A Portable Scanning Electron Microscope Column Design Based on the Use of Permanent Magnets
|
|
- Samuel Buck Francis
- 5 years ago
- Views:
Transcription
1 SCANNING VOL. 20, (1998) Received October 8, 1997 FAMS, Inc. Accepted with revision November 9, 1997 A Portable Scanning Electron Microscope Column Design Based on the Use of Permanent Magnets A. KHURSHEED, J. C. PHANG, J. T. L. THONG The Centre for Integrated Circuit Failure Analysis and Reliability (CICFAR), Electrical Engineering Department, National University of Singapore, Singapore Summary: A portable scanning electron microscope (SEM) design is presented which makes use of permanent magnets. Simulation results predict that such an SEM is feasible and that it can be compact. The height is typically less than 12 cm. The is designed to be modular, so that it can fit onto a wide range of different specimen chamber types, and can also be readily replaced. Key words: s, miniature, permanent magnets, scanning electron microscope Introduction The microminiaturization of electron optical systems has received considerable attention over the last decade (Chang et al. 1990, Chisholm et al. 1997). At present, the scanning electron microscope (SEM) is a bulky device which, when installed at a specific location, is not designed to be moved again. There are several advantages to making the SEM portable, where the overall height from the specimen chamber to the electron gun is less than 12 cm. First, more compact SEM s give rise to the possibility of making the SEM modular. The bulky items such as vacuum pumps, drive, and display electronics are fixed, and the SEM s electron optics section, that is, the part comprising the, can be easily removed and replaced. Figure 1a depicts a schematic diagram of this concept. In principle, the user will be able to choose from a range of different SEM s, each suited to different applications, while employing the Address for reprints: A. Khursheed The Centre for Integrated Circuit Failure Analysis and Reliability (CICFAR) Electrical Engineering Department National University of Singapore 10 Kent Ridge Crescent Singapore eleka@nus.sg same vacuum and electronics system. By simply replacing the, the SEM can, in principle, be converted into a different analytical tool. At present for instance, quantitative voltage contrast or critical dimension topographic measurements, or Auger spectrometer analysis require the use of different SEMs, each designed to meet different requirements. The portable SEM provides the opportunity of having an SEM cluster, where each is designed for a different application. The cluster of SEM s will be compact enough to fit on a normal shelf or table top. The second advantage of a portable SEM is that it can be used on a variety of different specimen chambers. This possibility is illustrated in Figure 1b. Since specimen chambers greatly vary in their size, shape, and vacuum conditions, the ability to use the same SEM for them all greatly reduces the cost and space requirements for SEM systems in general. At present, if a different specimen chamber is required, it necessitates the use of a different SEM. The third advantage of the portable SEM comes from its reduced volume. The smaller volume leads to faster pump down times and simplifies vacuum requirements. Fourth, since the on-axis aberrations scale with the size of the electron lenses, the spatial resolution of the final primary beam is expected to be significantly better than that of conventional SEMs. Fifth, the shorter primary beam path length means that it will experience less variation from stray fields. A shorter path length is also expected to result in smaller probe sizes, since it reduces the energy spread in the primary beam due to the effect of electron electron interactions. Finally, a light-weight can be scanned or moved over large distances. This might be useful where the specimen is large and cannot be easily moved, or where its movement is restricted by a bulky item, such as another piece of test or measurement equipment. The main difficulty in scaling down the size of present SEM s relates to the current coil density in the magnetic electron lenses. The increase in current density is proportional to the square of the factor of reduction. Already at their present size, special considerations of cooling need to be taken into account. The use of permanent magnets is a way of overcoming this problem. Permanent magnets, unlike conventional SEM electromagnetic magnets, do not operate by employing current carrying coils and so avoid potential heating problems.
2 88 Scanning Vol. 20, 2 (1998) Permanent Magnets Efforts to introduce permanent magnet lenses into electron microscopy date back from the very beginnings of electron microscope design. At that time, they were advocated as a way of overcoming power supply instability problems (Rieke 1982), but it soon became apparent that the field strength of permanent magnets was not strong enough for many common applications in electron optics. It was not until the 1960s, when new permanent magnet materials based upon rare earth metals became available, that permanent magnet technology rapidly improved. At present, the coercive force (H c ) of a neodymium-ironboron permanent magnet can be as high as A/m (Magnet Sales and Manufacturing, Inc.), which, for a cylindrical tube shaped magnet 2.5 cm high, corresponds to an equivalent coil strength of 22,500 AT. This excitation strength is much higher than that required by conventional SEMs in their normal mode of operation. The coil excitation in a standard objective magnetic lens required to focus a 25 kev beam at a working distance of 5 mm is around 1,000 AT, a factor of 22.5 times smaller than that provided by the permanent magnet. Recently, the possible use of permanent magnets to make miniature electron lenses has been proposed (Adamec et al. 1995). Replacement Specimen chamber SEM Display Permanent magnets are very stable; after 100,000 hours of operation, the change in the field strength is only a few fractions of a percent. Another attractive feature is that they can be made in a variety of different shapes and sizes and with different orientations of magnetisation. Since magnetic electron lenses have cylindrical symmetry, the two possible directions of magnetisation are in the longitudinal direction, parallel to the electron optic axis, or perpendicular to it, in the radial direction. Both possibilities have been advocated for electron optics applications (Adamec et al. 1995, Rieke 1982). In this study, the direction of magnetisation is chosen to be parallel to the electron beam. Although it is possible to manufacture the permanent magnet directly in the required cylindrical tube shape, this is not usually done in many different sizes for high coercive force materials; however, this does not present a major difficulty, since the permanent magnet can be made of curvilinear blocks. Typically, four blocks can be used, one for each quadrant. Furthermore, it is quite common for permanent magnet manufacturers to make custom designs. The permanent magnet lens designs presented in this paper were made through the use of the KEOS finite element programs (Khursheed 1992). These programs are able to include the effect of saturation in the iron circuit and calculate thirdorder deflection aberrations on the primary beam. The incorporation of permanent magnets in KEOS was done by modelling the coercive force in terms of two thin current-carrying coils at the iron-air interface, as shown in Figure 2. The effective relative permeability of a permanent magnet is around unity, and this value is input into the simulation program. It is relatively straightforward to show from Maxwell s magnetostatic equation for the field intensity H and current density J in cylindrical coordinates that the magnitude of the excitation current at each interface NI is given by H c L, where L is the Table Vacuum system Drive electronics Z (optic axis) SEM NI North NI H C L Specimen chamber 1 Specimen chamber 2 South FIG. 1 Scanning electron microscope (SEM) replacement. The portable SEM and different specimen chambers. FIG. 2 Simulation model of the permanent magnet. R (radius)
