MAN Revision 002. InSight Mini C-arm Imaging System Technical Reference Manual
|
|
- Joshua Wood
- 6 years ago
- Views:
Transcription
1 MAN Revision 002 InSight Mini C-arm Imaging System Technical Reference Manual
2 April 2007 The information contained in this Manual is confidential and proprietary to Hologic, Inc. This information is provided only to authorized representatives of Hologic s customers solely for the purpose of facilitating the use of Hologic s products. No information contained herein may be disclosed to any unauthorized person for any purpose whatsoever without the prior written consent of Hologic, Inc. The information in this document is subject to change without notice by Hologic, Inc. All rights reserved. InSight and Fluoroscan are trademarks of Hologic, Inc. Hologic and the Hologic logo are registered trademarks of Hologic, Inc. All other trademarks, registered trademarks, and product names used within this document are the property of their respective owners. The Hologic products described in this Manual are protected by at least some of the following U.S. Patents and foreign counterparts: 6,234,672; 6,236,712; PCT Intl. Application US98/25281 This manual originally drafted in the English language. Printed in the U.S.A. Hologic, Inc. 35 Crosby Drive Bedford, MA USA Tel: (781) Fax: (781) Service: (800) 321-HOLX ( ) User Support: (800) 321-HOLX ( Hologic Europe Hologic N.V. Leuvensesteenweg 250A 1800 Vilvoorde Belgium Tel: Fax:
3 Table of Contents Product Specifications...1 Power Requirements...1 Environmental Requirements...1 Weights/Dimensions...1 Space Requirements...1 X-Ray Specifications...2 Performance...2 Imaging Specifications...3 Laser Performance Specifications...3 Equipment Classification...4 Symbols...5 InSight Entrance Exposure Rate Data...6 Entrance Exposure Rate Data, Auto Mode...7 Scatter Radiation Survey...8 Insight System Labels...11 Manufacturer Serial Number Label...11 Patent Label...11 X-Ray System Warning Label...12 Electric Shock Warning Label...12 Class 1 Laser Label...12 Caution on Incline Label...12 Transport Label...12 X-Ray Controller Compliance Label...13 X-Ray Source Compliance Label...13 Collimator Compliance Label...14 Image Intensifier Compliance Label...14 C-Arm Compliance Label...15 AC Input Power Fuse Label...15 Accessory Fuse Label...15 Printer Fuse Label...16 Explosion Risk Label...16 Extremities Only Label...16 Field of View...16 Computer On/Standby Label...17 Ground Labels...17 Back Panel Labels...17 iii
4 iv InSight Technical Reference Manual
5 Product Specifications Power Requirements Parameter AC Input Voltage Wattage AC Input Frequency Computer System Backup Time Value Nominal 100/120/220/240 V Single Phase 750 watts maximum Nominal 50/60 Hz ~ 2 minutes with full charge Environmental Requirements Parameter Value Temperature Range (Operating) to +32 C (+60 to +90 F) Temperature Range (Storage) -20 to +50 C (-4 to +122 F) Relative Humidity (Operating) Relative Humidity (Storage and Transport) Atmospheric Pressure 20-80% non-condensing 20-85% non-condensing hpa ( mm Hg) Weights/Dimensions Height Width Depth Weight 65 in. (165.1 cm) 35 in. (88.9 cm) 35in. (88.9 cm) 500 pounds (227 kg) Space Requirements Required Floor Space: 5 ft. x 6.6 ft. (1.5 m x 2.0 m) Door Size Ground Clearance: 30 in. (88.9 cm) minimum in. (9.8 cm) minimum Floor Capacity lb/ft 2 (737.2 kg/m 2 ) Page 1 of 17
6 X-Ray Specifications Source - Image receptor distance 44 cm (17.5") X-ray source X-ray tube window Beam filtration Focal spot Field of View Rated peak tube potential Tube kvp range Tube current range Tolerance Maximun Duty cycle at 75 kvp/0.100 ma Grounded tungsten anode X-ray tube with custom designed high voltage generator " (0.127mm) Beryllium Stainless steel with aluminum equivalence 2.5mm mm 7.5 watts Operator selectable collimation of 4" or 6" diameter (10.16 cm or cm) FOV at the plane of the image intensifier 75 kvp 40 to 75 kvp to ma kvp ± 5% of displayed value ma ± 8% ma of displayed value ma ± 5% >0.035 ma of displayed value 50% Performance Parameter Acquisition Frame Rate Value Non-cine recording 30 FPS Cine recording 25 FPS Fluoroscopic Time Alarm Range 15 seconds 5 minutes, adjustable in 15 second increments (default is 5 minutes) Image Storage Capacity Cine Loop Capacity Laser Alignment Guide Timeout 4000 Images (maximum) 10 minutes maximum 20 seconds Page 2 of 17
7 Imaging Specifications Image intensifier C-arm dimensions Image processing Video system Resolution (using Nuclear Associates test tool , located at the minimum SSD in normal (6") mode) Contrast (using Nuclear Associates test tool , located at the surface of the image intensifier housing in normal (6") mode) High sensitivity Cesium Iodide Source to detector distance: 17.5" (44 cm) Free space: 14" (36 cm) Depth of arc = 14" (36 cm) Digital, with 2 to 8 frames of video averaging and last frame freeze 21" (53 cm) LCD touch screen monitor Solid state video camera DVI-I output 7.0 lp/mm, minimum Must resolve, at minimum, 30# mesh At least 3 of 4 low contrast masses (2,4,6,8 mm) visible Low contrast inserts in both monitor adjustment squares visible Laser Performance Specifications Parameter Condition Value Laser Power Device on IEC Class 1 Laser Wavelength Device on 650 nm Projected Line Width At image intensifier surface 0.05 ± 0.03 inches Projected Line Length Line Intersection Location Accuracy At center of image intensifier surface At center of image intensifier surface 7.00 ± 0.50 inches ± inches Auto-off delay Following turn on 20.0 ± 2.0 seconds Input Power Voltage N/A 5.00 ± 0.25 volts Input Power Current Device off 10 ma, max. Input Power Current Device on 100 ma, max. Page 3 of 17
