JEOL 5DII. Operation introduction. By Serge Charlebois

Size: px
Start display at page:

Download "JEOL 5DII. Operation introduction. By Serge Charlebois"

Transcription

1 JEOL 5DII Operation introduction By Serge Charlebois July 2003

2 General procedure Loading the cassette in the load lock Selecting EOS mode, table and aperture Setting and maximising current Observation mode Using the 4 th lens in 5 th lens modes Moving the stage (1) Setting focus, centering the aperture and correcting astigmatism Loading the cassette onto the exposure stage Using the CALIB program Moving the stage (2) Getting closer to exposure time Correcting for stage rotation and image distortion Setting the table for mark detection Mark detection adjustment

3 Loading the cassette in the load lock (1) Turn the load lock door handle and press the flashing red VENT button When the door opens depress the VENT button to stop the N 2 flow The cassette must be inserted in the load lock with the copper handle facing to the right of the of the load lock, i.e. towards the manipulator arm. Close the door and turn the door handle into locking position Cassette F JEOL In the JEOL, the sample will be viewed in the same way as seen in the cassette when held with the copper handle to the right (the F in the illustration will be viewed as is in the JEOL) Load Lock Manipulator Cassette User

4 Loading the cassette in the load lock (2) To be able to open the load lock, the elevator must be in initial position Once loaded and inserted in the load lock the filled cassette position should be marked by a green LED in the cassette column The STEP UP green LED should also be lit to indicate that the cassette holder in is the right position (must be tightly screwed in) These LED indicate the used cassette slots STEP UP The AUTO LOADER section is located at the upper far right of the console STAGE INITIAL

5 Selecting EOS mode and table Log on to the FYVAX7 in the DEC terminal window Start the CALIB program enter DEC as the terminal type enter the EOS mode as chosen in the conversion process enter the EOS table assigned to you for this EOS mode (see user table) Start the RESTOR routine to demagnetise the lens and return to a known/previous state of the column Select the proper aperture Function of the desired current (see beam current graphs)

6 Setting and maximising current (1a) The column optimisation is done by maximising the current with the help of various alignment coils The current is measured using a Faraday cup that must be placed under the beam Use the CURREN routine in CALIB to have the Faraday cup positioned and current measured automatically. This is the normal current measurement procedure

7 Setting and maximising current (1b) If the current value obtained is closed to the desired value The column is close to optimal alignment Activate the NORMAL current measurement mode and go to Setting and maximising current (3) If the current value obtained is far from that desired The column is not well aligned The tables might not have been used for a long time The JEOL might have been realigned, etc. Activate the NORMAL current measurement mode and go to Setting and maximising current (2) Current value View of the picoammeter on the upper console with the door open NORMAL mode

8 Setting and maximising current (2a) Setting the reference parameters Example of reference parameter file entry 29-jun-03 Bengt Nilsson BIAS 46 FIL 15 G-AL XH = 104 XT = -305 YH = 119 YT = AL XH = 65 XT = -191 YH = 59 YT = nA 2nd = XH Enter the alignment coil parameters Select G-AL and using the 4 knobs enter the values XH, XT, YH and YT Select 1-AL and using the 4 knobs enter the values XH, XT, YH and YT YH Alternatively selected the G-AL coil and the 1-AL coil to enter the 8 parameters XT YT

9 Setting and maximising current (2b) Setting the reference parameters Set the current Unselect 4TH and 5TH lens Press 2/3 then press 2ND Use the COARSE and FINE button above to set the value as in the reference file The current should now be approximately that stated in the reference file It may vary if you are not using the same aperture The lens panel is located in the middle of the lower console In all cases, do not use any value below ~ You might find proper current (low range) but it will be unstable and the beam spot will be deformed and potentially multiple First unselect 4TH and 5TH Then press 2/3 and then 2ND

10 Setting and maximising current (3) Set the current to the desired value See Setting and maximising current (2b) Select G-AL Use the two left buttons to maximise Select 1-AL Use the two right buttons to maximise Readjust the current to the desired value and repeat the maximisation cycle For large currents you may have to repeat the cycle numerous times increasing the current by small amount every time When satisfied set to REMOTE measurement. See Setting and maximising current (1b) Start CURREN routine on the computer and save the value Use the left buttons with the 1-AL coil Use the right buttons with the 1-AL coil

11 Beam blanker SEI - Secondary electron detector Observation mode (1) Overview of the buttons used Focus adjustment Contrast and brightness with SEI Lens selection Magnification Scanning speed Selection of the SEI or BEI detector Scanning area: Normal wide area Rapid small area Contrast and brightness for the BEI detector

12 Observation mode (2) You will occasionally need to use the JEOL to observe part of the stage/sample as one would do with a SEM Move the stage to the location you want to observe ex. start INIMSH routine Press the desired lens 4TH has a wider field of view and is therefore easier to use to locate objects Press SEI and confirm that the BLK (beam blanker) is off You should see a the beam scanning on the screen Adjust the brightness and contrast to obtain an image Use the RAPID SCAN switch (separate control panel) To scan rapidly a small area: better for focussing, etc. To scan slowly a large area: best for searching objects To use the BEI detector Press SEI (main console) Flip the SEI/BEI switch (separate control panel) to BEI Use the BRIGHTNESS and CONTRAST button on that panel

13 Using the 4 th lens in 5 th lens modes In any mode, observation can be done using either the 4 th of the 5 th lens The focus might be wrong as settings are independent for both lens The exposure will be done with the lens designated by the EOS mode The is a displacement between images observed by the two lenses 5 th 4 th x,y x+~140µm, y-~55µm An object placed at this position as viewed with the 4 th lens will be at center when viewed with the 5 th lens

14 Moving the stage (1) To move the stage Select a direction Note that the referential is inverted X and Y directions can be selected by pressing simultaniously Press an increment button ONE µm, 10µm, 100µm, 1mm and 5mm The stage moves and a the position is indicated on the upper console Use the CONT button to move the stage constinously in the selected direction Ex. to move to exchange position, select X and Y and press CONT Select a direction Position and direction indicator on the upper console Increments Continuous motion in the chosen direction

