Sensitivity calibration of digital colour cameras for auroral imaging

Size: px
Start display at page:

Download "Sensitivity calibration of digital colour cameras for auroral imaging"

Transcription

1 Sensitivity calibration of digital colour cameras for auroral imaging Fred Sigernes 1,*, Jeffrey Morgan Holmes 1, Margit Dyrland 1, Dag Arne Lorentzen 1, Trond Svenøe 2, Karsten Heia 3, Takehiko Aso 4, Sergey Chernouss 5 and Charles Sterling Deehr 6 1 The University Centre in Svalbard (UNIS), N-9171 Longyearbyen, Norway 2 The Norwegian Polar Institute, Ny-Ålesund, Norway 3 Norwegian Institute of Fisheries and Aquaculture, Tromsø, Norway 4 National Institute of Polar Research, Tokyo, Japan 5 Polar Geophysical Institute, Murmansk Region, Apatity, Russia 6 Geophysical Institute, University of Alaska, Fairbanks, USA * Corresponding author: fred@unis.no Abstract: A method to sensitivity calibrate Digital Single Lens Reflective (DSLR) cameras is outlined. A low intensity calibrated light source tunable in wavelength is described. 31 monochromatic lines from 4000 to 7000 Å with a bandpass of ~12 Å were used to find the spectral responses for the D70 and the D200 cameras manufactured by Nikon. The source radiance ranged from about 300 to 1.6k R/Å. The cameras were operated in manual mode with 4 seconds exposure time at ISO 1600, which are typical settings required for night time photography of the aurora. For the Nikon D200 camera, the blue, green and red spectral responsivities peak at 4600, 5300 and 5900 Å, respectively. The response was high for the blue colour channel with a clear cut-off at 4100 Å for the UV part of the spectrum. The red channel response indicates low sensitivity above 6600 Å. The D70 shows similar spectral responsivity, except that it peaks in the green colour channel and it is more sensitive to both UV and NIR radiation. Both cameras are capable of detecting night - and dayside auroral at 4 second exposure time. For optimal auroral imaging capability, the green and red spectral responsivities need to be shifted up by about 300 and 400 Å in wavelength Optical Society of America OCIS codes: ( ) Optical Standards and testing; ( ) Spectroscopy, visible; ( ) Calibration; ( ); Astronomical optics; ( ) Space optics. References and links 1. N. Partamies, M. Syrjäsuo, and E. Donovan, Using colour in auroral imaging, Can. J. Phys. 85, (2007). 2. J.W. Chamberlain, Physics of the Aurora and Airglow, (Academic Press, New York and London, 1961). 3. D. Baker and G. Romick, The Rayleigh: interpretation of the unit in terms of column emission rate or apparent radiance expressed in SI units, Appl. Opt. 15, (1976). 4. F. Sigernes, J.M. Holmes, M. Dyrland, D.A. Lorentzen, S. Chernous, T. Svenøe, J. Moen, and C.S. Deehr, Absolute calibration of optical devices with a small field of view, J. Opt. Technol. 74, (2007). 5. G.W. Chang, and Y.C. Chen, Spectral Responsivity Estimator for Color Vision Systems: Filter Selection and Noise Effect, Proceedings of National Sci. Council, Part A 25, (2001). 6. C.S. Deehr, G.G. Sivjee, A. Egeland, K. Henriksen, P.E. Sandholt, R. Smith, P. Sweeney, C. Duncan, and J. Gilmer, Ground-Based Observations of F Region Aurora Associated With the Magnetospheric Cusp, J. Geophys. Res. 85, (1980). 7. M.H. Rees, Physics and chemistry of the upper atmosphere, (Cambridge University Press, 1989). (C) 2008 OSA 29 September 2008 / Vol. 16, No. 20 / OPTICS EXPRESS 15623

2 1. Introduction The arrival of the Digital Single Lens Reflex (DSLR) camera opens up new possibilities in the field of night sky colour photography of aurora [1]. Its ability to make long exposures with high signal-to-noise levels and the variety of objectives to choose from makes the DSLR camera a cost-effective choice compared to many of the scientific grade camera systems used today. In auroral imaging, monochromatic sensors are used in combination with optical interference filters to obtain the different wavelength emission intensities. The design requires first that light from the front lens needs to be collimated onto the filters. The filters are usually mounted on a rotating wheel called a filter wheel. The wheel needs to position each filter in alignment to the optical axis of the system before any exposure can take place. Secondly, the narrowband light from the filters needs to be focused onto the detector plane. Furthermore, if a light intensifier is used, then an additional relay lens must be included to transfer the intensified image onto the detector plane. The whole system becomes complex with several optical elements that need to be aligned for maximum throughput. In addition, three successive exposures with different centered wavelength filters rotated into position are necessary to create a color composite image. The operation consumes time which will diminish the spatial resolution since auroras tend to move quickly over the sky. The DSLR camera on the other hand uses only one lens element or objective together with a Colour Filter Array (CFA), or Colour Filter Mosaic (CFM), in front of the image sensor. A common CFM is the Bayer filter, which contains a sub mosaic of two green, one blue and one red transparent filter on top of each pixel of the sensor, respectively. The colour value of an image pixel is a combination of how much light that passes through each filter in the sub mosaic. But the details on how each camera manufacturer carries out the color reconstruction or demosaicing are proprietary information. The processed colour- or even the raw format images of the cameras do not give accurate information about the intensity in absolute physical units. It is therefore necessary to develop a procedure on how to conduct an intensity calibration. This study describes a method to obtain the spectral responsivity of a DSLR camera. Results from two Nikon cameras (D70 and D200) are presented. Finally, an evaluation is given on their ability to image aurora. Fig. 1. Experimental setup in the optical laboratory at UNIS. (1) fiber bundle from Leica 150 W Tungsten Halogen lamp, (2) camera mount, (3) order sorting filter wheel, (4) HR320 monochromator, (5) exit slit holder, (6) baffle, (7) fiber bundle to FICS spectrograph, (8) Lambertian screen, (9) FICS spectrograph, (10) steel mount rod. (C) 2008 OSA 29 September 2008 / Vol. 16, No. 20 / OPTICS EXPRESS 15624