8 Equipment Classification InSight is classified* as follows: Type of protection against electric shock CLASS 1 Degree of protection against electric shock TYPE B Laser per IEC CLASS 1 Degree of protection against harmful ingress of water Mode of operation IPXO Continuous Operation InSight is also classified* as: Equipment not suitable for use in the presence of a flammable anesthetic mixture with air or with oxygen or nitrous oxide. * Medical Electrical Equipment - UL Page 4 of 17
9 LBL Rev 001 InSight Technical Reference Manual Symbols The following symbols are found on the InSight system: R CSA listed device 0044 CE marked device Dangerous voltage Attention - Read the Caution or Warning statement that follows Type B equipment X-ray filtration Time delay fuse Protective ground Functional ground Equipotential ground USB connector X-ray on switch Laser on switch Print Store Rotate image Increase or decrease kv/ma Electrostatic sensitive device Dispose of electrical/electronic equipment properly Computer On/Standby switch Rear wheels free to move in all directions Rear wheels move forward and back only Rear wheels locked Cycle through the noise suppression settings Page 5 of 17
10 InSight Entrance Exposure Rate Data Typical entrance exposure rates, measured at a point in center of image intensifier, 2 cm above surface of the housing. kv ua ugy/s Page 6 of 17
11 kv ua ugy/s Entrance Exposure Rate Data, Auto Mode Typical entrance exposure rates with Auto kv/ma ON. 6" mode 8.9 ugy/s 4" mode 9.6 ugy/s Measured using RTI Barracuda dosimeter with Multi-Purpose Detector, positioned 2 cm above input surface of the image intensifier housing. There is no additional attenuation of the x-ray beam between the source and the dosimeter detector (open beam). Page 7 of 17
12 Scatter Radiation Survey Table 1: Air Kerma (mg/h) Rate Parallel to Cathode - Anode Axis using ANSI Knee Phantom; 65 kvp/.088 Ma Distance (cm) mg/h mr/h mg/h mr/h mg/h mr/h mg/h mr/h mg/h mr/h N/A N/A N/A N/A Table 2: Air Kerma (mg/h) Rate Parallel to Cathode - Anode Axis using ANSI Knee Phantom; 75 kvp/.100 ma Distance (cm) mg/h mr/h mg/h mr/h mg/h mr/h mg/h mr/h mg/h mr/h N/A N/A N/A N/A Table 3: Air Kerma (mg/h) Rate Perpendicular to Cathode - Anode Axis using ANSI Knee Phantom; 65 kvp/.088 ma Distance (cm) mg/h mr/h mg/h mr/h mg/h mr/h mg/h mr/h mg/h mr/h N/A N/A N/A N/A Page 8 of 17
13 Table 4: Air Kerma (mg/h) Rate Perpendicular to Cathode - Anode Axis using ANSI Knee Phantom; 75 kvp/.100 ma Distance (cm) mg/h mr/h mg/h mr/h mg/h mr/h mg/h mr/h mg/h mr/h N/A N/A N/A N/A Table 5: Air Kerma (mg/h) Rate Parallel to Cathode - Anode Axis using Anthropomorphic Hand Phantom; 50 kvp/.058 ma Distance (cm) mg/h mr/h mg/h mr/h mg/h mr/h mg/h mr/h mg/h mr/h N/A N/A N/A N/A Table 6: Air Kerma (mg/h) Rate Parallel to Cathode - Anode Axis using Anthropomorphic Hand Phantom; 75 kvp/.100 ma Distance (cm) mg/h mr/h mg/h mr/h mg/h mr/h mg/h mr/h mg/h mr/h N/A N/A N/A N/A Page 9 of 17
14 Table 7: Air Kerma (mg/h) Rate Perpendicular to Cathode - Anode Axis using Anthropomorphic Hand Phantom; 50 kvp/.058 ma Distance (cm) mg/h mr/h mg/h mr/h mg/h mr/h mg/h mr/h mg/h mr/h N/A N/A N/A N/A Table 8: Air Kerma (mg/h) Rate Perpendicular to Cathode - Anode Axis using Anthropomorphic Hand Phantom; 75 kvp/.100 ma Distance (cm) mg/h mr/h mg/h mr/h mg/h mr/h mg/h mr/h mg/h mr/h N/A N/A N/A N/A Page 10 of 17
15 Insight System Labels Manufacturer Serial Number Label The Manufacturer Serial Number Label is on the back of the base cabinet. System Model: InSight 2 Manufactured by: Hologic, Inc. 35 Crosby Drive Bedford, MA USA R 0044 Electrical rating: 100/120/220/240V 50/60 Hz, 750VA 50% Max Duty Cycle at 75kVp, 100uA LBL Rev. 001 Patent Label The Patent Label is on the back of the base cabinet. LBL Rev 001 Page 11 of 17
16 X-Ray System Warning Label The X-Ray System Warning Label is on the front of the base cabinet below the Computer On/ Standby switch. WARNING: This X-Ray unit may be dangerous to patient and operator unless safe exposure factors and operating instructions are observed REV 001 Electric Shock Warning Label The Electric Shock Warning Label is located near the fasteners for panels covering high voltage components. LBL REV 001 Class 1 Laser Label The Class 1 Laser Label is on the High Voltage Power Supply cover near the laser aperture. Class 1 Laser Product Complies with FDA performance standards for laser products except for deviations pursuant to Laser Notice number 50 dated July 26, LBL REV 002 Caution on Incline Label The Caution on Incline Label is on the back of the unit, above the handle. USE CAUTION WHEN MOVING SYSTEM ON INCLINE LBL Rev 001 Transport Label The Transport Label is on the back of the unit, above the handle. TRANSPORT UNIT WITH C-ARM LOCKED LBL Rev 002 Page 12 of 17
