1 DICOM Conformance Statement SagiPlan 2.0 Concerned Product: SagiPlan Version 2.0 Document Date: November 26, 2015 Document Version: 01
2 Overview SagiPlan is a software product to perform the dosimetry for the treatment of cancerous tissue with temporarily applied radioactive radiation sources, i.e. high dose rate (HDR) brachytherapy. SagiPlan uses DICOM services to import images and structures and to export structures, treatment plans and dose information. Furthermore SagiPlan allows to Query/Retrieve other DICOM compliant applications for specific DICOM Objects. SagiPlan imports the images and structures from DICOM Media Files, which may be located on local disk or on exchangeable media such as CD or DVD. A TCP/IP Network Handler (called SagiPlanDS) can be set up to handle the import of images and structures from remote applications to the SagiPlan Application via network. This TCP/IP Handler acts as a receiver for DICOM Objects from network and stores these incoming files as DICOM Media Files on local disk. Once saved, these Files may be accessed by the Sagi- Plan application. SagiPlan exports the structures, treatment plans and dose information to DICOM Media Files or transfers these information objects to other systems via network by launching the Store SCU service. 2/179
3 1 Table Of Contents OVERVIEW TABLE OF CONTENTS INTRODUCTION REVISION HISTORY AUDIENCE REMARKS LEGAL AND COPYRIGHT INFORMATION TERMS AND DEFINITIONS BASICS OF DICOM COMMUNICATION ABBREVIATIONS REFERENCES NETWORKING IMPLEMENTATION MODEL Application Data Flow Diagram Functional Definition of AE s Functional Definition of SagiPlanDS Functional Definition of SagiPlan Sequencing of Real World Activities AE SPECIFICATIONS SagiPlanDS Application Entity SOP Classes Association Policies General Number of Associations Asynchronous Nature Implementation Identifying Information Association Initiation Policy Association Acceptance Policy Real World Activity Remote Storage SCU SagiPlan Application Entity SOP Classes Association Policies General Number of Associations Asynchronous Nature Implementation Identifying Information Association Initiation Policy Activity Send Files (Store SCU) / Export RT files Activity Query/Retrieve Association Acceptance Policy NETWORK INTERFACES Physical Network Interface Supported communication stacks TCP/IP stack Physical media support CONFIGURATION AE Title/Presentation Address Mapping /179
4 Local AE Titles Remote AE Title/Presentation Address Mapping SagiPlanDS SagiPlan Storage SCU SagiPlan Query/Retrieve SCU Find Parameters MEDIA INTERCHANGE IMPLEMENTATION MODEL Application Data Flow Diagram Functional definitions of AE s Sequencing of Real World Activities File Meta Information for Implementation Class and Version AE SPECIFICATIONS Application Entity SagiPlan File Meta Information for the SagiPlan Application Entity Real-World Activities Real-World Activity Export DICOM RT Files Real-World Activity Import DIOCOM RT Struct Real-World Activity Import DICOM Images AUGMENTED AND PRIVATE APPLICATION PROFILES MEDIA CONFIGURATION SUPPORT OF CHARACTER SETS SECURITY SECURITY PROFILES ASSOCIATION LEVEL SECURITY APPLICATION LEVEL SECURITY ANNEXES IOD CONTENTS Created SOP Instances RT Plan IOD RT Structure Set IOD RT Dose IOD Patient Module General Study Module Frame Of Reference Module General Equipment Module General Image Module Image Plane Module Image Pixel Module Multi Frame Module RT Series Module RT Dose Module Dose Summation Type Referenced Fraction Group Sequence RT DVH Module Structure Set Module Structure Set Module for the RT Structure Set IOD: Structure Set Module for the RT Dose IOD: ROI Contour Module Structure Set Module for the RT Structure Set IOD: Structure Set Module for the Dose IOD: RT Dose ROI Module Dose Value RT ROI Observations Module /179
5 RT General Plan Module RT Fraction Scheme Module RT Brachy Application Setups Module SOP Common Module Approval Module RT Prescription Module Usage of Attributes from received IOD s Computed Tomography Image IOD Magnetic Resonance Image IOD Positron Emission Tomography Image IOD RT Image IOD Secondary Capture Image IOD Ultrasound Image IOD Computed Radiography Image IOD X-Ray RF Image IOD X-Ray Angiographic Image IOD Digital X-Ray Image IOD Patient Module General Study Module General Series Module Frame Of Reference Module General Image Module Image Plane Module Image Pixel Module CT Image Module SOP Common Module RT Series Module RT Image Module SC Image Module CR Image Module X-Ray Image Module X-Ray Acquisition Module DX Detector Module DX Image Module DATA DICTIONARY OF PRIVATE ATTRIBUTES /179
6 List Of Tables Table 3-1 SOP CLASSes for SagiPlanDS Table 3-2 Number of Associations as an association acceptor for SagiPlanDS Table 3-3 DICOM Implementation Class and Version for SagiPlanDS Table 3-4 SagiPlanDS accepted Presentation Contexts Table 3-5 Storage C-STORE Response Status Table 3-6 SOP Classes for SagiPlan Storage SCU Table 3-7 SOP Classes for SagiPlan Query/Retrieve SCU Table 3-8 Number of Associations as an association initiator for SagiPlan Table 3-9 DICOM Implementation Class and Version for SagiPlan Table 3-10 Proposed Presentation Contexts for SagiPlan AE as Store SCU Table 3-11 SagiPlan AE as Store SCU Response Status Handling Table 3-12 Proposed Presentation Contexts for SagiPlan AE as Query/Retrieve SCU Table 3-13 SagiPlan AE as Query/Retrieve SCU Response Status Handling for C-FIND Requests Table 3-14 SagiPlan AE as Query/Retrieve SCU Response Status Handling For C-MOVE Requests Table 3-15 AE Title Configuration Table Table 3-16 Configuration Parameters Table for SagiPlanDS Application Entity Table 4-1 File Meta Information for Implementation Class and Version Table 4-2 AE Related Application Profiles, Real-World Activities, and Roles Table 4-3 Supported SOP Classes and Transfer Syntaxes for Export DICOM RT Files 33 Table 4-4 Supported SOP Classes and Transfer Syntaxes for Import DICOM RT Struct Table 4-5 Supported SOP Classes and Transfer Syntaxes for Import DIOCOM Images 34 Table 7-1 RT PLAN IOD MODULES Table 7-2 RT STRUCTURE SET IOD MODULES Table 7-3 RT DOSE IOD MODULES Table 7-4 PATIENT MODULE ATTRIBUTES Table 7-5 GENERAL STUDY MODULE ATTRIBUTES Table 7-6 FRAME OF REFERENCE MODULE ATTRIBUTES Table 7-7 GENERAL EQUIPMENT MODULE ATTRIBUTES Table 7-8 GENERAL IMAGE MODULE ATTRIBUTES Table 7-9 IMAGE PLANE MODULE ATTRIBUTES Table 7-10 IMAGE PIXEL MODULE ATTRIBUTES Table 7-11 MULTI-FRAME MODULE ATTRIBUTES Table 7-12 RT SERIES MODULE ATTRIBUTES Table 7-13 RT DOSE MODULE ATTRIBUTES Table 7-14 Dose Summation Type Table 7-15 Referenced Fraction Group Sequence Table 7-16 RT DVH MODULE ATTRIBUTES Table 7-17 STRUCTURE SET MODULE ATTRIBUTES Table 7-18 STRUCTURE SET MODULE ATTRIBUTES Table 7-19 ROI CONTOUR MODULE ATTRIBUTES Table 7-20 ROI CONTOUR MODULE ATTRIBUTES Table 7-21 RT DOSE ROI MODULE ATTRIBUTES Table 7-22 RT ROI OBSERVATIONS MODULE ATTRIBUTES Table 7-23 RT GENERAL PLAN MODULE ATTRIBUTES Table 7-24 RT FRACTION SCHEME MODULE ATTRIBUTES Table 7-25 RT BRACHY APPLICATION SETUPS MODULE ATTRIBUTES Table 7-26 SOP COMMON MODULE ATTRIBUTES /179
7 Table 7-27 Approval MODULE ATTRIBUTES Table 7-28 Approval MODULE ATTRIBUTES Table 7-29 CT IMAGE IOD MODULES Table 7-30 MR IMAGE IOD MODULES Table 7-31 PET IMAGE IOD MODULES Table 7-32 RT IMAGE IOD MODULES Table 7-33 SC IMAGE IOD MODULES Table 7-34 US IMAGE IOD MODULES Table 7-35 CR IMAGE IOD MODULES Table 7-36 XRF IMAGE IOD MODULES Table 7-37 X-RAY ANGIOGRAPHIC IMAGE IOD MODULES Table 7-38 DIGITAL X-RAY IMAGE IOD MODULES Table 7-39 PATIENT MODULE ATTRIBUTES Table 7-40 GENERAL STUDY MODULE ATTRIBUTES Table 7-41 GENERAL SERIES MODULE ATTRIBUTES Table 7-42 FRAME OF REFERENCE MODULE ATTRIBUTES Table 7-43 GENERAL IMAGE MODULE ATTRIBUTES Table 7-44 IMAGE PLANE MODULE ATTRIBUTES Table 7-45 IMAGE PIXEL MODULE ATTRIBUTES Table 7-46 CT IMAGE MODULE ATTRIBUTES Table 7-47 SOP COMMON MODULE ATTRIBUTES Table 7-48 Table C RT SERIES MODULE ATTRIBUTES Table 7-49 Table C.8-38 RT IMAGE MODULE ATTRIBUTES Table 7-50 SC IMAGE MODULE ATTRIBUTES Table 7-51 BASIC PIXEL SPACING CALIBRATION MACRO ATTRIBUTES Table 7-52 CR IMAGE MODULE ATTRIBUTES Table 7-53 X-RAY IMAGE MODULE ATTRIBUTES Table 7-54 X-RAY ACQUISITION MODULE ATTRIBUTES Table 7-55 DX DETECTOR MODULE ATTRIBUTES Table 7-56 DX IMAGE MODULE ATTRIBUTES Table 7-57 DATA DICTIONARY OF PRIVATE ATTRIBUTES /179
8 List of Figures Figure 3-1 Application Data Flow Diagram Figure 4-1 APPLICATION DATA FLOW DIAGRAM /179
