NEW DESIGN FOR S.F.A STENT : Dual Component
|
|
- Karin Kelly
- 6 years ago
- Views:
Transcription
1 NEW DESIGN FOR S.F.A STENT : Dual Component 1 Max Amor Louis MAX Pasteur AMOR Clinic Above-Knee Essey and les Below-Knee Nancy Data France
2 Dual Component Stent Design Designed to: Maximize flexibility while minimizing risk of stent fracture Allow axial compression while resisting stent elongation Conforms and allows vessel movement
3 Gore Solution: Dual Component Stent Design CARMEDA and CBAS are trademarks of Carmeda AB, a wholly owned subsidiary of W. L. Gore & Associates, Inc. GORE and designs are trademarks of W. L. Gore & Associates W. L. Gore & Associates, Inc.
4 Fluoropolymer Enables Axial Compressibility GORE and designs are trademarks of W. L. Gore & Associates W. L. Gore & Associates, Inc.
5 Progression of Evaluation Bench-top Testing Pre-Clinical In Vivo Testing Limited Release Clinical Evaluation
6 Elongation Testing Test 6 mm x 80 mm stents Clamp with 70 mm exposed Apply 70 grams force Measure elongation Significance Less elongation provides deployment accuracy and may decreased potential for stent fracture GORE and designs are trademarks of W. L. Gore & Associates W. L. Gore & Associates, Inc.
7 Elongation Results LifeStent, and LUMINEXX, are trademarks and / or registered trademarks of C. R. Bard, Inc. S.M.A.R.T. and CONTROL are trademarks of Cordis Corporation. Protégé and EverFlex are trademarks of Covidien. ABSOLUTE is a trademark of Abbott Laboratories. sinus-superflex is a trademark of OptiMed. ZILVER and PTX are trademarks of COOK Medical. SUPERA is a trademark of IDEV. GORE, TIGRIS, and designs are trademarks of W. L. Gore & Associates W. L. Gore & Associates, Inc.
8 Impact of Elongation LifeStent is a trademark and / or registered trademarks of C. R. Bard, Inc. GORE, TIGRIS, and designs are trademarks of W. L. Gore & Associates W. L. Gore & Associates, Inc.
9 Straightening Force Testing Test 6 mm x 80 mm stents Clamp one end of the stent Displace the free end 15 mm down Measure upward force of stent Significance Less straightening force allows the stent to conform to vessel movement GORE and designs are trademarks of W. L. Gore & Associates W. L. Gore & Associates, Inc.
10 Straightening Force Results LifeStent, and LUMINEXX, are trademarks and / or registered trademarks of C. R. Bard, Inc. S.M.A.R.T. and CONTROL are trademarks of Cordis Corporation. Protégé and EverFlex are trademarks of Covidien. ABSOLUTE is a trademark of Abbott Laboratories. sinus-superflex is a trademark of OptiMed. ZILVER and PTX are trademarks of COOK Medical. SUPERA is a trademark of IDEV. GORE, TIGRIS, and designs are trademarks of W. L. Gore & Associates W. L. Gore & Associates, Inc.
11 Flexibility Testing GORE TIGRIS Vascular Stent BARD LifeStent Vascular Stent IDEV SUPERA Stent Cordis S.M.A.R.T. CONTROL Stent Covidien Protege EverFlex Stent LifeStent are trademarks and / or registered trademarks of C. R. Bard, Inc. S.M.A.R.T. and CONTROL are trademarks of Cordis Corporation. Protégé are trademarks of Covidien. SUPERA is a trademark of IDEV. GORE, TIGRIS, and designs are trademarks of W. L. Gore & Associates W. L. Gore & Associates, Inc.
12 Flexibility Testing (continued) Cook Zilver PTX Stent OptiMed sinus-superflex Stent Bard Lumminex Stent LUMINEXX, are trademarks and / or registered trademarks of C. R. Bard, Inc. sinus-superflex is a trademark of OptiMed. ZILVER and PTX are trademarks of COOK Medical.
13 Flexibility Testing (continued) GORE TIGRIS Vascular Stent Bard Lumminex Stent LUMINEXX is a trademark and / or registered trademarks of C. R. Bard, Inc. GORE, TIGRIS, and designs are trademarks of W. L. Gore & Associates W. L. Gore & Associates, Inc.
14 Longitudinal Compression Test 6 mm x 80 mm stents Compress 12 mm (15%) Measure force of stent resisting compression Significance Less resistance to compression allows the stent to naturally conform to vessel shortening and decreases potential for stent fracture GORE and designs are trademarks of W. L. Gore & Associates W. L. Gore & Associates, Inc.
15 Longitudinal Compression Testing Results LifeStent, and LUMINEXX, are trademarks and / or registered trademarks of C. R. Bard, Inc. S.M.A.R.T. and CONTROL are trademarks of Cordis Corporation. Protégé and EverFlex are trademarks of Covidien. ABSOLUTE is a trademark of Abbott Laboratories. sinus-superflex is a trademark of OptiMed. ZILVER and PTX are trademarks of COOK Medical. SUPERA is a trademark of IDEV. GORE,TIGRIS, and designs are trademarks of W. L. Gore & Associates W. L. Gore & Associates, Inc.
