8730 Electric Parameter Meter User manual
|
|
- Donald Walters
- 5 years ago
- Views:
Transcription
1 8730 Electric Parameter Meter User manual ( Ver 2.10 ) Qingdao QINGZHI Instruments Co., Ltd. Address: 5th Floor, Building 1, Shenghe Building, No.58 Shandongtou Road, Laoshan District, Qingdao, China, Post code Telephone/ Fax (Multi-lines), (Multi-lines) Technical phone: (0) Website: 1
2 Chapter 1 Features 1. Test accuracy Test condition: sine wave, line filter and frequency filter should be closed. Table 1 Parameter Range Fundamental error Remark Voltage 15/30/60/150 /300/600 Current 0.5/1/2/5/ 10/20 Active power U*I Power factor 0.1 ~ 1 DC ±(0.16%rdg+0.04%rng) 0.5 Hz F <100Hz ±(0.16%rdg+0.04%rng) 100 Hz F <1kHz ±(0.2%rdg+0.05%rng) 1k Hz F <10kHz ±(0.28%rdg+0.07%rng) 10k Hz F <100kHz ±(( 0.1%+ F/10k*0.1%)rdg+0.15%rng) DC ±(0.16%rdg+0.04%rng) 0.5 Hz F <100Hz ±(0.16%rdg+0.04%rng) 100 Hz F <1kHz ±(0.2%rdg+0.05%rng) 1k Hz F <10kHz ±(0.28%rdg+0.07%rng) 10k Hz F <100kHz ±(( 0.1%+F/10k*0.1%)rdg+0.2 %rng) 0.5 Hz F <100Hz ± Hz F <1kHz ± k Hz F <10kHz ± k Hz F <100kHz ± 0.1 Overload 1.2 times PF=1.0 Frequency Energy cumulate 0.5 ~ 100kHz 0 ~ MWh Counting way: 0.1%*rdg (when signal value is more than10% nowaday range) DC ±(0.16%rdg+0.04%rng) 0.5 Hz F <100Hz ±(0.16%rdg+0.04%rng) 100 Hz F <1kHz ±(0.2%rdg+0.05%rng) 1k Hz F <10kHz ±(0.28%rdg+0.07%rng) 10k Hz F <100kHz ±(( 0.1%+F/10k*0.1%)rdg+0.2 % rng) Energy timing 999h.59m.59s ± 2seconds/hour Harmonic 2. Other parameters Input way: 40Hz-70Hz: test 1~99 orders harmonic, sample 4 cycles each time 70Hz-130Hz: test 1~99 orders harmonic, sample 4 cycles each time 130Hz-250Hz: test 1~99 orders harmonic, sample 4 cycles each time 250Hz-440Hz: test 1~99 orders harmonic, sample 4 cycles each time Accuracy grade: B Voltage and current are both floating input. Voltage input impedance is about 660kΩ and current input impedance is about 4mΩ.External shunt signal input impedance is about 100kΩ. Test signal peak value: voltage and current are both 1.6 times of range. 2
3 A/D transition: speed orders/second, 16 bits. Voltage and current sample meanwhile. Harmonic analyse: 1024 points, floating point, rectangular window. Calibration zero way: automatic. It calibrates nowaday range zero every 30 minutes when energy cumulate and every10 minutes when others. It takes about 0.1 second every time. Line filter: cut-off frequency is 600Hz. After opening Line filter, input signal frequency range which can be ensured is 45-65Hz. The other frequency accuracy can not be ensured. Display update: DC: about 3 times/second. Virtual value: about 3 times/second to 1 time/ 3 seconds. Display update period increases as input signal frequency reduces. Power consumption: < 10VA. Weight: about 6.5 kg. Dimension: 400 x130 x380 mm (width x height x depth). 3. Working condition Air pressure: (86~106) kpa Temperature:(0~40) Relative humidity: 85%RH Power supply:ac: ( ) V 50/60Hz 4. Secure need Insulation resistance: Between input terminals to case, between power supply to case, between power supply to input terminals. Above: 2MΩ Withstanding voltage: Between input terminals to case, between power supply to case, between power supply to input terminals. Endure: AC 2000V frequency 50Hz sine wave voltage. *Complying standard: GB/T
4 Ⅰ.8730 Front panel operation Chapter 2 Instruction 9 10 Instruction: Figure 1 1. State indicator light 2. Range transition 3. Test mode 4. Harmonic 5. Energy 6. Basic function 7. System parameter 8. Power supply switch 9. Display window 10. Test parameter 1. Display window Three windows can display following parameters. (Table 2) Order Indicator light Table 2 Parameter Unit Order Indicator light Parameter 1 V Voltage volt 7 Var Reactive power var 2 A Current amp 8 DEG Degree degree 3 W Active power watt 9 h Combination use 4 Hz Frequency Hz 10 % Relative harmonic content 5 PF Power factor 11 Time Energy cumulate 6 VA Apparent power volt ampere Wh time Remark 1: m, k, M transition: 1M = 1000k; 1k = 1000; 1m = Remark 2: Harmonic analyse instruction of voltage and current: Unit Wh hh:mm:ss It can have harmonic analyse when fundamental wave frequency is within Hz range. a)40-70hz: test 99 orders harmonic, sample 4 cycles each time. 4
5 b)70hz-130hz: test 1~50 orders harmonic, sample 8cycles each time. c)130hz-250hz: test 1~30 orders harmonic, sample 16 cycles each time. d)250hz-440hz: test 1~10 orders harmonic, sample 32 cycles each time. Note: It can not have harmonic analyse when frequency is without up range. Displaying 0 order harmonic is total harmonic. Remark 3: Display particular character meaning: A-LOW: Prompt nowaday current input signal is too small to test current frequency. U-LOW: Prompt nowaday voltage input signal is too small to test voltage frequency. -OL-: Prompt nowaday display data exceeds display range. U-OL: Prompt nowaday window test value is unknowable because input voltage exceeds range. A-OL: Prompt nowaday window test value is unknowable because input current exceeds range. UF-ERR: In harmonic state: prompt it can not have harmonic analyse because nowaday frequency of input voltage signal is out of range. Other state: prompt it can not test nowaday voltage frequency or frequency< 0.5Hz. AF-ERR: In harmonic state: prompt it can not have harmonic analyse because nowaday frequency of input current signal is out of range. Other state: prompt it can not test nowaday current frequency or frequency< 0.5Hz. UF-OL: Prompt frequency of nowaday voltage input signal exceeds 120 khz. AF-OL: Prompt frequency of nowaday voltage input signal exceeds 120 khz. ZERO: Prompt calibration zero. It calibrates nowaday range zero point every 10 minutes if meter is in a steady range. ----: Prompt to change range. It need wait circuit steady. ORXX: Display is at window A. Prompt nowaday harmonic orders are XX. 2. State indicator light Table 3 Order Indicator light Instruction Remark 1 SAMPLE Working state indicator light It flickers continually when working normally. 2 V OVER Voltage over range Input voltage exceeds nowaday range 3 A OVER Current over range Input current exceeds nowaday range 4 A EXT External shunt Start external shunt 5 COMM Communication It flickers continually when working normally. 5