3 2. Experimental setup A fiber optic illuminator connected to a Czerney Turner monochromator is used as a light source in front of a white diffuse screen (Lambertian surface). Figure 1 shows the experimental setup. The fiber optical illuminator (Leica CLS 150) uses a 150 W Tungsten lamp. The fiber bundle from the lamp house is connected by a lens to the monochromator. The lens is used to match the field of view of the fiber to the monochromator (f-matching). The lamp house is located in a separate room with a hole in the wall for the fiber bundle. Fig. 2. Optical diagram. (1) lamp input fiber bundle, (2) f-matching lens, (3) order sorting filter wheel, (4) entrance slit, (5) flat mirror, (6) concave mirror (collector), (7) reflective grating, (8) focusing mirror, (9) flat mirror, (10) exit slit, (11) Lambertian diffuse re-emitting surface, (12) DSLR camera, (13) input fiber bundle to FICS spectrograph. The monochromator or the HR320 Czerney Turner spectrometer is made by Jobin Yvon. It uses two concave mirrors of focal length 320 mm. The grating has 1200 grooves / mm with an effective area of 68 x 68 mm 2. The entrance and exit slit widths are fixed at 0.5 mm, which results in a spectral band pass of ~ 12 Å. The slits are easily changed if needed. The Lambertian surface (SRT ) is located 40 cm from the exit slit. It is made of Spectralon and is produced by Labsphere, Inc. The diffuse reflectance factors of the screen are nearly constant (ρ λ = 0.98) throughout the visible and near infrared regions of the spectrum. The intensity of the screen in absolute units is monitored be a Fixed Imaging Compact Spectrograph (FICS SN 7743) made by ORIEL. It uses a concave holographic grating (230 grooves / mm) with a nominal spectral range of Å. The detector is a 16-bit dynamic range thermoelectric cooled CCD from Hamamatsu (model INSTASPEC IV). The field of view of FICS is 22 o, which f-matches the fused silica fiber bundle used as entrance optics. The bandpass is approximately 96 Å. The spectral resolving power of the instrument is moderate, but for our purposes it is enough. The instrument s main task is to locate the centre or peak intensity wavelength of the output spectral line and to measure its intensity in units of R/Å. The unit Rayleigh, R, is used by the auroral and astronomical community. The generalized definition is R 1/ 4π 10 # photons cm sec sr, (1) ( ) 1 and it applies directly to radiance [2, 3]. The spectrograph is calibrated by the method outlined by [4]. The optical diagram of the system is shown in Fig. 2. The angular position of the grating determines the central wavelength illuminating the screen. The procedure is to first position the grating at the desired start central wavelength at 4000 Å, and measure its intensity by the FICS spectrograph. Secondly, the DSLR takes an image of the screen in raw format. (C) 2008 OSA 29 September 2008 / Vol. 16, No. 20 / OPTICS EXPRESS 15625

4 The grating is then rotated to the next position at 4100 Å, and the procedure is repeated up to an end wavelength of 7000 Å. The screen is then illuminated by 31 spectral lines throughout the visible spectrum. 3. Spectral responsivity In order to measure intensity with a DSLR camera it is necessary to obtain the total raw counts of the image from each color channel, k, separately. The equation is given as [ CTS] ( k ) ( k ) ui = Ci ( λ) S ( λ) dλ, (2) where λ is the wavelength, C i the intensity of the source in absolute units, and S (k) is the spectral responsivity [cf. 5]. The subscript i tags the specific observation of the source. In our case i refers to the 31 different spectra from the monochromator. Eq. (2) can be written in vector form as ( k ) ˆ T ˆ ( k ) u = C S. (3) The spectral responsivity S ˆ ( k ) where the matrix C is given as i i can now be solved if we form a set of observations ( k ) ˆ ˆ ( k ) u = C S, (4) [ C C C C ] T C =. (5) ( ) Thus matrix C forms a library of 31 spectral lines. The vector u contains 31 numbers of ( ) total counts and Sˆ k has 31 unknown response values for each colour channel, k, respectively. Originally, each vector C i contains 1024 elements covering the spectral region from 4000 to Å. The dimension is reduced to 31 by sampling the vector at each center wavelength setting of the monochromator. Singular Value Decomposition (SVD) may now be applied to solve Eq. (4). ˆ k Fig. 3. Source functions. C i (λ) is the set of observations that consists of 31 spectra from the monochromator (HR320) illuminating the diffuse screen (Lambertian surface SRT ). (C) 2008 OSA 29 September 2008 / Vol. 16, No. 20 / OPTICS EXPRESS 15626

5 4. Calibration Results Our library of spectra from the monochromator or set of observations is shown in Fig. 3. The exposure time of the FICS spectrograph is 4 seconds. The Nikon D70 and D200 cameras use the same exposure time at high sensitivity settings (ISO 1600) without being over exposed by the monochromator lamp. The lens aperture is set to f/1.4. Note that we used the same lens (AF Nikkor 50 mm f/1.4d) on both cameras houses. The intensity rises from ~300 R / Å at 4000 Å to ~1600 R /Å at 5900 Å, and ends at ~900 R / Å at 7000 Å. This is expected since the result is a convolution between the transmission / reflection profiles of each active component, the grating efficiency and the lamp spectrum. The spectrum of the tungsten lamp should slowly increase over the wavelength region while the monochromator grating efficiency starts to decrease beyond the red part of the spectrum. The drops in intensities at 4600 Å and 6700 Å are due to dips in the grating efficiency. Fig. 4. Block diagram of data handling. (k ) α is the total raw counts of the decomposed k i colour plane from observation i. (k ) β is the corresponding total dark counts. The Nikon raw images of format.nef are read by a program called MaxDSLR by Diffraction Limited, and converted to 16 bit uncompressed FITS images. FITS stands for Flexible Image Transport System and is a standard astronomical data format. The images are decomposed into its 3 primary colour planes by the Interactive Data Language (IDL) from the company ITT Visual Information Solutions. Figure 4 shows a block diagram of the data handling process. The raw data from both cameras are shown in Fig. 5. The counts follow a general behavior with overlap between the filters and higher count rates for the green filter. The filters are designed to resemble the human eye colour response. The D200 has twice as many pixels as the D70. This is the reason for the D200 s overall higher total count rates. In addition, D200 has a sharper blue and red cut off than the D70. The palettes in Fig. 5 are the automatically processed JPEG images of the cameras for each observation. It is hard to visually see any difference between in the RGB colour values obtained by the two cameras. Note that we use total counts and not counts per pixel for the raw data images. The reason is that the screen is not completely uniformly illuminated by the monochromator. We need to apply an integrating sphere instead of a flat screen to achieve a uniform source. These spheres are expensive, but we aim to get one for future studies. Therefore as a start, the raw data are (C) 2008 OSA 29 September 2008 / Vol. 16, No. 20 / OPTICS EXPRESS 15627

6 treated as the total summed counts per sensor with units of 10 6 counts [MCTS]. The assumption is that both the cameras and the FICS spectrograph detect the same illuminated area of the screen. The source area is well within the field of views of all instruments. (k ) The total dark count β is subtracted from each observation. This is done by one 4 second exposure in total darkness. Correspondingly, the bias of the detector is found by setting the exposure time to shortest possible (1/8000 seconds). The bias is found to have no effect on our measurements due to the long exposure time of 4 seconds. Fig. 5. Raw data. Panel (A) shows raw counts from the Nikon D70 camera for each color channel. The palette in the upper left corner is the processed JPEG images from the camera for each observation. Panel (B) shows the corresponding results from the Nikon D200 camera. Fig. 6. The spectral responsivity. Panel (A) shows the spectral responsivity of the Nikon D70 camera for each color channel. Panel (B) shows the corresponding results from the Nikon D200 camera. The spectral responsivity may now be solved according to Eq. (4). The result is shown in Fig. 6 for both cameras. The responsivities for the cameras look quite similar in shape with peaks for the blue, green and red channels at 4600, 5300 and 5900 Å, respectively. The main difference is that the D200 sensitivity peaks in the blue instead of the green part of the spectrum. It is also clear that the D70 has higher sensitivity in the far red and violet regions of the spectrum. These extended wings favor the D70 when it comes to taking near infra-red or ultraviolet photos. On the other hand, the D200 will in general produce sharper images due to less detected chromatic aberration from the wings. The D200 spectral responsivity is squeezed in range ( Å) compared to the D70, and any out of wavelength range lens aberrations will not be detected. (C) 2008 OSA 29 September 2008 / Vol. 16, No. 20 / OPTICS EXPRESS 15628