16 Longitudinal Compression Testing GORE TIGRIS Vascular Stent Cordis S.M.A.R.T. CONTROL Stent BARD LifeStent Vascular Stent Covidien Protege EverFlex Stent LifeStent are trademarks and / or registered trademarks of C. R. Bard, Inc. S.M.A.R.T. and CONTROL are trademarks of Cordis Corporation. Protégé and EverFlex are trademarks of Covidien. GORE,TIGRIS, and designs are trademarks of W. L. Gore & Associates W. L. Gore & Associates, Inc.
17 Blood Loop Test Test 6 mm x 80 mm stents Deploy in 5 mm tubing Circulate human blood at 37 C for two hours Significance Less thrombus formation maximizes potential for clinical patency GORE and designs are trademarks of W. L. Gore & Associates W. L. Gore & Associates, Inc.
18 Blood Loop Testing GORE TIGRIS Vascular Stent Covidien Protege EverFlex Stent IDEV SUPERA Stent Bard Lumminex Stent Abbott Absolute Pro LL Stent BARD LifeStent Vascular Stent LifeStent, and LUMINEXX, are trademarks and / or registered trademarks of C. R. Bard, Inc. Protégé and EverFlex are trademarks of Covidien. ABSOLUTE is a trademark of Abbott Laboratories. SUPERA is a trademark of IDEV. GORE,TIGRIS, and designs are trademarks of W. L. Gore & Associates W. L. Gore & Associates, Inc.
19 Blood Loop Testing (continued) OptiMed sinus-superflex Stent Cordis S.M.A.R.T. CONTROL Stent Cook Zilver PTX Stent S.M.A.R.T. and CONTROL are trademarks of Cordis Corporation. sinus-superflex is a trademark of OptiMed. ZILVER and PTX are trademarks of COOK Medical. GORE and designs are trademarks of W. L. Gore & Associates W. L. Gore & Associates, Inc.
20 Chronic In Vivo: 90 day histology Lumen surface of overlapped section is widely patent with a thin endothelialized neointima. Neointimal consists entirely of organized, fibromuscular neointima. Virtually no inflammatory cells present. GORE and designs are trademarks of W. L. Gore & Associates W. L. Gore & Associates, Inc.
21 First in Man Cases Park Krankenhaus, Leipzig, Germany Mid SFA Lesion 7 x 40 GORE TIGRIS Vascular Stent Popliteal Lesion 6 x 80 GORE TIGRIS Vascular Stent GORE,TIGRIS, and designs are trademarks of W. L. Gore & Associates W. L. Gore & Associates, Inc.
22 Conclusion The GORE TIGRIS Vascular Stent dual component design provides the unique combination of: Minimal stent elongation Low straightening force The ability to absorb longitudinal compression It mimics vessel movement and minimizes potential for stent fractures GORE,TIGRIS, and designs are trademarks of W. L. Gore & Associates W. L. Gore & Associates, Inc.
23
24 Images GORE and designs are trademarks of W. L. Gore & Associates W. L. Gore & Associates, Inc.
25 Images GORE and designs are trademarks of W. L. Gore & Associates W. L. Gore & Associates, Inc.
26 Images GORE and designs are trademarks of W. L. Gore & Associates W. L. Gore & Associates, Inc.
27 Images GORE and designs are trademarks of W. L. Gore & Associates W. L. Gore & Associates, Inc.
28 GORE, TIGRIS, VIABAHN, and designs are trademarks of W. L. Gore & Associates W. L. Gore & Associates, Inc. AR0149-EU1 APRIL 2012
Wire Technology: How to Choose Based on Anatomy and Strategy
Wire Technology: How to Choose Based on Anatomy and Strategy Craig M. Walker, MD, FACC, FACP Clinical Professor of Medicine Tulane University School of Medicine New Orleans, LA Clinical Professor of Medicine
More informationOptical Characteristics of Next Generation Dual Optic IOL
Optical Characteristics of Next Generation Dual Optic IOL Scott Evans, MD Sanjeev Kasthurirangan, PhD Val Portney, PhD Financial Disclosures Scott Evans is an employee of Abbott Medical Optics Inc. Sanjeev
More informationELVAREX CUSTOM MADE MEASURING INTRODUCTORY WORKSHOP
ELVAREX CUSTOM MADE MEASURING INTRODUCTORY WORKSHOP Learning Objectives Ability to recognise the most appropriate time to measure for garment Understand the differences between flat knit and circular knit,
More informationInterdisciplinarity on the Bench Top
Interdisciplinarity on the Bench Top Model-based reasoning in bio-science and -engineering university research laboratories Nancy J. Nersessian Georgia Institute of Technology www.cc.gatech.edu/~nersessian
More informationLeading the Way in Guide Wire Technology
1/13 Leading the Way in Comprehensive portfolio of guide wires Advanced Technology Abbott Vascular Guide Wires / Indications: This HI-TORQUE Guide Wire is intended to facilitate the placement of balloon
More informationEnhanced Functionality of High-Speed Image Processing Engine SUREengine PRO. Sharpness (spatial resolution) Graininess (noise intensity)
Vascular Enhanced Functionality of High-Speed Image Processing Engine SUREengine PRO Medical Systems Division, Shimadzu Corporation Yoshiaki Miura 1. Introduction In recent years, digital cardiovascular
More informationBionic Fistula Needles. The gentle puncture
Bionic Fistula Needles The gentle puncture Oozing at the puncture site leads to an increased risk of infection for patients and staff. The Bionic needle design prevents these side-effects by its cylindrical
More informationProstar XL Percutaneous Vascular Surgical Device
Clinical leader for percutaneous large access site closure Close Access Sites Up to 24F Secure Closure Prostar XL Percutaneous Vascular Surgical Device Improved Outcomes 2009 Abbott Laboratories. All rights
More information160-slice CT SCANNER / New Standard for the Future
TECHNOLOGY HISTORY For over 130 years, Toshiba has been a world leader in developing technology to improve the quality of life. Our 50,000 global patents demonstrate a long, rich history of leading innovation.