6 3. Function key Range transition key: there are 6 keys to transit range of voltage and current. Automatic key: transit range automatically. It is effective when indicator light lightens. Repeating operation can relieve automatic transition. Increase key: increase range manually. Range indicator light moves right one bit when operates once. Decrease key: decrease range manually. Range indicator light moves right one bit when operates once. Note: 1. Current range is immutable when there is external shunt input. 2. In energy cumulate state, current and voltage range can be set but not be changed. 3. In harmonic state, range is alterable. 3.2 Test mode key: there are 4 test mode keys for different test need. It is effective when indicator light lightens. RMS virtual value: True Root Mean Square mode. Apply AC & DC. It is effective when indicator light lightens. DC direct current: DC mode. Apply DC. Use mean value. Switch: from DC to virtual value. If input signal is 0 or frequency is too low, the first display update period will be more than 3 seconds. For meter need judge if input signal frequency is less than 0.5 Hz. It is normal. Baud rate : display data includes test data multiplication by baud rate. Apply elementary data of instrument transformer. It is effective when indicator light lightens. Frequency filter : start lowpass filter (cut-off frequency is 600Hz). Only apply frequency signal of voltage and current. Line filter : start lowpass filter (cut-off frequency is 600Hz). Apply all signals except frequency signal of voltage and current. 3.3 Harmonic key: there are 2 keys. Apply fundamental phase setup of harmonic test function and enter / exit harmonic test state. Setup : build fundamental phase of harmonic analyse. In combination with system parameter key (Refer to Table 4 & Table 6). 6
7 Table 4 Harmonic setup Window A Window B Window C Instruction Har00 code Code:1234 Enter setup code(code= 1234 ) Har01 SynC Fundamental phase of harmonic analyse --U- in step with voltage signal, --A- in step with current signal Analyse : enter / exit harmonic analyse state. It is effective when indicator light lightens. Repeating operation can exit. Note: Harmonic analyse and energy cumulate can not work meanwhile. 3.4 Energy key: there are 4 keys. Apply energy cumulate operation. Start : start energy cumulate. Start energy cumulate of nowaday test item. It is effective when indicator light lightens. Stop : stop energy cumulate. Stop energy cumulate of nowaday test item. It is effective when indicator light lightens. Setup : set time need of automatic stop energy cumulate. If the time is 0, it must make manual stop or energy cumulate to Wh. In combination with system parameter key (Refer to Table 5 & Table 6). Reset : clear all energy cumulate data. It is effective after energy cumulate stops. Table 5 Energy time setup Window A Window B Window C Instruction int00 code Code enter value Enter setup code(code= 1234 ). If code is wrong, it can only check the data but not set it. int00 code Err Appears only when code is wrong. int01 Hour Hour 1. Time need of energy cumulate. int02 Minute Minute 2. Max: 999 hours 59 minutes 59 seconds. 3. Energy cumulate stops automatically when reaching int03 Second Second setup time. int04 U_rng Voltage range int05 A_rng Current range Set up voltage range of energy cumulate. It can not transit range automatically in energy cumulate state. Set up current range of energy cumulate. It can not transit range automatically in energy cumulate state. 3.5 Basic function key: there are 2 keys. Apply meter basic function operation. Print : the key function is invalid. Lock : lock communication data and display data. In system setup state, press the key can exit setup quickly and not save setup data. 7
8 3.6 System parameter key: there are 6 keys. Apply meter system basic parameter setup. Setup : enter parameter setup. After enter, window A displays SYSXX ; window B displays nowaday parameter; window C displays nowaday parameter value. Repeating operation to choose setup parameter. (Table 6) key: change decimal point place of setup value. key: circular right shift to change nowaday nixie tube (flicker point). < key: circular left shift left to change nowaday nixie tube (flicker point). key: increase flicker point value circularly. key: decrease flicker point value circularly. Table 6 System parameter setup Key Window A Window B Window C Instruction Setup SYS00 code Code enter value Enter setup code(code= 1234 ) If code is wrong, it can only check the data but not setup data. Setup SYS00 code Err Appears only when code is wrong. Setup SYS 01 UR Voltage ratio Default: 1 Setup SYS 02 I-AR Internal shunt current ratio Default: 1 Setup SYS 03 E-AR External shunt current ratio Default: 1 inner use internal shunt Setup SYS 04 Shunt Voltage ratio outer use external shunt (Press to choose) Default: inner Setup SYS 05 updat Display update speed 1-6 default: 1, set data update speed of voltage, current and power. Setup SYS 06 addr Setup SYS 07 bps Communication address Communication baud rate Default: Default: 9600 Special instruction: current access choice and use. Inner internal shunt ---test current with internal shunt. Set current ratio I-AR ---access CT transformation ratio. Example: when use 200A/5A CT: 200/5= outer internal shunt ---test current with external shunt. (Note terminal) Set current ratio E-AR ---access CT transformation ratio. Example: when use 150A/75mV E-AR=150/75=
9 Please note connection length and screen when using shunt because voltage signal of shunt is too low and easy to be disturbed. Input signal of shunt can not exceed 75mV, or A-OVER indicator light lightens and current displays A-OL. Display update speed updat range is 1-6. It uses root mean square to virtual value U & I. It uses arithmetic mean to direct voltage, direct current and power. Example: Meter samples every 0.3 second (50Hz) to get a series of virtual value data. As follows: U0,U1,U2,U3,U4,U5,U6,U7,U8 Un, so the nth display value Meter samples every 0.3 second (50Hz) to get a series of direct voltage data. As follows: U0,U1,U2,U3,U4,U5,U6,U7,U8 Un, so the nth display value Reactive power and apparent power of the nth display are both educe value of nowaday voltage, current and power. If input signal flops strongly, you can decrease display speed value to make it steady. Setup example: set input code 1)Operate system parameter setup key to confirm that Window A displays SYS00, Window B displays code. 2)Operate <, > key to make Window C to leftmost flicker. 3)Operate key once to make flicker place display number 1. 4)Operate >, key once to make flicker place display number 2. 5)Operate repeatedly 2)3)4)to enter code Meter manufactory setup information 1)System parameter setup refers to default value. 2)Meter national configuration of communication serial port is RS232C. RS485 is as selective expansion item. Ⅱ.8730 Back panel operation 9
