INFRARED RADIATION THERMOMETER KT15 II

Size: px
Start display at page:

Download "INFRARED RADIATION THERMOMETER KT15 II"

Transcription

1 INFRARED RADIATION THERMOMETER KT15 II Operating Instructions /10/16e (valid from Serial Number 10301) All rights reserved! The reproduction, transmission or use of this document or its contents is not permitted without express written authority. Offenders will be liable for damages. All rights, including rights created by patent grant or registration of a utility model or design, are reserved. Heitronics GmbH reserves the right to make modifications and improvements in their design without prior notice. (c) HEITRONICS Infrarot Messtechnik GmbH Kreuzberger Ring 40 D Wiesbaden Tel.: +49 (0) Fax: +49 (0) info@heitronics.com Internet:

2 SAFETY INSTRUCTIONS Please read carefully the instructions given in TECHNICAL SPECIFICATIONS, especially the connection and operating instructions as well as the connection instructions and configurations dealt with in SYSTEM START-UP. ATTENTION In case of wrong connection, the device might be destroyed. The Infrared Radiation Thermometer is an optical measuring instrument. Dirty lenses will lead to measuring errors. Please pay attention to the instruction given in MAINTENANCE AND CALIBRATION. HEITRONICS production meets the latest technical development thereby applying sophisticated components. Nevertheless, functional errors may occur in exceptional cases. The failure of an instrument may cause a measuring value being issued which seems to be correct, but is wrong. Please pay attention to the instructions given in MAINTENANCE AND CALIBRATION. Only when using a female connector with cable or a dummy plug attached to the plug of the device - in addition to the equipment sealing - the specified system of protection can be achieved. Operation with Laser: Devices completely mounted and equipped with integrated laser meet the safety requirements of Laser class 2. Before handling the lens in any way, the voltage supply is to be switched off or the connecting plug is to be pulled out to ensure that the laser will not switch on automatically. ATTENTION The Laser must not be switched on, if the lens of the Infrared Radiation Thermometer is removed. The integrated laser can only operate up to the upper limit of the ambient temperature of the Infrared Radiation Thermometer (60 C). ATTENTION The Laser will be switched off at an ambient temperature 60 C to defend it from destruction SAFETY INSTRUCTIONS KT 15 II 03/03/14e

3 Erklärung über die Konformität DECLARATION OF CONFORMITY Diese Erklärung gilt für folgende Erzeugnisse: This declaration is valid for the following products: Geräteart: Type of instrument: Typenbezeichnung: Designation of model: Infrarot Strahlungspyrometer Infrared Radiation Pyrometer KT15 II Serie KT15 II Series Diese Erklärung wird abgegeben durch This declaration is issued by Kreuzberger Ring Wiesbaden, Germany Hiermit wird bestätigt, dass die Produkte gemäß den Richtlinien des Rates zur Angleichung der Rechtsvorschriften der Mitgliedsstaaten über die elektromagnetische Verträglichkeit (89/336/EWG) mit den unten genannten Normen übereinstimmen: In accordance with the EU-Directive of Electro-magnetic-compatibility (89/336/EWG) the manufacturer declare, that the device described above is conform to the essential requirements of the EU-Directives: EN Class B EN Wiesbaden, 15. Juli 2003 Wiesbaden, July 15, 2003

4 PREFACE The remarkable features of the HEITRONICS devices are a design suitable for specific application as well as an uncomplicated operation. Nevertheless, it is recommended to read the entire instruction manual before putting the instruments into operation. Please observe the safety instructions as well as the particularly emphasized annotations when reading the pertinent instruction to the operation. The operating instructions are primarily directed to the user. They contain the information and instructions necessary to successfully use the instruments including all options. If you still have any questions after having read this instruction manual, please contact us. Our staff will be at your disposal. RoHS-CONFORMITY This product is in conformity with Directive 2011/65/EU of the European Parliament and of the Council of 8 June 2011 on the restriction of the use of certain hazardous substances in electrical and electronic equipment PREFACE KT15 II 02/01/15e

5 CONTENTS SAFETY INSTRUCTIONS DECLARATION OF CONFORMITY PREFACE RoHS-CONFORMITY CONTENTS DATA SHEET GENERAL Measuring temperatures using Infrared Radiation Thermometers Infrared Radiation Thermometers for GENERAL APPLICATION Infrared Radiation Thermometers for measuring METALS Infrared Radiation Thermometers for measuring PLASTICS Infrared Radiation Thermometers for measuring GLASS Infrared Radiation Thermometers for measuring GASES TECHNICAL SPECIFICATIONS Basics Technical Information 'Protective Cooling Jacket' Assignment for 12-pin connector for RS232-Interface Assignment for 12-pin connector for RS485-Interface Assignment for 7-pin connector Legend table "Temperature resolution" Temperature resolution KT15.01 II / KT15.02 II KT15.21 II KT15.23 II / KT15.24 II KT15.25 II KT15.41 II / KT15.42 II KT15.43 II KT15.45 II KT15.62 II KT15.63 II KT15.69 II KT15.81 II / KT15.82 II KT15.83 II / KT15.85 II START-UP Mounting Fig. Mounting KT15 II Protective Cooling Jacket Electrical Connection Operation without serial interface Operation with serial interface Optical Adjustment Aiming via focus laser Aiming via pilot laser Aiming via Laserpointer LP Installation Protective Cooling Jacket WK CONTENTS KT15 II 17/01/15e

6 OPERATION AND APPLICATION Application Check routines Continuous measurement of ambient temperature Use of digital input Use of digital outputs Monitoring the device function during operation Use of maximum and minimum value memories Operation via keyboard Operation of the Menu Main Menu Submenu: Temp / Info / Emi Submenu: Tamb Submenu: Resp / Memo Submenu: HiAI / LoAI Submenu: Laser / Unit Submenu: Aout Submenu: Uart Submenu: Cal Submenu: Conf Submenu: Test Communication via interface Preparing the device for operation with interface RS232C Using the interface Control of communication Data transfer from KT15 II for evaluation purposes Commands Delimiter Input buffer Description of commands... separate instructions MAINTENANCE AND CALIBRATION General information Cleaning the lens Checking the display accuracy FIGURES AND TABLES Spectral emissivity of various materials Total emissivity of some materials at 20 C Emissivity of foils Spectral emissivity of Infrared Radiation Thermometers and transmission curves of various synthetic materials Spectral emissivity ε, transmissivity τ and reflectivity ρ of glass Housing dimensions KT15II Field of view Transmission curves (illustrated in a diagram) ff WARRANTY CONDITIONS SERVICE ADDRESSES CONTENTS KT15 II 17/01/15e 0-2

7 1 DATA SHEET Type... Serial No.... Spectral response... Temperature range Calibration factor... Lens... Spacer... Detector Digital Interface... Options Accessories Code... Further data Hereby it is confirmed that the Infrared Radiation Thermometer specified above conforms to the data indicated in the specifications. Tester: Wiesbaden, DATA SHEET KT15 II 03/01/15e 1-1

8 2 GENERAL 2.1 Measuring temperatures using Infrared Radiation Thermometers Above a temperature of absolute zero, which is approx C or 0 K, all bodies emit electromagnetic radiation, the wavelength and density of which depend on the temperature. Up to a temperature of approx. 600 C, the wavelength of the radiation can be found exclusively within the infrared range (thermal radiation). It is only at higher temperatures where part of this radiation is emitted in the form of visible light. The radiation emitted by the body (radiation density) also depends on the surface of the individual body. At a fixed temperature, the maximum radiation density is emitted by a "blackbody" source of radiation. All real bodies only emit a fraction of this radiation density at the same temperature. The ratio of this fractional part to the maximum radiation density is called the emissivity ε. Naturally, the emissivity is always < 1. It depends on the nature of the surface of the material, on the material itself and on the wavelength. If the emissivity is known, the temperature of the object can be determined by measuring the infrared radiation emitted by the very object. Devices used to measure this kind of radiation are called Infrared Radiation Thermometers. As the measurement is effected without the Infrared Radiation Thermometer being in contact with the object, no distortion of the temperature field due to heat dissipation need to be feared, as is the case, e.g., with probe-type thermometers. The Infrared Radiation Thermometer is a measuring transducer, which receives the infrared radiation emitted by the measuring object itself and transforms it into a standardized output signal. All optic and electronic components are located in a small housing of a solid construction, which is made from die-cast metal so that the installation of a Infrared Radiation Thermometer is possible even in narrow spaces. By selecting a number of various lenses and detectors, it is possible to generously modify the measuring field, if the measuring distance is given. For the application of the Infrared Radiation Thermometer under high ambient temperature, water-cooling, air purge fittings and vacuum tight lenses can be delivered as accessories GENERAL KT15 II 02/03/14e 2-1

9 2.2 Infrared Radiation Thermometers for GENERAL APPLICATION The Infrared Radiation Thermometers series KT15.81 II / KT15.82 II /KT15.83 II are used to measure surface temperatures within a temperature range of - 50 C to 1000 C. The spectral sensitivity can be found within the range of the atmospheric window, which is between 8 and 14 µm. As the atmospheric transmissivity is very high in this spectral range, no weakening of the infrared radiation due to CO 2 or to water vapor contained in the air is to be expected. To meet especially high requirements as to the transmission of the atmosphere as it is a case e.g. when carrying out meteorological measurements Infrared Radiation Thermometer KT15.85 II is suitable. The KT15.81 II / KT15.82 II / KT15.83 II can be employed in the measuring of temperatures during the production and processing of plastics, rubber, paper, textiles, paint, ceramics etc. The emissivity of various metals is shown in figure 10 (chapter 8: Illustrations). The devices can also be used to measure the temperature of objects exposed to infrared radiation, since the radiation maximum of infrared radiators is within the range of shorter wavelengths, which do not cause the Infrared Radiation Thermometers KT15.81 II / KT15.82 II / KT15.83 II to react. The Infrared Radiation Thermometer KT15.81 II is used in the spectral range of 8 to 10 μm. Besides being used for large-surface measurement, this Infrared Radiation Thermometer type is also suitable for measuring thick foils. The Infrared Radiation Thermometer KT15.82 II is applied for the standard measuring range of 8 to 14 µm thereby guaranteeing a good signal resolution. The Infrared Radiation Thermometer KT15.83 II provides a particularly high signal resolution owing to its large spectral range. Therefore, it is very well suited for measuring low temperatures. For high requirements on the transmission of the atmosphere, such as found in meteorological measurements, the KT15.85 II type is available. This device type operates within the spectral ranges of 9.6 to 11.5 µm, where the transmission of the atmosphere is extremely high GENERAL KT15 II 02/03/14e 2-2

10 2.3. Infrared Radiation Thermometers for measuring METALS The spectral emissivity of the Infrared Radiation Thermometers KT15.01 II / KT15.02 II is 2 to 2.7 µm and 2 to 4.5 µm. Within this spectral range, metals and metal-oxides show a relatively high spectral emissivity. This is why the device is particularly well suited to measure the temperatures of these materials. Since the atmospheric transmissivity is high within the spectral range of 2 to 2.7 µm, distortions of the measured value owing to absorption of water vapor and CO 2 in the air can be ignored (Fig. 11, ILLUSTRATIONS). The Infrared Radiation Thermometer type KT15.01 II is used for measuring temperatures from 300 C onwards; the Infrared Radiation Thermometer type KT15.02 II is suitable for measuring temperatures from 200 C onwards. Temperatures from 500 C respectively 700 C onwards are preferably measured by means of the spectral pyrometers of the series KT18S or the quotient pyrometers of the series KT18R. 2.4 Infrared Radiation Thermometers for measuring PLASTICS These types of Infrared Radiation Thermometers are employed for the measuring of surface temperatures of plastics within a temperature range of 0 to 600 C. These Infrared Radiation Thermometers are equipped with narrow-band infrared filters. KT15.21 II measures radiation at 3.43 µm, KT15.23 II measures radiation at 6.8 µm, KT15.24 II measures radiation at 7.93 µm, KT15.25 II measures radiation at 8.05 µm. Within these spectral ranges, virtually all synthetic materials have strong absorption bands so that even thin foils show high emissivity. The Infrared Radiation Thermometer KT15.21 II is suitable for most foils from 100 C onwards. The Infrared Radiation Thermometer KT15.23 II measures thin foils made from polyethylene, polypropylene, polyisobutane, polynitrile, polystyrene and similar materials from 0 C onwards. The Infrared Radiation Thermometer KT15.24 II and KT II measure thin foils polyester as well as foils made from - acrylic, - cellulose, - fluorine compounds, - polycarbonate, - polyamide, - polyurethane, - polyvinylchloride and similar materials after analysis from 0 C onwards GENERAL KT15 II 02/03/14e 2-3

11 2.5 Infrared Radiation Thermometers for measuring GLASS KT15.42 II is suitable for measuring glass and quartz. The spectral sensitivity is 4.9 to 5.5 µm. The emissivity of glass is very similar to the emissivity of a blackbody radiator within this spectral range. Furthermore, this spectral range eliminates any disturbing influences resulting from the strong absorption bands of water vapor at 6.2 µm. The Infrared Radiation Thermometers KT15.01 II and KT15.41 II are not measuring the surface but the average value of temperatures up to a certain penetration depth (glass volume). The lower limits are 300 C for KT15.01 II and 400 C for KT15.41 II. The types KT15.42 II and KT15.43 II measure the surface temperature from 100 C respectively from 0 C onwards. 2.6 Infrared Radiation Thermometers for measuring GASES By means of narrow-band filters, the selective measurement of temperature of certain gases is possible. A rather long response time is to be chosen since the radiation energy is very low. KT15.21 II measures hydrocarbon from 100 C onwards. KT15.62 II measures materials within the absorption range of carbon dioxide and carbon monoxide, KT15.63 II measures carbon monoxide selectively, KT15.64 II is provided for nitrogen oxide and KT15.69 II has been especially developed for determination of combustion temperatures. Using KT15.41 II, objects can be measured through gases and flames from 400 C onwards GENERAL KT15 II 02/03/14e 2-4

