Nature Protocols: doi: /nprot Supplementary Figure 1. Principle effects of LSFM illumination and detection on image quality.

Size: px
Start display at page:

Download "Nature Protocols: doi: /nprot Supplementary Figure 1. Principle effects of LSFM illumination and detection on image quality."

Transcription

1 Supplementary Figure 1 Principle effects of LSFM illumination and detection on image quality. (A) Top view of a scheme showing the arrangement of illumination objective, sample chamber and detection objective in a light sheetbased fluorescence microscope. The inset depicts the illumination and detection of a single line in a specimen. (B) Top view of the objective arrangement in the mdslm. In the inset, image quality indicators are provided as colored arcs. The green arc marks the quarter with an expected high quality, the orange arcs the quarters with an expected medium quality and the red arc the quarter with an expected low quality. (C) Y maximum projection of a Z stack with the image quality indicators introduced in B.

2 Supplementary Figure 2 Example quantification of the serosa window area over time during serosa window closure. (A) Z maximum projection of a Tribolium embryo during gastrulation that commences serosa window closure seen from the ventral side. (B) Time lapse of single planes, 26 µm below the ventral surface of the embryo, showing the process of serosa window closure. The first row depicts the raw planes. In the second row, an entropy filter was applied to the raw planes. Then, an intensity threshold was set to obtain the binary images. The binary images were inverted and further processed by morphological operations to extract the serosa window area in each image, as shown in the third row. The fourth row illustrates the raw planes marked with the extracted serosa window area. (C) A plot of the serosa window area as a function of development time, providing evidence that the serosa window area during serosa window closure follows an exponential decay. ZP, Z maximum projection with image processing; PA, plane with intensity adjustment.

3 Supplementary Figure 3 Chronological staging table of Tribolium embryogenesis. The left column contains Z maximum projections along four orientations (0 (ventral), 90 (lateral), 180 (dorsal), and 270 (lateral)) of a Tribolium embryo at the beginning of five major embryogenetic events. In the middle column, an outline of the most important developmental processes with a temporal indication in hours (developments at 35 C and at room temperature) and stage indication in percent of full development. Please note that 00:00 h relates to our standard imaging starting point (that is after a 1 h egg laying period / 15 h of incubation at 25 C and 1 h of preparation at room temperature), whereas the stage indication in percent is provided for the whole of embryogenesis. The right column shows Z maximum projections of developmental processes during the respective embryogenetic events that have been the focus of previous research. Red arrows indicate the location of the process mentioned in the bullet points. RT, room temperature (23±1 C); ap., approximately.

4 Supplementary Figure 4 Scheme of the recommended control experiments. In addition to the imaging procedure of the fluorescent embryo (left column), we suggest two control experiments that can be run in parallel: a fluorescent control embryo (middle column) and a wild type control embryo (right column). All steps in which the control experiments have to be considered are indicated and also mentioned in the Procedure section.

5 Supplementary Figure 5 Operating principle of the digital scanned laser light sheet-based fluorescence microscope (DSLM). In contrast to other types of light sheet-based fluorescence microscopes, the DSLM does not have a static light sheet. Instead, a dynamic light sheet is generated by scanning the laser beam with a piezo-driven two-axes scanning mirror (see green annotations). The laser beam, originating from the laser source, is guided by the scanning mirror through an f-theta and a tube lens into the illumination objective. The dynamic light sheet illuminates a thin slice of the Tribolium embryo, and the emitted fluorescence is guided by a perpendicularly arranged detection objective through an appropriate filter and a tube lens into the camera. The detail on the right shows the sample holder with the embryo in front of the detection objective. Adapted and modified from ref. 18 with permission, Nature Publishing Group.

6 Supplementary Figure 6 Perfusion system. (A) Overview of mdslm with connected perfusion system. The main elements of the circulatory perfusion system are labelled. Controlled transfer of PBS from the reservoir is achieved via the inflow pump through the heating tube into the sample chamber. From the chamber, excessive PBS is transferred via the outflow pump back into the reservoir. Upper detail box shows a top view of the sample chamber. The perfusion system connects to the periphery of the sample chamber and, therefore, does not impair the imaging process. Lower detail box shows the front view of the sample chamber with connected inflow and outflow tubes. The inflow tube connects to the lower and the outflow tube to the upper joint of the sample chamber. Therefore, the PBS outflow rate is limited by the inflow rate. Directions of flow are indicated with white arrowheads. A third tube is connected to a catch basin located beneath the sample chamber in case of a leakage, as indicated with a gray arrowhead. (B) Setup and equilibrium of perfusion system. A constant temperature in the sample chamber is maintained via a combination of PBS flow rate and the temperature generated in the heating tube. Prior to the experiment, the perfusion system is given two hours to ensure a stable temperature in the sample chamber. The circulatory flow of PBS and relative warmth are indicated with colored arrowheads.

7 Supplementary Figure 7 Illustration of equipment setup. (A) Front (left) and side view (right) of the glass cover slip holder. (A ) Front (left panel) and side view (right panel) of the holder in A with attached glass cover slip. (A ) Glass cover slip holder in the LSFM sample chamber. From left to right: top view of glass cover slip holder perpendicular to Z (the detection axis), top, front and side view of the glass cover slip holder rotated by 45 around Y (the rotation axis). (B) Capillary holder (left panel) with capillary attached (right panel). (C) Syringe components with cut lines indicated in orange (left panel). Components with indicated parts cut off (right panel). (C ) Assembled syringe (top panel), with rubber hose (middle panel) and capillary (bottom panel) attached. (D) Humidity chamber. (D ) Scaffold created from serological pipette fragments without (left panel) and with object slide (right panel).

8 Supplementary Figure 8 Test recording of agarose-embedded beads column for microscope calibration. The agarose-embedded beads column is shown along four orientations as a Z (first row), X (second row) and Y (third row) maximum projections. The maximum projections illustrate correct calibration of the microscope without any imaging artifacts. The rotated Y maximum projections (fourth row) demonstrate a correctly set up recording of the bead agarose column. Individual beads can be identified along all directions as shown in the detail images (fifth row). Due to the point spread function, beads are expected to appear as bright, round dots in the Z maximum projection and elongated along X and Y maximum projections. ZP, Z maximum projections with image processing; XP, X maximum projections with image processing; YP, Y maximum projections with image processing.