17 X-Ray Controller Compliance Label The X-Ray Controller Compliance Label is on the X-Ray Controller assembly and on the back of the base cabinet. This X-ray controller complies with DHHS rules 21 CFR Subchapter J applicable on date of manufacture. Manufactured by: Hologic, Inc. 35 Crosby Drive Bedford, MA USA Serial Number: Date of Manufacture: LBL Rev. 001 X-Ray Source Compliance Label The X-Ray Source Compliance Label is on the X-Ray source module and on the back of the base cabinet. Page 13 of 17
18 Collimator Compliance Label The Collimator Compliance Label is on the X-Ray source module (near the collimator assembly) and on the back of the base cabinet. This X-ray collimator complies with DHHS rules 21 CFR Subchapter J applicable on date of manufacture. Filtration: 2.5 mm Al 70 kvp Manufactured by: Hologic, Inc. 35 Crosby Drive Bedford, MA USA Date of Manufacture: LBL Rev. 002 Image Intensifier Compliance Label The Image Intensifier Compliance Label is on the back of the Radiological Imaging Unit cover. This Image Intensifier conforms to all applicable standards under Title 21 of the U.S. Code of Federal Regulations (21 CFR) Subchapter J. Manufactured by: Hologic, Inc. 35 Crosby Drive Bedford, MA USA Serial Number: Date of Manufacture: LBL Rev. 003 Page 14 of 17
19 C-Arm Compliance Label The C-Arm Compliance Label is on the back of the High Voltage Power Supply cover. This C-Arm Assembly conforms to all applicable standards under Title 21 of the U.S. Code of Federal Regulations (21 CFR) Subchapter J. Manufactured by: Hologic, Inc. 35 Crosby Drive Bedford, MA USA Serial Number: Date of Manufacture: LBL Rev. 001 AC Input Power Fuse Label The AC Input Power Fuse Label is on the back of the base cabinet near the AC power connector. Input Voltage Fuse Type Size V T10.0A/250V 5x20mm V T6.3A/250V 5x20mm (fuses are slow blow) LBL Rev. 002 Accessory Fuse Label The Accessory Fuse Label is on the back of the base cabinet near the accessory outlet. Accessory Outlet 120V 0.8A max Fuse Type: 5x20mm 1.0A 250V TD (Slow blow) LBL Rev. 001 Page 15 of 17
20 Printer Fuse Label The Printer Fuse Label is located at the back of the printer compartment near the printer outlet. Printer Outlet 120V Fuse Type: 5x20mm (Slow blow) 2.6A max 3.15A 250V TD LBL Rev. 001 Explosion Risk Label The Explosion Risk Label is located on the back of the base cabinet. Attention Risk of explosion. Do not use in the presence of flammable anesthetics. Risque d explosion. Ne pas employer en présence d anesthésiques inflammables. LBL Rev. 001 Extremities Only Label The Extremities Only Label is located on the back of the base cabinet. For examination of extremities only. LBL REV 001 Field of View Field of View is represented by raised circles on the underside of the High Voltage Power Supply. When the Field of View selection lever is positioned at the large circle (shown below) full field of view is selected. When the Field of View selection lever is positioned at the small circle reduced field of view is selected. Page 16 of 17
21 Computer On/Standby Label The Computer On/Standby Label is on the front of the base cabinet above the Computer On/ Standby switch. Ground Labels Each system grounding point, depending upon type of ground, has one of the Ground Labels shown below. Protective Ground Functional Ground Equipotential Ground Back Panel Labels The Ethernet and DVI labels are located in the middle of the system back panel below their respective connectors. ETHERNET DVI Page 17 of 17
Digital Panoramic X- Ray unit. Product Data
Digital Panoramic X- Ray unit Product Data Standard examination programs Standard panoramic: adult/child panoramic exam TMJ open/close mouth: 4 slices are taken in the same image (left/right condyle, open/close
More informationAtomscope PORTABLE VETERINARY X-RAY EQUIPMENT INSTRUCTION MANUAL. Version 1.0
Atomscope PORTABLE VETERINARY X-RAY EQUIPMENT INSTRUCTION MANUAL Version 1.0 Thank you for purchasing our HFX90V portable veterinary x-ray unit. We are confident that you will be pleased with the radiographs
More informationRadView Manual Collimator
OPERATOR MANUAL RadView Manual Collimator DEL MEDICAL IMAGING CORPORATION 11550 WEST KING STREET FRANKLIN PARK, IL. 60131 TECHNICAL SUPPORT (847) 288-7000 FAX (847) 288-7011 THIS DOCUMENT CONTAINS INSTALLATION,
More informationNuclear Associates
Nuclear Associates 07-706 Patient Phantom/Penetrometer System Users Manual March 2005 Manual No. 07-706-1 Rev. 2 2004, 2005 Fluke Corporation, All rights reserved. Printed in U.S.A. All product names are
More informationNeo Ultrasound Module Manual
Neo Ultrasound Module Manual Installation Instructions For complete User Operating Instructions, including Cautions, Warnings, Dangers, Indications, and Contraindications, refer to the User s Manuals.
