz Measured, VIRC, MP1, 2MHz Figure 19: PDF for 8 MHz M easured, 1GHz, VIRC Figure 15: PDF for 2 MHz Figure 2: PDF for 1 GHz

6 The theory of the stochastic behaviour of reverberation chambers is described in detail in [4]. In [4, 9] it is shown that the magnitude of the electric field strength in any direction is Χ (Chi) distributed with two degrees of freedom, or Rayleigh distributed. This is based on the assumption that there is no direct path between the transmitting antenna and receiving sensor. If there is a direct path between the antenna and sensor then the Rayleigh distribution has to be modified resulting in the Rayleigh-Rice distribution. This distribution results in a longer tail for the total electric field strength distribution. From the comparison between measured and theoretical PDF in figures 14 to 2 we can observe that the measured PDF has a longer tail than the theoretical PDF. Because the dimension of the VIRC was small with respect to the transmitting antenna, the longer tail could be caused by a direct transmission path between transmitting antenna and field strength sensor. Cumulative probability density function The cumulative probability density function (CDF) for the measurement frequencies 15 MHz, 2 MHz, 3 MHz, 4 MHz, 6 MHz, 8 MHz and 1 GHz are shown in Figure 21. To estimate the quality of the chamber the experimental data has been compared to the theoretical curve. A similar measurement has been carried out in a MSC. The CDFs are drawn in Figure 22. The CDF of 2 MHz deviates considerably while the CDF of 15 MHz is nearly equal to the theoretical curve. Cumulative Probability Cumulative Probability % 8% 6% 4% 2% % % 8% 6% 4% 2% 1 GHz CHI (6DOF) VIRC, MP1, 15MHz VIRC, MP1, 2MHz VIRC, MP1, 3MHz VIRC, MP1, 4MHz VIRC, MP1, 6MHz VIRC, MP1, 8MHz VIRC, MP1, MHz Field Strength w.r.t. Mean Figure 21: CDF for the VIRC Theoretical CDF 2 MHz Theoretical CDF 15 MHz 2 MHz Theory, Chi(6dof) MSC 1, 2MHz MSC 1, 4MHz MSC 1, 6MHz MSC 1, 8MHz MSC 1, MHz % Field Strength w.r.t. Mean Figure 22: CDF for a MSC 1 GHz Furthermore the CDFs for higher frequencies are lower than the theoretical CDF, for both the VIRC and the MSC. This could be caused by the direct ray path and corresponding Rayleigh-Rice distribution as described in the previous paragraph. A comparison of the VIRC and MSC CDFs, as shown in Figure 21 and 22, shows that the VIRC CDFs are approaching the theoretical CDF better than the MSC CDF. Therefore we can conclude that an estimation of the field strength in the VIRC is much more reliable than an estimation of the field strength in this MSC. We have to admit that the MSC used for the experiments is not optimized as reverberation chamber, and a properly operated MSC should result in a better CDF. Conclusion A new type of reverberation chamber to create a spatial uniform and isotropic electromagnetic field is the Vibrating Intrinsic Reverberation Chamber (VIRC). Several measurements have been performed within the VIRC. It was shown that the VIRC can even change the lowest resonance frequency by more than MHz, whereas the conventional mode stirred chamber (MSC) resonance frequency changes only a few MHz. The stirring ratio was more than 4 db for low frequencies, which proves that the VIRC is capable to change the boundary conditions adequately. The field strength at several frequencies has been measured as function of the time and the samples of the total electric field strength have been sorted from the minimum field strength to the maximum field strength, which yields the experimental probability density function. The corresponding cumulative probability density function (CDF) has been compared to the theoretical CDF and shows good resemblance. References [1] P. Corona, G. Latmiral, E. Paolini, L. Piccioli, Use of a reverberating enclosure for measurements of radiated power in the microwave range, IEEE Transactions on Electromagnetic Compatibility, vol.emc-18, no.2, May 1976, p.54-9 [2] M.L. Crawford, G.H. Koepke, Operational Considerations of a Reverberation Chamber for EMC immunity Measurements - Some Experimental Results, IEEE Symposium on EMC, 1984 [3] M.L. Crawford, G.H. Koepke, Design, Evaluation and Use of a Reverberation Chamber for Performing Electromagnetic Susceptibility/Vulnerability Measurements, NBS Technical Note 92, April [4] D.A. Hill, Electromagnetic theory of reverberation chambers, National Institute of Standards and Technology Technical note 156, dec [5] D.I. Wu, D.C. Chang, The Effect of an Electrically Large Stirrer in a Mode-Stirred Chamber, IEEE Transactions on Electromagnetic Compatibility, May 1989, pp [6] F.B.J. Leferink, W. van Etten, The Vibrating Intrinsic Reverberation Chamber, IEEE Transactions on Electromagnetic Compatibility, 2 [7] Mélopée-ET23, 3-axis Electric Field Sensor, Thomson-CSF [8] M. Petirsch, A. Schwab, Optimizing Shielded Chambers Utilizing Acoustic Analogies, IEEE Symposium on EMC, 1997, pp [9] J.G. Kostas, B. Boverie, Statistical model for a mode stirred chamber, IEEE Transactions on Electromagnetic Compatibility, vol. 33, no. 4, nov. 1991, pp [] C.W. Helstrom, Probability and Stochastic Processes for Engineers, 2- ed., MacMillan, 1984

Experimental Results Obtained in the Vibrating Intrinsic Reverberation Chamber

Experimental Results Obtained in the Vibrating Intrinsic Reverberation Chamber Experimental Results Obtained in the Vibrating ntrinsic Reverberation Chamber Frank Leferink Hollandse Signaalapparaten B.V. University of Twente Research, Development & Faculty of Electrical Engineering