10 Instruction: Figure 2 Back panel 1. Voltage connection terminal 2. Current connection terminal 3. Shunt connection terminal 4. Power supply socket 5. Communication serial port 6. Expansion interface Note 1. Back panel includes power supply socket, connection terminal, communication serial port and expansion interface. 2. Power supply socket is for power input. The socket down must put fuse and specification is 250V Connection terminal is for connecting test loop. For different test ways, the meter external way of connection terminal are different. Ⅲ. Connection for DC & power frequency AC signal Note 1. The meter can test AC & DC signal. 2. Test signal must keep terminal consistency. 3. Current terminal and shun can not access meanwhile. 4. Access signal can not exceed range. 5. Practical connection refers to back appearance please. Figure 3 Load connection Ⅳ. Verification connection way Note 1. Verify AC & DC signal respectively. 2. Test signal must keep terminal consistency. 3. Current terminal and shun can not access meanwhile. 4. Access signal can not exceed range. 5. Practical connection refers to Figure 4 please. Figure 4 Verification connection 10
11 Chapter 3 Serial port instruction Ⅰ. Serial port instruction 1. There are RS232 and RS485 two choice. Hardware interface of serial port is 9 needle D type socket. 2. Serial interface pin is identified as: RS232: 2---RXD 3---TXD 5---GND; RS485: 1---A 4---B. The word near interface indicates serial terminal type. RS232 or RS485 pane means effective serial terminal. 3. When connecting mainframe to meter with serial cable, turn off the power for meter and mainframe or it may destroy the meter. Ⅱ. Communication test procedure instruction 1. Read README.TXT file carefully to find right communication test procedure. (8730 file) 2. Copy 8730 file to target HD. Open the file and double click 8730.exe to enter communication procedure operation interface. 3. Choose computer communication port. Make sure communication address & baud rate are the same with product setup. 4. Click get to get test data. Ⅲ. Serial port communication failure check 1. Check if communication address & baud rate are the same with upper monitor setup; if different, change it. 2. Disconnect meter and upper monitor to test meter and upper monitor signal wire. For RS232 port: TXD to GND port voltage of meter and upper monitor should be -8V~-12V. For RS485 port: A to B port voltage of upper monitor should be +2V~+5V. If the up test signal is not normal, there is error of interface or hardware connection. 3. Serial interface communication can receive data but there are errors often. Please examine if the communication connection of meter and upper monitor is OK. Serial interface wire should use shielding conductor and connect shielding layer with ground. 11
12 Chapter 4 Meter packing list Order Name Quantity Remark 1 Electric parameter meter 1 2 Power supply wire 1 3 Power fuse 2 0.5A 4 User s manual 1 5 Qualification 1 6 Measurement report 1 7 Warranty list 1 8 Feedback form of unpack check-up 1 9 Communication software CD 1 Chapter 5 Use note and fault clearance Ⅰ. Use note 1. The meter enclosure must earth well. 2. The analyzer should work in recommendatory environment. 3. The load power supply should be turned off when wiring load port. 4. Don t exceed the measurement limit. 5. The meter should electrify preheating 30 minutes. 6. Please avoid other disturb when using shun input current. 7. External shunt rated input 75 mv. Prohibit inputting crest value > 1V, or may destroy the meter. Ⅱ. Fault clearance 1. There is no display when turning on and power supply indicator light does not lighten. Check if power supply connects; voltage is normal; fuse is OK. 2. Test data has obvious deviation or power display negative value. Check if connection terminal is OK. Note voltage and current are in phase terminal. 12
13 3. Way to replace fuse Fuse is under power supply electrical outlet. Please put out plug when change. 2. Pry fuse outlet from power supply electrical outlet with screwdriver. 3. Put out the fuse from outlet when fuse shows. 4. Change new fuse and push fuse outlet back to power supply electrical outlet by hand. 13
MS8268 HANDHELD DIGITAL MULTIMETER OPERATOR S INSTRUCTION MANUAL
MS8268 HANDHELD DIGITAL MULTIMETER OPERATOR S INSTRUCTION MANUAL Table of Contents TITLE PAGE 1. GENERAL INSTRUCTIONS 1 1.1 Precaution safety measures 1 1.1.1 Preliminary 1 1.1.2 During use 2 1.1.3 Symbols
More informationZHUHAI PILOT TECHNOLOGY CO., LTD.
SPM32 Multifunctional Power Meter SPM32 Multifunction Power Meter Installation & Operation Manual V1.0 ZHUHAI PILOT TECHNOLOGY CO., LTD. Danger and warning! This device can be installed only by professionals.
More informationPower analyzers and Energy Meters Power Analyzer Type WM14-96
Power analyzers and Energy Meters Power Analyzer Type WM14-96 Optional RS422/485 serial port Alarms (visual only) V LN, An Class 2 (active energy) Class 3 (reactive energy) Accuracy ±0.5 F.S. (current/voltage)
More informationPM6000. Voltech. Power Analyzer
Voltech PM6000 Power Analyzer A New Standard in Power Analysis Up to 6 wattmeter channels Basic accuracy: 0.02% of reading Bandwidth: 10MHz Easy-to-use in all applications The Voltech PM6000 Power Analyzer
More informationPower analyzers and Energy Meters Power Analyzer Type WM14-DIN
Power analyzers and Energy Meters Power Analyzer Type WM14-DIN Optional RS422/485 serial port Alarms (visual only) V LN, An Class 2 (active energy) Class 3 (reactive energy) Accuracy ±0.5 F.S. (current/voltage)
More informationETHOS 5030 TRUE-RMS DIGITAL MULTIMETER OPERATION MANUAL
ETHOS 5030 TRUE-RMS DIGITAL MULTIMETER OPERATION MANUAL 1 1. SAFETY INFORMATION SAFETY SYMBOLS Warning! Dangerous Voltage (Risk of electric shock). Caution! Refer to the user s manual before using this
More informationPage 1 of IR Port 5 External power port 6 Calibration button port 7 DUT + jack 8 DUT - jack 9 Shielding grounding jack
1. General instructions: Thank you for purchasing a MS5308 LCR digital bridge meter. The MS5308 LCR digital bridge meter is a professional instrument for measuring inductance, capacitance and resistance.