More informationAccepts Wire Guide Diameter inch. Stent Diameter mm
Order Number Reference Part Number Accepts Wire Guide inch Length Minimum ID inch Fr Formula 418 80 cm Over-the-Wire Delivery System G56605 FOR418-18-80-4-12.018 4 12.068 6.0/5.0 G56606 FOR418-18-80-4-16.018
More information(12) United States Patent (10) Patent No.: US 7.556,644 B2
USOO7556644B2 (12) United States Patent (10) Patent No.: US 7.556,644 B2 Burpee et al. (45) Date of Patent: Jul. 7, 2009 (54) FLEXIBLE STENT 5,545,210 A * 8/1996 Hess et al.... 128/898 5,669,932 A * 9/1997
More informationSpeedTip CCS 5.0, 7.0
SURGICAL TECHNIQUE STEP BY STEP SpeedTip CCS 5.0, 7.0 Cannulated Compression Screws APTUS 2 SpeedTip CCS 5.0, 7.0 Cannulated Compression Screws SpeedTip CCS 5.0, 7.0 Cannulated Compression Screw 3 SpeedTip
More informationEUROPEAN PATENT OFFICE U.S. PATENT AND TRADEMARK OFFICE CPC NOTICE OF CHANGES 537 DATE: AUGUST 1, 2018 PROJECT MP0184. Action Subclass Group(s)
EUROPEAN PATENT OFFICE U.S. PATENT AND TRADEMARK OFFICE CPC NOTICE OF CHANGES 537 The following classification changes will be effected by this Notice of Changes: Action Subclass Group(s) SCHEME: Titles
More informationSpeedTip CCS 5.0, 7.0
SURGICAL TECHNIQUE STEP BY STEP SpeedTip CCS 5.0, 7.0 Cannulated Compression Screws APTUS 2 SpeedTip CCS 5.0, 7.0 Cannulated Compression Screws Contents 3 Introduction Product Materials Indications Contraindications
More informationEBI 10 High Precision Wireless Data Loggers
8 EBI 10 High Precision Wireless Data Loggers The Perfect Solution for Your Applications To fit each of your specific processes, ebro offers a wide variety of EBI 10 temperature, pressure and humidity
More informationCalcaneus StepPlate. Surgical Technique. Calcaneus StepPlate
Calcaneus StepPlate Surgical Technique Calcaneus StepPlate Medializing Calcaneal Osteotomy (MCO) with Calcaneus StepPlate 1 2 Perform an incision on the lateral side of the calcaneus. An L -shaped incision
More informationThe Current CTO Toolbox
The Current CTO Toolbox R. Michael Wyman, MD Director, Cardiac Catheterization Laboratory and Cardiovascular Interventional Research Torrance Memorial Medical Center Torrance, CA Disclosure Statement of
More informationFOCUSED INNOVATIONS IN LIFE SCIENCES
FOCUSED INNOVATIONS IN LIFE SCIENCES INNOVATIONS IN LIFE SCIENCES Gore first entered the Life Sciences industry in the mid-1970s with a life-saving blood vessel replacement engineered to work in harmony
More informationAnatomy of Contemporary CTO Guidewires and Selection Criteria
Anatomy of Contemporary CTO Guidewires and Selection Criteria Craig A. Thompson, M.D., MMSc. Director, Invasive Cardiology and Vascular Medicine Yale University School of Medicine New Haven, CT USA Disclosure
More informationSuccessful endovascular treatment for BTK lesion using wire rendezvous technique and retrograde knuckle wire technique by collateral approach
Successful endovascular treatment for BTK lesion using wire rendezvous technique and retrograde knuckle wire technique by collateral approach Katsutoshi Takayama, MD, Ph.D Department of Radiology and Interventional
More informationForward Looking Statements
NYSE MKT: NSPR February 2015 Forward Looking Statements This presenta-on contains forward- looking statements. Such statements may be preceded by the words intends, may, will, plans, expects, an-cipates,
More informationHow to survive the MDR
How to survive the MDR Louis Habets LifetecZONe, 21 September 2017 1 Agenda About the Medical Device Regulation Specific project! Highlights only showing where to find Discussion and Questions LifetecZONe,
More informationThermo Scientific SPECTRONIC 200 Visible Spectrophotometer. The perfect. teaching instrument
Thermo Scientific SPECTRONIC 200 Visible Spectrophotometer The perfect teaching instrument Designed for the Teaching Laboratory Thermo Scientific SPECTRONIC spectrophotometers have served as core analytical
More information(12) Patent Application Publication (10) Pub. No.: US 2004/ A1
(19) United States US 20040073141A1 (12) Patent Application Publication (10) Pub. No.: US 2004/0073141 A1 Hartley et al. (43) Pub. Date: Apr. 15, 2004 (54) GUIDE WIRE (75) Inventors: David Ernest Hartley,
More informationThe thin wall ESI tip appears unusually. advantageous for ensuring stable ESI. conditions and efficient ion transfers.