9 Supplementary Figure 9 Test recording of the embryo along four directions prior to time lapse imaging. The embryo is shown in four orientations in the transmission light channel (first row) and as Z (second row), X (third row) and Y (fourth row) maximum projections. The maximum projections illustrate a correctly set up recording along four directions without any imaging artifacts. Some preparation issues and misconfigurations of the imaging setup (e.g. insufficient amount of agarose in the stability layer leading to a drift of the embryo, incorrect Z spacing, incorrect configured volume of view), might be more apparent in X and Y maximum projections than in the Z maximum projections. Indicators for a correctly set up recording are: X and Y maximum projections depict smooth object boundaries and show no motion blur; Z maximum projections and the corresponding X maximum projections exhibit high similarity; and in the Y maximum projections, the shape of the embryo should be roughly circular. ZP, Z maximum projections with image processing; XP, X maximum projections with image processing; YP, Y maximum projections with image processing.

10 Supplementary Figure 10 Qualitative coverage of the embryo. The embryo is shown along four orientations as Z and Y maximum projections. The Y maximum projections (second row) illustrate that the quality of the recording strongly depends on the amount and type of biological material passed by the illumination and emission light paths. If no biological material obstructs the light paths, the resulting image quality is high (lower left area in all orientations). However, if the amount of material that has to be passed increases, the image quality drops (upper left and lower right areas in all orientations). The image quality is worst in areas where both light paths have to pass the complete embryo (upper right areas in all orientations). The rotated Y maximum projections (third row) highlight the fact that features of the embryo might only be partly visible in some orientations. The green lines outline the clearly detectable germ band borders from the respective direction. ZP, Z maximum projections with image processing; YP, Y maximum projections with image processing.

11 Supplementary Table 1 Metadata and parameters for the three example datasets. Dataset (DS) DS0001 DS0002 DS0003 Species Line Line Genotype Stock Stock Medium Stock Conditions Egg Laying Period Egg Laying Medium Pre-imaging Incubation Tribolium castaneum (Herbst) / Arthropoda Insecta Coleoptera Tenebrionidae EFA-nGFP transgenic line 10 (background strain: vermilion white) one insert / homozygote ~300 adults, less than 4 months old full grain wheat flour ( , Demeter, Darmstadt, Germany) supplemented with 5% (wt/wt) inactive dry yeast (62-106, Flystuff, San Diego, CA, USA) 12:00 h light / 12:00 h darkness at 25 C and 70% relative humidity (DR-36VL, Percival Scientific, Perry, IA, USA) 01:00 h at 25 C and 70% relative humidity exposed to light 405 fine wheat flour ( , Demeter, Darmstadt, Germany) supplemented with 5% (wt/wt) inactive dry yeast (62-106, Flystuff, San Diego, CA, USA) 15:00 h at 25 C and 70% relative humidity in darkness LSFM Type mdslm based on DSLM 18,72 with structured illumination 19 Laser Lines Excitation Objective Emission Objective Emission Filters Camera Dataset File Type (Raw) Dechorionation Mounting Agarose Imaging Buffer 488 nm / 60 mw diode laser (PhoxX , Omicron Laserprodukte GmbH, Rodgau-Dudenhofen, Germany) 2.5x NA 0.06 EC Epiplan-Neofluar objective ( , Carl Zeiss, Göttingen, Germany) 10x NA 0.3 W N-Achroplan objective ( , Carl Zeiss, Göttingen, Germany) 525/50 single-band bandpass filter (FF03-525/50-25, Semrock/AHF Analysentechnik AG, Tübingen, Germany) High-resolution CCD (Clara, Andor, Belfast, United Kingdom) TIFF, 16 bit grayscale (planes are deposited as Z stacks, indicated as PL(ZS) in the file name) ~30 s in 10% (vol/vol) sodium hypochlorite ( ML, Sigma Adlrich, Taufkirchen, Germany) in PBS ph 7.4 ( , Gibco Life Technologies GmbH, Darmstadt, Germany) 1% (wt/vol) low-melt agarose (6351.2, Carl Roth, Karlsruhe, Germany) in PBS ph 7.4 ( , Gibco Life Technologies GmbH, Darmstadt, Germany) PBS ph 7.4 ( , Gibco Life Technologies GmbH, Darmstadt, Germany) Imaging Temperature 35 C (perfusion system) room temperature (23±1 C) room temperature (23±1 C) 92 Gigabyte (raw) 642 Gigabyte (raw) 193 Gigabyte (raw) Dataset Size 37 Gigabyte (ZIP) 273 Gigabyte (ZIP) 77 Gigabyte (ZIP) 0.47 Gigabyte (JPEG2000) 5.54 Gigabyte (JPEG2000) 1.62 Gigabyte (JPEG2000) Retrieval survived, developed to healthy adult, produced fertile progeny survived, developed to healthy adult, produced fertile progeny survived, developed to healthy adult, produced fertile progeny Figures 10 (upper row) 4D-F; 7; 8; 9; 10 (lower row); 12 (upper row) 11 Supplementary Figures - 2; 3; 9; 10 - Supplementary Videos - 3; 4; 5 6

12 Time Points (TP) 49 (TP0001-TP0049) 241 (TP0001-TP0241) 49 (TP0001-TP0049) TP Interval 00:30 h 00:30 h 00:30 h Total Time (TP TP Interval) 24:00 h 120:00 h 24:00 h Directions (DR) 4 (DR0001-DR0004) 4 (DR0001-DR0004) 4 (DR0001-DR0004) DR Orientations 0, 90, 180, 270 0, 90, 180, 270 0, 90, 180, 270 Channels (CH) 1 (CH0001) 1 (CH0001) 2 (CH0001-CH0002) CH0001 Excitation 488 nm 488 nm 488 nm CH0001 Power 135 µw (close to the embryo) 135 µw (close to the embryo) 135 µw (close to the embryo) CH0001 Exposure Time 50 ms 50 ms 50 ms CH0001 Emission Filter 525/50 single-band bandpass filter 525/50 single-band bandpass filter 525/50 single-band bandpass filter CH0002 Excitation nm, structured illumination 50 Hz CH0002 Power µw (close to the embryo) CH0002 Exposure Time ms CH0002 Emission Filter /50 single-band bandpass filter Planes (PL) 175 (PL0001-PL0175) 198 (PL0001-PL0198) 183 (PL0001-PL0183) Z Spacing 2.58 µm 2.58 µm 2.58 µm Z Distance (PL Z Spacing) µm µm µm X-Dimensions (XD) 1040 pixels (raw) 600 pixels (cropped) 1040 pixels (raw) 600 pixels (cropped) 1040 pixels (raw) 600 pixels (cropped) X Spacing µm µm µm X Length (XD X Spacing) µm (raw) µm (cropped) µm (raw) µm (cropped) µm (raw) µm (cropped) Y-Dimensions (YD) 1392 pixels (raw) 1000 pixels (cropped) 1392 pixels (raw) 1000 pixels (cropped) 1392 pixels (raw) 1000 pixels (cropped) Y Spacing µm µm µm µm (raw) µm (raw) µm (raw) Y Length (YD Y Spacing) µm (cropped) µm (cropped) µm (cropped) The original datasets, associated figures and videos in full quality are available at