3 A. Khursheed et al.: Portable SEM design using permanent magnets 89 magnet height. The directions of these interface currents are in the azimuthal direction and are of opposite sign. Note that this formula is only valid if the thickness of each coil is small relative to the magnet s height. A suitable way of choosing the thickness of each coil is to make it equal to the local numerical mesh spacing of the finite element program. The accuracy of the finite element programs was investigated for a two-gap permanent magnet test lens. The axial field distribution for this lens was measured experimentally by the use of a Hall probe. The experimental results agreed to within 10% of the simulation predictions (Khursheed et al. 1997). The Basic Scanning Electron Microscope Column Design Figure 3 shows a schematic of the portable SEM. The SEM has an overall height of around 12 cm. Both the objective lens and condenser lens were designed to avoid saturation problems by increasing the thickness of the iron in the magnetic circuit and placing the position of the permanent magnets far from the electron optic axis. The condenser lens uses two gaps in the same magnetic circuit, and so, in effect, is 80 mm 14 FIG. 3 Portable scanning electron microscope schematic. 1 = Magnetic specimen holder; 2 = nonmagnetic specimen chamber; 3 = magnetic circuit of the objective lens; 4 = deflection coils; 5 = permanent magnet of the objective lens; 6 = magnetic slip ring of the objective lens; 7 = electron detector; 8 = predeflection coils; 9 = magnetic circuit of the condenser lens; 10 = permanent magnet for second air gap of the condenser lens; 11 = permanent magnet for the first air gap of the condenser lens; 12 = upper magnetic slip ring of the condenser lens; 13 = lower magnetic slip ring of the condenser lens; 14 = electron gun equivalent to two condenser lenses. Magnetic slip rings are used to short the magnetic circuit and so control the strength of the magnetic field on axis. These slip rings provide a way of varying the demagnification of the primary beam spot size. The outer magnetic slip ring on the objective lens can be used to focus the primary beam onto the specimen. Focusing can also be achieved by varying the specimen height. The objective lens design is based upon a single pole magnetic immersion lens. The peak magnetic field typically lies around 2 mm below the lower pole piece tip. Deflection coils are set above and within the objective lens and operate together to scan the primary beam in a moving objective lens configuration. The electron detector is placed above the objective lens and collects the secondary and backscattered electrons. The detector may be either a microchannel plate or a scintillator detector. The trajectory paths of the secondary electrons will be collimated due to the sharply decreasing magnetic field of the objective lens, and their transport efficiency from the specimen to the electron detector is expected to be high, better than the transport efficiency for secondary electrons in conventional SEMs. The kind of collection system described here is known as a through-the-lens secondary detection system and is commonly used in high resolution SEM systems (Electron Beam Testing Technology, 1993). The proposed permanent magnet is NdFeB grade 35 (H c = A/m) (Magnet Sales and Manufacturing, Inc.). Various types of iron with different saturation characteristics were considered for use in the magnetic circuit of the lenses. Obviously, high saturation material such as Permendur is preferable, but the simulations showed that the results were not critical on the material type. For the size considered, ordinary mild steel produced results that were only marginally worse than the Permendur material. The flux lines for both the condenser lens and objective lens are shown in Figure 4a and b, respectively. As a first approximation, each lens was analysed separately. The open boundaries around the magnets were substituted with a zero vector potential boundary, which has the effect of lowering the field strength on axis by around 5 to 10%, so the on-axis field strengths presented in this paper are slightly underestimated. It should also be noted that the flux lines at the top of the objective lens and at the bottom of the condenser lens unit are inaccurate since, in practice, the objective lens is connected to the objective lens. This error does not significantly affect the field distribution inside the SEM. In practice, flux lines will connect the top of the condenser lens with the bottom of the objective lens. The flux lines in Figure 4a and b show that the field distribution for the portable SEM is quite different from conventional SEM s. For conventional SEMs, almost all the field distribution is contained within the microscope, but for the portable SEM, the field distribution extends into the region around the. For most applications, this should not affect the operation of the SEM. The equivalent excitation current for the permanent magnets of the condenser lens is 7,200 AT. The permanent mag-
4 90 Scanning Vol. 20, 2 (1998) nets produce a maximum axial field strength of 0.71 Tesla at the centre of each gap in the condenser lens, which is able to produce a focal length of 1.9 mm and a demagnification of 20 for a 25 kev beam. The total demagnification of the condenser lens is thus typically 400. When the slip rings are pushed to the positions which short circuit the permanent magnets, the maximum field strength falls to 0.14 Tesla, giving a focal length of 1.9 mm and a demagnification of 20 at each gap for a 1 kev beam. The axial flux density for the condenser lens is shown in Figure 5. The two narrow peaks correspond to the two gaps in the magnetic circuit. Each gap spans 1 mm. In principle, the condenser lens can be made at least two to three times smaller in size since the height of the magnets is only 8 mm. Figure 6a shows a schematic representation of the pole tip of the objective lens. It is mainly the pole tip dimensions that determine the primary beam optics of the lens. These dimensions include the effective on-axis gap size G in the magnetic circuit, the radius of the lower pole tip R, and the working distance W. The allowable thickness of the specimen is obviously given by G-W. In the following design, R=2.4 mm, which allows sufficient space for in-lens deflection coils to be mounted on the lower pole tip of the objective lens (Zhao and Khursheed et al. 1997). The use of a magnetic specimen holder helps to increase the field strength at the specimen. As the gap size G decreases, the peak magnetic field strength on the axis increases, which in turn lowers the on-axis aberrations. Significant improvement occurs for gap sizes which are < 15 mm; beyond this amount, the results are similar to those obtained for a nonmagnetic specimen holder. For a gap size of mm, a 1 kev primary beam focuses at a working distance of W=1.16 Axial magnetic field strength (Tesla) Gap 1 Gap FIG Distance along the axis (mm) Axial flux density for the condenser lens. Pole tip R W G Axis 0.80 Axial magnetic field strength (Tesla) Lens gap G = mm G = 2 mm FIG. 4 Flux lines for the condenser lens unit; flux lines for the objective lens Pole tip Distance along the axis (mm) FIG. 6 Pole tip and lens gap schematic of the immersion objective lens. Axial field variation around the pole tip for two different gap sizes in the immersion objective lens.