12 3 TECHNICAL SPECIFICATIONS Spectral sensitivity: Temperature measuring range: Temperature resolution: DATA SHEET DATA SHEET Tables (Page 3-7ff) Accuracy (provided that the emissivity has been correctly set, after a warm-up period of 15 min.): ± 0.5 C plus 0.7 % of the temperature difference between the housing containing the measuring instruments and the object to be measured or: value of temperature resolution. The higher value shall prevail. Long-time stability: Lens used: Possible lenses: Target diameter (95 %): Field of view: Radiation detector: better than 0.1 of the absolute measuring temperature in Kelvin/month DATA SHEET OPTIONS AND ACCESSORIES The target diameter depends on the lens and the detector used in each individual case. When short-focus lenses are used, the distance within which the minimum measuring field can be found varies by ± 4%. Marking the field of view is carried out by means of different accessories. OPTICAL ADJUSTMENT (chapter 4.5) HEITRONICS pyroelectric detector Permissible ambient temperature: C For higher ambient temperatures, cooling and protecting jackets are available. OPTIONS AND ACCESSORIES and TECHNICAL INFORMATION (page 3-3) Storage temperature: C Weight: approx kg Dimensions: FIGURES: Fig. 15 Operating voltages: AC: 12 (-10 %) V (+ 10 %) Hz DC: V power consumption: 3.5 W Type of protection: IP TECHNICAL SPECIFICATIONS KT15 II 21/03/12e

13 Analog output: Lower temperature value (T.low): Upper temperature value (T.end): Minimum temperature difference: possible signal outputs (variable by programming) ma, ma V, V minimum temp. value for analog signal maximum temp. value for analog output depends on final temperature (T.end) The following values will result: Minimum temperature Final temperature difference 150 C 50 C 200 C 100 C 1000 C 200 C > 1000 C 400 C Load of analog output: current output: load 550 Ohm voltage output: load 10 kohm Resolution of analog output: 16 bit Digital interface V24 (RS232C): Response time (90 %): kbaud variable by programming KT15 IIP: 0,005; 0,01; 0,03; 0,1; 0,3; 1; 3; 10; 30; 60; 120; 240; 360, 480 und 600 s KT15 IID: 0,03; 0,1; 0,3; 1; 3; 10 s Protection against oscillation: Connector type: DIN Bl. 8; test: Fc vibration strength: A B1 E frequency range: 10 to 55 Hz amplitude: +/- 0.2 mm duration of test/position: 30 min miniature circular connector, Franz Binder GmbH, type , series 423, 12 pin and type , series 423, 7 pin Options: Thermal switch: switching temperature: > 70 C load: voltage: 48 V, current: 0.5 A alternative: Analog input: 0 10 V alternative: Digital input: floating contact or 0 30 V (low 1 V, high 4 V) Laser: SAFETY INSTRUCTIONS Laser type Pilot laser Focus laser Protection class Output Wavelength Form Diameter Max. ambient temperature 2 < 1 mw 650 nm round beam Ø 3 1 m distance 60.0 C 2 < 1 mw 650 nm cross lines with centering circle adapted to field of view in the focus distance 60.0 C TECHNICAL SPECIFICATIONS KT15 II 21/03/12e 3-2

14 Technical Information Cooling Jacket WK15 The HEITRONICS Infrared Radiation Thermometers can be operated at an environmental temperature of up to 60 C without any additional coolants. If the environmental temperatures are higher than that mentioned above, cooling fittings can be used. Basically, the cooling jacket WK15 can be cooled with air or water. The possible maximum environmental temperatures are higher for water cooling than for air cooling ( individual Data Sheet). If the cooling process is too strong, i.e. if the cooling air / cooling water is too cold, this may result in condensate as soon as cooling medium falls below the dew point. To prevent this from occurring, the cooling air / cooling water must have a minimum temperature as a function of the relative humidity. The following table specifies the minimum temperature of the cooling air / cooling water as a function of the environmental temperature and the relative humidity of the environmental air. Minimum temperature of the cooling air / cooling water Relative air humidity (of the environmental air) Environmental air temperature / C 2% 4% 10% 20% 30% 50% 70% 30,0 5,0 5,0 5,0 6,0 11,0 19,0 25,0 C 40,0 5,0 5,0 5,0 13,0 20,0 28,0 34,0 C 50,0 5,0 5,0 10,0 21,0 28,0 38,0 45,0 C 60,0 5,0 5,0 18,0 28,0 38,0 47,0 54,0 C 70,0 5,0 9,0 24,0 38,0 45,0 57,0 nm C 80,0 5,0 15,0 32,0 45,0 55,0 nm nm C 90,0 10,0 21,0 38,0 52,0 nm nm nm C >100,0 15,0 27,0 45,0 60,0 nm nm nm C Legend: nm * Operation not possible since the minimum temperature is above 60 C Table: Minimum temperature of the coolant TECHNICAL SPECIFICATIONS KT15 II 25/01/15e 3-3

15 Assignment for 12-pin connector - for RS232-Interface ATTENTION In case of wrong connection, the device might be destroyed. Cores color Code DIN IEC 757 plug-in contacts Function RS232 9-pin RS pin red RD A CTS 7 4 white WH B supply voltage (AC or DC) green/white - alternative: grey/pink or colorless GNWH GYPK colorless C + thermal switch analog input digital input (note 1) isolated digital input (DI) 24V-logic / 5V-logic (note 2) open-collector output 2 (DO2) (note 3) grey GY D thermal switch analog input digital input (note 1) DI/DO1-zero (Gnd) (note 2) yellow YE E + analog output DI/DO1/DO2-zero (Gnd) (note 3) brown/white - alternative: red/blue or orange BNWH RDBU OR F DTR 6 6 open-collector output (DO1) (note 2) pink PK G TXD 2 3 violet VT H RTS 8 5 blue BU J RXD 3 2 black BK K Datacom (data ground) 5 7 brown BN L + supply voltage (AC or DC) green GN M analog output Notes: 1: Hardware-Programming see Data Sheet page 1-1 2: Valid for Interface map with open-collector output. 3: Valid for Interface map with 2 open-collector output. Further information next page 12-pin plug connector 12-pin socket-outlet TECHNICAL SPECIFICATIONS KT15 II 25/01/15e 3-4

16 Assignment valid for RS232-Interface Only 1 variant possible (Version Data Sheet page 1-1) TECHNICAL SPECIFICATIONS KT15 II 25/01/15e 3-5

17 Assignment for 12-pin connector - for RS485-Interface ATTENTION In case of wrong connection, the device might be destroyed. Cores color Code DIN IEC 757 plug-in contacts Function red RD A BUS + input (RXD+) (note 1, 4) not connected (note 1, 5) white WH B supply voltage (AC or DC) green/white - alternative: grey/pink or colorless GNWH GYPK colorless C + thermal switch analog input digital input isolated digital input (DI) 24V-logic / 5V-logic (note 1) (note 1, 2) open-collector output 2 (DO 2) (note 1, 3) grey GY D thermal switch analog input digital input (note 1) DI/DO1-zero (Gnd) (note 1, 2) yellow YE E + analog output DI/DO1/DO2-zero (Gnd) (note 1, 3) brown/white - alternative: red/blue or orange BNWH RDBU OR F open-collector output (DO1) (note 1, 2) pink PK G violet VT H blue BU J BUS - output (TXD-) (note 1, 4) BUS - input / BUS - output (DATA-) (note 1, 5) BUS + output (TXD+) (note 1, 4) BUS + input / BUS + output (DATA+) (note 1, 5) BUS - input (RXD-) (note 1, 4) not connected (note 1, 5) black BK K BUS-GND brown BN L + supply voltage (AC or DC) green GN M analog output Notes: 1: Hardware-Programming see Data Sheet 1-1 2: Valid for Interface map with isolation and open-collector output. 3: Valid for Interface map with isolation and 2 open-collector output. 4: Valid for RS485, Full-duplex 5: Valid for RS485, Half-duplex 12-pin plug connector 12-pin socket-outlet TECHNICAL SPECIFICATIONS KT15 II 25/01/15e 3-6

18 Assignment for 7-pin connector ATTENTION When connecting the Infrared Radiation Thermometer to the Power Supply Unit T24 please note the connection conditions in the manual. Cores color Code DIN IEC 757 plug-in contacts brown BN 4 + white WH 2 yellow YE 5 + green GN 1 supply voltage (AC or DC) analog output blue BU 3 housing pink PK 6 grey GY 7 + thermal switch analog input digital input (note 1) thermal switch analog input digital input (note 1) Note 1: Hardware-Programming see Data Sheet page pin plug connector 7-pin socket-outlet The load resistors for the analog outputs are: for voltage output 0 to 1 V 10 kohm, for voltage output 0 to 10 V 10 kohm, for current output 0 to 20 ma 550 Ohm for current output 4 to 20 ma 550 Ohm For the signal evaluation, differential inputs are recommended TECHNICAL SPECIFICATIONS KT15 II 25/01/15e

19 Temperaturauflösung/Temperature resolution KT15.01 II / KT15.02 II Temperaturauflösung (NET) in ± K (bei Emissionsgrad = 1; σ = 2) für Standardobjektive. Temperature resolution (NET) in ± K (emissivity-setting = 1; σ = 2) for standard lenses. Strahlertemperatur Radiation temperature Einstellzeit Response time KT15.01 II Detektortyp Detector type Einstellzeit Response time KT15.02 II Detektortyp Detector type A B C G A B C G 200 C 300 C 500 C 700 C 1000 C 5 ms ** ** ** ** 5 ms ** ** 10 ms ** ** ** ** 10 ms ** ** 30 ms ** ** ** ** 30 ms ** 100 ms ** ** ** 100 ms ms 6.70 ** ** ** 300 ms s ** ** 1 s s ** 3 s s ** 10 s ms ** ** ** 5 ms ** 10 ms ** ** ** 10 ms ** 30 ms ** ** 30 ms ms ** 100 ms ms ** 300 ms s s s s s s ms ** 5 ms ms ** 10 ms ms ms ms ms ms ms s s s s s s ms ms ms ms ms ms ms ms ms ms s s s s s s ms ms ms ms ms ms ms ms ms ms s s s s s s ** Value is > 20 C. This setting is not recommended. Table: Temperature resolution Infrared Radiation Thermometer KT15.01 II / KT15.02 II TECHNICAL SPECIFICATIONS KT15 II 26/01/16e 3-8

20 Temperaturauflösung/Temperature resolution KT15.21 II Temperaturauflösung (NET) in ± K (bei Emissionsgrad = 1; σ = 2) für Standardobjektive. Temperature resolution (NET) in ± K (emissivity-setting = 1; σ = 2) for standard lenses. Strahlertemperatur Radiation temperature Einstellzeit Response time KT15.21 II Detektortyp Detector type A B C G 20 C 100 C 200 C 300 C 400 C 5 ms ** ** ** ** 10 ms ** ** ** ** 30 ms ** ** ** ** 100 ms ** ** ** ** 300 ms ** ** ** ** 1 s ** ** ** ** 3 s ** ** ** ** 10 s ** ** ** ** 5 ms ** ** ** ** 10 ms ** ** ** ** 30 ms ** ** ** ** 100 ms ** ** ** ** 300 ms ** ** ** ** 1 s 9.10 ** ** ** 3 s 5.45 ** ** ** 10 s ** ** 5 ms ** ** ** ** 10 ms ** ** ** ** 30 ms ** ** ** 100 ms ** ** 300 ms ** ** 1 s ** 3 s ** 10 s ms ** ** ** 10 ms ** ** ** 30 ms ** ** 100 ms ** 300 ms ** 1 s s s ms 6.05 ** ** ** 10 ms ** ** 30 ms ** 100 ms ** 300 ms s s s ** Value is > 20 C. This setting is not recommended. Table: Temperature resolution Infrared Radiation Thermometer KT15.21 II TECHNICAL SPECIFICATIONS KT15 II 26/01/16e 3-9

21 Temperaturauflösung/Temperature resolution KT15.23 II / KT15.24 II Temperaturauflösung (NET) in ± K (bei Emissionsgrad = 1; σ = 2) für Standardobjektive. Temperature resolution (NET) in ± K (emissivity-setting = 1; σ = 2) for standard lenses. Strahlertemperatur Radiation temperature Einstellzeit Response time KT15.23 II Detektortyp Detector type Einstellzeit Response time KT15.24 II Detektortyp Detector type A B C G A B C G 20 C 100 C 200 C 300 C 400 C 5 ms ** ** ** ** 5 ms ** ** ** ** 10 ms ** ** ** ** 10 ms ** ** ** ** 30 ms ** ** ** ** 30 ms ** ** ** 100 ms ** ** ** 100 ms 5.60 ** ** ** 300 ms 6.80 ** ** ** 300 ms ** ** 1 s ** ** 1 s ** 3 s ** 3 s ** 10 s ** 10 s ms ** ** ** ** 5 ms ** ** ** ** 10 ms ** ** ** ** 10 ms ** ** ** ** 30 ms 8.65 ** ** ** 30 ms 6.00 ** ** ** 100 ms ** ** 100 ms ** ** 300 ms ** ** 300 ms ** 1 s ** 1 s ** 3 s s s s ms ** ** ** ** 5 ms ** ** ** ** 10 ms ** ** ** ** 10 ms ** ** ** ** 30 ms ** ** 30 ms ** ** 100 ms ** 100 ms ** 300 ms ** 300 ms ** 1 s s s s s s ms ** ** ** ** 5 ms ** ** ** ** 10 ms ** ** ** ** 10 ms ** ** ** ** 30 ms ** ** 30 ms ** ** 100 ms ** 100 ms ** 300 ms ** 300 ms s s s s s s ms ** ** ** ** 5 ms ** ** ** ** 10 ms ** ** ** ** 10 ms ** ** ** ** 30 ms ** 30 ms ** 100 ms ** 100 ms ** 300 ms ms s s s s s s ** Value is > 20 C. This setting is not recommended. Table: Temperature resolution Infrared Radiation Thermometer KT15.23 II / KT15.24 II TECHNICAL SPECIFICATIONS KT15 II 26/01/16e

