HF-400. Electrosurgical Analyzer for function tests of HF Surgical Equipment

Size: px
Start display at page:

Download "HF-400. Electrosurgical Analyzer for function tests of HF Surgical Equipment"

Transcription

1 User manual HF-400 English Electrosurgical Analyzer for function tests of HF Surgical Equipment Instructions for the user: This manual is supplied in the form of a loose-leaf publication. Any necessary updates and additional pages can thus be included without difficulty later so that you always have the latest version of the manual. We provide you with supplements and/or changes to your manual along with technical modifications that are made to your testing system. The DIN/VDE symbols and drawings shown in this manual have been reproduced with the permission of the Deutsches Institut für Normung e.v. (DIN: German Institute for Standardization) and the Verband Deutscher Elektrotechniker e.v. (VDE: German Association of Electrical Engineers). In applying the relevant standards, it is important to use the versions with the most recent date of issue. We reserve the right to make any changes to the information in this manual without prior notice. Elektronik accepts no responsibility in this regard. The software that comes with this product is supplied within the framework of a License Agreement. The software may be used and copied only in accordance with the terms and conditions of the License Agreement. All rights pertaining to reprinting and reproducing the manual, or parts thereof, are reserved. No part of the manual may be reproduced in any form whatsoever (photocopy, microfilm or any other procedure) or processed, reproduced or distributed using electronic systems without prior written approval from Elektronik. Other products mentioned by name are trademarks or registered trademarks of their respective companies. Elektronik Forstweg 1 D Brieselang GERMANY Tel Fax Edition

2 European Union Declaration of Conformity EG-Konformitätserklärung Nr Hersteller: Produktbeschreibung: Modell: Elektronik, Forstweg 1, Brieselang Bundesrepublik Deutschland HF-Chirurgie Tester zur Funktionsprüfung von HF-Chirurgiegeräten HF-400 Konform mit den Bestimmungen folgender EU-Richtlinien: Richtlinie 2014/35/EU des Europäischen Parlaments und des Rates vom 26. Februar 2014 (Niederspannungsrichtlinie) Richtlinie 2011/65/EU des Europäischen Parlaments und des Rates vom 8. Juni 2011 zur Beschränkung der Verwendung bestimmter gefährlicher Stoffe in Elektro- und Elektronikgeräten (RoHS Richtlinie) Richtlinie 2014/30/EU des Europäischen Parlaments und des Rates vom 26. Februar 2014 (Elektromagnetische Verträglichkeit) Angewendete Normen und technische Spezifikationen: EN : , Sicherheitsanforderungen an elektrische Mess-, Steuer-, Regel-, und Laborgeräte, Teil1 Allgemeine Anforderungen (IEC : Cor. :2011) EN : Elektrische Mess-, Steuer-, Regel- und Laborgeräte - EMV-Anforderungen Teil 1: Allgemeine Anforderungen (IEC :2012); Deutsche Fassung EN :2013 EN : EN : EN : EN : EN : EN : EN : EN : Brieselang, den Bernd Schulz Elektronik 2

3 1 INTRODUCTION TECHNICAL SPECIFICATION MEASUREMENT ACCURACY FURTHER INFORMATION WARNING SIGNS AND SYMBOLS FUNCTIONAL DESCRIPTION PHYSICAL FUNDAMENTALS OF THE POWER MEASUREMENT AT ELECTRO SURGERY UNITS POWER MEASUREMENT AT HF ELECTROSURGICAL DEVICES LEAKAGE CURRENT MEASUREMENT AT HF ELECTROSURGICAL DEVICES FUNDAMENTAL MEASUREMENT SETUP NEUTRAL ELECTRODE TEST CREST FACTOR OPERATION INSTRUCTIONS HOW TO USE MAIN MENU POWER HF LEAKAGE CURRENT SETUP PC-OPERATION MEASUREMENT SERVICE AND CALIBRATION

4 1 Introduction The Electrosurgical Analyzer HF-400 serves to test the function of electro surgery units. The measurements and examinations are accomplished after the valid regulations as well as appropriate manufacturer recommendations. You can accomplish individual measurements with the device or let run off automatic examinations over the PC software. The operation of the device is cursor controlled and with a few YES / NO decisions the different functions of the device could be activated. The last choice you made will remain, also after switching the device off, so that you do not need to select everything again. In the stand-alone-operation, the device is used with 6 foil keys, which are integrated in the front panel. To control the HF-400, a 100% compatible IBM computer with industrial standards is needed. The communication between the PC and the HF-400 follows through the USB or RS-232 interface. Attention Safety instructions! During and after the measurements an increased danger of injury exits. During the measurements, suitable steps must guarantee the protection against dangerous voltage. 4

5 2 Technical specification Line voltage: 83 V 264 V ac, 50 / 60 Hz Power consumption: 50 VA Safety class: 1 ambient temperature: C Storage temperature: C Measuring ranges HF current : HF-Peak Current (V2): Discrimination: or HF output power: Crest Faktor (V2) HF leakage current: Discrimination: Neutral electrode test: Range: Measuring principle: Load resistors: Swing in time: Rated voltage: Output power: ma ma 0,1 ma Watt 1-10 (at > 1000 ma) ma 0,1 ma Ohm 0,3-10 MHz Thermal electric converter 10 Ohm, 25 Ohm Ohm in steps of 25 Ohm < 3 sec. max V 500 watt 1 min on, 5 min off Permanent max. 200 watt at 25 C environmental temperature ( Ohm) 5

6 2.1 Measurement accuracy Measurement Range Error HF output power: watt 1 watt or 2,5 % of measurement value HF-leakage current: ma 2 ma or 5 % of measurement value HF current : ma 2 ma or 4 % of measurement value HF-Peak Current (V2): ma 2 ma or 4 % of measurement value Load resistors: 10 Ohm, 25 Ohm 6375 Ohm 5 % (The specified measuring accuracy refers to the measuring element. Technical modifications and errors reserved. 05/2016) 2.2 Further information Keyboard: Display: Interfaces: Testing plugs: Accessories: Mechanical data: 6 key foil keyboard 4 x 20 char LCD B/W-Display 1 x USB for PC 1 x RS-232 for PC 1 x RS-232 for additional test devices 2 safety plugs 4 mm for HF output power 2 safety plugs 4 mm for HF leakage current 1 safety plug 4 mm for PE 1 x aquipotential 1 x potential balance cable 1 x RS-232 cable Light weight metal case 340 x 87 x 290 mm (W x H x D) Approx. 3,8 kg 6

7 3 Warning signs and symbols Please read this documentation before using the device. Within this manual potential hazards are advised and the actions necessary to avoid risks described. HINT: Labels and texts on the device are part of the documentation! All warning labels and symbols on the HF-400 are explained in the following: general warning signs (IEC :2011) Disregard documentation, read the manual! terminal for earth (IEC :2011) functional earth terminal replaced: OLD symbol HINT: Detachable power cords may not be replaced by poorly dimensioned power cords. 7

