Keysight Technologies How to Make the Best Switch Mode Power Supply Measurements. Application Note

Size: px
Start display at page:

Download "Keysight Technologies How to Make the Best Switch Mode Power Supply Measurements. Application Note"

Transcription

1 Keysight Technologies How to Make the Best Switch Mode Power Supply Measurements Application Note

2 Introduction Characterizing the operation of switch mode power supplies requires a broad range of measurements. Most of the required measurements can be performed using an oscilloscope. However, many engineers today still perform these measurements manually on their scopes, which is very time consuming. Many of today s midand high-end oscilloscopes offer power supply measurement options that can help automate the process of the characterizing a power supply. This application note will look at how to make the following switch mode power supply measurements with an oscilloscope s automated power measurement application: Input analysis Power quality Current harmonics Inrush current Switching analysis Rds(on) and Vce(sat) Switching loss Slew rate Modulation Output analysis Output ripple Turn-on/off time Transient response Efficiency Frequency response analysis Power supply rejection ratio (PSRR) Control loop response (Bode gain and phase) Quite unique in this set of measurements, for an oscilloscope that is, are the frequency response measurements. These types of measurements typically require a vector network analyzer (VNA), which is also called a frequency response analyzer (FRA). This application note begins with a brief discussion on probing tips, and then provides step-by-step instructions on how to perform the various power supply characterization measurements using an InfiniiVision 3000T or 4000 X-Series oscilloscope with the Power Measurement option (DSOX3PWR/ DSOX4PWR). This application note also provides explanations of each measurement and why it is important for complete power supply characterization.

3 03 Keysight How to Make the Best Switch Mode Power Supply Measurements - Application Note Table of Contents Introduction... 2 Probing Tips... 4 Deskewing the Probes... 6 Power Quality Analysis... 7 Current Harmonics Analysis... 8 Inrush Current Analysis... 9 Switching Loss Analysis Rds (on) and Vce (sat) Analysis Slew Rate Analysis Modulation Analysis Output Ripple Analysis Turn-on/Turn-off Analysis Transient Response Analysis PSRR Analysis Control Loop Response (Bode Gain and Phase) Analysis Efficiency Analysis Conclusion Figure 1. The Keysight Technologies, Inc. switch mode power supply measurements training kit. Required equipment Keysight 3000T or 4000 X-Series oscilloscope with the power measurements option (DSOX3PWR or DSOX4PWR) U1880A deskew fixture with USB cable for power N2790A high-voltage differential active probe, or equivalent 1147B 15A current probe, or equivalent 10:1 passive voltage probe Your switch mode power supply (SMPS) or Keysight s SMPS measurements training kit (Figure 1)

4 04 Keysight How to Make the Best Switch Mode Power Supply Measurements - Application Note Probing Tips If using Keysight s N2790A high-voltage differential active probe (Figure 2) for your power measurements, this probe can be manually set for either 50:1 attenuation for up to 140 V (DC plus peak AC) measurements, or 500:1 attenuation for up to 1400 V measurements. If you are using Keysight s SMPS training kit as your DUT, the 50:1 setting is appropriate. But, if you are performing measurements on your own switching power supply that may have higher input and switching voltages, then you may need to use the 500:1 setting. When you initially connect the N2790A probe to one of the input channels of the Keysight 3000T/4000/6000 X-Series oscilloscope, the scope will automatically detect and set its own probe attenuation factor for that channel to 50:1, but if you have the probe attenuation manually set on the probe to 500:1, then you will need to manually enter a 500:1 attenuation factor in the scope s probe menu for that particular channel. The 1147B or N2893A current probe (Figure 3) is a 0.1 V/A probe (10:1 attenuation). When this probe is connected to any input channel of a Keysight 3000T/4000/6000 X-Series oscilloscope, the scope will automatically detect that a current probe is connected to provide measurements in units of Amps. It will also automatically detect and compensate for the 10:1 probe attenuation. No user settings are required. Figure 2. Keysight s N2790A 100-MHz high-voltage differential active probe. When you connect the current probe to a current loop in your device under test, make sure to fully close AND lock the probe clamp. To lock it in place, slide the locking mechanism fully forward until you feel and hear it click in place. Figure 3. Keysight s 1147B 50-MHz, 15-A AC/DC current probe.

5 05 Keysight How to Make the Best Switch Mode Power Supply Measurements - Application Note Probing Tips (Continued) When using the 1147B or N2893A current probe to make current measurements on Keysight s SMPS training kit, it is very easy to connect to various designed-in PC board current loops. But, if you are making measurements on your own switching power supply, rarely are designed-in current loops available. This means you will probably need to create temporary current loops in your prototype power supply s circuitry in order to measure current and power. Figure 4 shows an example of a wire-loop in series with the Source terminal on a FET switching device in order to measure drain-to-source current. To measure input AC current, engineers will often just strip back the insulation on the power cord to access the line current wire. Note that although this may be quick and easy, this is NOT a Keysight-recommended technique. After extended use, a magnetic field can build up in a current probe. When you are making power supply measurements using a current probe, you should occasionally degauss (de-magnify) the probe. To do this, simply disconnect the current probe from the device under test, close and lock the current probe clamp, and then press the DEMAG button near the base of the probe where it connects to the scope. Note that you can also calibrate the offset of the probe (and scope) at the same time by rotating the ZERO ADJ thumb-wheel until the baseline current waveform aligns with the ground indicator on the scope s display. Figure 4. Creating a special current loop in your power supply to measure current.

6 06 Keysight How to Make the Best Switch Mode Power Supply Measurements - Application Note Deskewing the Probes When using dissimilar type probes, such as a voltage probe and current probe, it is important to perform a probe deskew calibration. This automatic procedure using the recommended U1880A deskew fixture (Figure 5) will null-out the difference in propagation delay between your probes to provide more accurate power measurements. This is especially important for the switching loss measurement where just a few nanoseconds can sometimes make a big difference in power and energy loss measurements during the turn-on and turn-off phases of transistor switching. 1. Connect the U1880A s USB cable between the deskew fixture and the USB port on the back of the oscilloscope. 2. Press the [Default Setup] front panel key. 3. Press the [Analyze] front panel key; then press the Features softkey and select the Power Application. 4. Press the Analysis softkey; then select Auto Deskew. 5. Set the S1 switch on the deskew fixture to the Small Loop setting as shown in the on-screen diagram in Figure 6. Select Setup Diagram to view. 6. Connect the N2790A high-voltage differential active probe between the scope s channel-1 input to J6 (red lead to red test point) and J7 (black lead to black test point) on the U1880A deskew fixture as illustrated in the on-screen diagram. 7. Connect the current probe from the scope s channel-2 input to the Small Loop in the direction indicated on the deskew fixture (current flow towards the top of the board). 8. Press the Auto Deskew softkey on the scope. When the automatic deskew calibration completes, your scope s display should look similar to Figure 7. The deskew calibration factor will remain in the channel-2 probe settings menu and is non-volatile. But you can manually change the deskew factor, or perform a factory default setup within the scope s [Save/Recall] menu to reset it back to zero. Performing a standard front panel [Default Setup] will NOT reset the deskew calibration factor. Figure 5. Keysight s U1880A deskew fixture. Figure 6. Deskew connections diagram. Figure 7. Deskew display after completing automatic deskew between voltage and current probes.

7 07 Keysight How to Make the Best Switch Mode Power Supply Measurements - Application Note Power Quality Analysis Power quality analysis measures the quality of the input AC line signal that supplies power to your switching power supply while operating. This measurement provides the following input signal quality parameters: Real power (P = VInstantaneous x I Instantaneous averaged over N cycles) Apparent power (S = VRMS x IRMS over N cycles) Reactive power (Q = SQRT(S 2 -P 2 )) Power factor (PF = Real/Apparent) Voltage crest factor (CF V = VPeak/VRMS) Current crest factor (CF I = IPeak/IRMS) Phase angle (Φ = ACOS(PF)) 1. If using Keysight s SMPS training kit, set the S2 load switch to the ON position (high-load, maximum current). 2. Press the [Default Setup] front panel key. 3. Press the [Analyze] front panel key; then select the Power Application under the Features softkey. 4. Press the Analysis softkey; then select the Power Quality measurement. 5. Press the Signals softkey. 6. Ensure that Voltage is set to 1 (channel-1), and Current is set to 2 (channel-2). 7. Connect the N2790A high-voltage differential active probe from the scope s channel-1 input to Line (red lead) and Neutral (back lead) as illustrated in the connections diagram shown in Figure 8. If using Keysight s SMPS training kit, connect the probe to TP2 (red lead) and TP1 (black lead). 8. Connect the current probe from the scope s channel-2 input to a wire-loop of the input line signal. If using Keysight s SMPS training kit, connect the current probe to the J1 loop in the direction indicated. 9. Press the AutoSetup softkey; then press Apply. Apparent Power S (VA) Real Power P (W) Φ Figure 7. Apparent, real, and reactive power. Reactive Power Q (VAR) Figure 8. Connections diagram for the power quality measurement. When AutoSetup is pressed, the scope optimally scales the voltage (yellow trace) and current (green trace) waveforms, turns on the waveform math power waveform (purple trace), and then sets the timebase to display two cycles (default setting) across the screen. Note that if you observe the current waveform (green trace) out of phase relative to the voltage waveform (yellow trace), which will result in negative power waveform pulses (purple trace), then your current probe is probably connected backwards. When Apply is pressed, the InfiniiVision X-Series scope will automatically measure all of the power quality parameters as shown in Figure 9. Figure 9. Input power quality measurements.

8 08 Keysight How to Make the Best Switch Mode Power Supply Measurements - Application Note Current Harmonics Analysis Current harmonics analysis measures the amplitude of frequency components that can be injected back into the AC lines. Endproducts must often meet specific standards of compliance in order not to disturb other equipment connected to the AC supply grid. This measurement will perform an FFT measurement on the current waveform, compare the results of the amplitudes of odd and even harmonics against a user-selected IEC standard, and also provides color-coded pass/fail indicators for each tested frequency up to the 40th harmonic. 1. If using Keysight s SMPS training kit, set the S2 load switch to the ON position (high-load, maximum current). 2. Press the [Analyze] front panel key; then select the Power Application under the Features softkey. 3. Press the Analysis softkey; then select the Current Harmonics measurement. 4. Press the Signals softkey. 5. Connect the N2790A voltage probe from the scope s channel-1 input to Line (red lead) and Neutral (back lead) as illustrated in the connections diagram shown In Figure 10. If using Keysight s SMPS training kit, connect the probe to TP2 (red lead) and TP1 (black lead). 6. Connect the current probe between the scope s channel-2 input to a wire-loop of the input line signal. If using Keysight s SMPS training kit, connect the current probe to the J1 loop in the direction indicated. 7. Ensure that Voltage is set to 1 (channel-1), and Current set to 2 (channel-2). 8. Press AutoSetup. If you use the default settings, the scope will display 20 cycles of the input line voltage (yellow trace) and current (green trace) waveforms on the scope s display. 9. Press the Settings softkey; then press the Line Freq softkey and set the frequency to either 50 Hz, 60 Hz, or 400 Hz depending upon the line frequency in your part of the world or application. Note that you can also select the appropriate IEC standard to test against in this menu as well. 10. Press the [Back] softkey to return to the previous menu; then press the Apply softkey to begin the Current Harmonics measurement. Figure 10. Connections diagram for the current harmonics measurement. Figure 11. Current harmonics measurement in a tabular format. When Apply is pressed, the scope performs an FFT waveform math operation (purple trace) on the current waveform with the results shown in a tabular format in the upper half of the scope s display as shown in Figure 11. The scope measures up to the 40th harmonic and compares against the selected IEC standard. The scope also measures and reports total harmonic distortion (THD). To view the higher harmonics, press the Scroll Harmonics softkey and then rotate the knob. 11. To view the results in a bar chart format, press the Settings softkey, then press the Display softkey and change from the Table setting to the Bar Chart setting as shown in Figure 12. Figure 12. Current harmonics measurement in a bar chart format.