The thin wall ESI tip appears unusually advantageous for ensuring stable ESI conditions and efficient ion transfers. Valaskovic, Kelleher, McLafferty Science (1996), 273, 1199 perfect zero-dead-volume
More informationPerclose ProGlide. Deployment Steps. Suture-Mediated Closure System Abbott. All rights reserved. ANZ /11
Deployment Steps SUTURE TRIMMER Handle Needle Plunger Collar Lever Marker Lumen Body Foot QuickCut Link Foot (Deployed) Proximal Guide Guide Wire Exit Port Distal Guide Marker Port Sheath posterior side
More informationDefibrillation/Surge/ESD Protector
MAX334 General Description The MAX334 is a patent-pending protection device intended to (with the help of external, energy-rated resistors) absorb repetitive defibrillation and other high-energy pulses
More informationThermo Scientific SPECTRONIC 200 Education
molecular spectroscopy Thermo Scientific SPECTRONIC 200 Education Part of Thermo Fisher Scientific Designed for the Teaching Laboratory Classroom Friendly Sample Compartment Whether you measure in 10 mm
More informationPre-clinical Development Support Services
Pre-clinical Development Support Services Enabling device development with unique understanding Our Company - Vascular Flow Technologies Vascular Flow Technologies (VFT) has evolved from a tiny surgeon
More informationExperimental and numerical investigation of tube sinking of rectangular tubes from round section
International Journal of Engineering and Technology sciences (IJETS) ISSN 2289-4152 Academic Research Online Publisher Research Article Experimental and numerical investigation of tube sinking of rectangular
More informationAn Acoustic Fluid-structure Simulation of a Therapeutic Ultrasound Wire Waveguide Apparatus
Dublin Institute of Technology ARROW@DIT Conference Papers School of Manufacturing and Design Engineering 2005-01-01 An Acoustic Fluid-structure Simulation of a Therapeutic Ultrasound Wire Waveguide Apparatus
More informationThe OPUS AutoCuff System for Rotator Cuff Repair
The OPUS AutoCuff System for Rotator Cuff Repair A revolutionary new system specifically designed for rotator cuff repair surgery System Includes: SmartStitch Suturing Device The Magnum PI Implant with
More informationIntroduction. Parametric Imaging. The Ultrasound Research Interface: A New Tool for Biomedical Investigations
The Ultrasound Research Interface: A New Tool for Biomedical Investigations Shelby Brunke, Laurent Pelissier, Kris Dickie, Jim Zagzebski, Tim Hall, Thaddeus Wilson Siemens Medical Systems, Issaquah WA
More information(12) United States Patent
USOO9474532B2 (12) United States Patent Yamanaka et al. (10) Patent No.: (45) Date of Patent: US 9.474,532 B2 Oct. 25, 2016 (54) EMBOLIZATION COIL (75) Inventors: Yasushi Yamanaka, Kanagawa (JP); Atsushi
More informationPART. Maxim Integrated Products 1
19-1999; Rev 4; 7/04 3.2Gbps Adaptive Equalizer General Description The is a +3.3V adaptive cable equalizer designed for coaxial and twin-axial cable point-to-point communications applications. The equalizer
More informationMethod of Declaration. Hospitality/Interest/Sponsorship
Hospitality/Interest/Sponsorship Date Received Method of Declaration Abbott Laboratories Limited 23/03/12 Sponsorship Abbott Laboratories Limited 28/05/12 Sponsorship Abbott Laboratories Limited 23/03/12
More informationBlood Vessel Detection in Images from Laser-Heated Skin
Blood Vessel Detection in Images from Laser-Heated Skin Abstract Alireza Kavianpour, Simin Shoari, Behdad Kavianpour CEIS Dept. DeVry University, Pomona, CA 91768 A computer method for recognizing blood
More informationIntegre Pro Scan combines pattern scanning and multi-color photocoagulation in our unique all-in-one laser/slit lamp design.