5 A. Khursheed et al.: Portable SEM design using permanent magnets 91 mm. For these low-voltage conditions, the focal length is calculated to be 1.53 mm, the spherical aberration coefficient is mm, and the chromatic aberration coefficient is mm. Conventional magnetic lenses have chromatic aberration coefficients which are typically > 2 mm. Since, at low voltages, it is chromatic aberration which mainly determines the final primary beam spot size, the spatial resolution of the miniature is predicted to be typically a factor of two better than that of conventional SEMs, which is in keeping with the performance expected for magnetic immersion objective lens systems (Reimer 1993). Obviously, for thin specimens the lens gap size G can be reduced. Figure 6b shows that, for G=10.22 mm, the maximum axial flux density is 0.24 Tesla, while for G=2 mm it rises to Tesla and is able to focus a 20 kv beam at a working distance of 2.19 mm. For these operating conditions, the focal length is 1.76 mm, the spherical aberration coefficient is 1 mm, and the chromatic aberration coefficient is 1.24 mm, giving a spot resolution of 0.63 nm for a Schottky gun emitter (with an assumed energy spread of 0.8 ev). These results show that the portable SEM should be capable of operating over a wide range of beam-operating voltages. Further increases in the axial magnetic field strength can be obtained by using a longer permanent magnet. For low-beam voltages, such as 1 kv, the outer slip ring is used to short the magnetic circuit. There are, in principle, many ways in which a number of portable SEM s can function together. One possibility is to use several SEM s for simultaneous inspection of different parts of a large specimen, such as a 150 mm diameter integrated circuit wafer. In this case, the specimen is placed on a large magnetic table, and instead of moving the specimen to examine the circuits on it, each SEM is moved. The diameter of each SEM can be designed to be < 50 mm, making it possible to have an array of at least four SEM s simultaneously inspecting the specimen. For other applications, it may be possible to create an SEM cluster, where an array of SEM s collect information in parallel. Conclusion A portable SEM design has been presented which is based upon the use of permanent magnets. The SEM s height is designed to be below 12 cm. Computer simulations predict that the SEM should be able to provide high spatial resolution. The is designed to be modular, so that it can be fitted onto a wide variety of different specimen chambers, and can also be easily replaced. A patent application on the portable SEM design has been made, and a project involving the National University of Singapore and the Institute of Materials Research and Engineering in Singapore has been initiated to provide the experimental verification of the simulation results presented in this paper. References Adamec P, Delong A, Lencova B: Miniature magnetic electron lenses with permanent magnets. J Microsc 179, Pt. 2, (1995) Chang THP, Kern DP, Muray LP: Microminiaturization of electron optical systems. J Vac Sci Technol B8 (6) (1990) Chisholm T, Liu H, Munro E, Rouse J, Zhu X: A compact electrostatic lithography for nanoscale exposure. Proc SPIE, Charged Particle Optics III, 3155, (1997) Electron Beam Testing Technology (Ed. Thong JTL), Ch. 2, 3, and 5. Plenum Press, New York (1993) Khursheed A: KEOS: Khursheed Electron Optics Software. Department of Electrical Engineering, University of Edinburgh, Mayfield Road, Edinburgh EH9 3JL, Scotland, UK (1992) Khursheed A, Thong JTL, Phang JC, Ong IP: Miniature scanning electron microscope design based upon the use of permanent magnets. Proc SPIE, Charged Particle Optics III, 3155, (1997) Magnet Sales and Manufacturing, Inc., Playa Court, Culver City, CA Reimer L: Image formation in low voltage scanning electron microscopy. Ch. 1, Tutorial texts. SPIE (1993) Rieke WD: Practical lens design. In Magnetic Electron Lenses, section 4.1 and 4.6, (Ed. Hawkes PW). Springer-Verlag, New York (1982) Zhao Y, Khursheed A: In-lens deflectors for a LVSEM magnetic immersion objective lens. Proc SPIE, Charged Particle Optics III, 3155, (1997)
Mohammed A. Hussein *
International Journal of Physics, 216, Vol. 4, No. 5, 13-134 Available online at http://pubs.sciepub.com/ijp/4/5/3 Science and Education Publishing DOI:1.12691/ijp-4-5-3 Effect of the Geometrical Shape
More informationNumerical analysis to verifying the performance of condenser magnetic lens in the scanning electron microscope.