15 Setting focus Focussing using routines in CALIB Start INIMSH to move to the mesh marks, locate one of them and establish stage reference system If the mark is not found you will have to adjust the mark detection parameters (see Mark detection adjustment) Start the FOCUS routine to focus on the mark present in the center of the field The computer automatically select the EOS mode lens To focus manually Select the desired lens 4TH or 5TH (focus settings are specific to each lens) Press SEI to enter observation mode and set proper brightness and contrast Use the COARSE and FINE button to set the focus Look for sharp features and small details. Start at small magnification and increase it gradually looking at smaller details

16 Centering the aperture 1 st : unselect all The aperture is a small hole that has to be centered in the column using To do so, you enter the WOB mode (wobbling) Unselect all lenses and coils Press WOB Select the desired lens The focus will now change in an oscillatory manner (wobble) resulting in a displacement of the image if the aperture is not centered 2 nd : WOB Center the aperture using the 4 arrow buttons of the stepping motor remote control To turn off wobbling mode Unselect the lens 3 rd : select the proper lens Press CLR on the right of WOB

17 Correcting astigmatism The astigmatism of the beam spot is corrected using coils To correct the astigmatism Enter observation mode et locate small objects with sharp angles or details Press ASTIG Note the values indicates and the value of the focus Use the X and Y buttons to correct any astigmatism You can also slightly change the focus around optimal value while correcting It is also possible to use the wobbling mode while correcting astigmatism Unselect all lenses and coils Press WOB Select 4TH or 5TH Select ASTIG Note the values indicates Use the X and Y buttons to correct any astigmatism The focus is now wobbled automatically

18 Loading the cassette onto the exposure stage Verify that the load lock is adequately pumped Bring the stage to the loading position The STAGE INITIAL LED will light up and the stage arrows (upper console) will flash Bring the elevator to the proper cassette Press STEP UP as needed to have a red LED facing the cassette to be loaded Press OPEN to open the gate valve Press TRANS to transfer the cassette onto the stage Press OPEN to close the gate valve The unloading procedure is similar To bring the elevator to initial position (ex. for loading cassette) Line up the elevator with the desired cassette Indicates that the stage is in exchange position and can be loaded Indicates that a cassette is on stage To open and close the gate valve To transfer the cassette from/to the stage To bring the elevator to the proper cassette

19 Using the CALIB program (1) The program has two modes toggled by PF4 In TEMPORARY mode the routines are executed In PARAMETER mode the parameter used by the routines can be modified To execute a routine, Select it with the arrow keys and press ENTER Answer N to the following question To modify the parameters for one run only Select it with the arrow keys and press ENTER Enter Y and the desired parameters. Press PF1 (EXIT) to store and execute the routine Press PF2 (QUIT) to quit without modifying the parameters and executing the routine If you have selected a wrong routine, there is no direct way to cancel Answer Y to the Modify parameters Then press PF2 (QUIT) to exit without executing the routine To leave the CALIB program without saving any modification to the table press PF2 (QUIT) To leave the CALIB program and save the modified tables Select the SAVE routine, then GO, then Y and confirm with Y at the end Press PF1 (EXIT) and then Y to save the table The SAVE routine and the save option when exiting do not operate the same way. Only the SAVE routine seems to save the whole table

20 Moving the stage (2) The stage can be moved using two routines in CALIB Start STGCOM Enter absolute motions by typing MV [x value],[y value] Enter relative displacements by typing RS [x value],[y value] For other commands see CALIB manual (use HE for help) Exit the routine to regain control of the console STGMOV Offers a selection of predefined positions on the stage 0 : EXIT 1 : MESH MARK 2 : WIRE MARK 3 : FARADAY CUP 4 : SEM SPECIMEN 5 : KNIFE EDGE 6 : RESOLUTION SPECIMEN 7 : AUXILIARY-1 8 : AUXILIARY-2 9 : CASSETTE EXCHANGE 10 : BEAM CATCHING HOLE Using this routine can lock up the console in remote mode Start the LOCAL routine in CALIB to regain control of the JEOL

21 Getting closer to exposure time You now have Aligned the column and set the current to the desired value Set a proper focus on the stage or sample Located the sample and noted the position for the exposure You should now Verify that the mesh mark detection works (INIMSH) If not, set the mark detection parameters Verify that the rotation correction works (PDEF) Correct for distortion (DISTOR) and save the results If alignment to a previous pattern is required Set the mark size, type and positions in SETWFR Execute SETWFR to verify that the marks are properly detected If not, adjust the mark detection parameters and enter them in You should now SAVE the tables and exit CALIB Validate and compile (CMPL) the JOBDECK file (.JDF) Validate and schedule (SCHD) the schedule file (.SDF) Expose the magazin file

22 Mark detection adjustment (1) If a mark detection does not work, the parameter table has to be adjusted Ex. you use a different current as the usual one Ex. the structure/substrate contrast is low Position yourself on the desired mark Press SEI, then MD, then ADD and use normal (wide) scan Set the MAG and SPD to Adjust ATTEN to get peeks of equal height Adjust LEVEL to get a proper logical waveform Note the values and report them in the CALIB settings ADD ATTEN LEVEL MAG Coarse Fine SPD Coarse Fine 4 th th SEI MD

23 Mark detection adjustment (1) The SETMDG routine The mark detection settings used in routines INIMSH, PDEF, DISTOR, FOCUS and others are set independently The routine SETMDG can be used to set all these parameters at once Start SETMDG Enter the parameters suitable for INIMSH The routine will then use the same values for other routines Follow the same logic for parameters for wafer mark detection This set the parameters for wafer mark detection and chip mark detection REMARK: the settings are not saved properly by using only the PF1 to exit CALIB Use the SAVE routine Then PF1 and save again

24 Coordinate systems Stage coordinates The uses a left handed Cartesian system The center of the cassette is at coordinates 87000,60000 (100,100) Exchange position Mesh marks Cassette +Y +X The exposure is done using an other coordinate system It is a right handed Cartesian system In this system, the center of the cassette is at (0,0) Use program OFFSET to convert between coordinates in the stage referential (as read on the console) and the exposure coordinates (as entered in the OFFSET command of JDF or SDF files) (87000,60000) Faraday cup

Standard Operating Procedure for the Amray 1810 Scanning Electron Microscope Version: 29 NOVEMBER 2014