9 2 Introduction 2.1 Revision History SagiPlan Version 2.0: Rev 01 September 29, Audience This document is written for users that need to understand how SagiPlan will integrate into their healthcare facility. This includes both those responsible for overall imaging network policy and architecture, as well as integrators who need to have a detailed understanding of the DICOM features of the product. This document contains some basic DICOM definitions so that any reader may understand how this product implements DICOM features. However, integrators are expected to fully understand all the DICOM terminology, how the tables in this document relate to the product s functionality, and how that functionality integrates with other devices that support compatible DICOM features. 2.3 Remarks The scope of this DICOM Conformance Statement is to facilitate integration between SagiPlan and other DICOM products. The Conformance Statement should be read and understood in conjunction with the DICOM Standard. DICOM by itself does not guarantee interoperability. The Conformance Statement does, however, facilitate a first-level comparison for interoperability between different applications supporting compatible DICOM functionality. This Conformance Statement is not supposed to replace validation with other DICOM equipment to ensure proper exchange of intended information. In fact, the user should be aware of the following important issues: The comparison of different Conformance Statements is just the first step towards assessing interconnectivity and interoperability between the product and other DICOM conformant equipment. Test procedures should be defined and executed to validate the required level of interoperability with specific compatible DICOM equipment, as established by the healthcare facility. It is the responsibility of the user (or the user s agent) to verify DICOM connectivity within the environment in which the SagiPlan application shall be used. It is the responsibility of the user to specify the appropriate test suite and to carry out the additional validation tests. Another device matching this Conformance Statement based on the comparison with its own Conformance Statement, will have a chance, but no guarantee that they will interoperate. DICOM only deals with communication, it is not a standard which specifies what is needed for certain applications to run on a device. 9/179
10 2.4 Legal and Copyright Information This document and the information therein are proprietary to Eckert & Ziegler BEBIG GmbH, Robert-Rössle-Str.10, Berlin, Germany. The document and the information therein shall not be reproduced in whole or in part without the express written permission of Eckert & Ziegler BEBIG GmbH. The information in this document may be altered or modified at any time without prior notice to the user. Eckert & Ziegler BEBIG GmbH does not assume any responsibility for any kind of errors in this manual or for any kind of damages caused by such errors. 2.5 Terms and Definitions Informal definitions are provided for the following terms used in this Conformance Statement. The DICOM Standard is the authoritative source for formal definitions of these terms. Abstract Syntax the information agreed to be exchanged between applications, generally equivalent to a Service/Object Pair (SOP) Class. Examples : Verification SOP Class, Modality Worklist Information Model Find SOP Class, Computed Radiography Image Storage SOP Class. Application Entity (AE) an end point of a DICOM information exchange, including the DICOM network or media interface software; i.e., the software that sends or receives DICOM information objects or messages. A single device may have multiple Application Entities. Application Entity Title the externally known name of an Application Entity, used to identify a DICOM application to other DICOM applications on the network. Application Context the specification of the type of communication used between Application Entities. Example: DICOM network protocol. Association a network communication channel set up between Application Entities. Attribute a unit of information in an object definition; a data element identified by a tag. The information may be a complex data structure (Sequence), itself composed of lower level data elements. Examples: Patient ID (0010,0020), Accession Number (0008,0050), Photometric Interpretation (0028,0004), Procedure Code Sequence (0008,1032). Information Object Definition (IOD) the specified set of Attributes that comprise a type of data object; does not represent a specific instance of the data object, but rather a class of similar data objects that have the same properties. The Attributes may be specified as Mandatory (Type 1), Required but possibly unknown (Type 2), or Optional (Type 3), and there may be conditions associated with the use of an Attribute (Types 1C and 2C). Examples: MR Image IOD, CT Image IOD, Print Job IOD. Media Application Profile the specification of DICOM information objects and encoding exchanged on removable media (e.g., CDs) 10/179
11 Module a set of Attributes within an Information Object Definition that are logically related to each other. Example: Patient Module includes Patient Name, Patient ID, Patient Birth Date, and Patient Sex. Negotiation first phase of Association establishment that allows Application Entities to agree on the types of data to be exchanged and how that data will be encoded. Presentation Context the set of DICOM network services used over an Association, as negotiated between Application Entities; includes Abstract Syntaxes and Transfer Syntaxes. Protocol Data Unit (PDU) a packet (piece) of a DICOM message sent across the network. Devices must specify the maximum size packet they can receive for DICOM messages. Security Profile a set of mechanisms, such as encryption, user authentication, or digital signatures, used by an Application Entity to ensure confidentiality, integrity, and/or availability of exchanged DICOM data Service Class Provider (SCP) role of an Application Entity that provides a DICOM network service; typically, a server that performs operations requested by another Application Entity (Service Class User). Examples: Picture Archiving and Communication System (image storage SCP, and image query/retrieve SCP), Radiology Information System (modality worklist SCP). Service Class User (SCU) role of an Application Entity that uses a DICOM network service; typically, a client. Examples: imaging modality (image storage SCU, and modality worklist SCU), imaging workstation (image query/retrieve SCU) Service/Object Pair (SOP) Class the specification of the network or media transfer (service) of a particular type of data (object); the fundamental unit of DICOM interoperability specification. Examples: Ultrasound Image Storage Service, Basic Grayscale Print Management. Service/Object Pair (SOP) Instance an information object; a specific occurrence of information exchanged in a SOP Class. Examples: a specific x-ray image. Tag a 32-bit identifier for a data element, represented as a pair of four digit hexadecimal numbers, the group and the element. If the group number is odd, the tag is for a private (manufacturer-specific) data element. Examples: (0010,0020) [Patient ID], (07FE,0010) [Pixel Data], (0019,0210) [private data element] Transfer Syntax the encoding used for exchange of DICOM information objects and messages. Examples: JPEG compressed (images), little endian explicit value representation. Unique Identifier (UID) a globally unique dotted decimal string that identifies a specific object or a class of objects; an ISO-8824 Object Identifier. Examples: Study Instance UID, SOP Class UID, SOP Instance UID. 11/179
12 Value Representation (VR) the format type of an individual DICOM data element, such as text, an integer, a person s name, or a code. DICOM information objects can be transmitted with either explicit identification of the type of each data element (Explicit VR), or without explicit identification (Implicit VR); with Implicit VR, the receiving application must use a DICOM data dictionary to look up the format of each data element. 2.6 Basics of DICOM Communication This section describes terminology used in this Conformance Statement for the nonspecialist. The key terms used in the Conformance Statement are highlighted in italics below. This section is not a substitute for training about DICOM, and it makes many simplifications about the meanings of DICOM terms. Two Application Entities (devices) that want to communicate with each other over a network using DICOM protocol must first agree on several things during an initial network handshake. One of the two devices must initiate an Association (a connection to the other device), and ask if specific services, information, and encoding can be supported by the other device (Negotiation). DICOM specifies a number of network services and types of information objects, each of which is called an Abstract Syntax for the Negotiation. DICOM also specifies a variety of methods for encoding data, denoted Transfer Syntaxes. The Negotiation allows the initiating Application Entity to propose combinations of Abstract Syntax and Transfer Syntax to be used on the Association; these combinations are called Presentation Contexts. The receiving Application Entity accepts the Presentation Contexts it supports. For each Presentation Context, the Association Negotiation also allows the devices to agree on Roles which one is the Service Class User (SCU - client) and which is the Service Class Provider (SCP - server). Normally the device initiating the connection is the SCU, i.e., the client system calls the server, but not always. The Association Negotiation finally enables exchange of maximum network packet (PDU) size, security information, and network service options (called Extended Negotiation information). The Application Entities, having negotiated the Association parameters, may now commence exchanging data. Common data exchanges include queries for worklists and lists of stored images, transfer of image objects and analyses (structured reports), and sending images to film printers. Each exchangeable unit of data is formatted by the sender in accordance with the appropriate Information Object Definition, and sent using the negotiated Transfer Syntax. There is a Default Transfer Syntax that all systems must accept, but it may not be the most efficient for some use cases. Each transfer is explicitly acknowledged by the receiver with a Response Status indicating success, failure, or that query or retrieve operations are still in process. Two Application Entities may also communicate with each other by exchanging media (such as a CD-R). Since there is no Association Negotiation possible, they both use a Media Application Profile that specifies pre-negotiated exchange media format, Abstract Syntax, and Transfer Syntax. 12/179