Integre Pro Scan combines pattern scanning and multi-color photocoagulation in our unique all-in-one laser/slit lamp design. Multi-color scanning photocoagulation takes on a new look. Integre Pro Scan
More informationMedicaLight SURGICAL AND EXAMINATION LIGHTING FOR VETERINARY FACILITIES
MedicaLight SURGICAL AND EXAMINATION LIGHTING FOR VETERINARY FACILITIES Celestial Star The Celestial Star offers ceiling, mobile, and wall configurations. This light offers maximum reach and range of motion,
More informationSmart Electromechanical Systems Modules
Smart Electromechanical Systems Modules A.E. Gorodetskiy Abstract The article considers design features of standard modules of smart electromechanical systems (SM SEMS). Also, shows that a variety of structures
More informationAltitude and Airspeed Measurements Temperature Range. Range psid psig psia 0 to 1 0 to 5 0 to 15 0 to 30 0 to 50 0 to to to 300
PC Board Mountable Pressure Sensor 0-100 mv Output Voltage Excitation Differential, Gage, and Absolute Temperature Compensated DESCRIPTION The MS4425 is a temperature compensated, piezoresistive silicon
More informationThe Perfect Close. to your surgical procedure. PER F ORM ANC E through innovation
The Perfect Close to your surgical procedure PER F ORM ANC E through innovation GORE-TEX Suture is a unique, microporous, nonabsorbable monofilament made of expanded polytetrafluoroethylene (eptfe), the
More informationCIRCULAR POLARIZED UHF ANTENNA SlimLine A5020
AB O U T TIMES -7 Pushing the boundaries of RFID technology worldwide Times-7 are leaders in RFID antenna design and manufacture. Our patented award winning UHF antennas meet the needs 150 mm / 5.9 in.
More informationBEAM DISTORTION IN DOPPLER ULTRASOUND FLOW TEST RIGS: MEASUREMENT USING A STRING PHANTOM
BEAM DISTORTION IN DOPPLER ULTRASOUND FLOW TEST RIGS: MEASUREMENT USING A STRING PHANTOM R. Steel, P. J. Fish School of Informatics, University of Wales, Bangor, UK Abstract-The tube in flow rigs used
More informationArterial Tissue Perforation Using Ultrasonically Vibrating Wire Waveguides
Dublin Institute of Technology ARROW@DIT Doctoral Engineering 2011 Arterial Tissue Perforation Using Ultrasonically Vibrating Wire Waveguides Mark Wylie Dublin Institute of Technology, Mark.wylie@dit.ie
More informationC A R D I O VA S C U L A R P R O D U C T S C ATA L O G
C A R D I O VA S C U L A R P R O D U C T S C ATA L O G MYO/WIRE TM AT R I A L T E M P O R A RY C A R D I A C PA C I N G W I R E S MYO/WIRE TM V E N T R I C U L A R T E M P O R A RY C A R D I A C PA C I
More informationCook Medical Peripheral Intervention MRI Information
Page 1 of 13 Home My Account Help Careers Text size: A A A Home / Clinical / Peripheral Intervention / Products / I Information Browse by Product Family Disease State Alphabetical Cook Medical adheres
More information(Biomedical Engineering Design) Spring 2009
BME 3013 (Biomedical Engineering Design) Spring 2009 Project #27# 27: Manipulatable Intracoronary Wire MID- SEMESTER REPORT March 11, 2009 TEAM MEMBERS: Allison McArton (Team Leader) Angwei Law (Communicator)
More informationKey Issues for Successful E-Health Solutions
Key Issues for Successful E-Health Solutions Prof. Nilmini Wickramasinghe, PhD MBA Epworth Chair Health Information Management RMIT University, School of Business IT and Logistics Email: nilmini.work@gmail.com
More informationMagnum Marketing.
+91-8048022364 Magnum Marketing https://www.indiamart.com/magnummarketing-delhi/ We, Magnum Marketing are leading Importer, Wholesaler and Supplier of Introducer Sheaths Machine, Angiography Catheters,
More informationPrecision Vascular Robotics. Corindus Vascular Robotics (CVRS) January 2017
Precision Vascular Robotics Corindus Vascular Robotics (CVRS) January 2017 1 Forward Looking Statements THIS PRESENTATION CONTAINS FORWARD-LOOKING STATEMENTS (AS SUCH TERM IS DEFINED IN SECTION 27A OF
More informationReflex TM Surgical Technique. Anterior Cervical Plate
Reflex TM Surgical Technique Anterior Cervical Plate Surgical Technique Acknowledgement: Stryker Spine extends their thanks to the following surgeons for their participation in the development of the Reflex
More informationPublication number: A1. int. Ci.5; A61M 25/00, A61 M 25/01
Europaisches Patentamt European Patent Office Office europeen des brevets Publication number: 0 532 109 A1 EUROPEAN PATENT APPLICATION Application number: 92202725.5 int. Ci.5; A61M 25/00, A61 M 25/01
More informationRevolutionary. Heart Valve Durability Tester. VDT-3600i. Medical device testing equipment Testing lab Simulation platforms Silicone mock vessels
Revolutionary Heart Valve Durability Tester VDT-3600i Medical device testing equipment Testing lab Simulation platforms Silicone mock vessels The revolutionary heart valve durability tester from BDC Labs
More informationMagnetic Micro Testing System Microservo MMT Series C225-E014A
Magnetic Micro Testing System Microservo MMT Series C225-E014A Microservo MMT Series Magnetic Micro Testing System In recent years strength evaluation of micro materials and micro parts is increasing its
More informationThe Perfect Close. to your surgical procedure. PER F ORM ANC E through innovation
The Perfect Close to your surgical procedure PER F ORM ANC E through innovation GORE-TEX Suture is a unique, microporous, nonabsorbable monofilament made of expanded polytetrafluoroethylene (eptfe), the
More informationMicroscopes & cells. 2. arm. 3. ocular lens. 4. objective lenses. 5. stage. 6. slide clamp. 7. stage controls
Microscopes & cells Objectives: At the end of this lab you should be able to: o demonstrate the safe and proper handling of a microscope, including carrying a microscope, slide placement, and storage.