Standard Operating Procedure for the Amray 1810 Scanning Electron Microscope Version: 29 NOVEMBER 2014 Standard Operating Procedure for the Amray 1810 Scanning Electron Microscope Version: 29 NOVEMBER 2014 1. Utility Requirements a. System power is supplied by two 120 VAC/20 A circuits. When doing maintenance

More information

OPERATION OF THE HITACHI S-450 SCANNING ELECTRON MICROSCOPE. by Doug Bray Department of Biological Sciences University of Lethbridge

OPERATION OF THE HITACHI S-450 SCANNING ELECTRON MICROSCOPE. by Doug Bray Department of Biological Sciences University of Lethbridge OPERATION OF THE HITACHI S-450 SCANNING ELECTRON MICROSCOPE by Doug Bray Department of Biological Sciences University of Lethbridge Revised September, 2000 Note: The terms in bold in this document represent

More information

JEOL 6500 User Manual

JEOL 6500 User Manual LOG IN to your session on the computer to the left of the microscope. Starting Conditions 1. Press Ctrl-Alt-Del and log on to the microscope computer. Click on JEOL PC SEM 6500 icon. Click yes if message

More information

JEOL 6700 User Manual 05/18/2009

JEOL 6700 User Manual 05/18/2009 JEOL 6700 User Manual 05/18/2009 LOG IN to your session on the computer to the right of the microscope. Starting Conditions 1. Click the button and read the Penning Gauge to ensure that the microscope

More information

JSM 6060 LV SCANNING ELECTRON MICROSCOPE STANDARD OPERATING PROCEDURES

JSM 6060 LV SCANNING ELECTRON MICROSCOPE STANDARD OPERATING PROCEDURES JSM 6060 LV SCANNING ELECTRON MICROSCOPE STANDARD OPERATING PROCEDURES RULES All users must go through a series of standard operation procedure training. For more information contact: Longlong Liao Teaching

More information

Title: Amray 1830 SEM#2 Semiconductor & Microsystems Fabrication Laboratory Revision: D Rev Date: 03/18/2016

Title: Amray 1830 SEM#2 Semiconductor & Microsystems Fabrication Laboratory Revision: D Rev Date: 03/18/2016 Approved by: Process Engineer / / / / Equipment Engineer 1 SCOPE The purpose of this document is to detail the use of the Amray 1830 SEM. All users are expected to have read and understood this document.

More information

Basic Operating Instructions for Strata Dual Beam 235 FIB/SEM

Basic Operating Instructions for Strata Dual Beam 235 FIB/SEM Basic Operating Instructions for Strata Dual Beam 235 FIB/SEM Warning Always adjust your specimen height before closing the chamber door to make sure your specimen will not hit the bottom of the lens;

More information

FE-SEM SU-8020 Operating manual (Preliminary version)

FE-SEM SU-8020 Operating manual (Preliminary version) FE-SEM SU-8020 Operating manual (Preliminary version) 2016/04/11 Seimitsu Bunseki sitsu lab. Starting up 1.Turn on the Display switch. Windows OS is starting up 2. Select the user SU-8000. 3. Click the

More information

Model SU3500 Scanning Electron Microscope

Model SU3500 Scanning Electron Microscope Model SU3500 Scanning Electron Microscope Modified and Parts taken from Hitachi Easy Operation Guide. Before using the Model SU3500 SEM, be sure to read the [GENERAL SAFETY GUIDELINES] in the instruction

More information

SEM OPERATION IN LOW VACUUM MODE

SEM OPERATION IN LOW VACUUM MODE SEM OPERATION IN LOW VACUUM MODE Instructions for JEOL 5800 LV The EVAC light of the SEM specimen chamber should be already lit when you approach the SEM & the SEM will have been left in the high vacuum

More information

Operating Checklist for using the Scanning Electron. Microscope, JEOL JSM 6400.

Operating Checklist for using the Scanning Electron. Microscope, JEOL JSM 6400. Smith College August 2009 Operating Checklist for using the Scanning Electron Microscope, JEOL JSM 6400. CONTENT, page no. Pre-Check 1 Startup 1 Specimen Insertion 2 Filament Saturation 2 Beam Alignment

More information

RAITH e-line OPERATING INSTRUCTIONS

RAITH e-line OPERATING INSTRUCTIONS RAITH e-line OPERATING INSTRUCTIONS 1) LOADING A SAMPLE a. Start the system i. On the Column PC (Right side monitor [R]), select the SmartSEM icon to on the desktop to begin the column software. ii. On

More information

1. Preliminary sample preparation

1. Preliminary sample preparation FEI Helios NanoLab 600 standard operating procedure Nicholas G. Rudawski ngr@ufl.edu (352) 392 3077 (office) (805) 252-4916 (cell) Last updated: 03/02/18 What this document provides: an overview of basic

More information

Dickinson College Department of Geology

Dickinson College Department of Geology Dickinson College Department of Geology Title: Equipment: BASIC OPERATION OF THE SCANNING ELECTRON MICROSCOPE (SEM) JEOL JSM-5900 SCANNING ELECTRON MICROSCOPE Revision: 2.2 Effective Date: 1/29/2003 Author(s):

More information

Operating the Hitachi 7100 Transmission Electron Microscope Electron Microscopy Core, University of Utah

Operating the Hitachi 7100 Transmission Electron Microscope Electron Microscopy Core, University of Utah Operating the Hitachi 7100 Transmission Electron Microscope Electron Microscopy Core, University of Utah Follow the procedures below when you use the Hitachi 7100 TEM. Starting Session 1. Turn on the cold

More information

1.3. Before loading the holder into the TEM, make sure the X tilt is set to zero and the goniometer locked in place (this will make loading easier).

1.3. Before loading the holder into the TEM, make sure the X tilt is set to zero and the goniometer locked in place (this will make loading easier). JEOL 200CX operating procedure Nicholas G. Rudawski ngr@ufl.edu (805) 252-4916 1. Specimen loading 1.1. Unlock the TUMI system. 1.2. Load specimen(s) into the holder. If using the double tilt holder, ensure

More information

Scanning electron microscope

Scanning electron microscope Scanning electron microscope 6 th CEMM workshop Maja Koblar, Sc. Eng. Physics Outline The basic principle? What is an electron? Parts of the SEM Electron gun Electromagnetic lenses Apertures Chamber and

More information

Operating Checklist for using the Scanning Electron Microscope, JEOL JSM 6400.