9 09 Keysight How to Make the Best Switch Mode Power Supply Measurements - Application Note Inrush Current Analysis Inrush current analysis measures the peak input current (positive or negative) when power is initially turned on. Since this measurement is based on the acquisition of a single-shot event, there is not an AutoSetup selection for this measurement (AutoSetup requires a repetitive input signal). For this reason, you must enter the expected peak current surge, as well as the steady-state peak-to-peak line voltage so that the scope can establish initial vertical scaling. The scope then provides step-by-step instructions on how to perform this single-shot measurement. 1. If using Keysight s Power Measurements training kit, set the S2 load switch to the ON position (high-load, maximum current). 2. Press the [Analyze] front panel key; then select the Power Application under the Features softkey. 3. Press the Analysis softkey; then select the Inrush Current measurement. 4. Press the Signals softkey; then connect the N2790A voltage probe between the scope s channel-1 input to Line (red lead) and Neutral (back lead) as illustrated in the connections diagram shown in Figure 13. If using Keysight s SMPS training kit, connect to TP2 (red lead) and TP1 (black lead) on the demo board. 5. Connect the current probe between the scope s channel-2 input to a wire-loop of the input Line signal. If using Keysight s training kit, connect the current probe to the J1 loop in the direction indicated. 6. Ensure that Voltage is set to 1 (channel-1), and Current is set to 2 (channel-2). 7. Set Max Vin and the approximate Expected surge current for your device under test. If using Keysight s SMPS training kit, then the default settings should be appropriate for this measurement. Note that the default Expected surge current and Max Vin settings have been optimized (default setup) for Keysight s SMPS training kit. Although the steady-state peak current of the demo board is approximately 1 A (with the S2 load switch set ON), the peak inrush current is much higher and can sometimes exceed ± 30 A. Determining the expected peak current of your device under test will be an iterative/trial & error process. The value that you enter for Expected should be higher than the actual peak current that will be captured and measured. 8. After accepting or changing the settings in the Signals menu, press the [Back] front panel key (left side of scope) to return to the previous menu. 9. Press the Apply softkey; then follow the step-by-step on-screen instructions, which are repeated below. 10. Turn OFF power; then press Next. 11. Turn ON power; then press Next. 12. Repeat several times in order to obtain the worst-case peak inrush current as shown in Figure 14. Figure 13. Setting expected inrush current and peak-to-peak line voltage for a single-shot inrush current measurement. Figure 14. Inrush current measurement. If the Peak Current measurement indicates either > or <, this is an indication that the current waveform is clipped and that you should then increase the expected current setting in the Signals menu for a more accurate measurement. The absolute peak inrush current will typically occur if the power switch is turned on at the same instance in time that the input line voltage is either at its absolute positive or negative peak value. This means that you should repeat this measurement several times in order to measure the worst-case inrush current.

10 10 Keysight How to Make the Best Switch Mode Power Supply Measurements - Application Note Switching Loss Analysis Switching loss analysis measures power and energy losses of your switching device (typically an FET). In a switch mode power supply (SMPS), most power and energy losses occur during the switching phases of the transistor when it turns on and turns off. During these phases the switching transistor goes in and out of saturation and momentarily operates in its linear region. Power and energy losses also occur during the conduction phase of the switching transistor. This is when voltage is at the transistor s saturated minimum and current flows. Losses during the non-conduction phase are typically insignificant, and should theoretically be zero. The Switching Loss measurement automatically measures losses over one switching cycle. But you can then optimize the scope s horizontal and vertical settings to perform more accurate power and energy loss measurements during particular switching phases. 1. If using Keysight s SMPS training kit, set the S2 load switch to the ON position (high-load, maximum current). 2. Press the [Analyze] front panel key; then select the Power Application under the Features softkey. 3. Press the Analysis softkey; then select the Switching Loss measurement. 4. Press the Signals softkey. 5. Connect the N2790A high-voltage differential probe between the scope s channel-1 input to the drain/ collector (red lead) and source/emitter (black lead) of your switching transistor as illustrated in the on-screen connections diagram of Figure 15. If using Keysight s SMPS training kit, connect to TP3 (red lead) and TP4 (black lead). 6. Connect the current probe between the scope s channel-2 input to a wire-loop from the source/emitter trace. If using Keysight s SMPS training kit, connect the current probe to the J2 current loop in the direction indicated. Note that before connecting the current probe to the current loop, this might be a good time to de-gauss the current probe. 7. Ensure that Voltage is assigned to 1 (channel-1), and Current to 2 (channel-2) within the Signals menu. 8. Press the AutoSetup softkey; then press Apply. Figure 15. Connecting the voltage and current probes to the transistor switching device. When AutoSetup is pressed, the scope will establish optimum vertical scaling on the voltage (yellow trace) and current (green trace) waveforms, sets the main timebase (upper display) to show two switching cycles, and sets the zoom timebase (lower display) to show one switching cycle.

11 11 Keysight How to Make the Best Switch Mode Power Supply Measurements - Application Note Switching Loss Analysis (Continued) At this point we are almost ready to begin making measurements. However, notice the on-screen message in red that says Offset Calibration: Required. Determining the losses during the conduction phase (when the switching transistor is fully turned on and the voltage between the drain and source of the transistor is near zero volts), requires that this voltage level be measured very accurately in order to accurately compute the power waveform (V x I) during this phase. Unfortunately, scopes do not typically have sufficient dynamic range to accurately measure the high switching voltage and the low saturation voltage at the same time. In other words, you may be asking the scope to accurately measure a low-level voltage while the scope s V/div setting is set to a high value. However, the biggest source of error is not insufficient bits of resolution as many would think. The biggest source of error is offset error induced by the scope s front-end circuitry and probe. This is true even though the scope s offset may be well within its stated specification. For instance, a typical offset specification may be something like ± 0.1 division. This may not seem like much error for most oscilloscope measurement applications, but ± 0.1 division can be huge when attempting to measure losses during the conduction phase. You should now see a green message saying the offset has been calibrated. The precision offset calibration that you just performed was for a single set of vertical setup conditions (channel-1 V/div and offset, and channel-2 V/div and offset). Minor offset errors that are within the scope s specification can vary slightly at different vertical and offset setup conditions. So at this point, do not change any vertical or offset settings on the scope. If you do, the precision offset calibration will be nullified. To now perform the switching loss measurement across one switching cycle (zoom timebase window), proceed with the following step. 14. Press Apply. When Apply is pressed, the scope turns on the multiply waveform math function (V x I) to create the power waveform (purple trace), and also measures the following parameters as shown in Figure 16: Switching frequency (counter measurement) Power loss (average power across timebase zoom window Power loss/cycle Energy loss (integration of power loss) Keysight s InfiniiVision oscilloscope provides for a highprecision offset calibration that be quickly performed once the oscilloscope is setup for a switching loss measurement. During this calibration, the scope measures its own error based on a 0.0 volt and 0.0 ampere input. The scope then uses the measured error to automatically correct the computed power waveform, which is based on multiplying the instantaneous voltage by the instantaneous current. Continuing from Step 8 above, let s now perform a precision offset calibration. 9. Press Offset Calibration, and then OK to confirm that the scope s vertical settings have been established for the switching loss measurement. 10. Following on-screen directions, disconnect your differential voltage probe from the DUT and short the tips together (0.0 V), then press OK. 11. When the voltage offset measurement has completed, then disconnect the current probe for the DUT, clamp the jaws fully closed, then press the DEMAG button on the probe. 12. Next press OK to perform the current offset error measurement at 0.0 amps. 13. Reconnect probes to the DUT and press OK. Figure 16. Power and Energy loss measurement over one switching cycle.

12 12 Keysight How to Make the Best Switch Mode Power Supply Measurements - Application Note Switching Loss Analysis (Continued) The scope should now be measuring the power and energy losses across one switching cycle. To measure losses during specific phases of switching, simply adjust the zoom window for the desired phase, such as the conduction phase or the turn-off phase. Figure 17 shows an example of measuring the losses during the conduction phase. The power loss parameter is the average of the power waveform over the time-span of the zoom timebase window. If the zoom timebase is zoomed into just a portion of the switching cycle, then power loss has little meaning. The PLoss/Cyc provides the portion of power loss relative to a full cycle. Energy Loss is the integral of the power waveform over the zoom timebase window. Another method to measure switching losses that may provide more accurate results is to enter a specified value of Rds(on), or measure Rds(on) and then use that measured value as shown in Figure 18. To measure power and energy losses based on an entered Rds(on) value of 200-mΩ, continue with the following steps: Figure 17. Measuring power and energy loss during the conduction phase. 15. Press the Settings softkey. 16. Press the Conduction softkey and then select Rds(on). 17. Press the Rds(on) softkey and enter a value of 200-mΩ. When Rds(on) has been selected as the conduction method, the scope computes power losses during the conduction phase as I 2 x Rds(on). The scope automatically determines the conduction phase based on the V Ref setting (default = 5%). During other switching phases, including the turn-off phase (when transistor switches from on to off), and the non-conduction phase (when transistor is fully turned off), the scope computes power losses based on V x I. However, during the non-conduction phase the scope uses I = 0.0 amps in the equation if current is below the I Ref setting (default = 5%). Any measured current during the non-conduction phase is typically due to oscilloscope/current probe offset error. Figure 18. Measuring power and energy losses based on using an entered value of Rds(on).