22 Temperaturauflösung/Temperature resolution KT15.25 II Temperaturauflösung (NET) in ± K (bei Emissionsgrad = 1; σ = 2) für Standardobjektive. Temperature resolution (NET) in ± K (emissivity-setting = 1; σ = 2) for standard lenses. Strahlertemperatur Radiation temperature Einstellzeit Response time KT15.25 II Detektortyp Detector type A B C G 20 C 100 C 200 C 300 C 400 C 5 ms ** ** ** ** 10 ms ** ** ** ** 30 ms ** ** ** 100 ms 5.40 ** ** ** 300 ms ** ** 1 s ** 3 s ** 10 s ms ** ** ** ** 10 ms ** ** ** ** 30 ms 5.80 ** ** ** 100 ms ** ** 300 ms ** 1 s s s ms ** ** ** ** 10 ms ** ** ** ** 30 ms ** ** 100 ms ** 300 ms ** 1 s s s ms ** ** ** ** 10 ms ** ** ** ** 30 ms ** ** 100 ms ** 300 ms s s s ms ** ** ** ** 10 ms ** ** ** ** 30 ms ** 100 ms ** 300 ms s s s ** Value is > 20 C. This setting is not recommended. Table: Temperature resolution Infrared Radiation Thermometer KT15.25 II TECHNICAL SPECIFICATIONS KT15 II 26/01/16e

23 Temperaturauflösung/Temperature resolution KT15.41 II / KT15.42 II Temperaturauflösung (NET) in ± K (bei Emissionsgrad = 1; σ = 2) für Standardobjektive. Temperature resolution (NET) in ± K (emissivity-setting = 1; σ = 2) for standard lenses. Strahlertemperatur Radiation temperature Einstellzeit Response time KT15.41 II Detektortyp Detector type Einstellzeit Response time KT15.42 II Detektortyp Detector type A B C G A B C G 20 C 100 C 300 C 700 C 1000 C 5 ms ** ** ** ** 5 ms ** ** ** ** 10 ms ** ** ** ** 10 ms ** ** ** ** 30 ms ** ** ** ** 30 ms ** ** ** ** 100 ms ** ** ** ** 100 ms ** ** ** 300 ms ** ** ** ** 300 ms 8.40 ** ** ** 1 s ** ** ** ** 1 s ** ** 3 s ** ** ** ** 3 s ** 10 s ** ** ** 10 s ** 5 ms ** ** ** ** 5 ms ** ** ** ** 10 ms ** ** ** ** 10 ms ** ** ** 30 ms ** ** ** ** 30 ms 6.40 ** ** ** 100 ms ** ** ** 100 ms ** ** 300 ms 8.75 ** ** ** 300 ms ** 1 s ** ** 1 s ** 3 s ** 3 s s ** 10 s ms 7.90 ** ** ** 5 ms ** ** 10 ms 6.40 ** ** ** 10 ms ** ** 30 ms ** ** 30 ms ** 100 ms ** 100 ms ms ms s s s s s s ms ** 5 ms ** 10 ms ** 10 ms ** 30 ms ms ms ms ms ms s s s s s s ms ** 5 ms ** 10 ms ** 10 ms ** 30 ms ms ms ms ms ms s s s s s s ** Value is > 20 C. This setting is not recommended. Table: Temperature resolution Infrared Radiation Thermometer KT15.41 II / KT15.42 II TECHNICAL SPECIFICATIONS KT15 II 26/01/16e

24 Temperaturauflösung/Temperature resolution KT15.43 II Temperaturauflösung (NET) in ± K (bei Emissionsgrad = 1; σ = 2) für Standardobjektive. Temperature resolution (NET) in ± K (emissivity-setting = 1; σ = 2) for standard lenses. Strahlertemperatur Radiation temperature Einstellzeit Response time KT15.43 II Detektortyp Detector type A B C G 20 C 100 C 300 C 700 C 1000 C 5 ms ** ** ** 10 ms ** ** ** 30 ms ** ** 100 ms ** 300 ms ** 1 s s s ms 6.30 ** ** ** 10 ms ** ** 30 ms ** 100 ms ** 300 ms s s s ms ** ** 10 ms ** 30 ms ** 100 ms ms s s s ms ** 10 ms ** 30 ms ms ms s s s ms ** 10 ms ** 30 ms ms ms s s s ** Value is > 20 C. This setting is not recommended. Table: Temperature resolution Infrared Radiation Thermometer KT15.43 II TECHNICAL SPECIFICATIONS KT15 II 26/01/16e

25 Temperaturauflösung/Temperature resolution KT15.45 II Temperaturauflösung (NET) in ± K (bei Emissionsgrad = 1; σ = 2) für Standardobjektive. Temperature resolution (NET) in ± K (emissivity-setting = 1; σ = 2) for standard lenses. Strahlertemperatur Radiation temperature Einstellzeit Response time KT15.45 II Detektortyp Detector type A B C G 20 C 100 C 300 C 700 C 1000 C 5 ms ** ** ** ** 10 ms ** ** ** ** 30 ms 12,55 ** ** ** 100 ms 5,15 19 ** ** 300 ms 3,4 12,65 ** ** 1 s 1,45 5,25 12,85 ** 3 s 0,85 3,15 7,7 ** 10 s 0,55 2,1 5,15 17,1 5 ms 17 ** ** ** 10 ms 13,75 ** ** ** 30 ms 5,05 18,65 ** ** 100 ms 2,05 7,65 18,6 ** 300 ms 1,4 5,1 12,4 ** 1 s 0,55 2,1 5,15 17,2 3 s 0,35 1,25 3,1 10,3 10 s 0,25 0,85 2,05 6,9 5 ms 6,4 ** ** ** 10 ms 5,15 19,1 ** ** 30 ms 1,9 7 17,05 ** 100 ms 0,8 2,85 6,95 ** 300 ms 0,5 1,9 4,65 15,5 1 s 0,2 0,8 1,95 6,45 3 s 0,15 0,5 1,15 3,85 10 s 0,1 0,3 0,75 2,6 5 ms 4 14,05 ** ** 10 ms 3,15 11,4 ** ** 30 ms 1,25 4,2 10,15 ** 100 ms 0,5 1,7 4,15 13,8 300 ms 0,3 1,15 2,75 9,2 1 s 0,15 0,45 1,15 3,85 3 s 0,1 0,3 0,7 2,3 10 s 0,05 0,2 0,45 1,55 5 ms 3,95 12,5 ** ** 10 ms 2,9 10,05 ** ** 30 ms 1,25 3,75 8,95 ** 100 ms 0,45 1,5 3,65 12, ms 0,3 1 2,45 8,1 1 s 0,15 0,4 1 3,4 3 s 0,1 0,25 0,6 2,05 10 s 0,05 0,15 0,4 1,35 ** Value is > 20 C. This setting is not recommended. Table: Temperature resolution Infrared Radiation Thermometer KT15.45 II TECHNICAL SPECIFICATIONS KT15 II 26/01/16e

26 Temperaturauflösung/Temperature resolution KT15.62 II Temperaturauflösung (NET) in ± K (bei Emissionsgrad = 1; σ = 2) für Standardobjektive. Temperature resolution (NET) in ± K (emissivity-setting = 1; σ = 2) for standard lenses. Strahlertemperatur Radiation temperature 400 C 500 C 700 C 1000 C 1500 C Einstellzeit Response time KT15.62 II Detektortyp Detector type A B C G 5 ms ** ** ** ** 10 ms ** ** ** ** 30 ms ** 100 ms ms s s s ms ** ** ** ** 10 ms ** ** ** ** 30 ms ** 100 ms ms s s s ms ** ** ** ** 10 ms ** ** ** ** 30 ms ms ms s s s ms ** ** ** ** 10 ms ** ** ** ** 30 ms ms ms s s s ms ** ** ** ** 10 ms ** ** ** ** 30 ms ms ms s s s ** Value is > 20 C. This setting is not recommended. Table: Temperature resolution Infrared Radiation Thermometer KT15.62 II TECHNICAL SPECIFICATIONS KT15 II 26/01/16e

27 Temperaturauflösung/Temperature resolution KT15.63 II Temperaturauflösung (NET) in ± K (bei Emissionsgrad = 1; σ = 2) für Standardobjektive. Temperature resolution (NET) in ± K (emissivity-setting = 1; σ = 2) for standard lenses. Strahlertemperatur Radiation temperature Einstellzeit Response time KT15.63 II Detektortyp Detector type A B C G 400 C 500 C 700 C 1000 C 1500 C 5 ms ** ** ** ** 10 ms ** ** ** ** 30 ms ** 100 ms ms s s s ms ** ** ** ** 10 ms ** ** ** ** 30 ms ** 100 ms ms s s s ms ** ** ** ** 10 ms ** ** ** ** 30 ms ms ms s s s ms ** ** ** ** 10 ms ** ** ** ** 30 ms ms ms s s s ms ** ** ** ** 10 ms ** ** ** ** 30 ms ms ms s s s ** Value is > 20 C. This setting is not recommended. Table: Temperature resolution Infrared Radiation Thermometer KT15.63 II TECHNICAL SPECIFICATIONS KT15 II 26/01/16e

28 Temperaturauflösung/Temperature resolution KT15.69 II Temperaturauflösung (NET) in ± K (bei Emissionsgrad = 1; σ = 2) für Standardobjektive. Temperature resolution (NET) in ± K (emissivity-setting = 1; σ = 2) for standard lenses. Strahlertemperatur Radiation temperature Einstellzeit Response time KT15.69 II Detektortyp Detector type A B C G 400 C 500 C 700 C 1000 C 1500 C 5 ms ** ** ** ** 10 ms ** ** ** ** 30 ms ** 100 ms ms s s s ms ** ** ** ** 10 ms ** ** ** ** 30 ms ** 100 ms ms s s s ms ** ** ** ** 10 ms ** ** ** ** 30 ms ms ms s s s ms ** ** ** ** 10 ms ** ** ** ** 30 ms ms ms s s s ms ** ** ** ** 10 ms ** ** ** ** 30 ms ms ms s s s ** Value is > 20 C. This setting is not recommended. Table: Temperature resolution Infrared Radiation Thermometer KT15.69 II TECHNICAL SPECIFICATIONS KT15 II 26/01/16e

29 Temperaturauflösung/Temperature resolution KT15.81 II / KT15.82 II Temperaturauflösung (NET) in ± K (bei Emissionsgrad = 1; σ = 2) für Standardobjektive. Temperature resolution (NET) in ± K (emissivity-setting = 1; σ = 2) for standard lenses. Strahlertemperatur Radiation temperature Einstellzeit Response time KT15.81 II Detektortyp Detector type Einstellzeit Response time KT15.82 II Detektortyp Detector type A B C G A B C G 25 C 20 C 100 C 300 C 700 C 5 ms 7.70 ** ** ** 5 ms ** ** 10 ms 6.25 ** ** ** 10 ms ** ** 30 ms ** ** 30 ms ** 100 ms ** 100 ms ms ms s s s s s s ms ** ** 5 ms ** 10 ms ** ** 10 ms ** 30 ms ** 30 ms ms ms ms ms s s s s s s ms ** 5 ms ** 10 ms ** 10 ms ** 30 ms ms ms ms ms ms s s s s s s ms ** 5 ms ms ** 10 ms ms ms ms ms ms ms s s s s s s ms ** 5 ms ms ms ms ms ms ms ms ms s s s s s s ** Value is > 20 C. This setting is not recommended. Table: Temperature resolution Infrared Radiation Thermometer KT15.81 II / KT15.82 II TECHNICAL SPECIFICATIONS KT15 II 26/01/16e

30 Temperaturauflösung/Temperature resolution KT15.83 II / KT15.85 II Temperaturauflösung (NET) in ± K (bei Emissionsgrad = 1; σ = 2) für Standardobjektive. Temperature resolution (NET) in ± K (emissivity-setting = 1; σ = 2) for standard lenses. Strahlertemperatur Radiation temperature Einstellzeit Response time KT15.83 II Detektortyp Detector type Einstellzeit Response time KT15.85 II Detektortyp Detector type A B C G A B C G 25 C 20 C 100 C 300 C 700 C 5 ms ** ** 5 ms ** ** ** 10 ms ** ** 10 ms ** ** ** 30 ms ** 30 ms 5.80 ** ** ** 100 ms ms ** ** 300 ms ms ** 1 s s s s s s ms ** 5 ms ** ** ** 10 ms ** 10 ms 9.50 ** ** ** 30 ms ms ** ** 100 ms ms ** 300 ms ms ** 1 s s s s s s ms ** 5 ms 6.70 ** ** ** 10 ms ** 10 ms 5.45 ** ** ** 30 ms ms ** 100 ms ms ** 300 ms ms s s s s s s ms ms ** ** 10 ms ms ** ** 30 ms ms ** 100 ms ms ms ms s s s s s s ms ms ** ** 10 ms ms ** ** 30 ms ms ** 100 ms ms ms ms s s s s s s ** Value is > 20 C. This setting is not recommended. Table: Temperature resolution Infrared Radiation Thermometer KT15.83 II / KT15.85 II TECHNICAL SPECIFICATIONS KT15 II 26/01/16e

31 4 START-UP 4.1 Mounting There are four ways of mechanically mounting the Infrared Radiation Thermometer: direct mounting via two tapped holes located on the bottom side of the device ( figure below) direct mounting via the tripod thread BSW 1/4" located on the bottom side of the device ( figure below) mounting by ball joint via thread BSW 1/4" or spring flange. The device permits an easy adjustment of the Infrared Radiation Thermometer to the measured object. ( OPTIONS AND ACCESSORIES) mounting via adapter ( OPTIONS AND ACCESSORIES) START-UP KT15 II 05/03/14e 4-1

32 4.2 Electrical Connection The Infrared Radiation Thermometer is delivered with PVC- or PTFE-cable with free ends. The Infrared Radiation Thermometer is delivered with a connector which is watertight when closed (IP65). The 12-pole connector is intended for the connection of the supply voltage of the analog output signal, the thermal switch, as well as in addition the connection of a serial interface (RS232C) as well as two alarm contacts. ATTENTION In case of wrong connection, the device might be destroyed. Connection of supply voltage T24 MS30 MS35 Connecting cable with free ends Connecting cable with free ends Connecting cable with plug connector Assignment of the individual cores, respectively plug-in contacts tables "Assignment" Option: 7-pole connector for the connection of the supply voltage of the analog output signal and the thermal switch 4.3 Operation without serial interface When operating without a serial interface the emissivity will be set as described in chapter Operation with serial interface Chapter : Description of the commands ATTENTION: 12-pole connection required START-UP KT15 II 05/03/14e 4-2