8 4 Functional description The HF-400 is a micro controller controlled testing device to test the function of electro surgery units in accordance with regulations and recommendations of the manufactures. The load resistor is adjustable to 10 Ohm and from Ohm in steps of 25 Ohm. The test parameters for testing can be laid down in a test instruction and can be tested automatically with a PC. This makes it possible to reduce the time for testing. On the front plate are 5 safety laboratory sockets. 2 safety laboratory sockets serve for the measurement of the HF resulting load current over the adjustable load resistance, which can be set from 10 ohms and 25 ohms 6375 ohms in 25 ohms steps. 2 safety laboratory sockets exclusively serve for the measurement of the HF leakage current through a fixed 200 ohms resistance. The 5-th Safety laboratory socket serves for the additional contacting of the equipotential-connection of the HF electro surgery devices to the measuring instrument. The normative demanded combinations can be made now by the operator by using safety laboratory lines which can he connected freely. This function mode offers a maximum of flexibility for the operator. Who is now in the situation to fulfill each also still so complex circuit instruction of the HF surgery devices manufacturers at minimum expenditure. At the same time the influence of the measuring instrument to the controller action of the HF electro surgery device is minimized and thus an improvement of the measuring behaviour is reached. The external variants have to be made as described in the appendix The following parameters can be determined: SPL551 SPL552 SPL553 SPL554 SPL555 SPL556 (V2) SPL557 (V2) SPL558 (V2) SPL559 HF output power HF leakage current HF-U RMS HF-I RMS HF-P M3 HF-U Peak HF-I Peak Crest Faktor Neutral Test 8

9 4.1 Physical fundamentals of the power measurement at electro surgery units Power output P L of electro surgery units is, as it is with all electrical generators, the product of the voltage U L at the load resistor R L and the current I L by the load resistor R L. (1) P L = U L x I L R i I L U 0 R L U L electrosurgery units I L and U L are dependent, as it is shown in the alternate circuit diagram, on the internal resistor R i of the surgical device and from the neutral voltage U 0 from the HF-generator within the HF surgical device: U0 (2) IL R R i L (3) U I R L L L U0 R R R i L L If you take the right sides of the equations (2) and (3) and put them in equation (1), then the dependence of power output P L from the electro surgery units on internal resistor Ri and on the load resistor R L is recognizable: (4) P L U0 R R i L 2 R L 9

10 With R L = 0 is P L = 0. This means that in case of short circuit, the power output is zero too. With R L = is P L = 0. This means that in case of neutral operation, the output power is zero. With R L = R i is P L = P Lmax. This means that the power P L reaches its maximum if the load resistor R L is equal to the internal resistor R i of the generator. With R L R i is P L < P Lmax. This means if the load resistor R L is larger or smaller than the internal resistor R i of the electro surgery units, the output power P L is smaller than its maximum output power P Lmax. This condition is called mismatching. Attention: The measurement of power only results in the maximum power in case of a power adaption, i.e. if the load resistor R L is equal to the internal resistor R i of the generator. If R L should be larger or smaller than R i, or is unknown, then the maximum power must be looked for by variation by R L. 4.2 Power measurement at HF Electrosurgical Devices The HF voltage of the HF Electrosurgical Analyzer is applied to a selectable load resistance. The current resulting from this HF voltage produces a HF voltage over a special current transformer proportional to the HF current. This HF voltage is loaded by a voltage divider and the signal resulting in such a way is down-divided to the input level of the RMS transducer. The RMS transducer works according to the thermoelectric principle. It consists of two equal measuring circles. Each measuring circle consists of a heating resistor and an infrared diode. The HF voltage arrived over a voltage divider on the heating resistor of the entrance circle and produces a temperature those for the rms of the HF voltage put on is proportional. The temperature of the heating resistor controls the resistor of the infrared diode. In the load circuit of the RMS transducer the second heating resistor will be heated with DC voltage, these change the resistor of the second infrared diode. A control circuit compares both of the heating power and regulates the secondary DC-voltage as long as both heating circles have the same temperature. DC voltage on the secondary winding, adjusting thereby, is proportional to the primary RMS voltage. The RMS value is digitized with an AD transducer, processed in Controller and shown on the display. Picture 6.1 and

11 4.3 Leakage current measurement at HF Electrosurgical Devices The HF leakage current of the HF Electrosurgical Devices is set to a 200 ohms load resistance. This current produces proportional HF voltage to the HF current over a special current transformer. This HF voltage is loaded by a voltage divider and the signal resulting in such a way is down-divided to the input level of the RMS transducer. Picture Fundamental measurement setup To start-up the device, connect it with the main voltage and if necessary start the connection with the PC over the provided RS-232 cable. At the PC put the RS-232 cable into RS-232 interface. The plug next to the PC plug (on the backside) of HF-400 with the description test adapter could be used for the connection with another testing device, e.g. modular test system GS-X. With the power switch the HF-400 is turned on. After switching on the self-check of the device begins. Afterwards the equipment is set into the mode of operation condition stand-alone. 4.5 Neutral electrode test The neutral electrode test is used to check the neutral electrode monitoring on RFsurgery devices. The neutral electrodes are connected to the Bi safety laboratory sockets according to picture After the selection of the function step (SPL559), the resistance is increased incrementally. While changing the resistance, the current in the RF surgery unit of the neutral electrodes changes too. This change has resulted, that the RF-surgery unit detects whether the neutral electrodes are correctly attached. Otherwise, the control refers to. The threshold of monitoring can check with this function step. Picture

12 4.6 Crest factor The crest factor describes the ratio of peak to RMS and serves as a characteristic value for the assessment of the waveform of the RF signal. High Crest factors mean a high proportion of harmonics. A sinusoidal alternating current has, for example, a crest factor of 1,414. The crest faktor measured under load, with a current rating of greater than 1000 ma. 12

13 5 Operation instructions 5.1 How to use The device works with six foil keys. ESC up left right down Enter The arrow keys serve the selection in the menu or to the parameter selection. The ENTER key serves as a start button. With the ESC key, the respective program can be broken off. The latest selection made is stored into the device, so it must not be selected again. 13

14 5.2 Main menu After switching on the device, you will see following announcement: * * * S. P. L. * * * E l e k t r o n i k H 2. 0 V After a short time, the main menu appears with following items: H F > H F p o w e r H F l e a k. c u r r. S e t u p H F > H F - U R M S H F - I R M S H F - P R M S V2: H F > H F - U P e a k H F - I P e a k C r e s t H F > N e u t r a l T e s t B i - B i 14

15 5.3 Power When you choose power with the ENTER-key, you can select the measurement sorted after different resistor values. S e l e c t l o a d r e s i s t Ω > 2 5 Ω 7 5 Ω Ω Ω Ω Ω With arrow keys RIGHT and LEFT you can select the value of the resistor you want. With arrow keys UP and DOWN you can change the selected resistor value in steps of 25 Ohm. With the ENTER key you start the measurement. With the ESC key, you will stop and leave the measurement. If you press the ESC key twice, you will be sent to the main menu. S P L H F o u t p u t p o w e r R L : 7 5 Ω H F p o w e r : W On the display the test step number, the test step description, the actual value, the measuring unit and the used load resistor are shown. The test step number has a three-figure number. 15