More information

In-situ EMI Measurements Using a Vibrating Intrinsic Reverberation Chamber

In-situ EMI Measurements Using a Vibrating Intrinsic Reverberation Chamber In-situ EMI Measurements Using a Vibrating Intrinsic Reverberation Chamber Frank B.J. Leferink *+, Dick J. Groot Boerle *, Fons A.G. Sogtoen *, Geert H.L.M Heideman +, Wim C. van Etten + * Thales Naval

More information

REVERBERATION CHAMBER FOR EMI TESTING

REVERBERATION CHAMBER FOR EMI TESTING 1 REVERBERATION CHAMBER FOR EMI TESTING INTRODUCTION EMI Testing 1. Whether a product is intended for military, industrial, commercial or residential use, while it must perform its intended function in

More information

High Field Strength in a Large Volume: The Extended Double Ridged Guide vs. the Biconical and Log-Periodic Antenna

High Field Strength in a Large Volume: The Extended Double Ridged Guide vs. the Biconical and Log-Periodic Antenna High Field Strength in a Large Volume: The Extended Double Ridged Guide vs. the Biconical and Log-eriodic Antenna Frank B.J. Leferink, Dick J. Groot Boerle Hollandse Signaalapparaten B.V. Environmental

More information

Correlation Between Measured and Simulated Parameters of a Proposed Transfer Standard

Correlation Between Measured and Simulated Parameters of a Proposed Transfer Standard Correlation Between Measured and Simulated Parameters of a Proposed Transfer Standard Jim Nadolny AMP Incorporated ABSTRACT Total radiated power of a device can be measured using a mode stirred chamber

More information

pel.com Microwave & RF, March 23 rd, 2016.

pel.com Microwave & RF, March 23 rd, 2016. www.siep pel.com Generating high EM fields using mode-stirred reverberation chambers for RTCA DO 160 applications Jean-François ROSNARHO Microwave & RF, March 23 rd, 2016. TABLE OF CONTENTS 1. RTCA DO

More information

A Comparison of the Microwave Oven and Reverberation Chamber

A Comparison of the Microwave Oven and Reverberation Chamber A Comparison of the Microwave Oven and Reverberation Chamber Yi Huang, Xu Zhu and Binoy Nair The University of Liverpool, L69 3GJ, UK Abstract The reverberation chamber is a facility mainly for electromagnetic

More information

Immunity Test System RIS 3000 / RIS 6000 acc. to IEC/EN

Immunity Test System RIS 3000 / RIS 6000 acc. to IEC/EN Description The setup of a radiated immunity test system can be done in the conventional way with many separate instruments or in a more comfortable and less risky way with our new EMC control unit, type

More information

Rationale for RC Testing. Vignesh Rajamani, PhD. Research Engineer Oklahoma State University

Rationale for RC Testing. Vignesh Rajamani, PhD. Research Engineer Oklahoma State University Rationale for RC Testing Vignesh Rajamani, PhD. Research Engineer Oklahoma State University Overview What is a reverberation chamber? Immunity testing Emissions testing Shielding effectiveness measurements

More information

CHRISTIAN S. LÖTBÄCK PATANÉ. Master of Science Thesis

CHRISTIAN S. LÖTBÄCK PATANÉ. Master of Science Thesis Reverberation Chamber Performance and Methods for Estimating the Rician K-factor Evaluation of Reverberation Chamber Measurements at the National Institute of Standards and Technology in Boulder, Colorado,

More information

A Novel Method for Determining the Lower Bound of Antenna Efficiency

A Novel Method for Determining the Lower Bound of Antenna Efficiency A Novel Method for Determining the Lower Bound of Antenna Efficiency Jason B. Coder #1, John M. Ladbury 2, Mark Golkowski #3 # Department of Electrical Engineering, University of Colorado Denver 1201 5th

More information

Numerical Study of Stirring Effects in a Mode-Stirred Reverberation Chamber by using the Finite Difference Time Domain Simulation

Numerical Study of Stirring Effects in a Mode-Stirred Reverberation Chamber by using the Finite Difference Time Domain Simulation Forum for Electromagnetic Research Methods and Application Technologies (FERMAT) Numerical Study of Stirring Effects in a Mode-Stirred Reverberation Chamber by using the Finite Difference Time Domain Simulation

More information

Practical Considerations for Radiated Immunities Measurement using ETS-Lindgren EMC Probes

Practical Considerations for Radiated Immunities Measurement using ETS-Lindgren EMC Probes Practical Considerations for Radiated Immunities Measurement using ETS-Lindgren EMC Probes Detectors/Modulated Field ETS-Lindgren EMC probes (HI-6022/6122, HI-6005/6105, and HI-6053/6153) use diode detectors

More information

CHARACTERISATION OF IN -HOUSE EMC TESTING FACILITIES FOR PRODUCT DESIGNERS. Paul Kay* and Andrew Nafalski**

CHARACTERISATION OF IN -HOUSE EMC TESTING FACILITIES FOR PRODUCT DESIGNERS. Paul Kay* and Andrew Nafalski** CHARACTERISATION OF IN -HOUSE EMC TESTING FACILITIES FOR PRODUCT DESIGNERS Paul Kay* and Andrew Nafalski** *Austest Laboratories, Adelaide **University of South Australia School of Electrical and Information

More information

Selecting the right antenna for the

Selecting the right antenna for the Zhong Chen ETS-Lindgren EMC Antenna Fundamentals Selecting the right antenna for the job can be a difficult task. In many cases, manufacturer terminologies and specifications are so varied that it is difficult

More information

STSM: On-Site Emission Measurements

STSM: On-Site Emission Measurements On-Site Emission Measurements Based on Reverberation Chamber Techniques Robert Vogt, STSM - SP, Boras, Sweden 20.09 03.10.2015 CONTENTS On-site emission measurements (Motivation) Test site classification

