Simultaneous two-photon calcium imaging at different depths with spatiotemporal multiplexing

Size: px
Start display at page:

Download "Simultaneous two-photon calcium imaging at different depths with spatiotemporal multiplexing"

Transcription

1 Nature Methods Simultaneous two-photon calcium imaging at different depths with spatiotemporal multiplexing Adrian Cheng, J Tiago Gonçalves, Peyman Golshani, Katsushi Arisaka & Carlos Portera-Cailliau Supplementary Figure 1 Supplementary Figure 2 Supplementary Figure 3 Supplementary Figure 4 Supplementary Figure 5 Effect of tissue depth on light scattering as measured by camera-type detection or photomultiplier tube. Multifocal in vivo calcium imaging and simultaneous electrophysiology. Optical configuration for multiple beam imaging. Read-out electronics. Optical transmission through various components of the prototype spatiotemporally multiplexed microscope. Supplementary Figure 6 Supplementary Figure 7 Supplementary Figure 8 Supplementary Note 1 Negligible beam cross-talk for Fluo-4 with 4-beam temporal multiplexing. Negligible beam cross-talk with OGB-1 AM with two-beam temporal multiplexing. Multiple beam scanning increases brightness and scanning speed for constant average laser power per beam. Longer fluorescence lifetimes, improving acquisition rates, and additional fluorophores.

2 Supplementary Figure 1 Effect of tissue depth on light scattering as measured by camera-type detection or photomultiplier tube. Cooled CCD PMT (HPD) Axial depth (µm) Radial distance (µm) µm -100 µm -200 µm -300 µm Cortical slice thickness 20 0 Intensity (db) Supplementary Figure 1 Effect of tissue depth on light scattering as measured by camera-type detection or photomultiplier tube. Resolution is measured as a function of brain depth. Fluorescence spheres (5 µm in diameter) were placed under fixed brain slices of varying thickness and imaged with 2PLSM (single beam), using either a cooled CCD camera (left) or a PMT (right) for detection. Note the degradation in resolution radially (laterally) seen with camera-type detection, due to visible wavelength scattering. Axial resolution is preserved with camera-type detection, demonstrating the degradation of the detection point spread function (PSF) rather than the excitation PSF. Dilution of signal with depth requires longer integration times for the camera (over minutes for 300 µm in depth, vs. seconds for PMT detection).

3 Supplementary Figure 2 Multifocal in vivo calcium imaging and simultaneous electrophysiology. bm. 4 bm. 3 bm. 2 bm. 1 a b µm c d 0.5 F/F 1 10 Cell Number s 20 s 10 pa Cell-attached recording 0.2 F/F e Simulated calcium imaging (from Ephys) Actual calcium imaging 2s Supplementary Figure 2 Multifocal in vivo calcium imaging and simultaneous electrophysiology. (a) Field of view scanned with 4 laser beams. The tip of the patch-clamp electrode is outlined in yellow and the recorded cell is indicated by an arrow. (b) Cell contours of 91 individual neurons. (c) Raw calcium traces from the first 10 cells. The simultaneous cell-attached electrophysiology trace is shown at the bottom (blue). The shaded area is shown with finer temporal resolution in e. (d) Raster plot of the calcium transient events for all 91 cells. (e) Detail of simultaneous electrophysiology and calcium imaging. From top to bottom: cellattached recording trace (black), simulated (convolved) calcium imaging trace based on electrophysiology (red), actual calcium imaging trace (green).

4 Supplementary Figure 3 Optical configuration for multiple beam imaging. a c SM1 SM1 PBS1 θ +0 ns +3 ns +6 ns +9 ns +0 ns p-pol L4 b SM1 M3 M2 PBS1 SE-M1 PBS2 M4 L1 L2 L3 +0 ns p-pol +3 ns s-pol +6 ns s-pol +9 ns p-pol L1 +6 ns s-pol L2 +3 ns s-pol L3 +9 ns p-pol d Telescope (T) Back aperture image Critical point (x) e ~ f o 2 T / d for Td / f o >> 1 x o o Back focal plane Back aperture d 1 f o d 2 o x o d 1 d 2 Back focal space Object space focus d critical point zero focal shift Back focal space focus d Object space Supplementary Figure 3 Optical configuration for multiple beam imaging. (a) To generate a vertical pattern of beam spots for single plane multibeam imaging, beams were converged onto scanning mirror (SM1) with an angular spacing θ = M x / f o where M is the demagnification by relay lens telescopes, x is the desired pitch at sample, and f o is the focal length of the objective lens, as approximated by f t / M o, where f t is the manufacturer tube lens focal length and M o is the objective magnification rating. (b) To generate a scan of four imaging planes, beams of varying divergence introduced by lenses (L1, L2, and L3) were combined by polarizing beam splitters (PBS1 and PBS2) and further combined by a sharp edge mirror (SE-M1). This led to a shift of two beams by ~ 20 µm. Note this could be done with a pinhole mirror of sufficient precision as well, to increase co-linearity. A pinhole mirror method would allow for a generalization of this configuration to N beams. (c) To generate a scan of two imaging planes with two beams each, a single polarizing beam splitter (PBS1) was used to combine four beams. Lenses introduced varying divergence or convergence (L1-L4). In all cases a-c, the back aperture was filled uniformly by maintaining uniform beam diameter at SM1. (d) An idealized optical train is shown for generating axial focal shift, consisting of the object space imaged by the objective lens, and a back focal space after a relay telescope with magnification T. To examine focal shift, let the objective focal length f o << f t, the tube length. Then, the back focal plane (BFP) will approximately Nature coincide Methods: with doi: /nmeth.1552 the objective exit pupil. Small displacements δ from the BFP will be imaged by the telescope at approximately Tδ from the BFP image (critical point x). (e) Ultimately the focal shift d 1 - f o = will be approximately f o2 T / d 1 where d 1 is the beam focus distance from x, making the approximation Tδ/f o >> 1.

5 Supplementary Figure 4 Read-out electronics. Photocathode 8kV APD Bias 300V Scanner synchronization Color select/shutter/piezo/etc γ e- HAPD Preamplifier Laser monitor Color select/ interleave V REF Demultiplexer Transimpedance/ charge-sensitive amplifier To display To disk array Frame grabber PCIe DMA transfer Gate generator Delay lines Host PCU Supplementary Figure 4 Read-out electronics. HPD signals are amplified and color channel is selected or interleaved by a high-speed analog switch (blue). The signals from the four separate beams are de-multiplexed by analog multiplication by four delayed gates with 3 ns width and 12 ns period (yellow). The delayed gates are generated by a laser monitor signal (green). The resulting current-voltage signal is then shaped for digitization by a commercial four-channel high-speed frame grabber (orange). PCI-express DMA transfer using the supplied frame grabber software library allows for real-time display, processing, and disk array recording (purple).

6 Supplementary Figure 5 Optical transmission through various components of the prototype spatiotemporally multiplexed microscope. 1 1 Power Transmission Spatiotemporal multiplexing prototype 2PLSM Zeiss 40x 1.0NA Olympus 40x 0.8NA Laser After delay lines After scanners After objectives 0.1 Supplementary Figure 5 Optical transmission through various components of the prototype spatiotemporally multiplexed microscope. Measurements were made before the system (at the laser output), after the delay lines section, after the scanning mirrors, and after the objective lens.