Numerical analysis to verifying the performance of condenser magnetic lens in the scanning electron microscope. Mohammed Abdullah Hussein Dept. of mechanization and agricultural equipment, College of agriculture
More informationA Parallel Radial Mirror Energy Analyzer Attachment for the Scanning Electron Microscope
142 doi:10.1017/s1431927615013288 Microscopy Society of America 2015 A Parallel Radial Mirror Energy Analyzer Attachment for the Scanning Electron Microscope Kang Hao Cheong, Weiding Han, Anjam Khursheed
More informationDesign and fabrication of a scanning electron microscope using a finite element analysis for electron optical system
Journal of Mechanical Science and Technology 22 (2008) 1734~1746 Journal of Mechanical Science and Technology www.springerlink.com/content/1738-494x DOI 10.1007/s12206-008-0317-9 Design and fabrication
More informationFunctions of the SEM subsystems
Functions of the SEM subsystems Electronic column It consists of an electron gun and two or more electron lenses, which influence the path of electrons traveling down an evacuated tube. The base of the
More informationINTRODUCTION THIN LENSES. Introduction. given by the paraxial refraction equation derived last lecture: Thin lenses (19.1) = 1. Double-lens systems
Chapter 9 OPTICAL INSTRUMENTS Introduction Thin lenses Double-lens systems Aberrations Camera Human eye Compound microscope Summary INTRODUCTION Knowledge of geometrical optics, diffraction and interference,
More informationDesign and Application of a Quadrupole Detector for Low-Voltage Scanning Electron Mcroscopy
SCANNING Vol. 8, 294-299 (1986) 0 FACM. Inc. Received: August 29, 1986 Original Paper Design and Application of a Quadrupole Detector for Low-Voltage Scanning Electron Mcroscopy R. Schmid and M. Brunner"
More informationSoftware for Electron and Ion Beam Column Design. An integrated workplace for simulating and optimizing electron and ion beam columns
OPTICS Software for Electron and Ion Beam Column Design An integrated workplace for simulating and optimizing electron and ion beam columns Base Package (OPTICS) Field computation Imaging and paraxial
More informationScanning electron microscope
Scanning electron microscope 6 th CEMM workshop Maja Koblar, Sc. Eng. Physics Outline The basic principle? What is an electron? Parts of the SEM Electron gun Electromagnetic lenses Apertures Chamber and
More informationScanning electron microscope
Scanning electron microscope 5 th CEMM workshop Maja Koblar, Sc. Eng. Physics Outline The basic principle? What is an electron? Parts of the SEM Electron gun Electromagnetic lenses Apertures Detectors
More informationNANO 703-Notes. Chapter 9-The Instrument
1 Chapter 9-The Instrument Illumination (condenser) system Before (above) the sample, the purpose of electron lenses is to form the beam/probe that will illuminate the sample. Our electron source is macroscopic
More informationLow Voltage Electron Microscope
LVEM5 Low Voltage Electron Microscope Nanoscale from your benchtop LVEM5 Delong America DELONG INSTRUMENTS COMPACT BUT POWERFUL The LVEM5 is designed to excel across a broad range of applications in material
More informationMeshing Challenges in Simulating the Induced Currents in Vacuum Phototriode
Meshing Challenges in Simulating the Induced Currents in Vacuum Phototriode S. Zahid and P. R. Hobson Electronic and Computer Engineering, Brunel University London, Uxbridge, UB8 3PH UK Introduction Vacuum
More informationDesign of a high brightness multi-electron-beam source
vailable online at www.sciencedirect.com Physics Procedia00 1 (2008) 000 000 553 563 www.elsevier.com/locate/procedia www.elsevier.com/locate/xxx Proceedings of the Seventh International Conference on
More informationIntroduction of New Products
Field Emission Electron Microscope JEM-3100F For evaluation of materials in the fields of nanoscience and nanomaterials science, TEM is required to provide resolution and analytical capabilities that can
More informationEffect of Shot Noise and Secondary Emission Noise in Scanning Electron Microscope Images
SCANNING VOL. 26, 36 40 (2004) Received: March 7, 2003 FAMS, Inc. Accepted with revision: October 14, 2003 Effect of Shot Noise and Secondary Emission Noise in Scanning Electron Microscope Images K. S.
More information2.Components of an electron microscope. a) vacuum systems, b) electron guns, c) electron optics, d) detectors. Marco Cantoni 021/
2.Components of an electron microscope a) vacuum systems, b) electron guns, c) electron optics, d) detectors, 021/693.48.16 Centre Interdisciplinaire de Microscopie Electronique CIME Summary Electron propagation
More informationElectron Sources, Optics and Detectors
Thomas LaGrange, Ph.D. Faculty Lecturer and Senior Staff Scientist Electron Sources, Optics and Detectors TEM Doctoral Course MS-637 April 16 th -18 th, 2018 Summary Electron propagation is only possible
More informationMODULE I SCANNING ELECTRON MICROSCOPE (SEM)
MODULE I SCANNING ELECTRON MICROSCOPE (SEM) Scanning Electron Microscope (SEM) Initially, the plan of SEM was offered by H. Stintzing in 1927 (a German patent application). His suggested procedure was
More informationSCANNING ELECTRON MICROSCOPY AND X-RAY MICROANALYSIS
SCANNING ELECTRON MICROSCOPY AND X-RAY MICROANALYSIS Robert Edward Lee Electron Microscopy Center Department of Anatomy and Neurobiology Colorado State University P T R Prentice Hall, Englewood Cliffs,
More informationTransmission Electron Microscopy 9. The Instrument. Outline
Transmission Electron Microscopy 9. The Instrument EMA 6518 Spring 2009 02/25/09 Outline The Illumination System The Objective Lens and Stage Forming Diffraction Patterns and Images Alignment and Stigmation
More informationLow Voltage Electron Microscope. Nanoscale from your benchtop LVEM5. Delong America
LVEM5 Low Voltage Electron Microscope Nanoscale from your benchtop LVEM5 Delong America DELONG INSTRUMENTS COMPACT BUT POWERFUL The LVEM5 is designed to excel across a broad range of applications in material
More information(Refer Slide Time: 00:10)
Fundamentals of optical and scanning electron microscopy Dr. S. Sankaran Department of Metallurgical and Materials Engineering Indian Institute of Technology, Madras Module 03 Unit-6 Instrumental details
More informationTutorial: designing a converging-beam electron gun and focusing solenoid with Trak and PerMag
Tutorial: designing a converging-beam electron gun and focusing solenoid with Trak and PerMag Stanley Humphries, Copyright 2012 Field Precision PO Box 13595, Albuquerque, NM 87192 U.S.A. Telephone: +1-505-220-3975
More informationWIEN Software for Design of Columns Containing Wien Filters and Multipole Lenses
WIEN Software for Design of Columns Containing Wien Filters and Multipole Lenses An integrated workplace for analysing and optimising the column optics Base Package (WIEN) Handles round lenses, quadrupoles,
More informationLow Voltage Electron Microscope
LVEM 25 Low Voltage Electron Microscope fast compact powerful Delong America FAST, COMPACT AND POWERFUL The LVEM 25 offers a high-contrast, high-throughput, and compact solution with nanometer resolutions.