More informationMinnesota Rules, Chapter 4732 X-ray Revision
Minnesota Rules, Chapter 4732 X-ray Revision DRAFT FLUOROSCOPIC X-RAY SYSTEMS, 1.0 Subpart 1. Applicability. Subpart 2. Limitation of the useful beam. Subpart 3. Measuring compliance; primary protective
More informationNew spectral benefi ts, proven low dose
New spectral benefi ts, proven low dose Philips MicroDose mammography SI, technical data sheet Philips MicroDose SI with single-shot spectral imaging is a fullfi eld digital mammography solution that delivers
More informationTechnical data CAMARGUE CS-VH50/300. VARIABLE Height Bucky Table With Ceiling Suspension
Technical data VARIABLE Height Bucky Table With Ceiling Suspension Model Variations CAMARGUE FH (Fixed Height) CAMARGUE FH Tomo CAMARGUE FH Ceiling suspension CAMARGUE VH (Variable Height) CAMARGUE VH
More informationGE AMX 4+ Portable X-Ray
GE AMX 4+ Portable X-Ray Typical Manufacturer s Picture GE Healthcare s AMX-4+ analog X-ray system provides high-performance in a compact, easy-to-maneuver package. The rotating arm and tube simplify positioning
More informationSPECIFICATION. Kilovoltage X-ray calibration system for protection and diagnostic level dosimetry. Prepared by
SPECIFICATION Kilovoltage X-ray Prepared by Igor Gomola, Technical Officer, Project ECU6023, Date 2015-Oct-06 Revision Date Status Comments 0.1 2015-Oct-06 Draft Igor Gomola Page 1 of 12 1. Scope This
More informationRotating Anode X-ray Tube Assembly
RADII H18X H18Y Datasheet Rotating Anode X-ray Tube Assembly The tube unit is so constructed that an X-ray tube is sealed in a diagnostic type protective tube housing of electric shockproof, radiation
More informationRotating Anode X-ray Tube
Datasheet RADII KL74-1.0/2.0-125 Rotating Anode X-ray Tube Rotating anode X-ray tube for the purpose of general diagnostic X-ray procedures. Specially processed Tungsten faced molybdenum target of 74 mm
More informationNuclear Associates
Nuclear Associates 07-649 CDRH Fluoroscopic Phantom Users Manual March 2005 Manual No. 07-649-1 Rev. 2 2004, 2005 Fluke Corporation, All rights reserved. Printed in U.S.A. All product names are trademarks
More informationPractix. Mobile Radiography. Constantly on the move, and yet never tired. Practix 160
Practix 160 Specifications Practix Mobile Radiography Constantly on the move, and yet never tired Practix 160 The mobile X-ray system Practix 160 offers ideal solutions for all routine examinations in
More informationI. PERFORMANCE OF X-RAY PRODUCTION COMPONENTS FLUOROSCOPIC ACCEPTANCE TESTING: TEST PROCEDURES & PERFORMANCE CRITERIA
FLUOROSCOPIC ACCEPTANCE TESTING: TEST PROCEDURES & PERFORMANCE CRITERIA EDWARD L. NICKOLOFF DEPARTMENT OF RADIOLOGY COLUMBIA UNIVERSITY NEW YORK, NY ACCEPTANCE TESTING GOALS PRIOR TO 1st CLINICAL USAGE
More informationFluoroscan InSight FD Mini C-Arm Imaging System
Extremity Imaging Fluoroscan InSight FD Mini C-Arm Imaging System Redefining Mini C-arm Technology More Imaging Options to suit your needs Building on Hologic s tradition of innovation in imaging, the
More informationX-RAYS - NO UNAUTHORISED ENTRY
Licencing of premises Premises Refer Guidelines A radiation warning sign and warning notice, X-RAYS - NO UNAUTHORISED ENTRY must be displayed at all entrances leading to the rooms where x-ray units are
More informationEmprint Ablation System with Thermosphere Technology
Emprint Ablation System with Thermosphere Technology Specification Guide + + + = FIELD CONTROL THERMAL CONTROL WAVELENGTH CONTROL THERMOSPHERE TM TECHNOLOGY Product Information REF / SKU Product Length
More informationRotating Anode X-Ray Tube Housing Assembly. General Data
Display Devices & Components Company TECHNICAL DATA ROTANODE TM E7252X Rotating Anode X-Ray Tube Housing Assembly! High speed rotating anode X-ray tube housing assembly for high energy radiographic and
More informationRotating Anode X-Ray Tube Housing Assembly. General Data. IEC Classification... Class I
Technical Data TD ROTANODE E7239X E7239FX E7239GX 197 Rotating Anode X-Ray Tube Housing Assembly Rotating anode X-ray tube housing assembly for the purpose of general diagnostic X-ray procedures. Specially
More informationData. microcat +SPECT
Data microcat +SPECT microcat at a Glance Designed to meet the throughput, resolution and image quality requirements of academic and pharmaceutical research, the Siemens microcat sets the standard for
More informationNuclear Associates
Nuclear Associates 76-700 Digital Subtraction Angiography Phantom Users Manual March 2005 Manual No. 76-700-1 Rev. 2 2004, 2005 Fluke Corporation, All rights reserved. Printed in U.S.A. All product names
More informationROTANODE TM E7252X E7252FX E7252GX Rotating Anode X-Ray Tube Housing Assembly. General Data. IEC Classification... Class I Electrical :
Technical Data TD ROTANODE TM E7252X E7252FX E7252GX Rotating Anode X-Ray Tube Housing Assembly High speed rotating anode X-ray tube housing assembly for high energy radiographic and cine-fluoroscopic
More informationMILADY. Product Data. Page 1 of 8
Page 1 of 8 The MILADY Mammographic Unit offers the best quality-to-price ratio to our customers worldwide. The unit advanced technology together with the application of industrial production standards,
More informationCHAPTER 6 QC Test For Fluoroscopic Equipment. Prepared by:- Kamarul Amin bin Abu Bakar School of Medical Imaging KLMUC
CHAPTER 6 QC Test For Fluoroscopic Equipment Prepared by:- Kamarul Amin bin Abdullah @ Abu Bakar School of Medical Imaging KLMUC Lesson Outcomes Describe the objectives of each QC test done. Identify QC
More informationHigh-Power 2.0 W 830 nm Fiber-Coupled Diode Laser 2486-L3 Series
COMMERCIAL LASERS High-Power 2.0 W 830 nm Fiber-Coupled Diode Laser 2486-L3 Series Key Features Up to 2 W output power at 0.22 NA 830 ± 15 nm 60 µm Step Index fiber Highly reliable Highly efficiency Applications
More informationY11-DR Digital Radiography (DR) Image Quality
Y11-DR Digital Radiography (DR) Image Quality Image quality is stressed for all systems in Safety Code 35. In the relevant sections Health Canada s advice is the manufacturer s recommended test procedures
More informationRotating Anode X-Ray Tube Housing Assembly. General Data. IEC Classification... Class I Type B
Technical Data TD ROTANODE E7239X E7239FX E7239GX 0197 Rotating Anode X-Ray Tube Housing Assembly Rotating anode X-ray tube housing assembly for the purpose of general diagnostic X-ray procedures. Specially
More informationNuclear Associates
Nuclear Associates 07-647 R/F QC Phantom Operators Manual March 2005 Manual No. 07-647-1 Rev. 2 2004, 2005 Fluke Corporation, All rights reserved. All product names are trademarks of their respective companies
More informationAcceptance Testing of a Digital Breast Tomosynthesis Unit
Acceptance Testing of a Digital Breast Tomosynthesis Unit 2012 AAPM Spring Clinical Meeting Jessica Clements, M.S., DABR Objectives Review of technology and clinical advantages Acceptance Testing Procedures
More informationPROVECTA V. Veterinary X-Ray System. Operator s Manual
PROVECTA V Veterinary X-Ray System Operator s Manual TABLE OF CONTENTS Introduction....................................................... 4 Warranty..........................................................