7 Supplementary Figure 6 Negligible beam cross-talk for Fluo-4 with 4-beam temporal multiplexing. a z y z x y Beam 2 only x Beam 1 only z y x Beam 2 (with beam 1) b 9 25 µm -0.2% 4-1.7% % 5 A(x,y) 10 B(x,y) +2.8% C(x,y) = A(x,y) + γb(x,y) Supplementary Figure 6 Negligible beam cross-talk for Fluo-4 with 4-beam temporal multiplexing. (a) Field of view scanned with beam 2 alone (left), beam 1 alone (middle), keeping other beams blocked, and field of view of beam 2 while simultaneously scanning with beam 1 (right). These images were acquired using the 3-D spatiotemporal multiplexing configuration and are average time intensity (xyt) projections of 45 sec-long in vivo calcium imaging movies. Regions of interest (ROI; yellow outlines) were selected around somas of presumed neurons (ROIs 1-9) and neuropil (ROI 10) in the beam 2 image. Note that bright astrocytes seen in layer 1 (beam 1 image) are not present in the beam 2 image. The rectangular region outlined in the left panel (red box) is shown at higher zoom in b. (b) Higher magnification detail of boxed region in panel a. Examples of cross talk values are shown for 3 representative cell body ROIs and 1 neuropil ROI. The average intensity change in beam 2 ROIs in the presence of beam 1 excitation was 3.0 ± 4.1 % (mean ± st. Nature Methods: doi: /nmeth.1552 dev.), suggesting that fluorescence leak-through between beams is not a dominant noise factor (see Online Methods).

8 Supplementary Figure 7 Negligible beam cross-talk with OGB-1 AM with 2-beam temporal multiplexing. Beam 1 only 10.5% 7.1% 4.1% 3.3% 5.6% 6-12 ns Cell 1 Cell 3 Cell 4 Cell 5 Cell 2 75 µm 0-6 ns Supplementary Figure 7 Negligible beam cross-talk with OGB-1 AM with 2-beam temporal multiplexing. A single beam was used to scan the field of view (top), with images at 0-6 ns (bottom) and 6-12 ns (top) after the excitation pulse shown images were acquired at 100 Hz and summed. Representative cells were selected (yellow, bottom) and the corresponding % cross-talk (i.e., contribution in the 6-12 ns image) for each cell is shown (top). The average static cross-talk for OGB was calculated to be 6.1 ± 2.9% (mean ± st. dev.). Because this percentage corresponds to somatic OGB-1 that on average already exhibits its longer lifetime of ~2.5 ns, action potential-induced calcium transients of ~ 10% F/F would be expected to result in cross-talk transients of 0.6%, which is on the order of Poisson fluctuations.

9 Supplementary Figure 8 Multiple beam scanning increases brightness and scanning speed for constant average laser power per beam. Beam 1 (0-12 ns) Beam 1 (0-6 ns) Beam 2 (6-12 ns) 75 µm 75 µm 7.13 ADC counts per pixel 1.43 x 10 6 ADC counts per frame (mean over 4000 frames) 800 x 250 pixels, 100 fps 1.5 W into system 7.98 ADC counts per pixel 3.19 x 10 6 ADC counts per frame (mean over 4000 frames) 800 x 500 pixels, 100 fps 1.5 W per beam into system Supplementary Figure 8 Multiple beam scanning increases brightness and scanning speed for constant average laser power per beam. To demonstrate specific advantages of multiple beam scanning in vivo, a field of view of mouse barrel cortex (layer 2/3) was stained with OGB-1 AM and imaged first with a single beam (left) and then with two beams 6 ns apart (right). The average power per beam was held constant and the laser repetition rate was 80 MHz. This implies a constant laser pulse energy, with 80 MHz pulse delivery at the sample for the case of a single beam, and 160 MHz pulse delivery for the case of two beams. Not only was scanning speed increased in the case of two beams (800 x 250 pixels at 100 frames per second versus 800 x 500 pixels at 100 frames per second), but fluorescence yield was approximately doubled (3.19 x 10 6 vs x 10 6 ADC counts). Thus, in a hypothetical situation where laser pulse energy is limited by non-linear photodamage, increasing laser pulse delivery rates is an option for increasing signal level. However, rather than merely increasing delivery rate to a single point, spatially distributing and resolving beams also provides an increase in scanning rate. Laser pulse energy may not necessarily be limited in all types of imaging, and non-linear damage limits (in terms of input pulse energy at the brain surface) certainly will have a functional dependence on depth. This functional dependence and the precise ideal pulse energies versus depth remain to be explored.

10 SUPPLEMENTARY NOTE Spatiotemporal multiplexing for fluorophores with longer lifetimes The number of spatiotemporally multiplexed beams could be greater than 4 provided one uses a laser with a lower repetition rate, which is determined by the length of the laser cavity or additional components such as regenerative amplifiers 1 or cavity dumpers 2. In addition, the lifetime of the calcium indicator used for temporally multiplexed excitation must be carefully considered to avoid cross-talk between beams. Much of in vivo 2PCI these days is done with OGB, so there may be great interest in the potential suitability of this dye for spatiotemporal multiplexing. OGB has both short and long lifetimes, (~350 ps and ~2.5 ns in live cells, respectively), the relative proportion of which appears to change in the presence of calcium as determined by its dissociation constant (K d ~ 200 nm) 3-4. This is due to a change in the radiative quantum yield in the presence of calcium. Fluo-4, whose excitation efficiency (but not its quantum yield) increases in the presence of calcium, has a relatively constant lifetime of ~1 ns. Thus, Fluo-4 was the preferred dye for our calcium imaging experiments using a 80 MHz laser and a 4-beam design. In contrast, at spiking levels ([Ca 2+ ] > 1 µm) 5, OGB was not ideally suited for our 4-beam multiplexing approach. However, using only 2 beams separated in time by 6 ns, we demonstrate in vivo calcium imaging with OGB with negligible cross-talk (Supplementary Fig. 7). In the future, any of the above mentioned approaches for lowering the laser repetition rate would permit the use of spatiotemporal multiplexing for 2PCI of dyes with longer lifetimes (~3 ns), including GFP-based genetically encoded calcium indicators. Additionally, in light of its calcium-dependent lifetime, OGB would also benefit from time-gated detection by excluding photons detected for the duration of its short [Ca 2+ ] free lifetime.