7 5. Auroral imaging In order to relate the spectral responsivity of the cameras to auroral imaging it is necessary to obtain information about the spectrum C i (λ) of the aurora in absolute units (see Eq. (2)). Figure 7 shows spectral measurements from the Kjell Henriksen Observatory (KHO) on Svalbard (78 o N, 16 o E) together with the obtained spectral responsivity of the cameras. The two spectra represent night- and dayside aurora, and are displayed to visualize the main (k) emissions that are contributing to the total counts u i in each channel k detected by the cameras. Note that the subscript i now refer to night- or dayside aurora. In general, the overall strongest emissions lines of the aurora are from the green 5577 Å and the red 6300 / 6364 Å of atomic oxygen [OI]. On the night side the green emission is stronger than the red, while on the dayside the red dominates over the green. This is due to the Fig. 7. Samples of night- and dayside spectra of the aurora from the Kjell Henriksen Observatory (KHO) on Svalbard, Norway. Both spectra are obtained by the Auroral Spectrograph (ASG) owned by the National Institute of Polar Research in Japan (NIPR). The spectral range is 4172 to 7334 Å with a medium spectral resolution of ~25 Å. The exposure time is 1 second for both samples. The dayside spectrum (blue curve) is taken 18th of January 2008 at 07:29:41 UT. The night side spectrum (red curve) is from 3rd of February 2008 at 21:46:47 UT. All of the known atomic- and molecular bands of the aurora are marked as a function of wavelength according to [2]. The right y-axis shows the intensity in R/Å. The colored dash - dotted and dotted lines show the spectral responsitivities of the Nikon D200 and the D70, respectively. difference in high versus low energy of the incoming electrons that produce the emissions on the night- versus the dayside, respectively [6]. The transition or border between low and high energy of the electrons is at ~500 ev. High energy electrons produce the green colour of an auroral arc. The upper part of the arc is red, produced by low energetic electrons. In addition, the green emission appears more structured than the red, which is more diffuse in appearance. The explanation is that after the electron has collided with the oxygen atom, the lifetime of the excited state is ~1 second for the green and ~2 minutes for the red emission. The rapid movement of an auroral arc over the sky is therefore seen in the green. The red is more washed out. A typical auroral [OI] arc is located in the altitude range from 100 up to 600 km. Note the absence of the molecular band emissions in the dayside spectrum. The bulk flux of the electrons does not have enough energy to reach the molecular part of the upper (C) 2008 OSA 29 September 2008 / Vol. 16, No. 20 / OPTICS EXPRESS 15629

8 atmosphere (altitudes < 100 km). However on the night side, the high energy tail of the electron flux may easily reach up to several kev and thus penetrate below ~100 km [7]. The result is seen in Fig. 7. N 2 + and N 2 band emissions appear in the blue and red wings of the spectrum, respectively. These emissions are seen as fast moving regions along the lower border of the auroral arcs. Figure 8 shows a typical auroral arc with its main auroral emissions as a function of altitude. The intensities of the auroral emissions in Fig. 7 are in the same order of magnitude as the source functions from the monochromator in Fig. 3. As a consequence, both the cameras are sensitive enough to detect both night- and dayside aurora with only 4 second exposure time. Also note that the two brightest emission lines in aurora, the green [OI] 5577 Å and the red [OI] 6300 / 6364 Å, are each located in wavelength above the peak spectral responsivity of the green and red colour channels of the cameras, respectively. This is off course not an ideal situation. A shift of the spectral responsivity of about 277 Å for the green and 400 Å for the red channel up in wavelength would improve the auroral sensitivity of both cameras substantially. Fig. 8. The main colours of the aurora. The illustration is based on a photograph taken by a Fujifilm S2 Pro camera from the SVALSAT satellite station close to Longyearbyen, Norway, at 16:04:48 UT 8 th of November, The exposure time was 30 seconds and the sensitivity was set to ISO On the other hand, there are little overlaps between the respective colour channels and the auroral emissions. Or in other words, the blue and green channels of the cameras are more or less blind to the auroral red emission lines. The same applies to the auroral green emission line. Both cameras are therefore able to detect and separate the red and green emissions of the aurora. Table 1 shows the total counts in each channel of the cameras calculated according to Eq. (2), using to the measured spectra in Fig. 7 as input. The numbers in Table 1 is the percentage of the total counts of the cameras. As seen in Fig. 7, the response is low for the blue channel (<10%). For nightside aurora the green and red channels detect close to 60 and 30 % of the total counts, respectively. The D200 seems to be more suitable to image red aurora (dayside) than the D70. The lack of red and blue signal could be compensated with taking a sequence of images with increasing exposure time to obtain higher count rates. The green channel will then most likely be over exposed in this process. The composite colour image may then be (C) 2008 OSA 29 September 2008 / Vol. 16, No. 20 / OPTICS EXPRESS 15630

9 Table 1. The Nikon D70 and D200 total counts for each colour channel in %. The counts are calculated for both night- and dayside aurora. Dayside aurora Nightside aurora D70 D200 D70 D200 Red u (1) Green u (2) Blue u (3) composed by choosing the optimum colour channels in the sequence. The main disadvantage would be loss of temporal resolution and spatial resolution if the aurora moves quickly. One other solution to increase the red and blue sensitivity would be to remove the internal UV-IR blocker filter in front of the CCDs. The N + 2 and the N 2 emissions will then be easier to detect together with other auroral emissions such at the strong near infrared OI emission line at 8446 Å. The above outlined calibration routine must then be repeated before any conclusion can be made. 6. Concluding remarks The principal results obtained by this study can be summarized as follows. (1) A fiber optical lamp is connected to the input of a monochromator which is tunable in wavelength. The output of the monochromator illuminates a Lambertian screen with a bandpass of ~12Å. (2) The brightness of the screen is monitored by an intensity calibrated spectrograph. A library of source functions consisting of 31 monochromatic lines is obtained in the visible part of the electromagnetic spectrum. The intensities range from 300 to 1.6 R/Å. (3) A method that retrieves the spectral responsivity of DSLR cameras has been outlined based on the above measured spectral library. (4) Two DSLR cameras, the Nikon D70 and D200, have been calibrated. Both cameras have peak spectral responsivities in the blue, green and red channels at 4600, 5300 and 5900 Å, respectively. The D200 have higher blue response compared to the D70. The D200 also cuts its response to ultraviolet and near infrared wavelengths more sharply compared to the D70. (5) As candidates for auroral imaging the cameras are able to detect both night- and dayside aurora with exposure times in the order of a few seconds for both the auroral green [OI] 5577 Å and the red [OI] 6300 / 6364 Å emissions. The blue sensitivity is low due to lack of strong emission lines in the aurora from this wavelength region. (6) The D200 is found to be slightly more sensitive than the D70 to the red auroral [OI] emission at 6300 / 6364 Å. (7) A shift in the spectral responsivity to higher wavelengths would result in optimal auroral performance for both cameras. The shift is 277 Å for the green and 400 Å for the red channel. (8) For future studies we aim to evaluate other cameras and replace the Lambertian screen with an integrating sphere. The spectral responsivity will then be deduced for (C) 2008 OSA 29 September 2008 / Vol. 16, No. 20 / OPTICS EXPRESS 15631

10 each pixel of the image sensors. Consequently, more detailed spatial information will be achieved of the aurora and its colours. Acknowledgement This work was financially supported by The Research Council of Norway through the project named: Norwegian and Russian Upper Atmosphere Co-operation On Svalbard #178911/S30 (NORUSCA). (C) 2008 OSA 29 September 2008 / Vol. 16, No. 20 / OPTICS EXPRESS 15632

How to take color pictures of aurora

How to take color pictures of aurora How to take color pictures of aurora Fred Sigernes 1,2,3 1 The University Centre in Svalbard (UNIS), N-9171 Longyearbyen, Norway 2 The Kjell Henriksen Observatory (KHO), Breinosa, Norway 3 Birkeland Centre