13 2.7 Abbreviations AE Application Entity AET Application Entity Title CD-R Compact Disk Recordable CR Computed Radiography CT Computed Tomography DICOM Digital Imaging and Communications in Medicine FSC File-Set Creator FSU File-Set Updater FSR File-Set Reader IOD Information Object Definition LUT Look-up Table MPEG Moving Picture Experts Group MR Magnetic Resonance Imaging NTP Network Time Protocol PET Positron Emission Tomography RT Radiotherapy SC Secondary Capture SCP Service Class Provider SCU Service Class User SOP Service-Object Pair TCP/IP Transmission Control Protocol/Internet Protocol U Unique (Key Attribute) US Ultrasound VL Visible Light 2.8 REFERENCES 1. NEMA The DICOM Standard Digital Imaging and Communications in Medicine (DICOM) Standard, available free at 2. SagiPlan-2.0.chm SagiPlan User s Guide 13/179
14 3 Networking 3.1 Implementation Model Application Data Flow Diagram The diagram below illustrates the Data Flow Diagram of the SagiPlan and SagiPlanDS Application Entities. 14/179
15 DICOM Media File (Temporary Object Storage on local disk for later import to the SagiPlan Application Entity) Application Entity SagiPlanDS (Storage SCP) Remote Storage SCU Real World Activity Query / Retrieve Application Entity SagiPlan Remote Query / Retrieve SCP Real World Activity Send Files and Export RT files Remote Storage SCP DICOM Media File (probably created by invoking the Export Functionality of the SagiPlan Application Entity) DICOM Standard Interface Figure 3-1 Application Data Flow Diagram 15/179
16 3.1.2 Functional Definition of AE s Functional Definition of SagiPlanDS The SagiPlanDS Application Entity implements a Service Class Provider (SCP) for the Storage Service Class. It listens on a specific TCP/IP port for incoming association requests from a Storage SCU and can receive images and other DICOM Information Objects according to the Storage Service Class. Received objects will be stored as DICOM Media Files on local disk, from which they can be imported later on to the SagiPlan Application. The SagiPlanDS Application Entity also supports the Verification Service Class as an SCP. The SagiPlanDS runs continuously as a service and may be setup via the SagiPlan application Functional Definition of SagiPlan The SagiPlan Application Entity implements the following functions: A Service Class User (SCU) for the Storage Service Class. In this role it initiates an association to a Remote AE as a Storage Service Class Provider to send DICOM Information Objects. If the remote AE accepts the association, SagiPlan will send the selected Information Object(s) by invoking a C-STORE request for each object on the same association. A Service Class User (SCU) for the Query / Retrieve Service Class. In this role it initiates an association to a Remote AE to Query for a list of DICOM objects. If the remote AE accepts the association, SagiPlan will send C-FIND requests to the remote AE to query for certain Patients, Studies or Series. Once the responses are received and the user has selected one or more of these Patients, Studies or Series, the SagiPlan AE will send C-MOVE requests to the remote AE to transfer the DICOM objects to a Store SCP. SagiPlan can be configured to choose the Sagi- PlanDS AE as the Store SCP destination of the C-MOVE requests Sequencing of Real World Activities Not applicable. 3.2 AE Specifications SagiPlanDS Application Entity SOP Classes The SagiPlanDS AE provides Standard Conformance to the following SOP Classes: 16/179
17 Table 3-1 SOP CLASSes for SagiPlanDS SOP Class Name SOP Class UID SCU SCP Verification SOP Class No Yes Stored Print Storage No Yes Hardcopy Grayscale Image Storage No Yes Hardcopy Color Image Storage No Yes Computed Radiography Image Storage No Yes Digital X-Ray Image Storage - For No Yes Presentation Digital X-Ray Image Storage - For Processing No Yes Digital Mammography X-Ray Image No Yes Storage For Presentation Digital Mammography X-Ray Image No Yes Storage For Processing Digital Intra Oral X-Ray Image Storage No Yes For Presentation Digital Intra Oral X-Ray Image Storage No Yes For Processing Standalone Modality LUT Storage No Yes Encapsulated PDF Storage No Yes 1 Standalone VOI LUT Storage No Yes Grayscale Softcopy Presentation State No Yes Storage Color Softcopy Presentation State Storage No Yes Pseudo Color Softcopy Presentation No Yes State Storage Blending Softcopy Presentation State No Yes Storage X-Ray Angiographic Image Storage No Yes Enhanced XA Image Storage No Yes X-Ray Fluoroscopy Image Storage No Yes Enhanced XRF Image Storage No Yes 1 PET Image Storage No Yes PET Curve Storage No Yes CT Image Storage No Yes Enhanced CT Image Storage No Yes 17/179
18 Nuclear Medicine Image Storage No Yes Ultrasound Multiframe Image Storage No Yes MR Image Storage No Yes Enhanced MR Image Storage No Yes MR Spectroscopy Storage No Yes RT Image Storage No Yes 1 RT Dose Storage No Yes 2 RT Structure Set Storage No Yes 3 RT Beams Treatment Record Storage No Yes 4 RT Plan Storage No Yes 5 RT Brachy Treatment Record Storage No Yes 6 RT Treatment Summary Record Storage No Yes 7 Ultrasound Image Storage No Yes Raw Data Storage No Yes Spatial Registration Storage No Yes Spatial Fiducials Storage No Yes Real World Value Mapping Storage No Yes Secondary Capture Image Storage No Yes Multiframe Single Bit Secondary Capture Image Storage No Yes Multiframe Grayscale Byte Secondary No Yes Capture Image Storage Multiframe Grayscale Word Secondary No Yes Capture Image Storage Multiframe True Color Secondary Capture No Yes Image Storage VL Endoscopic Image Storage No Yes 1 Video Endoscopic Image Storage No Yes 1.1 VL Microscopic Image Storage No Yes 2 Video Microscopic Image Storage No Yes VL Slide Coordinates Microscopic No Yes Image Storage 3 18/179
19 VL Photographic Image Storage No Yes 4 Video Photographic Image Storage No Yes Ophthalmic Photography 8 Bit Image No Yes Storage 5.1 Ophthalmic Photography 16 Bit Image No Yes Storage 5.2 Stereometric Relationship Storage No Yes 5.3 Standalone Overlay Storage No Yes Basic Text SR No Yes 1 Enhanced SR No Yes 2 Comprehensive SR No Yes 3 Procedure Log Storage No Yes 0 Mammography CAD SR Storage No Yes 0 Key Object Selection Document No Yes 9 Chest CAD SR Storage No Yes 5 X-Ray Radiation Dose SR Storage No Yes 7 Standalone Curve Storage No Yes Twelve Lead ECG Waveform Storage No Yes General ECG Waveform Storage No Yes Ambulatory ECG Waveform Storage No Yes Hemodynamic Waveform Storage No Yes Cardiac Electrophysiology Waveform No Yes Storage Basic Voice Audio Waveform Storage No Yes Notice: Although the SagiPlanDS supports many SOP Classes, only some of them are supported by the SagiPlan application via the SagiPlan Application Entity for later import (see Application Entity ) Association Policies General The maximum PDU length of the AE is bytes. The AE does not support any SOP class extended negotiations. 19/179
20 Number of Associations Table 3-2 Number of Associations as an association acceptor for SagiPlanDS Maximum number of simultaneous associations SagiPlanDS will attempt only one association at a time Asynchronous Nature Asynchronous mode of operation is not supported Implementation Identifying Information Table 3-3 DICOM Implementation Class and Version for SagiPlanDS Implementation Class UID Implementation Version Name SagiPlanDS Association Initiation Policy The SagiPlanDS Application Entity never initiates associations Association Acceptance Policy When SagiPlanDS accepts an association, it will answer a C-ECHO request or receive any DICOM Information Objects transmitted on that association and store the Objects on disk. It places no limitations on how may connect to it Real World Activity Remote Storage SCU Description and Sequencing of Activities Not applicable Accepted Presentation Contexts 20/179