11 Improving acquisition rates by combining spatiotemporal multiplexing with AODs and targeted path scanning In principle, spatiotemporal multiplexing can also be combined with other approaches that address the closely related issues of signal to noise ratio (SNR), field of view (FOV) in two and three dimensions, and time resolution for in vivo 2PCI, including random access point scanning with acousto-optical deflectors (AODs) or targeted-path scanning with closed loop mirrors 6-8. Both approaches seek to increase time resolution and SNR by restricting laser delivery to a region of interest (e.g., neuronal cell bodies, but not the surrounding neuropil). AODs allow millisecond time resolution by scanning a limited number of points coinciding with preselected subsets of labeled neurons over a large FOV. Unfortunately, the number of AODs necessary to implement 3-D scanning introduces considerable spatial and temporal dispersion, as well as power loss, although this can be easily compensated for using prism correction or pulse pre-chirping 7. In addition, the axial extent of scanning with AODs is less than 50 µm 8. The arbitrary targeted path scanning technique takes advantage of closed-loop mirrors to scan optimized contours coinciding with cell bodies 6. For 3-D imaging, this approach can be combined with a piezo-electric objective scanner, resulting in acquisition rates of several Hz for small volumes 9. More importantly, an advantage of the raster scanning approach using multiple beams that we used is that it preserves full frame image scanning in 3-D, which allows a full range of image processing methods for cell identification and image registration. To combine these approaches with spatiotemporal multiplexing would require multiple sets of scanners and simultaneous projection of their apertures on to the objective back aperture. One solution would be to simply project the apertures of multiple sets of AODs or closed-loop scanning mirrors on to the objective back aperture with different access angles. This would allow each individual beam to scan

12 distinct FOVs. To achieve independent beam scanning in the same FOV, polarization optics generally provides a passive solution for combining the optical paths of two sets of scanners. The availability of more elegant solutions will depend on the specific scan method and pattern. Fluorescence lifetime-induced cross-talk We measured the potential beam cross-talk due to the finite fluorescence lifetime of the dye compared to the 3 ns beam delays during normal experimental imaging conditions. A region of layer 2/3 in barrel cortex was stained with Fluo-4 AM and in vivo calcium imaging was carried out as in Fig. 3, with different beams scanning different axial planes in 3-D (Supplementary Fig. 6). First, a 45 s time-lapse sequence of images was obtained with only beams 2 scanning the tissue (the other beams were blocked). Next, 1 minute later, another 45 s time-lapse sequence of images was obtained with only beam 1. A final 45 s time-lapse sequence was taken with beams 1 and 2 scanning the tissue at different planes (30 µm apart in the z direction). This last calcium movie is subject to crosstalk between beams. Regions of interest around Fluo-4 labeled cells in the beam 2 image plane were manually selected and their average pixel intensity throughout the movie was calculated. The average leakage (γ) of beam 1 into the beam 2 image was calculated using the following linear combination, C(x,y) = A(x, y) + γ(x,y) B(x,y) where A is the pixel intensity in the beam 2 alone image, B is the pixel intensity in the beam 1 alone image, and C the pixel intensity in the beam 2 image when simultaneously scanning with beam 1. Cross talk varied for different ROIs, but the average leak-through was 3.0 ± 4.1% (mean ± st. dev.). Cross-talk likely also varies slightly both spatially due to other fluorophore compositions (e.g., autofluorescence),

13 and also temporally (e.g., neuronal activity), which is why these measurements were done over 45 sec. For example, if cross-talk occurs during a 10% ΔF/F calcium transient, the dynamic contribution of this transient to the adjacent beam would be small (a ~0.3% transient). This effect will be the same order or less than Poisson fluctuations in our system and is ameliorated by the fact that the fluorescence lifetime of Fluo-4 decreases during a calcium transient 5, which brings somatic calcium concentration to the dye dissociation constant k 10 d. In the case of OGB, using two beams spaced by 6 ns in time, the average signal leak-through for cellular regions was 6.1 ± 2.9% (mean ± st. dev.; Supplementary Fig. 7). REFERENCES 1 Theer, P., Hasan, M. T. & Denk, W. Two-photon imaging to a depth of 1000 microns in living brains by use of a Ti:Al2O3 regenerative amplifier. Opt. Lett. 28, (2003). 2 Ramaswamy, M., Ulman, M., Paye, J. & Fujimoto, J. G. Cavity-dumped femtosecond Kerr-lens mode-locked Ti:A1(2)O(3)laser. Opt. Lett. 18, , doi:12002 [pii] (1993). 3 Wilms, C. D., Schmidt, H. & Eilers, J. Quantitative two-photon Ca2+ imaging via fluorescence lifetime analysis. Cell Calcium 40, 73-79, doi:s (06) [pii] /j.ceca (2006). 4 Wilms, C. D. & Eilers, J. Photo-physical properties of Ca2+-indicator dyes suitable for two-photon fluorescence-lifetime recordings. J. Microsc. 225, , doi:jmi1746 [pii] /j x (2007).

14 5 Helmchen, F., Imoto, K. & Sakmann, B. Ca2+ buffering and action potential-evoked Ca2+ signaling in dendrites of pyramidal neurons. Biophys. J. 70, , doi:s (96) [pii] /S (96) (1996). 6 Lillis, K. P., Eng, A., White, J. A. & Mertz, J. Two-photon imaging of spatially extended neuronal network dynamics with high temporal resolution. J. Neurosci. Methods 172, , doi:s (08) [pii] /j.jneumeth (2008). 7 Kremer, Y. et al. A spatio-temporally compensated acousto-optic scanner for two-photon microscopy providing large field of view. Opt. Express 16, , doi: [pii] (2008). 8 Duemani Reddy, G., Kelleher, K., Fink, R. & Saggau, P. Three-dimensional random access multiphoton microscopy for functional imaging of neuronal activity. Nat. Neurosci. 11, , doi:nn.2116 [pii] /nn.2116 (2008). 9 Gobel, W., Kampa, B. M. & Helmchen, F. Imaging cellular network dynamics in three dimensions using fast 3D laser scanning. Nat. Methods 4, (2007). 10 Yoshiki, K., Azuma, H., Yoshioka, K., Hashimoto, M. & Araki, T. Finding of optimal calcium ion probes for fluorescence lifetime measurement. Optical Review 12, (2005).

Nature Neuroscience: doi: /nn Supplementary Figure 1. Optimized Bessel foci for in vivo volume imaging.

Nature Neuroscience: doi: /nn Supplementary Figure 1. Optimized Bessel foci for in vivo volume imaging. Supplementary Figure 1 Optimized Bessel foci for in vivo volume imaging. (a) Images taken by scanning Bessel foci of various NAs, lateral and axial FWHMs: (Left panels) in vivo volume images of YFP + neurites

More information

3D light microscopy techniques

3D light microscopy techniques 3D light microscopy techniques The image of a point is a 3D feature In-focus image Out-of-focus image The image of a point is not a point Point Spread Function (PSF) 1D imaging 1 1 2! NA = 0.5! NA 2D imaging

More information

Nature Methods: doi: /nmeth Supplementary Figure 1

Nature Methods: doi: /nmeth Supplementary Figure 1 . Supplementary Figure 1 Schematics and characterization of our AO two-photon fluorescence microscope. (a) Essential components of our AO two-photon fluorescence microscope: Ti:Sapphire laser; optional

More information

Akinori Mitani and Geoff Weiner BGGN 266 Spring 2013 Non-linear optics final report. Introduction and Background

Akinori Mitani and Geoff Weiner BGGN 266 Spring 2013 Non-linear optics final report. Introduction and Background Akinori Mitani and Geoff Weiner BGGN 266 Spring 2013 Non-linear optics final report Introduction and Background Two-photon microscopy is a type of fluorescence microscopy using two-photon excitation. It

More information

Characteristics of point-focus Simultaneous Spatial and temporal Focusing (SSTF) as a two-photon excited fluorescence microscopy