16 5.4 HF Leakage current If you choose HF leakage current with the ENTER key, you can select the measurement sorted after resistors. S e l e c t l o a d r e s i s t Ω > 2 5 Ω 7 5 Ω Ω Ω Ω Ω With arrow keys RIGHT and LEFT you select the value of the resistor you want. With arrow keys UP and DOWN you can change the selected resistor value in steps of 25 Ohm. With the ENTER key you start the measurement. With the ESC key, you will stop and leave the measurement. If you press the ESC key twice, you will be sent to the main menu. S P L H F L e a k a g e c u r r e n t R L : 7 5 Ω L e a k c u r r. : m A On the display the test step number, the test step description, the actual value, the measuring unit and the used load resistor are shown. The test step number has a three-figure number. 16

17 5.5 Setup If you choose current with the ENTER key, you can select the setup of date, time and language. H F t i m e & d a t e n e x t C a l i b r a t i o n c h o o s e l a n g u a g e With these setting you can set the real-time clock. H F t i m e & d a t e > : 0 8 : 2 6 H F c h o o s e l a n g u a g e > d e u t s c h With these settings, you can choose between following languages: german, english, france, polski, spanish, italian, portugese and turkish H F S e r i a l n u m b e r : n e x t C a l i b r a t i o n The date of the next recommended calibration and the Serial number is shown too. 17

18 5.9 PC-operation The communication between the PC and the HF-400 follows through the USB / RS-232 interface with a special PC-Software. Function steps from HF-400 SPL551 SPL552 SPL553 SPL554 SPL555 SPL556 (V2) SPL557 (V2) SPL558 (V2) SPL559 HF output power HF leakage current HF-U RMS HF-I RMS HF-P M3 HF-U Peak HF-I Peak Crest Faktor Neutral Test PC-Parameter from HF-400 Load resistor 10 Ohm, 25 Ohm to 6375 Ohm in steps of 25 Ohm 18

19 6 Measurement Picture 6. 1 Picture

20 Picture 6. 3 Picture

21 Picture 6. 5 Picture

22 Picture 6. 7 Picture

23 Picture 6. 9 Picture

24 Picture Picture

25 Picture Picture

26 Picture

27 7 Service and calibration The measuring - and test devices quality, manufactured from Elektronik Brieselang, is guaranteed by a multi-level quality control - and quality assurances in accordance to DIN EN ISO After final inspection and calibration each device gets an examine plaque and a calibration certificate. The calibration is enforced with standards, which are restorable to the standards of the Physical- Technical Federal Institution. In agreement with TÜV CERT, business-group TÜV Rheinland / Berlin- Brandenburg, is recommended for normal, accordingly to the corresponding use of the measuring and test devices form Elektronik Brieselang an annually calibration interval. The calibration includes a maintenance, update of the firmware, complete control of the accuracy of measurements for of all test steps by control with one or several reference-values and calibration. By an update of the firmware for example, changes in standards, expansions of the function of the devices etc. are automatically integrated. By the implemented inspection and calibration, the validity is fixed by the implemented measurement list in the associated user guide. The warranted characteristics for the PC-operation of the measuring - and test devices of Elektronik are guaranteed through the offered or authorized softwareprograms. For cleaning, you should wipe off your device with a moist cloth, which is impregnated with a mild detergent. It is important after that to rub your device dry. Attention Hint! Before cleaning always pull of the power plugs, otherwise danger for life exists Please take note in the use with the test system that it is despite of protected measuring channel inputs a sensitive measuring instrument. Waste disposal: Through the used parts of the measuring - and test devices of the Elektronik, etc. accumulators, Lithium batteries, and similar things, a professional waste disposal is necessary. This can be done through Elektronik, ear , if demand is enforced. 27

Portable Appliance Testers. OmegaPAT MI 2140 BetaPAT MI 2141 User Manual Ver Code No

Portable Appliance Testers. OmegaPAT MI 2140 BetaPAT MI 2141 User Manual Ver Code No Portable Appliance Testers OmegaPAT MI 2140 BetaPAT MI 2141 User Manual Ver. 1.2. Code No. 20 750 684 Distributor: Producer: METREL d.d. Ljubljanska 77 SI-1354 Horjul E-mail: metrel@metrel.si http://www.metrel.si

More information

Digital Function Generator

Digital Function Generator Digital Function Generator 13654-99 PHYWE Systeme GmbH & Co. KG Robert-Bosch-Breite 10 37079 Göttingen Germany Tel. +49 (0) 551 604-0 Fax +49 (0) 551 604-107 E-mail info@phywe.de Operating Instructions

More information

Precipitation Monitor ,

Precipitation Monitor , THE WORLD OF WEATHER DATA - THE WORLD OF WEATHER DATA - THE WORLD OF WEATHER DATA Instruction for use 021197/11/09 Precipitation Monitor 5.4103.10.000, 5.4103.10.700 ADOLF THIES GmbH & Co. KG Hauptstraße

More information

Operating Instructions PROFITEST H+E TECH. Diagnostics Unit for Electric Charging Stations (Type 2 Connector Socket and Plug) /3.

Operating Instructions PROFITEST H+E TECH. Diagnostics Unit for Electric Charging Stations (Type 2 Connector Socket and Plug) /3. Diagnostics Unit for Electric Charging Stations (Type 2 Connector Socket and Plug) 3-349-878-03 1/3.16 Opening the Instrument / Repairs The instrument may only be opened by authorized, trained personnel

More information

3B SCIENTIFIC PHYSICS

3B SCIENTIFIC PHYSICS 3B SCIENTIFIC PHYSICS Digital Multimeter E 1018832 Instruction sheet 12/16 SD/UD 1 probe 1a Finger guards 2 Measurement socket 10 A for current measurement in 10-A (positive) 3 Measurement socket COM (negative)

More information

Rigel 377. Instruction Manual 367A550 Issue 2.0. Electrosurgical Analyser. August Seaward Group Issue 2.0

Rigel 377. Instruction Manual 367A550 Issue 2.0. Electrosurgical Analyser. August Seaward Group Issue 2.0 Rigel 377 Electrosurgical Analyser August 2011 2011 Seaward Group Issue 2.0 Instruction Manual 367A550 Issue 2.0 Limited Warranty & Limitation of Liability Rigel Medical, part of the SEAWARD GROUP guarantees

More information

Standard Rocking Platform WT15 Code No , Manual August 2004

Standard Rocking Platform WT15 Code No , Manual August 2004 Standard Rocking Platform WT15 Code No. 042-400, 042-490 Manual August 2004!! Warning!! Please read these instructions carefully before using this apparatus! Biometra GmbH Rudolf-Wissell-Str. 30 D-37079

More information

Precipition Sensor with analogue Intensity Output ,

Precipition Sensor with analogue Intensity Output , THE WORLD OF WEATHER DATA - THE WORLD OF WEATHER DATA - THE WORLD OF WEATHER DATA Instruction for use 021335/11/09 Precipition Sensor with analogue Intensity Output 5.4103.20.041, 5.4103.20.741 ADOLF THIES

More information

Operating Instructions / Voltage Tester per IEC with Phase and Polarity Tester and Phase Sequence Indicator

Operating Instructions / Voltage Tester per IEC with Phase and Polarity Tester and Phase Sequence Indicator Operating Instructions 3-349-298-15 3/11.09 METRAVOLT 7A Voltage Tester per IEC 61243-3 with Phase and Polarity Tester and Phase Sequence Indicator CAT IV GMC-I Messtechnik GmbH Südwestpark 15 90449 Nürnberg

More information

SUPPLY NETWORK ANALYZER. CVMk SERIES. CVMk-4C SERIES INSTRUCTION MANUAL. ( M / 00 B - Manual 1 / 2 ) (c) CIRCUTOR S.A.