More informationLVEM 25. Low Voltage Electron Mictoscope. fast compact powerful
LVEM 25 Low Voltage Electron Mictoscope fast compact powerful FAST, COMPACT AND POWERFUL The LVEM 25 offers a high-contrast, high-throughput, and compact solution with nanometer resolutions. All the benefits
More informationPerformance evaluation of fractional-slot tubular permanent magnet machines with low space harmonics
ARCHIVES OF ELECTRICAL ENGINEERING DOI 10.1515/aee-2015-0049 VOL. 64(4), pp. 655-668 (2015) Performance evaluation of fractional-slot tubular permanent magnet machines with low space harmonics Jiabin Wang
More informationChapter 2 Instrumentation for Analytical Electron Microscopy Lecture 7. Chapter 2 CHEM Fall L. Ma
Chapter 2 Instrumentation for Analytical Electron Microscopy Lecture 7 Outline Electron Sources (Electron Guns) Thermionic: LaB 6 or W Field emission gun: cold or Schottky Lenses Focusing Aberration Probe
More informationOct. 30th- Nov. 1st, 2017
Thomas LaGrange, Ph.D. Faculty Lecturer and Senior Staff Scientist Electron Sources, Optics and Detectors SEM Doctoral Course MS-636 Oct. 30th- Nov. 1st, 2017 Summary Electron propagation is only possible
More informationA DIGITIZING DEVICE FOR FILMLESS VISUAL DETECTORS. F. Villa Stanford Linear Accelerator Center ABSTRACT
-1- SS-7S 2100 A DIGITIZING DEVICE FOR FILMLESS VISUAL DETECTORS F. Villa Stanford Linear Accelerator Center ABSTRACT We describe a device for eliminating film as data storage for visual detectors. The
More informationConfocal chromatic sensors and confocal microscope Micrometer measurement of thickness, displacement, position
Confocal chromatic sensors and confocal microscope Micrometer measurement of thickness, displacement, position 2 optoncdt 2401 Confocal displacement measurement system - Non-contact measurement principle
More information2.5D Finite Element Simulation Eddy Current Heat Exchanger Tube Inspection using FEMM
Vol.20 No.7 (July 2015) - The e-journal of Nondestructive Testing - ISSN 1435-4934 www.ndt.net/?id=18011 2.5D Finite Element Simulation Eddy Current Heat Exchanger Tube Inspection using FEMM Ashley L.
More informationCHAPTER 2 ELECTROMAGNETIC FORCE AND DEFORMATION
18 CHAPTER 2 ELECTROMAGNETIC FORCE AND DEFORMATION 2.1 INTRODUCTION Transformers are subjected to a variety of electrical, mechanical and thermal stresses during normal life time and they fail when these
More informationReflecting optical system to increase signal intensity. in confocal microscopy
Reflecting optical system to increase signal intensity in confocal microscopy DongKyun Kang *, JungWoo Seo, DaeGab Gweon Nano Opto Mechatronics Laboratory, Dept. of Mechanical Engineering, Korea Advanced
More informationS200 Course LECTURE 1 TEM
S200 Course LECTURE 1 TEM Development of Electron Microscopy 1897 Discovery of the electron (J.J. Thompson) 1924 Particle and wave theory (L. de Broglie) 1926 Electromagnetic Lens (H. Busch) 1932 Construction
More informationNo part of this material may be reproduced without explicit written permission.
This material is provided for educational use only. The information in these slides including all data, images and related materials are the property of : Robert M. Glaeser Department of Molecular & Cell
More informationScanning Electron Microscopy. EMSE-515 F. Ernst
Scanning Electron Microscopy EMSE-515 F. Ernst 1 2 Scanning Electron Microscopy Max Knoll Manfred von Ardenne Manfred von Ardenne Principle of Scanning Electron Microscopy 3 Principle of Scanning Electron
More informationScanning Electron Microscopy Basics and Applications
Scanning Electron Microscopy Basics and Applications Dr. Julia Deuschle Stuttgart Center for Electron Microscopy MPI for Solid State Research Room: 1E15, phone: 0711/ 689-1193 email: j.deuschle@fkf.mpg.de
More informationVISUAL PHYSICS ONLINE DEPTH STUDY: ELECTRON MICROSCOPES
VISUAL PHYSICS ONLINE DEPTH STUDY: ELECTRON MICROSCOPES Shortly after the experimental confirmation of the wave properties of the electron, it was suggested that the electron could be used to examine objects
More informationPROCEEDINGS OF A SYMPOSIUM HELD AT THE CAVENDISH LABORATORY, CAMBRIDGE, Edited by
X - R A Y M I C R O S C O P Y A N D M I C R O R A D I O G R A P H Y PROCEEDINGS OF A SYMPOSIUM HELD AT THE CAVENDISH LABORATORY, CAMBRIDGE, 1956 Edited by V. E. COSSLETT Cavendish Laboratory, University
More informationOct. 30th- Nov. 1st, 2017
Thomas LaGrange, Ph.D. Faculty Lecturer and Senior Staff Scientist Electron Sources, Optics and Detectors SEM Doctoral Course MS-636 Oct. 30th- Nov. 1st, 2017 Summary Electron propagation is only possible
More informationPicture perfect. Electromagnetic simulations of transformers
38 ABB review 3 13 Picture perfect Electromagnetic simulations of transformers Daniel Szary, Janusz Duc, Bertrand Poulin, Dietrich Bonmann, Göran Eriksson, Thorsten Steinmetz, Abdolhamid Shoory Power transformers
More informationEDDY CURRENT INSPECTION FOR DEEP CRACK DETECTION AROUND FASTENER HOLES IN AIRPLANE MULTI-LAYERED STRUCTURES
EDDY CURRENT INSPECTION FOR DEEP CRACK DETECTION AROUND FASTENER HOLES IN AIRPLANE MULTI-LAYERED STRUCTURES Teodor Dogaru Albany Instruments Inc., Charlotte, NC tdogaru@hotmail.com Stuart T. Smith Center