33 4.5 Optical Adjustment By means of Infrared Radiation Thermometer KT15 II, the temperature of target can be measured at any distance on the condition that the desired target area is larger than the measured target diameter of the Infrared Radiation Thermometer ( ILLUSTRATIONS, Field of view diagram). For applications desiring a minimum target area to be measured, the measuring field diameter will define at what distance the smallest measured target diameter will be. For optical adjustment of the measuring object further accessories are available ( OPTIONS AND ACCESSORIES) Aiming via focus laser The illumination of the measured target diameter via the focus laser is switched on/off the "keyboard". At works the focus laser is set for the mounted lens. This means that there will be a reticule with circle at the minimum target diameter for the mounted objective indicating the target size and position. In case other objectives are attached to the Infrared Radiation Thermometer, the focus laser is no longer adjusted, i.e. neither the position nor the size of the target diameter is indicated correctly Aiming via pilot laser ATTENTION Please observe the SAFETY INSTRUCTIONS. A pilot laser can be integrated into the Infrared Radiation Thermometer KT15 II in case a lens is used which is transparent to visible light. It marks the centre of the target diameter for the objective which has been mounted at works. It is not possible to show the minimum target diameter with this option Aiming via Laserpointer LP15 For lenses of type Kx, a laserpointer is required ( OPTIONS AND ACCESSORIES), which is to be attached to lens Kx. The laser marks the centre of the measuring field within a distance between 0.5 m and 5 m START-UP KT15 II 05/03/14e 4-3

34 4.6 Installation of Protective Cooling Jacket WK15 Installation of protective cooling jacket WK15 with integrated air purge for different applications in connection with adapter B4-B2 and B4-B7-B2 Fitting for Fitting for air purge * cooling jacket * Please, use only the marked connection for the respective combination! START-UP KT15 II 04/07/13e 4-4

35 5 OPERATION AND APPLICATION Configuration The Infrared Radiation Thermometers series KT15 II can be adapted to the process by means of parameterization. The adaptation is carried out via the keyboard or via the digital interface. The operation via keyboard is described in the following. The communication via interface is carried out analog to it. 5.0 Application Check routines Check routines detect errors and indicate such errors. Interface error Errors arising with the communication via the serial interface, are automatically output via the latter as text. Error of measuring range Errors arising due to the overflow or underflow of the range limits are also given via the analog output. Overflow and underflow is indicated at the display. If KT15 II interface is programmed for temperature output, an error message is output instead of a temperature value. The configured signal of the analog output is switched to the maximum value in case of overflow (OF) (approx. 5 % above range limit value), and to the minimum value in case of underflow (UF) (approx. 5 % below range limit value at voltage output, approx % at current output). The following cases are monitored: Overflow (OF) Underflow (UF) 1. The measuring temperature is above the instrument measuring range. 1. The measuring temperature is below the instrument measuring range. 2. The internal temperature is above the permissible sensor temperature. 2. The internal temperature is below the permissible sensor temperature. 3. The external ambient temperature is above the instrument measuring range. 3. The external ambient temperature is below the instrument measuring range. 4. The voltage at the external temperature input is above the operating range (0 to 10 V). 4. The voltage at the external temperature input is below the operating range (0 to 10 V). 5. Data memory is partly destroyed. 5. Rotational speed of the modulator is outside the limit. 6. Initialization error of the instrument. 6. Internal miscalculation OPERATION AND APPLICATION KT15 II 07/02/15e

36 5.0.2 Continuous measurement of ambient temperature For continuous measurement of the ambient temperature, a 0 to 10 V signal which is proportional to the ambient temperature must be fed in at terminal C and D of the 12-pole connector. The configuration of the input must be programmed with EasyConfig: Minimum temperature value corresponding to 0 V, Maximum temperature value corresponding to 10 V. For the transmission of the measured temperature, a module which is recommended for the sensor is needed. OPTIONS AND ACCESSORIES. The figure shows the connection of a module with PT100 in three-wire technique to an Infrared Radiation Thermometer KT15 II. ATTENTION The device must be provided for 'Analog input' OPERATION AND APPLICATION KT15 II 06/03/14e 5-2

37 5.0.3 Use of the digital input If the device is equipped with the option 'digital input', this option can be used for remote control of one of the programmable functions as described below, or a connecting cable with integrated switch can be used. The option 'digital input' must be programmed at the factory. ( Data Sheet Page 1-1 ) Remote control functions via the digital input Remote control is done by means of a floating contact 'open collector' or a voltage. Standard is status control. The input is active (ON) at 0 V to 1 V and inactive (OFF) from 4 V to 30 V respectively in open condition Switch integrated in connecting cable Alternatively for the remote control a switch can be integrated in the connecting cable. Then remote control via the contact is no longer possible Programmable functions Turning on the laser If the laser is installed and enabled by the interface command 'LASER...' (ENABLE), the laser can be switched on via digital input. Example: 0 V or short circuit switches the laser on. Reset of the peak values The device continuously updates a maximum and a minimum value memory. With the digital input the values can be reset to the actual value. The function is helpful in case the analog output is programmed by the command 'OUT ' for reproduction of a peak value. Example: 0 V resets the maximum value memory to the actual value. Reset of the digital outputs If it is configured, the digital outputs can be reset via the digital input to the inactive state Use of the digital outputs DESCRIPTION OF COMMANDS: Command 2.14: Configuration of the digital outputs Example: 0 V will reset the digital output OPERATION AND APPLICATION KT15 II 05/03/12e 5-3

38 5.0.4 Use of the digital outputs The digital outputs display the alarm status of the Infrared Radiation Thermometer via the active level. The digital outputs can be configured that either 'Low' alarm or 'High' alarm or both alarms are transmitted. The digital outputs are also configurable as a synchronization output for the function 'Self-reset' of the maximum and minimum value memories. 'Inverse Logic' or 'Not Inverse Logic' can also be selected. For the reset of the digital output the following configurations can be selected: Reset by command Reset by digital input Reset by lower deviation of alarm level Reset by expiration of a programmable time Reset after lower deviation or exceed of threshold temperature Use of the digital input DESCRIPTION OF COMMANDS: Command 2.12: Configuration of digital input Command 2.14: Configuration of digital outputs Monitoring the device function during operation Monitoring the operating temperature The series KT15 II Infrared Radiation Thermometers have a built-in temperature switch which opens at 70 C. This switch is connected to the flange-plug ( chapter 3). Monitoring the analog output signal The type of the output signal can be selected for the series KT15 II Infrared Radiation Thermometers. For monitoring the output signal the '4-20 ma' type is recommended. At this setting a measuring temperature within the set temperature range will supply a signal between 4 and 20 ma. If the value falls below the calibrated temperature range, refer to the type sheet, 3.95 ma is supplied, and 21 ma is supplied if it is exceeded. Monitoring the power consumption The series KT15 II Infrared Radiation Thermometers draw a current of 90 ma to 150 ma at 24 V during operation. If this value is out of range, this indicates misbehaviour. Monitoring during operation with a serial interface For monitoring the Infrared Radiation Thermometer we recommend that the serial interface should be switched to the 'send the measured temperature value repeatedly'. As long as the Infrared Radiation Thermometer functions, the temperature value will be sent at the defined time interval. If it is out of range, the Infrared Radiation Thermometer will provide an error message OPERATION AND APPLICATION KT15 II 06/03/14e 5-4

39 5.0.6 Use of the maximum and minimum value memories The configuration of these parameters can be done by the software EasyConfig or EasyMeas. The maximum and minimum value memories are continually updated by the Infrared Radiation Thermometer. The stored data can be read by command, and also switched to the analog output. Commands for 'Memory functions' Commands for 'Configuration of analog output' If cyclically the temperature of moving material has to be determined by using the maximum or minimum value memory, the relevant memory must be reduced by a decay rate to the current measured value, or reset Decay rate With the command 'OUT..' the decay rate can be programmed Reset of memory (manually controlled) Reset with the command 'MEM RESET' Reset via digital input Self-reset For this the digital input must be programmed at the factory. Furthermore, the function memory reset must be programmed with the command 'CONFig DIGin..'. If an external reset signal is not available, the Infrared Radiation Thermometer generates this automatically. It offers alternatively the self-reset after a reset time, as well as after falling below a temperature threshold Reset after lower deviation of threshold temperature If there is a space (gap) between the targets, and the background temperature deviates to a sufficient extent of the temperature of the target, the Infrared Radiation Thermometer can generate the reset signal automatically. Therefore in the case of the maximum value memory the background temperature must be lower than the temperature of the target, in the case of the minimum value memory the background temperature must be higher than the temperature of the target. For synchronization of the evaluation, the generated reset signal can be switched to the digital output. The edge change occurs by exceeding (reset point maximum value memory) and falling below T lim ; the minimum value memory accordingly reversed (see figures 1 + 2). By selecting 'MANUAL' or 'AUTO' mode, the self-reset of the memories is activated. The device detects a 'temperature gap' between two targets and resets maximum or minimum value memory at the end of the gap OPERATION AND APPLICATION KT15 II 08/07/16e 5-5

40 Mode 'MANUAL' uses a programmable temperature threshold (T lim ) for detecting the gap. Mode 'AUTO' determines T lim automatically with the average of the maximum and minimum values. In this mode a reset time t res can be configured additionally. If during that time no target has been detected, the maximum resp. minimum value memory will be reset. If the time-controlled reset is not needed, the maximum reset time of 600 s should be adjusted. If the maximum value memory is enabled, the reset is done when T lim is exceeded again. If the minimum value memory is enabled, the reset is done when T lim has fallen below (see figures 1 + 2). Commands 'Configuration of digital output' in Universal Protocol. Please note: The time constant of the device may be set to auto-reset function only as large as the measurement is not distorted by integration! The extended memory configuration can be provided both via the digital interface and the analog output. Via the extended configuration of the analog output, the maximum or minimum value memory can be switched to the analog output. Commands 'Configuration of analog output' in Universal Protocol OPERATION AND APPLICATION KT15 II 08/07/16e 5-6

41 Example: Self-reset with maximum value memory, MANUAL mode Background Thick product Thin product Figure 1: Measurement Figure 2: Signal curve Reset after reset time The maximum or minimum value memories can be reset alternatively after the expiration of a programmable time. The reset time is retriggered automatically, when the value of the activated maximum value memory rises, resp. minimum value memory falls. This feature can be selected for only one of the memory types OPERATION AND APPLICATION KT15 II 08/07/16e 5-7

42 5.1 Operation via keyboard With the configuration menu, the most important parameters are set or interrogated. For full access to all parameters the software EasyConfig or EasyMeas is needed. LED on: up-/down-button is active menu- and value-display page up menu increase the value divide into submenu confirm the change back to the previous menu back without change Temp ESC Change Enter RS 232 RS 232 serial interface change someything page down menu reduce the value Operation of the Menu To activate the display press any button. The main menu starts (Temp). The LED between the up-/down-buttons shines. That means: these buttons are activated. With these buttons you can choose other menus or change values. Press Enter to go to a submenu in order to indicate settings and values. Press ESC to go back to the main menu without any change. Press Change to change settings or values. The submenu is flashing and the LED shines. Use the up-/down-buttons for selecting and changing. Confirm the selection with Enter. Feed the Code if required. Confirm the correctness of the Code with ok. Press ESC several times to go back to main menu or to switch off the menu. The menu automatically switches off after two minutes. The Code is valid until you leave the menu (display off) OPERATION AND APPLICATION KT15 II 05/03/12e

43 Main Menu Temp Lase 1.1 Enter 1.9 Enter ESC Change ESC Change RS 232 RS 232 Info Unit 1.2 Enter 2.0 Enter ESC Change ESC Change RS 232 RS 232 Emi AOut 1.3 Enter 2.1 Enter ESC Change ESC Change RS 232 RS 232 TAmb Uart 1.4 Enter 2.2 Enter ESC Change ESC Change RS 232 RS 232 Resp Cal 1.5 Enter 2.3 Enter ESC Change ESC Change RS 232 RS 232 Memo Conf 1.6 Enter 2.4 Enter ESC Change ESC Change RS 232 RS 232 HiAl Code 1.7 Enter 2.5 Enter ESC Change ESC Change RS 232 RS 232 LoAl 1.8 Enter ESC Change RS OPERATION AND APPLICATION KT15 II 05/03/12e 5-9

44 Submenu Temp Temp Enter Enter 350 ESC ESC Change C RS 232 Temp Temperature indicator Submenu Info Info Enter Enter KT15 SeNr ESC ESC Change.82A 1111 C C RS 232 Info device information KT15 pyrometer series.82a spectral range, detector type -50 temperature range beginning 1000 temperature range end Submenu Emi Emi Enter Enter Emi Change Enter Emi ESC ESC Change ESC RS 232 Emi emissivity OPERATION AND APPLICATION KT15 II 05/03/12e

45 Submenu Tamb TAmb Enter Enter Int Change Int Enter Int ESC ESC Change ESC ESC RS 232 C C Only with option ext. temp. input Ext Enter Ext ESC or no ESC ESC 150 Ext C Configuration of the external input with HEITRONICS Software EASY CONFIG via RS232 serial interface. Man Enter Valu Enter Man Change 75 Enter Man ESC Tamb ambient temperature flash Int internal temperature ESC ESC ESC ESC Man manual input C C Ext external sensor OPERATION AND APPLICATION KT15 II 05/03/12e

46 Submenu Resp Resp ESC Change Resp response time Enter Enter 5ms Change 5ms ESC. RS s Enter 1s ESC. ESC.. 600s Submenu Memo Memo Enter Enter Hold Change Hold ESC Change 100 Hardware-reset only with option Digital Input RS 232 C ESC Max Enter Max ESC Min 100 C Memo memory Diff Hold momentary value Max maximal value Actu Min minimum value Diff difference value Aktu actual value (In this menu the button ENTER is also button for HOLD and RESET) OPERATION AND APPLICATION KT15 II 05/03/12e