More information

Observational Astronomy

Observational Astronomy Observational Astronomy Instruments The telescope- instruments combination forms a tightly coupled system: Telescope = collecting photons and forming an image Instruments = registering and analyzing the

More information

Improving the Collection Efficiency of Raman Scattering

Improving the Collection Efficiency of Raman Scattering PERFORMANCE Unparalleled signal-to-noise ratio with diffraction-limited spectral and imaging resolution Deep-cooled CCD with excelon sensor technology Aberration-free optical design for uniform high resolution

More information

Applications of Steady-state Multichannel Spectroscopy in the Visible and NIR Spectral Region

Applications of Steady-state Multichannel Spectroscopy in the Visible and NIR Spectral Region Feature Article JY Division I nformation Optical Spectroscopy Applications of Steady-state Multichannel Spectroscopy in the Visible and NIR Spectral Region Raymond Pini, Salvatore Atzeni Abstract Multichannel

More information

Chemistry 524--"Hour Exam"--Keiderling Mar. 19, pm SES

Chemistry 524--Hour Exam--Keiderling Mar. 19, pm SES Chemistry 524--"Hour Exam"--Keiderling Mar. 19, 2013 -- 2-4 pm -- 170 SES Please answer all questions in the answer book provided. Calculators, rulers, pens and pencils permitted. No open books allowed.

More information

MS260i 1/4 M IMAGING SPECTROGRAPHS

MS260i 1/4 M IMAGING SPECTROGRAPHS MS260i 1/4 M IMAGING SPECTROGRAPHS ENTRANCE EXIT MS260i Spectrograph with 3 Track Fiber on input and InstaSpec IV CCD on output. Fig. 1 OPTICAL CONFIGURATION High resolution Up to three gratings, with

More information

POTENTIAL OF MULTISPECTRAL TECHNIQUES FOR MEASURING COLOR IN THE AUTOMOTIVE SECTOR

POTENTIAL OF MULTISPECTRAL TECHNIQUES FOR MEASURING COLOR IN THE AUTOMOTIVE SECTOR POTENTIAL OF MULTISPECTRAL TECHNIQUES FOR MEASURING COLOR IN THE AUTOMOTIVE SECTOR Meritxell Vilaseca, Francisco J. Burgos, Jaume Pujol 1 Technological innovation center established in 1997 with the aim

More information

ECEN. Spectroscopy. Lab 8. copy. constituents HOMEWORK PR. Figure. 1. Layout of. of the

ECEN. Spectroscopy. Lab 8. copy. constituents HOMEWORK PR. Figure. 1. Layout of. of the ECEN 4606 Lab 8 Spectroscopy SUMMARY: ROBLEM 1: Pedrotti 3 12-10. In this lab, you will design, build and test an optical spectrum analyzer and use it for both absorption and emission spectroscopy. The

More information

Spectroscopy in the UV and Visible: Instrumentation. Spectroscopy in the UV and Visible: Instrumentation

Spectroscopy in the UV and Visible: Instrumentation. Spectroscopy in the UV and Visible: Instrumentation Spectroscopy in the UV and Visible: Instrumentation Typical UV-VIS instrument 1 Source - Disperser Sample (Blank) Detector Readout Monitor the relative response of the sample signal to the blank Transmittance

More information

Presented by Jerry Hubbell Lake of the Woods Observatory (MPC I24) President, Rappahannock Astronomy Club

Presented by Jerry Hubbell Lake of the Woods Observatory (MPC I24) President, Rappahannock Astronomy Club Presented by Jerry Hubbell Lake of the Woods Observatory (MPC I24) President, Rappahannock Astronomy Club ENGINEERING A FIBER-FED FED SPECTROMETER FOR ASTRONOMICAL USE Objectives Discuss the engineering

More information

OPAL Optical Profiling of the Atmospheric Limb

OPAL Optical Profiling of the Atmospheric Limb OPAL Optical Profiling of the Atmospheric Limb Alan Marchant Chad Fish Erik Stromberg Charles Swenson Jim Peterson OPAL STEADE Mission Storm Time Energy & Dynamics Explorers NASA Mission of Opportunity

More information

DESIGN NOTE: DIFFRACTION EFFECTS

DESIGN NOTE: DIFFRACTION EFFECTS NASA IRTF / UNIVERSITY OF HAWAII Document #: TMP-1.3.4.2-00-X.doc Template created on: 15 March 2009 Last Modified on: 5 April 2010 DESIGN NOTE: DIFFRACTION EFFECTS Original Author: John Rayner NASA Infrared

More information

Spectral and Polarization Configuration Guide for MS Series 3-CCD Cameras

Spectral and Polarization Configuration Guide for MS Series 3-CCD Cameras Spectral and Polarization Configuration Guide for MS Series 3-CCD Cameras Geospatial Systems, Inc (GSI) MS 3100/4100 Series 3-CCD cameras utilize a color-separating prism to split broadband light entering

More information

Performance Comparison of Spectrometers Featuring On-Axis and Off-Axis Grating Rotation

Performance Comparison of Spectrometers Featuring On-Axis and Off-Axis Grating Rotation Performance Comparison of Spectrometers Featuring On-Axis and Off-Axis Rotation By: Michael Case and Roy Grayzel, Acton Research Corporation Introduction The majority of modern spectrographs and scanning

More information

NIRCam optical calibration sources

NIRCam optical calibration sources NIRCam optical calibration sources Stephen F. Somerstein, Glen D. Truong Lockheed Martin Advanced Technology Center, D/ABDS, B/201 3251 Hanover St., Palo Alto, CA 94304-1187 ABSTRACT The Near Infrared

More information

Chapter Ray and Wave Optics

Chapter Ray and Wave Optics 109 Chapter Ray and Wave Optics 1. An astronomical telescope has a large aperture to [2002] reduce spherical aberration have high resolution increase span of observation have low dispersion. 2. If two

More information

Chemistry Instrumental Analysis Lecture 7. Chem 4631

Chemistry Instrumental Analysis Lecture 7. Chem 4631 Chemistry 4631 Instrumental Analysis Lecture 7 UV to IR Components of Optical Basic components of spectroscopic instruments: stable source of radiant energy transparent container to hold sample device

More information

Physics 308 Laboratory Experiment F: Grating Spectrometer

Physics 308 Laboratory Experiment F: Grating Spectrometer 3/7/09 Physics 308 Laboratory Experiment F: Grating Spectrometer Motivation: Diffraction grating spectrometers are the single most widely used spectroscopic instrument. They are incorporated into many

More information

Application Note (A11)

Application Note (A11) Application Note (A11) Slit and Aperture Selection in Spectroradiometry REVISION: C August 2013 Gooch & Housego 4632 36 th Street, Orlando, FL 32811 Tel: 1 407 422 3171 Fax: 1 407 648 5412 Email: sales@goochandhousego.com

More information

Maya2000 Pro Spectrometer

Maya2000 Pro Spectrometer now with triggering! Maya2000 Pro Our Maya2000 Pro Spectrometer offers you the perfect solution for applications that demand low light-level, UV-sensitive operation. This back-thinned, 2D FFT-CCD, uncooled

More information

Section 1: SPECTRAL PRODUCTS

Section 1: SPECTRAL PRODUCTS Section 1: Optical Non-dispersive Wavelength Selection Filter Based Filter Filter Fundamentals Filter at an Incidence Angle Filters and Environmental Conditions Dispersive Instruments Grating and Polychromators