Characteristics of point-focus Simultaneous Spatial and temporal Focusing (SSTF) as a two-photon excited fluorescence microscopy Characteristics of point-focus Simultaneous Spatial and temporal Focusing (SSTF) as a two-photon excited fluorescence microscopy Qiyuan Song (M2) and Aoi Nakamura (B4) Abstracts: We theoretically and experimentally

More information

Multifluorescence The Crosstalk Problem and Its Solution

Multifluorescence The Crosstalk Problem and Its Solution Multifluorescence The Crosstalk Problem and Its Solution If a specimen is labeled with more than one fluorochrome, each image channel should only show the emission signal of one of them. If, in a specimen

More information

Technology Note ZEISS LSM 880 with Airyscan

Technology Note ZEISS LSM 880 with Airyscan Technology Note ZEISS LSM 880 with Airyscan Introducing the Fast Acquisition Mode ZEISS LSM 880 with Airyscan Introducing the Fast Acquisition Mode Author: Dr. Annette Bergter Carl Zeiss Microscopy GmbH,

More information

FEMTOSMART. Benefits. Features

FEMTOSMART. Benefits. Features FEMTOSMART Extremely large space under the objective For in vivo studies Field upgradability Patented imaging technologies Flexible scanning methods Maximal photon collection Elevated, column-based body

More information

Why and How? Daniel Gitler Dept. of Physiology Ben-Gurion University of the Negev. Microscopy course, Michmoret Dec 2005

Why and How? Daniel Gitler Dept. of Physiology Ben-Gurion University of the Negev. Microscopy course, Michmoret Dec 2005 Why and How? Daniel Gitler Dept. of Physiology Ben-Gurion University of the Negev Why use confocal microscopy? Principles of the laser scanning confocal microscope. Image resolution. Manipulating the

More information

Practical work no. 3: Confocal Live Cell Microscopy

Practical work no. 3: Confocal Live Cell Microscopy Practical work no. 3: Confocal Live Cell Microscopy Course Instructor: Mikko Liljeström (MIU) 1 Background Confocal microscopy: The main idea behind confocality is that it suppresses the signal outside

More information

Single-photon excitation of morphology dependent resonance

Single-photon excitation of morphology dependent resonance Single-photon excitation of morphology dependent resonance 3.1 Introduction The examination of morphology dependent resonance (MDR) has been of considerable importance to many fields in optical science.

More information

Nature Methods: doi: /nmeth Supplementary Figure 1. Schematic of 2P-ISIM AO optical setup.

Nature Methods: doi: /nmeth Supplementary Figure 1. Schematic of 2P-ISIM AO optical setup. Supplementary Figure 1 Schematic of 2P-ISIM AO optical setup. Excitation from a femtosecond laser is passed through intensity control and shuttering optics (1/2 λ wave plate, polarizing beam splitting

More information

3D light microscopy techniques

3D light microscopy techniques 3D light microscopy techniques The image of a point is a 3D feature In-focus image Out-of-focus image The image of a point is not a point Point Spread Function (PSF) 1D imaging 2D imaging 3D imaging Resolution

More information

Nature Structural & Molecular Biology: doi: /nsmb Supplementary Figure 1

Nature Structural & Molecular Biology: doi: /nsmb Supplementary Figure 1 Supplementary Figure 1 Supplemental correlative nanomanipulation-fluorescence traces probing nascent RNA and fluorescent Mfd during TCR initiation. Supplemental correlative nanomanipulation-fluorescence

More information

Point Spread Function. Confocal Laser Scanning Microscopy. Confocal Aperture. Optical aberrations. Alternative Scanning Microscopy

Point Spread Function. Confocal Laser Scanning Microscopy. Confocal Aperture. Optical aberrations. Alternative Scanning Microscopy Bi177 Lecture 5 Adding the Third Dimension Wide-field Imaging Point Spread Function Deconvolution Confocal Laser Scanning Microscopy Confocal Aperture Optical aberrations Alternative Scanning Microscopy

More information

Leica TCS SP8 Quick Start Guide

Leica TCS SP8 Quick Start Guide Leica TCS SP8 Quick Start Guide Leica TCS SP8 System Overview Start-Up Procedure 1. Turn on the CTR Control Box, Fluorescent Light for the microscope stand. 2. Turn on the Scanner Power (1) on the front

More information

TCSPC at Wavelengths from 900 nm to 1700 nm

TCSPC at Wavelengths from 900 nm to 1700 nm TCSPC at Wavelengths from 900 nm to 1700 nm We describe picosecond time-resolved optical signal recording in the spectral range from 900 nm to 1700 nm. The system consists of an id Quantique id220 InGaAs

More information

Boulevard du Temple Daguerrotype (Paris,1838) a busy street? Nyquist sampling for movement

Boulevard du Temple Daguerrotype (Paris,1838) a busy street? Nyquist sampling for movement Boulevard du Temple Daguerrotype (Paris,1838) a busy street? Nyquist sampling for movement CONFOCAL MICROSCOPY BioVis Uppsala, 2017 Jeremy Adler Matyas Molnar Dirk Pacholsky Widefield & Confocal Microscopy

More information

Resolution. Diffraction from apertures limits resolution. Rayleigh criterion θ Rayleigh = 1.22 λ/d 1 peak at 2 nd minimum. θ f D

Resolution. Diffraction from apertures limits resolution. Rayleigh criterion θ Rayleigh = 1.22 λ/d 1 peak at 2 nd minimum. θ f D Microscopy Outline 1. Resolution and Simple Optical Microscope 2. Contrast enhancement: Dark field, Fluorescence (Chelsea & Peter), Phase Contrast, DIC 3. Newer Methods: Scanning Tunneling microscopy (STM),

More information

Multicolor 4D Fluorescence Microscopy using Ultrathin Bessel Light sheets

Multicolor 4D Fluorescence Microscopy using Ultrathin Bessel Light sheets SUPPLEMENTARY MATERIAL Multicolor 4D Fluorescence Microscopy using Ultrathin Bessel Light sheets Teng Zhao, Sze Cheung Lau, Ying Wang, Yumian Su, Hao Wang, Aifang Cheng, Karl Herrup, Nancy Y. Ip, Shengwang

More information

Confocal Microscopy. Kristin Jensen

Confocal Microscopy. Kristin Jensen Confocal Microscopy Kristin Jensen 17.11.05 References Cell Biological Applications of Confocal Microscopy, Brian Matsumoto, chapter 1 Studying protein dynamics in living cells,, Jennifer Lippincott-Schwartz

More information

INTRODUCTION TO MICROSCOPY. Urs Ziegler THE PROBLEM

INTRODUCTION TO MICROSCOPY. Urs Ziegler THE PROBLEM INTRODUCTION TO MICROSCOPY Urs Ziegler ziegler@zmb.uzh.ch THE PROBLEM 1 ORGANISMS ARE LARGE LIGHT AND ELECTRONS: ELECTROMAGNETIC WAVES v = Wavelength ( ) Speed (v) Frequency ( ) Amplitude (A) Propagation

More information

Observational Astronomy

Observational Astronomy Observational Astronomy Instruments The telescope- instruments combination forms a tightly coupled system: Telescope = collecting photons and forming an image Instruments = registering and analyzing the

More information

Shaping light in microscopy:

Shaping light in microscopy: Shaping light in microscopy: Adaptive optical methods and nonconventional beam shapes for enhanced imaging Martí Duocastella planet detector detector sample sample Aberrated wavefront Beamsplitter Adaptive

More information

Fast two-layer two-photon imaging of neuronal cell populations using an electrically tunable lens

Fast two-layer two-photon imaging of neuronal cell populations using an electrically tunable lens Fast two-layer two-photon imaging of neuronal cell populations using an electrically tunable lens Benjamin F. Grewe, 1,3 Fabian F. Voigt, 1,3 and Marcel van t Hoff, 1 and Fritjof Helmchen 1,* 1 Department

More information

FLUORESCENCE MICROSCOPY. Matyas Molnar and Dirk Pacholsky

FLUORESCENCE MICROSCOPY. Matyas Molnar and Dirk Pacholsky FLUORESCENCE MICROSCOPY Matyas Molnar and Dirk Pacholsky 1 The human eye perceives app. 400-700 nm; best at around 500 nm (green) Has a general resolution down to150-300 μm (human hair: 40-250 μm) We need

More information

Opterra II Multipoint Scanning Confocal Microscope. Innovation with Integrity

Opterra II Multipoint Scanning Confocal Microscope. Innovation with Integrity Opterra II Multipoint Scanning Confocal Microscope Enabling 4D Live-Cell Fluorescence Imaging through Speed, Sensitivity, Viability and Simplicity Innovation with Integrity Fluorescence Microscopy The

More information

Instant super-resolution imaging in live cells and embryos via analog image processing

Instant super-resolution imaging in live cells and embryos via analog image processing Nature Methods Instant super-resolution imaging in live cells and embryos via analog image processing Andrew G. York, Panagiotis Chandris, Damian Dalle Nogare, Jeffrey Head, Peter Wawrzusin, Robert S.

More information

Acoustic resolution. photoacoustic Doppler velocimetry. in blood-mimicking fluids. Supplementary Information

Acoustic resolution. photoacoustic Doppler velocimetry. in blood-mimicking fluids. Supplementary Information Acoustic resolution photoacoustic Doppler velocimetry in blood-mimicking fluids Joanna Brunker 1, *, Paul Beard 1 Supplementary Information 1 Department of Medical Physics and Biomedical Engineering, University

More information

University of California Los Angeles. Spatio temporal Excitation emission Multiplexing for Multifocal 2 photon Calcium Imaging in Deep Tissue

University of California Los Angeles. Spatio temporal Excitation emission Multiplexing for Multifocal 2 photon Calcium Imaging in Deep Tissue University of California Los Angeles Spatio temporal Excitation emission Multiplexing for Multifocal 2 photon Calcium Imaging in Deep Tissue A dissertation submitted in partial satisfaction of the requirements

More information

ADVANCED METHODS FOR CONFOCAL MICROSCOPY II. Jean-Yves Chatton Sept. 2006

ADVANCED METHODS FOR CONFOCAL MICROSCOPY II. Jean-Yves Chatton Sept. 2006 ADVANCED METHODS FOR CONFOCAL MICROSCOPY II Jean-Yves Chatton Sept. 2006 Workshop outline Confocal microscopy of living cells and tissues X-Z scanning Time series Bleach: FRAP, photoactivation Emission

More information

PCS-150 / PCI-200 High Speed Boxcar Modules

PCS-150 / PCI-200 High Speed Boxcar Modules Becker & Hickl GmbH Kolonnenstr. 29 10829 Berlin Tel. 030 / 787 56 32 Fax. 030 / 787 57 34 email: info@becker-hickl.de http://www.becker-hickl.de PCSAPP.DOC PCS-150 / PCI-200 High Speed Boxcar Modules

More information

Multiphoton Microscopy

Multiphoton Microscopy Multiphoton Microscopy A. Neumann, Y. Kuznetsova Introduction Multi-Photon Fluorescence Microscopy is a relatively novel imaging technique in cell biology. It relies on the quasi-simultaneous absorption

More information

Multiphoton Detection Unit (MDU)

Multiphoton Detection Unit (MDU) Experts in Electrophysiology Microscope Equipment Multiphoton Detection Unit (MDU) For integration with the Scientifica SliceScope Pro motorised microscopy system 2 Multiphoton Detection Unit Two-photon

More information

Confocal, hyperspectral, spinning disk

Confocal, hyperspectral, spinning disk Confocal, hyperspectral, spinning disk Administrative HW 6 due on Fri Midterm on Wed Covers everything since previous midterm 8.5 x 11 sheet allowed, 1 side Guest lecture by Joe Dragavon on Mon 10/30 Last

More information

Maria Smedh, Centre for Cellular Imaging. Maria Smedh, Centre for Cellular Imaging

Maria Smedh, Centre for Cellular Imaging. Maria Smedh, Centre for Cellular Imaging Nonlinear microscopy I: Two-photon fluorescence microscopy Multiphoton Microscopy What is multiphoton imaging? Applications Different imaging modes Advantages/disadvantages Scattering of light in thick

More information

Development of a High-speed Super-resolution Confocal Scanner

Development of a High-speed Super-resolution Confocal Scanner Development of a High-speed Super-resolution Confocal Scanner Takuya Azuma *1 Takayuki Kei *1 Super-resolution microscopy techniques that overcome the spatial resolution limit of conventional light microscopy

More information

Imaging Retreat - UMASS Customized real-time confocal and 2-photon imaging

Imaging Retreat - UMASS Customized real-time confocal and 2-photon imaging Imaging Retreat - UMASS 2012 Customized real-time confocal and 2-photon imaging Mike Sanderson Department of Microbiology and Physiological Systems University of Massachusetts Medical School Thanks for

More information

Very short introduction to light microscopy and digital imaging

Very short introduction to light microscopy and digital imaging Very short introduction to light microscopy and digital imaging Hernan G. Garcia August 1, 2005 1 Light Microscopy Basics In this section we will briefly describe the basic principles of operation and

More information

Pupil Planes versus Image Planes Comparison of beam combining concepts

Pupil Planes versus Image Planes Comparison of beam combining concepts Pupil Planes versus Image Planes Comparison of beam combining concepts John Young University of Cambridge 27 July 2006 Pupil planes versus Image planes 1 Aims of this presentation Beam combiner functions

More information

Shreyash Tandon M.S. III Year

Shreyash Tandon M.S. III Year Shreyash Tandon M.S. III Year 20091015 Confocal microscopy is a powerful tool for generating high-resolution images and 3-D reconstructions of a specimen by using point illumination and a spatial pinhole

More information

Leica TCS SP8 Quick Start Guide

Leica TCS SP8 Quick Start Guide Leica TCS SP8 Quick Start Guide Leica TCS SP8 System Overview Start-Up Procedure 1. Turn on the CTR Control Box, EL6000 fluorescent light source for the microscope stand. 2. Turn on the Scanner Power

More information

Electron-Bombarded CMOS

Electron-Bombarded CMOS New Megapixel Single Photon Position Sensitive HPD: Electron-Bombarded CMOS University of Lyon / CNRS-IN2P3 in collaboration with J. Baudot, E. Chabanat, P. Depasse, W. Dulinski, N. Estre, M. Winter N56:

More information

TRAINING MANUAL. Multiphoton Microscopy LSM 510 META-NLO

TRAINING MANUAL. Multiphoton Microscopy LSM 510 META-NLO TRAINING MANUAL Multiphoton Microscopy LSM 510 META-NLO September 2010 Multiphoton Microscopy Training Manual Multiphoton microscopy is only available on the LSM 510 META-NLO system. This system is equipped

More information

Applications of Optics

Applications of Optics Nicholas J. Giordano www.cengage.com/physics/giordano Chapter 26 Applications of Optics Marilyn Akins, PhD Broome Community College Applications of Optics Many devices are based on the principles of optics

More information

EE119 Introduction to Optical Engineering Spring 2003 Final Exam. Name:

EE119 Introduction to Optical Engineering Spring 2003 Final Exam. Name: EE119 Introduction to Optical Engineering Spring 2003 Final Exam Name: SID: CLOSED BOOK. THREE 8 1/2 X 11 SHEETS OF NOTES, AND SCIENTIFIC POCKET CALCULATOR PERMITTED. TIME ALLOTTED: 180 MINUTES Fundamental

More information

Mesoscale Two-Photon Microscopy: Engineering a Wide Field of View with Cellular Resolution

Mesoscale Two-Photon Microscopy: Engineering a Wide Field of View with Cellular Resolution Mesoscale Two-Photon Microscopy: Engineering a Wide Field of View with Cellular Resolution Jeffrey N. Stirman, PhD, and Spencer L. Smith, PhD Neuroscience Center and Department of Cell Biology and Physiology

More information

Training Guide for Leica SP8 Confocal/Multiphoton Microscope

Training Guide for Leica SP8 Confocal/Multiphoton Microscope Training Guide for Leica SP8 Confocal/Multiphoton Microscope LAS AF v3.3 Optical Imaging & Vital Microscopy Core Baylor College of Medicine (2017) Power ON Routine 1 2 Turn ON power switch for epifluorescence

More information

High Resolution BSI Scientific CMOS

High Resolution BSI Scientific CMOS CMOS, EMCCD AND CCD CAMERAS FOR LIFE SCIENCES High Resolution BSI Scientific CMOS Prime BSI delivers the perfect balance between high resolution imaging and sensitivity with an optimized pixel design and

More information

Zeiss 780 Training Notes

Zeiss 780 Training Notes Zeiss 780 Training Notes Turn on Main Switch, System PC and Components Switches 780 Start up sequence Do you need the argon laser (458, 488, 514 nm lines)? Yes Turn on the laser s main power switch and

More information

Confocal Imaging Through Scattering Media with a Volume Holographic Filter

Confocal Imaging Through Scattering Media with a Volume Holographic Filter Confocal Imaging Through Scattering Media with a Volume Holographic Filter Michal Balberg +, George Barbastathis*, Sergio Fantini % and David J. Brady University of Illinois at Urbana-Champaign, Urbana,

More information

NIH Public Access Author Manuscript Opt Lett. Author manuscript; available in PMC 2010 August 9.

NIH Public Access Author Manuscript Opt Lett. Author manuscript; available in PMC 2010 August 9. NIH Public Access Author Manuscript Published in final edited form as: Opt Lett. 2010 January 1; 35(1): 67 69. Autoconfocal transmission microscopy based on two-photon induced photocurrent of Si photodiodes

More information

Operation Guide for the Leica SP2 Confocal Microscope Bio-Imaging Facility Hunter College October 2009

Operation Guide for the Leica SP2 Confocal Microscope Bio-Imaging Facility Hunter College October 2009 Operation Guide for the Leica SP2 Confocal Microscope Bio-Imaging Facility Hunter College October 2009 Introduction of Fluoresence Confocal Microscopy The first confocal microscope was invented by Princeton

More information

Spectral phase shaping for high resolution CARS spectroscopy around 3000 cm 1

Spectral phase shaping for high resolution CARS spectroscopy around 3000 cm 1 Spectral phase shaping for high resolution CARS spectroscopy around 3 cm A.C.W. van Rhijn, S. Postma, J.P. Korterik, J.L. Herek, and H.L. Offerhaus Mesa + Research Institute for Nanotechnology, University

More information

plasmonic nanoblock pair

plasmonic nanoblock pair Nanostructured potential of optical trapping using a plasmonic nanoblock pair Yoshito Tanaka, Shogo Kaneda and Keiji Sasaki* Research Institute for Electronic Science, Hokkaido University, Sapporo 1-2,

More information

Supplementary Figure 1

Supplementary Figure 1 Supplementary Figure 1 Technical overview drawing of the Roadrunner goniometer. The goniometer consists of three main components: an inline sample-viewing microscope, a high-precision scanning unit for

More information

Supplementary Information. Stochastic Optical Reconstruction Microscopy Imaging of Microtubule Arrays in Intact Arabidopsis thaliana Seedling Roots

Supplementary Information. Stochastic Optical Reconstruction Microscopy Imaging of Microtubule Arrays in Intact Arabidopsis thaliana Seedling Roots Supplementary Information Stochastic Optical Reconstruction Microscopy Imaging of Microtubule Arrays in Intact Arabidopsis thaliana Seedling Roots Bin Dong 1,, Xiaochen Yang 2,, Shaobin Zhu 1, Diane C.

More information

Aberrations and adaptive optics for biomedical microscopes

Aberrations and adaptive optics for biomedical microscopes Aberrations and adaptive optics for biomedical microscopes Martin Booth Department of Engineering Science And Centre for Neural Circuits and Behaviour University of Oxford Outline Rays, wave fronts and

More information

Basics of confocal imaging (part I)

Basics of confocal imaging (part I) Basics of confocal imaging (part I) Swiss Institute of Technology (EPFL) Faculty of Life Sciences Head of BIOIMAGING AND OPTICS BIOP arne.seitz@epfl.ch Lateral resolution BioImaging &Optics Platform Light

More information

Multi-channel imaging cytometry with a single detector

Multi-channel imaging cytometry with a single detector Multi-channel imaging cytometry with a single detector Sarah Locknar 1, John Barton 1, Mark Entwistle 2, Gary Carver 1 and Robert Johnson 1 1 Omega Optical, Brattleboro, VT 05301 2 Philadelphia Lightwave,

More information

Rapid Adaptive Optical Recovery of Optimal Resolution over Large Volumes

Rapid Adaptive Optical Recovery of Optimal Resolution over Large Volumes SUPPLEMENTARY MATERIAL Rapid Adaptive Optical Recovery of Optimal Resolution over Large Volumes Kai Wang, Dan Milkie, Ankur Saxena, Peter Engerer, Thomas Misgeld, Marianne E. Bronner, Jeff Mumm, and Eric

More information

DCS laser for Thomson scattering diagnostic applications

DCS laser for Thomson scattering diagnostic applications DCS laser for Thomson scattering diagnostic applications Authors Jason Zweiback 10/6/2015 jzweiback@logostech.net 1 Summary Motivation DCS laser Laser for Thomson scattering diagnostics 2 What is the Dynamic

More information

Nikon Instruments Europe

Nikon Instruments Europe Nikon Instruments Europe Recommendations for N-SIM sample preparation and image reconstruction Dear customer, We hope you find the following guidelines useful in order to get the best performance out of