More informationDECLARATION OF CONFORMITY
DECLARATION OF CONFORMITY Manufacturer's Name: Transmille Ltd. Manufacturer's Address: Unit 4, Select Business Centre Lodge Road Staplehurst TN12 0QW. United Kingdom. Declares, that the product Product
More informationMODEL 9041 EXTENDED SPECIFICATIONS. 25ppm TRANSPORTABLE CALIBRATOR
MODEL 9041 EXTENDED SPECIFICATIONS 25ppm TRANSPORTABLE CALIBRATOR 9041 EXTENDED SPECIFICATIONS General Specifications TRANSMILLE LTD Warm Up Time Double the time since last used up to 20 minutes maximum
More informationPEN TYPE DIGITAL MULTIMETER OPERATION MANUAL T8211D
PEN TYPE DIGITAL MULTIMETER OPERATION MANUAL T8211D T8211D 1 1. SAFETY INFORMATION BE EXTREMELY CAREFUL IN THE USE OF THIS METER. Improper use of this device can result in electric shock or destroy of
More informationEnergy Management Modular Power Analyzers Type WM2-96
ISE, Inc. - 10100 Royalton Rd. - Cleveland, OH 44133 USA Energy Management Modular Power Analyzers Type WM2-96 - Tel: (440) 237-3200 - Fax: (440) 237-1744 - http://instserv.com Class 1 Modular power analyzer
More informationMCTS II Multi Channel Current Transducer System. Installation Manual
MCTS II Multi Channel Current Transducer System Installation Manual MA MCTS II, Ver. 1.0 1 10/2017 Contents 1. Introduction 2. Receiving of goods 3. Hardware Installation 3.1. Connection of the current
More informationMultimeter 500CVD21 RTU500 series
Remote Terminal Units - Data sheet Multimeter 500CVD21 RTU500 series CT/VT interface with 4 voltage and 24 current inputs for direct monitoring of 3/4 wire 0 300 V AC (line to earth), 0...500 V AC (phase
More information770E DIGITAL MULTIMETER OPERATOR S MANUAL
770E DIGITAL MULTIMETER OPERATOR S MANUAL TABLE OF CONTENTS 1.Overview... 1 2.Panel Layout...1/2 3. Safety information... 2 4. Specical cautions for operation.... 2 5.General specifications...2/3 6.Testing
More informationOPERATION MANUAL OF MODEL TH2681/A HIGH RESISTANCE TESTER
OPERATION MANUAL OF MODEL TH2681/A HIGH RESISTANCE TESTER Changzhou Tonghui Electronic Co., Ltd Addr: No.3 Tianshan Rd., New District, Changzhou Zip: 213022 Tel: 86-519-5132222 Fax: 86-519-59972 Http://
More informationEnergy Management Modular Smart Power Transducer Type SPT-90
Energy Management Modular Smart Power Transducer Type SPT-90 Class 0.5 (current/voltage) 16-bit µp-based modular smart power transducer Measurements of: W, Wavg, VA, VAr, PF, Wh, VAh, VArh, Amax (among
More informationModel PA3000 Power Analyzer
Model PA3000 Power Analyzer INSTRUMENT SPECIFICATIONS SPECIFICATION CONDITIONS This document contains specifications and supplemental information for the Model PA3000 Power Analyzer. Specifications are
More informationBENNING. CM7 Electrical Tester. Service Information
BENNING CM7 Electrical Tester Service Information TABLE OF CONTENTS Page Introduction 1 Precautions and Safety Information 1 Symbols 1 Safety 2 Specifications 3 General Specification 3 Voltage Specification
More informationPower analyzers and Energy Meters. Multifunction indicator. Type WM12-DIN. Product Description. How to order WM12-DIN AV5 3 D X.
Power analyzers and Energy Meters Multifunction indicator Type WM12-DIN Accuracy ±0.5 F.S. (current/voltage) Multifunction indicator Display of instantaneous variables: 3x3 digit Variable system and phase
More informationGSC57 Rel of 257/10/05
Integrated meter for test verify and Power Quality analysis Pag 1 of 7 1. ELECTRICAL SPECIFICATIONS VERIFY TESTS Accuracy is indicated as ± (% readings + no. of digits) at 23 C ± 5 C, con relative humidity
More informationMeters. Ampere Meter. Volt Meter. Watt Meter. Var. Meter. Power Factor Meter. Frequency Meter. Demand Meter. Meter Relay.
Meters Ampere Meter Volt Meter Watt Meter Var. Meter Power Factor Meter Frequency Meter Outside Connection (Watt, PF, Var) 3 2 7 Receive Indicator Meter Demand Meter Meter Relay Meter Dimensions Meter
More informationFor ultra-high precision measurement of current: DC, AC, pulsed..., with galvanic separation between primary and secondary. Applications.