47 Submenu HiAl HiAl Enter Enter on Change on ESC Change ESC off Enter off ESC RS 232 HiAl high alarm Valu Enter 20 Change Enter 55 ESC C C On switched on ESC ESC Off switched off Valu temperature value Submenu LoAl LoAl Enter Enter on Change on ESC Change ESC off Enter off ESC RS 232 LoAl low alarm Valu Enter 20 Change Enter 45 ESC On switched on C C Off switched off ESC ESC Valu temperature value OPERATION AND APPLICATION KT15 II 05/03/12e

48 Submenu Laser Lase ESC Change Enter Enter on Change on ESC off Enter off ESC RS 232 Lase Laser ESC on switched on off switched off Submenu Unit Unit ESC Change Enter Enter C Change C ESC K Enter K ESC RS 232 flasch Unit Units F ESC C Celsius K Kelvin F Fahrenheit OPERATION AND APPLICATION KT15 II 05/03/12e

49 Minimum temperature difference: depends on final temperature The following values will result: Minimum temperature Final temperature difference 150 C 50 C 200 C 100 C 1000 C 200 C > 1000 C 400 C Submenu Aout AOut ESC Change Enter Enter TeRa Enter Temp Change Temp ESC ESC Rad Enter Rad ESC RS 232 ESC Rang Enter 0 Change low Change 0 20 Enter 20 ESC ESC ESC C C high Change Enter 0 ESC 400 C ma/v Enter 0_20 Change 0_20 ma ESC ESC 4_20 0_1 Enter 0_1 ESC 0_10 V Out Enter Actu Change Actu Aout analogue output ESC ESC Max Actu actual value DeRa decay rate high high temperature range Min Enter Min ESC K/s Kelvin/second low low temperature range ma/v current-/voltage output Max maximum value Min minimum value DeRa Enter Rate Change 0 2 Enter Rate ESC Out output Rad radiation 0 2 Rang temperature range ESC Rate decay rate value K/s K/s Temp temperature radiant TeRa temperature radiant OPERATION AND APPLICATION KT15 II 05/03/12e

50 Submenu Uart Uart Enter Enter BauR Change 9,6 ESC Change 9,6 Enter 19,2 ESC ESC 19 RS 232 ESC Data Change 8 Enter 8 ESC 8 ESC 7 ESC Stop Change 1 Enter 1 ESC 1 ESC 2 ESC Pari Change even Enter even ESC Uart digital output even BauR baudrate ESC odd ESC Data data bits even even Hand handshake odd odd Pari parity Hand ESC Stop stopbits flasch Hand with hardware handshake yes OPERATION AND APPLICATION KT15 II 05/03/12e

51 Submenu Cal Cal Enter Enter show Enter CalF Change Enter 1250 ESC ESC Change CalF 1254 ESC ESC RS 232 Cal calibration show show Enter CalT Change 348o 350o Enter 350o ESC CalF indication of cal-factor Temp 348o show ESC ESC Temp indication of temperature value C OPERATION AND APPLICATION KT15 II 05/03/12e

52 Submenu Conf Conf Enter Enter Mode Enter Mode Change Stnd ESC Change Stnd SC12 Enter Mode ESC RS 232 SC12 ESC ESC Pmtr Enter Dflt Enter???? ESC Conf Configuration Mode mode Stnd standard mode SC12 scanner mode Tamb ambient temperature Pmtr parameter Dflt delivery data OPERATION AND APPLICATION KT15 II 05/03/12e

53 Code test Code ESC Change Enter Enter 0 1 Enter Enter Enter Enter OK RS 232 back to menu Code Code input Such as example: Code 1234 Press ENTER when Code is questioned. At first feed the left numeral (f.e.: 1) of the code with the buttons and confirm with ENTER. Then the next numeral and so on. To go back use button ESC OPERATION AND APPLICATION KT15 II 05/03/12e

54 5.2 Communication via interface Preparing the device for operation with interface RS232C The interface signals are available at the 12-pin plug of the Infrared Radiation Thermometer KT15II. The signals are connected to the computer interface as follows: KT15II computer Name plug-in contact 9 pole 25 pole name RXD J 3 2 TXD TXD G 2 3 RXD RTS H 8 5 CTS CTS A 7(4) 4(20) RTS(DTR) DTR F 6 6 DSR GND K 5 7 GND The transfer is controlled by means of hardware handshake (RTS/CTS). Some RS232 drivers use DTR instead of RTS for the handshake. In this case, CTS of Infrared Radiation Thermometer KT15II must be connected to DTR of the computer. The transfer parameters baud rate, number of data bits, parity, stop bits and delimiter are set by means of the operating keys at the Infrared Radiation Thermometer KT15II. These parameters are to be set to equal values as for the Infrared Radiation Thermometer KT15II and the computer. Selectable baud rates: 9.6 kbaud 19.2 kbaud 38.4 kbaud 57.6 kbaud kbaud Possible data formats: Number of data bits: 7 or 8 Number of stop bits: 1 or 2 Parity: even, uneven or none The Infrared Radiation Thermometer KT15II is able to receive data with one stop bit independent of the setting. For transmission, the selected number of stop bits is inserted Using the interface Control of communication Data transfer from evaluation to Infrared Radiation Thermometer KT15II The ready-to-receive state of the Infrared Radiation Thermometer KT15II is signalized via line RTS. The coding shown in the following table shall apply OPERATION AND APPLICATION KT15II 15/03/14e

55 Function ready-to-receive evaluation of a command receive buffer overflow RTS-state on (high) off (low) off (low) Data which are sent during the RTS-state is OFF are lost. Data transfer from Infrared Radiation Thermometer KT15II for evaluation purposes Before the Infrared Radiation Thermometer KT15II sends a message, the CTS input is checked. If the CTS-state is on (high), the whole message including the selected delimiter is sent. Commands The individual commands are transmitted as a text in ASCII characters. As for instruction characters, only the first three characters will be checked, numerical values are accepted with or without fractional parts, blanks must not be filled in between operational sign and numerical value. An answer is only given in case of interrogation (?). If a command cannot be read, an error message will be output. If parameters are not found to be within the permissible range of values (e.g. left-hand end of extent < right-hand end of extent), the command will not be executed and an error message will be output. Delimiter Infrared Radiation Thermometer KT15II identifies the ASCII control characters carriage return (0D) and line feed (0A) as end of instruction. As soon as the end of instruction has been recognized, KT15II will reset the ready-toreceive state (RTS) until the command is evaluated. The answer of the Infrared Radiation Thermometer KT15II is (0D). Input buffer Infrared Radiation Thermometer KT15II has an input buffer of 40 characters. When a longer command has been received, KT15II issues an error message Description of the commands The commands for programming the parameters are described in separate instructions: Universal Protocol Communication via Serial Interface for all digital HEITRONICS Infrared Radiation Thermometers OPERATION AND APPLICATION KT15II 15/03/14e

56 6 MAINTENANCE AND CALIBRATION 6.1 General information The design of HEITRONICS Infrared Radiation Thermometers provides for a long period of service without the necessity of special maintenance. It is advisable to check the display accuracy approximately every three months. Measuring errors may arise, for instance, owing to pollution of the lens. For this reason, this component should first of all be cleaned. 6.2 Cleaning the lens For this purpose a cleaning- and servicing kit is available ( OPTIONS AND ACCES- SORIES). Fine dust can be brushed off by means of a lens brush or blown off the lens carefully. Do not use compressed air or canned compressed air which could cause residues on the objective. If the lens is heavily soiled and has greasy deposits, a microfiber cloth, cotton tabs and the special lens cleaner will help cleaning it, after having used the lens brush. 6.3 Testing the display accuracy The display accuracy of the Infrared Radiation Thermometer can be checked by means of a blackbody calibration source. This test is carried out within the range indicated on the type plate. It is advisable to carry out the test under high temperature. For this purpose, the Infrared Radiation Thermometer is placed in front of a black body calibration source so that the integrated optical equipment is focused on the radiator. The temperature of the black body calibration source must be determined. If a black body calibration Source of type HEITRONICS SW15 ( OPTIONS AND ACCESSORIES) is used, it has to be mounted to the lens. The Infrared Radiation Thermometer has to be connected to an appropriate power supply. Approximately 15 minutes after appliance of the supply voltage the Infrared Radiation Thermometer is ready for calibration. For checking we recommend to use the following minimum temperatures: Model Recommended minimum temperature C KT15.01 II 580 to 600 KT15.02 II 580 to 600 KT15.21 II 280 to 300 KT15.23 II 130 to 150 KT15.24 II 130 to 150 KT15.25 II 130 to 150 KT15.41 II KT15.42 II KT15.43 II KT15.62 II KT15.63 II KT15.69 II KT15.81 II KT15.82 II KT15.83 II KT15.85 II 580 to to to to to to to to to to MAINTENANCE AND CALIBRATION KT15 II 03/07/16e

57 8 FIGURES Nr. Page 10a Spectral emissivity of various materials b Total emissivity of some materials at a temperature of 20 C (spectral range µm) Emissivity of various foils Spectral emissivity of Infrared Radiation Thermometers and transmission curves of various synthetic materials Spectral emissivity ε, transmissivity ε and reflection ρ of glass Dimensions KT15 II Target Diameter Transmission curves (illustrated in a diagram): Spectral Response µm Spectral Response µm Spectral Response 3.43 µm Spectral Response 6.8 µm Spectral Response 7.93 µm Spectral Response 8.05 µm Spectral Response 3.87 µm Spectral Response µm Spectral Response µm Spectral Response 4.24 µm Spectral Response 4.48 µm Spectral Response 4.66 µm Spectral Response 8 10 µm Spectral Response 8 14 µm Spectral Response 8 20 µm Spectral Response µm FIGURES KT15 II 04/03/14e 8-1

58 Fig. 10a Spectral emissivity of various materials Fig. 10b Total emissivity of some materials at a temperature of 20 C (spectral range µm) FIGURES KT15 II 02/06/08e 8-2

59 Fig. 12 Emissivity of various foils FIGURES KT15 II 02/06/08e 8-3

60 PE: Polyäthylen PET: Polyester PTFE: Polytetrafluoräthylen PVC: Polyvinylchlorid Fig. 13 Spectral emissivity of Infrared Radiation Thermometers and transmission curves of various synthetic materials FIGURES KT15 II 04/03/14e 8-4

61 Fig. 14 Spectral emissivity ε, transmissivity ε and reflection ρ of glass FIGURES KT15 II 02/06/08e 8-5

62 Fig. 15 Housing dimensions Infrared Radiation Thermometer KT15 II FIGURES KT15 II 04/03/14e 8-6

63 Fig. 30 no longer required FIGURES KT15 II 03/07/13e 8-7

64 Example: Detector Type A µm µm µm 1 lens K6 2 lens L6 3 lens M6 4 lens N6 Fig. 31 Target Diameter KT15.8x II Detector Type A FIGURES KT15 II 02/06/08e 8-8

65 Spektrale Empfindlichkeit SP01 ( µm) Spectral Response SP01 ( µm) 30 Relative Empfindlichkeit [%] Relative Sensitivity [%] ,5 1 1,5 2 2,5 3 3,5 Wellenlänge [µm] Wavelength [µm] Spektrale Empfindlichkeit SP02 ( µm) Spectral Response SP02 ( µm) Relative Empfindlichkeit [%] Relative Sensitivity [%] Wellenlänge [µm] Wavelength [µm] FIGURES KT15 II 02/06/08e 8-9

66 Spektrale Empfindlichkeit SP21 (3.43 µm) Spectral Response SP21 (3.43 µm) 35 Relative Empfindlichkeit [%] Relative Sensitivity [%] ,5 2,7 2,9 3,1 3,3 3,5 3,7 3,9 4,1 4,3 4,5 Wellenlänge [µm] Wavelength [µm] Spektrale Empfindlichkeit SP23 (6.8 µm) Spectral Response SP23 (6.8 µm) 30 Relative Empfindlichkeit [%] Relative Sensitivity [%] ,2 6,4 6,6 6,8 7 7,2 7,4 7,6 7,8 8 Wellenlänge [µm] Wavelength [µm] FIGURES KT15 II 02/06/08e

67 Spektrale Empfindlichkeit SP24 (7.93 µm) Spectral Response SP24 (7.93 µm) 70 Relative Empfindlichkeit [%] Relative Sensitivity [%] ,2 7,4 7,6 7,8 8 8,2 8,4 8,6 8,8 9 Wellenlänge [µm] Wavelength [µm] Spektrale Empfindlichkeit SP25 (8.05 µm) Spectral Response SP25 (8.05 µm) Relative Empfindlichkeit [%] Relative Sensitivity [%] ,2 7,4 7,6 7,8 8 8,2 8,4 8,6 8,8 9 Wellenlänge [µm] Wavelength [µm] FIGURES KT15 II 02/06/08e

68 Spektrale Empfindlichkeit SP41 (3.87 µm) Spectral Response SP41 (3.87 µm) Relative Empfindlichkeit [%] Relative Sensitivity [%] ,2 3,4 3,6 3,8 4 4,2 4,4 4,6 4,8 5 Wellenlänge [µm] Wavelength [µm] Spektrale Empfindlichkeit SP42 ( µm) Spectral Response SP42 ( µm) 35 Relative Empfindlichkeit [%] Relative Sensitivity [%] ,2 4,4 4,6 4,8 5 5,2 5,4 5,6 5,8 6 Wellenlänge [µm] Wavelength [µm] FIGURES KT15 II 02/06/08e

69 Spektrale Empfindlichkeit SP43 ( µm) Spectral Response SP43 ( µm) Relative Empfindlichkeit [%] Relative Sensitivity [%] ,2 7,4 7,6 7,8 8 8,2 8,4 8,6 8,8 9 Wellenlänge [µm] Wavelength [µm] Spektrale Empfindlichkeit SP61 (4.24 µm) Spectral Response SP61 (4.24 µm) Relative Empfindlichkeit [%] Relative Sensitivity [%] ,2 3,4 3,6 3,8 4 4,2 4,4 4,6 4,8 5 Wellenlänge [µm] Wavelength [µm] FIGURES KT15 II 02/06/08e