Operating Checklist for using the Scanning Electron Microscope, JEOL JSM 6400. Smith College August 2005 Operating Checklist for using the Scanning Electron Microscope, JEOL JSM 6400. CONTENT, page no. Pre-Check, 1 Specimen Insertion, 1 Startup, 2 Filament Saturation, 2 Beam Alignment,

More information

Please follow these instructions for use of the Philips CM100 TEM. Adopted from website below.

Please follow these instructions for use of the Philips CM100 TEM. Adopted from website below. Please follow these instructions for use of the Philips CM100 TEM. Adopted from website below. http://staff.washington.edu/wpchan/if/cm100_inst.shtml Instructions for the Philips CM100 TEM and peripherals

More information

Figure 1 The Raith 150 TWO

Figure 1 The Raith 150 TWO RAITH 150 TWO SOP Figure 1 The Raith 150 TWO LOCATION: Raith 150 TWO room, Lithography area, NanoFab PRIMARY TRAINER: SECONDARY TRAINER: 1. OVERVIEW The Raith 150 TWO is an ultra high resolution, low voltage

More information

Scanning Electron Microscope FEI INSPECT F50. Step by step operation manual

Scanning Electron Microscope FEI INSPECT F50. Step by step operation manual Scanning Electron Microscope FEI INSPECT F50 Step by step operation manual Scanning Electron Microscope, FEI Inspect F50 FE-SEM-F Observation Flow Saving Data And Analysis Specimen preparation Error check

More information

This document assumes the user is already familiar with basic operation of the instrument in TEM mode and use of the Microscope Control interface.

This document assumes the user is already familiar with basic operation of the instrument in TEM mode and use of the Microscope Control interface. FEI Tecnai F20 S/TEM: imaging in STEM mode Nicholas G. Rudawski ngr@ufl.edu (805) 252-4916 (352) 392-3077 Last updated: 05/10/18 This document assumes the user is already familiar with basic operation

More information

Basic Users Manual for Tecnai-F20 TEM

Basic Users Manual for Tecnai-F20 TEM Basic Users Manual for Tecnai-F20 TEM NB: This document contains my personal notes on the operating procedure of the Tecnai F20 and may be used as a rough guide for those new to the microscope. It may

More information

Check that the pneumatic hose is disconnected!!!! (unless your using the BSE detector, of course)

Check that the pneumatic hose is disconnected!!!! (unless your using the BSE detector, of course) JEOL 7000F BASIC OPERATING INSTRUCTIONS-Ver.-2.0 Note: This is minimal operation checklist and does not replace the other reference manuals. Read the manual for Specimen Exchange (JEOL 7000 Specimen Exchange

More information

2 How to operate the microscope/obtain an image

2 How to operate the microscope/obtain an image Morgagni Operating Instructions 50079 010912 2-1 2 ow to operate the microscope/obtain an image 2.1 Starting the microscope 2.1.1 Starting the microscope with several manually-operated steps 1. Turn on

More information

MSE 460 TEM Lab 2: Basic Alignment and Operation of Microscope

MSE 460 TEM Lab 2: Basic Alignment and Operation of Microscope MSE 460 TEM Lab 2: Basic Alignment and Operation of Microscope Last updated on 1/8/2018 Jinsong Wu, jinsong-wu@northwestern.edu Aims: The aim of this lab is to familiarize you with basic TEM alignment

More information

FEI Tecnai G 2 F20 Operating Procedures

FEI Tecnai G 2 F20 Operating Procedures FEI Tecnai G 2 F20 Operating Procedures 1. Startup (1) Sign-up in the microscope log-sheet. Please ensure you have written an account number for billing. (2) Log in to the computer: Login to your account

More information

SEM Training Notebook

SEM Training Notebook SEM Training Notebook Lab Manager: Dr. Perry Cheung MSE Fee-For-Service Facility Materials Science and Engineering University of California, Riverside March 8, 2018 (rev. 3.5) 1 Before you begin Complete

More information

User Operation of JEOL 1200 EX II

User Operation of JEOL 1200 EX II **Log onto Computer** Open item program Start Up Procedure User Operation of JEOL 1200 EX II 1. If scope is not running, locate an electron microscopy technician (EMT) to find out why not. 2. Turn up brightness

More information

LEO 912 TEM Short Manual. Prepared/copyrighted by RH Berg Danforth Plant Science Center

LEO 912 TEM Short Manual. Prepared/copyrighted by RH Berg Danforth Plant Science Center LEO 912 TEM Short Manual Prepared/copyrighted by RH Berg Danforth Plant Science Center Specimen holder [1] Never touch the holder (outside of the O-ring, double-headed arrow) because finger oils will contaminate

More information

1.1. Log on to the TUMI system (you cannot proceed further until this is done).

1.1. Log on to the TUMI system (you cannot proceed further until this is done). FEI DB235 SEM mode operation Nicholas G. Rudawski ngr@ufl.edu (805) 252-4916 1. Sample loading 1.1. Log on to the TUMI system (you cannot proceed further until this is done). 1.2. The FIB software (xp)

More information

Nanoscale Fabrication & Characterization Facility. Raith e-line EBL Users Guide (updated:aug 2 nd, 2017)

Nanoscale Fabrication & Characterization Facility. Raith e-line EBL Users Guide (updated:aug 2 nd, 2017) Nanoscale Fabrication & Characterization Facility Raith e-line EBL Users Guide (updated:aug 2 nd, 2017) Overview: The Raith e-line EBL system is designed to write features with critical dimensions as small

More information

SEM Training Notebook

SEM Training Notebook SEM Training Notebook Lab Manager: Dr. Perry Cheung MSE Fee-For-Service Facility Materials Science and Engineering University of California, Riverside December 21, 2017 (rev. 3.4) 1 Before you begin Complete

More information

Scanning electron microscope

Scanning electron microscope Scanning electron microscope 5 th CEMM workshop Maja Koblar, Sc. Eng. Physics Outline The basic principle? What is an electron? Parts of the SEM Electron gun Electromagnetic lenses Apertures Detectors