13 13 Keysight How to Make the Best Switch Mode Power Supply Measurements - Application Note Rds(on) and Vce(sat) Analysis Rds(on) is the effective resistance from drain to source of a FETtype transistor when it is fully turned on during the conduction phase of switching. Vce(sat) is the saturation voltage of a bipolar type transistor when it is fully turned on during the conduction phase. To measure Rds(on) or Vce(sat), do the following: 1. If using Keysight s SMPS training kit, set the S2 load switch to the ON position (high-load, maximum current). 2. Press the Analyze front panel key; then select the Power Application under the Features softkey. 3. Press the Analysis softkey; then select the Rds(on) & Vce(sat) measurement. 4. Press the Signals softkey. 5. Connect the N2790A high-voltage differential probe between the scope s channel-1 input to the drain (red lead) and source (black lead) of the switching transistor as illustrated in the on-screen connection diagram of Figure 19. If using Keysight s SMPS training kit, connect to TP3 (red lead) and TP4 (black lead). 6. Connect the current probe between the scope s channel-2 input to a wire-loop from the source/emitter trace. If using Keysight s SMPS training kit, connect the current probe to the J2 current loop in the direction indicated. 7. Ensure that Voltage is assigned to 1 (channel-1), and Current to 2 (channel-2) within the Signals menu. 8. Press the AutoSetup softkey. When AutoSetup is pressed, the scope will establish optimum vertical scaling on the voltage (yellow trace) and current (green trace) waveforms, sets the main timebase (upper display) to show two switching cycles, and sets the zoom timebase (lower display) within the middle 50% of the conduction phase. At this point you should observe an on-screen message that says Offset Calibration: Required. Just as performing switching loss measurements accurately requires that minor offset errors be calibrated out, measuring Rds(on) also requires that minor offset errors be calibrated out. Refer to the previous section in this document on Switching Loss Analysis for a more detailed discussion of why this is important. Proceed with performing a precision offset calibration as described previously, or simply follow the on-screen instruction after pressing Offset Calibration. After performing the offset calibration, continue with the following step to measure Rds(on): When Apply is pressed, the scope measures the average corrected voltage and average corrected current during the zoom timebase window and then computes Rds(on) as V/I as shown in Figure 20. In this example, Rds(on) measured 129-mΩ. If you now press the Use Rds(on) softkey, the scope will rescale the zoom timebase to a one full switching cycle, plug in the average measured value of Rds(on), and then begin to perform a power and energy switching loss measurement based on using Rds(on) during the conduction phase. Refer to the previous section of this document for a more detailed discussion on this topic. Figure 19. Connecting the voltage and current probes to the transistor switching device for the Rds(on) measurement. 9. Press Apply. Figure 20. Measuring Rds(on).

14 14 Keysight How to Make the Best Switch Mode Power Supply Measurements - Application Note Slew Rate Analysis Slew rate analysis measures the rate of change of the voltage and/or current waveform when the switching transistor of your power supply switches on and off. 1. If using Keysight s SMPS training kit, set the S2 load switch to the ON position (high-load, maximum current). 2. Press the [Analyze] front panel key; then select the Power Application under the Features softkey. 3. Press the Analysis softkey; then select the Slew Rate measurement. 4. Press the Signals softkey; then connect the N2790A voltage probe between channel-1 of the scope to the drain/collector (red lead) and source/emitter (black lead) of your switching transistor as illustrated in the on-screen connections diagram of Figure 21. If using Keysight s SMPS training kit, connect to TP3 (red lead) and TP4 (black lead). 5. Connect the current probe to a wire-loop of the source/ emitter trace. If using Keysight s SMPS training kit, connect the current probe to the J2 current loop in the direction indicated. 6. Ensure that Voltage is set to 1 (channel-1), and Current is set to 2 (channel-2). 7. Press the AutoSetup softkey; then press Apply. Your scope s display should now look similar to Figure 22. AutoSetup will optimally scale the voltage and current waveforms, sets the main timebase (upper display) to show one switching cycle across screen, and sets the zoom timebase (lower display) windowed-in on the turn-off phase of the switching voltage and current waveforms. When Apply is pressed, the scope turns on the differentiate (dv/dt) math function (purple trace) based on the voltage waveform in order to dynamically plot the slew rate of the voltage waveform. Apply also turns on the Max and Min measurements in order to automatically measure the maximum and minimum slew rates of this waveform. Figure 21. Slew rate measurement connections diagram. dv/dt Figure 22. Slew rate measurement of the voltage waveform during the transistor s turn-off phase. To measure the maximum and minimum slew rates of the current waveform, press the Source softkey and change the source setting from Voltage to Current. To measure the slew rates of the voltage and current waveform during the transistor turn-on phase, adjust the horizontal delay/ position knob until the zoom timebase is windowed-in on the turn-on phase (when voltage drops to near zero) as shown in Figure 23. Note that for a more stable display of the turn-on phase, you may want to leave the delay setting at zero, but then change the trigger edge from rising to falling. This will establish a more stable trigger during the turn-on phase with the zoom display centered on-screen. Figure 23. Slew rate measurement of the current waveform during the transistor turn-on phase.

15 15 Keysight How to Make the Best Switch Mode Power Supply Measurements - Application Note Modulation Analysis Modulation analysis is typically used to characterize the pulse-width modulation (PWM) feedback from the DC output to the switching transistor s gate terminal for dynamic voltage regulation. Unfortunately, this measurement cannot be demonstrated using Keysight s SMPS training kit since the FET s Gate terminal is not accessible for probing. 1. Press the [Analyze] front panel key; then select the Power Application under the Features softkey. 2. Press the Analysis softkey; then select the Modulation measurement. 3. Press the Signals softkey. 4. Connect the N2790A voltage probe between channel-1 of the scope to the gate/base (red lead) and source/ emitter (black lead) of your switching transistor as illustrated in the on-screen connections diagram of Figure 24. Note that although the connection diagrams also shows a current probe connected to the drain, this probe connection is not necessary for this measurement. 5. Ensure that Voltage is set to 1 (channel-1). 6. Press the AutoSetup softkey; then press Apply. Figure 24. Modulation measurement connections diagram. Measurement Trend waveform of Duty Cycle When AutoSetup is pressed the scope optimally scales the voltage waveform (gate-to-source signal) and sets the timebase based on the user-defined duration setting. When Apply is pressed, the scope turns on the measurement trend waveform math function (purple trace) based on sequential duty cycle measurements of the gate-to-source signal (yellow trace) across the scope s display as shown in Figure 25. This is basically a plot of duty cycle on the vertical axis and time on the horizontal axis. In this example, we have rescaled the timebase to 200 μs/div (2 ms duration) in order to optimally observe duty cycle modulation of the gate signal. The duty cycle of this switching device s gate signal appears to be modulating at a 1 khz rate with a maximum measured duty cycle of approximately 15% and minimum measured duty cycle of approximately 13%. The modulation measurement can also be very useful for characterizing turn-on characteristics of your switching power supply. Figure 26 shows an example of a single-shot measurement of the frequency of the switching transistor s gate signal when power is initially applied. In this case we have changed the measurement type from duty cycle to frequency, and then manually set up the scope to perform a single-shot acquisition. In this example we can observe the amplitude of the gate-to-source signal (yellow trace) ramp up to its steady-state condition in approximately 1.8 ms. The measurement trend waveform (purple trace), which is based on sequential frequency measurements of the gate signal across the display, shows that the transistor quickly ramps up to a steady-state switching frequency of 69 khz in 600 μs. Figure 25. Modulation measurement of duty cycle on the switching transistor s Gate input. Measurement Trend waveform of Frequency Figure 26. Modulation measurement of frequency on the switching transistor s Gate input during power-up.

16 16 Keysight How to Make the Best Switch Mode Power Supply Measurements - Application Note Output Ripple Analysis Output ripple analysis measures the peak-to-peak extremes of the output DC signal of your power supply. It also measures AC-RMS of the output DC signal. The AC-RMS measurement is the same as standard deviation (σ), which is a commonly-used measurement to characterize random noise. Output ripple is typically dominated by switching noise, but can also include other random noise and signal coupling from various sources in your system. This measurement basically measures the quality of the power supply s voltage regulation and filtering to reject switching noise, as well as other noise/interference sources. 1. If using Keysight s SMPS training kit, set the S2 load switch to the ON position (high-load, maximum current). 2. Press the [Analyze] front panel key; then select the Power Application under the Features softkey. 3. Press the Analysis softkey; then select the Output Ripple measurement. 4. Press the Signals softkey. 5. Connect a standard 10:1 passive voltage probe between the scope s channel-3 input to the DC output signal and ground as illustrated in the on-screen connections diagram shown in Figure 27. If using Keysight s SMPS training kit, connect the 10:1 passive probe from channel-3 to TP5 (probe tip grabber) and TP6 (ground). 6. Press the Voltage softkey; then change setting to 3 (measured source will be channel-3). 7. Press the AutoSetup softkey; then press Apply. Your scope s display should now look similar to Figure 28. When AutoSetup is pressed the scope AC couples the input signal on the selected channel to eliminate the DC component, and then optimally scales the selected input channel s vertical setting to display just the output ripple of the DC output signal. When Apply is pressed, the scope measures the output ripple of the DC output signal in terms of peak-to-peak voltage, as well as AC-RMS (s). Note that for a well-regulated power supply that generates a very low level of output ripple, a 1:1 passive probe such as Keysight s N2870A or D, may be required for this measurement. To better measure high frequency components of output ripple, Keysight suggests using the N7020A 2-GHz active probe rail probe. Whichever probe you choose to use to measure output ripple, note that shorter ground connections will produce more accurate measurement results. To learn more about output ripple measurements, download the application note titled, Making Your Best Power Integrity Measurements listed at the end of this document. Figure 27. Output ripple measurement connections diagram. Figure 28. Output ripple measurement on the DC output signal.

17 17 Keysight How to Make the Best Switch Mode Power Supply Measurements - Application Note Turn-on/Turn-off Analysis Turn-on analysis measures the time from when power is initially switched on until the dc output reaches 90% (default setting) of its expected steady-state level. Turn-off analysis measures the time from when power is switched off until the DC output decays to 10% (default setting) of the steady-state on level. 1. If using Keysight s SMPS training kit, set the S2 load switch to the ON position (high-load, maximum current). 2. Press the [Analyze] front panel key; then select the Power Application under the Features softkey. 3. Press the Analysis softkey; then select the Turn On/ Turn Off measurement. 4. Press the Signals softkey. 5. Connect the N2790A differential active probe between the scope s channel-1 input to Line (red lead) and Neutral (black lead) as illustrated in the on-screen connections diagram shown in Figure 29. If using Keysight s power training kit, connect to TP2 (red lead) and TP1 (black lead). 6. Connect a standard 10:1 passive voltage probe between the scope s channel-3 input to the DC output of your DUT. If using Keysight s SMPS training kit, connect to TP5 (probe tip grabber) and TP6 (ground). 7. Press the Input V softkey; then set to 1 (channel-1). 8. Press the Output V softkey; then set to 3 (channel-3). Figure 29. Turn-on/turn-off connections diagram in the Signals menu. Note that the default settings for duration, Max Vin, and Steady Vout have been optimized for Keysight s SMPS training kit. If testing a different power supply, then enter the appropriate values for Max Vin (peak-to-peak) and Steady Vout. Also, you may want to begin your test with the default duration setting of 500 ms, which will determine the scope s timebase setting for this measurement. 9. Press the [Back] front panel key to return to the previous menu. 10. Press the Apply softkey to begin the test; then follow on-screen instructions, which are repeated below. 11. Turn OFF power; then press Next. 12. Turn ON power; then press Next. Figure 30. Turn-on test using a default duration time setting of 500 ms. If the turn-on time of your device under test is less than 400 ms, then your scope s display should look similar to Figure 30. If the initial duration setting is too low, then the measurement cannot be performed. If this is the case, then set the duration time to a higher value as indicated by the on-screen message; then try again. Alternatively, to improve the timing resolution of this measurement you may want to set the duration time to a lower value. After determining the best duration time for this measurement, repeat the test several times to determine bestand worst-case turn-on time. Figure 31 shows this same test with the duration time set to 50 ms. AC input DC output Figure 31. Turn-on time test using a duration time setting of 50 ms to improve measurement resolution and accuracy.