More informationARCO Rk.5 Digital Mobile C-Arm
Product Data ARCO Rk.5 Digital Mobile C-Arm Product Data 1 Easy to move INTRODUCTION Designed with the latest technology throughout, the ARCO Rk.5 set new standards for the excellence image quality ease
More informationCh. 223 VETERINARY MEDICINE CHAPTER 223. VETERINARY MEDICINE GENERAL PROVISIONS X-RAYS RADIOACTIVE MATERIAL. Authority
Ch. 223 VETERINARY MEDICINE 25 223.1 CHAPTER 223. VETERINARY MEDICINE Sec. 223.1. Purpose and scope. 223.2. [Reserved]. 223.2a. Definitions. 223.3 223.6. [Reserved]. 223.7. Structural shielding. 223.8.
More informationNuclear Associates , , CT Head and Body Dose Phantom
Nuclear Associates 76-414,76-414-4150,76-415 CT Head and Body Dose Phantom Users Manual March 2005 Manual No. 76-414-1 Rev. 2 2004, 2005 Fluke Corporation, All rights reserved. Printed in U.S.A. All product
More informationNuclear Associates
Nuclear Associates 07-591 Focal Spot Test Tool Users Manual February 2005 Manual No. 07-591-1 Rev. 2 2004, 2005 Fluke Corporation, All rights reserved. Printed in U.S.A. All product names are trademarks
More informationHigh-Power 8.0 W 9xx nm Fiber-Coupled Diode Laser 6397-L3 Series
COMMERCIAL LASERS High-Power 8.0 W 9xx nm Fiber-Coupled Diode Laser 6397-L3 Series Key Features 8.0 W output power 105 µm aperture 0.2 NA Highly reliable Applications Fiber laser pumping Material processing
More informationLASER ENHANCED REVOLVER GRIP OWNER S MANUAL RED LASER GREEN LASER
LASER ENHANCED RED LASER GREEN LASER REVOLVER GRIP OWNER S MANUAL LASER ENHANCED GRIP Installation Instructions Caution... 3 Safety Labels... 4 Installation...5-7 Programming...8-10 Batteries (Red Laser)...
More informationSINGLE-MODE LASER DIODES. Chip on Submount, QA-Mount. Laser Diodes
Laser QA 112/17 / V01 / IF / sheaumann/diodes/sm/qa_sm Chip on Submount, QA-Mount SINGLE-MODE LASER DIODES Laser DESCRIPTION High brightness, high quality, and high reliability are the foundation of our
More informationSPECIFIC TECHNICAL CRITERIA
Issue Date: 2009-04-10 Page 3 of 58 Report Reference # E116994-A48-UL-1 SPECIFIC TECHNICAL CRITERIA TEST REPORT UL 60601-1 Medical Electrical Equipment Part 1: General requirements for safety Report Reference
More informationImage Quality. HTC Grid High Transmission Cellular Grid provides higher contrast images
B R E A S T I M A G I N G S O L U T I O N S Setting the benchmark for mammography M-IV Series Innovations in breast imaging The Lorad M-IV Series exemplifies Hologic s commitment to developing advanced
More informationNuclear Associates &
Nuclear Associates 06-526 & 06-526-2200 RAD-CHECK PLUS Operators Manual March 2005 Manual No. 136201 Rev. 5 2003, 2005 Fluke Corporation, All rights reserved. Printed in U.S.A. All product names are trademarks
More informationPerformance Meets Simplicity
Performance Meets Simplicity Cone Beam 3D Imaging Systems P Panoramic X-ray Systems Intraoral X-ray Systems Digital Intraoral Sensors Digital X-ray Phosphor Plates Intraoral Cameras Imaging Software One
More informationMobileDaRt Evolution MX8 Version
PD504-021 Mobile X-ray System MobileDaRt Evolution MX8 Version MX8c type GENERAL The MobileDaRt Evolution is a general-purpose mobile digital X-ray system, which can be freely moved throughout a hospital
More informationSTEREOTACTIC BREAST BIOPSY EQUIPMENT SURVEYS
STEREOTACTIC BREAST BIOPSY EQUIPMENT SURVEYS JAMES A. TOMLINSON, M.S. Diagnostic Radiological Physicist American Board of Radiology Certified Medical Physics Consultants, Inc. Bio 28 yrs experience 100%
More informationL4 Series Operator Manual
L4 Series Operator Manual Access Laser Company 917 134 th St SW, Suite A1 Everett, WA 98204 425.582.8674 www.accesslaser.com L4 Series Operator Manual_Rev_J TABLE OF CONTENTS Section Page Company Statement...