More information/ Rudder Installation Guide
/ Rudder Installation Guide Parts List Part # Part Description Qty Part # Part Description Qty 1030 Plastic Bead 1 1195 Rope; 1/8" Black Nylon 1 @ 9 1044 1/4-20 x 5/8" Pan Head Bolt 2 1353 Sliding Track
More informationASAHI Peripheral Guide Wire
ASAHI Peripheral Guide Wire AMK-DT256 Ver.1.01 / 12TS017 SYMBOLS Legal Manufacturer LOT Lot number Do not use if package is damaged REF Catalogue number Do not reuse STERILE EO Sterilized using ethylene
More informationArthroscopic Protector Meniscus Suturing Surgical Techniques
Arthroscopic Protector Meniscus Suturing Surgical Techniques Protector Meniscus Suturing Protector Meniscus Suturing Technique I For the relatively rare meniscus tears located sufficiently anterior, to
More informationMedical Extrusion & Secondary Ops SPOTLIGHT Conference November 10, 2016 Arburg Technical Center Irvine, CA
Medical Extrusion & Secondary Ops SPOTLIGHT Conference November 10, 2016 Arburg Technical Center Irvine, CA Thursday, November 10, 2016 Irvine, CA 7:30-8:15AM Check in and continental breakfast SPOTLIGHT
More informationAbbott 36 TH ANNUAL J.P. MORGAN HEALTHCARE CONFERENCE. January 8, 2018 Brian Yoor, Executive Vice President, Finance and CFO
Abbott 36 TH ANNUAL J.P. MORGAN HEALTHCARE CONFERENCE January 8, 2018 Brian Yoor, Executive Vice President, Finance and CFO FORWARD LOOKING STATEMENT Some statements in this presentation may be forward-looking
More informationPresentation Temporary Presentation
Presentation 2017 Temporary Presentation 2014 12.08.2014 1 Company structure Sales & production subsidiary USA Sales subsidiary Germany Headquarters & production site Worldwide representation Custom designed
More informationDesigning an MR compatible Time of Flight PET Detector Floris Jansen, PhD, Chief Engineer GE Healthcare
GE Healthcare Designing an MR compatible Time of Flight PET Detector Floris Jansen, PhD, Chief Engineer GE Healthcare There is excitement across the industry regarding the clinical potential of a hybrid
More informationConnecting Systems for Cable Repair (Cable Splices)
p pproval roramms for U rcos orasato Mutual coto (M)No. 2.9 STANDARD FOR CERTIFICATION Approval Programmes Type Approval Programmes for EU recognised organisation Mutual Recognition (MR) No. MR-E015 Connecting
More informationProgrammable transmitter of temperature, relative humidity and other derived humidity values with 4-20 ma outputs Instruction Manual
T3111 TRANSMITTER Programmable transmitter of temperature, relative humidity and other derived humidity values with 4-20 ma outputs Instruction Manual Instruction manual for use of T3111 transmitters Transmitter
More informationPhysics Spring 2006 Experiment 9 TRAVELING WAVES
Physics 31210 Spring 2006 Experiment 9 TRAVELING WAVES Reference: Halliday, Resnick & Walker, 7th Ed., Sections 16-1 to 5, Sections 17-1 to 4 I. Introduction: Waves of all kinds, propagating through many
More informationEndovascular aortic repair Fenestrated
Endovascular aortic repair Fenestrated Zenith ENDOVASCULAR GRAFTS MEDICAL www.cookmedical.com/web/aortic-intervention/ Active Fixation Our anchoring barbs are the industry standard for design and migration
More informationTechnical Note. Technical Note - DLC. Dynamic Load Compensation
Technical Note Dynamic Load Compensation Technical Note - DLC Calibrate multiple, odd sized or large sensors without compromising the accuracy Improved and documented uncertainty budget calculations Accuracy
More informationSTEM Society Meeting, March 12, 2013
STEM Society Meeting, March 12, 2013 James Emery Last Edit: 4/3/2013 Contents 1 About the STEM Society and the STEM Society Website 1 2 The March Meeting Announcement 2 3 Jim Emery: A Biographical Sketch
More informationThermo Scientific SPECTRONIC 200 Visible Spectrophotometer. The perfect tool. for routine measurements
Thermo Scientific SPECTRONIC 200 Visible Spectrophotometer The perfect tool for routine measurements The Standard for Routine Measurements Thermo Scientific SPECTRONIC spectrophotometers have served as
More informationDescription. Order code Package Packing
TDA7391PD 32 W bridge car radio amplifier Features High power capability: 40 W/3.2 EIAJ 32 W/3.2 @ V S = 14.4 V, f = 1 khz, d = 10 % 26 W/4 @ V S = 14.4 V, f = 1 khz, d = 10 % Differential inputs (either
More informationECMF02-2BF3. Dual line IPAD, common mode filter with ESD protection for high speed serial interface. Description. Features.