SUPPLY NETWORK ANALYZER. CVMk SERIES. CVMk-4C SERIES INSTRUCTION MANUAL. ( M / 00 B - Manual 1 / 2 ) (c) CIRCUTOR S.A. SUPPLY NETWORK ANALYZER CVMk SERIES & CVMk-4C SERIES INSTRUCTION MANUAL ( M 981 171 / 00 B - Manual 1 / 2 ) (c) CIRCUTOR S.A. ----- Supply network analyzer CVMk and CVMk-ITF ------ Manual 1 / 2 --- Page

More information

RIGOL. User s Guide. RP1000D Series High Voltage Differential Probe. Feb RIGOL Technologies, Inc

RIGOL. User s Guide. RP1000D Series High Voltage Differential Probe. Feb RIGOL Technologies, Inc User s Guide RP1000D Series High Voltage Differential Probe Feb. 2013 RIGOL Technologies, Inc Guaranty and Declaration Copyright 2012 RIGOL Technologies, Inc. All Rights Reserved. Trademark Information

More information

RMO500 M I C R O O H M M E T E R

RMO500 M I C R O O H M M E T E R M I C R O O H M M E T E R Reference Manual IBEKO POWER AB Sweden 2 IBEKO POWER AB Manual Version:. MV.01 IBEKO POWER AB 2003 This Reference Manual is a publication of IBEKO POWER AB, 18 125 Lidingö, Sweden.

More information

MODEL 3810/2 Line Impedance Stabilization Network

MODEL 3810/2 Line Impedance Stabilization Network EMC TEST SYSTEMS FEBRUARY 1996 REV C PN 399197 MODEL 3810/2 Line Impedance Stabilization Network OPERATION MANUAL USA P.O. Box 80589 Austin, Texas 78708-0589 2205 Kramer Lane, Austin, Texas 78758-4047

More information

ELECTROSURGICAL UNIT ANALYZER

ELECTROSURGICAL UNIT ANALYZER ELECTROSURGICAL UNIT ANALYZER ESU-2000A USER MANUAL BC BIOMEDICAL ESU-2000A TABLE OF CONTENTS WARNINGS, CAUTIONS, NOTICES... ii DESCRIPTION... 1 OVERVIEW... 2 OPERATING INSTRUCTIONS... 3 MANUAL REVISIONS...

More information

Voltage Transient Emission Test

Voltage Transient Emission Test Manual For Operation AN 200 Series Voltage Transient Emission Test AN 200 AN 200B The AN 200 is used to evaluate automotive electrical and electronic components for conducted emissions of transients along

More information

Standard Power Pack P25T Order-No Manual May 2009

Standard Power Pack P25T Order-No Manual May 2009 Standard Power Pack P25T Order-No. 040-850 Manual May 2009!!Attention!! Please read this manual carefully prior to operating the device Biometra GmbH Rudolf-Wissell-Straße 30, D-37079 Göttingen P.O. Box

More information

Portable equipment for insulation fault location for unearthed and earthed systems (IT and TN systems) to be used in conjunction with or without an

Portable equipment for insulation fault location for unearthed and earthed systems (IT and TN systems) to be used in conjunction with or without an EDS309 Portable equipment for insulation fault location for unearthed and earthed systems (IT and TN systems) to be used in conjunction with or without an equipment for insulation fault location EDS309x_D00012_02_D_XXEN/11.2018

More information

GammaPAT MI 3311 Short instructions Ver. 1.4, Code no

GammaPAT MI 3311 Short instructions Ver. 1.4, Code no GammaPAT MI 3311 Short instructions Ver. 1.4, Code no. 20 751 626 Distributor: Manufacturer: METREL d.d. Ljubljanska cesta 77 1354 Horjul Slovenia E-mail: metrel@metrel.si http://www.metrel.si 2010 METREL

More information

METRATESTER 5+ Tester for DIN VDE /6.15

METRATESTER 5+ Tester for DIN VDE /6.15 3-348-817-03 15/6.15 Applications Electrical safety testing of electrical equipment: per DIN VDE 0701-0702: 2008 by means of measurement of: Protective conductor resistance Insulation resistance Protective

More information

DM-45 Digital Multimeter

DM-45 Digital Multimeter INSTRUCTION MANUAL DM-45 Digital Multimeter Read and understand all of the instructions and safety information in this manual before operating or servicing this tool. Description The Greenlee DM-45 Digital

More information

SUNNY BEAM REPEATER Transmission Range Increase for Sunny Beam

SUNNY BEAM REPEATER Transmission Range Increase for Sunny Beam SUNNY BEAM REPEATER Transmission Range Increase for Sunny Beam User Manual SBeamRep-BEN091911 98-0002611 Version 1.1 EN SMA Solar Technology AG Table of Contents Table of Contents 1 Notes on this Manual..............................

More information

SIMEAS-T. Operating Instructions Transducer without auxiliary power. 7KG6111 and 7KG6101. Operating Instructions

SIMEAS-T. Operating Instructions Transducer without auxiliary power. 7KG6111 and 7KG6101. Operating Instructions Operating Instructions SIMEAS-T s Operating Instructions Transducer without auxiliary power for alternating current for alternating voltage for alternating voltage with expanded end range 7KG6111 and 7KG6101

More information

Operating Instructions TIC 251/252. Profibus DP-Gateway PG 0009 BE/E (0903)

Operating Instructions TIC 251/252. Profibus DP-Gateway PG 0009 BE/E (0903) Operating Instructions TIC 251/252 Profibus DP-Gateway PG 0009 BE/E (0903) Contents Page 1. Safety Precautions... 3 2. Understanding The TIC 251/252... 4 2.1. For Your Orientation... 4 Delivery Consignment...

More information

QA-ES III Electrosurgical Analyzer

QA-ES III Electrosurgical Analyzer QA-ES III Electrosurgical Analyzer Technical Data The QA-ES III Electrosurgical Analyzer simplifies testing to ensure the performance and safety of electrosurgical units. With generator output current

More information

Portable equipment for insulation fault location for unearthed and earthed systems (IT and TN systems) to be used in conjunction with or without an

Portable equipment for insulation fault location for unearthed and earthed systems (IT and TN systems) to be used in conjunction with or without an EDS309 Portable equipment for insulation fault location for unearthed and earthed systems (IT and TN systems) to be used in conjunction with or without an equipment for insulation fault location EDS309x_D00012_00_D_XXEN/05.2015

More information

TETRIS 1000 High Impedance Active Probe. Instruction Manual

TETRIS 1000 High Impedance Active Probe. Instruction Manual TETRIS 1000 High Impedance Active Probe Instruction Manual Copyright 2015 PMK GmbH All rights reserved. Information in this publication supersedes that in all previously published material. Specifications

More information

QA-ES III Electrosurgical Analyzer

QA-ES III Electrosurgical Analyzer QA-ES III Electrosurgical Analyzer Technical Data The QA-ES III Electrosurgical Analyzer simplifies testing to ensure the performance and safety of electrosurgical units. With generator output current

More information

MAGNETIC FIELD METER Operator s Manual

MAGNETIC FIELD METER Operator s Manual Edition 4.4 September 2011 MAGNETIC FIELD METER 3000 Operator s Manual The MFM 3000 is a professional magnetic field instrument To make the best use of the instrument we recommend that you read this manual

More information

LT Pilot remote control system for Licht-Technik motoryokes and Click and Move

LT Pilot remote control system for Licht-Technik motoryokes and Click and Move LT Pilot remote control system for Licht-Technik motoryokes and Click and Move Users manual LT Pilot V2.0 Fabrication and Marketing Licht-Technik Hagenbach & Grill Osterwaldstr. 9-10 80805 München Tel.