More information

Test and Measurement for EMC

Test and Measurement for EMC Test and Measurement for EMC Bogdan Adamczyk, Ph.D., in.c.e. Professor of Engineering Director of the Electromagnetic Compatibility Center Grand Valley State University, Michigan, USA Ottawa, Canada July

More information

Characteristics of Biconical Antennas Used for EMC Measurements

Characteristics of Biconical Antennas Used for EMC Measurements Advance Topics in Electromagnetic Compatibility Characteristics of Biconical Antennas Used for EMC Measurements Mohsen Koohestani koohestani.mohsen@epfl.ch Outline State-of-the-art of EMC Antennas Biconical

More information

Compact MIMO Antenna with Cross Polarized Configuration

Compact MIMO Antenna with Cross Polarized Configuration Proceedings of the 4th WSEAS Int. Conference on Electromagnetics, Wireless and Optical Communications, Venice, Italy, November 2-22, 26 11 Compact MIMO Antenna with Cross Polarized Configuration Wannipa

More information

Electromagnetic Compatibility ( EMC )

Electromagnetic Compatibility ( EMC ) Electromagnetic Compatibility ( EMC ) Introduction EMC Testing 1-2 -1 Agenda System Radiated Interference Test System Conducted Interference Test 1-2 -2 System Radiated Interference Test Open-Area Test

More information

ISO INTERNATIONAL STANDARD

ISO INTERNATIONAL STANDARD INTERNATIONAL STANDARD ISO 11452-11 First edition 2010-09-01 Road vehicles Component test methods for electrical disturbances from narrowband radiated electromagnetic energy Part 11: Reverberation chamber

More information

E-Field Uniformity Test Volume In Gtem Cell Based On Labview

E-Field Uniformity Test Volume In Gtem Cell Based On Labview www.ijecs.in International Journal Of Engineering And Computer Science ISSN:2319-7242 Volume 4 Issue 4 April 215, Page No. 11646-1165 E-Field Uniformity Test Volume In Gtem Cell Based On Labview Dominic

More information

4GHz / 6GHz Radiation Measurement System

4GHz / 6GHz Radiation Measurement System 4GHz / 6GHz Radiation Measurement System The MegiQ Radiation Measurement System (RMS) is a compact test system that performs 3-axis radiation pattern measurement in non-anechoic spaces. With a frequency

More information

Shielding Effectiveness Summary Results for RadiaShield Technologies, Inc. RadiaShield Fabric

Shielding Effectiveness Summary Results for RadiaShield Technologies, Inc. RadiaShield Fabric Test Date(s): July 9 through July 19, 2010 UST Project Number: 10-0164 Summary Results for Product Description The Sample Under Test (SUT) is the. The SUT is a textile which is used as a protective shield

More information

A GTEM BEST PRACTICE GUIDE APPLYING IEC TO THE USE OF GTEM CELLS

A GTEM BEST PRACTICE GUIDE APPLYING IEC TO THE USE OF GTEM CELLS - 27-39 H1 A BEST PRACTICE GUIDE APPLYING IEC 61-4-2 TO THE USE OF CELLS A. Nothofer, M.J. Alexander, National Physical Laboratory, Teddington, UK, D. Bozec, D. Welsh, L. Dawson, L. McCormack, A.C. Marvin,

More information

Shielding Effectiveness Report HQDP

Shielding Effectiveness Report HQDP HQDP Mates with QSH-DP, QTH-DP Description: 0.50mm 100Ω Differential 30 AWG Twinax Cable Assembly Samtec, Inc. 2005 All Rights Reserved Table of Contents Product Overview... 1 Test Overview... 1 Shielded

More information

Statistical Analysis of Modern Communication Signals

Statistical Analysis of Modern Communication Signals Whitepaper Statistical Analysis of Modern Communication Signals Bob Muro Application Group Manager, Boonton Electronics Abstract The latest wireless communication formats like DVB, DAB, WiMax, WLAN, and

More information

Specification for Radiated susceptibility Test

Specification for Radiated susceptibility Test 1 of 11 General Information on Radiated susceptibility test Supported frequency Range : 20MHz to 6GHz Supported Field strength : 30V/m at 3 meter distance 100V/m at 1 meter distance 2 of 11 Signal generator

More information

Considerations about Radiated Emission Tests in Anechoic Chambers that do not fulfil the NSA Requirements

Considerations about Radiated Emission Tests in Anechoic Chambers that do not fulfil the NSA Requirements 6 th IMEKO TC Symposium Sept. -, 8, Florence, Italy Considerations about Radiated Emission Tests in Anechoic Chambers that do not fulfil the NSA Requirements M. Borsero, A. Dalla Chiara 3, C. Pravato,

More information

MICROWAVE MICROWAVE TRAINING BENCH COMPONENT SPECIFICATIONS:

MICROWAVE MICROWAVE TRAINING BENCH COMPONENT SPECIFICATIONS: Microwave section consists of Basic Microwave Training Bench, Advance Microwave Training Bench and Microwave Communication Training System. Microwave Training System is used to study all the concepts of

More information

FISCHER CUSTOM COMMUNICATIONS, INC.