More informationIndustrial X-ray Tube KL
Description stationary anode x-ray tube is a highly popular and proven side window x-ray tube, which has focus 1.0 mm, is designed for non-destructive x-ray inspection. stationary anode x-ray tube can
More informationD C 01/2019 3
D-0117968-C 01/2019 3 4 D-0117968-C 01/2019 Screw Driver Screw Driver Unplug both the Red & Blue connectors. (see above) Place a small flat head screw driver on the small orange tabs and push down while
More informationRead and be familiar with this manual before operating or servicing this device. To ensure operator, technician, and patient safety, use only as
Read and be familiar with this manual before operating or servicing this device. To ensure operator, technician, and patient safety, use only as specified in this manual. Product Manufacturer: Medical
More informationProX Intraoral X-ray. PLANMECA is proud to introduce a new intraoral X-ray unit to its comprehensive collection of imaging products- the ProX.
The premium intraoral X-ray unit... ProX Intraoral X-ray PLANMECA is proud to introduce a new intraoral X-ray unit to its comprehensive collection of imaging products- the ProX. This advanced unit provides
More informationDigital Radiography System
www.drgem.co.kr GXR-SD SERIES GXR-SD SERIES Digital Radiography System GXR-SD Series www.drgem.co.kr High Performance and Reliability The Optimal System for Digital Solution The GXR-SD Series is a diagnostic
More informationTECHNICAL SPECIFICATIONS for 300KV CONSTANT POTENTIAL X RAY EQUIPMENT (for industrial applications) S. No. PARTICULARS BHEL SPECIFICATIONS
TECHNICAL SPECIFICATIONS for 300KV CONSTANT POTENTIAL X RAY EQUIPMENT (for industrial applications) BIDDER s OFFER S. No. PARTICULARS BHEL SPECIFICATIONS 1.0 Area of Application The 300kV Constant potential
More informationAdvanced Test Equipment Rentals ATEC (2832)
Established 1981 Advanced Test Equipment Rentals www.atecorp.com 800-404-ATEC (2832) 6500 Series Loop Antennas User Manual ETS-Lindgren Inc. reserves the right to make changes to any product described
More informationEntry-Level Model for Laser Marking
CO2 Laser Marker LP-300 SERIES FDA Conforming to FDA regulations (Only LP-310-A) Conforming to Low Voltage and EMS Directive (Only LP-310-C) GB Conforming to GB standards (Only LP-310-B) Entry-Level Model
More informationImage Quality. HTC Grid High Transmission Cellular Grid provides higher contrast images
B R E A S T I M A G I N G S O L U T I O N S Setting the benchmark for mammography M-IV Series Innovations in breast imaging The Lorad M-IV Series exemplifies Hologic's commitment to developing advanced
More informationOPERATOR S INSTRUCTIONS (for USA & Canada)
MODEL 097 DENTAL X-RAY OPERATOR S INSTRUCTIONS (for USA & Canada) WARNING This X-ray equipment may be dangerous to patients and operators unless safe exposure factors and operating instructions are observed.
More informationQUAD 1000 OWNERS MANUAL OPERATION GUIDE
QUAD 1000 OWNERS MANUAL OPERATION GUIDE www.laserlinemfg.com REV. 4.30.18 COMPONENTS 4001-0000 Quad 1000 5 arc second zenith laser featuring electronic servo self-leveling with variable temperature compensation
More informationJoint ICTP/IAEA Advanced School on Dosimetry in Diagnostic Radiology and its Clinical Implementation May 2009
2033-6 Joint ICTP/IAEA Advanced School on Dosimetry in Diagnostic Radiology and its Clinical Implementation 11-15 May 2009 Dosimetry for Fluoroscopy Basics Renato Padovani EFOMP Joint ICTP-IAEA Advanced
More informationProduct Information PI
Rotating Anode X-ray Tube Assembly Product Information PI ROTANODE The TOSHIBA XH-183 housing is developed for use in surgical C-arm application. It provides ease for use at a nominal patient table height.
More informationCorsix R 32 Accu PRODUCT DATA
PRODUCT DATA Product data Rev. 1 (April 2011) Page Description and Configurations 3 Technical Features 4 Radiological characteristics 4 Operating characteristics 5 Mechanical characteristics 6 Size and
More informationOBIS CellX. The Universal Light Engine FEATURES
OBIS CellX The Universal Light Engine OBIS CellX is a multi-wavelength platform for use as the laser excitation Light Engine in applications requiring or laser wavelengths from a single module. CellX delivers
More informationPractix. Mobile Radiography. Changing its point of view comes naturally. Practix 300
Practix 300 Specifications Practix Mobile Radiography Changing its point of view comes naturally Practix 300 With its swiveling telescopic tube arm and rotatable column, the Practix 300 adds a new and
More informationELECTRONIC CONTROL CONCEPTS 160 Partition Street Saugerties, NY or local phone
ELECTRONIC CONTROL CONCEPTS 160 Partition Street Saugerties, NY 12477 (800)VIP-XRAY (845)247-9028 Fax or 800-847-9729 local phone 845-246-9013 http://www.eccxray.com sales@eccxray.com INSTRUCTION MANUAL