More information

SECULIFE SR PC Controllable Instrument for Measuring Safety-Relevant Characteristic Values of Electrical Devices /2.

SECULIFE SR PC Controllable Instrument for Measuring Safety-Relevant Characteristic Values of Electrical Devices /2. 3-349-443-03 3/2.12 Applications Testing for the Electrical Safety of Electrical Medical Devices in Accordance with the German Medical Product Law (MPG) / MDD The test instrument is used for quick and

More information

High power radio transmission module MR03 type

High power radio transmission module MR03 type High power radio transmission module MR03 type User s manual CONTENTS 1. APPLICATION...3 2. MR03 MODULE SET...4 3. INSTALLATION...4 3.1 Module assembly...4 3.2 Connection diagrams...5 3.3 Connection way

More information

User s Manual. Ground Resistance Clamp On Tester MODEL

User s Manual. Ground Resistance Clamp On Tester MODEL User s Manual Ground Resistance Clamp On Tester MODEL 382357 Warranty EXTECH INSTRUMENTS CORPORATION warrants the basic instrument to be free of defects in parts and workmanship for one year from date

More information

Power Meter. Measurement Guide. for Anritsu RF and Microwave Handheld Instruments BTS Master Site Master Spectrum Master Cell Master

Power Meter. Measurement Guide. for Anritsu RF and Microwave Handheld Instruments BTS Master Site Master Spectrum Master Cell Master Measurement Guide Power Meter for Anritsu RF and Microwave Handheld Instruments BTS Master Site Master Spectrum Master Cell Master Power Meter Option 29 High Accuracy Power Meter Option 19 Inline Peak

More information

Network Analyzer for Low-, Medium- and High-Voltage Networks

Network Analyzer for Low-, Medium- and High-Voltage Networks Technical data Network nalyzer for Low-, Medium- and High-Voltage Networks Model PQ-Box 150 1 Fault detection 1 Evaluation of voltage quality according to EN50160 and IEC61000-2-2 (2-4) 1 FFT nalysis up

More information

IDEAL INDUSTRIES, INC. TECHNICAL MANUAL MODEL: MODEL: Multimeter Service Information

IDEAL INDUSTRIES, INC. TECHNICAL MANUAL MODEL: MODEL: Multimeter Service Information IDEAL INDUSTRIES, INC. TECHNICAL MANUAL MODEL: 61-340 MODEL: 61-342 Multimeter Service Information The Service Information provides the following information: Precautions and safety information Specifications

More information

FREJA 306 Relay Testing System

FREJA 306 Relay Testing System Relay Testing System A Megger Group Company FREJA 306 Relay testing system FREJA 306 is the latest member of the relay testing equipment from Programma, quick and easy to use, like the FREJA 300. The rugged

More information

Voltage Sensors URV5-Z

Voltage Sensors URV5-Z Data sheet Version 05.00 Voltage Sensors URV5-Z May 2005 Universal voltage measurements from RF to microwaves The voltage sensors of the URV5-Z series are indispensable tools in RF and microwave laboratories,

More information

ISOMETER IRDH575. Approvals

ISOMETER IRDH575. Approvals Insulation monitoring device for unearthed AC, DC and AC/DC systems (IT systems) with control and display function for EDS insulation fault location systems IRDH575_D00089_02_D_XXEN/09.2018 Insulation

More information

DEUTSCH NORSK ENGLISH SVENSKA FRANÇAIS DANSK NEDERLANDS ČEŠTINA ITALANIO Hersteller UR190

DEUTSCH NORSK ENGLISH SVENSKA FRANÇAIS DANSK NEDERLANDS ČEŠTINA ITALANIO Hersteller UR190 DEUTSCH ENGLISH FRANÇAIS NEDERLANDS ITALANIO NORSK SVENSKA DANSK ČEŠTINA Hersteller Wörlein GmbH Tel.: +49 9103/71670 Gewerbestrasse 12 Fax.: +49 9103/716712 D 90556 Cadolzburg Email. info@woerlein.com

More information

U ACTIVE SAT DISTRIBUTION FIELD

U ACTIVE SAT DISTRIBUTION FIELD U 911... 946 ACTIVE SAT DISTRIBUTION FIELD Illustrations front view U 911 Controller port Indicator lamps Remote feed ON/OFF Labeling field Measurement points outputs Y outputs X input Y input X Bus port

More information

PQ-Box 100 Network Analyzer for Low-, Medium- and High-Voltage Networks

PQ-Box 100 Network Analyzer for Low-, Medium- and High-Voltage Networks GB Technical data Issue 06/2009 PQ-Box 100 Network Analyzer for Low-, Medium- and High-Voltage Networks Fault detection Evaluation of voltage quality according to EN50160 and IEC61000-2-2 Fault recorder

More information

DATA SHEET Selectable AC transducer TAS-311DG

DATA SHEET Selectable AC transducer TAS-311DG DATA SHEET Selectable AC transducer Measuring on AC networks Class 0.5 (IEC688) measurement Supply/measur. voltage up to 690 V Easy configuration via PC interface possible DEIF A/S Frisenborgvej 33 DK-7800

More information

MW3105 DIGITAL CLAMP MULTIMETER

MW3105 DIGITAL CLAMP MULTIMETER MW3105 DIGITAL CLAMP MULTIMETER 2 M MW3105 A 01 INTRODUCTION 1.1 - Unpacking and inspection Upon removing your new Digital Clamp Meter from its packing, you should have the following items: 1. Digital

More information

UMG 103-CBM. Universal measurement device for DIN rails

UMG 103-CBM. Universal measurement device for DIN rails Universal measurement device for DIN rails Harmonics GridVis Analysis software Modbus interface Measurement accuracy 0.5 Communication Protocols: Modbus RTU / Slave Interface RS485 Accuracy of measurement

More information

Operating Instruction

Operating Instruction Level and Pressure Operating Instruction VEGASEL 643 6 5 + + 3 5 + + max. min. on VEGASEL 643 5 6 7 8 9 10 11 12 13 14 in out Contents Contents Safety information... 2 1 Product description 1.1 Function

More information

Indeks: WMGBPAT815 Miernik bezpieczeństwa sprzętu elektrycznego

Indeks: WMGBPAT815 Miernik bezpieczeństwa sprzętu elektrycznego PAT-815 Indeks: WMGBPAT815 Miernik bezpieczeństwa sprzętu elektrycznego Descripción PAT-815 digital meter is used to measure the parameters of electrical equipment (power tools, white goods, etc.) which