18 18 Keysight How to Make the Best Switch Mode Power Supply Measurements - Application Note Turn-on/Turn-off Analysis (Continued) Let s now perform a turn-off time measurement. 13. Continuing from the previous Turn On time measurement, press the Test softkey; then change from Turn On to Turn Off. 14. Press the Signals softkey; then note the default settings of duration, Max Vin, and Steady Vout. These setting have been optimized for Keysight s SMPS training kit used in this demo procedure. If testing a different power supply, Max Vin and Steady Vout should already be set to the correct values based on the previous Turn On test. But the default duration time setting of 1.00 second may not be a valid setting for your power supply. 15. Press the [Back] front panel key to return to the previous menu; then press the Apply softkey and follow the step-by-step on-screen instructions, which are repeated below. 16. Turn ON power; then press Next. 17. Turn OFF power; then press Next. Figure 32. Turn-off time test. AC input DC output Your scope s display should now look similar to Figure 32. Repeat this measurement several times to determine the best- and worst-case turn-off times. Note that you may need to adjust the Duration time setting for this test. If using Keysight s SMPS training kit for this measurement, try setting the S2 switch to the OFF position (low-load, minimum current) and then repeat the Turn Off time measurement. You will probably need to increase the duration time to 2.0 seconds with the S2 load switch in the OFF position. With a decreased output load, the output will decay more slowly when power is turned off.

19 19 Keysight How to Make the Best Switch Mode Power Supply Measurements - Application Note Transient Response Analysis Transient response analysis measures the time for the output DC voltage to settle within a user-set percentage of the expected output level after a sudden change in output load (increase or decrease in output current). Before setting up this measurement, you must first determine what the approximate output current levels are for various load conditions. 1. If using Keysight s SMPS training kit, set the S2 load switch to the OFF position (low-load, minimum current). 2. Press the [Analyze] front panel key; then select the Power Application under the Features softkey. 3. Press the Analysis softkey; then select the Transient Response measurement. 4. Press the Signals softkey. 5. Connect the current probe between the scope s channel-2 input to the output current signal on your device under test as illustrated in the on-screen connections diagram shown in Figure 33. If using Keysight s SMPS training kit, connect to the J3 current loop in the direction indicated on the demo board. Note that the current-direction arrow on the current probe should be facing down towards your bench. 6. Connect a standard 10:1 passive voltage probe between the scope s channel-3 input to the output DC voltage signal. If using Keysight s power demo board, connect to TP5 (probe tip grabber) and TP6 (ground). 7. Press the Voltage softkey; then select 3 (channel-3). 8. Press the Current softkey; then select 2 (channel-2). 9. Set the Steady Vout for your DUT. If using Keysight s SMPS training kit, the default level of 12.0 V is appropriate. 10. Press the [Back] front panel to return to the previous menu. 11. Press the Settings softkey; then set Initial I to the approximate low-load output current level (assuming that the S2 load switch is in the OFF position (low- load/low current). Also set the New I to the approximate high-load output current level. If using Keysight s SMPS training kit, you can use the default settings of 60 ma and 215 ma respectively. 12. Press the [Back] front panel key to return to the previous menu. 13. Press the Apply softkey; then follow the on-screen stepby-step instructions, which are repeated below. 14. Increase the output load (higher current). If using Keysight s SMPS training kit, move the S2 load switch to the ON position (away from the J3 current loop). 15. Press Next. Figure 33. Transient response connections diagram in the Signals menu. V-out transient I-out increase Figure 34. Transient response settling time measurement when output load increases (output current suddenly increases). Your scope s display should now look similar to Figure 34. When output current measured by the channel-2 current probe (green trace) suddenly increases, output voltage (blue trace) momentarily drops and then settles back to within 10% of the expected output voltage level in approximately 12 ns.

20 20 Keysight How to Make the Best Switch Mode Power Supply Measurements - Application Note Transient Response Analysis (Continued) Let s now perform this transient response settling time measurement for when the output load decreases (lower output current). 16. If using Keysight s SMPS training kit, make sure that the S2 load switch is now in the ON position (away from the J3 current loop). This will put the training kit into an initial high-load (high output current) condition. 17. Press the Settings softkey; if using Keysight s SMPS training kit, then set Initial I to approximately 200 ma, and New I to approximately 60 ma. If making transient response measurements on your own power supply, then set to appropriate values. 18. Press the [Back] front panel key to return to the previous menu. 19. Press the Apply softkey; then follow the on-screen stepby-step instructions, which are repeated below. 20. Decrease the output load (lower current). If using Keysight s SMPS training kit, move the S2 load switch to the OFF position (towards the J3 current loop). 21. Press Next. V-out transient I-out decrease Figure 35. Transient response settling time measurement when output load decreases (output current suddenly decreases). Your scope s display should now look similar to Figure 35. When output current (green trace) suddenly drops, output voltage (blue trace) momentarily increases and then settles to within 10% of the expected output voltage level in approximately 21 ns. Note the transient response settling time measurement for both increasing and decreasing load conditions should be repeated several times in order to characterize best- and worst-case settling times.

21 21 Keysight How to Make the Best Switch Mode Power Supply Measurements - Application Note PSRR Analysis Power supply rejection ratio (PSRR) analysis, sometimes referred to as power supply ripple rejection, measures a voltage regulator s ability to reject coupling at various swept frequencies. Although this analysis is often performed using a network analyzer, this measurement can be performed automatically with Keysight s InfiniiVision X-Series oscilloscope using the scope s built-in waveform generator. To perform this measurement you must connect the output of the scope s WaveGen (Gen Out 1 only on the 4000 or 6000 X-Series) to the DC-to-DC converter s DC input via a summing network, such as the PicoTest J2120A line injector. PSRR is logarithmically plotted versus frequency as 20Log(Vi/Vo), which is the inverse of a standard Bode plot. Although a standard 10:1 passive voltage probe is typically used to measure V-in, because of the limited dynamic of an oscilloscope, a 1:1 passive voltage probe such as Keysight s N2870A or 10070D is recommended for measuring V-out. Note that this measurement cannot be demonstrated using Keysight s SMPS training kit. 1. Press the [Analyze] front panel key. 2. Press the Analysis softkey; then select the Power Supply Rejection Ratio (PSRR) measurement. 3. Press the Setup & Apply sofkey. 4. To view the connections diagram shown in Figure 36, touch the blue? symbol; then make the following connections: a. Connect the WaveGen s output (Gen Out 1 only if using a Keysight 4000 X-Series scope) to the modulation input of your AC + DC summing device, such as the Picotest J2120A. b. Connect your DC input to the DC input of the summing device (J2120A). c. Connect the modulated AC + DC output of the summing device (J2120) to the DC input of the DC-to- DC converter under test. d. Connect a standard 10:1 passive probe between the scope s channel-1 input to the DC input of the DC-to- DC converter under test. e. Connect a 1:1 passive probe (recommended) between the scope s channel-2 input to the DC output of the DC-to-DC converter under test. 5. Close the connections diagram dialog box. 6. In the Setup & Apply menu, establish desired test parameters (start frequency, stop frequency, test amplitude, etc.). 7. Touch Apply. Figure 36. PSRR connections diagram in the Signals menu. Figure 37 shows the results a PSRR measurement after the swept test has completed. In this example we tested from 100 Hz to 20 MHz. The blue trace is the PSRR plotted in db. After the measurement has completed, you can use the scope s tracking markers to measure the amount of rejection at various frequencies. In this test, we measure approximately 99 db of rejection at 37 khz. To learn more about the PSRR measurements, download the application note listed at the end of this document. Figure 37. Power supply rejection ratio (PSSR) measurement using the scope s built-in waveform generator as an automatic swept source.

22 22 Keysight How to Make the Best Switch Mode Power Supply Measurements - Application Note Control Loop Response (Bode Gain and Phase) Analysis A Control Loop Response measurement can be used to characterize the stability of your power supply. A vector network analyzer or a frequency response analyzer is typically used for this measurement. However, this measurement can also be performed using a Keysight 3000T or 4000A X-Series oscilloscope. This measurement basically provides a gain and phase plot versus frequency. Performing this measurement requires that you insert a low-value shunt/injection resistor (typically 5-Ω to 50-Ω) into the feedback network of your power supply. You then inject an error signal (using an isolation transformer) across the shunt resistor and measure the voltage on each side of the resistor across a band of frequencies. Note that this measurement cannot be demonstrated using Keysight s SMPS training kit. 1. Press the [Analyze] front panel key. 2. Press the Analysis softkey; then select the Control Loop Response (Bode) measurement. 3. Press the Setup & Apply sofkey. 4. To view the connections diagram shown in Figure 38, touch the blue? symbol; then make the following connections: a. Connect the WaveGen s output (Gen Out 1 only if using a Keysight 4000 X-Series scope) to the input of your isolation transformer, such as the Picotest J2101A Injection Transformer. b. Connect the output of your isolation transformer (J2101A) across the injection/shunt resistor that has been inserted into the feedback path of your power supply. c. Connect a 1:1 passive probe (recommended) between the scope s channel-1 input to the top of the feedback path. d. Connect a 1:1 passive probe (recommended) between the scope s channel-2 input to the DC output of your power supply. 5. Close the connections diagram dialog box. 6. In the Setup & Apply menu, establish desired test parameters (start frequency, stop frequency, test amplitude, etc.). 7. Touch Apply. Figure 38. Connections diagram for a Control Loop Response measurement. Figure 39. Gain and phase plot with automatic phase margin and gain margin measurements. Figure 39 shows the gain plot in blue and phase plot in pink. At the completion of this test, the scope also automatically measures the phase margin (PM) and the gain margin (GM). These are indicators of the stability of the power supply. To learn more about control loop response (Bode), refer the application note listed at the end of this document.