More informationHigh-Power 8xx nm Fiber-Coupled Diode Laser 2495-L3 Series
COMMERCIAL LASERS High-Power 8xx nm Fiber-Coupled Diode Laser 2495-L3 Series Key Features 4.0 W output power industrial 808 nm 4.5 W output power medical/dental 812 nm 105 µm aperture 0.2 NA Highly reliable
More informationPhilips MicroDose Mammography SI
English Technical Specification Philips MicroDose Mammography SI Model L50 Software 9.0 Technical Specification - Philips MicroDose Mammography SI Model L50 Software 9.0 Document Title Technical Specification
More informationQuality Design Because you pay more and more attention to the look and effectiveness of your practice
Quality Design Because you pay more and more attention to the look and effectiveness of your practice ACTEON has developed X-Mind unity with: - Uncompromising design - Clean lines - Proven quality - Superior
More informationCIRCLEX 0.3/0.8P324&0.6/1.2P324DK-85
PD53-012 p Rotating Anode X-ray tube Assembly 0.3/0.8P32&0.6/1.2P32DK-85 GENERAL The Shimadzu 0.3/0.8P32DK-85 & 0.6/1.2P32DK-85, Rotating Anode X-ray tube assemblies are rated to 150kV and feature a 100mm
More informationPET1606J2F. Pilani Electron Tubes & Devices Pvt. Ltd. Water Cooled Triode. For Industrial RF Heating. Drop in equivalent of BW1606J2F
Water Cooled Triode For Industrial RF Heating Drop in equivalent of BW1606J2F Output Power: 30 kw Anode voltage: 10 kv max Anode dissipation: 15 kw max Frequency up to 30 MHz Manufactured in India, in
More informationMammography is a radiographic procedure specially designed for detecting breast pathology Approximately 1 woman in 8 will develop breast cancer over
Mammography is a radiographic procedure specially designed for detecting breast pathology Approximately 1 woman in 8 will develop breast cancer over a lifetime Breast cancer screening programs rely on
More informationSPAC265-8W. AC-DC power supply module. Features. Description. Applications
AC-DC power supply module Datasheet production data Features Open frame switch mode power supply European input voltage range Single output 5 V, 8 W peak power, 4 W continuous operating mode EMC compliance
More informationMODEL 3810/2 Line Impedance Stabilization Network
EMC TEST SYSTEMS FEBRUARY 1996 REV C PN 399197 MODEL 3810/2 Line Impedance Stabilization Network OPERATION MANUAL USA P.O. Box 80589 Austin, Texas 78708-0589 2205 Kramer Lane, Austin, Texas 78758-4047
More informationNuclear Associates
Nuclear Associates 07-492 Mammographic Digital kvp Meter Operators Manual March 2005 Manual No. 164001 Rev. 3 2003, 2005 Fluke Corporation, All rights reserved. Printed in U.S.A. All product names are
More informationOEM s #1 Choice for X-Ray Sources and Generators. Focus on Performance, Reliability, Form Factor, and Versatility.
OEM s #1 Choice for X-Ray Sources and Generators Focus on Performance, Reliability, Form Factor, and Versatility www.vjtxray.com Applications Security, Industrial NDT, Food & Pharmaceutical, Quality &
More informationNJDEP Medical Physicist s Radiographic QC Survey Registration Number:
Facility Name NJDEP ID # NJDEP Medical Physicist s Radiographic QC Survey PLEASE PRINT Facility Information Unit Information Manufacturer Model Console Model # Console serial # Tube serial # Location (room)
More informationBCA 9SK. 60 Hz on request Absorbed current on Stationary Vac and Vac in fluoro mode
ELECTRICAL CHARACTERISTICS Voltage 230 Vac ±10% monophase standard 105 / 115 / 125 / 220 / 240 Vac ±10% monophase on request Frequency 50 Hz standard 60 Hz on request Absorbed current on Stationary 4.5
More informationTest Equipment for Radiology and CT Quality Control Contents
Test Equipment for Radiology and CT Quality Control Contents Quality Control Testing...2 Photometers for Digital Clinical Display QC...3 Primary Workstations...3 Secondary Workstations...3 Testing of workstations...3
More informationSAMSUNG SMART DIGITAL RADIOGRAPHY. XGEO GU60A series REDEFINING ERGONOMICS
SAMSUNG SMART DIGITAL RADIOGRAPHY XGEO GU60A series REDEFINING ERGONOMICS U-arm Positioner (Fully Automated) The XGEO GU60A/60A-65 is a universal, fully motorized system. Its unique U-arm rotates +120
More informationBCA 9RK. 60 Hz on request Absorbed current on Stationary Vac and Vac in fluoro mode
ELECTRICAL CHARACTERISTICS RADIOLOGICAL DATA MONOBLOC Voltage 230 Vac ±10% monophase standard 105 / 115 / 125 / 220 / 240 Vac ±10% monophase on request Frequency 50 Hz standard 60 Hz on request Absorbed
More informationDescription. Features Advanced Multiple Quantum Well (MQW) Fabry-Perot Laser Design Cost-effective Uncooled Laser Technology 5.6-mm TO-style package
Description The series of Multi-Quantum Well (MQW) Fabry-Perot (FP) lasers are well suited for low-cost high-speed transmitters. The devices feature high output power and wide operating temperature range.