Dual line IPAD, common mode filter with ESD protection for high speed serial interface Description Datasheet - production data The ECMF02-2BF3 is a highly integrated common mode filter designed to suppress
More informationThermo Scientific SPECTRONIC 200
molecular spectroscopy Thermo Scientific SPECTRONIC 200 Part of Thermo Fisher Scientific The New Standard for Routine Measurements Robust, Multifunction Sample Compartment Whether you measure in 10 mm
More informationVACUSEAL MODEL 300 & 400
VACUSEAL MODEL 300 & 400 G L HOT TUB PRODUCTS 2 Toelles Road, Suite 13 Wallingford, CT 06492 860-469-2580 www.hottubproducts.com J G F H H K E D C I A P B 10 9 8 7 6 5 4 3 2 1 0 0 1 Figure 1 2 3 4 SPAS
More informationwelcome Professionalism - Reliability - Committment
2016 our history C.C.I.A.A. - DEP.VA 0001200521 - ITALY - T.M. PROTECTION PAT - welcome Professionalism - Reliability - Committment Quality Safety Research Planning Service Price ISO 9001 ISO 11117:2008
More informationThe Quality of Light TM. LineLight. Metaphase Technologies Inc. LL040710
The Quality of Light TM LineLight Line Lights Offers the widest selection, the brightest, and the most uniformed and lowest cost line light. TM MetaBright Line Light High Power Line Light TM MetaLight
More informationGeneral catalog. BESMAK Servo-hydraulic Medium-Capacity fatigue test System with Furnace. 1 BESMAK Material Testing Machines
General catalog BESMAK Servo-hydraulic Medium-Capacity fatigue test System with Furnace 1 BESMAK Material Testing Machines www.besmaklab.com 2 BESMAK Material Testing Machines www.besmaklab.com General
More informationSPORTS MEDICINE: TECHNOLOGIES AND GLOBAL MARKETS
SPORTS MEDICINE: TECHNOLOGIES AND GLOBAL MARKETS HLC183A March 2015 Shalini Shahani Dewan Project Analyst ISBN: 1-62296-061-0 BCC Research 49 Walnut Park, Building 2 Wellesley, MA 02481 USA 866-285-7215
More informationWIRE, ELECTRIC, MODIFIED PERFLUOROALKOXY COPOLYMER RESIN-INSULATED, NICKEL-COATED COPPER
Raychem Wire and Cable 501 Oakside Avenue, Redwood City, CA 94063-3800 SPECIFICATION: WCD 3111 THIS ISSUE: Issue 5 DATE: 24 May 2012 REPLACES: Issue 4 PAGE: 1 of 14 WIRE, ELECTRIC, MODIFIED PERFLUOROALKOXY
More informationImplantable Drug Delivery Devices Market - Global Industry Analysis, Size, Share, Trends And Forecast,
Transparency Market Research Implantable Drug Delivery Devices Market - Global Industry Analysis, Size, Share, Trends And Forecast, 2012 2018 Buy Now Request Sample Published Date: Mar 2013 Single User
More informationAnalog Vs. Digital Weighing Systems
Analog Vs. Digital Weighing Systems When sizing up a weighing application there are many options to choose from. With modern technology and the advancements in A/D converter technology the performance
More informationDESIGN AND FOLDED FABRICATION OF NOVEL SELF-BRACED TRIANGULAR STRUCTURAL SECTIONS AND FRAMES
DESIGN AND FOLDED FABRICATION OF NOVEL SELF-BRACED TRIANGULAR STRUCTURAL SECTIONS AND FRAMES XiaoQiang Shi 1, Joseph M. Gattas 2,* 1 Master s student, School of Civil Engineering, University of Queensland,
More informationAre Wi-Fi Networks Harmful to Your Health?