More information

Integrating Spheres. Why an Integrating Sphere? High Reflectance. How Do Integrating Spheres Work? High Damage Threshold

Integrating Spheres. Why an Integrating Sphere? High Reflectance. How Do Integrating Spheres Work? High Damage Threshold 1354 MINIS Oriel Integrating Spheres Integrating spheres are ideal optical diffusers; they are used for radiometric measurements where uniform illumination or angular collection is essential, for reflectance

More information

QE65000 Spectrometer. Scientific-Grade Spectroscopy in a Small Footprint. now with. Spectrometers

QE65000 Spectrometer. Scientific-Grade Spectroscopy in a Small Footprint. now with. Spectrometers QE65000 Spectrometer Scientific-Grade Spectroscopy in a Small Footprint QE65000 The QE65000 Spectrometer is the most sensitive spectrometer we ve developed. Its Hamamatsu FFT-CCD detector provides 90%

More information

NIR SPECTROSCOPY Instruments

NIR SPECTROSCOPY Instruments What is needed to construct a NIR instrument? NIR SPECTROSCOPY Instruments Umeå 2006-04-10 Bo Karlberg light source dispersive unit (monochromator) detector (Fibres) (bsorbance/reflectance-standard) The

More information

An integral eld spectrograph for the 4-m European Solar Telescope

An integral eld spectrograph for the 4-m European Solar Telescope Mem. S.A.It. Vol. 84, 416 c SAIt 2013 Memorie della An integral eld spectrograph for the 4-m European Solar Telescope A. Calcines 1,2, M. Collados 1,2, and R. L. López 1 1 Instituto de Astrofísica de Canarias

More information

Optical Coherence: Recreation of the Experiment of Thompson and Wolf

Optical Coherence: Recreation of the Experiment of Thompson and Wolf Optical Coherence: Recreation of the Experiment of Thompson and Wolf David Collins Senior project Department of Physics, California Polytechnic State University San Luis Obispo June 2010 Abstract The purpose

More information

Supplementary Figure 1. Effect of the spacer thickness on the resonance properties of the gold and silver metasurface layers.

Supplementary Figure 1. Effect of the spacer thickness on the resonance properties of the gold and silver metasurface layers. Supplementary Figure 1. Effect of the spacer thickness on the resonance properties of the gold and silver metasurface layers. Finite-difference time-domain calculations of the optical transmittance through

More information

CHAPTER 7. Components of Optical Instruments

CHAPTER 7. Components of Optical Instruments CHAPTER 7 Components of Optical Instruments From: Principles of Instrumental Analysis, 6 th Edition, Holler, Skoog and Crouch. CMY 383 Dr Tim Laurens NB Optical in this case refers not only to the visible

More information

Better Imaging with a Schmidt-Czerny-Turner Spectrograph

Better Imaging with a Schmidt-Czerny-Turner Spectrograph Better Imaging with a Schmidt-Czerny-Turner Spectrograph Abstract For years, images have been measured using Czerny-Turner (CT) design dispersive spectrographs. Optical aberrations inherent in the CT design

More information

SPECTRAL SCANNER. Recycling

SPECTRAL SCANNER. Recycling SPECTRAL SCANNER The Spectral Scanner, produced on an original project of DV s.r.l., is an instrument to acquire with extreme simplicity the spectral distribution of the different wavelengths (spectral

More information

Spectroscopic Instrumentation

Spectroscopic Instrumentation Spectroscopic Instrumentation Theodor Pribulla Astronomical Institute of the Slovak Academy of Sciences, Tatranská Lomnica, Slovakia Spectroscopic workshop, February 6-10, 2017, PřF MU, Brno Principal

More information

Guide to SPEX Optical Spectrometer

Guide to SPEX Optical Spectrometer Guide to SPEX Optical Spectrometer GENERAL DESCRIPTION A spectrometer is a device for analyzing an input light beam into its constituent wavelengths. The SPEX model 1704 spectrometer covers a range from

More information

Oriel MS260i TM 1/4 m Imaging Spectrograph

Oriel MS260i TM 1/4 m Imaging Spectrograph Oriel MS260i TM 1/4 m Imaging Spectrograph MS260i Spectrograph with 3 Track Fiber on input and InstaSpec CCD on output. The MS260i 1 4 m Imaging Spectrographs are economical, fully automated, multi-grating

More information

SCCH 4: 211: 2015 SCCH

SCCH 4: 211: 2015 SCCH SCCH 211: Analytical Chemistry I Analytical Techniques Based on Optical Spectroscopy Atitaya Siripinyanond Office Room: C218B Email: atitaya.sir@mahidol.ac.th Course Details October 19 November 30 Topic

More information

MULLARD SPACE SCIENCE LABORATORY UNIVERSITY COLLEGE LONDON Authors: H. Kawakami, Alice Breeveld and John Fordham*

MULLARD SPACE SCIENCE LABORATORY UNIVERSITY COLLEGE LONDON Authors: H. Kawakami, Alice Breeveld and John Fordham* XMM Optical Monitor MULLARD SPACE SCIENCE LABORATORY UNIVERSITY COLLEGE LONDON Authors: H. Kawakami, Alice Breeveld and John Fordham* * Dept. Physics and Astronomy, UCL Characteristics of the FM intensifiers

More information

ECEN 4606, UNDERGRADUATE OPTICS LAB

ECEN 4606, UNDERGRADUATE OPTICS LAB ECEN 4606, UNDERGRADUATE OPTICS LAB Lab 2: Imaging 1 the Telescope Original Version: Prof. McLeod SUMMARY: In this lab you will become familiar with the use of one or more lenses to create images of distant

More information

TriVista. Universal Raman Solution

TriVista. Universal Raman Solution TriVista Universal Raman Solution Why choose the Princeton Instruments/Acton TriVista? Overview Raman Spectroscopy systems can be derived from several dispersive components depending on the level of performance

More information

Multispectral Image Capturing System Based on a Micro Mirror Device with a Diffraction Grating

Multispectral Image Capturing System Based on a Micro Mirror Device with a Diffraction Grating Multispectral Image Capturing System Based on a Micro Mirror Device with a Diffraction Grating M. Flaspöhler, S. Buschnakowski, M. Kuhn, C. Kaufmann, J. Frühauf, T. Gessner, G. Ebest, and A. Hübler Chemnitz

More information

TIME-PRESERVING MONOCHROMATORS FOR ULTRASHORT EXTREME-ULTRAVIOLET PULSES

TIME-PRESERVING MONOCHROMATORS FOR ULTRASHORT EXTREME-ULTRAVIOLET PULSES TIME-PRESERVING MONOCHROMATORS FOR ULTRASHORT EXTREME-ULTRAVIOLET PULSES Luca Poletto CNR - Institute of Photonics and Nanotechnologies Laboratory for UV and X-Ray Optical Research Padova, Italy e-mail:

More information

SOAR Integral Field Spectrograph (SIFS): Call for Science Verification Proposals

SOAR Integral Field Spectrograph (SIFS): Call for Science Verification Proposals Published on SOAR (http://www.ctio.noao.edu/soar) Home > SOAR Integral Field Spectrograph (SIFS): Call for Science Verification Proposals SOAR Integral Field Spectrograph (SIFS): Call for Science Verification

More information

PHYS 202 OUTLINE FOR PART III LIGHT & OPTICS

PHYS 202 OUTLINE FOR PART III LIGHT & OPTICS PHYS 202 OUTLINE FOR PART III LIGHT & OPTICS Electromagnetic Waves A. Electromagnetic waves S-23,24 1. speed of waves = 1/( o o ) ½ = 3 x 10 8 m/s = c 2. waves and frequency: the spectrum (a) radio red