21 The SagiPlanDS Application Entity will accept presentation contexts for all of the above mentioned supported SOP Classes using any of the transfer syntaxes: Table 3-4 SagiPlanDS accepted Presentation Contexts Abstract Syntax Transfer Syntax Role Extended Name UID Name UID Negotiation See See Explicit VR SCP None note note Little Endian See See Explicit VR SCP None note note Big Endian See See Implicit VR SCP None note note Little Endian Note: This applies to all supported Storage Abstract Syntaxes, i.e. to all supported SOP Classes The SagiPlanDS prefers the transfer syntaxes in the following order: 1. Explicit VR Little Endian 2. Implicit VR Little Endian 3. Explicit VR Big Endian The SagiPlanDS application does not support extended negotiation SOP Specific Conformance for SOP Classes Service Status Failure Further Meaning Refused: Out of Resources Error: Data Set does not match SOP Class Error: Cannot understand Table 3-5 Storage C-STORE Response Status Error Code A700 A900 C000 Success Success 0000 All o.k. Reason Application out of memory, file system or database write error (e. g. file system full) SOP class or instance UID in C-STORE-RQ does not match UIDs in the received dataset Received dataset without SOP class or instance UID; received Presentation State that failed syntax check; internal application error 21/179
22 3.2.2 SagiPlan Application Entity SOP Classes The SagiPlan AE as a Storage SCU provides Standard Conformance to the following SOP Classes: Table 3-6 SOP Classes for SagiPlan Storage SCU SOP Class Name SOP Class UID Computed Radiography Image Storage Digital XRay Image Storage For Presentation Digita lxray Image Storage For Processing Digital Mammography XRay Image Storage For Presentation Digital Mammography XRay Image Storage For Processing Digital Intra Oral XRay Image Storage For Presentation Digital Intra Oral XRay Image Storage For Processing Encapsulated PDF Storage Grayscale Softcopy Presentation State Storage Color Softcopy Presentation State Storage Pseudo Color Softcopy Presentation State Storage Blending Softcopy Presentation State Storage XRay Angiographic Image Storage Enhanced XA Image Storage XRay Fluoroscopy Image Storage Enhanced XRF Image Storage PET Image Storage PET Curve Storage CT Image Storage Enhanced CT Image Storage Nuclear Medicine Image Storage Ultrasound Multiframe Image Storage MR Image Storage Enhanced MR Image Storage MR Spectroscopy Storage RT Image Storage RT Dose Storage RT Structure Set Storage RT Beams Treatment Record Storage RT Plan Storage RT Brachy Treatment Record Storage RT Treatment Summary Record Storage Ultrasound Image Storage Raw Data Storage Spatial Registration Storage /179
23 Spatial Fiducials Storage Real World Value Mapping Storage Secondary Capture Image Storage Multiframe Single Bit Secondary Capture Image Storage Multiframe Grayscale Byte Secondary Capture Image Storage Multiframe Grayscale Word Secondary Capture Image Storage Multiframe True Color Secondary Capture Image Storage VL Endoscopic Image Storage VL Microscopic Image Storage VL Slide Coordinates Microscopic Image Storage VL Photographic Image Storage Ophthalmic Photography 8 Bit Image Storage Ophthalmic Photography 16 Bit Image Storage Stereometric Relationship Storage Basic Text SR Enhanced SR Comprehensive SR Procedure Log Storage Mammography CAD SR Key Object Selection Document Chest CAD SR XRay Radiation Dose SR Twelve Lead ECG Waveform Storage General ECG Waveform Storage Ambulatory ECG Waveform Storage Hemodynamic Waveform Storage Cardiac Electrophysiology Waveform Storage Basic Voice Audio Waveform Storage The SagiPlan AE as a Query / Retrieve SCU provides Standard Conformance to the following SOP Classes: SOP Class Name Table 3-7 SOP Classes for SagiPlan Query/Retrieve SCU SOP Class UID Study Root Query/Retrieve Information Model FIND Patient Root Query/Retrieve Information Model MOVE Study Root Query/Retrieve Information Model MOVE Association Policies General The maximum PDU length of the AE is bytes. The AE does not support any SOP class extended negotiations. 23/179
24 Number of Associations Table 3-8 Number of Associations as an association initiator for SagiPlan Maximum number of simultaneous associations SagiPlan will attempt only one association at a time Asynchronous Nature Asynchronous mode of operation is not supported Implementation Identifying Information Table 3-9 DICOM Implementation Class and Version for SagiPlan Implementation Class UID Implementation Version Name SagiPlan Association Initiation Policy Activity Send Files (Store SCU) / Export RT files Description and Sequencing of Activities The user can select one or more DICOM information objects to be transferred via the Store SCU service to another DICOM AE. The information objects might be DICOM Media files stored on hard disc or DICOM RT Plan, RT Dose and RT Struct information objects directly created from the plan data of a SagiPlan study. The SagiPlan AE as a Storage SCU will try to initiate an association to a remote Storage SCP. The user has to select a destination DICOM peer for that. If multiple objects shall be transferred then multiple C- STORE requests will be issued over the association Proposed Presentation Contexts Table 3-10 Proposed Presentation Contexts for SagiPlan AE as Store SCU Abstract Syntax Transfer Syntax Role Extended Name UID Name UID Negotiation See See Explicit VR SCU None note note Little Endian 24/179
25 See note See note See note See note Explicit VR Big Endian Implicit VR Little Endian SCU None SCU None Note: This applies to all supported Storage Abstract Syntaxes, i.e. to all supported SOP Classes SOP Specific Conformance for SOP Classes Service Status Failure Warning Table 3-11 SagiPlan AE as Store SCU Response Status Handling Further Meaning Refused: Out of Resources Error: Data Set does not match SOP Class Error: Cannot understand Coercion of Data Elements Data Set does not match SOP Class Elements Discarded A7xx A9xx Cxxx Error Code B000 B007 B006 Behaviour The user is informed that the C-STORE request has failed. All outstanding C-STORE requests are cancelled. The SCP has stored the SOP Instance. The user is informed that the C-STORE request has been performed with warnings. Success Success 0000 The SCP has successfully stored the SOP Instance. The user is informed that the C-STORE request has been performed successfully Activity Query/Retrieve Description and Sequencing of Activities The user may enter values into a search mask for querying a Query/Retrieve SCP. The search mask is based on entries for Patient Names, Patient Registration Numbers, Patient Birthdates, Study IDs or Study Dates. After the user invokes the Query command, an association will be initiated to a remote Query/Retrieve SCP. The user has to define/select the AE Address, AE Title and port of the remote Query/Retrieve SCP. After association establishment a C-FIND request will be performed at the STUDY Level for querying all Studies which attributes match those of the search mask. For every received response at STUDY level a new C-FIND request will be performed at the SERIES Level for querying all Series of the related Study. All responses are arranged and displayed in a hierarchical list (Patient with one or more studies, consisting of one or more series). The user has to select at least one Patient, Study or Series in this list in order to invoke the Transfer command. After that an association to the remote Query/Retrieve SCP will be initiated. For every selected patient a C-MOVE request at PATIENT level will be per- 25/179
26 formed, for every selected Study a C-MOVE request at STUDY level and for every selected series a C-MOVE request at SERIES level. The AE Title of the move destination has to be selected by the user. It is even possible to select the SagiPlanDS AE as the move destination in order to make the transferred objects accessible by the SagiPlan application Proposed Presentation Contexts Study Root Query/Retriev e Information Model FIND Explicit VR Little Endian SC U Table 3-12 Proposed Presentation Contexts for SagiPlan AE as Query/Retrieve SCU Abstract Syntax Transfer Syntax Role Name UID Name UID Extended Nego tiation None Study Root Query/Retriev e Information Model FIND Explicit VR Big Endian SC U None Study Root Query/Retriev e Information Model FIND Implicit VR Little Endian SC U None Patient Root Query/Retriev e Information Model MOVE Explicit VR Little Endian SC U None Patient Root Query/Retriev e Information Model MOVE Explicit VR Big Endian SC U None Patient Root Query/Retriev e Information Model MOVE Implicit VR Little Endian SC U None Study Root Query/Retriev e Information Model MOVE Explicit VR Little Endian SC U None Study Root Explicit SC None 26/179
27 Query/Retriev e Information Model MOVE 2 VR Big Endian U Study Root Query/Retriev e Information Model MOVE Implicit VR Little Endian SC U None SOP Specific Conformance for SOP Classes Table 3-13 SagiPlan AE as Query/Retrieve SCU Response Status Handling for C-FIND Requests Service Status Failure Cancel Further Meaning Refused: Out of Resources Identifier does not match SOP Class Unable to process Matching terminated due to Cancel request Error Code A700 A900 Cxxx FE00 Behaviour The user is informed that the C-FIND request has failed. No results are displayed in the Query/Retrieve results list. No more results are displayed. Success Matching is complete No final Identifier is supplied. Pending Matches are continuing Current Match is supplied and any Optional Keys were supported in the same manner as Required Keys. Matches are continuing Warning that one or more Optional Keys were not sup The SCP has completed the matches. Results are displayed in the Query/Retrieve results list. FF00 FF01 The result transferred with this C-FIND Response is collected for display in the Query/Retrieve results list once the C-FIND operation is completed. 27/179