FISCHER CUSTOM COMMUNICATIONS, INC. FISCHER CUSTOM COMMUNICATIONS, INC. Current Probe Catalog FISCHER CUSTOM COMMUNICATIONS, INC. Fischer Custom Communications, Inc., is a manufacturer of custom electric and magnetic field sensors for military

More information

Microwave Circuit Design and Measurements Lab. INTRODUCTION TO MICROWAVE MEASUREMENTS: DETECTION OF RF POWER AND STANDING WAVES Lab #2

Microwave Circuit Design and Measurements Lab. INTRODUCTION TO MICROWAVE MEASUREMENTS: DETECTION OF RF POWER AND STANDING WAVES Lab #2 EE 458/558 Microwave Circuit Design and Measurements Lab INTRODUCTION TO MICROWAVE MEASUREMENTS: DETECTION OF RF POWER AND STANDING WAVES Lab #2 The purpose of this lab is to gain a basic understanding

More information

YRS01 York Reference Source

YRS01 York Reference Source York Reference Source York Reference Source The YRS01 is a multi-mode reference source capable of producing a broadband noise or comb output up to 1GHz. Selectable noise or comb output - Flexibility across

More information

EMI T E S T R E P O R T

EMI T E S T R E P O R T EMI T E S T R E P O R T - FCC Part 15B - Test Report No. : T38935-00-02TK 27. November 2014 Date of issue Type / Model Name : One Touch Select Plus Flex Product Description : Blood glucose meter with Bluetooth

More information

ELEC 0017: ELECTROMAGNETIC COMPATIBILITY LABORATORY SESSIONS

ELEC 0017: ELECTROMAGNETIC COMPATIBILITY LABORATORY SESSIONS Academic Year 2015-2016 ELEC 0017: ELECTROMAGNETIC COMPATIBILITY LABORATORY SESSIONS V. BEAUVOIS P. BEERTEN C. GEUZAINE 1 CONTENTS: EMC laboratory session 1: EMC tests of a commercial Christmas LED light

More information

Reverberation Chambers

Reverberation Chambers Reverberation Chambers Andy Lambourne 29 th November 2007 Contents 01 Introduction 02 Uses of a Reverberation Chamber 03 Example Facilities 04 Chamber Properties 05 Standards and Test Methods 06 Calibration

More information

Mode-Stirred Method Implementation for HIRF Susceptibility Testing and Results Comparison with Anechoic Method

Mode-Stirred Method Implementation for HIRF Susceptibility Testing and Results Comparison with Anechoic Method NASA/TM-2001-211033 Mode-Stirred Method Implementation for HIRF Susceptibility Testing and Results Comparison with Anechoic Method Truong X. Nguyen and Jay J. Ely Langley Research Center, Hampton, Virginia

More information

7. Experiment K: Wave Propagation

7. Experiment K: Wave Propagation 7. Experiment K: Wave Propagation This laboratory will be based upon observing standing waves in three different ways, through coaxial cables, in free space and in a waveguide. You will also observe some

More information

FDTD CHARACTERIZATION OF MEANDER LINE ANTENNAS FOR RF AND WIRELESS COMMUNICATIONS

FDTD CHARACTERIZATION OF MEANDER LINE ANTENNAS FOR RF AND WIRELESS COMMUNICATIONS Progress In Electromagnetics Research, PIER 4, 85 99, 999 FDTD CHARACTERIZATION OF MEANDER LINE ANTENNAS FOR RF AND WIRELESS COMMUNICATIONS C.-W. P. Huang, A. Z. Elsherbeni, J. J. Chen, and C. E. Smith

More information

Shielding Effectiveness Report

Shielding Effectiveness Report VRDPC-050-01-S-D-RA Mates with VPDP/VPLSP/VPSTP Description: Data Rate I/O Cable Assemblies Samtec, Inc. 2005 All Rights Reserved Table of Contents Product Overview... 1 Shielded Room Noise Floor Verification...

More information

EMC/EMI MEASURING INSTRUMENTS & ACCESSORIES SHORT-FORM CATALOG 2009

EMC/EMI MEASURING INSTRUMENTS & ACCESSORIES SHORT-FORM CATALOG 2009 EMC/EMI MEASURING INSTRUMENTS & ACCESSORIES SHORT-FORM CATALOG 2009 Our trek started in a small laboratory over 25 years ago. Since then, we ve been focused on making EMC measurements easier and the measuring

More information

Company report to support the development of the Ecma MM-EMF technical report (attached)

Company report to support the development of the Ecma MM-EMF technical report (attached) Ecma/TC20/2009/019 DATE CIRCULATION: 2009-04-07 FROM: TO: DOC TYPE: TITLE: AUTHOR: P.A. Beeckman, Philips Applied Technologies Members Ecma TC20-MM-EMF Company report to support the development of the

More information

CHAPTER 6 EMI EMC MEASUREMENTS AND STANDARDS FOR TRACKED VEHICLES (MIL APPLICATION)

CHAPTER 6 EMI EMC MEASUREMENTS AND STANDARDS FOR TRACKED VEHICLES (MIL APPLICATION) 147 CHAPTER 6 EMI EMC MEASUREMENTS AND STANDARDS FOR TRACKED VEHICLES (MIL APPLICATION) 6.1 INTRODUCTION The electrical and electronic devices, circuits and systems are capable of emitting the electromagnetic

More information

EMI -- T E S T R E P O R T

EMI -- T E S T R E P O R T Registration No. DAT-P-207/05 EMI -- T E S T R E P O R T - FCC Part 15B - Test Report No. : T32619-00-04HU 24. July 2008 Date of issue Type / Model Name : R-PO7470 Product Description : Handheld Reader

More information

Regarding RF Isolation for small Enclosures

Regarding RF Isolation for small Enclosures Regarding RF Isolation for small Enclosures IEEE electromagnetic society and IEEE standard board has published standards for measuring the shielding effectiveness (SE) of chambers. The measurement methods

More information

Test specification: Section (e)(1), Radiated emissions below 40 GHz Test procedure: ANSI C63.4, Sections 8.3.2, 13.2, 13.4 Test mode: Compliance

