BRIGHTFIELD Olympus TH4-200 Olympus TH4-200 Please keep objectives off touching to avoid their loosening or damage.

Size: px
Start display at page:

Download "BRIGHTFIELD Olympus TH4-200 Olympus TH4-200 Please keep objectives off touching to avoid their loosening or damage."

Transcription

1 BRIGHTFIELD 1. Remove the microscope cover. 2. Dial the brightness adjustment knob (1) of Olympus TH4-200 halogen light source down to MIN. 3. Turn on the Olympus TH4-200 halogen light switch for transmitted light (2) to I (ON). 4. Check the light path. The light path selector knob (3) should be put to upper ( eye ) position. 5. Remove the ND5 and ND25 neutral density filters of filter pocket (4) from the light path of illumination turning them upwards. 6. While looking through eyepieces adjust the light intensity using the brightness adjustment knob (1). Try to avoid working at maximum of power of the lamp as that shortens life span of the lamp. 7. Adjust the interpupillary distance. While looking through the eyepieces, adjust the distance between the eyepieces (5) until the left and right fields of view coincide completely. 8. Make sure the 10x objective (6) is in place. 9. Place the slide with your specimen on the stage (7). 10. Find your specimen using the stage controls (8). 11. Focus specimen using fine/course focusing knobs (9). 12. Adjust the diopter: Close your right eye and focus on the specimen using the fine focus knob. Close your left eye and focus on the specimen using the diopter ring (10) on the right ocular. Open both eyes and confirm that the focus is comfortable. 13. If desired switch to the next objective by rotating the nosepiece (11) holding the nosepiece rim only and focus in the sample. Continue until you reach the desired magnification. Please keep objectives off touching to avoid their loosening or damage. 14. Establish Koehler illumination: Close field iris diaphragm (12) completely. Keep stage with sample on it fixed and avoid moving it in either of directions including focusing of the sample. Focus the image of the edge of the field diaphragm by raising or lowering the condenser using

2 the condenser focusing knob (13). Please note that focusing of sample and focusing of condenser is performed with different mechanisms. Check if the circle of light is centered in the field of view. Open field diaphragm till it start touching the round edge of field of view by its corners. If the diaphragm is not centered, use the two condenser centering screws (14) to move the field iris diaphragm image to the center of the field of view ( corners /edge gap is uniform). Open the field iris diaphragm until its image circumscribes the field of view. Match the opening of the condenser aperture iris diaphragm (15) with the N.A. of the objective in use in order to achieve the optimum objective performance. Start with the aperture diaphragm fully open. Start closing the aperture diaphragm and stop at the position of abrupt change in brightness of the image. This is the position of condenser objective aperture match. NOTE: Most specimen are usually low in contrast, reducing the diaphragm opening to 70%-80% of the N.A. value of the respective objective will generally provide an image of acceptable quality. That will correspond in roughly 20-30% decrease in brightness of the image. To check the opening (not recommended as it leads to contamination of optics especially if done by inexperienced users); a) remove one ocular and look down the tube, b) adjust the condenser iris so that 80% of the field is light, c) replace the ocular, or a) set condenser aperture iris scale to about 80% of the N.A. value of objective, b) example: N.A. of objective is 0.75, set the scale to 0.75 x 0.8 = Examine specimen and document image if necessary: To obtain still digitized images, use manual for Hamamatsu Digital Camera. 16. When finished: Remove the sample from the table and wipe it from oil with Whatman or Kimwhipes tissues if needed. Put 100x objective in upper position. Touch the lens carefully (no dragging through the lens!!) 3 or 4 times with dry and clean part of Whatman or Kimwhipes tissue first to remove oil from objective till no oil marks observed left on the tissue. Touch the lens carefully once with Whatman or Kimwhipes soaked with 70% alcohol.

3 Repeat the cleaning for 60x oil objective. Always use clean part of tissue or a new tissue to avoid cross-contamination between objectives. Turn the nosepiece (11) back until the empty position in turret is into place. Turn the light intensity down using the brightness adjustment knob (1). Turn off the halogen light switch (2) to 0 (OFF). Cover the microscope. Leave the room CLEAN and TIDY! FLUORESCENCE 1. Remove the microscope cover. 2. Put a record in Log Book indicating Life Time of the mercury bulb readings from OLYMPUS U-RFL-T Power Supply Unit (16). 3. Switch the UNIBLITZ VMM-3 Three Channel Shutter Driver ON. The switch is at far right corner of the device. Power and Driver active indicators for channels 2 and 3 (CH.# 2 and CH.# 3) are turned read. LED CH.#1 stays dark. Control over Channel 1 is performed through MetaMorph software interface. 4. In the case of using digital camera for acquisition switch HAMAMATSU camera driver ON. Find the switch at the lower left front corner of the controller. Power indicator of the controller will turn Green. 5. To turn on the mercury bulb, set the main switch (17) on the Supply Unit to I (ON). The light indicator (19) on the Power Supply Unit should go ON. Wait 5 to 10 minutes for the mercury bulb to stabilize. For safety reasons, the mercury bulb should stay ON for at least 30 minutes before turning OFF. 6. Setup microscope for brightfield imaging (see instructions on Brightfield # 2-14). 7. Engage the fluorescence mirror unit (18) until the proper filter cube is in place:

4 Number Excitation Emission Dichromatic in turret Mirror Unit filter filter mirror 1 Bright Field 2 U-MWU Applicable Fluorochrome 4, 6-diamidino-2-phenyl-indole HCI (DAPI), 7-amino-4-methylcoumarin-3-acetic acid (AMCA), 8-Anilino-1-naphthalene sulfonic acid Mg salt (ANS), Alexa Fluor 350, Aniline blue, bis-aminophenyl-oxadizole (BAO) -Feulgen, Calcein Blue, Calcofluor white, Cascade Blue, Coumarin maleimide (CPM), Diphenylhexatriene (DPH), Dansyl chloraide, Dansyl hydrazine, Dimethylaminonaphthalene-5-sulfonic acid (DANS), EBFP, ERTracker Blue-White DPX, Euchrysine (Acridine Orange), Fast Blue, Filipin, Fluorescamine, Granular Blue, Hoechst 33258, Hoechst 33342, Monochlorobimane, o-phtalaldehyde (OPT), SITS, Tetracycline, True Blue 3 U-MWIBA Applicable Fluorochrome Alexa Fluor 488, Alexa Fluor 488-Phalloidin, Alexa Fluor 532, BCECF/AM, BODIPY FL, BODIPY FL Ceramide, Calcein/AM, Carboxyfluorescein-diacetate (CFDA), Cy2, DiO C EGFP, EYFP, Eosin-isothiocyanate, Fluoresceindiacetate (FDA), Fluorescein-isothiocyanate (FITC), FluorX, GFP (WT), GFP (S65T), JC-1, MitoFluor Green, NBD-Phallacidin, Oregon Green 488, Rhodamine 123, Spectrum Green, SYTOX Green nucleic acid stain, YO PRO 1, YOYO 1 4 U-MWIG IF 565 Applicable Fluorochrome 7-Amino Actinomycin D, Acidic Fuchsin (Pararosaniline), Alexa Fluor 546, Alexa Fluor 568, Cy3, Dil, Ethidium Bromide, Ethidium homodimer-1, Evans Blue, LysoTracker Red DND-99, Merocyanine 540, MitoTracker Red CMXRos, NanoOrange, Nile Red, Pararosaniline-Feulgen, Phycoerythrin B (PE-B), Phycoerythrin R (PE-R), Primuline O, Propidium Iodide (PI), Pyronin Y, Resorufin, RFP (DsRed), Rhodamine B-isothiocyanate (RITC), Spectrum Orange, Tetramethylrhodamine-isothyocyanate (TRITC), Xylenol Orange 5 Empty 6 Empty 8. Turn the light intensity down all the way if you want to avoid bright field component appeared in the captured image using the brightness adjustment knob (1). 9. To allow light from the mercury bulb to reach the specimen, slide the shutter knob (19) to position marked as empty circle. 10. To improve image contrast and to prevent color fading of fluorescent light in other part than observed region, pull out the field iris diaphragm knob (20) so that the image of the field iris diaphragm just circumscribes the field of view. 11. Examine specimen and document image if necessary: Double-click MetaMorph icon in the folder Meta Imaging Series 7.1 from Desktop to run the acquisition software.

5 Select Active value from Illum: pop-up list. Click Live icon from the Toolbar to have the live image window appeared on the screen. Use Set Current Shutter Stage button to control electric shutter (21) when needed. Switch the shutter ON ( Set Current Shutter Stage icon background becomes bright) Select desired objective and focus in image looking through eyepieces. Switch the Light Path Selector towards Camera icon. Image may appear on the screen at this stage at sufficiently high fluorescence from the sample. Click Configure Acquisition icon from the Toolbar. Click Autoscale check box to have image appeared on the screen at low fluorescence from the sample. Adjust the exposure slide at the right most position of Toolbar Click Snap for capturing Black and White image Save the captured image in your sub-directory at D:\Users. It is recommended to keep the objective magnification information next to the sample number included in every image name. 12. To avoid bleaching, slide the shutter knob (19) to the position marked as filled circle whenever you are not observing the specimen. 13. When finished: Slide the shutter knob (19) to the position marked as filled circle. If mercury bulb has burned for at least 30 minutes, turn off the mercury bulb by setting the main switch (17) on the Power Supply Unit to O (OFF). The light indicator (22) on the Power Supply Unit should go out. NOTE: The mercury bulb needs to cool down for 2 hours before turning back on. Remove the sample from the table and wipe it from oil with Whatman or Kimwhipes tissues if needed. Put 100x objective in upper position. Touch the lens carefully (no dragging through the lens!!) 3 or 4 times with dry and clean part of Whatman or Kimwhipes tissue first to remove oil from objective till no oil marks observed left on the tissue. Touch the lens carefully once with Whatman or Kimwhipes soaked with 70% alcohol. Repeat the cleaning for 60x oil objective. Always use clean part of tissue or a new tissue to avoid cross-contamination between objectives. Turn the nosepiece (11) back until the empty position in turret is into place.

6 Engage the fluorescence mirror unit (18) back to BF (brightfield) - #1 Cover the microscope leaving the mercury burner housing uncovered. Leave the room CLEAN and TIDY! SIMULTANEOUS FLUORESCENCE AND BRIGHT FIELD 1. Remove the microscope cover. 2. Setup the microscope for Bright Field and Fluorescence imaging above Adjust the transmitted light for the best balance of fluorescence and DIC brightness using the brightness adjustment knob (1) Examine specimen and document image if necessary: 5. See #11 at Fluorescence section To avoid bleaching, slide the shutter knob (19) to the position marked as filled circle whenever you are not observing the specimen. 7. When finished: Slide the shutter knob (19) to the position marked as filled circle. If mercury bulb has burned for at least 30 minutes, turn off the mercury bulb by setting the main switch (17) on the Power Supply Unit to O (OFF). The light indicator (22) on the Power Supply Unit should go out. NOTE: The mercury bulb needs to cool down for 2 hours before turning back on. Remove the sample from the table and wipe it from oil with Whatman or Kimwhipes tissues if needed. Put 100x objective in upper position. Touch the lens carefully (no dragging through the lens!!) 3 or 4 times with dry and clean part of Whatman or Kimwhipes tissue first to remove oil from objective till no oil marks observed left on the tissue. Touch the lens carefully once with Whatman or Kimwhipes soaked with 70% alcohol. Repeat the cleaning for 60x oil objective. Always use clean part of tissue or a new tissue to avoid cross-contamination between objectives. Turn the nosepiece (11) back until the empty position in turret is into place. Engage the fluorescence mirror unit (18) back to BF (brightfield) - #1 Turn the light intensity down using the brightness adjustment knob (1). Turn off the halogen light switch (2) to 0 (OFF). Cover the microscope leaving the mercury burner housing uncovered. Leave the room CLEAN and TIDY!

7 APPENDIX 1: Outline Olympus BX71 Microscope

8 1 halogen light brightness adjustment knob 2 halogen light power switch 3 light path selector 4 filter pocket 5 eyepieces 6 objective 7 stage 8 stage controls 9 fine/course focusing knobs 10 diopter ring of the eyepiece 11 objectives nosepiece 12 field diaphragm 13 condenser focusing knob 14 condenser centering knobs 15 aperture diaphragm slider 16 mercury burner power supply 17 main switch of the power supply of the mercury burner 18 fluorescence mirror unit 19 UV light mechanical shutter knob 20 field diaphragm of UV illumination source 21 UV light electric shutter 22 Light indicator on the mercury burner power supply

Nikon E800 Operating Instructions.