23 23 Keysight How to Make the Best Switch Mode Power Supply Measurements - Application Note Efficiency Analysis Efficiency analysis measures the input real power and output power in order to compute the efficiency of your power supply (Efficiency = Pwr(out)/Pwr(in) x 100). Performing this measurement requires three active probes that require power (high-voltage differential active probe plus two active current probes). If two current probes are not available, then this measurement can be performed in a 2-step process: measure input power, measure output power, then compute efficiency. But let s first show how to perform this measurement using two current probes in just one step. 1. If using Keysight s SMPS training kit, set the S2 load switch to the ON position (high-load, maximum current). 2. Press the [Analyze] front panel key; then select the Power Application under the Features softkey. 3. Press the Analysis softkey; then select the Efficiency measurement. 4. Press the Signals softkey; then scroll to the bottom of the connections diagram. 5. Connect the N2790A high-voltage differential active probe between the scope s channel-1 input to Line (red lead) and Neutral (black lead) as illustrated in the onscreen connections diagram shown in Figure 40. If using Keysight s SMPS training kit, connect to TP2 (red lead) and TP1 (black lead). 6. Connect a current probe between the scope s channel-2 input to the input Line current loop. If using Keysight s SMPS training kit, connect to the J1 current loop in the direction indicated. 7. Connect a standard 10:1 passive voltage probe between the scope s channel-3 input to the DC output signal. If using Keysight s power demo board, connect to TP5 (probe tip grabber) and TP6 (ground). 8. If using a Keysight 4000 and 6000 X-Series oscilloscope, connect another current probe between the scope s channel-4 input to the output DC current loop. If using Keysight s SMPS training kit, connect to the J3 current loop in the direction indicated. If using a Keysight 3000T X-Series oscilloscope, you must use a second current probe that receives power from an external source since this scope can only power two active probes. 9. Press the Input V softkey; then set to 1 (channel-1). 10. Press the Input I softkey; then set to 2 (channel-2). 11. Press the Output V softkey; then set to 3 (channel-3). 12. Press the Output I softkey; then set to 4 (channel-4). 13. Press the AutoSetup softkey; then press Apply. Your scope s display should now look similar to Figure 41. Keysight s SMPS training kit power supply measured an approximate efficiency of just 76% in this example. Figure 40. Efficiency measurement connections diagram in the Signals menu. Figure 41. Power efficiency measurement results. If two current probes are not available for this measurement, then follow the above procedure to measure the input power with your current probe connected between channel-2 and the input current loop. Connect nothing to the scope s channel-4 input. After performing the above procedure, take note of the Input power results, but ignore the output power and efficiency measurements. Next, move the current probe that is connected to the scope s channel-2 input to channel-4. Then connect it to the output DC current loop (J3 if using Keysight s SMPS training kit). Manually reposition and rescale channel-3 (V out) and channel-4 until both DC traces can be observed on-screen. Now take note of the output power results. Efficiency = Pwr-out/Pwr-in x 100%.

24 24 Keysight How to Make the Best Switch Mode Power Supply Measurements - Application Note Conclusion Each of these measurements is important for testing and characterizing your switch mode power supplies. Click here to learn more about probes, oscilloscopes, and software for testing your SMPS. Related Literature Publication title InfiniiVision Oscilloscope Probes and Accessories - Selection Guide N2790A 100 MHz, N2791A 25 MHz and N2891A 70 MHz High-voltage Differential Probes - Data Sheet N2870A Series Passive Probes and Accessories - Data Sheet DSOX3PWR/DSOX4PWR/DSOX6PWR Power Measurement Options - Data Sheet InfiniiVision 4000 X-Series Oscilloscopes - Data Sheet InfiniiVision 6000 X-Series Oscilloscopes - Data Sheet InfiniiVision 3000T X-Series Oscilloscopes - Data Sheet Making Your Best Power Integrity Measurements - Application Note Power Supply Rejection Ratio (PSRR) Measurements - Application Note Power Supply Control Loop Response (Bode Plot) Measurements - Application Note Publication number EN EN EN EN EN EN EN EN EN EN To download these documents, insert the publication number in the URL: Product website For the most up-to-date and complete application and product information, please visit our product website at: Keysight Oscilloscopes Multiple form factors from 20 MHz to > 90 GHz Industry leading specs Powerful applications

25 25 Keysight How to Make the Best Switch Mode Power Supply Measurements - Application Note AdvancedTCA Extensions for Instrumentation and Test (AXIe) is an open standard that extends the AdvancedTCA for general purpose and semiconductor test. The business that became Keysight was a founding member of the AXIe consortium. ATCA, AdvancedTCA, and the ATCA logo are registered US trademarks of the PCI Industrial Computer Manufacturers Group. LAN extensions for Instruments puts the power of Ethernet and the Web inside your test systems. The business that became Keysight was a founding member of the LXI consortium. PCI extensions for Instrumentation (PXI) modular instrumentation delivers a rugged, PC-based high-performance measurement and automation system.

26 26 Keysight How to Make the Best Switch Mode Power Supply Measurements - Application Note Evolving Since 1939 Our unique combination of hardware, software, services, and people can help you reach your next breakthrough. We are unlocking the future of technology. From Hewlett-Packard to Agilent to Keysight. For more information on Keysight Technologies products, applications or services, please contact your local Keysight office. The complete list is available at: Americas Canada (877) Brazil Mexico United States (800) mykeysight A personalized view into the information most relevant to you. Register your products to get up-to-date product information and find warranty information. Keysight Services Keysight Services can help from acquisition to renewal across your instrument s lifecycle. Our comprehensive service offerings onestop calibration, repair, asset management, technology refresh, consulting, training and more helps you improve product quality and lower costs. Keysight Assurance Plans Up to ten years of protection and no budgetary surprises to ensure your instruments are operating to specification, so you can rely on accurate measurements. Keysight Channel Partners Get the best of both worlds: Keysight s measurement expertise and product breadth, combined with channel partner convenience. Asia Pacific Australia China Hong Kong India Japan 0120 (421) 345 Korea Malaysia Singapore Taiwan Other AP Countries (65) Europe & Middle East Austria Belgium Finland France Germany Ireland Israel Italy Luxembourg Netherlands Russia Spain Sweden Switzerland Opt. 1 (DE) Opt. 2 (FR) Opt. 3 (IT) United Kingdom For other unlisted countries: (BP ) DEKRA Certified ISO9001 Quality Management System Keysight Technologies, Inc. DEKRA Certified ISO 9001:2015 Quality Management System This information is subject to change without notice. Keysight Technologies, 2017 Published in USA, December 7, EN

Switch Mode Power Supply Measurements

Switch Mode Power Supply Measurements Switch Mode Power Supply Measurements Application Note Using an Agilent InfiniiVision 3000/4000 X-Series Oscilloscope With the Power Measurements Option Introduction Agilent s 3000 and 4000 X-Series oscilloscopes

More information

Keysight Technologies DSOX3PWR/DSOX4PWR/DSOX6PWR Power Measurement Options

Keysight Technologies DSOX3PWR/DSOX4PWR/DSOX6PWR Power Measurement Options Keysight Technologies DSOX3PWR/DSOX4PWR/DSOX6PWR Power Measurement Options Data Sheet For InfiniiVision 3000, 4000 and 6000 X-Series Oscilloscopes 02 Keysight DSOX3PWR/DSOX4PWR/DSOX6PWR Power Measurement

More information

Keysight DSOX3PWR/DSOX4PWR/DSOX6PWR Power Measurement Options

Keysight DSOX3PWR/DSOX4PWR/DSOX6PWR Power Measurement Options Keysight DSOX3PWR/DSOX4PWR/DSOX6PWR Power Measurement Options For InfiniiVision 3000, 4000 and 6000 X-Series Oscilloscopes Data Sheet 02 Keysight DSOX3PWR/DSOX4PWR/DSOX6PWR Power Measurement Options -

More information

Keysight Technologies Alliance for Wireless Power (A4WP) Measurements Using an Oscilloscope (Part 3)

Keysight Technologies Alliance for Wireless Power (A4WP) Measurements Using an Oscilloscope (Part 3) Keysight Technologies Alliance for Wireless Power (A4WP) Measurements Using an Oscilloscope (Part 3) Power and Efficiency Measurements Application Note Introduction One of the primary instruments used

More information

Today most of engineers use oscilloscope as the preferred measurement tool of choice when it comes to debugging and analyzing switching power

Today most of engineers use oscilloscope as the preferred measurement tool of choice when it comes to debugging and analyzing switching power Today most of engineers use oscilloscope as the preferred measurement tool of choice when it comes to debugging and analyzing switching power supplies. In this session we will learn about some basics of

More information

Keysight Technologies DSOX3PWR/DSOX4PWR/DSOX6PWR Power Measurement Options

Keysight Technologies DSOX3PWR/DSOX4PWR/DSOX6PWR Power Measurement Options Keysight Technologies DSOX3PWR/DSOX4PWR/DSOX6PWR Power Measurement Options Data Sheet For InfiniiVision 3000, 4000 and 6000 X-Series Oscilloscopes Achieve cost-effective analysis of your switching mode

More information

Agilent U1881A and U1882A Power Measurement Application for Agilent InfiniiVision and Infiniium Oscilloscopes

Agilent U1881A and U1882A Power Measurement Application for Agilent InfiniiVision and Infiniium Oscilloscopes Agilent U1881A and U1882A Power Measurement Application for Agilent InfiniiVision and Infiniium Oscilloscopes Data Sheet Fast, automatic and reliable characterization of switching mode power devices Today

More information

Measurement and Analysis for Switchmode Power Design

Measurement and Analysis for Switchmode Power Design Measurement and Analysis for Switchmode Power Design Switched Mode Power Supply Measurements AC Input Power measurements Safe operating area Harmonics and compliance Efficiency Switching Transistor Losses

More information

U1881A and U1882A Power Measurement Application for InfiniiVision and Infiniium Oscilloscopes

U1881A and U1882A Power Measurement Application for InfiniiVision and Infiniium Oscilloscopes U1881A and U1882A Power Measurement Application for InfiniiVision and Infiniium Oscilloscopes Data Sheet Fast, automatic and reliable characterization of switching mode power devices Today s power supply

More information

U1881A and U1882A Power Measurement Application for InfiniiVision and Infiniium Oscilloscopes

U1881A and U1882A Power Measurement Application for InfiniiVision and Infiniium Oscilloscopes U1881A and U1882A Power Measurement Application for InfiniiVision and Infiniium Oscilloscopes Data Sheet Fast, automatic and reliable characterization of switching mode power devices Today s power supply

More information

Switched Mode Power Supply Measurements

Switched Mode Power Supply Measurements Power Analysis 1 Switched Mode Power Supply Measurements AC Input Power measurements Safe operating area Harmonics and compliance Efficiency Switching Transistor Losses Measurement challenges Transformer

More information

ME 365 EXPERIMENT 1 FAMILIARIZATION WITH COMMONLY USED INSTRUMENTATION

ME 365 EXPERIMENT 1 FAMILIARIZATION WITH COMMONLY USED INSTRUMENTATION Objectives: ME 365 EXPERIMENT 1 FAMILIARIZATION WITH COMMONLY USED INSTRUMENTATION The primary goal of this laboratory is to study the operation and limitations of several commonly used pieces of instrumentation:

More information

Measuring Power Supply Switching Loss with an Oscilloscope

Measuring Power Supply Switching Loss with an Oscilloscope Measuring Power Supply Switching Loss with an Oscilloscope Our thanks to Tektronix for allowing us to reprint the following. Ideally, the switching device is either on or off like a light switch, and instantaneously