More informationBuilding a New Software of Electromagnetic Lenses (CADTEL)
International Letters of Chemistry, Physics and Astronomy Online: 2013-03-03 ISSN: 2299-3843, Vol. 9, pp 46-55 doi:10.18052/www.scipress.com/ilcpa.9.46 2013 SciPress Ltd., Switzerland Building a New Software
More informationElectron Beam Lithography. Adam Ramm
Electron Beam Lithography Adam Ramm Why use electrons? Negligible diffraction limitations: R = k λ NA With current optical technology, this equates to about 45nm resolution. For an electron, wavelength
More informationParallel Mode Confocal System for Wafer Bump Inspection
Parallel Mode Confocal System for Wafer Bump Inspection ECEN5616 Class Project 1 Gao Wenliang wen-liang_gao@agilent.com 1. Introduction In this paper, A parallel-mode High-speed Line-scanning confocal
More informationIntroduction to Scanning Electron Microscopy
Introduction to Scanning Electron Microscopy By: Brandon Cheney Ant s Leg Integrated Circuit Nano-composite This document was created as part of a Senior Project in the Materials Engineering Department
More informationTechnical Explanation for Displacement Sensors and Measurement Sensors
Technical Explanation for Sensors and Measurement Sensors CSM_e_LineWidth_TG_E_2_1 Introduction What Is a Sensor? A Sensor is a device that measures the distance between the sensor and an object by detecting
More informationLenses Design Basics. Introduction. RONAR-SMITH Laser Optics. Optics for Medical. System. Laser. Semiconductor Spectroscopy.
Introduction Optics Application Lenses Design Basics a) Convex lenses Convex lenses are optical imaging components with positive focus length. After going through the convex lens, parallel beam of light
More information3. What is hysteresis loss? Also mention a method to minimize the loss. (N-11, N-12)
DHANALAKSHMI COLLEGE OF ENGINEERING, CHENNAI DEPARTMENT OF ELECTRICAL AND ELECTRONICS ENGINEERING EE 6401 ELECTRICAL MACHINES I UNIT I : MAGNETIC CIRCUITS AND MAGNETIC MATERIALS Part A (2 Marks) 1. List
More informationMODELLING OF A MAGNETIC ADHESION ROBOT FOR NDT INSPECTION OF LARGE METAL STRUCTURES
MODELLING OF A MAGNETIC ADHESION ROBOT FOR NDT INSPECTION OF LARGE METAL STRUCTURES G. SHIRKOOHI and Z. ZHAO School of Engineering, London South Bank University, 103 Borough Road, London SE1 0AA United
More informationMAGNETOSCOP Measurement of magnetic field strengths in the range 0.1 nanotesla to 1 millitesla
MAGNETOSCOP Measurement of magnetic field strengths in the range 0.1 nanotesla to 1 millitesla Extremely high sensitivity of 0.1 nanotesla with field and gradient probe Measurement of material permeabilities
More informationChapter 1. Basic Electron Optics (Lecture 2)
Chapter 1. Basic Electron Optics (Lecture 2) Basic concepts of microscope (Cont ) Fundamental properties of electrons Electron Scattering Instrumentation Basic conceptions of microscope (Cont ) Ray diagram
More informationTarget Temperature Effect on Eddy-Current Displacement Sensing
Target Temperature Effect on Eddy-Current Displacement Sensing Darko Vyroubal Karlovac University of Applied Sciences Karlovac, Croatia, darko.vyroubal@vuka.hr Igor Lacković Faculty of Electrical Engineering
More informationCs-corrector. Felix de Haas
Cs-corrector. Felix de Haas Content Non corrector systems Lens aberrations and how to minimize? Corrector systems How is it done? Lens aberrations Spherical aberration Astigmatism Coma Chromatic Quality
More information2.Components of an electron microscope. a) vacuum systems, b) electron guns, c) electron optics, d) detectors. Marco Cantoni, 021/
2.Components of an electron microscope a) vacuum systems, b) electron guns, c) electron optics, d) detectors Marco Cantoni, 021/693.48.16 Centre Interdisciplinaire de Microscopie Electronique CIME MSE-603
More informationChapter 18 Optical Elements
Chapter 18 Optical Elements GOALS When you have mastered the content of this chapter, you will be able to achieve the following goals: Definitions Define each of the following terms and use it in an operational
More informationELECTRON MICROSCOPY. 13:10 16:00, Oct. 6, 2008 Institute of Physics, Academia Sinica. Tung Hsu
ELECTRON MICROSCOPY 13:10 16:00, Oct. 6, 2008 Institute of Physics, Academia Sinica Tung Hsu Department of Materials Science and Engineering National Tsing Hua University Hsinchu 300, TAIWAN Tel. 03-5742564
More informationELECTRON OPTICS. Prof. John G. King Dr. John W. Coleman Dr. Edward H. Jacobsen. Graduate Students. Steven R. Jost Norman D. Punsky
II. ELECTRON OPTICS Academic and Research Staff Prof. John G. King Dr. John W. Coleman Dr. Edward H. Jacobsen Graduate Students Steven R. Jost Norman D. Punsky A. HIGH-RESOLUTION HIGH-CONTRAST ELECTRON
More informationNotes on the VPPEM electron optics
Notes on the VPPEM electron optics Raymond Browning 2/9/2015 We are interested in creating some rules of thumb for designing the VPPEM instrument in terms of the interaction between the field of view at
More informationLenses. Optional Reading Stargazer: the life and times of the TELESCOPE, Fred Watson (Da Capo 2004).