Nikon E800 Operating Instructions. Nikon E800 Operating Instructions. You can request electronic copies of this manual by contacting lshats@jhsph.edu Copies are also available on the JHU MMI Department web site. Please send your comments

More information

Zeiss Axio Imager.A1 manual

Zeiss Axio Imager.A1 manual Zeiss Axio Imager.A1 manual Power-up protocol 1. Mercury lamp 2. Power strip on shelf 3. Computer The Mercury lamp should always be first-on and last-off. This prevents any electrical surges caused by

More information

Nikon E800 Microscope. Operating Instructions

Nikon E800 Microscope. Operating Instructions Nikon E800 Microscope Operating Instructions B Watson 12/2005 Table of contents: 1. The Nikon E800 Microscope 2. Turning the system ON and OFF 3. Selecting the light path 4. Operating in transmitted light

More information

Nikon E800 Operating Instructions.

Nikon E800 Operating Instructions. Nikon E800 Operating Instructions. You can request electronic copies of this manual by contacting imaging@fhcrc.org. Copies are also available on the Scientific Imaging web site. Please send your comments

More information

BX-61: Brightfield Instruction /Continue to scroll for Fluorescent Instuctions

BX-61: Brightfield Instruction /Continue to scroll for Fluorescent Instuctions BX-61: Brightfield Instruction /Continue to scroll for Fluorescent Instuctions Starting up: Schematic of Olympus BX-61. 1. Turn on Olympus microscope power box (left of microscope) with toggle switch on

More information

Using the Nikon TE2000 Inverted Microscope

Using the Nikon TE2000 Inverted Microscope Wellcome Trust Centre for Human Genetics Molecular Cytogenetics and Microscopy Core Using the Nikon TE2000 Inverted Microscope Fluorescence image acquisition using Scanalytic s IPLab software and the B&W

More information

Olympus Fluoview 1000S Spectral Confocal Microscope Introduction to the NRI-MCDB Microscopy Facility Spectral Confocal Microscope

Olympus Fluoview 1000S Spectral Confocal Microscope Introduction to the NRI-MCDB Microscopy Facility Spectral Confocal Microscope Olympus Fluoview 1000S Spectral Confocal Microscope Introduction to the NRI-MCDB Microscopy Facility Spectral Confocal Microscope Improved Optics More Lasers 405 diode 440 diode 488 Argon 515 Argon 559

More information

Instructions for Making On-Line Reservations for Microscopes in NB11-204

Instructions for Making On-Line Reservations for Microscopes in NB11-204 Instructions for Making On-Line Reservations for Microscopes in NB11-204 1. Log into Mail using Mail.swmed.edu 2. Log in using your university id and password. 3. Click the Calendar Tab at the top right

More information

SHORT INSTRUCTIONS FOR OPERATING LSM1/2 (Zeiss LSM510) AT CIAN Version 1.4, September 2014

SHORT INSTRUCTIONS FOR OPERATING LSM1/2 (Zeiss LSM510) AT CIAN Version 1.4, September 2014 CIAN LSM1 or LSM2 short instructions, version 1.4, September 2014 page 1 of 6 SHORT INSTRUCTIONS FOR OPERATING LSM1/2 (Zeiss LSM510) AT CIAN Version 1.4, September 2014 Before starting To work with LSM1

More information

b. Turn the power switch and key to on position for blue laser.

b. Turn the power switch and key to on position for blue laser. OLYMPUS FLUOVIEW 300 CONFOCAL MICOSCOPE OPERATION PROCEDURE 1. Turn ON microscope in this order: 1) Turn on mercury lamp (Note: once the mercury lamp is turned off, DO NOT turn it back on for at least

More information

Characterization Microscope Nikon LV150

Characterization Microscope Nikon LV150 Characterization Microscope Nikon LV150 Figure 1: Microscope Nikon LV150 Introduction This upright optical microscope is designed for investigating up to 150 mm (6 inch) semiconductor wafers but can also

More information

EPIFLUORESCENCE &/OR BRIGHTFIELD MICROSCOPY

EPIFLUORESCENCE &/OR BRIGHTFIELD MICROSCOPY EPIFLUORESCENCE &/OR BRIGHTFIELD MICROSCOPY TURN ON THE FOLLOWING EQUIPMENT The fluorescent light (if needed) The power strip for the microscope and accessories The CoolSNAP HQ camera on the right (Turn

More information

Zeiss Axioskop II. The AIF's "routine" light microscope. (Installed 8/24/04)AxioCam installed July 11th 2005

Zeiss Axioskop II. The AIF's routine light microscope. (Installed 8/24/04)AxioCam installed July 11th 2005 Zeiss Axioskop II The AIF's "routine" light microscope. (Installed 8/24/04)AxioCam installed July 11th 2005 Featuring: Phase Contrast Darkfield DIC/Nomarski Brightfield Fluorescent filters for Dapi, FITC,Rhodamine

More information

Olympus IX71 Microscope and DP71 Camera Instructions

Olympus IX71 Microscope and DP71 Camera Instructions Olympus IX71 Microscope and DP71 Camera Instructions Microscopy in Medicine (MiM) Core Emory University Department of Medicine 1 Olympus IX71 Image Capture Procedure 2 3 1. STARTING-UP PROCEDURE: Remove

More information

START-UP PROCEDURE 1 THE MICROSCOPE STAND 3 OBJECTIVES 5 STARTING WITH LAS (SOFTWARE) AND SETTING UP THE MICROSCOPE STAND 7

START-UP PROCEDURE 1 THE MICROSCOPE STAND 3 OBJECTIVES 5 STARTING WITH LAS (SOFTWARE) AND SETTING UP THE MICROSCOPE STAND 7 Leica DMI AF6000LX Table of contents START-UP PROCEDURE 1 THE MICROSCOPE STAND 3 OBJECTIVES 5 STARTING WITH LAS (SOFTWARE) AND SETTING UP THE MICROSCOPE STAND 7 ACQUIRE MODULE 6 SETTING THE LIGHTPATH 6