13 Supplementary Note Troubleshooting of common problems during LSFM calibration and subsequent imaging processing. Please note that in some figures the problem and in others the reason is shown. A green checkmark in the upper right corner indicates the correct execution, whereas a red cross indicates an incorrect execution. (A-A ) Bright dot not visible when glass cover slip holder is placed in microscope chamber (Step 9). (A) The correct glass cover slip holder position reflects the laser light directly into the detection objective. (A -A ) Glass cover slip holders that are positioned offset either left (A ) or right (A ) cause the reflected laser light to miss the detection objective. Also ensure that the glass cover slip holder is positioned at the correct height (front view in detail box in A). (B-B ) Light sheet does not span straight across the complete field of view along Y (Step 11). (B) The light sheet should extend approximately 20% over the field of view along Y, so that the visible part of the light sheet is homogeneous. (B ) If the scanning range of the activated scanning mirror is too small, the light sheet does not span across the complete field of view along Y. (B ) If the scanning mirror is not adjusted correctly, the light sheet is tilted. (C-C ) Light sheet does not appear focused (Step 11). (C) The light sheet should be a straight focused line with minimal width highlighting surface irregularities on the inserted glass cover slip. (C ) The light sheet appears broader when the waist is not positioned in the field of view. (C ) It appears blurry when the focal plane is not in the center of the light sheet.

14 Supplementary Note (continued) (D-D ) Beads are missing or appear blurry on either end of Y (Step 14). (D) The beads should appear focused and homogeneously distributed along the vertical direction across the full height of the acquired image. (D ) If beads are missing on either end or on both ends of the image, the light sheet height has to be readjusted, the glass capillary is in the field of view, or the agarose-embedded beads column is too short. (D ) If beads appear focused in the image center and blurry on the outer edges in the vertical direction, the light sheet is tilted. The light sheet orientation needs to be readjusted using the glass cover slip holder. (E-E ) Beads appear blurry across the complete field of view (Step 14). (E) A circle of homogenously distributed beads should appear and beads should be focused across the full Z distance of the Y maximum projection. (E ) If beads appear dim and elongated, the light sheet waist is not positioned correctly and the position of the illumination objective needs to be readjusted using the piezo nano-positioners. (E ) If beads appear blurry, the light sheet does not overlap with the focal plane. The position of the detection objective needs to be readjusted using the piezo nano-positioners.

15 Supplementary Note (continued) (F-F ) Difficulties in mounting the embryo on the agarose hemisphere (Step 28 ). (F) The semi axes of the agarose hemisphere should be at best identical (see also Figure 3G). (F -F ) If the pole radius is smaller (F ) or greater (F ), the embryo tends to tip over during the mounting procedure. (G-G ) The embryo moves after being placed on the pole of the agarose hemisphere (Step 28). (G-G ) During the mounting process, the posterior end of the embryo should be attached to the brush tip, while the anterior end remains accessible. (G ) If the embryo is attached vice versa, either reposition the embryo with a second brush, or release the embryo into the PBS and restart the picking process. (H-H ) Embryo detaches when sample holder is inserted into sample chamber (Step 31). (H) The drop of liquid agarose to fix the embryo has to be correctly placed on the agarose hemisphere. (H ) If it is placed on the steel pipe, the agarose does not reach the embryo. (H ) If the drop is positioned directly on the embryo, the embryo might detach. (I-I ) Incorrect amount of agarose for the mounting and stability layers (Steps 31 / 40 / 44). (I) The mounting and stability layers should cover approximately between a quarter and half of the embryo surface at maximum and the agarose flanks should be as steep as possible. The embryo shown here survived the imaging process and developed into a fertile adult. (I ) If too little agarose is applied, the embryo detaches during the insertion into the sample chamber. (I ) If too much agarose is applied, the survival chance of the embryo and the imaging quality are decreased. The embryo shown here did not survive the imaging process.

16 Supplementary Note (continued) (J-J ) Blurred or dark regions at anterior or posterior regions of the embryo (Step 36). (J) The embryo should be visible and in focus in the complete acquired image. (J ) If the light sheet height is insufficient, the embryo is only partially visible. (J ) A tilted light sheet results in blurry regions at the top and bottom along Y. (K-K ) During long-term imaging, the whole embryo appears blurred or certain regions appear dark after some time (Step 36). (K) The embryo should remain focused in the image volume, even during long-term imaging. (K ) Insufficient amount of agarose in the stability layer might lead to a drift of the embryo out of the image volume. Besides, the Z stack should be large enough to cover slight movements of the embryo over time. (K ) If the chamber sealing for example generates excessive force on the detection objective, a focus drift may occur over time.

Fast, high-contrast imaging of animal development with scanned light sheet based structured-illumination microscopy

Fast, high-contrast imaging of animal development with scanned light sheet based structured-illumination microscopy nature methods Fast, high-contrast imaging of animal development with scanned light sheet based structured-illumination microscopy Philipp J Keller, Annette D Schmidt, Anthony Santella, Khaled Khairy,

More information

Practical work no. 3: Confocal Live Cell Microscopy

Practical work no. 3: Confocal Live Cell Microscopy Practical work no. 3: Confocal Live Cell Microscopy Course Instructor: Mikko Liljeström (MIU) 1 Background Confocal microscopy: The main idea behind confocality is that it suppresses the signal outside

More information

Comparing FCS and FRAP as methodologies for calculating diffusion

Comparing FCS and FRAP as methodologies for calculating diffusion Bi/BE 227 Winter 2018 Assignment #4 Comparing FCS and FRAP as methodologies for calculating diffusion Schedule: Jan 29: Assignment Jan 29-Feb 14: Work on assignment Feb 14: Student PowerPoint presentations.

More information

Bi/BE 227 Winter Assignment #3. Adding the third dimension: 3D Confocal Imaging

Bi/BE 227 Winter Assignment #3. Adding the third dimension: 3D Confocal Imaging Bi/BE 227 Winter 2016 Assignment #3 Adding the third dimension: 3D Confocal Imaging Schedule: Jan 20: Assignment Jan 20-Feb 8: Work on assignment Feb 10: Student PowerPoint presentations. Goals for this

More information

Figure 1: A detailed sketch of the experimental set up.