28 ported for existence and/or matching for this Identifier. A C-CANCEL-FIND request is issued after 1000 responses per C-FIND request have been received. Relational-queries are not supported. Table 3-14 SagiPlan AE as Query/Retrieve SCU Response Status Handling For C-MOVE Requests Service Status Failure Cancel Further Meaning Refused: Out of Resources Unable to calculate number of matches Refused: Out of Resources Unable to perform sub-operations Refused: Move Destination unknown Identifier does not match SOP Class Unable to Process Sub-operations terminated due to Cancel Indication Error Code A701 A702 A801 A900 Cxxx FE00 Behaviour None of the requested SOP Instances could be transferred (moved). The user will be informed about the failure. No more SOP Instances are transferred. Warning Success Pending Sub-operations Complete One or more Failures Sub-operations Complete No Failures Sub-operations are continuing B000 Some or all SOP Instances have not been transmitted successfully. The user will be informed about the Warning response All SOP Instances have successfully been transmitted. The user is informed that all FF00 objects have been retrieved successfully. Transferring requested SOP Instances is continuing. No C-CANCEL-MOVE requests are ever issued. 28/179
29 Association Acceptance Policy The SagiPlan Application Entity as a Query/Retrieve SCU never accepts associations. 3.3 Network Interfaces Physical Network Interface Supported communication stacks The Application provides TCP/IP Network Communication Support as defined in PS 3.8 (part 8 of the DICOM standard) TCP/IP stack The Application uses the TCP/IP stack of Microsoft Windows Operating System, which is the underlying operating system of the application Physical media support The Application can run on any physical network media that is supported by the underlying hardware and operating system (i.e. standard PCs and Windows operating system). 3.4 Configuration AE Title/Presentation Address Mapping Local AE Titles The local AE Titles for the SagiPlan as well as the SagiPlanDS Application Entities are configurable. The configuration is performed via the SagiPlan Setup Dialog and via the SagiPlanDS Service Controller Window, respectively. Table 3-15 AE Title Configuration Table Application Entity Default AE Title Default TCP/IP Port SagiPlan SagiPlan Not Applicable SagiPlanDS SagiPlanDS 104 (configurable) 29/179
30 Remote AE Title/Presentation Address Mapping SagiPlanDS Not Applicable SagiPlan Storage SCU The TCP/IP Address, the AE Title and the Port Number of the remote Storage SCP are configurable. Multiple configurations can be prepared via the corresponding SagiPlan setup menu. These can then be selected in the subsequent Send Files requests SagiPlan Query/Retrieve SCU Find The TCP/IP Address, the AE Title and the Port Number of the remote Query/Retrieve SCP are configurable. Multiple configurations can be prepared via the appropriate SagiPlan setup menu. These can then be selected in the subsequent Query requests. The AE Title of the move destination for the Query/Retrieve Move request is configurable too. Multiple AE Titles can be prepared via the appropriate SagiPlan setup menu. These can then be selected for the subsequent Transfer requests Parameters The following table shows DICOM relevant configuration parameters. Parameter Table 3-16 Configuration Parameters Table for SagiPlanDS Application Entity Configurable (Yes/No) General Parameters Time-out waiting for acceptance or rejection Response No to an Association Open Request. (Application Level timeout) Default Value 30 seconds General DIMSE level time-out values No Unlimited Time-out waiting for response to TCP/IP connect No Unlimited request. (Low-level timeout) Max PDU Size No /179
31 4 Media Interchange 4.1 Implementation Model Application Data Flow Diagram Export DICOM RT Files Application Entity SagiPlan FSC DICOM Media File (on any Windows Volume) Import DICOM Images Import DICOM RT Struct Application Entity SagiPlan FSR DICOM Standard Interface DICOM Media File (on any Windows Volume) Figure 4-1 APPLICATION DATA FLOW DIAGRAM Functional definitions of AE s The SagiPlan Application Entity is responsible for the import of DICOM Information objects (images and structure sets) to the SagiPlan application. It is also responsible for the export of DICOM Information Objects (treatment plans, dose information and structure sets) to other applications. Those functions are operated by the user invoking the Import / Export commands in the SagiPlan application. The SagiPlan Application Entity acts as a FSC for the export of DICOM Information Objects and as a FSR for the import of DICOM Information Objects. The SagiPlan Application Entity can write DICOM RT Plan, RT Dose and RT Struct Information Objects to any attached Windows Volume as a DICOM File which conforms to the 31/179
32 DICOM File Format as described in PS 3.10 of the DICOM standard. It can read DICOM Files (which conforms to the DICOM File Format) from any Windows Volume Sequencing of Real World Activities All FSR and FSC activities are sequentially initiated in the user interface of the SagiPlan application, and another activity may not be initiated until the prior activity has completed File Meta Information for Implementation Class and Version Table 4-1 File Meta Information for Implementation Class and Version File Meta Information Version 00, 01 Implementation Class UID Implementation Version Name Not Implemented 4.2 AE Specifications The SagiPlan AE provides standard conformance to the DICOM Interchange Option of the Media Storage Service Class Application Entity SagiPlan Table 4-2 AE Related Application Profiles, Real-World Activities, and Roles Supported Application Profile Real-World Activity Roles STD-GEN-CD Export DICOM RT Files FSC STD-GEN-DVD-RAM Import DICOM RT Struct FSR Import DICOM Images FSR File Meta Information for the SagiPlan Application Entity The SagiPlan Application Entity Title is not stored. 32/179
Tmypacs. DICOM Conformance Statement 1.0. Document Version: Tmypacs. Product Name(s): 2.0. Release:
Tmypacs DICOM Conformance Statement Document Version: Product Name(s): Release: 1.0 Tmypacs 2.0 Date: January 19, 2014 1. COMFORMANCE STATEMENT OVERVIEW The TmypacsServer is a DICOM server. The ImageServer
More informationDICOM Conformance Statement
Rogan-Delft B.V. Customer Information Bulletin Title DICOM Conformance Statement Scope Rogan OnLine XS Archiver Target Group Service Engineers Page 2 of 33 How To Contact Rogan-Delft BV Wiltonstraat 41
More informationDigiMam Conformance Statement for DICOM V3.0
DigiMam Conformance Statement for DICOM V3.0 Copyright 2004 by I.M.S. s.r.l. DOCUMENT VERSIONS Version Date Author Changes 1.00 15-Feb-05 IMS s.r.l. First Version DOCUMENT VERSIONS Page 2 of 29 TABLE OF
More informationKretztechnik AG. Voluson 730 Ultrasound System
Kretztechnik Voluson 730 Ultrasound System DICOM Conformance Statement Rev 1.02 Voluson 730 DICOM Coformance Rev 1.02 2001-07-18 Kretztechnik AG Zipf/Austria KRETZTECHNIK AG TIEFENBACH 15 Telefon: +43
More informationQuantitative Coronary Angiography on the Cardiovascular Measurement System. QCA-CMS Version 5.2 for Windows-NT, 2000 and XP
Quantitative Coronary Angiography on the Cardiovascular Measurement System QCA-CMS Version 5.2 for Windows-NT, 2000 and XP Conformance Statement September 1, 2002 MEDIS medical imaging systems Date: September
More informationDICOM Conformance Statement
GE Medical Systems Kretz Ultrasound DICOM Conformance Statement 105952 Revision 0 VOLUSON 730Expert/Pro V. 4.0.x 0123 Copyright 2000, 2001, 2002, 2003, 2004 by GE Medical Systems Kretztechnik GmbH & Co
More informationDICOM Conformance. DICOM Detailed Specification for Diagnostic Labs and Radiology Center Connectivity
DICOM Detailed Specification for Diagnostic Labs and Radiology Center Connectivity Authored by Global Engineering Team, Health Gorilla April 10, 2014 Table of Contents About Health Gorilla s Online Healthcare
More informationConformance Statement for DICOM Viewer
MedDream Conformance Statement for DICOM Viewer (version 5.1) 2015, Softneta UAB, Kaunas All rights reserved in the event of granting of patents or registration as a utility patent. All names of companies
More informationInteleViewer and lat er DI CO M Conf ormance St at ement
InteleViewer 4. 11. 1 and lat er DI CO M Conf ormance St at ement COPYRIGHT 2004-2017 Intelerad Medical Systems Incorporated. All Rights Reserved. No portion of the contents of this publication may be
More informationDICOM3.0 Conformance Statement
DICOM3.0 Conformance Statement INTRODUCTION 1 IMPLEMENTATION MODEL 2 AE Specifications 3 DIRECT DIGITIZER Communication Profiles 4 Extensions/Specializations/Privatizations 5 Configuration 6 0197 Support
More informationDICOM Conformance Statement
DICOM Conformance Statement Application Annex: MultiModality Applications on Philips IntelliSpace Portal V7.0 Koninklijke Philips N.V. 2014 All rights are reserved. Document Number: ICAPPF.0013673 Issued
More informationDICOM Conformance Statement