Current Transducer IT 605-S ULTRASTAB I PN = 600 A For ultra-high precision measurement of current: DC, AC, pulsed..., with galvanic separation between primary and secondary. Features Wide operating temperature
More informationVFSC9 ELECTRONIC SPEED CONTROLLER. Mounting and operating instructions
ELECTRONIC SPEED CONTROLLER Mounting and operating instructions Table of contents SAFETY AND PRECAUTIONS 3 PRODUCT DESCRIPTION 4 ARTICLE CODES 4 INTENDED AREA OF USE 4 TECHNICAL DATA 4 STANDARDS 5 WIRING
More informationEnergy Management Energy Meter GM3T E GM3T-RS485 E
Energy Management Energy Meter GM3T E09 814 96 GM3T-RS485 E09 816 65 Certified according to MID Directive Product Description Three-phase energy meter with removable front LCD display unit. The same unit
More informationWT210/WT at rated. 5 ma range. input Low current (WT210) DC 0.5 Hz. to100 khz. Digital Power Meters ±0.2% WT210/WT230 Digital Power Meters
WT2/WT23 Digital Power Meters WT2/WT23 Digital Power Meters Digital Sampling Power Meters with Superior Cost Performance Low power measurement (IEC623) Harmonic measurement Compact design (half-rack size)
More informationTPS TRANSDUCER POWER SUPPLY & LEM IT CURRNET TRANSDUCER
INSTALLATION MANUAL TPS TRANSDUCER POWER SUPPLY & LEM IT CURRNET TRANSDUCER Version 06/2010 ENG CONTENTS 1. INTRODUCTION 1.1 MAXIMUM RANGES 1.2 MINIMUM RANGES 2. RECEIVING OF GOODS 3. HARDWARE INSTALLATION
More informationMODEL 9050 EXTENDED SPECIFICATIONS. 50ppm TRANSPORTABLE CALIBRATOR
MODEL 9050 EXTENDED SPECIFICATIONS Ü 50ppm TRANSPORTABLE CALIBRATOR 9050A EXTENDED SPECIFICATIONS General Specifications TRANSMILLE LTD Warm Up Time Double the time since last used up to 20 minutes maximum
More informationENERGY SAVING TOOLS. Digital Sampling Power Meters with Superior Cost Performance. Digital Power Meters
Established 98 Advanced Test Equipment Rentals www.atecorp.com 8-44-ATEC (2832) ENERGY SAVING TOOLS Digital Sampling Power Meters with Superior Cost Performance Digital Power Meters WT2/WT23 Basic power
More informationAPPLICATIONS. Oscilloscope Display. Harmonic Measurement. High Precision Measurement up to 0.1% Power Supplies Tests Motor Tests
YOUR POWER TESTING SOLUTION IT9100 Power Meter Oscilloscope Display Harmonic Measurement High Precision Measurement up to 0.1% IT9100 Power Meter APPLICATIONS Power Supplies Tests Motor Tests Household
More informationThank you again for choosing AstroAI, if you have any questions or concerns regarding your product, please contact us at
ASTROAI USER MANUAL DT132A 4000 Count Auto-Ranging Multimeter Thank you for purchasing the AstroAI DT132A 4000 Count Auto-Ranging Multimeter. It is a 3 ¾ digit, 3999 counts, auto-ranging digital multimeter.
More informationMT /6 Smart Digital Clamp Meter. User s Manual 1 st Edition, 2016 Copyright by Prokit s Industries Co., Ltd.
MT-3110 3 5/6 Smart Digital Clamp Meter User s Manual 1 st Edition, 2016 Copyright by Prokit s Industries Co., Ltd. 1.Safety Information Warnings Special attention shall be paid when using the meter, improper
More informationMultiCube Multi-Function Electricity Meter. Installation and Operation
MultiCube Multi-Function Electricity Meter Installation and Operation PREFACE MultiCube Operating Guide Revision 1.08 Jan 2007 This manual represents your meter as manufactured at the time of publication.
More informationDVM1190 DIGITAL MULTIMETER
DIGITAL MULTIMETER 1. Introduction Thank you for buying the. This digital multimeter has a large LCD, a data-hold function and a backlight. The device uses a very practical safety mechanism that keeps
More informationEnergy Management Energy Meter Type EM21 72R Retro-Fit
Energy Management Energy Meter Type EM21 72R Retro-Fit Including 3 miniature split-core current sensors 10mm (90A), 16mm (150A) and 24mm (250A) diameter holes Class A (kwh) according to EN50470-3 Class
More informationEnergy Management Energy meters Type EM2-DIN
Energy Management Energy meters Type 6-dgt µp-based indicator Manual scrolling of partial and total energies: kwh, kvarh. TRMS measurement of distorted waves (voltage/current) All configuration functions
More informationCONTENTS. SAFETY PRECAUTIONS: Before use, read the following safety precautions
CONTENTS [1] SAFETY PRECAUTIONS: Before use, read the following safety precautions 2 [2] APPLICATION AND FEATURES 5 [3] NAME OF COMPONENT UNITS 6 [4] DESCRIPTION OF FUNCTIONS 8 [5] MEASUREMENT PROCEDURE
More informationEnergy Management. Type WM14-96 Basic Version. Product Description. How to order WM14-96 AV5 3 D PGX. Type Selection. Input specifications.
Energy Management Power Analyzer Type WM-96 Basic Version Optional dual pulse output Alarms (visual only) V, An Optional galvanically insulated measuring inputs Product Description 3-phase power analyzer
More informationIC800SSD Hardware Manual Pub 348R5. for models. A publication of
IC800SSD Hardware Manual Pub 348R5 for models IC800SSD104S1A IC800SSD104RS1A IC800SSD107S1A IC800SSD107RS1A IC800SSD407RS1A IC800SSD216S1A IC800SSD216RS1A IC800SSD420RS1A IC800SSD228S1A IC800SSD228RS1A
More informationWatt Meter - Model 2335
MORE WATT METERS Watt Meter - Model 2335 Dc to more than 1MHz Inputs up to 10A and 1000V 0.2% Uncertainty Low Power Factor Accuracy Peak or RMS Readings Isolated Inputs Harmonic Analysis Multifunction
More informationFor ultra-high precision measurement of current: DC, AC, pulsed..., with galvanic separation between primary and secondary. Applications.
Current Transducer IT 205-S ULTRASTAB I PN = 200 A For ultra-high precision measurement of current: DC, AC, pulsed..., with galvanic separation between primary and secondary. Features Wide operating temperature
More informationFor ultra-high precision measurement of current: DC, AC, pulsed..., with galvanic separation between primary and secondary. Applications.