More information

Keysight U1882B Measurement Application for Infiniium Oscilloscopes. Data Sheet

Keysight U1882B Measurement Application for Infiniium Oscilloscopes. Data Sheet Keysight U1882B Measurement Application for Infiniium Oscilloscopes Data Sheet 02 Keysight U1882B Measurement Application for Infiniium Oscilloscopes - Data Sheet Fast, Automatic and Reliable Characterization

More information

Power Measurements for Switch-Mode Power Supplies SAVE Verona 2011

Power Measurements for Switch-Mode Power Supplies SAVE Verona 2011 Power Measurements for Switch-Mode Power Supplies SAVE Verona 2011 Agenda Power measurements tools Switch-mode power supplies Automated power measurements Summary Reference information 2 Switch-Mode Power

More information

Introduction to Oscilloscopes Instructor s Guide

Introduction to Oscilloscopes Instructor s Guide Introduction to Oscilloscopes A collection of lab exercises to introduce you to the basic controls of a digital oscilloscope in order to make common electronic measurements. Revision 1.0 Page 1 of 25 Copyright

More information

Oscilloscope Fundamentals. For Electrical Engineering and Physics Undergraduate Students

Oscilloscope Fundamentals. For Electrical Engineering and Physics Undergraduate Students Oscilloscope Fundamentals For Electrical Engineering and Physics Undergraduate Students Agenda What is an oscilloscope? Probing basics (low-frequency model) Making voltage and timing measurements Properly

More information

DPO3PWR, MDO3PWR and DPO4PWR Power Analysis Modules ZZZ User Manual

DPO3PWR, MDO3PWR and DPO4PWR Power Analysis Modules ZZZ User Manual x DPO3PWR, MDO3PWR and DPO4PWR Power Analysis Modules ZZZ User Manual *P071263101* 071-2631-01 xx DPO3PWR, MDO3PWR and DPO4PWR Power Analysis Modules ZZZ User Manual www.tektronix.com 071-2631-01 Copyright

More information

UCE-DSO212 DIGITAL OSCILLOSCOPE USER MANUAL. UCORE ELECTRONICS

UCE-DSO212 DIGITAL OSCILLOSCOPE USER MANUAL. UCORE ELECTRONICS UCE-DSO212 DIGITAL OSCILLOSCOPE USER MANUAL UCORE ELECTRONICS www.ucore-electronics.com 2017 Contents 1. Introduction... 2 2. Turn on or turn off... 3 3. Oscilloscope Mode... 4 3.1. Display Description...

More information

Probe Considerations for Low Voltage Measurements such as Ripple

Probe Considerations for Low Voltage Measurements such as Ripple Probe Considerations for Low Voltage Measurements such as Ripple Our thanks to Tektronix for allowing us to reprint the following article. Figure 1. 2X Probe (CH1) and 10X Probe (CH2) Lowest System Vertical

More information

The University of Jordan Mechatronics Engineering Department Electronics Lab.( ) Experiment 1: Lab Equipment Familiarization

The University of Jordan Mechatronics Engineering Department Electronics Lab.( ) Experiment 1: Lab Equipment Familiarization The University of Jordan Mechatronics Engineering Department Electronics Lab.(0908322) Experiment 1: Lab Equipment Familiarization Objectives To be familiar with the main blocks of the oscilloscope and

More information

Multiple Instrument Station Module

Multiple Instrument Station Module Multiple Instrument Station Module Digital Storage Oscilloscope Vertical Channels Sampling rate Bandwidth Coupling Input impedance Vertical sensitivity Vertical resolution Max. input voltage Horizontal

More information

Combinational logic: Breadboard adders

Combinational logic: Breadboard adders ! ENEE 245: Digital Circuits & Systems Lab Lab 1 Combinational logic: Breadboard adders ENEE 245: Digital Circuits and Systems Laboratory Lab 1 Objectives The objectives of this laboratory are the following:

More information

LAB I. INTRODUCTION TO LAB EQUIPMENT

LAB I. INTRODUCTION TO LAB EQUIPMENT LAB I. INTRODUCTION TO LAB EQUIPMENT 1. OBJECTIVE In this lab you will learn how to properly operate the basic bench equipment used for characterizing active devices: 1. Oscilloscope (Keysight DSOX 1102A),

More information

UCE-DSO210 DIGITAL OSCILLOSCOPE USER MANUAL. FATIH GENÇ UCORE ELECTRONICS REV1

UCE-DSO210 DIGITAL OSCILLOSCOPE USER MANUAL. FATIH GENÇ UCORE ELECTRONICS REV1 UCE-DSO210 DIGITAL OSCILLOSCOPE USER MANUAL FATIH GENÇ UCORE ELECTRONICS www.ucore-electronics.com 2017 - REV1 Contents 1. Introduction... 2 2. Turn on or turn off... 3 3. Oscilloscope Mode... 3 3.1. Display

More information

LAB I. INTRODUCTION TO LAB EQUIPMENT

LAB I. INTRODUCTION TO LAB EQUIPMENT 1. OBJECTIVE LAB I. INTRODUCTION TO LAB EQUIPMENT In this lab you will learn how to properly operate the oscilloscope Agilent MSO6032A, the Keithley Source Measure Unit (SMU) 2430, the function generator

More information

EXPERIMENT NUMBER 2 BASIC OSCILLOSCOPE OPERATIONS

EXPERIMENT NUMBER 2 BASIC OSCILLOSCOPE OPERATIONS 1 EXPERIMENT NUMBER 2 BASIC OSCILLOSCOPE OPERATIONS The oscilloscope is the most versatile and most important tool in this lab and is probably the best tool an electrical engineer uses. This outline guides

More information

External Source Control

External Source Control External Source Control X-Series Signal Analyzers Option ESC DEMO GUIDE Introduction External source control for X-Series signal analyzers (Option ESC) allows the Keysight PXA, MXA, EXA, and CXA to control

More information

P a g e 1 ST985. TDR Cable Analyzer Instruction Manual. Analog Arts Inc.

P a g e 1 ST985. TDR Cable Analyzer Instruction Manual. Analog Arts Inc. P a g e 1 ST985 TDR Cable Analyzer Instruction Manual Analog Arts Inc. www.analogarts.com P a g e 2 Contents Software Installation... 4 Specifications... 4 Handling Precautions... 4 Operation Instruction...

More information

TAKE THE MYSTERY OUT OF PROBING. 7 Common Oscilloscope Probing Pitfalls to Avoid

TAKE THE MYSTERY OUT OF PROBING. 7 Common Oscilloscope Probing Pitfalls to Avoid TAKE THE MYSTERY OUT OF PROBING 7 Common Oscilloscope Probing Pitfalls to Avoid Introduction Understanding common probing pitfalls and how to avoid them is crucial in making better measurements. In an

More information

Lab Reference Manual. ECEN 326 Electronic Circuits. Texas A&M University Department of Electrical and Computer Engineering

Lab Reference Manual. ECEN 326 Electronic Circuits. Texas A&M University Department of Electrical and Computer Engineering Lab Reference Manual ECEN 326 Electronic Circuits Texas A&M University Department of Electrical and Computer Engineering Contents 1. Circuit Analysis in PSpice 3 1.1 Transient and DC Analysis 3 1.2 Measuring

More information

Transient Current Measurement for Advance Materials & Devices

Transient Current Measurement for Advance Materials & Devices & Devices 8 May 2017 Brian YEO Application Engineer Keysight Technologies Agenda 2 High speed data acquisition basics Challenges & solutions for transient current measurement. Considerations when making

More information

Getting Started. MSO/DPO Series Oscilloscopes. Basic Concepts

Getting Started. MSO/DPO Series Oscilloscopes. Basic Concepts Getting Started MSO/DPO Series Oscilloscopes Basic Concepts 001-1523-00 Getting Started 1.1 Getting Started What is an oscilloscope? An oscilloscope is a device that draws a graph of an electrical signal.

More information

2-channel models. 4-channel models

2-channel models. 4-channel models 2-channel models DSO1052B DSO1072B DSO1102B DSO1152B 50 MHz 70 MHz 100 MHz 150 MHz 4-channel models DSO1004A DSO1014A DSO1024A 60 MHz 100 MHz 200 MHz DSO1022A 200 MHz Segment 1 Segment 2 Segment 1000

More information

AP034-OM-E Rev D ISSUED: January 2000 ²

AP034-OM-E Rev D ISSUED: January 2000 ² 3HUIRUPDQFH9HULILFDWLRQ 3HUIRUPDQFH9HULILFDWLRQ This procedure can be used to verify the warranted characteristics of the AP034 Active Differential Probe. The recommended calibration interval for the model

More information

Lab #5 Steady State Power Analysis

Lab #5 Steady State Power Analysis Lab #5 Steady State Power Analysis Steady state power analysis refers to the power analysis of circuits that have one or more sinusoid stimuli. This lab covers the concepts of RMS voltage, maximum power

More information

Active Power Factor Correction Verification Measurements with an Oscilloscope APPLICATION NOTE

Active Power Factor Correction Verification Measurements with an Oscilloscope APPLICATION NOTE Active Power Factor Correction Verification Measurements with an Oscilloscope AC-DC power supplies, especially those designed to comply with IEC61000-3-2 or ENERGY STAR standards, often include some form

More information

U1604A Handheld Oscilloscopes, 40 MHz

U1604A Handheld Oscilloscopes, 40 MHz Products & Services Technical Support Buy Industries About Agilent Search: All Test & Measurement Go United States Home >... > Oscilloscopes > U1600A Series handheld oscilloscopes (2 models) > U1604A Handheld

More information

LAB II. INTRODUCTION TO LAB EQUIPMENT

LAB II. INTRODUCTION TO LAB EQUIPMENT 1. OBJECTIVE LAB II. INTRODUCTION TO LAB EQUIPMENT In this lab you will learn how to properly operate the oscilloscope Keysight DSOX1102A, the Keithley Source Measure Unit (SMU) 2430, the function generator

More information

Jitter Analysis Techniques Using an Agilent Infiniium Oscilloscope

Jitter Analysis Techniques Using an Agilent Infiniium Oscilloscope Jitter Analysis Techniques Using an Agilent Infiniium Oscilloscope Product Note Table of Contents Introduction........................ 1 Jitter Fundamentals................. 1 Jitter Measurement Techniques......

More information

PXIe Contents. Required Software CALIBRATION PROCEDURE

PXIe Contents. Required Software CALIBRATION PROCEDURE CALIBRATION PROCEDURE PXIe-5160 This document contains the verification and adjustment procedures for the PXIe-5160. Refer to ni.com/calibration for more information about calibration solutions. Contents

More information

Measuring Power Supply Switching Loss with an Oscilloscope

Measuring Power Supply Switching Loss with an Oscilloscope Measuring Power Supply Switching Loss with an Oscilloscope Application Note Introduction With the demand for improving power efficiency and extending the operating time of battery-powered devices, the

More information

Keysight DSOXT3FRA/DSOX4FRA/DSOX6FRA Frequency Response Analyzer (FRA) Option

Keysight DSOXT3FRA/DSOX4FRA/DSOX6FRA Frequency Response Analyzer (FRA) Option Keysight DSOXT3FRA/DSOX4FRA/DSOX6FRA Frequency Response Analyzer (FRA) Option For Keysight 3000T, 4000A, and 6000A X-Series Oscilloscopes Data Sheet Introduction Frequency Response Analysis (FRA) is often

More information

Transient Data Acquisition System, TAS 4-40 Potential-free measurement of fast rise pulses:

Transient Data Acquisition System, TAS 4-40 Potential-free measurement of fast rise pulses: Transient Data Acquisition System, TAS 4-40 Potential-free measurement of fast rise pulses: High precision measurement of fast rising voltages and currents causes considerable problems in many spheres

More information

DSA-815 Demo Guide. Solution: The DSA 800 series of spectrum analyzers are packed with features.