More information

Spectral Analysis of the LUND/DMI Earthshine Telescope and Filters

Spectral Analysis of the LUND/DMI Earthshine Telescope and Filters Spectral Analysis of the LUND/DMI Earthshine Telescope and Filters 12 August 2011-08-12 Ahmad Darudi & Rodrigo Badínez A1 1. Spectral Analysis of the telescope and Filters This section reports the characterization

More information

instruments Solar Physics course lecture 3 May 4, 2010 Frans Snik BBL 415 (710)

instruments Solar Physics course lecture 3 May 4, 2010 Frans Snik BBL 415 (710) Solar Physics course lecture 3 May 4, 2010 Frans Snik BBL 415 (710) f.snik@astro.uu.nl www.astro.uu.nl/~snik info from photons spatial (x,y) temporal (t) spectral (λ) polarization ( ) usually photon starved

More information

Some Aspects of Light Pollution in the Near Infrared

Some Aspects of Light Pollution in the Near Infrared Some Aspects of Light Pollution in the Near Infrared Željko Andreić 1 and Doroteja Andreić 2 1 Faculty of Mining, Geology and Petroleum Eng., University of Zagreb, Pierottijeva 6, 10 000 Zagreb, Croatia,

More information

Image Slicer for the Subaru Telescope High Dispersion Spectrograph

Image Slicer for the Subaru Telescope High Dispersion Spectrograph PASJ: Publ. Astron. Soc. Japan 64, 77, 2012 August 25 c 2012. Astronomical Society of Japan. Image Slicer for the Subaru Telescope High Dispersion Spectrograph Akito TAJITSU Subaru Telescope, National

More information

Spectroscopy Lab 2. Reading Your text books. Look under spectra, spectrometer, diffraction.

Spectroscopy Lab 2. Reading Your text books. Look under spectra, spectrometer, diffraction. 1 Spectroscopy Lab 2 Reading Your text books. Look under spectra, spectrometer, diffraction. Consult Sargent Welch Spectrum Charts on wall of lab. Note that only the most prominent wavelengths are displayed

More information

CubeSat-Scale Hyperspectral Imager for Middle Atmosphere Investigations

CubeSat-Scale Hyperspectral Imager for Middle Atmosphere Investigations CubeSat-Scale Hyperspectral Imager for Middle Atmosphere Investigations Rick Doe 1, Steve Watchorn 2, John Noto 2, Robert Kerr 2, Karl van Dyk 1, Kyle Leveque 1, and Christopher Sioris 3 1 SRI International

More information

Dual-FL. World's Fastest Fluorometer. Measure absorbance spectra and fluorescence simultaneously FLUORESCENCE

Dual-FL. World's Fastest Fluorometer. Measure absorbance spectra and fluorescence simultaneously FLUORESCENCE Dual-FL World's Fastest Fluorometer Measure absorbance spectra and fluorescence simultaneously FLUORESCENCE 100 Times Faster Data Collection The only simultaneous absorbance and fluorescence system available

More information

Technical Notes. Integrating Sphere Measurement Part II: Calibration. Introduction. Calibration

Technical Notes. Integrating Sphere Measurement Part II: Calibration. Introduction. Calibration Technical Notes Integrating Sphere Measurement Part II: Calibration This Technical Note is Part II in a three part series examining the proper maintenance and use of integrating sphere light measurement

More information

Department of Electrical Engineering and Computer Science, Massachusetts Institute of Technology, 77. Table of Contents 1

Department of Electrical Engineering and Computer Science, Massachusetts Institute of Technology, 77. Table of Contents 1 Efficient single photon detection from 500 nm to 5 μm wavelength: Supporting Information F. Marsili 1, F. Bellei 1, F. Najafi 1, A. E. Dane 1, E. A. Dauler 2, R. J. Molnar 2, K. K. Berggren 1* 1 Department

More information

Applications of Optics

Applications of Optics Nicholas J. Giordano www.cengage.com/physics/giordano Chapter 26 Applications of Optics Marilyn Akins, PhD Broome Community College Applications of Optics Many devices are based on the principles of optics

More information

Examination, TEN1, in courses SK2500/SK2501, Physics of Biomedical Microscopy,

Examination, TEN1, in courses SK2500/SK2501, Physics of Biomedical Microscopy, KTH Applied Physics Examination, TEN1, in courses SK2500/SK2501, Physics of Biomedical Microscopy, 2009-06-05, 8-13, FB51 Allowed aids: Compendium Imaging Physics (handed out) Compendium Light Microscopy

More information

Components of Optical Instruments. Chapter 7_III UV, Visible and IR Instruments

Components of Optical Instruments. Chapter 7_III UV, Visible and IR Instruments Components of Optical Instruments Chapter 7_III UV, Visible and IR Instruments 1 Grating Monochromators Principle of operation: Diffraction Diffraction sources: grooves on a reflecting surface Fabrication:

More information

The University of Toledo R. Ellingson and M. Heben

The University of Toledo R. Ellingson and M. Heben focal length, f Spectral Measurement Using a Monochromator, Thermopile Detector, and Lock-In Amplifier September 18, 2012 The University of Toledo R. Ellingson and M. Heben Where are We, Where we are Going?

More information

arxiv: v1 [astro-ph.im] 26 Mar 2012

arxiv: v1 [astro-ph.im] 26 Mar 2012 The image slicer for the Subaru Telescope High Dispersion Spectrograph arxiv:1203.5568v1 [astro-ph.im] 26 Mar 2012 Akito Tajitsu The Subaru Telescope, National Astronomical Observatory of Japan, 650 North

More information

Improved Spectra with a Schmidt-Czerny-Turner Spectrograph

Improved Spectra with a Schmidt-Czerny-Turner Spectrograph Improved Spectra with a Schmidt-Czerny-Turner Spectrograph Abstract For years spectra have been measured using traditional Czerny-Turner (CT) design dispersive spectrographs. Optical aberrations inherent

More information

ROTATING SHADOWBAND SPECTRORADIOMETER MODEL RSS-1024/UVRSS-1024 BULLETIN RSS/UVRSS-1024

ROTATING SHADOWBAND SPECTRORADIOMETER MODEL RSS-1024/UVRSS-1024 BULLETIN RSS/UVRSS-1024 ROTATING SHADOWBAND SPECTRORADIOMETER MODEL RSS-1024/UVRSS-1024 BULLETIN RSS/UVRSS-1024 General Description The Rotating Shadowband Spectroradiometer (RSS) combines a high-performance 1024-pixel Charge

More information

Basic Components of Spectroscopic. Instrumentation

Basic Components of Spectroscopic. Instrumentation Basic Components of Spectroscopic Ahmad Aqel Ifseisi Assistant Professor of Analytical Chemistry College of Science, Department of Chemistry King Saud University P.O. Box 2455 Riyadh 11451 Saudi Arabia

More information

University of Wisconsin Chemistry 524 Spectroscopic Components *

University of Wisconsin Chemistry 524 Spectroscopic Components * University of Wisconsin Chemistry 524 Spectroscopic Components * In journal articles, presentations, and textbooks, chemical instruments are often represented as block diagrams. These block diagrams highlight

More information

\Ç à{x ÇtÅx Éy ALLAH à{x `xüv yâä

\Ç à{x ÇtÅx Éy ALLAH à{x `xüv yâä \Ç à{x ÇtÅx Éy ALLAH à{x `xüv yâä Ultraviolet Radiation from Some Types of Outdoor Lighting Lamps Dr.Essam El-Moghazy Photometry and Radiometry division, National Institute for Standards (NIS), Egypt.