Current Transducer IT 700-SB ULTRASTAB I PM = 700 A For ultra-high precision measurement of current: DC, AC, pulsed..., with galvanic separation between primary and secondary. Features ± 10 V voltage output
More informationInstruction Sheet UPS SERIES. Serial Control Protocol. I Rev E
Instruction Sheet UPS SERIES Serial Control Protocol I-00341 Rev E (THIS PAGE INTENTIONALLY LEFT BLANK) Page 1 TABLE OF CONTENTS 1 Protocol Overview...3 1.1 Signal characteristics...3 1.2 Primary DB9 Pin
More informationRAIL MOUNTED NETWORK METER TYPE N27D USER S MANUAL
RAIL MOUNTED NETWORK METER TYPE N27D USER S MANUAL 1 Contents 1. Application... 5 2. meter set... 6 3. BASIC REQUIREMENTS, OPERATIONAL SAFETy... 6 4. installation... 10 4.1. Mounting... 10 4.2. External
More informationPower Quality Analyser MEMOBOX 300 smart
Power Quality Analyser MEMOBOX 300 smart The MEMOBOX 300 smart is the universal tool for power quality analysis, disturbance investigation and network optimisation on low and medium voltage power networks.
More informationDM-III MULTITEST Power Quality Recorder
DM-III MULTITEST Power Quality Recorder Amprobe s full-featured Three-Phase Power Quality Recorders provide the essential functions and capabilities required to operate accurately and effectively in today
More information3 PHASE DIGITAL POWER CLAMP METER USERS MANUAL
PHASE DIGITAL POWER CLAMP METER USERS MANUAL CONTENTS Safety Precautions - - - - - - - - - - - - - - - - - - - - - - - - - - - Safety Information - - - - - - - - - - - - - - - - - - - - - - - - - - - Safety
More informationKMD-S04 Multímetro de bolsillo
www.grupotemper.com KMD-S04 Multímetro de bolsillo Table of Contents Title Page Overview ~~~~~~~~~~~~~~~~~~~~~~~~~~~~~~~~~~~~~~~~~~ 3 Unpacking Inspection ~~~~~~~~~~~~~~~~~~~~~~~~~~~~~~~~~ 4 Safety Information
More informationTRANSMILLE 3010A PRECISION MULTIPRODUCT CALIBRATOR EXTENDED SPECIFICATIONS
TRANSMILLE 3010A PRECISION MULTIPRODUCT CALIBRATOR EXTENDED SPECIFICATIONS www. transmille.com 3010A EXTENDED SPECIFICATIONS General Specifications TRANSMILLE LTD Warm Up Time Double the time since last
More informationEASTRON SDM630MCT-RJV / SDM630MCT-RJA User Manual V1.1. Three phase multifunction din rail energy meter
SDM630MCT-RJV-333mV SDM630MCT-RJA-00mA Three phase multifunction din rail energy meter Plug-in solution; labor saving; wiring mistake free Measures kwh KVarh, KW, Kvar, KVA, P, F, PF, Hz, dmd, V, A, THD,etc.
More informationEnergy Management. Type WM12-DIN. Product Description. How to order WM12-DIN AV5 3 D X. Type Selection. Input specifications
Energy Management Multifunction indicator Type WM12-DI Accuracy ±0.5 F.S. (current/voltage) Multifunction indicator Display of instantaneous variables: 3x3 digit Variable system and phase measurements:
More informationDM-III Multitest & DM-III Multitest F Power Quality Recorders
DM-III Multitest & DM-III Multitest F Power Quality Recorders CAT 600V TRUE RMSz Amprobe s full-featured Three-Phase Power Quality Recorders provide the essential functions and capabilities required to
More informationDVM98. True RMS Digital Multimeter. 1 Safety information. 1.1 Preliminary. 1.2 During use
True RMS Digital Multimeter DVM98 1 Safety information This multimeter has been designed according to IEC - 1010 concerning electronic measuring instruments with an overvoltage category (CAT II) and pollution
More informationExternal RF Driver. Electrical Driver for AOTF-crystal. Instruction Manual
External RF Driver Electrical Driver for AOTF-crystal Instruction Manual Table of Contents 1 General... 3 2 Compliance... 3 3 Labeling... 4 4 Interface... 5 4.1 Front Panel... 5 4.2 Back Panel... 6 4.2.1
More informationSPECIFICATIONS Digital Indicator 1. General
SPECIFICATIS Digital Indicator 1. General The instrument is a digital indicator for the purpose of instrumentation, and suitable for the application in hopper, tank scale, and so on. 2. Specifications
More informationEnergy Management Energy Analyzer Type EM210
Energy Management Energy Analyzer Type EM210 Multi-use housing: for both DIN-rail and panel mounting applications Current inputs AV option: CT 5A Current inputs MV option: current sensor 333 mv (CTV series)
More informationTrue RMS Bench Multimeter 5492B
Data Sheet True RMS Bench Multimeter 5492B True RMS Bench Multimeter The B&K Precision model 5492B is a versatile 5½-digit, 120,000-count bench multimeter suitable for applications in education, service
More informationWide-band Models from DC to 100 MHz
CLAMP SENSOR Wide-band Models from DC to 1 MHz 3273-5 - 3276 CLAMP ON PROBES http://www.hioki.co.jp/ HIOKI company overview, new products, environmental considerations and other information are available
More informationInstallation and Operating Instructions Quadratic Integra 1530 Digital Metering Systems
Installation and Operating Instructions Quadratic Integra 1530 Digital Metering Systems Tyco Electronics UK Limited Crompton Instruments Freebournes Road, Witham, Essex, CM8 3AH, UK Tel: +44 1376 509 509
More informationM-140 Multifunction Calibrator. Operation manual
M-140 Multifunction Calibrator Operation manual MEATEST M-140 Multifunction Calibrator Content Operation manual... 1 Basic Information... 3 Preparation for operation... 5 Inspecting package contents,
More informationFluke 8845A/8846A Digital Multimeters Extended Specifications
Fluke 8845A/8846A Digital Multimeters Extended Specifications The Fluke 8845A and 8846A 6.5 digit precision multimeters have the precision and versatility to handle your most demanding measurements, on
More informationPQube 3 Specifications
PQube 3 Specifications Reference conditions for factory tests: 19~25 C, 10%~70% RH MAINS VOLTAGE MEASURING CHANNELS Measurement Channels Power Configuration/ Range of Nominal Input Voltage 3 Line to Neutral,
More informationRSMFX-2R MULTIFUNCTIONAL
Mounting and operating instructions Table of contents SAFETY AND PRECAUTIONS 3 PRODUCT DESCRIPTION 4 ARTICLE CODES 4 INTENDED AREA OF USE 4 TECHNICAL DATA 4 STANDARDS 4 OPERATIONAL DIAGRAMS 5 WIRING AND
More informationFor the electronic measurement of current: DC, AC, pulsed..., with galvanic separation between the primary and the secondary circuit.