More information

PLA 33. Power line analyzer. User and service manual. version 2.4

PLA 33. Power line analyzer. User and service manual. version 2.4 PLA 33 Power line analyzer User and service manual version 2.4 Content. Front control panel and terminal plate...3 7.2.2. System frequency setting...0 2. Device description...4 7.2.3. Password protection...0

More information

INSTRUCTION MANUAL LKG 601 Electrical Safety Analyzer

INSTRUCTION MANUAL LKG 601 Electrical Safety Analyzer INSTRUCTION MANUAL LKG 601 Electrical Safety Analyzer 110 Toledo Street Farmingdale, NY 11735 USA http://www.netech.org 510-USER-Manual Rev3 10/29/2007 Dear User, We appreciate your purchase of the LKG

More information

DIGITAL DUAL DISPLAY AC/DC CLAMP METER MODEL-860A OPERATION MANUAL

DIGITAL DUAL DISPLAY AC/DC CLAMP METER MODEL-860A OPERATION MANUAL DIGITAL DUAL DISPLAY AC/DC CLAMP METER MODEL-860A OPERATION MANUAL DIGITAL DUAL DISPLAY AC/DC CLAMP METER MODEL-860A TABLE OF CONTENTS TITLE PAGE Safety Information Safety Symbols... 1 Meter Description...

More information

Mars-Energo. Volt/mA Calibrator CALMAR-S. User s manual Edition 3.1

Mars-Energo. Volt/mA Calibrator CALMAR-S. User s manual Edition 3.1 Mars-Energo Volt/mA Calibrator CALMAR-S User s manual Edition 3.1 Manufacturer: OOO NPP Mars-Energo Address: V.O. 13 Line, 6-8, office 41H, Saint-Petersburg, Russia, 199034 2017 MARS-ENERGO Table of contents

More information

CEC VIBRATION TRANSMITTER

CEC VIBRATION TRANSMITTER CEC 1-808 VIBRATION TRANSMITTER Operation & Maintenance Manual 746 Arrow Grand Circle Covina, CA 91722 United States of America Tel: (626) 938-0200 Fax: (626) 938-0202 Internet: http://www.cecvp.com E-mail:

More information

Technical Documentation

Technical Documentation Technical Documentation for metratec QuasarMX HF RFID Reader Date: April 2015 Version: 1.4 Technical Documentation metratec QuasarMX Page 1 of 13 Table of Contents 1 General Information / Security Advice...3

More information

79/26 Series III Multimeter

79/26 Series III Multimeter 79/26 Series III Multimeter Instruction Sheet W Read First: Safety Information Never use the meter if the meter or test leads look damaged. Be sure the test leads and switch are in the correct position

More information

TD 1000 PLUS. Secondary Injection Relay Test Set. Designed for testing relays and transducers

TD 1000 PLUS. Secondary Injection Relay Test Set. Designed for testing relays and transducers Secondary Injection Relay Test Set Designed for testing relays and transducers Two current outputs to test differential relays Convertible current and voltage generator With phase angle shifter Frequency

More information

High-precision process calibrator Model CED7000

High-precision process calibrator Model CED7000 Calibration technology High-precision process calibrator Model CED7000 WIKA data sheet CT 85.51 Applications Research and development laboratories Calibration service companies and service industry Industry

More information

Electronic AC Load ZSAC Series. 400 W up to 21,000 W 260 V up to 440 V 3 A up to 100 A. Rev. 3.03

Electronic AC Load ZSAC Series. 400 W up to 21,000 W 260 V up to 440 V 3 A up to 100 A. Rev. 3.03 Electronic AC Load Series 400 W up to 21,000 W 260 V up to 440 V 3 A up to 100 A Rev. 3.03 Electronic AC Loads, Series Interface overview RS-232 USB GPIB LAN System bus Analog / Analog isolated X X Standard

More information

Multi range AC current clamps (1000A / 100A / 5A / 0.5A) A 1281 Instruction manual Version 2.0, Code no

Multi range AC current clamps (1000A / 100A / 5A / 0.5A) A 1281 Instruction manual Version 2.0, Code no (1000A / 100A / 5A / 0.5A) A 1281 Instruction manual Version 2.0, Code no. 20 751 696 Distributor: Manufacturer: METREL d.d. Ljubljanska cesta 77 1354 Horjul Slovenia web site: http://www.metrel.si e-mail:

More information

Operating instructions Fail-safe delay timer AZS About this document. Content

Operating instructions Fail-safe delay timer AZS About this document. Content 8 Appendix 8.1 Wiring example...4 8.2 Integral System Diagnostics (ISD)....5 9 EU Declaration of conformity Operating instructions.............pages 1 to 6 Original x.000 / 11.2017 / v.a. - 101126753-

More information

T5-600/T Electrical Tester

T5-600/T Electrical Tester T5-600/T5-1000 Service Information Introduction This service information sheet provides the following information for the T5-600 and T5-1000 Electrical Testers (hereafter referred to as "the tester").

More information

ICARUS ELECTRONIC GAS VOLUME CONVERTER

ICARUS ELECTRONIC GAS VOLUME CONVERTER ICARUS ELECTRONIC GAS VOLUME CONVERTER 1. INTRODUCTION 2. TECHNICAL DETAILS ICARUS is an electronic volume converter which, combined with a gas meter, provides the gas measurement in Sm3, calculated at

More information

Operator s Manual. PP017 and PP018 Passive Probes

Operator s Manual. PP017 and PP018 Passive Probes Operator s Manual PP017 and PP018 Passive Probes 700 Chestnut Ridge Road Chestnut Ridge, NY, 10977-6499 Tel: (845) 425-2000, Fax: (845) 578 5985 teledynelecroy.com PP017 and PP018 Passive Probes Operator

More information

E205 Long Range Wireless Modem V1.0 Data Sheet

E205 Long Range Wireless Modem V1.0 Data Sheet E205 Long Range Wireless Modem V1.0 Data Sheet The Long Range Wireless Modem E205 is a radio transmitter / receiver designed for transparent wireless data transmission over long distances. It operates

More information

Mounting instruction and operating manual. Access Point (UK) HmIP-HAP-UK

Mounting instruction and operating manual. Access Point (UK) HmIP-HAP-UK Mounting instruction and operating manual Access Point (UK) HmIP-HAP-UK Package contents Quantity Description 1 Homematic IP Access Point (UK) 1 Plug-in mains adapter 1 Network cable 2 Screws 2 Plugs 1

More information

C4000. inline control. Photometric Converter. Real-time photometric converter. Factory zero for scattered light sensors

C4000. inline control. Photometric Converter.  Real-time photometric converter. Factory zero for scattered light sensors C4000 Photometric Converter inline control Real-time photometric converter Factory zero for scattered light sensors Simultaneous operation of up to four inline sensors Process control functions via Remote

More information

T 1000 PLUS Secondary Injection Relay Test Set

T 1000 PLUS Secondary Injection Relay Test Set Secondary Injection Relay Test Set Designed for testing relays and transducers Microprocessor controlled With phase angle shifter Frequency generator Test results and settings are saved into local memory