DICOM Conformance Statement Application Annex: MR-CT Roadmap R1.1 On Interventional Workspot R1.4 Koninklijke Philips N.V. 2017 All rights are reserved. ICAP-T-030001.09b Corresponds to ICAP-W-030001.02
More informationDICOM Conformance Statement
DICOM Conformance Statement Application Annex: 3D Roadmap R1.1.5 Koninklijke Philips N.V. 2015 All rights are reserved. Document Number: ICAP-PF.0015381 Issued by: Philips Medical Systems Nederland BV,
More informationDICOM Conformance Specifications for the HDI Ultrasound System (Release 0.9)
DICOM Conformance Specifications for the HDI 1500 Ultrasound System (Release 0.9) Table of Contents 0. Introduction...3 0.1 DICOM Background...3 1. Implementation Model...3 1.1 Application Data Flow Diagram...3
More informationREGIUS CONSOLE CS-3. DICOM 3.0 Conformance Statement CODE NO Manufacturer: 1 Sakura-machi, Hino-shi Tokyo , Japan
REGIS CONSOLE CS-3 DICOM 3.0 Conformance Statement CODE NO. 0862 Manufacturer: 1 Sakura-machi, Hino-shi Tokyo 191-8511, Japan Revision History Date Version Description 1 Contents 1 INTRODCTION... 4 1.1
More informationDICOM Conformance Statement
DICOM Conformance Statement Application Annex: Stentboost R4.2.5 Koninklijke Philips N.V. 2015 All rights are reserved. Document Number: ICAP-PF.0015387 Issued by: Philips Medical Systems Nederland BV,
More informationDigital Imaging and Communications in Medicine (DICOM) Supplement 39: Add Stored Print Media Storage - Retire Normalized Print Media Storage
Digital Imaging and Communications in Medicine (DICOM) Supplement 39: Add Stored Print Media Storage - Retire Normalized Print Media Storage DICOM Standards Committee, Working Group 300 N. 7th Street Rosslyn,
More informationAGFA MEDICAL IMAGING DICOM Conformance Statement
MED/RH/000428 Page 1 of 24 Agfa Medical Imaging 11 September, 2000 AGFA MEDICAL IMAGING DICOM Conformance Statement DICOM Print Server for Digital 400 V1.1 8 September, 2000 Page 2 of 24 MED/RH/000428
More informationicad, Inc. TotalLook TM DICOM Conformance Statement
icad, Inc. TotalLook T DICO Conformance Statement 5858-6005-2 Form 1 Proprietary Information of icad, Inc. Page 1 of 26 2007 icad, Inc. All rights reserved. Reproduction of this document without prior
More informationDICOM Conformance Statement
DICOM Conformance Statement Application Annex: US Applications on Philips IntelliSpace Portal V6.0 Koninklijke Philips Electronics N.V. 2013 All rights are reserved. Document Number: PIIOffc.0001323.01
More informationDICOM Enhancements Current and Future
RADIOLOGY RESEARCH Parts I & II 2 DICOM Enhancements Current and Future Part I An overview of the DICOM standard Parts of the standard The concept of objects Implementation of changes Donald Peck, PhD
More informationPrintlink5-IN. DICOM 3.0 Conformance Statement PRINT MANAGEMENT SYSTEM CODE NO Manufacturer:
PRINT MANAGEMENT SYSTEM Printlink5-IN DICOM 3.0 Conformance Statement CODE NO. 1116 Manufacturer: 2970 Ishikawa-machi, Hachioji-shi Tokyo 192-8505 Japan Important Notes - Konica Minolta Medical & Graphic,
More informationDICOM Conformance Statement
DICOM Conformance Statement Application Annex: US Applications on Philips IntelliSpace Portal V7.0 Koninklijke Philips N.V. 2014 All rights are reserved. Document Number: ICAP-PF.0013672 Issued by: Philips
More informationDICOM Conformance Statement
DICOM Conformance Statement Application Annex: NM Applications on Philips IntelliSpace Portal V5.0 Koninklijke Philips Electronics N.V. 2012 All rights are reserved. Document Number: PIIOffc.0000147.00
More informationDICOM Conformance Statement
DICOM Conformance Statement Application Annex: CT Applications on Philips IntelliSpace Portal V5.0 Koninklijke Philips Electronics N.V. 2012 All rights are reserved. Document Number: PIIOffc.0000143.01
More informationDICOM Educational Conference Brisbane, Australia
DICOM Educational Conference Brisbane, Australia SEPTEMBER 27-28, 2018 DICOM OVERVIEW & PROCESS KEVIN O DONNELL, CHAIR, DICOM WG10 (STRATEGIC) PAST-CHAIR, DICOM STANDARDS CMTE CANON MEDICAL RESEARCH USA
More informationCenova 2.1 DICOM Conformance Statement
BREAST IMAGING SOLUTIONS DICOM Standard Interface Cenova 2.1 DICOM Conformance Statement MAN-02008 Rev 001 Cenova 2.1 DICOM Conformance Statement MAN-02008 Rev 001 Technical Support For support in North
More informationA Module for Visualisation and Analysis of Digital Images in DICOM File Format
A Module for Visualisation and Analysis of Digital Images in DICOM File Format Rumen Rusev Abstract: This paper deals with design and realisation of software module for visualisation and analysis of digital
More informationDICOM Correction Item
DICOM Correction Item Correction Number CP-564 Log Summary: Type of Modification Correction Name of Standard PS 3.3, PS 3.6, PS 3.17 2004 Rationale for Correction A mammography CAD system often prefers
More informationDICOM Conformance Statement
DICOM Conformance Statement Application Annex: Nuclear Medicine Viewer on Xcelera R3.2L1 SP2 Koninklijke Philips Electronics N.V. 2011 All rights are reserved. Document Number: PIIOffc.0000081 Issued by:
More informationDICOM Correction Proposal
Tracking Information - Administration Use Only DICOM Correction Proposal Correction Proposal Number Status CP-1713 Letter Ballot Date of Last Update 2018/01/23 Person Assigned Submitter Name David Clunie
More informationDigital Imaging and Communications in Medicine (DICOM)
Digital Imaging and Communications in Medicine (DICOM) Supplement 197: Ophthalmic Optical Coherence Tomography for Angiographic Imaging Storage SOP Classes Prepared by: DICOM Standards Committee 1300 N.
More informationDICOM Conformance Statement
DV6950 DICOM Conformance Statement Product: DryView 6950 Laser Imaging System Document # 9J5911 This publication is protected by Federal Copyright law, with all rights reserved. Note: Publication of a
More informationDICOM Conformance Statement
Carestream DryView 5700 Laser Imager DICOM Conformance Statement May 10, 2018 Document # 8J5397 PART #: 8J5397 VERSION # 5.0 PAGE 1 of 40 This publication is protected by Federal Copyright law, with all
More informationDICOM Correction Proposal Form
DICOM Correction Proposal Form Tracking Information - Administration Use Only Correction Proposal Number CP-270 STATUS Assigned Date of Last Update 2001/06/20 Person Assigned Andrei Leontiev andrei_leontiev@idx.com
More informationDigital Imaging and Communications in Medicine (DICOM)
Digital Imaging and Communications in Medicine (DICOM) Supplement 197: Ophthalmic Tomography for Angiographic Imaging Storage SOP Classes Prepared by: DICOM Standards Committee 1300 N. 17 th Street Suite
More informationDICOM Conformance Statement
Carestream DryView 5850 Laser Imager DICOM Conformance Statement Document # 7F5608 All printed copies of this document are Uncontrolled. PAGE 1 of 40 Carestream Health Inc. reserves the right to change
More informationIHE Radiology Technical Framework Supplement. Stereotactic Mammography Image (SMI) Trial Implementation
Integrating the Healthcare Enterprise 5 IHE Radiology Technical Framework Supplement 10 Stereotactic Mammography Image (SMI) 15 Trial Implementation 20 25 Date: June 11, 2013 Author: IHE Radiology Technical
More informationDigital Imaging and Communications in Medicine (DICOM) Supplement 56: Ultrasound Waveform
Digital Imaging and Communications in Medicine (DICOM) Supplement 56: Ultrasound Waveform DICOM Standards Committee, Working Group 12 - Ultrasound 1300 N. 17 th Street, Suite 1847 Rosslyn, Virginia 22209
More informationScalable Data Storage Analysis and Solution for Picture Archiving and Communication Systems (PACS)
Scalable Data Storage Analysis and Solution for Picture Archiving and Communication Systems (PACS) Ayhan Akbal and Erhan Akbal II. PIURE ARCHIVING AND COMMUNICATION SYSTEM Abstract Information technologies
More informationDICOM Image And Data Management For Nuclear Medicine, Physiological Measurements, Radiotherapy And Ultrasound READ ONLINE
DICOM Image And Data Management For Nuclear Medicine, Physiological Measurements, Radiotherapy And Ultrasound READ ONLINE radiology information system (RIS) solutions, workstations for acquisition, non-dicom
More informationChin J Radiol 2003; 28:
Chin J Radiol 2003; 28: 175-180 175 modality console Image acquisition Quality assurance ISO Quality control PACS Picture Archiving and Communication System - Communication in Medicine [1] service WorkList
More information26 CP Correct order of reference to pixel spacing values in SR Image Library
26 CP-1526 - Correct order of reference to pixel spacing values in SR Image Library Page 1 1 Status JLetter Ballot 2 Date of Last Update 2016/01/18 3 Person Assigned David Clunie 4 mailto:dclunie@dclunie.com
More informationUser Manual Veterinary
Veterinary Acquisition and diagnostic software Doc No.: Rev 1.0.1 Aug 2013 Part No.: CR-FPM-04-022-EN-S 3DISC, FireCR, Quantor and the 3D Cube are trademarks of 3D Imaging & Simulations Corp, South Korea,
More informationGrubstr- 6-8 D Vörstetten Tel.: +49(0) Fax: +49(0) Steinhart Medizinsysteme GmbH
Hipax Diagnostic Workstation Table of Contents 1. PRODUCT DESCRIPTION... 2 2. HIPAX DW MODULES... 2 2.1 DW BASE MODULE... 2 2.1.1 DW BASE MODULE STANDARD (05-010)... 2 2.1.2 DW BASE MODULE LIGHT (05-020)...