Figure 1: A detailed sketch of the experimental set up. Electronic Supplementary Material (ESI) for Soft Matter. This journal is The Royal Society of Chemistry 2015 Supplementary Information Detailed Experimental Set Up camera 2 long range objective aluminum

More information

Nature Structural & Molecular Biology: doi: /nsmb Supplementary Figure 1

Nature Structural & Molecular Biology: doi: /nsmb Supplementary Figure 1 Supplementary Figure 1 Supplemental correlative nanomanipulation-fluorescence traces probing nascent RNA and fluorescent Mfd during TCR initiation. Supplemental correlative nanomanipulation-fluorescence

More information

Optical Sectioning Deep Inside Live Embryos by Selective Plane Illumination Microscopy

Optical Sectioning Deep Inside Live Embryos by Selective Plane Illumination Microscopy Supporting Online Material Optical Sectioning Deep Inside Live Embryos by Selective Plane Illumination Microscopy 1 Material and methods 1.1 Setup Jan Huisken, Jim Swoger, Filippo Del Bene, Joachim Wittbrodt,

More information

Instructions for the Experiment

Instructions for the Experiment Instructions for the Experiment Excitonic States in Atomically Thin Semiconductors 1. Introduction Alongside with electrical measurements, optical measurements are an indispensable tool for the study of

More information

Supplementary Information for. Surface Waves. Angelo Angelini, Elsie Barakat, Peter Munzert, Luca Boarino, Natascia De Leo,

Supplementary Information for. Surface Waves. Angelo Angelini, Elsie Barakat, Peter Munzert, Luca Boarino, Natascia De Leo, Supplementary Information for Focusing and Extraction of Light mediated by Bloch Surface Waves Angelo Angelini, Elsie Barakat, Peter Munzert, Luca Boarino, Natascia De Leo, Emanuele Enrico, Fabrizio Giorgis,

More information

Horiba LabRAM ARAMIS Raman Spectrometer Revision /28/2016 Page 1 of 11. Horiba Jobin-Yvon LabRAM Aramis - Raman Spectrometer

Horiba LabRAM ARAMIS Raman Spectrometer Revision /28/2016 Page 1 of 11. Horiba Jobin-Yvon LabRAM Aramis - Raman Spectrometer Page 1 of 11 Horiba Jobin-Yvon LabRAM Aramis - Raman Spectrometer The Aramis Raman system is a software selectable multi-wavelength Raman system with mapping capabilities with a 400mm monochromator and

More information

Supplementary Figure S1. Schematic representation of different functionalities that could be

Supplementary Figure S1. Schematic representation of different functionalities that could be Supplementary Figure S1. Schematic representation of different functionalities that could be obtained using the fiber-bundle approach This schematic representation shows some example of the possible functions

More information

User Manual for HoloStudio M4 2.5 with HoloMonitor M4. Phase Holographic Imaging

User Manual for HoloStudio M4 2.5 with HoloMonitor M4. Phase Holographic Imaging User Manual for HoloStudio M4 2.5 with HoloMonitor M4 Phase Holographic Imaging 1 2 HoloStudio M4 2.5 Software instruction manual 2013 Phase Holographic Imaging AB 3 Contact us: Phase Holographic Imaging

More information

Nikon Instruments Europe

Nikon Instruments Europe Nikon Instruments Europe Recommendations for N-SIM sample preparation and image reconstruction Dear customer, We hope you find the following guidelines useful in order to get the best performance out of

More information

Nature Methods: doi: /nmeth Supplementary Figure 1. Schematic of 2P-ISIM AO optical setup.

Nature Methods: doi: /nmeth Supplementary Figure 1. Schematic of 2P-ISIM AO optical setup. Supplementary Figure 1 Schematic of 2P-ISIM AO optical setup. Excitation from a femtosecond laser is passed through intensity control and shuttering optics (1/2 λ wave plate, polarizing beam splitting

More information

AgilEye Manual Version 2.0 February 28, 2007

AgilEye Manual Version 2.0 February 28, 2007 AgilEye Manual Version 2.0 February 28, 2007 1717 Louisiana NE Suite 202 Albuquerque, NM 87110 (505) 268-4742 support@agiloptics.com 2 (505) 268-4742 v. 2.0 February 07, 2007 3 Introduction AgilEye Wavefront

More information

User Manual. Digital Compound Binocular LED Microscope. MicroscopeNet.com

User Manual. Digital Compound Binocular LED Microscope. MicroscopeNet.com User Manual Digital Compound Binocular LED Microscope Model MD82ES10 MicroscopeNet.com Table of Contents i. Caution... 1 ii. Care and Maintenance... 2 1. Components Illustration... 3 2. Installation...

More information

3B SCIENTIFIC PHYSICS

3B SCIENTIFIC PHYSICS 3B SCIENTIFIC PHYSICS Equipment Set for Wave Optics with Laser 1003053 Instruction sheet 06/18 Alf 1. Safety instructions The laser emits visible radiation at a wavelength of 635 nm with a maximum power

More information

ECEN. Spectroscopy. Lab 8. copy. constituents HOMEWORK PR. Figure. 1. Layout of. of the

ECEN. Spectroscopy. Lab 8. copy. constituents HOMEWORK PR. Figure. 1. Layout of. of the ECEN 4606 Lab 8 Spectroscopy SUMMARY: ROBLEM 1: Pedrotti 3 12-10. In this lab, you will design, build and test an optical spectrum analyzer and use it for both absorption and emission spectroscopy. The

More information

Topics. - How to calibrate the LSM scanner. - How to clean the microscope. - How to adjust the pinhole alignment. - How to adjust the Collimator

Topics. - How to calibrate the LSM scanner. - How to clean the microscope. - How to adjust the pinhole alignment. - How to adjust the Collimator Topics - How to calibrate the LSM scanner - How to measure the PSF - How to clean the microscope - How to adjust the pinhole alignment - How to adjust the Collimator How to calibrate the LSM scanner The

More information

μscope Microscopy Software

μscope Microscopy Software μscope Microscopy Software Pixelink μscope Essentials (ES) Software is an easy-to-use robust image capture tool optimized for productivity. Pixelink μscope Standard (SE) Software had added features, making

More information

GXCapture 8.1 Instruction Manual

GXCapture 8.1 Instruction Manual GT Vision image acquisition, managing and processing software GXCapture 8.1 Instruction Manual Contents of the Instruction Manual GXC is the shortened name used for GXCapture Square brackets are used to

More information

LSM 780 Confocal Microscope Standard Operation Protocol

LSM 780 Confocal Microscope Standard Operation Protocol LSM 780 Confocal Microscope Standard Operation Protocol Basic Operation Turning on the system 1. Sign on log sheet according to Actual start time 2. Check Compressed Air supply for the air table 3. Switch

More information

3B SCIENTIFIC PHYSICS

3B SCIENTIFIC PHYSICS 3B SCIENTIFIC PHYSICS Equipment Set for Wave Optics with Laser U17303 Instruction sheet 10/08 Alf 1. Safety instructions The laser emits visible radiation at a wavelength of 635 nm with a maximum power