70 Spektrale Empfindlichkeit SP62 (4.48 µm) Spectral Response SP62 (4.48 µm) Relative Empfindlichkeit [%] Relative Sensitivity [%] ,5 3,7 3,9 4,1 4,3 4,5 4,7 4,9 5,1 5,3 5,5 Wellenlänge [µm] Wavelength [µm] Spektrale Empfindlichkeit SP63 (4.66 µm) Spectral Response SP63 (4.66 µm) Relative Empfindlichkeit [%] Relative Sensitivity [%] ,5 3,7 3,9 4,1 4,3 4,5 4,7 4,9 5,1 5,3 5,5 Wellenlänge [µm] Wavelength [µm] FIGURES KT15 II 02/06/08e

71 Spektrale Empfindlichkeit SP81 (8 10 µm) Spectral Response SP81 (8..10 µm) 60 Relative Empfindlichkeit [%] Relative Sensitivity [%] Wellenlänge [µm] Wavelength [µm] Spektrale Empfindlichkeit SP82 (8 14 µm) Spectral Response SP820 (8 14 µm) Relative Empfindlichkeit [%] Relative Sensitivity [%] Wellenlänge [µm] Wavelength [µm] FIGURES KT15 II 02/06/08e

72 Spektrale Empfindlichkeit SP83 (8 20 µm) Spectral Response SP83 (8 20 µm) Relative Empfindlichkeit [%] Relative Sensitivity [%] Wellenlänge [µm] Wavelength [µm] FIGURES KT15 II 05/10/16e

73 GARANTIEBEDINGUNGEN Die (nachfolgend HEITRONICS genannt) haftet unter Ausschluss weitergehender Ansprüche für Mängel an den von ihr gelieferten Infrarot-Strahlungsthermometern und deren Zubehör, und zwar für die Dauer von 24 Monaten nach Maßgabe folgender Bedingungen: 1. Die Mängelhaftung erstreckt sich ausschließlich auf kostenlosen Ersatz fehlerhafter Teile im Hause HEITRONICS. Die Mängelhaftung bezieht sich insbesondere nicht auf natürliche Abnutzung und nicht auf Schäden, die auf unsachgemäßer Bedienung oder Beanspruchung oder sonstigen von HEITRONICS nicht verschuldeten Umständen beruhen. Die Mängelhaftung gilt nicht für Batterien. Das Gerät ist in der Originalverpackung frachtfrei an HEITRONICS zu senden. Kosten für Steuern, Gebühren und Zölle trägt der Versender. Transportschäden gehen zu Lasten des Versenders. 2. Die Frist für die Mängelhaftung beginnt mit dem Tage des Geräteversandes aus dem Hause HEITRONICS. 3. Etwa auftretende Mängel sind HEITRONICS unverzüglich zu melden, um weitergehende Auswirkungen möglichst zu vermeiden. 4. Ersetzte Teile gehen in das Eigentum von HEITRONICS über. Für Ersatzteile leistet HEITRONICS bis zum Ablauf der für den ursprünglichen Liefergegenstand geltenden Frist in der vorgenannten Weise Gewähr. 5. Alleiniger Gerichtsstand für alle sich aus der Mängelhaftung ergebenden Streitigkeiten ist Wiesbaden, Deutschland. WARRANTY CONDITIONS Temperature measuring equipment delivered by (hereinafter referred to as HEITRONICS) is warranted against defects, excluding consequential liability, notably for a period of 24 months subject to the following conditions: 1. Warranty is limited to the free replacement of defective parts at its works. In particular, warranty does not cover normal wear and tear or damage due to improper use or overloading or other circumstances for which HEITRONICS is not responsible. Warranty does not include batteries. The device must be returned to HEITRONICS in original packing, carriage paid. Costs for taxes, fees and customs duties are to be paid by sender. Shipping damage is borne by the sender. 2. The warranty period starts from the date of delivery from its works. 3. Information concerning eventually encountered defects has to be forwarded to HEITRONICS immediately to preclude possible consequential damage. 4. Replaced parts or components are returned to the property of HEITRONICS. Replacements are warranted on the conditions mentioned above until the expiration of the warranty period for the originally delivered equipment. 5. Jurisdiction for any legal dispute arising from this warranty shall be limited to the Court District of Wiesbaden, Germany. CONDITIONS DE GARANTIE La garantie de (ci-après mentionnée HEITRONICS) couvre les défauts des radiomètres et accessoires livrés par elle, à l'exclusion de toute autre réclamation, pour une durée de 24 mois dans les conditions suivantes: 1. La responsabilité de HEITRONICS est limitée au remplacement gratuit des pièces défectueuses dans les usines de HEITRONICS. La garantie ne couvre pas le cas d'usure normale, non plus les dommages provoqués par fausse manoeuvre, par des conditions de travail trop dures ou des circonstances dont HEITRONICS n'est pas responsable. Les batteries ne sont pas sous garantie. L'appareil doit être retourné à HEITRONICS dans son emballage d'origine, port payé. L expéditeur paye pour les frais des impôts, des taxes et des droits de douane. Dommages de transport sont à la charge de l'expéditeur. 2. La période de garantie commence le jour d'expédition des appareils par les usines HEITRONICS. 3. Tout défaut doit être signalé à HEITRONICS de toute urgence pour éviter des conséquences plus graves. 4. Les pièces échangées deviennent propriété de HEITRONICS. Les pièces de rechange bénéficient de la garantie dans les conditions mentionnées ci avant, jusqu'à l'expiration de la période prévue pour la livraison d'origine. 5. Pour tous litiges qui pourraient naître de l'application de la garantie, la seule juridiction compétente sera celle de Wiesbaden, R.F.A. Garantiebedingungen 04/08/14 def

74 HEITRONICS INFRAROT MESSTECHNIK GMBH IRM SERVICE Lieferanschrift / Delivery address / Adresse de livraison / Dirección de entrega: Kreuzberger Ring WIESBADEN GERMANY Tel.: Fax: info@heitronics.com Internet: Vertriebsorganisation Angaben zu unseren regionalen Vertriebspartnern finden Sie im Internet. Sales Network For details about our regional representatives, please, refer to the internet. Réseau des ventes Vous pouvez trouver les coordonnées de nos représentants régionaux sur Internet. Organizacion de la venta Informaciónes referente a nuestros regionales colaboradores de venta encuentran en el internet.

Float switch For industrial applications, plastic version Model RLS-2000

Float switch For industrial applications, plastic version Model RLS-2000 Level measurement Float switch For industrial applications, plastic version Model RLS-2000 WIKA data sheet LM 50.04 Applications Level measurement of liquids in machine building Control and monitoring

More information

LASER. Analog Laser Displacement Transducer. LAM Series. Key-Features: Content:

LASER. Analog Laser Displacement Transducer. LAM Series. Key-Features: Content: LASER Analog Laser Displacement Transducer LAM Series Key-Features: Content: Overview, Measuring Principle...2 Installation Instructions...3 Technical Data...4 Technical Drawings.7 Electrical Connection...9

More information

RoHS. High shaft load capacity. Shock / vibration resistant

RoHS. High shaft load capacity. Shock / vibration resistant Due to their sturdy bearing construction in Safety Lock Design, the Sendix 5000 and 5020 offer high resistance against vibration and installation errors. The rugged housing, high protection level of up

More information

Absolute Encoders - Singleturn

Absolute Encoders - Singleturn The Sendix 5853 and Sendix 5873 singleturn encoders with SSI or BiSS interface and optical sensor technology can achieve a resolution of max. 7 bits. These encoders are also available with an optional

More information

Rotary Measurement Technology Absolute Encoders, Multiturn

Rotary Measurement Technology Absolute Encoders, Multiturn Mechanical drive Safety-LockTM High rotational speed -40 to 90 C IP Temperature High IP High shaft load capacity Shock/ vibration resistant Magnetic field proof Short-circuit proof Reverse polarity protection

More information

High-quality pressure transmitter For general industrial applications Model S-20

High-quality pressure transmitter For general industrial applications Model S-20 Electronic pressure measurement High-quality pressure transmitter For general industrial applications Model S-20 WIKA data sheet PE 81.61 Applications Critical industrial applications Demanding applications

More information

Metis H311 / H322. Highly Advanced, Full Featured 2-Color High-Speed Pyrometers

Metis H311 / H322. Highly Advanced, Full Featured 2-Color High-Speed Pyrometers Metis H311 / H322 Highly Advanced, Full Featured 2-Color High-Speed Pyrometers High-Speed 2-color pyrometer for non-contact temperature measurement in short wavelength spectral range, primarily for measurements

More information

Standard Sendix 5000 / 5020 (shaft / hollow shaft) Push-Pull / RS422 / Open collector. Robust performance. Many variants

Standard Sendix 5000 / 5020 (shaft / hollow shaft) Push-Pull / RS422 / Open collector. Robust performance. Many variants ue to their sturdy bearing construction in Safety-ock esign, the Sendix 5000 and 500 offer high resistance against vibration and installation errors. The rugged housing, high protection level of up to

More information

OEM pressure sensor For industrial applications Models O-10 (T), O-10 (5)

OEM pressure sensor For industrial applications Models O-10 (T), O-10 (5) Pressure OEM pressure sensor For industrial applications Models O-10 (T), O-10 (5) WIKA data sheet PE 81.65 for further approvals see page 11 Applications Hydraulics and pneumatics Pumps and compressors

More information

Distance sensors. System description. Distance measurement using triangulation

Distance sensors. System description. Distance measurement using triangulation Distance sensors System description Distance measurement using triangulation The measurement principle of optical triangulation is suitable for the precise determination of distances at close range. With

More information

Phone: Fax: Web: -

Phone: Fax: Web:  - ue to their sturdy bearing construction in Safety-ock esign, the Sendix 5000 and 500 offer high resistance against vibration and installation errors. The rugged housing, high protection level of up to

More information

Metis M311 / M Color Pyrometer Series

Metis M311 / M Color Pyrometer Series Metis M311 / M322 2-Color Pyrometer Series Self contained 2-color pyrometer for non-contact temperature measurement in short wavelength spectral range, primarily for measurements on metals and bright and

More information

LA-T SERIES. Fast and reliable minute granular flaw detection in winding wires. Winding Wire Granular Flaw Detector PARTICULAR USE SENSORS

LA-T SERIES. Fast and reliable minute granular flaw detection in winding wires. Winding Wire Granular Flaw Detector PARTICULAR USE SENSORS SERIES Winding Wire Granular Flaw Detector Orders accepted till September, 2003 Production to be discontinued from April, 2004 Fast and reliable minute granular flaw detection in winding wires Slim Reliable

More information

Rotary Measurement Technology Incremental Encoders

Rotary Measurement Technology Incremental Encoders -20 to 60 C Temperature Shock/vibration resistant Short-circuit protection Reverse polarity protection High rotational speed Rugged Balanced, stainless-steel clamping rings, special bearing-shaft connection

More information

Absolute Encoders Multiturn

Absolute Encoders Multiturn The Sendix F36 multiturn with the patented Intelligent Scan Technology is an optical multiturn encoder in miniature format, without gears and with 00% insensitivity to magnetic fields. With a size of just

More information

Absolute Encoders - Singleturn

Absolute Encoders - Singleturn The Sendix 5 and Sendix 7 singleturn encoders with SSI or BiSS-C interface and optical sensor technology can achieve a resolution of max. 7 bits. These encoders are also available with an optional SinCos

More information

High-quality pressure transmitter For general industrial applications Model S-10

High-quality pressure transmitter For general industrial applications Model S-10 Electronic pressure measurement High-quality pressure transmitter For general industrial applications Model S-10 WIKA data sheet PE 81.01 for further approvals see page 4 Applications Machine building

More information

Precipitation Monitor ,

Precipitation Monitor , THE WORLD OF WEATHER DATA - THE WORLD OF WEATHER DATA - THE WORLD OF WEATHER DATA Instruction for use 021197/11/09 Precipitation Monitor 5.4103.10.000, 5.4103.10.700 ADOLF THIES GmbH & Co. KG Hauptstraße

More information

M70LL Laser Distance Sensor

M70LL Laser Distance Sensor M7LL Laser Distance Sensor for automated manufacturing with Ethernet interface to connect with PLC The analog sensors of series M7LL use a PSD for their receiver optics. This assures a position measurement

More information

Metis M311 / M322. Highly Advanced, Full Featured 2-Color Pyrometer Series

Metis M311 / M322. Highly Advanced, Full Featured 2-Color Pyrometer Series Metis M311 / M322 Highly Advanced, Full Featured 2-Color Pyrometer Series Self contained 2-color pyrometer for non-contact temperature measurement in short wavelength spectral range, primarily for measurements

More information

Absolute Encoders Multiturn

Absolute Encoders Multiturn Absolute Encoders Multiturn Functional Safety, optical The absolute multiturn encoders Sendix 5863 SIL and 5883 SIL are perfectly suited for use in safety-related applications up to SIL3 according to DIN

More information

FEATURES. 15 m ft. 4 m ft. 2 m ft. 100 mm in. 600 mm in. 15 m ft. 4 m ft. 2 m ft. 100 mm 3.