DSA-815 Demo Guide. Solution: The DSA 800 series of spectrum analyzers are packed with features. FAQ Instrument Solution FAQ Solution Title DSA-815 Demo Guide Date:08.29.2012 Solution: The DSA 800 series of spectrum analyzers are packed with features. Spectrum analyzers are similar to oscilloscopes..

More information

EENG-201 Experiment # 4: Function Generator, Oscilloscope

EENG-201 Experiment # 4: Function Generator, Oscilloscope EENG-201 Experiment # 4: Function Generator, Oscilloscope I. Objectives Upon completion of this experiment, the student should be able to 1. To become familiar with the use of a function generator. 2.

More information

Spectrum Analyzer TEN MINUTE TUTORIAL

Spectrum Analyzer TEN MINUTE TUTORIAL Spectrum Analyzer TEN MINUTE TUTORIAL November 4, 2011 Summary The Spectrum Analyzer option allows users who are familiar with RF spectrum analyzers to start using the FFT with little or no concern about

More information

Measurement Bench. Accessories. Power supply. Wave form generator. Multimetre. Oscilloscope. Dr. L.Scucchia

Measurement Bench. Accessories. Power supply. Wave form generator. Multimetre. Oscilloscope. Dr. L.Scucchia Measurement Bench Accessories Power supply Wave form generator Multimetre Oscilloscope OSCILLOSCOPE Oscilloscope (1) The oscilloscope allows to display a voltage (vertical axis - Y axis) versus time (horizontal

More information

Multi-function Gain-Phase Analyzer (Frequency Response Analyzer) Model 2505

Multi-function Gain-Phase Analyzer (Frequency Response Analyzer) Model 2505 OTHER PRODUCTS.. Multi-function Gain-Phase Analyzer ( Response Analyzer) Model 2505 Standard Configurations Gain phase analyzer response analyzer Phase Angle Voltmeter (PAV) Fast dual channel wide-band

More information

Power Supply Measurement and Analysis with the MSO/DPO Series Oscilloscopes

Power Supply Measurement and Analysis with the MSO/DPO Series Oscilloscopes Power Supply Measurement and Analysis with the MSO/DPO Series Oscilloscopes Introduction Power supplies can be found in many different electronic devices, from children s toys to computers and office equipment

More information

UNIVERSITY OF CALIFORNIA, SANTA BARBARA Department of Electrical and Computer Engineering. ECE 2A & 2B Laboratory Equipment Information

UNIVERSITY OF CALIFORNIA, SANTA BARBARA Department of Electrical and Computer Engineering. ECE 2A & 2B Laboratory Equipment Information UNIVERSITY OF CALIFORNIA, SANTA BARBARA Department of Electrical and Computer Engineering ECE 2A & 2B Laboratory Equipment Information Table of Contents Digital Multi-Meter (DMM)... 1 Features... 1 Using

More information

User s Manual for Integrator Long Pulse ILP8 22AUG2016

User s Manual for Integrator Long Pulse ILP8 22AUG2016 User s Manual for Integrator Long Pulse ILP8 22AUG2016 Contents Specifications... 3 Packing List... 4 System Description... 5 RJ45 Channel Mapping... 8 Customization... 9 Channel-by-Channel Custom RC Times...

More information

Sampling and Reconstruction

Sampling and Reconstruction Experiment 10 Sampling and Reconstruction In this experiment we shall learn how an analog signal can be sampled in the time domain and then how the same samples can be used to reconstruct the original

More information

Oscilloscope Applications MREC Rural Energy Conference 2012 Prepared by Paul Ortmann, P.E. /

Oscilloscope Applications MREC Rural Energy Conference 2012 Prepared by Paul Ortmann, P.E. / MREC Rural Energy Conference 2012 Prepared by Paul Ortmann, P.E. / portmann@idahopower.com Introduction An oscilloscope can help the user get more detailed electrical measurements than can be obtained

More information

Agilent Technologies 3000 Series Oscilloscopes

Agilent Technologies 3000 Series Oscilloscopes Agilent Technologies 3000 Series Oscilloscopes Data Sheet Full-featured oscilloscopes for the smallest budgets Features: 60 to 200 MHz bandwidths 1 GSa/s maximum sample rate Large 15-cm (5.7-in) color

More information

Notes on Experiment #1

Notes on Experiment #1 Notes on Experiment #1 Bring graph paper (cm cm is best) From this week on, be sure to print a copy of each experiment and bring it with you to lab. There will not be any experiment copies available in

More information

U1571A Ni-MH Battery Pack for U1600A Handheld Oscilloscopes

U1571A Ni-MH Battery Pack for U1600A Handheld Oscilloscopes United States Home >... > Oscilloscope Accessories > U1600 Series Oscilloscope Accessories > U1571A Ni-MH Battery Pack for U1600A Handheld Oscilloscopes Key Specifications Features Ni-MH Battery Pack,

More information

Agilent N2740A Education Training Kit for 1000 Series Oscilloscopes

Agilent N2740A Education Training Kit for 1000 Series Oscilloscopes Agilent N2740A Education Training Kit for 1000 Series Oscilloscopes Lab Manual A Notices Agilent Technologies, Inc. 2008 No part of this manual may be reproduced in any form or by any means (including

More information

HP 16533A 1-GSa/s and HP 16534A 2-GSa/s Digitizing Oscilloscope

HP 16533A 1-GSa/s and HP 16534A 2-GSa/s Digitizing Oscilloscope User s Reference Publication Number 16534-97009 February 1999 For Safety Information, Warranties, and Regulatory Information, see the pages behind the Index Copyright Hewlett-Packard Company 1991 1999

More information

How to Get Clean DC Power FIND AND ELIMINATE NOISE

How to Get Clean DC Power FIND AND ELIMINATE NOISE How to Get Clean DC Power FIND AND ELIMINATE NOISE How to Get Clean DC Power FIND AND ELIMINATE NOISE Clean DC power is the backbone of stable systems. Without clean power, your device will experience

More information

User s Manual for Integrator Short Pulse ISP16 10JUN2016

User s Manual for Integrator Short Pulse ISP16 10JUN2016 User s Manual for Integrator Short Pulse ISP16 10JUN2016 Specifications Exceeding any of the Maximum Ratings and/or failing to follow any of the Warnings and/or Operating Instructions may result in damage

More information

Help Volume Agilent Technologies. All rights reserved. Instrument: Agilent Technologies 16533/34A Digitizing Oscilloscope

Help Volume Agilent Technologies. All rights reserved. Instrument: Agilent Technologies 16533/34A Digitizing Oscilloscope Help Volume 1992-2002 Agilent Technologies. All rights reserved. Instrument: Agilent Technologies 16533/34A Digitizing Oscilloscope Agilent Technologies 16533/34A Digitizing Oscilloscope The Agilent Technologies

More information

LABORATORY 4. Palomar College ENGR210 Spring 2017 ASSIGNED: 3/21/17

LABORATORY 4. Palomar College ENGR210 Spring 2017 ASSIGNED: 3/21/17 LABORATORY 4 ASSIGNED: 3/21/17 OBJECTIVE: The purpose of this lab is to evaluate the transient and steady-state circuit response of first order and second order circuits. MINIMUM EQUIPMENT LIST: You will

More information

B. Equipment. Advanced Lab

B. Equipment. Advanced Lab Advanced Lab Measuring Periodic Signals Using a Digital Oscilloscope A. Introduction and Background We will use a digital oscilloscope to characterize several different periodic voltage signals. We will

More information

Introduction to basic laboratory instruments

Introduction to basic laboratory instruments BEE 233 Laboratory-1 Introduction to basic laboratory instruments 1. Objectives To learn safety procedures in the laboratory. To learn how to use basic laboratory instruments: power supply, function generator,

More information

How to Setup a Real-time Oscilloscope to Measure Jitter

How to Setup a Real-time Oscilloscope to Measure Jitter TECHNICAL NOTE How to Setup a Real-time Oscilloscope to Measure Jitter by Gary Giust, PhD NOTE-3, Version 1 (February 16, 2016) Table of Contents Table of Contents... 1 Introduction... 2 Step 1 - Initialize

More information

ECE 2274 Lab 1 (Intro)

ECE 2274 Lab 1 (Intro) ECE 2274 Lab 1 (Intro) Richard Dumene: Spring 2018 Revised: Richard Cooper: Spring 2018 Forward (DO NOT TURN IN) The purpose of this lab course is to familiarize you with high-end lab equipment, and train

More information

Exercise 6. Range and Angle Tracking Performance (Radar-Dependent Errors) EXERCISE OBJECTIVE

Exercise 6. Range and Angle Tracking Performance (Radar-Dependent Errors) EXERCISE OBJECTIVE Exercise 6 Range and Angle Tracking Performance EXERCISE OBJECTIVE When you have completed this exercise, you will be familiar with the radardependent sources of error which limit range and angle tracking

More information

ET 304A Laboratory Tutorial-Circuitmaker For Transient and Frequency Analysis

ET 304A Laboratory Tutorial-Circuitmaker For Transient and Frequency Analysis ET 304A Laboratory Tutorial-Circuitmaker For Transient and Frequency Analysis All circuit simulation packages that use the Pspice engine allow users to do complex analysis that were once impossible to

More information

54645D. Mixed Signal Oscilloscope

54645D. Mixed Signal Oscilloscope 54645D Mixed Signal Oscilloscope Page 1 of 42 Instructions for the use of the 54645D Mixed Signal Oscilloscope This pamphlet is intended to give you (the student) an overview on the use of the 54645D Mixed

More information

Parts to be supplied by the student: Breadboard and wires IRLZ34N N-channel enhancement-mode power MOSFET transistor

Parts to be supplied by the student: Breadboard and wires IRLZ34N N-channel enhancement-mode power MOSFET transistor University of Utah Electrical & Computer Engineering Department ECE 1250 Lab 3 Electronic Speed Control and Pulse Width Modulation A. Stolp, 12/31/12 Rev. Objectives 1 Introduce the Oscilloscope and learn

More information

LeCroy PowerMeasure System. Get the Complete Picture

LeCroy PowerMeasure System. Get the Complete Picture LeCroy PowerMeasure System Get the Complete Picture Power Measurements Made Easy! COMPLETE PACKAGE! POWER DEVICE ANALYSIS: measures power device saturation voltage, instantaneous power loss, safe operating