Lenses Equipment optical bench, incandescent light source, laser, No 13 Wratten filter, 3 lens holders, cross arrow, diffuser, white screen, case of lenses etc., vernier calipers, 30 cm ruler, meter stick
More informationZero Focal Shift in High Numerical Aperture Focusing of a Gaussian Laser Beam through Multiple Dielectric Interfaces. Ali Mahmoudi
1 Zero Focal Shift in High Numerical Aperture Focusing of a Gaussian Laser Beam through Multiple Dielectric Interfaces Ali Mahmoudi a.mahmoudi@qom.ac.ir & amahmodi@yahoo.com Laboratory of Optical Microscopy,
More informationRecent results from the JEOL JEM-3000F FEGTEM in Oxford
Recent results from the JEOL JEM-3000F FEGTEM in Oxford R.E. Dunin-Borkowski a, J. Sloan b, R.R. Meyer c, A.I. Kirkland c,d and J. L. Hutchison a a b c d Department of Materials, Parks Road, Oxford OX1
More informationVery short introduction to light microscopy and digital imaging
Very short introduction to light microscopy and digital imaging Hernan G. Garcia August 1, 2005 1 Light Microscopy Basics In this section we will briefly describe the basic principles of operation and
More informationInsertion Devices Lecture 4 Undulator Magnet Designs. Jim Clarke ASTeC Daresbury Laboratory
Insertion Devices Lecture 4 Undulator Magnet Designs Jim Clarke ASTeC Daresbury Laboratory Hybrid Insertion Devices Inclusion of Iron Simple hybrid example Top Array e - Bottom Array 2 Lines of Magnetic
More informationNANO MODIFICATION OF THE W(100)/ZrO ELECTRON EMITTER TIP USING REACTIVE ION ETCHING
NANO MODIFICATION OF THE W(100)/ZrO ELECTRON EMITTER TIP USING REACTIVE ION ETCHING Miroslav HORÁČEK, František MATĚJKA, Vladimír KOLAŘÍK, Milan MATĚJKA, Michal URBÁNEK Ústav přístrojové techniky AV ČR,
More informationMIMS Workshop F. Hillion. MIMS Workshop
MIMS Workshop 23 - F. Hillion MIMS Workshop 1/ Practical aspects of N5 Tuning Primary column : small probe, high current, influence of Z. Dynamic Transfer and scanning. Cy and P2/P3. LF4, Q and chromatic
More informationScanning Electron Microscopy
Scanning Electron Microscopy For the semiconductor industry A tutorial Titel Vorname Nachname Titel Jobtitle, Bereich/Abteilung Overview Scanning Electron microscopy Scanning Electron Microscopy (SEM)
More informationA Numerical Study of Depth of Penetration of Eddy Currents
A Numerical Study of Depth of Penetration of Eddy Currents S.Majidnia* a,b, R.Nilavalan b, J. Rudlin a a. TWI Ltd, Cambridge,United Kingdom b Brunel University, London,United Kingdom shiva.majidnia@twi.co.uk
More informationA Practical Guide to Free Energy Devices
A Practical Guide to Free Energy Devices Part PatD14: Last updated: 25th February 2006 Author: Patrick J. Kelly This patent application shows the details of a device which it is claimed, can produce sufficient
More informationCHAPTER 11 HPD (Hybrid Photo-Detector)
CHAPTER 11 HPD (Hybrid Photo-Detector) HPD (Hybrid Photo-Detector) is a completely new photomultiplier tube that incorporates a semiconductor element in an evacuated electron tube. In HPD operation, photoelectrons
More informationPHY170: OPTICS. Things to do in the lab INTRODUCTORY REMARKS OPTICS SIMULATIONS
INTRODUCTORY REMARKS PHY170: OPTICS The optics experiments consist of two major parts. Setting up various components and performing the experiments described below. Computer simulation of images generated
More informationELECTRON MICROSCOPY AN OVERVIEW
ELECTRON MICROSCOPY AN OVERVIEW Anjali Priya 1, Abhishek Singh 2, Nikhil Anand Srivastava 3 1,2,3 Department of Electrical & Instrumentation, Sant Longowal Institute of Engg. & Technology, Sangrur, India.
More informationHeisenberg) relation applied to space and transverse wavevector
2. Optical Microscopy 2.1 Principles A microscope is in principle nothing else than a simple lens system for magnifying small objects. The first lens, called the objective, has a short focal length (a
More informationX-ray generation by femtosecond laser pulses and its application to soft X-ray imaging microscope
X-ray generation by femtosecond laser pulses and its application to soft X-ray imaging microscope Kenichi Ikeda 1, Hideyuki Kotaki 1 ' 2 and Kazuhisa Nakajima 1 ' 2 ' 3 1 Graduate University for Advanced
More informationMINIATURE X-RAY SOURCES AND THE EFFECTS OF SPOT SIZE ON SYSTEM PERFORMANCE
228 MINIATURE X-RAY SOURCES AND THE EFFECTS OF SPOT SIZE ON SYSTEM PERFORMANCE D. CARUSO, M. DINSMORE TWX LLC, CONCORD, MA 01742 S. CORNABY MOXTEK, OREM, UT 84057 ABSTRACT Miniature x-ray sources present
More informationSCIENTIFIC INSTRUMENT NEWS. Introduction. Design of the FlexSEM 1000
SCIENTIFIC INSTRUMENT NEWS 2017 Vol. 9 SEPTEMBER Technical magazine of Electron Microscope and Analytical Instruments. Technical Explanation The FlexSEM 1000: A Scanning Electron Microscope Specializing
More informationTopics 3b,c Electron Microscopy
Topics 3b,c Electron Microscopy 1.0 Introduction and History 1.1 Characteristic Information 2.0 Basic Principles 2.1 Electron-Solid Interactions 2.2 Electromagnetic Lenses 2.3 Breakdown of an Electron
More informationTransmissions Electron Microscopy (TEM)
Transmissions Electron Microscopy (TEM) Basic principles Diffraction Imaging Specimen preparation A.E. Gunnæs MENA3100 V17 TEM is based on three possible set of techniqes Diffraction From regions down
More informationCHAPTER 9 POSITION SENSITIVE PHOTOMULTIPLIER TUBES
CHAPTER 9 POSITION SENSITIVE PHOTOMULTIPLIER TUBES The current multiplication mechanism offered by dynodes makes photomultiplier tubes ideal for low-light-level measurement. As explained earlier, there