More informationDICOM Implementations for Digital Radiography
The Medicine Behind the Image DICOM Implementations for Digital Radiography Dr. David A. Clunie, MB.,BS., FRACR Chief Technology Officer RadPharm, Inc. Disclosures David Clunie, MB.,BS., FRACR CTO, RadPharm,
More information23 CP Clarify Enhanced US Volume Image and Frame Type Values 3 and 4
23 CP-1463 - Clarify Enhanced US Volume Image and Frame Type Values 3 and 4 Page 1 1 Status Finale Text 2 Date of Last Update 2015/11/10 3 Person Assigned David Clunie 4 mailto:dclunie@dclunie.com 5 Submitter
More informationDICOM Implementations for Digital Radiography
The Medicine Behind the Image DICOM Implementations for Digital Radiography David A. Clunie Princeton Radiology Pharmaceutical Research Disclosure & Acknowledgements CTO RadPharm Proprietor of PixelMed
More informationOperation Manual. Canon CXDI-1 System Digital Radiography
Canon CXDI-1 System Digital Radiography Operation Manual Before using the instrument, be sure to read this manual thoroughly. Also, read the manuals of other instruments in this system. Keep the manual
More informationConformance Statement for DICOM Viewer
MedDream Conformance Statement for DICOM Viewer (version 5.00) 2015, Softneta UAB, Kaunas All rights reserved in the event of granting of patents or registration as a utility patent. All names of companies
More informationDICOM & IHE : An Update On -> the DICOM Standard -> the IHE Radiology Framework
RADIOLOGY RESEARCH The DICOM Standard http://medical.nema.org/ DICOM & IHE : An Update On -> the DICOM Standard -> the IHE Radiology Framework Michael Flynn, PhD mikef@rad.hfh.edu 2 DICOM Parts DICOM Review
More informationA SURVEY ON DICOM IMAGE COMPRESSION AND DECOMPRESSION TECHNIQUES
A SURVEY ON DICOM IMAGE COMPRESSION AND DECOMPRESSION TECHNIQUES Shreya A 1, Ajay B.N 2 M.Tech Scholar Department of Computer Science and Engineering 2 Assitant Professor, Department of Computer Science
More informationBest and Worst Practices and Software Development Tools DICOM
DICOMWeb 2015 Conference & Hands-on Workshop Best and Worst Practices and Software Development Tools DICOM David Clunie (dclunie@dclunie.com) PixelMed Publishing Background & Disclosures l Owner, PixelMed
More information23 CP Clarify Enhanced US Volume Image and Frame Type Values 3 and 4
23 CP-1463 - Clarify Enhanced US Volume Image and Frame Type Values 3 and 4 Page 1 1 Status Letter Ballot 2 Date of Last Update 2015/09/16 3 Person Assigned David Clunie 4 mailto:dclunie@dclunie.com 5
More informationETSI TS V1.1.1 ( ) Technical Specification
TS 100 392-3-8 V1.1.1 (2008-04) Technical Specification Terrestrial Trunked Radio (TETRA); Voice plus Data (V+D); Part 3: Interworking at the Inter-System Interface (ISI); Sub-part 8: Generic Speech Format
More informationCOST EFFECTIVE FLAT PANEL DIGITAL RADIOGRAPHY UPGRADE SOLUTIONS
COST EFFECTIVE FLAT PANEL DIGITAL RADIOGRAPHY UPGRADE SOLUTIONS DRive is a digital imaging DR hardware & Software solution designed for General Radiography of anatomy. It intended to replace film/screen
More information35 CP JPEG-LS Planar Configuration constraints conflict with WSI, US, VL, Enhanced Color MR and Page 1 36 compressed RGB images
35 CP-1843 - JPEG-LS Planar Configuration constraints conflict with WSI, US, VL, Enhanced Color MR and Page 1 36 compressed RGB images 1 Status Jan 2019 Voting Packet 2 Date of Last Update 2018/11/12 3
More informationCANopen Programmer s Manual Part Number Version 1.0 October All rights reserved
Part Number 95-00271-000 Version 1.0 October 2002 2002 All rights reserved Table Of Contents TABLE OF CONTENTS About This Manual... iii Overview and Scope... iii Related Documentation... iii Document Validity
More informationKODAK Dental Imaging Software. Quick Start Guide
KODAK Dental Imaging Software Quick Start Guide Notice Congratulations on your purchase of The KODAK Dental Imaging Software. Thank you for your confidence in our products and we will do all in our power
More informationDigital Image Management: the Basics
Digital Image Management: the Basics Napapong Pongnapang, Ph.D. Department of Radiological Technology Faculty of Medical Technology Mahidol University Outline From screen/film to digital radiography PACS/Tele
More information13 Compressed RGB components (rather than YBR) really are used by some WSI vendors in order to avoid the loss in conversion of 14 color spaces.
18 CP-1841 - Allow compressed RGB for WSI Page 1 1 Status Jan 2019 Voting Packet 2 Date of Last Update 2018/11/12 3 Person Assigned David Clunie 4 mailto:dclunie@dclunie.com 5 Submitter Name Aaron Stearrett
More informationINTERNATIONAL TELECOMMUNICATION UNION SERIES T: TERMINALS FOR TELEMATIC SERVICES
INTERNATIONAL TELECOMMUNICATION UNION ITU-T T.4 TELECOMMUNICATION STANDARDIZATION SECTOR OF ITU Amendment 2 (10/97) SERIES T: TERMINALS FOR TELEMATIC SERVICES Standardization of Group 3 facsimile terminals
More informationIMPAX 6 DISPLAY TOOL LIST
IMPAX 6 DISPLAY TOOL LIST IMPAX 6.0 TOOLS INDEX A Advance by Image Allows you to scroll from one image or frame to the next Advance by Page Pages through images in a large series, one screen at a time
More informationSECTION I - CHAPTER 1 DIGITAL RADIOGRAPHY: AN OVERVIEW OF THE TEXT. Exam Content Specifications 8/22/2012 RADT 3463 COMPUTERIZED IMAGING
RADT 3463 - COMPUTERIZED IMAGING Section I: Chapter 1 RADT 3463 Computerized Imaging 1 SECTION I - CHAPTER 1 DIGITAL RADIOGRAPHY: AN OVERVIEW OF THE TEXT RADT 3463 COMPUTERIZED IMAGING Section I: Chapter
More informationi1800 Series Scanners
i1800 Series Scanners Scanning Setup Guide A-61580 Contents 1 Introduction................................................ 1-1 About this manual........................................... 1-1 Image outputs...............................................
More informationCopley ASCII Interface Programmer s Guide
Copley ASCII Interface Programmer s Guide PN/95-00404-000 Revision 4 June 2008 Copley ASCII Interface Programmer s Guide TABLE OF CONTENTS About This Manual... 5 Overview and Scope... 5 Related Documentation...