More information

Guide to Confocal 5. Starting session

Guide to Confocal 5. Starting session Guide to Confocal 5 Remember that when booking and before starting session you can check for any problems at https://www.bris.ac.uk/biochemistry/uobonly/cif/index.html Starting session Switch on microscope

More information

Motorized Axio Observer Start-up instructions

Motorized Axio Observer Start-up instructions Start-up instructions 1. If using fluorescence turn on Fluorescent light source. TL light Source (Hal 100) 2. Turn on microscope using switch on lower left side of the microscope. 3. If imaging, turn on

More information

Widefield-NikonEclipseTE200-ORCA Nikon Eclipse TE200 Inverted Microscope with Hamamatsu 1394 Orca-ER Cooled CCD Camera and Micromanager Software

Widefield-NikonEclipseTE200-ORCA Nikon Eclipse TE200 Inverted Microscope with Hamamatsu 1394 Orca-ER Cooled CCD Camera and Micromanager Software Widefield-NikonEclipseTE200-ORCA Nikon Eclipse TE200 Inverted Microscope with Hamamatsu 1394 Orca-ER Cooled CCD Camera and Micromanager Software September 2007 Check website for most current User Guide

More information

Quick Start Guide. Leica SP5 X

Quick Start Guide. Leica SP5 X Quick Start Guide Leica SP5 X Please note: Some of the information in this guide was taken from Leica Microsystems Leica TCS SP5 LAS AF Guide for New Users. This work is licensed under the Creative Commons

More information

Operation Guide for the Leica SP2 Confocal Microscope Bio-Imaging Facility Hunter College October 2009

Operation Guide for the Leica SP2 Confocal Microscope Bio-Imaging Facility Hunter College October 2009 Operation Guide for the Leica SP2 Confocal Microscope Bio-Imaging Facility Hunter College October 2009 Introduction of Fluoresence Confocal Microscopy The first confocal microscope was invented by Princeton

More information

User Manual. Digital Compound Binocular LED Microscope. MicroscopeNet.com

User Manual. Digital Compound Binocular LED Microscope. MicroscopeNet.com User Manual Digital Compound Binocular LED Microscope Model MD82ES10 MicroscopeNet.com Table of Contents i. Caution... 1 ii. Care and Maintenance... 2 1. Components Illustration... 3 2. Installation...

More information

STEINDORFF METALLURGICAL MICROSCOPE. NYMCS-620 Instruction Manual

STEINDORFF METALLURGICAL MICROSCOPE. NYMCS-620 Instruction Manual METALLURGICAL MICROSCOPE NYMCS-620 Instruction Manual It is recommended strongly that you study this manual thoroughly before using the microscope. Retain this manual in an easily accessible place near

More information

Zeiss AxioImager.Z2 Brightfield Protocol

Zeiss AxioImager.Z2 Brightfield Protocol Zeiss AxioImager.Z2 Brightfield Protocol 1) System Startup Please note put sign-up policy. You must inform the facility at least 24 hours beforehand if you can t come; otherwise, you will receive a charge

More information

CAPTURING IMAGES ON THE HIGH-MAGNIFICATION MICROSCOPE

CAPTURING IMAGES ON THE HIGH-MAGNIFICATION MICROSCOPE University of Virginia ITC Academic Computing Health Sciences CAPTURING IMAGES ON THE HIGH-MAGNIFICATION MICROSCOPE Introduction The Olympus BH-2 microscope in ACHS s microscope lab has objectives from

More information

Nikon Ti-E Microscope Manual. Rightmire Hall Ohio State University. Director: Tony Brown Rightmire

Nikon Ti-E Microscope Manual. Rightmire Hall Ohio State University. Director: Tony Brown Rightmire Nikon Ti-E Microscope Manual Rightmire Hall Ohio State University Director: Tony Brown Rightmire 060 292-1205 brown.2302@osu.edu Facility Manager: Paula Monsma Rightmire 062 293-0939 292-1367 monsma.1@osu.edu

More information

OMM300. Inverted Metallurgical Microscope

OMM300. Inverted Metallurgical Microscope OMM300 Inverted Metallurgical Microscope Instruction Manual Please read the instructions carefully before operating CONTENTS Safety 2 Parts List 2 Features 3 Assembly 5 Operation 7 Maintenance 9 Specifications

More information

Biological Microscope Manual

Biological Microscope Manual Version No.: V1.2 Series Biological Microscope Manual This manual expatiates the using method, troubleshooting and maintenance about MT-50 series biological microscope. Please study this manual thoroughly

More information

Things to check before start-up.

Things to check before start-up. Byeong Cha Page 1 11/24/2009 Manual for Leica SP2 Confocal Microscope Enter you name, the date, the time, and the account number in the user log book. Things to check before start-up. Make sure that your

More information

LSM 510 META in Chang Gung University

LSM 510 META in Chang Gung University Content LSM 510 META in Chang ung University LSM 510 META 路 理 The features and applications of LSM 510 META 01-09 Introduction of the hardware 10-12 Fluorescence observation in conventional microscope

More information

TEKSCOPE MICROSCOPE. Models N2 Series USER S MANUAL

TEKSCOPE MICROSCOPE. Models N2 Series USER S MANUAL TEKSCOPE MICROSCOPE Models N2 Series USER S MANUAL Contents Before use 1 1.Nomenclature. 2 2.Operation 4 2-1 Angle of observation.. 4 2-2 Set the specimen slide.. 4 2-3 Set illumination 4 2-4 Adjust focus

More information

Brightfield Microscopy and Image Acquisition on Spotcam1. by Ryan Taylor/Nancy Kleene Last modified 10/02/05 by Birgit Ehmer

Brightfield Microscopy and Image Acquisition on Spotcam1. by Ryan Taylor/Nancy Kleene Last modified 10/02/05 by Birgit Ehmer Brightfield Microscopy and Image Acquisition on Spotcam1 by Ryan Taylor/Nancy Kleene Last modified 10/02/05 by Birgit Ehmer Log onto the computer. Enter your username and password to log onto the server.