Test specification: Section (e)(1), Radiated emissions below 40 GHz Test procedure: ANSI C63.4, Sections 8.3.2, 13.2, 13.4 Test mode: Compliance Test specification: Section 15.253(e)(1), Radiated emissions below 40 GHz Test procedure: ANSI C63.4, Sections 8.3.2, 13.2, 13.4 Plot 7.2.7 Radiated emission measurements at frequency 7280 MHz Low channel

More information

Broadband Antenna FDTD Modeling for EMC Test

Broadband Antenna FDTD Modeling for EMC Test Broadband Antenna FDTD Modeling for EMC Test R. Jauregui, M. A. Heras and F. Silva Grup de Compatibilitat Electromagnètica (GCEM),Departament d Enginyeria Electrònica (DEE), Universitat Politècnica de

More information

Characterization of a Photonics E-Field Sensor as a Near-Field Probe

Characterization of a Photonics E-Field Sensor as a Near-Field Probe Characterization of a Photonics E-Field Sensor as a Near-Field Probe Brett T. Walkenhorst 1, Vince Rodriguez 1, and James Toney 2 1 NSI-MI Technologies Suwanee, GA 30024 2 SRICO Columbus, OH 43235 bwalkenhorst@nsi-mi.com

More information

Network Analysis Basics

Network Analysis Basics Adolfo Del Solar Application Engineer adolfo_del-solar@agilent.com MD1010 Network B2B Agenda Overview What Measurements do we make? Network Analyzer Hardware Error Models and Calibration Example Measurements

More information

EIA STANDARD TP-66A. EMI Shielding Effectiveness Test Procedure for Electrical Connectors EIA A EIA A

EIA STANDARD TP-66A. EMI Shielding Effectiveness Test Procedure for Electrical Connectors EIA A EIA A EIA STANDARD ANSI/-2000(R2007) Approved: May 5, 2000 Reaffirmed: March 1, 2007 TP-66A EMI Shielding Effectiveness Test Procedure for Electrical Connectors (Revision of EIA-364-66) MAY 2000 ELECTRONIC COMPONENTS,

More information

INTERNATIONAL STANDARD

INTERNATIONAL STANDARD INTERNATIONAL STANDARD IEC 61726 Second edition 1999-11 Cable assemblies, cables, connectors and passive microwave components Screening attenuation measurement by the reverberation chamber method Câbles,

More information

EMC/EMI MEASURING INSTRUMENTS & ACCESSORIES SHORT-FORM CATALOG 2011

EMC/EMI MEASURING INSTRUMENTS & ACCESSORIES SHORT-FORM CATALOG 2011 EMC/EMI MEASURING INSTRUMENTS & ACCESSORIES SHORT-FORM CATALOG 2011 All-in-one Digital EMI Analyzer 10 Hz - 3 GHz PMM 9010/30P EMI Analyzer 10 Hz - 3 GHz Our trek started in a small laboratory over 25

More information

Semi Anechoic Chamber (SAC)

Semi Anechoic Chamber (SAC) 1 of 9 Semi Anechoic Chamber (SAC) Approximate Dimensions of 3m Semi Anechoic Chamber (SAC) Length: 10m Width: 9m Height: 9m Frequency range of Semi Anechoic Chamber: 9 KHz to 40 GHz Emission test (EMI):

More information

Signal and Noise Measurement Techniques Using Magnetic Field Probes

Signal and Noise Measurement Techniques Using Magnetic Field Probes Signal and Noise Measurement Techniques Using Magnetic Field Probes Abstract: Magnetic loops have long been used by EMC personnel to sniff out sources of emissions in circuits and equipment. Additional

More information

Bird Model 7022 Statistical Power Sensor Applications and Benefits

Bird Model 7022 Statistical Power Sensor Applications and Benefits Applications and Benefits Multi-function RF power meters have been completely transformed since they first appeared in the early 1990 s. What once were benchtop instruments that incorporated power sensing

More information

Chalmers Publication Library

Chalmers Publication Library Chalmers Publication Library About Random LOS in Rician Fading Channels for MIMO OTA Tests This document has been downloaded from Chalmers Publication Library (CPL). It is the author s version of a work

More information

Large E Field Generators in Semi-anechoic Chambers for Full Vehicle Immunity Testing

Large E Field Generators in Semi-anechoic Chambers for Full Vehicle Immunity Testing Large E Field Generators in Semi-anechoic Chambers for Full Vehicle Immunity Testing Vince Rodriguez ETS-Lindgren, Inc. Abstract Several standards recommend the use of transmission line systems (TLS) as

More information

Report Of. Shielding Effectiveness Test For. DefenderShield. Test Date(s): September 1 October 2, 2012

Report Of. Shielding Effectiveness Test For. DefenderShield. Test Date(s): September 1 October 2, 2012 Report Of Test For Test Date(s): September 1 October 2, 2012 UST Project No: Total Number of Pages Contained Within This Report: 15 3505 Francis Circle Alpharetta, GA 30004 PH: 770-740-0717 Fax: 770-740-1508

More information

Future In Radiated Immunity Testing

Future In Radiated Immunity Testing Future In Radiated Immunity Testing Flynn Lawrence Flynn Lawrence is an Applications Engineer for AR RF/Microwave Instrumentation. At AR, Flynn is actively engaged in new application and product development

More information

Harmonizing the ANSI-C12.1(2008) EMC Tests. Harmonizing the ANSI-C12.1(2008) EMC Tests

Harmonizing the ANSI-C12.1(2008) EMC Tests. Harmonizing the ANSI-C12.1(2008) EMC Tests Harmonizing the ANSI-C12.1(2008) EMC Tests Subcommittee 1 (Emissions) Subcommittee 5 (Immunity) Joint Task Force on C12.1 June 17, 2013 1 The Accredited Standards Committee C63 presents Harmonizing the