More informationIntroduction to Light Microscopy. (Image: T. Wittman, Scripps)
Introduction to Light Microscopy (Image: T. Wittman, Scripps) The Light Microscope Four centuries of history Vibrant current development One of the most widely used research tools A. Khodjakov et al. Major
More informationOPERATION OF THE HITACHI S-450 SCANNING ELECTRON MICROSCOPE. by Doug Bray Department of Biological Sciences University of Lethbridge
OPERATION OF THE HITACHI S-450 SCANNING ELECTRON MICROSCOPE by Doug Bray Department of Biological Sciences University of Lethbridge Revised September, 2000 Note: The terms in bold in this document represent
More informationPerformance Factors. Technical Assistance. Fundamental Optics
Performance Factors After paraxial formulas have been used to select values for component focal length(s) and diameter(s), the final step is to select actual lenses. As in any engineering problem, this
More informationScanning Electron Microscopy SEM. Warren Straszheim, PhD MARL, 23 Town Engineering
Scanning Electron Microscopy SEM Warren Straszheim, PhD MARL, 23 Town Engineering wesaia@iastate.edu 515-294-8187 How it works Create a focused electron beam Accelerate it Scan it across the sample Map
More information6 Electromagnetic Field Distribution Measurements using an Optically Scanning Probe System
6 Electromagnetic Field Distribution Measurements using an Optically Scanning Probe System TAKAHASHI Masanori, OTA Hiroyasu, and ARAI Ken Ichi An optically scanning electromagnetic field probe system consisting
More informationSenderovich 1. Figure 1: Basic electrode chamber geometry.
Senderovich 1 Electrode Design Adjustments to a High Voltage Electron Gun Igor Senderovich Abstract In order to emit and accelerate electron bunches for the new ERL demanding small longitudinal emittance,
More informationIntroduction to Electron Microscopy-II
Introduction to Electron Microscopy-II Prof. David Muller, dm24@cornell.edu Rm 274 Clark Hall, 255-4065 Ernst Ruska and Max Knoll built the first electron microscope in 1931 (Nobel Prize to Ruska in 1986)
More informationIntroduction to Electron Microscopy
Introduction to Electron Microscopy Prof. David Muller, dm24@cornell.edu Rm 274 Clark Hall, 255-4065 Ernst Ruska and Max Knoll built the first electron microscope in 1931 (Nobel Prize to Ruska in 1986)
More informationattocfm I for Surface Quality Inspection NANOSCOPY APPLICATION NOTE M01 RELATED PRODUCTS G
APPLICATION NOTE M01 attocfm I for Surface Quality Inspection Confocal microscopes work by scanning a tiny light spot on a sample and by measuring the scattered light in the illuminated volume. First,
More informationAssembly and Experimental Characterization of Fiber Collimators for Low Loss Coupling
Assembly and Experimental Characterization of Fiber Collimators for Low Loss Coupling Ruby Raheem Dept. of Physics, Heriot Watt University, Edinburgh, Scotland EH14 4AS, UK ABSTRACT The repeatability of
More informationInduction heating of internal
OPTIMAL DESIGN OF INTERNAL INDUCTION COILS The induction heating of internal surfaces is more complicated than heating external ones. The three main types of internal induction coils each has its advantages
More informationOPTICAL SYSTEMS OBJECTIVES
101 L7 OPTICAL SYSTEMS OBJECTIVES Aims Your aim here should be to acquire a working knowledge of the basic components of optical systems and understand their purpose, function and limitations in terms
More informationFabrication of Probes for High Resolution Optical Microscopy
Fabrication of Probes for High Resolution Optical Microscopy Physics 564 Applied Optics Professor Andrès La Rosa David Logan May 27, 2010 Abstract Near Field Scanning Optical Microscopy (NSOM) is a technique
More informationANALYTICAL MICRO X-RAY FLUORESCENCE SPECTROMETER
Copyright(c)JCPDS-International Centre for Diffraction Data 2001,Advances in X-ray Analysis,Vol.44 325 ANALYTICAL MICRO X-RAY FLUORESCENCE SPECTROMETER ABSTRACT William Chang, Jonathan Kerner, and Edward
More informationChapter 9 - Ray Optics and Optical Instruments. The image distance can be obtained using the mirror formula:
Question 9.1: A small candle, 2.5 cm in size is placed at 27 cm in front of a concave mirror of radius of curvature 36 cm. At what distance from the mirror should a screen be placed in order to obtain
More informationELECTRON MICROSCOPY. 14:10 17:00, Apr. 3, 2007 Department of Physics, National Taiwan University. Tung Hsu
ELECTRON MICROSCOPY 14:10 17:00, Apr. 3, 2007 Department of Physics, National Taiwan University Tung Hsu Department of Materials Science and Engineering National Tsinghua University Hsinchu 300, TAIWAN
More information!"#$$%$&'()*"+,-$'.%",-/"-0*'-$'#'12%0'
!#$$%$&'()*+,-$'.%,-/-0*'-$'#'12%0' #$%&'()*+!#$%&'()*+,'-+%./%0123453%6%$3+*.-.,27%#.3%$-8'9'37%$%:;;:
More informationConformal optical system design with a single fixed conic corrector
Conformal optical system design with a single fixed conic corrector Song Da-Lin( ), Chang Jun( ), Wang Qing-Feng( ), He Wu-Bin( ), and Cao Jiao( ) School of Optoelectronics, Beijing Institute of Technology,
More information5. The Scanning Electron Microscope
Physical Principles of Electron Microscopy 5. The Scanning Electron Microscope Ray Egerton University of Alberta and National Institute of Nanotechnology Edmonton, Canada www.tem-eels.ca regerton@ualberta.ca
More information