More information

Operating Manual * * Differential pressure transmitter. Table of Contents. 1 Safety guidelines. 1.1 General Information

Operating Manual * * Differential pressure transmitter. Table of Contents. 1 Safety guidelines. 1.1 General Information *09005137* BA_EN_DE50 Rev.A 11/12 *09005137* d e v e l o p i n g s o l u t i o n s DE50 Operating Manual Differential pressure transmitter Table of Contents 1 Safety guidelines 2 Application purpose 3

More information

T 1000 PLUS. Secondary Injection Relay Test Set. Designed for testing relays and transducers

T 1000 PLUS. Secondary Injection Relay Test Set. Designed for testing relays and transducers Secondary Injection Relay Test Set Designed for testing relays and transducers Microprocessor controlled With phase angle shifter Frequency generator High power outputs Large graphical display Compact

More information

Earth 2/3 MI 2126 Instruction Manual Version 1.0, Code No

Earth 2/3 MI 2126 Instruction Manual Version 1.0, Code No Earth 2/3 MI 2126 Instruction Manual Version 1.0, Code No. 20 750 190 Distributor: Manufacturer: METREL d.d. Ljubljanska cesta 77 1354 Horjul Slovenia web site: http://www.metrel.si e-mail: metrel@metrel.si

More information

Earth Leakage Monitoring System IsoBase IsoHub IsoOut

Earth Leakage Monitoring System IsoBase IsoHub IsoOut 2014 11 21 Earth Leakage Monitoring System IsoBase IsoHub IsoOut User Manual Due to our policy of continual improvement, specifications may change without prior notice Page 2 (34) Contents Earth Leakage

More information

EC Declaration of Conformity

EC Declaration of Conformity EC Declaration of Conformity TABLE OF CONTENTS 1. SAFETY SUMMARY.. 1 2. INTRODUCTION 5 3. SPECIFICATION 7 EN 61326-1: Electrical equipment for measurement, control and laboratory use EMC requirements (1997+A1:

More information

GREISINGER electronic GmbH D Regenstauf, Hans-Sachs-Straße 26 phone: / , fax: / ,

GREISINGER electronic GmbH D Regenstauf, Hans-Sachs-Straße 26 phone: / , fax: / , H66.0.01.6C-05 Operating Manual Conductivity measuring instrument from Version 1.2 GLF 100 GREISINGER electronic GmbH D - 93128 Regenstauf, Hans-Sachs-Straße 26 phone: +49 9402 / 9383-0, fax: +49 9402

More information

KMT - Kraus Messtechnik GmbH

KMT - Kraus Messtechnik GmbH www.instrumentation.it INSTRUMENTATION DEVICES SRL Via Acquanera 29, 22100 COMO (Italy) tel. +39.031.525391 - fax +39.031.507984 - info@instrumentation.it KMT - Kraus Messtechnik GmbH Gewerbering 9, D-83624

More information

SPM-50 RF Spectrum Power Meter PC Software User Manual

SPM-50 RF Spectrum Power Meter PC Software User Manual SPM-50 RF Spectrum Power Meter PC Software User Manual Shineway Technologies, Inc. Notices Copyright 2014, ShinewayTech, All rights reserved. No part of this manual may be reproduced in any form or by

More information

2015 RIGOL TECHNOLOGIES, INC.

2015 RIGOL TECHNOLOGIES, INC. Service Guide DG000 Series Dual-channel Function/Arbitrary Waveform Generator Oct. 205 TECHNOLOGIES, INC. Guaranty and Declaration Copyright 203 TECHNOLOGIES, INC. All Rights Reserved. Trademark Information

More information

POWER ANALYZER OPERATION MANUAL. Model : DW-6090A

POWER ANALYZER OPERATION MANUAL. Model : DW-6090A POWER ANALYZER Model : DW-6090A Your purchase of this POWER ANALYZER marks a step forward for you into the field of precision measurement. Although this POWER ANALYZER is a complex and delicate instrument,

More information

C.A 8332 C.A Advanced Test Equipment Rentals ATEC (2832) Display your network analysis all in picture format!

C.A 8332 C.A Advanced Test Equipment Rentals ATEC (2832) Display your network analysis all in picture format! Established 1981 Advanced Test Equipment Rentals www.atecorp.com 800-404-ATEC (282) Display your network analysis all in picture format! OBSERVE, DIAGNOSE, MONITOR... C.A 82 C.A 84 Three Phase Power Quality

More information

Electrical devices may only be mounted and connected by electrically skilled persons.

Electrical devices may only be mounted and connected by electrically skilled persons. enet Server Order No. : 5301 00 Power adapter enet Server 12 V DC Operating instructions 1 Safety instructions Electrical devices may only be mounted and connected by electrically skilled persons. Serious

More information

NADIR USER'S GUIDE. September 2009 MADE - V 3.20

NADIR USER'S GUIDE. September 2009 MADE - V 3.20 NADIR USER'S GUIDE September 2009 MADE - V 3.20 CONTENTS 1 SAFETY INFORMATION... 3 1.1 SAFETY RECOMMENDATIONS :... 3 1.2 FOLLOWING THE SAFETY RECOMMENDATIONS :... 3 1.3 WARNING LABELS... 3 1.4 DANGEROUS

More information

PKT 512A-RO High Impedance Passive Cable Divider

PKT 512A-RO High Impedance Passive Cable Divider PKT 512A-RO High Impedance Passive Cable Divider Instruction Manual Copyright 2011 PMK GmbH All rights reserved. Information in this publication supersedes that in all previously published material. Specifications

More information

Model UT511 OPERATING MANUAL

Model UT511 OPERATING MANUAL Model UT511 OPERATING MANUAL TITLE PAGE Introduction Unpacking the Meter Safety Information International Electrical Symbols Battery Saver (Sleep Mode) Battery Indication The Meter Structure Display Key

More information

T 1000 PLUS. Secondary Injection Relay Test Set. Designed for testing relays and transducers

T 1000 PLUS. Secondary Injection Relay Test Set. Designed for testing relays and transducers Secondary Injection Relay Test Set Designed for testing relays and transducers Microprocessor controlled With phase angle shifter Frequency generator Test results and settings are saved into local memory

More information

C.A 8332 C.A Display your network analysis all in picture format! OBSERVE, DIAGNOSE, MONITOR... Three Phase Power Quality Analysers

C.A 8332 C.A Display your network analysis all in picture format! OBSERVE, DIAGNOSE, MONITOR... Three Phase Power Quality Analysers C.A 8332 C.A 8334 Display your network analysis all in picture format! OBSERVE, DIAGNOSE, MONITOR... Three Phase Power Quality Analysers Site ALSTOM POWER Service - Massy (France) Ready-to-use measurements

More information

Power Quality Analyzer / High Frequency Recorder

Power Quality Analyzer / High Frequency Recorder Technical data Power Quality nalyzer / High Frequency Recorder Model PQ-Box 300 1 Fault detection 1 Evaluation of voltage quality according to EN50160 und IEC61000-2-2/-2-4 1 Permanent FFT nalysis from