More information

2. Raise HT to 200kVby following the procedure explained in 1.6.

2. Raise HT to 200kVby following the procedure explained in 1.6. JEOL 2100 MANUAL Quick check list 1. If needed, fill the reservoir with LN2 2. Raise HT to 200kVby following the procedure explained in 1.6. 3. Insert specimen holder into TEM (Insert holder in airlock,

More information

Section 1: TEM parts and functions... 2

Section 1: TEM parts and functions... 2 Introduction The set of instructions below are written by Charlie Sanabria within the first few sessions of his TEM training process, and are intended for anyone interested in viewing the TEM operation

More information

STEM alignment procedures

STEM alignment procedures STEM alignment procedures Step 1. ASID alignment mode 1. Write down STD for TEM, and then open the ASID control window from dialogue. Also, start Simple imager viewer program on the Desktop. 2. Click on

More information

Using the Hitachi 3400-N VP-SEM

Using the Hitachi 3400-N VP-SEM Using the Hitachi 3400-N VP-SEM Opening the Chamber to Load Specimens (This may also be done later using the software) 1. Click the AIR button on the front of the machine: 2. Wait a few minutes until you

More information

1.2. Make sure the viewing screen is covered (exposure to liquid N 2 may cause it to crack).

1.2. Make sure the viewing screen is covered (exposure to liquid N 2 may cause it to crack). FEI Tecnai F20 S/TEM: imaging in TEM mode Nicholas G. Rudawski ngr@ufl.edu (805) 252-4916 (352) 392-3077 Last updated: 01/21/18 1. Filling the cold trap (if needed) 1.1. Prior to use, the cold trap needs

More information

FEI Falcon Direct Electron Detector. Best Practice Document

FEI Falcon Direct Electron Detector. Best Practice Document FEI Falcon Direct Electron Detector Best Practice Document 2 1. Introduction FEI Falcon Direct Electron Detector Best Practice Application Guide The FEI Falcon Detector is based on direct electron detection

More information

Jeol JEM Responsible personell: Endy ( ) Online booking is compulsory!

Jeol JEM Responsible personell: Endy ( ) Online booking is compulsory! Jeol JEM 1230 Responsible personell: Endy (45279377) Online booking is compulsory! After training you will have access to working alone on the instrument. All insertion of samples is done by responsible

More information

R PROFLAME Instruction Book Collection

R PROFLAME Instruction Book Collection 9.956.028 R00 584 PROFLAME Instruction Book Collection 4-17 18-29 584 PROFLAME System 30-39 Appendix: DIP SWITCH NUMBER (0=ON 1=OFF) 40-41 4-17 Fig. 1 The SIT is a device that allows, in conjunction with

More information

Transmission Electron Microscopy 9. The Instrument. Outline

Transmission Electron Microscopy 9. The Instrument. Outline Transmission Electron Microscopy 9. The Instrument EMA 6518 Spring 2009 02/25/09 Outline The Illumination System The Objective Lens and Stage Forming Diffraction Patterns and Images Alignment and Stigmation

More information

SAMUEL ROBERTS NOBLE ELECTRON MICROSCOPY LABORATORY. Operating Procedures for the Zeiss 9 S-2. Transmission Electron Microscope

SAMUEL ROBERTS NOBLE ELECTRON MICROSCOPY LABORATORY. Operating Procedures for the Zeiss 9 S-2. Transmission Electron Microscope 1 SAMUEL ROBERTS NOBLE ELECTRON MICROSCOPY LABORATORY Operating Procedures for the Zeiss 9 S-2 Transmission Electron Microscope Prepared by Dr. Scott D. Russell Department of Botany and Microbiology September,

More information

MSE 595T Transmission Electron Microscopy. Laboratory III TEM Imaging - I

MSE 595T Transmission Electron Microscopy. Laboratory III TEM Imaging - I MSE 595T Basic Transmission Electron Microscopy TEM Imaging - I Purpose The purpose of this lab is to: 1. Make fine adjustments to the microscope alignment 2. Obtain a diffraction pattern 3. Obtain an

More information

Amray 3600 FESEM. Standard Operating Procedure. v2.2 modified by Bryan Cord. General Notes...3. Sample Loading...5. System Loading...

Amray 3600 FESEM. Standard Operating Procedure. v2.2 modified by Bryan Cord. General Notes...3. Sample Loading...5. System Loading... Amray 3600 FESEM Standard Operating Procedure v2.2 modified 5.13.13 by Bryan Cord Contents General Notes...3 Sample Loading...5 System Loading...8 Imaging...12 Saving Data...16 System Unloading...18 Troubleshooting...20

More information

SOP for Hitachi S-2150 Scanning Electron Microscope For review purposes only

SOP for Hitachi S-2150 Scanning Electron Microscope For review purposes only SOP for Hitachi S-2150 Scanning Electron Microscope For review purposes only Version 1.0 Prepared by D. Turnbull February 21, 2007. Please submit any omissions to the Author Note: This SEM is a recent

More information

DISCO DICING SAW SOP. April 2014 INTRODUCTION

DISCO DICING SAW SOP. April 2014 INTRODUCTION DISCO DICING SAW SOP April 2014 INTRODUCTION The DISCO Dicing saw is an essential piece of equipment that allows cleanroom users to divide up their processed wafers into individual chips. The dicing saw

More information

STRUCTURE OF THE MICROSCOPE

STRUCTURE OF THE MICROSCOPE STRUCTURE OF THE MICROSCOPE Use the word list to label the microscope below: Light Source Coarse adjustment knob Diaphragm Stage Clips Objectives Fine Adjustment Knob Base Stage Stage Clips Arm Revolving

More information

ZEISS EVO SOP. May 2017 ELECTRON OPTICS

ZEISS EVO SOP. May 2017 ELECTRON OPTICS ZEISS EVO SOP May 2017 ELECTRON OPTICS The patented EVO column is the area of the SEM, where electrons are emitted, accelerated, deflected, focused, and scanned. Main characteristics of the EVO optics

More information

University of MN, Minnesota Nano Center Standard Operating Procedure

University of MN, Minnesota Nano Center Standard Operating Procedure Equipment Name: Atomic Force Microscope Badger name: afm DI5000 PAN Revisionist Paul Kimani Model: Dimension 5000 Date: October 6, 2017 Location: Bay 1 PAN Revision: 1 A. Description i. Enhanced Motorized

More information

This document assumes the user is already familiar with basic operation of the instrument in TEM mode and use of the digital camera.