More information

UltraGraph Optics Design

UltraGraph Optics Design UltraGraph Optics Design 5/10/99 Jim Hagerman Introduction This paper presents the current design status of the UltraGraph optics. Compromises in performance were made to reach certain product goals. Cost,

More information

CONFIGURING. Your Spectroscopy System For PEAK PERFORMANCE. A guide to selecting the best Spectrometers, Sources, and Detectors for your application

CONFIGURING. Your Spectroscopy System For PEAK PERFORMANCE. A guide to selecting the best Spectrometers, Sources, and Detectors for your application CONFIGURING Your Spectroscopy System For PEAK PERFORMANCE A guide to selecting the best Spectrometers, s, and s for your application Spectral Measurement System Spectral Measurement System Spectrograph

More information

Powerful DMD-based light sources with a high throughput virtual slit Arsen R. Hajian* a, Ed Gooding a, Thomas Gunn a, Steven Bradbury a

Powerful DMD-based light sources with a high throughput virtual slit Arsen R. Hajian* a, Ed Gooding a, Thomas Gunn a, Steven Bradbury a Powerful DMD-based light sources with a high throughput virtual slit Arsen R. Hajian* a, Ed Gooding a, Thomas Gunn a, Steven Bradbury a a Hindsight Imaging Inc., 233 Harvard St. #316, Brookline MA 02446

More information

Noise Analysis of AHR Spectrometer Author: Andrew Xiang

Noise Analysis of AHR Spectrometer Author: Andrew Xiang 1. Introduction Noise Analysis of AHR Spectrometer Author: Andrew Xiang The noise from Spectrometer can be very confusing. We will categorize different noise and analyze them in this document from spectrometer

More information

Spectrophotometer. An instrument used to make absorbance, transmittance or emission measurements is known as a spectrophotometer :

Spectrophotometer. An instrument used to make absorbance, transmittance or emission measurements is known as a spectrophotometer : Spectrophotometer An instrument used to make absorbance, transmittance or emission measurements is known as a spectrophotometer : Spectrophotometer components Excitation sources Deuterium Lamp Tungsten

More information

Astrophotography. An intro to night sky photography

Astrophotography. An intro to night sky photography Astrophotography An intro to night sky photography Agenda Hardware Some myths exposed Image Acquisition Calibration Hardware Cameras, Lenses and Mounts Cameras for Astro-imaging Point and Shoot Limited

More information

Supplementary Materials

Supplementary Materials Supplementary Materials In the supplementary materials of this paper we discuss some practical consideration for alignment of optical components to help unexperienced users to achieve a high performance

More information

Diffraction. Interference with more than 2 beams. Diffraction gratings. Diffraction by an aperture. Diffraction of a laser beam

Diffraction. Interference with more than 2 beams. Diffraction gratings. Diffraction by an aperture. Diffraction of a laser beam Diffraction Interference with more than 2 beams 3, 4, 5 beams Large number of beams Diffraction gratings Equation Uses Diffraction by an aperture Huygen s principle again, Fresnel zones, Arago s spot Qualitative

More information

Astro-photography. Daguerreotype: on a copper plate

Astro-photography. Daguerreotype: on a copper plate AST 1022L Astro-photography 1840-1980s: Photographic plates were astronomers' main imaging tool At right: first ever picture of the full moon, by John William Draper (1840) Daguerreotype: exposure using

More information

TSBB09 Image Sensors 2018-HT2. Image Formation Part 1

TSBB09 Image Sensors 2018-HT2. Image Formation Part 1 TSBB09 Image Sensors 2018-HT2 Image Formation Part 1 Basic physics Electromagnetic radiation consists of electromagnetic waves With energy That propagate through space The waves consist of transversal

More information

BEAM HALO OBSERVATION BY CORONAGRAPH

BEAM HALO OBSERVATION BY CORONAGRAPH BEAM HALO OBSERVATION BY CORONAGRAPH T. Mitsuhashi, KEK, TSUKUBA, Japan Abstract We have developed a coronagraph for the observation of the beam halo surrounding a beam. An opaque disk is set in the beam

More information

COLOUR INSPECTION, INFRARED AND UV

COLOUR INSPECTION, INFRARED AND UV COLOUR INSPECTION, INFRARED AND UV TIPS, SPECIAL FEATURES, REQUIREMENTS LARS FERMUM, CHIEF INSTRUCTOR, STEMMER IMAGING THE PROPERTIES OF LIGHT Light is characterized by specifying the wavelength, amplitude

More information

Modern Instrumental Methods of Analysis Prof. Dr. J.R. Mudakavi Department of Chemical Engineering Indian Institute of Science, Bangalore

Modern Instrumental Methods of Analysis Prof. Dr. J.R. Mudakavi Department of Chemical Engineering Indian Institute of Science, Bangalore Modern Instrumental Methods of Analysis Prof. Dr. J.R. Mudakavi Department of Chemical Engineering Indian Institute of Science, Bangalore Module No. # 02 Lecture No. # 08 Ultraviolet and Visible Spectrophotometry

More information

Sharpness, Resolution and Interpolation

Sharpness, Resolution and Interpolation Sharpness, Resolution and Interpolation Introduction There are a lot of misconceptions about resolution, camera pixel count, interpolation and their effect on astronomical images. Some of the confusion

More information

SPECTRAL IRRADIANCE DATA

SPECTRAL IRRADIANCE DATA The radiometric data on the following pages was measured in our Standards Laboratory. The wavelength calibrations are based on our spectral calibration lamps. Irradiance data from 250 to 2500 nm is based

More information

Gemini 8m Telescopes Instrument Science Requirements. R. McGonegal Controls Group. January 27, 1996

Gemini 8m Telescopes Instrument Science Requirements. R. McGonegal Controls Group. January 27, 1996 GEMINI 8-M Telescopes Project Gemini 8m Telescopes Instrument Science Requirements R. McGonegal Controls Group January 27, 1996 GEMINI PROJECT OFFICE 950 N. Cherry Ave. Tucson, Arizona 85719 Phone: (520)

More information

On-line spectrometer for FEL radiation at

On-line spectrometer for FEL radiation at On-line spectrometer for FEL radiation at FERMI@ELETTRA Fabio Frassetto 1, Luca Poletto 1, Daniele Cocco 2, Marco Zangrando 3 1 CNR/INFM Laboratory for Ultraviolet and X-Ray Optical Research & Department

More information

Detection and application of Doppler and motional Stark features in the DNB emission spectrum in the high magnetic field of the Alcator C-Mod tokamak

Detection and application of Doppler and motional Stark features in the DNB emission spectrum in the high magnetic field of the Alcator C-Mod tokamak Detection and application of Doppler and motional Stark features in the DNB emission spectrum in the high magnetic field of the Alcator C-Mod tokamak I. O. Bespamyatnov a, W. L. Rowan a, K. T. Liao a,

More information

A stray light corrected array spectroradiometer for complex high dynamic range measurements in the UV spectral range.

A stray light corrected array spectroradiometer for complex high dynamic range measurements in the UV spectral range. A stray light corrected array spectroradiometer for complex high dynamic range measurements in the UV spectral range Mike Clark Gigahertz-Optik GmbH m.clark@gigahertz-optik.de Array spectroradiometers

More information

Chapter 16 Light Waves and Color

Chapter 16 Light Waves and Color Chapter 16 Light Waves and Color Lecture PowerPoint Copyright The McGraw-Hill Companies, Inc. Permission required for reproduction or display. What causes color? What causes reflection? What causes color?