More information

Life Science Instrumentation. New Generation. Light Sheet Fluorescence Microscope. Alph

Life Science Instrumentation. New Generation. Light Sheet Fluorescence Microscope. Alph Life Science Instrumentation Light Sheet Fluorescence Microscope New Generation Alph Modular Light Sheet Microscope Alpha 3 is a new generation of light sheet fluorescence microscope addressing the needs

More information

LOS 1 LASER OPTICS SET

LOS 1 LASER OPTICS SET LOS 1 LASER OPTICS SET Contents 1 Introduction 3 2 Light interference 5 2.1 Light interference on a thin glass plate 6 2.2 Michelson s interferometer 7 3 Light diffraction 13 3.1 Light diffraction on a

More information

Procedures for Performing Cryoelectron Microscopy on the FEI Sphera Microscope

Procedures for Performing Cryoelectron Microscopy on the FEI Sphera Microscope Procedures for Performing Cryoelectron Microscopy on the FEI Sphera Microscope The procedures given below were written specifically for the FEI Tecnai G 2 Sphera microscope. Modifications will need to

More information

Rapid Non linear Image Scanning Microscopy, Supplementary Notes

Rapid Non linear Image Scanning Microscopy, Supplementary Notes Rapid Non linear Image Scanning Microscopy, Supplementary Notes Calculation of theoretical PSFs We calculated the electrical field distribution using the wave optical theory developed by Wolf 1, and Richards

More information

Technical Explanation for Displacement Sensors and Measurement Sensors

Technical Explanation for Displacement Sensors and Measurement Sensors Technical Explanation for Sensors and Measurement Sensors CSM_e_LineWidth_TG_E_2_1 Introduction What Is a Sensor? A Sensor is a device that measures the distance between the sensor and an object by detecting

More information

Zeiss LSM 780 Protocol

Zeiss LSM 780 Protocol Zeiss LSM 780 Protocol 1) System Startup F Please note the sign-up policy. You must inform the facility at least 24 hours beforehand if you can t come; otherwise, you will receive a charge for unused time.

More information

Nature Neuroscience: doi: /nn Supplementary Figure 1. Optimized Bessel foci for in vivo volume imaging.

Nature Neuroscience: doi: /nn Supplementary Figure 1. Optimized Bessel foci for in vivo volume imaging. Supplementary Figure 1 Optimized Bessel foci for in vivo volume imaging. (a) Images taken by scanning Bessel foci of various NAs, lateral and axial FWHMs: (Left panels) in vivo volume images of YFP + neurites

More information

Nikon. King s College London. Imaging Centre. N-SIM guide NIKON IMAGING KING S COLLEGE LONDON

Nikon. King s College London. Imaging Centre. N-SIM guide NIKON IMAGING KING S COLLEGE LONDON N-SIM guide NIKON IMAGING CENTRE @ KING S COLLEGE LONDON Starting-up / Shut-down The NSIM hardware is calibrated after system warm-up occurs. It is recommended that you turn-on the system for at least

More information

CHAPTER1: QUICK START...3 CAMERA INSTALLATION... 3 SOFTWARE AND DRIVER INSTALLATION... 3 START TCAPTURE...4 TCAPTURE PARAMETER SETTINGS... 5 CHAPTER2:

CHAPTER1: QUICK START...3 CAMERA INSTALLATION... 3 SOFTWARE AND DRIVER INSTALLATION... 3 START TCAPTURE...4 TCAPTURE PARAMETER SETTINGS... 5 CHAPTER2: Image acquisition, managing and processing software TCapture Instruction Manual Key to the Instruction Manual TC is shortened name used for TCapture. Help Refer to [Help] >> [About TCapture] menu for software

More information

plasmonic nanoblock pair

plasmonic nanoblock pair Nanostructured potential of optical trapping using a plasmonic nanoblock pair Yoshito Tanaka, Shogo Kaneda and Keiji Sasaki* Research Institute for Electronic Science, Hokkaido University, Sapporo 1-2,

More information

Basic Users Manual for Tecnai-F20 TEM

Basic Users Manual for Tecnai-F20 TEM Basic Users Manual for Tecnai-F20 TEM NB: This document contains my personal notes on the operating procedure of the Tecnai F20 and may be used as a rough guide for those new to the microscope. It may

More information

Laser Speckle Reducer LSR-3000 Series

Laser Speckle Reducer LSR-3000 Series Datasheet: LSR-3000 Series Update: 06.08.2012 Copyright 2012 Optotune Laser Speckle Reducer LSR-3000 Series Speckle noise from a laser-based system is reduced by dynamically diffusing the laser beam. A

More information

Optika ISview. Image acquisition and processing software. Instruction Manual

Optika ISview. Image acquisition and processing software. Instruction Manual Optika ISview Image acquisition and processing software Instruction Manual Key to the Instruction Manual IS is shortened name used for OptikaISview Square brackets are used to indicate items such as menu

More information

Before you start, make sure that you have a properly calibrated system to obtain high-quality images.

Before you start, make sure that you have a properly calibrated system to obtain high-quality images. CONTENT Step 1: Optimizing your Workspace for Acquisition... 1 Step 2: Tracing the Region of Interest... 2 Step 3: Camera (& Multichannel) Settings... 3 Step 4: Acquiring a Background Image (Brightfield)...

More information

Training Guide for Carl Zeiss LSM 5 LIVE Confocal Microscope

Training Guide for Carl Zeiss LSM 5 LIVE Confocal Microscope Training Guide for Carl Zeiss LSM 5 LIVE Confocal Microscope AIM 4.2 Optical Imaging & Vital Microscopy Core Baylor College of Medicine (2017) Power ON Routine 1 2 Verify that main power switches on the

More information

1. Preliminary sample preparation

1. Preliminary sample preparation FEI Helios NanoLab 600 standard operating procedure Nicholas G. Rudawski ngr@ufl.edu (352) 392 3077 (office) (805) 252-4916 (cell) Last updated: 03/02/18 What this document provides: an overview of basic

More information

Supplementary Material (ESI) for Lab on a Chip This journal is The Royal Society of Chemistry 2011

Supplementary Material (ESI) for Lab on a Chip This journal is The Royal Society of Chemistry 2011 Supplemental Figure 1: A. Injection of oil (side channel) and cell extract (central channel) in a FFD sealed with a hydrophilic slide (clean glass surface): due to wetting of the cell extract with the

More information

Swept-Field User Guide

Swept-Field User Guide Swept-Field User Guide Note: for more details see the Prairie user manual at http://www.prairietechnologies.com/resources/software/prairieview.html Please report any problems to Julie Last (jalast@wisc.edu)