More information

Application Note #60 Harmonic Measurement for IEC And other Radiated Immunity Standards

Application Note #60 Harmonic Measurement for IEC And other Radiated Immunity Standards Application Note #60 Harmonic Measurement for IEC 61000-4-3 And other Radiated Immunity Standards By: Applications Engineering In the rush to complete RF immunity testing on schedule, it is not all that

More information

NSA Calculation of Anechoic Chamber Using Method of Moment

NSA Calculation of Anechoic Chamber Using Method of Moment 200 Progress In Electromagnetics Research Symposium 2006, Cambridge, USA, March 26-29 NSA Calculation of Anechoic Chamber Using Method of Moment T. Sasaki, Y. Watanabe, and M. Tokuda Musashi Institute

More information

Exercise 1-3. Radar Antennas EXERCISE OBJECTIVE DISCUSSION OUTLINE DISCUSSION OF FUNDAMENTALS. Antenna types

Exercise 1-3. Radar Antennas EXERCISE OBJECTIVE DISCUSSION OUTLINE DISCUSSION OF FUNDAMENTALS. Antenna types Exercise 1-3 Radar Antennas EXERCISE OBJECTIVE When you have completed this exercise, you will be familiar with the role of the antenna in a radar system. You will also be familiar with the intrinsic characteristics

More information

Shielding Effectiveness Report HQCD

Shielding Effectiveness Report HQCD HQCD Mates with QSH, QTH, QSH-EM Description: 0.50mm Q Strip High Speed Coax Cable Assembly Samtec, Inc. 2005 All Rights Reserved Table of Contents Product Overview... 1 Test Overview... 1 Shielded Room

More information

Fully Anechoic Room Validation Measurements to CENELEC pren

Fully Anechoic Room Validation Measurements to CENELEC pren Fully Anechoic Room Validation Measurements to CENELEC pren517-3 M.A.K.Wiles*,W.Muellner** *ETS,Rochester,UK **Austrian Research Center,Seibersdorf,Austria Abstract Many small to medium sized EMC anechoic

More information

SHIELDING EFFECTIVENESS MEASUREMENTS ON ENCLOSURES WITH VARIOUS APERTURES BY BOTH MODE-TUNED REVERBERATION CHAMBER AND GTEM CELL METHODOLOGIES

SHIELDING EFFECTIVENESS MEASUREMENTS ON ENCLOSURES WITH VARIOUS APERTURES BY BOTH MODE-TUNED REVERBERATION CHAMBER AND GTEM CELL METHODOLOGIES Progress In Electromagnetics Research B, Vol. 2, 103 114, 2008 SHIELDING EFFECTIVENESS MEASUREMENTS ON ENCLOSURES WITH VARIOUS APERTURES BY BOTH MODE-TUNED REVERBERATION CHAMBER AND GTEM CELL METHODOLOGIES

More information

EN :2007+A1:2011 Electromagnetic compatibility Emission standard for residential, commercial and light-industrial environments

EN :2007+A1:2011 Electromagnetic compatibility Emission standard for residential, commercial and light-industrial environments EMC Page 3 / 33 Test report No.: EN 61000-6-3:2007+A1:2011 Electromagnetic compatibility Emission standard for residential, commercial and light-industrial environments Date of measurement: 2013-10-16

More information

Saturation of Active Loop Antennas

Saturation of Active Loop Antennas Saturation of Active Loop Antennas Alexander Kriz EMC and Optics Seibersdorf Laboratories 2444 Seibersdorf, Austria Abstract The EMC community is working towards shorter test distances for radiated emission

More information

RECOMMENDATION ITU-R SM Method for measurements of radio noise

RECOMMENDATION ITU-R SM Method for measurements of radio noise Rec. ITU-R SM.1753 1 RECOMMENDATION ITU-R SM.1753 Method for measurements of radio noise (Question ITU-R 1/45) (2006) Scope For radio noise measurements there is a need to have a uniform, frequency-independent

More information

Some Consideration on Shielding Effectiveness Testing by Means of the Nested Reverberation Chambers

Some Consideration on Shielding Effectiveness Testing by Means of the Nested Reverberation Chambers 766 W.J. KRZYSZTOFIK, R. BOROWIEC, B. BIEDA, SOME CONSIDERATION ON SHIELDING EFFECTIVENESS TESTING Some Consideration on Shielding Effectiveness Testing by Means of the Nested Reverberation Chambers Wojciech

More information

Overview of EMC Regulations and Testing. Prof. Tzong-Lin Wu Department of Electrical Engineering National Taiwan University

Overview of EMC Regulations and Testing. Prof. Tzong-Lin Wu Department of Electrical Engineering National Taiwan University Overview of EMC Regulations and Testing Prof. Tzong-Lin Wu Department of Electrical Engineering National Taiwan University What is EMC Electro-Magnetic Compatibility ( 電磁相容 ) EMC EMI (Interference) Conducted

More information

Measurement of RF Emissions from a Caterpillar Inc. MSS3s RF ID Key Fob

Measurement of RF Emissions from a Caterpillar Inc. MSS3s RF ID Key Fob Measurement of RF Emissions from a Caterpillar Inc. MSS3s RF ID Key Fob For Caterpillar Inc. 330 S.W. Adams Street Peoria, IL 61630 P.O. Number JBL 11260 Date Tested May 11, 2016 Test Personnel Mark Longinotti

More information

Measurement Environment Influence Compensation to Reproduce Anechoic Chamber Measurements with Near Field Scanning

Measurement Environment Influence Compensation to Reproduce Anechoic Chamber Measurements with Near Field Scanning Measurement Environment Influence Compensation to Reproduce Anechoic Chamber Measurements with Near Field Scanning Denis Rinas, Alexander Zeichner, Stephan Frei TU Dortmund University Dortmund, Germany