More information

CHAPTER 6 Exposure Time Calculations

CHAPTER 6 Exposure Time Calculations CHAPTER 6 Exposure Time Calculations In This Chapter... Overview / 75 Calculating NICMOS Imaging Sensitivities / 78 WWW Access to Imaging Tools / 83 Examples / 84 In this chapter we provide NICMOS-specific

More information

Compact Dual Field-of-View Telescope for Small Satellite Payloads

Compact Dual Field-of-View Telescope for Small Satellite Payloads Compact Dual Field-of-View Telescope for Small Satellite Payloads James C. Peterson Space Dynamics Laboratory 1695 North Research Park Way, North Logan, UT 84341; 435-797-4624 Jim.Peterson@sdl.usu.edu

More information

UV/Optical/IR Astronomy Part 2: Spectroscopy

UV/Optical/IR Astronomy Part 2: Spectroscopy UV/Optical/IR Astronomy Part 2: Spectroscopy Introduction We now turn to spectroscopy. Much of what you need to know about this is the same as for imaging I ll concentrate on the differences. Slicing the

More information

Light, Color, Spectra 05/30/2006. Lecture 17 1

Light, Color, Spectra 05/30/2006. Lecture 17 1 What do we see? Light Our eyes can t t detect intrinsic light from objects (mostly infrared), unless they get red hot The light we see is from the sun or from artificial light When we see objects, we see

More information

Calibration of a High Dynamic Range, Low Light Level Visible Source

Calibration of a High Dynamic Range, Low Light Level Visible Source Calibration of a High Dynamic Range, Low Light Level Visible Source Joe LaVeigne a, Todd Szarlan a, Nate Radtke a a Santa Barbara Infrared, Inc., 30 S. Calle Cesar Chavez, #D, Santa Barbara, CA 93103 ABSTRACT

More information

Sensors and Sensing Cameras and Camera Calibration

Sensors and Sensing Cameras and Camera Calibration Sensors and Sensing Cameras and Camera Calibration Todor Stoyanov Mobile Robotics and Olfaction Lab Center for Applied Autonomous Sensor Systems Örebro University, Sweden todor.stoyanov@oru.se 20.11.2014

More information

INTRODUCTION TO CCD IMAGING

INTRODUCTION TO CCD IMAGING ASTR 1030 Astronomy Lab 85 Intro to CCD Imaging INTRODUCTION TO CCD IMAGING SYNOPSIS: In this lab we will learn about some of the advantages of CCD cameras for use in astronomy and how to process an image.

More information

Chapter 5 Nadir looking UV measurement.

Chapter 5 Nadir looking UV measurement. Chapter 5 Nadir looking UV measurement. Part-II: UV polychromator instrumentation and measurements -A high SNR and robust polychromator using a 1D array detector- UV spectrometers onboard satellites have

More information

HR2000+ Spectrometer. User-Configured for Flexibility. now with. Spectrometers

HR2000+ Spectrometer. User-Configured for Flexibility. now with. Spectrometers Spectrometers HR2000+ Spectrometer User-Configured for Flexibility HR2000+ One of our most popular items, the HR2000+ Spectrometer features a high-resolution optical bench, a powerful 2-MHz analog-to-digital

More information

Horiba LabRAM ARAMIS Raman Spectrometer Revision /28/2016 Page 1 of 11. Horiba Jobin-Yvon LabRAM Aramis - Raman Spectrometer

Horiba LabRAM ARAMIS Raman Spectrometer Revision /28/2016 Page 1 of 11. Horiba Jobin-Yvon LabRAM Aramis - Raman Spectrometer Page 1 of 11 Horiba Jobin-Yvon LabRAM Aramis - Raman Spectrometer The Aramis Raman system is a software selectable multi-wavelength Raman system with mapping capabilities with a 400mm monochromator and

More information

Spectral Transmission Measurements on various Astronomical Filters.

Spectral Transmission Measurements on various Astronomical Filters. Spectral Transmission Measurements on various Astronomical Filters. Andreas Bartels - June 2008 Thanks to my friend Olivier, who provided the Spectrometer, I was able to do some spectral transmission measurements

More information

Vision. The eye. Image formation. Eye defects & corrective lenses. Visual acuity. Colour vision. Lecture 3.5

Vision. The eye. Image formation. Eye defects & corrective lenses. Visual acuity. Colour vision. Lecture 3.5 Lecture 3.5 Vision The eye Image formation Eye defects & corrective lenses Visual acuity Colour vision Vision http://www.wired.com/wiredscience/2009/04/schizoillusion/ Perception of light--- eye-brain

More information

High Speed Hyperspectral Chemical Imaging

High Speed Hyperspectral Chemical Imaging High Speed Hyperspectral Chemical Imaging Timo Hyvärinen, Esko Herrala and Jouni Jussila SPECIM, Spectral Imaging Ltd 90570 Oulu, Finland www.specim.fi Hyperspectral imaging (HSI) is emerging from scientific

More information

ALISEO: an Imaging Interferometer for Earth Observation

ALISEO: an Imaging Interferometer for Earth Observation ALISEO: an Imaging Interferometer for Earth Observation A. Barducci, F. Castagnoli, G. Castellini, D. Guzzi, C. Lastri, P. Marcoionni, I. Pippi CNR IFAC Sesto Fiorentino, ITALY ASSFTS14 Firenze - May 6-8,

More information

Miniaturized Spectroradiometer

Miniaturized Spectroradiometer Miniaturized Spectroradiometer Thomas Morgenstern, Gudrun Bornhoeft, Steffen Goerlich JETI Technische Instrumente GmbH, Jena, Germany Abstract This paper describes the basics of spectroradiometric instruments

More information

Spectrophotometric calibration system for DECam

Spectrophotometric calibration system for DECam Spectrophotometric calibration system for DECam J.-P. Rheault*, D. L. DePoy, T. W. Behm, E. W. Kylberg, Kris Cabral, Rick Allen, J. L. Marshall, Department of Physics and Astronomy, Texas A&M University,

More information

ISUAL Imager Science Performance Test Report Author: Stewart Harris

ISUAL Imager Science Performance Test Report Author: Stewart Harris ISUAL Imager Science Performance Test Report Author: Stewart Harris Revision History: Level Date Description A 17 Mar 2003 initial release B 31 Mar 2003 included additional set up information, radiance

More information

Make Your Own Digital Spectrometer With Diffraction Grating

Make Your Own Digital Spectrometer With Diffraction Grating Make Your Own Digital Spectrometer With Diffraction Grating T. Z. July 6, 2012 1 Introduction and Theory Spectrums are very useful for classify atoms and materials. Although digital spectrometers such

More information

Astr 535 Class Notes Fall

Astr 535 Class Notes Fall Astr 535 Class Notes Fall 2017 86 4. Observing logs: summary program informtion, weather information, calibration data, seeing information, exposure information. COMMENTS are critical. READABILITY is critical

More information

Reikan FoCal Aperture Sharpness Test Report

Reikan FoCal Aperture Sharpness Test Report Focus Calibration and Analysis Software Reikan FoCal Sharpness Test Report Test run on: 10/02/2016 19:57:05 with FoCal 2.0.6.2416W Report created on: 10/02/2016 19:59:09 with FoCal 2.0.6W Overview Test

More information