More information

Leica SP8 TCS Users Manual

Leica SP8 TCS Users Manual Version : 07/08/0 Leica SP8 TCS Users Manual Start up:. Turn the PC Microscope, Scanner Power, Laser Power, and the Laser Emission key to on (bottom right of desk).. Turn on the fluorescent lamp (top left

More information

ZEISS LSM510META confocal manual

ZEISS LSM510META confocal manual ZEISS LSM510META confocal manual Switching on the system 1) Switch on the Remote Control button located on the table to the right of the microscope. This is the main switch for the whole system including

More information

TABLE OF CONTENTS. Safety notes i. Care and Maintenance. ii. 1. Components Illustration Installation of Components.. 4

TABLE OF CONTENTS. Safety notes i. Care and Maintenance. ii. 1. Components Illustration Installation of Components.. 4 TABLE OF CONTENTS Safety notes i Care and Maintenance. ii 1. Components Illustration... 1 2. Installation of Components.. 4 2.1 Installation Diagram... 4 2.2 Installation Procedures 5 3. Operation...11

More information

TRAINING MANUAL. Olympus FV1000

TRAINING MANUAL. Olympus FV1000 TRAINING MANUAL Olympus FV1000 September 2014 TABLE OF CONTENTS A. Start-Up Procedure... 1 B. Visual Observation under the Microscope... 1 C. Image Acquisition... 4 A brief Overview of the Settings...

More information

AxioVision 4.5 Brightfield Image Capture Procedure

AxioVision 4.5 Brightfield Image Capture Procedure AxioVision 4.5 Brightfield Image Capture Procedure 1. STARTING-UP PROCEDURE: Remove blue dust cover and place on shelf under microscope. Turn on the halogen lamp by pushing the switch at the back right

More information

Simplified Instructions: Zeiss Brightfield Microscope S1000

Simplified Instructions: Zeiss Brightfield Microscope S1000 Contents General Microscope Set-Up Adjust Illumination Focus Condenser Open Software Image Capture Settings Shading & Color Corrections Image Capture & Viewing Scaling and Measurements Synopsis of Other

More information

Zeiss LSM 880 Protocol

Zeiss LSM 880 Protocol Zeiss LSM 880 Protocol 1) System Startup Please note put sign-up policy. You must inform the facility at least 24 hours beforehand if you can t come; otherwise, you will receive a charge for unused time.

More information

Operating Instructions for Zeiss LSM 510

Operating Instructions for Zeiss LSM 510 Operating Instructions for Zeiss LSM 510 Location: GNL 6.312q (BSL3) Questions? Contact: Maxim Ivannikov, maivanni@utmb.edu 1 Attend A Complementary Training Before Using The Microscope All future users

More information

Microscope ECLIPSE 80i Instructions <Reference>

Microscope ECLIPSE 80i Instructions <Reference> M318E 03.12.CF.1(2/2) Microscope ECLIPSE 80i Instructions Introduction Thank you for purchasing this Nikon product. This instruction manual, which describes basic microscope operations, is

More information

OM FL400. Reflected Light Fluorescence Microscope. Instruction Manual. Please read instructions carefully before using microscope.

OM FL400. Reflected Light Fluorescence Microscope. Instruction Manual. Please read instructions carefully before using microscope. OM FL400 Reflected Light Fluorescence Microscope Instruction Manual Please read instructions carefully before using microscope. Contents Safety ---------------------------------------------- 2 Parts List

More information

Zeiss Axiovert 135 Fluorescence Microscope Quick Guide / Operations Manual (v. 1.0 February 09)

Zeiss Axiovert 135 Fluorescence Microscope Quick Guide / Operations Manual (v. 1.0 February 09) University of Chicago Integrated Light Microscopy Core Dr. Vytas Bindokas, Director http://digital.bsd.uchicago.edu By: Christine Labno, Assistant Director Room: AB-129 Phone: 4-9040 Zeiss Axiovert 135

More information

Nikon Eclipse Ti A1-A Confocal Operating Manual. Start-up. Microscope

Nikon Eclipse Ti A1-A Confocal Operating Manual. Start-up. Microscope Nikon Eclipse Ti A1-A Confocal Operating Manual Start-up 1. Turn on Excite Fluorescent light power supply- metal halide. a. Cool down as for mercury bulb b. Wheel closed liquid light guide 2. Turn on power

More information

User Manual. Trinocular Metallurgical Microscope. MicroscopeNet.com

User Manual. Trinocular Metallurgical Microscope. MicroscopeNet.com User Manual Trinocular Metallurgical Microscope Model M83MPTR MicroscopeNet.com Table of Contents i. Caution.. 1 ii. Care and Maintenance... 2 1. Components Illustration..... 3 2. Installation...4 3. Operation

More information

Widefield 1. Switching on

Widefield 1. Switching on Widefield 1 Switching on 1. Ignite DG5 lamp - must be switched on first (if previous user has switched off, wait 30 min before igniting) 2. Wait 5s and then turn on the main DG5 controller switch. 3. DG5

More information

Nikon C1si Spectral Laser Scanning Confocal Microscope. User Guide

Nikon C1si Spectral Laser Scanning Confocal Microscope. User Guide Nikon C1si Spectral Laser Scanning Confocal Microscope User Guide Contents: C1Si Turn-On/ShutDown Procedures... 2 Overview... 4 Setup for epi-illumination to view through the eyepieces:... 5 Setup for

More information

Therefore, all descriptions and illustrations in this instruction manual, including all specifications are subject to change without notice.

Therefore, all descriptions and illustrations in this instruction manual, including all specifications are subject to change without notice. We are constantly endeavouring to improve our instruments and to adapt them to the requirements of modern research techniques and testing methods. This involves modification to the mechanical structure

More information

LSM 710 Confocal Microscope Standard Operation Protocol

LSM 710 Confocal Microscope Standard Operation Protocol LSM 710 Confocal Microscope Standard Operation Protocol Basic Operation Turning on the system 1. Switch on Main power switch 2. Switch on System / PC power button 3. Switch on Components power button 4.