Probably Not, But Why Not Lower Radiation in Them Anyway? A GoNet Systems ebrief With almost every communication and computing function going wireless, consumers and device users are understandably concerned
More informationGE Healthcare. Precision 500D. Digital radiography and fluoroscopy system
GE Healthcare Precision 500D Digital radiography and fluoroscopy system What do you look for in a classical digital R&F system? Exceptional image quality at low dose, certainly. Increased clinical productivity
More informationNumerical analysis of the process of pipe connection with a clamping ring using the finite element method
Numerical analysis of the process of pipe connection with a clamping ring using the finite element method Kang-Yul Bae #1, Young-Soo Yang #2 and Sungun Go #3 #1 Department of Mechatronics Engineering,
More informationThe pattern scanning multi-color photocoagulator that puts you in control
The pattern scanning multi-color photocoagulator that puts you in control RETINAL PHOTOCOAGULATION LASER TRABECULOPLASTY LASER IRIDOTOMY Helping the world see clearly 2 INTEGRE PRO SCAN FROM ELLEX Transforming
More informationX-ray phase-contrast imaging
...early-stage tumors and associated vascularization can be visualized via this imaging scheme Introduction As the selection of high-sensitivity scientific detectors, custom phosphor screens, and advanced
More informationHigh-precision inch guide units Made in Germany
High-precision inch guide units Made in Germany cm 1 2 3 4 5 6 7 8 9 10 11 12 13 14 15 16 17 18 19 20 cm inch inch 1 2 3 4 5 6 7 8 STEINEL All you need for tools Your partner for punching and bending
More informationPower that changes the world.
Power that changes the world. 3M Display Materials & Systems Division Display Solutions for Tablets Create amazing displays. 3M Display Enhancement Films Improve visual quality. Wide viewing angle and
More informationKeywords: Thermography, Diagnosis, Image analysis, Chronic wound, Burns
Blucher Mechanical Engineering Proceedings May 2014, vol. 1, num. 1 www.proceedings.blucher.com.br/evento/10wccm THE APPLICATION OF PASSIVE THERMOGRAPHY AND MEASUREMENT OF BURN SURFACE AREA FOR THE ASSESSMENT
More informationEBI 10 High Precision Wireless Data Loggers
EBI 10 High Precision Wireless Data Loggers The Perfect Solution for Your Applications To fit each of your specific processes, ebro offers a wide variety of EBI 10 temperature, pressure and humidity data
More informationSOFTSYNCHRO HIGH PERFORMANCE TAP H O L D E R S
SOFTSYNCHRO HIGH PERFORMANCE TAP H O L D E R S You Know Taps Now take them to the next level Triple your tap life and more with Emuge Softsynchro Tap Holders! E muge is widely recognized in the industry
More informationBeyond Performance and Value
Putting the back in ultrasound Beyond Performance and Value Esaote s new ultra-performance MyLab 9 exp ultrasound system is designed to support a full range of shared service diagnostic imaging environments.
More informationTDA7231A 1.6W AUDIO AMPLIFIER OPERATING VOLTAGE 1.8 TO 15 V LOW QUIESCENT CURRENT HIGH POWER CAPABILITY LOW CROSSOVER DISTORTION SOFT CLIPPING
1.6 AUDIO AMPLIFIER OPERATING VOLTAGE 1.8 TO 15 V LO QUIESCENT CURRENT. HIGH POER CAPABILITY LO CROSSOVER DISTORTION SOFT CLIPPING DESCRIPTION The is a monolithic integrated circuit in 4 + 4 lead minidip
More informationproduct specifications
product specifications SEAMLESS WAVE FRONT EXTREMELY HIGH SPL FOR USE IN 3- or 4-WAY SYSTEMS PRECISE ANGULAR CONTROL The Adamson Y18 - the original and most powerful enclosure of the Y-Axis series - continues
More information(12) United States Patent (10) Patent No.: US 6,238,421 B1. Ginther et al. (45) Date of Patent: May 29, 2001
USOO6238421B1 (12) United States Patent (10) Patent No.: Ginther et al. (45) Date of Patent: May 29, 2001 (54) INDUCTION HEATING DEVICE AND 4,633,875 1/1987 Turner... 607/76 METHOD FOR METALLC MPLANTS
More informationOTFI 0285 XLM Plus High CRI LED Fiber Optic Light Module with Electronics Key Features
DATASHEET Lighting Solutions XLM Plus High CRI LED Fiber Optic Light Module with Electronics Key Features Color Rendering Index (CRI) over 90 with R9 values typically over 80 Bright Light intensity is
More informationPolyvinyl Alcohol Yarn Prepared into a Biodegradable Vascular Stent: Effects of Plied Number, Twist Factor on Yarn Structure and Mechanical Properties
2017 Asia-Pacific Engineering and Technology Conference (APETC 2017) ISBN: 978-1-60595-443-1 Polyvinyl Alcohol Yarn Prepared into a Biodegradable Vascular Stent: Effects of Plied Number, Twist Factor on
More informationTDA7384A. 4 x 46 W quad bridge car radio amplifier. Features. Description
4 x 46 W quad bridge car radio amplifier Datasheet production data Features High output power capability: 4 x 46 W / 4 max. 4 x 27 W / 4 @ 14.4 V, 1 khz, 10 % Low distortion Low output noise Standby function
More information