More informationExternal RF Driver. Electrical Driver for AOTF-crystal. Instruction Manual
External RF Driver Electrical Driver for AOTF-crystal Instruction Manual Table of Contents 1 General... 3 2 Compliance... 3 3 Labeling... 4 4 Interface... 5 4.1 Front Panel... 5 4.2 Back Panel... 6 4.2.1
More informationPEAK OUTPUT CURRENT 1. MAP V 4.7V to 5.8V 8A 11A 0.2% ±1% 1% 4.9V to 5.1V PEAK OUTPUT CURRENT 4 LINE REGULATION
The Bel Power Solutions MAP30/40/42 Series of power supplies combines low cost and universal input in a board-only power solution to meet commercial and industrial requirements. Full international safety,
More informationProduct Information PI
Rotating Anode X-ray Tube Assembly Product Information PI ROTANODE E7252X E7252FX E7252GX High speed rotating anode X-ray tube assembly for high energy radiographic and cine-fluoroscopic operations. The
More informationMG7095 Tunable S-Band Magnetron
MG7095 Tunable S-Band Magnetron The data should be read in conjunction with the Magnetron Preamble and with British Standard BS9030 : 1971. ABRIDGED DATA Mechanically tuned pulse magnetron intended primarily
More informationMobile Laser Scanning. High-Performance LiDAR Sensor for KINEMATIC Laser Scanning. visit our website
High-Performance LiDAR Sensor for KINEMATIC Laser Scanning RIEGL VUX-1HA very high measurement rate up to 1,, meas./sec very high scan speed up to 25 scans / second 5 mm survey-grade accuracy field of
More informationInstruction Manual. Self-Leveling Cross-Line Laser Level Model No and
3728H_Manuals 9/2/10 8:02 AM Page 1 Self-Leveling Cross-Line Laser Level Model No. 40-6620 and 40-6625 Instruction Manual Congratulations on your choice of this Self-Leveling Cross-Line Laser Level. We
More informationTable of Contents. Polytel GMA Glucose Meter Accessory Model PWR USER MANUAL Version: 2
Table of Contents Polytel GMA Glucose Meter Accessory Model PWR-08-03 USER MANUAL Version: 2 Polymap Wireless 310 S. Williams Blvd. Ste. 346 Tucson, Arizona 85711 (520) 747-1811 www.polymapwireless.com
More informationDIGITAL LASER DISTANCE METER
DIGITAL LASER DISTANCE METER LD05-A10GF with glass-fiber coupled remote optical head The RIEGL LD05-A10GF is a multi-purpose laser distance meter based on precise timeof-flight laser range measurement
More informationOperating instructions Diffuse reflection sensor with background suppression. OJ51xx laser / / 2010
Operating instructions Diffuse reflection sensor with background suppression OJ51xx laser 704811 / 00 05 / 2010 Contents 1 Preliminary note3 1.1 Symbols used 3 2 Safety instructions 3 3 Functions and features
More informationTHE PANTAK THERAPAX SXT 150
THE PANTAK THERAPAX SXT 150 combines the safety and ease-of-use of photon radiation therapy with state-of-the-art technology, including microprocessor control and solid-state electronics. Using Therapax,
More informationFORTUS 360mc/400mc and FDM 360mc/400mc 3D Production System. Site Preparation Guide Rev E
FORTUS 360mc/400mc and FDM 360mc/400mc 3D Production System Site Preparation Guide 107167-0005 Rev E Liability Statement The information in this document is subject to change without notice. Stratasys,
More informationInovision Model 451P Pressurized µr Ion Chamber Survey Meter
Inovision Model 451P Pressurized µr Ion Chamber Survey Meter High sensitivity µr measurements of exposure and exposure rate Available with dose equivalent energy response (SI units) Fast response to measure
More informationDose Reduction and Image Preservation After the Introduction of a 0.1 mm Cu Filter into the LODOX Statscan unit above 110 kvp
Dose Reduction and Image Preservation After the Introduction of a into the LODOX Statscan unit above 110 kvp Abstract: CJ Trauernicht 1, C Rall 1, T Perks 2, G Maree 1, E Hering 1, S Steiner 3 1) Division
More informationTable of Contents. Polytel GMA Glucose Meter Accessory for Bayer Ascensia Contour Model PWR USER MANUAL Version: B
Table of Contents Polytel GMA Glucose Meter Accessory for Bayer Ascensia Contour Model PWR-08-07 USER MANUAL Version: B Polymap Wireless 310 S. Williams Blvd. Ste. 350 Tucson, Arizona 85711 (520) 747-1811
More informationABOVETABLE X-RAY SOURCE FLUOROSCOPIC AND SPOT-FILM SYSTEMS
PART VI ABOVETABLE X-RAY SOURCE FLUOROSCOPIC AND SPOT-FILM SYSTEMS FORM FDA 3069 REPRINTED April 2000 ROUTINE COMPLIANCE TESTING ABOVETABLE X-RAY SOURCE FLUOROSCOPIC AND SPOT-FILM SYSTEMS (Test Procedure
More informationSonicator Plus 994 Specifications
Sonicator Plus 994 Specifications General Specifications: Input: Certification: Domestic model ETL and C-ETL Listed: Domestic model Classification: CE model Year 2000 Compliant Weight: Dimensions: Operating
More informationXRC X-Ray Calibration System
XRC X-Ray Calibration System Technical Description Contents 1. Equipment overview 2. X-ray Beam Specifications 3. The Control Console 4. Radiation Safety 5. Filters 6. X-ray Generator & Heat Exchange Equipment
More informationAgilent 81662A DFB Laser Agilent 81663A DFB Laser Agilent Fabry-Perot Lasers
Agilent 81662A DFB Laser Agilent 81663A DFB Laser Agilent Fabry-Perot Lasers Technical Specifications May 2003 The Agilent 81662A low power and 81663A high power DFB Laser Source modules are best suited
More informationProduct Bulletin. SDL-2400 Series 2.0 & 3.0 W, 798 to 800/808 to 812 nm High-brightness Laser Diodes
Product Bulletin SDL-24 Series 2. & 3. W, 798 to 8/88 to 812 nm High-brightness Diodes The SDL-24 series laser diodes represent a breakthrough in high continuous wave (CW) optical power and ultra-high
More informationHigh Power Outdoor. GaN Solid State Power Amplifiers
Teledyne Paradise Datacom s newly packaged High Power Outdoor (H) series of Solid State Power Amplifiers is packaged with the latest Gallium Nitride, GaN, based SSPA modules. Utilizing the latest in linearized
More informationHigh Power Outdoor. GaAs Solid State Power Amplifiers
Teledyne Paradise Datacom s newly packaged High Power Outdoor (H) series of Solid State Power Amplifiers represent the latest in High Power Microwave Amplifier Technology. The SSPA package achieves the
More informationTable of Contents. Polytel GMA Glucose Meter Accessory for Abbott FreeStyle Lite and FreeStyle Freedom Lite Model PWR USER MANUAL Version: B
Table of Contents Polytel GMA Glucose Meter Accessory for Abbott FreeStyle Lite and FreeStyle Freedom Lite Model PWR-08-09 USER MANUAL Version: B Polymap Wireless 310 S. Williams Blvd. Ste. 350 Tucson,
More information