More information

MASSACHUSETTS INSTITUTE OF TECHNOLOGY Department of Electrical Engineering and Computer Science

MASSACHUSETTS INSTITUTE OF TECHNOLOGY Department of Electrical Engineering and Computer Science Student Name Date MASSACHUSETTS INSTITUTE OF TECHNOLOGY Department of Electrical Engineering and Computer Science 6.161 Modern Optics Project Laboratory Laboratory Exercise No. 3 Fall 2005 Diffraction

More information

Supplementary information, Figure S1A-S1H The thickness and the uniformity of the light sheet at different DOFs. By

Supplementary information, Figure S1A-S1H The thickness and the uniformity of the light sheet at different DOFs. By Supplementary information, Figure S1A-S1H The thickness and the uniformity of the light sheet at different DOFs. By imaging FITC-containing solution, the thickness of the light sheet generated by the P3A-DSLM

More information

X rays X-ray properties Denser material = more absorption = looks lighter on the x-ray photo X-rays CT Scans circle cross-sectional images Tumours

X rays X-ray properties Denser material = more absorption = looks lighter on the x-ray photo X-rays CT Scans circle cross-sectional images Tumours X rays X-ray properties X-rays are part of the electromagnetic spectrum. X-rays have a wavelength of the same order of magnitude as the diameter of an atom. X-rays are ionising. Different materials absorb

More information

Multi-resolution Cervical Cell Dataset

Multi-resolution Cervical Cell Dataset Report 37 Multi-resolution Cervical Cell Dataset Patrik Malm December 2013 Centre for Image Analysis Swedish University of Agricultural Sciences Uppsala University Uppsala 2013 Multi-resolution Cervical

More information

Zeiss LSM 510 Confocor III Training Notes. Center for Cell Analysis & Modeling

Zeiss LSM 510 Confocor III Training Notes. Center for Cell Analysis & Modeling Zeiss LSM 510 Confocor III Training Notes Center for Cell Analysis & Modeling Confocor 3 Start Up Go to System Module Turn on Main Switch, System/ PC, and Components Switches Do you need the arc lamp?

More information

MSE 595T Transmission Electron Microscopy. Laboratory III TEM Imaging - I

MSE 595T Transmission Electron Microscopy. Laboratory III TEM Imaging - I MSE 595T Basic Transmission Electron Microscopy TEM Imaging - I Purpose The purpose of this lab is to: 1. Make fine adjustments to the microscope alignment 2. Obtain a diffraction pattern 3. Obtain an

More information

FRAUNHOFER AND FRESNEL DIFFRACTION IN ONE DIMENSION

FRAUNHOFER AND FRESNEL DIFFRACTION IN ONE DIMENSION FRAUNHOFER AND FRESNEL DIFFRACTION IN ONE DIMENSION Revised November 15, 2017 INTRODUCTION The simplest and most commonly described examples of diffraction and interference from two-dimensional apertures

More information

Quality Performance, Innovative Design

Quality Performance, Innovative Design Dimensions Confocal Laser Scanning Biological Microscope Table size (mm): 1400(W) 800(D) * Table is not available from Olympus. Avoid placing the controller directly on the floor. Dimensions / Weight /

More information

Supplemental Method Information Zeiss LSM710

Supplemental Method Information Zeiss LSM710 Supplemental Method Information Zeiss LSM710 1 Under the Light Path window set up the confocal for imaging a green dye (Alexa488-EGFP). For example, set up the light path as shown here using the 488 nm

More information

Be aware that there is no universal notation for the various quantities.

Be aware that there is no universal notation for the various quantities. Fourier Optics v2.4 Ray tracing is limited in its ability to describe optics because it ignores the wave properties of light. Diffraction is needed to explain image spatial resolution and contrast and

More information

DIGITAL-MICROSCOPY CAMERA SOLUTIONS USB 3.0

DIGITAL-MICROSCOPY CAMERA SOLUTIONS USB 3.0 DIGITAL-MICROSCOPY CAMERA SOLUTIONS USB 3.0 PixeLINK for Microscopy Applications PixeLINK will work with you to choose and integrate the optimal USB 3.0 camera for your microscopy project. Ideal for use

More information

Single-photon excitation of morphology dependent resonance

Single-photon excitation of morphology dependent resonance Single-photon excitation of morphology dependent resonance 3.1 Introduction The examination of morphology dependent resonance (MDR) has been of considerable importance to many fields in optical science.

More information

Leica DMi8A Quick Guide

Leica DMi8A Quick Guide Leica DMi8A Quick Guide 1 Optical Microscope Quick Start Guide The following instructions are provided as a Quick Start Guide for powering up, running measurements, and shutting down Leica s DMi8A Inverted

More information

IC 2 S High Performance Objectives

IC 2 S High Performance Objectives M i c r o s c o p y f r o m C a r l Z e i s s IC 2 S igh Performance Objectives for Biomedical Applications with Laser Based Imaging Systems LSM,, ConfoCor, TIRF and ELYRA Carl Zeiss offers a large range

More information

Deep and fast live imaging with two-photon scanned light-sheet microscopy

Deep and fast live imaging with two-photon scanned light-sheet microscopy Nature Methods Deep and fast live imaging with two-photon scanned light-sheet microscopy Thai V Truong, Willy Supatto, Davis S Koos, John M Choi & Scott E Fraser Supplementary Figure 1 Supplementary Figure

More information

Zeiss LSM 880 Protocol

Zeiss LSM 880 Protocol Zeiss LSM 880 Protocol 1) System Startup Please note put sign-up policy. You must inform the facility at least 24 hours beforehand if you can t come; otherwise, you will receive a charge for unused time.

More information

Things to check before start-up.

Things to check before start-up. Byeong Cha Page 1 11/24/2009 Manual for Leica SP2 Confocal Microscope Enter you name, the date, the time, and the account number in the user log book. Things to check before start-up. Make sure that your

More information

ScanArray Overview. Principle of Operation. Instrument Components

ScanArray Overview. Principle of Operation. Instrument Components ScanArray Overview The GSI Lumonics ScanArrayÒ Microarray Analysis System is a scanning laser confocal fluorescence microscope that is used to determine the fluorescence intensity of a two-dimensional

More information

Basic Optics System OS-8515C

Basic Optics System OS-8515C 40 50 30 60 20 70 10 80 0 90 80 10 20 70 T 30 60 40 50 50 40 60 30 70 20 80 90 90 80 BASIC OPTICS RAY TABLE 10 0 10 70 20 60 50 40 30 Instruction Manual with Experiment Guide and Teachers Notes 012-09900B

More information

Description of options, upgrades and accessories for the laser beam stabilization system Compact

Description of options, upgrades and accessories for the laser beam stabilization system Compact Description of options, upgrades and accessories for the laser beam stabilization system Compact The basic configuration of the Compact laser beam stabilization system is fully equipped for stabilization

More information

Supplementary Figure 2. Schematic of the outer ring and glass base incorporated with glass microfluidics.