This document assumes the user is already familiar with basic operation of the instrument in TEM mode and use of the digital camera. FEI Tecnai F20 S/TEM: acquiring diffraction patterns Nicholas G. Rudawski ngr@ufl.edu (805) 252-4916 (352) 392-3077 Last updated: 10/18/17 This document assumes the user is already familiar with basic

More information

Supplementary Materials

Supplementary Materials Supplementary Materials In the supplementary materials of this paper we discuss some practical consideration for alignment of optical components to help unexperienced users to achieve a high performance

More information

Heidelberg µpg 101 Laser Writer

Heidelberg µpg 101 Laser Writer Heidelberg µpg 101 Laser Writer Standard Operating Procedure Revision: 3.0 Last Updated: Aug.1/2012, Revised by Nathanael Sieb Overview This document will provide a detailed operation procedure of the

More information

1.1. In regular TEM imaging mode, find a region of interest and set it at eucentric height.

1.1. In regular TEM imaging mode, find a region of interest and set it at eucentric height. JEOL 2010F operating procedure Covers operation in STEM mode (See separate procedures for operation in TEM mode and operation of EDS system) Nicholas G. Rudawski ngr@ufl.edu (805) 252-4916 NOTE: this operating

More information

Using the Microscope for a NANSLO Remote Web-based Science Lab Activity

Using the Microscope for a NANSLO Remote Web-based Science Lab Activity Using the Microscope for a NANSLO Remote Web-based Science Lab Activity MICROSCOPE RWSL LAB INTERFACE INSTRUCTIONS The Remote Web-based Science Lab (RWSL) microscope is a high quality digital microscope

More information

General information. If you see the instrument turned off, notify MIC personnel. MIC personnel will help you insert your samples into the instrument.

General information. If you see the instrument turned off, notify MIC personnel. MIC personnel will help you insert your samples into the instrument. JEOL JSM-7400F Table of contents General information.. 3 The operation panel. 4 The different sample holders and inserting the samples.. 5 Turning on the beam... 6 Stage map control... 8 Correcting astigmatism...

More information

Introduction of New Products

Introduction of New Products Field Emission Electron Microscope JEM-3100F For evaluation of materials in the fields of nanoscience and nanomaterials science, TEM is required to provide resolution and analytical capabilities that can

More information

Nikon AZ100. Laser Scanning Macro Confocal Microscope. Jordan Briscoe Adam Fries Kyle Marchuk Kaitlin Corbin. May 2017.

Nikon AZ100. Laser Scanning Macro Confocal Microscope. Jordan Briscoe Adam Fries Kyle Marchuk Kaitlin Corbin. May 2017. Nikon AZ100 Laser Scanning Macro Confocal Microscope Jordan Briscoe Adam Fries Kyle Marchuk Kaitlin Corbin May 2017 Contents 1 Introduction 2 2 Hardware - Startup 2 3 Software/Operation 4 3.1 Multidimensional

More information

Nikon Ti-E Microscope Manual. Rightmire Hall Ohio State University. Director: Tony Brown Rightmire

Nikon Ti-E Microscope Manual. Rightmire Hall Ohio State University. Director: Tony Brown Rightmire Nikon Ti-E Microscope Manual Rightmire Hall Ohio State University Director: Tony Brown Rightmire 060 292-1205 brown.2302@osu.edu Facility Manager: Paula Monsma Rightmire 062 293-0939 292-1367 monsma.1@osu.edu

More information

Procedures for Performing Cryoelectron Microscopy on the FEI Sphera Microscope

Procedures for Performing Cryoelectron Microscopy on the FEI Sphera Microscope Procedures for Performing Cryoelectron Microscopy on the FEI Sphera Microscope The procedures given below were written specifically for the FEI Tecnai G 2 Sphera microscope. Modifications will need to

More information

How to use the Jeol 1010 TEM of GI (Liesbeth own GI version)

How to use the Jeol 1010 TEM of GI (Liesbeth own GI version) How to use the Jeol 1010 TEM of GI (Liesbeth own GI version) 1.Load the specimen Load a grid into the rod holder: USE ONLY THE TOP POSITION (blue arrow), Specimen selection on 1 (The rear one is only a

More information

Zeiss LSM 780 Protocol

Zeiss LSM 780 Protocol Zeiss LSM 780 Protocol 1) System Startup F Please note the sign-up policy. You must inform the facility at least 24 hours beforehand if you can t come; otherwise, you will receive a charge for unused time.

More information

Instructions for Tecnai a brief start up manual

Instructions for Tecnai a brief start up manual Instructions for Tecnai a brief start up manual Version 3.0, 8.12.2015 Manual of Tecnai 12 transmission electron microscope located at Aalto University's Nanomicroscopy Center. More information of Nanomicroscopy

More information

User Manual. Digital Compound Binocular LED Microscope. MicroscopeNet.com

User Manual. Digital Compound Binocular LED Microscope. MicroscopeNet.com User Manual Digital Compound Binocular LED Microscope Model MD82ES10 MicroscopeNet.com Table of Contents i. Caution... 1 ii. Care and Maintenance... 2 1. Components Illustration... 3 2. Installation...