More information33 CP Clarify handling of ICC profiles in WADO-RS for encapsulated images such as JPEG Page 1
33 CP-1804 - Clarify handling of ICC profiles in WADO-RS for encapsulated images such as JPEG Page 1 1 Status Assigned 2 Date of Last Update 2018/03/25 3 Person Assigned David Clunie 4 mailto:dclunie@dclunie.com
More informationDOCUMENT SCANNER INSTRUCTIONS. Space. Backup. Count Only. New File. Scanner. Feeding Option Manual Auto Semi-Auto
E FILM F Scanner A Space Count Only New File Feeding Option Manual Auto Semi-Auto Backup DOCUMENT SCANNER INSTRUCTIONS NOTICE q Copyright 2001 by CANON ELECTRONICS INC. All rights reserved. No part of
More informationAISG Specification Layer 2 Array Identification
AISG Specification Layer 2 Array Identification Revision History DATE ISSUE NOTES 31 January 2013 1.0 First release Copyright AISG Ltd 2013 Layer 2 Array Identification v1.0 Page 1 of 9 CONTENT Revision
More informationEleva Workspot 1.x - Training. Advanced User Functions. State: April 2007
Eleva Workspot 1.x - Training Advanced User Functions State: April 2007 Purpose of this Presentation General Get an overview about all "" functions in Eleva Workspot 1.x These are available in the System
More informationProgeny Imaging Veterinary
Progeny Imaging Veterinary User Guide V1.14 and higher 00-02-1605 Rev. K1 ECN: ECO052875 Revision Date: 5/17/2017 Contents 1. About This Manual... 6 How to Use this Guide... 6 Text Conventions... 6 Getting
More informationINTERNATIONAL TELECOMMUNICATION UNION DATA COMMUNICATION NETWORK: INTERFACES
INTERNATIONAL TELECOMMUNICATION UNION CCITT X.21 THE INTERNATIONAL (09/92) TELEGRAPH AND TELEPHONE CONSULTATIVE COMMITTEE DATA COMMUNICATION NETWORK: INTERFACES INTERFACE BETWEEN DATA TERMINAL EQUIPMENT
More informationThings you may want to know
Things you may want to know 1 / 9 Last updated 18-Nov-05 Why medical monitors are special? Unlike commercial displays, Medical displays are used in the hospital and they are displaying contain life critical
More informationMEDICAL X-RAY 2D AND 3D IMAGE VIEWER:ROLE FOR THE MEDICAL IMAGE IN DICOM STANDARD
MEDICAL X-RAY 2D AND 3D IMAGE VIEWER:ROLE FOR THE MEDICAL IMAGE IN DICOM STANDARD Mrs.B.A.Khivsara Mr.Shakadwipi Amol J. Mr. Nagare Sachin N. Mr. Phophaliya Abhijeet Mr.Gujrathi Apurv N. Abstract : A variety
More informationGet more from your images with Symphony Image Processing
DIRECT RADIOGRAPHY The user-friendly DelWorks image acquisition and processing software provides a wide range of tools for a variety of image enhancements. Its user interface simplifies every step of the
More informationISO INTERNATIONAL STANDARD. Graphic technology Prepress digital data exchange Tag image file format for image technology (TIFF/IT)
INTERNATIONAL STANDARD ISO 12639 Second edition 2004-05-15 Graphic technology Prepress digital data exchange Tag image file format for image technology (TIFF/IT) Technologie graphique Échange de données
More informationWiMOD LR Base Plus Firmware
WiMOD LR Base Plus Firmware Feature Specification Version 1.0 Document ID: 4000/40140/0137 IMST GmbH Carl-Friedrich-Gauß-Str. 2-4 47475 KAMP-LINTFORT GERMANY Overview Document Information File name WiMOD_LR_Base_Plus_Feature_Spec.docx
More informationModular Metering System ModbusTCP Communications Manual
Modular Metering System Manual Revision 7 Published October 2016 Northern Design Metering Solutions Modular Metering System ModbusTCP 1 Description The multicube modular electricity metering system simultaneously
More informationSolutions Page Content ImagePilot. Primary keywords: Digital radiography Computed radiography Image viewing and storage
Solutions Page Content Primary keywords: Digital radiography Computed radiography Image viewing and storage Solution Description For facilities with medium volume imaging, Konica Minolta s original all-in-one
More informationZView ver.2 DR Console Software. User Manual. (for HUMAN) Manual Version : 2.1 Revision : 002
ver.2 DR Console Software User Manual (for HUMAN) Manual Version : 2.1 Revision : 002 Table of Contents 1. Introduction... 4 2. Purpose... 4 3. System requirements... 5 3-1. Workstation... 5 3-2. Monitor...
More informationDRG-Series. Digital Radio Gateway. Tait P25 CCDI Tier-2 (TM9400 Series Mobile Radio) Digital Radio Supplement
DRG-Series Digital Radio Gateway Tait P25 CCDI Tier-2 (TM9400 Series Mobile Radio) Digital Radio Supplement DRG-Series Digital Radio Gateway Tait P25 CCDI Tier-2 (TM9400 Series Mobile Radio) Digital Radio
More informationProgeny Imaging. User Guide V x and Higher. Part Number: ECN: P1808 REV. F
Progeny Imaging User Guide V. 1.6.0.x and Higher Part Number: 00-02-1598 ECN: P1808 REV. F Contents 1 About This Manual... 5 How to Use this Guide... 5 Text Conventions... 5 Getting Assistance... 6 2 Overview...
More informationimedviewer Operation Manual V1.4 Shimadzu Software Development Canada inc.
imedviewer Operation Manual V1.4 Shimadzu Software Development Canada inc. 30 March, 2012 IMedViewer Operation Manual P a g e 2 Table of Contents 1 IMPORTANT USER INTERFACE CHANGE FROM PREVIOUS VERSION...
More informationINTERNATIONAL STANDARD
INTERNATIONAL STANDARD IEC 62320-2 Edition 2.0 2016-10 colour inside Maritime navigation and radiocommunication equipment and systems Automatic identification system (AIS) Part 2: AIS AtoN Stations Operational
More informationThe DICOM standard for medical thermal imaging by J. Ruminski*
July 2-5, 2008, Krakow - Poland The DICOM standard for medical thermal imaging *Department of Biomedical Engineering, Gdansk niversity of Technology, Gdansk, Poland Abstract by J. Ruminski* In imaging
More informationSmartRAD. Advanced Digital Radiography System
SmartRAD Advanced Digital Radiography System SmartRAD Expanding The Horizons Of Digital Radiography CMT introduces the SmartRAD Digital Radiography system, featuring an integrated flat panel digital detector
More informationi800 Series Scanners Image Processing Guide User s Guide A-61510
i800 Series Scanners Image Processing Guide User s Guide A-61510 ISIS is a registered trademark of Pixel Translations, a division of Input Software, Inc. Windows and Windows NT are either registered trademarks
More informationETSI ETR TECHNICAL July 1998 REPORT
ETSI ETR 300-5 TECHNICAL July 1998 REPORT Source: TETRA Reference: DTR/TETRA-01011-5 ICS: 33.020 Key words: TETRA, dialling, addressing Terrestrial Trunked Radio (TETRA); Voice plus Data (V+D); Designers'
More informationDICOM-compatible compression of WSI and diagnostic evaluation
of WSI and diagnostic evaluation R. Zwönitzer, H. Hofmann, A. Roessner, T. Kalinski 2nd European Workshop in Tissue Imaging and Analysis June 25-26, 2010 - Heidelberg 1 GPWL / GP-PPS Introduction Overview
More information1. Queries are issued to the image archive for information about computed tomographic (CT)
Appendix E1 Exposure Extraction Method examinations. 1. Queries are issued to the image archive for information about computed tomographic (CT) 2. Potential dose report screen captures (hereafter, dose
More informationImage optimization guide
Image Optimization guide for Image Submittal Images can play a crucial role in the successful execution of a book project by enhancing the text and giving the reader insight into your story. Although your
More informationETSI EN V1.1.1 ( )
EN 300 390-2 V1.1.1 (2000-09) Candidate Harmonized European Standard (Telecommunications series) Electromagnetic compatibility and Radio spectrum Matters (ERM); Land Mobile Service; Radio equipment intended
More informationGet more from your images with Symphony Image Processing
DIRECT RADIOGRAPHY The user-friendly DelWorks image acquisition and processing software possesses a wide range of tools for a variety of image manipulations. Its user interface simplifies every step of
More informationTable of Contents. Sound Technologies, Inc.
Table of Contents DICOM Conformance Statements and Version Histories are released regularly. For additional assistance with our product s latest DICOM and interconnectivity statements, in addition to or
More informationProjects Connector User Guide
Version 4.3 11/2/2017 Copyright 2013, 2017, Oracle and/or its affiliates. All rights reserved. This software and related documentation are provided under a license agreement containing restrictions on
More information745 Transformer Protection System Communications Guide
Digital Energy Multilin 745 Transformer Protection System Communications Guide 745 revision: 5.20 GE publication code: GEK-106636E GE Multilin part number: 1601-0162-A6 Copyright 2010 GE Multilin GE Multilin
More informationSECTION I - CHAPTER 2 DIGITAL IMAGING PROCESSING CONCEPTS
RADT 3463 - COMPUTERIZED IMAGING Section I: Chapter 2 RADT 3463 Computerized Imaging 1 SECTION I - CHAPTER 2 DIGITAL IMAGING PROCESSING CONCEPTS RADT 3463 COMPUTERIZED IMAGING Section I: Chapter 2 RADT
More informationChina Resources Wandong Medical Equipment Co., Ltd. High Frequency 50kW Digital RF System - HF51-5
China Resources Wandong Medical Equipment Co., Ltd. High Frequency 50kW Digital RF System - HF51-5 #3, No.9, Jiuxianqiaodong Road, Chaoyang District, Beijing 100015, P.R. China E-mail: international@wandong.com.cn
More information29 CP Define CT Reconstruction Diameter more precisely and correct Enhanced CT illustration Page 1
29 CP-1569 - Define CT Reconstruction Diameter more precisely and correct Enhanced CT illustration Page 1 1 Status Final Text 2 Date of Last Update 2016/09/08 3 Person Assigned David Clunie 4 mailto:dclunie@dclunie.com
More informationGenesis Channel Manager
Genesis Channel Manager VPI 160 Camino Ruiz, Camarillo, CA 93012-6700 (Voice) 800-200-5430 805-389-5200 (Fax) 805-389-5202 www.vpi-corp.com 1 Contents Genesis Channel Manager -----------------------------------------------------------------------------------------------------------------
More information