Supplementary Figure 2. Schematic of the outer ring and glass base incorporated with glass microfluidics. Supplementary Figure 1. Image of compressor mount ~CM! and bottom side of coverslip compressor ~CC!. a: The CC will thread into the CM and will push down on a 25 mm coverslip that is nestled within the

More information

Quintic Hardware Tutorial Camera Set-Up

Quintic Hardware Tutorial Camera Set-Up Quintic Hardware Tutorial Camera Set-Up 1 All Quintic Live High-Speed cameras are specifically designed to meet a wide range of needs including coaching, performance analysis and research. Quintic LIVE

More information

DIC Imaging using Laser Scanning Microscopes (LSMs) on Axio Imager Stands

DIC Imaging using Laser Scanning Microscopes (LSMs) on Axio Imager Stands DIC Imaging using Laser Scanning Microscopes (LSMs) on Axio Imager Stands Differential Interference Contrast (DIC) imaging is a technique used to increase contrast in brightfield images. In confocal systems,

More information

Image Capture TOTALLAB

Image Capture TOTALLAB 1 Introduction In order for image analysis to be performed on a gel or Western blot, it must first be converted into digital data. Good image capture is critical to guarantee optimal performance of automated

More information

AxioCam HR Success Through Performance

AxioCam HR Success Through Performance Microscopy from Carl Zeiss AxioCam HR Success Through Performance The high-resolution camera for digital documentation Superior performance for research and routine work brilliant quality documentation

More information

MorphCol Supplement 5

MorphCol Supplement 5 - Page 1 - MorphCol Supplement 5 Michael Knappertsbusch, 31.5.7 3.2.1.1 Analysis of variation in shape and size due to repeated automatic positioning of a single microfossil into the same position using

More information

Exp No.(8) Fourier optics Optical filtering

Exp No.(8) Fourier optics Optical filtering Exp No.(8) Fourier optics Optical filtering Fig. 1a: Experimental set-up for Fourier optics (4f set-up). Related topics: Fourier transforms, lenses, Fraunhofer diffraction, index of refraction, Huygens

More information

Super High Vertical Resolution Non-Contact 3D Surface Profiler BW-S500/BW-D500 Series

Super High Vertical Resolution Non-Contact 3D Surface Profiler BW-S500/BW-D500 Series Super High Vertical Resolution Non-Contact 3D Surface Profiler BW-S500/BW-D500 Series Nikon's proprietary scanning-type optical interference measurement technology achieves 1pm* height resolution. * Height

More information

LDA Laser-Doppler-Anemometry

LDA Laser-Doppler-Anemometry Related topics Interference, Doppler effect, scattering of light by small particles (Mie scattering), high and low-pass filters, sampling theorem, spectral power density, turbulence. Principle and task

More information

Stitching MetroPro Application

Stitching MetroPro Application OMP-0375F Stitching MetroPro Application Stitch.app This booklet is a quick reference; it assumes that you are familiar with MetroPro and the instrument. Information on MetroPro is provided in Getting

More information

Full-screen mode Popup controls. Overview of the microscope user interface, TEM User Interface and TIA on the left and EDS on the right

Full-screen mode Popup controls. Overview of the microscope user interface, TEM User Interface and TIA on the left and EDS on the right Quick Guide to Operating FEI Titan Themis G2 200 (S)TEM: TEM mode Susheng Tan Nanoscale Fabrication and Characterization Facility, University of Pittsburgh Office: M104/B01 Benedum Hall, 412-383-5978,

More information

End-of-Chapter Exercises

End-of-Chapter Exercises End-of-Chapter Exercises Exercises 1 12 are conceptual questions designed to see whether you understand the main concepts in the chapter. 1. Red laser light shines on a double slit, creating a pattern

More information

Instructions for Tecnai a brief start up manual

Instructions for Tecnai a brief start up manual Instructions for Tecnai a brief start up manual Version 3.0, 8.12.2015 Manual of Tecnai 12 transmission electron microscope located at Aalto University's Nanomicroscopy Center. More information of Nanomicroscopy

More information

Akinori Mitani and Geoff Weiner BGGN 266 Spring 2013 Non-linear optics final report. Introduction and Background

Akinori Mitani and Geoff Weiner BGGN 266 Spring 2013 Non-linear optics final report. Introduction and Background Akinori Mitani and Geoff Weiner BGGN 266 Spring 2013 Non-linear optics final report Introduction and Background Two-photon microscopy is a type of fluorescence microscopy using two-photon excitation. It

More information

Zeiss 780 Training Notes

Zeiss 780 Training Notes Zeiss 780 Training Notes Turn on Main Switch, System PC and Components Switches 780 Start up sequence Do you need the argon laser (458, 488, 514 nm lines)? Yes Turn on the laser s main power switch and

More information

INSTRUCTIONS FOR COURSE WORK 4 (AxioVert) Instructor: Anne Vaahtokari (MIU) 1. Purpose of the work

INSTRUCTIONS FOR COURSE WORK 4 (AxioVert) Instructor: Anne Vaahtokari (MIU) 1. Purpose of the work INSTRUCTIONS FOR COURSE WORK 4 (AxioVert) Instructor: Anne Vaahtokari (MIU) 1. Purpose of the work In this work, you will get familiar with an inverted epifluorescence microscope. Also, you will learn

More information

Eric B. Burgh University of Wisconsin. 1. Scope

Eric B. Burgh University of Wisconsin. 1. Scope Southern African Large Telescope Prime Focus Imaging Spectrograph Optical Integration and Testing Plan Document Number: SALT-3160BP0001 Revision 5.0 2007 July 3 Eric B. Burgh University of Wisconsin 1.

More information

INTERFEROMETER VI-direct

INTERFEROMETER VI-direct Universal Interferometers for Quality Control Ideal for Production and Quality Control INTERFEROMETER VI-direct Typical Applications Interferometers are an indispensable measurement tool for optical production

More information

LSM 510 Meta Training Notes

LSM 510 Meta Training Notes LSM 510 Meta Training Notes Turning on the system Turn on X-Cite power supply. This supplies light for epifluorescence for viewing your samples through the microscope. Turn on the remote control switch.