FEATURES. 15 m ft. 4 m ft. 2 m ft. 100 mm in. 600 mm in. 15 m ft. 4 m ft. 2 m ft. 100 mm 3. 273 Cylindrical Photoelectric Sensor SERIES Related Information General terms and conditions... F-3 Glossary of terms... P.1549~ guide... P.231~ General precautions... P.1552~ PHOTO PHOTO MEASURE -LS200

More information

Cable transducer Analog output / redundant Measuring length absolute up to 5 m

Cable transducer Analog output / redundant Measuring length absolute up to 5 m Features Magnetic sensing method Resolution: 12 bit Interface analog 0...10 V / 0.. V /...20 ma Measuring length up to m Removable stickers for water outlet Teach inputs Extremely light thanks to housing

More information

ENCODER. SSI Multiturn. Series , Key-Features: Content:

ENCODER. SSI Multiturn. Series , Key-Features: Content: ENCOER SSI Multiturn Series 8.5863, 8.5883 Content: Order Code Shaft... Order Code Hollow Shaft...3 Order Code Blind Hollow Shaft...3 Connection...4 Technical rawing...5 Technical ata...8 Key-Features:

More information

Absolute encoders - SSI

Absolute encoders - SSI with through hollow shaft Features Encoder multiturn / SSI Optical sensing method Resolution: singleturn 14 bit, multiturn 12 bit Compact design Cost-efficient mounting High reliability by self-diagnostics

More information

XXXX e. X d.. X X X a

XXXX e. X d.. X X X a The Heavy Duty incremental encoder type 0H boasts a high degree of ruggedness in a very compact design. Its special construction makes it perfect for all applications in very harsh environments. / RoHS

More information

Precipition Sensor with analogue Intensity Output ,

Precipition Sensor with analogue Intensity Output , THE WORLD OF WEATHER DATA - THE WORLD OF WEATHER DATA - THE WORLD OF WEATHER DATA Instruction for use 021335/11/09 Precipition Sensor with analogue Intensity Output 5.4103.20.041, 5.4103.20.741 ADOLF THIES

More information

Technical data. General specifications V DC No-load supply current I 0. typ. 50 ma Power consumption P 0

Technical data. General specifications V DC No-load supply current I 0. typ. 50 ma Power consumption P 0 Model Number SYNCHRON SERIELLES INTERFACE Cable pull rotary encoder with SSI interface Features Solid yet lightweight plastic construction Compact, slim design (the shaft of the mounted rotary encoder

More information

XXXX e. X d.. X X X a

XXXX e. X d.. X X X a The Heavy uty incremental encoder type 0H boasts a high degree of ruggedness in a very compact design. Its special construction makes it perfect for all applications in very harsh environments. / RoHS

More information

CD4 Series with linear image sensor and electronic shutter provides accurate measurement.

CD4 Series with linear image sensor and electronic shutter provides accurate measurement. 3 series Series with linear image sensor and electronic shutter provides accurate measurement. Laser displacement sensor features easy setup and operation. High accuracy of.1µm resolution and ±.1% F.S.

More information

Z500. High-Precision Sensor that Measures and Displays an Object's Profile. Profile Measuring System

Z500. High-Precision Sensor that Measures and Displays an Object's Profile. Profile Measuring System Profile Measuring System High-Precision Sensor that Measures and Displays an Object's Profile. Features OMRON's original line beam method provides a complete solution to profile measurement problems. Conventional

More information

Absolute Encoders Singleturn

Absolute Encoders Singleturn The absolute singleturn encoders Sendix 5853 SIL and 5873 SIL are perfectly suited for use in safety-related applications up to SIL3 according to DIN EN ISO 6800-5- or PLe to DIN EN ISO 3849. The extra

More information

Rotary Position Technology Incremental Encoders

Rotary Position Technology Incremental Encoders -40 to 80 C Temperature Shock/vibration resistant Short-circuit protected Reverse polarity protection High rotational speed Rugged Balanced, stainless-steel clamping rings, special bearing-shaft connection

More information

Pressure transmitter For general industrial applications Model A-10

Pressure transmitter For general industrial applications Model A-10 Electronic pressure measurement Pressure transmitter For general industrial applications Model A-10 WIKA data sheet PE 81.60 for further approvals see page 9 Applications Machine building Shipbuilding

More information

Rotary Measuring Technology

Rotary Measuring Technology SSI or parallel interface Highest shock resistance on the market ( 2500 m/s2, 6 ms acc. to DIN IEC 68-2-27) Divisions: up to 16384 (14 bits), singleturn Housing Ø 58 mm IP 66 Various options (e.g. LATCH,

More information

Wind Direction Transmitter First Class analogue output x0.1xx

Wind Direction Transmitter First Class analogue output x0.1xx THE WORLD OF WEATHER DATA - THE WORLD OF WEATHER DATA - THE WORLD OF WEATHER DATA Instruction for Use 021514/03/09 Wind Direction Transmitter First Class analogue output 4.3150.x0.1xx ADOLF THIES GmbH

More information

200W 500W, Air Cooled QUBE Fiber Lasers

200W 500W, Air Cooled QUBE Fiber Lasers Technical Specification 200W 500W, Air Cooled QUBE Fiber Lasers CONTENTS 1.1 SCOPE 2 1.2 OPTICAL SPECIFICATION 2 1.3 BEAM DELIVERY FIBER SPECIFICATION 3 1.4 ALIGNMENT LASER 4 1.5 POWER DISTRIBUTION 4 1.6

More information

ENCODER. Analog Multiturn Encoder. Serie 8.M5861. Key-Features:

ENCODER. Analog Multiturn Encoder. Serie 8.M5861. Key-Features: ENCODER Multiturn Encoder Serie 8.M5861 Key-Features: Content: Order Code...2 Technical Data...3 Outputs & Connection... Dimensions...5 - Shaft with max. 10 mm diameter - 0...10 V,...20 ma, 0...5 V analog

More information

BML-S1F_-A/Q _-M _-_0-KA Magnetic Linear Encoder System. User's Guide. Magnetic Linear Encoder System BML-S1F_-A/Q _-M _-_0-KA

BML-S1F_-A/Q _-M _-_0-KA Magnetic Linear Encoder System. User's Guide. Magnetic Linear Encoder System BML-S1F_-A/Q _-M _-_0-KA User's Guide BML-S1F_-A/Q _-M _-_0-KA Balluff GmbH Schurwaldstraße 9 73765 Neuhausen a.d.f. Deutschland Tel. +49 7158 173-0 Fax +49 7158 5010 balluff@balluff.de www.balluff.com 1 Table of Contents 1 Safety

More information

Distance Sensors BOD 26K.... with optional Master-Slave-Mode

Distance Sensors BOD 26K.... with optional Master-Slave-Mode Distance... with optional Master-Slave-Mode Laser Distance The Balluff distance sensor not only detects the presence of objects, but can also indicate their precise position or location. This means the

More information

LASER. Analog Laser Displacement Transducer. LAM Series. Key-Features: Content:

LASER. Analog Laser Displacement Transducer. LAM Series. Key-Features: Content: LASER Analog Laser Displacement Transducer LAM Series Key-Features: Content: Measuring Principle...2 Installation Instructions...3 Technical Data LAM-S...4 Technical Data LAM-F...5 Technical Drawing...6

More information

OEM pressure transmitter for general industrial applications Models O-10 (T), O-10 (5)

OEM pressure transmitter for general industrial applications Models O-10 (T), O-10 (5) Electronic pressure measurement OEM pressure transmitter for general industrial applications Models O-10 (T), O-10 (5) WIKA data sheet PE 81.65 Applications Hydraulics and pneumatics Pumps and compressors

More information

SITRANS F flowmeters SITRANS F US. Transmitter FUS060. Overview

SITRANS F flowmeters SITRANS F US. Transmitter FUS060. Overview Overview Function Displays and keypad Operation of the SITRANS FUS060 transmitter can be carried out using: Keypad and display unit HART communicator PC/laptop and SIMATIC PDM software via HART communication

More information

Preliminary Specification

Preliminary Specification Preliminary Specification CONTENTS 1.1 SCOPE 2 1.2 OPTICAL SPECIFICATION 2 1.3 BEAM DELIVERY FIBER SPECIFICATION 3 1.4 ALIGNMENT LASER 3 1.5 POWER DISTRIBUTION 3 1.6 WATER COOLING REQUIREMENTS 4 1.7 CONTROL

More information

LASER. Analog Laser Displacement Transducer. LAS Series. Key-Features:

LASER. Analog Laser Displacement Transducer. LAS Series. Key-Features: LASER Analog Laser Displacement Transducer LAS Series Key-Features: Content: Measurement Principle & Installation...2 Overview of the Series...3 Technical Data.4 Technical Drawing...7 Linearity & Resolution...8

More information

Laser diffuse reflection light scanner with background suppression. Dimensioned drawing

Laser diffuse reflection light scanner with background suppression. Dimensioned drawing Specifications and description Laser diffuse reflection light scanner with background suppression Dimensioned drawing We reserve the right to make changes DS_HRTL3B_en.fm en 01-2010/12 50114049 10-30 V

More information

PD30CNB25xxPS. Photoelectrics, Background suppression reflective - PointSpot. Main features. Main functions. Description

PD30CNB25xxPS. Photoelectrics, Background suppression reflective - PointSpot. Main features. Main functions. Description Photoelectrics, Background suppression reflective - PointSpot Main features Miniature sensor range The visible red PointSpot light makes alignment very easy Range: 250 mm Sensitivity adjustment by potentiometer

More information

TRANSFER RADIATION THERMOMETER

TRANSFER RADIATION THERMOMETER HEITRONICS Infrarot Messtechnik TRANSFER RADIATION THERMOMETER COVERING THE TEMPERATURE RANGE OF -50 C TO 1000 C Contents General Information Temperature Ranges and Spectral Ranges Field of Views and SSE

More information

Temperature Sensor G1/2" Standard

Temperature Sensor G1/2 Standard SENSORS FOR FOOD AND BIOPHARMA. Product Information TFP-40, -50, -60, -160, -180 FOOD Temperature Sensor G1/2" Standard Application / Specified Usage Temperature measurement in vessels and pipes No product

More information

OEM pressure transmitter For mobile working machines Model MH-3

OEM pressure transmitter For mobile working machines Model MH-3 Electronic pressure measurement OEM pressure transmitter For mobile working machines Model MH-3 WIKA Data sheet PE 81.59 Applications Load monitoring Load moment limitation Hydraulic drive control Special

More information

Inductive Standard Displacement Transducers

Inductive Standard Displacement Transducers WA Inductive Standard Displacement Transducers Special features - Available as displacement probe or with detachable plunger - Good thermal stability in the event of temperature gradients - Space saving,

More information

Operator Manual. Transmitters Series KAT-... Analog transmitters in 2-wire technique

Operator Manual. Transmitters Series KAT-... Analog transmitters in 2-wire technique Operator Manual Transmitters Series KAT-... Analog transmitters in 2-wire technique Important Safety Notes! In the case of non-observance of the installation and safety notes or improper use of the transmitter

More information

Distance sensors. System description. Distance measurement using triangulation

Distance sensors. System description. Distance measurement using triangulation Distance sensors System description Distance measurement using triangulation The measurement principle of optical triangulation is suitable for the precise determination of distances at close range. With

More information

PD30ETB20xxIS. Photoelectrics, Background Suppression reflective with IR light. Main features. Description

PD30ETB20xxIS. Photoelectrics, Background Suppression reflective with IR light. Main features. Description Photoelectrics, Background Suppression reflective with IR light Main features Description The PD30ET... stainless steel sensors are built with high-quality materials and designed for harsh environments.

More information

Temperature Sensor G1/2" Standard

Temperature Sensor G1/2 Standard 1 Product Information TfP-40, -50, -60, -160, -180 Temperature Sensor G1/2" Standard f o o d Application / Specified Usage Authorizations Temperature measurement in vessels and pipes No product contacting

More information

Ultra-slim Body Picking Sensor NA1-PK5 SERIES NA1-5 SERIES. Even a slim hand is detectable by the 25 mm in pitch beam area sensor

Ultra-slim Body Picking Sensor NA1-PK5 SERIES NA1-5 SERIES. Even a slim hand is detectable by the 25 mm in pitch beam area sensor 487 Ultra-slim Sensor NA-PK5 SERIES SERIES General terms and conditions... F-7 Sensor selection guide... P.475~ Related Information Glossary of terms... P.359~ General precautions... P.45 PHOTO PHOTO Conforming

More information

maxon document number:

maxon document number: maxon document number: 791272-04 1 Table of contents... 2 2 Table of figures... 3 3 Introduction... 4 4 How to use this guide... 4 5 Safety Instructions... 5 6 Performance Data... 6 6.1 Motor data... 6

More information

Operating Instructions for Microwave Level Switch. Model: LNM

Operating Instructions for Microwave Level Switch. Model: LNM Operating Instructions for Microwave Level Switch Model: LNM 1. Contents 1. Contents...2 2. Note...3 3. Instrument Inspection...3 4. Regulation Use...3 5. Operating Principle...4 6. Mechanical Connection...5

More information

Ultra-slim Body Picking Sensor NA1-PK5 SERIES NA1-5 SERIES. Even a slim hand is detectable by the 25 mm in pitch beam area sensor

Ultra-slim Body Picking Sensor NA1-PK5 SERIES NA1-5 SERIES. Even a slim hand is detectable by the 25 mm in pitch beam area sensor 43 PHOTO PHOTO Ultra-slim Sensor NA-PK5 SERIES SERIES General terms and conditions... F-3 guide... P.49~ Related Information Glossary of terms... P.549~ General precautions... P.55~ Recognition Make sure

More information

Absolute encoders - SSI Solid shaft with clamping or synchro flange Optical multiturn encoders up to 14 bit ST / 16 bit MT

Absolute encoders - SSI Solid shaft with clamping or synchro flange Optical multiturn encoders up to 14 bit ST / 16 bit MT Features Encoder multiturn / SSI Optical sensing method Resolution: max. singleturn 14 bit, multiturn 16 bit Clamping or synchro flange Electronic setting of zero point Counting direction input Available

More information

DATA SHEET. OD Precision. Displacement sensors for precise and comprehensive measurement tasks

DATA SHEET. OD Precision. Displacement sensors for precise and comprehensive measurement tasks DATA SHEET OD Precision Displacement sensors for precise and comprehensive measurement tasks OD Precision Displacement sensors for precise and comprehensive measurement tasks Displacement sensors from

More information

Process transmitter Model UPT-20, with pressure port Model UPT-21, with flush diaphragm

Process transmitter Model UPT-20, with pressure port Model UPT-21, with flush diaphragm Electronic pressure measurement Process transmitter Model UPT-20, with pressure port Model UPT-21, with flush diaphragm WIKA data sheet PE 86.05 Applications Process technology Machine building and plant

More information

RS /E-EV BLU SERIES

RS /E-EV BLU SERIES The RS/E and RS/EV series burners are characterised by a modular monoblock structure that means all necessary components can be combined in a single unit thus making installation easier, faster and, above

More information

Multiturn absolute encoder

Multiturn absolute encoder SYNCHRON SERIELLES INTERFACE Multiturn absolute encoder Release date: 2016-07-12 16:14 Date of issue: 2016-07-12 t4917_eng.xml Model Number Features Industrial standard housing Ø58 mm 0 Bit multiturn Data

More information

Distance sensors. Measurement from 20 mm to 250 m. Optical distance sensors from Page 146. from Page 152. from Page 152.