More information

Confocal Microscopy and Related Techniques

Confocal Microscopy and Related Techniques Confocal Microscopy and Related Techniques Chau-Hwang Lee Associate Research Fellow Research Center for Applied Sciences, Academia Sinica 128 Sec. 2, Academia Rd., Nankang, Taipei 11529, Taiwan E-mail:

More information

Application Note. The New 2D Superresolution Mode for ZEISS Airyscan 120 nm Lateral Resolution without Acquiring a Z-stack

Application Note. The New 2D Superresolution Mode for ZEISS Airyscan 120 nm Lateral Resolution without Acquiring a Z-stack The New 2D Superresolution Mode for ZEISS Airyscan 120 nm Lateral Resolution without Acquiring a Z-stack The New 2D Superresolution Mode for ZEISS Airyscan 120 nm Lateral Resolution without Acquiring a

More information

CHAPTER 9 POSITION SENSITIVE PHOTOMULTIPLIER TUBES

CHAPTER 9 POSITION SENSITIVE PHOTOMULTIPLIER TUBES CHAPTER 9 POSITION SENSITIVE PHOTOMULTIPLIER TUBES The current multiplication mechanism offered by dynodes makes photomultiplier tubes ideal for low-light-level measurement. As explained earlier, there

More information

Study of self-interference incoherent digital holography for the application of retinal imaging

Study of self-interference incoherent digital holography for the application of retinal imaging Study of self-interference incoherent digital holography for the application of retinal imaging Jisoo Hong and Myung K. Kim Department of Physics, University of South Florida, Tampa, FL, US 33620 ABSTRACT

More information

Optical Sectioning Deep Inside Live Embryos by Selective Plane Illumination Microscopy

Optical Sectioning Deep Inside Live Embryos by Selective Plane Illumination Microscopy Supporting Online Material Optical Sectioning Deep Inside Live Embryos by Selective Plane Illumination Microscopy 1 Material and methods 1.1 Setup Jan Huisken, Jim Swoger, Filippo Del Bene, Joachim Wittbrodt,

More information

Lab4 Hanbury Brown and Twiss Setup. Photon Antibunching

Lab4 Hanbury Brown and Twiss Setup. Photon Antibunching Lab4 Hanbury Brown and Twiss Setup. Photon Antibunching Shule Li Abstract Antibunching is a purely quantum effect and cannot be realized from the classical theory of light. By observing the antibunching

More information

LSM 710 Confocal Microscope Standard Operation Protocol

LSM 710 Confocal Microscope Standard Operation Protocol LSM 710 Confocal Microscope Standard Operation Protocol Basic Operation Turning on the system 1. Switch on Main power switch 2. Switch on System / PC power button 3. Switch on Components power button 4.

More information

High-speed 1-frame ms scanning confocal microscope with a microlens and Nipkow disks

High-speed 1-frame ms scanning confocal microscope with a microlens and Nipkow disks High-speed 1-framems scanning confocal microscope with a microlens and Nipkow disks Takeo Tanaami, Shinya Otsuki, Nobuhiro Tomosada, Yasuhito Kosugi, Mizuho Shimizu, and Hideyuki Ishida We have developed

More information

Fast, high-contrast imaging of animal development with scanned light sheet based structured-illumination microscopy

Fast, high-contrast imaging of animal development with scanned light sheet based structured-illumination microscopy nature methods Fast, high-contrast imaging of animal development with scanned light sheet based structured-illumination microscopy Philipp J Keller, Annette D Schmidt, Anthony Santella, Khaled Khairy,

More information

Training Guide for Carl Zeiss LSM 5 LIVE Confocal Microscope

Training Guide for Carl Zeiss LSM 5 LIVE Confocal Microscope Training Guide for Carl Zeiss LSM 5 LIVE Confocal Microscope AIM 4.2 Optical Imaging & Vital Microscopy Core Baylor College of Medicine (2017) Power ON Routine 1 2 Verify that main power switches on the

More information

Wide-Field TCSPC FLIM with bh SPC-150 N TCSPC System and Photek FGN Detector

Wide-Field TCSPC FLIM with bh SPC-150 N TCSPC System and Photek FGN Detector Wide-Field TCSPC FLIM with bh SPC-150 N TCSPC System and Photek FGN 392-1000 Detector Abstract: We present a wide-field TCSPC FLIM system consisting of a position-sensitive MCP PMT of the delay-line type,

More information

Megapixel FLIM with bh TCSPC Modules

Megapixel FLIM with bh TCSPC Modules Megapixel FLIM with bh TCSPC Modules The New SPCM 64-bit Software Abstract: Becker & Hickl have recently introduced version 9.60 of their SPCM TCSPC data acquisition software. SPCM version 9.60 not only

More information

Confocal Microscopy. (Increasing contrast and resolu6on using op6cal sec6oning) Lecture 7. November 2017

Confocal Microscopy. (Increasing contrast and resolu6on using op6cal sec6oning) Lecture 7. November 2017 Confocal Microscopy (Increasing contrast and resolu6on using op6cal sec6oning) Lecture 7 November 2017 3 Flavours of Microscope Confocal Laser Scanning Problem: Out of Focus Light Spinning disc 2-Photon

More information

Information & Instructions

Information & Instructions KEY FEATURES 1. USB 3.0 For the Fastest Transfer Rates Up to 10X faster than regular USB 2.0 connections (also USB 2.0 compatible) 2. High Resolution 4.2 MegaPixels resolution gives accurate profile measurements

More information

Supplementary Figure S1. Schematic representation of different functionalities that could be

Supplementary Figure S1. Schematic representation of different functionalities that could be Supplementary Figure S1. Schematic representation of different functionalities that could be obtained using the fiber-bundle approach This schematic representation shows some example of the possible functions

More information

DCS-120. Confocal Scanning FLIM Systems. Based on bh s Multidimensional Megapixel FLIM Technology

DCS-120. Confocal Scanning FLIM Systems. Based on bh s Multidimensional Megapixel FLIM Technology Based on bh s Multidimensional Megapixel FLIM Technology Complete Laser Scanning FLIM Microscopes FLIM Upgrades for Existing Conventional Microscopes Multidimensional TCSPC technique High throughput dual-channel

More information

A novel tunable diode laser using volume holographic gratings

A novel tunable diode laser using volume holographic gratings A novel tunable diode laser using volume holographic gratings Christophe Moser *, Lawrence Ho and Frank Havermeyer Ondax, Inc. 85 E. Duarte Road, Monrovia, CA 9116, USA ABSTRACT We have developed a self-aligned

More information

Instructions for the Experiment

Instructions for the Experiment Instructions for the Experiment Excitonic States in Atomically Thin Semiconductors 1. Introduction Alongside with electrical measurements, optical measurements are an indispensable tool for the study of

More information

Be aware that there is no universal notation for the various quantities.