More information

Panther. Installation instructions. Receivers PN-RX-MN5 (PN-R8-1) PN-RX-MN5E (PN-R8-10) PN-RX-MD11 (PN-R8-6) PN-RX-MD11E (PN-R8-11)

Panther. Installation instructions. Receivers PN-RX-MN5 (PN-R8-1) PN-RX-MN5E (PN-R8-10) PN-RX-MD11 (PN-R8-6) PN-RX-MD11E (PN-R8-11) Provided by: www.hoistsdirect.com Panther Installation instructions Receivers PN-RX-MN5 (PN-R8-1) PN-RX-MN5E (PN-R8-10) PN-RX-MD11 (PN-R8-6) PN-RX-MD11E (PN-R8-11) IM-PN-RX101-A01-EN Language: English

More information

1 Safety instructions. 2 Device components. Radio bus system. Radio socket outlet adapter for dimming Order No. : Operating instructions

1 Safety instructions. 2 Device components. Radio bus system. Radio socket outlet adapter for dimming Order No. : Operating instructions Order No. : 1185.. Operating instructions 1 Safety instructions Electrical equipment may only be installed and fitted by electrically skilled persons. Serious injuries, fire or property damage possible.

More information

WHALETEQ. ESU Neutral Electrodes Impedance Tester. Model: HFPA150. User Manual

WHALETEQ. ESU Neutral Electrodes Impedance Tester. Model: HFPA150. User Manual WHALETEQ ESU Neutral Electrodes Impedance Tester Model: HFPA150 User Manual Version2014-10-30 Hardware Version1.3.x 1. Introduction HFPA150 is a unique tester specifically designed for the testing of neutral

More information

DEIF A/S. TAS-311DG Selectable AC transducer J (UK) Installation and start up instructions

DEIF A/S. TAS-311DG Selectable AC transducer J (UK) Installation and start up instructions Installation and start up instructions TAS-311DG Selectable AC transducer Voltage, current, frequency or phase angle transducer Supply voltage up to 690V DEIF A/S Configuration via PC-interface possible

More information

Always there to help you. Register your product and get support at AJ3200. Question? Contact Philips.

Always there to help you. Register your product and get support at  AJ3200. Question? Contact Philips. Always there to help you Register your product and get support at www.philips.com/support Question? Contact Philips AJ3200 User manual Contents 1 Important 2 2 Your clock radio 2 Introduction 2 What's

More information

400Amp True RMS AC Clamp Meter + NCV

400Amp True RMS AC Clamp Meter + NCV User Guide 400Amp True RMS AC Clamp Meter + NCV Model MA410T Introduction Thank you for selecting the Extech MA410 Clamp Meter. This meter measures AC Current, AC/DC Voltage, Resistance, Capacitance, Frequency,

More information

Electronic Heat Cost Allocator GRADUS User Manual

Electronic Heat Cost Allocator GRADUS User Manual Electronic Heat Cost Allocator GADUS User Manual ev. 1 02/15 Table of Contents 1 APPLICATION AND FUNCTION... 3 2 CONTENTS OF PACKAGE... 3 3 GENEAL INFOMATION... 3 4 GENEAL MOUNTING INFOMATION... 3 5 DISPLAY

More information

RF converter WTX660-E05060

RF converter WTX660-E05060 RF converter WTX660-E05060 The WTX660.. RF converter receives and processes data from heat cost allocators, heat and water meters within a M-bus radio system. Storage for up to 500 metering devices Repeater

More information

AC-SENSOR 63 UND 250 AC-SENSOR 63 AND 250

AC-SENSOR 63 UND 250 AC-SENSOR 63 AND 250 AC-SENSOR INSTALLATIONSANLEITUNG INSTALLATION GUIDE DE EN AC-SENSOR 63 UND 250 INSTALLATIONSANLEITUNG AC-SENSOR 63 AND 250 INSTALLATION GUIDE 1 Scope of delivery Check the delivery for completeness and

More information

Fluke 1740 Series. Three-Phase Power Quality Loggers Memobox. Assess power quality and conduct long-term studies with ease

Fluke 1740 Series. Three-Phase Power Quality Loggers Memobox. Assess power quality and conduct long-term studies with ease Fluke 1740 Series Three-Phase Power Quality Loggers Memobox Technical Data Assess power quality and conduct long-term studies with ease Compact and rugged, the Fluke 1740 Series three-phase power quality

More information

User s Guide. RP7000 Series Active Probe. Dec RIGOL Technologies, Inc.

User s Guide. RP7000 Series Active Probe. Dec RIGOL Technologies, Inc. User s Guide RP7000 Series Active Probe Dec. 2012 RIGOL Technologies, Inc. Guaranty and Declaration Copyright 2011 RIGOL Technologies, Inc. All Rights Reserved. Trademark Information RIGOL is a registered

More information

ENA33LCD. Power line analyzer. User and service manual. Obrezija 5 SI-1411 Izlake

ENA33LCD. Power line analyzer. User and service manual.   Obrezija 5 SI-1411 Izlake ENA33LCD Power line analyzer User and service manual version 2.9 (FW version 6.8 and newer) ETI, d.o.o. Obrezija 5 SI-1411 Izlake www.etigroup.eu/products-services 1. Front control panel and terminal plate

More information

Wireless Receiver E28Q Mounting and Operating Instructions (Original operating instructions)

Wireless Receiver E28Q Mounting and Operating Instructions (Original operating instructions) EN Wireless Receiver E28Q Wireless Receiver E28Q Mounting and Operating Instructions (Original operating instructions) Always read before initial operation! 28509900_E28Q_EN_2012-05-08.doc 1 / 12 1 Data

More information

UNIVERSAL-DDS-VFO UDV ( 1 Hz to 10 MHz)

UNIVERSAL-DDS-VFO UDV ( 1 Hz to 10 MHz) UNIVERSAL-DDS-VFO UDV ( 1 Hz to 10 MHz) Connection and operating instructions 1. Introduction The UDV is the ideal device to adapt older, VFO-controlled transceivers to modern requirements regarding frequency

More information

Technical specification FLUXUS F401 H2O

Technical specification FLUXUS F401 H2O Technical specification Ultrasonic flowmeter for water Portable, very robust and easy-to-use ultrasonic flowmeter for the water and wastewater industry Features Very high bi-directional measuring accuracy

More information

OPERATION & SERVICE MANUAL FOR FC 110 AC POWER SOURCE

OPERATION & SERVICE MANUAL FOR FC 110 AC POWER SOURCE OPERATION & SERVICE MANUAL FOR FC 100 SERIES AC POWER SOURCE FC 110 AC POWER SOURCE VERSION 1.3, April 2001. copyright reserved. DWG No. FC00001 TABLE OF CONTENTS CHAPTER 1 INTRODUCTION... 1 1.1 GENERAL...

More information

INSTRUCTIONS FOR QUALIFIED PERSONNEL ONLY!

INSTRUCTIONS FOR QUALIFIED PERSONNEL ONLY! KMT - Kraus Messtechnik GmbH Gewerbering 9, D-83624 Otterfing, Germany, 08024-48737, Fax. 08024-5532 Home Page: http://www.kmt-telemetry.com, Email: info@kmt-telemetry.com CT2-Mini User manual INSTRUCTIONS

More information