More information

Multicolor 4D Fluorescence Microscopy using Ultrathin Bessel Light sheets

Multicolor 4D Fluorescence Microscopy using Ultrathin Bessel Light sheets SUPPLEMENTARY MATERIAL Multicolor 4D Fluorescence Microscopy using Ultrathin Bessel Light sheets Teng Zhao, Sze Cheung Lau, Ying Wang, Yumian Su, Hao Wang, Aifang Cheng, Karl Herrup, Nancy Y. Ip, Shengwang

More information

DIC Imaging using Laser Scanning Microscopes (LSM) on Inverted Stands

DIC Imaging using Laser Scanning Microscopes (LSM) on Inverted Stands DIC Imaging using Laser Scanning Microscopes (LSM) on Inverted Stands Differential Interference Contrast (DIC) imaging is a technique used to increase contrast in brightfield images. In confocal systems,

More information

LSM 510 Training Notes

LSM 510 Training Notes LSM 510 Training Notes Turning on the system Turn on the arc lamp, found on the bench top left of the microscope. This supplies light for epifluorescence for viewing your samples through the microscope.

More information

EE119 Introduction to Optical Engineering Spring 2003 Final Exam. Name:

EE119 Introduction to Optical Engineering Spring 2003 Final Exam. Name: EE119 Introduction to Optical Engineering Spring 2003 Final Exam Name: SID: CLOSED BOOK. THREE 8 1/2 X 11 SHEETS OF NOTES, AND SCIENTIFIC POCKET CALCULATOR PERMITTED. TIME ALLOTTED: 180 MINUTES Fundamental

More information

Optical Coherence: Recreation of the Experiment of Thompson and Wolf

Optical Coherence: Recreation of the Experiment of Thompson and Wolf Optical Coherence: Recreation of the Experiment of Thompson and Wolf David Collins Senior project Department of Physics, California Polytechnic State University San Luis Obispo June 2010 Abstract The purpose

More information

880 Quantum Electronics Optional Lab Construct A Pulsed Dye Laser

880 Quantum Electronics Optional Lab Construct A Pulsed Dye Laser 880 Quantum Electronics Optional Lab Construct A Pulsed Dye Laser The goal of this lab is to give you experience aligning a laser and getting it to lase more-or-less from scratch. There is no write-up

More information

Leica SP8 TCS Users Manual

Leica SP8 TCS Users Manual Version : 07/08/0 Leica SP8 TCS Users Manual Start up:. Turn the PC Microscope, Scanner Power, Laser Power, and the Laser Emission key to on (bottom right of desk).. Turn on the fluorescent lamp (top left

More information

Supplementary Information. Stochastic Optical Reconstruction Microscopy Imaging of Microtubule Arrays in Intact Arabidopsis thaliana Seedling Roots

Supplementary Information. Stochastic Optical Reconstruction Microscopy Imaging of Microtubule Arrays in Intact Arabidopsis thaliana Seedling Roots Supplementary Information Stochastic Optical Reconstruction Microscopy Imaging of Microtubule Arrays in Intact Arabidopsis thaliana Seedling Roots Bin Dong 1,, Xiaochen Yang 2,, Shaobin Zhu 1, Diane C.

More information

Reading: Lenses and Mirrors; Applications Key concepts: Focal points and lengths; real images; virtual images; magnification; angular magnification.

Reading: Lenses and Mirrors; Applications Key concepts: Focal points and lengths; real images; virtual images; magnification; angular magnification. Reading: Lenses and Mirrors; Applications Key concepts: Focal points and lengths; real images; virtual images; magnification; angular magnification. 1.! Questions about objects and images. Can a virtual

More information

Supplementary Figure 1

Supplementary Figure 1 Supplementary Figure 1 Technical overview drawing of the Roadrunner goniometer. The goniometer consists of three main components: an inline sample-viewing microscope, a high-precision scanning unit for

More information

Introduction of New Products

Introduction of New Products Field Emission Electron Microscope JEM-3100F For evaluation of materials in the fields of nanoscience and nanomaterials science, TEM is required to provide resolution and analytical capabilities that can

More information

ECEN 4606, UNDERGRADUATE OPTICS LAB

ECEN 4606, UNDERGRADUATE OPTICS LAB ECEN 4606, UNDERGRADUATE OPTICS LAB Lab 3: Imaging 2 the Microscope Original Version: Professor McLeod SUMMARY: In this lab you will become familiar with the use of one or more lenses to create highly

More information

Reference and User Manual May, 2015 revision - 3

Reference and User Manual May, 2015 revision - 3 Reference and User Manual May, 2015 revision - 3 Innovations Foresight 2015 - Powered by Alcor System 1 For any improvement and suggestions, please contact customerservice@innovationsforesight.com Some

More information

2 How to operate the microscope/obtain an image

2 How to operate the microscope/obtain an image Morgagni Operating Instructions 50079 010912 2-1 2 ow to operate the microscope/obtain an image 2.1 Starting the microscope 2.1.1 Starting the microscope with several manually-operated steps 1. Turn on

More information

Microscopy from Carl Zeiss

Microscopy from Carl Zeiss Microscopy from Carl Zeiss Contents Page Contents... 1 Introduction... 1 Starting the System... 2 Introduction to ZEN Efficient Navigation... 5 Setting up the microscope... 10 Configuring the beam path

More information

AP Physics Problems -- Waves and Light

AP Physics Problems -- Waves and Light AP Physics Problems -- Waves and Light 1. 1974-3 (Geometric Optics) An object 1.0 cm high is placed 4 cm away from a converging lens having a focal length of 3 cm. a. Sketch a principal ray diagram for

More information

SUPPLEMENTARY INFORMATION

SUPPLEMENTARY INFORMATION pubs.acs.org/acssensors SUPPLEMENTARY INFORMATION Whole-Cell Photoacoustic Sensor Based on Pigment Relocalization Antonella Lauri 1,2,4,, Dominik Soliman 1,3,, Murad Omar 1, Anja Stelzl 1,2,4, Vasilis

More information

Supporting Information 1. Experimental

Supporting Information 1. Experimental Supporting Information 1. Experimental The position markers were fabricated by electron-beam lithography. To improve the nanoparticle distribution when depositing aqueous Ag nanoparticles onto the window,

More information

Kit for building your own THz Time-Domain Spectrometer

Kit for building your own THz Time-Domain Spectrometer Kit for building your own THz Time-Domain Spectrometer 16/06/2016 1 Table of contents 0. Parts for the THz Kit... 3 1. Delay line... 4 2. Pulse generator and lock-in detector... 5 3. THz antennas... 6

More information

PICO MASTER 200. UV direct laser writer for maskless lithography

PICO MASTER 200. UV direct laser writer for maskless lithography PICO MASTER 200 UV direct laser writer for maskless lithography 4PICO B.V. Jan Tinbergenstraat 4b 5491 DC Sint-Oedenrode The Netherlands Tel: +31 413 490708 WWW.4PICO.NL 1. Introduction The PicoMaster

More information