Current Transducer IN 1000-S N = 1000 A For the electronic measurement of current: DC, AC, pulsed..., with galvanic separation between the primary and the secondary circuit. Features Closed loop (compensated)
More informationFluke 1740 Series. Three-Phase Power Quality Loggers Memobox. Assess power quality and conduct long-term studies with ease
Fluke 1740 Series Three-Phase Power Quality Loggers Memobox Technical Data Assess power quality and conduct long-term studies with ease Compact and rugged, the Fluke 1740 Series three-phase power quality
More informationFluke 1740 Series. Three-Phase Power Quality Loggers Memobox. Assess power quality and conduct long-term studies with ease
Fluke 1740 Series Three-Phase Power Quality Loggers Memobox Technical Data Assess power quality and conduct long-term studies with ease Compact and rugged, the Fluke 1740 Series three-phase power quality
More informationEnergy Management Energy Meter with plug-in Output Modules Type EM4-DIN
Energy Management Energy Meter with plug-in Output Modules Type EM4-DIN Degree of protection (front): IP 40 Front dimensions: 9 DIN modules RS 422/485 Serial port by means of optional module Dual pulse
More informationIMC-316P Hardware Manual for model IMC-316P-X-D
IMC-316P Hardware Manual for model IMC-316P-X-D Pub 352r0 A publication of 1050 Highland Drive Ann Arbor, Michigan 48108 Phone: (734) 665-5473 Fax: (734) 665-6694 www.whedco.com 2000 Whedco Incorporated,
More informationModel Hz to 10MHz Precision Phasemeter. Operating Manual
Model 6610 1Hz to 10MHz Precision Phasemeter Operating Manual Service and Warranty Krohn-Hite Instruments are designed and manufactured in accordance with sound engineering practices and should give long
More informationAC / DC STROOMTANG MULTIMETER NI 30 AC / DC CLAMP MULTIMETER NI 30. Handleiding / Manual
AC / DC STROOMTANG MULTIMETER NI 30 AC / DC CLAMP MULTIMETER NI 30 Handleiding / Manual INTRODUCTION 1-1 Unpacking and Inspection Upon removing your new Digital Clamp Multimeter from its packing, you should
More informationEnergy Management Three-phase energy meter with output modules Type EM4-DIN
Energy Management Three-phase energy meter with output modules Type EM4-DIN Class 1 (active energy) Class 2 (reactive energy) Three-phase multi-function energy meter Back-lighted LCD display 3 1 / 2 DGT
More informationFor ultra-high precision measurement of current: DC, AC, pulsed..., with galvanic separation between primary and secondary. Applications.
Current Transducer IT 200-S ULTRASTAB I PM = 200 A For ultra-high precision measurement of current: DC, AC, pulsed..., with galvanic separation between primary and secondary. Features Closed loop (compensated)
More informationDigital Power Meter. IM E 4th Edition
Digital Power Meter 4th Edition Product Registration Thank you for purchasing YOKOGAWA products. YOKOGAWA provides registered users with a variety of information and services. Please allow us to serve
More informationUser's Manual. Model f/1f Power Clamp-on Meter
User's Manual Model 382075 3f/1f Power Clamp-on Meter Introduction Congratulations on your purchase of the Extech Model 382075 Auto Range Power Clamp-On Meter. Careful use of this device, will provide
More informationAMM-1022 Digital Multimeter USER`S MANUAL
Digital Multimeter USER`S MANUAL www.tmatlantic.com CONTENTS 1. SAFETY INFORMATION.3 2. DESCRIPTION..6 3. SPECIFICATIONS.8 4. OPERATING INSTRUCTION..11 4.1 Voltage measurement...11 4.2 Current measurement
More informationTrue RMS Bench Multimeter 5492B
Data Sheet 5492B T The B&K Precision model 5492B is a versatile 5½-digit, 120,000-count bench multimeter suitable for applications in education, service repair, and manufacturing. The instrument enhances
More informationDVM645BI BENCH MULTIMETER TAFELMULTIMETER MULTIMETRE DE TABLE BANCO MULTÍMETRO TISCHMULTIMETER. User Manual. Gebruikershandleiding
BENCH MULTIMETER TAFELMULTIMETER MULTIMETRE DE TABLE BANCO MULTÍMETRO TISCHMULTIMETER User Manual Gebruikershandleiding Manuel d'utilisation Gebrauchsanleitung Introduction BENCH MULTIMETER This manual
More informationDigital Function Generator
Digital Function Generator 13654-99 PHYWE Systeme GmbH & Co. KG Robert-Bosch-Breite 10 37079 Göttingen Germany Tel. +49 (0) 551 604-0 Fax +49 (0) 551 604-107 E-mail info@phywe.de Operating Instructions
More informationRTU560. Multimeter 560CVD03. Power Measuring Display. Characteristics. Application. Data Sheet Multimeter 560CVD03
Multimeter 560CVD03 Without LCD Display There are several versions available: With LCD Display Figure 1 Type Display LO HI 1A 5A Versions R0031 3U3I x x x R0035 3U3I x x x R0051 3U3I x x x R0055 3U3I x
More informationArbiter Systems, Inc Vendels Circle, Suite 121 Paso Robles, CA USA
Model 931A Power System Analyzer Model 930A Three Phase Power Analyzer Model 929A Three Phase Power Meter with Digital Signal Analysis Model 931A shown with included accessories Arbiter Systems, Inc. Models
More informationIDEAL INDUSTRIES, INC. TECHNICAL MANUAL MODEL: MODEL: Multimeter Service Information
IDEAL INDUSTRIES, INC. TECHNICAL MANUAL MODEL: 61-340 MODEL: 61-342 Multimeter Service Information The Service Information provides the following information: Precautions and safety information Specifications
More informationKelba (Australia) Pty Ltd. PA8101S Manual ~ TABLE OF CONTENTS
Kelba (Australia) Pty Ltd Tel: Int 61 2 9476 4544 Web: www.kelba.com Email: Kelba@bigpond.net.au PA8101S Manual ~ TABLE OF CONTENTS Chapter 1: Introduction to the PA8101S Indicator (1) Chapter 2: Panel
More information1. SAFETY INFORMATION.1 2. DESCRIPTION SPECIFICATIONS.6 4. OPERATING INSTRUCTION Voltage measurement Current measurement 10
CONTENTS 1. SAFETY INFORMATION.1 2. DESCRIPTION..4 3. SPECIFICATIONS.6 4. OPERATING INSTRUCTION..9 4.1 Voltage measurement...10 4.2 Current measurement 10 4.3 Resistance measurement...12 4.4 Diode test.12
More information79/26 Series III Multimeter
79/26 Series III Multimeter Instruction Sheet W Read First: Safety Information Never use the meter if the meter or test leads look damaged. Be sure the test leads and switch are in the correct position
More informationBENNING. Service Information MODEL: MM1-1 / MM1-2 / MM1-3
BENNING MODEL: MM1-1 / MM1-2 / MM1-3 Service Information The Service Information provides the following information: Precautions and safety information Specifications Basic maintenance (cleaning, replacing
More informationOperation. Displayed channel. Measuring range. Status indication/ remote control Key lock Measuring mode/ time constant. Scale.