Distance sensors. Measurement from 20 mm to 250 m. Optical distance sensors from Page 146. from Page 152. from Page 152. Distance sensors Measurement from 0 mm to 50 m Optical distance sensors from Page 4 from Page 5 FT 50 Operating range (scanning distance) from 30 to 300 mm High absolute accuracy High measurement frequency

More information

Wind Transmitter First Class Advanced Classified according to IEC ( )

Wind Transmitter First Class Advanced Classified according to IEC ( ) THE WORLD OF WEATHER DATA - THE WORLD OF WEATHER DATA - THE WORLD OF WEATHER DATA Instruction for Use 021519/01/09 Wind Transmitter First Class Advanced Classified according to IEC 61400-12-1 (2005-12)

More information

EQUIPMENT INFORMATION

EQUIPMENT INFORMATION OPTICAL (NIR) MEASURING SYSTEM UR 5500 LED EQUIPMENT INFORMATION Robert-Bosch-Straße 5 D-56566 Neuwied Tel. +49 (0) 26 31 / 96 40 00 Fax. +49 (0) 26 31 / 96 40 40 Internet: www.sensor-control.de E-Mail:

More information

ph or ORP Transmitter

ph or ORP Transmitter ph or ORP Transmitter Programmable outputs: two transistor and single or dual analog 4-20 ma (Process + Temp) Removable backlighted display Universal process connection Compatible with 120 mm ph/ ORP probes

More information

Absolute encoders - SSI Solid shaft with clamping or synchro flange Optical multiturn encoders 18 bit ST / 12 bit MT

Absolute encoders - SSI Solid shaft with clamping or synchro flange Optical multiturn encoders 18 bit ST / 12 bit MT Features High resolution encoder multiturn / SSI Optical sensing method Resolution: singleturn 18 bit, multiturn 12 bit Electronic setting of zero point Counting direction input Available with additional

More information

Ultra-slim Body Picking Sensor NA1-PK5 SERIES NA1-5 SERIES. Even a slim hand is detectable by the 25 mm in pitch beam area sensor

Ultra-slim Body Picking Sensor NA1-PK5 SERIES NA1-5 SERIES. Even a slim hand is detectable by the 25 mm in pitch beam area sensor 473 PHOTO PHOTO Ultra-slim Sensor NA-PK5 SERIES SERIES General terms and conditions... F-7 Sensor selection guide... P.4~ Related Information Glossary of terms... P.455~ General precautions... P.458~ Conforming

More information

RS /E-EV BLU SERIES

RS /E-EV BLU SERIES The RS/E and RS/EV series burners are characterised by a modular monoblock structure that means all necessary components can be combined in a single unit thus making installation easier, faster and, above

More information

DIGITAL LASER DISTANCE METER

DIGITAL LASER DISTANCE METER DIGITAL LASER DISTANCE METER LD05-A10GF with glass-fiber coupled remote optical head The RIEGL LD05-A10GF is a multi-purpose laser distance meter based on precise timeof-flight laser range measurement

More information

DMP 343. Industrial Pressure Transmitter. Without Media Isolation. accuracy according to IEC 60770: 0.35 % FSO. Nominal pressure

DMP 343. Industrial Pressure Transmitter. Without Media Isolation. accuracy according to IEC 60770: 0.35 % FSO. Nominal pressure DMP 4 Industrial Pressure Transmitter Without Media Isolation accuracy according to IEC 60770: 0.5 % FSO Nominal pressure from 0... 0 mbar up to 0... 000 mbar Product characteristics excellent linearity

More information

Operating Instructions TIC 251/252. Profibus DP-Gateway PG 0009 BE/E (0903)

Operating Instructions TIC 251/252. Profibus DP-Gateway PG 0009 BE/E (0903) Operating Instructions TIC 251/252 Profibus DP-Gateway PG 0009 BE/E (0903) Contents Page 1. Safety Precautions... 3 2. Understanding The TIC 251/252... 4 2.1. For Your Orientation... 4 Delivery Consignment...

More information

Data Sheet. Absolute pressure transducer Nominal (rated) pressure 10 bar to 3000 bar. Special features

Data Sheet. Absolute pressure transducer Nominal (rated) pressure 10 bar to 3000 bar. Special features P3 Top Class P3MB version with fixed cable P3MBP version with plug connection TEDS: Transducer Electronic Data Sheet P3 Absolute pressure transducer Nominal (rated) pressure 10 bar to 3000 bar Special

More information

Product Specification. 6.1 Gb/s Short-Wavelength SFP+ Transceiver FTLF8526P3BNL

Product Specification. 6.1 Gb/s Short-Wavelength SFP+ Transceiver FTLF8526P3BNL Product Specification 6.1 Gb/s Short-Wavelength SFP+ Transceiver FTLF8526P3BNL PRODUCT FEATURES Up to 6.1 Gb/s bi-directional data links Hot-pluggable SFP+ footprint Built-in digital diagnostic functions

More information

Melt pressure Transmitter GREENLINE Type DTAI with Temperature element

Melt pressure Transmitter GREENLINE Type DTAI with Temperature element Melt pressure Transmitter GREENLINE Type DTAI with Temperature element Operation and Maintenance Manual English Mercury free NaK free Seite 1 / 9 Contents: 1. Introduction 2. Operating range and field

More information

Absolute encoders - SSI Solid shaft with clamping or synchro flange Optical multiturn encoders up to 14 bit ST / 16 bit MT

Absolute encoders - SSI Solid shaft with clamping or synchro flange Optical multiturn encoders up to 14 bit ST / 16 bit MT Features Encoder multiturn / SSI Optical sensing method Resolution: max. singleturn 14 bit, multiturn 16 bit Clamping or synchro flange Electronic setting of zero point Counting direction input Available

More information

Optical laser distance sensors. Dimensioned drawing

Optical laser distance sensors. Dimensioned drawing Dimensioned drawing en 02-2012/11 50118552 50 100mm 18-30 V DC We reserve the right to make changes DS_ODSL9L100_en_50118552.fm High resolution 0.01mm Reflection-independent distance information Highly

More information

Digital temperature transmitter Model T15.H, head mounting version Model T15.R, rail mounting version

Digital temperature transmitter Model T15.H, head mounting version Model T15.R, rail mounting version Electrical temperature measurement Digital temperature transmitter Model T15.H, head mounting version Model T15.R, rail mounting version WIKA data sheet TE 15.01 Applications Process industry Machine building

More information

Laser light scanner with background suppression. Dimensioned drawing

Laser light scanner with background suppression. Dimensioned drawing Dimensioned drawing We reserve the right to make changes DS_HT10_2M_en_0130293.fm en 01-201/09 0130293 18-30 V DC 0 2000mm The laser light scanner, based on the principle of light propagation time measurement,

More information

Infrared Temperature Measurement System

Infrared Temperature Measurement System Infrared Temperature Measurement System Modline 3 - the most advanced micro computer based non-contact temperature measurement system, featuring high performance, reliability and versatility. The Modline

More information

VariTrans P Compact automation solutions can be implemented thanks to the 17.5 mm modular housing and operation at temperatures up to 70 C.

VariTrans P Compact automation solutions can be implemented thanks to the 17.5 mm modular housing and operation at temperatures up to 70 C. ProLine Interface Technology Transducers for High Voltage / Shunt Applications Compact high voltage transducers with VariPower broad-range power supply and genuine calibrated range selection. The Task

More information

SCHMIDT Flow Sensor SS SS FB

SCHMIDT Flow Sensor SS SS FB Product description Flow sensor with integrated microcontroller transducer for the measurement of air-flow velocity and for the determination of the flow rate and flow volume. The aerodynamically styled

More information

OEM miniature resistance thermometer Models TR31-3 and TR31-K, thread-mounted

OEM miniature resistance thermometer Models TR31-3 and TR31-K, thread-mounted Electrical temperature measurement OEM miniature resistance thermometer Models TR31-3 and TR31-K, thread-mounted WIKA data sheet TE 60.31 further approvals see page 10 Applications Machine building, plant

More information

Incremental Encoder Model: FOI58

Incremental Encoder Model: FOI58 Incremental Encoder Model: FI58 Document no. FI 188 BE Date: 1.01.201 Robust mechanical construction, high vibration and shock resistance 1 to 16,8 pulses/ revolution perating temperature - 0 C to + 85

More information

Laser Displacement Sensor Z4M. Ordering Information

Laser Displacement Sensor Z4M. Ordering Information Laser Displacement Sensor Compact Displacement Sensor with 1.5-micron Resolution Offers Long-distance Measurement for In-line Production Inspection High resolution with 3 user-selectable response speed

More information

INPROX sensors. displacement MLS compact ccd-laser distance sensor

INPROX sensors. displacement MLS compact ccd-laser distance sensor compact ccd- distance sensor MLS7-250 high vibration and shock resistant -CCD specifications measuring range 250mm resolution >50µm measuring frequency 1000Hz small dimensions (mm) (65x50x20) high shock

More information

Z4D-F. LED-type Micro Displacement Sensor with Many Features in a Compact Body. Micro Displacement Sensor. Ordering Information.

Z4D-F. LED-type Micro Displacement Sensor with Many Features in a Compact Body. Micro Displacement Sensor. Ordering Information. / Length Micro Sensor LED-type Micro Sensor with Many Features in a Compact Body Compact body only /4th the volume of our previous product. % cost reduction over our previous product. High performance

More information

Pressure transmitter For general industrial applications Model A-10

Pressure transmitter For general industrial applications Model A-10 Electronic pressure measurement Pressure transmitter For general industrial applications Model A-10 WIKA data sheet PE 81.60 Applications Machine building Ship building Measurement and control technology

More information

NOVOTURN Multiturn Sensor non-contacting. Series RSM-2800

NOVOTURN Multiturn Sensor non-contacting. Series RSM-2800 NOVOTURN Multiturn Sensor non-contacting Series RSM-2800 Special features Non-contacting, magnetic Long life Electrical range 720 up to 5760 in 360 -steps available (2 to 16 turns) True-Power-On System:

More information

T10FS. Data Sheet. Torque Flange. Special features. Installation example T10FS. B en

T10FS. Data Sheet. Torque Flange. Special features. Installation example T10FS. B en T10FS Torque Flange Data Sheet Special features Nominal (rated) torques: 100 NVm, 200 NVm, 500 NVm, 1 knvm, 2 knvm, 3 knvm, 5 knvm, 10 knvm Nominal speed from 12,000 rpm to 24,000 rpm Low rotor weights

More information

Absolute encoders multiturn

Absolute encoders multiturn The Sendix M6 with Energy Harvesting Technology is an electronic multiturn encoder in miniature format, without gear and without battery. With a size of just 6 x 5 mm it offers a blind hollow shaft of

More information

ENCODER. Analog Multiturn Encoder. Serie 8.M3661/M3681. Key-Features:

ENCODER. Analog Multiturn Encoder. Serie 8.M3661/M3681. Key-Features: ENCODER Multiturn Encoder Serie 8.M366/M368 Key-Features: Content: Order Code...2 Technical Data...3 Outputs & Connections...5 Dimensions...6 - Shaft with max. 0 mm diameter - 0...5/0V, 4...20 ma analog

More information

Rotary Measuring Technology

Rotary Measuring Technology Our ATEX encoders now also carry approval for Dust Flameproof-enclosure type of construction with approval for Zone 1 and 21 ExII2GEExdIICT6 and ExII2DIP6xT85 C Shaft ø 12 mm One type for every situation:

More information

INTRODUCTION 提夫自控技术 ( 上海 ) 有限公司 LAS laser sensors cover measurement ranges from 1 to mm. The integrated micro-controller delivers an accur

INTRODUCTION 提夫自控技术 ( 上海 ) 有限公司 LAS laser sensors cover measurement ranges from 1 to mm. The integrated micro-controller delivers an accur LASER Analog Laser Displacement Transducer LAS Series Key-Features: Content: Measurement Principle & Installation...2 Overview of the Series...3 Technical Data.4 Technical Drawings...9 Diagrams...10 Order

More information

Triple Beam Adjustable Range Reflective Photoelectric Sensor Amplifier Built-in

Triple Beam Adjustable Range Reflective Photoelectric Sensor Amplifier Built-in 371 Triple Beam Adjustable Range Reflective Photoelectric SERIES Related Information General terms and conditions... F-7 Glossary of terms... P.1455~ selection guide... P.271~ General precautions... P.1458~

More information

Pressure transmitter for general industrial applications Model A-10

Pressure transmitter for general industrial applications Model A-10 Electronic pressure measurement Pressure transmitter for general industrial applications Model A-10 WIKA data sheet PE 81.60 Applications Machine building Machine tools Measurement and control technology

More information

Sensing method Appearance Sensing distance Operation mode Model. Emitter E3S-2LE4 Through-beam *1. 2 m. 0.1 to 2 m. 300 mm. 2 m

Sensing method Appearance Sensing distance Operation mode Model. Emitter E3S-2LE4 Through-beam *1. 2 m. 0.1 to 2 m. 300 mm. 2 m Photoelectric Sensor with Built-in Amplifier CSM DS_E 1 General-purpose Photoelectric Sensor for High Quality and Reliable Detection Be sure to read Safety Precautions on page 8. Ordering Information General-purpose

More information

Encoder WDGI 58E.

Encoder WDGI 58E. Encoder WDGI E www.wachendorffautomation.com/wdgie Wachendorff utomation... systems and encoders Complete systems Industrial rugged encoders to suit your application Standard range and customer versions

More information