More information

required. The inside enclosure is then covered with a radiowave absorber to reduce reflections. Such an

required. The inside enclosure is then covered with a radiowave absorber to reduce reflections. Such an APPENDIX II PERFORNMNOE EVALUATION OF A MICROWAVE ANECHOIC CHAMBER.'.v Antenna measurements, Ii scattering experiments etc. have to be conducted in an environment free from radio signal interference. Generally

More information

GTEM cells. Emissions and immunity testing in a single, shielded environment

GTEM cells. Emissions and immunity testing in a single, shielded environment GTEM cells Emissions and immunity testing in a single, shielded environment GTEM cells Emissions and immunity testing in a single, shielded environment Function A GTEM (Gigahertz Transverse Electro Magnetic)

More information

Ave output power ANT 1(dBm) Ave output power ANT 2 (dbm)

Ave output power ANT 1(dBm) Ave output power ANT 2 (dbm) Page 41 of 103 9.6. Test Result The test was performed with 802.11b Channel Frequency (MHz) power ANT 1(dBm) power ANT 2 (dbm) power ANT 1(mW) power ANT 2 (mw) Limits dbm / W Low 2412 7.20 7.37 5.248 5.458

More information

Performance Evaluations and Comparative Electromagnetic Compatibility Measurements on Compact Fluorescent Lamps

Performance Evaluations and Comparative Electromagnetic Compatibility Measurements on Compact Fluorescent Lamps 10 th International Conference on DEVELOPMENT AND APPLICATION SYSTEMS, Suceava, Romania, May 27-29, 2010 Performance Evaluations and Comparative Electromagnetic Compatibility Measurements on Compact Fluorescent

More information

SCOPE OF ACCREDITATION TO ISO/IEC 17025:2005 & ANSI/NCSL Z

SCOPE OF ACCREDITATION TO ISO/IEC 17025:2005 & ANSI/NCSL Z SCOPE OF ACCREDITATION TO ISO/IEC 17025:2005 & ANSI/NCSL Z540-1-1994 ETS-LINDGREN INC. 1301 Arrow Point Drive Cedar Park, TX 78613 Ron Bethel Phone: 512 531 6400 CALIBRATION Valid To: December 31, 2018

More information

TEST REPORT FROM RADIO FREQUENCY INVESTIGATION LTD.

TEST REPORT FROM RADIO FREQUENCY INVESTIGATION LTD. TEST REPORT FROM RADIO FREQUENCY INVESTIGATION LTD. Test Of: Wood & Douglas Ltd ST500 Transmitter Test Report Serial No: RFI/EMCB2/RP39403B This Test Report supersedes RFI Test Report No.: RFI/EMCB1/RP39403B

More information

You will need the following pieces of equipment to complete this experiment: Wilkinson power divider (3-port board with oval-shaped trace on it)

You will need the following pieces of equipment to complete this experiment: Wilkinson power divider (3-port board with oval-shaped trace on it) UNIVERSITY OF TORONTO FACULTY OF APPLIED SCIENCE AND ENGINEERING The Edward S. Rogers Sr. Department of Electrical and Computer Engineering ECE422H1S: RADIO AND MICROWAVE WIRELESS SYSTEMS EXPERIMENT 1:

More information

EFFECT OF SHIELDING ON CABLE RF INGRESS MEASUREMENTS LARRY COHEN

EFFECT OF SHIELDING ON CABLE RF INGRESS MEASUREMENTS LARRY COHEN EFFECT OF SHIELDING ON CABLE RF INGRESS MEASUREMENTS LARRY COHEN OVERVIEW Purpose: Examine the common-mode and differential RF ingress levels of 4-pair UTP, F/UTP, and F/FTP cables at an (RJ45) MDI port

More information

MEASUREMENTS OF COUPLING THROUGH BRAIDED SHIELD VIA NEW CONDUCTED IMMUNITY TECH- NIQUE

MEASUREMENTS OF COUPLING THROUGH BRAIDED SHIELD VIA NEW CONDUCTED IMMUNITY TECH- NIQUE Progress In Electromagnetics Research C, Vol. 11, 61 68, 2009 MEASUREMENTS OF COUPLING THROUGH BRAIDED SHIELD VIA NEW CONDUCTED IMMUNITY TECH- NIQUE M. Ghassempouri College of Electrical Engineering Iran

More information

Data Sheet. Peak, CW & Average. Power Sensors. Taking performance to a new peak

Data Sheet. Peak, CW & Average. Power Sensors. Taking performance to a new peak Data Sheet Peak, CW & Average Power Sensors Taking performance to a new peak Peak, CW & Average Power Sensors The overall performance of a power meter dependents on the power sensor employed. Boonton has

More information

Double-Ridged Waveguide Horn

Double-Ridged Waveguide Horn Model 3106 200 MHz 2 GHz Uniform Gain Power Handling up to 1.6 kw Model 3115 1 GHz 18 GHz Low VSWR Model 3116 18 GHz 40 GHz Quality Construction M O D E L 3 1 0 6 Double-Ridged Waveguide Horn PROVIDING

More information

Broadband Circular Polarized Antenna Loaded with AMC Structure

Broadband Circular Polarized Antenna Loaded with AMC Structure Progress In Electromagnetics Research Letters, Vol. 76, 113 119, 2018 Broadband Circular Polarized Antenna Loaded with AMC Structure Yi Ren, Xiaofei Guo *,andchaoyili Abstract In this paper, a novel broadband

More information

Reverberation Chambers Design and Construction Considerations for Aerospace and Military Test Requirements

Reverberation Chambers Design and Construction Considerations for Aerospace and Military Test Requirements Reverberation Chambers Design and Construction Considerations for Aerospace and Military Test Requirements REVERBERATION CHAMBERS: DESIGN AND CONSTRUCTION CONSIDERATIONS FOR AEROSPACE AND MILITARY TEST