More information

Microscope Confocal LSM510 META

Microscope Confocal LSM510 META Microscope Confocal LSM510 META Welcome to the Zeiss LSM 510 Meta Confocal tutorial. Before using the LSM 510 META, Log off any other computer that is open with your personal login. You will need to put

More information

Quick Guide for Zeiss 710 Laser Scanning Confocal MGH Cancer Center

Quick Guide for Zeiss 710 Laser Scanning Confocal MGH Cancer Center Quick Guide for Zeiss 710 Laser Scanning Confocal MGH Cancer Center For any questions or concerns, please contact: Linda Nieman lnieman@mgh.harvard.edu Office: (617) 643-9684 Cell: (512) 565-8076 Chenyue

More information

Delphi-X Observer Superior imaging. microscopes

Delphi-X Observer Superior imaging. microscopes Delphi-X Observer Superior imaging microscopes for ADVANCED LIFE SCIENCES State-of-the-art microscopes for advanced Life Science applications The ergonomic design and outstanding state-of-the-art objectives

More information

Zeiss AxioObserver with ApoTome

Zeiss AxioObserver with ApoTome Zeiss AxioObserver with ApoTome Quick Start User Guide LSU Health Sciences Center-Shreveport Research Core Facility (RCF) Microscopy Table of Contents 1 Start up the system.. Page 3 2 Touch screen controller

More information

Delphi-X Observer Superior imaging. microscopes

Delphi-X Observer Superior imaging. microscopes Delphi-X Observer Superior imaging microscopes for ADVANCED LIFE SCIENCES State-of-the-art microscopes for advanced Life Science applications The ergonomic design and outstanding state-of-the-art objectives

More information

Simplified Instructions: Olympus Widefield Microscope S1230

Simplified Instructions: Olympus Widefield Microscope S1230 Contents General Microscope Operation Simple Image Capture Multi-Wavelength Capture Z-Series Timelapse Combining Capture Modes Synopsis of Other Functions Pages 2-23 24-40 41-47 48-56 57-59 60-68 69-83

More information

Manual for BMS E1 eplan series, compound microscope

Manual for BMS E1 eplan series, compound microscope Manual for BMS E1 eplan series, compound microscope The compound microscope allows it to study, at cell level, structures of textures of botanical and zoological nature. (e.g. slides of roots, leaves and

More information

Imaging Introduction. September 24, 2010

Imaging Introduction. September 24, 2010 Imaging Introduction September 24, 2010 What is a microscope? Merriam-Webster: an optical instrument consisting of a lens or combination of lenses for making enlarged images of minute objects; especially:

More information

MICROSCOPY FOR THE DEVELOPMENTAL BIOLOGY STUDENT...

MICROSCOPY FOR THE DEVELOPMENTAL BIOLOGY STUDENT... MICROSCOPY FOR THE DEVELOPMENTAL BIOLOGY STUDENT... You will be using two configurations of microscope during the course of the semester to observe specimens and record your results: compound microscopes

More information

Nikon Eclipse Ti2-E Widefield/Spinning Disk Confocal Microscope Standard Operation Protocol

Nikon Eclipse Ti2-E Widefield/Spinning Disk Confocal Microscope Standard Operation Protocol Nikon Eclipse Ti-E Widefield/Spinning Disk Confocal Microscope Standard Operation Protocol Please sign on the log sheet before switching on system. Turn on system Turn on A only if confocal mode or laser

More information

Contents STARTUP MICROSCOPE CONTROLS CAMERA CONTROLS SOFTWARE CONTROLS EXPOSURE AND CONTRAST MONOCHROME IMAGE HANDLING

Contents STARTUP MICROSCOPE CONTROLS CAMERA CONTROLS SOFTWARE CONTROLS EXPOSURE AND CONTRAST MONOCHROME IMAGE HANDLING Operations Guide Contents STARTUP MICROSCOPE CONTROLS CAMERA CONTROLS SOFTWARE CONTROLS EXPOSURE AND CONTRAST MONOCHROME IMAGE HANDLING Nikon Eclipse 90i Operations Guide STARTUP Startup Powering Up Fluorescence

More information

Easy Kohler Illumination Method

Easy Kohler Illumination Method Easy Kohler Illumination Method ACADEMIC SKILLS CENTRE (ASC) A. Silverberg Completion of a Kohler illumination method is required before a microscope can be used efficiently. The Kohler method is designed

More information

TouchBright Ver. 7.51

TouchBright Ver. 7.51 TouchBright Ver. 7.51 High-Performance LED Excitation System Efficient Use Long Lifetime Brightest LEDs Compact Design High-Performance Live Cell Instrument Co., LTD www.touchbrightled.com TouchBright

More information

Industrial Microscope ECLIPSE LV100DA Instructions

Industrial Microscope ECLIPSE LV100DA Instructions M371 E 06.1.NF.1 (2/3) Industrial Microscope ECLIPSE LV100DA Instructions Thank you for purchasing the Nikon product. This instruction manual is written for the users of the Nikon Industrial Microscope

More information

Quick Guide for Zeiss 710 Laser Scanning Confocal MGH Cancer Center

Quick Guide for Zeiss 710 Laser Scanning Confocal MGH Cancer Center Quick Guide for Zeiss 710 Laser Scanning Confocal MGH Cancer Center For any questions or concerns, please contact: Linda Nieman lnieman@mgh.harvard.edu Office: (617) 643-9684 Cell: (512) 565-8076 Chenyue

More information

INSTRUCTIONS FOR COURSE WORK 4 (AxioVert) Instructor: Anne Vaahtokari (MIU) 1. Purpose of the work

INSTRUCTIONS FOR COURSE WORK 4 (AxioVert) Instructor: Anne Vaahtokari (MIU) 1. Purpose of the work INSTRUCTIONS FOR COURSE WORK 4 (AxioVert) Instructor: Anne Vaahtokari (MIU) 1. Purpose of the work In this work, you will get familiar with an inverted epifluorescence microscope. Also, you will learn

More information

ML7520 ML7530 DIOPTER ADJUSTMENT RING BINOCULAR BODY, INCLINED 30. (a) Field Iris Control Lever. (c) Filter Slots EYEPIECES, KHW10X

ML7520 ML7530 DIOPTER ADJUSTMENT RING BINOCULAR BODY, INCLINED 30. (a) Field Iris Control Lever. (c) Filter Slots EYEPIECES, KHW10X JAPAN DIOPTER ADJUSTMENT RING BINOCULAR BODY, INCLINED 30 (a) Field Iris Control Lever (c) Filter Slots EYEPIECES, KHW10X ANALYZER CONTROL LEVER (b) Aperture Iris Control Lever LIGHT SOURCE HOUSING VERTICAL

More information

Operating Checklist for using the Laser Scanning Confocal Microscope. Leica TCS SP5.