More information

PHYS 3153 Methods of Experimental Physics II O2. Applications of Interferometry

PHYS 3153 Methods of Experimental Physics II O2. Applications of Interferometry Purpose PHYS 3153 Methods of Experimental Physics II O2. Applications of Interferometry In this experiment, you will study the principles and applications of interferometry. Equipment and components PASCO

More information

Operating F20/F30 with SerialEM

Operating F20/F30 with SerialEM Chen Xu xuchen@brandeis.ede $BrandeisEM: ~emdoc-xml/en_us.iso8859-1/articles/operating-f20-or-f30/article.xml, 1 2013-01-19 01:42:20 xuchen Exp$ This is a quick check list for the Tecnai F20 or Tecnai

More information

AF Area Mode. Face Priority

AF Area Mode. Face Priority Chapter 4: The Shooting Menu 71 AF Area Mode This next option on the second screen of the Shooting menu gives you several options for controlling how the autofocus frame is set up when the camera is in

More information

Functions of the SEM subsystems

Functions of the SEM subsystems Functions of the SEM subsystems Electronic column It consists of an electron gun and two or more electron lenses, which influence the path of electrons traveling down an evacuated tube. The base of the

More information

STEM Spectrum Imaging Tutorial

STEM Spectrum Imaging Tutorial STEM Spectrum Imaging Tutorial Gatan, Inc. 5933 Coronado Lane, Pleasanton, CA 94588 Tel: (925) 463-0200 Fax: (925) 463-0204 April 2001 Contents 1 Introduction 1.1 What is Spectrum Imaging? 2 Hardware 3

More information

Contents STARTUP MICROSCOPE CONTROLS CAMERA CONTROLS SOFTWARE CONTROLS EXPOSURE AND CONTRAST MONOCHROME IMAGE HANDLING

Contents STARTUP MICROSCOPE CONTROLS CAMERA CONTROLS SOFTWARE CONTROLS EXPOSURE AND CONTRAST MONOCHROME IMAGE HANDLING Operations Guide Contents STARTUP MICROSCOPE CONTROLS CAMERA CONTROLS SOFTWARE CONTROLS EXPOSURE AND CONTRAST MONOCHROME IMAGE HANDLING Nikon Eclipse 90i Operations Guide STARTUP Startup Powering Up Fluorescence

More information

P20 Zoom Flash Zoom Flash P20 P20 Zoomblitz Flash con zoom P20 Flash externo P20 Flash Zoom P20

P20 Zoom Flash Zoom Flash P20 P20 Zoomblitz Flash con zoom P20 Flash externo P20 Flash Zoom P20 P20 Zoom Flash Zoom Flash P20 P20 Zoomblitz Flash con zoom P20 Flash externo P20 Flash Zoom P20 User s Guide Guide d'utilisation Benutzerhandbuch Manuale per l'utente Guía del usuario Guia do usuário 4J6021

More information

IRIS \ IRIS-I QUICK SET-UP GUIDE STEP 1 INSTALL

IRIS \ IRIS-I QUICK SET-UP GUIDE STEP 1 INSTALL IRIS \ IRIS-I QUICK SET-UP GUIDE STEP 1 INSTALL Confirm contents of package: 1 sensor, 1 cable, 1 wide lens (default), 1 narrow lens, mounting template, User s Guide. Install the sensor at the desired

More information

LASER ENHANCED REVOLVER GRIP OWNER S MANUAL RED LASER GREEN LASER

LASER ENHANCED REVOLVER GRIP OWNER S MANUAL RED LASER GREEN LASER LASER ENHANCED RED LASER GREEN LASER REVOLVER GRIP OWNER S MANUAL LASER ENHANCED GRIP Installation Instructions Caution... 3 Safety Labels... 4 Installation...5-7 Programming...8-10 Batteries (Red Laser)...

More information

RENISHAW INVIA RAMAN SPECTROMETER

RENISHAW INVIA RAMAN SPECTROMETER STANDARD OPERATING PROCEDURE: RENISHAW INVIA RAMAN SPECTROMETER Purpose of this Instrument: The Renishaw invia Raman Spectrometer is an instrument used to analyze the Raman scattered light from samples

More information

Standard Operating Procedure

Standard Operating Procedure Standard Operating Procedure Nanosurf Atomic Force Microscopy Operation Facility NCCRD Nanotechnology Center for Collaborative Research and Development Department of Chemistry and Engineering Physics The

More information

MICROSCOPY MICROSCOPE TERMINOLOGY

MICROSCOPY MICROSCOPE TERMINOLOGY 1 MICROSCOPY Most of the microorganisms that we talk about in this class are too small to be seen with the naked eye. The instruments we will use to visualize these microbes are microscopes. The laboratory

More information

Exercise 6. Range and Angle Tracking Performance (Radar-Dependent Errors) EXERCISE OBJECTIVE

Exercise 6. Range and Angle Tracking Performance (Radar-Dependent Errors) EXERCISE OBJECTIVE Exercise 6 Range and Angle Tracking Performance EXERCISE OBJECTIVE When you have completed this exercise, you will be familiar with the radardependent sources of error which limit range and angle tracking

More information

Guide to Confocal 5. Starting session

Guide to Confocal 5. Starting session Guide to Confocal 5 Remember that when booking and before starting session you can check for any problems at https://www.bris.ac.uk/biochemistry/uobonly/cif/index.html Starting session Switch on microscope

More information

HydroFORM Phased Array Corrosion Mapping System Part 2 Mechanical Setup

HydroFORM Phased Array Corrosion Mapping System Part 2 Mechanical Setup HydroFORM Phased Array Corrosion Mapping System Part 2 Mechanical Setup HydroFORM Corrosion Mapping System Mechanical Setup The HydroFORM Phased Array System is designed to perform corrosion mapping in

More information

Full-screen mode Popup controls. Overview of the microscope user interface, TEM User Interface and TIA on the left and EDS on the right

Full-screen mode Popup controls. Overview of the microscope user interface, TEM User Interface and TIA on the left and EDS on the right Quick Guide to Operating FEI Titan Themis G2 200 (S)TEM: TEM mode Susheng Tan Nanoscale Fabrication and Characterization Facility, University of Pittsburgh Office: M104/B01 Benedum Hall, 412-383-5978,

More information

Filter & Spectrometer Electron Optics

Filter & Spectrometer Electron Optics Filter & Spectrometer Electron Optics Parameters Affecting Practical Performance Daniel Moonen & Harold A. Brink Did Something Go Wrong? 30 20 10 0 500 600 700 800 900 1000 1100 ev 1 Content The Prism

More information

ScanArray Overview. Principle of Operation. Instrument Components

ScanArray Overview. Principle of Operation. Instrument Components ScanArray Overview The GSI Lumonics ScanArrayÒ Microarray Analysis System is a scanning laser confocal fluorescence microscope that is used to determine the fluorescence intensity of a two-dimensional