Be aware that there is no universal notation for the various quantities. Fourier Optics v2.4 Ray tracing is limited in its ability to describe optics because it ignores the wave properties of light. Diffraction is needed to explain image spatial resolution and contrast and

More information

Examination, TEN1, in courses SK2500/SK2501, Physics of Biomedical Microscopy,

Examination, TEN1, in courses SK2500/SK2501, Physics of Biomedical Microscopy, KTH Applied Physics Examination, TEN1, in courses SK2500/SK2501, Physics of Biomedical Microscopy, 2009-06-05, 8-13, FB51 Allowed aids: Compendium Imaging Physics (handed out) Compendium Light Microscopy

More information

Adaptive optics two-photon fluorescence microscopy

Adaptive optics two-photon fluorescence microscopy Adaptive optics two-photon fluorescence microscopy Yaopeng Zhou 1, Thomas Bifano 1 and Charles Lin 2 1. Manufacturing Engineering Department, Boston University 15 Saint Mary's Street, Brookline MA, 02446

More information

Spectroscopy in the UV and Visible: Instrumentation. Spectroscopy in the UV and Visible: Instrumentation

Spectroscopy in the UV and Visible: Instrumentation. Spectroscopy in the UV and Visible: Instrumentation Spectroscopy in the UV and Visible: Instrumentation Typical UV-VIS instrument 1 Source - Disperser Sample (Blank) Detector Readout Monitor the relative response of the sample signal to the blank Transmittance

More information

Components of confocal and two-photon microscopes

Components of confocal and two-photon microscopes Components of confocal and two-photon microscopes Internal training 07/04/2016 A. GRICHINE Platform Optical microscopy Cell imaging, IAB, ISdV Plan Confocal laser scanning microscope o o o Principle Main

More information

Supplementary Figure 1. Effect of the spacer thickness on the resonance properties of the gold and silver metasurface layers.

Supplementary Figure 1. Effect of the spacer thickness on the resonance properties of the gold and silver metasurface layers. Supplementary Figure 1. Effect of the spacer thickness on the resonance properties of the gold and silver metasurface layers. Finite-difference time-domain calculations of the optical transmittance through

More information

Digital Camera Technologies for Scientific Bio-Imaging. Part 2: Sampling and Signal

Digital Camera Technologies for Scientific Bio-Imaging. Part 2: Sampling and Signal Digital Camera Technologies for Scientific Bio-Imaging. Part 2: Sampling and Signal Yashvinder Sabharwal, 1 James Joubert 2 and Deepak Sharma 2 1. Solexis Advisors LLC, Austin, TX, USA 2. Photometrics

More information

ChromoFlex. The flexibility of four colour multiphoton imaging. Experts in Electrophysiology & Imaging

ChromoFlex.   The flexibility of four colour multiphoton imaging. Experts in Electrophysiology & Imaging Experts in Electrophysiology & Imaging 1 Multiphoton Imaging ChromoFlex The flexibility of four colour multiphoton imaging www.scientifica.uk.com/chromoflex www.scientifica.uk.com 2 ChromoFlex Multiple

More information

PROCEEDINGS OF SPIE. Measurement of low-order aberrations with an autostigmatic microscope

PROCEEDINGS OF SPIE. Measurement of low-order aberrations with an autostigmatic microscope PROCEEDINGS OF SPIE SPIEDigitalLibrary.org/conference-proceedings-of-spie Measurement of low-order aberrations with an autostigmatic microscope William P. Kuhn Measurement of low-order aberrations with

More information

Laser Speckle Reducer LSR-3000 Series

Laser Speckle Reducer LSR-3000 Series Datasheet: LSR-3000 Series Update: 06.08.2012 Copyright 2012 Optotune Laser Speckle Reducer LSR-3000 Series Speckle noise from a laser-based system is reduced by dynamically diffusing the laser beam. A

More information

2013 LMIC Imaging Workshop. Sidney L. Shaw Technical Director. - Light and the Image - Detectors - Signal and Noise

2013 LMIC Imaging Workshop. Sidney L. Shaw Technical Director. - Light and the Image - Detectors - Signal and Noise 2013 LMIC Imaging Workshop Sidney L. Shaw Technical Director - Light and the Image - Detectors - Signal and Noise The Anatomy of a Digital Image Representative Intensities Specimen: (molecular distribution)

More information

Spectral Imaging with the Opterra Multipoint Scanning Confocal

Spectral Imaging with the Opterra Multipoint Scanning Confocal Spectral Imaging with the Opterra Multipoint Scanning Confocal Outline Opterra design overview Scan Modes Light Path Spectral Imaging with Opterra Drosophila larva heart. Opterra Design Overview Supravideo

More information

Supplemental Information. Light Prior to Eye Opening Promotes. Retinal Waves and Eye-Specific Segregation

Supplemental Information. Light Prior to Eye Opening Promotes. Retinal Waves and Eye-Specific Segregation Neuron, Volume 100 Supplemental Information Light Prior to Eye Opening Promotes Retinal Waves and Eye-Specific Segregation Alexandre Tiriac, Benjamin E. Smith, and Marla B. Feller Supplementary Figure

More information

Supplemental Figure 1: Histogram of 63x Objective Lens z axis Calculated Resolutions. Results from the MetroloJ z axis fits for 5 beads from each

Supplemental Figure 1: Histogram of 63x Objective Lens z axis Calculated Resolutions. Results from the MetroloJ z axis fits for 5 beads from each Supplemental Figure 1: Histogram of 63x Objective Lens z axis Calculated Resolutions. Results from the MetroloJ z axis fits for 5 beads from each lens with a 1 Airy unit pinhole setting. Many water lenses

More information

3.0 Alignment Equipment and Diagnostic Tools:

3.0 Alignment Equipment and Diagnostic Tools: 3.0 Alignment Equipment and Diagnostic Tools: Alignment equipment The alignment telescope and its use The laser autostigmatic cube (LACI) interferometer A pin -- and how to find the center of curvature

More information

High-speed, low-photodamage nonlinear imaging using passive pulse splitters

High-speed, low-photodamage nonlinear imaging using passive pulse splitters High-speed, low-photodamage nonlinear imaging using passive pulse splitters Na Ji, Jeffrey C Magee & Eric Betzig Supplementary figures and text: Supplementary Figure 1 A monolithic 1x pulse splitter with

More information

NanoSpective, Inc Progress Drive Suite 137 Orlando, Florida

NanoSpective, Inc Progress Drive Suite 137 Orlando, Florida TEM Techniques Summary The TEM is an analytical instrument in which a thin membrane (typically < 100nm) is placed in the path of an energetic and highly coherent beam of electrons. Typical operating voltages

More information

Bio 407. Applied microscopy. Introduction into light microscopy. José María Mateos. Center for Microscopy and Image Analysis

Bio 407. Applied microscopy. Introduction into light microscopy. José María Mateos. Center for Microscopy and Image Analysis Center for Microscopy and Image Analysis Bio 407 Applied Introduction into light José María Mateos Fundamentals of light Compound microscope Microscope composed of an objective and an additional lens (eyepiece,

More information

MULTIPHOTON MICROSCOPY. Matyas Molnar Dirk Pacholsky

MULTIPHOTON MICROSCOPY. Matyas Molnar Dirk Pacholsky MULTIPHOTON MICROSCOPY Matyas Molnar Dirk Pacholsky Information Information given here about 2 Photon microscopy were mainly taken from these sources: Background information on 2-Photon microscopy: http://micro.magnet.fsu.edu/primer/techniques/fluorescence/multiphoton/

More information