Operating Checklist for using the Laser Scanning Confocal Microscope. Leica TCS SP5. Smith College August 2010 Operating Checklist for using the Laser Scanning Confocal Microscope Leica TCS SP5. CONTENT, page no. Startup, 1 Initial set-up, 1 Software, 2 Microscope Specimen observation

More information

Nikon TE300 Eclipse Wide-Field Microscope

Nikon TE300 Eclipse Wide-Field Microscope Nikon TE300 Eclipse Wide-Field Microscope User Guide LSU Health Science Center-Shreveport Research Core Facility 1 User manual for Nikon Elements software Equipment: Nikon TE300 Eclipse microscope Photometrics

More information

Olympus Time-lapse Microscope Basic operation

Olympus Time-lapse Microscope Basic operation Olympus Time-lapse Microscope Basic operation To start up the microscope 1. Switch on the Olympus UCB. (label as ) 1 Power Switch 2 2. Switch on the MT10. (label as ) Power Switch Page 1 of 18 3 3. Switch

More information

DIC Imaging using Laser Scanning Microscopes (LSMs) on Axio Imager Stands

DIC Imaging using Laser Scanning Microscopes (LSMs) on Axio Imager Stands DIC Imaging using Laser Scanning Microscopes (LSMs) on Axio Imager Stands Differential Interference Contrast (DIC) imaging is a technique used to increase contrast in brightfield images. In confocal systems,

More information

Everest System / Slidebook Operating Procedures

Everest System / Slidebook Operating Procedures Everest System / Slidebook Operating Procedures NOTICE: This guide is meant to supplement training, not replace it. All users must be trained first hand by a core employee. Training of others in your lab

More information

DIC Imaging using Laser Scanning Microscopes (LSM) on Inverted Stands

DIC Imaging using Laser Scanning Microscopes (LSM) on Inverted Stands DIC Imaging using Laser Scanning Microscopes (LSM) on Inverted Stands Differential Interference Contrast (DIC) imaging is a technique used to increase contrast in brightfield images. In confocal systems,

More information

User Manual. Digital Compound Binocular LED Microscope. MicroscopeNet.com

User Manual. Digital Compound Binocular LED Microscope. MicroscopeNet.com User Manual Digital Compound Binocular LED Microscope Model MD827S30L series MicroscopeNet.com Table of Contents i. Caution... 1 ii. Care and Maintenance... 2 1. Components Illustration... 3 2. Installation...

More information

STEINDORFF. Inverted Fluorescence Microscope NYMCS-702. Instruction Manual

STEINDORFF. Inverted Fluorescence Microscope NYMCS-702. Instruction Manual Inverted Fluorescence Microscope NYMCS-702 Instruction Manual This manual is written for Inverted Fluorescence Microscope NYMCS-702. For safety and for keeping the best performance, making you familiar

More information

QUICKSTART GUIDE: WIDEFIELD WF3 Zeiss Cell Observer Live Cell Imaging System (SAF, ROOM 409) Imperial College London

QUICKSTART GUIDE: WIDEFIELD WF3 Zeiss Cell Observer Live Cell Imaging System (SAF, ROOM 409) Imperial College London Imperial College London Facility for Imaging by Light Microscopy QUICKSTART GUIDE: WIDEFIELD WF3 Zeiss Cell Observer Live Cell Imaging System (SAF, ROOM 409) Observing Life As It Happens Startup procedure...

More information

Confocal 510 Tutorial

Confocal 510 Tutorial Confocal 510 Tutorial You will have to log on in order to use the microscope. You will be charged according the time you are logged in, so please don t forget to log out after you are done. If you don

More information

Brief manual how to start and close the Leica sp2 Confocal. (TCS SP2 AOBS system mounted on a DM IRE2)

Brief manual how to start and close the Leica sp2 Confocal. (TCS SP2 AOBS system mounted on a DM IRE2) Brief manual how to start and close the Leica sp2 Confocal (TCS SP2 AOBS system mounted on a DM IRE2) A. Switching on hardware B. Acquiring and saving images C. Switching off the microscope D. Good working

More information

Reichert Univar Manual

Reichert Univar Manual Reichert Univar Manual Translated from the 11/1975 German language edition, with slight modifications. William R. Porter San Marcos CA USA 2017 v 1.3 Notes This is a very slightly-modified, new (2017)

More information

Zoom Stereo Microscope NYMCS-360 Instruction Manual

Zoom Stereo Microscope NYMCS-360 Instruction Manual Zoom Stereo Microscope NYMCS-60 Instruction Manual This manual is written for stereo microscope NYMCS-60. To ensure the safety, obtain optimum performance and to familiarize yourself fully with the use

More information

User Manual. Trinocular Infinity Compound LED Microscope. MicroscopeNet.com

User Manual. Trinocular Infinity Compound LED Microscope. MicroscopeNet.com User Manual Trinocular Infinity Compound LED Microscope Model M8333Z series MicroscopeNet.com Table of Contents i. Caution... 1 ii. Care and Maintenance... 2 1. Components Illustration... 3 2. Installation...

More information

VERTICAL FLUORESCENCE ILLUMINATOR ATTACHMENT

VERTICAL FLUORESCENCE ILLUMINATOR ATTACHMENT VERTICAL FLUORESCENCE ILLUMINATOR ATTACHMENT This instruction manual has been written for use of the Vertical Fluorescence Illuminator Model AH-RF L-LB in conjunction with the Universal Research Microscope

More information

Zeiss Deconvolution Microscope: A Quick Guide

Zeiss Deconvolution Microscope: A Quick Guide Zeiss Deconvolution Microscope: A Quick Guide Start-up Uncover microscope. Do not put dust cover on the floor. Plug in both cameras. The default camera is the AxioCam HRm (monochrome camera) for fluorescence

More information

STEINDORFF NYMC Polarizing Microscope

STEINDORFF NYMC Polarizing Microscope NYMC38000 Polarizing Microscope In order to exert performance of this microscope and to ensure the safety, please read the operating instruction carefully before use. 1 I. APPLICATION: NYMC38000 series

More information

Leica SP8 TCS Users Manual

Leica SP8 TCS Users Manual Leica SP8 TCS Users Manual Follow the procedure for start up and log on as posted in the lab. Please log on with your account only and do not share your password with anyone. We track and confirm usage