More information

Physics Requirements for the CXI 0.1 micron Sample Chamber

Physics Requirements for the CXI 0.1 micron Sample Chamber PHYSICS REQUIREMENT DOCUMENT (PRD) Doc. No. SP-391-000-20 R1 LUSI SUB-SYSTEM Coherent X-Ray Imaging Physics Requirements for the Sébastien Boutet CXI Scientist, Author Signature Date Paul Montanez CXI

More information

JEOL 5800LV SCANNING ELECTRON MICROSCOPE OPERATOR'S MANUAL

JEOL 5800LV SCANNING ELECTRON MICROSCOPE OPERATOR'S MANUAL JEOL 5800LV SCANNING ELECTRON MICROSCOPE OPERATOR'S MANUAL Michael N. Spilde Christopher Adcock Original version: October 1999 Revised: January 2006 Department of Earth and Planetary Sciences and Institute

More information

CAPTURING IMAGES ON THE HIGH-MAGNIFICATION MICROSCOPE

CAPTURING IMAGES ON THE HIGH-MAGNIFICATION MICROSCOPE University of Virginia ITC Academic Computing Health Sciences CAPTURING IMAGES ON THE HIGH-MAGNIFICATION MICROSCOPE Introduction The Olympus BH-2 microscope in ACHS s microscope lab has objectives from

More information

EXPRIMENT 3 COUPLING FIBERS TO SEMICONDUCTOR SOURCES

EXPRIMENT 3 COUPLING FIBERS TO SEMICONDUCTOR SOURCES EXPRIMENT 3 COUPLING FIBERS TO SEMICONDUCTOR SOURCES OBJECTIVES In this lab, firstly you will learn to couple semiconductor sources, i.e., lightemitting diodes (LED's), to optical fibers. The coupling

More information

R I T. Title: Wyko RST Plus. Semiconductor & Microsystems Fabrication Laboratory Revision: A Rev Date: 05/23/06 1 SCOPE 2 REFERENCE DOCUMENTS

R I T. Title: Wyko RST Plus. Semiconductor & Microsystems Fabrication Laboratory Revision: A Rev Date: 05/23/06 1 SCOPE 2 REFERENCE DOCUMENTS Approved by: Process Engineer / / / / Equipment Engineer 1 SCOPE The purpose of this document is to detail the use of the Wyko RST Plus. All users are expected to have read and understood this document.

More information

Match the microscope structures given in the left column with the statements in the right column that identify or describe them.

Match the microscope structures given in the left column with the statements in the right column that identify or describe them. 49 Prelab for Name Match the microscope structures given in the left column with the statements in the right column that identify or describe them. Key: a. coarse adjustment knob f. turret or nosepiece

More information

Development of JEM-2800 High Throughput Electron Microscope

Development of JEM-2800 High Throughput Electron Microscope Development of JEM-2800 High Throughput Electron Microscope Mitsuhide Matsushita, Shuji Kawai, Takeshi Iwama, Katsuhiro Tanaka, Toshiko Kuba and Noriaki Endo EM Business Unit, JEOL Ltd. Electron Optics

More information

Build your own. Pack. Stages 19-22: Continue building Robi s left arm

Build your own. Pack. Stages 19-22: Continue building Robi s left arm Build your own Pack 06 Stages 19-22: Continue building Robi s left arm Build your own All rights reserved 2015 Published in the UK by De Agostini UK Ltd, Battersea Studios 2, 82 Silverthorne Road, London

More information

SUSS Mask Aligner. Purpose: To expose photoresist on a wafer using a photomask

SUSS Mask Aligner. Purpose: To expose photoresist on a wafer using a photomask SUSS Mask Aligner Purpose: To expose photoresist on a wafer using a photomask Overview This SOP will go over how to use the machine for basic exposures. This will include commonly used controls and frequently

More information

Zeiss LSM 880 Protocol

Zeiss LSM 880 Protocol Zeiss LSM 880 Protocol 1) System Startup Please note put sign-up policy. You must inform the facility at least 24 hours beforehand if you can t come; otherwise, you will receive a charge for unused time.

More information

Recording EPR Spectra using ER 4102ST Resonator

Recording EPR Spectra using ER 4102ST Resonator Recording EPR Spectra using ER 4102ST Resonator This protocol gives step-by-step instructions for recording an EPR spectrum using the high sensitivity Bruker SHQE cavity (assuming the SHQE is already in

More information

Zeiss AxioImager.Z2 Brightfield Protocol

Zeiss AxioImager.Z2 Brightfield Protocol Zeiss AxioImager.Z2 Brightfield Protocol 1) System Startup Please note put sign-up policy. You must inform the facility at least 24 hours beforehand if you can t come; otherwise, you will receive a charge

More information

Figure 1. Oil-immersion objectives available for use with the Lionheart FX.

Figure 1. Oil-immersion objectives available for use with the Lionheart FX. Tech Note Oil Objective Introduction The Lionheart FX automated imager is compatible with high numerical aperture oil immersion objectives. These objectives offer magnification up to 100X and significantly

More information

Marine Invertebrate Zoology Microscope Introduction

Marine Invertebrate Zoology Microscope Introduction Marine Invertebrate Zoology Microscope Introduction Introduction A laboratory tool that has become almost synonymous with biology is the microscope. As an extension of your eyes, the microscope is one

More information

X-Mag. High-Magnification Workstation User s Guide Revision A

X-Mag. High-Magnification Workstation User s Guide Revision A X-Mag High-Magnification Workstation User s Guide www.fsinspection.com 440831-001 Revision A Freedom Scientific, Inc., 11800 31st Court North, St. Petersburg, Florida 33716-1805, USA www.fsinspection.com

More information

FOLLOWSPOT CONTROLLER

FOLLOWSPOT CONTROLLER SOFTWARE VERSION 2.0 WWW.PRG.COM FOLLOWSPOT CONTROLLER USER MANUAL AutoPar, Bad Boy, Best Boy, Followspot Controller, Mbox, Mini Node, Nocturne, Series 400, ReNEW, Super Node, V476, V676, and Virtuoso

More information

1. Specimen Holder Removal, Loading, and Insertion

1. Specimen Holder Removal, Loading, and Insertion OPERATION OF THE PHILIPS CM-200 FEG-TEM When not in use, the CM-200 should be in the MICROSCOPE ON configuration with the HIGH TENSION ON (illuminates green when the high tension is on).. The microscope

More information