More information

Basic Communication Laboratory Manual. Shimshon Levy&Harael Mualem

Basic Communication Laboratory Manual. Shimshon Levy&Harael Mualem Basic Communication Laboratory Manual Shimshon Levy&Harael Mualem September 2006 CONTENTS 1 The oscilloscope 2 1.1 Objectives... 2 1.2 Prelab... 2 1.3 Background Theory- Analog Oscilloscope...... 3 1.4

More information

Publication Number ATFxxB Series DDS FUNCTION WAVEFORM GENERATOR. User s Guide

Publication Number ATFxxB Series DDS FUNCTION WAVEFORM GENERATOR. User s Guide Publication Number 101201 ATFxxB Series DDS FUNCTION WAVEFORM GENERATOR User s Guide Introduction This user's guide is used for all models of ATFxxB series of DDS function generator. xx in the model number

More information

Fluke 192/196/199. MS 190 and MA 190. Users Manual Supplement

Fluke 192/196/199. MS 190 and MA 190. Users Manual Supplement Fluke 192/196/199 MS 190 and MA 190 Users Manual Supplement 4822 872 00979 April 2000, Rev.2, 9/00 2000 Fluke Corporation. All rights reserved. Printed in the Netherlands. All product names are trademarks

More information

Power Supply Rejection Ratio Measurement

Power Supply Rejection Ratio Measurement Power Supply Rejection Ratio Measurement Using the Bode 100 and the Picotest J2120A Line Injector www.telesplicing.com.tw +886-2-27053146 sales@telesplicing.com.tw Page 2 of 10 Table of Contents 1 EXECUTIVE

More information

Analog Arts SG985 SG884 SG834 SG814 Product Specifications [1]

Analog Arts SG985 SG884 SG834 SG814 Product Specifications [1] www.analogarts.com Analog Arts SG985 SG884 SG834 SG814 Product Specifications [1] 1. These models include: an oscilloscope, a spectrum analyzer, a data recorder, a frequency & phase meter, and an arbitrary

More information

Divide. MHz models) waveform record

Divide. MHz models) waveform record The 2550 series digital storage oscilloscopes provide high performance and value in 2-channel and 4-channel configurations. With bandwidth from 70 MHz to 300 MHz and 2 GSa/s sample rates, these oscilloscopes

More information

OSCILLOSCOPES. Oscilloscopes CS-5300 SERIES RS-232C OPTION OPTION CS-5370P/5370/5350 FEATURES OUTLINE. Photo: CS-5370P

OSCILLOSCOPES. Oscilloscopes CS-5300 SERIES RS-232C OPTION OPTION CS-5370P/5370/5350 FEATURES OUTLINE. Photo: CS-5370P Oscilloscopes 100MHz 2-Channel Programmable Oscilloscope ( With Digital Readout / Cursor) CS-5370P CS-5370 100MHz 3-Channel Oscilloscope ( With Digital Readout / Cursor) 50MHz 3-Channel Oscilloscope (

More information

Noise Measurements Using a Teledyne LeCroy Oscilloscope

Noise Measurements Using a Teledyne LeCroy Oscilloscope Noise Measurements Using a Teledyne LeCroy Oscilloscope TECHNICAL BRIEF January 9, 2013 Summary Random noise arises from every electronic component comprising your circuits. The analysis of random electrical

More information

EE 1210 Op Amps, Gain, and Signal Integrity Laboratory Project 6

EE 1210 Op Amps, Gain, and Signal Integrity Laboratory Project 6 Objective Information The purposes of this laboratory project are for the student to observe an inverting operational amplifier circuit, to demonstrate how the resistors in an operational amplifier circuit

More information

ECE3204 D2015 Lab 1. See suggested breadboard configuration on following page!

ECE3204 D2015 Lab 1. See suggested breadboard configuration on following page! ECE3204 D2015 Lab 1 The Operational Amplifier: Inverting and Non-inverting Gain Configurations Gain-Bandwidth Product Relationship Frequency Response Limitation Transfer Function Measurement DC Errors

More information

LeCroy 9304A, 9304AM Digital Oscilloscopes 200 MHz Bandwidth, 100 MS/s. Main Features

LeCroy 9304A, 9304AM Digital Oscilloscopes 200 MHz Bandwidth, 100 MS/s. Main Features LeCroy 9304A, 9304AM Digital Oscilloscopes 200 MHz Bandwidth, 100 MS/s Main Features Four Channels 50k and 200k Point Records DOS Compatible Floppy Disk, PCMCIA portable hard drive and Memory Card Options

More information

DC/DC Converter Stability Measurement

DC/DC Converter Stability Measurement Strongly supported by By Stephan Synkule, Lukas Heinzle & Florian Hämmerle 2018 by OMICRON Lab V3.3 Visit www.omicron-lab.com for more information. Contact support@omicron-lab.com for technical support.

More information

CHAPTER 6. Motor Driver

CHAPTER 6. Motor Driver CHAPTER 6 Motor Driver In this lab, we will construct the circuitry that your robot uses to drive its motors. However, before testing the motor circuit we will begin by making sure that you are able to

More information

HAMEG Modular System Series 8000

HAMEG Modular System Series 8000 HAMEG Modular System Series 8000 In many years of practical application the HAMEG Modular System Series 8000 has proven its value to the customer. The advantages of this Modular System have been demonstrated

More information

PXIe Contents CALIBRATION PROCEDURE

PXIe Contents CALIBRATION PROCEDURE CALIBRATION PROCEDURE PXIe-5632 This document contains the verification and adjustment procedures for the PXIe-5632 Vector Network Analyzer. Refer to ni.com/calibration for more information about calibration

More information

PGT313 Digital Communication Technology. Lab 3. Quadrature Phase Shift Keying (QPSK) and 8-Phase Shift Keying (8-PSK)

PGT313 Digital Communication Technology. Lab 3. Quadrature Phase Shift Keying (QPSK) and 8-Phase Shift Keying (8-PSK) PGT313 Digital Communication Technology Lab 3 Quadrature Phase Shift Keying (QPSK) and 8-Phase Shift Keying (8-PSK) Objectives i) To study the digitally modulated quadrature phase shift keying (QPSK) and

More information

PDN Probes. P2100A/P2101A Data Sheet. 1-Port and 2-Port 50 ohm Passive Probes

PDN Probes. P2100A/P2101A Data Sheet. 1-Port and 2-Port 50 ohm Passive Probes P2100A/P2101A Data Sheet PDN Probes 1-Port and 2-Port 50 ohm Passive Probes power integrity PDN impedance testing ripple PCB resonances transient step load stability and NISM noise TDT/TDR clock jitter

More information

Fourier Theory & Practice, Part II: Practice Operating the Agilent Series Scope with Measurement/Storage Module

Fourier Theory & Practice, Part II: Practice Operating the Agilent Series Scope with Measurement/Storage Module Fourier Theory & Practice, Part II: Practice Operating the Agilent 54600 Series Scope with Measurement/Storage Module By: Robert Witte Agilent Technologies Introduction: This product note provides a brief

More information

Lab #1 Lab Introduction

Lab #1 Lab Introduction Cir cuit s 212 Lab Lab #1 Lab Introduction Special Information for this Lab s Report Because this is a one-week lab, please hand in your lab report for this lab at the beginning of next week s lab. The

More information

Experiment #2: Introduction to Lab Equipment: Function Generator, Oscilloscope, and Multisim

Experiment #2: Introduction to Lab Equipment: Function Generator, Oscilloscope, and Multisim SCHOOL OF ENGINEERING AND APPLIED SCIENCE DEPARTMENT OF ELECTRICAL AND COMPUTER ENGINEERING ECE 2110: CIRCUIT THEORY LABORATORY Experiment #2: Introduction to Lab Equipment: Function Generator, Oscilloscope,

More information

Models 296 and 295 combine sophisticated

Models 296 and 295 combine sophisticated Established 1981 Advanced Test Equipment Rentals www.atecorp.com 800-404-ATEC (2832) Models 296 and 295 50 MS/s Synthesized Multichannel Arbitrary Waveform Generators Up to 4 Independent Channels 10 Standard

More information

PSM Soft. Features and Functions January PC Software Guide. Getting connected and Communication

PSM Soft. Features and Functions January PC Software Guide. Getting connected and Communication PSM Soft PC Software Guide Features and Functions January 2010 The PSM series Phase Sensitive Multimeters provide a wide range of exceptionally accurate and versatile instrumentation in one unique package.

More information

DSOXEDK Educator s Oscilloscope Training Kit. Lab Guide and Tutorial for Agilent 4000 X-Series Oscilloscopes

DSOXEDK Educator s Oscilloscope Training Kit. Lab Guide and Tutorial for Agilent 4000 X-Series Oscilloscopes DSOXEDK Educator s Oscilloscope Training Kit Lab Guide and Tutorial for Agilent 4000 X-Series Oscilloscopes s1 Notices Agilent Technologies, Inc. 2008-2012 The copyright on this instructional material

More information

OSCILLOSCOPES. Oscilloscopes CS-5400 SERIES CS-5400/5450 FEATURES OUTLINE CS-5400

OSCILLOSCOPES. Oscilloscopes CS-5400 SERIES CS-5400/5450 FEATURES OUTLINE CS-5400 99 Washington Street Melrose, MA 02176 Fax 781-665-0780 TestEquipmentDepot.com Oscilloscopes 100MHz 3-Channel Oscilloscope (With Digital Readout / Cursor) CS-5400 100MHz 3-Channel Oscilloscope CS-5405

More information

U1600 Series Handheld Digital Oscilloscopes

U1600 Series Handheld Digital Oscilloscopes U1600 Series Handheld Digital Oscilloscopes Data Sheet Delivering more functionality and performance with a handheld digital oscilloscope Features Three-in-one solution: Dual channel oscilloscope, True

More information

Educator s Training Kit for 1000 X-Series Oscilloscopes

Educator s Training Kit for 1000 X-Series Oscilloscopes Educator s Training Kit for 1000 X-Series Oscilloscopes Lab Guide and Tutorial for Undergraduate Electrical Engineering and Physics Students Notices Keysight Technologies, Inc. 2008-2017 No part of this

More information

Digital Storage Oscilloscopes Models 2540B, 2542B, 2540B-GEN, 2542B-GEN

Digital Storage Oscilloscopes Models 2540B, 2542B, 2540B-GEN, 2542B-GEN Data Sheet Digital Storage Oscilloscopes Models 2540B, 2542B, 2540B-GEN, 2542B-GEN The 2540B, 2542B, 2540B-GEN, and 2542B-GEN dual channel 60 MHz and 100 MHz digital storage oscilloscopes deliver performance

More information

TEST & MEASURING INSTRUMENTS. Analyzer. (4 Ports) 4 Ports

TEST & MEASURING INSTRUMENTS. Analyzer. (4 Ports) 4 Ports TEST & MEASURING INSTRUMENTS Analyzer (4 Ports) 4 Ports Key Features Frequrncy Range : 100kHz ~ 8GHz, 16 Parameters support (S11 ~ S44) Measurement time per point : 100us per point Wide Output Power Range

More information