Electronics & Software Type 5080A... Multichannel Laboratory This universal laboratory charge amplifier can be used for force and torque measurements with piezoelectric dynamometers or force plates. Piezoelectric
More informationMC133C/133Ci. Electric Power/Energy Calibrator. Operation manual
MC133C/133Ci Electric Power/Energy Calibrator Operation manual MC133C Electric Power/Energy Calibrator Powertek Content 1 Basic information... 5 2 Preparation for operation... 6 2.1 Inspecting package
More information5700A/5720A. Operator Guide. Series II Multi-Function Calibrator
5700A/5720A Series II Multi-Function Calibrator Operator Guide PN 3474006 May 1996, Rev. 1, 5/09 2009 Fluke Corporation. All rights reserved. Printed in USA. All product names are trademarks of their respective
More informationModel Range code System Power supply Option. variables. Housing for panel mounting, (front) protection degree IP50 and optional RS485 serial output.
Power analyzers and Energy Meters Multifunction indicator Type WM12-96 Accuracy ±0.5 F.S. (current/voltage) Multifunction indicator Display of instantaneous variables: 3x3 digit Variable system and phase
More informationPOWER CLAMP-ON METER 405
POWER CLAMP-ON METER 405 ENGLISH User Manual 2 CONTENTS RECEIVING YOUR SHIPMENT... 8 ORDERING INFORMATION... 8 1 PRESENTATION... 9 1.1 THE ROTARY SWITCH... 10 1.2 THE FUNCTION BUTTONS... 11 1.3 THE DISPLAY...
More informationWide-band Models from DC to 100 MHz
CLAMP ON PROBE 327 series AC/DC CURRENT SENSOR CT686 series 99 Washington Street Melrose, MA 2176 Phone 781-665-14 Toll Free 1-8-517-8431 Visit us at www.testequipmentdepot.com Wide-band Models from DC
More informationDigital Proportional Amplifier. Type: RT-VSPA2-50. Series: 3X. Table of contents. Features. RD13763/11.07 Replaces:
Replaces: 09.12 Electro-hydraulic Control Digital Proportional Amplifier Type: RT-VSPA2-50 Series: 3X Table of contents Contents Page Features 1 Ordering code 2 Functional description 2 Block circuit diagram
More informationTechnical Specifications for Portable 3-Ø Power Analyzer
Technical Specifications for Portable 3-Ø Power Analyzer SCOPE: This specification covers the general specifications of 3-phase power analyzer that can be used to measure the no load and full load losses
More informationPOWER ANALYZER OPERATION MANUAL. Model : DW-6090A
POWER ANALYZER Model : DW-6090A Your purchase of this POWER ANALYZER marks a step forward for you into the field of precision measurement. Although this POWER ANALYZER is a complex and delicate instrument,
More information5700A/5720A Series II Multi-Function Calibrator
5700A/5720A Series II Multi-Function Calibrator Operator Guide PN 601648 May 1996 1996 Fluke Corporation, Inc. All rights reserved. Printed in U.S.A. Contents What is in this Guide?... 2 Safety Summary...
More informationI PM. Current Transducer ITZ 5000-SB FLEX ULTRASTAB = 5000 A
urrent Transducer ITZ 5000-SB FLEX ULTRASTAB I PM = 5000 A For ultra-high precision measurement of current: D, A, pulsed..., with galvanic separation between primary and secondary. Separate magnetic head
More informationOPERATOR S INSTRUCTION MANUAL
2 OPERATOR S INSTRUCTION MANUAL CLAMP METER HOLD 20 A 200 A 2 A 600 600 LIGHT ON/OFF Digital Clamp Meter 2 20 200 A khz COM CAT.II MAX 600 CONTENTS 1. General instruction.....1 2. Description...1 2.1 Precautions
More informationContents 1. General instructions. 1.1 Precautions safety measures Protection mechanisms. 2. Description. 2.1 Instrument Familiarization. 2.
Contents 1. General instructions. 1.1 Precautions safety measures... 1.2 Protection mechanisms. 2. Description. 2.1 Instrument Familiarization. 2.2 LCD Display 2.3 Keypad. 3. Function description. 3.1
More informationPower Analyzer PA4000 Datasheet
Power Analyzer PA4000 Datasheet Tektronix PA4000 Power Analyzers deliver highly accurate, multi-channel power, energy, and efficiency measurements. Precisely-matched inputs, unique Spiral Shunt technology,
More informationDIGITAL MULTIMETER CONTENTS DIGITAL MULTIMETER CONTENTS
CONTENTS CONTENTS CONTENTS 1. SAFETY INFORMATION...1 1.1 Preliminary...1 1.2 Dos and don ts...2 1.3 Symbols...3 1.4 Precautions...4 2. DESCRIPTION...5 2.1 Names of parts...6 2.2 Switches, buttons and input
More informationGAMMA 12 DIGITAL MULTIMETER
Gamma 12 www.sifamtinsley.co.uk DATASHEET Issue 1.0 Multifunction Meters Transducers & Isolators Temperature Controllers Converters & Recorders Digital Panel Meters Current Transformers Analogue Panel
More information