More information

EMC Antenna Parameters and Their Relationships

EMC Antenna Parameters and Their Relationships EMC Antenna Parameters and Their Relationships Author : John D. M. Osburn, EMC Test Systems 04/05/2012 *originally published in June 1997. Introduction The basics of the EMC profession often get buried

More information

Vector Network Analyzer Application note

Vector Network Analyzer Application note Vector Network Analyzer Application note Version 1.0 Vector Network Analyzer Introduction A vector network analyzer is used to measure the performance of circuits or networks such as amplifiers, filters,

More information

OPEN TEM CELLS FOR EMC PRE-COMPLIANCE TESTING

OPEN TEM CELLS FOR EMC PRE-COMPLIANCE TESTING 1 Introduction Radiated emission tests are typically carried out in anechoic chambers, using antennas to pick up the radiated signals. Due to bandwidth limitations, several antennas are required to cover

More information

A New TEM Horn Antenna Designing Based on Plexiglass Antenna Cap

A New TEM Horn Antenna Designing Based on Plexiglass Antenna Cap Journal of Applied Science and Engineering, Vol. 21, No. 3, pp. 413 418 (2018) DOI: 10.6180/jase.201809_21(3).0012 A New TEM Horn Antenna Designing Based on Plexiglass Antenna Cap Lin Teng and Jie Liu*

More information

On the Design of Software and Hardware for a WSN Transmitter

On the Design of Software and Hardware for a WSN Transmitter 16th Annual Symposium of the IEEE/CVT, Nov. 19, 2009, Louvain-La-Neuve, Belgium 1 On the Design of Software and Hardware for a WSN Transmitter Jo Verhaevert, Frank Vanheel and Patrick Van Torre University

More information

SITE-TO-SITE REPRODUCIBILITY IN CONDUCTED IMMUNITY TESTS ON PC-BASED DATA ACQUISITION SYSTEMS

SITE-TO-SITE REPRODUCIBILITY IN CONDUCTED IMMUNITY TESTS ON PC-BASED DATA ACQUISITION SYSTEMS SITE-TO-SITE REPRODUCIBILITY IN CONDUCTED IMMUNITY TESTS ON PC-BASED DATA ACQUISITION SYSTEMS G.Betta 1, D.Capriglione 1, C.Spataro 2, G.Tinè 3 1 DAEIMI University of Cassino, Via G.Di Biasio 43, 03043

More information

KULLIYYAH OF ENGINEERING

KULLIYYAH OF ENGINEERING KULLIYYAH OF ENGINEERING DEPARTMENT OF ELECTRICAL & COMPUTER ENGINEERING ANTENNA AND WAVE PROPAGATION LABORATORY (ECE 4103) EXPERIMENT NO 3 RADIATION PATTERN AND GAIN CHARACTERISTICS OF THE DISH (PARABOLIC)

More information

Residual Phase Noise Measurement Extracts DUT Noise from External Noise Sources By David Brandon and John Cavey

Residual Phase Noise Measurement Extracts DUT Noise from External Noise Sources By David Brandon and John Cavey Residual Phase Noise easurement xtracts DUT Noise from xternal Noise Sources By David Brandon [david.brandon@analog.com and John Cavey [john.cavey@analog.com Residual phase noise measurement cancels the

More information

Performance Analysis of Different Ultra Wideband Planar Monopole Antennas as EMI sensors

Performance Analysis of Different Ultra Wideband Planar Monopole Antennas as EMI sensors International Journal of Electronics and Communication Engineering. ISSN 09742166 Volume 5, Number 4 (2012), pp. 435445 International Research Publication House http://www.irphouse.com Performance Analysis

More information

Investigation of wave diffusers to scatter the unstirred radiation of an antenna in a reverberation chamber

Investigation of wave diffusers to scatter the unstirred radiation of an antenna in a reverberation chamber Investigation of wave diffusers to scatter the unstirred radiation of an antenna in a reverberation chamber By Evangelia Aikaterini Karadimou Master by Research University of York, UK Department of Electronics

More information

A 2 to 4 GHz Instantaneous Frequency Measurement System Using Multiple Band-Pass Filters

A 2 to 4 GHz Instantaneous Frequency Measurement System Using Multiple Band-Pass Filters Progress In Electromagnetics Research M, Vol. 62, 189 198, 2017 A 2 to 4 GHz Instantaneous Frequency Measurement System Using Multiple Band-Pass Filters Hossam Badran * andmohammaddeeb Abstract In this

More information

MANUAL. PCD - Precision Conical Dipole Antenna

MANUAL. PCD - Precision Conical Dipole Antenna MANUAL PCD - Precision Conical Dipole Antenna RF Engineering MANUAL Precision Conical Dipole Antenna 12.10.2009 Version 2.0 Notice Seibersdorf Labor GmbH reserves the right to make changes to any product

More information

RADIOMETRICS Midwest Corporation

RADIOMETRICS Midwest Corporation RADIOMETRICS Midwest Corporation Shielding Effectiveness Test Report Tests Performed on an IMS-AMCO Shielded Rack Test Unit #2 Part Number S40469 Radiometrics Document RP-5760B Test Specifications MIL-STD-285

More information

Keywords: ISM, RF, transmitter, short-range, RFIC, switching power amplifier, ETSI

Keywords: ISM, RF, transmitter, short-range, RFIC, switching power amplifier, ETSI Maxim > Design Support > Technical Documents > Application Notes > Wireless and RF > APP 4929 Keywords: ISM, RF, transmitter, short-range, RFIC, switching power amplifier, ETSI APPLICATION NOTE 4929 Adapting

More information