More information

Leica DB LB Research microscope and Studo Lite Imaging software

Leica DB LB Research microscope and Studo Lite Imaging software Leica DB LB Research microscope and Studo Lite Imaging software Room B523 User Guide Molecular Imaging Unit University of Helsinki www.miu.helsinki.fi 9.4.2008 1 GENERAL USER INFORMATION... 1 2 SETTINGS

More information

personal DELTAVISION (pdv)

personal DELTAVISION (pdv) GUIDELINES AND HINTS Version 1.3 (March 2015) personal DELTAVISION (pdv) Epifluorescence microscope from Applied Precision Inc.: The microscope can be found in room 1.320. For details see the architectural

More information

MAKE SURE YOUR SLIDES ARE CLEAN (TOP & BOTTOM) BEFORE LOADING DO NOT LOAD SLIDES DURING SOFTWARE INITIALIZATION

MAKE SURE YOUR SLIDES ARE CLEAN (TOP & BOTTOM) BEFORE LOADING DO NOT LOAD SLIDES DURING SOFTWARE INITIALIZATION Olympus VS120-L100 Slide Scanner Standard Operating Procedure Startup 1) Red power bar switch (behind monitor) 2) Computer 3) Login: UserVS120 account (no password) 4) Double click: WAIT FOR INITIALIZATION

More information

Instruction Manual T Binocular Acromat Research Scope T Trinocular Acromat Research Scope

Instruction Manual T Binocular Acromat Research Scope T Trinocular Acromat Research Scope Research Scope Instruction Manual T-29031 Binocular Acromat Research Scope T-29041 Trinocular Acromat Research Scope T-29032 Binocular Semi-Plan Research Scope T-29042 Trinocular Semi-Plan Research Scope

More information

A masterpiece in microscopy

A masterpiece in microscopy English A masterpiece in microscopy microscopes for Life and biomedical Sciences The ideal microscope for Life and Biomedical Sciences The microscope designers of Euromex have combined their 50 years of

More information

Components of the Microscope

Components of the Microscope Swift M3 Microscope The Swift M3 is a versatile microscope designed for both microscopic (high magnification, small field of view) and macroscopic (low magnification, large field of view) applications.

More information

User manual for Nikon Elements software

User manual for Nikon Elements software User manual for Nikon Elements software Equipment: Nikon TE300 Eclipse microscope ANDOR Neo/Zyla B&W camera (default) DS Fi2 color camera Sign in on the sign in sheet; please use both your given name and

More information

Artisan Technology Group is your source for quality new and certified-used/pre-owned equipment

Artisan Technology Group is your source for quality new and certified-used/pre-owned equipment Artisan Technology Group is your source for quality new and certified-used/pre-owned equipment FAST SHIPPING AND DELIVERY TENS OF THOUSANDS OF IN-STOCK ITEMS EQUIPMENT DEMOS HUNDREDS OF MANUFACTURERS SUPPORTED

More information

Laboratory Introduction

Laboratory Introduction Laboratory Introduction There are two basic categories of microscopes: light microscopes and electron microscopes. Light, or optical, microscopes require light waves to provide the illumination while electron

More information

MIF ZEISS VIOLET CONFOCAL ZEN 2009 PROTOCOL

MIF ZEISS VIOLET CONFOCAL ZEN 2009 PROTOCOL MIF ZEISS VIOLET CONFOCAL ZEN 2009 PROTOCOL START-UP On the Switchbox, turn both black switches to the ON position. Wait for the microscope to boot up completely (watch the screen on the side of the microscope).

More information

Olympus Confocal Microscope User Guide Last updated

Olympus Confocal Microscope User Guide Last updated 1 Olympus Confocal Microscope User Guide Last updated 11-22-2017 Contents: Laser Safety Training Requirement... 1 Mercury Lamp Precautions... 1 Biosafety requirements and rules for work in the MIC... 2

More information

OPERATING INSTRUCTIONS

OPERATING INSTRUCTIONS Zeiss LSM 510 M eta Confocal M icroscope OPERATING INSTRUCTIONS Starting the System: 1. Turn the black knob on the laser box one-quarter turn from Off to On. You will hear the laser cooling mechanisms

More information

Dante (Microscope) & Beatrice (Guide) Orth Lab

Dante (Microscope) & Beatrice (Guide) Orth Lab Dante (Microscope) & Beatrice (Guide) Orth Lab Olympus IX81 Widefield Microscope User Guide v. 1.2 (11/2014) Objectives 4x/0.13NA UPLFLN Semi Apo 10x/0.4NA PH UPLAPO Plan Apo 20x/0.8NA PH UPLAPO Plan Apo

More information

Zeiss Axioplan 2 imaging microscope and Axiovision software

Zeiss Axioplan 2 imaging microscope and Axiovision software Zeiss Axioplan 2 imaging microscope and Axiovision software Microscopes 1 and 2 in room B501b User Guide Molecular Imaging Unit University of Helsinki www.miu.helsinki.fi 20.5.2010 1 GENERAL... 1 1.1...

More information

Biology 29 Cell Structure and Function Spring, 2009 Springer LABORATORY 1: THE LIGHT MICROSCOPE

Biology 29 Cell Structure and Function Spring, 2009 Springer LABORATORY 1: THE LIGHT MICROSCOPE Biology 29 Cell Structure and Function Spring, 2009 Springer LABORATORY 1: THE LIGHT MICROSCOPE Prior to lab: 1) Read these instructions (p 1-6) 2) Go through the online tutorial, the microscopy pre-lab

More information

Ⅰ. Application. Ⅱ. Main Technical Specification. AS1 Biological microscope,which is widely used in medical and

Ⅰ. Application. Ⅱ. Main Technical Specification. AS1 Biological microscope,which is widely used in medical and Ⅰ. Application Objective Eyepiece Monocular Head The Light Above Coarse Adjustment Knob AS1 Biological microscope,which is widely used in medical and hygienic establishments for conventional microscopic

More information

CONFOCAL MICROSCOPE (Zeiss LSM 510 META v4.2)

CONFOCAL MICROSCOPE (Zeiss LSM 510 META v4.2) Wellcome Trust Centre for Human Genetics Molecular Cytogenetics and Microscopy Core CONFOCAL MICROSCOPE (Zeiss LSM 510 META v4.2) 1) STARTING THE SYSTEM Abridged INSTRUCTIONS Switch on the mercury bulb

More information